Standard Practices for Force Calibration and Verification of Testing

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 14

This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles

for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

Designation: E4 − 21 American Association State


Highway and Transportation Officials Standards
AASHTO No: T67

Standard Practices for


Force Calibration and Verification of Testing Machines1
This standard is issued under the fixed designation E4; the number immediately following the designation indicates the year of original
adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A superscript
epsilon (´) indicates an editorial change since the last revision or reapproval.
This standard has been approved for use by agencies of the U.S. Department of Defense.

1. Scope* retransmitted—which are verified with provisions of 1.2.1,


1.1 These practices cover procedures for the force calibra- 1.2.2, or 1.2.3, and are within the specifications stated in
tion and verification, by means of force measurement Section 15, comply with Practices E4.
standards, of tension or compression, or both, static or quasi- 1.6 The requirements of these practices limit the major
static testing machines (which may, or may not, have force- components of measurement uncertainty when calibrating
indicators). These practices are not intended to be complete testing machines. These Standard Practices do not require the
purchase specifications for testing machines. allowable force measurement error to be reduced by the
1.2 Testing machines may be verified by one of the three amount of the measurement uncertainty encountered during a
following methods or combination thereof. Each of the meth- calibration. As a result, a testing machine verified using these
ods require a specific measurement uncertainty, displaying practices may produce a deviation from the true force greater
metrological traceability to The International System of Units than 61.0 % when the force measurement error is combined
(SI). with the measurement uncertainty.
1.2.1 Use of standard weights, 1.7 This standard does not purport to address all of the
1.2.2 Use of equal-arm balances and standard weights, or safety concerns, if any, associated with its use. It is the
1.2.3 Use of elastic force measurement standards. responsibility of the user of this standard to establish appro-
priate safety, health, and environmental practices and deter-
1.3 The procedures of 1.2.1–1.2.3 apply to the calibration
and verification of the force-measuring systems associated with mine the applicability of regulatory limitations prior to use.
the testing machine, including the force indicators such as a 1.8 This international standard was developed in accor-
scale, dial, marked or unmarked recorder chart, digital display, dance with internationally recognized principles on standard-
etc. In all cases the buyer/owner/user must designate the ization established in the Decision on Principles for the
force-measuring system(s) to be verified and included in the Development of International Standards, Guides and Recom-
certificate and report of calibration and verification. mendations issued by the World Trade Organization Technical
Barriers to Trade (TBT) Committee.
1.4 Units—The values stated in either SI units or inch-
pound units are to be regarded separately as standard. The 2. Referenced Documents
values stated in each system are not necessarily exact equiva- 2.1 ASTM Standards:2
lents; therefore, to ensure conformance with the standard, each E6 Terminology Relating to Methods of Mechanical Testing
system shall be used independently of the other, and values E74 Practices for Calibration and Verification for Force-
from the two systems shall not be combined. Measuring Instruments
1.4.1 Other non-SI force units may be used with this E467 Practice for Verification of Constant Amplitude Dy-
standard such as the kilogram-force (kgf) which is often used namic Forces in an Axial Fatigue Testing System
with hardness testing machines 2.2 BIPM Standard:3
1.5 Forces indicated on displays/printouts of testing ma- JCGM 100 : Evaluation of measurement data - Guide to the
chine data systems—be they instantaneous, delayed, stored, or Expression of Uncertainty in Measurement.

2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
1
These practices are under the jurisdiction of ASTM Committee E28 on contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
Mechanical Testing and is the direct responsibility of Subcommittee E28.01 on Standards volume information, refer to the standard’s Document Summary page on
Calibration of Mechanical Testing Machines and Apparatus. the ASTM website.
3
Current edition approved June 1, 2021. Published August 2021. Originally Available from BIPM - Pavillon de Breteuil F-92312 Sèvres Cedex FRANCE.
approved in 1923. Last previous edition approved in 2020 as E4 – 20. DOI: this document is available free-of-charge at https://www.bipm.org/en/publications/
10.1520/E0004-21. guides/vim.html

*A Summary of Changes section appears at the end of this standard


Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States

North Carolina A , Thu Aug 26 06:10:58 GMT 2021


E4 − 21
JCGM 200 : International vocabulary of metrology — Basic 3.2.6.1 Discussion—A force measurement standard is a
and general concepts and associated terms (VIM). specific type of “measurement standard” as defined in JCGM
200: International vocabulary of metrology — Basic and
3. Terminology general concepts and associated terms (VIM).
3.1 For definitions of terms used in this practice, refer to 3.2.7 force-measuring system, n—of a testing machine, a
Terminology E6. component of a testing machine that measures and indicates the
3.2 Definitions: force applied by the testing machine.
3.2.1 calibration, n—operation that, under specified 3.2.8 force-sensing device, n—of a testing machine, a com-
conditions, in a first step, establishes a relation between the ponent of the force-measuring system, that measures through
quantity values with measurement uncertainties provided by deformation or other means the force applied by the testing
measurement standards and corresponding indications with machine.
associated measurement uncertainties and, in a second step, 3.2.8.1 Discussion— Examples of a force-sensing device
uses this information to establish a relation for obtaining a include a strain-gage force transducer (commonly called a load
measurement result from an indication. cell) and a pressure transducer.
3.2.1.1 Discussion—A calibration may be expressed by a 3.2.9 measurement accuracy, n—closeness of agreement
statement, calibration function, calibration diagram, calibration between a measured quantity value and a true quantity value of
curve, or calibration table. In some cases, it may consist of an a measurand.
additive or multiplicative correction of the indication with
associated measurement uncertainty. 3.2.9.1 Discussion—The concept “measurement accuracy”
3.2.1.2 Discussion—Calibration should not be confused is not a quantity and is not given a numerical quantity value. A
with adjustment of a measuring system, often mistakenly measurement is said to be more accurate when it offers a
called “self-calibration”, nor with verification of calibration. smaller measurement error.
3.2.1.3 Discussion—Often, the first step alone in the above 3.2.9.2 Discussion—The term “measurement accuracy”
definition is perceived as being calibration. JCGM 200:2012 4 should not be used for measurement trueness and the term
“measurement precision” should not be used for ‘measurement
3.2.2 elastic force measurement standard, n—system con-
accuracy’, which, however, is related to both these concepts.
sisting of an elastic member combined with an appropriate
3.2.9.3 Discussion—“Measurement accuracy” is sometimes
device for indicating the magnitude (or a quantity proportional
understood as closeness of agreement between measured
to the magnitude) of deformation of the member under an
quantity values that are being attributed to the measurand.
applied force.
JCGM 200:2012 4
3.2.3 exercise, v—apply the maximum force to be used in 3.2.10 metrological traceability, n—property of a measure-
the calibration to either an elastic force measurement standard ment result whereby the result can be related to a reference
or the force-sensing device of a testing machine, or to both, to through a documented unbroken chain of calibrations, each
reestablish the hysteresis pattern that tends to disappear during contributing to the measurement uncertainty.
periods of disuse, or with the change of mode of force
application, as from compression to tension. 3.2.10.1 Discussion—For this definition, a “reference” can
be a definition of a measurement unit through its practical
3.2.4 force indicator, n—of a testing machine, a component
realization, or a measurement procedure including the mea-
of a force-measuring system that presents, in force units, the
surement unit for a non-ordinal quantity, or a measurement
force measured by the force-measuring system.
standard.
3.2.5 force measurement error, E, n—in the case of a testing 3.2.10.2 Discussion—Metrological traceability requires an
machine, the difference obtained by subtracting the force established calibration hierarchy.
indicated by the force measurement standard from the indi- 3.2.10.3 Discussion—Specification of the reference must
cated force of the testing machine. include the time at which this reference was used in establish-
3.2.5.1 Discussion—In a certificate and report of calibration ing the calibration hierarchy, along with any other relevant
and verification, “force measurement error” shall be used with metrological information about the reference, such as when the
numerical values, for example, “At a force of 300 kN first calibration in the calibration hierarchy was performed.
[60 000 lbf], the force measurement error of the testing ma- 3.2.10.4 Discussion—For measurements with more than one
chine was + 67 N [+ 15 lbf].” input quantity in the measurement model, each of the input
3.2.6 force measurement standard, n—a standard weight, an quantity values should itself be metrologically traceable and
equal-arm balance and a standard weight, or an elastic force the calibration hierarchy involved may form a branched
measurement standard used as a reference, with associated structure or a network. The effort involved in establishing
measurement uncertainty, in compliance with these practices metrological traceability for each input quantity value should
and Practices E74. be commensurate with its relative contribution to the measure-
ment result.
4
3.2.10.5 Discussion—Metrological traceability of a mea-
This definition is reproduced here from JCGM 200:2012 International vocabu-
surement result does not ensure that the measurement uncer-
lary of metrology – Basic and general concepts and associated terms (VIM) with
permission from the Director of BIPM. The text has been put in ASTM Interna- tainty is adequate for a given purpose or that there is an
tional’s form and style. absence of mistakes.

