Professional Documents
Culture Documents
VD_UM_MagicMaXUniversal users guide_small
VD_UM_MagicMaXUniversal users guide_small
User's Guide
Notice
This User’s Guide is an integral part of the MagicMaX Universal. To ensure proper use of this product,
please read this User’s Guide carefully and keep it for the future reference. Do not carry out any
adjustment or procedure other than those described in this User’s Guide. The attempt to do so may
result in hazardous situation such as fire, explosion or electric shock to patient, operator or service
engineer.
MagicMaX Universal and its accessories must not be used for any other purpose than described in the
accompanying documentation (intended use). Violation will result in loss of warranty.
IBA Dosimetry GmbH does not accept liability for injury to personnel or damage to equipment that may
result from misuse of this equipment, failure to observe the hazard notices contained in this User’s
Guide, or failure to observe local health and safety regulations.
IBA Dosimetry GmbH shall under no circumstances be liable for incidental or coincidental damage
arising from use of the equipment described in this document.
No part of the accompanying documentation may be translated or reproduced without written
permission of IBA Dosimetry GmbH, unless reproduction is carried out for the sole purpose to be used
by several people in the same department.
The user must treat the accompanying documentation like any other copyrighted material. Especially, if
part of the accompanying documentation is provided in electronic form, these files shall not be modified
in any way. IBA Dosimetry GmbH and its suppliers retain title and all ownership rights to the
accompanying documentation including those of translation into other languages, either in electronic or
printed form.
Several key functions of the MagicMaX Universal and its associated components are protected by
international patents.
All brands, trademarks and trade names mentioned in this documentation are the property of the
applicable manufacturers and companies. IBA Dosimetry GmbH renounces all claims of possession to
rights in brands, trademarks and trade names mentioned in this documentation, which are not their
property.
1. Introduction 1
1.1. Intended use 1
1.2. Non-intended use 1
1.3. General Description 1
1.4. Documentation 2
1.4.1. About this User’s Guide 2
1.5. Symbols 3
2. Health and Safety Information 5
2.1. General 5
2.2. Operator 5
2.3. Safety Precautions 5
2.3.1. Device Handling 5
2.3.2. Protection from Fluids 6
2.3.3. Use of Accessories 6
2.4. Environmental Requirements 6
2.5. Electromagnetic Compatibility (EMC) 7
2.6. Regulatory Requirements 10
3. System components 11
3.1. Parts overview 11
3.1.1. Basic Components 11
3.1.2. Optional accessories 11
3.2. MagicMaX Application Sets 11
4. Hardware Description 13
4.1. Device connection 13
4.2. MagicMaX Universal Basic Unit 13
4.3. MagicMaX Universal Detectors 14
4.3.1. Detectors for the single-detector channel 14
4.3.2. Detectors for the multi-detector channel 15
4.4. PC (optional) 16
4.5. Galvanic Isolator (optional) 16
5. Installation 17
5.1. Requirements 17
5.2. Computer Installation 17
6. Software Description 21
6.1. Topbar - Main Menu, device status, and settings 22
6.2. Measurement List panel 24
6.2.1. Exporting Results to Excel 25
1.4. Documentation
The documentation supplied with the MagicMaX Universal consists of this User's Guide and the factory
calibration certificates depending on the model and configuration.
IM PO RT ANT N O T IC E
PICTURES AND SCREENSHOTS
All numbers and selections displayed in pictures and screenshots are only
examples and no recommendations for settings or entries.
Conventions
The functions of the device, dialog captions and dialog text are indicated by bold font.
Examples: Starting Time, Dose Rate.
Referrals to chapter and section headings in this User’s Guide are indicated by italic font.
Examples: Notice, Technical Specifications.
Throughout this user User’s Guide, hazardous situations or operations are identified by DANGER,
WARNING, CAUTION and NOTICE. They are indicated by specific signs and colors, described below:
Sign Meaning
CAUTION, without the safety alert symbol, used to address issues related to
possible hardware damage.
