DVD1XN1J Ver.23.

00-00 (China markets only)


<Applicable to S/N> <Release Date>
Refer to ECR 322G056 February 2023

<Parts Information>
CV891460: CDR_ASSY NO.527 (DVD1XN1J(23.00-00))

Changed Specifications
1 The NEUT/EO classification algorithm of XN-10x has been changed to be the same as that of XN-10.
2 Now IG has been added as a reportable parameter on XN-10x.
3 On XN-10x, now PLT-F license can be added, and IPF has been changed to a reportable parameter.
4 Decision of RET Abn Scattergram has beefed up.
5 Decision of PLT Abn Scattergram has beefed up.
6 When a new stain cartridge is loaded, the opening date is written on the RFID, but this process
overlaps with writing the number of remaining tests, which can cause the analyzer to aberrate.
Therefore, the software has been modified to prevent double writing to RFID.
7 In Ver.22.19-00, service settings to use barcode reader F-100 were added. In addition to these
settings, adoption logic for sample ID read by barcode reader has been added to further increase the
reading rate of low quality labels. New barcode reader setting below has also been added in service
menu.

Barcode reader setting (1) Barcode reader setting (2)

Resolved Program issues

1 Resolved the issue that the suspended sampler measurement could not be resumed when the mode
switch was pressed to switch to the sampler measurement mode after the manual measurement was
completed only when SA-01 was used as the sampler.
Remarks
Module Version
[IPU] 00-22 Build205
[Device Control Manager] 00-22 Build205
[Analysis] 00-22 Build89
[Analyzer Sequencer] 00-22 Build205
[Sampler Sequencer] 00-22 Build205

Sequence Program Version


[Analyzer] 00-69
[Sampler] 00-16

ECR 322G056
DVD1XN1J Ver.22.11-00 (China markets only)
<Applicable to S/N> <Release Date>
Refer to ECR 322B052 September 2022

<Parts Information>
BV341672: CDR_ASSY NO.377 (DVD1XN1J(22.11-00))

<FDS File Name>


P_XNSWSTCN_22110001.cab (Including installer)
P_XNSWSTCN_22110000.cab (VHD only)

Changed Specifications
[Horizontal development of Global version Ver.22.09-00]
1 The XN-1590 configuration has been added.
2 The analysis channel, used as adopted PLT and WBC result (“switching”), can now be displayed on
“Cumulative” and HPC screen.
Adopted channels (performed “switching”) can be identified with a marking starting with”&” on the
screen as follows.
These markings are displayed on “Numerical”, “Graph” screen, and printed results for “Cumulative”
and “HPC”.
Markings added to the result indicate the channel. The explanation is as follows.

Meaning Mark
Adopted result on WBC-D &D
Adopted result on PLT-O &O
Adopted result on PLT-F &F

[Horizontal development of Global version Ver.22.10-00]


3 The logon function using an IC card reader has been added only for Windows 10.
4 An adjustment function for MiniCollect microtube has been added.
5 Detail codes are now output for “Cannot recognize Fluorocell --- information” errors to parameters for
error history screen, error history list printing, error history CSV output, and SNCS error report.
Following errors are scope of this change:
- Cannot recognize Fluorocell WNR information
- Cannot recognize Fluorocell WDF information
- Cannot recognize Fluorocell WPC information
- Cannot recognize Fluorocell RET information
- Cannot recognize Fluorocell PLT information
Detail codes to be output are as follows:

Param1 Param2
Top 2byte Bottom 2byte 4byte
Error occurrence Error type Error status
position
01: Sequencer 01: Discrepancy on decrypted 0001: REAGENT_ERROR_UID
side UID at an event other than
startup check
02: Different tag being read 0001: REAGENT_ERROR_UID
from the tag read previously
03: Unregistered third party 0001: REAGENT_ERROR_UID
reagent tag being read
04: Failure on password 0002: REAGENT_ERROR_PASS
authentication
05: ISO15963 error -
occurrence
02: DCM side 06: Discrepancy on remaining 0003: REAGENT_ERROR_COUNT
test count in UID existing in
record
07: Expired (Sysmex genuine 0006:
reagent) REAGENT_ERROR_EXPIR_GENUINE
07: Expired (Registered third 0007:
party reagents whose coded REAGENT_ERROR_EXPIR_REGISTER
UID and UID matches) ED
07: Expired (Registered third 0008:
party reagents whose coded REAGENT_ERROR_EXPIR_THIRDPART
UID and UID do not match) Y
08: SLIX2 authentication 0004: REAGENT_ERROR_AUTHE
failure (signature verification
error)
08: SLIX2 authentication 0005: REAGENT_ERROR_MANU
failure (manufacturer error)
09: Registered third party -
reagent being set again
6 The configuration including CV-55 can now be selected in “ConfigTool”.
7 A service setting for the smart indicator (SI-14) has been added.