North Carolina A , Thu Aug 26 06:10:58 GMT 2021


E4 − 21
3.2.10.6 Discussion—A comparison between two measure- 3.3.3 force, n—in the case of testing machines, a force
ment standards may be viewed as a calibration if the compari- measured in units such as pound-force, newton, or kilogram-
son is used to check and, if necessary, correct the quantity force.
value and measurement uncertainty attributed to one of the 3.3.3.1 Discussion—The newton is that force which acting
measurement standards. JCGM 200:2012 4 on a 1-kg mass will give to it an acceleration of 1 m/s2. The
3.2.11 testing machine, n—force-measuring type, a me- pound-force is that force which acting on a [1-lb] mass will
chanical device for applying and measuring forces on a give to it an acceleration of 9.80665 m/s2 [32.1740 ft/s2]. The
specimen being tested. kilogram-force is that force which acting on a 1-kg mass will
3.2.11.1 Discussion—A testing machine generally consists give to it an acceleration of 9.80665 m/s2[32.1740 ft/s2].
of two components, a mechanism for applying forces to a 3.3.4 percent error of force EP, n—in the case of a testing
specimen being tested and a force-measuring system for machine, the ratio, of the force measurement error to the
measuring the applied forces. applied force as measured by the force measurement standard,
3.2.11.2 Discussion—Some testing machines do not have a expressed as a percent.
force indicator such as some creep testing machines which 3.3.4.1 Discussion—The indicated force of the testing
apply a force utilizing weights and a lever mechanism. machine, and the applied force, as measured by the force
3.2.12 verification, n—provision of objective evidence that measurement standard, shall be recorded at each calibration
a given item fulfils specified requirements. force. The force measurement error, E, and the percent error of
3.2.12.1 Discussion—EXAMPLE 1 Confirmation that a forces, EP, shall be calculated from these data as follows:
given reference material as claimed is homogeneous for the E5A2B (1)
quantity value and measurement procedure concerned, down to
a measurement portion having a mass of 10 mg. E p 5 @ ~ A 2 B ! /B # 3 100
3.2.12.2 Discussion—EXAMPLE 2 Confirmation that per- where:
formance properties or legal requirements of a measuring A = force indicated by the testing machine being verified, N
system are achieved. [or lbf, etc.], and
3.2.12.3 Discussion—EXAMPLE 3 Confirmation that a tar- B = value of the applied force, N [or lbf, etc.], as measured
get measurement uncertainty can be met. by the force measurement standard, in the same units as
3.2.12.4 Discussion—When applicable, measurement un- A.
certainty should be taken into consideration. 3.3.5 permissible variation, n—in the case of testing
3.2.12.5 Discussion—The item may be, for example, a machines, the maximum allowable force measurement error in
process, measurement procedure, material, compound, or mea- the value of the quantity indicated.
suring system. 3.3.5.1 Discussion—It is convenient to express permissible
3.2.12.6 Discussion—The specified requirements may be, variation in terms of percent error of force. The numerical
for example, that a manufacturer’s specifications are met. value of the permissible variation for a testing machine is so
3.2.12.7 Discussion—Verification in legal metrology, as de- stated hereafter in these practices.
fined in VIML5, and in conformity assessment in general, 3.3.6 resolution of the force-measuring system, n—smallest
pertains to the examination and marking and/or issuing of a change of force that can be estimated or ascertained on the
verification certificate for a measuring system. force indicating apparatus of the testing machine, at any
3.2.12.8 Discussion—Verification should not be confused applied force.
with calibration. Not every verification is a validation. 3.3.6.1 Discussion—Appendix X1 describes a method for
3.2.12.9 Discussion—In chemistry, verification of the iden- determining resolution.
tity of the entity involved, or of activity, requires a description
3.3.7 resolution of analog force-measuring systems (scales,
of the structure or properties of that entity or activity. JCGM
dials, recorders, etc.), n—the resolution is the smallest change
200:2012 4
in force indicated by a displacement of a pointer, or pen line.
3.3 Definitions of Terms Specific to This Standard: 3.3.7.1 Discussion—The resolution is calculated by multi-
3.3.1 calibration force, n—A force selected where the indi- plying the force corresponding to one graduation by the ratio of
cated force of the testing machine is compared with the applied the width of the pointer or pen line to the center-to-center
force as indicated by the force measurement standard. distance between two adjacent graduation marks. The typical
ratios used are 1:1, 1:2, 1:5, or 1:10. A spacing of 2.5 mm [0.10
3.3.1.1 Discussion—Calibration forces shall be selected in
in.] or greater is recommended for the ratio of 1:10. A ratio less
accordance with these Practices E4, see Section 11.
than 1:10 should not be used.
3.3.2 capacity range, n—in the case of testing machines, the
3.3.7.2 Discussion—If a force indicating dial has gradua-
range of forces for which it is designed.
tions spaced every 2.0 mm [0.080 in.], the width of the pointer
3.3.2.1 Discussion—Some testing machines have more than is approximately 1.0 mm (0.040 in.), and one graduation
one capacity range, that is, multiple ranges. represent 25N [5 lbf]. The ratio used would be 1:2 and the
resolution would be equal to 12-1⁄2 N [2-1⁄2 lbf].
3.3.7.3 Discussion—If the indicated force fluctuates by
5
OIML, International Vocabulary of Terms in Legal Metrology (VIML). more than twice the resolution, as described in 3.3.7, the

North Carolina A , Thu Aug 26 06:10:58 GMT 2021


E4 − 21
resolution, expressed as a force, shall be equal to one-half the research laboratory to measure material properties, or in a
range of the fluctuation. production line to qualify a product for shipment. No matter
3.3.8 resolution of digital force measuring systems what the end use of the testing machine may be, it is necessary
(numeric, displays, printouts, etc.), n—the resolution is the for users to know that the amount of force applied and
smallest change in force that can be displayed on the force indicated is traceable to the International System of Units (SI)
indicator, at any applied force. through a National Metrology Institute (NMI). The procedures
3.3.8.1 Discussion—A single digit or a combination of in Practices E4 may be used to calibrate these testing machines
digits may be the smallest change in force that can be so that the measured forces are traceable to the SI. A key
indicated. element of traceability to the SI is that the force measurement
3.3.8.2 Discussion—If the indicated force fluctuates by standards used in the calibration have known force
more than twice the resolution, as described in 3.3.8, the characteristics, and have been calibrated in accordance with
resolution, expressed as a force, shall be equal to one-half the Practice E74.
range of the fluctuation. 5.2 The procedures in Practices E4 may be used by those
3.3.9 verified range of forces, n—in the case of testing using, manufacturing, and providing calibration service for
machines, the range of indicated forces for which the testing testing machines and related instrumentation.
machine gives results within the permissible variations speci-
fied. 6. Elastic Force Measurement Standards
3.3.9.1 Discussion—This term is also defined in Practice
E74 and has a different meaning. If the term "verified range of 6.1 When calibrating testing machines, elastic force mea-
forces" is preceded by "Class A", the Practices E 74 definition surement standards shall be only used within their Class A
shall apply. verified range of forces as determined by Practice E74.