Area Expertise
Installation Experts
Maintenance Experts
Troubleshooting Experts
1.5. Symbols
The following symbols are used in this User’s Guide or the device.
Important! Observe the warnings and safety precautions given in the enclosed
documents.
The device meets the essential requirements of Council EC Directive 93/42/EEC
concerning medical devices.
Recycling
IM PO RT ANT N O T IC E
ALL PERSONNEL MUST READ THIS CHAPTER
All personnel must read this chapter and be fully aware of its contents before
commencing installation work, operating, or servicing the MagicMaX Universal.
If the MagicMaX Universal is used in a way not specified in this User’s Guide, the
protection provided by the equipment may be reduced.
2.2. Operator
The MagicMaX Universal should only be used by people who are
■ aware of and understand the limitations of the device as they relate to the measurement of radiation
output,
■ knowledgeable about safety procedures to be observed when working with radiation sources.
C AUT IO N
WET DEVICES
Do not use devices on which moisture (condensation) has developed because of
temperature changes unless they have completely dried.
IM PO RT ANT N O T IC E
NO OTHER EQUIPMENT CLOSE TO THE OPERATING DEVICE
The equipment or system should not be used adjacent to or stacked with other
equipment and that if adjacent or stacked use is necessary, the equipment or
system should be observed to verify normal operation in the configuration in
which it will be used.
The use of cables with the MagicMaX Universal other than those described in this document may result
in an increased emission or decreased immunity of the MagicMaX Universal.
The MagicMaX Universal is intended for use in the electromagnetic environment specified below. The
customer or the user of the MagicMaX Universal should assure that it is used in such an environment.
Electromagnetic environment -
Emissions test Compliance
guidance
The MagicMaX Universal is intended for use in the electromagnetic environment specified below. The
customer or the user of the MagicMaX Universal should assure that it is used in such an environment.
<5 % UT
(>95 % dip in UT)
for 5 sec
Power frequency magnetic fields
Power frequency (50/60 Hz)
should be at levels characteristic
magnetic field
3 A/m 3 A/m of a typical location in a typical
commercial or hospital
IEC 61000-4-8
environment.
Note: UT is the a.c. mains voltage prior to application of the test level.
The MagicMaX Universal is intended for use in the electromagnetic environment specified below.
The customer or the user of the MagicMaX Universal should assure that it is used in such an
environment.
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is afflicted by absorption
and reflection from structures, objects and people
a
Field strengths from fixed transmitters, such as base stations from radio (cellular / cordless) telephone and land
mobile radios, amateur radio, AM, FM broadcast, and TV cannot be predicted theoretically with accuracy. To
assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the MagicMaX Universal is used exceeds the
applicable RF compliance level above, the MagicMaX Universal should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the
MagicMaX Universal.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
The MagicMaX Universal is intended for use in an electromagnetic environment in which radiated
RF disturbances are controlled. The customer or the user of the MagicMaX Universal can help
prevent electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the MagicMaX Universal as recommended
below, according to the maximum output power of the communications equipment.
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is afflicted by
absorption and reflection from structures, objects and people.
MagicMaX CT Case
A complete set for CTDI measurements
MagicMaX Universal, consisting of:
• MagicMaX Universal basic unit
• A USB stick (including MagicMaX software, drivers, user manual, etc.)
• USB cables, 1 m and 5 m
Mammography multidetector XR including a detector holder
Illuminance Detector MM-LS
3 part PMMA CT phantom (head /body / pediatric)
CT chamber 10 cm (DCT10-MM), 30 cm (DCT30-MM) upon request
Trolley case MM-CT
RQA, RQM,
C1
current probe, or
USB
MagicMaX
Illuminance detector
C2
software Galvaic
connector*
XM, XR, or DCT10-
C1: Triax Lemo plug
C2: 7-pin Multi plug
MM
* The Galvaic connector is needed only when current probe is used and the PC is connected to mains power, See Section 4.5 for
details.