Setting name Setting value Default value


External indicator light 0: SI-10/11/12/13 1: SI-14 (In case
1: SI-14 of upgrading, the
default is “0”)

8 The “Blast/AbnLympho?” flagging specificity has been improved.


9 Improvements of Q-flagging judgment value calculation for “Blast?”, “AbnLympho?”, and “Blast/
AbnLympho?”, which were made in Ver.21.11-00 (for European markets only), have been applied to
the other markets, AM, JP, AP, and CN.
10 McAfee Ver.7 can now be used on Windows 7.

[Horizontal development of Global version Ver.22.12-00]


11 A message displayed on “Aspiration Sensor Cal” has been changed.

Before change Execute Aspiration Sensor Calibration after Aspiration Sensor was
executed.
After change Execute Aspiration Sensor Calibration after Aspiration Sensor
Adjustment was executed.
[Horizontal development of Global version Ver.22.15-00]
12 The versions of VisualStudio and Microsoft SQLServer have been changed. The installed applications
depend on Windows OS.
The service life for previous versions of Visual Studio and Microsoft SQL Server will end before the
end of the sales of XN analyzers. To resolve the situation, the versions of Visual Studio and Microsoft
SQL Server for Windows 10 have been changed.

Before change After change


VisualStudio2008 -> VisualStudio2017
SQLServer2014 Express SP2 SQLServer2016 Express SP2

The new VisualStudio and SQLServer are installed automatically when the installer runs. .NET
Framework4.7.2 is installed as well.

Note:
(1) Once upgraded to Ver.22.11-00, downgrading to previous versions cannot
be performed.
(2) On Windows7, VisualStudio2017 is installed. However, SQL Server 2016 is
not installed.

[Horizontal development of Global version Ver.22.16-00]


13 Outputting QC comment in each plot in QC chart to a HOST computer is supported.
(ASTM format only)

QC Comment

14 Canceling the database optimization process has been supported.

[Cancel] button

15 Adding service setting “Sensitivity adjustment of the PLT-F channel in QC analysis”. Plotting position
of PLT cluster on PLT-F ch can be adjusted by this setting. (Setting range is 0-14 and default value is
0)
16 Message displayed when reagent supply source is changed between RU and DCL has been
changed. Previously when RU-20 has any problem and reagent source is changed to DCL, and then
changed back to RU-20 again, at that time the message displayed was confusing and customer can
connect DST directly. To address the issue, message has been changed.

Ver.22.10-00 and before Ver.22.11-00

17 Memory size IPU can use has been changed from 2GB to 4GB. (Only for Windows10 64bit OS)
Previously there was “Out of Memory” error can occur because IPU can access up to 2GB of memory.
The memory size that IPU can access is extended to address this issue.
18 QC data is no longer written directly to the folder accessed by Support Manager.
QC data was directly written to the folder monitored by Support Manager.
However, to prevent zero byte (no content) of data from being sent to the SNCS server, QC data is no
longer written directly to the folder monitored by Support Manager now.

[Horizontal development of Global version Ver.22.18-00]


19 XR can now control XR-9000 in which BT-50 is included in the configuration.
20 Users who do not have the right to change settings can now switch the reagent from RU to
CELLPACK DCL.
21 A service setting has been added to set the number of tests to be taken before performing the "Drain
RBC Isolation Chamber". (Setting name: Analysis interval for drain RBC Isolation Chamber,
Execution timing: During startup sequence at next startup)
22 The determination of the WBC Abn Scattergram has been enhanced when the fluorescence
intensity (SFL) of the WBC decreased.

