Professional Documents
Culture Documents
AE series SME-v202
AE series SME-v202
02
Contents
0 ELECTRICAL FEATURES AND LINE DISTURBANCE ................................... 1
4 MAINTENANCE.................................................................................................... 37
4.1 Maintenance operation ................................................................................. 38
c.1
AE series SME Ver. 2.02
4.1.1 Daily Maintenance............................................................................................. 38
4.1.2 Monthly Maintenance ....................................................................................... 38
4.1.3 Six Month Maintenance ................................................................................... 38
4.1.4 One Year Maintenance..................................................................................... 38
4.1.5 Cuvette Maintenance ....................................................................................... 38
4.2 Lamp Replacement ....................................................................................... 39
4.3 Probe Replacement ...................................................................................... 39
c.2
AE series SME Ver. 2.02
Serious power supply line disturbances have negative effects on the system.
An efficient ground connection is necessary to reduce line disturbances and guarantee the
maximum safety to the operator.
Symbol Table
Prohibited Action
Necessary Action before using
No warranty applied if you
instrument.
do this action.
1
AE series SME Ver. 2.02
1 GENERAL DESCRIPTION
1.1 Introduction
AE-1000 AE-3000
Assay Method End Point, Fixed Time, Linearity & Non-Linearity Analysis, Kinetics
Speed 100 tests/hour 200 tests/hour
2 reagents 2 reagents
Assay Item Max 45 items simultaneously
Edit Item Any item, no forbidden
Sample Positions 49 60
Reagent Positions R1=28, R2=28 45
Reagent Bottle 50ml and 23ml 70ml and 20ml
Sample Volume 3-50ul 3-100ul
Reagent Volume R1=25-350ul R1=1-350ul
R2=10-150ul R2=0-200ul
Reagent Minimal
180ul 180ul
Volume
Calibration Linearity/Non-Linearity calibration
Control Add control at any time. Six kinds of control at most for one item
Filter(nm) 340, 405,450, 510, 546, 578, 630 ,700
Light Source 12V 35W halogen lamp
Acceptor 8 pcs imported
Linearity Range 0 - 3.0000Abs.
Temperature Control Cuvette 37±0.2 ℃
Cuvette 120 high quality cuvettes
Washing System 4 nozzles 7 nozzles
Dimensions (mm) 740(W)x570(D)x480(H) 840(W)x740(D)x500(H)
Weight (kgs) 46 70
2 2
3 3
5 5
2
AE series SME Ver. 2.02
1.4 Block Diagram
The following figure shows a summarized scheme of functional parts of the Instrument:
3
AE series SME Ver. 2.02
1.5 Mechanical Device i
All the mechanical devices are identified in the following table which provides brief descriptions of
their functions.
Location Description
1 Center Axis
2 Encoder
3 Photo Micro Sensor
4 Sample Cup
5 Sample Plate Holder
6 Shell of Sample Plate
7 Sample Holder Lower Plate
8 Mechanical Floor for Sample Plate
9 Gear for Sample Plate
10 Motor
4
AE series SME Ver. 2.02
1.5.2 Reagent Plate
Location Description
1 Center Axis
2 Encoder
3 Photo Micro Sensor
4 Reagent Bottle
5 Reagent Plate Holder
6 Shell of Reagent Plate
7 Reagent Holder Lower Plate
8 Mechanical Floor for Reagent Plate
9 Gear for Reagent Plate
10 Motor
5
AE series SME Ver. 2.02
1.5.3 Reaction Plate
Location Description
1 Center Axis
