Professional Documents
Culture Documents
DVD1XN1X_VersionHistory_E
DVD1XN1X_VersionHistory_E
20-00
<Applicable to S/N> <Release Date> JP: Ver.22.20-00
Refer to ECR 322G056 February 2023 EU: Ver.22.20-00
AP: Ver.22.20-00
<Parts Information> AM: Ver.22.20-00
CG723992: CDR_ASSY NO.532(DVD1XN1X(22.20-00)) CN: Ver.22.08.08-00
AN616770: CDR_ASSY NO.478(DVD1XN2J(22.08.08-00)) (For China market only)
Changed Specifications
1 Now GP Customize function can be used in BB mode.
2 GDI function has been expanded.
3 A service setting has been added to prevent the message "Aspiration Sensor is OFF" from being
displayed when the blood aspiration sensor is not used.
4 Platelet CHECK Level2 has been removed from QC list. (EU market only)
5 In XN BB mode, Ret ch information which is not displayed as QC measurement result has been
removed from browser display. (EU market only)
6 Now XN CAL can be used for WBC calibration in the BB mode.
7 WBC calibration formula in BB mode has been changed.
8 Decision of RET Abn Scattergram has beefed up.
9 Decision of PLT Abn Scattergram has beefed up.
10 When a new stain cartridge is loaded, the opening date is written on the RFID, but this process
overlaps with writing the number of remaining tests, which can cause the analyzer to aberrate.
Therefore, the software has been modified to prevent double writing to RFID.
11 In Ver.22.19-00, service settings to use barcode reader F-100 were added. In addition to these
settings, adoption logic for sample ID read by barcode reader has been added to further increase the
reading rate of low quality labels. New barcode reader setting below has also been added in service
menu.
1 Resolved the issue that the suspended sampler measurement could not be resumed when the mode
switch was pressed to switch to the sampler measurement mode after the manual measurement was
completed only when SA-01 was used as the sampler.
2 The UI list of Brazilian Portuguese has been corrected.
Remarks
Module Version
[IPU] 00-22 Build205
[Device Control Manager] 00-22 Build205
[Analysis] 00-22 Build89
[Analyzer Sequencer] 00-22 Build205
[Sampler Sequencer] 00-22 Build205
ECR 322G056
DVD1XN1X Ver.22.19-00
<Applicable to S/N> <Release Date> JP: Ver.22.19-00
Refer to ECR 322B052 September 2022 EU: Ver.22.19-00
AP: Ver.22.19-00
<Parts Information> AM: Ver.22.19-00
CQ759907: CDR_ASSY NO.525(DVD1XN1X(22.19-00)) CN: Ver.22.08.07-00
Changed Specifications
1 IPU now works in an environment with Firewall Function enabled.
2 As SNCS integrates into Caresphere, IPU no longer uses the word "SNCS".
3 New service setting "Analysis interval for Exchange Piercer" to change the number of measurements
before piercer replacement message occurs has been added.
4 The default value for the service setting "Barcode reader type" has been changed.
Was Is
NFT-7175 (OPTOELECTRONICS) BL-185 (KEYENCE)
5 Add the new service setting to use barcode reader F-100 (OPTOELECTRONICS). (Each analyzer
system settings)
・Number of barcode reader decode matches
・Minimum reading digits for barcode reader
6 Norwegian and Swedish have been added as display languages in IPU. (for EU market only)
1 Resolved the issue that log off history would not be kept if Windows exited before writing log off
history to the database in the shutdown process.
2 Resolved the issue that some of the conditions in rule 1 of PLT-I/PLT-O switching were incorrect and
could not be switched when MicroR was 5.0%.
3 Resolved the issue that the management range can not be set correctly when QC file is restored and
QC limit setting is set as "%".
4 Resolved the issue that cleaning operation can not be performed properly when 2 units are started
simultaneously to be cleaned by CELLCLEAN AUTO on XN-9100/XR-9000 with BT-50 in the
configuration.
5 Resolved the issue that grade value of Q-Flag does not become blank when measurement result is
blank data.
6 Resolved the issue that inability to restore older versions of smp files in Ver.22-18-00.
Remarks
Module Version
[IPU] 00-22 Build194
[Device Control Manager] 00-22 Build194
[Analysis] 00-22 Build80
[Analyzer Sequencer] 00-22 Build194
[Sampler Sequencer] 00-22 Build194
ECR 322B052
DVD1XN1X Ver.22.18-00
<Applicable to S/N> <Release Date> JP: Ver.22.18-00
Refer to ECR 321A036 March 2022 EU: Ver.22.18-00
AP: Ver.22.18-00
<Parts Information> AM: Ver.22.18-00
AK910980: CDR_ASSY NO.504 (DVD1XN1X(22.18-00)) CN: Ver.22.10.00-00
Changed Specifications
1 XR can now control XR-9000 in which BT-50 is included in the configuration.
2 Users who do not have the right to change settings can now switch the reagent from RU to
CELLPACK DCL.
3 A service setting has been added to set the number of tests to be taken before performing the "Drain
RBC Isolation Chamber". (Setting name: Analysis interval for drain RBC Isolation Chamber,
Execution timing: During startup sequence at next startup)
4 The determination of the WBC Abn Scattergram has been enhanced when the fluorescence
intensity (SFL) of the WBC decreased.
1 Resolved the issue that IPU stops when shutdown is complete when the host computer is
turned off during shutdown.
2 Resolved the issue that the screen lock is re-activated when the screen is not operated after the
screen lock is released by IC card.
3 Resolved the issue that expiration date of reagent is not displayed correctly if the analyzer is
shutdown without measurement after staining solution replacement.
4 The classification algorithm has been modified so that MN and PMN could be correctly classified
when measuring the sample that the population of multinucleated cells appear in area with high
fluorescence intensity (SFL) in BF mode. (An algorithm is added to detect the misclassification of a
part of PMN as MN in the macrophage region.)
5 The classification algorithm has been modified so that LYMPH and MONO could be correctly
classified when measuring the sample with aberration cells appearing in area with high fluorescence
sensitivity in WB mode.
