GEHC-SP_DISCOVERY-RF180_2-70F

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 16

Typical

----
----
REV DATE MODIFICATIONS
01 - C1 - Cover Sheet 10 - S3 - Structural Details (1) ----
02 - C2 - Disclaimer - Site Readiness 11 - M1 - HVAC
03 - A1 - General Notes
04 - A2 - Equipment Layout
12 - E1 - Electrical Notes GE Healthcare ----
13 - E2 - Electrical Layout ----
05 - A3 - Section Views 14 - E3 - Electrical Elevations
06 - A4 - Equipment Details 15 - E4 - Details-Interconnections
07 - A5 - Equipment Details & Delivery (2)
08 - S1 - Structural Notes
16 - E5 - Power Requirements DISCOVERY RF180
09 - S2 - Structural Layout FINAL STUDY
Drawn by Verified by Concession S.O. (GON) PIM Manual Rev
A mandatory component of this drawing set is the GE Healthcare Pre Installation manual. Failure to reference the Pre Installation manual will result in
incomplete documentation required for site design and preparation.
Pre Installation documents for GE Healthcare products can be accessed on the web at: www.gehealthcare.com/siteplanning GSP PMM ---- - 5793724-1EN 10
GE does not take responsibility for any damages resulting from changes on drawings made by others. Errors may occur by not referring to the complete Format Scale File Name Date Sheet
set of final issue drawing. GE cannot accept responsibility for any damage due to the partial use of GE final issue drawings, however caused. All
dimensions are in millimeters unless otherwise specified. Do not scale from printed pdf files. GE accepts no responsibility or liability for defective work
due to scaling from these drawings. A3 1/4"=1'-0" EN-RF-TYP-DISCOVERY_RF180-NF.DWG 07/Mar/2024 01/16
DISCLAIMER CUSTOMER SITE READINESS REQUIREMENTS

GENERAL SPECIFICATIONS
REQUIRED MANUALS FOR SYSTEM PRE-INSTALLATION
· GE is not responsible for the installation of developers and associated equipment, lighting, cassette trays and Description Document Number*
protective screens or derivatives not mentioned in the order.
Product specific Pre-installation Manual Refer to cover page
· The final study contains recommendations for the location of GE equipment and associated devices, electrical
wiring and room arrangements. When preparing the study, every effort has been made to consider every *documents can be accessed in multiple languages at https://www.gehealthcare.com/support/manuals
aspect of the actual equipment expected to be installed.
· The layout of the equipment offered by GE, the dimensions given for the premises, the details provided for · A mandatory component of this drawing set is the GE HealthCare Pre-installation manual. Failure to reference
the pre-installation work and electrical power supply are given according to the information noted during the Pre-installation manual will result in incomplete documentation required for site design and preparation.
on-site study and the wishes expressed by the customer.
· The room dimensions used to create the equipment layout may originate from a previous layout and may not · The items on the GE HealthCare Site Readiness Checklists listed below are REQUIRED to facilitate equipment
be accurate as they may not have been verified on site. GE cannot take any responsibility for errors due to delivery to the site. Equipment will not be delivered if these requirements are not satisfied.
lack of information.
· Dimensions apply to finished surfaces of the room. REQUIRED SITE-READINESS CHECKLISTS FOR SYSTEM PRE-INSTALLATION
· Actual configuration may differ from options presented in some typical views or tables.
· If this set of final drawings has been approved by the customer, any subsequent modification of the site must Modality Document Number*
be subject to further investigation by GE about the feasibility of installing the equipment. Any reservations Computerized Tomography DOC2949059
must be noted.
Radiology, Radiology and Fluouroscopy, Mammography, Bone Mass
· The equipment layout indicates the placement and interconnection of the indicated equipment components. Densitometry DOC2949063
There may be local requirements that could impact the placement of these components. It remains the
All modality Customer/Contractor Worksheet DOC2949068
customer's responsibility to ensure that the site and final equipment placement complies with all applicable
local requirements. *documents can be accessed in multiple languages at https://www.gehealthcare.com/support/manuals
· All work required to install GE equipment must be carried out in compliance with the building regulations and
the safety standards of legal force in the country concerned. · Any deviation from these drawings must be communicated in writing to and reviewed by your local GE
· These drawings are not to be used for actual construction purposes. The company cannot take responsibility HealthCare installation project manager prior to making changes.
for any damage resulting therefrom.
· Make arrangements for any rigging, special handling, or facility modifications that must be made to
CUSTOMER RESPONSIBILITIES deliver the equipment to the installation site. If desired, your local GE HealthCare installation project
manager can supply a reference list of rigging contractors.
· It is the responsibility of the customer to prepare the site in accordance with the specifications stated in the
final study. A detailed site readiness checklist is provided by GE. It is the responsibility of the customer to · New construction requires the following;
ensure all requirements are fulfilled and that the site conforms to all specifications defined in the checklist and 1. Secure area for equipment,
final study. The GE Project Manager of Installation (PMI) will work in cooperation with the customer to follow 2. Power for drills and other test equipment,
up and ensure that actions in the checklist are complete, and if necessary, will aid in the rescheduling of the 3. Restrooms.
delivery and installation date.
· Prior to installation, a structural engineer of record must ensure that the floor and ceiling is designed in such a · Provide for refuse removal and disposal (e.g. crates, cartons, packing)
way that the loads of the installed system can be securely borne and transferred. The layout of additional
structural elements, dimensioning and the selection of appropriate installation methods are the sole · For CT systems it is required to minimize vibrations within the scan room. It is the customer's
responsibility of the structural engineer. Execution of load bearing structures supporting equipment on the responsibility to contract a vibration consultant/engineer to implement site design modifications to
ceiling, floor or walls are the customer's responsibility. meet the GE vibration specification. Refer to the system Pre-installation manual for vibration
specifications.
RADIO-PROTECTION
· Suitable radiological protection must be determined by a qualified radiological physicist in conformation with
local regulations. GE does not take responsibility for the specification or provision of radio-protection.

