Professional Documents
Culture Documents
1F095028G_14-EN_USER_MANUAL_SI610-1_MKZ2026-1
1F095028G_14-EN_USER_MANUAL_SI610-1_MKZ2026-1
TABLE OF CONTENTS 3
1. PRINCIPLE OF OPERATION 5
3. OPERATION CONDITIONS 6
4. TECHNICAL PARAMETERS 6
5. EXPLANATION OF SYMBOLS 8
6. SAFETY PRECAUTIONS 8
7. PRODUCT DESCRIPTION 19
9.4 SWITCHING THE AIR TEMPERATURE MODE ON AND SETTING THE AIR TEMPERATURE
38
10.2 TRENDS 66
10.3 ALARMS 68
14. COMPLETION 98
The climatic conditions are controlled by an electronic temperature regulator. This regulator regulates
heating of the patient space according to either air or body temperature. The regulation mode is operator
selectable. Temperature is measured by a built-in thermometer and is displayed on the incubator display.
When oxygen is applied its concentration is regulated automatically according to operator’s requirements
and displayed on the incubator display.
A built-in humidifier module enables automatic relative humidity regulation according to operator’s
requirements. Its value is displayed on the incubator display.
Air is also sucked into the patient chamber from the outside. Air purity is ensured by a built-in
bacteriological filter.
The incubator may be optionally equipped with a module for SpO2 monitoring. The module enables
measurement of haemoglobin saturation of arterial blood with oxygen SpO2, pulse, PI, PVI, SpHb.
All the important incubator parameters are watched by a system of alarms. Deviations from these
parameters are indicated optically and acoustically. In case of more serious faults the incubator is
equipped by a circuit that disconnects heating, oxygen input and humidification. The alarm functions are
also backed against power failure.
In order to reduce the loss of the desired values in the patient area (temperature, relative humidity and
oxygen concentration), after opening the front or rear door, the incubator is provided with an automatic
airflow increase system in the patient area.
No counter indications are known from the point of view of provision of diagnostic and treatment care in
neonatology and paediatrics. There is a risk of retrolental fibroplasia in application of higher oxygen
concentration in the patient space.
The incubator may only be operated under the following operation conditions:
- For indoor use (healthcare facility room)
Incubator location - In dust free environment without aggressive fumes
and gases out of direct sunshine
20 – 30 °C, the outside temperature must be at least
Room temperature by 3 C° lower than the required temperature inside
the patient space
Pressure 500 – 1100 hPa
5 - 95 % (non-condensing humidity)
Air humidity - the ambient humidity must at least by 3% lower
than the required humidity inside the patient space
Incubator operational position horizontal
4. Technical Parameters
Unit type identification SI-610-1
version 1: 230 V (±10%) / 50 – 60 Hz
Power voltage version 2: 110-115 V (±10%) / 50 – 60 Hz
version 3: 120-127 V (±10%) / 50 – 60 Hz
Power input
With fixed height: 550 VA
With adjustable height: 600 VA
TEMPERATURE MODE
Two operation modes enabling temp. measurement and regulation according to air or body
temperature:
Air temperature mode 20.0 °C - 37.0 °C, in 0.1 °C increments
(adjustment range) 37.1 °C - 39.0 °C, in 0.1 °C increments
34.0 °C - 37.0 °C , in 0.1 °C increments
Body temperature mode (adjustment range)
37.1 °C - 39.0 °C , in 0.1 °C increments
WEIGHING MODE
Enables patient weighing:
PATIENT DATA
Enables patient data (trends of values and alarms) to be stored and printed
Number of patients in the memory 20*
WARNING:
Use the incubator upon instructions of a physician, who is aware of the so far
known risks and advantages of incubator use.
CAUTION!
1. The incubator may only be operated and cleaned by a person trained in
operation and cleaning of the particular incubator type. Training is provided
by the manufacturer or by an authorized service organization. The trainer will
make an entry about the training in the incubator technical log and into the
handing-over report.
2. Do not use the incubator with an electric or mechanical defect. Call the
service if you find a defect.
3. The incubator may increase patient fluid loss. It is necessary to control the
patient's body temperature while increasing the patient's fluid intake.
CAUTION!
1. Do not use this probe rectally.
2. Is it possible to use only probes with a valid temperature calibration.
WARNING:
1. A faulty skin temperature probe may only be replaced with the same type,
supplied with the incubator.
2. The body temperature probes are classified as applied parts of BF type. The
patient is not electrically connected with the ground during application.
CAUTION!
1. Oxygen application inside the incubator increases fire risk.
2. Do not place any auxiliary devices not approved for operation in explosive
environment inside the incubator.
3. Any parts in the patient space, the O2 gaskets, O2 fittings and O2 hoses must
not get in touch with lubricants.
