Professional Documents
Culture Documents
Analysis of Volatile Organic Compounds and Petroleum Hydrocarbons (C6-C9) in Soil and Sediment with an Agilent 8890 GC_5977B MSD System and a Purge and Trap Sampler
Analysis of Volatile Organic Compounds and Petroleum Hydrocarbons (C6-C9) in Soil and Sediment with an Agilent 8890 GC_5977B MSD System and a Purge and Trap Sampler
Author Abstract
Youjuan Zhang
Agilent Technologies
HJ 605-2011 is primarily used by environmental labs in China for the analysis of
(Shanghai) Co. Ltd,
volatile organic compounds (VOCs) in soil and sediment, while HJ 1020-2019 is
Shanghai 200131 P.R. China
used for the determination of petroleum hydrocarbons (C6 to C9) in the same matrix.
This application note is targeted at labs that analyze both VOCs and petroleum
hydrocarbons (C6 to C9) in soil and sediment. The Teledyne Tekmar Atomx XYZ
purge and trap (P&T) system and an Agilent 8890 gas chromatograph (GC)
configured with a flame ionization detector (FID) and a mass spectrometric detector
(MSD) enable a unified GC configuration that provides the flexibility to run HJ 605
and HJ 1020 on a single instrument. The results of this application note meet all
performance criteria outlined by HJ 605-2011 and HJ 1020-2019 methods.
Introduction Experimental to split the column effluent 1:1 between
the MSD and FID. Figure 1 shows the
Soil contamination has been a growing An Agilent 8890 GC, an Agilent 5977B configuration of the system. Tables 1
concern because it can be a source MSD equipped with FID, and a Teledyne and 2 show the parameter details.
of groundwater (drinking water) Tekmar Atomx XYZ sampler were P&T, GC, and MSD parameters are
contamination and contaminated soils used for this series of experiments. A compatible with HJ 605 and HJ 1020
can also reduce the usability of land for purged two-way CFT device was used method guidelines.
development. In support of its waste
management program, the Chinese
Ministry of Environmental Protection has
developed various methods to test for
PSD FID
contaminants in environmental samples. SSL
VOCs and petroleum hydrocarbons are
Agilent DB-624 UI
frequently detected pollutants in soil Agilent Atomx XYZ
5977B 30 m × 0.25 mm, 1.4 µm
and sediments. HJ 605-2011 details
MSD
the determination of VOCs in soil
and sediment using P&T and GC/MS
Splitter
systems.1 HJ 1020-2019 describes the
0.7 m × 0.15 mm
analysis of petroleum hydrocarbons
(C6 to C9) in the same matrix using
P&T and GC/FID.2 Many environmental 1.9 m × 0.15 mm
VOCs and petroleum hydrocarbons. A Transfer Line Temperature 140 °C Sweep Needle Time 0.25 min
purged two-way CFT device was used to Sample Mount Temperature 90 °C Desorb Preheat Temperature 245 °C
split the column effluent 1:1 between the Water Heater Temperature 90 °C GC Start Signal Begin desorb
MSD and FID, forming a versatile system Sample Cup Temperature 20 °C Desorb Time 2 min
suited for routine analysis of the two Soil Valve Temperature 100 °C Drain Flow 300 mL/min
types of compounds in soil samples. Standby Flow 10 mL/min Desorb Temperature 250 °C
and petroleum hydrocarbon (C6 to C9) Water Volume 10 mL Bake Flow 200 mL/min
analysis, respectively, with one GC Purge Mix Speed Medium Bake Temperature 280 °C
system configured for both FID and Purge Time 11 min MCS Bake Temperature 180 °C
MSD. The linearity, repeatability, method Purge Flow 40 mL/min Trap No. 9
detection limit (MDL), and recovery for Purge Temperature 20 °C
the targeted 65 VOCs and petroleum MCS Purge Temperature 20 °C
hydrocarbons were tested on the new, Dry Purge Time 2 min
combined platform. Dry Purge Flow 100 mL/min
2
Chemicals, standards and sample Table 2. Agilent 8890 GC and 5977B MSD system conditions.
