DVD1XNVX Version History E
DVD1XNVX Version History E
<Part No./Name>
BF860669: CDR_ASSY NO.315(DVD1XNVX)
Changed Specifications
1 A new animal species "(022) BoneMarrow" has been added.
2 The logon name at the time of measurement can be confirmed from the measurement result.
3 In the measurement of camel blood, the setting range of PLT CAL in the animal species setting has
been changed to 0.0 ~ 999.9.
When measuring camel blood platelets with RBC/PLTch, normal calibration alone was not sufficient to
measure them correctly, and it was necessary to multiply them by a special factor in the animal
species setting, and an upper limit of 150 was insufficient.
The setting range has been changed to address this issue.
“Service Species CAL” will be displayed in “Animal Species Setting” -> “Service”, if “Service Setting” -
> “IPU settings” -> “Miscellaneous” -> “Enable Species PLTI_CAL” is set to “1: Enable” and IPU is
restarted.
Resolved issues
1 When samples of dogs and cats were measured, the fluorescence intensity of WBC tended to be
higher than the measurement of human blood, and NRBC and WBC were not fractionated correctly.
To resolve this issue, analysis algorithms for WNRch has been changed.
2 When the instrument was shut down without measuring sample after exchanging stain solution, the
expiration date of reagent was not displayed correctly.This problem has been resolved.
Remarks
Module Version
IPU 00-12
Device Control Manager 00-12
Analysis 00-12
Analyzer Sequencer 00-12
Sampler Sequencer 00-12
Analyzer 03-61
Sampler 00-16
ECR: 321G059
<Release Date>
<Part No./Name> February 2021
BF860669: CDR_ASSY NO.315(DVD1XNVX)
Changed Specifications
1 The versions of VisualStudio and Microsoft SQLServer have been changed. The installed applications
depend on Windows OS.
The service life for previous versions of Visual Studio and Microsoft SQL Server will end before the
end of the sales of XN analyzers. To resolve the situation, the versions of Visual Studio and Microsoft
SQL Server for Windows 10 have been changed.
The new VisualStudio and SQLServer are installed automatically when the installer runs. .NET
Framework4.7.2 is installed as well.
Note:
(1) Once upgraded to Ver.3.06-00, downgrading to previous versions cannot
be performed.
(2) On Windows7, VisualStudio2017 is installed. However, SQL Server 2016 is
not installed.
Before change Execute Aspiration Sensor Calibration after Aspiration Sensor was
executed.
After change Execute Aspiration Sensor Calibration after Aspiration Sensor
Adjustment was executed.
Resolved issues
1 The problem that "0" is output to the measurement data when the measurement item without the
order is printed in GP has been resolved. (When a measurement item with no order is printed in GP,
"Blank" is output to the measurement data.)
(2) When opening database failed at startup, thresholds for all Q-Flags were 0 since thresholds for Q-
Flags could not be read. If measurement was performed with this condition, all Q-Flag results were
“Positive”. To resolve this issue, when opening database failed, retrial is executed. If retrial failed, the
following message box appears and measurements cannot be performed until restarting IPU.
Massage box
(2) After the QC file whose QC file number had been deleted was registered with different lot number, the
QC file could not be deleted. This issue has been resolved.
(3) When starting the IPU program, mounting VHD could fail and “MountError” would occur. This issue
has been resolved.
(4) On Windows10, there was a case that an error dialogue appeared and IPU locked up when the SNCS
router was connected. This issue has been resolved.
(7) After the “Positive ID check error” occurred, SP did not make smear when the rack with error was re-
entered. This issue has been resolved.
(2) With the configuration that there are two main units are used (i.e XN-2000) there was a possibility not
to show restart button after shutdown. This issue has been resolved.
In case that left analyzer is turned OFF and right analyzer is turned ON, restart button could not be
displayed when shutting down the analyzer with CELLCLEAN on rack. This issue has been resolved.
(3) In APD Parameter Setting menu, service engineers could not input voltage value as they intended.
This issue has been resolved.
