<Parts Information>
CV891460: CDR_ASSY NO.527 (DVD1XN1J(23.00-00))
Changed Specifications
1 The NEUT/EO classification algorithm of XN-10x has been changed to be the same as that of XN-10.
2 Now IG has been added as a reportable parameter on XN-10x.
3 On XN-10x, now PLT-F license can be added, and IPF has been changed to a reportable parameter.
4 Decision of RET Abn Scattergram has beefed up.
5 Decision of PLT Abn Scattergram has beefed up.
6 When a new stain cartridge is loaded, the opening date is written on the RFID, but this process
overlaps with writing the number of remaining tests, which can cause the analyzer to aberrate.
Therefore, the software has been modified to prevent double writing to RFID.
7 In Ver.22.19-00, service settings to use barcode reader F-100 were added. In addition to these
settings, adoption logic for sample ID read by barcode reader has been added to further increase the
reading rate of low quality labels. New barcode reader setting below has also been added in service
menu.
1 Resolved the issue that the suspended sampler measurement could not be resumed when the mode
switch was pressed to switch to the sampler measurement mode after the manual measurement was
completed only when SA-01 was used as the sampler.
Remarks
Module Version
[IPU] 00-22 Build205
[Device Control Manager] 00-22 Build205
[Analysis] 00-22 Build89
[Analyzer Sequencer] 00-22 Build205
[Sampler Sequencer] 00-22 Build205
ECR 322G056
DVD1XN1J Ver.22.11-00 (China markets only)
<Applicable to S/N> <Release Date>
Refer to ECR 322B052 September 2022
<Parts Information>
BV341672: CDR_ASSY NO.377 (DVD1XN1J(22.11-00))
Changed Specifications
[Horizontal development of Global version Ver.22.09-00]
1 The XN-1590 configuration has been added.
2 The analysis channel, used as adopted PLT and WBC result (“switching”), can now be displayed on
“Cumulative” and HPC screen.
Adopted channels (performed “switching”) can be identified with a marking starting with”&” on the
screen as follows.
These markings are displayed on “Numerical”, “Graph” screen, and printed results for “Cumulative”
and “HPC”.
Markings added to the result indicate the channel. The explanation is as follows.
Meaning Mark
Adopted result on WBC-D &D
Adopted result on PLT-O &O
Adopted result on PLT-F &F
Param1 Param2
Top 2byte Bottom 2byte 4byte
Error occurrence Error type Error status
position
01: Sequencer 01: Discrepancy on decrypted 0001: REAGENT_ERROR_UID
side UID at an event other than
startup check
02: Different tag being read 0001: REAGENT_ERROR_UID
from the tag read previously
03: Unregistered third party 0001: REAGENT_ERROR_UID
reagent tag being read
04: Failure on password 0002: REAGENT_ERROR_PASS
authentication
05: ISO15963 error -
occurrence
02: DCM side 06: Discrepancy on remaining 0003: REAGENT_ERROR_COUNT
test count in UID existing in
record
07: Expired (Sysmex genuine 0006:
reagent) REAGENT_ERROR_EXPIR_GENUINE
07: Expired (Registered third 0007:
party reagents whose coded REAGENT_ERROR_EXPIR_REGISTER
UID and UID matches) ED
07: Expired (Registered third 0008:
party reagents whose coded REAGENT_ERROR_EXPIR_THIRDPART
UID and UID do not match) Y
08: SLIX2 authentication 0004: REAGENT_ERROR_AUTHE
failure (signature verification
error)
08: SLIX2 authentication 0005: REAGENT_ERROR_MANU
failure (manufacturer error)
09: Registered third party -
reagent being set again
6 The configuration including CV-55 can now be selected in “ConfigTool”.
7 A service setting for the smart indicator (SI-14) has been added.
Before change Execute Aspiration Sensor Calibration after Aspiration Sensor was
executed.
After change Execute Aspiration Sensor Calibration after Aspiration Sensor
Adjustment was executed.
[Horizontal development of Global version Ver.22.15-00]
12 The versions of VisualStudio and Microsoft SQLServer have been changed. The installed applications
depend on Windows OS.