[Horizontal development of Global version Ver.22.19-00]


23 IPU now works in an environment with Firewall Function enabled.
24 As SNCS integrates into Caresphere, IPU no longer uses the word "SNCS".
25 New service setting "Analysis interval for Exchange Piercer" to change the number of measurements
before piercer replacement message occurs has been added.
26 The default value for the service setting "Barcode reader type" has been changed.
Was Is
NFT-7175 (OPTOELECTRONICS) BL-185 (KEYENCE)

27 Add the new service setting to use barcode reader F-100 (OPTOELECTRONICS). (Each analyzer
system settings)
・Number of barcode reader decode matches
・Minimum reading digits for barcode reader
Resolved Program issues

[Horizontal development of Global version Ver.22.08-00]


1 With program version 22.07, RFID tag issue had been addressed, however when upgrading the
device driver was not updated and the countermeasure was not activated. This issue has been
resolved.
2 When viewing IFU in the IPU, the application program has been changed from Acrobat to “PDF
viewer” with version 22.05 and “PDF viewer” installation program was added. However, at version up,
that program was not registered in the “White List” for McAfee and error message “IPU.exe has
stopped working….” appeared and IPU program did not start up at next power on. This issue has
been resolved.
3 With the Rerun/Reflex/Comment Rule, when Retest:Eo was judged as “True” (means “Retest:Eo”
appears on the screen), Retest: PLT was also judged as “True” and Action that was set in rule screen
was applied. This issue has been resolved.

[Horizontal development of Global version Ver.22.09-00]


4 In the XN-1500 configuration, if the following steps were performed, SA-21 could not be used. This
issue has been resolved.
(1) XN was powered on by powering SP-50 ON with WOL.
(2) XN IPU was shutdown.
(3) Although SP-50 was being used, XN IPU was powered OFF and SA-21 could not be used.
5 Previously, due to the failure of IPU startup process, the IPU program could fail to start normally and a
Mounter.exe termination error could occur. This issue has been resolved.
6 When a new sample ID was read while multiple samples were selected on “Worklist”, a lot of
additional samples could be selected erroneously and IPU operation could be slow. This issue has
been resolved.
7 The “Aspiration unit left-right motor error” could occur on XN analyzers which has a countermeasure
kit against erroneous high HGB. This issue has been resolved.
8 Previously, Q-Flag value “0” was not displayed (left blank) for Q-Flag even though parameters are
judged as “Negative”. This issue has been resolved and Q-Flag value “0” is now displayed.
9 On WNRch, WBC could erroneously be differentiated as NRBC. This issue has been resolved.
10 BASO result measured on WDFch is now adopted if BASO result measured on WNRch is found
abnormal.
11 The sum of WBC was not 100% in some specific cases. This issue has been resolved.
12 When “Auto output” configuration was set and analysis result was automatically sent to the host
computer, the action messages for the result of delta check could not be sent to the host computer.
This issue has been resolved.

[Horizontal development of Global version Ver.22.10-00]


13 In XN-1500, there was a possibility that a sample tube was not returned to the sample rack when
performing sampler analysis after correcting an error categorized as “Sampler Analysis Stop 3”, which
is errors relating to the analysis line such as a rack move error. This issue has been resolved.
14 When SA-01 was used, there was a possibility that a background check would not finish and the
instrument status did not become “Ready” if there was a lot of files (more than 5000 data files). This
issue has been resolved.
15 “Low count error” and “PLT sampling error” can now be output correctly.
16 If the birth date was not registered in the “Register Patient Information” dialogue, the current date was
registered as the birth date. This issue has been resolved and the patient birth date is no set if the
birth date is left blank. Additionally, the usability has been improved.
17 When replacing reagents on XN-2000 or etc., instrument status could not become “Ready” if the
mode was switched between registering new reagents and executing replacement. This issue has
been resolved.

[Horizontal development of Global version Ver.22.12-00]


18 All logon names displayed on the “Audit Log” screen was "(Logoff)”. This issue has been resolved.
19 When the number of measured samples exceeded 100,000 after installation, data bacame
fragmented and database volume unintendedly increased. When the size of database exceeded 4
GB, new measurement results could not be added. To resolve this issue, the following measure has
been implemented.
(1) IPU will perform the database defragmentation (optimization) as soon as the database uses the
specified volume. When the optimization is being performed, the following progress bar appears:

(2) A setting to select frequency for database optimization from High, Middle, and Low has been
added to “System” in “IPU Setting”. The frequency depends on the setting value and the number
of samples per day. The rough indication is as follows:

Frequency of optimization
Setting Required
value 600 analyses 400 analyses 200 analyses time
per day per day per day
High Every 12 days Every 19 days Every 37 days 10 minutes
Middle Every 37 days Every 56 days Every 112 days 15 minutes
Low Every 87 days Every 131 days Every 262 days 25 minutes
In addition, a database optimization tool (DBDefrag.exe) to defragment database manually and a
database repair tool (XN_DatabaseRepairer.exe) have been added to “C:\Program
Files\Sysmex\Tools” folder.