4. Summary of Practice 7. Advantages and Limitations of Methods


4.1 Practices E4 calibration consists of comparing the indi-
7.1 Calibration by Standard Weights—Calibration by the
cated force of the testing machine (or the testing machine’s
direct application of standard weights to the weighing mecha-
applied force in the case of testing machines that do not have
nism of the testing machine, where practicable, is the most
force indicators) to a force measurement standard at various
accurate method. Its limitations are: (1) the small range of
calibration forces. These comparisons are used to establish the
forces that can be calibrated, (2) the nonportability of any large
force measurement error at each calibration force at least two
amount of standards weights, and (3) its nonapplicability to
times. The metrological requirements of these Practices E4
horizontal testing machines or vertical testing machines having
intrinsically account for measurement uncertainty by limiting
weighing mechanisms that are not designed to be actuated by
the major contributions to measurement uncertainty such as
a downward force.
requirements for the force measurement standard used,
resolution, repeatability, and measurement accuracy. As a 7.2 Calibration by Equal-Arm Balance and Standard
result, the Practices E4 calibration and verification procedure Weights—The second method of calibration of testing ma-
along with the certificate and report of calibration and verifi- chines involves measurement of the force by means of an
cation provide metrological traceability to the SI for the equal-arm balance and standard weights. This method is
force-measuring system of the testing machine. limited to a still smaller range of forces than the foregoing
4.1.1 Although Practices E4 do not require reporting mea- method and is generally applicable only to certain types of
surement uncertainty of the calibration, it may be calculated hardness testing machines in which the force is applied through
and included in the certificate and report of calibration and an internal lever system.
verification.
7.3 Calibration by Elastic Force Measurement Standards—
4.2 Practices E4 verification consists of using the force The third method of calibration of testing machines involves
measurement errors determined along with resolution and measurement of the elastic strain or deflection under force of a
return-to-zero readings as evidence that the force indicator(s) ring, loop, tension or compression bar, or other elastic force
of a testing machine indicates values, or that the testing measurement standard. The elastic force measurement standard
machine applies forces, that meet the requirements of these is free from the limitations referred to in 7.1 and 7.2.
Practices E4 in terms of percent error of force, repeatability,
resolution, and return-to-zero at the calibration forces directed 8. System Calibration
by these Practices E4.
4.3 If the force-measuring system of the testing machine 8.1 A testing machine shall be calibrated and verified as a
fails to meet any of these requirements and is adjusted, a full system with the force-sensing device and force indicator (see
calibration and verification in accordance with these Practices 1.3 and 1.5) in place and operating as in actual use.
E4 shall be conducted after the adjustment is made. 8.1.1 If this is not technically possible, refer to Annex A1,
Calibrating the Force-Measuring System out of the Test Ma-
5. Significance and Use chine. Out of the testing machine calibrations shall be in
5.1 Testing machines that apply and indicate force are used accordance with the main body of Practices E4 and its Annex
in many industries, in many ways. They might be used in a A1.

North Carolina A , Thu Aug 26 06:10:58 GMT 2021


E4 − 21
8.2 System calibration and verification is invalid if the or balance-arm systems, correct the force for the local value of
devices are removed and checked independently of the testing gravity and for nominal air buoyancy.
machine unless calibration is performed according to Annex 9.1.1 The force exerted by a weight in air is obtained by:
A1.
8.3 Many testing machines are designed to be able to F 5 M 3g 1 2S d
D D (2)
interchange force-sensing devices (usually these are force
where:
transducers commonly called load cells). Usually these force-
sensing devices vary in capacity range. Lower capacity range F = Force, N
force-sensing devices are used for better resolution and accu- M = true mass of the weight, kg
g = local acceleration due to gravity, m/s2,
racy at lower test forces and higher capacity range force-
d = air density (1.2 kg/m3), and
sensing devices are used to apply and measure higher forces. D = density of the weight in the same units as d.
During use of a testing machine of this type, the force-sensing
devices may be routinely installed and uninstalled, which 9.1.2 For the purposes of this standard, g can be calculated
effectively creates multiple force-measuring systems. For such with a sufficient uncertainty using the following formula.
force-sensing devices, interchangeability shall be established g 5 9.7803@ 1 1 0.0053 ~ sin [ ! 2 # 2 0.000001967h (3)
during the original calibration and shall be reestablished after
an adjustment is performed. This is accomplished by perform- where:
ing a Practices E4 calibration with the force-sensing device in Ø = latitude
place as during normal use. It is advisable that orientation be h = elevation above sea level in metres
kept consistent, such as by noting the direction of the cable NOTE 1—Eq 3 corrects for the shape of the earth and the elevation
connector so that when reinstalling the force-sensing device, above sea level. The first term, which corrects for the shape of the earth,
is a simplification of the World Geodetic System 84 Ellipsoidal Gravity
the orientation will be repeated. Remove and reinstall the
Formula. The results obtained with the simplified formula differ from
force-sensing device between the two calibration runs to those in the full version by less than 0.0005%. The second term combines
demonstrate interchangeability. Repeat the procedure for each a correction for altitude, the increased distance from the center of the
interchangeable force-sensing device used in the testing ma- earth, and a correction for the counter-acting Bouguer effect of localized
chine. increased mass of the earth. The second term assumes a rock density of
2.67 g/cm3. If the rock density changed by 0.5 g/cm3, an error of 0.003 %
8.3.1 Some testing machines are designed with multiple would result.
force-sensing devices permanently mounted usually with dif-
ferent test areas for each force sensing device. Section 8.3 does 9.2 In some cases, a mass might not be designated in
not apply to such testing machines unless the force-sensing kilograms, for instance it might be denoted in pounds and it
devices are interchanged as described in 8.3. might be desired to know the force exerted in pound-force
8.3.2 Introduction of a new interchangeable force sensing units. In other cases, it might be desired to know the force
device(s) shall require that interchangeability be established exerted in kilogramforce units where the mass is designated in
per 8.3. kilograms. In these cases, the force in non–SI units exerted by
a weight in air is calculated as follows:
8.4 A Practices E4 calibration consists of at least two
calibration runs of the calibration forces selected in the verified
range(s) of forces. See 11.1 to 11.3.
Fc 5
M 3g
9.80665S1 2
d
D D (4)
where:
8.4.1 If the initial calibration run produces values within the
Practices E4 requirements of Section 15, the data may be used where:
“as found” for calibration run one of the two required for the Fc = force expressed in non-SI units, such as, pound
new certificate and report of calibration and verification. force or kilogram-force,
8.4.2 If the initial calibration run produces any values which M = true mass of the weight, in the corresponding
are outside of the Practices E4 requirements, the “as found” mass units of the, Fc is being expressed, such as,
data may be reported and may be used in accordance with pound or kilogram,
applicable quality control programs. Calibration adjustments g = local acceleration due to gravity, m/s2,
shall be made to the force-measuring system(s), after which the d = air density (1.2 kg/m3),
two required calibration runs shall be conducted and reported D = density of the weight in the same units as d, and
9.80665 = the factor converting SI units of force into non-SI
in the new certificate and report of calibration and verification.
units of force; this factor is equal to the value for
8.4.3 Calibration adjustments may be made to improve the standard gravity, 9.80665 m/s2.
measurement accuracy of the system. They shall be followed
by the two required calibration runs, and issuance of a new If M, the mass of the weight is in pounds, the force will be
certificate and report of calibration and verification and certifi- in pound-force units [lbf]. If M is in kilograms, the force will
cate. be in kilogram-force units (kgf). These non-SI force units are
related to the newton (N), the SI unit of force, by the following
9. Gravity and Air Buoyancy Corrections relationships:
9.1 In the calibration of testing machines, where standard 1 lbf 5 4.448222N (5)
weights are used for applying forces directly or through lever 1 kgf =9.80665 N ~exact! (6)

