Download as pdf or txt
Download as pdf or txt
You are on page 1of 200

x

| Contents |

Contents

1 Important
2 Introduction
3 Description of functions
4 Disassembling and assembling
5 Service procedures
6 Troubleshooting
7 Preventive maintenance
8 Index
9 Revision history
10 Diagrams

SERVO-air Ventilator System, Service Manual 0-1


| Contents |

0-2 SERVO-air Ventilator System, Service Manual


| Important | 1 |

1 Important
Responsibility for the safe functioning of the equipment
1.1 General reverts to the owner or user in all cases in which
service or repair has been done by a non-professional
The information in this Service Manual is based on or by persons who are not employed by or authorized
the SERVO-air Ventilator System of System by Getinge, and when the equipment is used for other
version 4.2. Unless otherwise stated, the Service than its intended purpose.
Manual is valid also for System versions below 4.2.
The system complies with standards and requirements
Maquet Critical Care AB, part of Getinge, is the legal as stated in the User's Manual.
manufacturer of the SERVO ventilator systems.
WARNING! Do not modify this
Service documentation for the SERVO-air Ventilator
equipment without authorization of the
System consists of:
manufacturer.
• User'slist
bullet Manual including Cleaning and Maintenance
manual. The User's Manual is an indispensable
complement to the Service Manual for proper System version and System software version
servicing.
The System version number can be found in the
• Service
bullet listManual
SYSTEM STATUS window on the User interface.
• Installation
bullet list Instructions Make sure that the version of the User’s Manual
• Sparelist
bullet parts information corresponds to the System version.
• Documentation
bullet list for optional equipment included in
The System SW version number can be found in the
the system
SYSTEM STATUS window on the User interface. For
• Product
bullet listEnd-of-Life - Disassembly Instructions functionality enhancement, the latest released System
(Disassembly and Disposal Instructions).
SW version is always recommended.

The SERVO-air Ventilator System is referred to as


'the system' in this document. Serial number
TTTTTT
The User’s Manual for the SERVO-air Ventilator There are two serial number labels on the unit:
System is referred to as ‘the User's Manual’ in this
document.
• One label
bullet list is attached to the Patient unit above the
mains power supply connector. The serial number
stated on this label is the ID number of the Patient
AAAAFAFFFAFF
Maquet Headquarter Support Center is referred to as
‘MCC HSC’ in this document. Note that MCC HSC is unit. The serial number is also stored in the software
available only for the Getinge service organization. memory as the 'System ID'.
For other service organizations, for example hospital • One label
bullet list is attached on the rear side of the User
biomedical departments, technical support is provided interface. The serial number stated on this label is
by the local Getinge representative. the ID number of the User interface.

In addition to the Important information given here and


in the related documents, always pay attention to
applicable local and national regulations.
R R

SERVO-air Ventilator System, Service Manual 1-1


| 1 | Important |

1.2 Symbols used in this manual 1.3 Hazard notices


Before disassembling or assembling the system, make
WARNING! Indicates critical
sure that:
information about a potential serious
outcome to the patient or the user.
• The On/Off
bullet list switch is set to Off.
• Mainslist
bullet power cable is disconnected.
CAUTION: Indicates instructions that
• External
bullet list 12 V battery is disconnected.
must be followed in order to ensure the
proper operation of the equipment.
• Backup
bullet listbattery modules are disconnected. Note
that the backup battery modules supply power to
parts of the system also when the system is
Note: ESD sensitive components. switched Off.
When handling ESD-sensitive devices, • Gas supply
bullet list is disconnected, central gas supply and
established procedures must be backup gas supply.
observed to prevent damage.
• The system,
bullet list including all gas conveying parts, is
cleaned. Refer to instructions in the User’s Manual.
Note: Special waste. This product
WARNING! With gas supply connected
contains electronic and electrical
to the system, there are pressurized
components. Discard disposable,
components inside the unit. All
replaced and left-over parts in
personnel must exercise extreme
accordance with appropriate industrial
caution if fault tracing or adjustments
and environmental standards.
are performed with gas supply
connected and with covers removed.
Note: Recyclable material. Recycling
must be performed in accordance with
WARNING! With power supply
appropriate industrial and
connected to the system, there are
environmental standards.
energized electrical components inside

Note: Trained and authorized


personnel. Only personnel trained and
authorized by Getinge shall be
permitted to perform installation,
AAAAAA the unit. All personnel must exercise
extreme caution if fault tracing or
adjustments are performed with power
supply connected and with covers
removed.
DDRDRDDDRRRR
service or maintenance of the system.
CAUTION: Do not expose the batteries
Note: Text inside a box is used to highlight to water, fire or excessive heat. Do not
important information. crush, disassemble, puncture or short
circuit the connector terminals.

1-2 SERVO-air Ventilator System, Service Manual


| Important | 1 |

1.4 Installation 1.6 Service

Note: Trained and authorized WARNING! The system must not be


personnel. Only personnel trained and serviced or maintained while in use with
authorized by Getinge shall be a patient.
permitted to install the system. The
installation and handing-over
Note: ESD sensitive components.
procedures are described in the
When working with ESD sensitive
Installation instructions.
components, always use a grounded
wrist band and a grounded work
surface. Adequate service tools must
1.5 Functional check always be used.

After any installation, maintenance or service


intervention in the system, perform a Functional check. Note: ESD sensitive components.
The Functional check is divided into two steps: PC boards (spare parts) must always
be kept in a package for sensitive
1. Internal leakage test: The Internal leakage test
electronic devices. Getinge will not
must be performed without O2 connected to the
otherwise assume responsibility for the
system. The test can be performed either as a materials used, the work performed or
separate Internal leakage test in Service & any possible consequences of same.
Settings or by running a Pre-use check. The
Internal leakage test must be successfully passed
before starting step 2 of the Functional check. Note: Special waste. This system
Note: If running a Pre-use check, make sure that contains electronic and electrical
the Internal leakage test was successfully passed. components. Discard disposable,
Other tests can be cancelled as they may fail replaced and left-over parts in
when O2 is not connected. accordance with appropriate industrial
and environmental standards.
2. Pre-use check: Connect O2 and perform a
complete Pre-use check according to instructions
in the User’s Manual. Note: Recyclable material. Recycling
must be performed in accordance with
appropriate industrial and
environmental standards.

Note: Trained and authorized


personnel. The system must be
serviced at regular intervals by
personnel trained and authorized by
Getinge. It is recommended that service
and maintenance is done as a part of
a service contract. Any maintenance
or service must be noted in a log book.

SERVO-air Ventilator System, Service Manual 1-3


| 1 | Important |

Note: All repair or service requests resulting from


1.8 Environmental declaration
a device failure to perform their intended function 1.8.1 Purpose
shall be submitted as complaints.
Please report a potential complaint to your local This environmental declaration applies to a SERVO-air
Getinge representative as soon as possible for basic system including the Mobile cart and one Battery
further processing. module.

Maintenance of the system must be performed by 1.8.2 Components with special environmental
Getinge authorized personnel with the following concern
intervals:
Components listed below shall be disposed of in
• Preventive
bullet list maintenance at least once a year, or accordance with appropriate industrial and
every 5000 hours of operation, whichever comes environmental standards.
first.
• The Battery
bullet list modules shall be replaced after Printed circuit boards
300 charge cycles or five years (60 months) from
manufacturing date, whichever comes first. • PC 1781
bullet list Pressure transducer, 2 pcs

Note: For system version 4.1 or lower the battery


• PC 1991
bullet list Control, including battery

life time is four years (48 months) from • PC 1992


bullet list Monitoring, including battery
manufacturing date. • PC 1993
bullet list Ethernet switch
• PC 1994
bullet list Nebulizer
• The Memory
bullet list backup batteries shall be replaced • PC 2004/PC
bullet list 2024 Expiratory channel
every five years.
• PC 2020
bullet list Main backplane
• The Turbine
bullet list module shall be replaced every 10
• PC 2021
bullet list Rear connectors
years.
• PC 2023
bullet list Panel

in the system.
TT
Only original spare parts from Getinge must be used •


PC 2025
bullet
PC 2026
bullet
PC 2027
bullet
list Battery back-plane
list Battery extension
list Motor control
AAFFAAFF
1.7 To the responsible service personnel • PC 2028
bullet list DC/DC & Battery control

Note: Trained and authorized


• PC 2029
bullet list Inspiratory channel extension

personnel. Only personnel trained and • PC 2030


bullet list Inspiratory channel
authorized by Getinge shall be • PC 2031/PC
bullet list 1785 Expiratory channel connector
permitted to perform installation,
service or maintenance of the system. Other electronics
T
The contents of this document are not binding. If any • LCD display
bullet list
T

RARARARAAAAA
Touch panel (glass)
DDRR

significant difference is found between the system and bullet list


this document, please contact Getinge for further • Turbine
bullet listmodule
information.
• O2 cell,
bullet listcontaining caustic lime and lead (Pb)
We reserve the right to modify the system without • O2 gaslistmodule, containing multiple PC boards
bullet
amending this document or advising the user. • Expiratory
bullet list cassette, containing PC board and other
electronics
• Expiratory
bullet list valve coil

1-4 SERVO-air Ventilator System, Service Manual


| Important | 1 |

• Safetylist
bullet valve pull magnet • Battery
bullet listmodule Li-ion
• AC/DC
bullet listconverter, containing PC board
Others – very small amounts
• Battery
bullet listmodule Li-ion
• Mainslist
bullet power supply connector • Filter paper
bullet list of fiberglass
• Fans, list
bullet 2 pcs
• Loudspeaker
bullet list 1.8.4 Articles of consumption

• Cablelist
bullet harness (handle as electronic waste)
1. Bacteria filter
2. Dust filter for the Air inlet filter
1.8.3 Construction materials
3. Filter for the O2 gas module
The construction materials used in the system, in % 4. Filter for the Inspiratory pressure transducer
of the total weight. 5. Nozzle unit for the O2 gas module

Metal – total 70% 6. Air inlet filter for the Turbine module

TTTTTTTTTTTTRTTR TTTTTTT
7. Battery modules
• Aluminium
bullet list 40%
8. Turbine module
• Steel,list
bullet zink, brass 30%
9. Memory backup batteries
Polymeric material – total 13% 10. Expiratory cassette
AFAFAFAFFFAFAFFFAAAAAFAFAFFFFF 11. Expiratory cassette membrane
• ABS (Acrylicnitrilebutadienstyrene)
bullet list
12. O2 cell including filter
• EPDMlist
bullet (Ethylenepropylenedienemonomer)
Item 1: Consumption approx. 250 pcs/year.
• FPM (Fluororubber)
bullet list

AA
• LDPElist
bullet (Low-density polyethylene) Item 2: Consumption approx. 15 pcs/year.

• NBR (Nitrilerubber)
bullet list Items 3 – 6: Replaced once a year or every 5000 hours
• PA (Polyamide)
bullet list of operation, whichever comes first.
• PBT (Polybutylene
bullet list terephthalate)
R
Item 7: Replaced after 300 charge cycles or five years
• PC (Polycarbonate)
bullet list (60 months) from manufacturing date, whichever
R
• PE (Polyethylene)
bullet list comes first.
R

• PEEKlist
bullet (Polyether ether ketone)
R

Item 8: Replaced every 10 years.


• POM (Polyoxymethylene)
bullet list
R
D
D
• PP (Polypropylene)
bullet list Item 9: Replaced every five years.
DR

• PTFElist
bullet (Polytetrafluoroethylene)
R

Items 10 – 12: Replaced when needed.


D

• PS (Polystyrene)
bullet list
D
D
• PUR (Polyurethane)
bullet list Articles related to clinical applications, e.g., patient
tubing and nebulizer medication cups are not included
D

• PVC (Polyvinyl
bullet list chloride)
in the list above.
D

• SI (Silicone)
bullet list
D

• TPE (Thermoplastic
bullet list elastomer) 1.8.5 Power consumption

Electronics – total 15% The power consumption depends on the operating


R

mode and Patient unit settings.


• Printed
bullet listcircuit boards, cables etc.
R

• Typical
bullet listminimum power consumption (standby):
Battery – total 2% 26 W
SERVO-air Ventilator System, Service Manual 1-5
| 1 | Important |

• Typical
bullet listpower consumption: 36 W 1.8.8 Product End-of-Life
Pressure control (intended delivered volume 500
ml) with PEEP=5 cmH2O, PAP=10 cmH2O, RR=20 For disassembly and disposal (scrapping) information,
breaths/min, FiO2=30%. Pressure/volume settings, refer to the document SERVO-air Ventilator System
Product End-of-Life Disassembly Instructions.
respiratory rate, air/oxygen mix, leakages and/or
other factors will affect power consumption and
battery backup time. See section Battery modules
for approximate recharge times.
• Typical
bullet listmaximum power consumption: 185 W

1.8.6 Noise level

A-weighted sound pressure level (LpA): <49 dB,


measured at a distance of 1 m.

FFTTTFFFTTT FTFT
A-weighted sound power level (LWA): <57 dB.

1.8.7 Packaging material

Materials for packing:




Loading
bullet listpallet.
Corrugated
bullet
bullet
list cardboard.
Shock-absorbing
list
A
material of expanded
polypropylene, EPP.

Materials added when packing for tropical climatic


conditions:
DRDR

• Plywood
bullet list container with metal reinforcements.
• Plasticlist
bullet wrapping.
• Desiccant
bullet list (humidity absorber).
A
The packaging material complies with the demands
stated by the International Plant Protection Convention
(IPPC).
AAA

1-6 SERVO-air Ventilator System, Service Manual


| Introduction | 2 |

2 Introduction

Table of Contents
2.1 Main units | 2-2
2.2 Patient unit | 2-4
2.3 User interface | 2 - 10
2.4 Software structure | 2 - 12
2.5 Internal communication | 2 - 14

SERVO-air Ventilator System, Service Manual 2-1


| 2 | Introduction |

2.1 Main units


The system is configured for different patient The system can be divided into the following main
categories: units:
• Pediatric
bullet list • Patient
bullet listunit. The Patient unit contains pneumatics
and electronics for gas supply to the patient. Power
• Adult list
bullet
supply and battery backup is also contained in the
These patient categories are as standard equipped
Patient unit.
with a number of ventilation modes suitable for each
category. Additional ventilation modes can be installed • User interface.
bullet list The User interface contains all
via a Software option installation. controls used to set the ventilation and monitoring
parameters. Ventilation parameters as well as other
important information are shown on the User
interface display.
The User interface cable connects the User interface
and the Patient unit.

2.1.1 System mounted on the optional Mobile cart

DDRR TT s

d
AAFF
f h

AA a g k l

;
j
DDRR

S20002

1. Patient unit 7. Battery compartments


2. User interface 8. Patient circuit
3. Expiratory inlet with moisture trap 9. AC mains power
4. Inspiratory outlet 10. O2 supply
5. Emergency air intake 11. Wheel lock
6. Air inlet for turbine

2-2 SERVO-air Ventilator System, Service Manual


| Introduction | 2 |

2.1.2 Optional accessories

There are also several accessories to be used with


the system, e.g.:

• Alarmlist
bullet output
• Battery
bullet listmodule (extra)
• Battery
bullet listcharger
• Mobilelist
bullet cart
• Mobilelist
bullet cart drawer kit
• Cabinet
bullet list(for Mobile cart)
• Gas cylinder
bullet list restrainer kit
• Shelf base
bullet list
• Support
bullet listArm 179
• Y piece
bullet listholder and hook
• Cablelist
bullet holder
• Aerogen
bullet list nebulizers (Aerogen Pro and
Aerogen Solo)




bullet
bullet
bullet

TTTTTTTTTTTTTTTTTT
Servolist
DuoGuard and Servo Guard filters
Expiratory
list heater, Servo DuoGuard
Thermoshell
list for expiratory cassette
AAAAFAFFAFAFAFAAFAFAFAFAFAFAFAFFFFF
Humidifier
bullet list holder
• Waterbag/IV
bullet list pole for humidifier holder
• MSynclist(adapter for HL7 communication)
bullet

For further information regarding accessories, contact


your Getinge representative.
DDRDRDRDRDRDRDRDRRRRR

SERVO-air Ventilator System, Service Manual 2-3


| 2 | Introduction |

2.2 Patient unit


Items accessible from the outside of the Patient unit are shown in this illustration.

s
F G
d D K

f S H

h j k l ; A J

S20003

1. On/Off switch. The switch must be pulled 11. Battery compartments


downwards before it can be switched. 12. USB ports
2. Expiratory outlet 13. Gas inlet for O2
3. Power indicators 14. Gas inlet for air including air inlet filter
4. RS-232 connectors 15. Inspiratory outlet
5. Equipotentiality terminal 16. Emergency air intake
6. AC mains power supply connector with fuse 17. Nebulizer connector
7. Alarm output connector 18. Expiratory inlet
8. External 12 V battery 19. Cooling fan air inlet with filter (on both sides)
9. Fuse for External 12 V battery 20. Expiratory cassette
10. Ethernet connector 21. Expiratory inlet with moisture trap

2-4 SERVO-air Ventilator System, Service Manual


| Introduction | 2 |

2.2.1 Main section

The main section is accessed by removing the top cover and the lid on the PC board rack.

s
g
d

f
h

j
S20004

No. Description Remark


1 PC 1993 Ethernet switch Part of subsystem Monitoring MON. Controls the Ethernet communication
in the system.
2 PC 1992 Monitoring Part of subsystem Monitoring MON. Main functions are alarm and
monitoring, calculations, trending of measured values and is also CAN-bus
master. Includes a Memory backup battery.
3 PC 1991 Control Part of subsystem Breathing BRE. Main functions are responsibility for
the patient treatment, i.e. how to regulate the inspiratory and expiratory
gas flow. Includes a Memory backup battery.
4 PC 2004 Expiratory channel/ Part of subsystem Breathing BRE. Main functions are expiratory flow
PC 2024 Expiratory channel measurement and expiratory cassette handling. PC 1781 Expiratory
pressure transducer is connected onto this board. PC 2024 replaces
PC 2004. Introduced in production Q1 2016.
5 Expiratory valve coil connector Cable connector for the Expiratory valve coil.
6 PC 1781 Expiratory pressure Expiratory pressure transducer connected to the Expiratory cassette with
transducer the Expiratory pressure tube.
7 Expiratory channel connector cable Cable connector for PC 2031/PC 1785 Expiratory channel connector.

SERVO-air Ventilator System, Service Manual 2-5


| 2 | Introduction |

2.2.2 PC 2020 Main back-plane

S20005

No. Description Remark


1 PC 2020 Main back-plane The PC boards, as listed above, are directly or indirectly connected to
the PC 2020 Main back-plane.

2-6 SERVO-air Ventilator System, Service Manual


| Introduction | 2 |

2.2.3 Inspiratory section

The left-hand side of the Patient unit contains the Inspiratory section.

d f

s
g

l
S20006

No. Description Remark


1 Turbine module The Turbine module generates compressed air and delivers air to the
inspiratory gas.
2 O2 gas module The O2 gas module is connected to the O2 supply and delivers O2 to the
inspiratory gas.
3 Inspiratory gas docking Connects the Turbine module and the O2 gas module. Contains the
Turbine module one-way valve.
4 Inspiratory flowmeter The inspiratory flowmeter measures the combined air and O2 flow, as
well as the temperature, of the inspiratory gas from the O2 gas module
and Turbine module.
5 Inspiratory flowmeter cable Cable connector for the Inspiratory flowmeter.
6 PC 1781 Inspiratory pressure Inspiratory pressure transducer connected to the inspiratory pipe with
transducer the inspiratory pressure tube.
7 O2 cell cable Cable connector for the O2 cell.
8 PC 2030 Inspiratory channel Connector board for the inspiratory section. PC 1781 Inspiratory pressure
transducer is connected onto this board.
9 O2 evacuation fan Forces cooling air through the Inspiratory section and evacuates O2 gas
in case of O2 leakage.

SERVO-air Ventilator System, Service Manual 2-7


| 2 | Introduction |

2.2.4 AC/DC section

The right-hand side of the Patient unit contains the AC/DC converter and Mains power inlet, as well as the
Expiratory valve coil.

a s

d
g

S20007

No. Description Remark


1 AC/DC Converter Converts the connected AC Power (mains power) to the internal DC
supply voltage.
2 Expiratory valve coil Controls the opening of the expiratory valve by pushing the valve
membrane in the cassette into desired position.
3 PC 2028 DC/DC & Battery control The main functions are On/Off control, control of the battery modules,
control of fans and connection to the User interface. It also supplies
required internal voltages.
4 Main fan Forces cooling air through the AC/DC section and the main section of
the Patient unit.
5 Mains power supply inlet Mains power connector including fuses.

2-8 SERVO-air Ventilator System, Service Manual


| Introduction | 2 |

2.2.5 Expiratory cassette

a s

S20008

No. Description Remark


1 Expiratory cassette
2 PC 2031/PC 1785 Expiratory channel connector PC 2031 must be used together with PC 2004.
PC 1785 must be used together with PC 2024.

The expiratory cassette is a complete unit. It contains


the following parts:

• Expiratory
bullet list inlet with moisture trap.
• PC 1786
bullet list Expiratory channel cassette.
• Ultrasonic
bullet list flowmeter.
• Heating
bullet listfoil to avoid condensation in the Ultrasonic
flowmeter.
• Pressure
bullet list transducer connection, incl. bacteria filter,
to connect the expiratory pressure transducer.
• Expiratory
bullet list valve incl. valve membrane.

PC 1786 Expiratory channel cassette inside the


Expiratory cassette is electrically connected to
PC 2004/PC 2024 Expiratory channel via
PC 2031/PC 1785 Expiratory channel connector.

The Expiratory valve coil, mounted under the


TTTTTT
Expiratory cassette compartment, controls the valve
membrane in the cassette.
AAFAAFAFFFF
R

SERVO-air Ventilator System, Service Manual 2-9


| 2 | Introduction |

2.3 User interface


The User interface is mounted onto the Patient unit
using a 2-joint swivel arm.

The User interface can be tilted up and down into a


suitable position.

For information regarding operation of the User


interface, refer to the User's Manual.

S20009

The illustration below shows the backside of the User The illustration below shows the inside of the User
interface after it is detached from the User interface interface rear cover.
holder.

a s d S20011

No. Description Remark


1 Loudspeaker grid
2 Loudspeaker connector
3 Loudspeaker The loudspeaker is attached with double-coated adhesive tape.

2 - 10 SERVO-air Ventilator System, Service Manual


| Introduction | 2 |

a s

S20012

No. Description Remark


1 Alarm indicator Multicolor LEDs on PC 2023 will flash when an alarm is triggered in the
system
2 PC 2023 Panel Part of subsystem Panel PAN.
3 LCD Display/Touch screen Complete unit including LED backlight.

SERVO-air Ventilator System, Service Manual 2 - 11


| 2 | Introduction |

2.4 Software structure 2.4.2 Breathing BRE

2.4.1 General The Breathing subsystem’s main function is the patient


treatment, i.e. how to regulate and measure the
The System software will contain all available system inspiratory and expiratory gas flow:
functionality. The software is distributed across the
subsystems. The installation procedure ensures that • Controls
bullet list all ventilation modes, ventilator settings
all subsystems are properly updated. and delivery of gas mixtures.

The System software is divided into the following main • Controls


bullet list the expiratory cassette handling including
flow measurement performed by the flowmeter in
subsystems:
the cassette.
• Breathing
bullet list BRE • Nebulizer
bullet list control. Controls the Nebulizer function.
• Monitoring
bullet list MON • Controls
bullet list Pre-use check operations on the Breathing
subsystem
• Panellist
bullet PAN
• Powerlist
bullet POW • Maintain
bullet list a persistent data storage area used to
save parameter settings.
• BlowerlistBLO
bullet
• System
bullet listID
• Controls
bullet list some of the system settings, e.g. storing
default patient category and default mode of
ventilation. The system settings will be stored in
persistent memory.
User interface The Breathing software is stored on PC 1991 Control
PC 2023 and PC 2004/PC 2024 Expiratory Channel. The
software must be re-installed if PC 1991 or
PAN SW
PC 2004/PC 2024 is replaced.

The Breathing software is executed by the


DDRDRDRRDRRTTTTTT PC 2028 POW SW
PC 1994
microprocessors on PC 1991 and PC 2004/PC 2024.

Patient unit

AAAAAA
PC 1786

PC 2029

PC 2020
AAAAAFAFFFFF
SYSTEM ID SW PC 2030
PC 2004/ PC 1781Exp

PC 1781
MON SW
BRE SW

Insp.

PC 2031/
PC 1785
PC 1993
PC 1991

PC 1992
PC 2024

BLO SW
PC 2026

PC 2027
PC 2033 (Turbine
module)

PC 2021 PC 2025
DDDRDRRRRR

S20135

2 - 12 SERVO-air Ventilator System, Service Manual


| Introduction | 2 |

2.4.3 Monitoring MON 2.4.5 Power POW

The Monitoring subsystem’s main function is alarm The Power subsystem has the following main
and monitoring, including trends of measured values: responsibilities:

• Detects
bullet listand handles alarms. Contains a backup • Communication
bullet list with plug-in battery modules to
sound buzzer. determine battery status.
• Monitors
bullet list the system behavior. • Controlling
bullet list management of, and voltage to, the
Inspiratory channel.
• Maintains
bullet list a persistent data storage area to save
Alarm settings. • Controlling
bullet list charging/discharging of plug-in battery
modules.
• Monitors
bullet list some of the system settings, e.g. default
alarm limits. • Selection
bullet list of power source to be used for the
system.
• CAN communication
bullet list with Expiratory channel.
• Battery
bullet listmodule and power source supervision. • Controlling
bullet list Main fan and O2 evacuation fan.

• Pre-use
bullet listcheck: Pre-use check operations on the The Power software is stored on PC 2028. The
Monitoring subsystem software must be re-installed if PC 2028 is replaced.

• External
bullet list communication protocol SCI (SERVO The Power software is executed by the microprocessor
Communication Interface). on PC 2028.
• Optional
list functionality such as ventilation mode and

T RRRRTRTTTTTTTT
bullet
other features. All options are stored in the
persistent memory.
The Monitoring software is stored on PC 1992
Monitoring. The software must be re-installed if
F
PC 1992 is replaced.
F
The Monitoring software is executed by the
F
microprocessor on PC 1992.
2.4.6 Blower BLO

The Blower subsystem is responsible for the


infrastructure, hardware, and services for control of
the blower. It also supplies the BRE and MON
subsystems with sensor data for inspiratory flow and
temperature as well as ambient air humidity and air
inlet filter pressure drop:
RRRARRARAARARARAAA

AAAAA
2.4.4 Panel PAN • Controlling
bullet list and monitoring the Turbine module.
F
• Managing
bullet list flow measurement for the Inspiratory flow
The Panel subsystem controls all user interaction, meter.
F
e.g.:
• Managing
bullet list ambient air temperature and air humidity
F
measurements.
• Presentation
bullet list and changes of ventilator settings:
F
• Presentation
bullet list of metrics, e.g. real-time values from • Enable
bullet listAir inlet filter supervision by measuring air
inlet filter pressure drop.
the Monitoring subsystem.
F
The Blower software is stored on PC 2027 Motor
• Presentation
bullet list of alarms and their corresponding
controller board. The software must be re-installed if
D
D
alarm sounds from all subsystem.
F
Turbine module is replaced.
The Panel software is stored on PC 2023. The
D
D
software must be re-installed if PC 2023 is replaced.
D

The Panel software is executed by the microprocessor


D
on PC 2023.
D
D
D
D
D

SERVO-air Ventilator System, Service Manual 2 - 13


| 2 | Introduction |

2.4.7 System ID 2.5 Internal communication


The System ID is stored on PC 2020 Main back-plane The internal communication is separated into different
and is unique for each system. communication protocols:

When installing software options, the System ID is • ETHERNET


bullet list bus using TCP/IP running at
used to verify that the option is valid for that system. 100 Mbit/s. ETHERNET connects the main
subsystems BRE, MON and PAN using the Ethernet
When replacing PC 2020 Main back-plane, the switch.
replacement (spare part) must be factory programmed
with the correct System ID to ensure proper ventilation • CAN bus
bullet list (Controller Area Network) running at
500 kbit/s. The CAN is primarily used for the
performance, compatibility with future updates and
Expiratory flowmeter.
correct options configuration.
• SPI bus
bullet list(Serial Peripheral Interface) primarily used
for communication between BRE and BLO.
• I2C bus
bullet list(Inter-Integrated Circuit) primarily used for
communication with:
- Local E2PROMS.
- The O2 cell.
- The expiratory cassette.
- The inspiratory flowmeter
- The ambient temperature and humidity sensor.
• Digitallist
bullet and analogue point-to-point signals
connecting subsystems and components.
• The RS-232
bullet list is used for the SERVO Communication
Interface SCI.
• The RS-485
bullet list is used for communication between
the POW and the MON subsystems.

AA AAAA
• The SMBus
bullet list (System Management Bus) is used for
communication with the POW subsystem and the
battery modules.
DDRDRDRR

2 - 14 SERVO-air Ventilator System, Service Manual


| Description of functions | 3 |

3 Description of functions

Table of Contents
3.1 O2 gas module (1) | 3-4
3.2 Turbine module including PC 2027 Motor | 3-6
control (2)
3.3 Inspiratory gas docking (3) | 3-7
3.4 Inspiratory pipe (4) | 3-7
3.5 Inspiratory flowmeter (5) | 3-7
3.6 Inspiratory channel (6) | 3-7
3.7 O2 cell (7) | 3-8
3.8 Safety valve including pull magnet (8) | 3-8
3.9 O2 evacuation fan (9) | 3-9
3.10 PC 2030 Inspiratory channel (10) | 3-9
3.11 PC 1781 Inspiratory pressure transducer (11) | 3 - 10
3.12 PC 2029 Inspiratory channel extension (12) | 3 - 10
3.13 Expiratory cassette (13) | 3 - 10
3.14 Expiratory valve (14) | 3 - 11
3.15 Expiratory valve coil (15) | 3 - 11
3.16 Expiratory outlet (16) | 3 - 11
3.17 PC 1786 Expiratory channel cassette (17) | 3 - 11
3.18 PC 2031/PC 1785 Expiratory channel | 3 - 11
connector (18)
3.19 PC 2020 Main back-plane (19) | 3 - 12
3.20 PC 1781 Expiratory pressure transducer (20) | 3 - 12
3.21 PC 2004/PC 2024 Expiratory channel (21) | 3 - 12
3.22 PC 1991 Control (22) | 3 - 13
3.23 PC 1992 Monitoring (23) | 3 - 13
3.24 PC 1993 Ethernet switch (24) | 3 - 14
3.25 PC 1994 Nebulizer (25) | 3 - 14
3.26 Mains inlet (26) | 3 - 14
3.27 AC/DC Converter (27) | 3 - 14
3.28 PC 2028 DC/DC & Battery control (28) | 3 - 14
3.29 On/Off switch (29) | 3 - 15
3.30 PC 2021 Rear connectors (30) | 3 - 15
3.31 PC 2026 Battery extension (31) | 3 - 16
3.32 PC 2025 Battery back-plane (32) | 3 - 16
3.33 Battery module (33) | 3 - 16
3.34 Main fan (34) | 3 - 17

SERVO-air Ventilator System, Service Manual 3-1


| 3 | Description of functions |

3.35 User interface cable (35) | 3 - 17


3.36 Touch screen (36) | 3 - 17
3.37 LCD Display (37) | 3 - 17
3.38 PC 2023 Panel (38) | 3 - 17
3.39 Alarm indicator (39) | 3 - 18
3.40 Loudspeaker (40) | 3 - 18
3.41 Optional equipment | 3 - 18

3-2 SERVO-air Ventilator System, Service Manual


| Description of functions | 3 |

About this chapter The System electronics contains all PC boards and
electrical components that are not included in the other
The information in this chapter refers to the System main blocks:
overview diagram in chapter Diagrams. Note the
numbering in each heading (in brackets), that 19. PC 2020 Main back-plane
corresponds to the position number in the System 20. PC 1781 Expiratory pressure transducer
overview diagram.
21. PC 2004/PC 2024 Expiratory channel
22. PC 1991 Control
Main blocks
23. PC 1992 Monitoring
The information in this chapter is divided into main 24. PC 1993 Ethernet switch
blocks:
25. PC 1994 Nebulizer
The Inspiratory section conveys the breathing gas 26. Mains inlet
from the gas inlets to the patient breathing system. It 27. AC/DC Converter
contains the following main components: 28. PC 2028 DC/DC & Battery control
1. O2 gas module 29. On/Off switch
2. Turbine module 30. PC 2021 Rear connectors
3. Inspiratory gas docking 31. PC 2026 Battery extension
4. Inspiratory pipe 32. PC 2025 Battery back-plane
5. Inspiratory flowmeter 33. Battery module
6. Inspiratory channel 34. Main fan
7. O2 cell 35. User interface cable

8. Safety valve including pull magnet


9. O2 evacuation fan
The User interface is the main user interface during
10. PC 2030 Inspiratory channel operation of the system. The User interface contains
11. PC 1781 Inspiratory pressure transducer the following main components:
12. PC 2029 Inspiratory channel extension
36. Resistive Touch screen.
37. LCD Display, 12” TFT display with LED backlight.