THE UNDERSIGNED, HEREBY CERTIFIES THAT I HAVE READ AND APPROVED THE PLANS IN THIS DOCUMENT.

DATE NAME SIGNATURE

Typical DISCOVERY RF180 EN-RF-TYP-DISCOVERY_RF180-NF.DWG - Rev B Date 07/Mar/2024 C2 - Disclaimer - Site Readiness 02/16
ENVIRONMENTAL SPECIFICATIONS

MAGNETIC INTERFERENCE
To guarantee specified imaging performance :
X-ray tubes and control console equipment must be located in ambient static field of less than 10 gauss.

ACOUSTIC OUTPUT
Measured 1 m from any point in system.
In-use: less than 65 dBA
Stand-by: less than 45 dBA

ALTITUDE AND ATMOSPHERIC PRESSURE


Operating atmospheric pressure: 800-1013 hPa
Operating altitude: +1949m to 0m [+6394 ft to 0 ft]
Storage atmospheric pressure: 800-1013 hPa
Storage altitude: +1949m to -50m [+6394 ft to -164 ft]
Refer to the Pre-installation Manual for detailed information about individual components.

Typical DISCOVERY RF180 EN-RF-TYP-DISCOVERY_RF180-NF.DWG Rev B Date 07/Mar/2024 A1 - General Notes 03/16
LEGEND
A GE SUPPLIED D AVAILABLE FROM GE
B GE SUPPLIED/CONTRACTOR INSTALLED E EQUIPMENT EXISTING IN ROOM
CUSTOMER/CONTRACTOR SUPPLIED AND ITEM TO BE REINSTALLED FROM
C *
INSTALLED ANOTHER SITE
MAX MAX
HEAT WEIGHT HEAT WEIGHT
BY ITEM DESCRIPTION OUTPUT (lbs) OUTPUT (kg)
(BTU/h) (W)

A'
A 1 TABLE WITH OVERFLOOR PLATE 2389 3197 700 1450
A 2 GENERATOR CABINET 3501 201 1026 91
13'-11" 6'-7" A 3 DIGITAL SYSTEM CABINET 1092 243 320 110
B 4 15kVA STEP-DOWN TRANSFORMER - 209 - 95
A 5 OPERATOR CONSOLE 194 18 57 8
A 6 MONITOR SUSPENSION BRIDGE - - - -
CONTROL
3 A 7 SINGLE MONITOR SUSPENSION 194 202 57 91.6
12
3'-5"

B 8 POWER DISTRIBUTION BOX - 176 - 80


5 A 9 4410 mm LONG RAILS - - - -
2

6'-0"
13 A 10 OTS WITH 3 m BRIDGE 1194 833 350 378
A 11 NON-TILTING WALLSTAND - 377 - 171
9 9 14 A 12 UPS - 33 - 14.8
15 A 13 TIMMS 2000 ON CART (TPC) 781 55 229 25
B' 10 B COUNTER TOP FOR EQUIPMENT- PROVIDE GROMMETED OPENINGS AS REQUIRED TO
C 14
ROUTE CABLES
C 15 CONTROL WALL TO CEILING WITH LEAD GLASS VIEWING WINDOW
1 16
19'-4"

R/F EXAM MINIMUM OPENING FOR EQUIPMENT DELIVERY IS 1200 mm x 1890 mm


C 16 [47 in x74 in], CONTINGENT ON A 2600 mm [102 in] CORRIDOR WIDTH
ROOM

EXAM ROOM HEIGHT

13'-4"
1'-10" FINISHED FLOOR TO SLAB HEIGHT TBD
FALSE CEILING HEIGHT 9'-10"
7 6
8'-6"

PLEASE NOTE THAT YOUR DISCOVERY RF180 INSTALLATION IN THE SELECTED ROOM DOES NOT
MEET THE FOLLOWING MINIMAL REQUIREMENT:
8
- 530 mm REQUIRED DISTANCE BETWEEN THE TUBE HEAD AND ANY STATIONARY OBJECT.