4. If disinfection agents based on ether are used these substances have to be
thoroughly aired before operating the incubator.
5. Using open fire and smoking is forbidden inside the incubator and close to it.
Notes:
1. A functional tester cannot be used to assess the accuracy of the pulse co-
oximeter.
WARNING:
1. Handling other controls than those located on the control box panel is
forbidden.
2. An operator may also change the bed inclination, control box position and the
height adjustment of the incubator.
CAUTION!
There must be no obstacles to air exhaustion and sucking in the patient space
(e.g. by napkins or blankets on the incubator bed). The air temperature
measurement might not correspond with the real air temperature over the bed
if the air openings are covered.
CAUTION!
1. The patient space (with a patient inside) may only be open in permanent
supervision of healthcare staff.
2. The baby has to be prevented from falling when the incubator is open.
3. After closing the front or the rear door or the oval port holes the operator has
to check whether they are well closed by means of the hinges.
WARNING:
1. The incubator must not be open longer than necessary.
2. The temperature, oxygen concentration and relative humidity drop when the
incubator is open.
CAUTION!
When closing the door, it is necessary to monitor whether any hose or cable
conducted through hose inputs was not loosened and squeezed in a door after
closing. This could cause malfunction of additional appliances or devices.
WARNING:
1. Withdrawn or tilted bed may only be loaded with the patient’s weight, max.
10 kg.
2. Further bed loading, e.g. with devices, or leaning against the bed is forbidden.
3. Before putting an infant into the incubator the operator has to check whether
the bed with the mattress is correctly located.
WARNING:
Material of maximum weight of 10 kg can be put on the shelf.
CAUTION!
1. Do not lean against the withdrawn drawers or load them more than mentioned
above.
2. The drawers must be closed when the incubator is transported.
WARNING:
1. The maximum load capacity of the big drawer is 5 kg.
2. The maximum load capacity of the small drawer is 2.5 kg.
In terms of electric shock protection the incubator is a Class I device in the sense of the above standards.
It may only be connected to the mains with power disconnection protection. Three-wire power cord
supplied by the manufacturer may only be used.
To improve the safety in the environment of a health-care facility the stationary incubator
SI-610-1 may be connected through a cable to terminals of complementary protection circuit as
per ČSN 33 20-00-4-41. The incubator is equipped with a standardized protective terminal.
The incubator may be connected via a special cable to a terminal of a surgery distribution system.
Both ends of the cable are equipped with angle plug contacts 0299-0-0032.
CAUTION!
1. Always disconnect the incubator from the wall socket when cleaning it.
2. Always disconnect the incubator from the wall socket when changing the fuses.
CAUTION !
Never pull the power cord plug out of the socket without pressing the lock
buttons.
CONTROL
BOX
UPPER PART
BED
LOWER
PART
CAUTION!
The knobs have to be in the locked position when the door is closed.
Water tank
A water tank for the humidifier module is located at
the front of the incubator.
See chapter SETTING THE INCUBATOR INTO
OPERATION for water filling instructions.
Toolbar
A toolbar for hanging additional instruments
is located at the side of the incubator.
Vertical bar
The vertical bar is attached to the bottom of the
incubator. It is used to hang additional devices.
FIX
The incubator electronic system enables the bed height adjustment on one side only. If the bed is inclined
on the left side for example and height adjustment on the right side is required, the bed has first to be
levelled on the left and then the position on the right can be adjusted.
CAUTION!
The patient’s position on the bed must be monitored when the bed is being
positioned. No patient body part must exceed the bed outlines.
Use the left keys to position the left side of the bed.
Press keys No. 10 and No. 11 to position the left side of the bed.
Press the appropriate key for the bed movement and hold it until the bed reaches the required position.
Use the right keys to position the right side of the bed.
Press keys No. 12 and No. 13 to position the right side of the bed.
Press the appropriate key for the bed movement and hold it until the bed reaches the required position.
The whole bed can be withdrawn from the incubator upper part by 245 mm.
WARNING: The patient bed may only be withdrawn in the horizontal position.
CAUTION!
Check that the bed is secured against overturning during rotation.
SI-610-1 Accessories
WARNING:
1. The maximum load capacity of the additional shelf is 5 kg.
2. The maximum load capacity of the infusion holder is 2 kg.
3. The maximum load capacity of the Euro rail is 5 kg.
5. Toolbar shelf
(Art. code 51FA410044)
6. Hose holder
(Art. code 21FK846093)
9. Gel mattress
(Art. code 51FA810003)
Additional devices
Incubator possible combinations with other devices: phototherapy lamps, suction pumps, devices for
monitoring vital functions.
CAUTION !
1. Additional devices must be approved according to ČSN EN 60601-1 and must
be approved as a medical device.
