Download as pdf or txt
Download as pdf or txt
You are on page 1of 6

5220

Chemical Oxygen Demand (COD)

Approved by Standard Methods Committee, 2022. Joint Task Group: Robin Parnell (chair), Christina Baker, Kerri Williams.

5220 A. Introduction

Chemical oxygen demand (COD) is defined as the amount of 2. Interferences and Limitations
a specified oxidant that reacts with the sample under controlled
conditions. The quantity of oxidant consumed is expressed in Oxidation of most organic compounds is 95% to 100% of the
terms of its oxygen equivalence. Because of its unique chemical theoretical value. Pyridine and related compounds resist oxida-
properties, the dichromate ion (Cr2O72-) is the specified oxidant tion and some volatile organic compounds react in proportion
in Methods 5220 B, C, and D; it is reduced to the chromic ion to their contact with the oxidant. Straight-chain aliphatic com-
(Cr3+) in these tests. Both organic and inorganic components of pounds can be more effectively oxidized using a silver sulfate
a sample are subject to oxidation, but in most cases the organic (Ag2SO4) catalyst.
component predominates and is of the greater interest. To mea- The most common interferent is the chloride ion. Chloride
sure either organic or inorganic COD alone, additional steps not reacts with silver ion to precipitate silver chloride, and thus inhib-
described here must be taken to distinguish one from the other. its the catalytic activity of silver. Bromide, iodide, and any other
COD is a defined test; the extent of sample oxidation can be reagent that inactivates the silver ion can interfere similarly. Such
affected by digestion time, reagent strength, and sample COD interferences are negative in that they tend to restrict the oxidizing
concentration. action of the dichromate ion itself. However, under the rigorous
COD often is used as a measurement of pollutants in waste- digestion procedures for COD analyses, chloride, bromide, or
water and natural waters. Other related analytical parameters iodide can react with dichromate to produce the elemental form of
are biochemical oxygen demand (BOD), total organic carbon the halogen and the chromic ion. Results then are in error on the
(TOC), and total oxygen demand (TOD). In many cases it is high side. The difficulties caused by the presence of the chloride
possible to correlate two or more of these values for a given can be overcome largely, though not completely, by complexing
sample. BOD is a measure of oxygen consumed by microor- with mercuric sulfate (HgSO4) before the refluxing procedure.
ganisms under specific conditions, TOC is a measure of organic Although 1 g HgSO4 is specified for a 50 mL sample, a lesser
carbon in a sample, and TOD is a measure of the amount of amount may be used where sample chloride concentration is
oxygen consumed by all elements in a sample when complete known to be less than 2000 mg/L, as long as a 10:1 weight ratio
(total) oxidation is achieved. of HgSO4:Cl- is maintained. For samples containing more than
In a COD analysis, hazardous wastes of mercury, hexavalent 2000 mg/L chloride concentration, sample dilution can be used if
chromium, sulfuric acid, silver, and acids are typically generated. the COD concentration is high enough to avoid dilution below the
Methods 5220 C and D reduce these waste problems, but may be detection limit. Techniques are available for measuring COD in
less accurate and less representative. waters with high chloride concentration and low levels of COD.2-4
These modified techniques include a titration procedure that pro-
1. Selection of Method vides COD values for samples containing greater than 2000 mg/L
Cl-. Determine the amount of chloride in the sample, and prepare
The open reflux method (5220 B) is suitable for a wide range a chloride solution at that concentration. Digest the prepared chlo-
of sample matrices where a large sample size is preferred. The ride solution with potassium dichromate solution and sulfuric acid
closed reflux methods (5220 C and D) are more economical in reagent, and analyze for COD. Analyze the sample the same way
the use of metallic salt reagents and generate smaller quantities and subtract the chloride COD value from the total sample COD
of hazardous waste, but require homogenization of samples con- value to determine the final sample COD concentration.
taining suspended solids to obtain reproducible results. Ampoules Ammonia and its derivatives, in the sample or generated from
and culture tubes with similar premeasured reagents are available nitrogen-containing organic matter, are sometimes oxidized.
commercially. If commercial premeasured tubes or ampoules are However, elemental chlorine reacts with these compounds.
used, follow manufacturer’s direction for sample size. Measure- Hence, corrections for chloride interferences are difficult. Ammo-
ments of sample volumes as well as reagent volumes and concen- nia has been shown to have a positive interference on COD when
trations are critical to obtain consistent results.1 using high range method 5220 D.4.5
Determine COD values of >50 mg/L O2 by using procedures Nitrite (NO2-) exerts a COD of 1.1 mg O2 per mg NO2--N.
5220 B.4a, C.4, or D.4. Procedure 5220 D.4 can also be used for Because concentrations of NO2- in waters rarely exceed 1 or
COD values <50 mg/L O2. Use procedure 5220 B.4b to deter- 2 mg/L NO2--N, the interference is considered insignificant and
mine, with lesser accuracy, COD values from 5 to 50 mg/L O2. usually is ignored. To eliminate a significant interference due to
NO2-, add 10 mg sulfamic acid for each mg NO2--N present in