Remarks
Module Version
IPU 00-11
Device Control Manager 00-11
Analysis 00-11
Analyzer Sequencer 00-11
Sampler Sequencer 00-11
Analyzer 03-61
Sampler 00-16
ECR: 320C010
<Release Date>
<Part No./Name> March 2020
BF860669: CDR_ASSY NO.315(DVD1XNVX)
Changed Specifications
1 A new animal species "(021) Bird" has been added.
Note:
(1) Measurements for “(021) Bird” can be performed only on analyzers with the “PLT-
F” license.
(2) Measurements for “(021) Bird” can only be performed in WB mode.
The measurement channels and parameters are shown in the table below:
Param1 Param2
Top 2byte Bottom 2byte 4byte
Error occurrence Error type Error status
position
01: Sequencer 01: Discrepancy on decrypted 0001: REAGENT_ERROR_UID
side UID at an event other than
startup check
02: Different tag being read 0001: REAGENT_ERROR_UID
from the tag read previously
03: Unregistered third party 0001: REAGENT_ERROR_UID
reagent tag being read
04: Failure on password 0002: REAGENT_ERROR_PASS
authentication
05: ISO15963 error -
occurrence
02: DCM side 06: Discrepancy on remaining 0003: REAGENT_ERROR_COUNT
test count in UID existing in
record
07: Expired (Sysmex genuine 0006:
reagent) REAGENT_ERROR_EXPIR_GENUINE
07: Expired (Registered third 0007:
party reagents whose coded REAGENT_ERROR_EXPIR_REGISTER
UID and UID matches) ED
07: Expired (Registered third 0008:
party reagents whose coded REAGENT_ERROR_EXPIR_THIRDPART
UID and UID do not match) Y
08: SLIX2 authentication 0004: REAGENT_ERROR_AUTHE
failure (signature verification
error)
08: SLIX2 authentication 0005: REAGENT_ERROR_MANU
failure (manufacturer error)
09: Registered third party -
reagent being set again
(4) “Blast/AbnLympho?” flagging specificity has been improved. * For (000)Other only
(5) Q-flagging judgment value calculations have been improved for “Blast?”, “AbnLympho?”, and
“Blast/AbnLympho?” * For (000)Other only
(6) In addition to Windows 10, McAfee Ver.7 (SECURITY MASTER (3) COMPLETE) can now be used
on Windows 7.
Resolved issues
1 If manual gate on the PLTch was not used, PLTI_CAL worked correctly. However, in case where
analysis with manual gate was performed on PLTch, PLTI_CAL did not work correctly. This issue has
been resolved.
Correct Incorrect
Remarks
Module Version
IPU 00-10
Device Control Manager 00-10
Analysis Module 00-10
Analysis Sequencer Module 00-10
Sampler Sequence 00-10
Analyzer 03-61
Sampler 00-16
ECR:320A022
<Release Date>
<Part No./Name> January 2019
BF860669: CDR_ASSY NO.315(DVD1XNVX)
Changed Specifications
1 Added service setting which calibrates PLT-I for each animal species.
Enable Species PLTI_CAL [0: Disable, 1: Enable]
Factory default setting: [0]
“Service Species CAL” will be displayed in “Species Setting” if “Enable Species PLTI_CAL” is set to
“1: Enable” and IPU is restarted.
2 Added service setting to select Mono calibration mode in two steps for the facility which requires
analysis result of Mono to match with visual inspection.
Select Analysis Mode [0: Mode1, 1: Mode2]
Factory default setting: [0]
Resolved issues
1 Corrected issue where buttons for switching displayed sample (Up and Down buttons) will not be
shown when the screen key board is turned ON and manual analysis screen is displayed.
3 Corrected issue where SFL(X) histogram shifts to the right when SFL(X) Gain magnification is
measured by [x2], [x4].
4 Corrected issue where IPU abnormally terminates during measurement in following cases.
(1) When measurement is performed under high DB load condition such as XN-9100 configuration.
(2) When multiple measurement results are selected within the explorer in Ver.3.03-00.
5 Corrected issue where analysis module abnormally terminates when analysis module without RET
license is measured using Low Aspiration (LA) mode.
Was Is
(2) Measure for Cat Eo fake high value, Neut fake low value
Was Is
(3) Measure for Cat Mono fake high value, Lymph fake low value
Was Is
(4) Measure for Mouse Neut fake high value, Lymph fake low value
Was Is
Was Is
7 Corrected issue where species setting backed up in Ver.3.01-00 cannot be restored to Ver.3.03-00
properly.