6 Resolved the issue that NRBC is not calculated correctly (false high or false low) due to a problem of
NRBC classification algorithm.
Remarks
Module Version
[IPU] 00-22 Build182
[Device Control Manager] 00-22 Build182
[Analysis] 00-22 Build76
[Analyzer Sequencer] 00-22 Build182
[Sampler Sequencer] 00-22 Build182
ECR 321A036
DVD1XN1X Ver.22.17-00 Upgrade Procedure
<Applicable to S/N> <Release Date>
Refer to ECR 321A035 November 2021
Changed Specifications
1 The classification algorithm has been changed for the residual RBC measurement in BB (Blood Bank)
mode as a countermeasure for false high in residual RBC.
2 The QC parameters for BB mode can be output to the external QC via online.
3 The RBC-O parameters, displayed as “RBC” in “BB Res Control Level1” and “BB Res Control Level2”
materials, are deleted from the QC parameters for BB mode.
4 The QC functions in BB mode become available for European and Asia Pacific markets.
5 For North American markets, WBC(BB) is changed from the research parameters to the reportable
parameters.
1 Even if the analysis mode was set to “Normal” after a “Precision Check” analysis was performed in
any analysis modes other than “Normal” in the Service mode setting, the analysis mode stayed the
same as it was.
This issue has been resolved. The “Precision Check” analysis can now be performed correctly even
after the analysis mode is switched from other modes than “Normal”.
2 The classification algorithm for WBC and NRBC has been changed to resolve the following 3 issues.
(1) NRBC could be false low if abnormal nucleated RBCs interfered with the WBC region on WNRch
and this caused incorrect WBC/NRBC classification.
(2) NRBC could be false high if unknown particles interfered with the TNC region on WNRch.
(3) NRBC could be false low when nucleated RBCs did not appear in the expected position on
WNRch and this caused incorrect NRBC classification.
After this change, the following operation are performed when the algorithm decides that the
classification cannot be performed normally.
(1) The NRBC+WBC cluster is grayed out.
(2) The WBC related items are masked.
(3) The “WBC Abn Scattergram” message is output.
3 When “RBC Abn Distribution”, indicating the possibility of abnormality on RBCch, was output, the low
reliability mark [*] was appended to the RET%, IRF, LFR, MFR, HFR, and RET-He items calculated
only in RETch.
This issue has been resolved, and the low reliability mark is not appended to the RET%, IRF, LFR,
MFR, HFR, and RET-He items when “RBC Abn Distribution” is output.
4 When the system was XN-9000 configuration, it was possible that the XN analyzers and CV-50 would
adopt sample numbers read by BT-40 instead of those read by the XN analyzer after BT-40 read the
sample barcodes. This would cause adopting incorrect sample numbers if user changed sample
position in the sample rack.
This issue has been resolved, and sample numbers read by XN analyzers are adopted at the time of
measurement.
5 Units for some parameters were not output correctly when measurement items were output via GDI.
This issue has been resolved.
Remarks
ECR 321A035
DVD1XN1X Ver.22.16-00 Upgrade Procedure
<Applicable to S/N> <Release Date>
Refer to ECR 320B032 February 2021
Changed Specifications
1 Assay value for PLT CHECK can be used when accessing “Read Assay File” in QC registration
screen, so that it can use PLT CHECK with barcode reading in BB mode.
2 Outputting QC comment in each plot in QC chart to a HOST computer is supported.
(ASTM format only)
QC Comment
3 WBC(BB) parameter has been changed from reportable parameter to research parameter.
(For North America market only)
4 Canceling the database optimization process has been supported.
[Cancel] button
5 “RE-MONO” has been added as a service parameter and they can be output to the host computer.
Parameters listed below are output as service parameter to host computer. (Not displayed in IPU)
(1) RE-MONO#
(2) RE-MONO%
(3) RE-MONO%M
This function is supported in ASTM format and Standard format only.
6 Adding service setting “Sensitivity adjustment of the PLT-F channel in QC analysis”. Plotting position
of PLT cluster on PLT-F ch can be adjusted by this setting. (Setting range is 0-14 and default value is
0)
7 Message displayed when reagent supply source is changed between RU and DCL has been
changed. Previously when RU-20 has any problem and reagent source is changed to DCL, and then
changed back to RU-20 again, at that time the message displayed was confusing and customer can
connect DST directly. To address the issue, message has been changed.
8 Memory size IPU can use has been changed from 2GB to 4GB. (Only for Windows10 64bit OS)
Previously there was “Out of Memory” error can occur because IPU can access up to 2GB of memory.
The memory size that IPU can access is extended to address this issue.
9 When display language is changed from English to local language, some wordings have been
changed to local language.
10 QC data is no longer written directly to the folder accessed by Support Manager.
QC data was directly written to the folder monitored by Support Manager.
However, to prevent zero byte (no content) of data from being sent to the SNCS server, QC data is no
longer written directly to the folder monitored by Support Manager now.
11 Previously when starting the “GP customize” program there was a possibility to have “.NET
Framework error” in the region where comma is used as decimal point. This issue is solved.
2 In APD Parameter Setting menu, service engineers could not input voltage value as they intended.
This issue has been resolved.
3 Some of functionalities could not be used such as reagent replacement, measurement mode change,
when logging on to the IPU using an IC card while the power of the main unit is turned on.
This issue has been resolved.
4 WBC count could become 1/3 when analyzing a whole blood sample in the LW mode if following
steps below are taken. This issue has been resolved.
(1) Select Low WBC mode, and click OK to close the dialog.
(2) Input sample No, check “Query to Host” and close the dialog by clicking [OK] button.
(3) Receive analysis order for LW mode from host computer.
(4) In “Change Measurement Mode” select Whole Blood..
(5) Press start switch to run a sample, without accessing “Manual Analysis” dialog.
5 With the configuration that there are two main units are used (i.e XN-2000) there was a possibility not
to show restart button after shutdown. This issue has been resolved.
In case that left analyzer is turned OFF and right analyzer is turned ON, restart button could not be
displayed when shutting down the analyzer with CELLCLEAN on rack. This issue has been resolved.
