Professional Documents
Culture Documents
Qrg Harmony Pro
Qrg Harmony Pro
Always perform a skin test on the intended treatment before the first
treatment session according to the skin test parameters on the table below.
Skin Test Parameters:
Deep 15 - 30
I – VI (-2) , (-1) 3 1800-2400 500 20
Wrinkles hours
Photoaging / 15 - 30
I – VI (+1) , (+2) 3 1800-2400 500 20
pigmentation hours
15 - 30
I – VI Skin Laxity (0) 3 1800-2400 500 20
hours
(*) per 20 cm2 area
• Slow movement inside the grid area while stacking several times
• End point: Mild local Erythema, no of treatments:3-5, Treatment intervals:2-3 weeks
• Caution: The appropriate optical density of protective eyewear for patient and
operator is: OD7+
• For detailed information, please refer to the Harmony XL PRO Operator’s Manual
2
Suggested Treatment Parameters:
Pulse Treatment
Skin Gauged Energy Accumulative Energy
Indication Repetition Area Grid
Type Tip (mJ) Joules/Grid*
Rate (Hz) (cm2)
I – VI Photoaging /
pigmentation (+1) , (+2) 3 1800-2400 600 20
• Slow movement inside the grid area while stacking several times
• End point: Mild local Erythema, no of treatments:3-5, Treatment intervals:2-3 weeks
• Caution: The appropriate optical density of protective eyewear for patient and
operator is: OD7+
• For detailed information, please refer to the Harmony XL PRO Operator’s Manual
3
2. H igh Power Nd:YAG Q-Switch
1064/532nm Module
1-5mm and CVD tip
• Tattoo Removal (Q-Switch 1064nm) - Blue, Black
• Tattoo Removal (Q-Switch 532nm) - Red
Always perform a skin test on the intended treatment before the first
treatment session according to the skin test parameters on the table below.
Skin Test Parameters Q-Switch 1064nm:
Pulse
Spot Size
Skin Type Repetition Energy (mJ) Waiting Period
(mm)
Rate (Hz)
Pulse Repetition
Skin Type Spot Size Energy (mJ/P)
Rate (Hz)
I - VI 1, 2, 3 1, 2, 3 600 - 1000
I - VI 4, 5, 6 1, 2, 3 700 - 1200
4
Always perform a skin test on the intended treatment before the first
treatment session according to the skin test parameters on the table below.
Skin Test Parameters Q-Switch 532nm:
Pulse
Skin Type Spot Size (mm) Repetition Energy (mJ) Waiting Period
Rate (Hz)
IV - VI 2, 3, 4 1, 2, 3 600 24 - 48 hours
Pulse Repetition
Skin Type Spot Size Energy (mJ/P)
Rate (Hz)
IV - VI 2, 3, 4 1, 2, 3 700 - 1000
5
3. Er:YAG 2940nm Module
7X7 pixel tip
• Skin Resurfacing
• Wrinkles
• Keratose
Always perform a skin test on the intended treatment before the first
treatment session according to the skin test parameters on the table below.
Skin Test Parameters:
Number of
Skin Type Energy (mJ/P) Pulse Mode Frequency (Hz)
Passes/Stacks*
6
4. Er:YAG 2940nm Module
1mm, 4mm tips
• Skin Resurfacing
• Wrinkles
• Keratose
Always perform a skin test on the intended treatment before the first
treatment session according to the skin test parameters on the table below.
Skin Test Parameters:
Fluence Frequency
Skin Type Mode Tip Size Intensity
(mJ/Pulse) (Hz)
Energy Frequency
Skin Type Mode Tip Size Intensity
(mJ/Pulse) (Hz)
7
5. Er:YAG 2940nm Module
Pixel roller 1X7 tip
• Skin Resurfacing
• Scar Revisions
Always perform a skin test on the intended treatment before the first
treatment session according to the skin test parameters on the table below.
