Download as pdf or txt
Download as pdf or txt
You are on page 1of 4

Anachem Lab Final Exam notes

Activity 5 | Calibration of Glasswares - Fill the buret with distilled water at room
temperature.
Calibration of glassware is one of the most effective
- Displace any air bubbles from the tip.
ways to decrease systematic error.
- Adjust the water level to near the zero mark.
Procedure:
Calculation Formula
Checking Pipet Use:
o Volume delivered = final volume – initial volume
- Weigh the amount of water delivered in o True volume = mass /density
successive pipettings to practice and check o Correction factor, 10 mL = true volume – volume
precision. delivered
- Determine the precision of delivery for different o Corrected volume = volume read from buret +
volumes. correction factor at a particular volume

Similar Experiments with Buret: Activity No. 6: Serial Dilution

- Conduct similar experiments to determine Serial dilution is a series of dilutions, with the dilution
precision and for volume calibration. factor staying the same for each step.
- Measure water temperature for accurate volume
The concentration factor is the initial volume divided
calibration.
by the final solution volume.
Cleaning Glassware:
The dilution factor is the inverse of the concentration
- Rinse the buret and pipets with distilled water to factor.
check cleanliness.
- Clean glassware retains a continuous, unbroken
film of water when emptied.
- Use detergent and a buret brush for cleaning if
necessary.
- For pipets, use warm water first, then detergent
with warm water, and finally a commercial
cleaning solution if needed.
- Rinse thoroughly with tap water and then
distilled water. Formula:
- Leave the buret filled with distilled water when
not in use.
- Always check and clean volumetric glassware
before use.
- Burets, if properly rinsed and filled with distilled
water, remain clean for weeks.
- Pipets need frequent cleaning due to easy
contamination. weight percent (w/v):

Using the Buret:


- Ensure the stopcock is properly lubricated and
the bore and tip are clear.
✓ Limit of detection and limit of quantitation
✓ Mathematical derivations
Selectivity/specificity
- The selectivity of an analytical method is its
ability to measure accurately an analyte in the
presence of interferences that may be expected
to be present in the sample matrix.
Preparation of a Standard Curve - Selectivity is checked by examining
chromatographic blanks (from a sample that is
known to contain no analyte) in the expected
time window of the analyte peak.
Precision
- is the degree of agreement among individual
test results when the procedure is applied
repeatedly to multiple samplings
- Precision is measured by injecting a series of
standards or analyzing series of samples from
Analytical Method Validation multiple samplings from a homogeneous lot.
- From the measured standard deviation (SD)
Method validation and Mean values, precision as relative standard
- is a key element in the establishment of deviation (% rsd) is calculated.
reference methods and within the assessment
of a laboratory’s competence in generating
dependable analytical records. - Precision can be also calculated by using
- is a procedure of performing numerous Horwitz equation: an exponential relationship
assessments designed to verify that an between the among-laboratory relative standard
analytical test system is suitable for its intended deviation (RSDR) and Concentration (C):
reason and is capable of providing beneficial
and legitimate analytical data. - For estimation of repeatability (RSDr), is
modified to:
Analytical method validation
- is an essential requirement to perform the
Table 1. Details of Horwitz values.
chemical evaluation
method validation examine is product-specific.
Procedure
Parameters to be checked for method validation:
✓ Selectivity/specificity
✓ Precision
✓ Accuracy
✓ Linearity Accuracy
✓ Range - is the degree of agreement of test results
✓ Stability generated by the method to the true value.
- Accuracy is measured by spiking the sample - Limit of Detection (LOD) – lowest
matrix of interest with a known concentration of concentration at which the instrument can
analyte standard and analyzing the sample detect but not quantify and the noise-to-signal
using the “method being validated.” ratio for LOD should be 1:3.
- The procedure and calculation for Accuracy - Limit of Quantitation (LOQ) – lowest
(as% recovery) will be varied from matrix to concentration at which the instrument can
matrix and it will be given in respective study detect and quantify. The noise-to-signal ratio for
plan or amendment to the study plan LOQ should be 1:10
- LOD and LOQ values are calculated manually
Linearity
by taking Noise to signal ratio of the
- is its capability to elicit check consequences lowest/known concentration of linearity samples.
which might be at once, or with the aid of well- It will be expressed in μg/ml or ppm.
described mathematical adjustments, - To calculate in %, values of LOD and LOQ will
proportional to the concentration of analytes in be multiplied by 100/lowest or known
within a given range. concentration of test item (mg/L) taken to
- Linearity is determined by injecting a series of analyze that particular a.i. or impurity analysis.
standards of stock solution/diluted stock solution - Calculations of LOD and LOQ values for
using the solvent/mobile phase, at a minimum of instrument sensitivity:
five different concentrations in the range of 50–
150% of the expected working range.
- linearity graph (Concentration vs. Peak Area
Response) - Calculations of LOD and LOQ values for
method:
Range
- is the interval between the upper and lower
levels that have been demonstrated to be
Mathematical derivations
determined with precision, accuracy, and
linearity using the set method. - Prepare a series of standard solutions
- This range will be the concentration range in (minimum five concentrations covering working
which the Linearity test is done concentrations used for routine analysis)
analyze each solution minimum twice and
Stability
record the instrument response
- Many analytes readily decompose prior to
Determination of limit of detection (LOD) and limit of
chromatography investigations, for example
quantitation (LOQ):
during the preparation of the sample solutions,
during extraction, clean-up, phase transfer, and ✓ linear regression equation be Y = a + bX
during storage of prepared vials. Under these ✓ X is called the independent variable
circumstances, method development should (concentration on the X-axis)
investigate the stability of the analyte. ✓ Y is the dependent variable. (instrument
- Accuracy test takes care of stability response on Y-axis)
- It is required to mention in the method how long ✓ “a” and “b” are the regression constants
a sample after extraction can be stored before ✓ “a” is known as the intercept, b
final analysis, based on the duration taken for ✓ “b,” the slope of the line, m
accuracy test ✓ Let (X1, Y1), (X2, Y2), (X3, Y3)…(Xn, Yn)
be the set of values required to be fit in the
Limit of detection (LOD) and limit of quantitation
linear equation.
(LOQ)
- Application of a, b, and Sa to obtain limit of
detection and limit of quantitation
- When Sa is obtained for a linear calibration line,
then it provides a clear information on the
standard deviation of the “Blank” (or Control)
response from the instruments.
- The LOD and LOQ can be worked out, as given
below:

You might also like