The Expiratory section conveys the breathing gas 38. PC 2023 Panel
from the patient breathing system to the Expiratory 39. Illuminated alarm indicator on top of the User
outlet. It contains the following main components: interface frame.
40. Loudspeaker
13. Expiratory cassette
14. Expiratory valve
15. Expiratory valve coil
16. Expiratory outlet
17. PC 1786 Expiratory channel cassette (integrated
into the Expiratory cassette)
18. PC 2031/PC 1785 Expiratory channel connector

SERVO-air Ventilator System, Service Manual 3-3


| 3 | Description of functions |

Memory types used in the system 3.1 O2 gas module (1)


There are four different types of memories used in the The O2 gas module regulates the inspiratory O2 gas
system: flow.

• Flashlist
bullet memory. For System software storage.
Present on PC 1991, PC 1992, PC 2004/PC 2024, g f d
PC 2023, PC 2027, PC 2028. The System software
can be re-installed/updated using a System h
software installation.
s
• RAM.list
bullet For temporary storage of software and data.
Present on PC 1991, PC 1992, PC 2004/PC 2024,
PC 2023, PC 2027, PC 2028.
j a
• RAM with
bullet list battery backup. For settings, trends
and logs. Present on PC 1991 and PC 1992.
• EEPROM.
bullet list For PC board information, configuration,
calibration data, etc. Present on several PC boards, k S20014

and in the O2 cell.


1. Gas inlet
2. Filter

TTT T 3.
4.
5.
Inspiratory valve temperature sensor
Supply pressure transducer
Flow transducer (Delta pressure transducer and
AAFAFFAF
wire mesh)
6. Nozzle unit with valve diaphragm
7. Inspiratory solenoid
8. Gas module key

Note: The O2 gas module is factory calibrated. The


O2 gas module must not be disassembled further
than described in chapter Preventive maintenance.
DRDRDRR

3.1.1 Gas inlet

O2 gas supply is connected to the system’s O2 inlet


nipple. The design of the O2 gas inlet nipple varies
according to the standard chosen.

O2 gas is to be connected from hospital central gas


supply or from an O2 gas cylinder.

Refer to Technical data in the User's Manual for O2


gas quality specifications.

3-4 SERVO-air Ventilator System, Service Manual


| Description of functions | 3 |

3.1.2 Filter The complete plastic nozzle unit must be replaced


during the Preventive maintenance. After replacement,
The Filter protects the system from particles in the gas allow the diaphragm to adapt to the valve seat by the
delivered to the O2 gas module. The Filter must be spring tension during approximately 10 minutes before
replaced during the Preventive maintenance. gas pressure is connected to the O2 gas module.

A non-return valve for the O2 gas inlet is located in the


filter cover. This valve will suppress short pressure 3.1.7 Inspiratory solenoid
drops in the O2 gas supply.
The gas flow through the O2 gas module is regulated
The non-return valve is also designed to slowly by the Inspiratory solenoid via the Nozzle unit.
evacuate compressed O2 gas from the module, if the
The current supplied to the solenoid is regulated so
O2 gas supply to the O2 gas module is disconnected. that the O2 gas module will deliver a gas flow
according to the settings on the User interface.
3.1.3 Inspiratory valve temperature sensor
3.1.8 Gas module key
The temperature of the supplied O2 gas is measured
by the Inspiratory valve temperature sensor. This The O2 gas module is provided with a mechanical key
sensor is situated in the gas flow. to prevent that gas modules adapted for other gases
are connected to the system.
The output signal from this sensor is used to
compensate for the O2 gas density variations due to The key consists of a plastic guide mounted
temperature. underneath the module and a corresponding guide
mounted in the Patient unit.
3.1.4 Supply pressure transducer
3.1.9 ID PROM
The pressure of the supplied O2 gas is measured by
the Supply pressure transducer. The O2 gas module is provided with an ID-PROM. The
ID information can be read by the system.
The output signal from this transducer is amplified. It
is then used to calculate the absolute pressure of the
O2 gas to compensate for gas density variations due
to pressure.

3.1.5 Flow transducer

The O2 gas flows through a wire mesh (resistance)


which causes a pressure drop. The pressure is
measured on both sides of this net and the differential
pressure value is then amplified.

3.1.6 Nozzle unit

The plastic Nozzle unit contains a valve diaphragm.


The valve diaphragm, controlled by the Inspiratory
solenoid, regulates the gas flow through the O2 gas
module.

SERVO-air Ventilator System, Service Manual 3-5


| 3 | Description of functions |

3.2 Turbine module including PC 2027 Motor 7. Nipple for air inlet pressure measurement for
control (2) supervision of pressure drop over the Air inlet
filter.
3.2.1 Turbine
The Turbine module shall be replaced every 10 years.
The Turbine generates the inspiratory air gas flow
(max. 240 l/min). It generates the air flow and pressure Note: The Turbine is factory calibrated. The Turbine
from the surrounding air. This air flow is then mixed must not be disassembled.
together with the O2 flow from the O2 gas module.

The maximum Turbine output pressure is dependent 3.2.2 PC 2027 Motor control
on the ambient pressure and may therefore vary
PC 2027 Motor control’s main purpose is to control
between 65 cmH2O (high altitude) and
and supervise the Turbine:
approx. 100 cmH2O (sea level).
• Powering
bullet list the Turbine. The power to the motor driver
s g is supplied through a step-up regulator that elevates
the voltage from the system voltage (approximately
11-17 V) to approximately 28 V. The energy created
when the motor is decelerated is stored in a
a capacitor buffer to enable quick acceleration.
Further excess energy will be delivered to other
parts of the system.
h • Data transferred
bullet list to BRE via SPI communication for
X T
further dispatching to MON:
P1 PC 2027 Motor control P - Inspiratory flow and gas temperature from the
Inspiratory flowmeter.
- Ambient air temperature and humidity from
d f h j
S20103
sensors on PC 2030.
- Air inlet filter pressure drop measured by a
The Turbine module includes: differential pressure sensor. Will indicate if the
Air inlet filter is occluded or clogged. Will also
1. Brushless DC motor powered by a motor driver
detect if the Air inlet filter is missing during
on PC 2027 Motor control. Maximum speed
Pre-use check.
70 000 r/min.
2. Turbine wheel.
• Passing
bullet listdata from the ambient air inlet filter
differential pressure sensor.
3. Hall effect sensors to measure motor speed and
phase.
3.2.3 Air inlet filter including dust filter
4. Temperature sensor to monitor the Turbine
temperature. The reading is used to detect
AA
The airflow to the Turbine, and to the patient, is
overheating and to log running conditions. protected by an Air inlet filter. The Air inlet filter
5. Silencing foam and rubber suspensions to consists of a HEPA (High Efficiency Particulate Air)
minimize vibrations and sound level. filter and a dust filter.
DR
6. Cooling air flow for the motor. Routed from the
The Air inlet filter must be replaced during the
Turbine unit out through the Patient unit via a
Preventive maintenance. Check the dust filter and
hose.
replace it once a month or more often if needed.

3-6 SERVO-air Ventilator System, Service Manual


| Description of functions | 3 |

3.3 Inspiratory gas docking (3)


The Inspiratory gas docking connects the O2 gas
module, and Turbine module, to the Inspiratory
flowmeter connector muff. The Inspiratory gas docking Gas flow
also includes:
T T
• The Gas
bullet list module silicone muff connects the O2 gas
module to the Inspiratory gas docking. a s d s
• The Turbine
bullet list module one-way valve that prevents S20102

gas flow backwards through the Turbine module. 1. Flow conditioner


2. Temperature sensors
3.4 Inspiratory pipe (4) 3. Heater
The Inspiratory pipe leads the gas from the Inspiratory The flowmeter is equipped with a flow conditioner on
gas docking to the Inspiratory channel. the inlet to create a laminar flow to enhance the
accuracy.
The Inspiratory pipe contains:
3.6 Inspiratory channel (6)
• Inspiratory
bullet list flowmeter connector muff.
• Inspiratory
bullet list flowmeter. The Inspiratory channel connects the Inspiratory pipe
• Inspiratory
bullet list channel connector muff. to the patient breathing system. The Inspiratory
FFFTTT FFTT
The Inspiratory pipe is equipped with a flow conditioner
ahead of the flowmeter, to ensure that the flowmeter
makes reliable measurements.
channel contains:

• The Inspiratory
bullet list pressure tube that connects the
Inspiratory channel with the Inspiratory pressure
transducer. The tube is connected to the bacteria
3.5 Inspiratory flowmeter (5) filter that protects the pressure transducer PC 1781
Inspiratory pressure transducer from contamination.
The Inspiratory flowmeter (inspiratory flow transducer)
The bacteria filter must be replaced during
measures the combined air and O2 flow, as well as
Preventive maintenance.
temperature, of the inspiratory gas from the O2 gas
RRAAARRAA
module and Turbine module. The flow measurements • The Safety
bullet list valve membrane.
are used as input to control the gas module and the • Housing
bullet list for the O2 cell.
Turbine, creating a feedback loop. • The Inspiratory
bullet list outlet, a 22 mm/15 mm tube
connector for the Inspiratory tube of the patient
The flowmeter uses a thermal measuring principle, in
breathing system.
this case a differential temperature measurement. The
basic principle is that two temperature sensors, one The Inspiratory channel can easily be removed using
mounted before and one after a small heater in the the quick coupling.
flow direction:
D

• At no list
bullet flow through the flowmeter, the temperature
will be equal on both temperature sensors.
D

• With alist
bullet flow through the flowmeter, the temperature
on the two temperature sensors will differ,
consequently the higher flow the bigger temperature
difference.
A
SERVO-air Ventilator System, Service Manual 3-7
| 3 | Description of functions |

3.7 O2 cell (7) 3.8 Safety valve including pull magnet (8)

The O2 concentration is measured by an O2 cell. The The movable axis of the Safety valve pull magnet
O2 cell is mounted in a housing on the Inspiratory controls the opening and closing of the Safety valve
channel and is protected by a bacteria filter. membrane in the Inspiratory channel. The pull magnet
is controlled (electrically activated = closed) by
O2 PC 2004/PC 2024 Expiratory Channel.
O2 gas module

O2
Air
O2 cell O2 gas module
Turbine module

Air
Turbine module
S20015

Maintenance, including exchange of bacteria filter, is


performed according to the User´s Manual.

The O2 cell gives an output voltage proportional to the


partial pressure of oxygen inside the Inspiratory pipe.
At constant ambient pressure this output is
a s
proportional to the O2 concentration in percent.

The life time of the cell is affected by the O2


concentration. With a concentration (at the cell) in % S20016

and expected cell life time in hours the following


applies at 25 ºC (77 ºF): 1. During startup, the pull magnet is electrically
activated and the pull magnet axis is pushed up,
Expected cell life time = O2 Conc (%) x Time (h) creating a clicking sound. The Safety valve is now
= 500 000 (%hours) closed. This is the normal operating mode, when
the pull magnet is electrically activated and the
The O2 cell is automatically calibrated each time a
safety valve membrane is closed.
Pre-use check is performed (if O2 is connected to the
2. When the Safety valve is not electrically activated,
system).
the weight of the pull magnet axis combined with
If the system has been in use continually for a long the design of the valve membrane causes the pull
time, the measured O2 concentration may drop due magnet axis to be pulled down to the bottom
to normal degradation of the O2 cell. This will activate position. When the pull magnet axis is in the
bottom position, the Safety valve is open and the
a nuisance alarm. For further information, refer to the
inspiratory gas is released from the Inspiratory
User's Manual.
pipe via the Emergency air intake resulting in
An ID PROM is integrated into each O2 cell. Its ID decreased inspiratory pressure. The procedure
information and remaining lifetime can be read by the described is a normal safety (pop-off) function.
system. The Safety valve is opened when:

Notes: • The ventilator


bullet list is switched Off or to Standby.

• Perform
bullet lista Pre-use check to calibrate the O2 cell. • The pressure
bullet list inside the inspiratory pipe is 5 cmH2O
above the preset Upper pressure alarm limit. This
• The Olist
bullet 2 Sensor cannot be installed in SERVO-air.
condition is controlled by the Monitoring subsystem.

3-8 SERVO-air Ventilator System, Service Manual


| Description of functions | 3 |

• The pressure
bullet list inside the inspiratory pipe is 7 cmH2O The O2 evacuation fan air inlet (1) is located at the
above the preset Upper pressure alarm limit. This bottom of the side cover. The air outlets (2) are located
condition is controlled by the Breathing subsystem. above and below the Air inlet filter on the back of the
Patient unit. The air inlet is protected by a filter that
• The pressure
bullet list inside the inspiratory pipe is above
117 ±7 cmH2O. This is an extra safety function and must be cleaned or replaced during the Preventive
this situation does not normally occur. maintenance.

• The safety
bullet list valve also opens by other alarms, for The O2 evacuation fan is monitored by a temperature
example the Airway pressure continuously sensor on PC 2030, and controlled by PC 2028.
high-alarm.
The O2 evacuation fan is always running:
The safety valve opening pressure is calibrated to
117 ±3 cmH2O during each Pre-use check. • Whenlist
bullet the system is turned on. This is to make sure
that no O2 gas is present in the Inspiratory section
The Emergency air intake (Safety valve outlet) is while the power is on in case of O2 leakage.
covered by a plastic grid. The Emergency air intake
• Whenlist
bullet the batteries are charging, and for 2.5 hours

DRRTT TTT
must be checked for obstructions during the Preventive
after the charging is complete.
maintenance.
Depending on the temperature conditions, the fan can
3.9 O2 evacuation fan (9) also run when the system is turned off but connected
AAFF AAFAFF
The O2 evacuation fan has two purposes:
to mains power supply.

The fan always runs on half effect, and with full effect
• Forcelist
bullet cooling air through the Inspiratory section. at approx. 43 °C (109 °F). When the temperature drops
• If needed,
bullet list evacuate O2 gas from the Inspiratory below approx. 37 °C (99 °F), the fan turns to half
section, to prevent O2 gas from reaching too high effect.
concentrations.
3.10 PC 2030 Inspiratory channel (10)
The cooling air flow inside the Inspiratory section is
indicated in the System overview. Interconnecting board including connectors for:
DDRDRR

• Turbine
bullet listmodule
• O2 gaslistmodule
bullet
• O2 celllist
bullet

s
• O2 evacuation
bullet list fan

AA •
Inspiratory
bullet list flowmeter
Safetylist
bullet Valve

PC 2030 Inspiratory channel is also connected to


PC 2020 Main back-plane via the PC 2029 Inspiratory
a
channel extension for communication with components
S20017 in the center part of the Patient unit.

1. O2 evacuation fan inlet The ambient air temperature and humidity sensor is
2. O2 evacuation fan outlet located on PC 2030. The gas temperature is used as
input to volume calculations and for supervision of the
temperature delivered to the patient.

SERVO-air Ventilator System, Service Manual 3-9


| 3 | Description of functions |

3.11 PC 1781 Inspiratory pressure 1. Ultrasonic transducer


transducer (11) 2. PC 1786 Expiratory channel cassette

PC 1781 Inspiratory pressure transducer is mounted 3. Bacteria filter


onto PC 2030 Inspiratory channel. 4. Expiratory pressure tube connector
5. Heating foil
The pressure, conveyed via the pressure tube
connected to this block, is led to and measured by its
differential pressure transducer. With differential 3.13.1 Expiratory inlet
reference to the ambient pressure, the output signal
A 22 mm/15 mm tube connector for the patient
is proportional to the measured pressure thus giving
breathing system’s Expiratory tube. The inlet is
a linear measurement in the range -40 cmH2O to
designed to make condensed water drip out and allow
+160 cmH2O.
use of a water trap for such water to be collected.
Technical limitation: Pressures exceeding ±400 cmH2O Expiratory inlet bacteria filter can be connected to
protect the cassette from contamination.
must be avoided.

Includes an ID PROM. The ID information can be read 3.13.2 Heating foil


by the system.
An electrical Heating foil is applied on the outside of
the Expiratory pipe where the Ultrasonic flowmeter is
3.12 PC 2029 Inspiratory channel extension
situated. The purpose of the Heating foil is to reduce
(12)
condensation and maintain a stable temperature in
PC 2029 Inspiratory channel extension is a connector the expiratory gas.
board. It connects PC 2030 Inspiratory channel to
PC 2020 Main back-plane. There are no active 3.13.3 Ultrasonic flowmeter
components on PC 2029.
The Ultrasonic flowmeter (expiratory flow transducer)
is a measuring device for the expiratory gas flow, using
3.13 Expiratory cassette (13)
ultrasound technique with two ultrasonic
The expiratory gas conveying parts and PC 1786 transducers/receivers. The measuring process is
Expiratory Channel Cassette are integrated into one controlled from PC 2004/PC 2024 Expiratory channel.
part – the Expiratory Cassette – which can be easily
removed for cleaning or exchange, see the User’s
Manual.

The Expiratory cassette can be interchanged between


different systems. A Pre use check is always required
after exchanging the Expiratory cassette.

a s a

S20019

g d f The left hand side transducer is sending out an


S20018
ultrasonic sound that is reflected against the inner wall
of the Expiratory channel. The ultrasonic sound is
3 - 10 SERVO-air Ventilator System, Service Manual
| Description of functions | 3 |

received by the right hand side transducer now acting 3.15 Expiratory valve coil (15)
as a receiver. The time from sending to receiving
ultrasonic sound in downstream expiratory gas flow The movable axis of the Expiratory valve coil controls
is measured. the opening of the Expiratory valve by pushing the
valve membrane into desired position. The power
Then the right hand side transducer (earlier receiving) supply to the coil is regulated so that the remaining
is sending out ultrasonic sound upstream the pressure in the patient system, towards the end of the
expiratory gas flow. The ultrasonic sound is received expiration time, is kept on the PEEP level according
by the left hand side transducer now acting as a to User interface setting.
receiver. The time from sending to receiving ultrasonic
sound in upstream expiratory gas flow is measured. 3.16 Expiratory outlet (16)
The time difference between the downstream and the The gas from the patient system leaves the system
upstream time measurements provides flow via this Expiratory outlet. A thermo sleeve can be
information. attached to the outlet to reduce condensation.

A temperature sensor inside the cassette measures


the expiratory gas temperature. This temperature
3.17 PC 1786 Expiratory channel cassette
measurement is also used when calculating the
(17)
expiratory flow. The PC 1786 Expiratory channel cassette is a
connection board, integrated into the Expiratory
3.13.4 Bacteria filter and expiratory pressure tube cassette, for the Ultrasonic flowmeter and for the
Heating foil. It connects to PC 2031/PC 1785 mounted
Via a Bacteria filter inside the cassette, the Expiratory
in the Expiratory cassette compartment.
pressure tube connects the cassette to the Expiratory
pressure transducer. The filter and the connector are Includes an ID PROM. The ID information can be read
integrated parts of the cassette. The filter protects the by the system.
transducer on PC 1781 Pressure transducer from
contamination. 3.18 PC 2031/PC 1785 Expiratory channel
connector (18)
3.14 Expiratory valve (14)
Note: PC 1785 replaces PC 2031. Introduced in
The Expiratory valve consists of a membrane in the
production during Q1 2016.
cassette that is operated by the axis of the Expiratory
valve coil. The valve is fully open as long as no power
is supplied to the coil. The PC 2031/PC 1785 Expiratory channel connector
is a connector board, including signal filters (and a
Operating capacity for the membrane is estimated to power regulator on PC 2031). It is mounted in the
10 000 000 breathing cycles. When this limit is passed Expiratory cassette compartment, and connects to
or if the membrane for some reason has become PC 1786 mounted in the Expiratory cassette when the
defective, it must be replaced. Refer to instructions in cassette is docked to the Expiratory cassette
chapter Disassembling and assembling. compartment.

Membrane capacity (counter) can be shown in the For software and hardware requirements and
System status window. Select SYSTEM STATUS > compatibility, see the Compatibility chart in chapter
Expiratory cassette to check Membrane capacity. The Revision history.
operating capacity meter must be reset after
replacement of the membrane. Note that the reset
function is not implemented in System version 1.0.

SERVO-air Ventilator System, Service Manual 3 - 11


| 3 | Description of functions |

3.19 PC 2020 Main back-plane (19) 3.21 PC 2004/PC 2024 Expiratory channel
(21)
PC 2020 is the interconnection board for PC boards
and components.
Note: PC 2024 replaces PC 2004. Introduced in
PC 2020 Main back-plane is also connected to production during Q1 2016.
PC 2030 Inspiratory channel via the PC 2029
Inspiratory channel extension for communication with PC 2004/PC 2024 Expiratory channel is part of
components in the Inspiratory section. subsystem Breathing BRE. The main functions are
expiratory flow measurement and expiratory cassette
The System ID (Serial No.), operating time, etc., is handling. Expiratory flow measurement is performed
stored in an EEPROM on PC 2020. Thus, when by the flowmeter in the cassette. PC 2004/PC 2024
replacing PC 2020, a spare part that is factory is connected to the cassette via PC 2031/PC 1785
programmed for the concerned system must be used. Expiratory channel connector.

As the Preventive maintenance time stamp will be lost PC 2004/PC 2024 Expiratory channel contains
when replacing PC 2020, a new time stamp must be electronics including microprocessor for handling of:
set. In order to make this new time stamp correct,
there are two alternatives: • Expiratory
bullet list flow measurement performed by the
flowmeter inside the cassette. The output signal
• Perform
bullet listPreventive maintenance on the system Exp. Flow is used in sub-systems Control and
and report completed Preventive maintenance in Monitoring.
Service & Settings.
• All electronic
bullet list connections to and from the cassette.
or
• Measurement
bullet list of expiratory pressure via pressure
the PC 1781 Expiratory pressure transducer.
• If correct
bullet list value (hours) until next Preventive
PC 2004/PC 2024 contains holder and electrical
maintenance is known, enter this value in Service
connector for the pressure transducer board.
& Settings.
• Expiratory
bullet list valve control (with input from Control).

3.20 PC 1781 Expiratory pressure transducer • Safetylist


bullet valve control (with input from other
subsystems).
(20)
Includes an ID PROM. The ID information can be read
PC 1781 Expiratory pressure transducer is mounted by the system.
onto PC 2004/PC 2024 Expiratory channel.
For software and hardware requirements and
Function identical to PC 1781 Inspiratory Pressure
FTFT
compatibility, see the Compatibility chart in chapter
Transducer. Revision history.

Note: If PC 2004/PC 2024 is replaced, the System


software must be re-installed.
AAA
RAA
R
3 - 12 SERVO-air Ventilator System, Service Manual
| Description of functions | 3 |

3.22 PC 1991 Control (22) • Monitoring


bullet list of alarm limits. Activates pressure
reducing mechanisms, including activation of the
PC 1991 Control is part of subsystem Breathing BRE. safety valve, in case of excessive breathing system
The main responsibility is functions for the patient pressure.
treatment, i.e. how to regulate the inspiratory and
expiratory gas flow. • Supervision
bullet list of delivered pressure and flow.
• Calculation
bullet list of parameters. Providing the system
PC 1991 Control comprises electronics including with different measured data.
microprocessor for handling of:
• Monitors
bullet list battery module status.

• O2 gaslistmodule and Turbine module.


bullet • Barometric
bullet list pressure. PC 1992 contains a
barometric transducer and the measured barometric
• Airwaylist
bullet pressure control with input from values
pressure is supplied to the other sub-units in the
measured by the Inspiratory and Expiratory
system.
pressure transducer.
• Nebulizer
bullet list control.
• Real time
bullet list clock, RTC.

A Memory backup battery on PC 1991 power supplies


• Temperature
bullet list monitoring. A thermistor on PC 1992
monitors the temperature inside the Patient unit.
the internal memory on the PC board. If the battery
An alarm is activated if the temperature is 77 ±5 °C
on PC 1991 is disconnected or if the battery voltage
(170 ±9 °F) or higher.
is too low, startup ventilation configurations made via
the Service & Settings may be erased. The Memory • System
bullet listvoltage monitoring: The following voltages
backup batteries must be replaced after five years. are supervised:
• +24 Vlist
bullet
Includes an ID PROM. The ID information can be read
by the system.
• +12 Vlist
bullet
• -12 V list
bullet

AAAAAAAAAAAA
DRDRDRDDDRRRDRRTTT
For software and hardware requirements and
compatibility, see the Compatibility chart in chapter
Revision history.


+5 V list
bullet
+3.3 Vlist
bullet
The buzzer on PC 1992 generates the alarm signal
in case of +5 V or +3.3 V power failures. The buzzer
AAAFFF
Note: The System software must be re-installed if
and +5 V/+3.3 V failure logic is powered by backup
PC 1991 is replaced.
capacitors in case of power failure.

3.23 PC 1992 Monitoring (23) • Maintains


bullet list a persistent data storage area to save
logs. The logs are saved once per minute.
PC 1992 Monitoring is part of subsystem Monitoring • RS-232
bullet listcommunication. The system is equipped
MON. The Monitoring subsystem features alarm and with two RS-232 ports. The RS-232 communication
monitoring functions, calculations, trending of is handled by PC 1992. The communication protocol
measured values and is also CAN-bus master. SCI (SERVO Communication Interface) is described
DDDRRR

in the 'SERVO Communication Interface –


PC 1992 Monitoring comprises electronics including Reference Manual'.
microprocessor for handling of:
• Remote
bullet listalarm function. The Remote alarm
• Control
bullet listthat all alarms are displayed on the User connector is available on PC 2021 Rear connectors.
interface and that the alarm sound is generated. In The alarm signal used by this option is generated
case of malfunction in the loudspeakers in the User on PC 1992.
interface, a backup sound generating device A Memory backup battery on PC 1992 power supplies
(buzzer) on PC 1992 will be activated. The backup the internal memory on the PC board. If the battery
buzzer is monitored by a microphone and tested at on PC 1992 is disconnected or if the battery voltage
startup and during Pre-use check. is too low, startup alarm configurations made via the
SERVO-air Ventilator System, Service Manual 3 - 13
| 3 | Description of functions |

Service & Settings, all logs, Pre-use check results, 3.27 AC/DC Converter (27)
date and time setting and service reports will be
erased. The Memory backup batteries must be The AC/DC Converter, including PC 2033, is a
replaced after five years. complete unit adapted for the system. The AC/DC
Converter converts the connected AC Power (mains
Includes an ID PROM. The ID information can be read power) to the internal DC supply voltage
by the system. +12V_AC-DC.

For software and hardware requirements and Mains power is supplied to the AC/DC Converter via
compatibility, see the Compatibility chart in chapter the Mains inlet.
Revision history.
3.28 PC 2028 DC/DC & Battery control (28)
Note: The System SW must be re-installed if PC
1992 is replaced. This PC includes routing of Nebulizer signals to/from
PC 1991, and Ethernet/USB signals to the PC 2023
Panel board. The PC 2025 Battery back-plane is
3.24 PC 1993 Ethernet switch (24) connected to PC 2028, via the PC 2026 Battery
extension board. A secondary temperature sensor on
PC 1993 Ethernet switch controls the Ethernet
the PC 2028 regulates the Main fan and the O2
communication in the system. Ethernet is used by the
evacuation fan.
PAN, BRE and MON subsystems.
The main functions for the PC 2028 DC/DC & Battery
Includes an ID PROM. The ID information can be read
control are:
by the system.
• On/Offlist
bullet control.
3.25 PC 1994 Nebulizer (25)
• Control
bullet listcharging/discharging of the Battery
modules.
PC 1994 Nebulizer is the integrated interface for
connection of the Aerogen® Nebulizer Systems, • Switchlist
bullet to battery power supply when Mains
Aerogen Pro and Aerogen Solo. PC 1994 Nebulizer power/External 12 V battery is lost, and control
contains electronics required to control the nebulizer switching between Mains power and External 12 V
units. battery.

For electrical safety test of the Nebulizer connector,


• Control
bullet listof Main fan through one thermistor on PC
2020, and one on PC 2028.
refer to chapter Preventive maintenance section Safety
inspection. • Control
bullet listof O2 evacuation fan through a thermistor
on PC 2030.

3.26 Mains inlet (26) • Supplylistrequired internal voltages.


bullet
• Connects
bullet list the User interface control cable.
Inlet for mains power supply including grounding
The ID information is stored in flash memory in the
connection.
CPU and can be read by the system.
The system will automatically adjust to the connected
mains power if the mains power is within specified 3.28.1 On/Off control
range. No voltage or frequency setting is required.
The power modes in the system are:
The mains inlet is equipped with two mains power
fuses, F11 and F12, rated 4 A.
AA
3 - 14 SERVO-air Ventilator System, Service Manual
| Description of functions | 3 |

• At Power
bullet list up, i. e. when the On/Off switch is turned • +12V_AC-DC
bullet list from the AC/DC converter.
On, all internal voltages will be enabled. When the
• +12V_UNREG_EXT_DC
bullet list from the External 12 V
On/Off switch is toggled from Off to On, the O2 battery.
evacuation fan is activated. The fan will run so that
any O2 build up is evacuated, prior to enabling
• +12V_UNREG
bullet list from the Battery modules.

voltages in the inspiratory compartment. PC 2028 converts the power supply voltages into the
• At Power
bullet list down, the Power supply system will following internal DC supply voltages:
deactivate the hardware signal Power_Good.H, in
order to let the different subsystems save their • +24 Vlist
bullet
current settings in non-volatile memory. Power • +12 Vlist
bullet UNREG_INTERNAL
down can be caused by: • +12 Vlist
bullet UNREG_VALVES
- Turning the On/Off switch Off. • +12 Vlist
bullet
- Mains failure resulting in a switch to battery, • -12 V list
bullet
where the backup battery voltage is too low for
proper operation of the system or if there is no
• +5 V list
bullet

DDDDRDRDRRRRDRDDRRR T T
backup battery connected. • +3.3 Vlist
bullet

- Automatic power down when the battery voltage


becomes too low for operation of the system. 3.29 On/Off switch (29)
AFAF
In the Off mode, only charging of Battery modules is
enabled (if the system is connected to mains). To allow
Switch to Power up or Power down the system. Refer
to section Power supply.
adequate cooling during battery charging, the Main
fan and O2 evacuation fan can operate also in Off
3.30 PC 2021 Rear connectors (30)
mode. All other circuitry is un-powered.

AAAAAAAAAA
The following connectors are located on PC 2021 Rear
Note: At least one Battery module must always be connectors:
installed to ensure proper operation.
• N26 –list
bullet External 12 V battery. The connector is
equipped with a fuse F1, rated 10 A. There are no
DRDR

Note: A Power system restart might be necessary alarms indicating power supply failure related to the
to reset power related alarms. To restart, switch Off External 12 V battery.
the system and disconnect both the mains power
supply and the External 12 V battery.
• N27 –list
bullet USB (x2). For USB service key, data copy,
software updating and mouse.