Skin Test Parameters:
Energy Number of
Intensity Skin Type Pulse Mode
(mJ/Pulse) Passes
Mild I - IV 600 Medium 1 (XY)
Moderate I - IV 700 Medium 1 (XY)
Aggressive I - IV 800 Medium 1 (XY)
8
6. Cooled Laser Long Pulsed Nd:YAG
1064nm Module
6mm, 2mm tips
• Vascular Lesions
Always perform a skin test on the intended treatment before the first
treatment session according to the skin test parameters on the table below.
Skin Test Parameters for Vascular Lesions:
Skin Type Pulse Width Fluence
Tip Size Waiting Period
(Fitzpatrick I-VI) (msec) (J/cm2)
6 mm 12, 15, 45 80 - 100 2 - 4 hours
I - III
2 mm 10 200 2 - 4 hours
6 mm 12, 15, 45 60 - 80 48 - 72 hours
IV - VI
2 mm 10 150 48 - 72 hours
9
7. Cooled Laser Long Pulsed Nd:YAG
1064nm Module
6mm, 10mm tips
• Hair Removal
Always perform a skin test on the intended treatment before the first
treatment session according to the skin test parameters on the table below.
Skin Test Parameters for Hair Removal:
Skin Type Pulse Width Fluence Waiting
Tip Size
(Fitzpatrick I-VI) (msec) (J/cm2) Period
10mm 15 40 - 50 24 - 48 hours
V
10mm 15 30 - 40 24 - 48 hours
10mm 15 40 - 50 24 - 48 hours
VI
10mm 15 30 - 40 24 - 48 hours
• End point: Peri follicular Erythema or Edema; no. of treatments 6-8, treatment
intervals; 4-8 weeks
• Caution: The appropriate optical density of protective eyewear for patient and
operator is:OD7+
• For detailed information, please refer to the Harmony XL PRO Operator’s Manual
10
Suggested Treatment Parameters for Hair Removal:
Skin Type
Hair Color Fluence (J/cm2) Fluence (J/cm2)
(Fitzpatrick)
Light 40 - 50 15**
VI
Dark 30 - 40 15**
• End point: Peri follicular Erythema or Edema; no. of treatments 6-8, treatment
intervals; 4-8 weeks
• Caution: The appropriate optical density of protective eyewear for patient and
operator is:OD7+
• For detailed information, please refer to the Harmony XL PRO Operator’s Manual
11
8. Cooled Laser Long Pulsed Nd:YAG
1064nm Module
6mm tip
• Facial Wrinkles
Always perform a skin test on the intended treatment before the first
treatment session according to the skin test parameters on the table below.
Skin Test Parameters for Facial Wrinkles:
Skin Type Pulse Width Fluence
Tip Size Waiting Period
(Fitzpatrick I-VI) (msec) (J/cm2)
12
Suggested Treatment Parameters for Facial Wrinkles:
Skin Type No. of Pulses Pulse Width
Tip Size Fluence (J/cm2)
(Fitzpatrick I-VI) per Spot (msec)
I - III 1 45 Up to 150
6 mm
IV - VI 1 45 Up to 120
13
9. Clear Skin- Er:Glass 1540nm Module
4mm,10mm, 7x7 pixel tips
• Acne Vulgaris
• Acne Scar
• Skin Rejuvenation
Always perform a skin test on the intended treatment before the first
treatment session according to the skin test parameters on the table below.