2 Gear for Reaction Plate
3 Shell of Reaction Plate
4 Reaction Plate Holder
5 Hole for Filter Holder
6 Main Body for Reaction Plate (Heating System inside)
7 Mechanical Floor for Reaction Plate
6
AE series SME Ver. 2.02
1.5.4 Reaction & Sample Plate (Double Plate)
Location Description
1 Center Axis
2 Encoder
3 Photo Micro Sensor
4 Reagent Plate Holder
5 Reagent Bottle
6 Serum Tube
7 Shell for Reagent & Sample Double Plate
8 Reagent Holder Lower Plate
9 Sample Holder Lower Plate
10 Gear for Sample Plate
11 Motor
7
AE series SME Ver. 2.02
1.5.5 Reagent & Sample Arm
Location Description
1 Motor for Probe Arm Up and Down
2 Motor for Probe Arm Left and Right
3 Gear for Motor(Left and Right)
4 Photo Micro Sensor for Arm Left and Right
5 Linear Bearing for Arm Up and Down
6 Arm Body
7 Beam for Arm
8 Probe
9 Probe Holder
10 Big Gear for Arm Left and Right
11 Bearing Holder Up
12 Belt for Arm Up and Down
13 Gear for Motor (Up and Down)
14 Bearing Holder Down
15 Washing Unit
16 Holder for Washing Unit
8
AE series SME Ver. 2.02
1.5.6 Mixer Arm
Location Description
1 Motor for Mixer Arm Up and Down
2 Motor for Mixer Arm Left and Right
3 Gear for Motor(Left and Right)
4 Photo Micro Sensor for Arm Left and Right
5 Linear Bearing for Arm Up and Down
6 Arm Body
7 Beam for Arm (Mixer Motor Holder)
8 Mixer Holder
9 Mixer Stick
10 Big Gear for Arm Left and Right
11 Bearing Holder Up
12 Belt for Arm Up and Down
13 Gear for Motor(Up and Down)
14 Bearing Holder Down
9
AE series SME Ver. 2.02
1.5.7 Washing Station
Location Description
1 Motor for Washing Arm Up and Down
2 Linear Guide for Washing Arm Left and Right
3 Belt Holder
4 Arm Body
5 Washing Nozzle Holder
6 Beam for Arm
7 Washing Nozzle
8 Fixed Screw Up
9 Fixed Screw Down
10 Gear Guide
11 Belt for Arm Up and Down
12 Gear for Motor (Up and Down)
10
AE-1000 Whole Piping Diagram (Ver2.02)
1.5*3.0(Teflon)
XB2 4 3*6
1.4*3.5
Diaphragm pump 1.4*3.5
(Type B 600-650ml)
1.6 Liquid Circuit
1.4*3.5 1.4*3.5
Wash station
7*10 3*6 3*6 1.4*3.5
7*10 Syringe
Internal tank (Waste)
Check valve
Solenoid valve
3*6
XB1 3
11
Transit Block
Diaphragm pump 3*6
(Type B 600-650ml) XA2
3*6
3*6 3*6 3*6 2 C
XB3
3*6
WASTE WATER
3*6 5 R
Diaphragm pump
(Type B 600-650ml)
AE series SME Ver. 2.02
AE-3000 Whole Piping Diagram (Ver2.02)
1.4*3.5
AE-3000
1.4*3.5
Sample
Reagent Nozzle cleaning unit 1.4*3.5 1.4*3.5 1.4*3.5 1.4*3.5
Mixing
Sample
&
7*10 Reagent
Solenoid valve Syringe
Wash station
Internal tank (Waste)
Transit Block
12
3*6 3*6 1 XA1
Diaphragm pump
(Type A 600-650ml)
3*6 3*6 3*6 3*6
3*6
Diaphragm pump
(Type A 600-650ml)
External joint
3*6 3*6 2
7
XB2 XA2
3 XB1 3*6 Solenoid valve
Diaphragm pump 4
(Type B 600-650ml)
Diaphragm pump XB3 5
3*6
(Type B 600-650ml)
3*6 Diaphragm pump
(Type B 600-650ml) 4*7 4*7 4*7 4*7
3*6
XB4
3*6
WASTE WATER
6 R
3*6
Diaphragm pump
(Type B 600-650ml)
7*10
AE series SME Ver. 2.02
AE series SME Ver. 2.02
1.7 Electrical Circuits
The Main Board is the brain of the instrument. All electronic parts are linked to and controlled by
the Main Board.
The Main Board receives commands from external computer. IAP microprocessor takes control of
all actions in the machine. Also, the program for this board is stored inside.
All the movement of mechanical parts is driven by step motor unit. The microprocessor drives 8
stepper motor control and 8 OPB of the mechanical parts.
The microprocessor drives amplifier board in the optical unit. The resolution of the amplifier is 16
bit.