The service life for previous versions of Visual Studio and Microsoft SQL Server will end before the
end of the sales of XN analyzers. To resolve the situation, the versions of Visual Studio and Microsoft
SQL Server for Windows 10 have been changed.
The new VisualStudio and SQLServer are installed automatically when the installer runs. .NET
Framework4.7.2 is installed as well.
Note:
(1) Once upgraded to Ver.22.11-00, downgrading to previous versions cannot
be performed.
(2) On Windows7, VisualStudio2017 is installed. However, SQL Server 2016 is
not installed.
QC Comment
[Cancel] button
15 Adding service setting “Sensitivity adjustment of the PLT-F channel in QC analysis”. Plotting position
of PLT cluster on PLT-F ch can be adjusted by this setting. (Setting range is 0-14 and default value is
0)
16 Message displayed when reagent supply source is changed between RU and DCL has been
changed. Previously when RU-20 has any problem and reagent source is changed to DCL, and then
changed back to RU-20 again, at that time the message displayed was confusing and customer can
connect DST directly. To address the issue, message has been changed.
17 Memory size IPU can use has been changed from 2GB to 4GB. (Only for Windows10 64bit OS)
Previously there was “Out of Memory” error can occur because IPU can access up to 2GB of memory.
The memory size that IPU can access is extended to address this issue.
18 QC data is no longer written directly to the folder accessed by Support Manager.
QC data was directly written to the folder monitored by Support Manager.
However, to prevent zero byte (no content) of data from being sent to the SNCS server, QC data is no
longer written directly to the folder monitored by Support Manager now.
27 Add the new service setting to use barcode reader F-100 (OPTOELECTRONICS). (Each analyzer
system settings)
・Number of barcode reader decode matches
・Minimum reading digits for barcode reader
Resolved Program issues
(2) A setting to select frequency for database optimization from High, Middle, and Low has been
added to “System” in “IPU Setting”. The frequency depends on the setting value and the number
of samples per day. The rough indication is as follows:
Frequency of optimization
Setting Required
value 600 analyses 400 analyses 200 analyses time
per day per day per day
High Every 12 days Every 19 days Every 37 days 10 minutes
Middle Every 37 days Every 56 days Every 112 days 15 minutes
Low Every 87 days Every 131 days Every 262 days 25 minutes
In addition, a database optimization tool (DBDefrag.exe) to defragment database manually and a
database repair tool (XN_DatabaseRepairer.exe) have been added to “C:\Program
Files\Sysmex\Tools” folder.
Note:
To execute DBDefrag.exe and XN_DatabaseRepairer.exe, logon as Windows administrator
is necessary.
20 On the PLT-F ch, particles which were not PLT were mistakenly counted as PLT, resulting in falsely
high PLT. To resolve this issue, when the abnormal scattergram appears, re-classification will be
performed and the “PLT Abn Scattergram” IP message will be shown. In addition, the low reliability
marks will be placed next to the PLT related parameters.
21 Backup files containing comma “,” and square bracket “[” could not be restored. This issue has been
resolved. With this change, the version for QC backup files has been upgraded to Ver.1.2.
22 When opening database failed at startup, thresholds for all Q-Flags were 0 since thresholds for Q-
Flags could not be read. If measurement was performed with this condition, all Q-Flag results were
“Positive”. To resolve this issue, when opening database failed, retrial is executed. If retrial failed, the
following message box appears and measurements cannot be performed until restarting IPU.
Massage box
26 After the QC file whose QC file number had been deleted was registered with different lot number, the
QC file could not be deleted. This issue has been resolved.
27 When starting the IPU program, mounting VHD could fail and “MountError” would occur. This issue
has been resolved.
28 The issue of DB restoration tool has been resolved.
29 When SA-01 was used on the XN-1000 configuration, the internal error could occur after performing
the followings:
(1) Sample measurement started on the sampler mode.
(2) After the analyzer picked up and mixed a sample tube, the mode switch was pressed and
switched to the manual measurement mode.