3.28.2 Power supply selection Note: Using a mouse with the system is only
permitted for service purposes.
If the internal DC supply voltage drops, due to power
supply failure, the power supply source will
automatically switch. The following power supply • N28 –list
bullet Ethernet. For Remote Services.

source priority is used: • N29 –list


bullet RS-232 (1). Isolated RS-232 connector
(identical to N30). For SCI communication or
1. Mains power MSync.
2. External 12 V battery (if connected) • N30 –list
bullet RS-232 (2). Isolated RS-232 connector
3. Battery modules (identical to N29). For SCI communication or
MSync.

3.28.3 Internal voltages

There are three power supply voltages to the system:


SERVO-air Ventilator System, Service Manual 3 - 15
| 3 | Description of functions |

• N67 –list
bullet Alarm output. Isolated alarm output • Expiration
bullet list date.
connector. The Alarm output software option is
• Charge
bullet listcycles.
required to enable the Alarm output function. For a
technical description of the function Alarm output,
• If the operational
bullet list capacity is too poor for continued
usage.
refer to Alarm output in section Optional equipment.
In all cases, battery replacement information will be
Pin configuration and signal names can be found in
shown on the User interface.
chapter Diagrams.
Even with this technology, the only time at which the
PC 2021 also hosts the On/Off switch and LEDs for
battery charge status is absolutely reliable is when it
power-on and mains power connection indication:
is either completely full or completely empty. What’s
more, if the battery only sees partial charges and
• A bluelist
bullet mains LED that is lit when mains power is
discharges during its application, then it may not get
connected.
the benefit of a 'full' or 'empty' reference point for some
• A green
bullet listpower on LED that is lit when the On/Off time, and must rely more and more on its calculated
switch is On.

DDDRRRTT T
figure.

3.31 PC 2026 Battery extension (31) Select SYSTEM STATUS > Batteries on the User
interface to check battery status. For further
PC 2026 Battery extension is a connector board. It information, refer to chapter Service procedures,
AAFF AF
connects PC 2025 Battery back-plane to PC 2028
DC/DC & Battery control. There are no active
section Battery modules.

components on PC 2026. The battery life time is 300 charge cycles or five years
(60 months) from manufacturing date, whichever

AAA
3.32 PC 2025 Battery back-plane (32) comes first. Remaining charge cycles and time
(months) are shown in the STATUS > Battery-window.
PC 2025 Battery back-plane has connectors for the
battery modules. PC 2025 is connected to PC 2028 Note: For system version 4.1 or lower the battery
DC/DC & Battery control via the PC 2026 Battery life time is four years (48 months) from
DDRR DR

extension board. manufacturing date.

3.33 Battery module (33) The remaining charge cycles are ‘full charge cycles’
i.e. from an empty battery (0 %) to a full battery
The Battery module is a 14.4 V/6.6 Ah Li-ion
(100 %). If the battery is charged in shorter cycles,
rechargeable 'smart battery'. Two Battery modules
e.g. from 75 % to 100 %, the cycles are added until
can be connected to the Battery compartment.
they reach 100 % and then calculated as one full
charge cycle.
Note: At least one Battery module must always be
installed for the system to work. Example: Four charge cycles from 75 % to 100 % is
counted as one full charge cycle.
To calculate its own status, the battery uses an internal
Typical recharge time approximately 2 h/battery
highly accurate voltmeter, amperemeter and time clock
(90 %), up to 3 h (100 %) if battery is completely
to measure actual charge in and out of the battery. In
discharged. If a battery is fully discharged, e.g. due to
addition, there are algorithms to compensate for the
long storage time, each battery may require up to
effects of discharge rate, discharge temperature,
6 hours charging time.
selfdischarge and charging efficiency, etc.
Each battery includes an ID PROM. The ID information
The system will, among others, monitor:
can be read by the system.

3 - 16 SERVO-air Ventilator System, Service Manual


| Description of functions | 3 |

The air inlet is protected by a filter that must be


CAUTION: Do not expose the batteries
cleaned or replaced during the Preventive
to water, fire or excessive heat. Do not
maintenance.
crush, disassemble, puncture or short
circuit the connector terminals.
3.35 User interface cable (35)
This User interface cable connects the Patient unit
3.34 Main fan (34) and the User interface. The cable is hidden in the
panel arm behind a cover.
The Main fan forces cooling air through the AC/DC
section, and the Main section of the Patient unit. The
Note: The User interface cable must only be
cooling air flow inside the Patient unit, is indicated in
connected or disconnected following the instructions
the System overview.
given in section Hazard notices in chapter
Disassembling and assembling.

3.36 Touch screen (36)

s The projected resistive Touch screen implements the


touch function of the User interface screen and is
interactive with information displayed on the LCD
Display. The Touch screen can be recalibrated in
Service & Settings.

a
3.37 LCD Display (37)
S20055

The LCD Display is a 12″ 4:3 TFT display with LED


1. Main fan inlet backlight for display of picture- and alphanumeric data.
2. Main fan outlet
It is estimated that the LCD Display will maintain an
The Main fan inlet (1) is located at the bottom of the acceptable brightness level during the unit’s lifetime.
side cover. The Main fan outlet (2) is located above Using Service & Settings, a backlight time meter can
the connectors on the backside of the Patient unit. be shown.

The Main fan is monitored by the temperature sensors 3.38 PC 2023 Panel (38)
on PC 2020, and PC 2028, and controlled by PC 2028.
The value from the temperature sensor, giving the The PC 2023 Panel contains drivers for:
highest temperature reading, will be used.
• Touchlist
bullet and display
The Fan is always on when the batteries are charging, • Loudspeaker
bullet list
and for 2.5 hours after the charging is complete.
• Microphone
bullet list
The Fan will start with half effect at approximately • Microcontroller
bullet list
33 °C (91 °F) and with full effect at approximately The multicolor LEDs, visible in the Alarm indicator,
43 °C (109 °F). When the temperature drops below are located on PC 2023.
approximately 37 °C (99 °F), the Fan turns to half
effect and when the temperature drops below
approximately 27 °C (81 °F), the Fan stops.

SERVO-air Ventilator System, Service Manual 3 - 17


| 3 | Description of functions |

3.39 Alarm indicator (39) system and the ventilator. Furthermore, if both the
NO- and the NC-contacts are utilized, the highest
The Alarm indicator is located on the upper part of the safety level for unforeseen failures can be achieved.
User interface frame. The multicolor LEDs on PC 2023 Note that, in both cases, even an intentional
are visible in the Alarm indicator. The LEDs will flash disconnection of the ventilator will trigger the external
when an alarm is triggered in the system: alarm.

• Blue LED
bullet list – Low priority alarm The Alarm output software option is required to enable
• Yellowlist
bullet LED – Medium priority alarm the Alarm output function. The Alarm output software
• Red LED
bullet list – High priority alarm is individually created for each specific system and
can only be installed on this system. The Alarm output
function must be enabled in the configuration software.
3.40 Loudspeaker (40)
Loudspeakers for, e.g., alarm. Connected to the 3.41.2 Battery module
PC 2023 Panel.
The Module compartment allows up to two Battery
There is one loudspeaker in the User interface. The modules. To guarantee safe battery backup, always
User interface is mechanically designed to achieve a use at least one battery.
good sound output.
For further information, refer to section Battery module.
At startup and during Pre-use check the functionality
of the loudspeakers are monitored by a microphone. 3.41.3 Battery charger

monitored.

TTT
During operation the loudspeaker is continuously

3.41 Optional equipment


The Battery charger is an external smart battery
charger used to charge the SERVO-air Battery
modules. The charger has one slot and can charge
one battery at a time. For further information, see the
AAAFFF
3.41.1 Alarm output Battery charger manual.

The Alarm output connector, N67 on PC 2021 Rear


3.41.4 Mobile cart
connectors, enables connection of an external alarm
signal system to the system. The alarm output signal The Mobile cart is designed for carrying the Patient
is active as long as the alarm audio is active on the unit, the User interface and all required optional
system. equipment.

For pin configuration and signal names in the Alarm The Mobile cart has standard 10x25 mm rails to carry
output connector N67, see Connectors in chapter accessories such as Support arm 179 and Humidifier
DDDRRR

Diagrams. For further information, refer to the 'Alarm holder. The column has vertical slots for installation
output connector – Reference Manual'. of e.g. Gas cylinder restrainers. The optional Drawer
kit or Cabinet can be mounted on the Mobile cart.
The Alarm output connector has two contact functions:
NO (Normally Open) and NC (Normally Closed). In
3.41.5 Gas cylinder restrainer kit
an alarm situation the open contact will close and the
closed one will open. The contacts are independent The Gas cylinder restrainers are mounted on each
of polarity and can be used both with AC and DC side of the Mobile cart. It is intended for two gas
systems. cylinders, maximum volume 4.5 litres each.

Preferably the NC contact should be used, as this will


detect any interruption in the alarm system or
unintentional disconnection of the cable between the
3 - 18 SERVO-air Ventilator System, Service Manual
| Description of functions | 3 |

3.41.6 Shelf base 3.41.11 MSync

The Shelf base is a mounting device for the Patient MSync is intended for information transfer from the
unit. It allows positioning of the Patient unit on a shelf. Getinge devices to the hospitals information system
by:
3.41.7 Support arm 179
• Importing
bullet list real-time clinical data from the device
Support arm 179 is equipped with a clamp intended using its respective communication protocol.
for standard 10x25 mm rails. The support arm has a • Converting
bullet list device data to HL7 compliant data.
central locking mechanism that secures the selected
position. A twin tube holder (15-27 mm diameter) and
• Exporting
bullet list device data in HL7 format to the
information system.
a filter/tube support (20/25 mm diameter) can be
An MSync unit is connected to each Getinge device
mounted on the support arm. Maximum load with fully
intended to be integrated with the information system.
extended arm is 1 kg.

3.41.8 Aerogen nebulizers

The Aerogen® Nebulizer Systems, Aerogen Pro and


Aerogen Solo, can be connected to the Nebulizer
connector. PC 1994 Nebulizer contains electronics
required to control the nebulizer units.

Aerogen Pro and Aerogen Solo are devices intended


to aerosolize physician-prescribed medications, for
inhalation, that are approved for use with a general
purpose nebulizer.

• Aerogen
bullet list Pro: Reusable nebulizer unit.
• Aerogen
bullet list Solo: Single use nebulizer unit.
The Aerogen Pro/Aerogen Solo are designed to
operate in-line with the ventilator circuit in acute and
sub acute care environments. It operates without
changing patient ventilator parameters.
RRRAAA
3.41.9 Humidifier and Humidifier holder

Recommended active (heated) humidifiers are Fisher


& Paykel MR 810 and MR 850. For further information,
refer to manufacturer's documentation.

The Humidifier holder is equipped with a clamp


FFDDTTD
intended for standard 10x25 mm rails. Maximum load
is 5 kg.

3.41.10 Waterbag/IV pole

The Waterbag/IV pole is mounted on the optional


Humidifier holder. It includes two hooks. Maximum
load is 1 kg.

SERVO-air Ventilator System, Service Manual 3 - 19


| 3 | Description of functions |

3 - 20 SERVO-air Ventilator System, Service Manual


| Disassembling and assembling | 4 |

4 Disassembling and assembling

Table of Contents
4.1 General | 4-2
4.2 Hazard notices | 4-2
4.3 PC boards and electrical components | 4-2
4.4 Information stored in the system | 4-3
4.5 Assembling guidelines | 4-5
4.6 Front cover | 4-6
4.7 Top cover | 4-6
4.8 Chassis | 4-7
4.9 Main section | 4-8
4.10 Inspiratory section | 4-9
4.11 AC/DC section | 4 - 16
4.12 Front section | 4 - 21
4.13 Rear section | 4 - 22
4.14 PC 2020 Main back-plane | 4 - 24
4.15 Expiratory cassette | 4 - 25
4.16 User interface | 4 - 28

SERVO-air Ventilator System, Service Manual 4-1


| 4 | Disassembling and assembling |

4.1 General CAUTION: Do not expose the batteries


Disassembling of the system is described in this to water, fire or excessive heat. Do not
chapter. If not stated otherwise, the assembling crush, disassemble, puncture or short
procedure is the reverse of the described circuit the connector terminals.
disassembling procedure.

The illustrations in the Spare Parts Lists are useful 4.3 PC boards and electrical components
when disassembling and assembling the system.
4.3.1 Replacing electrical components
After any installation, maintenance or service
intervention in the system, perform a 'Functional check’ As stated in section Hazard notices, mains power and
according to instructions in chapter Important. battery power supply must be disconnected and On/Off
switch set to Off when replacing electrical components.
4.2 Hazard notices If energized electrical components are disconnected
or connected, this may interfere with the operation of
Before disassembling or assembling the system, make the unit.
sure that:
After replacement of PC boards or other electrical
• The system
bullet list is switched Off. components, it is recommended to perform an
• Mainslist
bullet power cable is disconnected. electrical safety test. For electrical safety tests, refer
to chapter Preventive maintenance, section Safety
• External
bullet list 12 V battery is disconnected.
inspection.
• Backup
bullet listbattery modules are disconnected. Note
that the backup battery modules will supply power
4.3.2 Handling PC boards
to parts of the system also when the system is
switched Off. Those who come into contact with circuit boards
• Gas supply
bullet list is disconnected, central gas supply and containing sensitive components must take certain
backup gas supply. precautions to avoid damaging the components (ESD
• The system,
bullet list including all gas conveying parts, is protection).
cleaned. Refer to instructions in the User’s Manual.
PC boards (spare parts) must always be kept in
WARNING! With gas supply connected protective packaging for sensitive electronic device.
to the system, there are pressurized
components inside the unit. All PC boards must not be inserted or removed while the
FFTFTFFFTTTT
personnel must exercise extreme mains power or battery power is applied to the PC
caution if fault tracing or adjustments boards.
are performed with gas supply
Remove and insert the PC boards very carefully to
connected and with covers removed.
avoid damage to the connectors.

WARNING! With power supply Note: The PC boards contain


connected to the system, there are components that are highly sensitive
energized electrical components inside to static electricity. When working with
RRAAAA

the unit. All personnel must exercise ESD sensitive components, always use
extreme caution if fault tracing or a grounded wrist band and a grounded
adjustments are performed with power work surface. Adequate service tools
supply connected and with covers must always be used.
removed.

4-2 SERVO-air Ventilator System, Service Manual


| Disassembling and assembling | 4 |

4.3.3 Replacing PC boards Other technical errors may occur after a software
installation. These errors will disappear after restarting
The System SW is distributed to the following PC the system.
boards and modules:

• PC 1991
bullet list Control
4.4 Information stored in the system
• PC 1992
bullet list Monitoring All electronic components equipped with an ID PROM
• PC 2004/PC
bullet list 2024 Expiratory channel store the information that is specific for the component,
for example article number, version number and serial
• PC 2023
bullet list Panel
number.
• PC 2027
bullet list Motor control
• PC 2028
bullet list DC/DC & Battery control Furthermore, some of the electronic components store
information related to the complete system, for
Additional actions when replacing PC boards example system software, system info, configurations,
logs and so forth. When replacing these components,
When delivered as spare parts, these components the stored information may be lost. See table below.
are equipped with a System SW version that may differ
from the version on the system to be repaired. To keep
the System SW version used prior to the replacement,
the applicable System SW version must be available

TTTTTT
for reinstallation purposes.

For functionality enhancement, the latest released


System SW version is always recommended.
AAAAAFAFFFFF
Before installing a new System SW version, ensure
that the software is fully compatible with all hardware,
software and mechanical components in the system.
Refer to the Compatibility chart in chapter Revision
history.

When replacing PC 1991 Control or PC 1992


Monitoring, turn on the unit and leave the unit on for
two minutes before restart. A certain time is required
for the options and settings to be transferred to the
DDDDDRDRRRRR

new PC board.

When replacing PC 1991 Control or PC 1992


Monitoring, the set time and date may need to be
adjusted.

When replacing PC 1992 Monitoring, Technical error


25 may occur. Adjust the Barometric pressure in
Service & Settings to solve this error (Service key
required).

When replacing PC 2004/PC 2024 Expiratory Channel,


two Pre-use checks may be required. The first check
will calibrate the Safety valve, but fail in other tests.
The second check will pass.
SERVO-air Ventilator System, Service Manual 4-3
| 4 | Disassembling and assembling |

PC 1781 Pressure transducer

PC 1786 Expiratory cassette


PC 2028 DC/DC & Battery
PC 2020 Main back-plane

PC 2027 Motor control


PC 1992 Monitoring

Expiratory channel
PC 2004/PC 2024
PC 1991 Control
Stored information

PC 2023 Panel
control

O2 cell
System software X X X X X X
System info: X
§ Serial number
§ Total operating and
ventilation time
§ Next preventive
maintenance
System info: X X
§ Installed software
options
System info: X
§ System version
Configurations X
§ Startup ventilation
configuration
Configurations X
§ Startup alarm
configuration
Configurations X
§ Units, language, date
format, etc.
Date and time setting X
Logs X X2 X
Trends X
Recordings X
Pre-use check results X X X X X
including transducer result calibration calibration calibration calibration
calibrations
Barometer calibration X
Touch screen calibration X
Performed preventive X1
maintenance
Replaced expiratory X1 X
membrane counter
Replaced backlight X1 X3
counter
Backlight operating time X3
Language X
Expiratory cassette: X
§ Date of first use
§ Total operating time
§ Remaining membrane
capacity
O2 cell: X
§ Date of first use
§ Estimated remaining
capacity (O2 cell)
S20104

1
Included in Service report as part of the Service log.
2
Copied to PC 1992 at Power On and continuously when the system is powered.

4-4 SERVO-air Ventilator System, Service Manual


| Disassembling and assembling | 4 |

3 Excluded in later System versions.

4.5 Assembling guidelines 4.5.3 Special grease

All parts of the system assembled with screws and Lubricants should normally not be used when servicing
nuts are tightened with a specified torque. Thread the system. If lubricants must be used, use only very
locking adhesives are used if required. small amounts of grease with P/N 95 73 700. This
special grease is O2 compatible.
In order to maintain these specifications over time, it
must be ensured that after any service intervention
WARNING! Due to the increased O2
removed parts are re-assembled and secured
concentration in valves and gas
according to instructions. Make sure to follow the
channels, only special grease intended
guidelines stated below.
for use in high O2 concentrations, are
allowed.
4.5.1 Tightening torque

• Thread
bullet listsize M3: 0.95 Nm ±15%
• Thread
bullet listsize M4: 3.1 Nm ±15%
• Thread
bullet listsize M4 (in plate thread joints):
1.4 Nm ±15%. These screw joints are indicated with
a ’1.4 Nm symbol’ in the concerned illustrations.
• Thread
bullet listsize M5: 5.0 Nm ±15%
• Thread
bullet listsize M6: 7.0 Nm ±15%
• Gas inlet
bullet list nipple: 20 Nm ±15%

Note: Do not fasten the screws too tightly in the


screw joints with plate threads. When tightening the
screws, it is important to turn each alternatively and
gradually. Do not apply specified torque until all the
screws have been positioned correctly into the plate
threads.
TTTTTT
4.5.2 Thread locking adhesives

• Loctitelist
bullet 243® on the expiratory valve coil screws.
AAAFAFAFAFFF
RRR

SERVO-air Ventilator System, Service Manual 4-5


| 4 | Disassembling and assembling |

4.6 Front cover

• Tilt back
bullet list the User interface.
• Release
bullet list the Front cover hatches, and then release
and lift off the Front cover.

S20020

4.7 Top cover


• FFTT FFTT
Remove
bullet list the screw (1).
Note that the screw remains in the top cover.
Release
bullet list and lift off the Top cover.
RAA RRAA
a

1.4 Nm
S20021
D
D

4-6 SERVO-air Ventilator System, Service Manual


| Disassembling and assembling | 4 |

4.8 Chassis
4.8.1 Removing the chassis

• Remove
bullet list the Front cover.
• Remove
bullet list the Top cover.
• Remove
bullet list the O2 evacuation fan filter (1) by pressing
the catch in and down.
• Remove
bullet list the Main fan filter (2) by pressing the catch
a
in and down.
• Remove
bullet list the Expiratory cassette.

S20022


FTFFTFFTFTTT
Remove
bullet list the five screws (positions 1-4) that secure
the Chassis.
Note that there are two screws at position (1), one
on each side of the Patient unit.

s
RARRRARRAAAA
a
a

1.4 Nm f d
S20023

• Pull the
bullet listChassis carefully to the sides (1), push s
backwards (2) and lift it off.
DDD
D

a a
D

S20024

SERVO-air Ventilator System, Service Manual 4-7


| 4 | Disassembling and assembling |

4.8.2 Replacing the handle

• Remove
bullet list the Chassis.
• Remove
bullet list the screws (1).
• Pull apart
bullet list the chassis to release the handle (2).
s

0.6 Nm

S20134

TT TTT
4.9 Main section
4.9.1 Main section lid

• Remove
bullet list the Top cover. a
AAFF AAAFFF
• Remove
bullet list the three screws (1-2) that secure the Main
section lid to the Patient unit.
DDDRRR

1.4 Nm

S20025

4-8 SERVO-air Ventilator System, Service Manual


| Disassembling and assembling | 4 |

4.9.2 PC board rack

• Remove
bullet list the Main section lid. a s d f
• Carefully
bullet list pull out the PC board:
- PC 1993 Ethernet switch (1)
- PC 1992 Monitoring (2)
- PC 1991 Control (3) g
- PC 2004/PC 2024 Expiratory channel (4)
including PC 1781 Expiratory pressure
transducer (5)
Note: Before removing PC 2004/PC 2024:
- Disconnect Expiratory valve coil cable
connector (6). h j k
S20032
- Disconnect Pressure transducer tube (7).
- Disconnect PC 2031/PC 1785 cable
connector (8).

TT
For software and hardware requirements and
compatibility, see the Compatibility chart in chapter
Revision history.
AAFF
Note: Additional actions may be required after
replacement of PC boards. Refer to section
Replacing PC boards in this chapter.

4.10 Inspiratory section


4.10.1 Inspiratory section lid
DDRR

• Remove
bullet list the Chassis. 1
• Remove
bullet list the eight screws (1-3) that secure the
Inspiratory section lid to the Patient unit. On systems
manufactured from Q3 2017 it is only required to
loosen (not remove) the three screws (2). The holes
in the lid are changed to slots as shown in the 4
illustration. 6
• Loosen
bullet listthe two screws (4).

Note: Note that the O2 evacuation fan mounted on


the Inspiratory section lid is connected to the Patient 2
unit by the fan cable. Carefully fold out the lid to 3
disconnect the cable.
1.4 Nm
TT

S20026

SERVO-air Ventilator System, Service Manual 4-9


| 4 | Disassembling and assembling |

• Disconnect
bullet list the O2 evacuation fan cable (5).

When assembling:
•Do notlist
bullet forget to connect the fan cable.
• Make list
bullet sure that the Inspiratory gas docking kit
connects properly to the Inspiratory section lid at
(6). g S20056

4.10.2 O2 evacuation fan

• Remove
bullet list the Inspiratory section lid.
• Disconnect
bullet list the O2 evacuation fan cable
connector (1) from the PC 2030 Inspiratory channel.

T TT TTT TTT
• Release
bullet list the four rubber straps (2), and lift off the
fan.

Note: When mounting the fan, make sure to point


the fan in the right direction. The cooling air flow
AFFAAAFFFAAAFFF
must be directed into the Patient unit. a s
S20039

4.10.3 Inspiratory flowmeter

• Remove
bullet list the Inspiratory section lid.
• Disconnect
bullet list the cable (1) from PC 2030.

Note: Be careful with the connection (2) on the


flowmeter. Make sure not to break it when removing
DDRRRDDDRRR

the flowmeter from the Inspiratory channel muff.

• Remove
bullet list the flowmeter from the Inspiratory channel
muff.

S20057

4 - 10 SERVO-air Ventilator System, Service Manual


| Disassembling and assembling | 4 |

4.10.4 Safety valve

• Remove
bullet list the Inspiratory section lid.
• Disconnect
bullet list the O2 cell cable (1).
• Loosen
bullet listand fold out the locking latch (2).
a

S20058

• Remove
bullet list the O2 cell (3).

TT T TTT S20059
AAFF AFAAAFFF
• Disconnect
bullet list the Inspiratory pressure transducer
tube (4).
• Remove
bullet list the Inspiratory channel (5).
DRDDDRRR

f S20109

SERVO-air Ventilator System, Service Manual 4 - 11


| 4 | Disassembling and assembling |

• Disconnect
bullet list the Safety valve cable (6) from PC 2030.
• Remove
bullet list the two screws (7) that secure the Safety
valve.
• Remove
bullet list the Safety valve and cable (8).

TTT
h
AFFF AAAFFF S20110

4.10.5 Gas module lid

• Remove
bullet list the Top cover. a
• Remove
bullet list the screw on the top of the Patient unit (1).
• Remove
bullet list the two screws on the side of the
Inspiratory section (2). On systems manufactured
from Q3 2017 it is only required to loosen (not
DDDRRR

d
remove) the two screws (2). The holes in the side
wall are changed to slots as shown in the illustration. s
• Pull out
bullet listthe Air inlet filter (4) from the socket.
• Remove
bullet list the Gas module lid (3). f
g
Note: When attaching the Gas module lid to the
Patient unit, make sure the Battery extraction tag (5)
is not stuck under the lid. This will make it hard to
1.4 Nm
pull out the Battery module.
TT TTT
S20029

4 - 12 SERVO-air Ventilator System, Service Manual


| Disassembling and assembling | 4 |

4.10.6 O2 gas module

• Remove
bullet list the Gas module lid. a 1.4 Nm
• Remove
bullet list the screw (1).
• Pull out
bullet listthe O2 gas module (2). s

S20111

4.10.7 Turbine module

TTT TT TTT
Remove
bullet list the Gas module lid.
• Pull out
bullet listthe Turbine module (1).

a
Note: Additional actions may be required after
F
replacement of PC boards (PC 2027 is included in
F
the Turbine module). Refer to section Replacing PC
F
boards in this chapter.
RRAA RRRAAA
S20033

4.10.8 Inspiratory gas docking

• Remove
bullet list the Inspiratory section lid.
• Remove
bullet list the Gas module lid. s
• Remove
bullet list the O2 module.
• Remove list the Turbine module.
D

bullet
F
• Disconnect
list the Inspiratory flowmeter connector
D

bullet
F
muff (1).
D

• Lift offlist
bullet the Inspiratory gas docking (2).
a

Note: Be careful with the connection on the


flowmeter, make sure not to break it when S20034

disconnecting the Inspiratory flowmeter muff.

SERVO-air Ventilator System, Service Manual 4 - 13


| 4 | Disassembling and assembling |

4.10.9 Turbine module one-way valve

• Remove
bullet list the Inspiratory gas docking.
g f d s a
• Remove
bullet list the four screws (1), and lift off the cap (2)
with O-ring (3).
• Pull out
bullet listthe one-way valve (4) from the Inspiratory
gas docking housing (5).

0.5 Nm

S20035

To insert the Turbine module one-way valve:

FT FFFTTT
Pull the
bullet liststrap through the hole in the Inspiratory gas
docking until it snaps into place.

Note: Make sure all parts are clean before inserting


the Turbine module one-way valve.

S20036
RA RRRAAA
DD
D

4 - 14 SERVO-air Ventilator System, Service Manual


| Disassembling and assembling | 4 |

4.10.10 PC 2030 Inspiratory channel

• Remove
bullet list the Inspiratory gas docking.
• Disconnect
bullet list the O2 cell cable (1) from PC 2030.
• Disconnect
bullet list the Inspiratory transducer tube (2) from
PC 1781.
• Remove
bullet list the four screws (3) that secure PC 2030.
Note that one screw must be removed through the
Safety valve channel.

FT FFFTFTTTA d
s
a

d
A
A
S20112

• Lift offlist
PC 2030 (4).
A
bullet h
DDDDRRRR

Note that PC 2029 (5) Inspiratory channel extension


may come loose when lifting off PC 2030.

4.10.11 PC 1781 Inspiratory pressure transducer


f
• Open list
bullet or cut the cable tie. g
• Lift offlist
bullet PC 1781 (6).
S20113

SERVO-air Ventilator System, Service Manual 4 - 15


| 4 | Disassembling and assembling |

4.11 AC/DC section


4.11.1 AC/DC section lid

• Remove
bullet list the Chassis.

Open the AC/DC section lid:


• Remove
bullet list the four screws (1) that secure the AC/DC
section lid to the Patient unit.

a 1.4 Nm
S20027

TT
Note: Note that the AC/DC section lid is still
attached to the hinge at the bottom.
AAFF
S20120
DDRR

4 - 16 SERVO-air Ventilator System, Service Manual


| Disassembling and assembling | 4 |

Remove the AC/DC section lid:

• Disconnect
bullet list the Main fan cable (1).

a
S20114

• Open list
bullet the AC/DC section lid to maximum, slide it
towards the back and then lift it off.

FFFTTT FT FT
4.11.2 Main fan
A
S20121

• Open list
bullet the AC/DC section lid. a s
• Disconnect
bullet list the Main fan cable connector (1) from
the Main back-plane.
• Release
bullet list the four rubber straps (2), and lift off the
DRDR

fan.

Note: When mounting the fan, make sure to point


A
the fan in the right direction. The cooling air flow
must be directed into the Patient unit. S20038

SERVO-air Ventilator System, Service Manual 4 - 17


| 4 | Disassembling and assembling |

4.11.3 AC/DC converter

• Open list
bullet the AC/DC section lid.
• Disconnect
bullet list the internal power cable connector (1).
• Disconnect
bullet list the signal cable connector (2).
• Disconnect
bullet list the mains power cable connector (3).
• Remove
bullet list the screws (4) and lift off the AC/DC
converter.
f a s d f
1.4 Nm

S20037

4.11.4 Mains inlet

• Remove
bullet list the AC/DC converter.
• Remove
bullet list the two screws (1) that secure the mains


• FFTFFTTTFFFFTFTTTT
power inlet.
Remove
bullet list the two nuts (2) that secure the grounding
cables to the grounding screw (3).
Remove
bullet list the mains power inlet.

a
RRARAAARRRRRAAAAA
s d

S20122
D
DD
DD

4 - 18 SERVO-air Ventilator System, Service Manual


| Disassembling and assembling | 4 |

4.11.5 Expiratory cassette holder

• Remove
bullet list the Chassis.
• Remove
bullet list the Main section lid.
• Remove
bullet list the four screws (1) that secure the User
interface cable lid (2) to the Patient unit.
• Lift offlist
bullet the User interface cable lid (2).
• Disconnect
bullet list the User interface cable (3) from the
connector. h
• While list
bullet holding the User interface, remove the four
hinge screws (4) that secure the User interface on
g
the Chassis. f
• Remove
bullet list the User interface (6).
• Remove
bullet list the screw (5).


FTTFFTTT FFTFTFFTFTFTFTTT
Open list
bullet
A
the AC/DC section lid (1).
d

a
s

S20046

s d f g

A
Remove
bullet list the Expiratory valve coil cable (2) from
PC 2004/PC 2024.
A
• Remove
bullet list the Expiratory flowmeter ribbon cable (3)
from PC 2004/PC 2024.
A
DDDRRDDRDDRDRRRR

• Remove list the Expiratory transducer tube (4) from


A
bullet
PC 2004/PC 2024.
• Remove list the plastic cover (5) at the Expiratory
A
bullet
cassette outlet by bending it towards the right and
up.
A A

a
S20116

SERVO-air Ventilator System, Service Manual 4 - 19


| 4 | Disassembling and assembling |

• Remove
bullet list the four screws that secure the cassette
holder.
• Carefully
bullet list lift at position (1) to release the Expiratory
cassette holder from the metal bracket.
• Carefully
bullet list lift off the Expiratory cassette holder and
at the same time release the two cables and tube
from the silicone guide (2).