https://doi.org/10.2105/SMWW.2882.103 1
5220 Chemical Oxygen Demand (COD) - B. Open Reflux Method

the sample volume used. Add the same amount of sulfamic acid (homogenize) all samples containing suspended solids before
to the reflux vessels containing the reagent water blank and analysis. If COD is to be related to BOD or TOC, ensure that all
standards. tests receive identical pretreatment. Make preliminary dilutions
Reduced inorganic species such as ferrous iron, sulfide, and for wastes containing a high COD to reduce the error inherent in
manganous manganese are oxidized quantitatively under the test measuring small sample volumes.
conditions. For samples containing significant levels of these spe-
cies, stoichiometric oxidation can be assumed from the known References
initial concentration of the interfering species and corrections
can be made to the COD value obtained. The method of standard 1. ASTM D1252-06(2020). Standard test methods for chemical oxygen
addition could be used if background concentrations of interfer- demand (dichromate oxygen demand) of water. West Conshohocken
ing inorganic species are not known. (PA): ASTM International; 2020.
The silver, hexavalent chromium, and mercury salts used in 2. Burns ER, Marshall C. Correction for chloride interference in
the COD determinations generate hazardous wastes. The greatest the chemical oxygen demand test. J Water Pollut Control Fed.
problem is in the use of mercury. If the chloride contribution to 1965;37(12):1716–1721.
3. Baumann FJ. Dichromate reflux chemical oxygen demand: Proposed
COD is negligible, HgSO4 can be omitted. Smaller sample sizes
method for chloride correction in highly saline wastes. Anal Chem.
(see Table 5220:1) reduce the waste. Recovery of the waste mate- 1974;46(9):1336–1338.
rial may be feasible if allowed by regulatory authority.6 4. Saral A, Goncaloğlu BI. Determination of real COD in highly chlori-
nated wastewaters. Clean Soil Air Water. 2008;36(12):996–1000.
3. Sampling and Storage 5. Kim BR. Effect of ammonia on COD analysis. Journal Water Pollut
Control Fed. 1989;61(5):614–617.
Samples may be collected in plastic or glass bottles. Test unsta- 6. Holm TR. Treatment of spent chemical oxygen demand solutions for
ble samples without delay. If a delay before analysis is unavoid- safe disposal. Technical research report prepared for Illinois State
Water Survey. Champaign (IL): Illinois; Hazardous Waste Research
able, preserve the sample by acidification to pH <2 using H2SO4.
and Information Center; 1996. http://hdl.handle.net/2142/2035
The holding time for acidified COD samples is 28 days. Blend

