2 FINAL CHECKS
Verify that sensor lead is between 23 and 24 cm long (see Section 5.2 above) EL
Verify that both face and bore seal o-rings are fitted to OEM units /
Verify that interrupted supply current is between 15 and 20mA at 12V (…15…..mA) EL
For OEM units, vacuum test for 20 minutes at 0.2Bar below atmospheric (~2m depth) /
For housed units, check that resistance between 0V supply (Pin 6) and body is >1MΩ EL
3 SIGNATURE
I verify that the instrument referred to in this document has been assembled and tested and checked according to the
procedure defined herein, and is ready for shipment to the customer.
Signature:… ..….......................