2. The leakage currents may increase after connecting additional devices.
3. Always follow the recommendations set by the appliance manufacturer when
using accessories.
the main power source with the power switch, auxiliary sockets, an equipotentionality terminal
and the main fuses: FU1,2 250V / T1.6AL, version 110V – 127V 250V / T3.15AL
FU3,4 250V / F630mAL, version 110V – 127V 250V / T1.25AL
FU5,6 250V / F1.6AL, version 110V – 127V 250V / T2.5AL
FU7,8 250V / F5AL, version 110V – 127V 250V / F5AL
- the air temperature sensor and the air temperature emergency sensor over 38/40 °C (the
sensors are located together under the plastic cover in the corner of the patient space)
- relative humidity sensor (located under the plastic cover with the temperature sensors)
- emergency heating sensor (located in the bath under the heating element)
2 body (skin) temperature probe sockets, an equipotentionality connector (protective connection), the
incubator power switch, the power cable socket and the connector for auxiliary sockets are located on the
left side of the upper part.
In the incubator version with the monitor of SpO2 this block also includes sockets for connecting the SpO2
probe.
Skin Skin
temperature temperature
O2 saturation
(SpO2) probe
socket
Main switch
Protection
interconnection
socket connector
Power cord
socket
Auxiliary
sockets
CAUTION !
1. Devices with a total maximum power of 100W can be connected to the
auxiliary socket connector.
2. The leakage currents may increase after connecting additional devices.
A quick connector for pressurized O2, connection is located at the bottom of the upper part of the incubator
O2 quick
connector
Lower part
- The version without the height adjustment contains no electronics.
- The version with the height adjustment – the electronic system is in the lower part:
Opens the mini trend menu of the incubator. In combination with keys 17, 18, 19, 20.
Key No. 19 - + / UP
Increases the adjusted values of:
- air temperature
- body temperature
- O2
- humidity
- movement within menus
To set the required values in the individual modes press the appropriate key (air, body, O2,
humidity) and then press +.
Key No. 20 - - / DOWN
Decreases the adjusted values of:
- air temperature
- body temperature
- O2
- humidity
- movement within menus
To set the required values in the individual modes press the appropriate key (air, body, O2,
humidity) and then press -.
Key No. 21 - LEFT
Movement within menus.
A window of an active mode shows the current measured value in green (in red if it is out of tolerance)
and the corresponding adjusted value in grey.
A window of an active mode also shows an indicator of the regulator performance (air or body
temperature, humidifier, oxygen dosage). The indicator active part size represents the percentage of
performance of the appropriate mode. Full bar = 100%, no bar = 0%. If any of the modes is inactive
there is OFF displayed in the corresponding window.
is a special connector marked on the left side of the incubator. Push the special cable on
the incubator connector and on the distribution connector. The protective connection cable
ABB – 0299-0-0032/3 is optional and is not supplied with the incubator.
WARNING:
The protective connection cable has to be protected against mechanical
damage and against disconnection.
If the incubator is switched on within 15 minutes from being switched off a screen with the last
adjusted incubator values appears.
Press Key No. 19 to set the values displayed in the PARAMETERS table.
Press key No. 2 AIR, to open the air temperature setting screen.
The air temperature setting screen also displays the current measured value of the temperature inside
the incubator.
The right column of the screen contains
instructions for the particular screen.
Measured patient’s body temperatures T1, T2 may be displayed simultaneously in the body T. Window
in the AIR mode after connecting the body temperature sensor to the appropriate socket.
During temperature reduction in the AIR mode by >2 ºC from the originally set temperature the front
door or the port holes of the patient space may be opened for the necessary time.
CAUTION!
When the door or the port holes of the patient space are open the baby has to
be watched to prevent it from falling from the incubator.
To reduce the temperature press key No. 2 AIR and set the required temperature with key No.
20 . Press key No. 18 to confirm the change. Unless is pressed the new value adjustment
will not be performed.
the key .
CAUTION!
Proper attachment of the probe to the patient’s body is necessary for correct
function of the body temperature mode. Probe separation during operation may
lead to wrong function of the body temperature mode.
When setting the desired body temperature in the range between 34ºC and 37ºC, the air temperature
in the incubator ranges between 27ºC and 37,5ºC. When setting the body temperature in the range
between 37ºC and 39ºC, the air temperature in the incubator ranges between 27ºC and 38,5ºC. In the
selected body temperature mode, the current air temperature in the incubator is dependent on the
desired temperature of the body and the heat capacity of the patient.
During temperature reduction in the BODY mode by >1 ºC from the originally set temperature the
front door or the port holes of the patient space may be opened for the necessary time.
CAUTION!
When the door or the port holes of the patient space are open the baby has to be
watched to prevent it from falling from the incubator.
