MEDISTER is the ideal disinfection device for laboratory waste, blood bags, drainage
systems, and filled or empty tubes (e.g. dialysis systems with dialyzers, etc.) The
residual moisture in used tubing is sufficient for safe disinfection. Completely dried
out tubing is not suited for disinfection in the MEDISTER. Therefore it is not advisable
to store used tubing (catheter sets, for instance) at room temperature for too long a
time before disinfection. We recommend to put them into the MEDITAINER container
immediately after use and disinfect the entire waste in the MEDISTER within 72
hours.
Exceptions:
WARNING!
Rigid impermeably sealed gas tight containers (e.g. glass bottles) must always
be opened before they are put into the MEDISTER device. EXPLOSION
HAZARD!
Small amounts of waste may be collected in a “special METEKA bio waste bag“ and
can be added to the other waste inside the MEDITAINER also in closed condition.
Do not fill the MEDITAINER more than three fourth and not more than one half if
you are processing mainly liquid waste in plastic bags (volume increase by inflation
of closed bags).
Maximum volume of liquids filled into the MEDITAINER – 1 liter
Maximum weight of MEDITAINER – 3 kg
Moisture in the waste is the basic prerequisite for safe disinfection and fire
prevention.
- Moist waste: (e.g. blood tubes, laboratory waste): pour 300 millilitres of water
evenly over the waste. (Hot water reduces the heating time!)
Special cases:
- Small and thin-walled closed rigid receptacles (in case they cannot be opened)
- Dry waste
NOTICE!
In case of closed containers with lyophilised germs/spores, field testing is required to
determine the efficacy of disinfection in the MEDISTER device.
It is possible to disinfect organ waste and animal corcasses in the MEDISTER 10.
After the treatment, however, these items must be disposed of according to legal
provisions (animal carcasses processing, burial, incineration)
When the MEDITAINER container has been filled according to the waste collecting
guidelines, put the protruding part of the waste bag inside the MEDITAINER, close
the lid and lock it by the straps.
If you want to treat materials/wastes not covered in this manual, please consult
METEKA company first.