North Carolina A , Thu Aug 26 06:10:58 GMT 2021


E4 − 21
9.2.1 For use in calibrating testing machines, corrections for force to any given calibration force and then by decreasing the force to
local values of gravity and air buoyancy to standard weights that calibration force, might not agree. Testing machines are usually used
under increasing forces, but if a testing machine is to be used under
calibrated in pounds can be made with sufficient precision decreasing forces, it should be calibrated under decreasing forces as well
using the multiplying factors from Table 1. Alternatively, the as under increasing forces.
following formula may be used to find the multiplying factor,
10.2 Testing machines that contain a single test area and
MF. Multiply MF times the mass of the weight given in pounds
possess a bidirectional loading and weighing system must be
to obtain the value of force in pounds-force, corrected for local
verified separately in both modes of weighing.
gravity and air buoyancy.
10.3 High-speed testing machines used for static testing
9.7803@ 1 1 0.0053 ~ sin [ ! 2 # 2 0.000001967h
MF 5 3 0.99985 must be verified in accordance with Practices E4. Warning—
9.80665
Practices E4 calibration values are not to be assumed valid for
(7) high-speed or dynamic testing applications (see Practice
where: E467).
Ø = latitude NOTE 4—The force measurement error of a testing machine of the
h = elevation above sea level in metres hydraulic-ram type, in which the ram hydraulic pressure is measured,
NOTE 2—Eq 7 and Table 1 correct for the shape of the earth, elevation might vary significantly with ram position. To the extent possible such
above sea level, and air buoyancy. The correction for the shape of the earth testing machines should be verified at the ram positions used.
is a simplification of the World Geodetic System 84 Ellipsoidal Gravity
Formula. The results obtained with the simplified formula differ by less 11. Selection of Calibration Forces
than 0.0005 %. The term that corrects for altitude, corrects for an
increased distance from the center of the earth and the counter-acting 11.1 Determine the upper and lower limits of the verified
Bouguer effect of localized increased mass of the earth. The formula range of forces of the testing machine to be verified. All
assumes a rock density of 2.67 g/cc. If the rock density changed by 0.5 calibration forces in the verified range of forces shall be at least
g/cc, an error of 0.003 % would result. The largest inaccuracy to be 200 times larger than the resolution of the force-measuring
expected, due to extremes in air pressure, temperature, and humidity when
using steel weights, is approximately 0.01 %. If aluminum weights are system at that calibration force.
used, errors on the order of 0.03 % can result. 11.2 If the lower limit of the verified range of forces is
9.3 Standard weights are typically denominated in a unit of greater than or equal to one-tenth of the upper limit, five or
mass. When a standard weight has been calibrated such that it more different calibration forces shall be selected such that the
exerts a specific force under prescribed conditions, the weight difference between two adjacent calibration forces is greater
will exert that force only under those conditions. When used in than or equal to one twentieth and less than or equal to
locations where the acceleration of gravity differs from the one one-third the difference between the upper and lower limits of
in the calibration location, it is necessary to correct the the verified range of forces. One calibration force shall be the
calibrated force value by multiplying the force value by the lower limit of the verified range of forces and another
value for local gravity and dividing by the value of gravity for calibration force shall be the upper limit. (Fewer calibration
which the weight was calibrated. Any required air buoyancy forces are required for testing machines designed to measure
corrections must also be taken into account. only a small number of discrete forces, such as certain hardness
testing machines, creep testing machines, etc.)
10. Application of Force
11.3 If the lower limit of the verified range of forces, is less
10.1 In the calibration of a testing machine, approach the than one-tenth the upper limit, calibration forces shall be
calibration force by increasing the force from a lower force. selected as follows:
NOTE 3—For any testing machine the force measurement errors 11.3.1 Starting with the lower limit of the verified range of
observed at corresponding calibration forces taken first by increasing the forces, establish overlapping force decades such that the

TABLE 1 Multiplying Factor, MF, in Air at Various Latitudes, see Eq 7


Elevation Above Sea Level, h, m (ft)
Latitude, Ø,° 0 500 1000 1500 2000 2500
(0) (1640) (3280) (4920) (6560) (8200)
0 0.9972 0.9971 0.9970 0.9969 0.9968 0.9967
5 0.9972 0.9971 0.9970 0.9969 0.9968 0.9967
10 0.9973 0.9972 0.9971 0.9970 0.9969 0.9968
15 0.9975 0.9974 0.9973 0.9972 0.9971 0.9970
20 0.9978 0.9977 0.9976 0.9975 0.9974 0.9973
25 0.9981 0.9980 0.9979 0.9978 0.9977 0.9976
30 0.9985 0.9984 0.9983 0.9982 0.9981 0.9980
35 0.9989 0.9988 0.9987 0.9986 0.9985 0.9984
40 0.9993 0.9992 0.9991 0.9990 0.9989 0.9988
45 0.9998 0.9997 0.9996 0.9995 0.9994 0.9993
50 1.0003 1.0002 1.0001 1.0000 0.9999 0.9998
55 1.0007 1.0006 1.0005 1.0004 1.0003 1.0002
60 1.0011 1.0010 1.0009 1.0008 1.0007 1.0006
65 1.0015 1.0014 1.0013 1.0012 1.0011 1.0010
70 1.0018 1.0017 1.0016 1.0015 1.0014 1.0013