To reduce the temperature press key No. 3 BODY and set the required temperature with key No.
20 . Press key No. 18 to confirm the change. Unless is pressed the new value adjustment
will not be performed.
CAUTION!
1. Only demineralized water may be used for humidification.
2. The demineralised water used must meet the requirements on the bacterial
purity of water specified in ISO 15883-4.
4. Add demineralized water as necessary through the opening in the top cover.
6. Insert the chamber back inside the incubator. Press it slightly inside, the chamber will get locked
and the cover will close.
WARNING:
1. The water level must not exceed the MAX line on the humidifier chamber front
wall.
2. Change the demineralized water in the humidifier tank every day.
Maximum possible relative humidity (RH) values cannot be achieved under all operating conditions.
Maximum available relative humidity depends on the set air temperature in the incubator,
When the set RH value is reduced by >25% from the originally adjusted value the front door or the
ports of the patient space may be opened.
CAUTION!
When the door or the port holes of the patient space are open the baby has to be
watched to prevent it from falling from the incubator.
To reduce the RH value press key No. 4 RELATIVE HUMIDITY and set the required RH value by
pressing key No. 20 . Press key No. 18 to confirm the change. Unless is pressed the
new value adjustment will not be performed.
CAUTION!
1. The water must not be drained from the humidifier chamber
earlier than in 30 minutes from switching the humidifier off.
2. Unless the mode is activated, the water must not remain in the humidifier
chamber.
If the above conditions are met, you can switch the oxygen concentration mode on.
Unless is pressed within 5 seconds the O2 CONCENTRATION MODE ON screen gets closed.
When the adjusted O2 concentration is reduced by >3% O2 from the originally adjusted O2
concentration the patient space cover or the port holes can be opened.
CAUTION!
When the door or the port holes of the patient space are open the baby has to be
watched to prevent it from falling from the incubator.
To reduce the O2 concentration value press key No. 5 O2 CONCENTRATION and set the required
O2 concentration value by pressing key No. 20 . Press key No. 18 to confirm the change.
pressing .
WARNING:
To achieve measurement accuracy the sensor has always to be calibrated at the
temperature that will be set in the incubator during oxygen dosage.
During the sensor calibration at altitude exceeding 1000 m, it is necessary to use
air with content O2 20,9 % and oxygen with content O2 100 % for sensor
calibration.
Press key No. 18 and perform the steps required on the LCD during the calibration. The steps
have to be performed as fast as possible as after a longer delay the window closes automatically and
the calibration has to be started again!
CAUTION!
After successful calibration put the O2 sensor holder back into the incubator
without delay.
After calibration of O2 sensor, insert the O2 sensor holder in to the incubator to its original position
and close the port hole.
Key No. 18 can only be pressed after installation of the O2 probes inside the incubator.
If the O2 mode was active before the O2 calibration, a screen with the latest adjusted O2 value is
offered. After confirmation of the value by key No. 18 the oxygen dosage continues at this level.
CAUTION!
Before recalibration take the O2 sensor holder out or the incubator and let it
stabilize in the air oxygen for at least 30 minutes
CAUTION!
After successful calibration put the O2 sensor holder back into the incubator
without delay.
If the O2 mode was active before the O2 calibration, a screen with the latest adjusted O2 value is
offered. After confirmation of the value by key No. 18 the oxygen dosage continues at this level.
CAUTION!
When you finish oxygen dosage disconnect the pressurized oxygen hose from the
incubator O2 quick connector.
The built-in module for monitoring Masimo rainbow SET enables measurement of:
- arterial blood haemoglobin saturation with oxygen SpO2
- pulse frequency
- perfusion index - PI
- non-invasive and continuous assessment of body fluid status - PVI (Pleth Variability Index)
- non-invasive and continuous monitoring of haemoglobin levels - SpHb
Alternatively also measurement of SpHb + special sensor for SpHb measurement can be ordered.
Learn more about the built-in Masimo rainbow SET here: www.masimo.com/home/rainbow-pulse-
co-oximetry
Apart from O2 saturation, patient’s pulse is also measured in the SpO2 mode. Pulse rate measurement
is based on the optical detection of a peripheral flow.
After switching the SpO2 mode on, the operator may switch on an audible signal accompanying the
pulse indication. This audible signal changes depending on the SpO2 value. The lower SpO2, the lower
the tone. The sound signal accompanying the heartbeat rhythm may be switched off.
PI indicates the strength of the arterial pulse signal. Pi may be used as a diagnostic tool during low
perfusion for the accurate prediction of illness severity.
PVI provides a non-invasive way to determine whether the patient responds to fluids.
In addition, PVI allows the detection of dehydration in patients. PVI may also monitor functional
haemodynamics or forces involved in blood circulation in patients.