Multi plug
USB connector
Lemo plug
The MagicMaX Universal basic unit has a channel for single detector and a channel for multi-detector
and three connectors:
■ USB connector
It connects the basic unit to a PC through an IBA delivered USB cable.
■ Lemo plug
It connects the single detector channel to a single semiconductor detector, RQA, RQM, or a MagicMaX
Current Probe, mAs-Clamp.
■ Multi plug
It connects the multi-detector channel a semiconductor multi-detector, XM or XR, or an ionization
chamber.
RQA RQM
The circled area is the reference points for DOSE and kV calibration, respectively.
■ MagicMaX Current Probe (optional): for invasive and non-invasive measurements of the tube currents.
1. Transformer jaws
Used for non-invasive measurements. Enclose the conductor by the jaws to measure DC/AC
current.
4. Power-on LED
When the unit is turned on, this LED will be lighted.
XR multi-detector XM multi-detector
The two circled areas are the reference points for DOSE and kV calibration, respectively
The printed circle (s) on the semiconductor detectors indicates the active area of detectors. Make sure
that the X-ray beam covers the circled area during the measurements.
DCT10-MM
The high voltage for the chamber is supplied through one of the 7 pins of the connector.
4.4. PC (optional)
Any PC fulfils the requirements listed in Section 5.1 can be used for the MagicMaX Universal.
An EeePC Asus laptop (see photo below) with installed software and needed drivers can be purchased
together with MagicMaX Universal.
EeePC
5.1. Requirements
Computer operating system:
■ Operating system: Windows XP 32 bit SP 3, Windows Vista 32 and 64 bit, Windows 7 Professional 32
and 64 bit
■ Pentium processor (or equivalent), minimum 1.6 GHz
■ 1 GB RAM.
■ 600 MB free hard disk space
■ DirectX 9.29.1973 or later
■ Internet Explorer version 6 or later
■ Microsoft .Net 4 Client Framework
■ Windows updates
■ Minimum one USB 2.0 port
Webserver:
Browser compatibilities:
■ Internet Explorer 9
■ Firefox 8
■ iOS Safari 4 and 5
■ Android 2.2 and 3.x
IM PO RT ANT N O T IC E
ADEMINISTRATIIV RIGHTS REQUIRED
For a successful installation and setup of the software, full administration rights
are required.
Select Search for the best driver in these locations and checkmark Include this location in the search;
click Browse to select the desired directory.
Click Next > in Found New Hardware Wizard, then click the Continue Anyway button in Hardware
Installation.
3
2
Splitter Splitter
i ii iii iv v vi
ii: Click this button to connect and disconnect the detector to the software. Click this button
during the measurement stops the measurement.
iii: Here the information of device status is displayed:
Offline – no device connected to the MagicMaX software;
Connecting ... – establishing the communication between software and the device;
Initializing ... – setting up the detector settings;
Init Background Compensation ... – measuring the background signal for background
compensation;
Ready – the system is ready for measurements.
Multi Plug group – settings for the detector connected to the 7-pin connector.
Lemo Plug group - settings for detector connected to the triax connector
Calibration – choices of calibration according to the beam quality. For example, if Auto Cal. Is
selected, the software calculates automatically the calibration factor for the used beam quality.
Tube - choices of the anode-filter combinations of the x-ray device for mammography application. For
other applications, Standard is selected by default.
Sensor is Trigger – if selected, the signal from this detector is (also) used for triggering. The
measurement will start or stop when the triggering value (already pre-set in the software) is reached. If
only one detector is used, it is automatically selected for triggering. The Current probe cannot be used
as a trigger.
ii iii iv v
i: All measured results.
The measured parameters are depended on the selected application in the Matrix Value
Subset drop-down list (see 6.3 for details).
