Download as pdf or txt
Download as pdf or txt
You are on page 1of 5

964

Stripping Voltammetry of Pb(II), Cu(II), and Hg(II) at a


Na®on-Coated Glassy Carbon Electrode Modi®ed by Neutral
Ionophores
Zuliang Chen, Zahra Pourabedi, and D. B. Hibbert*
Department of Analytical Chemistry, The University of New South Wales, Sydney, 2052, Australia

Received: January 14, 1999


Accepted: April 14, 1999

Abstract
The analysis of Pb2‡, Cu2‡ and Hg2‡ by stripping voltammetry using a Na®on-coated glassy carbon electrode modi®ed with diethyl-
dithiocarbamic acid (DDTC) or 1,4,7,10,13,16-hexaoxacyclooctadecane(18-crown-6) is described. Metal ions were accumulated at the
electrode by complexing with the ionophore, reduction and detected by differential pulse anodic stripping voltammetry. A ®ve-fold increase
in the voltammetric response was achieved at the modi®ed electrode compared to a bare glassy electrode. Parameters and conditions such as
the modi®er composition, the pH of supporting electrolyte, deposition potential, deposition time, and stripping mode were optimized. For
®ve minutes preconcentration at DDTC in 100 mM KNO3, pH 2.5, linear calibration plots in the range 0.5 mM to 50 mM were obtained with
a detection limit of 10 nM. For 18-crown-6, after accumulation at open circuit for 1 minute, a linear plot between 2 and 40 mM was achieved.
The lifetimes of each electrode were more than three weeks.
Keywords: Na®on modi®ed electrode, Heavy metal analysis, Stripping voltammetry

1. Introduction potential interfering ions have no electrochemistry (alkali and


alkaline earth metals) or are discriminated in the stripping
Stripping voltammetry, employing various chemically modi- voltammogram. Prabhu et al. have demonstrated the use of
®ed electrodes (CMEs), is widely used in the determination of crowns (dibenzo crowns) and cryptands for the analysis of lead
trace metals and organic compounds [1]. Many CMEs have been [22]. Dithiocarbamates form stable complexes with various metal
reported that are based on the speci®c accumulation of a target ions such as Pb2‡, Cu2‡ and Hg2‡ [23]. Polymeric ion-exchangers
compound at the electrode prior to the voltammetric determina- such as Na®on have attracted particular interest for CMEs. Na®on
tion. Accumulation occurs by reaction between modi®er and the is inert and conducting, and provides both hydrophobic regions to
analyte in several ways, e.g., ion-exchange, complexation, bioac- stabilize ligands, and hydrophilic regions to allow the permeation
cumulation, covalent attachment and hydrophobic interaction, as of aqueous cations [24]. Its ability to reject anions is useful when
described in a review by Arrigan [2]. Recently, there have been analyzing cations in different matrix solutions. Na®on has been
several reports of CMEs used for accumulationystripping voltam- used as an active matrix in electrochemical analysis, for example
metric determination of trace metals based on the use of macro- the voltammetric analysis of metal cations [25, 26] and for
cyclic reagents. These reagents offer a de®ned cavity for metal ion monitoring nitric oxide in physiological conditions [27]. Combin-
complexation and hence the possibility of enhanced sensitivity and ing the complexing ability of ionophores and the sensitivity of
selectivity. A new CME using a-benzoinoxime has been devel- stripping voltammetry requires suf®ciently strong binding to
oped for the determination of Cu2‡ [3]. CMEs modi®ed with accumulate the metal, and a matrix in which the metal ion may
Krypto®x-222 [4] and mercaptoacetic acid [5] were used for be reduced, and subsequently re-oxidized. Na®on ful®ls these
selective accumulation of Hg2‡. CMEs for the determination of requirements, and the model ionophores, 1,4,7,10,13,16-hexa-
Pb2‡ were constructed with ligands such as diphenylthiocarba- oxacyclooctadecane(18-crown-6) and diethyldithiocarbamic acid
zone [6] and benzoin [7]. (DDTC) are both available, stable in the proposed environment,
The use of macrocyclic reagents for stripping voltammetric and have the requisite complexing ability. It must also be observed
analysis of metals has received little attention. Recently, a CME that as the use of mercury for anodic stripping voltammetry is
modi®ed polymeric calixarene has been successfully used for the rapidly coming to a close on health and safety grounds, for the
accumulation of Pb2‡, Cu2‡ and Hg2‡ ion with subsequent anodic technique to continue as a cheap, rapid method for the analysis of
stripping voltammetry [8]. Humic acids have also been incorpo- trace metals in solution, new electrode systems must be found.
rated into modi®ed carbon paste electrode and employed for In this article we demonstrate the use of the electrodes based on
stripping voltammetry of Pb2‡, Cu2‡ and Hg2‡ [9]. Similarly, glassy carbonyNa®ony18-crown-6 or DDTC for the anodic strip-
glyoxal bis(2-hydroxyanil) has been used in a CME for voltam- ping voltammetry of lead, copper and mercury. Parameters
metry of Hg2‡ and Ag‡ [10]. Macrocyclic compounds that can affecting the voltammetric response are optimized.
bind heavy metals including lead [11±17], cadmium [18], mercur- The reaction at a Na®on-coated glassy carbon electrode may be
y(II) [17, 19, 20] have been reported. The simple oxygen crowns, written schematically:
such as 18-crown-6, bind a number of ions, including Hg2‡ [21].
However, when used in ion selective electrodes their lack of
selectivity (in particular their tendency to bind alkali metal ions) ‰CarbonÿÿNafion SO3 Na†2 ÿÿIonophoreŠ ‡ Hg2‡ >
is considered a drawback. In voltammetry, however, many of the 1†
‰CarbonÿÿNafion SOÿ 2‡
3 †2 Ionophore Hg Š ‡ 2 Na
ÿ
ÿ ‡

