Techi Op 3d Pro Installation Manual 216363 r4 en Ww
Techi Op 3d Pro Installation Manual 216363 r4 en Ww
ENGLISH
216363 rev. 4
0.805.5080
Copyright Copyright © 2018 by PaloDEx Group Oy.
All rights reserved.
4 PC requirements ........................................................................................................ 27
4.1 Minimum PC requirements.................................................................................. 27
4.2 The dental imaging software ............................................................................... 29
rev 1
6 Panoramic and 3D calibrations and adjustments................................................... 63
6.1 Introduction ......................................................................................................... 63
6.2 When to calibrate the unit ................................................................................... 64
6.3 Reset maintenance counter ................................................................................ 65
6.4 Preparing for calibration ...................................................................................... 65
6.5 The calibration sequence .................................................................................... 68
6.5.1 Calibration of the preheat of the tube....................................................... 68
6.5.2 Calibration of the tube current.................................................................. 68
6.5.3 3D collimator calibration........................................................................... 69
6.5.4 3D Beam size verification (optional)......................................................... 71
6.5.5 Panoramic collimator calibration .............................................................. 72
6.5.6 Panoramic geometry calibration............................................................... 76
6.5.7 3D geometry calibration ........................................................................... 78
6.5.8 SFOV (Small panel) 3D lasers alignment ................................................ 80
6.5.9 MFOV (Medium panel) 3D lasers alignment ............................................ 82
6.5.10 Panoramic laser alignment....................................................................... 86
6.5.11 3D pixel calibration................................................................................... 87
6.5.12 Panoramic pixel calibration ...................................................................... 88
6.5.13 3D Quality Check ..................................................................................... 90
6.5.14 Panoramic Quality Check (optional)......................................................... 95
2 rev
8.9.1 Programming a R3220 Main CPU board having wrong unit
type configured....................................................................................... 148
8.9.2 Programming a non-responsive cephalostat CPU ................................. 150
8.10 Troubleshooting ................................................................................................ 154
8.10.1 Problem: No connection to the X-ray unit / R3220 circuit board ............ 154
8.10.2 Problem: Automatic software update fails.............................................. 155
8.10.3 Problem: Manual update fails................................................................. 155
8.10.4 Problem: S2terminal is unable to find binary files .................................. 155
8.11 Configurable panoramic mAs limit .................................................................... 156
rev 3
4 rev
1 Introduction
1 Introduction
1.1 ORTHOPANTOMOGRAPH™ OP 3D PRO
OP 3D PRO is a dental X-ray system for producing high
quality digital images of dentition, TM-joints and skull. In
order to take images with OP 3D PRO you need a suitable
PC hardware connected to the OP 3D PRO unit and
CLINIVIEW™ imaging software to capture and manage
images. The OP 3D PRO performs the following
procedures:
Panoramic
• Standard panoramic
• Pediatric panoramic
• Wide arch panoramic
• Bitewing
• TMJ, posterior-anterior (PA) projection
• Ortho TMJ, axial corrected lateral projection
• Maxillary sinus
• Ortho Zone enhanced panoramic
• Orthogonal panoramic
Cephalometric (optional)
• 61 x 41 mm Field of View
• 61 x 78 mm Field of View
• 50 x 50 mm Field of View
1
• 61 x 78 mm Field of View
• 78 x 78 mm Field of View
• 78 x 150 mm Field of View
• 130 x 150 mm Field of View (optional)
2
1 Introduction
1.4 References
The following instructions are delivered with in the
OP 3D PRO installation manual:
FH = Frankfort-Horizontal
H = Horizontal
3
1 Introduction
4
1 Introduction
5
1 Introduction
6
1 Introduction
7
1 Introduction
8
1 Introduction
Do not connect any equipment to the unit that has not been
supplied with the unit or that is not recommended by the
manufacturer. The use of accessory equipment not
complying with the equivalent safety requirements of this
equipment may lead to a reduced level of safety of the
resulting system.