5220 B. Open Reflux Method

1. General Discussion dried at 150 °C for 2 h, in reagent water and dilute to 1000 mL.
This reagent undergoes a 6-electron reduction reaction; the equiv-
a. Principle: Most types of organic matter are oxidized by a boil- alent concentration is 6 × 0.04167 M or 0.2500 N.
ing mixture of chromic and sulfuric acids. A sample is refluxed in a b. Sulfuric acid reagent: Add Ag2SO4, reagent or technical
strong acidic solution with a known excess of potassium dichromate grade, crystals or powder, to conc H2SO4 at the rate of 10.1 g
(K2Cr2O7). After digestion, the remaining unreduced K2Cr2O7 is Ag2SO4 per liter of H2SO4. Let stand 1 to 2 d to dissolve. Mix.
titrated with ferrous ammonium sulfate to determine the amount Caution: Solution is highly reactive. Use appropriate personal
of K2Cr2O7 consumed and the oxidizable matter is calculated in protective equipment and a chemical fume hood.
terms of oxygen equivalent. Keep ratios of reagent weights, vol- c. Ferroin indicator solution: Dissolve 1.485 g 1,10-phenanthroline
umes, and strengths constant when using sample volumes other monohydrate and 695 mg FeSO4 ⋅ 7H2O in reagent water and dilute
than 50 mL. The standard 2-h reflux time may be reduced if it has to 100 mL. This indicator solution may be purchased already pre-
been shown that a shorter period yields the same results. pared.
b. Quality control (QC): The QC practices considered to be d. Standard ferrous ammonium sulfate (FAS) titrant, approxi-
an integral part of each method are summarized in Table 5020:1. mately 0.25 M: Dissolve 98 g Fe(NH4)2(SO4)2 ⋅ 6H2O in reagent
water. Add 20 mL conc H2SO4, cool, and dilute to 1000 mL.
2. Apparatus Standardize this solution daily against a standard K2Cr2O7 solu-
tion as follows:
a. Reflux apparatus, consisting of 500-or 250-mL Erlenmeyer Dilute 25.00 mL standard K2Cr2O7 to about 100 mL. Add
flasks with ground-glass 24/40 neck and 300-mm jacket Liebig, 30 mL conc H2SO4 and cool. Titrate with FAS titrant using 0.10
West, or equivalent condenser with 24/40 ground-glass joint, and to 0.15 mL (2 to 3 drops) ferroin indicator.
a hot plate having sufficient power to produce at least 1.4 W/cm2
of heating surface, or equivalent. molarity of FAS solution =
b. Blender or homogenizer. volume 0.04167 M K 2Cr2 O7 solution titrated, mL
c. Pipets, Class A and wide-bore. ´0.2500
volume FAS used in titration, mL
d. Buret.
e. Mercuric sulfate (HgSO4), crystals or powder.
3. Reagents f. Sulfamic acid: Required only if the interference of nitrites is
to be eliminated (see 5220 A.2).
a. Standard potassium dichromate solution, 0.04167 M: Dis- g. Potassium hydrogen phthalate (KHP) standard, HOOC-
solve 12.259 g K2Cr2O7, primary standard grade, previously C6H4COOK: Lightly crush and then dry KHP to a constant

https://doi.org/10.2105/SMWW.2882.103 2
5220 Chemical Oxygen Demand (COD) - C. Closed Reflux, Titrimetric Method