THEREFORE WE MUST APPLY A WARNING LABEL ON BOTH TUBE HEAD SIDES TO REMIND THE
11 OPERATOR ABOUT ENTRAPMENT HAZARD DURING GANTRY MOTIONS.
1'-5"

6'-11"
10'-5"
20'-10"
A

For Accessory Sales: (866) 281-7545 Options 1, 2, 1, 2 or mail to: gehcaccessorysales@ge.com


Typical DISCOVERY RF180 EN-RF-TYP-DISCOVERY_RF180-NF.DWG 1/4"=1'-0" Rev B Date 07/Mar/2024 A2 - Equipment Layout 04/16
FRONT VIEW A-A' EXAM ROOM CEILING HEIGHTS

RECOMMENDED AND MINIMUM ROOM HEIGHTS


CEILING HEIGHT
CONFIGURATION SPECIFICATION
mm ft
9'-10" Table Minimum 2600 8'-6"

6 10 Recommended 3000 9'-10"


9 Overhead Tube Suspension Minimum 2600 8'-6"
7
Recommended 2920 9'-7"
Monitor Suspension Minimum 2600 8'-6"
11 Tilting Wall Stand* Minimum 2600 8'-6"
Recommended 2650 8'-8"
8
Non-tilting Wall Stand Minimum 2400 7'-8"
11 Recommended 2800 9'-2"
2 3
4 1
0'-0"
Note : Minimum room height for the table must take into consideration the most protruding object
from the ceiling that is in the system area (for example the rails for OTS).

*Not available in USA

SIDE VIEW B-B'

9 10 9
9'-10"

11
4

1 0'-0"

Typical DISCOVERY RF180 EN-RF-TYP-DISCOVERY_RF180-NF.DWG 1/4"=1'-0" Rev B Date 07/Mar/2024 A3 - Section Views 05/16
TYPICAL FRONT VIEW AND TUBE ANGULATION TYPICAL FRONT VIEW AND TUBE TRAVEL

40° 40°
2460
[96.9 in]
255 975 975 255
[10 in] [38.4 in] [38.4 in] [10 in]
Rec. 3000
[9'-10"]

Rec. 3000
[9'-10"]
500 500
[19.7 in] [19.7 in]

[Min. 18 in - Max. 51 in]


Min. 956 - Max. 1020

Min. 463 - Max. 1290


Min. 463 - Max. 1290

[37.6 in - 40.1 in]


[18.3 in - 50.8 in]

1200 380 850 770 460


Center of gravity [47.2 in] [15 in] [33.5 in] [30.3 in] [18.1 in]

TYPICAL FRONT VIEW WITH TABLE 90° TYPICAL SIDE VIEW

2070
[19.7 in]
500

[81.5 in]
650
[25.6 in]

[25.6 in]
650
Rec. 3000

[Max. 116 in]


Rec. 3000

[9'-10"]
Max. 2950
[9'-10"]

[45.3 in]
Min. 482 - Max. 545

1150
175 175
[19 in - 21.5 in] [6.9 in] [6.9 in]
[99.2 in]
2520

803
[89.2 in]

[31.6 in]
2265

Min. 956 - Max. 1020


[37.6 in - 40.1 in]
[50.6 in]
1284

[18.7 in]

Center of gravity
475

Center of gravity
(with tube centered)
1204
[47.4 in]
[2.4 in]
[0.5 in]

80 1549 576.5
60
12

[3.1 in] [61 in] [22.7 in]

Typical DISCOVERY RF180 EN-RF-TYP-DISCOVERY_RF180-NF.DWG Rev B Date 07/Mar/2024 A4 - Equipment Details 06/16
GENERATOR DELIVERY

4x 13mm [0.5 in] Seismic THE CUSTOMER/CONTRACTOR SHOULD:

[11.6 in]
anchor locations.

295
· Provide an area adjacent to the installation site for delivery and unloading of the GE equipment.
· Ensure that the dimensions of all doors, corridors, ceiling heights are sufficient to accommodate the
movement of GE equipment from the delivery area into the definitive installation room.
23 459 · Ensure that access routes for equipment will accommodate the weights of the equipment and any

[2.1 in]
[.9 in] [18.1 in] transportation, lifting and rigging equipment.

54
BOTTOM VIEW · Ensure that all necessary arrangements for stopping and unloading on public or private property not
Seismic mounting plate and anchors are provided by GE. belonging to the customer have been made.
410 508
[16.1 in] [20 in]
The unit (table main frame and body) can be shipped through a 1200 mm [47 in] door width and 1200 mm [47 in]
Wiring access minimum hallway width. The unit can be divided and the major assemblies can be maneuvered through a 1020
mm [40 in] doorway and 1300 mm [51 in] hallway width (Order has to be placed accordingly, table is shipped in
one piece by default). The OTS can be shipped through a 900 mm [35 in] door width and 2600 mm [102.4 in]
205 240
minimum hallway width or through a 1400 mm [55 in] door width and 1800 mm [71 in] minimum hallway width.
[8.1 in] [9.4 in]

