Ceiling Pendants - Operation and Maintenance Manual
Ceiling Pendants - Operation and Maintenance Manual
Ceiling Pendants - Operation and Maintenance Manual
Exemple of configuration
1
NOVAIR
GENERAL DESCRIPTION
As a leading specialist group in operating theatre departments and medical gas pipeline
systems, NOVAIR is uniquely qualified to understand both the user and construction
requirements.
The range of pendants has been developed in response to the following criteria:
Increased Safety
Increased Hygiene and Sterility
Ease of Use
2
NOVAIR
No more trailing leads or hoses to restrict movement within the department, all equipment
is attached via short connecting hoses or cables to integral hook up points.
Fully encapsulated system, incorporating modular service boxes, makes maintenance and
general use easier.
Single or two pieces arm assemblies, motorised or static, equipment can now be located at
point of use with the greatest of ease.
Fully equipped as desired with, high and low voltage supplies, medical gas outlet points,
communication points, miscellaneous.
Generally, there are two types of Operating theatre Pendant;
General Description
The unit is a vertically mounted Services Column that is rigidly mounted to the Ceiling
Structure of the Building Fabric. These units normally contain the Medical Gas, AGSS and
Electrical Services such as standard sockets together with UPS Sockets that are required at the
point of use. Configurations of these services will vary depending on the Customer
Specifications and requirements.
Care should be taken to ensure that the unit is securely fixed to ensure that there is a minimum
of deflection at the head/console.
The fixing method and supporting steelwork will vary according the building structure and the
depth of ceiling void.
Practical Considerations
The standard recommended height of the lower face of a rigid pendant with a horizontal shelf
is 1.4 meter, which will provide access to the equipment mounted on the Shelf.
3
NOVAIR
However, Pendants that have no shelf may be mounted at a higher level and which must take
into account the differing heights of theatre personnel who will have to make the connections
to the services, or conditions needed for particular operations.
The pendant should be located near the head of the operating table. However, note must be
taken of any moving equipment in the theatre, such as operating theatre lamps, to ensure
that the location of the pendant does not foul this equipment.
Finish
The pendant is normally supplied in white epoxy finish (RAL 9002), but other colours can be
provided. The unit can also be supplied with a special non‐reflective finish suitable for use with
laser equipment.
Electrical Requirements
The standard supply to the pendant is 220 volts, single phase, 50 Hz. Other voltages and
frequencies are available on request.
Drawing
340
4
NOVAIR
The simple design of the Terminal Unit enables ease of installation to suit most applications.
If your particular application is not listed please contact us for specific details.
The Terminal Unit consists of a terminal block with integral shut off valve and stub of 12‐mm
copper pipe (150‐mm or 200‐mm) and a back‐plate or bed‐head back‐plate.
The different gases are clearly identified on the ring socket by the use of bold permanent
colours to BS 5682 – 1992. An anti rotational pin is fitted to all horizontally mounted units to
ensure the correct orientation of downstream equipment.
5
NOVAIR
The terminal unit exceeds the criteria given in the table below:
* Maximum pressure drop across terminal unit with NIST connectors and 2M of flexible
hose.
6
NOVAIR
Electrical Sockets
The Electrical Sockets installed may conform to differing standards to suit the user. The
operation of the sockets and the electrical connections should be checked on a regular basis
to ensure fault free conditions apply. Full safety checks of these electrical sockets and any
other electrical sockets in the Operating Theatre should be checked by the Hospital staff
frequently.
Pressure Gauges
Pressure gauges/indicators are provided in the Console to give an approximate indication of
the pressure at the gases within the Pendant. These units are not to be used as a finite
pressure indication and they are to be used only as an aid to fault identification.
General Description
The Multi‐movement pendant is a modular unit, which can be supplied with or without height
adjustment. The pendant consists of a ceiling mount, upon which a wide selection of arms can
be mounted. Flexible connectors carry services. Gas services are physically separated from the
electrical services. The bottom panel of the pendant can be removed to provide access for
maintenance.