AFFTFTT
A
a
s
S20117

4.11.6 PC 2031/PC 1785 Expiratory channel


connector
A
DDDRRR

• Remove
bullet list the Expiratory cassette holder.
• Remove
bullet list the two screws that secure the s
PC 2031/PC 1785 to the Expiratory cassette holder.
• Carefully
bullet list pull out PC 2031/PC 1785 (1). a
For software and hardware requirements and
compatibility, see the Compatibility chart in chapter
Revision history.

4.11.7 Expiratory valve coil

• Remove
bullet list the Expiratory cassette holder. S20119

• Remove
bullet list the four screws that secure the Expiratory
valve coil to the Expiratory cassette holder.
• Remove
bullet list the Expiratory valve coil (2).
FTTT

4 - 20 SERVO-air Ventilator System, Service Manual


| Disassembling and assembling | 4 |

4.12 Front section


4.12.1 Nebulizer

• Remove
bullet list the Chassis.
• Remove
bullet list the five screws (1) that secure the
Nebulizer lid.
• PC 1994
bullet list Nebulizer (2) is mounted inside the
Nebulizer lid and can now be removed from the lid.

s a
S20031

4.12.2 Front section lid including PC 2028 DC/DC

T TTTT TTT
& Battery control

• Remove
bullet list the Nebulizer.
• Remove
bullet list the four screws (1) that secure the User
interface cable lid (2) to the Patient unit. d s
AAAAFFFF AAAFFF
• Remove
bullet list the User interface cable lid.
• Disconnect
bullet list the User interface cable (3) from the
connector.

a
S20124

• Remove
bullet list the seven screws (1) that secure the Front
panel lid (incl. PC 2028) to the Patient unit.
DDDRRR

Note: Some of the screws may have been removed


earlier.

• Carefully
bullet list pull out the lower part of the Front section
lid to release the PC board connectors, then slide
the lid and the board down to remove.

S20028

SERVO-air Ventilator System, Service Manual 4 - 21


| 4 | Disassembling and assembling |

4.12.3 PC 2026 Battery extension board

• Remove
bullet list the Front section lid including PC 2028
DC/DC & Battery control
• Carefully
bullet list pinch the plastic spacer (1) that secures
PC 2026, i.e. with pliers.
a
• Remove
bullet list PC 2026 (2).

4.12.4 PC 2025 Battery back-plane

• Remove
bullet list the Front section lid including PC 2028
d
DC/DC & Battery control
• Remove
bullet list the four screws that secure PC 2025. s
• Lift outlist
bullet PC 2025 (3).

FTFTTT FFFTTTFFTFTT
4.13 Rear section
A
S20115

4.13.1 Rear connectors lid including PC 2021 Rear


connectors
A
• Remove
bullet list the Top cover.
A
• Remove
bullet list the two screws (1).
DDRRDRDDRRR

• Remove
bullet list the mains power cable restrainer (2). a
Not shown in the illustration.
• Remove
bullet list the screw (3).
• Remove
bullet list the five screws (4) that secure the Rear
A
connectors lid (incl. PC 2021) to the Patient unit.
d
• Carefully
bullet list slide out the Rear connectors lid.
A

s
A

f
S20030

4 - 22 SERVO-air Ventilator System, Service Manual


| Disassembling and assembling | 4 |

4.13.2 On/Off switch

• Remove
bullet list the Rear connectors lid including PC 2021
Rear connectors.
a
• Disconnect
bullet list the On/Off switch cable (1) from
PC 2021.
• Remove
bullet list the nut and washer (2) that secure the
On/Off switch.
• Lift offlist
bullet the On/Off switch.
s

TTTTT TTTT
S20123

4.13.3 Battery guides


AAAFAFFFF AAAFAFFF
• Remove
bullet list the chassis.
• Remove
bullet list the gas module lid.
• Remove
bullet list the two screws (1).
• Insert list
bullet a thin blade between the plastic battery guide
and the metal wall. This will release the guides.
a
• Pull offlistboth plastic guides (2 and 3).
bullet
1
When assembling, make sure that the leaf springs (4) 1.4 Nm
2

are mounted in the left-side guide.


DDDRDRRR

f d s S20131

SERVO-air Ventilator System, Service Manual 4 - 23


| 4 | Disassembling and assembling |

4.14 PC 2020 Main back-plane

• Remove
bullet list the Chassis.
• Remove
bullet list the Main section lid and the PC boards.
• Remove
bullet list the Inspiratory section lid, and remove:
- Inspiratory flowmeter
- Safety valve
- Gas module lid
- O2 module
- Turbine module
- Inspiratory gas docking
- PC 2030 Inspiratory channel incl. PC 1781
Inspiratory pressure transducer
- PC 2029 Inspiratory channel extension
• Remove
bullet list the Nebulizer and the Front section lid.

TTTTT TTT
• Remove
bullet list the AC/DC section lid.
• Remove
bullet list the Rear connectors lid incl. PC 2021 Rear
connectors.
• PC 2020list Main back-plane is mounted with seven
AAFAAAFFFF AAAFFF
bullet
clips (1). To release the board from the clips, pull it
carefully approximately 3 mm towards the front.
• Lift thelist
bullet board slightly and then slide it out from the
front.
RRR

a
D

S20118
R
RR D
D

4 - 24 SERVO-air Ventilator System, Service Manual


| Disassembling and assembling | 4 |

4.15 Expiratory cassette


The Expiratory cassette is a complete unit and must
not be disassembled. There are two spare parts though
that may need to be replaced:

• Moisture
bullet list trap including O-ring
• Expiratory
bullet list cassette valve membrane

4.15.1 Moisture trap

• Use alist
bullet screwdriver and carefully bend on both sides
of the Moisture trap to remove it from the Expiratory
cassette.

TTT
AAFAFF
S20126

4.15.2 Expiratory cassette valve membrane

Operating capacity for the membrane is estimated to


10.000.000 breathing cycles. When this limit is passed
or if the membrane for some reason has become
DDDRRR

defective, it must be replaced.

Membrane capacity (counter) can be shown in the


System status window. Tap SYSTEM STATUS >
Expiratory cassette to check Membrane capacity. The
operating capacity meter must be reset after
replacement of the membrane.

Note: It is very important for the function of the


expiratory valve that the valve membrane is
removed and mounted correctly as described below.

To remove the valve membrane from the cassette

SERVO-air Ventilator System, Service Manual 4 - 25


| 4 | Disassembling and assembling |

• Carefully
bullet list remove the membrane including retaining
ring using a suitable retaining ring pliers.

S20041

To mount the valve membrane into the cassette

• Place list
bullet the retaining ring (1) correctly into the
membrane. Make sure that the retaining ring (1) and


FT FT FT
the washer (2) are correctly mounted.

Note: It is very important that the valve membrane


and the membrane seat in the cassette are clean.
Dirt particles can create leakage in the cassette.
A
Place list
bullet the membrane onto the cassette as shown in
S20042

the illustration.
DR DR

S20043

• Squeezelist the retaining ring pliers and carefully press


A

bullet
the membrane in place into the cassette as shown
in the illustration.

S20044

4 - 26 SERVO-air Ventilator System, Service Manual


| Disassembling and assembling | 4 |

• Carefully
bullet list release and remove the pliers.
• Checklist
bullet that the membrane is not deformed by the
retaining ring. If necessary, remove the membrane
and redo the complete mounting procedure.
• Mountlist
bullet the expiratory cassette onto the Patient unit.
• Resetlist
bullet the operating capacity meter after
replacement of the membrane. To access the reset
button, tap SERVICE & SETTINGS > SERVICE
REPORT > Report replacement of exp. cassette
membrane.
S20045

TTTT
AAAFAFFF
DDDRDRRR

SERVO-air Ventilator System, Service Manual 4 - 27


| 4 | Disassembling and assembling |

4.16 User interface


Remove the User interface from the Panel arm and
place it on an ESD protected work space, following
the instructions below.

4.16.1 Removing the User interface

• Remove
bullet list the Front cover.
• Remove
bullet list the four screws (1) that secure the User
interface cable lid (2) to the Patient unit.
s
• Lift offlist
bullet the User interface cable lid. d
• Disconnect
bullet list the User interface cable (3) from the
connector.

S20125

• Tilt back
bullet list the User interface. a s
• Remove
bullet list the four screws (2) that secure the Panel

T TT TTTT
arm lid (1), and then lift off the lid.
AAFFAAAAFFFF
S20047

• Remove
bullet list the four hinge screws (1).
DDDDRRRR

S20048

4 - 28 SERVO-air Ventilator System, Service Manual


| Disassembling and assembling | 4 |

4.16.2 Opening the User interface

• Remove
bullet list the caps (1) covering the screws on the
heat sink.
• Remove
bullet list the five screws (2) and lift off the User
interface from the holder.

T TT
a s
S20049

• Disconnect
bullet list the User interface cable (1) from the
AF AAFF
User interface.
DDRR

S20050

SERVO-air Ventilator System, Service Manual 4 - 29


| 4 | Disassembling and assembling |

• Disconnect
bullet list the loudspeaker cable (1). a s
• Remove
bullet list the eight screws (2) and lift off the backside
of the panel chassis.

S20051

FTT FT FFTT
• The loudspeaker
bullet list (1) is attached to the backside of
the panel chassis using double-coated adhesive
tape.
RA RRAA
a
S20052

• Remove
bullet list the Alarm indicator (1).
s a
• Remove
bullet list the eight screws (2) and lift off the metal
cover.
D
D

S20053

4 - 30 SERVO-air Ventilator System, Service Manual


| Disassembling and assembling | 4 |

• Disconnect
bullet list the cables (2-4), and then remove the
seven Panel board screws (1). a
• Remove
bullet list the LCD display (6) and PC 2023 Panel (5),
and place them on an ESD protected work space.
g s
Note: Additional actions may be required after
replacement of PC boards. Refer to section
'Replacing PC boards' in this chapter. h d

f
S20054

TT
AAFF
DDRR

SERVO-air Ventilator System, Service Manual 4 - 31


| 4 | Disassembling and assembling |

4 - 32 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

5 Service procedures

Table of Contents
5.1 Service key | 5-2
5.2 Service & Settings | 5-3
5.3 Software installation | 5 - 34
5.4 Leakage detection | 5 - 39
5.5 Battery modules | 5 - 45
5.6 Memory backup batteries | 5 - 47

SERVO-air Ventilator System, Service Manual 5-1


| 5 | Service procedures |

5.1 Service key 5.1.1 Connecting a USB mouse to the system

The service key is a USB memory stick provided to If the Touch screen does not function properly, due
personnel trained and authorized by Getinge to to a faulty panel or not being well enough calibrated,
perform installation, service or maintenance of the a mouse can be connected to the system. The mouse
system. can then be used to, e.g., logging in to Service &
Settings to start the Touch screen calibration.
With the service key it is possible to access different
user levels of the built-in Service & Settings software.
Note: Using a mouse with the system is only
There are two user levels intended for service that are
permitted for service purposes.
described in this manual:

• Trained
bullet listbiomed. User level intended for hospital 1. Connect the mouse to the USB port when the
biomed trained and authorized by Getinge. system is turned Off.

• Field service.
bullet list User level intended for trained and 2. Start the system.
authorized Getinge and Getinge partner’s 3. Perform your tasks, e.g., logging in to Service &
personnel. Settings and starting the Touch screen calibration.
The service key, and its associated access code, are 4. Disconnect the mouse, and then restart the
personal and must not be handed-over to anyone else. system.
The access code is unique for each service key.

Logs can be exported to a USB memory stick, refer


to section Save logs to USB: in chapter Service &
Settings. With a service key, logs can be exported
during ventilation and in Standby.

Notes:
FT T FFTT
• Whenlist
bullet saving information to a USB memory stick
(logs, screenshots, etc), allow this procedure to
complete before removing the memory stick.
Information will be shown on the control panel
when completed.
• Make list
bullet sure to log out (select EXIT) from Service
& Settings when completed. If the Service key is
removed while Service & Settings is open,
access remains until manual exit.
A
A
DDRR

5-2 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

5.2 Service & Settings


Service & Settings (S&S) is a built-in software provided
to facilitate troubleshooting, service and maintenance
of the system.

WARNING! Service & Settings must


not be activated with a patient
connected to the system.

Notes:
• It is only
bullet list possible to access Service & Settings
with the system in Standby mode.
• The Service
bullet list & Settings menus diagram in chapter
Diagrams gives an overview of the Service &
Settings functions.

5.2.1 To access Service & Settings

• Tap SERVICE
bullet list & SETTINGS in the extended menu.
The Service & Settings login menu is displayed.

There are five options in the login menu:


TTTT TT
BIOMED.
bullet list Tap BIOMED and enter the 4 digit code
1973 (default) to access this user level. The User’s
Manual describes user level BIOMED.
• SERVICE.
bullet list Only for personnel trained and authorized
AFFAAFF
by Getinge. A Service key connected to the system
and the associated access code is required.
• REMOTE
bullet list SERVICE. Remote Services start button.
• LICENSES.
bullet list Licenses for third party software
packages. Listed for legal reasons.
• EXIT. list
bullet To exit Service & Settings.
DDRR

SERVO-air Ventilator System, Service Manual 5-3


| 5 | Service procedures |

5.2.2 To access the BIOMED user level:

• Insert list
bullet the Service key.
• Tap BIOMED.
bullet list
• Enter list
bullet the access code and verify by tapping the
green check mark.
• The Service
bullet list & Settings start screen is displayed.

The start screen shows the actual user level.

The following choices are available:

• STATUS
bullet list
• LOGSlist
bullet
• SERVICE
bullet list REPORT
• CONFIGURATION
bullet list

TTTTTTTT TTTT
COPYlist
bullet CONFIGURATION
• INSTALLATION
bullet list
• CALIBRATION
bullet list
• EXIT
bullet list
AAAAAFAFAFAFFFFFAAAAFFFF
These Service & Settings main functions are described
later in this chapter.
DRDRDRRRRRDDDDRRRR

5-4 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

Status

STATUS is used for viewing system information,


installed software options and parts data.

The following information is available for System info:

• SYSTEM
bullet list INFORMATION
• O2 CELL
bullet list
• EXPIRATORY
bullet list CASSETTE
• BATTERY
bullet list STATUS

The following information is available for Options:


• INSTALLED
bullet list OPTIONS (software options)
FTFFFFTTTT
RARRRRAAAA
D
D
DD

SERVO-air Ventilator System, Service Manual 5-5


| 5 | Service procedures |

Logs

LOGS is used for viewing Event logs for a certain


period of time. A date interval can also be set and a
search function is available.

The following search filters are available for Event logs:

• Alarms
bullet list
• Settings
bullet list
• Functions
bullet list
• Configuration
bullet list

Notes:
• Logs can
bullet list be saved to a USB memory stick.
• Only alist
bullet USB memory stick may be connected to
FTFTTFFTT FFFFTTTT
the USB port.

Save logs to USB:

• Insert list
bullet a USB memory stick.
• Tap Save
bullet list to USB. It is not necessary to select logs
before saving to USB. All required logs will be
copied to the USB memory stick.
A
• The message
bullet list SAVE LOGS TO USB will be shown
A
during the file transfer. When completed, all log files
are saved in a ‘ventilatorData’ folder on the USB
A
memory.
A
• Remove
bullet list the USB memory stick.
DDDDRRRR
AA

5-6 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

Service report

SERVICE REPORT is used for reporting service tasks.

The following completed service tasks can be


confirmed:

• Reportlist
bullet replacement of exp. cassette membrane
• Reportlist
bullet completed preventive maintenance
• Reportlist
bullet replaced backlight. Not active from System
version 3.0 and not available from System version
4.0.
Tap Open keyboard to enter comments and a
signature.

With the button Edit, it is possible to set a shorter time


until next Preventive Maintenance. This function can
be used if the time stamp for the previous Preventive
Maintenance has been lost, for example after
replacement of the Main back-plane.

TTT
AAAFFF
DDDRRR

SERVO-air Ventilator System, Service Manual 5-7


| 5 | Service procedures |

Configuration

CONFIGURATION is used for viewing and editing the


systems configuration.

The following choices are available and explained


below:

• SETTINGS
bullet list
• STARTUP
bullet list CONFIGURATION
• ALARMS
bullet list
• NETWORK
bullet list
• SET DATE
bullet list & TIME
• BIOMED
bullet list CODE

TTTTTT
AAAAAFAFFFFF
DDDDDRDRRRRR

5-8 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

Settings

SETTINGS is divided into two sections; GENERAL


and UNITS.

The following settings are available in GENERAL:

• Remote
bullet listServices after pre-use check. To set if log
files should be sent via Remote Services after
pre-use checks.

The unit settings for the following parameters are


available in UNITS:

• Heightlist
bullet
• Weight
bullet list

TT T
AAFFAF
DDRRDR

SERVO-air Ventilator System, Service Manual 5-9


| 5 | Service procedures |

Startup configuration

STARTUP CONFIGURATION is used to set the


systems startup configuration.

In this menu, it is also possible to restore factory default


startup configuration.

Refer to the User’s Manual for information on startup


configuration.

5 - 10 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

Alarms

ALARMS is used to set the system alarm limits for:

• Patient
bullet listcategory and type of ventilation:
- Pediatric
- Pediatric NIV
- Adult
- Adult NIV
• Restore
bullet listdefault alarm limits.

• It is also
bullet list possible to set Alarm sound level or
Restore default audio level.

TT T
AAF FAF
DDRRDR

SERVO-air Ventilator System, Service Manual 5 - 11


| 5 | Service procedures |

Network

NETWORK is used to display and configure network


parameters when connecting the system to Remote
Services.

In the Network menu, it is possible to select


Auto (DHCP) or Manual configuration.

For installation and setup instructions, refer to the


‘MCare Remote Services, SERVO-air ventilator system
– Installation Instructions’.

5 - 12 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

Set date & time

SET DATE & TIME is used to set the systems real time
clock.

SERVO-air Ventilator System, Service Manual 5 - 13


| 5 | Service procedures |

Biomed code

BIOMED CODE is used to set the four-digit code used


to log in as Biomed to Service & Settings.

The default biomed code is 1973.

5 - 14 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

Copy configuration

COPY CONFIGURATION is used to export or import


the systems configuration using a USB memory stick.

This function can for example be used to quickly install


the same configuration to several systems.

Note: It is not possible to copy the startup


configuration from system version 4.2 to a ventilator
with a lower system version.

SERVO-air Ventilator System, Service Manual 5 - 15


| 5 | Service procedures |

Installation

INSTALLATION is used to install:

• PERMANENT
bullet list OPTIONS. This function will install
Software options.
• SOFTWARE
bullet list INSTALLATION. This function will
install a new System software version.
Installation is described separately in section ‘Software
installation’.

TT
AAFF
DDRR

5 - 16 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

Exit

EXIT is used to log out from Service & Settings.

Make sure to log out (select EXIT) from Service &


Settings when completed. If the Service key is removed
with Service & Settings open, access to Service &
Settings remains.

5.2.3 To access the SERVICE user level:

• Insert list
bullet the Service key.
• Tap SERVICE.
bullet list
• Enter list
bullet the access code and verify by tapping the
green check mark.
• The Service
bullet list & Settings start screen is displayed.

The start screen shows the actual user level.

The following choices are available:

• STATUS
bullet list
• LOGS
bullet list
• TESTS
bullet list
• SERVICE
bullet list REPORT

FTTFTFTTTTFFFTFTTT
CONFIGURATION
bullet list
• COPY CONFIGURATION
bullet list
• INSTALLATION
bullet list
• CALIBRATION
bullet list
• EXIT list
bullet
These Service & Settings main functions are described
later in this chapter.
RRRAAAA

SERVO-air Ventilator System, Service Manual 5 - 17


| 5 | Service procedures |

Status

STATUS is used for viewing system information,


installed software options and parts data.

The following information is available for System info:

• SYSTEM
bullet list INFORMATION
• O2 CELL
bullet list
• EXPIRATORY
bullet list CASSETTE
• BATTERY
bullet list STATUS

The following information is available for Options:


• INSTALLED
bullet list OPTIONS (software options)

TTT T TTTTF
FFF
RARRRRAAAA
The following information is available for Parts data
(parts with ID-PROM):
• PARTlist
bullet (name)
• ARTICLE
bullet list NUMBER
• VERSION
bullet list NUMBER
• SERIAL
bullet listNUMBER
F
DD
D
D

5 - 18 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

Logs

LOGS is used for viewing Service logs and Event logs


for a certain period of time. A date interval can also be
set and a search function is available.

The following search filters are available for Service


logs:

• Test results
bullet list
• Test details
bullet list
• Preventive
bullet list maintenance
• Service
bullet listreport
• Technical
bullet list alarms
• Installation
bullet list

The following search filters are available for Event logs:

• Alarms
bullet list
• Settings
bullet list
• Functions
bullet list
• Configuration
bullet list

TT TTTTTTTTTTTT
Notes:


Logs can
bullet list be saved to a USB memory stick.
Only alist
bullet USB memory stick may be connected to
the USB port.
AAFAFAAFFAFAFAAAFAFFFFF
Save logs to USB:

• Insert list
bullet a USB memory stick.
• Tap Save
bullet list to USB. It is not necessary to select logs
before saving to USB. All required logs will be
copied to the USB memory stick.
• The message
bullet list SAVE LOGS TO USB will be shown
during the file transfer. When completed, all log files
DRDRDDDRDRRRRR

are saved in a ‘ventilatorData’ folder on the USB


memory.
• Remove
bullet list the USB memory stick.

SERVO-air Ventilator System, Service Manual 5 - 19


| 5 | Service procedures |

Tests

TESTS is used to perform a Pre-use check, either the


complete check, or any of the separate tests included
in the Pre-use check.

• Tap Select
bullet list PUC tests to perform a complete Pre-use
check.

All Pre-use checks are described in chapter


Troubleshooting.

The TEST menu also contains choices to:

• Selectlist
bullet breathing tests (selected in the screenshot)
• Selectlist
bullet monitoring tests
• Selectlist
bullet panel tests
These tests are included in Pre-use check, but when
FFFTTT FT
performed in this menu, they are limited to each
subsystems function.
RRRAAA RA
D

5 - 20 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

The tool Manual panel tests is primarily intended for


manufacturing tests, but can also be used during
service or maintenance.

The Manual panel tests requires interaction with the


user and performs the following subtests:

• Devicelistdata. Shows device data.


bullet
• Calibration
bullet list test. Tap on the circles to check the
touch screen calibration.
• Backlight
bullet list test. To check the backlight at different
levels.
• LED test.
bullet list Tap the swatches and check that the
corresponding color on the alarm indicator is lit.
• Pixel RGB
bullet list test. The screen switches between
different colors to make it possible to check for
defective pixels.
• Soundlist
bullet test. To check different sound levels.
• Ethernet
bullet list test. Checks the Ethernet communication
between the User interface and the Patient unit.
Tap Run to start the test. To continue the test

TTTTTTT
sequence, swipe diagonally on the screen.

After each test, the user can select Yes if the test
passed, No if the test failed or Cancel to cancel the
AAAFFAAFAFAFFF
test. These results can be saved as a log file on a USB
memory stick.
DRRDDRDRDRRR

SERVO-air Ventilator System, Service Manual 5 - 21


| 5 | Service procedures |

Service report

SERVICE REPORT is used for reporting service tasks.

The following completed service tasks can be


confirmed:

• Reportlist
bullet replacement of exp. cassette membrane
• Reportlist
bullet completed preventive maintenance
• Reportlist
bullet replaced backlight. Not active from System
version 3.0 and not available from System version
4.0.
Tap Open keyboard to enter comments and a
signature.

With the button Edit, it is possible to set a shorter time


until next Preventive Maintenance. This function can
be used if the time stamp for the previous Preventive
Maintenance has been lost, for example after
replacement of the Main back-plane.

TTT
AAAFFF
DDDRRR

5 - 22 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

Configuration

CONFIGURATION is used for viewing and editing the


systems configuration.

The following choices are available and explained


below:

• SETTINGS
bullet list
• STARTUP
bullet list CONFIGURATION
• ALARMS
bullet list
• NETWORK
bullet list
• SET DATE
bullet list & TIME
• BIOMED
bullet list CODE

TTTTTT
AAAAAFAFFFFF
DDDDDRDRRRRR

SERVO-air Ventilator System, Service Manual 5 - 23


| 5 | Service procedures |

Settings

SETTINGS is divided into two sections; GENERAL


and UNITS.

The following settings are available in GENERAL:

• Date format
bullet list
• bullet list separator
Decimal
• Language
bullet list
• PM reminder
bullet list (On/Off)
• Enable
bullet listbiomed service report (On/Off). Will
enable/disable the SERVICE REPORT function in
Biomed user level. Available in Field service user
level only.
• Remote
bullet listServices after pre-use check. To set if log
files should be sent via Remote Services after
pre-use checks.
• Flashing
bullet list standby message (On/Off)

The unit settings for the following parameters are


• TTTTTT TTTTTTT
available in UNITS:

Airwaylist
bullet pressure
Flow list
bullet
AFAFFFFFAAFAFAAAAFFFFF
• Gas supply
bullet list pressure
• Barometric
bullet list pressure
• Heightlist
bullet
• Weight
bullet list
R DD
RDDR
DRR
RR

5 - 24 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

Startup configuration

STARTUP CONFIGURATION is used to set the


systems startup configuration.

In this menu, it is also possible to restore factory default


startup configuration.

Refer to the User’s Manual for information on startup


configuration.

SERVO-air Ventilator System, Service Manual 5 - 25


| 5 | Service procedures |

Alarms

ALARMS is used to set the system alarm limits for:

• Patient
bullet listcategory and type of ventilation:
- Pediatric
- Pediatric NIV
- Adult
- Adult NIV
• Restore
bullet listdefault alarm limits.

• It is also
bullet list possible to set Alarm sound level or
Restore default audio level.

TT T
AAF FAF
DDRRDR

5 - 26 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

Network

NETWORK is used to display and configure network


parameters when connecting the system to Remote
Services.

In the Network menu, it is possible to select


Auto (DHCP) or Manual configuration.

For installation and setup instructions, refer to the


‘MCare Remote Services, SERVO-air ventilator system
– Installation Instructions’.

SERVO-air Ventilator System, Service Manual 5 - 27


| 5 | Service procedures |

Set date & time

SET DATE & TIME is used to set the systems real time
clock.

5 - 28 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

Biomed code

BIOMED CODE is used to set the four-digit code used


to log in as Biomed to Service & Settings.

The default biomed code is 1973.

SERVO-air Ventilator System, Service Manual 5 - 29


| 5 | Service procedures |

Copy configuration

COPY CONFIGURATION is used to export or import


the systems configuration using a USB memory stick.

This function can for example be used to quickly install


the same configuration to several systems.

Note: It is not possible to copy the startup


configuration from system version 4.2 to a ventilator
with a lower system version.

5 - 30 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

Installation

INSTALLATION is used to install:

• PERMANENT
bullet list OPTIONS. This function will install
Software options.
• SOFTWARE
bullet list INSTALLATION. This function will
install a new System software version.
Installation is described separately in section ‘Software
installation’.

TT
AAFF
DDRR

SERVO-air Ventilator System, Service Manual 5 - 31


| 5 | Service procedures |

Calibration

CALIBRATION is used for calibration of:


• Barometer
bullet list
• Touchlist
bullet screen

Barometer

Barometer calibration is used to calibrate the internal


barometer. A barometer or information about the actual
barometric pressure at the installation site is required.

Note: The actual barometric pressure is affected by


altitude and weather. Make sure that the absolute
barometric pressure at your current location and
elevation is used. Do not use barometric pressure
readings from a weather service as these typically
report barometric pressure at sea level.

FTFFFTTT FFTT
The barometer must be calibrated:




If the Barometer
bullet
bullet
bullet
list
If a Technical
list
list
test in Pre-use check fails.
error code indicates Barometer error.
If the barometric pressure check performed during
installation or Preventive maintenance shows that
calibration is required, that is if the Barometric
pressure shown in the SYSTEM STATUS > General
A
window differs more than ±5% from the actual
A
barometric pressure.
• After replacement
bullet list of PC 1992 Monitoring or in case
of malfunction in the memory backup battery on
DDRR

PC 1992.

To calibrate the barometer:

• Tap CALIBRATION
bullet list and then tap BAROMETER.
• Tap Update.
bullet list
• Enter list
bullet the actual barometric pressure and tap the
OK button.
After calibration, check that the new Barometric
pressure value corresponds to the set value. Note that
the new Barometric pressure value may differ 1-2 hPa
from the set value.
AA
A

5 - 32 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

Touch screen

The touch screen must be calibrated:

• If an inaccuracy
bullet list is seen in the touch screen function.
• After replacement
bullet list of PC 2023 Panel.

To calibrate the touch screen:

• Tap CALIBRATION
bullet list and then tap TOUCHSCREEN.
• Tap Start
bullet list calibration.

T T TTTT
• Use alist
bullet narrow-tipped tool to tap each of the nine line
intersections that are presented on the screen.

Note: Do not use sharp tools on the screen.


FF
RARRRARAAA
FF
• Tap each
bullet list of the nine circles presented on the
screen.

When the calibration is completed, the Start calibration


menu appears.
D

Note: If the touch screen is not calibrated well


enough, it can be difficult to enter Service & Settings
D
F
to start the calibration procedure. If so, a mouse can
be connected to the system to facilitate the
navigation. For instructions, refer to Connecting a
D

USB mouse to the Patient unit in chapter Service


D

procedures, section Service key.

SERVO-air Ventilator System, Service Manual 5 - 33


| 5 | Service procedures |

Exit

EXIT is used to log out from Service & Settings.

Make sure to log out (select EXIT) from Service &


Settings when completed. If the Service key is removed
with Service & Settings open, access to Service &
Settings remains.

reinstalled from the USB memory stick as part of


5.3 Software installation the software installation procedure. For lower
system versions, copy the Startup configuration to
5.3.1 General a USB memory stick prior to a software installation
and reinstall the configuration afterwards.
• This chapter
bullet list describes installation of System
software and Software options. • Make list
bullet sure that the software is fully compatible with
all hardware, software and mechanical components
• Only personnel
bullet list trained and authorized by Getinge
in the system. See the Compatibility chart in chapter
shall be permitted to install System software and
Revision history.
Software options.
• Software
bullet list installation is performed in Service &
Settings. The Software installation function is
available in BIOMED and SERVICE user levels.
With BIOMED user level, Service key is not
required.
• For functionality
bullet list enhancement, the latest released
System software version is always recommended.

5.3.2 Check before installation


T TTT
• It is not
bullet listrecommended to install System software
with lower version number than already installed in
the system:
- Check installed System software version, see
AAFAFF

System SW version in SYSTEM STATUS >


General.
- Check the System software version stored on
the USB memory stick.
A
• Somelist
bullet parts of the configuration may be changed
during the System software installation. With
System version 4.0 or higher, the Startup
configuration will automatically be copied to and
R

5 - 34 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

5.3.3 Check after installation

• A software
bullet list installation may change the system
functionality and therefore require a new version of
the User's Manual.
• After any
bullet list software installation in the system, perform
a Pre-use check according to instructions in the
User's Manual.

CAUTION: The system must not be


switched off during the software
installation process. Such interruption
may cause malfunctions on PC boards.