Electroanalysis 1999, 11, No. 13 # WILEY-VCH Verlag GmbH, D-69469 Weinheim, 1999 1040±0397/99/1309±0964 $17.50‡:50=0
Stripping Voltammetry of Lead(II), Copper(II), and Mercury(II) 965

2eÿ ‡ ‰CarbonÿÿNafion SOÿ 2‡ ‡


3 †2 Ionophore Hg Š ‡ 2 Na !
ÿ
ÿ 18-crown-6 electrodes was carried out at open circuit by dipping
‰CarbonÿÿNafion SO3 Na†2 ÿÿIonophore 0
Hg Š the electrode into 20 mL of a stirred, buffered solution of nitrates
for a given time (the accumulation time). The electrode was
2† removed from the cell, washed thoroughly with distilled water
and dried with a soft tissue. It was then transferred to a cell
‰CarbonÿÿNafion SO3 Na†2 ÿÿIonophore Hg0 Š ! containing fresh electrolyte (0.1 M acetate buffer, pH 4.0). The
2eÿ ‡ ‰CarbonÿÿNafion SOÿ 2‡ solution was deaerated by bubbling nitrogen through it while a
3 †2 Ionophore Hg Š
ÿ
ÿ
potential of ÿ1.1 V was applied to the electrode for about 20