6 In BF mode, the sum of WBC was not 100% in some specific cases. This issue has been resolved.
Remarks
ECR 320B032
DVD1XN1X Ver.22.15-00 Upgrade Procedure
<Applicable to S/N> <Release Date>
S/N: Not fixed yet March 2020
Changed Specifications
1 The versions of VisualStudio and Microsoft SQLServer have been changed. The installed applications
depend on Windows OS.
The service life for previous versions of Visual Studio and Microsoft SQL Server will end before the
end of the sales of XN analyzers. To resolve the situation, the versions of Visual Studio and Microsoft
SQL Server for Windows 10 have been changed.
The new VisualStudio and SQLServer are installed automatically when the installer runs. .NET
Framework4.7.2 is installed as well.
Note:
(1) Once upgraded to Ver.22.15-00, downgrading to previous versions cannot
be performed.
(2) On Windows7, VisualStudio2017 is installed. However, SQL Server 2016 is
not installed.
2 The “Blood Bank (BB) mode (IVD)” whose analysis parameters are “reportable” has been added for
XN-10 and XN-20 only. *License is required.
The BB mode (IVD) can be applied to only XN-10 and XN-20 analyzers with the PLT-F channel such
as XN-10 [B1] and XN-20 [A1] in the XN-1000 or XN-2000 configuration.
Note:
(1) In case that BB mode (RUO) has been used in previous versions
(Ver.22.09-00 thru Ver.22.13-00) and the BB mode (IVD) is introduced, the
IPU program needs to be re-installed. And then, the license key for the BB
mode (IVD) needs to be entered.
(2) In case of the XN-2000 configuration, 2 licenses for both analyzers are
required to use the BB mode (IVD).
1 When WBC aggregation occurred on the WNR ch, “BASO#” and “BASO%” research parameters
were not masked. This has been changed so that those parameters are now masked.
2 When the data form is “0.XXXX” on the “Target/Limit Setting” screen, the data could be input only to
three decimal places and could not be input to the four decimal places. This issue has been resolved.
This issue occurred on the following items:
3 After the QC file whose QC file number had been deleted was registered with different lot number, the
QC file could not be deleted. This issue has been resolved.
4 When a manual measurement was performed in BB mode, the QC sample taken into the analyzer
could not be taken out and the analyzer would not start measurement until restarting.
5 When starting the IPU program, mounting VHD could fail and “MountError” would occur. This issue
has been resolved.
6 The French and Spanish messages have been modified.
7 The stability of WBC classification function has been improved for the BB mode quality control.
8 On Windows10, there was a case that an error dialogue appeared and IPU locked up when the SNCS
router was connected. This issue has been resolved.
9 On Windows 10, the execution of DB restoration tool could fail when upgrading from Ver.22.11-00 or
earlier. This issue has been resolved.
10 When SA-01 was used on the XN-1000 configuration, the internal error could occur after performing
the followings:
(1) Sample measurement started on the sampler mode.
(2) After the analyzer picked up and mixed a sample tube, the mode switch was pressed and
switched to the manual measurement mode.
(3) After the analyzer returned the sample tube to the rack, the rack was removed by hand.
(3) The mode switch was pressed again to change to the sampler measurement mode.
(5) The rack removed on the previous step was entered again and the measurement started.
(6) The internal error occurred when the analyzer detected the first sample tube.
This issue has been resolved.
11 Error messages displayed on XN IPU for the RU-20 pressure errors were incorrect. This issue has
been resolved so that the same messages as shown in the IFU will now appear.
12 Previously, analysis results for WBC related parameters were not masked, depending on the “Mask of
analysis result with misclassification for WBC differentials” service setting, when NRBC interference
to WBC area was detected and “WBC Abn Scattergram” was output. This issue has been resolved
and WBC related parameters are now masked.
Applicable parameters
NEUT#, NEUT%, LYMPH#, LYMPH%, MONO#, MONO%, EO#, EO%, BASO#,
BASO%, NRBC#, NRBC%, IG#, IG%, WBC-N, WBC-D, WBC-P, HFLC#, HFLC%,
NEUT#&, NEUT%&, BA-N#, BA-N%, BA-D#, BA-D%
13 When the “Limit Setting” was set as “Ratio(%) for the “QC Setting” and QC lot was registered, “Limit
range(%)” for sensitivity items (WNR-X, WNR-Y, WNR-Z etc.) were 100%. However, normal users
cannot change the sensitivity items and the limit were not managed properly. (Only the “Service” user
can change the sensitivity items)
This issue has been resolved by using “Differential(#)” regardless if limit is set to “Ratio(%)” or
“Differential (#)” for sensitivity items.
14 After the “Positive ID check error” occurred, SP did not make smear when the rack with error was re-
entered. This issue has been resolved.
15 On XN-2000 configuration, if only one of 2 analyzers performs measurement repeatedly, the piercer
mechanism unit in the other analyzer did not return to the initial position and continually moved to
right. This issue has been resolved.
On analyzers with Ver.22.09-00 or later, the piercer mechanism unit moved to right when the
pneumatic unit turned to “Sleep” after the instrument status changed from “Working” to “Ready”.
However, initialization of the piercer mechanism unit was not performed until starting measurement.
Therefore, if the instrument status changed from “Working” to “Ready” without performing
measurement repeatedly, the piercer mechanism unit continually moved to right.
To resolve this issue, initialization of the piercer mechanism unit is now performed when the
pneumatic unit turns to “Sleep”.
In addition, the piercer mechanism unit does not move to right when the “Parts information KIT
NO.458” setting value is set as “Not Installed” in the “NVram” setting.
16 When the “Blasts/Abn Lympho?” Q-Flag positive level was set as 90 and “Blasts?” and “Abn
Lympho?” are set as default value (100), there was a case that the “Abn Lympho?” Q-flag was
incorrectly judged as “Positive” though the Q-Flag value of the measurement result for “Abn
Lympho?” was 90.
This issue has been resolved and the Q-Flag judgment is always performed based on the positive
level configured in “Service Settings”.