4-bromofluorobenzene, 2-methylpentane, Oven 2. 38 °C (1 min), then 3.8 °C/min to 80 °C (1 min), then 10 °C/min to 105 °C
(5 min), then 10 °C/min to 150 °C (1 min), then 10 °C/min to 180 °C (5 min),
and n-decane were also purchased from recommended by HJ1020 method
ANPEL. Three soil samples (sample 1, FID Restrictor 0.7 m × 0.15 mm id deactivated fused silica tubing (p/n 160-2625-10)
sample 2, and sample 3) were obtained MSD Restrictor 1.9 m × 0.15 mm id deactivated fused silica tubing (p/n 160-2625-10)
from one customer’s lab for tests. Temperature: 250 °C, hydrogen: 30 mL/min, air: 300 mL/min
FID
Samples 1 and 2 were used for recovery Column + make-up gas (N2) = constant: 25 mL/min
testing, while sample 3 was used for real MSD Transfer Line Temperature 250 °C
3
Results and discussion methanol. Therefore, working solutions performance were verified based on
for these compounds should be freshly concentration quantitation results
Volatile organic compounds analysis prepared before injection. The amount instead of the absolute response of
of methanol in the working solution may targeted analytes. Seven consecutive
The MSD data for VOCs were collected
affect the peak shape and response analyses of calibration solution were
in SCAN mode and analyzed using
of the analytes. To obtain a stable run at 20 µg/L. The %RSD of detected
Agilent MassHunter software. Figure 2
and reliable response, the volume of concentrations for 65 VOCs were
shows a chromatogram of 65 target
methanol in each working solution in the range of 0.6 to 5.5%, which
compounds at a concentration of
should be kept constant. demonstrated excellent quantitation
50 µg/L. Excellent peak shape was
The ISTD quantitative method is used precision. Table 3 and Figure 3 show the
obtained for most analytes. The six
in the HJ 605-2011 method, and the details for all compounds, which met the
early eluting compounds are extremely
instrument repeatability and linearity criteria of less than 20% RSD required in
volatile and coelute with water and
the HJ 605 method.
×106
4.0 40,41
3.6
42,43
3.2 59,60
51,52 57
2.8 38,39 50 54 55
49 53 Istd3
Counts
45
2.4
58
2.0 56 62
36 63 65
20,21 31,32 47,48
surr3
1.6 29 64
19 22,23 Istd2
surr2
surr1
1.2 Istd1
14,15,16
9 24 37
0.8 25 30 33 46
8 18
12 13 34
6 11 17 26 44
0.4 7 27 35
1 23 45 10 28 61
0
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19
Acquisition time (min)
4
Table 3. Analysis results for VOCs.
Mean Recovery % (n = 4)
MDL Concentration
No. Name RT m/z CF R2 (µg/kg) RSD% (n = 7) Sample 1 Sample 2
1 Dichlorodifluoromethane 2.182 85.0 0.99984 0.7 2 91.1 90.6
5
Mean Recovery % (n = 4)
MDL Concentration
No. Name RT m/z CF R2 (µg/kg) RSD% (n = 7) Sample 1 Sample 2
Surr 3 4-Bromofluorobenzene 12.146 174.0 0.99924 0.7 1.7 97.9 89.6
6
The calibration curve was developed below 1.1 µg/kg, except four compounds in Figure 4 and Table 3. As Table 3
based on an ISTD method for six (chloromethane, chloroethane, illustrates, the recovery rate for sample
concentration levels from 5 µg/L to 4-methyl-2-pentanone, and naphthalene), 1 was between 85.3 and 113.6%, and
200 µg/L. The correlation coefficient R2 which have MDLs below 3.6 μg/kg . for sample 2 was between 71.0 and
values for 65 VOCs are shown in Table 3. The method recovery was measured 119.4%. Considering the different matrix
Excellent linearity was demonstrated by spiking 25 µL of 10 mg/L calibration in samples 1 and 2, the difference in
for all compounds with an R2 range of standard into 5 g real soil samples recovery rate is acceptable. Sample 2
0.99446 to 1. The method detection limit (corresponding to 50 µg/kg VOCs in has a more complicated matrix, and
(MDL) for 65 VOCs was calculated from the soil sample). For this work, two the recovery results of late effluent
the standard deviation of seven replicate soil samples (samples 1 and 2) were components in this sample were
runs of the lowest calibration standard at obtained from the customer’s lab. relatively low because those compounds
5 µg/L, then translated into MDL (µg/kg) Four runs of each sample’s duplicates have a high boiling point and are hard to
followed by the HJ 605-2011 method with spiking were tested and the average purge from real samples.
requirement. Note that all MDLs are recovery rates were calculated, shown
Sample 1
Mean recovery rate % of 65 VOCs for samples 1 and 2 (n = 4) Sample 2
140
130
120
110
100
90
80
70
60
50
Figure 4. The mean recovery rates for 65 VOCs for sample 1 and sample 2.