seconds in order to reduce complexed metals. The differential
If the stripping (Eq. 3) is performed in electrolyte not containing pulse voltammograms were recorded from 0 to ‡600 mV or
Hg2‡ the equilibrium of Equation 1 will reverse with removal of ‡1.0 V against AgyAgCl at a scan rate of 20 mV sÿ1 and a pulse
Hg2‡ from the electrode. This is important for the regeneration of amplitude of 50 mV. During measurements a continuous stream of
the electrode for subsequent determinations. nitrogen gas was passed through the solution. After each measure-
ment the 18-crown-6 electrode was regenerated for its next use by
cycling the potential between 0.0 and ‡600 mV at 20 mV sÿ1 until
no peaks appeared (usually three cycles) con®rming the complete
2. Experimental removal of the electroactive material from the electrode surface.
Ions were accumulated and simultaneously reduced at the DDTC
2.1. Apparatus and Reagents electrode by holding the electrode at ÿ1.0 V in 0.1 M KNO3 pH
Voltammetry measurements were performed with a BAS-100B 2.0 solution for the required accumulation time (30 s to 1 min). The
electrochemical analyzer (BAS, Lafayette, USA). The three- electrode was regenerated in fresh 0.1 M KNO3 pH 2.0 solution by
electrode system consisted of a modi®ed glassy carbon disk cycling at 100 mV sÿ1 between ÿ1.0 V and ‡1.0 V.
working electrode (3 mm diameter, BAS, Lafayette, USA), a
AgyAgCl reference electrode (BAS) and a platinum wire
(0.5 mm) auxiliary electrode.
All solutions were prepared with doubly distilled water. The 3. Results and Discussion
ionophores 18-crown-6 and DDTC were supplied by Aldrich
(Gillingham, Dorset, UK). A stock solution of 10 mM of iono- 3.1. Voltammetric Behavior
phore was made in ethanol and added to the polymer solution as
required. The Na®on polymer used in this study was per¯uorinated Initial studies showed that differential pulse voltammetry gave
ion-exchange powder 5 % wyw solution in a mixture of lower superior peaks and baseline than linear sweep voltammetry, and all
aliphatic alcohols and 10 % water (Aldrich) which was then work reported here was performed in the differential pulse mode.
diluted to 1 % with methanol for modi®cation of the electrode. Figure 1 shows the differential pulse stripping voltammograms for
The ionophore and Na®on solutions were prepared in HPLC-grade the test metals using the procedure described above with a bare
solvent and mixed in appropriate ratio to give the modifying glassy electrode, a Na®on-coated electrode and an electrode
solution. All metal nitrate solutions and pH buffer solutions modi®ed with Na®on and DDTC. The sensitivity increased
were made in doubly distilled water and metal nitrate solutions dramatically when the ionophore was introduced into the modify-
were stored in polyethylene bottles. The supporting electrolyte ing coat. We note that the voltammetric responses to Hg2‡ and
was 0.1 M acetate buffer made of sodium acetate and acetic acid
made by dilution of glacial acetic acid (Ajax).

2.2. Preparation of Electrodes


The glassy carbon electrode was polished, rinsed with deio-
nized water, sonicated in dilute nitric acid solution for 3 minutes,
washed using deionized water, and then air-dried prior to being
coated with the modifying solution. A solution containing 1 %
Na®on was prepared by dilution of the stock solution with ethanol.
This was mixed with an equal volume of a 10 mM solution of the
crown or a 2 mM solution of DDTC. 5 mL was applied to a
polished glassy carbon electrode and left to dry in air. A freshly
prepared modi®ed electrode was tested by performing ®ve, 60 s
accumulationy voltammetry cycles in a solution containing 50 mM
of Pb2‡, Cu2‡ and Hg2‡ to ensure a repeatable response. The
electrode was then suitable for use in the various experiments
undertaken.

2.3. Procedure Fig. 1. Differential pulse anodic stripping voltammograms (100 mV sÿ1)
of Pb2‡, Cu2‡ and Hg2‡ at a) glassy carbon, b) glassy carbon ‡ Na®on,
Two procedures were employed to accumulate ions at the and c) glassy carbon ‡ Na®on ‡ DDTC. 30 s accumulation time at
electrodes. Chemical accumulation of the test metal ions on the ÿ1.0 V. Electrolyte 0.1 M KNO3 pH 2.0 plus 50 mM of each ion.

Electroanalysis 1999, 11, No. 13


966 Z. Chen et al.

Cu2‡ were also slightly improved at an electrode modi®ed only


with Na®on, which re¯ects the ion-exchange properties of Na®on.
This con®rms the results seen in the cyclic voltammetry of Hg2‡
and Cu2‡ at a Na®on-modi®ed electrode [25].

