DigiEye280 Mechanical structure

Radiography product global technical support department

Service@mindray.com
Version number: V5.1
No.:

© 2011 Mindray Confidential


DigiEye280 Mechanical structure
⚫ Course profiles:
➢ This course is first lesson of DigiEye280 products professional classes, mainly introduces
the mechanical structure of the machine, including machine overview, mechanical
movement, mechanical adjustment and the product change history. The most important
part is Mechanical module structure which including Tube stand, Floating patient table
and so on.
➢ After learning this course, engineers can grasp the name of each spare parts and function
of the DigiEye280 system, know the method of disassembly and be able to solve the
general issues.
➢ Before learning this course, should have to learn the following courses or the relevant
knowledge:
《MobiEye700 Product training course》

2 © 2011 Mindray Confidential


DigiEye280 Mechanical structure

1 DigiEye280 overview
2 Mechanical module structure
➢ Tube stand
➢ Floating patient table
➢ Digital wall stand
➢ Operator console
3 Mechanical adjustment
4 Spare part list and Product manual
5 Product change History (EFQ023)

3 © 2011 Mindray Confidential


1 DigiEye280 overview
1.1 The main part of machine
1.2 The movement direction
1.3 The clinical application of machine
1.4 The Demo video

Product model: DigiEye280


Product code: DR15
DrocApp software access:
1. For administrator: admin, mindray99!
2. For engineer: drserver, “dynamic password”

4 © 2011 Mindray Confidential


1.1 The main part of machine
3 Digital wall stand 1 Tube stand 4 Generator

5 Operator console
2 Floating patient table
5 © 2011 Mindray Confidential
1.2 The movement direction
Movement features:
Total 10 movement directions: Full-manual movement, brake.
linear motion (6) :
1, Left-right of tube stand (1720mm)
2, Up-down of the tube head (560~1800mm)
3, Left-right of table plate (900mm)
4, Backward-forward of table plate (260mm)
5, Left-right of under table detector box (580mm)
6,Up-down of digital wall stand detector box
(350~1800mm)

Rotary motion (4) :


1, Vertical rotation of tube head (±180@90 °)
2, Pitching motion of tube head (-20~17 °)
3, Rotating of collimator (±180@15 ° )
4, Rotating of tube stand (±180@90 °)

6 © 2011 Mindray Confidential


1.2 The movement direction
⚫ Movement feature:

± °
° °
± °
± °

7 © 2011 Mindray Confidential


1.3 The clinical application of machine

SID:
Source Image Distance

1.77m

SID:1.8m
3cm

SID:1.8m
8 © 2011 Mindray Confidential
1.3 The clinical application of machine

SID:1.1m

9 © 2011 Mindray Confidential


1.3 The clinical application of machine

SID:1.0m

10 © 2011 Mindray Confidential


1.3 The clinical application of machine

11 © 2011 Mindray Confidential


1.3 The clinical application of machine

12 © 2011 Mindray Confidential


1.3 The clinical application of machine

13 © 2011 Mindray Confidential


1.4 The Demo video

14 © 2011 Mindray Confidential


1.4 The Demo video

15 © 2011 Mindray Confidential


1.4 The Demo video

16 © 2011 Mindray Confidential


1.4 The Demo video

17 © 2011 Mindray Confidential


1.4 The Demo video

18 © 2011 Mindray Confidential


1.4 The Demo video

19 © 2011 Mindray Confidential


1.4 The Demo video

20 © 2011 Mindray Confidential


1 DigiEye280 overview -- Practice
⚫ Make a exposure at the chest position and get the image. (80KV, 320mA, 4ms)
⚫ Make a exposure at the lying position and get the image. (80KV, 320mA, 20ms)
⚫ Check the dose when closing the collimator light field and at the lowest power exposure
parameters (40KV, 10mA, 10ms).