An optional motor can be provided and gives 300mm of height adjustment. Other travel
lengths are available, depending on the installation. Operation is by an electrically activated
remote control push‐button located at the pendant, or if required, at a theatre wall panel.
The pendant meets the requirements of the UK DHSS Health Technical memorandum No.
2022. The standard medical gas terminals installed in the pendant conform to HTM No. 2022
and to BS.5682. The pendant can be constructed to other standards, and other makes of
terminal can be fitted.
It is recommended that the terminal units fitted into the pendant be of a similar type to those
in other parts of the hospital.
7
NOVAIR
Services
Services are installed to suit the particular installation. Typical services are:
1. Medical Gas Services
2. Electrical Outlets
3. Waste Anaesthetic Gas Outlets
4. Vacuum (Suction) Outlet
Medical Gases
These services are to BS 5682, and include 02, N2O, O2/N2O mixture (50%/50%), vacuum, air
(4 bar) and air (7 bar). Maximum number of medical gases that can be accommodated: 8
Electrical outlets
These can be high or low voltage. Outlets can be splash proof. Maximum number of 13 amp
sockets that can be accommodated: single: 6 twin: 3
Practical Consideration
The standard recommended height of the lower face of a pendant is 1.7 meter. However,
this does not take into account the differing heights of theatre personnel, or conditions
needed for particular operations.
8
NOVAIR
The pendant allows the unit to be lifted above working height when the unit is not required,
but also allows the height to be adjusted to suit particular circumstances. The pendant should
be located near the head of the operating table. However, note must be taken of any moving
equipment in the theatre, such as operating theatre lamps, to ensure that the location of the
pendant does not foul this equipment.
Finish
The pendant can be supplied in white epoxy finish (RAL 9002), or in stainless steel. The unit
can also be supplied with a special non‐reflective finish suitable for use with laser equipment.
Electrical Requirements
The standard supply to operate the pendant is 240 volts, single phase, 50 Hz. Other voltages
and frequencies are available on request.
Drawing
15
Z
X
Electrical Box NIST Connectors
Cover
630
m
0m
100
500
Y
530
1100
9
NOVAIR
Operation
Pressing a switch in the remote control unit operates the pendant. One switch controls the
rise and another controls the fall. Components fitted into the pendant should be operated in
accordance with the manufacturer instructions.
Maintenance
Maintenance is important for this pendant to ensure that the flexible connectors have not
become worn. Generally, these connectors should be replace regularly. ‘Competent’ persons
as defined by HTM 2022 should carry out maintenance.
Maintenance of components should be in accordance with the manufacturer instructions.
It is recommended that the maintenance requirements for the components be combined into
one program of maintenance for the pendant and incorporated into the planned maintenance
program for the hospital.
Safety
The pendant should be installed by Competent’ persons as defined by HTM 2022. Particular
care should be taken when installing pendants with several gases to ensure that the pipeline
to each gas terminal outlet is connected to the correct gas supply pipeline.
Checks should be made using gas flow and gas identity tests to ensure that the connections
are correct.
Care should be taken to ensure that electrical earth connections are not disturbed and to
ensure that the pendant itself is correctly electrically bonded to the gas and electricity supply
to avoid hazard to patients and theatre personnel.
10
NOVAIR
Lengths of Arms used in Pendants are standard and 1000mm for single horizontal arm pendant
and 600 and 800mm for twin horizontal arm pendant. Optionally, other sizes can be
manufactured upon customer request.
Finish
The pendant is normally supplied in white epoxy finish (RAL 9002), but other colours can be
provided. The unit can also be supplied with a special non‐reflective finish suitable for use with
laser equipment.
Safety
The pendant should be installed by Competent’ persons as defined by HTM 2022. Particular
care should be taken when installing pendants with several gases to ensure that the pipeline
to each gas terminal outlet is connected to the correct gas supply pipeline.
Checks should be made using gas flow and gas identity tests to ensure that the connections
are correct.
Care should be taken to ensure that electrical earth connections are not disturbed and to
ensure that the pendant itself is correctly electrically bonded to the gas and electricity supply
to avoid hazard to patients and theatre personnel.