The Main Board is driven by three power supports: +12V, -12V and 27V.
From Thermistor
From Relay board
J8 J1 From
Switching supply
+12V
G12V
Wavelength sensor adjustment -12V
12V Adjust 24V Adjust
+27V
8 5 4 1
G27V
7 6 3 2
12V 24V 5V 5V
JT7 JT6 JT4 JT2
13
AE series SME Ver. 2.02
1.7.2 Motor Driver Board
14
AE series SME Ver. 2.02
1.7.5 Level Sensor Board
1. Usually the green LED is on. Turn the volume anticlockwise until the LED goes out.
2. After the LED is out, set a finger to the position approx. 10mm from the tip of the probe and turn
the volume clockwise until the LED is on. After the LED is on, turn clockwise about 3 more times
(attention – the probe will be unable to detect the liquid level if the volume is turned too much).
If adjusted properly, the LED will go out for a moment when the tip of probe reaches liquid level.
LED
1. Confirm the current level of sensitivity. Short-circuit the blue-arrowed pins in the diagram
below for a moment. As LED on the board starts to blink on and off, count the number of
times it blinks. The number of the blinks directly shows the current level of sensitivity.
Front Back
15
AE series SME Ver. 2.02
2. When increasing the level (decrease sensitivity), the following procedure must be followed.
Short-circuit the blue-arrowed pins in the diagram below for a moment. As soon as the LED
on the board starts to blink, short-circuit the yellow-arrowed pins. Keep the shorted status
while LED continues to blink. This procedure lowers one senstivity level.
Front Back
3. When decreasing the level (increasing sensitivity), the following procedure must be followed.
Short-circuit the blue-arrowed pins in the diagram below for a moment. As soon as the LED
on the board starts to blink, short-circuit the yellow-arrowed pins. Keep the shorted status
while LED continues to blink. This procedure raises one sensitivity level.
Front Back
4. After adjusting the level, go back to the procedure (1) to confirm the changed sensitivity.
LED
1. Confirm the current level of sensitivity. Short-circuit the blue-arrowed pins in the diagram
below for a moment. As LED on the board starts to blink on and off, count the number of
times it blinks. The number of the blinks directly shows the current level of sensitivity.
Front Back
16
AE series SME Ver. 2.02
2. When increasing the level (deacreasing sensitivity), the following procedure must be followed.
Short-circuit the blue-arrowed pins in the diagram below for a moment. As soon as the LED
on the board starts to blink, short-circuit the yellow-arrowed pins. Keep the shorted status
while LED continues to blink. This procedure lowers one senstivity level.
Front Back
3. When decreasing the level (inceasing sensitivity), the following procedure must be followed.
Short-circuit the blue-arrowed pins in the diagram below for a moment. As soon as the LED
on the board starts to blink, short-circuit the yellow-arrowed pins. Keep the shorted status
while LED continues to blink. This procedure raises one senstivity level.
Front Back
4. After adjusting the level, go back to the procedure (1) to confirm the changed sensitivity.
17
AE series SME Ver. 2.02
The password is 666666. After entering it the service menu will be displayed as follows:
Adjustment differs by model (depending on the number of Arms). Refer to 2-4 for detail.
18
AE series SME Ver. 2.02
19
AE series SME Ver. 2.02
2.1.2 Open Address
When double click the upper right corner, the Open Adressing window appears.
Double click
20
AE series SME Ver. 2.02
21
AE series SME Ver. 2.02
2.2 Machine Config
CAUTION
This menu is very important for instrument. All parameters are
stored in the main board. The engineer can backup and restore
the parameters from this menu. Correct settings will help the
machine to work in the right way.
When installing the instrument in hospital for the first time, be sure
to backup the parameters in this menu with the connecting PC.
Do not modify the settings unless instructed by ERMA.
22
AE series SME Ver. 2.02
Single Cycle Column The cycle time for Testing with single
reagent mode.
AE-1000: 24s
AE-3000: 15s
*There are some differences among
individual units.
Double Cycle Column The cycle time for Testing with double
reagent mode.
AE-1000: 33s
AE-3000: 19s
*There are some differences among
individual units.