5.3.4 Materials/documents required

TTT TT
USB memory
bullet list stick with the System software
version/Software option to be installed. Only
Getinge approved USB memory sticks must be
used.
• User’slist
Manual.
AAFAFF AAFF
bullet
• Gas supply
bullet list required for the Pre-use check.
R DDRR

SERVO-air Ventilator System, Service Manual 5 - 35


| 5 | Service procedures |

5.3.5 Software information

• System
bullet listsoftware: A System software installation

Software option
installs a new System software version. A System Software option

installation
software installation is not depending on the serial SW Option
Software option
number of the system and does not alter the SW Option
installed Software options. Software option

• Software
bullet list option: A Software option installation
installs Software options ordered for the concerned
system. A Software option is individually created for System software

System software
installation
each system and can only be installed on the
intended system. Serial number of the system must Corrections
and function
be stated when ordering a Software option.
improvement

S25085

TT
AFAF
DDRR

5 - 36 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

5.3.6 System software installation

To perform a System software installation:

• Turn the
bullet list system On.
• Enter list
bullet Service & Settings (BIOMED or SERVICE
user level).
• With System
bullet list version 4.0 or higher, the Startup
configuration will automatically be copied to the USB
memory stick as part of the software installation
procedure. For lower system versions, copy the
configuration to the USB memory stick.
• Connect
bullet list the USB memory stick, with the System
software version to be installed.
• Tap INSTALLATION.
bullet list
• Tap SOFTWARE
bullet list INSTALLATION.
• The software
bullet list installation activation box will now
open. Check that correct System SW version to be

FFTFTFTFTFTFFTFTTTFFFTTT



installed is stated in this box.
Tap Yes
bullet listto start the installation. The software
installer screen will now open and the installation
will start.
The installer
bullet list
installation.
Whenlist
bullet
screen now shows the progress of the

the installation is completed (after 10–11


minutes), this is notified in the installer screen. See
A
Note 1 below.
A
• Restart
bullet listthe system. See Note 2 below.
• System
bullet listversion 4.0 or higher, the Startup
A
configuration will automatically be copied from the
USB memory stick as part of the software installation
DDRRRDRDDRRR

procedure. For lower system versions, connect the


USB memory stick containing the configuration and
copy the configuration from the USB memory stick.
A
• Perform
bullet lista Pre-use check according to instructions
in the User's Manual.
A

Note 1: Very occasionally, it may seem that the Note 2: After a software installation, technical alarms
A

installation is interrupted and the Service & Settings may be activated due to reset internal memory on the
A

login menu will appear. If so, do not cancel the PC boards. Restart the system once more and check
installation. The installation will proceed in the that no technical alarms are activated.
background. Wait 15 minutes before restarting the
A

system.
A

SERVO-air Ventilator System, Service Manual 5 - 37


| 5 | Service procedures |

5.3.7 Software option installation

To perform a Software option installation:

• Turn the
bullet list system On.
• Enter list
bullet Service & Settings (BIOMED or SERVICE
user level).
• Connect
bullet list the USB memory stick with the Software
options to be installed.
• Tap INSTALLATION.
bullet list
• Tap PERMANENT
bullet list OPTIONS.

• The software
bullet list installation activation box will now
open. Furthermore, all software options that will be
available in the system after the installation will be
listed on the screen.





FFFTTFFTTFTFTFFTTFFTTT
Tap Yes
bullet listto start the installation.
Whenlist
bullet

Wait two
bullet

bullet
Perform
bullet
the installation is completed (after a few
seconds), this is notified in the installer screen.
list minutes to allow the Service log to be
updated.
Restartlistthe system.
lista Pre-use check according to instructions
in the User's Manual.
RARAARARARRARARAAA
D DDD D D

5 - 38 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

5.4 Leakage detection Test case 3 – Inspiratory section:

5.4.1 General • Nozzlelistunit in O2 gas module


bullet

When a leakage is detected in the Internal leakage • Gas module


bullet list silicone connector muff

test, the Pre-use check will stop and display either • Inspiratory
bullet list gas docking incl. one-way valve
Leakage or Excessive leakage. To locate the leakage, • Inspiratory
bullet list flowmeter connector muff
the three leakage tests described below can be used. • Inspiratory
bullet list flowmeter
Test equipment: • Inspiratory
bullet list channel connector muff
• Inspiratory
bullet list pressure silicone tube
• Test tube
bullet list
• Safetylist
bullet valve membrane
• bullet list detector tool, P/N 68 81 146
Leakage
• Safetylist
bullet valve pull magnet
• Manometer,
bullet list local purchase (e.g. a Cuff filling
• O2 celllist
bullet bacteria filter seal
measuring device with manometer)
• Inspiratory
bullet list pressure silicone tube

5.4.2 Parts tested during the check • Inspiratory


bullet list pressure bacteria filter
• Inspiratory
bullet list bacteria filter holder incl. tube
Test case 1 – Test tube:
• Inspiratory
bullet list pressure transducer

• Test tube
bullet list

Test case 2 – Expiratory channel:

• Expiratory
bullet list inlet incl. moisture trap with O-ring
• Expiratory
bullet list cassette
FDDFDFDFTDFTDFTDTDTDFTDDTFFFTTT




RRRRRARARARARARARARARARAAAAA
Expiratory
bullet
bullet
list valve membrane
Expiratory
list pressure transducer tube
Expiratory
bullet list pressure transducer
Expiratory
bullet list valve coil
RAAARRRAAA
D D D

SERVO-air Ventilator System, Service Manual 5 - 39


| 5 | Service procedures |

5.4.3 Starting the leakage detection

• Start the
bullet list manual leakage detection when the
Pre-use check has stopped during the 'Internal
leakage test' and displayed either Leakage or
Excessive leakage.
• Do notlist
bullet cancel the Pre-use check, the Redo function
will be used during the leakage detection.

5.4.4 Test case 1 - Test tube

1. Check the test tube for leakage. Replace if


required.
2. Reconnect the test tube properly.
3. Tap Redo.
If the leakage remains, continue with Test case 2.

TT
Insp.
Exp.
AAFF
S20087
DDRR

5 - 40 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

5.4.5 Test case 2 – Expiratory channel

Expiratory channel

H PC 1786 Exp. channel cassette


Expiratory inlet Expiratory B A
including moisture cassette PC 2031/
trap with O-ring PC 1785 Expiratory
Expiratory valve
channel membrane
connector

PC 2031/ Expiratory pressure


PC 1785 transducer tube

Expiratory
valve coil
Part of leakage
detector kit

Expiratory
pressure
transducer PC 1781

P40 N40
P50
MOD1

Manometer

PC 2004/PC 2024 Expiratory channel

S20088

1. Connect the test equipment to the expiratory inlet.


2. Tap Redo.
3. When the expiratory valve closes, pressurize to
approx. 50 cmH2O. This must be done quickly
because the expiratory valve coil is closed only
for 2-3 seconds. A pressure drop indicates a
leakage.
If a leakage is detected in Expiratory channel,
continue with Expiratory valve coil. If not, go to Test
case 3.

SERVO-air Ventilator System, Service Manual 5 - 41


| 5 | Service procedures |

Expiratory valve coil

1. Remove the expiratory cassette.


2. Tap Redo.
3. Make sure the expiratory valve coil axis is
activated (pushed up).
If the axis is activated, continue with Expiratory
cassette. If not, check/replace the items listed below,
one at a time:
• Checklist
bullet the expiratory valve coil including cable and
connector.
• Replace
bullet list PC 2004/PC 2024 Expiratory Channel.
• Replace
bullet list PC 1991 Control.

Expiratory cassette

1. Connect the test equipment to the expiratory inlet


Press
(while the cassette is removed).
2. Block the expiratory pressure transducer channel
hole on the cassette and close the expiratory valve
by hand, see illustration.
3. Pressurize to approx. 50 cmH2O. If a leakage is
FTFTT TTT
detected check/replace the following parts (one at
a time):
• Expiratory
bullet list inlet incl. moisture trap with O-ring.
• Expiratory
bullet list valve membrane. The membrane
AAAFFF
may not close properly due to cleaning residues
on the membranes sealing surface and/or on
the corresponding sealing surface in the
S20089
cassette.
• Expiratory
bullet list cassette.

Expiratory pressure transducer and tube


DDDRRR

1. If a leakage is detected in the Expiratory channel


but still remains after performed tests Expiratory
valve coil and Expiratory cassette, check the
Expiratory pressure transducer and Expiratory
pressure transducer tube.

5 - 42 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

5.4.6 Test case 3 – Inspiratory section

O2 cell O2 gas module


O2 cell
bacteria Nozzle unit O2
Inspiratory filter seal in gas inlet
outlet Inspiratory module
Inspiratory channel Inspiratory gas
pipe docking
Insp. Ins
flo pi
pressure wm rat
tube eteory
Part of leakage Safety valve r
detector kit membrane Insp. Turbine module
pressure PC 1781
Manometer Inspiratory Air
bacteria inlet
Safety valve filter and holder pressure
pull magnet transducer

N17P17
S20090

Complete Inspiratory section Safety valve piston

1. Connect the test equipment to the Inspiratory If no leakage was found in the above mentioned parts,
outlet. remove the Inspiratory channel and restart the Pre-use
2. Pressurize to approx. 50 cmH2O. A pressure drop check. Check that the Safety valve piston is activated
indicates a leakage. If a leakage is indicated, the during the internal leakage test.
parts below should be checked/replaced:
If the piston is activated, continue with Inspiratory gas
• Inspiratory
bullet list channel. docking and Inspiratory pipe. If not, check/replace
Note: In order to remove the Inspiratory as follows (one at a time):
channel, the safety valve must be opened (it is
closed during the test). This is done by • Checklist
bullet the safety valve piston including cable and
cancelling the test and setting the ventilator to connector.
Standby. • Replace
bullet list PC 2030 Inspiratory Channel.
• O2 celllist
bullet bacteria filter seal. • Replace
bullet list PC 2004/PC 2024 Expiratory Channel.
• Safetylist
bullet valve membrane. Clean if necessary.
• Inspiratory
bullet list pressure tube.
• Inspiratory
bullet list pressure bacteria filter and holder.
FFTTTTFT
A
A
SERVO-air Ventilator System, Service Manual 5 - 43
| 5 | Service procedures |

Inspiratory gas docking and Inspiratory pipe

Gas module silicone O2 gas module


Inspiratory connector muff
channel
connector Nozzle unit O2
muff in O2 gas inlet
Inspiratory module
gas
docking
In
flo spi
Part of leakage wm rat
detector kit eteory
r
Turbine module
Inspiratory pipe
Manometer Air
inlet
Inspiratory
flowmeter
connector
muff
S20091

1. Connect the test equipment to the inspiratory


channel connector muff outlet, i.e. the inspiratory
channel removed, see picture.
2. Pressurize to approx. 50 cmH2O. A pressure drop
indicates a leakage. If a leakage is indicated, the
inspiratory section must be opened for further
investigation. See chapter Disassembling and
Assembling. Check/replace the following parts in
the Inspiratory section:
• Nozzlelistunit in the O2 gas module.
bullet
• Gas module
bullet list silicone connector muff.
• Inspiratory
bullet list gas docking incl. one-way valve.
Clean the valve if necessary.
• Inspiratory
bullet list flowmeter connector muff.
• Inspiratory
bullet list flowmeter.
• Inspiratory
bullet list channel connector muff.

Note: The Inspiratory pressure transducer and the


holder for inspiratory pressure bacteria filter are
bypassed in this test. They should be checked
visually while the section is opened.
FFTFTFTFTTT
5.4.7 Completing the Leakage detection

1. Restore the unit including test tube.


2. Select Redo and make sure the test passes.
A

5 - 44 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

5.5 Battery modules


5.5.1 General

• The batteries
bullet list condition and capacity are verified
during the Pre-use check and also continuously
during operation of the unit. No further function
checks of the batteries are required.
• Typical
bullet listrecharge time approximately 2 h/battery
(90 %), up to 3 h (100 %) if battery is completely
discharged. If a battery is fully discharged, e.g. due
to long storage time, each battery may require up
to 6 hours charging time.
• The System
bullet list software will monitor:
- Remaining life time. The Battery life time is
300 charge cycles or five years (60 months) from
manufacturing date, whichever comes first.

Note: For system version 4.1 or lower the

T TT
battery life time is four years (48 months) from
manufacturing date.

- If the operational capacity is too poor for


AFAAFF
continued usage.
In both cases, battery replacement information will
be shown on the User interface.

CAUTION: Do not expose the batteries


to water, fire or excessive heat. Do not
crush, disassemble, puncture or short
circuit the connector terminals.
DDRDRR

SERVO-air Ventilator System, Service Manual 5 - 45


| 5 | Service procedures |

5.5.2 Battery status

Select SYSTEM STATUS > Batteries to open the


Battery status window.

The Battery status window shows:

• Battery
bullet listcapacity. Calculated as the sum of the
estimated backup time for the connected battery
modules reduced with 20 minutes as an extra safety
feature. This estimate may differ from the actual
usable backup time during running. Usable backup
time depends on set mode and selected ventilation
settings.
• Slot number.
bullet list
• Battery
bullet listmodule S/N.
• Remaining
bullet list operating time in minutes for each
battery.
• Charge
bullet listindicator for each battery. If the ventilator
system is running on battery power, the indicator
for the active battery turns yellow.

TTTTT T
Remaining
bullet list life time (months) and charge cycles for
each battery.
If Replace battery-message is displayed, the battery
is unreliable, regardless of the operating time displayed
AAFAAAFFFFAF
under Batteries. In this situation, replace the battery
even when the STATUS window indicates that
significant operating time remains.

At least one battery module must always be installed.


DRRRRDR

5 - 46 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

5.6 Memory backup batteries


5.6.1 General

• The Memory
bullet list backup batteries must be replaced
after five years. A Technical error message will
appear on the screen if the battery voltage level is
too low.
• Always
bullet listreplace both batteries at the same time to
keep the same replacement date for both batteries.
• After any
bullet list maintenance or service of the system,
perform a Pre-use check. Refer to the User's
Manual.

Note: ESD sensitive components.


The Memory backup batteries are
mounted on ESD sensitive PC boards.
Refer to chapter Disassembling and
assembling, section Handling PC

TTTT TTT
boards for further information regarding
ESD sensitive components.
AFFFF AAFAFF
5.6.2 Preparations

• Turn Off
bullet list the system.
• Disconnect
bullet list the mains power cable.
• Disconnect
bullet list the O2 gas supplies (wall and/or
cylinder).
• Remove
bullet list patient tubing.
DDRDRR

SERVO-air Ventilator System, Service Manual 5 - 47


| 5 | Service procedures |

5.6.3 Replacing Memory backup battery

• Remove
bullet list the Patient unit front cover. Refer to chapter
Disassembling and assembling. s d
• Carefully
bullet list pull out PC 1991 and PC 1992.
Note: Do not remove the cable connector (1).
Certain information stored in the PC backup is
disconnected:
- The battery backup on PC 1992 supplies power
to the real time clock and to the memory
containing logs and settings for the last minutes,
warm and cold start information, all of which is
lost at disconnection.
- The battery backup on PC 1991 supplies power
to the memory containing startup ventilation
configuration, which is lost at disconnection.

TTTTT TT
Therelist
bullet are two equal battery connectors (2 and 3)
on the PC boards. Connect the new battery to the
a f
un-used connector. The memory functions are now S20093

secured by the new battery.


• Cut thelistcable ties holding the old battery (4) to the
AAFAFFAFF AAFF
bullet
PC board.
• Disconnect
bullet list and remove the old battery.
• Mountlist
bullet the new battery onto the PC board using new
cable ties as shown in the illustration.
• Insert list
bullet the PC board into the correct PC board slot
and reassemble the Patient unit.
R D DRR

5 - 48 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

SERVO-air Ventilator System, Service Manual 5 - 49


| Troubleshooting | 6 |

6 Troubleshooting

Table of Contents
6.1 General | 6-2
6.2 Start-up test | 6-2
6.3 Pre-use check | 6-3
6.4 Other tests | 6-6
6.5 Test logs | 6-7
6.6 Technical error codes and messages | 6 - 23

SERVO-air Ventilator System, Service Manual 6-1


| 6 | Troubleshooting |

6.1 General
The information in this chapter is applicable to System The troubleshooting guides below focus on the
version 1.0 – 4.2. For functionality enhancement, the technical problems. Information about clinically related
latest System software version is recommended. problems can be found in the User's Manual.

This chapter contains: Possible causes to malfunction, not mentioned in the


• Descriptions
bullet list of the Start-up test and the Pre-use following troubleshooting guides, that need to be
check. considered during troubleshooting:
• Information
bullet list about how to interpret the Test log and • Incorrectly
bullet list assembled system after cleaning,
the Technical error codes as well as recommended maintenance or service.
actions to eliminate malfunctions reported by the • Disconnected
bullet list or incorrectly connected cable
system. connectors, PC board connectors or interconnection
Before starting troubleshooting, try to eliminate all boards.
possibilities of operational errors. If the malfunction • Disconnected
bullet list or defective gas tubes.
remains, use the troubleshooting guides below as well
as the information in chapter Description of functions In the descriptions and troubleshooting guides below,
to locate the faulty part. Perform actions step by step the tested subsystems are referred to as:
and check that the malfunction is eliminated. • BRE =list
bullet Breathing

When the fault is corrected, carry out a complete


• MON list
bullet = Monitoring

Pre-use check as described in the User's Manual.


FDFDTDT FFTT • PAN =list
bullet Panel.

6.2 Start-up test

Internal technical tests Recommended action if the test fails


• Software
bullet listcheck 1. Restart the system. Do not touch the User interface
during system start-up. Interfering with the touch screen,
• bullet listEEPROM
Reading
loudspeaker grid and so forth, may affect the internal
• Checksum
bullet list EEPROM technical tests.

A
Audio test.
bullet list 2. Reinstall the System software.
(BRE + MON + PAN)
RRARAARARAA
A
DDRR

6-2 SERVO-air Ventilator System, Service Manual


| Troubleshooting | 6 |

6.3 Pre-use check Temperature and Moist sensor test (MON)

The system requests the user to start the automatic Checks the following sensors in the inspiratory section:
Pre-use check at every startup of the system. It is also
possible to select the Pre-use check via the Standby • Humidity
bullet list (0.0–100.0 %). Sensor on PC 2030
menu. Inspiratory channel.
• Turbine
bullet listtemperature (0.0–70.0 °C). Sensor inside
The User's Manual describes how to perform this Turbine module.
Pre-use check. The Pre-use check description on the
following pages gives more detailed information about
• Ambient
bullet list temperature (0.0–55.0 °C). Sensor on
PC 2030 Inspiratory channel.
the Pre-use check.
• Inspiratory
bullet list gas temperature. (0.0–60.0 °C). Sensor
Some of the recommended actions described below inside Inspiratory flowmeter.
refer to Service & Settings (S&S). The Service key is
required to access S&S. Troubleshooting can of Blower inlet test (MON)
course be performed without access to the S&S, but
Checks the turbine module inlet:
for some of the recommended actions, S&S will make
troubleshooting faster and easier.
• Inlet resistance.
bullet list 1.00–3.30 cmH2O/l/s at
2.70–3.30 l/s inspiratory flow.
Check if the fault remains after each performed service
action. Re-run the complete Pre-use check or run the • Inlet pressure
bullet list sensor. Pressure diff. sensor test
concerned test from S&S. offset steps 1586–1690.

In the Standby menu it is possible to perform a Alarm output connector test (MON)
separate Patient circuit test to evaluate circuit leakage
and measure circuit compliance. This separate test Remote alarm test in the S&S Test menu.
does not replace the complete Pre-use check.
Checks that the Alarm output connector relay switches
between open and closed.
6.3.1 Internal test sequence

Panel audio test (PAN)

Checks that the loudspeakers in the User interface


generate an adequate sound level. A sound level is
generated and a microphone mounted nearby the
loudspeakers measures the sound. PAN calculates
RRAARRRARAAA
Power failure test (MON)

Generates failures on different system voltages.


Checks that these failures are detected.

6.3.2 Barometer test sequence

the measured value to verify that the sound level Barometer test (MON)
meets the requirement.
Checks that the barometric pressure measured by the
Monitor audio test (MON)
D
internal barometer is within 650–1070 hPa (mbar).

Checks that the buzzer in the Patient unit generates Checks that the measured barometric pressure values
D
an adequate sound level. A power failure is generated between BRE and MON differ less than 8 hPa (mbar).
in order to activate the buzzer. A microphone mounted
D
nearby the buzzer measures the sound. MON
calculates the measured value to verify that the sound
level meets the requirement.
D

SERVO-air Ventilator System, Service Manual 6-3


| 6 | Troubleshooting |

6.3.3 Internal leakage test sequence • Checks


bullet listthat speed is 38000–42000 r/min.

Internal leakage test (BRE) • Checks


bullet listthat deceleration (fall time) from approx.
40000 to 5000 r/min is <0.150 s.
Checks the internal leakage, with the test tube
connected, using the inspiratory and expiratory 6.3.5 Pressure transducer test sequence
pressure transducers:
Breathing pressure transducer test (BRE)
• Generates
bullet list 50 cmH2O pressure.
Checks that the offset values for the pressure
• Checks
bullet listthat the measured pressure values differ
transducers do not differ more than ±6 cmH2O (±0.3 V)
less than 10 cmH2O between inspiratory and
from factory calibration.
expiratory pressure transducers.
• Checks
bullet listthat the system volume required to generate Monitor pressure transducer test (MON)
50 cmH2O is 0.2000–0.4000 l.
Calibrates and checks the inspiratory and expiratory
• Checks
bullet listthat the leakage at 50 cmH2O is maximum
pressure transducers:
0.0200 l/min.
• The new
bullet list offset values for the inspiratory and
To locate a leakage, a leakage detection test can be expiratory pressure transducers may not differ more
performed, see Leakage detection in chapter Service than ±6 cmH2O (±0.3 V) from factory calibration.
procedures. The Leakage detection test is a
complement to the recommended actions below. • With the
bullet list inspiratory pressure transducers factory
default gain used as a reference, a new gain factor
is set for the expiratory pressure transducer. The

TRTRTT RRTTTT
6.3.4 Turbine and Gas supply test sequence
new gain factor may not differ more than ±8% from
factory calibration.

AAA AA
Gas supply test (MON)
• Checks
bullet listthat the difference between BRE and MON
Checks gas supply: is less than ±0.8 cmH2O at 50 cmH2O during the
F
test.
• Start with
bullet list a short gas flush.
F
• Checks
bullet listthat the O2 gas supply pressure measured
• Checks
bullet listthat the pressure generated for the Monitor
pressure transducer tests is 60 cmH2O (±5 cmH2O).
by the internal gas supply pressure transducer is
within 200–600 kPa (2000–6000 mbar).
A
F
Expiratory valve test (BRE)
• Checks
bullet listthat the O2 supply pressure differs less than
A
100 hPa (mbar) during the test. Checks the expiratory valve at four different pressures:
DF
40, 20, 10 and 5 cmH2O. Measured values must be
• Checks
bullet listthat the measured supply gas pressure
within ±15 cmH2O.
D
values between MON and BRE differ less than
A
DDRDDRRR

20 hPa (mbar).
Checks offset and gain in the expiratory valve:
A

This test requires O2 gas supply.


• Offsetlist
bullet (0.06768–0.15684 A).

Turbine accuracy test (MON) • Gain (0.00730–0.01130


bullet list A/cmH2O).

Checks the turbine module: This test requires O2 gas supply.

• Checks
bullet listthat operating hours is <40000 h.
• Checks
bullet listthat speed is 4750–5250 r/min.
• Checks
bullet listthat acceleration (rise time) from approx.
F

5000 to 40000 r/min is <0.150 s.


F

6-4 SERVO-air Ventilator System, Service Manual


| Troubleshooting | 6 |

6.3.6 Safety valve test sequence • Checks


bullet listthat measured inspiratory flow differs less
than 5% from the flow delivered from the O2 gas
Monitor safety valve test (MON) module at a flow of 0.5 l/s (30 l/min).
Checks the hardware signals related to the safety • Checks
bullet listthat the difference between MON and BRE
valve functions. subsystems is less than 0.005 l/s (<1%).

Breathing safety valve test (BRE) This test requires O2 gas supply.

Checks that the safety valves opening pressure is Breathing flow transducer test (BRE)
approx. 117 cmH2O and calibrates the valve if
required. The test can be repeated up to five times to Checks the offset values for:
achieve a Passed result:
• Inspiratory
bullet list flow (22.021–24.609 V).
• Acceptance
bullet list interval for the first test is • O2 flow
bullet list(0.000–0.081 V).
115.50–118.50 cmH2O.
• Expiratory
bullet list flow (0.093–0.142 V).
• Acceptance
bullet list interval for the following tests is
114.00–120.00 cmH2O. Flow test (MON)

• The test
bullet list fails if an opening pressure within Monitoring flow test in the S&S Test menu.
114.00–120.00 cmH2O is not reached after five
attempts. Checks the offset values for:

This test requires O2 gas supply. • O2 flow


bullet list(0.000–0.033 V).
• Expiratory
bullet list flow (0.024–0.072 V).
6.3.7 O2 cell test sequence

AA AAA AA
With 60 % O2 concentration and 1 l/s gas flow:
O2 cell test (MON)
• Checks
bullet listinspiratory flow (0.9798–1.0198 l/s).
Checks and calibrates the O2 cell:
• Checks
bullet listO2 flow (0.4842–0.5040 l/s).
FDTDTDTT DFDFTFTT
• Measures
list expiratory flow and calculates gain



Checks
bullet

bullet
Checks
bullet
R R
listif the O2 cell is consumed. Expected life
time 500 000 %hours.
Checks listwith 100 % O2 (>70.0 %).
listwith 21 % O2 (<25.0 %).
bullet
(k value) for the expiratory cassette (0.90–1.15).

With 100 % O2 concentration and 0.5 l/s gas flow:

• Checks
bullet listinspiratory flow BRE and MON
• Checks
bullet listoffset (4–24 mV).
(0.4750–0.5250 l/s).
• Calculates
bullet list gain.
• Checks
bullet listO2 flow BRE and MON (0.4900–0.5100 l/s).
A
This test requires O2 gas supply. • Checks
bullet listthat the difference between O2 flow BRE
and MON is less than 0.005 l/s (<1 %).
A

6.3.8 Flow transducer test sequence • Checks


bullet listexpiratory flow MON (±5 % of delivered
R
flow).
Insp flow sensor test (MON)
A


R
DRRR

Checks
bullet listthat the difference between expiratory flow
Sensirion flow sensor test in the S&S Test menu. BRE and MON is less than 0.005 l/s (<1 %).
R
Checks the inspiratory flowmeter:

• Checks
bullet listoffset (9920–10080).

SERVO-air Ventilator System, Service Manual 6-5


| 6 | Troubleshooting |

With 21 % O2 concentration and 0.5 l/s gas flow: 6.3.10 Patient circuit test sequence

• Checks
bullet listO2 flow BRE and MON (-0.0016–0.0016 l/s). Patient circuit test (BRE)

• Checks
bullet listinspiratory flow BRE and MON Checks the system with patient tubing connected:
(0.4900–0.5100 l/s).
• Checks
bullet listdifference between inspiratory flow BRE • Checks
bullet listthat system volume is 0.1500–6.0000 l at
and MON is less than 0.005 l/s (<1 %). 50 cmH2O (>40 cmH2O).

• Checks
bullet listexpiratory flow MON (±5 % of delivered • Checks
bullet listthat compressible volume is
flow). 0.02–6.00 ml/cmH2O.
• Checks
bullet listdifference between expiratory flow BRE • Checks
bullet listthat leakage is <0.0800 l/min at 50 cmH2O
and MON is less than 0.005 l/s (<1 %). (>40 cmH2O).

With 0 l/s gas flow:


• Checks
bullet listflow resistance at 10 l/min.
• Checks
bullet listflow resistance at 60 l/min.
• Checks
bullet listinspiratory flow BRE and MON
(-0.0016–0.0016 l/s). The system volume and compressible volume checks
can be repeated up to three times to achieve a stable

FTDFTDTDTDD FTFFTFTFFTTT
If O2 is not connected, the check is performed with
Air. The Flow test passes if the result of this check
corresponds to the old calibration factor from the
previous Pre-use check. The same expiratory cassette
must be used.

6.3.9 Battery switch test sequence

Battery switch test (MON)


A compressible volume value.

6.3.11 Alarm state test sequence

External alarm system test (MON)

If the option Alarm output is enabled, the user can test


the external alarm system.

For pin configuration and signal names in the Alarm


output connector N67, see Connectors in chapter
A
Checks the battery module(s):
Diagrams. For further information, refer to the 'Alarm
• Checks listthat a Getinge battery module is connected. output connector – Reference Manual'.
RRRARARAAA
bullet
A
• Checks
bullet listthe status, age and number of charge Alarm state test (MON)
DDRDDRDRDRRR

cycles.
A
• Checks
bullet listthat there is a battery module in slot 2 (main Checks that no Technical error alarms are active
slot). during the Pre-use check.
A
• Checks
bullet listthat the power supply switches to battery
when mains power is disconnected. 6.4 Other tests
• Checks
bullet listthat the power supply switches back to 6.4.1 Patient circuit test sequence
mains power when reconnected.
A

Patient circuit test (BRE)


This test is not performed if:
This test performs the Patient circuit test separately,
• Less than
bullet list 10 minutes backup time remains in the as described above.
connected battery modules.
• No battery
bullet list module is connected.

6-6 SERVO-air Ventilator System, Service Manual


| Troubleshooting | 6 |

6.5 Test logs


This section describes how to interpret the Test logs
related to the Pre-use check.

6.5.1 Interpreting the Test result logs

The Pre-use check, complete test as well as separate Some checks will be repeated if they fail. If they pass
tests, is logged in the Service log. Select Test results when repeated, the check and the complete test
and Test details to show all test result logs. A Service sequence will pass. This will be shown in the log.
key is required to access the Service log. The
description below shows how to interpret the When a check fails due to a user error; a Dialog Choice
Service log. (DIALOG) appears and gives the user the choice to
redo the check by choosing Redo or, to bypass the
If one or more checks within a test fail, the log will show check by choosing Cancel.
which check that failed and also include an error code.
These error codes are called Check Failure Identifier Always perform recommended actions in a
(CFI). All checks that fail during the Pre-use check chronological order. A test that fails can cause
have a unique CFI. Possible causes and recommended subsequent tests to fail.
actions for checks that fail permanently or intermittently
are listed in the following CFI tables.

Acronyms in test logs


Acronym Test log entry type Acronym Test log entry type
DIALOG Dialog choise SFIN Test sequence result
FIN Test result SSTA Test sequence start
INF Info STA Test start
MEA Measurement value

SERVO-air Ventilator System, Service Manual 6-7


| 6 | Troubleshooting |

Test log example 1

The log example below shows an Insp flow sensor


test. This test is included in the Flow transducer test
sequence:

1. The Insp flow sensor test starts.


2. The measured Insp flow sensor total flow (0.5279
l/s) is outside the tolerance interval.
3. The check of measured Insp flow sensor total flow
failed with CFI 4 in the log post.
4. The test sequence result is failed with CFI 4 in
the log post.
5. The complete Insp flow sensor test result is failed
with CFI 4 in the log post.
2017-10-20 19:37:38 [STA] Insp flow sensor test
2017-10-20 19:37:40 [CHK_SSTA] Insp. flow transducer test sequence
2017-10-20 19:37:40 [CHK_STA] Insp flow sensor test offset check
2017-10-20 19:37:40 [MEA] Insp flow sensor test offset steps = 10001 - Tolerance Interval = [ 9920; 10080 ] ( Unit : - )
2017-10-20 19:37:40 [CHK_FIN] PASSED
2017-10-20 19:37:40 [CHK_SFIN] PASSED
2017-10-20 19:37:43 [CHK_SSTA] Insp flow sensor test O2 measurement check sequence
2017-10-20 19:37:43 [CHK_STA] Insp flow sensor test Insp total flow check
2017-10-20 19:37:43 [MEA] Insp flow sensor test total flow = 0.5279 - Tolerance Interval = [ 0.4736; 0.5235 ] ( Unit : l/s )
2017-10-20 19:37:43 [CHK_FIN] FAILED : Insp flow sensor test, sensirion measurement failure [Error Code = 4]
2017-10-20 19:37:43 [CHK_STA] Insp flow sensor test Bre - Mon Insp total flow check
2017-10-20 19:37:43 [MEA] Insp flow sensor test Bre - Mon Insp total flow: Mon flow = 0.5279 - set flow = 0.5000 - Tolerance interval =
[ 0.5223; 0.5323 ] ( Unit : l/s )
2017-10-20 19:37:43 [CHK_FIN] PASSED
2017-10-20 19:37:43 [CHK_SFIN] FAILED - [Error Code = 4]
2017-10-20 19:37:43 [FIN] Insp flow sensor test FAILED - [Error Code = 4] S20138

6-8 SERVO-air Ventilator System, Service Manual


| Troubleshooting | 6 |

Test log example 2

The log example below shows the Turbine and Gas


supply test sequence:

1. The Turbine and Gas supply test sequence starts.


2. The Gas supply test starts.
3. The measured O2 supply pressure (2 mbar) is
outside the tolerance interval.
4. The O2 supply pressure is too low. The O2 supply
pressure test failed with CFI 2 in the log post
5. The Gas supply test failed with CFI 2 in the log
post.
6. The different turbine test passed, but the complete
Turbine and Gas supply test sequence failed due
to the low O2 pressure in the Gas supply test.