Skin Test Parameters - Acne Vulgaris:
Skin Type Pulse Number
Fluence Number Waiting
(Fitzpatrick Tip* Repetition of stacked
(mJ/P) of passes Period
I-VI) Rate (Hz) pulses
• End points (Acne Vulgaris and Acne Scars) : Mild Erythema; No. Of treatments: 4-6
Treatment Intervals: 1-3 weeks
• For Er:Glass-The appropriate optical density of protective eyewear for patient and
operator is: OD7+
• For detailed information, please refer to the Harmony XL PRO Operator’s Manual
14
Skin Test Parameters - Acne Scars / Skin Rejuvenation
Skin Type Pulse Number
Fluence Number Waiting
(Fitzpatrick Tip** Repetition of stacked
(mJ/P) of passes Period
I-VI) Rate (Hz) pulses
I – III 7x7 Pixel 3/Stack 6-8 800 - 1000 2-3 15-30 min
• End points (Acne Vulgaris and Acne Scars) : Mild Erythema; No. Of treatments: 4-6
Treatment Intervals: 1-3 weeks
• For Er:Glass-The appropriate optical density of protective eyewear for patient and
operator is: OD7+
• For detailed information, please refer to the Harmony XL PRO Operator’s Manual
15
10. DYE-VL Pro Module (500-600nm)
• Vascular Lesions
• Pigmented Lesions
Always perform a skin test on the intended treatment before the first
treatment session according to the skin test parameters on the table below.
Skin Test Parameters - DYE-VL Mode (Stationary Mode)
Ultrasonic gel must be applied to the skin
16
Skin Test Parameters - DYE-SSR Mode (In Motion Mode)
Ultrasonic gel must be applied to the skin
• End point: vessel discoloring, vessel shrinking or blanching; mild Erythema; darkening
of the pigmented lesions. No. of treatments: 3-6; Treatment intervals: 2-3 weeks
• Caution: The appropriate optical density of protective eyewear for patient is: OD 5 + and
for operator: OD 3 +
• For detailed information, please refer to the Harmony XL PRO Operator’s Manual
17
11. DYE SR Pro Module (550-650nm)
• Vascular Lesions
• Pigmented Lesions
Always perform a skin test on the intended treatment before the first
treatment session according to the skin test parameters on the table below.
Skin Test Parameters - SR Mode (Stationary Mode)
Ultrasonic gel must be applied to the skin
Fitzpatrick
Pulse Width (msec) Cooling (%) Fluence (J/cm2)
Skin Type
18
Always perform a skin test on the intended treatment before the first
treatment session according to the skin test parameters on the table below.
Skin Test Parameters - DYE-SSR Mode (In Motion Mode)
Ultrasonic gel must be applied to the skin
• End point: vessel discoloring, vessel shrinking or blanching; mild Erythema; darkening of
the pigmented lesions. No. of treatments: 3-6; treatment intervals: 2-3 weeks
• Caution: The appropriate optical density of protective eyewear for patient is: OD 5 + and
for operator: OD 3 +
• For detailed information, please refer to the Harmony XL PRO Operator’s Manual
19
12. S PEED AFT (6.4 cm²)
SHR MODE
• Hair Removal
Always perform a skin test on the intended treatment before the first
treatment session according to the skin test parameters on the table below.
Suggested Skin Test Parameters:
Ultrasonic gel must be applied to the skin
Skin Type
Fluence (J/cm2) Cooling [On/Off] Total Energy [kJ]*
(Fitzpatrick I-VI)
I 7 On 12
II 7 On 12
III 7 On 12
IV 6 On 12
V 5 On 12
VI 5 On 12
(*) In-Motion technique; the total energy is adjusted for 10x30 cm (~300cm2) area
• End point: Perifollicular, Erythema and Edema, no. of treatments 6-10, treatment
intervals; 4-8 weeks
• Caution: The appropriate optical density of protective eyewear for patient is: OD
5 + and for operator: OD 3 +
• For detailed information, please refer to the Harmony XL PRO Operator’s Manual
20
Suggested Treatment Parameters:
Ultrasonic gel must be applied to the skin
Skin Type
Fluence (J/cm2) Cooling [On/Off] Total Energy [kJ]*
(Fitzpatrick I-VI)
I 7 On 12 - 15
II 7 On 12 - 15
III 7 On 12 - 15
IV 6 On 12 - 15
V 5 On 12 - 15
VI 6 On 12 - 15
(*) The total energy is adjusted for 10x30 cm (~300cm2) area
• End point: Perifollicular, Erythema and Edema, no. of treatments 6-10, treatment
intervals; 4-8 weeks
• Caution: The appropriate optical density of protective eyewear for patient is: OD
5 + and for operator: OD 3 +
• For detailed information, please refer to the Harmony XL PRO Operator’s Manual
21
13. S PEED AFT (6.4 cm²)
HR MODE
• Hair Removal
Always perform a skin test on the intended treatment before the first
treatment session according to the skin test parameters on the table below.