11
NOVAIR
Drawing
NIST CONNECTORS
1000 mm
600 mm
12
NOVAIR
on the consoles and during movement of the consoles buttons should be pushed. The
specifications of services (medical gas services and electric services) on pendant are same as
the services defined in previous sections.
Finish
The pendant is normally supplied in white epoxy finish (RAL 9002), but other colours can be
provided. The unit can also be supplied with a special non‐reflective finish suitable for use with
laser equipment.
Drawing
1000 mm
1000 mm 1000 mm
600 mm 600 mm
13
NOVAIR
Electrical Requirements
The standard supply to the pendant is 220 volts, single phase, 50 Hz. Other voltages and
frequencies are available on request.
The services are run to a first‐fix unit in the ceiling. For medical gases, these services will
terminate in a Non‐interchangeable screw connection (NIST) to BS 5682. The first fix
installation should be completed, fully tested and blanked until the second‐fix installation is
needed. The pendant is supplied as a second‐fix item for installation during the finishing of
the theatre and is normally supplied with all components installed and tested in the factory.
Gas services are via flexible hose connectors, which are attached to the first‐fix NIST
connectors. Electrical services in the pendant normally terminate in a junction (electrical) box,
at the top of the pendant for connection to the hospital electrical wiring system. The pendant
should be fitted to the solid ceiling of the theatre and not supported by a false ceiling only.
The pendant can be supplied as one long unit, which is secured to the main ceiling and then
passes through the false ceiling. Alternatively, a framework can be attached to the main
ceiling, which extends downwards to just above the false ceiling. The pendant can then be
attached to the framework. The pendant is supplied with a beveled fascia plate to provide a
neat junction at the ceiling. All securing bolts must be adequate for the load of the pendant.
Any framework must also be designed for rigidity. Even small loads, such as those imposed by
inserting gas probes can cause large deflections on a long pendant. Any framework should be
designed for maximum rigidity to avoid these deflections.
14
NOVAIR
Tighten the hexagonal nuts M12 (3) and be sure that the ceiling plate is fitted firmly to the
raw ceiling and leveled correctly.
After completing the installation of ceiling flange, interface flange should be installed.
Installation of interface flange should be as described below and shown in Figure 2;
The interface flange should be fixed to the ceiling flange by means of 4 pieces of 50 mm
diameter steel pipes, M20 threaded rods, washers and nuts which are manufactured for the
required length.
Screw the hexagonal nut M20 (10) approximately 25 mm on the threaded rod M20 (8) and
apply a lock washer (9) then fix on the nut (11) which is welded on the ceiling flange (1).
Screw the threaded rod M20 (8) to the ceiling flange (1) until it stops. And tighten the
hexagonal nut (10) securely.
15
NOVAIR
If the threaded rods (8) have not been screwed securely, the supporting system may fall
down. Hence, screwing of the threaded rods should be securely to the ceiling flange (1).
Supporting pipes (7) may cause heavy injuries. Since the interface flange (6) does not provide
safe fixing of the supporting pipes (7) on the threaded rod (8), nobody must be allowed under
the installation area during installation.
After installation of the supporting pipes (7) on the threaded rods (8) place the interface
flange and fix it to the ceiling plate (1) by tightening the hexagonal nuts M 20 (12) and
washers (9). Be sure that the interface flange (6) is fitted firmly to the ceiling plate (1) and is
leveled correctly.
After fixing the interface flange (6) locate the supporting bracket system (13) to the opening
of the interface flange (6) as shown in Figure 3.
Fix the supporting bracket system (13) to the interface flange (6) with 6 pieces of M 18 x
60 mm hexagonal bolts (14), washers (15) and M 18 nuts (16).
16
NOVAIR
The supporting bracket must be mantled in a horizontally aligned position to ensure light
mobility and safe positioning of the supporting bracket system (13)
After tightening nuts (16) check the horizontal alignment by means of a spirit level (17) and
if required adjust the horizontal alignment by using shims between the interface flange (6)
and the supporting bracket system (13)
The vertical arm (21) is mounted to the supporting bracket system (13) by means of twelve
M6x10mm screws (22).
