TOPSAFE Class II
Model 1.2, 1.5 and 1.8 BIOLOGICAL SAFETY CABINETS
TOPSAFE Class II Biological Safety Cabinets are specifically designed
for the Australian market. They Incorporate a new microprocessor
based control for the handling of low-moderate
risk-hazardous agents.
Features:
• MAC – Microprocessor Automatic Control air volume monitoring.
A new, innovative microprocessor control system. Offering state
of the art safety features the microprocessor-based module provides
automatic control of the pre set airflow volumes throughout the whole usable
life of the installed filters. When a deviation from the set airflow is detected, Materials:
be it from supply voltage increase-decrease or filter loading the automatic • Outer surfaces of cold rolled steel with stove paint finish RAL
control adjusts the airflows accordingly. 7035 e RAL 7036
• Ergonomic design with sloped front for user comfort • Inner exposed surfaces and spilt liquid sump of grade 3 stain-
• Stainless steel work zone chamber less steel and 2B finish
• Pneumatically assisted window • Front screen of safety laminated glass.
• LED display, monitoring airflow parameters. Elapsed hour meter
• Operator security lockout
Accessories:
• Floor stand (supplied on request)
• Quiet, low vibration operation
• Drawers set (supplied on request)
• Splash proof power outlet, Gas and Vacuum taps. Gas with solenoid valve
• Exhaust kits (supplied on request)
• Rear work zone mounted UV light with safety interlock and timer
• Formaline vaporizer (supplied on request).
• H 14 Class HEPA filters as per ISO EN 1822-1
• DOP ports located within negative plenum along with return air guard
Applications:
• Manufactured and designed to comply with Australian and International
standards: AS/NZS 2252.2 and EN 12469 • Hospitals (pharmacies, laboratories, surgical suites, general air
• Manufactured under an accredited ISO 9001 QMS. handling systems)
• Complete after market service provided by our NATA accredited test and • Haemotology, oncology and pathology clinics
certification laboratory. LAF technologies Accreditation No: 14775 • Pharmaceutical, electronics, chemical, medical,stem cell, IVF
• Jack socket for remote monitoring (free voltage max 2A, max 240V) and food manufacturing
• Audible and visible alarms (which cannot be muted). • Universities, animal and research facilities.
Alarms are activated in case of:
- ventilation parameters out of safety limits
- front window open
- fan failure
biocabinets
Biological Safety Cabinets | PCR Cabinets | Laminar Flow Cabinets | Fume Cabinets | Cytotoxic Safety Cabinets | Animal Change Stations
Associated Company:
TOPSAFE Class II
Model 1.2, 1.5 and 1.8 BIOLOGICAL SAFETY CABINETS
Specifications:
Top-Safe 1.2
POWER SUPPLY
• Mains: 240 V single phase – ± 10% 50 Hz ± 1%
• Power excluding internal power point: W 700
• Electrical class: I
DIMENSIONS AND WEIGHT
• Weight (Kg): 180
• Overall dimensions (mm): L. 1380 x H. 1450 x W. 795
• Working area dimensions (mm): L. 1230 x H. 700 x W. 580
• Front opening height (mm): 195
MAIN PARAMETRES
• Lighting: Lux >900
• Vibration at the centre of the work surface: <0.006 mm RMS
• Max increase in ambient temperature: <4°C
• Sound level: < 62 dB(A)
HEPA FILTERS
• LAF filter: 1219 x 610 x 70
• Exhaust filter: 610 x 457 x 90
Top-Safe 1.5
POWER SUPPLY
• Mains: 240 V single phase – ± 10% 50 Hz ± 1%
• Power excluding internal power point: W 1000
• Electrical class: I
DIMENSIONS AND WEIGHT
• Weight (Kg): 220
• Overall dimensions (mm): L. 1685 x H. 1450 x W. 795
• Working area dimensions (mm): L. 1530 x H. 700 x W. 580
• Front opening height (mm): 195
MAIN PARAMETRES
• Lighting: Lux >900
• Vibration at the centre of the work surface: <0.006 mm RMS
• Max increase in ambient temperature: <4°C
• Sound level: < 62 dB(A)
HEPA FILTERS
• LAF filter: 1525 x 610 x 70
• Exhaust filter: 610 x 610 x 90
Top-Safe 1.8
POWER SUPPLY
• Mains: 240 V single phase – ± 10% 50 Hz ± 1%
• Power excluding internal power point: W 1400
• Electrical class: I
DIMENSIONS AND WEIGHT
• Weight (Kg): 290
• Overall dimensions (mm): L. 1990 x H. 1450 x W. 795
• Working area dimensions (mm): L. 1840 x H. 700 x W. 580
• Front opening height (mm): 195
MAIN PARAMETRES
• Lighting: Lux >900
• Vibration at the centre of the work surface: <0.006 mm RMS
• Max increase in ambient temperature: <5°C
• Sound level: < 62 dB(A)
HEPA FILTERS
• LAF filter: 1829 x 610 x 70
• Exhaust filter: 762 x 610 x 70