[41.5 in]
1053
High tension DIMENSIONS OF DELIVERY WITH DOLLY TRANSPORT EQUIPMENT
cable access
MAJOR ASSEMBLIES LENGTH (mm) WIDTH (mm) HEIGHT (mm) WEIGHT (kg)
[18.7 in]

Table main frame and body 2460 [97 in] 1064 [41.9 in] 1859 [73.2 in] -
476

Table main frame 2460 [97 in] 650 [25.6] 1859 [73.2 in] 350 [772 lb]
AC input and
Table body 1269 [50 in] 1005 [39.6 in] 1633 [64.3 in] 620 [1367 lb]
DIN rail access
OTS (Optional) 3010 [118.5 in] 744.2 [29.3 in] - 328 [723 lb]
Center of gravity SIDE VIEW FRONT VIEW Monitor Suspension (Optional) 3010 [118.5 in] 744.2 [29.3 in] - 328 [723 lb]
SCALE 1:20

DIGITAL SYSTEM CABINET FOCAL SPOT TRAVEL


805
[31.7 in]
KALOS SUSPENSION
[4.9 in]

Seismic mounting plates and anchors are provided by GE.


125

314 X-Ray head


[12.4 in] travel area
Optional
[22 in]
560

cable inlet
[10.6 in]

Cable inlet
269

[4.9 in]
125

[85 in or 109* in]


[113 in or 136* in]
TOP VIEW

2170 or 2770*
2860 or 3460*
2x 12mm [0.47 in] Seismic
floor anchor locations. Column
250 250 travel area
[9.8 in] [9.8 in] 2x 12mm [0.47 in] Seismic
656
[25.8 in] wall anchor locations.

MIN. 100
[32.7 in]

[3.9 in]
[37.5 in]

[35.5 in]

830
952

901

3540 or 4240**
[139.4 in or 166.9** in]
WALL

[17.4 in]

4398 or 5098**
443

[173.2 in or 200.7** in]


*Distance depends on bridge length
**If equipped with 700 mm rail extension option.
Center of gravity SCALE 1:50
SIDE VIEW FRONT VIEW
SCALE 1:20
Typical DISCOVERY RF180 EN-RF-TYP-DISCOVERY_RF180-NF.DWG Rev B Date 07/Mar/2024 A5 - Equipment Details & Delivery (2) 07/16
STRUCTURAL NOTES CEILING REQUIREMENTS
· Methods of support for the steelwork that will permit attachment to structural steel or through bolts in
concrete construction should be favored. Do not use concrete or masonry anchors in direct tension. The construction and load capability of the ceiling must be sufficient for the installation of the OTS unit.
· All units that are wall mounted or wall supported are to be provided with supports where necessary. Wall
supports are to be supplied and installed by the customer or his contractors. See plan for suggested locations. It is the responsibility of the customer or structural engineer to evaluate the best fixation system to be
· Control walls shall be constructed to minimum 2130mm (7'-0") high. used for the ceiling rails/mounting systems as described in the following layout example under
· Dimensions are to finished surfaces of room. consideration of the options to be installed.
· Customers contractor must provide all penetrations in post tension floors.
· Customers contractor must provide and install any non-standard anchoring. Documents for standard
anchoring methods are included with GE equipment drawings for geographic areas that require such A good ceiling levelness is required. For OTS, a maximum of 7 mm height displacement along the ceiling
documentation. longitudinal rails is allowed and can be corrected with the use of shims to ensure the rails are as close
· Customers contractor must provide and install hardware for "through the floor" anchoring and/or any bracing
to perfect level as possible.
under access floors. This contractor must also provide floor drilling that cannot be completed because of an
obstruction encountered while drilling by the GE installer such as rebar etc.
· It is the customer's responsibility to perform any floor or wall penetrations that may be required. The It is recommended that sprinkler heads not be placed between the stationary rails. All sprinkler heads should be
customer is also responsible for ensuring that no subsurface utilities (e.g., electrical or any other form of mounted so they do not extend downward more than 6.35 mm from the ceiling while in the 'resting' position.
wiring, conduits, piping, duct work or structural supports (i.e. post tension cables or rebar)) will interfere or
come in contact with subsurface penetration operations (e.g. drilling and installation of anchors/screws)
performed during the installation process. To ensure worker safety, GE installers will perform surface
penetration operations only after the customer's validation and completion of the "GE surface penetration
permit".
· Different anchor types are used to install the components of the system. Refer to Structural Requirements
Section(s) of the Pre-Installation Manual for each anchor requirement.
· Refer to the Structural Requirements Section for the required minimum embedment.
· The ground surface must be flat and leveled, maximum tolerance for leveling is ±1.5 mm per 1 m (0.2 in per 10
feet). A grout pad provided by the contractor is required to meet this specification. The maximum pad
thickness is 6.3 mm (0.25 in).

Typical DISCOVERY RF180 EN-RF-TYP-DISCOVERY_RF180-NF.DWG Rev B Date 07/Mar/2024 S1 - Structural Notes 08/16
ITEM DESCRIPTION

(GE SUPPLIED / CONTRACTOR INSTALLED)


1 Area occupied by GE supplied table baseplate
2 Area occupied by GE supplied wall stand baseplate
3 Area occupied by GE supplied transformer
(CONTRACTOR SUPPLIED & INSTALLED)
4 Support backing, locate as shown.