weight at 110 °C. Dissolve 425 mg in reagent water and dilute to the refluxing flask open to the atmosphere without the condenser
1000 mL. KHP has a theoretical COD of 1.176 mg O2 per mil- attached. Compute the amount of HgSO4 to be added (before
ligram and this solution has a theoretical COD of 500 mg/L O2. concentration) on the basis of a weight ratio of 10:1, HgSO4:Cl-,
This solution is stable when refrigerated in an amber glass con- using the amount of Cl- present in the original volume of sample.
tainer for up to 3 months. Be alert to the development of visible Carry a blank reagent through the same procedure. This technique
biological growth and discard if present. has the advantage of concentrating the sample without significant
losses of easily digested volatile materials. Hard-to-digest volatile
4. Procedure materials such as volatile acids are lost, but an improvement is
gained over ordinary evaporative concentration methods.
a. Treatment of samples with COD of >50 mg/L O2: Blend the c. Determination of standard solution: Evaluate the technique
sample if necessary and pipet 50.00 mL into a 500-mL refluxing and quality of reagents by conducting the test on a standard potas-
flask. For samples with a COD of >900 mg/L O2, use a smaller sium hydrogen phthalate solution.
portion diluted to 50.00 mL. Add 1 g HgSO4, several glass beads,
and very slowly add 5.0 mL sulfuric acid reagent, with mixing to 5. Calculation
dissolve HgSO4. Cool while mixing to avoid a possible loss of
volatile materials. Add 25.00 mL 0.04167 M K2Cr2O7 solution
and mix. Attach the flask to the condenser and turn on the cooling ( A − B)× M ×8000
COD as mg/L O2 =
water. Add 70.0 mL sulfuric acid reagent through the open end of mL of sample
condenser. Continue swirling and mixing while adding the sulfu-
ric acid reagent. Caution: Mix thoroughly before applying heat where:
to prevent local heating of the flask bottom and a possible
B = mL FAS used for sample,
blowout of flask contents.
A = mL FAS used for blank,
Cover the open end of the condenser with a small beaker to
M = molarity of FAS, and
prevent foreign material from entering, and reflux for 2 h. Cool
8000 = milliequivalent weight of oxygen × 1000 mL/L.
and wash down the condenser with reagent water. Disconnect the
reflux condenser and dilute the mixture to about twice its volume
6. Precision and Bias
with reagent water. Cool to room temperature and titrate excess
K2Cr2O7 with FAS, using 0.10 to 0.15 mL (2 to 3 drops) ferroin
indicator. Although the quantity of ferroin indicator is not critical, A set of synthetic samples containing potassium hydrogen
use the same volume for all titrations. Take as the endpoint of phthalate and NaCl was tested by 74 laboratories. At a COD of
the titration the first sharp color change from blue-green to red- 200 mg/L O2 in the absence of chloride, the standard deviation was
dish brown that persists for 1 min or longer. In the same manner, ±13 mg/L (coefficient of variation, 6.5%). At a COD of 160 mg/L
reflux and titrate a blank containing the reagents and a volume of O2 and 100 mg/L Cl-, the standard deviation was ±14 mg/L (coef-
reagent water equal to that of sample. ficient of variation, 10.8%).
b. Alternate procedure for low-COD samples (<50 mg/L): Fol-
low the procedure of paragraph a above, with two exceptions: (i) Bibliography
use standard 0.004167 M K2Cr2O7 or 0.025 N, and (ii) titrate with
standardized 0.025 M FAS. Exercise extreme care with this pro- Moore WA, Kroner RC, Ruchhoft CC. Dichromate reflux method for
cedure because even a trace of organic matter on the glassware or determination of oxygen consumed. Anal Chem. 1949;21(8):953–957.
Medalia AI. Test for traces of organic matter in water. Anal Chem.
from the atmosphere may cause gross errors. If a further increase
1951;23(9):1318–1320.
in sensitivity is required, concentrate a larger volume of sample Moore WA, Ludzack FJ, Ruchhoft CC. Determination of oxygen-consumed
before digesting under reflux as follows: Add all reagents (100 mL values of organic wastes. Anal Chem. 1951;23(9):1297–1300.
in total) to a sample larger than 50 mL (final volume greater than Dobbs RA, Williams RT. Elimination of chloride interference in the
150 mL) and reduce the total volume to 150 mL by boiling in chemical oxygen demand test. Anal Chem. 1963;35(8):1064–1067.

5220 C. Closed Reflux, Titrimetric Method

1. General Discussion This procedure is applicable to COD values between 40 and


400 mg/L. Obtain higher values by dilution. Alternatively, use
a. Principle: See 5220 B.1a. higher concentrations of dichromate digestion solution to deter-
b. Interferences and limitations: See 5220 A.2. Volatile organic mine greater COD values. COD values of 100 mg/L or less can
compounds are more completely oxidized in the closed system be obtained by using a more dilute dichromate digestion solution
because of longer contact with the oxidant. Before each use or a more dilute FAS titrant. Overall accuracy can be improved by
inspect culture-tube caps for breaks in the PTFE liner. Select a using an FAS titrant that is less than the 0.10 M solution specified
culture-tube size according to manufacturer specifications, block below. Higher dichromate concentrations or reduced FAS con-
heater capacity, and degree of sensitivity desired. centrations probably require titrations be performed in a separate

https://doi.org/10.2105/SMWW.2882.103 3
5220 Chemical Oxygen Demand (COD) - C. Closed Reflux, Titrimetric Method

