4 NGM200 Specification
4 NGM200 Specification
Data Sheet
Version 03.01
AT A GLANCE
Thanks to their high accuracy and fast load recovery time, the R&S®NGM200 power supplies are perfect
for challenging applications. Their two-quadrant architecture allows them to function both as a source and
a sink to simulate batteries and loads. Their short recovery times enable them to handle fast load changes
that occur, for example, when mobile communications devices switch from sleep mode to transmit mode.
High-speed data acquisition and convenient battery simulation widen the range of applications.
The single-channel R&S®NGM201 and the two-channel Thanks to their fast recovery time of < 30 µs, their mini-
R&S®NGM202 deliver up to 60 W of output power per mum overshoot even during a demanding load change
channel. The output channels are floating, galvanically iso- and fast impedance regulation, the R&S®NGM200 power
lated and protected against overload and short circuits. supplies are ideal for powering IoT devices and other
battery-operated devices.
With four measurement ranges for current and a resolu-
tion of up to 6 ½ digits when measuring voltage, current With an acquisition rate of up to 500 000 sample/s, even
and power, the R&S®NGM200 power supplies are perfect extremely fast variations in voltage or current can be
for characterizing devices that have low power consump- captured.
tion in standby mode and high current in full load opera-
tion. In many cases, an additional digital multimeter is no The linear two-quadrant design of the output stages allows
longer necessary. the R&S®NGM200 power supply series to operate as a
source and sink with minimum residual ripple and noise,
ideally supporting the development of power amplifiers
and MMICs. The optional battery simulation f unctionality
provides test conditions that imitate the use of a real
battery.
2
BENEFITS
Technology for challenging tasks Easy operation
► page 5 ► page 10
Model overview
Parameter R&S®NGM201 R&S®NGM202
Number of output channels 1 2
Total output power 60 W 120 W
Maximum output power per channel 60 W
Output voltage per channel 0 V to 20 V
Maximum output current per channel ≤ 6 V: 6 A, > 6 V: 3 A
Load recovery time < 30 μs
Maximum power and current per channel when used as a load 60 W, 3 A
4
TECHNOLOGY FOR CHALLENGING TASKS
Fast load regulation Minimum residual ripple and low noise
Consumer electronics such as mobile phones and IoT Advanced electronic circuitry is often very complex and
devices require very little power in sleep mode. However, sensitive to interference on the supply lines. In order to
the current increases abruptly as soon as the device supply interference-free voltage to such sensitive DUTs,
switches to transmit mode. A power supply used to power the power supplies must provide extremely stable output
such DUTs must be capable of handling load changes voltages and currents. All types of ripple and noise need to
from a few µA to the ampere range without creating be avoided. The R&S®NGM200 power supplies have linear
voltage drops or overshoots. regulation and are ideal for sensitive DUTs.
Current
Voltage, R&S®NGM200
Time
Two channels can be connected together to supply bipolar circuits with, for example, +12 V/–12 V.
+12 V
–12 V
6
Variable output impedance High-speed acquisition (FastLog functionality)
A power supply should have an output impedance as The R&S®NGM200 power supplies offer the high-speed
low as possible to suppress loading effects on the DUT. FastLog functionality to capture voltage and current mea-
However, there are applications where certain battery surement results. This data can be stored on an external
types need to be simulated in a controlled manner or USB storage device or can be transferred to an exter-
where it is necessary to simulate the increase in internal nal PC via USB or LAN. With an acquisition rate of up to
impedance as the battery discharges. The R&S®NGM200 500 ksample/s, voltage and current results are available
power supplies support these applications due to their every 2 μs. Instead of the readback voltage, also the volt-
adjustable output impedance range. age measured by the optional digital voltmeter functional-
ity (R&S®NGM-K104) can be logged with 500 ksample/s.
Using the two-channel R&S®NGM202, the data acquisition
can run on both channels in parallel.
5 Standard
logging
4 Fast
logging
1
The FastLog functionality follows voltage/current variations
with a resolution of up to 2 μs. It detects spikes that would
0
Time go unnoticed with slower instruments.
Various parameters can be set at the outputs of the R&S®NGM200 power supplies, The high-speed FastLog functionality provides an acquisition rate up to
for example the output impedance, a delay to switch on the outputs and different 500 ksample/s.
trigger modes.