3.2. Optimization of Analytical Parameters


The electrode composition was optimized for the amount of
Na®on and the loading of the ionophore. As the mass of Na®on
increases at constant ionophore loading (1 mL of 10 mM) and
constant (2 mL) volume of coating solution, the voltammetric
peaks go through a maximum at about 1 wt. % in alcohol solution
(Fig. 2).
Similarly as the ionophore content is increased at constant
weight of Na®on the peaks also pass through a maximum at
about 15 wt. % of DDTC in ethanol (equivalent to 1 mM, see Fig.
3). There will be a general increase in the amount of ion
accumulated as both Na®on and ionophore are increased.
However, at high loadings of ionophore, the Na®on will be
disrupted by the nonconducting ionophore and it is possible that Fig. 3. The effect of increasing DDTC content of the coating solution at
not all the accumulated ion will be available for the electroche- constant Na®on concentration (1 mL of 1 wt.%). Deposition time: 30 s,
mical steps. As the total loading of the electrode increases the coat deposition potential: ÿ1200 V. Supporting electrolyte: 0.1 M KNO3, pH:
becomes thicker and eventually the transport of ions within the 4.0, sweep rate: 100 mV sÿ1; j) Pb2‡; u) Cu2‡, and s) Hg2‡ 50 mM.
electrode will become dif®cult with the complexation taking place
largely at the electrodeysolution boundary with electrons for the
electrochemical processes being available at the carbonycoat appears to be no adverse effects on the ionophore from this
boundary. procedure. A chemical accumulation at open circuit is necessary
The pH of the electrolyte is important when analyzing metal when the method is used in the ®eld, for example, analyzing water
ions with ionophores with groups that can be protonated. Too high samples. The electrode is simply dipped in the water for the
a pH leads to hydrolysis of the ion, particularly with Hg2‡, and as required length of time, and the subsequent analysis may be
the pH is lowered ionophores such as the crown ethers will done in a portable apparatus nearby, or the electrode may be
protonate at the oxygen atoms reducing the complexing ability stored and stripped back in the laboratory. When a potential was
of the ionophore. Figure 4 illustrates this for Hg2‡ at a DDTC applied during accumulation, the subsequent stripping current was
electrode and a 18-crown-6 electrode. It is seen that the DDTC greater as the accumulation potential was made more negative. A
electrodes could be operated in lower pH solutions with resulting potential of ÿ1.0 V was chosen for the experiments reported here.
higher peak currents. Figure 5 shows the effect of accumulation time of Hg2‡ at a
There are some advantages to depositing metal simultaneously DDTC electrode for simultaneous accumulation and reduction,
with accumulation. The analysis becomes a two step process and at a 18-crown-6 electrode for separate accumulation. The
(accumulationyreduction and stripping) as in conventional ASV, stripping currents increased with accumulation time to a plateau as
the deposition time and the overall time is shortened and there

Fig. 4. The voltammetric peak as a function of electrolyte pH. Filled


squares: Na®onyDDTC modi®ed electrode; 50 mM Hg2‡, accumulation
Fig. 2. The effect of increasing Na®on concentration at constant iono- time 30 s at ÿ1.0 V; electrolyte 0.1 M KNO3 (current60.1). Open
phore loading (5 mM 18-crown-6). The current is the peak current of a squares: a Na®ony18-crown-6 modi®ed electrode, 1 mM Hg2‡ accumu-
solution of 10 mM Hg2‡. Accumulation time: 1 min. lation time: 2 min; electrolyte 0.1 M acetate buffer.