■ Select Create a new file, and then click OK, the Excel window containing the contents in the current
Table View in MagicMaX opens. An example is shown below:
■ Select Append to a new file, and then click OK, browse and select a saved Excel file, click Open. By
clicking OK, the Excel window opens, containing the contents of the selected Excel file following by the
results in Table View in the MagicMaX window.
IM PO RT ANT N O T IC E
ONLY APPEND THE DATA SETS FROM THE SAME APPLICATION
TOGETHER
Please ensure that the data set are append together are from the same
application.
■ Click Open the Template Folder in Explorer … in the Export to Excel dialog, Windows Explorer
opens this template directory:
Click on the name of the desired template ("Template for Mammo" in the example above) , and then
OK.
The Excel window will open and the measurement results will be written in to the selected template.
IM PO RT ANT N O T IC E
ENSURE THE TEMPLATE MATCHES THE EXPORTED DATASET
Please ensure that the template is for the same application of the exported
dataset, e.g., the units in the template should be the same as the measured units
of the exported dataset.
An easy way to generate a template is to use an Excel file exported from MagicMaX, delete the data,
modify it to your needs, and save it as an Excel template file.
In Appendix, a list of tokens can be used as a cell name in an Excel template. After defining the needed
triples (only the cell names are mandatory) and formatting the template, save it as an Excel template file
(*.xltx). Make sure it is in the MagicMaX template folder,
Local Disk (C) - Users - [User's folder] - AppData - Roaming - Microsoft - Windows -Templates -
MagicMaX.
ii
i: Click this button and select an application setup from the list in the Matrix Value Subset for the
Measurement List panel and Favorite View. Proceed with OK.
i
ii
iii
iv
ii: Zoom out - click this button to reduce in the time dimension.
iv: Save - click this button to save the wave as a *.cvs file.
v: Maximize/reset - click this button to maximize the Waveform panel to the full window; Click this
button again to return to the original size.
1. Connect the detector(s) to the basic unit and the basic unit to the PC.
2. Start the program by double-clicking the MagicMaX shortcut icon on the PC desktop or the MagicMaX
program in the Start menu.
“Connecting ... ” is displayed on the top bar, and the Sensor Setup dialog opens.
3. Select the settings in the Sensor Setup dialog according to the application (for details, see Section 6.1), and
then click OK to finish the process.
The Top bar in the MagicMaX window displays “Initializing ...” and “Init Background Compensation ...”
in sequence. It takes a few seconds. Finally when “Ready” appears, the system is ready for measurements.
C AUT IO N
NO HOT PLUGGING
Do not connect or disconnect any MagicMaX Universal device when the
MagicMaX software is running.
C AUT IO N !
DO NOT PLUG / UNPLUG THE MULTI-PLUG CONNECTOR
One pin of the multi plug carries high voltage. Do not plug/unplug this connector
when the MagicMaX software is running.
Danger of electrical shock.
IM PO RT ANT N O T IC E
RESTRICTED TRIGGER FUNCTION
The Current Probe can be used as a trigger, thus, the start of the measurement
will be triggered from a Multidetector.
2. Choose the method of measurement, i.e. Invasive or Non-invasive with the rotary switch in the current
probe.
Invasive measurement:
Connect the invasive cables to the Invasive measurement connectors and the given bridges in the
generator.
W AR NI NG
INVASIVE MEASUREMENTS
Invasive measurements are only allowed for serivcing or trained and certified
employees from the X-ray manufacturer.
C AUT IO N
USING GALVANIC ISOLATOR
If the PC is connected to the mains power for Invasive mode measurement with
the Current Probe, a galvanic isolator must be used between MagicMaX basic
unit and PC: connect the PC to the galvanic isolator and the galvanic isolator to
the PC with two USB cables..
Cover
Shutter
Illuminace Detector MM-LS
Press to open the shutter and turened 45° clockwis e to lock it at the open-position, step back. turn the
shutter 45° counter-clockwise to close it. The meas ement is done.
The measement is done.Turn the shutter 45° counter -clockwise to close the shutter. Put the cover back
when it is not in measement.