Suggested Skin Test Parameters:
Ultrasonic gel must be applied to the skin
Skin Type
Pulse Width (msec) Fluence (J/cm2) Waiting Period
(Fitzpatrick I-VI)
I 30 15 30 minutes
II 30 15 30 minutes
III 40 12 30 minutes
IV 50 9 24 – 48 hours
V 50 7 24 – 48 hours
VI 50 5 24 – 48 hours
• End point: Perifollicular, Erythema and Edema, no. of treatments 6-10, treatment intervals;
4-8 weeks
• Caution: The appropriate optical density of protective eyewear for patient is: OD 5 + and
for operator: OD 3 +
• For detailed information, please refer to the Harmony XL PRO Operator’s Manual
22
Suggested Treatment Parameters:
Ultrasonic gel must be applied to the skin
Skin Type
Hair Color Fluence (J/cm2) Pulse Width (msec)
(Fitzpatrick I-VI)
Light 15 – 20 30
I
Dark 13 – 18 30
Light 15 – 20 30
II
Dark 13 – 18 30
Light 15 – 20 40
III
Dark 12 – 17 40
Light 15 – 17 40
IV
Dark 10 – 15 50
Light Up to 15 50
V
Dark Up to 13 50
Light* N/A N/A
VI
Dark Up to 12 50
(*) It is not recommended to treat hair that is lighter than the surrounding skin.
• End point: Perifollicular, Erythema and Edema, no. of treatments 6-10, treatment intervals;
4-8 weeks
• Caution: The appropriate optical density of protective eyewear for patient is: OD 5 + and
for operator: OD 3 +
• For detailed information, please refer to the Harmony XL PRO Operator’s Manual
23
14. AFT Cooled SHR Pro Module (5cm2)
SHR MODE
• Hair Removal
Always perform a skin test on the intended treatment before the first
treatment session according to the skin test parameters on the table below.
Suggested Skin Test Parameters:
Ultrasonic gel must be applied to the skin
I 7 30 100 10.0
II 7 30 100 10.0
III 7 30 100 10.0
IV 6 30 100 10.0
V 5 30 100 10.0
VI 5 30 100 10.0
• End point: Perifollicular Erythema and Edema, no. of treatments 6-10 ,treatment
intervals; 4-8 weeks
• Caution: The appropriate optical density of protective eyewear for patient is: OD 5 +
and for operator: OD 3 +
• For detailed information, please refer to the Harmony XL PRO Operator’s Manual
24
Suggested Treatment Parameters:
Ultrasonic gel must be applied to the skin
• End point: Perifollicular Erythema and Edema, no. of treatments 6-10 ,treatment
intervals; 4-8 weeks
• Caution: The appropriate optical density of protective eyewear for patient is: OD 5 +
and for operator: OD 3 +
• For detailed information, please refer to the Harmony XL PRO Operator’s Manual
25
15. A
FT Cooled SHR Pro Module (5cm2)
HR MODE
• Hair Removal
Always perform a skin test on the intended treatment before the first
treatment session according to the skin test parameters on the table below.
Suggested Skin Test Parameters:
Ultrasonic gel must be applied to the skin
I 30 100 15 30 minutes
II 30 100 15 30 minutes
III 40 100 12 30 minutes
IV 50 100 9 24 - 48 hours
V 50 100 7 24 - 48 hours
VI 50 100 5 24 - 48 hours
26
Suggested Treatment Parameters:
Ultrasonic gel must be applied to the skin
Light 100 15 - 20 30
I
Dark 100 13 - 18 30
Light 100 15 - 20 30
II
Dark 100 13 - 18 30
Light 100 15 - 20 40
III
Dark 100 12 - 17 40
Light 100 15 - 17 40
IV
Dark 100 10 - 15 50
Light 100 Up to 15 50
V
Dark 100 Up to 13 50
Light* 100 N/A N/A
VI
Dark 100 Up to 12 50
(*) It is not recommended to treat hair that is lighter than the surrounding skin.