1.7 Disclaimer
The manufacturer shall have no liability for consequential
damages, personal injury, loss, damage or expense
directly or indirectly arising from the use of its products. No
agent, distributor or other party is authorized to make any
warranty or other liability on behalf of the manufacturer with
respect to its products.
1.8 Disposal
The device, its spare parts, its replacement parts and its
accessories may include parts that are made of or include
materials that are non-environmentally friendly or
hazardous. These parts must be disposed of in
accordance with all local, national and international
regulations regarding the disposal of non-environmentally
friendly or hazardous materials.
9
1 Introduction
10
2 Unit description
2 Unit description
2.1 Main parts and controls
1. Column
2. Carriage
3. Main support
4. Rotating unit
5. On/off switch (rear of carriage)
and main fuses
6. Tubehead assembly
7. Touch screen display
8. Patient positioning panel
9. Sensor head
10.Head support
11.Chin rest
12.Handles
13.Cephalostat unit
14.Cephalostat sensor
15.Secondary collimator
16.Patient positioning panel
11
2 Unit description
12
2 Unit description
1. Midsagittal light
2. Frankfort horizontal (FH) light /
Horizontal light, top of 130 mm high FOV (Medium Panel
3D option only)
3. Image layer light
4. Cephalometric FH light
5. TMJ light
6. Horizontal light, top of 78 mm high FOV
7. Horizontal light, top of 61 mm high FOV
8. Horizontal light, top of 50 mm FOV
9. Horizontal light, bottom of FOV
13
2 Unit description
Panoramic lights
1. Midsagittal light
2. Frankfort horizontal (FH) light
1. Image layer
2. TMJ light
14
2 Unit description
3D lights (optional)
1. Midsagittal light
15
2 Unit description
1. Carriage UP
2. Carriage DOWN
3. Positioning lights ON/OFF
4. Patient positioning lights
5. Start position
6. Chin support UP
7. Chin support DOWN
8. Move the image layer anterior before exposure 3 mm,
with sinus program 10 mm
9. Normal occlusion/ reset position
10.Move the image layer posterior before exposure 3 mm,
with sinus program 10 mm.
2. Carriage DOWN
16
2 Unit description
17
2 Unit description
1. Main label
2. 10A & 15A Fuse labels (next to the fuse holder)
3. Laser class 1 warning label IEC 60825-1:2007
4. Ethernet label
5. Sensors
6. (Primary) collimator label
7. (Secondary) cephalostat collimator label
8. Tubehead label
(on the tubehead and on the tubehead cover)
9. Warning label for deadly voltages
(inside the tubehead cover)
10.Cephalostat main label
18
2 Unit description
(R)
(C)
(J)
(D)
19
2 Unit description
20
2 Unit description
21
2 Unit description
22
3 Pre-installation requirements
3.1 The unit
■ See chapter 9 Pre-sales check list.
Size (LxWxH) in: 67.3 x 25.5 x 38.1 in 46.4 x 25.5 x 38.1 in 55.1 x 30.7 x 34.6 in
Gross weight 154 kg/ 339.5 lbs 114 kg/ 251.3 lbs 66 kg/ 145.5 lbs
approx.:
Net weight approx.: 71 kg/156.5 lbs 83 kg/183 lbs 25 kg/55 lbs
■ Unit weights:
PAN CEPH 3D
194 kg 229 kg 196 kg
(428 lb.) (505 lb.) (432 lb.)
NOTICE! Mounting bolts for floor and wall fastening are not
included in the delivery. The fixing hardware used to
permanently attach the unit to the wall must be correct type
for the wall and wall material.
23
3 Pre-installation requirements
24
3 Pre-installation requirements
Pan/3D:
Cephalostat:
25
3 Pre-installation requirements
Fixing hardware
NOTICE! Mounting bolts for floor and wall fastening are not
included in the delivery.