PVI is a measure of dynamic changes in the perfusion (PI) that occur during one or more complete
respiratory cycles, and may exhibit changes that reflect physiological factors such as vascular tone,
circulating blood volume, and intravenous pressure excursion.
LIST OF AUTHORISED SENSOR TYPES AND SENSOR PATIENT CABLES FOR SpO2 MONITORING
CAUTION !
Sensors M-LNCS inf, M-LNCS Neo, M-LNCS Neo-Pt, R1 20L and R1 25L are for
single use only.
In case of re-use of these sensors, the following may occur:
- transmission of infection among patients
- faulty measurement due to poor sensor positioning
- faulty measurement due to sensor damage during washing.
Information on the range of wavelengths of light emitted by the probe can be very useful for
doctors performing photodynamic therapy.
The Pulse Oximeter is calibrated to show functional oxygen saturation.
CAUTION !
When installing a specific sensor and an extension cable, refer to the Masimo
package leaflet information.
Signal SIQ bar is a signal quality indicator, most useful during motion and low perfusion situations.
The bargraf rises and falls with the pulse, its height indicating signal quality.
confirm with .
9.8.3 SpO2 - switching the alarm on and selection of the alarm type
The operator may choose from 4 alarm modes
selection with .
18 .
No. 18 .
No. 18 .
The chosen SpO2 and pulse alarm type symbol , , , is displayed in the upper right hand
corner of the SpO2 window.
SpHb alarms are displayed on the bottom line of the SpO2 middle window.
In case of LOW / HIGH SpHb alarm, the text of the relevant alarm is displayed and the operator is
also notified of the deviation of the measured value by flashing the measured data (eg SpHb 15,0%).
The chosen PI and PVI / SpHb alarm type symbol , , , is displayed in the upper right hand
corner of the SpO2 window.
Other MASIMO alarms are displayed at the bottom of the SpO2 window.
No. 16 . The muting of the acoustic alarm is signalled by flashing the crossed-out speaker symbol
When a new alarm is triggered, the muting of the acoustic alarm is automatically cancelled.
Weighing modes:
WEIGHT - patient weighing.
This mode enables direct displaying of patient’s weight.
- measuring weight increase/decrease from the last tare setting.
This mode enables direct displaying of patient weight increase/decrease.
TARE - setting the weight on the bed to zero.
CAUTION!
1. When declining the bed in the WEIGHT mode the patient’s position on the bed
has to be watched. No part of the baby’s body must exceed the bed edges.
2. Handling the bed in the WEIGHT mode is not allowed, the scales system may
be damaged.
WARNING:
1. The maximum range of the scales is 6kg.
2. To comply with the high accuracy weight measurement in the NEWBORN
WEIGHING mode it is necessary to lay a newborn in the center of the bed.
3. To comply with the high accuracy weight measurement in the NEWBORN
WEIGHING mode it is necessary that the bed does not touch the front or rear
door.
Press key No. 7 (PATIENT WEIGHT) and hold it for about 2 seconds. The electronic system
automatically shifts the bed to the weighing position.
The instructions displayed in the individual screens of the WEIGHT menu have to be followed:
- LIFT PATIENT UP
- PUT PATIENT DOWN
The incubator electronic system sets the zero automatically before weighing.
WARNING:
To set the scales to zero the change of the weight on the bed must be at least
300 g.
.
The current measured patient’s weight is
displayed.
After selecting MANUAL it is necessary to press key No. 15 to lock the keyboard. To unlock the
After selecting AUTO 1min - 5min, the buttons will be locked automatically 1 - 5min after the last
WARNING:
The incubator may only be switched off of there is no patient inside.
Press the main switch on the side of the incubator. Switching off is signalled by illumination of the
main switch button.
selection with .
A field containing instructions for the particular window is displayed at the bottom of the menu
screens.
The incubator electronic system monitors the time from the last button pressing on all the menu
screens. Unless none of the keys , , , is pressed within approx. 10 sec an audible alarm
.
Choose and confirm SELECT PATIENT, to
choose a patient from the list in memory.
Choose and confirm SHOW PATIENT to choose
a patient from the list to display a patient
card.
Choose and confirm NEW PATIENT, to enter
data of a new patient.
Choose and confirm EDIT PATIENT, to modify
patient’s data
Choose and confirm DELETE PATIENT, to
delete a patient from the list.
Choose and confirm BACK, to return to the
previous screen.
Choose and confirm CANCEL, to close the
PATIENT MENU.
selection with .
After selection and confirmation the screen shows
the name, surname, date/time of birth.
, .
press .
with .
To return to the basic screen choose CANCEL and
confirm with .
with .