23
AE series SME Ver. 2.02
Button After setting all parameters in config
menu, press this button to Send the
parameters to Microprocessor of Main
Board.
Button If the parameters are accidentally cleared
or missed by special condition, press this
button to Get parameter from
Microprocessor of Main Board.
Button After all settings and adjustments, press
this button to Backup all the parameters
and settings to PC file.
Button If the parameters and adjustments are
accidentally cleared or missed by special
condition, press this button to Restore all
the parameters and adjustments from PC
file.
24
AE series SME Ver. 2.02
This menu allows adjustment of the optical unit.
25
AE series SME Ver. 2.02
Button After getting base for the corresponding
filter, click these two buttons to check all
line bases for each cuvette page by page.
26
AE series SME Ver. 2.02
How to adjust the line base?
27
AE series SME Ver. 2.02
2.4 Adjustment
Horiz ontal Le ve l
Mixe r Arm
Re action Plate
Re age nt Arm
Re age nt Plate
108
1
12
Sample Arm
Sample Plate
Horiz ontal Le ve l
28
AE series SME Ver. 2.02
Press Place 1button for Horizontal position.. Confirm that Sample Arm is right above the
Washing Unit. If it is not centered, slightly loosen the screw (see the photo below) to adjust the
Arm to the center. Press OK button after adjustment. +/- Step buttons will not be used.
Vertical Place 1
29
AE series SME Ver. 2.02
Place 3
Use +1, -1, +10, -10 step buttons to move the Arm for adjustment. After finishing adjustment,
click OK button to save the adjusted position or click Cancel button to cancel current
adjustment.
30
AE series SME Ver. 2.02
Press Place 1 button for Horizontal position.. Confirm that Reagent Arm is right above the
Washing Unit. If it is not centered, slightly loosen the screw (see the photo below) to adjust the
Arm to the center. Press OK button after adjustment. +/- Step buttons will not be used.
Vertical Place 1
31
AE series SME Ver. 2.02
Place 2
Vertical Place 3
32
AE series SME Ver. 2.02
2.4.3 Wash Arm
2mm
Washing
Station
Use +1, -1, +10, -10 step buttons to move the Arm for adjustment. After finishing
adjustment, click OK button to save the adjusted position or click Cancel button to cancel
current adjustment.
33
AE series SME Ver. 2.02
2.4.4 Sample Mix Arm (AE-3000)
Press Place 2 button for Horizontal position. Confirm that Sample Mix Arm is right above the Washing
Unit. If it is not centered, slightly loosen the screw (see the photo below) to adjust the Arm to the
center. Press OK button after adjustment. +/- Step buttons will not be used.
Vertical Place 1
34
AE series SME Ver. 2.02
Place 2
Use +1, -1, +10, -10 step buttons to move the Arm for adjustment. After finishing
adjustment, click OK button to save the adjusted position or click Cancel button to cancel
current adjustment.
2.4.5 Reagent and Sample Tray (*When adjustment of Reagent/Sample Arm is insufficient)
PLATE DESCRIPTION
Reagent Tray1 When pressing the button, Reagent Arm moves to the top
center of No.1 Reagent Bottle (the position adjusted in
2-4-2 Horizontal, Place 2). If the position is slightly off the
center of the bottle, adjust the position of Reagent Tray 1.
Sample Tray1 (AE-3000) When pressing the botton, Sample Arm moves to the top
center of No.1 Sample Cup (the position adjusted in 2-4-1
Horizontal, Place 2). If the position is slightly off the center
of the bottle, adjust the position of Sample Tray 1.
Reagent Tray2 (AE-1000) When pressing the botton, Reagent Arm moves to the top
center of No.29 Reagent Bottle (the position adjusted in
2-4-2 Horizontal, Place 5). If the position is slightly off the
center of the bottle, adjust the position of Reagent Tray 2.
Use +1, -1, +10, -10 step buttons to move the Arm for adjustment. After finishing adjustment,
click OK button to save the adjusted position or click Cancel button to cancel current
adjustment.