Analyzer 03-61
Sampler 00-16
ECR:318F055
<Release Date>
<Part No./Name> May 2018
BF860669: CDR_ASSY NO.315(DVD1XNVX(00-08))
Changed Specifications
1 Ferret, *Sheep, *Goat, and Hamster have been added in Analyzable animal species.
With this change, Chapter 1 (section 1.10.2), which was the list for animal species that are not
installed, has been excluded.
*For Sheep and Goat, PLT-I, MPV, PCT, P-LCR and PDW are not analyzed.
In addition, the following Error, IP, or Action Messages are not determined.
[Error Message]
RET channel error
[IP Message]
PLT Abnormal Distribution
RET Abnormal Scattergarm
[Action Messages]
PLT channels difference exists
2 MONO_CAL setting for each animal species (service setting) has been added.
3 XN-V series Instructions for Use (IFU) manuals have been updated.
Resolved issues
None
Remarks
Module Version
IPU 00-08
Device Control Manager 00-08
Analysis Module 00-08
Analysis Sequencer Module 00-08
Sampler Sequence 00-08
Analyzer 02-60
Sampler 00-16
ECR:317L029
<Release Date>
<Part No./Name> December 2017
BF860669: CDR_ASSY NO.315(DVD1XNVX(00-06))
Changed Specifications
1 The number of mixing at the time of sampler analysis using the RBT sample tube for IDEXX has been
changed to 10 times.
In order to make the throughput of the sampler analysis using the IDEXX RBT sample tube the same
as that in the case of the normal rack, service setting for the number of mixing at the time of sampler
analysis of IDEXX RBT sample tube has been added. (Set the initial value as 10 times.) * For details,
refer to Service Manual Chapter 5 Service Program.
2 Analysis settings for IDEXX have been optimized for the data obtained by IDEXX.
As a result of the evaluation using the CELLPACK for IDEXX, it was found that it was necessary to
change the analysis settings below, so the analysis settings for IDEXX were optimized for the data
obtained by IDEXX.
[Target algorithm]
- WDF analysis algorithm: (007) Horse
- RET analysis algorithm: (006) Cat
Setting Default
Setting name Description
range setting
Enable RET Set whether RET analysis of Cat for 0 to1 0 (Disable)
analysis of Cat IDEXX enable. Restart the IPU program
for IDEXX after changing this setting. [0: Disable,
1: Enable]
6 "Base animal species" has been added to Explorer displayable item (initial setting: Hide).
* It can be set on the layout change in the explorer menu.
9 WBC, PLT priority settings can now be changed for all animal species.
Initial value should be PLT-F > PLT-O > PLT-I.
10 In manual analysis and manual analysis (extension), the particle color in the RBC area of PLT-Fch
has been changed to "blue".
12 "Reanalysis date" and "Reanalysis time" have been added to the sort condition on the explorer
screen.
- “Password Mode” has been added to Service Settings. The default setting is “Service defined”.
14 As countermeasures against use of reagents not authorized by Sysmex, the following changes have
been made.
- It is now possible to use ICODE SLIX2 RFID tags, which feature increased security.
- A message box will now be displayed to inform users if an unauthorized reagent has been set in the
instrument.
- The name of manufacturer and the place of production read from the RFID tag will now be saved in
the IPU database and displayed in the archives.
18 When lymphocyte is incorrectly discriminated as monocyte, lymphocyte shows falsely low value. To
resolve this issue, discrimination algorithm has been improved.
(Applied only to analysis algorithm of animal species (000) Other)
19 WDFch LYMP/MONO replacement prevention algorithm has been added to the XN-V series.
(Applied only to analysis algorithm of animal species (000) Other)
21 When neutrophilic cells are incorrectly discriminated as eosinocyte, eosinocyte shows falsely high
value. To resolve this issue, discrimination algorithm has been improved.
(Applied only to analysis algorithm of animal species (000) Other)
22 As there was unnecessary movement of solenoid valve, the sequence has been revised so that WPC
homolyzer charging is executed only during WPC measurement sequence.
24 Previously, restarting sampler analysis was not possible after recovering from “Rack removed” error
on SP-50 in XN-1500 configuration. The XN-1500 can now restart sampler analysis after recovering
from SP-50 “Rack removed” error.