vessel, rather than in the digestion vessel, because of the volumes Table 5220:1. Example Sample and Reagent Quantities for Various
of titrant required. Digestion Vessels
c. Quality control (QC): The QC practices considered to be an Sulfuric
integral part of each method are summarized in Table 5020:1. Digestion Acid Total Final
Sample Solution Reagent Volume
2. Apparatus Digestion Vessel (mL) (mL) (mL) (mL)
Culture tubes:
a. Digestion vessels: Preferably use borosilicate culture tubes, 16 × 100 mm 2.50 1.50 3.5 7.5
16- × 100-mm, 20- × 150-mm, or 25- × 150-mm, with PTFE- 20 × 150 mm 5.00 3.00 7.0 15.
lined screw caps. Alternatively, use borosilicate ampoules, 10-mL 25 × 150 mm 10.0 6.00 14. 30.
capacity, 19- to 20-mm diam. Standard 10-mL 2.50 1.50 3.5 7.5
Digestion vessels with premixed reagents and other accessories ampoules
are available from commercial suppliers. Contact the supplier for
specifications.
b. Block heater or similar device to operate at 150 ± 2 °C, with and reagent volumes. The most critical volumes are of the sam-
holes to accommodate digestion vessels. The use of culture tubes ple and digestion solution. Make volumetric measurements as
probably requires the caps to be outside the vessel to protect the accurate as practical; use Class A volumetric ware. The use of
caps from heat. Caution: Do not use an oven because leak- hand-held pipettors with nonwetting (polyethylene) pipet tips is
ing samples may generate a corrosive and possibly explosive practical and adequate. Use a buret for titrations. Measure H2SO4
atmosphere. Also, culture tube caps may not withstand the to ±0.1 mL. Place the sample in a culture tube or ampoule and
150 °C temperature in an oven. add the digestion solution. Carefully add sulfuric acid reagent by
c. Microburet. running it down the inside of the vessel so an acid layer is formed
d. Ampoule sealer: Use only a mechanical sealer to ensure under the sample-digestion solution layer. Tightly cap the tubes
strong, consistent seals. or seal the ampoules, and invert each several times to mix com-
pletely. Caution: Solution is highly reactive. Use appropriate
personal protective equipment and a chemical fume hood.
3. Reagents
Mix thoroughly before applying heat to prevent local heating
of the vessel bottom and a possibly explosive reaction.
a. Standard potassium dichromate digestion solution,
Place tubes or ampoules in a block digester preheated to
0.01667 M or 0.10 N: To about 500 mL reagent water, add 4.903 g
150 °C and reflux for 2 h behind a protective shield. Caution:
K2Cr2O7 (primary standard grade, previously dried at 150 °C for
These sealed vessels may be under pressure from gases
2 h), 167 mL conc H2SO4, and 33.3 g HgSO4. Dissolve, cool to
generated during digestion. Wear face and hand protec-
room temperature, and dilute to 1000 mL.
tion when handling. If sulfuric acid is omitted or reduced
b. Sulfuric acid reagent: See 5220 B.3b.
in concentration, very high and dangerous pressures will be
c. Ferroin indicator solution: See 5220 B.3c. Dilute this reagent
generated at 150 °C. Cool to room temperature and place the
by a factor of 5 (1 + 4).
vessels in a test tube rack. Some mercuric sulfate may precip-
d. Standard ferrous ammonium sulfate titrant (FAS), approx-
itate, but this does not affect the analysis. Remove the culture
imately 0.10 M: Dissolve 39.2 g Fe(NH4)2(SO4)2 ⋅ 6H2O in
tube caps and add a small PTFE-covered magnetic stirring bar.
reagent water. Add 20 mL conc H2SO4, cool, and dilute to 1000 mL.
If ampoules are used, transfer the contents to a larger container
Standardize the solution daily against standard K2Cr2O7 diges-
for titrating. Add 0.05 to 0.10 mL (1 to 2 drops) ferroin indi-
tion solution as follows:
cator and stir rapidly on a magnetic stirrer while titrating with
Pipet 5.00 mL digestion solution into a small beaker. Add 10 mL
standardized 0.10 M FAS. The endpoint is a sharp color change
reagent water to substitute for sample. Cool to room temperature.
from blue-green to reddish brown, although the blue-green may
Add 1 to 2 drops diluted ferroin indicator and titrate with FAS
reappear within minutes. In the same manner, reflux and titrate
titrant.
a blank containing the reagents and a volume of reagent water
equal to that of the sample.
molarity of FAS solution =
volume 0.01667 M K 2Cr2 O7 solution titrated, mL 5. Calculation
´0.1000
volume FAS used in titration, mL

e. Sulfamic acid: See 5220 B.3f ( A − B)× M ×8000


COD as mg/L O2 =
f. Potassium hydrogen phthalate standard: See 5220 B.3g. mL of sample
g. Mercuric sulfate (HgSO4): Note that mercuric sulfate is not
required if chloride is present. where:
B = mL FAS used for sample,
4. Procedure A = mL FAS used for blank,
M = molarity of FAS, and
Wash culture tubes and caps with 20% H2SO4 before first use 8000 = milliequivalent weight of oxygen × 1000 mL/L.
to prevent contamination. Refer to Table 5220:1 for proper sample

https://doi.org/10.2105/SMWW.2882.103 4
5220 Chemical Oxygen Demand (COD) - D. Closed Reflux, Colorimetric Method