Installation tools
- Electric drill
- Spanners (wrenches) 10, 17 (x 2) mm AF
- Allen keys (Hexagon socket wrenches) 1.5-8mm
- Flat blade screwdrivers
- Spirit level
- Pliers and wire cutters
- Scissors/Knife
- Service tools set
- 3D calibration phantom
- QC phantom
26
4 PC requirements
4.1 Minimum PC requirements
Minimum PC requirements for 2D and 3D acquisition workstation
Processor Intel Core i5, i7 or Xeon, 4-cores or more
GPU (graphics NVIDIA Quadro M2000 4GB or GeForce GTX 1050 Ti
processing unit) 4GB
Memory 8 Gigabytes RAM, or more
Hard disk 1 TB or more
RAID 1 or RAID 5 recommended for data redundancy,
plus backup
Network Gigabit Ethernet 1000 Mb/s
Operating system Windows 10 Pro or Enterprise, 64-bit, Windows 8.1 Pro
or Enterprise, 64-bit, Windows 7 Professional, Ultimate or
Enterprise, 64-bit, with SP1
Display 1920 x 1080 resolution (Full HD) or higher, at least 300
cd/m2 brightness for typical room lighting, native contrast
ratio 100:1 or better, 8-bit panel strongly recommended.
Standard The PC must meet the IEC 60950 standard (minimum re-
quirements)
Other OpenCL 1.1 support
DVD-ROM drive
Anti-virus software
27
4 PC requirements
28
4 PC requirements
29
4 PC requirements
30
5 Installing the unit
NOTICE! Save the packaging materials as they may be
needed if you move the unit to a new location.
CEPHALOSTAT BOX
COLUMN BOX CARRIAGE BOX
(optional)
3. Lift off the top of the box and then the sides.
31
5 Installing the unit
6. Fold down the bottom end of the box and slide the unit
over the edge of the pallet.
32
5 Installing the unit
Adjustments:
1) angle
2) depth
3) lateral Additional wall bracket (lower)
4) vertical
33
5 Installing the unit
9. Erect the column by lifting from the top. Move the unit
to the place where it is installed and set it beside the
wall.
11. Attach the unit to the upper wall bracket on the wall with
screws and washers.
13. Remove the screws of the mains inlet cover and detach
the cover.
34
5 Installing the unit
14. Drill holes and attach the unit to the floor with four (4)
screws.
35
5 Installing the unit
36
5 Installing the unit
37
5 Installing the unit
3. Lift off the top of the box and then the sides. Remove
the plastic wrap.
38
5 Installing the unit
6. Lift the carriage and push the main supports onto the
column supports.
The lifting places are shown by arrows in the picture.
39
5 Installing the unit
40
5 Installing the unit
1
2
3 3
41
5 Installing the unit
42
5 Installing the unit
1 Gigabit ethernet
(JP004 marked with “Gigabit”)
2 Grounding point
3 Mains power cable
On the top
(Switch power cable pointed out)
On the rear
43
5 Installing the unit
44
5 Installing the unit
3. Fasten the touch screen display to the arm with the pre-
viously detached screws.
45
5 Installing the unit
46
5 Installing the unit
47
5 Installing the unit
48
5 Installing the unit
49
5 Installing the unit
50
5 Installing the unit
51
5 Installing the unit
3. Lift off the top of the box and then the sides.
Remove the plastic wrap.
52
5 Installing the unit
53
5 Installing the unit
6. To pull the cephalostat head from the box, grab the low-
er packaging styrofoam cells from both sides. Lift the
front end slightly and pull the unit out of the box careful-
ly, leaving the cephalostat arm to its place.
54
5 Installing the unit
55
5 Installing the unit
10. Remove the upper cover (A) from the cephalostat head
and put it aside. Take the cephalostat head out of the
styrofoam packaging and lift it carefully to the hook on
the other end of the cephalostat arm.