When the Illiminance Detector is connected, the Matrix Value Suset is automatically selects
Illuminace.
IM PO RT ANT N O T IC E
OBSERVE REGULATIONS
The mentioned quality assurance tests are recommendations only. International
standards and national laws must be observed.
The label attached on the device indicates the date when the next calibration is due.
However, please ensure that all local regulations and laws applicable to the use of MagicMaX Universal
in your country are complied. For example, if the MagicMaX Universal used in Germany, the German
Medical Product Law applies.
IM PO RT ANT N O T IC E
RESPONSIBILITY
It is the user’s responsibility to follow the guidance given in this User’s Guide and
to use established professional practice to ensure the product is in usable
condition and is being used correctly.
9.4. Troubleshooting
Generally, the MagicMaX Universal has been designed to have a long reliable life. If a problem is
noticed with a measuring system, it may not just be caused by the detector assembly, but also by the
display, processor, Power Supply or interconnecting cables.
There are three classes of error indications:
■ Usually, class one can be corrected by the user since the error is caused by the wrong operation or
handling of the device.
■ Errors of class two are errors in the system, the software, or the module. A restart or re-connection may
be necessary to correct the problem. If the problem does not disappear, you need to contact the
manufacturer.
■ If an error of class three appears, a restart may solve the problem. Usually, your support/supplier has to
be contacted to fix it.
RE CY C LI NG
None of the parts mentioned above may be disposed by the general house or
hospital waste disposal system. Please observe all local regulations governing
the disposal of your system. The end-user is responsible for complying with all
local regulations regarding the removal of the product from service. To avoid
environmental damage and/or injury and if you have no facility to convert the
device to electronic waste at the end of its life cycle, please return it to IBA
Dosimetry GmbH. We will ensure an environmentally correct recycling.
To comply with legal requirements concerning the environmental compatibility of our products
(protection of natural resources, avoidance of waste), we endeavor to reuse components and to return
them to the production cycle. We guarantee the functioning, quality and life of these components by
taking extensive quality assurance measures, just as for factory-new components.
The MagicMaX Universal system contains materials which can be detrimental to the environment.
Therefore, the proper disposal of the MagicMaX Universal system according to the regulations of
national law and in view of protecting the environment must be ensured. This is guaranteed by returning
the MagicMaX Universal system to the manufacturer.
C AUT IO N
MAXIMUM DOSE SURFACE PRODUCT RATE
²
The maximum dose surface product rate of 3000.00 µGym /s must not be
exceeded.
10.4. XR Multidetector
Item Characteristic
Type of detector Silicon photodiode (PIN)
Measurement principle Ionization current in silicon semiconductor
Measurement values kVp, PPV , HVL, Dose, Dose Rate, Dose per Pulse,
Exposure Time, and Wave Form
kVp indicated value PPV (IEC 61676 compliant)
• Range • 40 – 160 kV
Radiography
• Accuracy • < ± 2 % or 0.7 kV
and
fluoroscopy • Total filtration range • 2 – 22 mm (0.08” – 0.87”)
• Sensitivity • 3 mm (0.11”) Al, 50 kV, 1 mA @ 50 cm (19.7”)
• Range • 75 – 160 kV
• Accuracy • < ± 2 % or 0.7 kV
CT
• Total filtration range • 2 – 22 mm (0.08” – 0.87”)
• Sensitivity • 6 mm (0.24”) Al, 90 kV, 0.5 mA @ 50 cm (19.7”)
Time:
Item Characteristic
Accuracy 1 % or 0.2 ms
± 1 pulse
Range 2 ms – 500 ms
6
1 – 0.3x10 pulse
Item Characteristic
Range 1.3 mm – 10 mm
Item Characteristic
Reference conditions:
12° Tungsten anode
Reference point:
Circled 1 cm diameter area, Dose and kV separated, midline – 4.8 mm from top surface.