To display the trend in the form of numeric
Division of the time line depends on the chosen trend graph zoom.
with .
To return to the basic screen choose CANCEL and
confirm with .
with .
To return to the basic screen choose CANCEL and
confirm with .
10.4 Data
The incubator enables transfer of patient data
on and from a USB flash disk. This function
serves for transferring data (trends, alarms)
between incubators so only the data of one
patient may be stored on one USB flash disc.
This enables uninterrupted data recording when
a patient is transferred from one SI-610-1 to
another.
confirm with .
WARNING:
Take the patient out of the incubator before performing the test.
The incubator test checks the important functions of the incubator. The incubator electronic system
is checked within the test. In the second test phase the operator has to press the appropriate keys
upon prompts displayed on the LCD display. The operator is guided through the test by messages on
the LCD display.
The incubator test may be performed by confirming the TEST item in the menu.
with .
To return to the basic screen choose CANCEL and
confirm with .
To store the new date and time values choose SET and press to confirm.
Choose the required setting option with or . Confirm your selection with .
If you choose FACTORY SETTING YES, the factory values will be set.
No. 19 or No. 20 .
key No. 16 .
The audible alarms LOW TEMPERATURE, LOW O2 CONCENTRATION, LOW RH are automatically muted
for 30 minutes: - during temperature, O2 concentration or relative humidity start
- after opening the chamber front/rear door
- after a change of temperature, O2 concentration or relative humidity (RH)
The incubator control system differentiates between severities of the individual alarms by
colour and alarm sound.
The visual signal colour identifies alarm priorities as follows:
Red: high priority = immediate operator’s intervention is necessary
Yellow: medium priority = immediate operator’s intervention is not necessary
Audible alarm corresponds with the alarm colour:
For red alarm: … .. … .. … .. = immediate operator’s intervention is necessary
For yellow alarm: … … … = immediate operator’s intervention is not necessary
AUDIBLE FAILURE
FAILURE VISUAL INDICATION REMEDY
INDICATION CAUSE
-Intermittent tone. Failure of the - Check power cord connection.
Red ALARM + -Audible alarm may be incubator internal - The patient has to be transferred
Power failure
POWER FAILURE muted for 5 min by power supply to another incubator, the incubator
SPEAKER key system switched off and a serviceman called
-Intermittent tone. The battery has -The patient has to be transferred to
Red ALARM + -Audible alarm may be not been charged
Battery failure another incubator, the incubator
BATTERY FAILURE muted for 5 min by since the initial 30
SPEAKER key minute charging switched off and a serviceman called
Red ALARM + - Indication
module failure
INCUBATOR - Beta module
FAILURE failure
Failure of the
Red ALARM +
indication module
CLOCK FAILURE clock
Red ALARM + Failure of data
DATABASE recording to the
FAILURE memory
Red ALARM +
Life function
RESUS MODULE module failure
FAILURE
Red ALARM +
INCUBATOR EEPROM failure.
FAILURE 100
Red ALARM +
-Intermittent tone -The patient has to be transferred to
INCUBATOR RTC failure
-Audible alarm may be another incubator, the incubator
System failure FAILURE 101 muted for 5 min by switched off and a serviceman
Red ALARM + pressing SPEAKER key called
Disconnection of
INCUBATOR inner connectors
FAILURE 102
Red ALARM + Ventilator speed
INCUBATOR measurement
FAILURE 103 failure
Red ALARM + Voltage
INCUBATOR measurement
FAILURE 104-108 failure
Red ALARM +
Battery converter
INCUBATOR failure
FAILURE 110
Red ALARM +
INCUBATOR Alarm unit failure
FAILURE 120
Red ALARM + Temperature
TEMPERATURE control failure
Red ALARM + - Intermittent tone. The speed of the - The patient has to be transferred to
Ventilator - Audible alarm may be ventilator has
INCUBATOR another incubator, the incubator
failure muted for 5 min by dropped below
FAILURE 109 pressing SPEAKER key 500 rpm. switched off and a serviceman called.
The SpO2 monitor is equipped with acoustic and visible signalling of several levels of parameters in
compliance with the states of sensors and the unit. The alarms have different audible component
(tone and modulation) according to the state of the patient and the device.
Operator can set alarm priority to 4 levels , in relation to the monitored parameters
(SpO2).
Acoustic signal of the appropriate alarm may be muted for 5 minutes by pressing Key No. 11 .
Acoustic alarm mute is indicated by flickering symbol (yellow loudspeaker), (red loudspeaker)
in the appropriate windows SpO2.
SpO2 alarms displayed at the top and middle part of the SpO2 window
VISUAL AUDIBLE FAILURE
FAILURE REMEDY
INDICATION INDICATION CAUSE
The following maintenance has to be performed to ensure correct operation of the incubator:
CAUTION!