25 The logging function has been enhanced so that the log can be recorded in detail when “Internal
Error” occurs*. (In order to get detailed information when a rare error occurs.)
* Internal Error occurs in following sequence error
- Unit activated twice
- Unit controlled illegally
- Motor activated twice
26 Previously, if XN-3000/3100 received smear order when “Command not to aspirate” was checked in
“Analysis Stop Setting” in standalone mode, a “Invalid analysis item is specified” error occurred. In
order to resolve this situation, sampler analysis will be stopped when “Command not to aspirate” was
checked.
1 Occurrence of WBC Abnormal Scattergram due to falsely high values of the Neut/Eo particle
discrimination has been resolved for the animal species performing Neut/Eo differentiation in the
WDF (SSC-FSC) channel.
2 CSV setting may fail to be backed up properly in the IPU setting backup file. This has been resolved.
In the CSV item setting, reagents not used in the system configuration are displayed as "blank
(nothing displayed)". (Example: "CELLPACK DFL", "Fluorocell RET" in the configuration without RET
license)
When backing up the setting with one of these items being "Output" and the other being "Do not
output", the problem that the reagents not used in the system configuration are backed up as "Output"
was found.
4 Lockup when suspending measurement while waiting for analysis order has been resolved.
6 Lockup when detecting a Micro tube while reagent is depleted during manual measurement has been
resolved.
8 “WBC Abn Sct.” is an error message which is displayed when WBC discrimination is not performed
correctly due to low fluorescence signal. However, there were cases where this message was not
displayed even though WBC discrimination was not performed correctly. To resolve this issue, WBC
discrimination algorithm has been improved.
(Applied only to analysis algorithm of animal species (000) Other)
10 Previously, there was a possibility that “Mount Error” (Windows error) occurred while starting up IPU
program. This has been resolved by extending monitoring time for starting up IPU program from 60
sec to 300 sec.
* This algorithm is detected by comparing LC-I and TNC-N. (LC-I is obtained from counted particles
in higher channels of RBC ch.)
Reference: Service Manual Chapter 5 Service Program
Remarks
Module Version
IPU 00-07
Device Control Manager 00-07
Analysis Module 00-07
Analysis Sequencer Module 00-07
Sampler Sequence 00-07
Analyzer 02-60
Sampler 00-16
ECR:317L020
<Release Date>
<Part No./Name> July 2017
BF860669: CDR_ASSY NO.315(DVD1XNVX(00-06))
Changed Specifications
None
Resolved issues
1 Countermeasure has been made for incorrect scattergram coloring when doing GP Print.
This phenomenon is only seen in GP printing. Measured values and scattergrams displayed on IPU
monitor are not affected.
(IPU monitor)
(GP Printing)
Remarks
Module Version
IPU 00-06
Device Control Manager 00-05
Analysis Module 00-05
Analysis Sequencer Module 00-05
Sampler Sequence 00-05
Analyzer 02-59
Sampler 00-16
ECR:317G027
<Release Date>
<Part No./Name> June 2016
AB833655: CDR_ASSY NO.271(DVD1XNVX(00-04))
Changed Specifications
<XN-V related>
1 XN-V now supports system configuration of XN-1500, XN-3100 and XN-9100. (Only for SP-50)
2 XN-V now supports species camel.
3 XN-V now supports SNCS. (Output files, data and timing for SNCS are same as ones for XN Ver.
22.01-00.)
4 License option for RET, PLT-F and BF have been added. (License option screen has been added with
this program version. It appears when the program version 3.00-00 is installed or upgraded from
previous version. Therefore, when these parameters are used, the license information must be
entered. Refer to “XN-V_ver.3.00-00_upgrade_proce.pdf” for detail.)
5 Following exclusive feature for IDEXX have been added.
(1)Scattergram color can now be changed by service setting "Enable IDEXX
function1"
(2)Analysis algorithm for CELLPACK (EPK) for IDEXX can now be selected by
service setting "Enable IDEXX function2"
(3)Using IDEXX test tube analysis function can now be changed by service set-
ting "Enable IDEXX function3"
6 Analysis algorithm for pre-diluted mode has been changed to improve relativity with whole blood
mode.