Electroanalysis 1999, 11, No. 13


Stripping Voltammetry of Lead(II), Copper(II), and Mercury(II) 967

neither did the 0.1 M Na‡ in the supporting electrolyte. The


determination of 100 nM of Pb2‡, Cu2‡ and Hg2‡ at a
Na®onyDDTC electrode was not affected by the addition of
10 mM of Co2‡, Ni2‡, Zn2‡, Cd2‡, Mn2‡, and Mg2‡. However,
the stripping currents were reduced by about 15 % in the presence
of the 10 mM Ag‡. The af®nity of sulfur bearing ionophores for
silver ions is well known [16].
A Na®ony18-crown-6 electrode was used each day for three
weeks. The calibration range and linearity for Hg2‡ was main-
tained but the calibration slope fell to 1.1 mA mMÿ1 after two
weeks and 0.8 mA mMÿ1 after three weeks indicating some
leaching of the ionophore.

4. Conclusion
Fig. 5. The peak current for stripping Hg2‡ at a DDTC electrode and a
18-crown-6 electrode as a function of accumulation time. Filled squares: We have demonstrated that glassy carbon electrodes modi®ed
Na®onyDDTC modi®ed electrode; 50 mM Hg2‡, accumulation: ÿ1.0 V; with Na®on and a neutral ionophore can be used for the stripping
electrolyte 0.1 M KNO3 pH 2 (current60.1). Open squares: a Na®ony18- voltammetry of mercury and other heavy metals. The Na®on acts
crown-6 modi®ed electrode, 1 mM Hg2‡, 0.1 M acetate buffer.
as both a suitable conducting medium and matrix for the iono-
phore, and an ion exchanger. The stability of the ionophores and
the ®lm became saturated, this occurring sooner when a potential the ability of the Na®on to extract cations into the membrane leads
was applied during accumulation. to an electrode that is precise and accurate. The analyte ion may be
preconcentrated by chemical interaction with the electrode at open
circuit potential, with subsequent electrochemical reduction, or by
applying a cathodic potential during accumulation for simulta-
3.3. Calibration
neous reduction.
Calibration plots were linear for solutions of Hg2‡ of 0.5 mM to A Na®ony18-crown-6 modi®ed electrode had a linear range of
100 mM at an 18-crown-6 electrode, for 1 minute accumulation 0.5 mM to 100 mM Hg2‡, for 1 minute accumulation time at open
time at open circuit. The slope was 1.2  0.03 (95 % con®dence circuit. The slope was 1.2  0.03 (95 % con®dence interval) mA
interval) mA mMÿ1. Similar results were obtained from the mMÿ1, with a detection limit of 1.9 mM, determined from the
Na®onyDDTC electrodes. For example the slopes of the calibra- calibration plot, or 100 nM determined from 3  sbaseline . Higher
tion lines for 5 minutes accumulation were 6.1, 5.8 and 4.5 mA slopes were found with Na®ony diethyldithiocarbamic acid for
mMÿ1 for Pb2‡, Cu2‡ and Hg2‡, respectively. The electrodes were Pb2‡, Cu2‡ and Hg2‡, namely 6.1, 5.8 and 4.5 mA mMÿ1,
regenerated between each run as described above. Relative respectively, for a 5 minute preconcentration with simultaneous
standard deviation for ®ve measurements was 6.1 % for one reduction at ÿ1.0 V.
minute accumulation of 10 mM Hg2‡ at a Na®ony18-crown-6
electrode and 3.0 to 5.0 % for 50 mM samples of Pb2‡, Cu2‡ and
Hg2‡ at a Na®onyDDTC electrode. 5. References
The detection limit, calculated from the standard error of the