Remarks
ECR 319E046
DVD1XN1X Ver.22.13-00
(North American Markets only)
<Applicable to S/N> <Release Date>
S/N: Not fixed yet December 2019
Changed Specifications
1 The following changes have been made for “BB mode”.
(1) Plasma measurement function has been deleted.
(2) The “RUO” warning message has been deleted.
(3) The following BB mode parameters now become reportable.
- RBC(BB)
- HGB (BB)
- HCT (BB)
- PLT (BB)
- WBC (BB)
None
Remarks
This program is only for BB-mode evaluation in the North American markets.
ECR 319F015
DVD1XN1X Ver.22.12-00 Upgrade Procedure
<Applicable to S/N> <Release Date>
S/N: Not fixed yet August 2019
Changed Specifications
1 A message displayed on “Aspiration Sensor Cal” has been changed.
Before change Execute Aspiration Sensor Calibration after Aspiration Sensor was
executed.
After change Execute Aspiration Sensor Calibration after Aspiration Sensor
Adjustment was executed.
2 For instruments with the “EU” destination, judgment calculation algorithms for “Blast?”, “Abn
Lympho?”, and “Blasts/Abn Lympho?” Q-flags have been changed. In addition, a service setting has
been added to decide whether the changed algorithms are used.
1 When the number of measured samples exceeded 100,000 after installation, data bacame
fragmented and database volume unintendedly increased. When the size of database exceeded 4
GB, new measurement results could not be added. To resolve this issue, the following measure has
been implemented.
(1) IPU will perform the database defragmentation (optimization) as soon as the database uses the
specified volume. When the optimization is being performed, the following progress bar appears:
(2) A setting to select frequency for database optimization from High, Middle, and Low has been
added to “System” in “IPU Setting”. The frequency depends on the setting value and the number
of samples per day. The rough indication is as follows:
Frequency of optimization
Setting Required
value 600 analyses 400 analyses 200 analyses time
per day per day per day
High Every 12 days Every 19 days Every 37 days 10 minutes
Middle Every 37 days Every 56 days Every 112 days 15 minutes
Low Every 87 days Every 131 days Every 262 days 25 minutes
In addition, a database optimization tool (DBDefrag.exe) to defragment database manually and a
database repair tool (XN_DatabaseRepairer.exe) have been added to “C:\Program
Files\Sysmex\Tools” folder.
Note:
To execute DBDefrag.exe and XN_DatabaseRepairer.exe, logon as Windows administrator
is necessary.
2 On the PLT-F ch, particles which were not PLT were mistakenly counted as PLT, resulting in falsely
high PLT. To resolve this issue, when the abnormal scattergram appears, re-classification will be
performed and the “PLT Abn Scattergram” IP message will be shown. In addition, the low reliability
marks will be placed next to the PLT related parameters.
Caution:
The QC backup files obtained on XN with Ver.22.12-00 cannot be restored on XN with
Ver.22.11-00 or lower. In addition, backup files obtained on XN with Ver.22.11-00 or lower
cannot be restored on XN with Ver.22.12-00 if the QC comment contains comma “,”.
4 When opening database failed at startup, thresholds for all Q-Flags were 0 since thresholds for Q-
Flags could not be read. If measurement was performed with this condition, all Q-Flag results were
“Positive”. To resolve this issue, when opening database failed, retrial is executed. If retrial failed, the
following message box appears and measurements cannot be performed until restarting IPU.
Massage box
5 On Ver.22.09-00 thru Ver.22.11-00, all logon names displayed on the “Audit Log” screen was
"(Logoff)”. This issue has been resolved.
6 On Ver.22.10-00 and Ver.22.11-00, a click of the [Left] button on the “Piercer Adjustment” screen
moved the piercer 4 pulses although the screen changed by only 1 pulse. This issue has been
resolved.
Remarks
None
ECR 319C033
DVD1XN2J Ver.22.08 (22.08.01-00) Upgrade Procedure
(For China market only)
<Applicable to S/N> <Release Date>
Refer to ECR 321I017 October 2021
Changed Specifications
Same contents as DVD1XN1X Ver.22.12-00.
Remarks
None
ECR 321I017
DVD1XN1X Ver.22.11-00 Upgrade Procedure
<Applicable to S/N> <Release Date>
S/N: Not fixed yet June 2019
Changed Specifications
1 To improve NEUT/ EO differentiation, the following changes have been made.
(1) The NEUT/ EO differentiation algorithm has been changed so that NEUT and EO values in EO
false low samples can be calculated correctly.
(2) The NEUT/ EO differentiation algorithm has been changed so that NEUT and EO values in EO
false high samples can be calculated correctly.
(3) The judgment conditions for the “WBC ABn Scattergram” flagging used to retest EO and the
“Confirm eosinophil” action message have been changed to improve the specificities.
None
Remarks
None
ECR 318J049
DVD1XN1X Ver.22.10-00 Upgrade Procedure
<Applicable to S/N> <Release Date>
S/N: Not fixed yet March 2019
Changed Specifications
1 The logon function using an IC card reader has been added only for Windows 10.
2 The GDI interface has been added. In addition, the GDI activation service setting has been added.
3 A service setting for the smart indicator (SI-14) has been added.
Param1 Param2
Top 2byte Bottom 2byte 4byte
Error occurrence Error type Error status
position
01: Sequencer 01: Discrepancy on decrypted 0001: REAGENT_ERROR_UID
side UID at an event other than
startup check
02: Different tag being read 0001: REAGENT_ERROR_UID
from the tag read previously
03: Unregistered third party 0001: REAGENT_ERROR_UID
reagent tag being read
04: Failure on password 0002: REAGENT_ERROR_PASS
authentication
05: ISO15963 error -
occurrence
02: DCM side 06: Discrepancy on remaining 0003: REAGENT_ERROR_COUNT
test count in UID existing in
record
07: Expired (Sysmex genuine 0006:
reagent) REAGENT_ERROR_EXPIR_GENUINE
07: Expired (Registered third 0007:
party reagents whose coded REAGENT_ERROR_EXPIR_REGISTER
UID and UID matches) ED
07: Expired (Registered third 0008:
party reagents whose coded REAGENT_ERROR_EXPIR_THIRDPART
UID and UID do not match) Y
08: SLIX2 authentication 0004: REAGENT_ERROR_AUTHE
failure (signature verification
error)
08: SLIX2 authentication 0005: REAGENT_ERROR_MANU
failure (manufacturer error)
09: Registered third party -
reagent being set again
1 In XN-1500, there was a possibility that a sample tube was not returned to the sample rack when
performing sampler analysis after correcting an error categorized as “Sampler Analysis Stop 3”, which
is errors relating to the analysis line such as a rack move error. This issue has been resolved.