7 Classification performance of analysis algorithm for whole blood mode has been improved.
8 Channel analysis including manual analysis has been improved as follows.
Result (except unable to be calculated) data will not be masked. ("----") When channel analysis
including manual analysis is performed, MONO_B_CAL calibrate will not performed.
9 Manual analysis gate for manual analysis screen and manual analysis screen (extended) except BF
mode can now be generated on RET, RET (EXT) and PLT-O scattergram.
10 LP print setting is now available on IPU setting by adding List Printer (LP) print format.
11 PLT-O setting for QC parameter has been added to service setting by displaying PLT-O or not
displaying.
12 Q-Flag settings for PLT Clumps? (WNR), PLT Clumps?(WDF), PLT Clumps?(PLT-F) by the species
has been added to service setting.
20 To support the RU-20 Upgrade Kit, the following changes have been made.
Four analyzers can be connected with the RU-20.
To connect four analyzers with the RU-20, the preparation volume of 9L/h is not sufficient.
Therefore, it is now possible to select between 9L/h and 12L/h in SystemConfig.
-"RU has stopped supplying reagent." has been changed to "RU-20 has stopped supplying reagent."
and "RU-21 has stopped supplying reagent." to distinguish RU-20 or RU-21.
(3) When XN-3100 or XN-1500 is selected in the initial setup wizard screen, the
following screen will be displayed and instrument ID will be stored.
(6) When the IPU starts up on the XN-1500, the SP-50 will start up together with
the analyzer. As for XN-1500, “SP-50 connection settings” button will be dis-
played in the System Configurator screen. (Refer to the picture in (5) listed
above.
(7) In the case of connecting XN-1500 or XN-3100, number of slide, smear con-
dition, replications, additional rinse in resist order, modify order screen have
been added.
Smear module in the work list has been changed to SP-50.
Number of slide, smear condition, replications, additional rinse and alarm can now be communicate
with the host computer in ASTM.
When CELLCLEAN AUTO is recognized, it will be delivered to XN and SP-50.
22 The speed of copying files during program upgrade has been increased.
23 Supporting XN-9100
(1)Barcode test can now be available.
(2)Previously, there was a possibility that the status of analyzers on CT-90 (icon)
was not correctly displayed. In order to resolve this situation, the status of the
icon can now always be updated.
(3)XN-9100 can now connect both CV-50 and CV-55. (Both CV-50 and CV55
can now be connected with the one IPU.)
24 For North American markets, analysis algorithm has been developed by modifying ver. 00-20 and ver.
21.00-00.
Resolved issues
1 Previously, piercer adjustment screen could not close even when "CLOSE" button was pressed. This
has been resolved.
2 Previously, lot registration information was not inherited when program was updated. This has been
resolved.
Remarks
Module Version
IPU 00-05
Device Control Manager 00-05
Analysis Module 00-05
Analysis Sequencer Module 00-05
Sampler Sequence 00-05
Analyzer 02-59
Sampler 00-16
ECR:317A081
<Release Date>
<Part No./Name> June 2016
AB833655: CDR_ASSY NO.271(DVD1XNVX(00-04))
Changed Specifications
Resolved issues
1 Previously, updown motion error on aspiration unit occurred when auto rinse was performed after the
analysis in control or calibrator mode. This issue has been resolved.
Remarks
Module Version
IPU 00-04
Device Control Manager 00-03
Analysis Module 00-04
Analysis Sequencer Module 00-03
Sampler Sequence 00-03
Analyzer 02-57
Sampler 00-15
ECR:316D049
<Release Date>
<Part No./Name> June 2016
AB833655: CDR_ASSY NO.271(DVD1XNVX(00-04))
Changed Specifications
1 Usability for creating manual analysis gate has been improved as follows.
1. When creating manual analysis gate, the gate can be closed by double-click.
2. At closing a gate, the program has been changed not to transition to the next manual analysis gate.
3. “Modify”, “Add”, and “Delete” menus are now displayed as context menu by right-clicking on
selected vertex coordinate display of manual analysis gate information.
4. Selected vertex on the manual analysis gate can now be moved by pressing arrow keys. However,
the vertex does not move over the scattergram area.
5. All selected manual analysis gates can now be moved by pressing “Shift” and arrow keys.
6. The coordinate which is the display position of gate name can now be saved.
2 Settings for extension parameter names and gate colors on the manual analysis (extended) have
been added as follows.