35
AE series SME Ver. 2.02
3 PARTS TESTING
Reagent Height Field Input reagent position in the No. box and click
check mark. Probe runs to check the liquid
36
AE series SME Ver. 2.02
level of reagent. Pulse number of the
remaining reagent level will be indicated In the
box below. The color of the box (background)
will be displayed in yellow if the remaining
level is low, and in red if there is no remaining
reagent.
Sample Height Field Input sample position in the No. box and click
check mark. Probe runs to check the liquid
level of sample. Pulse number of the
remaining sample level will be indicated in the
box below. The color of the box (background)
will be displayed in yellow if the remaining
level is low, and in red if there is no remaining
sample.
A/D Reading Field Before checking A/D reading, remove reaction
cells 61 – 120 while the power is off. Turn on
the power and let it idle for about 30 minutes.
Then go to Parts Testing menu and click the
check button for A/D reading. A/D value will
be displayed next to each wavelength.
Confirm that the values are around 50,000
with the condition that all the cells are
removed. Click the buttons in order of Set ‘0’,
Save ‘0’. Finish A/D reading by once again
clicking the check button.
If the values are below 20,000, the lamp must
be replaced.
4 MAINTENANCE
37
AE series SME Ver. 2.02
4.1 Maintenance operation
Professional maintenance will protect the machine and let it work well.
CAUTION
38
AE series SME Ver. 2.02
CAUTION
The replacement should be executed after the lamp cools.
Otherwise it will hurt the operator. Be sure not to touch the body
of the lamp directly when replacing lamps.
CAUTION
The position adjustment should be executed carefully and
precisely. Otherwise it will damage the probe.
39
AE series SME Ver. 2.02
6 TROUBLE SHOOTING
Movement errors happen when the software fails to find a reference during a movement.
Possible Causes:
Something is obstructing the movement
Movement flag is broken
Motor is broken or is in a self-safety status
Motor driver on the Main Board is broken
Bridge Board is broken
Motor cable is badly connected
Cable for Photo Micro Sensor is badly connected
Photo Micro Sensor is broken or some light noise is present in the environment
Other mechanical causes
Solutions:
Check mechanical system
Check flag
Check motor
Check Main Board temperature
Check driver on Main Board
Check any light source near Photo Micro Sensor
Check Bridge Board
Check cables on Main Board and Bridge Board
Check cables connecting Photo Micro Sensor to Bridge Board
Possible Causes:
Heating system is broken
Cable for heating system is badly connected
Temperature sensor is broken or is not working properly
Temperature sensor cable is badly connected
Heating control of Main Board is broken
Solutions:
Check Heating System
Check the cable connection
Check connection between temperature sensor and heating system
Check cable between temperature sensor and Main Board
Replace the board
40
AE series SME Ver. 2.02
6.3 Freeze Error
Possible Causes:
Peltier is broken
Power of freeze system is broken
Solution:
Check Peltier. If it is broken, replace it.
Check the power of freeze system. Check the voltage of 12V switching regulator.
Possible Causes:
Serial cables are connected to wrong ports or badly connected
Main Board is broken
Main Board has no power supply
Solution:
Check serial cable connections and choose the right ports
Check Main Board and power supply
Possible Causes:
Instantaneous current or voltage shocks led to this condition.
Solution:
Restore Adjustment from file with SERVICE MENU- MACHINE CONFIG.
Possible Causes:
Waste bottle is full
Water bottle or acid bottle or basic bottle is empty
Level sensors are broken or not working properly
Cables between level sensor and connector are badly connected
Cable between connector and Main Board is badly connected
Solution:
Check bottle level
Check level sensor
Check cables between level sensor and connector
Check cable between connector and Main Board
Possible Causes:
Some serum tubes and reagent bottles are empty
Faulty adjustment
Level sensor on Level Sensor Board is broken
Level Sensor Board is broken or badly power supplied
Solution:
Check reagent and serum levels
Check adjustment
Check level sensor
Check Level Sensor Board
41
AE series SME Ver. 2.02
6.8 Level Sensor Error
Possible Causes:
Bad ground connection
Bad connection of cable between the probe and Level Sensor Board
Level Sensor Board is broken or badly power supplied
Solution:
Check all ground connection for the instrument.
Check the cable between the probe and Level Sensor Board.
Check Level Sensor Board and power support.