7
Petroleum hydrocarbons (C6 to C9) by injecting the single standard solutions Figure 5 shows an overlay of three
analysis of 2-methylpentane (C6H14) and n-decane chromatogram (2-methylpentane single
The HJ 1020-2019 method specifies (C10H22) to identify the retention time standard, n-decane single standard and
organic compounds with a retention time range for C6 to C9. The integration starts C6 to C9 standard mixture) obtained
between 2-methylpentane (included) and at the beginning of 2-methylpentane and from FID signals using the same oven
n-decane (not included) on the FID signal ends before the beginning of n-decane. program as the HJ 605 method. The
as our target. It identified these peaks The Area Sum ON/OFF function in retention time range of C6 to C9 could
as the overall integral as C6 to C9. Before the MassHunter software was used easily be identified by comparing the
proceeding with the calibration and for integration. three chromatograms.
analysis, a qualitative test was performed
×105
8.2
7.8
7.4
7.0
6.6
6.2
5.8
5.4 C6 C6 to C9 C10
5.0
Response units
4.6
4.2
3.8
3.4
3.0
2.6
2.2
1.8
1.4
1.0
0.6
3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 7.5 8.0 8.5 9.0 9.5 10.0 10.5 11.0 11.5 12.0 12.5 13.0 13.5
Acquisition time (min)
Figure 5. An overlay of chromatograms obtained from the system using the HJ 605 oven program. The black line represents a single standard solution of
2-methylpentane (C6); the blue line represents a single standard solution of n-decane (C10); and the red line represents a calibration solution of petroleum
hydrocarbons. The peaks between C6 (included) and C10 (not included) were named C6 to C 9.
8
Figure 6 shows the calibration hydrocarbons in the soil). Samples 1 for sample 1 was 91.8%, and 89.4% for
performance. The correlation and 2 were used for the recovery tests. sample 2. The better recovery rate for
coefficient of 0.9997 indicates a good Three runs of each sample’s duplicates sample 1 also shows that the matrix
linear fit across the full calibration with spiking were tested and the of sample 2 is more complicated than
range. A repeatability test was done average recovery rates were calculated. sample 1.
by measuring the detector response Table 4 shows that the recovery rate
of seven sequential injections at the
concentration of 0.1 μg/mL and 1 μg/mL.
Table 4 and Figure 7 show the results. ×107 C6 to C9
4.0 y = 6838457.780186*x + 1375326.620808
The measured area relative standard 3.5 R2 = 0.9997
deviation (RSD) of 3.7% and 0.8% were Responses
3.0
far below the required maximum of 25% 2.5
2.0
RSD described in the HJ 1020 method.
1.5
The MDL for petroleum hydrocarbons 1.0
was also calculated from the standard 0.5
deviation of seven replicate runs of 0
0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5
the lowest calibration standard at
Concentration (ng/mL)
0.1 μg/mL. To fit with the HJ 1020-2019
method requirements, 5 g of sample Figure 6. A six-level linear calibration of petroleum hydrocarbons from 0 to 6 μg/mL.
was used for MDL calculation, and the
MDL corresponded to 0.02 mg/kg. The Table 4. Analysis results for petroleum hydrocarbons.
method recovery was measured by
RSD (n = 7) Mean Recovery % (n = 3)
spiking 10 µL of 500 mg/L calibration MDL
standard into 5 g of soil samples Name CF R 2
0.1 μg/mL 1 μg/mL (mg/kg) sample 1 sample 2
×105
×105
5.5 1.2
5.0 1.1
4.5 1.0
4.0
0.9
Response units
3.5
0.8
3.0
0.7
2.5
1.5
1.0
0.5
0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19
Acquisition time (min)
Figure 7. An overlay of seven chromatograms obtained from duplicate analysis at a concentration of 1 μg/mL.