6. Precision and Bias of 195 mg/L O2 in the absence of chloride, the standard deviation
was ±11 mg/L O2 (coefficient of variation, 5.6%). At an average
Sixty synthetic samples containing potassium hydrogen phthal- COD of 208 mg/L O2 and 100 mg/L Cl-, the standard deviation
ate and NaCl were tested by 6 laboratories. At an average COD was ±10 mg/L O2 (coefficient of variation, 4.8%).

5220 D. Closed Reflux, Colorimetric Method

1. General Discussion 3. Reagents

a. Principle: See 5220 B.1a. When a sample is digested, the a. Digestion solution, high range: Add to about 500 mL reagent
dichromate ion oxidizes COD material in the sample. This results water 10.216 g K2Cr2O7, primary standard grade, previously
in the change of chromium from the hexavalent (VI) state to the dried at 150 °C for 2 h, 167 mL conc H2SO4, and 33.3 g HgSO4
trivalent (III) state. Both of these chromium species are colored (0.21 N or 0.0035 M potassium dichromate solution). Dissolve,
and absorb in the visible region of the spectrum. The dichromate cool to room temperature, and dilute to 1000 mL.
ion (Cr2O72-) absorbs strongly in the 400-nm region, where the b. Digestion solution, low range: Prepare as in paragraph
chromic ion (Cr3+) absorption is much less. The chromic ion a above, but use only 1.022 g K2Cr2O7 creating a 0.021 N or
absorbs strongly in the 600-nm region, where the dichromate 0.0035 M solution of potassium dichromate.
has nearly zero absorption. In a 9 M sulfuric acid solution, the c. Sulfuric acid reagent: See 5220 B.3b.
approximate molar attenuation coefficients for these chromium d. Sulfamic acid: See 5220 B.3f.
species are as follows: e. Potassium hydrogen phthalate standard: See 5220 B.3g.
• Cr3+ – 50 L/mol cm at 604 nm; f. Mercuric sulfate (HgSO4): Note that mercuric sulfate is not
• Cr2O72- – 380 L/mol cm at 444 nm; required if chloride is present.
• Cr3+ – 25 L/mol cm at 426 nm.
Low range COD values of around 90 mg/L or less (depending 4. Procedure
on the manufacturer) can be determined by measuring a decrease
in Cr2O72- at 420 nm. The corresponding generation of Cr3+ gives Follow the manufacturer’s instructions for the treatment of
a small absorption increase at 420 nm, but this is compensated for samples, blanks and calibration when using a commercial COD
in the calibration procedure. product.
For high range COD values around 100 to 1500 mg/L (depending a. Treatment of samples: Measure a suitable volume of sample and
on the manufacturer), an increase in the Cr3+ absorbance at 600 nm reagents into the tube or ampoule as indicated in Table 5220:1. Pre-
is determined. Commercially premixed reagent digestion tubes are pare, digest, and cool the samples, blank, and one or more standards
available for low and high range COD that may provide a larger as directed in 5220 C.4. Note the safety precautions. It is critical that
working COD range due to varying reagent concentration ratio to the volume of each component be known and that the total volume
sample volume. Higher values can be obtained by sample dilution. be the same for each reaction vessel. If volumetric control is difficult,
b. Interferences and limitations: See 5220 C.1b. transfer the digested sample, dilute to a known volume, and read.
For this procedure to be applicable, all visible light-absorbing Premixed reagents in digestion tubes are available commercially.
interferents must be absent or be compensated for. This includes b. Measurement of dichromate reduction: Cool the sample to
insoluble suspended matter as well as colored components. If room temperature, and when samples are cooled, vent, if nec-
either type of interference occurs, the test is not necessarily lost essary, to relieve any pressure generated during digestion. Mix
because COD can be determined titrimetrically as in 5220 C. If the contents of the reaction vessels to combine condensed water
not corrected, chloride interference may result in cloudy digestate and dislodge insoluble matter. Let the suspended matter settle
inside the reagent tube, leading to inaccurate results. and ensure that the optical path is clear. Measure the absorption
c. Quality control (QC): The QC practices considered to be an of each sample blank and standard at the selected wavelength
integral part of each method are summarized in Table 5020:1. (420 nm or 600 nm). High range COD is measured at or near
600 nm and low range COD is measured at or near 420 nm.
2. Apparatus For high range COD, use a digested blank as the reference solu-
tion to zero the spectrophotometer. Measure all samples, blanks,
a. See 5220 C.2. Ensure that reaction vessels are of optical and standards against this solution. Monitor the absorbance of the
quality. Other types of absorption cells with varying path lengths digested blank by zeroing on reagent water added to an empty
may be used. Use the attenuation coefficients of the ions of inter- vial and then reading the digested blank. Any marked increase in
est for this approach. blank absorbance must be investigated. This is especially import-
b. Spectrophotometer, colorimeter, or photometer for use at ant when a new lot of reagents is used.
or near 600 nm or 420 nm with access opening adapter for an For low range COD, use reagent water as the reference solution
ampoule or 16-, 20-, or 25-mm tubes. Verify that the instrument to zero the spectrophotometer. Measure all samples, blanks, and
operates in the region of 420 nm and 600 nm. Values slightly dif- standards against this solution. Analyze a digested blank to ensure
ferent from these may be found, depending on the spectral band- the reagent quality and to determine the reagents’ contribution
pass of the instrument. to the decrease in absorbance during the digestion. Subtract the