North Carolina A , Thu Aug 26 06:10:58 GMT 2021


E4 − 21
maximum calibration force in each decade is ten times the machine or on trays or other supports suspended from the
lowest calibration force in the decade. The lowest calibration force-sensing device in place of the specimen. Use standard
force in the next higher decade is the same as the highest weights certified within five years to be accurate within 0.1 %.
calibration force in the previous decade. The highest decade Apply the standard weights in ascending increments. If data is
might not be a complete decade. to be taken in both ascending and descending directions,
11.3.2 Five or more different calibration forces shall be remove the standard weights in reverse order. Record the
selected per decade such that the difference between two forces, corrected for gravity and air buoyancy in accordance
adjacent calibration forces is greater than or equal to one- with Section 9.
twentieth and less than or equal to one-third the difference
NOTE 7—The method of calibration by direct application of standard
between the maximum and the minimum calibration force in weights can be used only on vertical testing machines in which the force
that decade. It is recommended that starting with the lowest on the weighing table, hydraulic support, or other weighing device is
calibration force in each decade, the ratios of the calibration downward. The total force is limited by the size of the platform and the
forces to the lowest calibration force in the decade are 1:1, 2:1, number of standard weights available. Twenty-five kg or [fifty lb] standard
weights are usually convenient to use. This method of calibration is
4:1, 7:1, 10:1 or 1:1, 2.5:1, 5:1, 7.5:1, 10:1. confined to small testing machines and is rarely used above 5000 N [1000
11.3.3 If the highest decade is not a complete decade, lbf].
choose calibration forces at the possible ratios and include the
13.2 Method B. Calibration of Hardness Testing Machines
upper limit of the verified range of forces. If the difference
by Equal-Arm Balance and Standard Weights:
between two adjacent calibration forces is greater than one-
13.2.1 Procedure:
third of the upper limit, add an additional calibration force.
13.2.1.1 Position the balance so that the indenter of the
NOTE 5—Example: A testing machine has a full-scale range of 5000 N testing machine being calibrated bears against a block centered
and the resolution of the force-measuring system is 0.0472 N. The lowest on one pan of the equal-arm balance, the balance being in its
possible calibration force is 9.44 N (0.0472 N × 200). Instead of decades
starting at 9.44 N, 94.4 N and 944 N, three decades, starting at 10 N, equilibrium position when the indenter is in that portion of its
100 N, and 1000 N are selected to cover the verified range of forces. travel normally occupied when making an impression. Place
Suitable calibration forces are 10 N, 20 N, 40 N, 70 N, 100 N, 200 N, standard weights complying with the requirements of Section
400 N, 700 N, 1000 N, 2000 N, 3000 N, 4000 N, 5000 N. Note that the 13 on the opposite pan to balance the force exerted by the
uppermost decade is not a complete decade and is terminated with the indenter.
upper limit of the verified range of forces. The 3000 N calibration force
was added because the difference between 2000 N and 4000 N was greater NOTE 8—This method can be used for the calibration of testing
than one-third of 5000 N. If the alternative distribution of forces is used, machines other than hardness-testing machines by positioning the force-
the calibration forces selected would be 10 N, 25 N, 50 N, 75 N, 100 N, applying member of the testing machine in the same way that the indenter
250 N, 500 N, 750 N, 1000 N, 2500 N, 3750 N, 5000 N. of a hardness-testing machine is positioned. For other methods of
11.4 All selected calibration forces shall be applied twice calibrating hardness testing machines see the applicable ASTM test
method.
during the procedure. Applied calibration forces on the second
calibration run are to be approximately the same as those on the 13.2.1.2 Since the permissible travel of the indenter of a
first calibration run. hardness-testing machine is usually very small, do not allow
the balance to oscillate or swing. Instead, maintain the balance
11.5 Approximately 30 s after removing the maximum force
in its equilibrium position through the use of an indicator such
in a range, record the return-to-zero reading of the force-
as an electric contact, which shall be arranged to indicate when
measuring system. The absolute value of the return-to-zero
the reaction of the indenter force is sufficient to lift the pan
reading shall be less than or equal to the greater of the absolute
containing the standard weights.
value of 0.1 % of the maximum force just applied or the
13.2.1.3 Using combinations of fractional standard weights,
absolute value of 1 % of the lowest calibration force in the
determine both the maximum value of the dead-weight force
verified range of forces.
that can be lifted by the testing machine indenter force during
12. Eccentricity of Force each of ten successive trials, and the minimum value that
cannot be lifted during any one of ten successive trials. Take
12.1 For the purpose of determining the verified range of the value of the indenting force as the average of these two
forces of the testing machine, apply all calibration forces so values. The difference between the two values shall not exceed
that the resultant force is as nearly along the axis of a testing 0.5 % of the average value.
machine as is possible.
13.3 Method C. Calibration by Elastic Force Measurement
NOTE 6—The effect of eccentric force on the measurement accuracy of Standard:
a testing machine can be determined by calibration readings taken with
force measurement standards placed so that the resultant force is applied 13.3.1 Temperature Equalization:
at definite distances from the axis of the testing machine, and the verified 13.3.1.1 When using an elastic force measurement standard
range of forces determined for a series of eccentricities. to calibrate the force-measuring system of a testing machine,
place the elastic force measurement standard near to, or
13. Methods of Calibration preferably in, the testing machine a sufficient length of time
13.1 Method A, Calibration by Standard Weights: before the calibration to ensure that the response of the elastic
13.1.1 Procedure: force measurement standard is stable.
13.1.1.1 Place standard weights of suitable design, finish, 13.3.1.2 During the calibration, measure the temperature of
and adjustment on the weighing platform of the testing the elastic force measurement standards within 61 °C [62 °F

North Carolina A , Thu Aug 26 06:10:58 GMT 2021


E4 − 21
] by placing a calibrated thermometer as close to the elastic 13.3.2.10 Record the indicated force of the testing machine
force measurement standard as possible. and the applied force from the elastic force measurement
13.3.1.3 Elastic force measurement standards not having an standard (temperature corrected as necessary), as well as the
inherent temperature-compensating feature must be corrected force measurement error and percent error of forces calculated
mathematically for the difference between ambient temperature from the readings.
and the temperature to which its calibration is referenced. 13.3.2.11 Under certain conditions, setups comprising mul-
Temperature-correction coefficients should be furnished (if tiple elastic force measurement standards may be used in
applicable) by the manufacturer of the elastic force measure- compression loading. All elastic force measurement standards
ment standard. Refer to Practice E74 for further information. to be loaded in parallel should be the same height (shims may
13.3.2 Procedure: be used) and the testing machine’s load axis should be
13.3.2.1 Place the elastic force measurement standard in the coincidental with the force axis of the multiple elastic force
testing machine so that its center line coincides with the center measurement standards setup. This is necessary so that a net
line of the heads of the testing machine. Record the Practice moment is not applied to the testing machine loading member.
E74 Class A verification value which establishes the lowest Setups using multiple elastic force measurement standards are
limit, or force level, allowable for the elastic force measure- not recommended unless the use of a single elastic force
ment standard’s Class A verified range of forces (see Practice measurement standard is not practicable.
E74). Each elastic force measurement standard shall be used
only within its Class A verified range of forces and identified 14. Lever-Type Creep-Rupture Testing Machines
with the calibration forces for which it is used. 14.1 Lever-type creep-rupture testing machines, which do
13.3.2.2 Exercise the force-sensing device of the testing not have a force-indicator, may be verified using standard
machine by applying the maximum calibration force to the weights or elastic force measurement standard(s), or both.
testing machine and returning to zero force. Zero the force- Standard weights used for calibration should conform to the
measuring system of the testing machine and repeat. Repeat as requirements of 13.1. In using an elastic force measurement
necessary, allowing the force measuring system of the testing standard, the requirements of 13.3 shall be met as applicable.
machine sufficient time to achieve stability in zero-force
14.2 Procedure:
indication.
14.2.1 Place the elastic force measurement standard in the
13.3.2.3 Exercise each elastic force measurement standard
testing machine and adjust the counterbalance (if the machine
to be used during calibration by applying the maximum
is so equipped) to compensate for the weight of the elastic
calibration force to be used with each elastic force measure-
force measurement standard.
ment standard and returning to zero force. Zero or record zero
14.2.2 Connect the lower crosshead of the testing machine
of the elastic force measurement standard and repeat. Repeat as
to the elastic force measurement standard, and apply forces
necessary, allowing the elastic force measurement standard
using standard weights in increments conforming to the pro-
sufficient time to achieve stability in zero-force indication.
visions of 11.1.
13.3.2.4 Each time the mode of force application is changed
14.2.3 Since many lever-type creep-rupture testing ma-
during the calibration for, example, compression to tension, the
chines do not have a provision for adjustment of the lever ratio
force-sensing device of the testing machine and the elastic
or tare, or both, it may be necessary to determine the “best fit”
force measurement standard(s) shall be re-exercised as de-
straight line through the calibration data, using the least
scribed above.
squares method. By doing this, the actual lever ratio and tare of
13.3.2.5 There are two methods for using elastic force each testing machine can be determined, and thus reduce force
measurement standards: measurement errors due to small variations of lever ratios.
13.3.2.6 Follow-the-Force Method—The force on the elastic Maximum percent error of force shall not exceed the require-
force measurement standard is followed until the force reaches ments stated in 15.1.
the desired calibration force on the force indicator of the testing
machine. Record the force measured by the elastic force CALCULATION AND REPORT
measurement standard.
13.3.2.7 Set-the-Force Method—The desired calibration 15. Basis of Verification
force is preset on the elastic force measurement standard, and 15.1 The percent error of forces within the verified range of
the testing machine’s force indicator is read when the desired forces of the testing machine shall not exceed 61.0 %. The
calibration force on the elastic force measurement standard is algebraic difference between force measurement errors of two
achieved. applications of the same calibration force (repeatability) shall
13.3.2.8 After selecting suitable calibration forces, obtain not exceed 1.0 % (see 11.1 and 11.3).
zero readings for both the testing machine and the elastic force
measurement standards, and apply forces slowly and smoothly 15.2 The verified range of forces stated shall not include
during all calibration measurements. forces outside the range of calibration forces applied during the
13.3.2.9 The calibration procedure must ensure that use of calibration.
the maximum force indicator, recorder, or other accessory 15.3 The Certificate and Report of Calibration and Verifi-
force devices does not cause testing machine percent error of cation shall state within what verified range of forces the
force to exceed the permissible variations of 15.1. testing machine may be used, rather than reporting a blanket