Structural support in ceiling for fastening ceiling supported equipment. Supports to run
continuous with no fittings extending below face of channel, run wall to wall, be parallel,
square, and in the same horizontal plane, flush with the finished ceiling. Rails are mounted to
5
3'-3" 6'-434" these supports every 26.0" (660mm) and require 452 lbs. (205 kg) per bolt load. Methods of
support that permit attachment to structural steel or through bolts in concrete should be
1'-6" 2'-6" favored. Do not use screw anchors in direct tension.

4'-2"
2'-0" 1
5
R/F EXAM
ROOM

1'-10" 9'-6"
11'-2"

5
4
5'-11"

2
3'-6"
2'-2"

3
0'-6"

0'-10"
0'-6"

9'-7"
11'-2"

Typical DISCOVERY RF180 EN-RF-TYP-DISCOVERY_RF180-NF.DWG 1/4"=1'-0" Rev B Date 07/Mar/2024 S2 - Structural Layout 09/16
TABLE FLOOR MOUNTING SUSPENSION RAILS MOUNTING SPECIFICATIONS

OVERFLOOR PLATE Unit weight: 1025 kg [2260 lb] STRUCTURAL RAILS PERPENDICULAR TO ANCHORING DETAILS
Plate weight: 180 kg [397 lb] 170
Total load: 1205 kg [2657 lb] SUSPENSION RAILS [6.69 in]
Loading area: 1.90 m² [20.5 ft²] 25 120 25
85 220 Floor loading: 805kg / m² [165 lb / ft²] 700 700 700 700 700 700 [0.98 in] [4.72 in] [0.98 in]
[3.3 in] [8.7 in] Screw M12x35 [27.56 in] [27.56 in] [27.56 in] [27.56 in] [27.56 in] [27.56 in]
TABLE BASE FIXATION Unistrut or
605 605 12 Spring washer equivalent rail

[0.5 in]

[1.77 in]
[23.8 in] [23.8 in] [0.5 in] Ø 12 [Ø 0.5 in] (not supplied

12

45
Plain washer Ø 12
by GE)
[Ø 0.5 in]
Unistrut
[min. 13 in/rec. 20 in]

Cable
min. 336/rec. 500

Bracket H-45

[23.6 in]
entrance

600

[0.4 in]
Bolt M10x30

10

[76.77 in]

[3.35 in]
1950
Max load on
[43.3 in]

[39.4 in]

85
Floor plate
1100

1000
Base fixation each bolt:

[5.51 in]
Critical Table base
(8 holes) 205kg [452 lb]

140
flatness NOT TO SCALE
1000 region Longitudinal
[39.4 in] 30 ceiling rails
1700 TABLE FLOOR PLATE FIXATION [1.18 in]
[66.9 in] Screw M10 85
50
[3.35 in]
[2 in]
Ø 15 Floor fixation (8 holes) SCALE 1:5 FASTENERS NOT SUPPLIED BY GE
· Rail extremities must be separated from the wall by a 150mm [6in] distance.

[5.3 in]
-Floor plate, M10 screws and washers are delivered with the system. Floor plate

135
[0.5 in]
Fire sprinklers, lights, air conditioning returns located in the suspension area must not be below the false ceiling.

[5.1 in]
·
-The floor thickness must be at least 200 mm [7.9 in].

130
12
· Unistrut or equivalent structural support must be fixed perfectly parallel and on the same plane.
-Critical flatness region must be flat and level, with deviations of no more than 1.5mm [1/16 in] · 660mm [26 in] steps are also possible.
· shims may be used along perimeter of baseplate to improve surface contact · A good ceiling levelness is required. For OTS, a maximum of 7mm [0.27 in] height displacement along the ceiling longitudinal rails is allowed
15 and can be corrected with the use of shims to ensure the rails are as close to perfect level as possible.
[0.6 in]
NOT TO SCALE SCALE 1:50
SCALE 1:20

WALL STAND FLOOR MOUNTING

WALL STAND BASE PLATE

Wall fixation D12-M6


680
[27 in] Boss fisher (2 holes)
38 72
[1 in] [3 in]

110
[4 in]
[14 in]

[10 in]
350

250

Ø 30 [Ø1 in] Floor fixation


(4 holes)

500
[20 in]

SCALE 1:10

Typical DISCOVERY RF180 EN-RF-TYP-DISCOVERY_RF180-NF.DWG Rev B Date 07/Mar/2024 S3 - Structural Details (1) 10/16
TEMPERATURE AND HUMIDITY SPECIFICATIONS

IN-USE CONDITIONS
EXAM ROOM CONTROL ROOM
Min Recommended Max Min Recommended Max
Temperature 15°C 23°C 35°C 15°C 23°C 35°C
[59°F] [73°F] [95°F] [59°F] [73°F] [95°F]
Relative humidity (1) 75% 75%
Heat dissipation max 2.5 kW [8490 BTU/hr] 0.41 kW [1399 BTU/hr]

STORAGE CONDITIONS
Temperature -10°C [14°F] to 50°C [122°F]
Relative humidity (1) 10% to 80%
Material should not be stored for more than 90 days.
(1) non-condensing

AIR RENEWAL
According to local standards.