2018-01-19 10:09:38 [SSTA] Turbine and Gas supply test sequence


2018-01-19 10:09:40 [STA] Gas supply test
2018-01-19 10:09:42 [CHK_STA] Checking O2 supply pressure.
2018-01-19 10:09:42 [MEA] O2 supply pressure = 2 - Tolerance Interval = [ 2000; 6000 ] ( Unit : mbar )
2018-01-19 10:09:42 [CHK_FIN] FAILED : O2 supply pressure too low. [Error Code = 2]
2018-01-19 10:09:42 [CHK_STA] Checking that O2 supply pressure (MAX) - O2 supply pressure (MIN) < Max tolerated difference.
2018-01-19 10:09:42 [MEA] O2 supply pressure (MAX) = 4 - Tolerance Interval ( -,O2 supply pressure (MIN) + 100 ) = [ -; 102 ] ( Unit : mbar )
2018-01-19 10:09:42 [CHK_FIN] PASSED
2018-01-19 10:09:42 [CHK_STA] Checking that |O2 supply pressure (MON) - O2 supply pressure (BRE)| < Tolerance.
2018-01-19 10:09:42 [MEA] O2 supply pressure (BRE) = 5 - Tolerance Interval (O2 supply pressure (MON) +/- 20) = [ -18; 24 ] ( Unit : mbar )
2018-01-19 10:09:42 [CHK_FIN] PASSED
2018-01-19 10:09:56 [ADD_INFO] Gas type: air.
2018-01-19 10:09:56 [FIN] Gas supply test FAILED - [Error Code = 2]
2018-01-19 10:09:56 [STA] Turbine accuracy test
2018-01-19 10:09:56 [CHK_STA] Checking turbine age
2018-01-19 10:09:56 [MEA] Turbine operating hours = 932 - Limit = [ -; 40000 ] ( Unit : hours )
2018-01-19 10:09:56 [CHK_FIN] PASSED
2018-01-19 10:09:58 [CHK_STA] Checking turbine rpm generation
2018-01-19 10:09:58 [MEA] Measured turbine rpm = 4996 - Tolerance Interval = [ 4750.00; 5250.00 ] ( Unit : rpm )
2018-01-19 10:09:58 [CHK_FIN] PASSED
2018-01-19 10:09:58 [CHK_STA] Checking turbine rise fall time
2018-01-19 10:09:58 [MEA] Measured rise time = 0.110 - Limits = [ -; 0.150 ] ( Unit : seconds )
2018-01-19 10:09:58 [CHK_FIN] PASSED
2018-01-19 10:09:59 [CHK_STA] Checking turbine rpm generation
2018-01-19 10:09:59 [MEA] Measured turbine rpm = 40043 - Tolerance Interval = [ 38000.00; 42000.00 ] ( Unit : rpm )
2018-01-19 10:09:59 [CHK_FIN] PASSED
2018-01-19 10:09:59 [CHK_STA] Checking turbine rise fall time
2018-01-19 10:09:59 [MEA] Measured fall time = 0.110 - Limits = [ -; 0.150 ] ( Unit : seconds )
2018-01-19 10:09:59 [CHK_FIN] PASSED
2018-01-19 10:09:59 [FIN] Turbine accuracy test PASSED
2018-01-19 10:10:00 [SFIN] Turbine and Gas supply test sequence FAILED S20137

SERVO-air Ventilator System, Service Manual 6-9


| 6 | Troubleshooting |

6.5.2 Pre-use check CFI

Internal test sequence

CFI Log description Recommended action


Panel audio test (PAN)
- If the test fails, the audio does not work. 1. Replace Loudspeaker.
2. Replace PC 2023 Panel.

Monitor audio test (MON)


1 The buzzer is permanently on. 1. Replace PC 1992 Monitoring.

2 The buzzer could not be turned on. 1. Replace PC 1992 Monitoring.

3 The buzzer could not be turned off. 1. Replace PC 1992 Monitoring.

Temperature and Moist sensor test


1 Humidity is not within accepted limits. 1. Replace PC 2030 Inspiratory channel.

Comment: 0.0–100.0 %.

2 Turbine temperature is not within accepted limits. 1. Check Air inlet filter.
2. Replace Turbine module.
Comment: 0.0–70.0 °C.

3 Ambient temperature is not within accepted 1. Replace PC 2030 Inspiratory channel.


limits.

Comment: 0.0–55.0 °C.

4 Insp gas temperature is not within accepted 1. Replace Inspiratory flowmeter.


limits.

Comment: 0.0–60.0 °C

Blower inlet test


1 Blower inlet resistance is too low. 1. Check that the Air inlet filter is connected.
2. Replace Turbine module.
Comment: <1.00 cmH2O/l/s.

2 Blower inlet resistance is too high. 1. Check Air inlet filter.


2. Replace Dust filter or complete Air inlet filter.
Comment: >3.30 cmH2O/l/s.
3. Replace Turbine module.

3 Pressure diff sensor test: pressure transducer 1. Replace Turbine module.


offset not within accepted limits.

Comment: 1586–1690.

Alarm output connector test (MON)


1 Remote Alarm is already active. 1. Replace PC 2021 Rear connectors.
2. Replace PC 1992 Monitoring.

2 Remote Alarm has not been activated. 1. Replace PC 2021 Rear connectors.
2. Replace PC 1992 Monitoring.

6 - 10 SERVO-air Ventilator System, Service Manual


| Troubleshooting | 6 |

CFI Log description Recommended action


3 Remote Alarm has not been deactivated. 1. Replace PC 2021 Rear connectors.
2. Replace PC 1992 Monitoring.

Power failure test (MON)


1 PFA alarm is inactive. 1. Replace PC 1992 Monitoring.

Comment: Check #1 for 5 V.

2 PFA alarm is inactive. 1. Replace PC 1992 Monitoring.

Comment: Check #2 for 5 V.

3 PFA alarm is inactive. 1. Replace PC 1992 Monitoring.

Comment: Check #1 for 3.3 V.

4 PFA alarm is inactive. 1. Replace PC 1992 Monitoring.

Comment: Check #2 for 3.3 V.

5 The buzzer has not been turned on. 1. Replace PC 1992 Monitoring.

6 The buzzer has not been turned off. 1. Replace PC 1992 Monitoring.

7 The remote alarm is active. 1. Replace PC 1992 Monitoring.

Comment: Failed to inactivate the remote alarm.

8 The remote alarm is active. 1. Replace PC 1992 Monitoring.

Comment: Failed to activate the remote alarm.

9 PFA alarm is active. 1. Replace PC 1992 Monitoring.

Barometer test sequence

CFI Log description Recommended action


Barometer test (MON)
1 Barometric Pressure (MON) is not within the 1. Calibrate the barometer (S&S).
accepted limits. 2. Replace PC 1992 Monitoring.
Comment: >1070 hPa (mbar). 3. Replace PC 2004/PC 2024 Expiratory channel.

2 Barometric Pressure (MON) is not within the 1. Calibrate the barometer (S&S).
accepted limits. 2. Replace PC 1992 Monitoring.
Comment: <650 hPa (mbar). 3. Replace PC 2004/PC 2024 Expiratory channel.

3 The difference between the values measured in 1. Replace PC 1991 Control.


BRE and MON is too high. 2. Replace PC 1992 Monitoring.
Comment: >8 hPa (mbar). 3. Replace PC 2004/PC 2024 Expiratory channel.
4. Replace O2 gas module.

SERVO-air Ventilator System, Service Manual 6 - 11


| 6 | Troubleshooting |

Internal leakage test sequence

CFI Log description Recommended action


Internal leakage test (BRE)
1 The difference between InspPress and ExpPress 1. Check that the pressure transducer tubes and the inspiratory
is too high. filter are correctly mounted.
2. Check that both PC 1781 Pressure transducer (Insp. and
Comment: >10 cmH2O Exp.) are correctly mounted.
3. If the ‘Pressure transducer test’ also fails, refer to the
recommended actions if ‘Pressure transducer test’ failed.

2 The System Volume is too low. 1. Replace Inspiratory flowmeter.

Comment: <0.2000 l

3 The System Volume is too high. 1. Check that the correct test tube is used during the Pre-use
check and check that the test tube is correctly connected.
Comment: >0.4000 l 2. If the ‘Flow transducer test’ also fails, replace Inspiratory
flowmeter.
3. Check that the expiratory cassette is correctly seated in the
cassette compartment.
4. If possible, replace the expiratory cassette and check if the
new cassette is accepted by the Pre-use check. If the new
cassette was accepted by the Pre-use check, the fault was
located to the cassette.
5. Check that the pressure transducer tubes/filters are correctly
mounted.
6. Check that the inspiratory gas docking, inspiratory flowmeter
and inspiratory channel are correctly mounted.
7. Check that the safety valve membrane is clean and correctly
seated in the inspiratory pipe.
8. Check that the safety valve closes properly when the Pre-use
check starts (distinct clicking sound from the valve). If the
safety valve not closes properly during this check:
- Replace safety valve pull magnet.
- Replace PC 2004/PC 2024 Expiratory channel.

6 - 12 SERVO-air Ventilator System, Service Manual


| Troubleshooting | 6 |

CFI Log description Recommended action


4 Inspiratory Pressure is too low. 1. The Leakage detection, described in chapter Service
procedures, can be used to locate the leakage. The Leakage
Comment: Excessive leakage. Required detection is a complement to the recommended actions
pressure 50 cmH2O not obtained, leakage test below.
cannot be performed. 1. Check that the connected gas supply pressure is within
the specified range.
2. Check that the test tube is correctly connected.
3. Check that the expiratory cassette is correctly seated in
the cassette compartment.
4. If possible, replace the expiratory cassette and check if
the new cassette is accepted by the Pre-use check. If
the new cassette was accepted by the Pre-use check,
the fault was located to the cassette.
5. Check that the pressure transducer tubes/filters are
correctly mounted.
6. Check that the inspiratory pipe is correctly mounted in
the inspiratory section.
7. Check that the safety valve membrane is clean and
correctly seated in the inspiratory pipe.
8. Check that the safety valve closes properly when the
Pre-use check starts (distinct clicking sound from the
valve). If the safety valve not closes properly during this
check:
- Replace safety valve pull magnet.
- Replace PC 2004/PC 2024 Expiratory channel.

SERVO-air Ventilator System, Service Manual 6 - 13


| 6 | Troubleshooting |

CFI Log description Recommended action


5 The Leakage Value is too high, or the safety 1. Check that the correct test tube is used during the Pre-use
valve is either not locked or its status could not check and check that the test tube is correctly connected.
be read. 2. Check that the expiratory cassette is correctly seated in the
cassette compartment.
Comment: >0.0200 l/min
3. If possible, replace the expiratory cassette and check if the
new cassette is accepted by the Pre-use check. If the new
cassette was accepted by the Pre-use check, the fault was
located to the cassette.
4. Check that the pressure transducer tubes/filters are correctly
mounted.
5. Check that the inspiratory gas docking, inspiratory flowmeter
and inspiratory channel are correctly mounted.
6. Check that the safety valve membrane is clean and correctly
seated in the inspiratory pipe.
7. Check that the safety valve closes properly when the Pre-use
check starts (distinct clicking sound from the valve). If the
safety valve opens during this test, the opening pressure
may not be correctly calibrated (see 'Safety valve test' below).
Run the 'Safety valve test' and repeat the complete Pre-use
check.
8. The Leakage detection, described in chapter Service
procedures, can be used to locate the leakage. The Leakage
detection is a complement to the recommended actions
below.
1. Check that the test tube is correctly connected.
2. Check that the expiratory cassette is correctly seated in
the cassette compartment.
3. If possible, replace the expiratory cassette and check if
the new cassette is accepted by the Pre-use check. If
the new cassette was accepted by the Pre-use check,
the fault was located to the cassette.
4. Check that the pressure transducer tubes/filters are
correctly mounted.
5. Check that the inspiratory pipe is correctly mounted in
the inspiratory section.
6. Check that the safety valve membrane is clean and
correctly seated in the inspiratory pipe.
7. Check that the safety valve closes properly when the
Pre-use check starts (distinct clicking sound from the
valve). If the safety valve not closes properly during this
check:
- Replace safety valve pull magnet.
- Replace PC 2004/PC 2024 Expiratory channel.

Turbine and Gas supply test sequence

CFI Log description Recommended action


Gas supply test (MON)
1 O2 supply pressure too high. 1. Check that the connected O2 supply pressure is within the
specified range.
Comment: >600 kPa (6000 mbar). 2. Replace O2 gas module.

6 - 14 SERVO-air Ventilator System, Service Manual


| Troubleshooting | 6 |

CFI Log description Recommended action


2 O2 supply pressure too low. 1. Check that the connected O2 supply pressure is within the
specified range.
Comment: <200 kPa (2000 mbar). 2. Replace O2 gas module.

3 The difference between O2 supply pressure 1. Check that the connected O2 supply pressure is stable.
(MAX) and O2 supply pressure (MIN) is too 2. Replace O2 gas module.
large.

Comment: O2 supply pressure variation too


large, >100 hPa (mbar).

4 The difference between O2 supply pressure 1. Replace PC 1991 Control.


(MON) and O2 supply pressure (BRE) is too 2. Replace PC 1992 Monitoring.
large.

Comment: >20 hPa (mbar).

11 Breathing disconnected. 1. Restart the system.


2. Replace PC 1991 Control.

Turbine accuracy test (MON)


1 Turbine age too old. 1. Replace Turbine module.

Comment: Max. 40000 h.

2 Turbine rpm accuracy error. 1. Replace Turbine module.

Comment: 4750–5250 r/min.

3 Turbine acceleration accuracy error. 1. Replace Turbine module.

Comment: Acceleration (rise time) from approx.


5000 to 40000 r/min - max. 0.150 s.

4 Turbine deceleration accuracy error. 1. Replace Turbine module.

Comment: Deceleration (fall time) from approx.


40000 to 5000 r/min - max. 0.150 s.

Pressure transducer test sequence

CFI Log description Recommended action


Breathing pressure transducer test (BRE)
1 Inspiration Offset could not be stored persistently 1. Replace PC 1781 Inspiratory pressure transducer.
- Value out of range. 2. Replace PC 1991 Control.
Comment: Differs more than ±6 cmH2O
(±0.300 V) from factory default.

2 Expiration Offset could not be stored persistently 1. Replace PC 1781 Expiratory pressure transducer.
- Value out of range. 2. Replace PC 1991 Control.
Comment: Differs more than ±6 cmH2O
(±0.300 V) from factory default.
SERVO-air Ventilator System, Service Manual 6 - 15
| 6 | Troubleshooting |

CFI Log description Recommended action


Monitor pressure transducer test (MON)
1 Inspiration offset >6 cmH2O from factory default. 1. Replace PC 1781 Inspiratory pressure transducer.
2. Replace PC 1992 Monitoring.
Comment: Differs more than ±6 cmH2O
(±0.300 V) from factory default.

2 Expiration offset >6 cmH2O from factory default. 1. Replace PC 1781 Expiratory pressure transducer.
2. Replace PC 1992 Monitoring.
Comment: Differs more than ±6 cmH2O
(±0.300 V) from factory default.

3 Expiration gain >8% from factory default. 1. Cancel the ‘Pressure transducer test’ and continue with the
remaining tests (this will calibrate the safety valve). Restart
the Pre-use check.
2. If possible, replace the expiratory cassette and check if the
new cassette is accepted by the Pre-use check. If the new
cassette was accepted by the Pre-use check, the fault was
located to the cassette.
3. Check/replace PC 1781 Pressure transducer (Insp. and Exp.).
To locate the faulty pressure transducer, replace one
transducer at a time.

4 Expiration offset is different at zero pressure and 1. Replace PC 1781 Expiratory pressure transducer.
high pressure.
5 The difference between the values measured in 1. Replace PC 1991 Control.
BRE and MON is too high. 2. Replace PC 1992 Monitoring.
Comment: Differs more than ±0.80 cmH2O.

6 Inspiration pressure transducer could not be 1. Replace PC 1781 Inspiratory pressure transducer.
locked or is faulty.
7 Expiration pressure transducer could not be 1. Replace PC 1781 Expiratory pressure transducer.
locked or is faulty.
8 Inspiratory pressure value out of bounds. 1. Replace Turbine module.

Comment: 45.80–54.20 cmH2O.

Expiratory valve test (BRE)


1 Expiratory Pressure too high. 1. If possible, replace the expiratory cassette and check if the
new cassette is accepted by the Pre-use check. If the new
Comment: Measures at 40, 20, 10 and 5 cmH2O. cassette was accepted by the Pre-use check, the fault was
Measured values must be within ±15 cmH2O. located to the cassette.
2. Replace Expiratory valve coil.

2 Expiratory Pressure too low. 1. Clean or replace Turbine module one-way valve.
2. If possible, replace the expiratory cassette and check if the
Comment: Measures at 40, 20, 10 and 5 cmH2O. new cassette is accepted by the Pre-use check. If the new
Measured values must be within ±15 cmH2O. cassette was accepted by the Pre-use check, the fault was
located to the cassette.
3. Replace Expiratory valve coil.

3 Voice Coil too strong. 1. If possible, replace the expiratory cassette and check if the
new cassette is accepted by the Pre-use check. If the new
Comment: Expiratory valve coil not within cassette was accepted by the Pre-use check, the fault was
specified limits; offset 0.06768–0.15684 A, gain located to the cassette membrane.
0.00730–0.01130 A/cmH2O. 2. Replace Expiratory valve coil.

6 - 16 SERVO-air Ventilator System, Service Manual


| Troubleshooting | 6 |

CFI Log description Recommended action


4 The expiratory valve constants were not stored 1. Replace PC 2004/PC 2024 Expiratory channel.
persistently.

SERVO-air Ventilator System, Service Manual 6 - 17


| 6 | Troubleshooting |

Safety valve test sequence

CFI Log description Recommended action


Monitor safety valve test (MON)
1 The Safety Valve could not be opened. 1. Replace PC 2004/PC 2024 Expiratory channel.
2. Replace PC 1992 Monitoring.

2 The Safety Valve could not be closed. 1. Replace Safety valve pull magnet.
2. Replace PC 2004/PC 2024 Expiratory channel.
3. Replace PC 1992 Monitoring.

3 The Safety Valve is open. 1. Replace Safety valve pull magnet.


2. Replace PC 2004/PC 2024 Expiratory channel.
3. Replace PC 1992 Monitoring.

4 Disable valves activation failed. 1. Replace PC 1992 Monitoring.

5 Disable valves deactivation failed. 1. Replace PC 1992 Monitoring.

Breathing safety valve test (BRE)


1 Pressure superior to allowed max and valve not 1. Replace Safety valve pull magnet.
open. 2. Replace PC 2004/PC 2024 Expiratory channel.
Comment: >120 cmH2O.

2 Pressure not rising. Timeout. 1. Check that the safety valve membrane is clean and correctly
seated in the inspiratory pipe.
Comment: Not possible to reach opening 2. Check that the inspiratory gas docking, inspiratory flowmeter,
pressure 117 cmH2O. inspiratory channel and turbine module one-way valve are
correctly mounted.

3 Calibration Impossible. 1. Replace Safety valve pull magnet.


2. Replace PC 2004/PC 2024 Expiratory channel.

4 The Safety Valve is not closed. 1. Replace Safety valve pull magnet.
2. Replace PC 2004/PC 2024 Expiratory channel.
3. Replace PC 1992 Monitoring.

5 The Safety Valve could not be disabled. 1. Replace PC 1991 Control.

6 The Safety Valve could not be enabled. 1. Replace PC 1991 Control.

O2 cell test sequence

CFI Log description Recommended action


O2 cell test (MON)
1 No O2 Transducer detected. 1. Check that the O2 cell is connected.
2. Replace O2 cell.
3. Replace O2 cell cable.
4. Replace PC 1992 Monitoring.

3 O2 Concentration in AIR Supply too low. 1. Replace O2 cell.

Comment: >25 %.
Error in the log text. Correct text is:
O2 Concentration in AIR Supply too high.
6 - 18 SERVO-air Ventilator System, Service Manual
| Troubleshooting | 6 |

CFI Log description Recommended action


4 O2 Concentration in O2 Supply too low. 1. Check that the connected O2 supply concentration is within
the specified range.
Comment: <70.0 % 2. Replace O2 cell.

6 O2 supply low. 1. Check that the connected O2 supply pressure is within the
specified range.
Comment: <200 kPa 2. Replace O2 gas module.

7 Gas supply not approved. 1. Gas supply test not passed. Redo the Gas supply test.

8 Invalid EEPROM data. 1. Replace O2 cell.

9 Lock failed. 1. Replace O2 cell.


2
Comment: Lock failed, I C problem

11 O2 Cell Calibration impossible. 1. Replace O2 cell.

12 The O2 Cell needs to be replaced. 1. Replace O2 cell.

Flow transducer test sequence

CFI Log description Recommended action


Insp flow sensor test (MON) (Sensirion flow sensor test)
1 Insp flow sensor test, measured O2 flow failed. 1. Replace O2 gas module.
2. Replace Inspiratory flowmeter.

2 Insp flow sensor test, Bre - Mon comparison 1. Replace PC 1991 Control.
failed. 2. Replace PC 1992 Monitoring.

3 Sensirion flow transducer offset not within 1. Replace Inspiratory flowmeter.


accepted limits. 2. Replace PC 1992 Monitoring.

4 Insp flow sensor test, sensirion measurement 1. Replace Inspiratory flowmeter.


failure. 2. Replace O2 gas module.

5 Insp flow sensor test, Bre - Mon Sensirion 1. Replace PC 1991 Control.
comparison failure. 2. Replace PC 1992 Monitoring.

6 No gas supply. 1. Check O2 supply.


2. Replace O2 gas module.

SERVO-air Ventilator System, Service Manual 6 - 19


| 6 | Troubleshooting |

CFI Log description Recommended action


Breathing flow transducer test (BRE)
1 Inspiration Air Offset could not be stored 1. Replace Inspiratory flowmeter.
persistently - Value out of range.

Comment: 22.021-24.609 V

2 Inspiration O2 Offset could not be stored 1. Replace O2 gas module.


persistently - Value out of range.

Comment: 0.000-0.081 V

3 Expiration Offset could not be stored persistently 1. Check that the expiratory cassette is correctly seated in the
- Value out of range. cassette compartment.
2. Replace the expiratory cassette.
Comment: 0.093-0.142 V
3. Replace PC 2004/PC 2024 Expiratory channel.

Flow test (MON)


2 The Inspiration O2 Offset is not within the 1. Replace O2 gas module.
accepted limits.

Comment: 0.000-0.033 V

3 The Expiration Offset is not within the accepted 1. Check that the expiratory cassette is correctly seated in the
limits. cassette compartment.
2. Replace the expiratory cassette.
Comment: 0.024-0.072 V
3. Replace PC 2004/PC 2024 Expiratory channel.

4 Expiration Gain is not within the specs. 1. Check that the connected O2 supply pressure is within the
specified range.
Comment: 0.90-1.15 (k value) 2. Check that the cassette is correctly seated in the cassette
compartment. If possible, replace the expiratory cassette and
check if the new cassette is accepted by the Pre-use check.
If the new cassette was accepted by the Pre-use check, the
fault was located to the cassette. The fail with the old cassette
may in this case be due to water collected inside the cassette.
3. Replace Inspiratory flowmeter.
4. Replace O2 gas module.
5. Replace PC 1785/PC 2031 Expiratory channel connector.
6. Replace PC 1991 Control.
7. Replace PC 1992 Monitoring.
8. Replace PC 2004/PC 2024 Expiratory channel.

6 Timeout getting Expiration Gain (k value) 1. Check that the expiratory cassette is correctly seated in the
cassette compartment.
2. Replace the expiratory cassette.

6 - 20 SERVO-air Ventilator System, Service Manual


| Troubleshooting | 6 |

Battery switch test sequence

CFI Log description Recommended action


Battery switch test (MON)
1 Mains is not on. 1. Connect mains power.

2 The batteries have not been switched on 1. Mains power not disconnected as demanded.
(TIMEOUT).
3 The batteries have been switched back off 1. Mains power not reconnected as demanded.
without mains being back on first. (TIMEOUT).
4 No battery operation detected by BRE. 1. Replace PC 1991 Control.

5 Remaining battery power is insufficient for the 1. Charge or replace Battery modules.
test to be performed.

Comment: <10 min.

6 No batteries attached. 1. Connect Battery module in slot 2 (main slot).


2. Replace PC 2028 DC/DC and Battery control.

7 The batteries have been switched on without 1. Replace PC 2028 DC/DC and Battery control.
mains loss first. (TIMEOUT). 2. Replace PC 1992 Monitoring.

8 The batteries have been switched back off 1. Replace PC 2028 DC/DC and Battery control.
without mains being back on first. (TIMEOUT). 2. Replace PC 1992 Monitoring.

9 Unknown battery detected. 1. Connect a Getinge battery module.

10 Power system external power source test failed. 1. Connect mains power.
(POW)

11 Power system battery switch test failed. 1. Connect Battery module in slot 2 (main slot).
(POW) 2. Replace PC 2028 DC/DC and Battery control.

12 Power system no external power source test 1. Replace PC 2028 DC/DC and Battery control.
failed.
(POW)

13 Bad battery detected. 1. Replace Battery module.


(POW)
14 No battery in main slot. 1. Connect Battery module in slot 2 (main slot).
(POW)
15 Battery has expired. 1. Replace Battery module.
(POW)

Comment: >300 charge cycles/>48 months from


manufacturing date.

16 Low capacity (FCC) battery detected. 1. Replace Battery module.


(POW)

SERVO-air Ventilator System, Service Manual 6 - 21


| 6 | Troubleshooting |

Patient circuit test sequence

CFI Log description Recommended action


Patient circuit test (BRE)
1 InspPress is too low. 1. If the internal leakage test has passed, the leakage is to be
located to the patient circuit. Check for leakage or replace
Comment: <40 cmH2O the patient circuit.

2 The Compressible Volume is too low. 1. Check or replace the patient circuit.

Comment: <0.02 ml/cmH2O.

3 The Compressible Volume is too high. 1. Check or replace the patient circuit.

Comment: >6.00 ml/cmH2O.

4 Leakage too high, or the safety valve is either 1. If the internal leakage test has passed, the leakage is to be
not locked or its status could not be read. located to the patient circuit. Check for leakage or replace
the patient circuit.
Comment: >0.0800 l/min

5 System Volume is too low. 1. Check that the internal leakage test passed.

Comment: <0.1500 l

6 System Volume is too high. 1. Check that the internal leakage test passed.
2. Improper patient circuit connected.
Comment: >6.0000 l

Alarm state test sequence

CFI Log description Recommended action


External alarm system test (MON) If this option is installed.
1 Remote External Alarm is not connected. 1. Check the external alarm system.
2. Replace PC 2021 Rear connectors.

2 Remote External Alarm is already active. 1. Check the external alarm system.
2. Replace PC 2021 Rear connectors.

Alarm state test (MON)


- cf alarms error codes Check Technical error alarms logs for further information.

6 - 22 SERVO-air Ventilator System, Service Manual


| Troubleshooting | 6 |

6.6 Technical error codes and messages


The table below shows recommended actions in case
of Technical errors.

6.6.1 Monitoring

Error
Error message / Possible cause Recommended action
code
1 Power error. 1. Replace PC 2028 DC/DC & Battery control.

-12 V too low, i.e. < -13.2 V.

A Power system restart might be necessary to reset power


related alarms. To restart, switch off the system and disconnect
both the mains power supply and the External 12 V battery.

2 Power error. 1. Check status of External 12 V battery (if


connected).
-12 V too high, i.e. > -10.8 V. 2. Replace PC 2028 DC/DC & Battery control.
A Power system restart might be necessary to reset power
related alarms. To restart, switch off the system and disconnect
both the mains power supply and the External 12 V battery.

3 Power error. 1. Check status of External 12 V battery (if


connected).
+12 V too low, i.e. < +10.8 V. 2. Replace PC 2028 DC/DC & Battery control.
A Power system restart might be necessary to reset power
related alarms. To restart, switch off the system and disconnect
both the mains power supply and the External 12 V battery.

4 Power error. 1. Replace PC 2028 DC/DC & Battery control.

+12 V too high, i.e. > +13.2 V.

A Power system restart might be necessary to reset power


related alarms. To restart, switch off the system and disconnect
both the mains power supply and the External 12 V battery.

5 Power error. 1. Replace PC 2028 DC/DC & Battery control.


2. Replace the O2 gas module.
+24 V too low, i.e. < 21.5 V.

A Power system restart might be necessary to reset power


related alarms. To restart, switch off the system and disconnect
both the mains power supply and the External 12 V battery.

6 Power error. 1. Replace PC 2028 DC/DC & Battery control.

+24 V too high, i.e. > 26.5 V.

A Power system restart might be necessary to reset power


related alarms. To restart, switch off the system and disconnect
both the mains power supply and the External 12 V battery.

SERVO-air Ventilator System, Service Manual 6 - 23


| 6 | Troubleshooting |

Error
Error message / Possible cause Recommended action
code
7 Software error. 1. Replace PC 1991 Control.
2. Replace PC 1992 Monitoring.
Patient category mismatch between Breathing and Monitoring
subsystems. 3. Replace the O2 gas module.

8 Insp. pause hold time exceeded. 1. Replace PC 1991 Control.

The max. Insp. pause hold time 30 s exceeded. The system


has not returned to ventilation after Insp. pause hold time.

9 Exp. pause hold time exceeded. 1. Replace PC 1991 Control.

The max. Exp. pause hold time 30 s exceeded. The system


has not returned to ventilation after Exp. pause hold time.

10 Ventilation stopped. 1. Replace PC 2004/PC 2024 Expiratory


channel.
The VALVES_DISABLED signal has stopped power supply to 2. Replace PC 1991 Control.
O2 gas module, Turbine module and Safety valve.
3. Replace PC 2028 DC/DC & Battery control.
4. Replace PC 1992 Monitoring.

11 Safety valve open. 1. Replace safety valve pull magnet.


2. Replace PC 2004/PC 2024 Expiratory
The Safety valve is detected as open without fulfilled opening channel.
conditions.

16 Ventilation error. 1. Replace PC 2004/PC 2024 Expiratory


channel.
PC 2004/PC 2024 Expiratory channel failure.

17 Ventilation stopped. 1. Report to MCC HSC or to your local


Getinge representative for further
Software conflict. Two PC 2004/PC 2024 Expiratory channel information.
detected on the CAN bus.

22 Backup audible alarm error. 1. Replace PC 1992 Monitoring.

Backup alarm buzzer error.

24 Backup audible alarm or Remote alarm error. 1. Replace PC 1992 Monitoring.

Error indicated in the backup alarm or remote alarm capacitor


on PC 1992 Monitoring.

6 - 24 SERVO-air Ventilator System, Service Manual


| Troubleshooting | 6 |

Error
Error message / Possible cause Recommended action
code
25 Communication error. Depending on ID # and Error # stated in the
Technical alarm log.
Communication error between ID PROMs/EEPOTs and
Monitoring subsystem. If repeated, replace the spare part indicated by
the ID # according to the list below:
The ID # is followed by an Error # in the Technical alarm log.
ID #
For Error 6, report to MCC HSC or to your local Getinge
representative for further information. 0: PC 2020 Main back-plane.

For all other Error #, replace the part identified by the ID #. 1: PC 1992 Monitoring.

Error # 2: PC 2028 DC/DC & Battery control.