North Carolina A , Thu Aug 26 06:10:58 GMT 2021


E4 − 21
acceptance or rejection. For testing machines with multiple- of force be greater than 61.0 % at any of the spot check force
capacity ranges, the verified range of forces for each range levels, calibrate and verify the testing machine immediately
shall be stated. (see 17.3).
15.4 Testing machines may be more or less accurate than 18.5 Maintain a record of the spot check tests which shall
the allowable 61.0 % percent error of force, or more or less include the name, serial number, spot check date, calibration
repeatable than 1.0 %, which are the Practices E4 verification agency if applicable, and the minimum value of the Class A
bases. Buyers/owners/users or product specification groups verified range of forces, of the elastic force measurement
might require or allow larger or smaller permissible variations. standard(s) used to make spot checks; also include the name of
Testing machines with percent error of forces larger than person making the spot checks.
61.0 % or repeatability errors larger than 1.0 % do not comply 18.6 The testing machine shall be considered verified up to
with Practices E4. the date of the last successful spot check (see 18.4), provided
16. Corrections that the testing machine is calibrated and verified in accordance
with Section 17 on a regular schedule.
16.1 The indicated force of a testing machine that exceeds
the permissible variation shall not be corrected either by 18.7 When spot checks are made, a clear, concise record
calculation or by the use of a calibration diagram in order to must be maintained as agreed upon between the supplier and
obtain values within the required permissible variation. the user. The record must also contain documentation of the
regular calibration data and schedule.
17. Time Interval Between Verifications
17.1 It is recommended that testing machines be calibrated 19. Certificate and Report of Calibration and Verification
and verified annually or more frequently if required. In no case 19.1 Prepare clear, complete, and error-free documentation
shall that time interval exceed 18 months (except testing (no alteration of data, dates, etc.) for each calibration and
machines in which a long-time test runs beyond the 18-month verification of a testing machine which shall include the
period). In such cases, the testing machine shall be calibrated following:
and verified after completion of the test. 19.1.1 Name of the calibration agency,
17.2 Testing machines shall be calibrated and verified im- 19.1.2 Date of calibration and verification,
mediately after repairs (this includes new or replacement parts, 19.1.3 Testing machine description, serial number, and
or mechanical or electrical adjustments) that may in any way location,
affect the operation of the force-measuring system or the values 19.1.4 Statement identifying the force-measuring system(s)
displayed. that were verified,
17.2.1 Examples of new or replacement parts which may 19.1.5 Text identifying the mode of calibration, for
not affect the operation of the force-measuring system are example, tension, compression, or universal,
printers, computer monitors, keyboards, and modems. 19.1.6 Verified range(s) of forces of each force-measuring
17.3 Calibration and verification is required immediately system of the testing machine and the associated resolution(s),
after a testing machine is relocated (except for machines 19.1.7 Indicated force of the testing machine and the applied
designed to be moved from place to place in normal use), and force as measured by the force measurement standard for each
whenever there is a reason to doubt the measurement accuracy calibration run at each calibration force,
of the force-measuring system, regardless of the time interval 19.1.8 Return-to-zero reading after each calibration run, for
since the last verification. each capacity range,
19.1.9 Testing machine force measurement error, percent
18. Measurement Accuracy Assurance Between error of forces, and the percent difference between the calibra-
Verifications tion runs (repeatability) at each calibration force,
18.1 Some product-testing procedures may require daily, 19.1.10 Maximum percent error of force for each force
weekly, or monthly spot checks to ascertain that a testing range verified,
machine is capable of producing accurate force values between 19.1.11 The method of calibration used,
the testing machine verifications specified in Section 17. 19.1.12 Statement that the calibration and verification has
18.2 Spot checks may be performed on ranges of interest or been performed in accordance with Practice E4-XX. It is
at force levels of interest utilizing a force measurement recommended that the calibration and verification be per-
standard that complies with Methods A, B, or C as applicable. formed in accordance with the latest published issue of Practice
Elastic force measurement standards shall meet Class A E4,
requirements of Practice E74 for the force level(s) at which the 19.1.13 Manufacturer, serial number, calibration agency,
spot checks are made. calibration date, calibration recall date, and the limits of the
Class A verified range of forces in accordance with Practice
18.3 Make spot checks at approximately 20 % and 80 % of E74 of all elastic force measurement standards used for the
a range unless otherwise agreed upon or stipulated by the calibration,
material supplier/user. 19.1.14 Temperature of the elastic force measurement
18.4 Testing machine percent error of force shall not exceed standard(s) used for the calibration and a statement that
61.0 % of the spot check applied forces. Should percent errors computed forces have been temperature corrected as necessary,

North Carolina A , Thu Aug 26 06:10:58 GMT 2021


E4 − 21
19.1.15 Manufacturer, serial number, calibration agency, 19.2 Each certificate and report of calibration and verifica-
calibration date, and the calibration recall date of all standard tion generated by the calibration agency shall be uniquely
weights or weight sets used for the calibration, identified. Include page numbers, the total number of pages or
19.1.16 The identification of the individual who performed a mark to signify the end of the document in order to ensure
the calibration and verification, that the pages are recognized as part of the certificate and
19.1.17 The name and signature of the person responsible, report of calibration and verification.
in charge of the calibration and verification, and
19.1.18 Optionally or if required, a statement of the mea- 20. Keywords
surement uncertainty of the calibration, see Appendix X2. 20.1 calibration; force range; resolution; verification

ANNEX

(Mandatory Information)

A1. CALIBRATING THE FORCE-MEASURING SYSTEM OUT OF THE TESTING MACHINE

A1.1 Significance and Use range of forces in conjunction with a testing machine or
A1.1.1 The following are the recognized reasons to perform mechanism for applying force (see Practice E74). Two or more
a force-measuring system calibration out of the testing ma- elastic force measurement standards of equal compliance
chine: maybe combined and loaded in parallel to meet special needs
A1.1.1.1 There is inadequate spacing within the testing for higher capacities.
application load train to allow placement of a force measure- A1.3 Calibration
ment standard.
A1.1.1.2 It is physically impossible to apply a primary A1.3.1 Out of testing machine calibrations shall include the
deadweight force in the compression mode without removal of force sensing device, conditioning electronics, force indicators,
the force-measuring system. and cables.
A1.1.1.3 It is a test rig that has no reaction frame. A1.3.2 Use either Method A or Method C or both as detailed
A1.1.2 Calibrating the force-measuring system out of the in Section 13, but center the force measurement standard on the
testing machine represents an independent and singular mea- centerline of force-sensing device.
surement uncertainty component of the total testing machine
A1.3.3 A minimum of two calibration runs is required per
system measurement uncertainty. Other measurement uncer-
mode (compression or tension). Rotate the position of the
tainty components within the testing machine system exist and
force-sensing device by approximately 120 degrees before
need to be identified and quantified to determine, or verify, the
repeating any series of forces. During the calibration, ensure
testing machine total performance and level of measurement
that the loading axis is on the center load axis of the
uncertainty. For example, mounting considerations, fixtures,
force-application mechanism. Introduce variations or any other
hardness, stiffness, alignment, flatness, and bending may con-
factors that are normally encountered in service.
tribute to the measurement uncertainty of the testing machine.
A1.1.3 Fixture and environment considerations should be A1.3.4 Repeatability between the two calibration runs shall
made, to the best degree possible, to simulate the environment be less than or equal to 0.5 %. If greater than 0.5 %, an
within the testing application (for example, duplicating a additional third calibration run is required. The force-sensing
preload). shall be rotated by approximately 240 degrees from the starting
position prior to performing the third calibration run. The
A1.1.4 Calibrating the force-measuring system out of the repeatability between the three calibration runs shall be less
testing machine can be performed: than 1.0 %. Refer to A1.1.2 to consider all the measurement
A1.1.4.1 On-site, removed from the testing machine, con- uncertainty issues in determining the total testing machine
sisting of a complete force-measuring system (force-sensing measurement uncertainty.
device, conditioning electronics, force indicators, and cables).
A1.1.4.2 Off-site, removed from the testing machine, con- A1.3.5 The percent error of forces within the verified range
sisting of a complete force-measuring system (force-sensing of forces of the testing machine shall not exceed 6 1.0 %.
device, conditioning electronics, force indictor, and cables).
A1.4 Calculation and Report
A1.2 Force Measurement Standard A1.4.1 Calibration of the force-measuring system out of a
A1.2.1 The force-measuring system shall be calibrated by testing machine shall be clearly noted on the certificate and
force measurement standards used over their Class A verified report of calibration and verification.