Background
Dose:
0.16 μSv/h
AP: anteroposterior: front to back
PA: posteroanterior: back to front
abdomen AP
21 © 2011 Mindray Confidential
DigiEye280 Mechanical structure

1 DigiEye280 overview
2 Mechanical module structure
➢ Tube stand
➢ Floating patient table
➢ Digital wall stand
➢ Operator console
3 Mechanical adjustment
4 Spare part list and Product manual
5 Product change History (EFQ023)

22 © 2011 Mindray Confidential


2 Mechanical movement structure

2.1 Tube stand


2.2 Floating patient table
2.3 Digital wall stand
2.4 Operator console

23 © 2011 Mindray Confidential


2 Mechanical movement structure 3 1

DigiEye280

1 2 Floating 3 Digital wall


Tube stand
patient table stand

Stand column
2
Stand column Sliding frame
Rope wheel Rope wheel
Table cabinet assembly
assembly
Pulley assembly Table plate Pulley assembly

Balancer Generator Balancer

Column base Under table Wall stand


detector box Detector box
Tube Manual brake

Collimator
Front cover
assembly 4 Operator console
24 © 2011 Mindray Confidential
Module structure-Tube stand
⚫ Tube stand




Stand base Tube head Stand column


25 © 2011 Mindray Confidential
2.1 Tube stand Rope wheel
assembly
⚫ Stand column Pulley
assembly

Stand column

Rope wheel
assembly

Cable clamper

Balance

Pulley assembly Balance


26 © 2011 Mindray Confidential
2.1 Tube stand
⚫ Rope wheel assembly

turntable spindle Electromagnet


fixture

Electromagnet
brake

27 © 2011 Mindray Confidential


2.1 Tube stand
⚫ Pulley assembly

Pulley of tube
stand
床边立柱滑车

Pulley bearing
滑车轴承组件
assembly
Side rigid
Side eccentric bearing
bearing Side
滑车侧偏心轴承组件 assembly
assembly eccentric
bearing
assembly

Side滑车侧轴承组件
rigid
bearing
assembly

28 © 2011 Mindray Confidential


2.1 Tube stand
⚫ Tube head

➢ movement

➢ The head front cover mainly provides the indication of rotation angle.

Head front cover X-ray tube

head installation block

Head
Collimator Angle pointer installation
block
29 © 2011 Mindray Confidential
2.1 Tube stand
⚫ Head installation block

Rotation plate metal

cupules

Align the gap and


hole

Cupules reset spring

30 © 2011 Mindray Confidential


2.1 Tube stand
⚫ Tube hoop and flange
Flange

Tube hoop 1

Rotation angle indicator


Tube hoop 2
Pitching
motion

31 © 2011 Mindray Confidential


2.1 Tube stand
⚫ Head front cover

32 © 2011 Mindray Confidential


2.1 Tube stand
⚫ Collimator

Y direction
X direction

33 © 2011 Mindray Confidential


2.1 Tube stand
⚫ Column base
➢ column base left-right
movement of tube stand.

Upper horizontal
eccentric wheel Upper horizontal
Upper vertical guide wheel
eccentric wheel Middle-sized
floating
Tapered roller electromagnetic
bearing brake

Carriage welded
piece
Lower horizontal
Positioning guide wheel
seat assy.
Lower horizontal
eccentric wheel

Travelling
wheel

34 © 2011 Mindray Confidential


2.1 Tube stand
⚫ Column base
➢ The rotation angle of stand column is free, but due to the limitation of cable winding, we usually
control the rotation angle within ±180°