NOTE
In case of using air conditioning systems that have a risk of water leakage it is recommended not to install it above
electric equipment or to take measures to protect the equipment from dropping water.

HEAT DISSIPATION DETAILS

STANDBY IN-USE STANDBY IN-USE


ROOM DESCRIPTION (kW) (kW) (BTU/hr) (BTU/hr)
Table 0.700 0.700 2388 2388
Generator 0.022 1.026 75 3500
Digital Systems Cabinet 0.320 0.320 1092 1092
Exam Room Partial UPS 0.050 0.050 N/A N/A
Monitor (single) 0.057 0.057 149 149
Advantech Monitor (single) 0.065 0.065 222 222
OTS 0.350 0.350 1190 1190

Typical DISCOVERY RF180 EN-RF-TYP-DISCOVERY_RF180-NF.DWG Rev B Date 07/Mar/2024 M1 - HVAC 11/16


CONNECTIVITY REQUIREMENTS ELECTRICAL NOTES

All Digital systems are equipped with Broadband fast Ethernet hardware for Service Diagnostics. The systems 1. All wires specified shall be copper stranded, flexible, thermo-plastic, color coded, cut 10 foot long at outlet
equipped with Digital Imaging are capable of placing electronic images on the Hospital image Ethernet Network boxes, duct termination points or stubbed conduit ends. All conductors, power, signal and ground, must be
(DICOM). run in a conduit or duct system. Electrical contractor shall ring out and tag all wires at both ends. Wire runs
must be continuous copper stranded and free from splices.
The Digital PC (part of the Digital subsystem) is the connectivity point between the system and the hospital. For a 1.1. Aluminum or solid wires are not allowed.
Broadband connection, it is the purchaser's responsibility to provide the connection at the Ethernet port on the 2. Wire sizes given are for use of equipment. Larger sizes may be required by local codes.
Digital PC via a Cat 5 Ethernet cable and the hospital Ethernet connection. 3. It is recommended that all wires be color coded, as required in accordance with national and local electrical
codes.
Note: System hardware is rated at 100/1000Mbs transfer rate. Hospital connections must be rated for 4. Conduit sizes shall be verified by the architect, electrical engineer or contractor, in accordance with local or
100/1000Mbs for optimal performance. One RJ45 Ethernet plus should be present in the room. national codes.
5. Convenience outlets are not illustrated. Their number and location are to be specified by others. Locate at
least one convenience outlet close to the system control, the power distritbution unit and one on each wall of
the procedure room. Use hospital approved outlet or equivalent.
6. General room illumination is not illustrated. Caution should be taken to avoid excessive heat from overhead
spotlights. Damage can occur to ceiling mounting components and wiring if high wattage bulbs are used.
Recommend low wattage bulbs no higher than 75 watts and use dimmer controls (except MR). Do not mount
lights directly above areas where ceiling mounted accessories will be parked.
7. Routing of cable ductwork, conduits, etc., must run direct as possible otherwise may result in the need for
greater than standard cable lengths (refer to the interconnection diagram for maximum usable lengths point
to point).
8. Conduit turns to have large, sweeping bends with minimum radius in accordance with national and local
electrical codes.
9. A special grounding system is required in all procedure rooms by some national and local codes. It is
recommended in areas where patients might be examined or treated under present, future, or emergency
conditions. Consult the governing electrical code and confer with appropriate customer administrative
personnel to determine the areas requiring this type of grounding system.
10. The maximum point to point distances illustrated on this drawing must not be exceeded.
11. Physical connection of primary power to GE equipment is to be made by customers electrical contractor with
the supervision of a GE representative. The GE representative would be required to identify the physical
connection location, and insure proper handling of GE equipment.
12. GEHC conducts power audits to verify quality of power being delivered to the system. The customer's
electrical contractor is required to be available to support this activity.

· All junction boxes, conduit, duct, duct dividers, switches, circuit breakers, cable tray, etc., are to be supplied
and installed by customers electrical contractor.
· Conduit and duct runs shall have sweep radius bends
· Conduits and duct above ceiling or below finished floor must be installed as near to ceiling or floor as possible
to reduce run length.
· Ceiling mounted junction boxes illustrated on this plan must be installed flush with finished ceiling.
· All ductwork must meet the following requirements:
1.Ductwork shall be metal with dividers and have removable, accessible covers.
2.Ductwork shall be certified/rated for electrical power purposes.
3.Ductwork shall be electrically and mechanically bonded together in an approved manner.
4.PVC as a substitute must be used in accordance with all local and national codes.
· All openings in raceway and access flooring are to be cut out and finished off with grommet material by the
customers contractor.
· General contractor to insert pull cords for all cable run conduits between the equipment room and the
operators control room.
· 10 foot pigtails at all junction points.
· Grounding is critical to equipment function and patient safety. Site must conform to wiring specifications
shown on this plan.