10.5. XM Multidetector
Total Time
Anode/Filter kV HVL [mm Al] Dose Dose Rate
Tube Filtration ±1% or 0.2
Combination ±2% or 0.7kV ±10% or 0.05mm ±5% ±5%
[um] ms
Reference conditions:
2.2 mm PMMA compression paddle in beam for all measurements (kVp, dose, hvl).
All beams calibrated with large focal spots.
Reference point:
Circled 1 cm diameter area, Dose and kV separated, midline – 4.8 mm from top surface.
RQM-Qualities
1,050
1,025
1,000
0,975
0,950
RQR-M1 RQR-M2 RQR-M3 RQR-M4 RQA-M1 RQA-M2 RQA-M3 RQA-M4
RQR-Qualities
1,050
1,025
1,000
0,975
0,950
RQR 3 RQR 4 RQR 5 RQR 6 RQR 7 RQR 8 RQR 9 RQR 10
RQA-Qualities
1,050
1,025
1,000
0,975
0,950
RQA 3 RQA 4 RQA 5 RQA 6 RQA 7 RQA 8 RQA 9 RQA 10
Features:
Air ionization chamber.
Homogeneous sensitivity over the entire active length of the chamber.
Provided with protective shielding (guarded).
Delivered with an individual calibration document.
Material:
External electrode: carbon fiber reinforced epoxy (CFRP)
Inner electrode: carbon fiber reinforced epoxy (CFRP)
Connector: 7-pin multi plug connector
Cable: 2 m flexible, low noise
max. Cable length: 10 m
Size
active volume: 4.9 cm³ (DCT10-MM), 14.7 cm³ (DCT30-MM)
Total active length: 100 mm (DCT10-MM), 300 mm (DCT30-MM)
Inner diameter of the external electrode: 8.0 mm
Diameter of the inner electrode: 1.0 mm
Operating data:
-15
Leakage current: < ± 4 x10 A
Measuring range for CT beam quality: 100 kV - 150 kV (RQR 8 - RQR 10)
Calibration factor (typical): ND, K = 72 mGy cm / nC
(120 kV / 4.5 mm Al HVL)
Sensitivity: 14.20 nC / Gycm
Dose measuring range: 0.1 mGycm - 999 Gycm
Dose rate measuring range: 1.00 mGycm / s - 0.5 Gycm/s
Uncertainty: <5%
Calibration reference: 120 kV, 2.5 mm Al; 1 m (RQR 9); in the chamber
axis at the center of the cavity volume
Beam incidence direction: The central beam axis is perpendicular to the
chamber axis.
Reference point: Calibration mark is oriented towards geared
towards the X-ray tube.
Beam Quality Voltage [kV] 1. HWD [mm Al] Filtering [mm Al] kq
CT-Qualitäten
CT qualities
1,050
1,030
1,010
kQ
0,990
0,970
0,950
100 120 150
Spannung
Voltage (kV)
Maximum current 2A
Measurement range 10 mA to 2 A
SensorTypeMultiItemTypeHVL
SensorTypeMultiItemTypeFiltration Multi Filtration
SensorTypeMultiItemTypekVp Multi kVp
SensorTypeMultiItemTypePPV Multi PPV
SensorTypeOptionalItemTypeDose Optional Dose
SensorTypeOptionalItemTypeTypicalPulseDose Optional TypicalPulseDose
52| Appendix – Tokens for Excel Templates | MagicMaX Universal User’s Guide
12. Index
Environmental Requirements ............................................6
Error Indications.............................................................. 39
A Experts .............................................................................3
Exposure Time ............................................................... 42
Accessories .................................................... i, 5, 6, 37, 51 External Input Connector .......................................... 41, 43
Accuracy .................................................................. 42, 43
Administration Rights ..................................................... 17
Altitude ........................................................................... 51 F
Factory Calibration Conditions ........................................ 42
B Filtration.............................................................. 46, 48, 49
Flicker Emissions ..............................................................7
Backscatter ........................................................ 45, 47, 50
Battery Indicator ............................................................. 51
Battery Type ................................................................... 51 H
Beam Quality...................................................... 46, 48, 49
Hardware Wizard..................................................... 18, 19
Harmonic Emissions .........................................................7
C Health and Safety Information...........................................5
Hot Plugging ................................................................... 29
Calibration ................................................................ 16, 35 HV Ripple ....................................................................... 42
Calibration & Quality Assurance ..................................... 35 HVL .............................................................. 42, 46, 48, 49
Calibration Certificate ....................................................... 2
Calibration Factor ............................................... 45, 47, 50
Calibration Reference Point ................................ 45, 47, 50 I
CISPR 11 ......................................................................... 7
Cleaning and Disinfection ............................................... 39 IBA Dosimetry GmbH............................... i, 5, 6, 37, 39, 57
Complaint Report ..................................................... 57, 59 IEC 60580 ........................................................................7