11. The pin (A) goes through the inlet. Attach the two bolts
(B) to lock the cephalostat head to its place.
B
B
56
5 Installing the unit
12. Check that the cephalostat arm is level and adjust the
tilt needed by loosening the two indicated screws. The
correct position of the arm may sometimes be deter-
mined only during calibrations.
Side horizontal balance in terms of the cephalostat Front horizontal balance in terms of the cephalostat sensor.
sensor.
57
5 Installing the unit
58
5 Installing the unit
1: Signal/Power cable
2: Ethernet cable
3: Grounding cable
1
3
59
5 Installing the unit
Joint cover for left and right handed cephalostat. Joint cover (A) screwed in place from below.
NOTICE! Only one cover is used according to the handedness of the cephalostat.
Upper cover of the cephalostat head. Fasten the upper cover to it's place with four screws
(M4x16) through the holes under the ceph head.
60
5 Installing the unit
D D
Top cover of the column. Attach the cover in place with 4 screws
Cephalostat arm end covers. Attach the covers in place with 2 screws
61
5 Installing the unit
17. Take the sensor out of its package and attach to the
sensor holder. Make sure the sensor is properly seated
down before pulling the lever.
18. Pull the lever down to lock the sensor to its place.
62
6 Panoramic and 3D calibrations and
adjustments
6.1 Introduction
The calibration procedure helps to maintain image quality
and correct operation of the unit.
63
6 Panoramic and 3D calibrations and adjustments
Tubehead
Collimator assembly
64
6 Panoramic and 3D calibrations and adjustments
3D Panel
Panoramic panel
_______________________________________
c) Press ENTER.
65
6 Panoramic and 3D calibrations and adjustments
_______________________________________
s2terminal 10.208.6.101
_______________________________________
66
6 Panoramic and 3D calibrations and adjustments
67
6 Panoramic and 3D calibrations and adjustments
2. Take an exposure.
3. Touch screen display and s2terminal inform you when
the calibration is done. Repeat the calibration until cali-
bration result “passed” is achieved.
2. Take an exposure.
3. Touch screen display informs when the calibration is
done. Repeat the calibration until calibration result
“passed” is achieved.
68
6 Panoramic and 3D calibrations and adjustments
3. Take an exposure.
2. Adjust the height with one screw on the bottom of the col-
limator.
69
6 Panoramic and 3D calibrations and adjustments
70
6 Panoramic and 3D calibrations and adjustments
4. Take an exposure.
5. The beam dimensions and calibration result are present-
ed in the result image.
First two lines are for DIN and the rest for the IEC standard.
71
6 Panoramic and 3D calibrations and adjustments
4. Take an exposure.
5. The status is displayed on the image captured in the im-
aging software, how much the height needs to be adjust-
ed and in which direction.
6. Adjust the height of the PAN sensor and check the tilt and
position of the collimator. Make adjustments and take
new exposures until calibration result “passed” is
achieved.
72
6 Panoramic and 3D calibrations and adjustments
73
6 Panoramic and 3D calibrations and adjustments
74
6 Panoramic and 3D calibrations and adjustments
75
6 Panoramic and 3D calibrations and adjustments
4. Take an exposure.
5. Repeat the calibration until calibration result “passed” is
achieved.
76
6 Panoramic and 3D calibrations and adjustments
77
6 Panoramic and 3D calibrations and adjustments
5. Take an exposure.
6. Repeat the calibration until calibration result “passed” is
achieved.
78
6 Panoramic and 3D calibrations and adjustments
79
6 Panoramic and 3D calibrations and adjustments
3
4
80
6 Panoramic and 3D calibrations and adjustments
A
A
A
A
6. Loosen the screw (B) of the upper laser. Align the tilt of
the laser by rotating it.
81
6 Panoramic and 3D calibrations and adjustments
7. Loosen the two screws (C) of the bottom laser. Align the
tilt of the laser by rotating.