2 When SA-01 was used, there was a possibility that a background check would not finish and the
instrument status did not become “Ready” if there was a lot of files (more than 5000 data files). This
issue has been resolved.
3 “Low count error” and “PLT sampling error” can now be output correctly.
4 The classification function for WNR ch has been improved for low RBC.
5 The motor operation on RBT aspiration has been improved since there was a possibility of a potential
motor failure.
6 The following issues on Ver.22.09-00 have been resolved.
(1) The “No Analysis Orders” error did not occur.
(2) Sample tubes were not transferred to the SP.
(3) The “Aspiration Sensor is OFF” error occurred though “Aspiration Sensor” setting was set with
“Use”.
7 If the birth date was not registered in the “Register Patient Information” dialogue, the current date was
registered as the birth date. This issue has been resolved and the patient birth date is no set if the
birth date is left blank. Additionally, the usability has been improved.
8 When replacing reagents on XN-2000 or etc., instrument status could not become “Ready” if the
mode was switched between registering new reagents and executing replacement. This issue has
been resolved.
9 When Hypo-he became high, a false low could occur in rare cases. This issue has been resolved.
10 Q-Flag judgment values could show 100 although flagging judgment result was “Negative”. This issue
has been resolved so that judgment values are calculated correctly.
The affected flags are as follows:
- RBC Agglutination?
- Turbidity/HGB Interf?
- Iron Deficiency?
- HGB Defect?
- Fragments?
Correct Incorrect
11 When SA-01 was used and following steps were performed, removing blood on bottom surface of the
tube cap could not be performed. This issue has been resolved.
(1) Sample was measured in the “Manual” mode with “Cap Open”.
(2) Sample was measured in the “Sampler” mode.
Remarks
None
ECR 318H059
DVD1XN1X Ver.22.09-00 Upgrade Procedure
<Applicable to S/N> <Release Date>
S/N: Not fixed yet October 2018
Changed Specifications
1 For Chinese markets only, the XN-1590 configuration has been added.
2 The analysis channel, used as adopted PLT and WBC result (“switching”), can now be displayed on
“Cumulative” and HPC screen.
Adopted channels (performed “switching”) can be identified with a marking starting with”&” on the
screen as follows.
These markings are displayed on “Numerical”, “Graph” screen, and printed results for “Cumulative”
and “HPC”.
Markings added to the result indicate the channel. The explanation is as follows.
Meaning Mark
Adopted result on WBC-D &D
Adopted result on PLT-O &O
Adopted result on PLT-F &F
2 The blood bank mode (Research Use Only) has been added. (To activate the blood bank mode,
license key is required.)
License keys to activate the blood bank mode can be only applied to XN-1000 or XN-2000
configuration whose XN-10 and XN-20 analyzer(s) have PLT-F channel.
1 In the XN-1500 configuration, if the following steps were performed, SA-21 could not be used. This
issue has been resolved.
(1) XN was powered on by powering SP-50 ON with WOL.
(2) XN IPU was shutdown.
(3) Although SP-50 was being used, XN IPU was powered OFF and SA-21 could not be used.
2 Previously, due to the failure of IPU startup process, the IPU program could fail to start normally and a
Mounter.exe termination error could occur. This issue has been resolved.
3 When a new sample ID was read while multiple samples were selected on “Worklist”, a lot of
additional samples could be selected erroneously and IPU operation could be slow. This issue has
been resolved.
4 The “Aspiration unit left-right motor error” could occur on XN analyzers which has a countermeasure
kit against erroneous high HGB. This issue has been resolved.
5 Previously, Q-Flag value “0” was not displayed (left blank) for Q-Flag even though parameters are
judged as “Negative”. This issue has been resolved and Q-Flag value “0” is now displayed.
6 On WNRch, WBC could erroneously be differentiated as NRBC. This issue has been resolved.
7 BASO result measured on WDFch is now adopted if BASO result measured on WNRch is found
abnormal.
8 The sum of WBC was not 100% in some specific cases. This issue has been resolved.
9 When using XN MemTool (initialization tool for XN analyzers and samplers) for the SA-21, rack feed-
out pulse value for the SA-30 was set instead. This issue has been resolved.
10 When “Auto output” configuration was set and analysis result was automatically sent to the host
computer, the action messages for the result of delta check could not be sent to the host computer.
This issue has been resolved.
Remarks
None
ECR 318C064
DVD1XN1X Ver.22.08-00 Upgrade Procedure
<Applicable to S/N> <Release Date>
S/N: Not fixed yet August 2018
Changed Specifications
None
1 With program version 22.07, RFID tag issue had been addressed, however when upgrading the
device driver was not updated and the countermeasure was not activated. This issue has been
resolved.
2 When viewing IFU in the IPU, the application program has been changed from Acrobat to “PDF
viewer” with version 22.05 and “PDF viewer” installation program was added. However, at version up,
that program was not registered in the “White List” for McAfee and error message “IPU.exe has
stopped working….” appeared and IPU program did not start up at next power on. This issue has
been resolved.
3 With the Rerun/Reflex/Comment Rule, when Retest:Eo was judged as “True” (means “Retest:Eo”
appears on the screen), Retest: PLT was also judged as “True” and Action that was set in rule screen
was applied. This issue has been resolved.
Remarks
None
ECR 318E043
DVD1XN1X Ver.22.07-00 Upgrade Procedure Upgrade Procedure
(FileDeliverySystem)
<Applicable to S/N> <Release Date>
S/N: Not fixed yet April 2018
Changed Specifications
None
“Add the rule for ‘WBC Abn Scattergram’ not to miss eosinophilia” setting screen
(3) A service setting “Display conditions of the analysis results for NEUT and EO” is no longer
displayed on “Service settings” screen.