Note:
To execute DBDefrag.exe and XN_DatabaseRepairer.exe, logon as Windows administrator
is necessary.

20 On the PLT-F ch, particles which were not PLT were mistakenly counted as PLT, resulting in falsely
high PLT. To resolve this issue, when the abnormal scattergram appears, re-classification will be
performed and the “PLT Abn Scattergram” IP message will be shown. In addition, the low reliability
marks will be placed next to the PLT related parameters.

PLT related parameters


PLT, PLT-F, PLT-F2, IPF, IPF#, H-IPF

Examples of abnormal PLT scattergram


Before change After change

21 Backup files containing comma “,” and square bracket “[” could not be restored. This issue has been
resolved. With this change, the version for QC backup files has been upgraded to Ver.1.2.

22 When opening database failed at startup, thresholds for all Q-Flags were 0 since thresholds for Q-
Flags could not be read. If measurement was performed with this condition, all Q-Flag results were
“Positive”. To resolve this issue, when opening database failed, retrial is executed. If retrial failed, the
following message box appears and measurements cannot be performed until restarting IPU.

Massage box

23 Resolved the issue about piercer adjustment

[Horizontal development of Global version Ver.22.15-00]


24 When WBC aggregation occurred on the WNR ch, “BASO#” and “BASO%” research parameters
were not masked. This has been changed so that those parameters are now masked.
25 When the data form is “0.XXXX” on the “Target/Limit Setting” screen, the data could be input only to
three decimal places and could not be input to the four decimal places. This issue has been resolved.
This issue occurred on the following items:

Measurement Mode Item and unit name


WB “10^6/uL” for RET#
WB “10^12/L” for RET#
WB “L/L” for PCT
WB “Ratio” for PCT
BB (PLT pack) “10^6/uL” for RBC
BB (PLT pack) “10^12/L” for RBC

26 After the QC file whose QC file number had been deleted was registered with different lot number, the
QC file could not be deleted. This issue has been resolved.
27 When starting the IPU program, mounting VHD could fail and “MountError” would occur. This issue
has been resolved.
28 The issue of DB restoration tool has been resolved.
29 When SA-01 was used on the XN-1000 configuration, the internal error could occur after performing
the followings:
(1) Sample measurement started on the sampler mode.
(2) After the analyzer picked up and mixed a sample tube, the mode switch was pressed and
switched to the manual measurement mode.
(3) After the analyzer returned the sample tube to the rack, the rack was removed by hand.
(3) The mode switch was pressed again to change to the sampler measurement mode.
(5) The rack removed on the previous step was entered again and the measurement started.
(6) The internal error occurred when the analyzer detected the first sample tube.
This issue has been resolved.
30 Error messages displayed on XN IPU for the RU-20 pressure errors were incorrect. This issue has
been resolved so that the same messages as shown in the IFU will now appear.
31 Previously, analysis results for WBC related parameters were not masked, depending on the “Mask of
analysis result with misclassification for WBC differentials” service setting, when NRBC interference
to WBC area was detected and “WBC Abn Scattergram” was output. This issue has been resolved
and WBC related parameters are now masked.

Applicable parameters
NEUT#, NEUT%, LYMPH#, LYMPH%, MONO#, MONO%, EO#, EO%, BASO#,
BASO%, NRBC#, NRBC%, IG#, IG%, WBC-N, WBC-D, WBC-P, HFLC#, HFLC%,
NEUT#&, NEUT%&, BA-N#, BA-N%, BA-D#, BA-D%

32 When the “Limit Setting” was set as “Ratio(%) for the “QC Setting” and QC lot was registered, “Limit
range(%)” for sensitivity items (WNR-X, WNR-Y, WNR-Z etc.) were 100%. However, normal users
cannot change the sensitivity items and the limit were not managed properly. (Only the “Service” user
can change the sensitivity items)
This issue has been resolved by using “Differential(#)” regardless if limit is set to “Ratio(%)” or
“Differential (#)” for sensitivity items.
33 After the “Positive ID check error” occurred, SP did not make smear when the rack with error was re-
entered. This issue has been resolved.
34 On XN-2000 configuration, if only one of 2 analyzers performs measurement repeatedly, the piercer
mechanism unit in the other analyzer did not return to the initial position and continually moved to
right. This issue has been resolved.