27
16. AFT cooled SSR Module
• Skin Rejuvenation
• Vascular Lesions
• Pigmented Lesions
Always perform a skin test on the intended treatment before the first
treatment session according to the skin test parameters on the table below.
Suggested Skin Test Parameters:
Ultrasonic gel must be applied to the skin
Skin Type
Fluence (J/cm2) Cooling [%] Total Energy [kJ]*
(Fitzpatrick I-VI)
Skin Type
Fluence (J/cm2) Cooling [%] Total Energy [kJ]*
(Fitzpatrick I-VI)
• End point: Mild vessel discoloring, vessel shrinking or blanching; mild Erythema;
darkening of the pigmented lesions. no of treatments: 3-6;
• Treatment intervals: 2-3 weeks
• Caution: The appropriate optical density of protective eyewear for patient is: OD 5 +
and for operator: OD 3 +
• For detailed information, please refer to the Harmony XL PRO Operator’s Manual
28
17. AFT Acne Module
• Mild to Moderate Acne Vulgaris
Always perform a skin test on the intended treatment before the first
treatment session according to the skin test parameters on the table below.
Skin Test Parameters-Stationary Mode
Ultrasonic gel must be applied to the skin
Skin Type
Pulse Width ( msec) Fluence ( J/cm2)*
( Fitzpatrick I-VI)
29
18. AFT Cooled SST Module
• Cutaneous Lesions
• Scars and Striae
• Wrinkles Reduction
• Skin Tightening
Always perform a skin test on the intended treatment before the first
treatment session according to the skin test parameters on the table below.
Skin Test Parameters - FACE
Ultrasonic gel must be applied to the skin
Skin Type
Fluence (J/cm2) Cooling (%) Total Energy [kJ]*
(Fitzpatrick I-VI)
(*) T he number of cycles and the total energy are adjusted for a 10x15 cm (150cm2) area
using In Motion technique
Skin Type
Fluence (J/cm2) Cooling (%) Total Energy [kJ]*
(Fitzpatrick I-VI)
(*) The number of cycles and the total energy are adjusted for a 10x15 cm (150cm2) area
using In Motion technique
• End point: Erythema, no. of treatments 4-6 ,treatment intervals; 3-4 weeks
• Caution: The appropriate optical density of protective eyewear for patient is: OD 5 +
and for operator: OD 3 +
• For detailed information, please refer to the Harmony XL PRO Operator's Manual
30
Skin Test Parameters - BODY
Ultrasonic gel must be applied to the skin
Skin Type
Fluence (J/cm2) Cooling (%) Total Energy [kJ]*
(Fitzpatrick I-VI)
(*) T he number of cycles and the total energy are adjusted for a 10x15 cm (150cm2) area
using In Motion technique
Skin Type
Fluence (J/cm2) Cooling(%) Total Energy [kJ]*
(Fitzpatrick I-VI)
(*) T he number of cycles and the total energy are adjusted for a 10x15 cm (150cm2) area
using In Motion technique
• End point: Erythema, no. of treatments 4-6 ,treatment intervals; 3-4 weeks
• Caution: The appropriate optical density of protective eyewear for patient is: OD 5 +
and for operator: OD 3 +
• For detailed information, please refer to the Harmony XL PRO Operator's Manual
31
19. NIR (Near Infrared) Body Module
• Muscles and Joints Pain Relief
Always perform a skin test on the intended treatment before the first
treatment session according to the skin test parameters on the table below.
