ELECTRONIC LENGTH MEASURING EQUIPMENT - ANALOGUE

✓ TESA GT 31 Lever Probes


Models with inclinable probe insert for measuring in two directions –
DIN 32876 Well suited for use where probes with measuring bolt moving lengthwise
Part 1 can not be conveniently operated.
± 0,3 mm • Ball-bearing balanced lever.
• Interchangeable probe insert fitted with a tungsten carbide ball tip,
inclinable through to 180°.
Any position
of use • Automatic reversal of the probing direction while that of the
2 dovetail indication remains unchanged.
attachments. • Protected against shocks by 2 safety clutches.
Both lower and • One-piece housing provided with 2 dovetails.
upper stops are
fixed.
Probe inserts with removable
stainless steel shaft. Also
with a 2 mm carbide ball tip.
For all other inserts, see Measuring range (mm) N*
under optional accessories 03210802 GT 31 ± 0,3 0,1 (standard)
on next pages.
03210801 GT 31 ± 0,3 0,02
2 m long cable.
5-pin DIN 45322 connector.
03210803 GT 31 ± 0,3 0,2
* Nominal value at electrical zero; max. deviation ±25%. Valid with probe housing and lever lying
All-metal housing, horizontally, as well as in static measuring.
matt-chromium
finish

Moved mass 01840105


12 g
13 kHz (± 5%)
ø 8h6
drive frequency.

20
Highest mechanical
frequency to 25 Hz
ø2

17
20 ± 0,5°C

5 °C to 60 °C
(90)

80% 40 50

IP40
(IEC 60529) 8,54
6,4 ø1
Equipped with ,5
1,65

a 2 mm dia.
probe insert
(No. 03260410) plus L
a 8 mm dia. fixing shaft
No. 01840105.

Shipping
packaging
mm µm µm µm* Technical
Identification data sheet
number
GT 31 0,7 0,1 0,25 0,2 + 50 · L2 03200266
Declaration * Linearity related max permissible error (L in mm).
of conformity
B Measuring direction

Measuring direction
A

Shorten lever

Note
With the insert lying parallel to the workpiece surface (Fig. A), the leverage matches
1:1. Therefore, no correction of the measured values is needed.
Any other position (angle α, Fig. B) will change the effective lever length, so that read
values must be corrected. With regard to this, also report to the instructions for use
that came with your electronic probes.