Fix the M6x10mm screws (22) and after checking the vertical alignment of the vertical arm
(21) tighten them.
After mounting the vertical arm slip on the cap (23) over the vertical arm (21) towards the
horizontal arm side to hide the screws and fix it by using the 102x3,5 O‐ring (24).
32
13
Electric (Junction) Box
6
NIST Connection
22
21
23
24
Installation of Pendant
17
NOVAIR
21
26
25
The electrical horizontal arm (25) should be mounted to the vertical arm (21) by means of
eight M6x10mm screws (26).
Fix the screws (26) and after checking the horizontal alignment of the electrical horizontal
arm (25) tighten them.
18
NOVAIR
After installation of horizontal arm, pendant console should be assembled to the horizontal
arm. During the assembly of console, please take care in handling the hoses and electric cables
which are connected to the console. Steps that should be followed are described below and
shown in Figure 6.
Before assembly of the console (26) to the electric horizontal arm (25) slide the hoses and
cables of the console carefully through the arms installed before and fix them to the interface
flange (6) temporarily.
The console (26) is mantled to the electrical horizontal arm (25) by means of eight M6x10mm
screws (27).
Fix the screws (27) and after checking the horizontal alignment of the console (26) tighten
them.
Console Side
21
25
Electric Motor
Side
19
NOVAIR
34
32
33
Electric Box
6
NIST
CONNECTIONS
20
NOVAIR
21
NOVAIR
After installation of the supporting pipes (6) on the threaded rods (7) mount the interface
flange (5) and fix it to the ceiling plate (1) by tightening the hexagonal nuts M 20 (11) and
washers (8). Be sure that the interface flange (5) is fitted firmly to the ceiling plate (1) and is
leveled correctly.
22
NOVAIR
After fixing the interface flange (5) locate the supporting bracket systems (12 & 13) to the
openings of the interface flange (5) as shown in Figure 11.
Fix the supporting bracket systems (12 & 13) to the interface flange (5) by 6 pieces of
M18x60 mm hexagonal bolts (14), washers (15) and M18 nuts (16).
The supporting brackets should be mantled in a horizontally aligned position to ensure light
mobility and safe positioning of the supporting bracket systems (12 & 13)
After tightening nuts (16) check the horizontal alignment by means of leveling device (17)
and if required adjust the horizontal alignment by using shims between the interface flange
(5) and the supporting bracket systems (12 & 13).
13
5
12
20
21
18 20
21
23
22
19
25
23
24
Figure 12 Installation of Horizontal and Vertical Arms of Dry and Wet Side Consoles
23
NOVAIR
After completing the assembly of supporting bracket systems (12&13), system is ready for the
assembly of horizontal and vertical arms of wet and dry side equipment consoles. Assembly
of arms are described below.
Initially the horizontal arms should be mounted. Installation of horizontal arms (18 &19) can
be mounted by means of sixteen M8x16 allen screws (20) and washers (21) to the supporting
bracket systems (12&13).
Covers (25) should be fitted to the vertical arms (22&24) as shown in Figure 7.4.
Then, vertical arms (22&24) can be mounted to horizontal arms (18&19) by means of
M6x12mm screws (23). For each arm twelve screws will be used.
Nous ne pouv ons pas afficher l’image.
22
27
26
24
25
27
28
DRY SIDE
WET SIDE
Figure 13 Assembly of Dry and Wet Side Equipment Consoles
24
NOVAIR
After installation of horizontal and vertical arms, wet and dry side equipment consoles should
be assembled to the vertical arms (22&24). During the assembly of consoles, please take care
in handling the hoses and electric cables which are connected to the consoles. Steps that
should be followed are described below and shown in Figure 13.
Before assembly of the consoles (26&28) to the vertical arms (22&24) slide the hoses and
cables of the console carefully through the arms installed before and fix them to the interface
flange (5) temporarily.