[Horizontal development of Global version Ver.22.16-00]


35 In APD Parameter Setting menu, service engineers could not input voltage value as they intended.
This issue has been resolved.
36 Some of functionalities could not be used such as reagent replacement, measurement mode change,
when logging on to the IPU using an IC card while the power of the main unit is turned on.
This issue has been resolved.
37 WBC count could become 1/3 when analyzing a whole blood sample in the LW mode if following
steps below are taken. This issue has been resolved.
(1) Select Low WBC mode, and click OK to close the dialog.
(2) Input sample No, check “Query to Host” and close the dialog by clicking [OK] button.
(3) Receive analysis order for LW mode from host computer.
(4) In “Change Measurement Mode” select Whole Blood..
(5) Press start switch to run a sample, without accessing “Manual Analysis” dialog.

38 With the configuration that there are two main units are used (i.e XN-2000) there was a possibility not
to show restart button after shutdown. This issue has been resolved.
In case that left analyzer is turned OFF and right analyzer is turned ON, restart button could not be
displayed when shutting down the analyzer with CELLCLEAN on rack. This issue has been resolved.
39 In BF mode, the sum of WBC was not 100% in some specific cases. This issue has been resolved.

[Horizontal development of Global version Ver.22.17-00]


40 Even if the analysis mode was set to “Normal” after a “Precision Check” analysis was performed in
any analysis modes other than “Normal” in the Service mode setting, the analysis mode stayed the
same as it was.
This issue has been resolved. The “Precision Check” analysis can now be performed correctly even
after the analysis mode is switched from other modes than “Normal”.
41 When “RBC Abn Distribution”, indicating the possibility of abnormality on RBCch, was output, the low
reliability mark [*] was appended to the RET%, IRF, LFR, MFR, HFR, and RET-He items calculated
only in RETch.
This issue has been resolved, and the low reliability mark is not appended to the RET%, IRF, LFR,
MFR, HFR, and RET-He items when “RBC Abn Distribution” is output.
42 When the system was XN-9000 configuration, it was possible that the XN analyzers and CV-50 would
adopt sample numbers read by BT-40 instead of those read by the XN analyzer after BT-40 read the
sample barcodes. This would cause adopting incorrect sample numbers if user changed sample
position in the sample rack.
This issue has been resolved, and sample numbers read by XN analyzers are adopted at the time of
measurement.
43 Units for some parameters were not output correctly when measurement items were output via GDI.
This issue has been resolved.

[Horizontal development of Global version Ver.22.18-00]


44 Resolved the issue that IPU stops when shutdown is complete when the host computer is
turned off during shutdown.
45 Resolved the issue that the screen lock is re-activated when the screen is not operated after the
screen lock is released by IC card.
46 Resolved the issue that expiration date of reagent is not displayed correctly if the analyzer is
shutdown without measurement after staining solution replacement.

[Horizontal development of Global version Ver.22.19-00]


47 Resolved the issue that log off history would not be kept if Windows exited before writing log off
history to the database in the shutdown process.
48 Resolved the issue that some of the conditions in rule 1 of PLT-I/PLT-O switching were incorrect and
could not be switched when MicroR was 5.0%.
49 Resolved the issue that the management range can not be set correctly when QC file is restored and
QC limit setting is set as "%".
50 Resolved the issue that cleaning operation can not be performed properly when 2 units are started
simultaneously to be cleaned by CELLCLEAN AUTO on XN-9100/XR-9000 with BT-50 in the
configuration.
51 Resolved the issue that grade value of Q-Flag does not become blank when measurement result is
blank data.
52 Resolved the issue that inability to restore older versions of smp files in previous version.