Apply treatment oil to the skin
Suggested setup
Body Skin Test Parameters
parameters
Skin Type Power Body area Total Energy Power Total Energy
( Fitzpatrick I-VI) setting (W) (cm2) (kj) Setting (w) (kj)
I 50-60 450 3.0-3.6 50-70 36-63
II 50-60 450 3.0-3.6 50-70 36-63
III 50-60 450 3.0-3.6 50-70 36-63
VI 50-60 450 3.0-3.6 50-70 36-63
V 50 450 2.4-3.0 50-70 30-50
IV 50 450 2.4-3.0 50-70 30-50
Suggested setup
Dorsal Hand Skin Test Parameters
parameters
Skin Type Power Total Energy Power Total Energy
Area (cm2)
(Fitzpatrick I-VI) setting (W) (kj) Setting (w) (kj)
I 50-60 60-100 3.0-3.6 60-70 16-18
II 50-60 60-100 3.0-3.6 60-70 16-18
III 50-60 60-100 3.0-3.6 60-70 16-18
IV 50-60 60-100 3.0-3.6 50-60 13-15
V 50 60-100 2.4-3.0 50-60 13-15
VI 50 60-100 2.4-3.0 50-60 13-15
* T he NIR module must be used In -Motion utilizing large brush strokes
* I f the patient is experiencing an uncomfortable level of heat, stop lasing for 5-10 second
and then continue treatment
*R each the therapeutic temperature of 40-44c before recording total energy
* Apply treatment oil to the skin
• End points: mild Erythema
• No. of treatments: 4-6
• Treatment intervals: 2 weeks
• Caution: The appropriate optical density of protective eyewear for patient is: OD 5 +
and for operator: OD 3.
• For detailed information, please refer to the Harmony XL PRO Operator’s Manual
32
20. NIR (Near Infrared) Face Module
• Muscles and Joints Pain Relief
Always perform a skin test on the intended treatment before the first
treatment session according to the skin test parameters on the table below.
Apply treatment oil to the skin
Suggested setup
Face Skin Test Parameters
parameters
Time of
Skin Type Power Total Power Total
Exposure
(Fitzpatrick I-VI) setting (W) Energy (kj) Setting (w) Energy (kj)
(sec)
* The Nir module must be used In-Motion utilizing large sweeping brush strokes
* Face treatment area is 10x15 cm (150cm2)
* If the patient is experiencing an uncomfortable level of heat, stop lasing for 5-10 seconds and
then continue treatment
* Reach the therapeutic temperature of 40-44°c before recording total energy
33
21. High Power UV Module
• Leukoderma-including Vitiligo
• Psoriasis
• Seborrheic Dermatitis
• Atopic Dermatitis (Eczema)
Always perform a MED on the intended treatment area a day or two before
the first treatment session according to the skin test parameters.
34
22. LED Module
• Topical Heating to Promote Increased Blood Flow
• Temporary Relaxation of Muscles
• Relief of Pain
• Reduction of Erythema
Before LED treatment procedures the patient should evaluated for the
presence of pre-existing conditions of skin disease that may cause skin
sensitivity or photosensitivity. The LED module can be operated in one of two
Modes: Continuous wave (CV) and Repeat (pulsed) wave.
VI 30/20 Repeat / CW 10
(*) Pulsed / cw
VI 30/20 Repeat / CW
35
Alma Lasers GmbH
Nordostpark 100-102 90411 Nürnberg, Germany Connect with
Tel. + 49-911 / 89-11-29-0 Fax + 49-911 / 89-11-29-99 Alma Lasers
info@alma-lasers.de | www.alma-lasers.de
© 2015 Alma Lasers Ltd.
All rights reserved. Alma Lasers, its logo, AFT, Alma, ClearLift,
ClearSkin, Dye-VL, Harmony, Impact, SHR, Speed, are
trademarks and/or registered trademarks of Alma Lasers Ltd.
in the United States and/or other countries.
info@almalasers.com | www.harmonyxlpro.com
PBHA25081404