This problem causes bad CV% on end point analysis. Possible causes can be divided in
four groups:
Preparation Problems
Reading Problems
Washing Problems
Mechanical Problems
Solution:
Check probe
Check Level Sensor Board and probe level on serum
Check liquid circuit of washing unit and be sure that water in the bottle is not dirty
Check probe up-down adjustment on washing unit (probe must be 2-3 mm below the
water level )
Check probe up-down adjustment on reaction cuvette (probe must be 2 cm below upper
cuvette level, and it must be in the center of the cuvette)
Check dilutor
Check motor cable of dilutor and its connection (see Movement Error section)
Check liquid circuit and all liquid circuit connection under pressure
Solution:
Check Amplifier Board
Check voltage for all wavelengths
42
AE series SME Ver. 2.02
Check 2.3 Base Adjust
Check lamp and lamp power supply
Check optical system at the wavelengths with poor results
Check silicone diodes
Check if filters are dirty
Check if environmental noises are present
Check cuvettes
Solution:
Check washing solution and water level
Check washing solution
Check water
Check washing station mechanical system (the far right Tip must reach the bottom of
cuvette)
Check washing station movement (see Movement Error section)
Check washing station adjustment (the far right Tip must fit cuvette)
Check washing station probes
Check washing liquid circuit
Check pumps
Solution:
Check movement of Reaction Plate
Check Mixer Stick
This problem may cause too high/low readings on kinetics analysis; possible causes can
be divided in two groups:
Reading Problems
Temperature Problems
Mechanical Problems
43
AE series SME Ver. 2.02
Lenses on the optical system are dirty
Silicone diodes are dirty
Filters are dirty
Environmental noises are present
Environmental light noises are present
Cuvettes are damaged
Solution:
Check Amplifier Board
Check voltage for 340 wavelength
Check 2.3 Base Adjust
Check lamp and lamp power supply
Check optical system at the wavelengths with poor results
Check and clean lenses
Check silicone diodes
Check if filters are dirty
Check if environmental noises are present
Check environmental light
Check cuvettes
Solution:
Check heating system
Check cable of heating system connection
Check connection between temperature sensor
Check cable between temperature sensor and Main Board
Solution:
Check movement of Reaction Plate
Check Mixer Stick
This problem may cause too high/low readings or bad CV% on Fixed Time analysis;
possible causes can be divided in two groups:
Reading Problems
Temperature Problems
Preparation Problems
Washing Problems
Mechanical Problems
44
AE series SME Ver. 2.02
Optical system is defective
Lenses on the optical system are dirty
Silicone diodes are dirty
Filters are dirty
Environmental noises are present
Environmental light noises are present
Cuvettes are damaged
Method parameters (Start point, End point) are not properly programmed
Solution:
Check Amplifier Board
Check voltage for 340 wavelength
Check 2.3 Base Adjust
Check lamp and lamp power supply
Check optical system at the wavelengths with poor results
Check and clean lenses
Check silicone diodes
Check if filters are dirty
Check if environmental noises are present
Check environmental light
Check cuvettes
Check method parameters
Solution:
Check heating system
Check cable of heating system connection
Check connection between temperature sensor
Check cable between temperature sensor and Main Board
Solution:
Check probe
Check Level Sensor Board and probe level on serum
Check liquid circuit of washing unit and be sure that water in the bottle is not dirty
45
AE series SME Ver. 2.02
Check probe up-down adjustment on washing unit (probe must be 2-3 mm below the
water level )
Check probe up-down adjustment on reaction cuvette (probe must be 2 cm below upper
cuvette level, and it must be in the center of the cuvette)
Check dilutor
Check motor cable of dilutor and its connection (see Movement Error section)
Check liquid circuit and all liquid circuit connection under pressure
Solution:
Check washing solution and water level
Check washing solution
Check water
Check washing station mechanical system (the far right Tip must reach the bottom of
cuvette)
Check washing station movement (see Movement Error section)
Check washing station adjustment (the far right Tip must fit cuvette)
Check washing station probes
Check washing liquid circuit
Check pumps
Solution:
Check movement of Reaction Plate
Check Mixer Stick
46
AE series SME Ver. 2.02
IVD