Electronic fuse with additional functions: “Fuse delay at output-on” specifies how The user can set safety limits to limit the adjustment range of the power supply and
long the fuse remains inactive after the channel is activated. The sensitivity of the prevent a DUT from being damaged due to accidentally using the wrong setting.
fuse is specified using the “Fuse delay time”.
8
BATTERY SIMULATION
Battery models Battery simulation
Real batteries show different characteristics depending on In particular, when battery-operated devices have to be
the type of battery and its charging condition. Capacity, optimized for lifecycle, the discharging behavior of the bat-
open circuit voltage (Voc) and equivalent series resistance tery type needs to be considered. The battery simulator
(ESR) are important battery characteristics that depend function of the R&S®NGM200 makes it possible to simu-
on its state of charge (SoC). The optional R&S®NGM-K106 late the real battery output performance. Testing can be
functionality allows users to simulate the behavior of based on a selected battery model, and battery capacity,
batteries under different charging conditions, e.g. when SoC and Voc can be set to any state to test the device
powering a DUT. under specific circumstances.
To define a battery model, the data of the battery can be The charging behavior of a battery can also be s imulated.
entered easily using the integrated battery model e ditor. This is particularly important when designing battery
Data sets for the common battery types Pb, lithium-ion, chargers. In this application, the R&S®NGM200 power
NiCd and NiMH are available as preconfigured files. These supply is used in sink mode.
can be easily modified according to the needs of a specific
application. Additional battery model data sets can Both cases provide dynamic simulation, meaning Voc,
be loaded from a USB device and stored on the ESR and SoC change according to charging/discharging
R&S®NGM200 power supplies. conditions like a real battery. The state of charge is shown
graphically; all other values are displayed numerically.
The battery simulation software includes data sets of common battery types that can Using the two-channel R&S®NGM202, battery simulation can be used simultaneously
be easily modified. on both channels.
All settings and operating modes are easy to read. When the power supply is in constant voltage mode, the numbers and the buttons light up green. Red is used for constant
current mode. The Output button lights up blue to indicate that channels are switched on (active).
10
QuickArb function EasyRamp function
Some applications require the voltage or the current to be Sometimes test sequences have to simulate operating
varied during a test sequence, for example when simu- conditions where the abrupt rise of the supply voltage
lating different charging conditions of a battery. The Arb has to be avoided. The EasyRamp function of the
function allows users to manually configure time/voltage R&S®NGM200 power supplies provides the solution. The
or time/current sequences via the user interface or to pro- output voltage can be increased continuously within a
gram them via external interfaces. time frame of 10 ms to 10 s. The EasyRamp function can
be operated both manually and remotely.
Other power supplies also offer an Arb function, but the
QuickArb function of the R&S®NGM200 power supplies Save and recall instrument settings
sets new standards. More points (4096 points) are sup- The save and recall functions make it easy to save and
ported per cycle. It is also possible to interpolate between recall frequently used settings.
the discrete points and select whether the sequence of
voltage values 1 V – 2 V – 3 V is to be run as steps, or
whether the voltage values are to be increased using linear
interpolation.
Numerical values can be entered using the virtual touchscreen keyboard or the rotary The large high-resolution display makes it easy to read the voltage and current values
knob. (even at great distances) and provides a lot of additional information.
12
SPECIFICATIONS
Definitions
General
Product data applies under the following conditions:
► Three hours storage at ambient temperature followed by 30 minutes warm-up operation
► All data is valid at +23 °C (–3 °C/+7 °C) after 30 minutes warm-up time.
► Specified environmental conditions met
► Recommended calibration interval adhered to
► All internal automatic adjustments performed, if applicable
Specification limit
Specification limit
X-axis
Measurement uncertainties Guard band
Uncertainties
Represent limits of measurement uncertainty for a given measurand. Uncertainty is defined with a coverage factor of 2 and has been calculated in line
with the rules of the Guide to the Expression of Uncertainty in Measurement (GUM), taking into account environmental conditions, aging, wear and tear.
Device settings and GUI parameters are indicated as follows: “parameter: value”.
Typical data as well as nominal and measured values are not warranted by Rohde & Schwarz.
In line with the 3GPP/3GPP2 standard, chip rates are specified in million chips per second (Mcps), whereas bit rates and symbol rates are specified in bil-
lion bit per second (Gbps), million bit per second (Mbps), thousand bit per second (kbps), million symbols per second (Msps) or thousand symbols per
second (ksps), and sample rates are specified in million samples per second (Msample/s). Gbps, Mcps, Mbps, Msps, kbps, ksps and Msample/s are not
SI units.