1 Mid-sagittal
2 Adjustable phantom holder
82
6 Panoramic and 3D calibrations and adjustments
83
6 Panoramic and 3D calibrations and adjustments
Adjust the lowest laser first. Loosen the upper screw. The
laser starts to rotate. Align the laser horizontally. Tighten
the screw.
84
6 Panoramic and 3D calibrations and adjustments
9. Adjust the FH-laser used for 130 x 150 FOV to the corre-
sponding grooves by mechanically adjusting the lasers.
10.Briefly press the exposure button (no X-rays are gener-
ated) to acknowledge that the check has been carried
out.
85
6 Panoramic and 3D calibrations and adjustments
86
6 Panoramic and 3D calibrations and adjustments
87
6 Panoramic and 3D calibrations and adjustments
4. Take an exposure.
5. The resulting image informs when the calibration is
passed.
5. Take an exposure.
6. Check that the calibration result is “passed”.
88
6 Panoramic and 3D calibrations and adjustments
89
6 Panoramic and 3D calibrations and adjustments
• Image noise
• Grey values
• Uniformity (Driver Update Package 15.1 or newer
required)
• Artefacts
90
6 Panoramic and 3D calibrations and adjustments
4. Take an exposure.
91
6 Panoramic and 3D calibrations and adjustments
Last lines of text on the image tell the file name of the
measurement results stored in a csv table. Path for this file
is typically C:\ProgramData\DXR120\. It is not necessary,
but the “3DROI” file can be copied to be archived as
constancy check results or imported to a spreadsheet
program to closer examine measured values. In case of
failed QC result, this file can be attached for technical
support request.
Noise measurement
92
6 Panoramic and 3D calibrations and adjustments
Field Uniformity
Artefacts
Requirements:
93
6 Panoramic and 3D calibrations and adjustments
Examples:
94
6 Panoramic and 3D calibrations and adjustments
4. Take an exposure.
95
6 Panoramic and 3D calibrations and adjustments
2
3 1
4
5
Subjects to be evaluated:
96
7 Cephalometric calibration and alignment
Take an exposure.
97
7 Cephalometric calibration and alignment
6. Take an exposure.
98
7 Cephalometric calibration and alignment
4. Take an exposure.
99
7 Cephalometric calibration and alignment
100
7 Cephalometric calibration and alignment
101
7 Cephalometric calibration and alignment
3. Take an exposure.
4. Follow the instructions in the calibration result image.
Possible instructions and references to further informa-
tion are found in table “instructions for adjustment”.
5. Take new exposures until the calibration result is
“Passed”.
102
7 Cephalometric calibration and alignment
103
7 Cephalometric calibration and alignment
104
7 Cephalometric calibration and alignment
105
7 Cephalometric calibration and alignment
106
7 Cephalometric calibration and alignment
107
7 Cephalometric calibration and alignment
108
7 Cephalometric calibration and alignment
4. Loosen the two upper screws and lift the cover to remove
it.
109
7 Cephalometric calibration and alignment
110
7 Cephalometric calibration and alignment
2. Adjust the tilt of the secondary collimator using the tilt ad-
justment screw.
111
7 Cephalometric calibration and alignment
3. Take an exposure.
4. Repeat the calibration until calibration result “passed” is
achieved.
112
7 Cephalometric calibration and alignment
The laser is on the secondary collimator, see chapter 2.2. Patient positioning
lights.
113
7 Cephalometric calibration and alignment
5. Take an exposure.
6. This calibration should always be passed.
114
7 Cephalometric calibration and alignment
4. Take an exposure.
115
7 Cephalometric calibration and alignment
Subjects to be evaluated:
The ear holder center pins and the x-ray focal spot must be
in line.
1. Turn the ear holders into the lateral view position and
move them completely apart.