Rotation axle
sleeve of tube
stand

35 © 2011 Mindray Confidential


2.1 Tube stand -- Practice
⚫ Replacing the Collimator

⚫ Steps

36 © 2011 Mindray Confidential


2.1 Tube stand -- Practice
⚫ Replacing the Head Installation Block

⚫ Steps

37 © 2011 Mindray Confidential


2.2 Floating patient table
⚫ Table plate
⚫ Table body
⚫ Rear sliding frame

38 © 2011 Mindray Confidential


2.2 Floating patient table
⚫ Floating frame

39 © 2011 Mindray Confidential


2.2 Floating patient table
⚫ Slide block assemble

40 © 2011 Mindray Confidential


2.2 Floating patient table
⚫ Stand column linking assembly

Stand column linking assembly

41 © 2011 Mindray Confidential


2.2 Floating patient table
⚫ Under table detector box

42 © 2011 Mindray Confidential


2.2 Floating patient table
⚫ Under table detector box

43 © 2011 Mindray Confidential


2.2 Floating patient table
⚫ Foot switch

44 © 2011 Mindray Confidential


2.2 Floating patient table
⚫ Generator




45 © 2011 Mindray Confidential


2.2 Floating patient table
⚫ Generator

46 © 2011 Mindray Confidential


2.2 Floating patient table
⚫ Generator
Resonant
inverter HV tank
Charge capacitor

EMC

fuse

Filament driver

HV controller

Assistant
Main control power
contactor Bucky
transformer board
47 © 2011 Mindray Confidential
2.2 Floating patient table
⚫ Table power box

48 © 2011 Mindray Confidential


2.2 Floating patient table
⚫ Detector power box

Connect detector Power


220VAC cable Connect To PC Communication indicator
power in (power and signal) (LAN) indicator Power button

Reset button
49 © 2011 Mindray Confidential
2.2 Floating patient table -- Practice
⚫ Replacing the table power box

⚫ Steps:

50 © 2011 Mindray Confidential


2.3 Digital wall stand
⚫ Function parameter:

➢ Patient armrest
➢ (optional)



3
1
Two backup holes
4

51 © 2011 Mindray Confidential


2.3 Digital wall stand
⚫ Drive mechanism: Rope
wheel

Pulley

Wall
stand

Balancer

52 © 2011 Mindray Confidential


2.3 Digital wall stand
⚫ Pulley and balancer

Side eccentric
Poise bearing
assembly
Guide pulley

Balancer
Side rigid bearing
assembly

53 © 2011 Mindray Confidential


2.3 Digital wall stand
⚫ Mechanical brake

Rubber brake

54 © 2011 Mindray Confidential


2.3 Digital wall stand
⚫ Limit structure
Limit structure is to limit the movement of detector pulley.

Lower limit
Upper limit

55 © 2011 Mindray Confidential


2.3 Digital wall stand
⚫ Detector box
There are two kinds of detector box for the wall stand.
➢ Right handed (default)
➢ Left handed Right handed

Detector box
assembly

Left handed

Wall Stand

56 © 2011 Mindray Confidential


2.3 Digital wall stand
⚫ Layout

Left-hand installation
(Reverse layout)
57 © 2011 Mindray Confidential
2.3 Digital wall stand
⚫ Detector box
Detector upper cover

Ionization chamber-grid
assy.

Drawer assy.

Detector lower cover

58 © 2011 Mindray Confidential


组件结构及传动机构介绍-立柱平板片盒组
2.3 Digital wall stand

⚫ Detector box Grid sliding rail
• 3 types of Ionization Chamber;
• 3 types of Grid(1.0m, 1.3m, 1.8m)
Grid

Ionization chamber

Ionization layer board

Grid handle

Ionization 3-filed solid Ionization 3-filed gas SIEMENS ionization chamber


59 © 2011 Mindray Confidential
2.3 Digital wall stand -- Practice
⚫ Change the right-hand detector box to left-hand detector box

⚫ Steps:

60 © 2011 Mindray Confidential


2.4 Operator console
⚫ PC
⚫ Monitor
⚫ System control box
⚫ Hand switch
⚫ Isolate transformer

61 © 2011 Mindray Confidential


2.4 Operator console
⚫ Computer and monitor

62 © 2011 Mindray Confidential


2.4 Operator console
⚫ System control box

63 © 2011 Mindray Confidential


DigiEye280 Mechanical structure

1 DigiEye280 overview
2 Mechanical module structure
➢ Tube stand
➢ Floating patient table
➢ Digital wall stand
➢ Operator console
3 Mechanical adjustment
4 Spare part list and Product manual
5 Product change History (EFQ023)

64 © 2011 Mindray Confidential


3 Mechanical adjustment
⚫ Review

Total 10 movement directions:

linear motion (6) :

Rotary motion (4) :