Electrical specifications
Outputs The channel outputs are galvanically isolated and not connected to ground.
Number of output channels R&S®NGM201 1
R&S®NGM202 2
Maximum output power R&S®NGM201 60 W
R&S®NGM202 120 W
Maximum output power per channel 60 W
Output voltage per channel 0 V to 20 V
Maximum output current per channel ≤ 6 V output voltage 6A
> 6 V output voltage 3A
Maximum voltage in serial operation R&S®NGM202 40 V
Maximum current in parallel operation R&S®NGM202, ≤ 6 V output voltage 12 A
R&S®NGM202, > 6 V output voltage 6A
Adjustable output impedance –50 mΩ to 100 Ω
Increments 1 mΩ
Recovery time ≤ 2 Ω, resistive load < 200 µs (meas.)
> 2 Ω, resistive load < 10 ms (meas.)
< 500 µV (RMS),
Voltage ripple and noise 20 Hz to 20 MHz
< 2 mV (peak-to-peak) (meas.)
Current ripple and noise 20 Hz to 20 MHz < 1 mA (RMS) (meas.)
Electronic load yes, R&S®NGM202: both channels
Sink voltage range 0 V to 20 V
Maximum sink power R&S®NGM201 60 W
R&S®NGM202 120 W (60 W per channel) 1)
Maximum sink current per channel 3A
constant voltage, constant current, constant
Sink modes
resistance
Constant resistance range 0 Ω to 10 kΩ (0.1 Ω increments)
Load regulation load change: 10 % to 90 %
Voltage ±(% of output + offset) < 0.01 % + 1 mV
Current ±(% of output + offset) < 0.01 % + 0.1 mA
Load recovery time regulation to within ±20 mV of the set voltage < 30 µs (meas.)
10 % to 90 % of rated output voltage, full load: < 125 µs (meas.)
Rise time
resistive load no load: < 125 µs (meas.)
90 % to 10 % of rated output voltage, full load: < 125 µs (meas.)
Fall time
resistive load no load: < 125 µs (meas.)
Programming resolution
Voltage 1 mV
Current 0.1 mA
Programming accuracy
Voltage ±(% of setting + offset) < 0.02 % + 3 mV
Current ±(% of setting + offset) < 0.05 % + 2 mA
1)
Time limited at an operating temperature of > +30 °C and total power > 90 W.
14
Output measurements
Measurement functions voltage, current, power, energy
Readback resolution
20 V range: 10 μV
Voltage
5 V range: 1 μV
10 A range: 10 μA
1 A range: 1 μA
Current
100 mA range: 100 nA
10 mA range: 10 nA
Readback accuracy
20 V range: < 0.02 % + 2 mV
Voltage ±(% of output + offset)
5 V range: < 0.02 % + 500 μV
10 A range: < 0.05 % + 250 μA
(< 0.05 % + 100 μA with R&S®NGM-B301 option
installed)
Current ±(% of output + offset)
1 A range: < 0.05 % + 1 mA
100 mA range: < 0.05 % + 100 μA
10 mA range: < 0.05 % + 15 μA
Temperature coefficient (per °C) +5 °C to +20 °C and +30 °C to +40 °C
Voltage 0.15 × specification/°C
Current 0.15 × specification/°C
Remote sensing yes, R&S®NGM202: both channels
Maximum sense compensation 2 V (meas.)
Ratings
Maximum voltage to ground 250 V DC
voltage with the same polarity connected
Maximum counter voltage 22 V
to the outputs
voltage with opposite polarity connected
Maximum reverse voltage 0.5 V
to the outputs
Maximum reverse current for 5 minutes max. 1A
Remote control
Command processing time < 6 ms (nom.)
Protection functions
Overvoltage protection adjustable, R&S®NGM202: both channels
Programming resolution 1 mV
Overpower protection adjustable, R&S®NGM202: both channels
Overcurrent protection (electronic fuse) adjustable, R&S®NGM202: both channels
Programming resolution 0.1 mA
Response time (Iload > Iresp × 2) at Iload ≥ 2 A < 1.5 ms (meas.)