(3) After the analyzer returned the sample tube to the rack, the rack was removed by hand.
(3) The mode switch was pressed again to change to the sampler measurement mode.
(5) The rack removed on the previous step was entered again and the measurement started.
(6) The internal error occurred when the analyzer detected the first sample tube.
This issue has been resolved.
30 Error messages displayed on XN IPU for the RU-20 pressure errors were incorrect. This issue has
been resolved so that the same messages as shown in the IFU will now appear.
31 Previously, analysis results for WBC related parameters were not masked, depending on the “Mask of
analysis result with misclassification for WBC differentials” service setting, when NRBC interference
to WBC area was detected and “WBC Abn Scattergram” was output. This issue has been resolved
and WBC related parameters are now masked.
Applicable parameters
NEUT#, NEUT%, LYMPH#, LYMPH%, MONO#, MONO%, EO#, EO%, BASO#,
BASO%, NRBC#, NRBC%, IG#, IG%, WBC-N, WBC-D, WBC-P, HFLC#, HFLC%,
NEUT#&, NEUT%&, BA-N#, BA-N%, BA-D#, BA-D%
32 When the “Limit Setting” was set as “Ratio(%) for the “QC Setting” and QC lot was registered, “Limit
range(%)” for sensitivity items (WNR-X, WNR-Y, WNR-Z etc.) were 100%. However, normal users
cannot change the sensitivity items and the limit were not managed properly. (Only the “Service” user
can change the sensitivity items)
This issue has been resolved by using “Differential(#)” regardless if limit is set to “Ratio(%)” or
“Differential (#)” for sensitivity items.
33 After the “Positive ID check error” occurred, SP did not make smear when the rack with error was re-
entered. This issue has been resolved.
34 On XN-2000 configuration, if only one of 2 analyzers performs measurement repeatedly, the piercer
mechanism unit in the other analyzer did not return to the initial position and continually moved to
right. This issue has been resolved.
38 With the configuration that there are two main units are used (i.e XN-2000) there was a possibility not
to show restart button after shutdown. This issue has been resolved.
In case that left analyzer is turned OFF and right analyzer is turned ON, restart button could not be
displayed when shutting down the analyzer with CELLCLEAN on rack. This issue has been resolved.
39 In BF mode, the sum of WBC was not 100% in some specific cases. This issue has been resolved.
Remarks
Module Version
[IPU] 00-22 Build194
[Device Control Manager] 00-22 Build194
[Analysis] 00-22 Build80
[Analyzer Sequencer] 00-22 Build194
[Sampler Sequencer] 00-22 Build194
ECR: 322B052
DVD1XN1J Ver.22.10-00 (China markets only)
<Applicable to S/N> <Release Date>
Refer to ECR 322C011 June 2022
Changed Specifications
1 Changed so that XN-10x can now select a configuration without RET license.
In following “3. Option selection” of “Initial setup wizard” screen, a selection “XN-10[-]” without RET
license has been added to [Channel] combo box shown in red frame below. And the XN-10[-] can be
selected from the combo box.
2 Changed so the system including XN-10x, XN-10, and XN-20 can be configured.
(1) System configuration patterns
1)The patterns (measurement module selection) included XN-10x can be selected from “Instrument
configuration” as shown below.
2) Module patterns
2-1) One module (Ver.22.09 and thereafter): XN-1000/XN-1500/XN-1590
XN-10x Only
Left Right
XN-20 XN-10x
XN-10 XN-10x
XN-10x XN-10x
2-3) Three modules (Ver.22.09 and thereafter): XN-9100 CV-50 CV-55, XN-9100 CV-55 x 3
Remarks
None
ECR: 322C011
DVD1XN1J Ver.22.09-00 (China markets only) Upgrade Procedure
<Applicable to S/N> <Release Date>
Refer to ECR 322B018 March 2022
Remarks
None
ECR 322B018
DVD1XN1J Ver.22.08-00 (China markets only)
<Applicable to S/N> <Release Date>
March 2022
Changed Specifications
- -
Remarks
Initial Version