(4) New action message “Confirm eosinophil and neutrophil count by other methods” has been added
to host output function.
(5) A pull-down menu option [Retest:Eo] has been added to “Action Message(s)” on “Register Rerun/
Reflex/Comment Rule” screen as a function for “ActionMessageDetail”.
2 A retry process for “Fluorocell RFID tag error”, which occurred due to writing failure, has been added.
Remarks
ECR 318D005
DVD1XN1X Ver.22.06-00 Upgrade Procedure Upgrade Procedure
(FileDeliverySystem)
<Applicable to S/N> <Release Date>
S/N: Not fixed yet March 2018
Changed Specifications
1 On XN-1500 and XN-3100 (standalone), two service settings, “Query to host wait time” and “Query
host computer for Rerun/Reflex analysis”, have been added.Therefore in ASTM format, using results
of initial or Rerun/Reflex analysis, the XN analyzer inquires the host computer for Rerun/Reflex
analysis and/or SP orders.
(1) Each time when “Query host computer for Rerun/Reflex analysis” setting value is changed, rule
expression will be evaluated and rule supplementary explanation will be changed.
(2) When the SA-01 is used in XN-1000 system, setting menus, “Query to host wait time” and “Query
host computer for Rerun/Reflex analysis”, do not appear as the SA-01 does not have a function to
perform reanalysis.
(3) On XN-3100(standalone), the rack is not fed out until XN receives analysis result because if the
rack is fed out, reanalysis cannot be performed when reanalysis is required as a result of
QueryToHost.
(4) On XN-1500, if XN received the order which cannot be measured, both XN and SP skip the
sample aspiration.
(5) Depending on “Query host computer for Rerun/Reflex analysis” setting value (0 through 2),
contents displayed on IPU screen and SP order adopted when preparing smear differ.
1) When “Query host computer for Rerun/Reflex analysis” setting value is 0. (Do Not Query) (XN
behaves same as previous versions)
(a) SP rule is executed.
(b) The SP rule flow chart and tab appear on the rule screen.
(c) The SP rule check box appears on rule “Initialize” screen box.
2) When “Query host computer for Rerun/Reflex analysis” setting value is 1 (Query) or 2 (Query
Using QueryToHost function in Rule Screen)
(a) SP rule is not executed.
Orders which are sent back to SP, corresponding to inquiry from SP, are prioritized as follows:
1. QueryToHost orders at reanalysis
2. QueryToHost orders at initial analysis
3. Orders registered manually
The order sent is eventually adopted as SP order.
(b) The SP rule flow chart and tab does not appear on the rule screen
(k) If SP order exists on Work List at the time of reanalysis, the result of QueryToHost is prioritized
and the order is sent to SP.
(l) Two Settings “SP Rule Setting” and “Default SP Order” do not appear on “Analysis Settings”
screen for “SP Setting”.
“Analyzer Settings” screen for SP setting when the setting value is “1” or “2”
2 On XN-9100 system without BT, a service setting “System behavior when RET is ordered in the
system in which RET cannot be analyzed” has been added. When this setting is configured as “1:
Skip”, sample analysis will be skipped if its order includes parameters which an XN analyzer cannot
analyze. Then, the sample will be transported to another XN analyzer which can analyze the
parameter.
Previously, if order containing RET was sent to an XN analyzer which could not analyze RET, the XN
analyzed parameters except for RET. After this change, either one of settings can be selected from
“analyzing sample without RET” or “skipping sample analysis”. When upgrading from previous
versions, the setting value is “0: Analyzer without RET discrete”. (The XN behaves as same as before
this change.)
In addition, to enable this setting, a service setting “ASTM exception order setting” needs to be “1:
Use”. Therefore, default value for “ASTM exception order setting” has been changed to “1”.
“System behavior when RET is ordered in the system in which RET cannot be analyzed”
setting screen
3 IPU settings, analyzer settings, service settings, and backup file for rules are now output to
RemoteMaintenance in SNCS folder when a service setting “Output report file” is configured as “1:
Output”.
4 A service setting “Send reagent information to the host computer has been added. When this setting
is configured as “1: Send”, reagent information used for analysis is sent to the host computer.
6 For EMEA markets, judgment for WBC IP messages has been modified, such as Blast/Abn Lympho
and others.
7 From Ver.22.04-00, WRP CHECK EX can be used. With this change, related error messages have
been changed as well.
Applicable parameters
PLT, PDW, MPV, P-LCR, PCT
Remarks
ECR 317H098
DVD1XN1X Ver.22.05-00
<Applicable to S/N> <Release Date>
S/N: Not applicable January 2018
Changed Specifications
1 The XN IPU program can now run on Windows 10 PC.
Remarks
Application of Ver.22.05-00 to the field has been skipped.
ECR 317H098
DVD1XN1X Ver.22.04-10 Upgrade Procedure Upgrade Procedure
(For North American Markets Only) (FileDeliverySystem)
Changed Specifications
None
1 In WB mode only: differentiation algorithm of neutrophils and eosinophils has been changed for cases
of eosinophil abnormalities.
With this change, an action message “Confirm eosinophil and neutrophil count by other methods” and
two service settings have been added to notify users of possible falsely-low of eosinophil or
eosinophil abnormality.
(1) An action message “Confirm eosinophil and neutrophil count by other methods” and a service
setting “Use ‘Confirm eosinophil’ action message” have been added.
When a service setting “Use ‘Confirm eosinophil’ action message” is set as “1: Use” and if the
analysis result is judged as a possible false-low caused by eosinophil abnormality, a message
“Confirm eosinophil and neutrophil count by other methods” is output and the low reliability mark is
applied to NEUT#, NEUT%, EO#, and EO%.
In addition, a message “Confirm eosinophil and neutrophil count by other methods” appears on
“Action” field in “Main” tab and on “List Box” in “User”, “Lab. Only”, and “Lab. Only(RUO) tabs.