9
Real sample analysis could be divided into two situations. the definition of the HJ 1020 method,
Using this unified GC configuration to The first situation is samples obtained the peaks between C6 (included) and
analyze VOCs gives accurate quantitative from suspected petroleum hydrocarbon C10 (not included) were named C6 to C9.
results because it is based on specific contamination sites, containing only the If using the HJ 605 oven program, the
quantification ions selected for each saturated hydrocarbons. The second is HJ 605 VOCs from methylene chloride
VOC compound on the MSD channel. samples with unknown contaminants to 2-chlorotoluene will therefore be
But the petroleum hydrocarbon analysis that may contain both petroleum identified as C6 to C9, together with other
is quantitated by FID signal. If there are hydrocarbon and VOCs. petroleum hydrocarbons. When using
VOCs in the soil samples and they elute The HJ 605 oven program can be applied the oven program recommended by the
within the C6 to C9 elution window, their to the C6 to C9 analysis of the first type HJ 1020 method, the retention ability
response on the FID will be counted in of soil samples. Special attention should of compounds with different polarities
the integrated area of C6 to C9 group be given to unknown soil samples that was different on an Agilent DB-624 Ultra
peak. It will therefore interfere with may be subject to VOC interference. Inert column, and 4-chlorotoluene and
the quantitation of targeted petroleum Figure 8 shows some VOCs in the 1,3,5-trimethylbenzene eluted earlier
hydrocarbons. As a result, the analysis of HJ 605 compound list that eluted within than C10.
real samples of petroleum hydrocarbons the C6 to C9 RT window. According to
C10
×106 C6
1.8
1.7
C6 to C9
1.6
1.5
1.4
1.3
1.2
Response units
1.1
1.0
2-Chlorotoluene
0.9
0.8 4-Chlorotoluene and
1,3,5-trimethylbenzene
0.7
0.6
0.5
0.4
Methylene chloride
0.3
0.2
0.1
0
5.0 5.5 6.0 6.5 7.0 7.5 8.0 8.5 9.0 9.5 10.0 10.5 11.0 11.5 12.0 12.5 13.0 13.5 14.0
Acquisition time (min)
Figure 8. An overlay of FID chromatograms obtained from the system. Black line represents a single standard solution of 2-methylpentane (C6); the blue line
represents a single standard solution of n-decane (C10); the red line represents a calibration solution of 65 VOCs. The peaks between C6 (included) and C10 (not
included) were named C6 to C 9 using the HJ 605 oven program.
10
Figure 9 shows that compounds targeted C6 to C9 hydrocarbons using the HJ 1020 be used during petroleum hydrocarbon
in the HJ 605 method from methylene oven temperature. We would therefore analysis with real samples containing
chloride to 1,3,5-trimethylbenzene eluted suggest using the HJ 605 oven program VOCs with relatively high boiling points.
within the C6 to C9 RT window using for petroleum hydrocarbon analysis An additional benefit with this unified
HJ 1020 oven program. Compared when there is no coexistence of VOCs or system is the ability to use the MSD
to the separation obtained under the VOCs with relatively high boiling points, signal to report a separate BTEX result.
HJ 605 oven temperature, two more such as 1,3,5-trimethylbenzene and The MSD signal can also help confirm
compounds, 4-chlorotoluene and 4-chlorotoluene. For reliable quantitative if there is VOC interference in the
1,3,5-trimethylbenzene, were identified as results, the HJ 1020 oven program can petroleum hydrocarbon analysis result.
C6 C10
×106
1.0
0.9 C6 to C9
0.8
0.7
Response units
0.6
0.5 1,3,5-trimethylbenzene
4-Chlorotoluene
0.4
2-Chlorotoluene
0.3 Methylene
chloride
0.2
0.1
5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27
Acquisition time (min)
Figure 9. An overlay of FID chromatograms obtained from the system. The black line represents a single standard solution of 2-methylpentane (C6); blue line
represents a single standard solution of n-decane (C10); red line represents a calibration solution of VOCs. The peaks between C6 (included) and C10 (not included)
were named C6 to C 9 using HJ 1020 oven program
11
Sample 3 was run to evaluate the results for the real samples. Figure 10 below the linear minimum (5 μg/L),
method under the same experimental shows an overlay of the MSD signal except for acetone, 2-hexanone, and
conditions. Sample 3 was divided into (black trace) and FID signal (red trace) 1,2,4-trichlorobenzene. Table 5 shows
two parts. One part with the internal produced from the analysis of sample 3. the calculated VOC results of sample 3.