2. Select the cephalometric exposure program.
116
7 Cephalometric calibration and alignment
3. Take an exposure.
4. Both ear holders have a small metal ball that should be
visible in the image: both balls should be displayed in the
same point so that they merge into a single ball. The larg-
er ball represents the tube side, and the smaller ball the
cephalostat sensor side.
5. If the balls do not merge into a single ball, the ear holders
must be adjusted.
117
7 Cephalometric calibration and alignment
Top view of ear holder assembly. To ease the loosening, lock the ear holder assembly to its place with the
locking switch.
118
7 Cephalometric calibration and alignment
119
7 Cephalometric calibration and alignment
120
8 Special procedures
8.1 General
When updating the unit, e.g. with a cephalometric or 3D
option, it is always recommended to update the unit with
latest firmware and the image capturing PC with the latest
driver software (KaVo Driver Update CD). Please contact
the manufacturer’s Technical Support for further
information.
121
8 Special procedures
122
8 Special procedures
123
8 Special procedures
1: Signal/Power cable
2: Ethernet cable
3: Grounding point
124
8 Special procedures
125
8 Special procedures
126
8 Special procedures
127
8 Special procedures
11. Move the bubble level to the other side of the cepha-
lostat head.
128
8 Special procedures
13. Move the prism assembly to the other side of the ceph-
alostat head.
129
8 Special procedures
14. Move the 2 stop screws (A) to the other side so that
when the patient is in the unit, they will be facing for-
wards to the room.
A
A
A
130
8 Special procedures
131
8 Special procedures
l: The GUI and/or cephalostat are on the left side of the unit
Usage examples:
132
8 Special procedures
133
8 Special procedures
134
8 Special procedures
Jumper
5. Make sure that you have correct main fuses (F1 & F2)
in the unit.
■ For 220-240Vac use two 10A fuses, type T10A H
250V (Littelfuse 326 010)
■ For 100-120Vac use two 15A fuses, type T15A H
250V (Littelfuse 326 015)
135
8 Special procedures
136
8 Special procedures
137
8 Special procedures
138
8 Special procedures
IP address: 10.208.6.102
139
8 Special procedures
xu R<ver_major>.<ver_minor>
Example:
xu R2.00
140
8 Special procedures
Example:
safe
xf R<ver_major>.<ver_minor>
141
8 Special procedures
Example:
xf R2.00
Example:
Example:
xcc r3300 R2.00
3. The new software is taken into use the next time the
unit is powered on.
Example:
142
8 Special procedures
3. The new software is taken into use the next time the
unit is powered on.
xfr OP3DPRO_CEPHR3210_R<ver_major>_<ver_mi-
nor>.srec core
Example:
xfr OP3DPRO_CEPHR3210_R2_00.srec core
xfs
OP3DPRO_CEPHRFW_R<ver_major>_<ver_minor>.sr
ec main
Example:
xfs OP3DPRO_CEPHFW_R2_00.srec main
143
8 Special procedures
login
Example:
pmode 0 2 00
144
8 Special procedures
145
8 Special procedures
7. Select the first row in the list and click the Change
File… button.
8. Select the correct sof file for the circuit board (e.g.
R3300_R2_00.sof for release R2.00 and R3300 circuit
board).
10. Click the Start button and wait for the download to com-
plete (takes a few seconds).
11. Volatile installation is typically followed by a permanent
install. Refer to chapter Manual update for instructions
on how to do this. The volatile install is voided if the unit
is powered off before the software is permanently in-
stalled.
146
8 Special procedures
8.8.1.2 R3220
6. Select the first row in the list and click the Change
File… button.
7. Select the correct sof file for the circuit board (e.g.
R3220_R2_00.jic for release R2.00 and R3220 circuit
board).
147
8 Special procedures
9. Click the Start button and wait for the download to com-
plete (takes a few seconds).