65 © 2011 Mindray Confidential


3 Mechanical adjustment
⚫ Adjust the level of patient table

Spirit level

66 © 2011 Mindray Confidential


3 Mechanical adjustment
⚫ Adjust the verticality of tube stand

67 © 2011 Mindray Confidential


3 Mechanical adjustment
⚫ Adjust the 0 degree of the tube stand

linearly

68 © 2011 Mindray Confidential


3 Mechanical adjustment
⚫ Adjust the 0 degree of the tube-collimator assy.

69 © 2011 Mindray Confidential


3 Mechanical adjustment
⚫ Adjust position of tube-collimator assy.

70 © 2011 Mindray Confidential


3 Mechanical adjustment
⚫ Adjust the verticality of digital wall stand

L R
up

Spacer
down

71 © 2011 Mindray Confidential


DigiEye280 Mechanical structure

1 DigiEye280 overview
2 Mechanical module structure
➢ Tube stand
➢ Floating patient table
➢ Digital wall stand
➢ Operator console
3 Mechanical adjustment
4 Spare part list and Product manual
5 Product change History (EFQ023)

72 © 2011 Mindray Confidential


4 Spare part list and Product manual

4.1 Spare part list


4.2 Product manual

73 © 2011 Mindray Confidential


4 Spare part list and Product manual
Instructions for each sheet

Definition version:
V: version
19: the year
10: the month
01: the first version of this mouth
Note: the last number is the version number,
not the date.

Including 21 types

Including 5 sheets

74 © 2011 Mindray Confidential


4 Spare part list and Product manual

➢ 1. Classification: DR material classification (21), using the pull-down box to select.


➢ 2. Order Number: CS engineer can apply the order number in CRM system.
➢ 3. 801 Number: The historical number, it’s special packaging material for 801 ordering number.
➢ 4. Part Number(PN) : Machine manufacturing BOM of the original code, usually it’s the same as the
Order Number.
➢ 5. Part Name in PI&PO: The name will be printed in the PI and PO.
➢ 6. Part Description: The name is in the CRM system, usually it’s the same as the name in PI&PO.
➢ 7. Application Range: The scope of application of ECR number or the machine serial number.
➢ 8. Location & Reference: The notes of the material.

75 © 2011 Mindray Confidential


4 Spare part list and Product manual

76 © 2011 Mindray Confidential


4 Spare part list and Product manual

77 © 2011 Mindray Confidential


4 Spare part list and Product manual

78 © 2011 Mindray Confidential


DigiEye280 Mechanical structure

1 DigiEye280 overview
2 Mechanical module structure
➢ Tube stand
➢ Floating patient table
➢ Digital wall stand
➢ Operator console
3 Mechanical adjustment
4 Spare part list and Product manual
5 Product change History (EFQ023)

79 © 2011 Mindray Confidential


5 Product change History
⚫ History
2014.2 2016.8 2020.9
Software:DROC V1.0 Software:DROC V1.0 Software:DrocApp V2.0
Tube:Toshiba (30/50kw) Tube:Toshiba (30/50kw) Tube:Toshiba (30/50/65kw)
Detector:Toshiba (wired) Detector:Mindray (wired) Detector:Mindray (Wired/Wireless)
Collimator: CX-03 (Tungsten lamp) Collimator: CX-03 (Tungsten lamp) Collimator: CX-05 (LED lamp)
Electric:Power box V1.0 Electric:Power box V2.0 Electric:Power box V3.0

Table Power box V3.0

80 © 2011 Mindray Confidential


5 Product change History

Product Code Product Model SN Remark
DigiEye 260 EB2 New
DigiEye 260E EB3 New
DigiEye 260P EB4 New
DigiEye 260T EB5 new
DigiEye 280 EB6 V2.0:B8
DigiEye 280E EB7 New
DR15 DigiEye 280P EB8 New
DigiEye 280T EB9 V2.0:B9
DigiEye 300 EC1 New
DigiEye 300E EC2 New
DigiEye 300P EC3 New
DigiEye 300T EC4 New

81 © 2011 Mindray Confidential


5 Product change History

✓ Mindray Wireless detector;


✓ Detector charge function;
✓ 65KW generator and new tube (65kw);
✓ New collimator;
✓ New electromagnet;
✓ …...