standards added was for VOC analysis, The two signals could be quantitated According to the calibration curve of
the other part, without ISTD, was used according to the calibration curves of petroleum hydrocarbons (C6 to C9), the
for petroleum hydrocarbons analysis. VOCs and petroleum hydrocarbons, concentration of C6 to C9 was well below
Stir bars in the vials are recommended to respectively. All VOC concentrations 0.1 μg/mL.
ensure thorough mixing and consistent were calculated and found to be well
×105
Istd 1 Istd 2 Istd 3
2.2 FID signal
2.0 MSD signal
1.8
1.6
1.4
Response
1.2
1.0
0.8
0.6
0.4
0.2
0
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19
Acquisition time (min)
Figure 10. GC/MS-SCAN and FID chromatogram of real sample (sample 3) using the HJ 605 oven program.
2-Hexanone 8
1,2,4-Trichlorobenzene 5
12
High-throughput real sample analysis Although the calibration curves of VOCs As mentioned in the HJ methods,
The analyses of VOCs and petroleum and petroleum hydrocarbons need to the total run time for VOCs is
hydrocarbons (C6 to C9) in real samples be established by sampling different 20 minutes and for petroleum
were performed by injecting samples standard solutions separately, for the hydrocarbons it is 35 minutes.
separately and then substituting MSD real soil sample, one sample injection It therefore takes approximately
and FID signals into different calibration can work for the analysis of two types 60 minutes for two separate VOC and
curves to obtain quantitative results. of compounds. The data can then be petroleum hydrocarbons analyses.
This configuration can acquire MSD and processed separately. As demonstrated in this application
FID signals simultaneously by sampling This high-throughput approach is note, it takes 20 or 35 minutes to
real samples in one injection. Figure 11 suitable for two situations: run the high‑throughput sample
shows simultaneous acquisition of MSD analysis. This new process greatly
• Samples containing both reduces pretreatment time and
and FID signals for samples containing
petroleum hydrocarbons and improves laboratory productivity. This
both petroleum hydrocarbons and
VOCs with relatively low boiling high-throughput sample analysis is
VOCs with relatively low boiling points,
points (compounds eluted before particularly attractive to laboratories
using the HJ 605 oven program. Since
4-chlorotoluene on a DB-624 Ultra trying to maximize sample throughput
VOCs are quantified using internal
Inert column), using the HJ 605 oven and minimize turn-around time.
standard methods, three internal
program (total run time: 20 minutes).
standards should be added into the
sample. The MSD signal was used for • Samples containing both petroleum
VOC analysis, while the FID signal was hydrocarbons and VOCs using the
used for petroleum hydrocarbons. For HJ 1020 oven program (total run
the FID signal, the peaks of the internal time: 35 minutes).
standards should be subtracted when
performing quantitative calculation.
×105
Istd 1 Istd 2 Istd 3
2.2 FID signal
2.0 MSD signal
1.8
1.6
1.4
Response
1.2
1.0
0.8
0.6
0.4
0.2
0
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19
Acquisition time (min)
Figure 11. GC/MS-SCAN and FID chromatogram of real sample (sample 3) obtained on the system simultaneously using the HJ 605 oven program.
13
Conclusion References
This application note demonstrates 1. Soil and Sediment-Determination of
a unified P&T/8890 GC/FID 5977B Volatile Organic Compounds-Purge
MSD configuration that provides and Trap Gas Chromatography/
the flexibility to run either VOCs or Mass Spectrometry Method. HJ
petroleum hydrocarbons (C6 to C9) on 605-2011.
a single instrument. If the HJ standard 2. Soil and Sediment-Determination
requirements need to be strictly obeyed, of Petroleum Hydrocarbons
VOCs and petroleum hydrocarbons (C6-C9)-Purge and Trap Gas
can be tested separately on the same Chromatography. HJ 1020-2019.
system, using the method parameters
recommended by the HJ 605 and
HJ 1020 methods. Fast analysis is
also mentioned in this study, and, for
samples containing both petroleum
hydrocarbons and VOCs with relatively
low boiling points, one injection with
the HJ 605 oven program can be used
for the analysis of the two kinds of
compounds simultaneously. The results
obtained from this versatile system
meet all performance criteria outlined
by HJ 605‑2011 and HJ 1020-2019
methods. The system is particularly
attractive to laboratories seeking low
costs and high flexibility.
www.agilent.com/chem
DE.2521527778