1. In the case of accidental contamination of the unit with biological material,
particularly with blood, it has to be disinfected with an agent with virucidous
effect immediately.
2. Because of oxygen atmosphere oils and greases must not be used.
WARNING:
1. When the incubator is put out of operation the water tank has to be drained
and cleaned.
2. The incubator has to be warmed at least to the ambient temperature when
disinfected.
3. Do not use disinfection agents containing alcohol, acetone or similar organic
solvents for decontamination and cleaning of the incubator, as they may
damage the materials used.
B/ Cleaning and disinfection of the parts getting in touch with the patient (to be
performed by trained operators)
WARNING:
The heating element inside the incubator bath may be hot after operation. Clean
and disinfect the heater area at least after 30 minutes after switching the unit off.
Clean and disinfect all the dismantled parts and the whole bath space.
The rubber sealing bushings have to be disinfected with a virucid agent at least once a week.
CAUTION!
1. Do not press on the optical sensor when cleaning and disinfecting.
2. No residues of cleaners or disinfectors must remain on the optical sensor after
cleaning and disinfection.
WARNING:
1. The water in the humidifier chamber and the heater inside the lower part of
the incubator humidifier chamber may be hot after operation.
2 The humidifier chamber may be drained no earlier than after 30 seconds from
switching the incubator off.
3. Clean and disinfect the heater area at least after 30 minutes after switching
the incubator off.
CAUTION !
1. It is forbidden to use hydrochloric acid products that can damage the stainless
steel heating board (corrosion).
2. When using a specific product, read the package leaflet. Do not use agents
damaging stainless steel.
3. The whole humidifier has to be disinfected with a virucid agent or in ethylenoxide vapour at least
once a week.
CAUTION !
Do not use chlorine disinfectants to disinfect the humidifier reservoir. Chlorine can
damage the stainless steel heating board (corrosion).
4. After the cleaning let the humidifier and reservoir parts air so as the remnants of disinfection and
cleaning agents evaporate.
5. Sterilize the humidifier tank with the upper lid and the bushings at 121 °C, for 10 – 15 min at 210
– 220 kPa.
After cleaning and disinfection assemble the humidifier tank in reverse order.
WARNING:
1. It is forbidden to immerse the probe in the disinfectant solution. The probe
can only be wiped with a disinfectant solution.
2. Be careful, cleaning and disinfection agents must not get into the probe plug.
WARNING:
Be careful, cleaning and disinfection agents must not get into the holes in the
probe plastic cover.
WARNING:
Be careful, cleaning and disinfection agents must not get into the holes in the
probes.
WARNING:
1. See the cleaning instructions in the directions for use for the Masimo re-
useable sensors.
2. Unless otherwise specified, do not sterilize sensors or patient cables by
irradiation, steam, autoclave or ethylene oxide.
3. It is forbidden to immerse the probe in the disinfectant solution. The probe
can only be wiped with a disinfectant solution.
4. Be careful, cleaning and disinfection agents must not get into the probe plug.
CAUTION!
1. Alcohol-containing products (ETHANOL, PROPANOL ...) should not be used to
disinfect the plexiglass parts of the incubator. This may cause cracking of these
parts.
2. It is permitted to use agents based on hydrogen peroxide and polyhexa
methyleneguanidine hydrochloride intended for plexiglass eg.
3. It is necessary to adhere to legal regulations of the country where the incubator
is being used and to use approved cleaning and disinfection agents.
4. When using disinfection and cleaning agents always adhere to application
regulations and recommendations defined by the manufacturer of the particular
agent.
5. Sterilization of components in a steam sterilizer is primarily intended for
cleaning and disinfection. The treated parts may not meet the requirements for
sterility (cannot be classified as a sterile).
6. Repeated steam sterilization causes degradation of used materials. Parts
intended for steam sterilization have a limited lifetime. The estimated service life
is 60 sterilization cycles (or 6 months).
WARNING:
The prescribed temperature distribution in the patient space might not be
achieved or the heater might get overheated without the ventilator cover.
WARNING:
In case of the wrong assembling of all parts, the bed cannot be pulled out.
CAUTION!
Never use the incubator without having checked the correct assembly of the
bed.
WARNING:
The required temperature distribution in the patient space might not be reached
if the upper plexiglass cover is not installed properly.
CAUTION! The incubator must not be used without the air filter.
WARNING:
Used batteries disposed with household waste may harm the environment
seriously. They release harmful substances, particularly so called heavy metals
after some time, which may contaminate soil or ground and surface water and
have demonstrably harmful influence on human health.
Recycling metal substances contained in batteries may bring substantial
energetic and material savings of raw materials.
