Installation
Kit Components (1) Check dimensions and Limit Switch operation
Part Number Description Qty. of the Actuator (Refer to Figure 1).
002-0727-00 Actuator Kit
• 029-2980-00 Actuator w/ Limit Switches ..........1 EQUIPMENT ALERT
• 015-1656-00 Harness (Up Limit, Red Wires) ...1 Assure Extend Limit Switch normally
• 015-1657-00 Harness (Dwn. Limit, Blk. Wires) 1 closed contact (C-NC) is in the OPEN
• 015-1113-00 Wire Nuts ....................................2 position when installing actuator onto table.
• 015-0013-01 Wire Tie (14 1/2" long) ................2 (2) Set the actuator so the contacts (C-N/C) on the
• 003-1334-00 Installation Instructions................1 Extend limit switch just trip to the OPEN posi-
tion.
Removal
(1) Remove the Tilt and/or Back Actuator and EQUIPMENT ALERT
Capacitor, refer to the appropriate Service and When installing the Actuator per instruc-
Parts Manual (Model 416 or 417) for proce- tions in the manual do not connect the
dures. table electrical leads and run the actuator unless they
have been changed to match those in Figure 2
(3) Install the Tilt and/or Back Actuator and Capac-
itor, refer to the appropriate Service and Parts
Manual (Model 416 or 417) for procedures.
EQUIPMENT ALERT
Do not extend or retract actu-
*4 1/2" ator past trip points of limit
(11.4 cm)
Extended switches when installing.
Limit Switch
(C/NC) Closed
*2.0"
(5.0) cm Retracted
Limit Switch
(C/NC) Closed
*8 5/16"
(21.1 cm)
NOTE*
Dimensions are preset at the
factory and approximate to
within 1/16" (0.16cm)
EQUIPMENT ALERT
Do not extend or retract actu-
ator past trip points of limit
switches when installing.
Figure 1 MA613500i
4 1
Red 8
Red Blk
Red 7
Wht
Blk
Red Red
Extended Red
Limit
Switch
Figure 2
Blk
Retracted
Red Limit
Switch
3
Wht
Blk
Red
Blk Wht
Actuator
Red 6
Blk
Capacitor
Wht
Red
5
Blk
MA613600i