“Add the rule for ‘WBC Abn Scattergram’ not to miss eosinophilia” setting screen
(3) A service setting “Display conditions of the analysis results for NEUT and EO” is no longer
displayed on “Service settings” screen.
(4) New action message “Confirm eosinophil and neutrophil count by other methods” has been added
to host output function.
(5) A pull-down menu option [Retest:Eo] has been added to “Action Message(s)” on “Register Rerun/
Reflex/Comment Rule” screen as a function for “ActionMessageDetail”.
2 A retry process for “Fluorocell RFID tag error”, which occurred due to writing failure, has been added.
Remarks
None
ECR 318D063
DVD1XN1X Ver.22.04-00 Upgrade Procedure Upgrade Procedure
(FileDeliverySystem)
<Applicable to S/N> <Release Date>
S/N: Not fixed yet November 2017
Changed Specifications
1 The name of flag for infection has been changed from “pRBC” to “iRBC”.
・pRBC includes the meaning of “Parasite” which may emphasizes “Malaria infection”.
Therefore, it has been changed to iRBC (inclusion RBC) to be easily accepted by users.
[Influenced parameter]
pRBC(R), pRBC-WNR#, pRBC-WDF#
2 WRP CHECK EX has become available.
・With the discontinuation for “RANGE CHECK X”, a linearity check product “WRP CHECK EX” has
become available as a replacement.
3 In the body fluid analysis mode, the error massage indicating “there is no sample tube on the sample
holder” has been added.
・In the body fluid analysis mode, if there is neither normal nor micro sample tube on the sample
holder, the Error Message “No tubes are in tube holder” will be displayed.
This update is only applicable to the XN-series with a SENSOR_ASSY NO.11 [AS261366], which
was applied by ECR 315F082.
Changes:
Check box “Tube Sensor” has been added on the screen at manual analysis mode (body fluid).
Default value of the check box is “checked” and will be initialized for individual sample.
In case that there is neither normal nor micro sample tube on the sample holder with check box is
“checked”, measurement can not be performed and the Error Massage “No tubes are in tube holder”
is displayed.
4 In the case of configuration including five or more CV-55, the error recovery method of XN analyzers
has been changed.
5 The countermeasure against a writing failure for the number of remaining tests to RFID tag in the
case of emergency stop has been added.
・When emergency stop occurs while writing the number of remaining tests into genuine reagent
RFID tag, the information of the remaining tests could not be written into the tag and a genuine
reagent was incorrectly recognized as an ingenuine rseagent.
Program has been changed to prevent this malfunction even in the state of emergency stop.
6 Writing function of the number of remaining tests to RFID tag has been enhanced.
・There were cases that genuine reagent was incorrectly recognized as ingenuine regent because the
actual remaining amount of fluorocell did not match with the number of remaining tests written in
RFID.
To resolve this issue, the number of times using fluorocell in the service menu mode (Diaphragm
Pump) has been written into RFID tag.
1 Differentiation algorithm of neutrophils and eosinophils in case there are degranulated eosinophils
has been changed.
・In the case of degranulated eosinophils, there are cases that eosinophil cluster is too close or
overlap to neutrophil cluster that causes false-classified as neutrophil.
To solve this issue, differentiation algorithm program has been changed.
2 The issue that the measurement can not be performed as order in the manual analysis mode has
been resolved.
・There was a possibility that the measurement could not be performed as host order when opening
the manual analysis mode display to order the next sample before absorbing.
Program has been changed to resolve this issue.
3 The countermeasure against “Fluorocell RFID tag error” which is frequently occurred when using new
RFID tag on Fluorocell bottles has been added.
Remarks
None
ECR 317G030
DVD1XN1X Ver.22.03-00 Upgrade Procedure Upgrade Procedure
(FileDeliverySystem)
<Applicable to S/N> <Release Date>
XN-10 S/N: 32971 and thereafter October 2017
XN-20 S/N: 13958 and thereafter
XN-1000 Supply Parts S/N: Not applicable
XN-1500 Supply Parts S/N: Not applicable
XN-2000 Supply Parts S/N: Not applicable
XN-3000 Supply Parts S/N: Not applicable
XN-3100 Supply Parts S/N: Not applicable
XN-9000 Supply Parts S/N: Not applicable
XN-9100 Supply Parts S/N: Not applicable
XN-IPU S/N: Not applicable
Changed Specifications
1 The monitoring method for “Abnormal pressure loss” has been changed.
(During startup of pneumatic unit, “Abnormal pressure loss” error is monitored.)
2 Previously, in ASTM communication, when an unavailable parameter (IPF: Research parameter)
order was received from the host computer, XN output this parameter erroneously as “0”. (The XN’s
action when receiving unavailable parameter was not clearly defined in the XN.) In order to resolve
this situation, the XN's action when receiving unavailable order, which is not described in “ASTM Host
Interface Specifications”, has been added in service setting.
•“ASTM exception order setting”: Setting whether recognizing measurement order or not when
receiving undescribed order in Specifications. The initial value is “0: Do Not Use”.
This setting action is same as before updating program. This means that when IPF order is received
from the host computer, IPF result will be “0”. When “1” is selected, only ordered parameters*
described in the specifications will be measured. The other parameters will be excluded and not be
measured.
(Refer to Service Manual Chapter 5 Service Program.)
* Refer to “Table 4.3.3.4.2: List of Analysis Parameters Ordered” (page 29: “Automated Hematology
Analyzer XN series ASTM Host Interface Specifications”).
3 (For North American market only) (RUO) IFU for RU-21 can now be viewed on Built-in IFU.
4 Previously, restarting sampler analysis was not possible after recovering from “Rack removed” error
on SP-50 in XN-1500 configuration. The XN-1500 can now restart sampler analysis after recovering
from SP-50 “Rack removed” error.
5 The logging function has been enhanced so that the log can be recorded in detail when “Internal
Error” occurs*. (In order to get detailed information when a rare error occurs.)