1. CHK_SUM_ERR 3: PC 2021 Rear connectors.


2. NR_OF_BYTES_IN_CHKSUM_ERR
4: PC 1781 Inspiratory pressure transducer.
3. CHKSUM_FILL_BYTE_ERR
4. WRITABLE_RECORDS_FAILURE 5: PC 1781 Expiratory pressure transducer.
5. WRITABLE_RECORDS_DIFFER
6: O2 cell or O2 cell cable.
6. COMM_ERR
7. VALIDATION_ERR 32: PC 2004/PC 2024 Expiratory channel.
8. WCR_VALIDATION_ERR
9. WRITE_NOT_PERMITTED 33: Turbine module.
10. OTHER_ERR 34: O2 gas module.

35: PC 1992 Monitoring.

256: PC 1991 Control.

512: PC 2023 Panel.

1024: PC 2004/PC 2024 Expiratory channel.

1025: Expiratory cassette.

1026: PC 1993 Ethernet switch.

Note: After replacement of PC 1992, TE 25


may occur. Adjust the barometric pressure in
Service & Settings to solve this error (Service
key required).

27 Backup audible alarm error. 1. Replace PC 1992 Monitoring.

Backup alarm buzzer error.

28 Audible alarm error. 1. Replace loudspeaker.


2. Replace PC 2023 Panel.
Loudspeaker error.

29 Memory backup battery depleted. 1. Replace Memory backup battery on


PC 1992 Monitoring.
The Memory backup battery on PC 1992 Monitoring is depleted.

SERVO-air Ventilator System, Service Manual 6 - 25


| 6 | Troubleshooting |

Error
Error message / Possible cause Recommended action
code
32 Software version conflict. 1. Reinstall the System software.
2. Report to MCC HSC or to your local
Conflict between the software versions installed. Getinge representative for further
information.

35 Ventilation disabled. 1. Replace PC 2004/PC 2024 Expiratory


channel.
Communication error or PC 2004/PC 2024 Expiratory channel
failure during startup.

37 Software error. 1. Replace PC 2004/PC 2024 Expiratory


channel.
PC 2004/PC 2024 Expiratory channel indicates wrong patient 2. Report to MCC HSC or to your local
category. Getinge representative for further
information.

38 Barometric pressure too high. 1. Calibrate barometer (S&S required).


2. Replace PC 1992 Monitoring.
Barometric pressure above allowed limit or barometer error.

39 Barometric pressure too low. 1. Calibrate barometer (S&S required).


2. Replace PC 1992 Monitoring.
Barometric pressure below allowed limit or barometer error.
Note: This alarm will be activated if the
ambient pressure is below 650 hPa, e.g. on
a high altitude.

40 Measured value invalid. 1. Restart the system.

Invalid measured metric detected. The invalid metric is not


displayed.

41 Real time clock error. 1. Replace PC 1992 Monitoring.

Internal processor real time clock error.

42 Checksum error. 1. Restart the system.


2. Replace Memory backup battery on
Software checksum error during startup. PC 1992 if together with Error code 29.
3. Replace PC 1992 Monitoring.

43 Battery communication error. 1. Update to System SW version 2.1.2 or


higher.
Battery module or PC 2028/PC 1992 communication error. 2. Replace the Battery module(s).
3. Replace PC 1992 Monitoring.
4. Replace PC 2028 DC/DC & Battery control.

46 Alarm output connector error. 1. Replace PC 2021 Rear connectors.


2. Replace PC 1992 Monitoring.
Alarm output connector error or communication error with
PC 1992.

6 - 26 SERVO-air Ventilator System, Service Manual


| Troubleshooting | 6 |

Error
Error message / Possible cause Recommended action
code
48 Pre-oxygen time exceeded. 1. Replace PC 1991 Control.

The max. Pre-oxygenation time during


DISCONNECTION/SUCTION exceeded. The system has not
returned to ventilation using the previous settings after the
Preoxygenation time.

50 Communication error. 1. Replace PC 1992 Monitoring.


2. Replace PC 2020 Main back-plane.
Communication error between ID PROM on PC 2020 Main
back-plane and Monitoring subsystem. System ID,
configuration, operating time, etc., not available.

51 On/Off switch error. 1. Replace On/Off switch including cable.

The On/Off switch is neither in On or Off position during more


than 3 seconds.

55 Communication error. Check for additional information in the Technical


log.
This error code is an indication of one or more of:
With log text
• Internallist
bullet Ethernet communication failure "CommunicationBrokenCondition:NodeStatusBreathing":
1. Replace PC 1991 Control.
• Batterylist
bullet status erroneously repeated
• Subsystem
bullet list mismatch of patient category, ventilation mode, With log text
and/or ventilation setting(s)
"CommunicationBrokenCondition:NodeStatusPanel":
• Certainlist
bullet problems with the Edi module. 1. Replace the Control cable.
2. Replace PC 2023 Panel or the complete
User Interface.

For all other errors:

1. Replace Control cable.


2. Update to latest released System SW
version.
3. Replace battery module.
4. Replace Edi module (if present).
5. Replace PC 1992 Monitoring.
6. Replace PC 1991 Control.
FFFTTT
7. Replace PC 1993 Ethernet Switch.
8. Report to MCC HSC or to your local
Getinge representative for further
information.

56 Missing SW option data. 1. Report to MCC HSC or to your local


Getinge representative for further
information.

57 Missing SW option data. 1. Reinstall software options.


2. Replace PC 1991 Control.
RAAA

58 Missing SW option data. 1. Reinstall software options.


2. Replace PC 2023 Panel.

59 SW option installation failure. 1. Reinstall software options.


2. Replace PC 1992 Monitoring

SERVO-air Ventilator System, Service Manual 6 - 27


| 6 | Troubleshooting |

Error
Error message / Possible cause Recommended action
code
60 Charger error. 1. Replace Battery module.
2. Replace PC 2028 DC/DC & battery control.
This error code is an indication of a charger error in the power
supply subsystem.

The cause is a hardware error for example:

• An SMB-bus
bullet list MUX error which may cause estimation of
remaining battery capacity to be inadequate.
• A charger
bullet list electronics error causing unreasonably high
current consumption.
A Power system restart might be necessary to reset power
related alarms. To restart, switch off the system and disconnect
both the mains power supply and the External 12 V battery.

61 Main fan error. 1. Replace Main fan.


2. Replace PC 2028 DC/DC & battery control.
Power supply to Main fan is monitored and this error code
indicates for example:

• Main fan
bullet listerror.
• Error inlist
bullet the power supply subsystem.
A Power system restart might be necessary to reset power
related alarms. To restart, switch off the system and disconnect
both the mains power supply and the External 12 V battery.

62

63

64
FT FFTT
Very high temperature in Turbine module. Ventilation stopped. 1.

Turbine module communication error. Ventilation stopped.

Turbine module stand still detected.

Turbine module has stopped, O2 is delivered instead.


1.
2.

1.
Replace Turbine module.

Replace Turbine module.


Replace PC 1992 Monitoring.

Replace Turbine module.

65 Turbine module failure. 1. Replace Turbine module.


RARRAA
Turbine module has stopped, O2 is delivered instead.

66 Turbine module under-voltage. 1. Replace Turbine module.

Turbine module has stopped, O2 is delivered instead.

67 Turbine module over-voltage. 1. Replace Turbine module.

Turbine speed control degraded. Ventilation continues but with


decreased gas delivery accuracy.
D

68 Turbine module input voltage failure. 1. Replace Turbine module.

Turbine module has stopped, O2 is delivered instead.


D

69 Turbine module failure. 1. Replace Turbine module.

Turbine module has stopped, O2 is delivered instead.

6 - 28 SERVO-air Ventilator System, Service Manual


| Troubleshooting | 6 |

Error
Error message / Possible cause Recommended action
code
70 Ambient air temperature sensor range error. 1. Replace PC 2030 Inspiratory channel.

71 Air humidity (RH) sensor range error. 1. Replace PC 2030 Inspiratory channel.

72 Inspiratory flowmeter temperature sensor range error. 1. Replace Inspiratory flowmeter.


2. Replace Turbine.

73 Ambient air temperature and humidity sensor error. 1. Replace PC 2030 Inspiratory channel.

74 Inspiratory flowmeter error. 1. Replace Inspiratory flowmeter.


2. Replace Turbine.

75 Power error. 1. Replace PC 2028 DC/DC & Battery control.

+12 V to Turbine module too high, i.e. > +18 V.

A Power system restart might be necessary to reset power


related alarms. To restart, switch off the system and disconnect
both the mains power supply and the External 12 V battery.

76 Power error. 1. Replace PC 2028 DC/DC & Battery control.


2. Replace Turbine module.
+12 V to Turbine module too low, i.e. < +10 V.

A Power system restart might be necessary to reset power


related alarms. To restart, switch off the system and disconnect
both the mains power supply and the External 12 V battery.

77 Power error. 1. Replace PC 2028 DC/DC & Battery control.

+12 V to Inspiratory flowmeter too high, i.e. > +13.5 V.

A Power system restart might be necessary to reset power


related alarms. To restart, switch off the system and disconnect
both the mains power supply and the External 12 V battery.

78 Power error. 1. Replace PC 2028 DC/DC & Battery control.


2. Replace Inspiratory flowmeter.
+12 V to Inspiratory flowmeter too low, i.e. < +10.8 V.

A Power system restart might be necessary to reset power


related alarms. To restart, switch off the system and disconnect
both the mains power supply and the External 12 V battery.

79 Power error. 1. Replace PC 2028 DC/DC & Battery control.

+5 V to Inspiratory flowmeter too high, i.e. > +5.5 V.

A Power system restart might be necessary to reset power


related alarms. To restart, switch off the system and disconnect
both the mains power supply and the External 12 V battery.

SERVO-air Ventilator System, Service Manual 6 - 29


| 6 | Troubleshooting |

Error
Error message / Possible cause Recommended action
code
80 Power error. 1. Replace PC 2028 DC/DC & Battery control.
2. Replace Inspiratory flowmeter.
+5 V to Inspiratory flowmeter too low, i.e. < +4.5 V.

A Power system restart might be necessary to reset power


related alarms. To restart, switch off the system and disconnect
both the mains power supply and the External 12 V battery.

81 Power error. 1. Replace PC 2028 DC/DC & Battery control.


2. Replace Inspiratory flowmeter.
-12 V to Inspiratory flowmeter too high, i.e. > -10.8 V.

A Power system restart might be necessary to reset power


related alarms. To restart, switch off the system and disconnect
both the mains power supply and the External 12 V battery.

82 Power error. 1. Replace PC 2028 DC/DC & Battery control.

-12 V to Inspiratory flowmeter too low, i.e. < -13.2 V.

A Power system restart might be necessary to reset power


related alarms. To restart, switch off the system and disconnect
both the mains power supply and the External 12 V battery.

84 O2 evacuation fan error. 1. Replace the O2 evacuation fan.


2. Replace PC 2028 DC/DC & Battery control.
A Power system restart might be necessary to reset power
related alarms. To restart, switch off the system and disconnect
both the mains power supply and the External 12 V battery.

6.6.2 Breathing

Error
Error message / Possible cause Recommended action
code
10001 Memory backup battery depleted. 1. Replace battery on PC 1991 Control.

The battery on PC 1991 Control is depleted.

10002 Communication error. 1. Replace PC 1991 Control.

Communication error with ID PROM on PC 1991 Control.

10003 Ventilation stopped. 1. Restart the system.


2. Replace PC 1991 Control.
Error in the internal memory detected.

This technical alarm may occur after a software installation.


Restart the system and check that no technical alarms are
activated.

If TE10003 remains, this indicates a HW error.

10004 Missing SW option data. 1. Reinstall software options.


2. Replace PC 1991 Control.

6 - 30 SERVO-air Ventilator System, Service Manual


| Troubleshooting | 6 |

6.6.3 Panel

Error
Error message / Possible cause Recommended action
code
20002 Backlight error. If the display is lit (and the error code is shown
on the display):
The backlight power consumption is measured. If the power 1. Replace PC 2023 Panel.
consumption is outside specified range, this technical error is If the display is not lit (the display is dark):
activated. 1. Replace Display incl. touchscreen.
This error message may also be logged immediately before 2. Replace PC 2023 Panel.
the No battery capacity alarm when the system is running in
Note: With a failure on the display, or likely
battery mode.
also on PC 2023 Panel, the User interface
display will become dark. This error code will
in such case not be possible to see (but will
be logged).

20003 Activate all pending alarm. Report to MCC HSC or to your local Getinge
representative for further information.
There is a software-related internal communications problem.

20004 Alarm sound level too low. 1. Restart the system. Do not touch the User
interface during system startup. Interfering
The loudspeakers are checked using a microphone during with the Touch screen, loudspeaker grid,
startup. etc., may affect the internal technical tests.
2. Replace loudspeaker.
3. Replace PC 2023 Panel.

20005 Communication error. 1. Update to latest released System SW


version.
Ethernet communication failure between PAN and MON. 2. Report to MCC HSC or to your local
Getinge representative for further
May occur with TE 55. information.

20006 Missing SW option data. 1. Restart the system.


2. Update to latest released System SW
If together with TE 56 or TE 10004, see recommended actions version.
for these errors.

SERVO-air Ventilator System, Service Manual 6 - 31


| 6 | Troubleshooting |

6.6.4 Expiratory flowmeter

Error
Error message / Possible cause Recommended action
code
For systems with PC 2004 Expiratory channel

40001 Expiratory flowmeter error. Technical alarm Expiration flow meter PC 2004
and Exp. cassette EEPROM read/write error:
See information logged together with error code 40001 in the 1. Replace PC 2004 Expiratory channel.
Service log / Technical alarms.
Technical alarm Expiration flow meter PC 2004
and Exp. cassette EEPROM checksum error:
1. Replace Expiratory cassette.

Technical alarm Expiration flow meter PC 2004


and PC 2004 thermistor failure:
1. Check Main fan operation for proper
cooling.
2. Replace PC 2004 Expiratory channel.

Technical alarm Expiration flow meter PC 2004


and PC 2004 60 V overrange:
1. Replace PC 2004 Expiratory channel.
2. Replace PC 2028 DC/DC & Battery control.

Technical alarm Expiration flow meter PC 2004


and PC 2004 60 V underrange:
1. Replace PC 1991 Control.
2. Replace PC 2028 DC/DC & Battery control.
3. Replace PC 2004 Expiratory channel.

For systems with PC 2024 Expiratory channel

40001 Internal communication error on PC 2024. 1. Replace PC 2024 Expiratory channel.

40002 Temperature sensor error on PC 2024. 1. Replace PC 2024 Expiratory channel.

Note that this error may occur if the system is running outside
operating temperature range.

40003 Power error. 1. Replace PC 2024 Expiratory channel.

+60 V supply voltage to the ultrasonic transducers in the


Expiratory cassette too high, i.e. > +75 V.

40004 Power error. 1. Replace PC 2024 Expiratory channel.


2. Replace PC 2028 DC/DC & Battery control.
+60 V supply voltage to the ultrasonic transducers in the
Expiratory cassette too low, i.e. < +45 V.

40005 Power error. 1. Replace PC 2024 Expiratory channel.

+1.8 V is more than 15 % outside nominal value.

40006 Power error. 1. Replace PC 2024 Expiratory channel.

+1.2 V is more than 15 % outside nominal value.

6 - 32 SERVO-air Ventilator System, Service Manual


| Troubleshooting | 6 |

Error
Error message / Possible cause Recommended action
code
40007 Power error. 1. Replace PC 2024 Expiratory channel.
2. Replace PC 2028 DC/DC & Battery control.
+5 V is more than 15 % outside nominal value.

40008 Power error. 1. Replace PC 2024 Expiratory channel.

-5 V is more than 15 % outside nominal value.

40009 Power error. 1. Replace PC 2024 Expiratory channel.

+2.5 V is more than 15 % outside nominal value.

40010 Internal temperature too high, > 77 ºC. 1. Ensure proper cooling from the Main fan:
- Check fan operation
The temperature is too high or temperature sensor error on
PC 2024. - Check the operating temperature
- Check the fan filter.
2. Replace PC 2024 Expiratory channel.

40011 Multiple CPU restarts on PC 2024. 1. Restart the system.


2. Replace PC 2024 Expiratory channel.

SERVO-air Ventilator System, Service Manual 6 - 33


| 6 | Troubleshooting |

6.6.5 Other Technical error alarms

Error message / Possible cause Recommended action


Airway pressure continuously high 1. Perform a Pre-use check and follow the recommended
actions if the check fails.
Obstruction leading to constant high airway pressure. 2. If the Pre-use check will pass, the alarm is related to
the patient circuit.

Expiratory cassette error 1. Replace the expiratory cassette.


2. Replace PC 2031/PC 1785 Expiratory channel
Technical error in Expiratory cassette or in the communication connector.
with the cassette. 3. Replace PC 2004/PC 2024 Expiratory channel.
Check the Service log > Technical alarms. If Expiration flow
meter Exp. cassette power failure is logged together with
error code Technical error in Expiratory cassette:

1. Replace PC 1991 Control.


2. Replace PC 2028 DC/DC & Battery control.

Nebulizer hardware error 1. Replace the nebulizer.


2. Replace connection cable.
Technical problem with nebulizer hardware. 3. Check internal PC board connections and cable
connections.
Technical problem with connection cable.
4. Replace PC1994 Nebulizer.
5. Replace PC 1991 Control.

Internal nebulizer error - TE 1. Restart the nebulizer.


2. Replace connection cable.
Technical problem with nebulizer hardware. 3. Check internal PC board connections and cable
connections.
Temperature in the nebulizer patient unit too high.
4. Replace PC1994 Nebulizer.
Technical problem with connection cable. 5. Replace PC 1991 Control.

6 - 34 SERVO-air Ventilator System, Service Manual


| Troubleshooting | 6 |

Error message / Possible cause Recommended action


O2 concentration low 1. Perform a Pre-use check and follow the recommended
actions if the check fails.
O2 concentration high 2. Connect an external gas concentration measuring
instrument to the inspiratory outlet to get a reference
There are two main reasons for these alarms: value:
- If the reference value differs from the set O2
• Gas delivery
bullet list error - Wrong gas concentration is delivered
from the system, but the gas concentration is measured
concentration, this indicates a gas delivery error.
correctly. - If the reference value corresponds to the set O2
concentration, this indicates a measurement error.
• Measurement
bullet list error - Correct gas concentration is
delivered from the system, but the gas concentration is
measured incorrectly.
Gas delivery error – Recommended actions

1. Check gas supply; concentration and pressure. See gas


supply requirements as stated in the User’s Manual.
Also note that:
- Gas supply with incorrect O2 concentration may
cause gas delivery error.
- Insufficient gas supply may cause short intermittent
pressure drops and instable O2 concentrations.
2. Replace nozzle units:
- If O2 concentration high – Start with nozzle unit in O2
gas module.
- If O2 concentration low – Start with nozzle unit in Air
gas module.

TT 3.

4.
Replace gas modules:
- If O2 concentration high – Start with O2 gas module.
- If O2 concentration low – Start with Air gas module.
Replace PC 1991 Control.
AAFF
Measurement error – Recommended actions

1. Replace O2 cell/sensor.
2. Replace PC 1992 Monitoring.

6.6.6 User interface errors

Error Recommended action


DDRR

LCD display not lit 1. Replace Display incl. touchscreen.


2. Replace PC 2023 Panel.

User interface will not start up 1. Check User interface cable.


2. Replace PC 2023 Panel.

Touch screen malfunction 1. Check internal and cable connections.


2. Replace Display incl. Touch screen.
3. Replace PC 2023 Panel.

Alarm indicator malfunction 1. Replace PC 2023 Panel.

SERVO-air Ventilator System, Service Manual 6 - 35


| 6 | Troubleshooting |

6 - 36 SERVO-air Ventilator System, Service Manual


| Preventive maintenance | 7 |

7 Preventive maintenance

Table of Contents
7.1 General | 7-2
7.2 Preventive maintenance intervals | 7-2
7.3 Required equipment | 7-2
7.4 Preventive maintenance calculation | 7-2
and notification
7.5 Maintenance kit SERVO-air 5000 h | 7-3
7.6 Performing the Preventive maintenance | 7-3

SERVO-air Ventilator System, Service Manual 7-1


| 7 | Preventive maintenance |

7.1 General 7.3 Required equipment


It is recommended that a complete cleaning of the • Standard
bullet list service tools.
system should be performed before carrying out
• Barometer
bullet list (or information about the actual
Preventive maintenance. The cleaning procedure is barometric pressure).
only permitted by personnel trained and authorized
by Getinge. Refer to instructions in the User's Manual
• Access
bullet listto the Service & Settings. Recommended
but not required.
(Cleaning and maintenance).
• Preventive
bullet list maintenance kit containing all parts
In some parts of the Preventive maintenance, as needed during the maintenance.
described in this chapter, access to Service & Settings • Isolation
bullet list test cable kit, P/N 68 83 300.
is required.

Optional equipment not covered by this Service 7.4 Preventive maintenance calculation
Manual, may also demand maintenance actions. Refer and notification
to the optional equipment’s documentation.
7.4.1 Calculation

WARNING! The system must not be The calculation of time until next Preventive
serviced or maintained while in use with maintenance (5000 hours) includes only operating
a patient. hours, standby hours are excluded.

Note: Trained and authorized 7.4.2 Notification


personnel. The system must be
The Preventive maintenance notification (PM
serviced at regular intervals by
reminder) is shown on the User interface. The PM
personnel trained and authorized by
reminder is shown after the Pre-use check dialogue
Getinge. It is recommended that service

AAAAA
(performed or cancelled).
and maintenance is done as a part of
a service contract. Any maintenance There are two different time limits for the PM reminder:
or service must be noted in a log book.
• The first
bullet list type of PM reminder will be shown when
there are 200 hours or less until next Preventive
T RRRRR
maintenance.
7.2 Preventive maintenance intervals
• The second
bullet list type of PM reminder will be shown
• Preventive
bullet list maintenance of the system must be when there are 0 hours until next Preventive
performed once a year, or every 5000 hours of maintenance, i.e. time limit for Preventive
operation, whichever comes first. maintenance is exceeded.
D
• The Battery
bullet list module life time is five years The PM reminder is active by default, but can be
D
(60 months) from manufacturing date or 300 charge disabled in Service & Settings CONFIGURATION >
cycles, whichever comes first. SETTINGS > GENERAL > PM reminder.
D
• The Turbine
bullet list module shall be replaced every 10
years.
D
• The Memory
bullet list backup batteries shall be replaced
every five years.
D

7-2 SERVO-air Ventilator System, Service Manual


| Preventive maintenance | 7 |

7.5 Maintenance kit SERVO-air 5000 h 7.6 Performing the Preventive maintenance
Only original parts from the manufacturer must be • Disassembling
bullet list and assembling of the system is
used. Spare parts and maintenance kits can be required when replacing parts included in the
ordered from your local Getinge representative. Maintenance kit SERVO-air 5000 h. If not stated
otherwise, refer to chapter Disassembling and
When performing this maintenance, the Maintenance assembling for instructions.
kit SERVO-air 5000 h should be used.
• All steps
bullet list during Preventive maintenance should be
The following parts, included in the Maintenance kit documented.
SERVO-air 5000 h, shall be replaced: • The letters
bullet list A – D in the text below refers to the
description of the Maintenance kit above.

7.6.1 Preparations

• Checklist
bullet and note Serial number, System software
C version and Operating time.
• Checklist
bullet that a User's Manual corresponding to the
installed System software version is present. Also
check that operating manuals for optional
equipment connected to the system are present.

A B D • Make list
bullet a general inspection/visual check of the
system for external defects or damages. Replace
S20094
defective or damaged parts.
Pos Description Remark • Make list
bullet sure that gas supply hoses and connectors
A Gas module filter For the O2 gas module. are correct.

AAAAAAA
B Gas module nozzle unit For the O2 gas module.
C Bacteria filter For the inspiratory
• Checklist
bullet if there are any unexpected Technical alarms
in the Service log available via Service & Settings.
pressure transducer.
D Air inlet filter For the Turbine module. • Perform
bullet lista Pre-use check to make sure that the
Complete with dust filter. system work properly before the maintenance.
DRDRDRDRR
• Set the
bullet listOn/Off switch to Off.
• Disconnect
bullet list the mains power cable.
• Disconnect
bullet list the O2 gas supply (wall and/or cylinder).
• Remove
bullet list patient tubing.
• If fitted,
bullet listremove bacteria filter from the Expiratory
inlet.

SERVO-air Ventilator System, Service Manual 7-3


| 7 | Preventive maintenance |

7.6.2 Gas module filter (A)

• Replace
bullet list the filter in the O2 gas module.
- When replacing the filter, move the rubber seal
from the old to the new filter.
- Make sure that the rubber seal is clean. Dirt
particles on the rubber seal may cause leakage.

Note: Check for moisture in the O2 gas module filter


housing. If moisture is found, this may indicate too
high moisture content in the O2 gas supply. Inform
the customer that the O2 gas supply must fulfill
requirements for medical grade gases according to
applicable standards. Refer to Technical data in the
User's Manual.
S20095

T T
7.6.3 Gas module nozzle unit (B)

• Replace
bullet list the nozzle unit in the O2 gas module.
AF AF
- Make sure that the O-ring on top of the nozzle
unit is correctly seated and clean. Dirt particles
on the nozzle unit may cause leakage.

Note: After replacement of the nozzle unit, wait


10 minutes before connecting pressure to the O2
gas module.
DR

S20096

7.6.4 Inspiratory pressure transducer filter (C)

• Replace
bullet list the filter for the inspiratory pressure
transducer.
- Make sure that the filter is correctly seated into
the rubber ring.
- Refer to instructions in the User's Manual
(Cleaning and maintenance).

Note: This filter may already be replaced during


cleaning performed in combination with the
Preventive maintenance. S20097

7-4 SERVO-air Ventilator System, Service Manual


| Preventive maintenance | 7 |

7.6.5 Air inlet filter for the turbine (D)

• Replace
bullet list the air inlet filter for the turbine.
- Pull out the filter from the socket.

S20098

- Push the new filter into the socket.

T
AF
S20099

7.6.6 Emergency air intake

• Checklist
bullet that there are no obstructions on the
DR

Emergency air intake (1). Clean if required.

7.6.7 Fan filters

• Remove
bullet list all fan filters:
a
- O2 evacuation fan inlet filter (2).
- Main fan inlet filter (3).
• Checklist
bullet if the filters are damaged. Replace damaged
s d
filters.
• If not damaged,
bullet list clean the filters. The filters can be
rinsed in water. Shake out and make sure that the
filters are free from excess water. S20100

• Mountlist
bullet the new/cleaned filters.
FT T

SERVO-air Ventilator System, Service Manual 7-5


| 7 | Preventive maintenance |

7.6.8 Fan tests

• O2 evacuation
bullet list fan (1) and Main fan (2):
- Turn the system On.
- Check that the O2 evacuation fan and Main fan
start and run for a few seconds.

a s
S20101

TT T T
7.6.9 Expiratory cassette

• Operating
bullet list capacity for the membrane is estimated
to 10 000 000 breathing cycles. When this limit is
passed or if the membrane for some reason has
AFFAF AF
become defective, it must be replaced. Refer to
instructions in chapter Disassembling and
assembling.
• Membrane
bullet list capacity (counter) can be shown in the
System status window. Select SYSTEM STATUS
> Expiratory cassette to check Membrane capacity.
• The operating
bullet list capacity meter must be reset after
replacement of the membrane. Note that the reset
function is not implemented in System version 1.0.
R D R

7-6 SERVO-air Ventilator System, Service Manual


| Preventive maintenance | 7 |

7.6.10 User interface • Selectlist


bullet SYSTEM STATUS > Batteries to check
battery status. For further information, refer to
• Checklist
bullet the screen readability: chapter Service procedures, section Battery
- Transparency modules.
- Surface
- Brightness (backlight). The LED backlight is 7.6.14 Turbine module
designed to maintain an acceptable brightness
level throughout the lifetime of the system. • The Turbine
bullet list module must be replaced every 10
years. Replacement is described in chapter
- Open a bright window, for example the Service
Disassembling and assembling.
& Settings login window, and check if pixels on
the screen are defective. A few defective pixels
can be accepted. Check that defective pixels are 7.6.15 Memory backup batteries
not concentrated to a small area which can
reduce the readability in this area. • The Memory
bullet list backup batteries must be replaced
every five years. Check manufacturing date on the
• Checklist
bullet the touch functions and make sure that the
Memory backup batteries. Replacement is
software responds to these actions. Tap on different
described in chapter Service procedures.

DRTDRTDTR T
parts of the touch screen.

7.6.16 Safety inspection


7.6.11 Barometric pressure
• Checklist
bullet all external cables and their connections for
• Selectlist
SYSTEM STATUS > General. Check that
bullet
F
the Barometric pressure value corresponds to the
actual barometric pressure value at the local site. •
damage.
Perform
bullet listan electrical safety test. Use of an Electrical
Safety Tester and the Isolation test cable kit (P/N

AA A A A
The value shown on the User interface may not
68 83 300) is recommended.
differ more than ±5% from the actual barometric
A
pressure. The electrical safety test is a standard procedure
according to IEC 60601-1:2005 or IEC 62353:2007
• If the value
bullet list shown on the User interface differs more
(or corresponding national standards). Allowable
than ±5% from the actual barometric pressure, the
values and test methods are defined in the
internal barometer must be calibrated. Barometer
standards.
calibration is described in chapter Service
R
procedures. The following tests shall be performed (IEC 62353
terminology in brackets):
A. Protective earthing (Protective earth resistance)
7.6.12 Gas supply pressure transducer
F
B. Touch current (Equipment leakage current –
• Connect
bullet list the gas supply (O2). direct method)
D

• Selectlist
bullet SYSTEM STATUS > General. C. Patient leakage current (Applied part leakage
• Disconnect
bullet list the O2 gas supply. current – direct method)
A

• Checklist
bullet that the O2 pressure value in the SYSTEM D. Insulation (Insulation resistance) – Measured
STATUS -window drops. between mains and protective earth.
F
E. Insulation (Insulation resistance) – Measured
between isolated connections and protective
7.6.13 Battery modules
earth.
R

• The Battery
bullet list modules shall be replaced after The system is classified within the standards listed
A

300 charge cycles or five years (60 months) from above, and shall be tested according to the
manufacturing date, whichever comes first. following:
F

• SERVO-air
bullet list base unit: Class 1, Type B.
SERVO-air Ventilator System, Service Manual 7-7
| 7 | Preventive maintenance |

Test A, B, D.
• Nebulizer
bullet list connector: Type BF. Test C, E.

7.6.17 Completing the Preventive maintenance

• Perform
bullet lista Functional check according to
instructions in chapter Important.
• Resetlist
bullet the Preventive maintenance time counter. In
Service & Settings, tap SERVICE REPORT >
Report completed preventive maintenance > Save
to log. This information will also be stored in the
log.
• Save system
bullet list log files. Remote Services is
recommended.
• Informlist
bullet the customer that Preventive maintenance
has been performed.