The consoles (26&28) are mantled to the vertical arms (22&24) by means of twelve
M6x10mm screws (27).
Fix the screws (27) and after checking the horizontal alignment of the consoles (26&28)
tighten them.
As shown in Figure 14 the ceiling flange of the pendants can be fixed by using male type steel
anchors placed to the holes drilled to the ceiling.
Drill 8 holes to the ceiling according to the fixing plan of the ceiling flange (1).
Mount M12 steel anchors (2) to the holes drilled and fix them.
25
NOVAIR
Locate the ceiling flange (1) to the anchors (2) and fix it by means of hexagonal nuts, M12 (3)
and washers (4).
Tighten the hexagonal nuts M12 (3) and be sure that the ceiling plate is fitted firmly to the
raw ceiling and leveled correctly.
10
10 1 8
7
8 X 9
6 9
8 5 DETAIL X
11
Figure 15 Installation of Interface Flange
After completing the installation of ceiling flange, interface flange should be installed.
Installation of interface flange should be as described below and shown in Figure 15;
The interface flange should be fixed to the ceiling flange by means of 4 pieces of 50 mm
diameter steel pipes, M20 threaded rods, washers and nuts which are manufactured for the
required length.
Screw the hexagonal nut M20 (9) approximately 25 mm on the threaded rod M20 (7) and
apply a lock washer (8) then fix on the nut (10) which is welded on the ceiling flange (1).
Screw the threaded rod M20 (7) to the ceiling flange (1) until it stops. And tighten the
hexagonal nut (9) securely.
If the threaded rods (7) have not been screwed securely, the supporting system may fall
down. Hence, screwing of the threaded rods should be securely to the ceiling flange (1).
Supporting pipes (6) may cause heavy injuries. Since the interface flange (5) does not provide
safe fixing of the supporting pipes (6) on the threaded rod (7), nobody should be allowed
under the installation area during installation.
After installation of the supporting pipes (6) on the threaded rods (7) place the interface
flange and fix it to the ceiling plate (1) by tightening the hexagonal nuts M 20 (11) and washers
26
NOVAIR
(8). Be sure that the interface flange (5) is fitted firmly to the ceiling plate (1) and is leveled
correctly.
After fixing the interface flange (5) locate the supporting bracket system (12) to the opening
of the interface flange (5) as shown in Figure 8.3.
Fix the supporting bracket system (12) to the interface flange (5) with 6 pieces of M 18 x
60 mm hexagonal bolts (13), washers (14) and M18 nuts (15).
The supporting bracket must be mantled in a horizontally aligned position to ensure light
mobility and safe positioning of the supporting bracket system (12).
After tightening nuts (15) check the horizontal alignment by means of leveling device (16)
and if required adjust the horizontal alignment by using shims between the interface flange
(5) and the supporting bracket system (12)
13
12
5
14
15
16
27
NOVAIR
After completing the assembly of supporting bracket systems (12), assembly of horizontal and
vertical arms should be carried out. Assembly of arms is described below.
Initially the horizontal arm (17) should be mounted. Installation of horizontal arm (17) can
be done by means of sixteen M8x16 allen screws (18) and washers (19) to the supporting
bracket systems (12).
Then, vertical arm (22) can be mounted to horizontal arm (17) by means of M6x12mm screws
(21).
Cover (20) should be fitted to the vertical arm (22) as shown in Figure 8.4.
28
NOVAIR
The electrical horizontal arm (24) should be mounted to the vertical arm (22) by means of
eight M6x10mm screws (23) as shown in Figure 17.
Fix the screws (23) and after checking the horizontal alignment of the electrical horizontal
arm (24) tighten them.
After installation of electrical horizontal arm, console should be assembled to the electrical
horizontal arm (24). During the assembly of console, please take care in handling the hoses
and electric cables which are connected to the console. Steps that should be followed are
described below and shown in Figure 18.
Before assembly of the console (26) to the electrical horizontal arm (24) slide the hoses and
cables of the console carefully through the arm installed before and fix them to the interface
flange (5) temporarily.