10

North Carolina A , Thu Aug 26 06:10:58 GMT 2021


E4 − 21

APPENDIXES

(Nonmandatory Information)

X1. DETERMINING RESOLUTION OF THE FORCE-MEASURING SYSTEM

X1.1 The resolution of a testing machine in general is a X1.4.3 Next apply as constant a force as possible at the
complex function of many variables including applied force, force where the resolution is to be ascertained to ensure that the
force range, electrical and mechanical components, electrical indicated force does not fluctuate by more than twice the
and mechanical noise, and software employed, to name a few. resolution determined in the previous step. If the indicated
force fluctuates by more than twice the resolution, the resolu-
X1.2 A variety of methods may be used to check the tion shall be equal to one-half the range of the fluctuation.
resolution of the system. Some suggested procedures are as
follows. X1.5 Procedure for Auto-Ranging Digital Force Measuring
Systems:
X1.3 Procedure for Analog Force-Measuring Systems: X1.5.1 This procedure is the same as that for non-auto-
X1.3.1 Typically these analog force-measuring system are ranging digital force-measuring systems except that the reso-
not auto-ranging. The resolution should be checked at the lution is checked at the lowest calibration force in each decade
lowest calibration force in each force range (typically 10 % of or at other forces to ensure that the force-measuring system
full scale of the force range(s)). resolution is 200 times smaller than the forces. Some examples
X1.3.2 Divide the pointer width by the distance between are as follows.
two adjacent graduation marks at the force where the resolution X1.5.1.1 A 150 000 N capacity testing machine is to be
is to be ascertained to determine the pointer to graduation ratio. verified from 300 N up to 150 000 N. The resolution should be
If the distance between the two adjacent graduation marks is determined at 300 N, 3000 N, and 30 000 N.
less than 2.5 mm [0.10 in.] and the ratio is less than 1:5, use 1:5 X1.5.1.2 A [60 000 lbf] capacity testing machine is to be
for the ratio. If the distance between the two adjacent gradua- verified from [240 lbf] up to [60 000 lbf]. The resolution
tion marks is greater than or equal to 2.5 mm [0.10 in.] and the should be determined at [240 lbf, 2400 lbf, and 24 000 lbf].
ratio is less than 1:10, use 1:10 for the ratio. If the ratio is X1.5.1.3 A 1000 N capacity testing machine is to be verified
greater than those given in these exceptions, use the ratio from 5 N up to 1000 N. The resolution should be determined
determined. Typical ratios in common usage are 1:1, 1:2, 1:5, at 5 N, 50 N, and 500 N.
and 1:10. X1.6 Procedure for Testing Machines with Discrete Forces
X1.3.3 Multiply the ratio determined above by the force Such as Certain Hardness Testing Machines and Creep Testing
represented by one graduation to determine the resolution. Machines:
X1.3.4 Apply as constant a force as possible where the X1.6.1 These testing machines generally incorporate fixed
resolution is to be ascertained to minimize the fluctuation of the lever ratios to apply force. The force applied is obtained by the
indicated force. It is recommended that the fluctuation be no weight applied on the lever multiplied by the lever ratio. They
more than twice the resolution determined in the previous step. do not have a resolution as described in the standard. This
procedure ensures that the sensitivity of the testing machine is
X1.4 Procedure for Non-Auto-Ranging Digital Force Mea- sufficient to apply accurate forces at the lowest calibration
suring Systems: force and may be substituted for reporting resolution.
X1.4.1 The resolution should be checked at the lowest X1.6.2 With an elastic force measurement standard
calibration force in each force range (typically 10 % of full mounted in the testing machine, apply the appropriate weight
scale of the force range(s). to the testing machine’s lever for the lowest calibration force.
X1.4.2 Apply a tension or compression force to a specimen X1.6.3 Gently add a weight to the weight on the lever
approximately equal to that at which the resolution is to be approximately equal to 1/200 of the weight of the weight on the
ascertained, and slowly change the applied force. Record the lever.
smallest change in force that can be ascertained as the X1.6.4 Ensure that at least one-half of the appropriate
resolution. Applying the force to a flexible element such as a change in force is detected by the elastic force measurement
spring or an elastomer makes it easier to change the force standard when the weight is added and when it is gently
slowly. removed.