Computer Configuration ........................................... 18, 20 IEC 60601 ................................................................ 7, 8, 9
Conductor Size............................................................... 51 IEC 61000 ................................................................ 7, 8, 9
Conventions ..................................................................... 2 IEC 61676 ...................................................................... 42
Cq ....................................................................... 46, 48, 49 Immunity Test ............................................................... 8, 9
Current Probe..................................................... 30, 31, 51 Input Current, max. ......................................................... 41
Intended Use ....................................................................1
Intended User ...................................................................3
D Interface Type................................................................. 41
Interference .................................................................. 3, 9
Data Rate, max. ............................................................. 41 Internal Detector for R & F .............................................. 43
Detector Cable Length........................................ 45, 47, 50 Ionization Current ......................................... 42, 45, 47, 50
Detector Patch Cord ........................................... 45, 47, 50
Device Handling ............................................................... 5
DIN EN 61267 ................................................................ 47 K
Direction of Incidence ......................................... 45, 47, 50
Disposal ......................................................................... 39 kVp ............................................................................ 42, 44
Documentation .............................................................. i, 2
Dose ..................................... 28, 41, 42, 43, 44, 45, 47, 50
Dose per Pulse......................................................... 42, 43 L
Dose Rate ................................... 28, 42, 43, 44, 45, 47, 50
Dose Rate Range ............................................... 45, 47, 50 Leakage Current ............................................................. 41
LEMO Connector ............................................................ 41
Lemosa Connector ................................................... 45, 47
E Liability .......................................................................... i, 5
EC Directive 2002/96/EC.................................................. 3
EC Directive 93/42/EEC ................................................... 3 M
Electrical Fast Transient / Burst ........................................ 8
Electromagnetic Compatibility (EMC) ............................... 7 Maintenance ......................................................... 3, 37, 38
Electromagnetic Emissions............................................... 7 Mammography Quality .................................................... 14
Electromagnetic Environment ............................. 7, 8, 9, 10 mAs Clamp ..................................................................... 28
Electromagnetic Environment - Guidance................. 7, 8, 9 Maximum Current ........................................................... 51
Electromagnetic Immunity ............................................ 8, 9 Measurement Principle ................................. 42, 45, 47, 50
Electromagnetic Interference .......................................... 10 Measurement Setup ....................................................... 21
Electromagnetic Propagation...................................... 9, 10 Measurement Uncertainty ................................... 45, 47, 50
Electrostatic Discharge (ESD) .......................................... 8 Measurement Values ...................................................... 42
Emissions Test ................................................................. 7 Measurement with Solid-state Detector........................... 30
Energy Dependence....................................................... 43 Multidetector............................................................... 30, 31
Environmental Conditions............................................... 42
S
X
Safety Precautions ........................................................... 5
Sampling Rate................................................................ 44 XD Multidetector ............................................................. 44
Semiconductor ............................................. 42, 45, 47, 50 XR Multidetector ............................................................. 42
Customer information
Company name: Contact person:
Address:
Product information
hardware version: Software version: Date of purchase ((DD/MM/YY):