The console (26) is mantled to the horizontal electrical arm (24) by means of twelve
M6x10mm screws (25).
Fix the screws (25) and after checking the horizontal alignment of the console (26) tighten
them.
29
NOVAIR
22
Finally, the covers of electrical horizontal arm should be mounted. Assembly of covers should
be as follows and shown in Figure 19;
Mount the plastic covers (27) to both ends of the electrical arm and fix them with screws.
Mount the metal covers (28) to both ends of the electrical arm and fix them with screws.
Locate the greater plastic covers (30) to the motor side of the electrical arm and fix them
with screws.
Locate the small plastic covers (29) to the console side of the electrical arm and fix them with
screws.
30
NOVAIR
Console Side
22
24
Electric Motor
Side
31
NOVAIR
As shown in Figure 20 the ceiling flange of the pendants can be fixed by using male type steel
anchors placed to the holes drilled to the ceiling.
Drill 8 holes to the ceiling according to the fixing plan of the ceiling flange (1).
Mount M12 steel anchors (2) to the holes drilled and fix them.
Locate the ceiling flange (1) to the anchors (2) and fix it by means of hexagonal nuts, M12 (3)
and washers (4).
Tighten the hexagonal nuts M12 (3) and be sure that the ceiling plate is fitted firmly to the
raw ceiling and leveled correctly.
10
10 1 8
7
8 X 9
6 9
8 5 DETAIL X
11
Figure 21 Installation of Interface Flange
After completing the installation of ceiling flange, interface flange should be installed.
Installation of interface flange should be as described below and shown in Figure 21;
The interface flange should be fixed to the ceiling flange by means of 4 pieces of 50 mm
diameter steel pipes, M20 threaded rods, washers and nuts which are manufactured for the
required length.
Screw the hexagonal nut M20 (9) approximately 25 mm on the threaded rod M20 (7) and
apply a lock washer (8) then fix on the nut (10) which is welded on the ceiling flange (1).
Screw the threaded rod M20 (7) to the ceiling flange (1) until it stops. And tighten the
hexagonal nut (9) securely.
If the threaded rods (7) have not been screwed securely, the supporting system may fall
down. Hence, screwing of the threaded rods should be securely to the ceiling flange (1).
32
NOVAIR
Supporting pipes (6) may cause heavy injuries. Since the interface flange (5) does not provide
safe fixing of the supporting pipes (6) on the threaded rod (7), nobody should be allowed
under the installation area during installation.
After installation of the supporting pipes (6) on the threaded rods (7) place the interface
flange and fix it to the ceiling plate (1) by tightening the hexagonal nuts M 20 (11) and washers
(8). Be sure that the interface flange (5) is fitted firmly to the ceiling plate (1) and is leveled
correctly.
14
12
13
5
15
16
17
After fixing the interface flange (5) locate the supporting bracket system (12) to the opening
of the interface flange (5) as shown in Figure 22.
Fix the supporting bracket system (12) to the interface flange (5) with 6 pieces of M 18 x 60
mm hexagonal bolts (14), washers (15) and M18 nuts (16).
The supporting bracket must be mantled in a horizontally aligned position to ensure light
mobility and safe positioning of the supporting bracket system (12).
33
NOVAIR
After tightening nuts (16) check the horizontal alignment by means of leveling device (17)
and adjust the horizontal alignment by using shims (13) between the interface flange (5) and
the supporting bracket system (12).
34
NOVAIR
After completing the assembly of supporting bracket system (12), horizontal and vertical arms
of the console should be done. Assembly of arm is described below.
Initially the horizontal arms (18) should be mounted. Installation of horizontal arms (18) can
be done by means of sixteen M8x16 allen screws (19) and washers (20) to the supporting
bracket system (12).
Covers (21) should be fitted to the vertical arm (23) as shown in Figure 9.4.
Then, vertical arm (23) can be mounted to horizontal arms (18) by means of twelve
M6x12mm screws (22).