https://doi.org/10.2105/SMWW.2882.103 5
5220 Chemical Oxygen Demand (COD) - D. Closed Reflux, Colorimetric Method

digested blank absorbance from the standards and samples absor- 6. Precision and Bias
bance to determine corrected absorbance. Plot the standard cor-
rected absorbance when preparing the calibration curve and use Forty-eight synthetic samples containing potassium hydrogen
the sample corrected absorbance to compare to the calibration phthalate and NaCl were tested by 5 laboratories. At an average
curve to determine the sample’s COD concentration. COD of 193 mg/L O2 in the absence of chloride, the standard
c. Preparation of calibration curve: Prepare at least 5 stan- deviation was ±17 mg/L O2 (coefficient of variation 8.7%). At
dards from the potassium hydrogen phthalate solution with COD an average COD of 212 mg/L O2 and 100 mg/L Cl-, the stan-
equivalents to cover each concentration range. Fill to desired vol- dard deviation was ±20 mg/L O2 (coefficient of variation, 9.6%).
ume with reagent water. Use the same reagent volumes, tube, or Additional QA/QC data for both high-and low-level procedures
ampoule size, and digestion procedure as for samples and mea- may be found elsewhere.1
sure absorbance. Prepare a calibration curve for each new lot of
tubes or ampoules. See Section 5020 1a and 1b for calibration Reference
criteria. When using a manufacturer’s spectrophotometer with a
factory calibration, verify the calibration daily using check stan- 1. ASTM D1252-06(2020). Standard test methods for chemical oxygen
dards of varying concentrations. Follow the manufacturer instruc- demand (dichromate oxygen demand) of water. West Conshohocken
tions for the use of the spectrophotometer. (PA): ASTM International; 2020.

5. Calculation Bibliography

To create a standard curve, plot absorbance versus COD concen- Jirka AM, Carter MJ. Micro semiautomated analysis of surface and waste-
tration in milligrams per liter. For low range calibration, correlation waters for chemical oxygen demand. Anal Chem. 1975;47(8):1397–
has a negative slope and for high range calibration, correlation has a 1402.
Himebaugh RR, Smith MJ. Semi-micro tube method for chemical oxy-
positive slope. Compare the instrument absorbance to the standard
gen demand. Anal Chem. 1979;51(7):1085–1087.
curve to determine the COD concentration of standards and sam-
ples. Apply the appropriate dilution factor when necessary.

Published Online: August 27, 2018


Revised: April 25, 2022
https://doi.org/10.2105/SMWW.2882.103 6

You might also like