Typical DISCOVERY RF180 EN-RF-TYP-DISCOVERY_RF180-NF.DWG Rev B Date 07/Mar/2024 E1 - Electrical Notes 12/16
Item Electrical Layout Item List
1 Power Distribution Box (PDB)
2 10" x 3 1/2" [250 x 100] Surface floor duct with minimum 2 dividers
3 10" x 3 1/2" [250 x 100] Surface wall duct with minimum 2 dividers
4 10" x 3 1/2" [250 x 100] duct thru wall with minimum 2 dividers
5 10" x 3 1/2" [250 x 100] Surface wall duct to bottom of PDB with minimum 2 dividers
6 Box above ceiling - size per local code
7 Box flush in ceiling - size per local code (OTS)
8 Box flush in ceiling - size per local code (Monitor)
9 2 1/2" [64] Conduit above ceiling
10 3 1/2" [89] Conduit above ceiling
11 4" x 4" x 4" [100 x 100 x 100] Box attached to duct (TIMS Readiness Kit)
3'-1"
12 Grommeted opening (Digital Systems Cabinet)
13 Grommeted opening (Generator Cabinet)
OTS
14 Grommeted opening (Table)
12 4 15 Grommeted opening (Console)

2'-10"
C

3'-2"
3
3'-3"

13 11 3 16 Grommeted opening (Wall Stand)


17 Grommeted opening (Transformer)
15
Electrical Outlet Legend
6 3 9 7 ITEM QTY Customer/contractor supplied and installed items unless otherwise specified.
Height above floor determined by local codes unless otherwise specified.
10 System emergency off (SEO), (recommended height 1.2m [48"] above floor)
X-Ray room warning light control panel
X-Ray ON lamp (L1) - 24V
A Door interlock switch (needed only if required by state/local codes)
9 Duplex hospital grade, dedicated wall outlet 120-v, single phase power
2 Network outlet
OTS on/off switch, (recommended height 1.2m [48"] above floor)
1'-2"

OTS

14
1'-10"

5 R/F EXAM
8'-6"

ROOM
1
17 8
B
1'-1"
3'-6"

3 16

3'-1"
10'-5"

Additional Conduit Runs


(Contractor Supplied and Installed)

From To Size
Qty
(Bubble # / Item) (Bubble # / Item) In. mm
3 phase power 1 As req'd As req'd
1 Power Distribution Box
16 Transformer 1 As req'd As req'd
Emergency off 1 1/2 16
OTS On/off switch 1 1/2 16
1 Power Distribution Box
6 Generator 1 As req'd As req'd
6 Transformer 1 As req'd As req'd
Warning light 1 1/2 16
1 phase power Warning light control 1 As req'd As req'd
1 1/2 16
6 Generator
Door Switch 1 1/2 16
Typical DISCOVERY RF180 EN-RF-TYP-DISCOVERY_RF180-NF.DWG 1/4"=1'-0" Rev B Date 07/Mar/2024 E2 - Electrical Layout 13/16
6
+9'-10"
+9'-10"

T.B.D.
8 7

3
1
5 12
14
3 4
5'-0"

17 2 3 13 15

± 0'-0" ± 0'-0"
0'-9"

0'-9"
A C

+9'-10"

16 3

± 0'-0"
0'-9"

B
Typical DISCOVERY RF180 EN-RF-TYP-DISCOVERY_RF180-NF.DWG 1/4"=1'-0" Rev B Date 07/Mar/2024 E3 - Electrical Elevations 14/16
INTERCONNECTION
CONFIGURATION WITH OTS AND WALL STAND
LEGEND
DIG Digital System
Cabinet
GEN Generator Cabinet
OTS Overhead Tube
Suspension
PDB Power Distribution
Box
UPS Uninterruptible
Power Supple
WS Wall Stand

11 m
[36.1 ft]
12 m 8m
[39.4 ft] PDB [26.3 ft]

10 m 9m
[32.8 ft] [29.5 ft]
17 m
[55.8 ft]

11 m 5m
[36.1 ft] GEN [16.4 ft] UPS
2m
[6.6 ft]

11 m
[36.1 ft]
11 m or 17 m or 21 m DIG
14 m
[45.9 ft]
TYPICAL CABLE MANAGEMENT
[36.1 ft or 55.8 ft or 68.9 ft] 11 m
[36.1 ft] 19 m
[62.3 ft] DUCT ON THE FLOOR WALL DUCT
12 m 15 m
[39.4 ft] [49.2 ft]
Access Point
14 m
[45.9 ft]
(option)
Table TIMS WS Dividers as required Removable coverplate Dividers as required Removable coverplate
(option) per local code per local code
In room 18 m
OTS secondary [59.1 ft]
console
(option)
20 m
[65.6 ft] Monitor
UPS cart/suspension
(only for USA)
Console
15 m
[49.2 ft]

