002-0727-00 Actuator Kit (Tilt & Back)

Installation Instructions (Models 416 / 417)

Installation
Kit Components (1) Check dimensions and Limit Switch operation
Part Number Description Qty. of the Actuator (Refer to Figure 1).
002-0727-00 Actuator Kit
• 029-2980-00 Actuator w/ Limit Switches ..........1 EQUIPMENT ALERT
• 015-1656-00 Harness (Up Limit, Red Wires) ...1 Assure Extend Limit Switch normally
• 015-1657-00 Harness (Dwn. Limit, Blk. Wires) 1 closed contact (C-NC) is in the OPEN
• 015-1113-00 Wire Nuts ....................................2 position when installing actuator onto table.
• 015-0013-01 Wire Tie (14 1/2" long) ................2 (2) Set the actuator so the contacts (C-N/C) on the
• 003-1334-00 Installation Instructions................1 Extend limit switch just trip to the OPEN posi-
tion.
Removal
(1) Remove the Tilt and/or Back Actuator and EQUIPMENT ALERT
Capacitor, refer to the appropriate Service and When installing the Actuator per instruc-
Parts Manual (Model 416 or 417) for proce- tions in the manual do not connect the
dures. table electrical leads and run the actuator unless they
have been changed to match those in Figure 2
(3) Install the Tilt and/or Back Actuator and Capac-
itor, refer to the appropriate Service and Parts
Manual (Model 416 or 417) for procedures.

EQUIPMENT ALERT
Do not extend or retract actu-
*4 1/2" ator past trip points of limit
(11.4 cm)
Extended switches when installing.
Limit Switch
(C/NC) Closed

*2.0"
(5.0) cm Retracted
Limit Switch
(C/NC) Closed
*8 5/16"
(21.1 cm)

NOTE*
Dimensions are preset at the
factory and approximate to
within 1/16" (0.16cm)

EQUIPMENT ALERT
Do not extend or retract actu-
ator past trip points of limit
switches when installing.

Figure 1 MA613500i

© Midmark Corporation - 2002 003-1334-00 (Rev. C) Page 1 of 2


Wire Connections (3) Install switch connector from black harness to
Retract (Down) limit switch.
(Refer to Figure 2)
(4) Install switch connector from red harness to
WARNING Extend (Up) limit switch.
Be sure supply power is off and capaci- (5) Connect black and red motor leads to black and
tor on actuator has been discharged red leads of harnesses that attach to Capacitor
before continuing to avoid personal injury. and common (C) terminal on actuator limit
switches.
(1) Turn off the power and discharge the capacitor
(6) Connect white motor lead terminal to white lead
then, cut terminal connections off leads that
coupler of P.C. board harness.
connected to capacitor.
Do not strip leads but install wire nuts (pro- (7) Connect red lead from P.C. board harness to
vided) then tape to secure. red lead from normally closed (NC) position on
Extend (Up) limit switch.
NOTE
(8) Connect black lead from P.C. board harness to
Terminal end of capacitor must be facing toward actu-
black lead from normally closed (NC) position
ator when positioned in mounting bracket.
on Retract (Down) limit switch.
(2) Connect flag terminal of red and black harness (9) Secure leads and harness to actuator and
lead to capacitor terminals, install cap and posi- motor with wire ties (supplied).
tion capacitor in mounting bracket.
(10) Run unit to check operation.

4 1
Red 8
Red Blk

Red 7
Wht
Blk

Red Red
Extended Red
Limit
Switch
Figure 2
Blk
Retracted
Red Limit
Switch
3

Wht

Blk

Red

Blk Wht
Actuator
Red 6

Blk
Capacitor
Wht
Red

5
Blk
MA613600i

© Midmark Corporation - 2002 Page 2 of 2