148
8 Special procedures
Example:
safe
xfp OP3DPRO_FW_R<ver_major>_<ver_minor>.srec
Example:
xfp OP3DPRO_FW_R2_00.srec
149
8 Special procedures
150
8 Special procedures
Example:
xfp OP3DPRO_CEPHRFW_R<ver_major>_<ver_minor>.srec
Example:
151
8 Special procedures
login
Example:
ngeoceph> login
main 1.4.4693/2012-08-10T15:32:43/TUULTESKELINEN
core 2012-01-31T12:17:34/EP2C70F672C6/AUTO
Example:
CEPH R3200c boot-2.18R2798>xcp r3210 xcp r3210
OP3DPRO_CEPHR3210_R<ver_major>_<ver_minor>.srec
_CEPHR3210_R2_00.srec
Transferring xcp R3210
OP3DPRO_CEPHR3210_R<ver_major>_<ver_minor>.srec
_CEPHR3210_R2_00.srec to Main CPU
Updating core...
transfer 864965 bytes from S-records...
update 0 864965 = 0x0..+864965
FRAMES:1205
Flashing image (864965 bytes) @0x0
152
8 Special procedures
Example:
CEPH R3210c boot-2.18R2798>xfp xcp R3210
OP3DPRO_CEPHR3210_R<ver_major>_<ver_minor>.srec
_CEPHFW_R2_00.srec
Reading program file "xcp R3210
OP3DPRO_CEPHR3210_R<ver_major>_<ver_minor>.srec
_CEPHFW_R2_00.srec" ...
Waiting for eraseErase time: 1342
Programming...
FRAMES:367
FRAMES:367
…
Checking embedded checksum...
checksum verified ok
;Flashing image (263148 bytes) @0x120000
Status command transfer error.
Verifying...
Error in login reply - command executed
verified ok
Done.
Connected to CBCT
CEPH R3210c boot-2.18R2798>
153
8 Special procedures
login
Example:
ngeoceph> login
main 1.4.4693/2012-08-10T15:32:43/TUULTESKELINEN
core 2012-01-31T12:17:34/EP2C70F672C6/AUTO
8.10 Troubleshooting
154
8 Special procedures
155
8 Special procedures
The unit type of the Main CPU board is incorrectly set. The
unit type for which the s2terminal is trying to find binary
files is indicated in the error message. See chapter Special
cases / Programming a R3220 Main CPU board having
wrong unit type configured for instructions on how to solve
this issue.
156
9 Pre-sales check list
Dealer
Clinic name
Clinic address
Clinic tel.
Clinic IT contact
End user
APPROVAL LEVEL
157
9 Pre-sales check list
158
9 Pre-sales check list
APPROVAL LEVEL
159
9 Pre-sales check list
IDENTIFIERS
AE Title Port number IP Address
Worklist SCP
Storage SCP
Print SCP
160
9 Pre-sales check list
YES NO DETAILS
Which software CLINIVIEW™ software (re- X
components are in- quired)
cluded in delivery?
OnDemand3D Dental
(3D visualization software)
161
9 Pre-sales check list
DICOM Conform- Have the DICOM conform- Can be downloaded from the manu-
ance statement ance statement(s) been deliv- facturer extranet
ered to the end-user?
Other
162
9 Pre-sales check list
163
9 Pre-sales check list
9.12 Dimensions
463
164
9 Pre-sales check list
165
9 Pre-sales check list
166
PaloDEx Group Oy reserves the right to make changes in specification and features shown
herein, or discontinue the product decribed at any time without notice or obligation.
Contact your PaloDEx Group Oy representative for the most current information.
216454-3
Headquarters
PaloDEx Group Oy
Nahkelantie 160 | FI-04300 Tuusula | FINLAND
Tel. +358 10 270 2000 | https://www.kavo.com/en/contact
www.kavokerr.com
USA
Dental Imaging Technologies Corporation
1910 North Penn Road | Hatfield, PA 19440 | USA
Tel: 1-215-997-5666 | Fax: 1-215-997-5665
https://www.kavo.com/en-us/contact-us