Fuse linking (FuseLink function) R&S®NGM202 yes
Fuse delay at output-on for R&S®NGM202: both channels 0 ms to 10 s (1 ms increments)
Fuse delay time for R&S®NGM202: both channels 0 ms to 10 s (1 ms increments)
yes, R&S®NGM202: independent for each
Overtemperature protection
channel
2)
20 V range or 10 A range: 12 500 measured values.
16
Display and interfaces
Display TFT 5" 800 × 480 pixel WVGA touch
4 mm safety sockets (channel outputs and
Front panel connections R&S®NGM201
remote sensing)
R&S®NGM202 4 mm safety sockets (channel outputs)
Rear panel connections 8-pin connector block per channel
Remote control interfaces standard USB-TMC, USB-CDC (virtual COM port),
LAN
R&S®NGM-K102,
WLAN
for serial numbers < 110 000 only
R&S®NGM-B105 IEEE-488 (GPIB)
Korea KC mark
EU: applied harmonized standard:
Electrical safety
in line with Low Voltage Directive 2014/35/EU EN 61010-1
USA, Canada CSA-C22.2 No. 61010-1
Austria, Belgium, Bulgaria, Croatia, Cyprus,
Czech Republic, Denmark, Estonia, Finland,
France, Germany, Greece, H ungary, Iceland,
Ireland, Italy, Latvia, Liechtenstein, L
ithuania,
WLAN approvals CE0682
Luxembourg, Malta, Netherlands, N orway,
Poland, Portugal, Romania, Slovakia, Slovenia,
Spain, Sweden, Switzerland, Turkey, United
Kingdom, for serial numbers < 110 000
Singapore, for serial numbers < 110 000 iMDA standards DB102020
USA, Canada, for serial numbers < 110 000 FCC, IC
RoHS in line with EU Directive 2011/65/EU EN IEC 63000
Mechanical resistance
5 Hz to 55 Hz, 0.3 mm (peak-to-peak)
Vibration sinusoidal 55 Hz to 150 Hz, 0.5 g const.,
in line with EN 60068-2-6
8 Hz to 500 Hz, acceleration: 1.2 g (RMS),
wideband noise
in line with EN 60068-2-64
40 g shock spectrum,
Shock in line with MIL-STD-810E, method 516.4,
procedure I
Mechanical data
222 mm × 97 mm × 436 mm
Dimensions W×H×D
(8.74 in × 3.82 in × 17.17 in)
Weight R&S®NGM201 7.2 kg (15.9 lb)
R&S®NGM202 7.4 kg (16.3 lb)
Rack installation R&S®HZN96 option ½ 19", 2 HU
operation 40 h/week over entire range of
Recommended calibration interval 1 year
specified environmental conditions
18
ORDERING INFORMATION
Designation Type Order No.
Base unit
Single-channel power supply R&S®NGM201 3638.4472.02
Two-channel power supply R&S®NGM202 3638.4472.03
Accessories supplied
Set of power cables, quick start guide
Options
Remote control via wireless LAN,
R&S®NGM-K102 3644.6367.02
for instruments with serial numbers < 110 000
Digital trigger I/O R&S®NGM-K103 3643.9904.02
Digital voltmeter functionality R&S®NGM-K104 3643.9927.02
Battery simulation R&S®NGM-K106 3636.6626.02
IEEE-488 (GPIB) interface R&S®NGM-B105 3641.6220.02
Higher accuracy in 10 A range R&S®NGM-B301 3689.6208.02
System components
19" rack adapter, 2 HU R&S®HZN96 3638.7813.02
Warranty
Base unit 3 years
All other items 1)
1 year
Options
Extended warranty, one year R&S®WE1
Extended warranty, two years R&S®WE2
Extended warranty with calibration coverage, one year R&S®CW1 Please contact your local
Extended warranty with calibration coverage, two years R&S®CW2 Rohde & Schwarz sales office.
Extended warranty with accredited calibration coverage, one year R&S®AW1
Extended warranty with accredited calibration coverage, two years R&S®AW2
Extended warranty with a term of one and two years (WE1 and WE2)
Repairs carried out during the contract term are free of charge 2) Necessary calibration and adjustments carried out
during repairs are also covered.
1)
For options installed, the remaining base unit warranty applies if longer than 1 year. Exception: all batteries have a 1 year warranty.
2)
Excluding defects caused by incorrect operation or handling and force majeure. Wear-and-tear parts are not included.
www.rohde-schwarz.com
ISO 9001
3609168532
3609.1685.32 03.01 PDP/PDW 1 en