82 © 2011 Mindray Confidential


5 Product change History
⚫ Generator
✓ Add 65KW generator
✓ There are two inverter boards in the inverter module

✓ The transformer changed


✓ Power increased: 400VA→500VA
✓ Add the fuses: for protection

V2.0 V3.0

83 © 2011 Mindray Confidential


5 Product change History
⚫ Detector box
✓ The silk screen changed

V2.0 V3.0

For different size detector


84 © 2011 Mindray Confidential
5 Product change History
⚫ Detector box
✓ Cancel the Wire clamp plate

V2.0 V3.0

85 © 2011 Mindray Confidential


5 Product change History
⚫ Detector box
✓ The grid guide rail changed

V2.0 V3.0

86 © 2011 Mindray Confidential


5 Product change History
⚫ Detector box
✓ The grid handle

V2.0 V3.0

87 © 2011 Mindray Confidential


5 Product change History
⚫ Detector box
✓ Siemens ionization chamber
✓ The fix brock changed

V2.0 V3.0

88 © 2011 Mindray Confidential


5 Product change History
⚫ Detector box
✓ Claymount ionization chamber
✓ The fix brock changed

V2.0 V3.0

89 © 2011 Mindray Confidential


5 Product change History
⚫ Detector box
◆ 1. increase the magnetic steel
✓ Only for wireless detector ◆ 2. increase the charging board

V2.0 V3.0

For different directions of detector


90 © 2011 Mindray Confidential
5 Product change History
⚫ Detector Draw
◆ 1. increase the charging cable
✓ Only for wireless detector

V2.0 V3.0

91 © 2011 Mindray Confidential


5 Product change History
⚫ Detector box under the table
✓ Delete the electromagnet and the button
✓ Increase the damper

Damper Cancel the button

V2.0 V3.0

92 © 2011 Mindray Confidential


5 Product change History
⚫ Electromagnet upgrade
✓ Tube head lifting electromagnet

Electromagnet

Power on: Lock Power on: Unlock


Power off: Unlock Power off: Lock

V2.0 V3.0

93 © 2011 Mindray Confidential


5 Product change History
⚫ Electromagnet upgrade
✓ Table stand left to right movement electromagnet

Power on: Lock Power on: Unlock


Power off: Unlock Power off: Lock

V2.0 V3.0

94 © 2011 Mindray Confidential


5 Product change History
⚫ Electromagnet upgrade
✓ Table floating electromagnet

Power on: Lock Power on: Unlock


Power off: Unlock Power off: Lock

V2.0 V3.0

95 © 2011 Mindray Confidential


5 Product change History
⚫ New collimator: CX-05
◆ 1. LED lamp
◆ 2. Power: 24VDC
◆ 3. adjustment filter:1.5/2.0mmAl

V2.0 V3.0

96 © 2011 Mindray Confidential


5 Product change History
⚫ Router

床体左侧路由器 床体右侧路由 床体侧面通过固定2


安装示意图 器安装示意图 个M8的六角头螺栓
固定路由器安装板.

97 © 2011 Mindray Confidential


5 Product change History
⚫ The wooden case changed
✓ Cancel the cushion

V2.0 V3.0

98 © 2011 Mindray Confidential


99 © 2011 Mindray Confidential

100 © 2011 Mindray Confidential


Version Editor Date Description
Number
1.0 Kennedy Jan.21.2014 Created

2.0 Alan Nov.16.2015 Update


PPT structure optimization

3.0 Alan Oct.16.2016 Update


(EFQ010/EFQ016/EFQ019)
FRU /configuration has been changed

4.0 Tristan Jun.11.2019 Audio

5.0 Tristan Aug.19.2020 Update(EFQ023)


5.1 Benson Mar.08.2021 Update

101 © 2011 Mindray Confidential


Version Reviewed Review Date Revision Details
Number by
1.0 Alfred Jan.21.2014 N.A.
2.0 Alfred Nov.16.2015 N.A.
3.0 Alfred Oct.16.2016 N.A.
4.0 Alfred Oct.10.2019 Audio
5.0 Alfred Aug.19.2020 Update(EFQ023)

102 © 2011 Mindray Confidential