NOT TO SCALE

Typical DISCOVERY RF180 EN-RF-TYP-DISCOVERY_RF180-NF.DWG Rev B Date 07/Mar/2024 E4 - Details-Interconnections 15/16


POWER REQUIREMENTS POWER DISTRIBUTION

GENERATOR TYPE 65 kW 80 kW 480V Main supply


SEO (4)
POWER SUPPLY Wye 3 PHASE + G 480V ±10% 3 phase + ground
MAINS FREQUENCY 50/60 Hz ± 2%
LINE INPUT REACTIVE POWER (PEAK) 95 kVA 119 kVA
LINE INPUT ACTIVE POWER 65 kW 80 kW
SDT 480 V OUT 100 A

400 V IN Power distribution


· Line supply should come into a power distribution box (PDB) containing the protective units and controls. The WL
PDB does not require a neutral line.
14 Black
14 White box (PDB)
· The section of the supply cable should be calculated in accordance with its length and the maximum 14 Green
permissible voltage drops. 480 V 24 V 230 V 400 V 230 V
· There must be discrimination between supply cable protective device at the beginning of the installation RML1 14 Black
14 White
(main low-voltage transformer side) and the protective devices in the PDB. 14 Green (2)
SUPPLY CHARACTERISTICS
14 Black 14 Black (1)
1 phase 14 White WLC 14 White OTS
· Power input must be separated from any others which may generate transients (elevators, air conditioning, power 14 Green GEN UPS TAB
14 Green (option)
radiology rooms equipped with high speed film changers...)
· All equipment (lighting, power outlets, etc...) installed with GE system components must be powered
separately. DIG Digital system cabinet
DLK1 Door Interlock Switch
GEN Generator cabinet OTS
GROUND SYSTEM WS 12 Black
OTS Overhead Tube Suspension (Option) On/Off 12 White
(option)
· Ground resistance lower than 2 Ohms is suggested. PDB Power Distribution Box (Option)
14 Black DLK1
· Equipotential: the equipotential link will be by means of an equipotential bar. This equipotential bar should be RML1 Room light dimmer switch 14 White
connected to the protective earth conductors in the ducts of the non GE cableways and to additional SEO Emergency OFF button, located 4 ft 14 Green
equipotential connections linking up all the conducting units in the rooms where GE units are located. [1.22m] above floor
TAB Patient Table
CABLES SDT Step-down Transformer, 15kVA
Cable supplied by customer
UPS Uninterruptible Power Supply
· Power and cable installation must comply with the distribution diagram below. WLC Warning Light Controller Cable supplied by GEHC
· All cables must be isolated and flexible, cable color codes must comply with standards for electrical WL Warning Light Cable supplied by GEHC,
installation. WS Wall Stand (Option) installed by customer
· Case PDB furnished by GE: The cables for signals and remote control (Y, SEO, L...) will go to PDB with a pigtail Equipment supplied by
length of 1.5m, and will be connected during installation. Each conductor will be identified and isolated (screw Notes :
customer
connector). (1) Dry contact: "X-Ray ON", released by the system.
Max. voltage = 30 V Equipment supplied by GEHC
CABLEWAYS (2) H07RN-F cable with 6.56 ft (2m) extra length on the floor behind the
back of generator cabinet (only if required per local requirements)
The general rules for laying cableways should meet the conditions laid down in current standards and regulations, (3) Max lug size 8 AWG
with regard to: (4) Fourth wire only needed with UPS option for USA
· Protecting cables against water (cableways should be waterproof) FEEDER TABLE
· Protecting cables against abnormal temperatures (proximity to heating pipes or ducts) MINIMUM FEEDER WIRE SIZE IN mm² AND (AWG)
· Protecting cables against temperature shocks
MINIMUM FEEDER WIRE LENGTH
· Replacing cables (cableways should be large enough for cables to be replaced) INPUT VOLTAGE
15m (50') 30m (100') 46m (150') 61m (200') 77m (250') 92m (300') 107m (350') 122m (400')
· Metal cableways should be grounded.
480 VAC *35 (3) *35 (3) *35 (3) *35 (3) 35 (2) 50 (1) 70 (1/0) 70 (1/0)
* MINIMUM WIRE SIZE FOR CIRCUIT BREAKER, BASED ON RECOMMENDED OVERCURRENT PROTECTION
GENERAL NOTES
In all cases qualified personnel must verify that the feeder (at the point of take-off) and the run to the Radiology system meet all the
requirements stated in the PIM.

For a single unit installation, the minimum transformer size is 112.5kva, synthesized power feed is not acceptable. Maximum allowable
transient voltage excursions are 2.5% of rated line voltage at a maximum duration of 5 cycles and frequency of 10 times per hour.
Ground wire will be same size as power cable. Ground will run from the equipment back to the power source/main grounding point and
always travel in the same conduit with the feeders and neutral. Neutral must be terminated inside the main disconnect panel and not at any
GE cabinet.

Typical DISCOVERY RF180 EN-RF-TYP-DISCOVERY_RF180-NF.DWG Rev B Date 07/Mar/2024 E5 - Power Requirements 16/16