Our batteries, their packaging or electric devices containing batteries are
marked with a crossed-out dustbin symbol. The symbol means that the battery
does not belong to a dustbin or a household waste container.
Throwing batteries away into household waste or in the country is strictly
forbidden.
WARNING:
After unpacking it is necessary to let the device warm up to the ambient
temperature min. 1 hour.
Optional accessories:
protection connection cable ABB – 0299-0-0032/3 50299000323
air filter PF10 597700131210
body temperature probe 21FF 846 314
oxygen sensor 5MLF16ITG
Consumables which will be replaced within technical safety inspections are paid for by the user of
the device.
Guarantee or after guarantee reports may only be performed by the manufacturer or a servicing
organization authorized by the manufacturer. The servicing organization must have the appropriate
certificate of servicing activities. Such a certificate is issued by the manufacturer.
Inspection may only be performed by the manufacturer or a servicing organization authorized by the
manufacturer.
SpO2 functionality is to be verified only by the manufacturer's authorized personnel or manufacturer-
approved service organization, meters and testers has to be approved by the manufacturer.
Any information on supplies and service may be obtained at the following addresses:
Manufacturer address
TSE spol. s r.o.
Mánesova 74
371 52 České Budějovice
Czech Republic
www.tse-medical.cz
WARNING !
1. Incubator is intended for use by healthcare professionals only.
2. Incubator may cause radio interference or interrupt operation of nearby
device. It may be necessary to re-orient or relocate close devices located
around the incubator or shielding the incubator site.
3. Do not use the incubator in close proximity to other devices. When used
in close proximity or placed on top of each other, the operator must
observe the behavior of the devices in order to verify normal operation in
the configuration in which they will be used.¨
4. Only the probes and cables supplied by the manufacturer or original spare
parts from the manufacturer may be connected to the incubator. Failure
to use original parts may result in increased EMC emissions or lower EMC
resistance of the incubator.
5. It is forbidden to use mobile phones within 2 meters of the incubator! In
a strong electromagnetic field, the function of the incubator electronics
may be interfered with.
Electromagnetic radiation
Applied part types BF- body temperature probes T1,T2, SpO2 probe
Appliance type Class I
TEMPERATURE MODE
Two operation modes enabling temp. measurement and regulation according to air or body
temperature:
Incubator warm-up time by 11 ºC
30 minutes
(according to ČSN IEC601-2-19)
WEIGHING MODE
Enables patient weighing:
Resolution 1g
±2 g ±1 digit (from 0 kg to 1 kg)
Scales accuracy ±5 g ±1 digit (from 1 kg to 3 kg)
±10 g ±1 digit (from 3 kg to 6 kg)
weight trends
Records
weight increments
1. SpCO and SpMet accuracy was determined by testing on healthy adult volunteers in the
range of 60-100% SpO2, 0-40% SpCO, and 0-15% SpMet against a laboratory CO-Oximeter.
SpO2 and SpMet accuracy was determined on 16 neonatal NICU patients ranging in age from
7-135 days old and weighing between 0.5-4.25 kg. Seventy-nine (79) data samples were
collected over a range of 70-100% Sp02 and 0.5-2.5% MetHb with a resultant accuracy of
2.9% SpO2 and 0.9% SpMet.
2. The Masimo sensors have been validated for no motion accuracy in human blood studies on
healthy adult male and female volunteers with light to dark skin pigmentation in induced
hypoxia studies in the range of 70-100% SpO2 against a laboratory CO-Oximeter and ECG
monitor. This variation equals plus or minus one standard deviation. Plus or minus one
standard deviation encompasses 68% of the population.
3. The Masimo sensors have been validated for motion accuracy in human blood studies on
healthy adult male and female volunteers with light to dark skin pigmentation in induced
hypoxia studies in the range of 70-100% SpO2 against a laboratory CO-oximeter and ECG
monitor. This variation equals plus or minus one standard deviation which encompasses 68%
of the population.
4. The Masimo SET Technology has been validated for low perfusion accuracy in bench top
testing against a Biotek Index 2 simulator and Masimo's simulator with signal strengths of
greater than 0.02% and transmission of greater than 5% for saturations ranging from 70 to I
00%. This variation equals plus or minus one standard deviation which encompasses 68% of
the population.
5. The Masimo sensors have been validated for pulse rate accuracy for the range of 25-240
bpm in bench top testing against a Biotek Index 2 simulator. This variation equals plus or
minus one standard deviation which encompasses 68% of the population.
6. SpHb accuracy has been validated on healthy adult male and female volunteers and on
surgical patients with light to dark skin pigmentation in the range of 8-17 g/dl SpHb against
a laboratory CO-oximeter. This variation equals plus or minus one standard deviation which
encompasses 68% of the population. The SpHb accuracy has not been validated with motion
or low perfusion.