Remarks
Module Version
[IPU] 00-22 Build194
[Device Control Manager] 00-22 Build194
[Analysis] 00-22 Build80
[Analyzer Sequencer] 00-22 Build194
[Sampler Sequencer] 00-22 Build194

Sequence Program Version


[Analyzer] 00-68
[Sampler] 00-16

ECR: 322B052
DVD1XN1J Ver.22.10-00 (China markets only)
<Applicable to S/N> <Release Date>
Refer to ECR 322C011 June 2022

<Part No./Name> Module Version


BV341672 [IPU] 00-22 Build105
CDR_ASSY NO.377 (DVD1XN1J(22.10-00)) [Device Control Manager] 00-22 Build105
[Analysis] 00-22 Build17
[Analyzer Sequencer] 00-22 Build105
<FDS File Name> [Sampler Sequencer] 00-22 Build105
P_XNSWSTCN_22100000.cab Sequence Program Version
[Analyzer] 00-60
[Sampler] 00-16

Changed Specifications
1 Changed so that XN-10x can now select a configuration without RET license.
In following “3. Option selection” of “Initial setup wizard” screen, a selection “XN-10[-]” without RET
license has been added to [Channel] combo box shown in red frame below. And the XN-10[-] can be
selected from the combo box.
2 Changed so the system including XN-10x, XN-10, and XN-20 can be configured.
(1) System configuration patterns
1)The patterns (measurement module selection) included XN-10x can be selected from “Instrument
configuration” as shown below.

2) Module patterns
2-1) One module (Ver.22.09 and thereafter): XN-1000/XN-1500/XN-1590

XN-10x Only

2-2) Two modules: XN-2000/XN-2800/XN-3000/XN-3100, XN-9100 CV-55, XN-9100 CV-55 x 2

Left Right
XN-20 XN-10x
XN-10 XN-10x
XN-10x XN-10x

2-3) Three modules (Ver.22.09 and thereafter): XN-9100 CV-50 CV-55, XN-9100 CV-55 x 3

Left Center Right


XN-10x XN-10x XN-10x

(2) Channel configuration patterns


Changed so the channel configuration (2 modules only) as shown in the table below can be
selected.
1) Two modules: XN-2000/XN-2800/XN-3000/XN-3100, XN-9100 CV-55, XN-9100 CV-50×2

Basic configuration Measurement channel option


None
XN-20 [PLT-F], XN-10x [RET]
XN-20, XN-10x
XN-20 [PLT-F], XN-10x [ - ]
XN-20 [ - ], XN-10x [RET]
None
XN-10 [RET, PLT-F], XN-10x [RET]
XN-10 [RET], XN-10x [RET]
XN-10, XN-10x
XN-10 [RET, PLT-F], XN-10x [ - ]
XN-10 [PLT-F], XN-10x [ - ]
XN-10 [RET], XN-10x [ - ]
None
XN-10x, XN-10x XN-10x [RET], XN-10x [RET]
XN-10x [RET], XN-10x [ - ]
2)The channel can be selected from the “Channel” combo box as shown below.

Resolved Program issues


- -

Remarks
None
ECR: 322C011
DVD1XN1J Ver.22.09-00 (China markets only) Upgrade Procedure
<Applicable to S/N> <Release Date>
Refer to ECR 322B018 March 2022

<Part No./Name> Module Version


[IPU] 00-22 Build92
[Device Control Manager] 00-22 Build92
[Analysis] 00-22 Build17
[Analyzer Sequencer] 00-22 Build92
<FDS File Name> [Sampler Sequencer] 00-22 Build92
P_XNSWSTCN_22090000.cab Sequence Program Version
[Analyzer] 00-60
[Sampler] 00-16

Resolved Program issues


1 Enhanced XN-10x to be able to configure transport system except XN-1000.
Added a function to be able to set the following models in the System Config of XN-10x:
XN-2000, XN-3000, XN-9000, XN-1000, XN-1500, XN-3100, XN-9100, XN-1590, XN-2800
(So far, only XN-10x has this setting function.)
2 Changed a function that the PLT-F license cannot be granted, when logging in as a production user
(Kakof).

Resolved Program issues


1 Fixed the problem that IPU may not be able to start, when forcibly turning off the power on the IPU
Windows 10 PC.

Remarks
None

ECR 322B018
DVD1XN1J Ver.22.08-00 (China markets only)
<Applicable to S/N> <Release Date>
March 2022

<Part No./Name> Module Version


[IPU] 00-22 Build73
[Device Control Manager] 00-22 Build72
[Analysis] 00-22 Build17
[Analyzer Sequencer] 00-22 Build72
[Sampler Sequencer] 00-22 Build72
Sequence Program Version
[Analyzer] 00-60
[Sampler] 00-16

Changed Specifications
- -

Resolved Program issues


- -

Remarks
Initial Version