TT TTT
F
AAFFAAFAF
R
DRDRDRDR

7-8 SERVO-air Ventilator System, Service Manual


| Index | 8 |

8 Index

Table of Contents
8.1 Alphabetic index | 8-2

SERVO-air Ventilator System, Service Manual 8-1


| 8 | Index |

8.1 Alphabetic index Environmental declaration 1-4


Equipotentiality terminal 2-4
A Ethernet bus 2-14
AC mains power 2-2 Ethernet connector 2-4
AC mains power supply connector with fuse 2-4 Exit 5-17, 5-34
AC/DC converter 2-8, 3-14, 4-18 Expiratory cassette 2-4, 2-9, 3-10, 4-25, 7-6
AC/DC section 2-8, 4-16 Expiratory cassette holder 4-19
AC/DC section lid 4-16 Expiratory cassette membrane 4-25
Aerogen nebulizers 2-3 Expiratory channel connector cable 2-5
Air inlet filter for turbine 3-6, 7-3, 7-5 Expiratory flowmeter 6-32
Air inlet for turbine 2-2 Expiratory heater 2-3
Alarm indicator 2-11, 3-18 Expiratory inlet 2-4
Alarm output 2-3, 3-18 Expiratory inlet with moisture trap 2-2, 2-4, 2-9
Alarm output connector 2-4 Expiratory outlet 2-4, 3-11
Alarms 5-11, 5-26 Expiratory section 3-3
Articles of consumption 1-5 Expiratory valve 2-9, 3-11
Assembling guidelines 4-5 Expiratory valve coil 2-8, 3-11, 4-19
Expiratory valve coil connector 2-5
B
External 12 V battery 2-4
Bacteria filter 2-9, 7-3
Barometer 5-32 F
Barometric pressure 7-7 Fan filters 7-5
Battery charger 2-3, 3-18 Fan tests 7-5
Battery compartments 2-2, 2-4 Fans 4-17
Battery module 3-16, 3-18, 5-45, 7-7 Filter 3-5
Battery module (extra) 2-3 Flash memory 3-4
Battery status 5-46 Flow transducer 3-5
Biomed code 5-14, 5-29 Front cover 4-6
Blower BLO 2-13 Front section lid 4-21
Breathing BRE 2-12, 6-2, 6-30 Functional check 1-3
Fuse for external 12 V battery 2-4
C
Cabinet 2-3 G
Cable holder 2-3 Gas cylinder restrainer kit 2-3, 3-18
Calibration 5-32 Gas inlet for air including air inlet filter 2-4
CAN bus 2-14 Gas inlet for O2 2-4, 3-4
Chassis 4-7 Gas module filter 7-3, 7-4
Completing the Preventive maintenance 7-8 Gas module key 3-5
Components with special environmental concern 1-4 Gas module lid 4-12
Configuration 5-8, 5-23 Gas module nozzle unit 7-3, 7-4
Connectors 10-3 Gas supply pressure transducer 7-7
Construction materials 1-5
Cooling fan air inlet with filter (on both sides) 2-4 H
Copy configuration 5-15, 5-30 Handling PC boards 4-2
Covers 4-6 Hazard notices 1-2
Heating foil 2-9
E Humidifier 3-19
EEPROM 3-4 Humidifier holder 2-3, 3-19
Emergency air intake 2-2, 2-4, 7-5

8-2 SERVO-air Ventilator System, Service Manual


| Index | 8 |

I N28 – Ethernet 10-3


I2C bus 2-14 N29/N30 – RS-232 10-3
ID PROM 3-5 N67 – Alarm output 10-3
Information stored in the system 4-3 Nebulizer 3-14, 3-19, 4-21
Inspiratory channel 3-7 Nebulizer connector 2-4
Inspiratory flowmeter 2-7, 3-7, 4-10 Network 5-12, 5-27
Inspiratory flowmeter cable 2-7 Noise level 1-6
Inspiratory gas docking 2-7, 3-7, 4-13 Nozzle unit 3-5
Inspiratory outlet 2-2, 2-4
Inspiratory pipe 3-7 O
Inspiratory pressure transducer filter 7-4 O2 cell 3-8, 4-11
Inspiratory section 2-7, 3-3, 4-9 O2 cell cable 2-7
Inspiratory section lid 4-9 O2 evacuation fan 2-7, 3-9, 4-10
Inspiratory solenoid 3-4, 3-5 O2 gas module 2-7, 3-4, 4-13
Inspiratory valve temperature sensor 3-5 O2 supply 2-2
Installation 5-16, 5-31 On/Off switch 2-4, 3-15, 4-23
Internal communication 2-14 Optional accessories 2-3
Optional equipment 3-18
L Other Technical error alarms 6-34
LCD Display 3-17
LCD Display/Touch screen 2-11 P
Leakage detection 5-39 Packaging material 1-6
Logs 5-6, 5-19 Panel PAN 2-13, 6-2, 6-31
Loudspeaker 2-10, 3-18 Patient circuit 2-2
Loudspeaker connector 2-10 Patient unit 2-2, 2-4
Loudspeaker grid 2-10 PC 1781 Expiratory pressure transducer 2-5, 3-12
PC 1781 Inspiratory pressure transducer 2-7, 3-10,
M 4-15
Main blocks 3-3 PC 1785 Expiratory channel connector 2-5, 2-9,
Main fan 2-8, 3-17, 4-17 3-11, 4-9
Main section 2-5 PC 1786 Expiratory channel cassette 2-9, 3-11
Main section lid 4-8 PC 1991 Control 2-5, 3-13, 4-9
Main units 2-2 PC 1992 Monitoring 2-5, 3-13, 4-9
Mains inlet 3-14, 4-18 PC 1993 Ethernet switch 2-5, 3-14, 4-9
Mains power supply inlet 2-8 PC 1994 Nebulizer 3-14
Maintenance kit SERVO-air 5000h 7-3 PC 2004 Expiratory channel 2-5, 2-9, 3-12, 4-9
Memory backup batteries 5-47, 7-7 PC 2020 Main back-plane 2-6, 3-12, 4-24, 10-4
Memory types 3-4 PC 2021 Rear connectors 3-15, 4-22
Mobile cart 2-3, 3-18 PC 2023 Panel 2-11, 3-17
Mobile cart drawer kit 2-3 PC 2024 Expiratory channel 2-5, 2-9, 3-12, 4-9
Moisture trap inlet 4-25 PC 2025 Battery back-plane 3-16, 4-21
Monitoring 6-23 PC 2026 Battery extension 3-16, 4-21
Monitoring MON 2-13, 6-2 PC 2027 Motor control 3-6
Mouse 5-2 PC 2028 DC/DC & Battery control 2-8, 3-14, 4-21
MSync (adapter for HL7 communication) 2-3, 3-19 PC 2029 Inspiratory channel extension 3-10, 4-15
PC 2030 Inspiratory channel 2-7, 3-9, 4-15, 10-8
N PC 2031 Expiratory channel connector 2-5, 2-9,
N26 – External 12 V battery 10-3 3-11, 4-9, 4-19
N27 – USB port 10-3
SERVO-air Ventilator System, Service Manual 8-3
| 8 | Index |

PC board rack 4-9 System ID 2-14


PC boards and electrical components 4-2 System overview 10-13
Power consumption 1-5 System software installation 5-37
Power indicators 2-4
Power POW 2-13 T
Pressure transducer connection 2-9 Technical error codes and messages 6-23
Preventive maintenance 7-1 Test case 1 – Test tube (Leakage detection) 5-39,
Preventive maintenance calculation and notification 5-40
7-2 Test case 2 – Expiratory channel (Leakage detection)
Pull magnet 3-8 5-39, 5-41
Test case 3 – Inspiratory section (Leakage detection)
R 5-39, 5-43
RAM 3-4 Tests 5-20
Rear connectors lid 4-22 Thermoshell for expiratory cassette 2-3
Replacing electrical components 4-2 Thread locking adhesives 4-5
Replacing PC boards 4-3 Tightening torque 4-5
RS-232 2-14 To the responsible service personnel 1-4
RS-232 connectors 2-4 Top cover 4-6
RS-485 2-14 Touch screen 3-17, 5-33
Turbine 3-6
S Turbine module 2-7, 3-6, 4-13, 7-7
Safety inspection 7-7 Turbine module incl PC 2027 Motor control 3-6
Safety valve 3-8, 4-11 Turbine module one-way valve 4-14
Service & Settings 5-3
Service & Settings menus 5-3, 10-12 U
Service key 5-2 Ultrasonic flowmeter 2-9
Service Manual revision history 9-5 USB mouse 5-2
Service report 5-7, 5-22 USB ports 2-4
Servo DuoGuard 2-3 User interface 2-2, 2-10, 3-3, 4-28, 7-7
Servo Guard filters 2-3 User interface cable 3-17
SERVO-air revision history 9-2 User interface errors 6-35
Set Date & Time 5-13, 5-28
Settings 5-9, 5-24 V
Shelf base 2-3, 3-19 Valve membrane (Expiratory valve) 2-9
SMBus 2-14
W
Software delivery 5-36
Waterbag/IV pole 2-3, 3-19
Software installation 5-34
Wheel lock 2-2
Software option installation 5-38
Software structure 2-12 Y
Special grease 4-5 Y piece holder and hook 2-3
SPI bus 2-14
Startup configuration 5-10, 5-25
Status 5-5, 5-18
Supply pressure transducer 3-5
Support arm 179 2-3, 3-19
Symbols used in this manual 1-2
System block diagram 10-2
System electronics 3-3

8-4 SERVO-air Ventilator System, Service Manual


| Revision history | 9 |

9 Revision history

Table of Contents
9.1 SERVO-air revision history | 9-2
9.2 Compatibility chart | 9-4
9.3 Service Manual revision history | 9-5

SERVO-air Ventilator System, Service Manual 9-1


| 9 | Revision history |

9.1 SERVO-air revision history

System version # Status Comments


System version 1.0 Not valid Replaced by System version 2.1 in
mandatory update MCC/17/004/IU.
System version 2.1 Valid Clinical improvements:
• GUI adapted
bullet list to the SERVO-U 2.1 GUI.
• Updated
bullet listPRVC algorithm to include
improvements for spontaneously
breathing patients.
• Pressure
bullet listlevel above PEEP maximum
value in NIV increased from 32 cmH2O
to 62 cmH2O.
• For pediatric
bullet list and adult patient category
outside the 130-200 cm height range,
the entered body weight is used and
presented as VT/BW, instead of
VT/PBW (predicted body weight).
• Swedish,
bullet list Norwegian and Danish
languages now included.

Service improvements:
• Now possible
bullet list to reset the expiratory
cassette membrane counter in Service
& Settings.
• SERVO
bullet Communication Interface (SCI)
list
updated to Protocol version 0004.
• Stability
bullet listenhancements to address the
occurrence of technical errors
Technical errors 43, 78, 80 and 81.
• Timinglist
bullet improvements in the internal
communication between Panel and
Monitoring to reduce occurrence of
Technical error 20006.
• Improvements
bullet list to reduce the
occurrence of buzzer alarms during
software installation (2.1.4).
• In Service
bullet list & Settings, STATUS > Parts
data, the new alternative version of PC
2024 Expiratory channel will be
presented with the correct name
(2.1.4).

System version 2.2 Valid Tryout option periods extended from 2 to


RRRRR
4.
D
DD
9-2 SERVO-air Ventilator System, Service Manual
| Revision history | 9 |

System version # Status Comments


System version 4.0 Valid Software improvements:
• Updates
bullet listin base software:
•Driving
bullet list pressure
•bullet list Flow in Volume Control
Decelerating
and the possibility to lock set values
•Short trends
bullet list
•Turbinelist
bullet
50000 hours
lifetime validated up to

•Servo list
bullet Communication Interface
(SCI) updated to protocol version
0007.
The configuration is copied
automatically during software
installation.
New options in Servo-air:
•High flow
bullet list therapy
•Servo compass
bullet list
Servo-air NIV is launched with the
release of system version 4.0.

System version 4.1 Valid • PEEP list


bullet can be set to 0 cm H2O.
• Maquet replaced by Getinge on the
bullet list
user interface.
• Servo list
bullet Communication Interface (SCI)
updated to protocol version 0008.

System version 4.2 Valid • Batterylist


bullet
years.
lifetime extended from 4 to 5

• PUC - Possible
bullet list to perform PUC without
performing the Patient circuit test.

DRRDDDRDRRR
SERVO-air Ventilator System, Service Manual 9-3
| 9 | Revision history |

9.2 Compatibility chart


The chart contains both software and hardware related
information such as:

• Software
bullet list version required to support hardware
changes.
• Minimum
bullet list software version required for a certain
option.

To find article numbers and revision numbers of the


PC boards:

• See the
bullet listlabels on the PC boards (part
number/revision number)
• Go to list
bullet STATUS > Parts data (ARTICLE
NUMBER/VERSION NUMBER) in the Service &
Settings menu. Service key required.
These article numbers cannot be used when ordering
spare parts. See the Spare parts list for correct

TT TT
ordering information.

PC 2004
PC 2024
Expiratory PC 2028
AAFFAAFF
PC 1991 PC 1992 Expiratory channel PC 2023 DC/DC &
Servo-air System SW version Control Monitoring channel 68 81 702 or Panel Battery control
System SW requirement for 66 89 785 66 89 790 66 92 477 68 87 462 68 81 737 68 81 577
version options revision # revision # revision # revision # revision # revision #

1.0.1 – ≤ 08 ≤ 09 ≥ 00 – ≥ 01 ≥ 04
1.0.2 – ≤ 08 ≤ 09 ≥ 00 – ≥ 01 ≥ 04
1.0.3 – ≤ 08 ≤ 09 ≥ 00 ≥ 01 ≥ 01 ≥ 04
2.1.2 – ≥ 06 ≥ 06 ≥ 00 ≥ 01 ≥ 01 ≥ 04
2.1.4 – ≥ 06 ≥ 06 ≥ 00 ≥ 01 ≥ 01 ≥ 04
2.2.0 – ≥ 06 ≥ 06 ≥ 00 ≥ 01 ≥ 01 ≥ 04
High Flow therapy
4.0.0 ≥ 06 ≥ 06 ≥ 00 ≥ 01 ≥ 01 ≥ 04
DDRRDDRR

Servo Compass
4.0.1 – ≥ 06 ≥ 06 ≥ 00 ≥ 01 ≥ 01 ≥ 04
4.1.0 – ≥ 06 ≥ 06 ≥ 00 ≥ 01 ≥ 01 ≥ 04
4.2.0 – ≥ 06 ≥ 06 ≥ 00 ≥ 01 ≥ 01 ≥ 04

PC 1785/PC 2031 Expiratory channel connector


limitations:

• PC 1785
bullet list Expiratory channel connector must be
used together with PC 2024.
• PC 2031
bullet list Expiratory channel connector must be
used together with PC 2004.

9-4 SERVO-air Ventilator System, Service Manual


| Revision history | 9 |

9.3 Service Manual revision history


9.3.1 Service Manual revision 01

• Intended
bullet list for internal use only.
• Not published.
bullet list

9.3.2 Service Manual revision 02

• Intended
bullet list for internal use only.
• Not published.
bullet list

9.3.3 Service Manual revision 03

• First released
bullet list and published version.

TTTTT
AAFAFAFAFF
DDDRDRDRRR

SERVO-air Ventilator System, Service Manual 9-5


| 9 | Revision history |

9.3.4 Service Manual revision 04

In the table below, the following keywords are used:

A = Added. Information about new functionality added.

B = Changed. Information about changed


functionality.

C = Corrected. Corrections made due to printing


errors or clarifications.

A B C Page Section Comment


x All chapters in the Service Manual New layout.
x All chapters in the Service Manual PC 2024 replaces PC 2004.
x All chapters in the Service Manual PC 1785 replaces PC 2031 (in systems with PC 2024).
x 1-1 - Important Restructured and adapted to new guidelines.
1-6
x 1-5 1.8.4 Articles of consumption Filter for O2 cell no longer included in the Preventive
maintenance kit.
x 2-10 - 2.3 User interface Description of User interface clarified.
2-11
x 3-3 Main blocks Description of Main blocks extended and corrected.
x 3-6 3.2 Turbine module including PC Description of Turbine module and PC 2027 extended.
2027 Motor control
x 3-7 3.5 Inspiratory flowmeter Description of Inspiratory flowmeter extended.
x 3-8 3.7 O2 cell Expected cell life time calculation clarified.
x 3-10 3.12 PC 2029 Inspiratory channel Information regarding PC 2029 introduced.
extension
x 3-16 3.31 PC 2026 Battery extension Information regarding PC 2026 introduced.
x 4-6 4.7.1 Tightening torque Tightening torque and procedure clarified.
x 4-7 4.8 Patient unit Disassembling and assembling instructions restructured and
clarified.
x 6-20 6.3.4 Expiratory flowmeter PC 2004: Recommended actions corrected.
x 6-20 6.3.4 Expiratory flowmeter PC 2024: Error code 40001-40011 introduced.
x 7-2 Required equipment An Isolation test cable kit is now available. To be used during
the electrical safety test.
x 7-7 7.5.16 Safety inspection Information regarding the electrical safety test improved.
x 10-12 10.5 System overview Illustration of Turbine module, Inspiratory gas docking and
Inspiratory pipe extended and clarified.

9-6 SERVO-air Ventilator System, Service Manual


| Revision history | 9 |

9.3.5 Service Manual revision 05

In the table below, the following keywords are used:

A = Added. Information about new functionality added.

B = Changed. Information about changed


functionality.

C = Corrected. Corrections made due to printing


errors or clarifications.

A B C Page Section Comment


x All chapters in the Service Manual The Service Manual now states Maquet Critical Care AB as
legal manufacturer but with Getinge as the local service
provider.
x All chapters in the Service Manual The information in the Service Manual is based on System
version 2.2.
x 2-12 2.4 Software structure Software structure illustration introduced.
x 3-4 3.1 O2 gas module (1) Illustration and list of included parts corrected.
CAUTION symbol regarding disassembly corrected to Note.
x 3-6 3.2 Turbine module including PC CAUTION symbol regarding disassembly corrected to Note.
2027 Motor control (2)
x 3-8 3.8 Safety valve including pull Illustration and description of safety valve clarified.
3-9 magnet (8)
x 3-11 3.18 PC 2031/PC 1785 Expiratory Reference to Compatibility chart clarified.
channel connector (18)
x 3-12 3.21 PC 2004/PC 2024 Expiratory Reference to Compatibility chart clarified.
channel (21)
x 3-13 3.22 PC 1991 Control (22) Reference to Compatibility chart clarified.
x 3-14 3.23 PC 1992 Monitoring (23) Reference to Compatibility chart clarified.
x 3-14 3.25 PC 1994 Nebulizer (25) Description reduced to cover only PC 1994. Description of
nebulizer functions moved to 3.41.8 Aerogen nebulizers.
x 3-15 3.30 PC 2021 Rear connectors (30) Information regarding USB connector changed, keyboard not
allowed to connect.
x 3-18 3.41.1 Alarm output Information regarding alarm priority corrected.
x 3-18 3.41.3 Battery charger Battery charger information added.
x 3-19 3.41.8 Aerogen nebulizers Description of nebulizer functions introduced, moved from
3.25 PC 1994 Nebulizer.
x 3-19 3.41.11 MSync MSync information added
x 4-3 4.3.3 Replacing PC boards Reference to Compatibility chart clarified.
x 4-5 4.5.1 Tightening torque Tightening torque for gas inlet nipple introduced.
x 4-8 4.8.2 Replacing the handle New spare part.
x 4-9 4.9.2 PC board rack Reference to Compatibility chart clarified.
x 4-9 4.10.1 Inspiratory section lid Changed disassembly/assembly instructions.
x 4-12 4.10.5 Gas module lid Changed disassembly/assembly instructions.
x 4-20 4.11.6 PC 2031/PC 1785 Expiratory Reference to Compatibility chart clarified.
channel connector

SERVO-air Ventilator System, Service Manual 9-7


| 9 | Revision history |

A B C Page Section Comment


x 4-23 4.13.3 Battery guides New spare part.
x 5-9 5.2.5 Service report Information clarified.
x 5-10 5.2.6 Configuration > Settings Available unit settings changed.
x 5-12 5.2.6 Configuration > Alarms Information clarified.
x 5-18 5.3 Software installation Information clarified.
5-19
x 5-20 5.3.6 System software installation Information clarified.
x 5-21 5.3.7 Software option installation Information clarified.
x 5-31 5.6.3 Replacing Memory backup Information regarding PC 1991 changed.
battery
x 6-3 to 6.3 Pre-use check Detailed description of all sub tests included in Pre-use check
6-6 introduced.
x 6-7 to 6.5 Test logs The previous troubleshooting guides replaced by information
6-20 on how to interpret Test result logs and use Check Failure
Identifier (CFI) to troubleshoot the system.
x 6-21 to 6.6 Technical error codes and Minor corrections of the following error codes:
6-31 messages TE 25
TE 37
TE 43
TE 44 (removed)
TE 55
TE 20005
TE 20006
TE 40001 (for PC 2004)
TE 40010
x 6-32 6.6.5 Other Technical error alarms Information clarified.
x 9-2 9.1 SERVO-air revision history Description of System versions 2.1 and 2.2 added.
x 9-3 9.2 Compatibility chart Compatibility chart introduced.
x 10-2 10.1 System block diagram System block diagram moved from chapter 2.
x 10-3 10.2 Connectors Information regarding N27 - USB port changed.
x 10-12 10.5 Service & Settings menus Information based on System version 2.2.
x 10-13 10.6 System overview MSync introduced.

9-8 SERVO-air Ventilator System, Service Manual


| Revision history | 9 |

9.3.6 Service Manual revision 06

In the table below, the following keywords are used:

A = Added. Information about new functionality added.

B = Changed. Information about changed


functionality.

C = Corrected. Corrections made due to printing


errors or clarifications.

A B C Page Section Comment


x All chapters in the Service Manual The information in the Service Manual is based on System
version 4.0.
x All chapters in the Service Manual Updated the Turbine extended lifetime to 10 years
x 1-4 1.6 Service Added a note with complaint handling information.
x 4-4 4.4 Information stored in the system Added a new note in the table.
x 4-5 4.5.2 Thread locking adhesives Updated with loctite 243.
x 4-25 4.15.2 25 Updated the text for expiratory cassette valve membrane.
x 5-4 5.2.2 4 Added service settings for BIOMED user.
x 5-37 5.2.3 17 Updated the text for service report.
x 5-38 5.3.2 Check before installation Updated text for new system version 4.0.
x 5-20 5.3.6 System software installation Updated text for new system version 4.0.
x 6-3 6.3 Pre-use check Updated Gas supply test and Monitor pressure transducer
test.
x 6-7 to 6.5 Test logs Updated different tables according to system version 4.0.
6-20
x 6-21 to 6.6 Technical error codes and Updated different tables according to system version 4.0.
6-31 messages
x 6-32 6.6.5 Other Technical error alarms Added O2 concentration error in the table.
x 7-7 7.6.10 User interface Updated text for touch function.
x 9-2 9.1 SERVO-air revision history Updated the table with system version 4.0.
x 9-3 9.2 Compatibility chart Updated the table with system version 4.0.
x 10-12 10.5 Service & Settings menus Updated the chart with system version 4.0.

SERVO-air Ventilator System, Service Manual 9-9


| 9 | Revision history |

9.3.7 Service Manual revision 07

A B C Page Section Comment


x 1-3, 1.6 Service, 1.7 Environmental Updated battery lifetime to five years (60 months)
1-5, declaration, 3.33 Battery module
3-16, (33), 5.5 Battery modules, 7.2
5-68, Preventive maintenance
7-2
x All chapters in the Service Manual Updated to "External 12 V battery".
x 5-2 5.1 Service key All logs can be exported with a service key.
x 5-23 5.2 Settings Compliance and resistance removed.
x 5-23 5.2 Copy configuration Added information that it is not possible to copy startup
configuration from system version 4.2 to a lower system
version.

9 - 10 SERVO-air Ventilator System, Service Manual


| Revision history | 9 |

SERVO-air Ventilator System, Service Manual 9 - 11


| Diagrams | 10 |

10 Diagrams

Table of Contents
10.1 System block diagram |10 - 2
10.2 Connectors |10 - 3
10.3 PC 2020 Main back-plane |10 - 4
10.4 PC 2030 Inspiratory channel |10 - 8
10.5 Service & Settings menus |10 - 12
10.6 System overview |10 - 13

SERVO-air Ventilator System, Service Manual 10 - 1


| 10 | Diagrams |

10.1 System block diagram

PC 2025 Baery back-plane


Subsystem allocaon

N34

N33
PC 2026 Baery extension

P34

P33
Baery module

P32
Breathing BRE
Slot 1
Monitoring MON
O₂ evacuaon fan
Safety
valve Panel PAN

P31
Baery module Power POW

T
Slot 2, main slot Blower BLO

Inspiratory
flowmeter
System ID
O₂ cell
P17

P23

P19

P18
P37

N15
Gas module O₂
PC 1781 Inspiratory RH
pressure transducer

Turbine module
PC 2027 Motor control

N16

P1
T

P2
Filter dP
PC 2030 Inspiratory channel
P21

P20

Loud-
speaker
N21

N20
P72

PC 2029 Inspiratory
channel extension
N11

N10
SD
N74

P11

P10

N12
P12

P39
On/Off switch
PC 2023 Panel

N77
N70

N36
HDD

ALM_LED

External

N26
PC 1993 Ethernet switch
N3

P3

12 V baery

Alarm output
N67

connector
N78
P76

P92

T
RS-232
N29

T
N2

P2

PC 1992 Monitoring connector


Touch screen

RS-232
PC 2021 Rear connectors
LCD display

N30

connector
Backlight

N9

P9

N27

USB port (x2)


N1

P1
PC 2028 DC/DC & Baery control

PC 1991 Control

Ethernet
N28

PC 1781 Expiratory PC 2004/PC 2024


pressure transducer Expiratory channel
N4

P4

T
PC 1994 Nebulizer

P40 P41
Nebulizer

Expiratory cassee
P35

PC 2020 Main P45


N5

P5

back-plane PC 1786 PC 2031/


Expiratory Expiratory Expiratory PC 1785
N46

P46

T valve coil flowmeter channel Expiratory


P14

P13

cassee channel
P7

connector

AC/DC converter
PC 2033 AC/DC Mains power
supply connector 110-230 VAC
150 W power supply
Main fan
S20013

10 - 2 SERVO-air Ventilator System, Service Manual


| Diagrams | 10 |

10.2 Connectors

N26 – External 12 V battery N27 – USB port (x2)

Non-isolated connector. USB 2.0 Type A.

S20106
1. GND
2. +12V_UNREG_EXT_DC
Recommended cable connector:
LEMO 736-FGG2B302CLAD82Z, Circular Push Pull
Connector.
Recommended cable:
Outer diameter 7.2 mm – 8.2 mm, two wire cable, area min.
2
2.5 mm (AWG13).

N28 – Ethernet N29/N30 – RS-232

Isolated connector. Unshielded RJ45.


Speed: 10 Mbit/s.

S20107

Isolated connector. 9 pin D-Sub.


RS-232C for data communication via the Servo
Communication Interface (SCI).
1. -
2. RXD_ISO
3. TXD_ISO
4. DTR_ISO
5. GND_ISO
6. -
7. RTS_ISO
8. -
9. -

N67 – Alarm output

S20108

Isolated connector. Modular plug 4/4 (4P4C)


Switching capability: Max 40 V DC, Max 500 mA, Max 20 W.
1. NO – Normally Open
2. NC – Normally Closed
3. Common
4. Common

SERVO-air Ventilator System, Service Manual 10 - 3


| 10 | Diagrams |

10.3 PC 2020 Main back-plane

10 - 4 SERVO-air Ventilator System, Service Manual


| Diagrams | 10 |

SERVO-air Ventilator System, Service Manual 10 - 5


| 10 | Diagrams |

10 - 6 SERVO-air Ventilator System, Service Manual


| Diagrams | 10 |

SERVO-air Ventilator System, Service Manual 10 - 7


| 10 | Diagrams |

10.4 PC 2030 Inspiratory channel

10 - 8 SERVO-air Ventilator System, Service Manual


| Diagrams | 10 |

SERVO-air Ventilator System, Service Manual 10 - 9


| 10 | Diagrams |

10 - 10 SERVO-air Ventilator System, Service Manual


| Diagrams | 10 |

SERVO-air Ventilator System, Service Manual 10 - 11


| 10 | Diagrams |

10.5 Service & Settings menus

10 - 12 SERVO-air Ventilator System, Service Manual


| Diagrams | 10 |

10.6 System overview


9 O2 evacuation fan
Inspiratory section

N19
P19
N15 P1
O2 P19 N19 T RH
7
cell
1 O2 Gas O2
ID #6 P18 N18 P23 N23
module inlet

ID #34
3
Inspiratory E2Pot
Inspiratory “0111””1111”
6 Inspiratory channel gas
outlet docking
5 In
Emergency flo spi
wm rat
air intake 8 eteory
4 r 2
Safety T
Safety valve valve Inspiratory Turbine
pull magnet pipe module
ID #32
Air
P17
T
inlet
X
N17
PC 1781 10 PC 2030 Inspiratory
ID #4 channel N16 P1 PC 2027 Motor control P
P20 P21
N20 N21 E2Prom “011”
11
37 PC 2029 Inspiratory P50 Inspiratory
12
channel extension MOD2 pressure
N10 N11 transducer

PC 1994 25
Nebulizer
Nebulizer N1 P32 33 Battery ID #xx
Aerogen 31 PC 2026 PC 2025 module Slot 1
interface Battery extension 32 Battery back-plane
User interface Battery Slot 2
P33 N33 P31 33 ID #xx
P34 module (main slot)
39
N34
P35

36 Touch screen P10 P11


Alarm 19 PC 2020 Main back-plane
indicator DC/DC
T E2Prom “00X”
37 LCD display +24 V
ID #0
+12 V N3
INTERN. P3
24 PC 1993 Ethernet switch MSync
+12 V (optional)
VALVES
+12 V Ethernet PC 2021 Rear
-12 V E2Prom “000” 30
switch E2Prom “100” ID #1026 connectors
+5 V
P72 N70 38 +3.3 V Mains LED
PC 2023 N2 Power On LED
T
Panel P2 23 PC 1992 Monitoring +12 V LAN
ID #2 Battery N22
X1
P22A/B N27 2xUSB
ID #512 E2Pot “00” E2Prom “101” ID #3
ID #35 CPU
X2 T E2Prom “000” ID #1
PS1 N28 Ethernet
N9
P9

P76
28 PC 2028 DC/DC & Battery control

40 N1 RS-232
Interface
Iso. N29
N78 P1 RS-232 (1)
Battery N22 22 PC 1991 Control
P22A/B RS-232
Interface
Iso. N30 RS-232 (2)
CPU
N12
P12

E2Prom “000” ID #256


35 User interface cable External
N36

12 V battery
N4 Fuse F1 N26
P4
PC 2004/PC 2024
User interface CPU
E2Prom “000”
21 Expiratory channel
P40 N40
Remote alarm
interface
Iso. N67 Alarm
PC 2023 T E2Pot “0010” ID #1024 output
MOD1 P41
N5
P5

P50 N41 N39 P39


P13 N13 P14 N14
20 PC 1781
Expiratory ID #5
pressure E2Prom “100” 29 On/Off
transducer switch
Power On/Off
PC 1994 N45
PC 2028 34 P45 26 Mains inlet
Main
Patient unit fan 18 P90 N90
PC 1786

PC 2033
PC 2031/ 27
PC 2029

PC 2020 PC 2030 AC/DC converter Fuse F11, F12


PC 1785
PC 1781

N7 P7 Expiratory
PC 1781
Exp.

channel
PC 2031/ connector
Insp.

PC 1785
P46 Expiratory section
PC 2004/

PC 1993

N46
PC 1991

PC 1992
PC 2024

PC 2026

H 17 PC 1786 Expiratory 13 Expiratory cassette


PC 2033 B channel cassette
A
E2Prom “X00”
ID #1025
PC 2021 PC 2025
14 Expiratory valve
Heating foil

Ultrasonic flowmeter 16 Expiratory


outlet
S20001 - 2017-09-12

Expiratory pressure tube Expiratory


Expiratory 15 valve
inlet coil

SERVO-air Ventilator System, Service Manual 10 - 13


| 10 | Diagrams |

10 - 14 SERVO-air Ventilator System, Service Manual


x
© Maquet Critical Care AB 2021. All rights reserved. • Maquet reserves the right to modify the design and specifications contained herein without prior notice.
• Order No. 68 84 130 • Service Manual • Printed in Sweden • Rev: 07 •

Getinge is a global provider of innovative solutions for operating rooms, intensive care units, sterilization departments and for life
science companies and institutions. Based on our firsthand experience and close partnerships with clinical experts, healthcare
professionals and medtech specialists, we are improving the everyday life for people - today and tomorrow.

Manufacturer:
Maquet Critical Care AB, Röntgenvägen 2, 171 54 Solna, Sweden.

www.getinge.com

You might also like