1. An item “Extension parameter setting” has been added to species setting window so that names
and colors of extension parameter for each channel have been configured.
2. Color and item name for the selected extension parameter can now be configured by right-click the
selected extension parameter gate information on the manual analysis (extended) window.
3 The displaying order of species on the work list screen has been changed to follow the settings of
displaying order on the species configuration.
4 Changing P/N (Positive/ Negative) function on the modification dialogue of analysis result can be
disabled by the service settings.
5 Printing layout for analysis result LP printing has been changed to the layout with large left page
margin.
6 Initial values for species CAL for PLT-I have been changed as follows. (Species: Mouse, Rabbit)
(001)Mouse: 1656
(005)Rabbit: 1177
7 WBC discrimination for WDFch has been improved. (Species : Mouse)
8 The numbers of digits for extension parameters RET-R0# through RET-R8# and RET-R1% through
RET-R8% have been changed so as to enable them to use with the same number of significant figure
for RET# and RET%.
9 Fixed password (c9.0) for “sysmex” account on Windows has now been changeable.
10 Error detection method for “Cannot recognize Fluorocell XXX information” has been changed. (XXX
stands for Dye parameters.)
11 Previously, users who can implement QC had the authority to delete the QC data. So as not to delete
the QC data by mistake, new type of authorized user that only can implement QC has been added.
12 Action messages for reagent replenish errors have been changed from “Open the Fluorocell XXX
cover and install new Fluorocell XX.” to “Fluorocell XXX is installed inside the instrument. Open front
cover of the instrument and install new Fluorocell XXX.”. (XXX stands for Dye parameters.)
13 Analysis results whose sample numbers containing the characters that cannot be used for naming on
Windows such as \, /, *, ?, “, <, >, and | can now be output in FCS files.
14 As a countermeasure against liposome interference, when instrument judges a liposome appearance
sample, the following processes can now be performed.
- Outputting WBC Abnormal Scattergram flag
- Attaching low reliable mark to WBC related parameters
- WBC re-discrimination
Resolved issues
1 Previously, the setting of ghost area on the PLT-F scattergram for human and Guinea pig were same.
This could result in falsely low value because PLT-F cluster tends to be depicted on lower value area
in SLF. To resolve this issue, ghost area on the PLT-F scattergram for Guinea pig has been modified.
(Species: Guinea Pig)
2 The initial value for "S_ANA_WNR_SLOPE_WNR" parameter has been changed as follows because
NBRC falsely high values Mouse WNRch could occur.
(001)Mouse: 203。
3 Discrimination error occurred on Non-Human Primates because distance between Neut cluster and
Eo cluster has been close on scattergram. To resolve this issue, the algorithm has been changed and
Neut/Eo re-discrimination process on the WDF (SSC-FSC) scatter gram has been added.
4 Previously, there was a possibility that giant platelets became falsely high value when these particles
interfered with WBC on WNR-ch. This has been resolved.
5 Piercer adjustment window could not be closed even after CLOSE button was pressed. This issue
has been resolved.
6 Previously, tube holder moved before checking upward and downward motion of piercer was
implemented and this could result in damaged piercer. To resolve this issue, startup sequence has
been changed to ascend piercer before the sample set unit moves at startup.
7 Previously, the piercer did not descend at cleaning RET/PLT-F chamber.
This has been changed to the piercer descending when cleaning the chamber.
8 The counted values for PLT-O and RBC-O decreased by 20% because FCM charging DP got tainted
by blood and could not discharge proper amount. To resolve this issue, cleaning sequence which
feeding CELLCLEAN into the FCM charging DP and cleaning has been added at shutdown or
cleaning.
9 Depending on the position of the piercer, air gap could not be created properly when switched to the
PD mode. To resolve this issue, the position where the air gap is created soon after switched to PD
mode, has been lowered 1.4mm from the previous position.
Remarks
Module Version
IPU 00-03
Device Control Manager 00-03
Analysis Module 00-03
Analysis Sequencer Module 00-03
Sampler Sequence 00-03
Analyzer 01-57
Sampler 00-15
ECR:316D049
<Part No./Name>
AU089393: CDR_ASSY NO.248
Changed Specifications
None
Resolved issues
None
Remarks