calibration curve by the recommended ISO procedure, was 1.9 mM [1] J. Wang, Stripping Analysis: Principles, Instrumentation and
[28]. This is considerably greater than that calculated from three Application, VCH, Deer®eld Beach, FL 1985.
times the baseline noise (100 nM), but it is considered to be more [2] D.W.M. Arrigan, Analyst 1994, 119, 1953.
realistic in rejecting type II errors (i.e., accepting that an excursion [3] T.Z. Peng, L.Q. Shen, G.S. Wang, Mikrochim. Acta 1996, 122, 125.
from the baseline is a peak when, in fact, it is not). Using the [4] I. Turyan, D. Mandler, Electroanalysis 1994, 6, 838.
[5] Y.X. Zhou, G.Y. Zhu, E.K. Wang, Electroanlysis 1994, 6, 903.
baseline method, for longer accumulation times the detection limit [6] T. Molinaholgado, J.M. Pinillamacias, L. Hernandezhernandez,
of the Na®ony18-crown-6 electrode was estimated to be 10 nM for Anal. Chim. Acta 1995, 309, 117.
10 minutes accumulation of Hg2‡ and about 1 nM for 5 minutes of [7] T.Z. Peng, Z. Tang, G.S. Wang, B.E. Shen, Electroanalysis 1994, 6,
accumulationydeposition of Pb2‡, Cu2‡ and Hg2‡ at a 597.
Na®onyDDTC electrode. [8] D.W.M. Arrigan, G. Svebla, S.J. Harris, M.A. McKervey, Electro-
analysis 1994, 6, 97.
[9] E.D. Jeong, M.S. Won, Y.B. Shim, Electroanalysis 1994, 6, 887.
[10] M.S. Won, D.W. Moon, Y.B. Shim, Electroanalysis 1995, 7, 1171.
3.4. Interferences [11] E. Malinowska, J. Jurczak, T. Stankiewicz, Electroanalysis 1993, 5,
489.
The effect of interfering species can be twofold, by competing [12] N. Tavakkoli, M. Shamsip, Anal. Lett. 1996, 29, 2269.
with the analyte for the ionophore, and by overlapping the [13] S.K. Srivastava, V.K. Gupta, S. Jain, Analyst 1995, 120, 495.
voltammetric wave of the analyte. The latter was not a problem [14] L.K. Shiigun, E.A. Novikov, Yu. A. Zolotov, Zh. Anal. Chem. 1986,
with the interfering ions investigated. Cd2‡ at ÿ800 mV is the 41, 482.
nearest to Pb2‡ at ÿ600 mV and these were easily resolved in the [15] S-R Sheen, J-S Shih, Analyst 1992, 117, 1691.
[16] X. Yang, N. Kumar, H. Chi, D.B. Hibbert, P.W. Alexander, Elec-
voltammetry. troanalysis 1997, 9, 549.
Addition of equimolar amounts of Cu2‡, Pb2‡, Zn2‡, Mn2‡, [17] X. Yang, D.B. Hibbert, P.W. Alexander, Talanta 1997, 45, 155.
Cd2‡ and Ag‡ to 10 mM Hg2‡ solution did not affect the [18] S.K. Srivastava, V.K. Gupta, S. Jain, Electroanalysis 1996, 8, 939.
voltammetric response of a Na®ony18-crown-6 electrode, [19] Z. Brzozka, M. Pietraszkiewicz, Electroanalysis 1991, 3, 855.

Electroanalysis 1999, 11, No. 13


968 Z. Chen et al.

[20] M-T Lai, J-S Shih, Analyst 1986, 111, 891. [25] Y. Nakabayashi, H. Imai, Bull. Chem. Soc. Japan 1994, 67, 113.
[21] I.M. Kolthoff, Anal. Chem. 1979, 51, 1R. [26] H.Y. Yang, I.W. Sun, Anal. Chim. Acta 1998, 358, 285.
[22] S.V. Prabhu, R.P. Baldwin and C. Kryger, Electroanalysis 1989, 1, [27] K.M. Mitchell, E.K. Michaelis, Electroanalysis 1998, 10, 81.
13. [28] International Organisation for Standardisation, Capability of
[23] A. Hulanicki, Talanta 1967, 14, 1371. Detection. Part 2: Methodology in the Linear Calibration Case,
[24] M.L. Iglehart, R.P. Buck, E. Pungor, Anal. Chem. 1988, 60, 290. ISOyDIS 11843±2 1997.

Electroanalysis 1999, 11, No. 13

You might also like