11

North Carolina A , Thu Aug 26 06:10:58 GMT 2021


E4 − 21

X2. IDENTIFYING AND DETERMINING MEASUREMENT UNCERTAINTY COMPONENTS


DURING AN ASTM E4 CALIBRATION

X2.1 The measurement uncertainty determined using this X2.2.2 A way of assessing the measurement uncertainty due
appendix is the measurement uncertainty of the errors reported to repeatability during calibration is to evaluate the differences
during calibration of a testing machine. It is not the measure- between the two calibration runs of data (the repeatability).
ment uncertainty of the testing machine or the measurement X2.2.2.1 For each calibration force, find the sum of the
uncertainty of test results determined using the testing ma- squares of the differences in force measurement error between
chine. the first and second calibration run of that calibration force and
the four calibration forces closest to that calibration force.
X2.2 Under normal conditions, the measurement uncer- Divide that sum by ten and take the square root of the result to
tainty of the reported force measurement errors of a testing obtain an estimate of the measurement uncertainty due to
machine determined during a calibration using Practice E4 is a repeatability during the calibration.
combination of three major components: the measurement
uncertainty associated with the calibration laboratory perform- NOTE X2.2—The sum is divided by ten because there are five pairs of
readings used, and the variance of each pair is equal to the difference
ing the calibration, the measurement uncertainty due to the divided by two.
repeatability of the testing machine during calibration, and
possibly the measurement uncertainty component due to the X2.2.2.2 Usually this type of assessment of measurement
resolution of the force-measurement system of the testing uncertainty due to repeatability will include the measurement
machine at the force the force measurement error is being uncertainty due to the resolution of the testing machine;
determined and at zero force. however, it is possible to repeat calibration forces without
seeing the effects of the resolution. At each calibration force,
X2.2.1 The measurement uncertainty associated with the test to see that the measurement uncertainty due to repeatability
calibration laboratory performing the calibration is a combina- is greater than the measurement uncertainty due to the resolu-
tion of factors such as, but not limited to: tion of the testing machine. If, at a given calibration force, the
X2.2.1.1 The measurement uncertainty of the calibration uncertainty due to repeatability is not greater than or nominally
laboratory’s force measurement standards per Practice E74, equal to the measurement uncertainty due to the resolution of
X2.2.1.2 Environmental effects such as temperature the testing machine, for that calibration force, include the
variations, components of measurement uncertainty due to the resolution
X2.2.1.3 Measurement uncertainty in the value used for the of the testing machine at that force and at zero force.
local acceleration of gravity at the site where the calibration is
performed when using standard weights, X2.2.3 The measurement uncertainty due to the resolution
X2.2.1.4 Drift in the force measurement standard, of the testing machine at each calibration force is the square
X2.2.1.5 Measurement uncertainty of the calibration of the root of the sum-of-the-squares of the following two compo-
force measurement standard, and nents.
X2.2.1.6 Reproducibility of the force measurement standard X2.2.3.1 The measurement uncertainty component due to
due to handling and fixturing. the resolution of the force-measuring system of the testing
machine being calibrated can be determined by dividing the
NOTE X2.1—A calibration laboratory’s measurement uncertainty resolution of the force-measuring system at the force where
should be based on the maximum measurement uncertainty of the force measurement uncertainty is being evaluated by the quantity of
measurement standards used and the worst environmental conditions
allowed. It can be advantageous to evaluate the measurement uncertainty two times the square root of three.
of the actual force measurement standard used at the actual force for X2.2.3.2 The measurement uncertainty component due to
which the measurement uncertainty of the force measurement error of the the resolution of the force-measuring system of the testing
testing machine is being determined. machine at zero force can be determined by dividing the
X2.2.1.7 If the calibration conditions are outside of the resolution of the force-measuring system at zero force by the
calibration laboratory’s normal operating parameters, addi- quantity of two times the square root of three.
tional components would need to be considered. For example,
a calibration laboratory can permit a 5 °C temperature variation X2.3 The two major components (or three if necessary) can
to occur during calibration and has factored this into their be combined by squaring each component, adding them
measurement uncertainty. When greater temperature variations together, and then taking the square root of the sum to
occur, the uncertainty due to this increased temperature varia- determine the combined measurement uncertainty of the force
tion should be included in the determination of measurement measurement error determined for the testing machine.
uncertainty. X2.4 The expanded measurement uncertainty can then be
X2.2.1.8 A calibration laboratory’s measurement uncer- determined by multiplying the combined measurement uncer-
tainty is usually expressed as an expanded measurement tainty by two, for a confidence level of approximately 95 %.
uncertainty using a coverage factor of two. If this is the case,
prior to combining it with the other measurement uncertainty X2.5 The following presents an example measurement un-
components, divide it by two to determine the standard certainty calculation. The measurement uncertainty of the
measurement uncertainty. reported force measurement error of a 10 000 N capacity

12

North Carolina A , Thu Aug 26 06:10:58 GMT 2021


E4 − 21
TABLE X2.1 Calibration Data
Calibration Run 1 Calibration Run 2 %
Testing Machine Force Measurement Percent Error Testing Machine Force Measurement Percent Error Repeatability
Indicated Force 1, Standard Reading, of Force Indicated Force 2, Standard Reading, of Force
N N (%) N N (%)
100 100.24 -0.24 100 100.02 -0.02 0.22
200 200.21 -0.11 200 200.23 -0.11 0.00
400 400.19 -0.05 400 400.37 -0.09 0.04
700 699.98 0.00 700 700.12 -0.02 0.02
1000 1000.15 -0.01 1000 1001.15 -0.11 0.10
2000 1998.84 0.06 2000 1995.33 0.23 0.17
4000 3994.31 0.14 4000 3988.20 0.30 0.16
7000 6981.97 0.26 7000 6979.86 0.29 0.03
10000 9989.00 0.11 10000 9967.54 0.32 0.21

testing machine is to be determined at 2000 N. X2.5.5 The measurement uncertainty component due to the
X2.5.1 The calibration laboratory’s measurement uncer- testing machine’s resolution at zero force ((see X2.2.3.2), uRZ
tainty expanded using a factor of 2 is 0.3 % of applied force. is:
The testing machine’s resolution at 2000 N is 0.5 N. The 0.5N
resolution of the testing machine at 0.0 N force is 0.5 N. Table u RZ 5 5 0.14 N (X2.4)
2 =3
X2.1 shows the results of two calibration runs:
X2.5.2 The measurement uncertainty component due to the X2.5.6 The total measurement uncertainty component due
calibration laboratory’s measurement uncertainty (see X2.2.1, to resolution at 2000 N is
uCL is: =~ 0.14 N ! 2 1 ~ 0.14 N ! 2 5 0.20 N (X2.5)
0.003 3 2000N
u CL 5 5 3.0 N (X2.1) X2.5.7 The combined measurement uncertainty (see X2.3)
2
can now determined. Since the measurement uncertainty due to
X2.5.3 The measurement uncertainty component due to the repeatability is greater than that due to resolution, the
repeatability (see X2.2.2) at 2000 N, ur is calculated as follows: component due to the resolution is not included.
The repeatability at 2000 N and the four closest forces to 2000
The combined measurement uncertainty of the force mea-
N are 0.02 % of 700 N, 0.10 % of 1000 N, 0.17 % of 2000 N,
surement error determined at 2000 N, u is:
0.16 % of 4000 N, and 0.03 % of 7000 N which respectively
are 0.14 N, 1.00 N, 3.40 N, 6.40 N, and 2.10 N. Therefore: u 5 =~ 3.0 N ! 2 1 ~ 2.4 N ! 2 5 3.8 N (X2.6)
u r5 (X2.2)
X2.5.8 The expanded measurement uncertainty of the force
measurement error determined at 2000 N, U using a coverage
Π~ 0.14 N ! 2 1 ~ 1.00 N ! 2 1 ~ 3.40 N ! 2 1 ~ 6.40 N ! 2 1 ~ 2.10 N ! 2
10
factor of two (see X2.4) is:
U 5 2 3 3.8 N 5 7.6 N (X2.7)
52.4N
7.6 N is 0.38 % of 2000 N.
X2.5.4 The measurement uncertainty component due to the
testing machine’s resolution at 2000 N, (see X2.2.3.1) uR2000 NOTE X2.3—For additional resources relating to measurement
is: uncertainty, refer to the JCGM 100:2008, Evaluation of measurement
data-Guide to the Expression of Uncertainty in Measurement.
0.5N
u R2000 5 5 0.14 N (X2.3)
2 =3

13

North Carolina A , Thu Aug 26 06:10:58 GMT 2021


E4 − 21
SUMMARY OF CHANGES

Committee E28 has identified the location of selected changes to this standard since the last issue (E4–20) that
may impact the use of this standard. (Approved June 1, 2021.)

(1) All sections were revised with respect to using consistent (4) 14.2.3 was revised.
terminology. (5) Section 4 was added.
(2) 1.5 was revised. (6) Appendix X2 was revised.
(3) Terms in 3.2 and 3.3 were added or revised.

Committee E28 has identified the location of selected changes to this standard since the last issue (E4–16) that
may impact the use of this standard. (Approved January 1, 2020.)

(1) Section 1 was revised. (11) Sections 15.1 and 15.4 were revised.
(2) Section 2.2 was added. (12) Section 17.3 was revised.
(3) Terms in sections 3.2 and 3.3 were revised. (13) Section 18.2 was revised.
(4) Section 6 was revised. (14) Section 19.1.9 was revised.
(5) Section 7.3 was revised. (15) Section 19.1.14 was revised.
(6) Section 8.4.3 was revised. (16) Section A1.2 was revised.
(7) Section 9.2.1 was revised (17) Section A1.3.2 was added.
(8) Section Note 6 was revised. (18) Section A1.3.5 was revised.
(9) Section 13.3.2 was revised. (19) Sections X1.6.2 and X1.6.4 were revised.
(10) Sections 14.1, 14.2.1, and 14.2.2 were revised.

ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.

This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.

This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org). Permission rights to photocopy the standard may also be secured from the Copyright Clearance Center, 222
Rosewood Drive, Danvers, MA 01923, Tel: (978) 646-2600; http://www.copyright.com/

14

North Carolina A , Thu Aug 26 06:10:58 GMT 2021

You might also like