Item Description
Software TM 2101
Load- elongation, Elongation-time, Time-elongation,
Curves display
Stress-strain
Max. force, speed, sample information, strength(Kpa, Mpa,
Data display
N/mm, N/mm2) etc…
E-Stop
Over-load protection
Safety features
Upper and lower limit switches
Load sensor with automatic retreat
APPLICATIONS
1. Deformation under defined load: Test deformation under defined load of tested sample
2. Load under defined deformation: Test load under defined deformation of tested sample
3. Software can issue report with results for max. force, leather adhesion,elongation, tensile strength,
peel strength, tear strength, compression strength, etc.
4. This machine can do different test as per customer requirement such as peel test, tear test,
compression test, bend test with different clamps.
Item Specification
Internal dimension (L*W*H) 400*400*700mm
Temperature range -40~+150℃
Temperature Accuracy ±1℃
Humidity accuracy ±3%R.H
Cooling speed 1℃/min in average (without loading)
Heating speed 2℃/min in average (without loading)
Internal chamber material SUS#304 stainless steel, mirror finished
External chamber material Stainless steel with paint spray
Cooling method Air cooling
Korea TEMI 880 controller
Controller LCD touch screen, programmable control temperature and humidity
Can set different parameter for cyclic test
Insulation material 50mm high density rigid Polyurethane foam
Heater Explosion-proof type SUS#304 stainless steel fins radiator pipe heater
Compressor France Tecumseh compressor
Lighting Heat resistance
Temperature sensor PT-100 dry and wet bulb sensor
Observation window Tempered glass
Testing hole Diameter 50mm, for cable routing
Protection for leakage
Over-temperature
Safety protection device Compressor overvoltage and overload
Heater short circuit
Water shortage
PROGRAMMABLE CONTROLLER
1. PIC controller for test chamber
2. Step types include: ramp, soak, jump, auto-start, and end
3. RS-232 interface to connect computer for output
INTERNAL CHAMBER
PARTNER