24
25
35
NOVAIR
After installation of horizontal and vertical arms, equipment console is going to be assembled
to the vertical arm (23). During the assembly of console, please take care in handling the
hoses and electric cables which are connected to the consoles. Steps that should be followed
are described below and shown in Figure 24.
Before assembly of the console (25) to the vertical arm (23) slide the hoses and cables of the
console carefully through the arm installed before and fix them to the interface flange (5)
temporarily.
The console (25) is mantled to the vertical arm (23) by means of twelve M6x10mm screws
(24).
Fix the screws (24) and after checking the horizontal alignment of the console (25) tighten
them.
LABELS USED ON THE PENDANTS
36
NOVAIR
W
Motor Coil Tip Label: Indicates the motor coil tip.
Gauge Label: Indicates the gauge which measures the pressure of Air 4
bar gas.
Gauge Label: Indicates the gauge which measures the pressure of Air 7
bar gas.
Gauge Label: Indicates the gauge which measures the level of Vacuum
PACKAGING
When opening the Pendant package, follow the caution marks on the package.
Otherwise, you can cause the unit to damage.
37
NOVAIR
CAT 6 CABLE
CAT 6 CABLE
CAT 6 CABLE
MICRO SWITCH
II
I
DOUBLE POLE
CAPASITY
SWITCH
M
1
38
NOVAIR
APPENDIX 2 Electrical Wiring Diagram of Surgical Pendants for Rooms N4 144 & N4 260
CAT 6 CABLE
CAT 6 CABLE
CAT 6 CABLE
CAT 6 CABLE
CAT 6 CABLE
MICRO SWITCH
5
3
1
6
5
II
I
4
2
DOUBLE POLE
CAPASITY
SWITCH
4
3
2
1
M
1
1
39
NOVAIR
APPENDIX 3 Electrical Wiring Diagram of Surgical Pendants for Rooms N4 144 & N4 260
CAT 6 CABLE
CAT 6 CABLE
CAT 6 CABLE
CAT 6 CABLE
CAT 6 CABLE
MICRO SWITCH
5
3
1
6
5
II
I
4
2
DOUBLE POLE
CAPASITY
SWITCH
4
3
2
1
M
1
1
40
NOVAIR
APPENDIX 4 Electrical Wiring Diagram of Wet Side of ICU/HDU Double Arm Pendants
CAT 6 CABLE
CAT 6 CABLE
CAT 6 CABLE
CAT 6 CABLE
CAT 6 CABLE
CAT 6 CABLE
41
42
CAT 6 CABLE
CAT 6 CABLE
CAT 6 CABLE
CAT 6 CABLE
CAT 6 CABLE
CAT 6 CABLE
CAT 6 CABLE
TELEPHONE CABLE
APPENDIX 5 Electrical Wiring Diagram of Dry Side of ICU/HDU Double Arm Pendants
NOVAIR
NOVAIR
Vertical and Rotational movement of pendants can be controlled by the buttons available on
front of the console as shown below. There are two pneumatic brakes which are located in
supporting bracket system and joints between two horizontal arms on anaesthesia pendants.
Brake button (1) operates the pneumatic brake in supporting bracket system and brake button
(2) operates the brake at the joints between two horizontal arms. Vertical (up & down)
movement of console can be adjusted by the button (3).
1 2 3
Same as anaesthesia pendants for surgical pendants vertical and rotational movements of
pendants can be controlled by the buttons available on front of the console as shown below.
There is single pneumatic brake on surgical pendant located in the supporting bracket system.
The button (1) operates the pneumatic brake and vertical (up & down) movement of console
can be adjusted by the button (2).
1 2
43
NOVAIR
Rotational movement of dry side can be controlled by the button available on front of the
console as shown below. There is single pneumatic brake which is located in supporting
bracket system and controlled by button (1). Button (2) is used to switch on/off night light on
the console.
Rotational movement of wet side can be controlled by the buttons available on front of the
console as shown below. There are two pneumatic brakes which are located in supporting
bracket system and joints between two horizontal arms. Brake button (1) operates the
pneumatic brake in supporting bracket system and brake button (2) operates the brake at the
joints between two horizontal arms.
44