* Internal Error occurs in following sequence error
• Unit activated twice
• Unit controlled illegally
• Motor activated twice
1 Previously, SP rule was not applied when measuring samples with raised bottom tube or Sarstedt
tube. As a result, smear was not prepared. This was because only normal sample tube was objective
sample tube. This has been resolved by adding raised bottom tube or Sarstedt tube as objective
sample tube.
2 Previously, there was a possibility that “Mount Error” (Windows error) occurred while starting up IPU
program. This has been resolved by extending monitoring time for starting up IPU program from 60
sec to 300 sec.
3 Countermeasures for false high WBC (-N) count have been made.
Symptom: WBC related parameters result became high.
Cause: Not specified
Countermeasures:
1 “Detection of falsely high WBC-N” algorithm* has been added to service setting. ([0: Do Not Use, 1:
Use] Initial value for EU destination is [1: Use]. The other markets are [0: Do Not Use].
2 A service setting to set the threshold of “Detection of the falsely high WBC-N” algorithm is added.
3 When falsely high WBC is detected, an action message “Suspect sample, check the sample.” is
displayed. Low reliability mark (*) is also attached to the WBC related item (WBC, NRBC#,
NRBC%).
* This algorithm is detected by comparing LC-I and TNC-N. (LC-I is obtained from counted particles
in higher channels of RBC ch.)
• Reference: Service Manual Chapter 5 Service Program
4 Previously, “Total Number of Blocks” was “2” regardless if outputting research parameters on DPS
host communication. In order to resolve this situation, “Total Number of Blocks” has been changed as
follows.
• Total Number of Blocks: “01” when outputting without research parameters.
• Total Number of Blocks: “02” when outputting with research parameters.
Remarks
None
ECR 317C001
DVD1XN1X Ver.22.02-00 Upgrade Procedure Upgrade Procedure
(FileDeliverySystem)
<Applicable to S/N> <Release Date>
S/N not applicable August 2017
Changed Specifications
1 Failure prediction function has been enhanced.
Data for items listed below has been monitored via SNCS to predict instrument failure.
- Draining time in startup sequence
- Values of each pressure in startup sequence
2 The performance of blood aspiration sensor has been improved.
To support market requests to increase the aspiration sensor stability, the calibration of blood aspira-
tion sensor has been added to correct the difference between discrete modes.
*Caution)
From Ver.22.02-00, blood aspiration sensor calibration is required when Service Engineer performs
actions like those shown below.
- Changing piercer or rinse cup
- Changing PCB NO.30001 (Refer to Service Manual Chapter 3)
3 Contents of files sent via SNCS have been changed.
4 Service setting for output of FCS file (used mainly for lab research) has been added to secure
additional HDD space.
- Setting value “0”: Not output FCS file (Default)
- Setting value “1”: Output FCS file
5 When lymphocyte is incorrectly discriminated as monocyte, lymphocyte shows falsely low value. To
resolve this issue, discrimination algorithm has been improved. (This algorithm has been adopted on
XN-L series as Ver.00-14)
6 Detecting condition has been modified to display “PLT-F channel error” when instrument detects lot
failure of Fluorosell PLT. (Refer to Service manual Chapter 6 "PLT-F channel error")
7 (For Chinese market only) SystemConfig has been revised so that XN-2800 can be installed.
8 Reagent management function has been improved.
- In order not to restart sample analysis on the instrument with simplified sampler, sampler analysis
method has been changed so that the restarting sampler analysis will not start while authenticating
stain tag (RFID tag). (Sampler analysis starts after completing authentication.)
9 When neutrophilic cells are incorrectly discriminated as eosinocyte, eosinocyte shows falsely high
value. To resolve this issue, discrimination algorithm has been improved.
10 As there was unnecessary movement of solenoid valve, the sequence has been revised so that WPC
homolyzer charging is executed only during WPC measurement sequence.
1 On XN-1000 SA-01 model, lockup when skipping sample aspiration has been resolved.
2 Lockup when suspending measurement while waiting for analysis order has been resolved.
3 Lockup when recovering from sleep mode has been resolved.
4 Lockup when detecting a Micro tube while reagent is depleted during manual measurement has been
resolved.
5 Error detecting function has been improved.
There was a possibility that residual waste liquid remained in RBC/HGB chamber due to a partial
clog. To detect residual waste liquid in the chamber, detect timing of the chamber draining has been
revised to move the timing ahead by 0.2 sec.
6 “WBC Abn Sct.” is an error message which is displayed when WBC discrimination is not performed
correctly due to low fluorescence signal. However, there were cases where this message was not
displayed even though WBC discrimination was not performed correctly. To resolve this issue, WBC
discrimination algorithm has been improved.
Remarks
None
ECR 316J044
DVD1XN1X Ver.22.01-00 Upgrade Procedure Upgrade Procedure
(FileDeliverySystem)
<Applicable to S/N> <Release Date>
S/N not applicable April 2017
Changed Specifications
1 Analysis algorithm for North American markets has been developed. (North American markets only)
• For North American markets, the setting range to select PLT Clumps algorithm has been set to “0 to
1”. For other markets, the setting range has been set to “0 to 2”.
1 Previously, when printing was done using GP Customize, the Russian language changed to the Latin
language in some regions. This issue has now been resolved.
2 Previously, there were some cases where the instrument stopped abnormally during startup. This
issue has now been resolved.
3 Previously, the analysis result of a sample number that starts with “QC” was not displayed if the lot
has not been registered. This issue has now been resolved.
Remarks
None
ECR 316I050
DVD1XN1X Ver.22.00-00 (Skipped)
<Applicable to S/N> <Release Date>
S/N not applicable April 2017
Changed Specifications
1 As for Sysmex user password on the IPU program, it is now possible to select between a one-day
password or a service-defined password, to prevent an unauthorized user from accessing the service
mode.
• When logged on as service user, “sysmex” will be displayed as “Logon Name”. The initial password
is “c9.0”. “sysmex” is displayed in “Logon Name” column, “Service” is displayed in “Operator Name”
column and “Build-in-User” is displayed in “Operator Info.” column.
• “Password Mode” has been added to Service Settings. The default setting is “Service defined”.
2 “Replace Piercer” has been moved from the User menu to the Service menu.
• Removed [Replace Piercer] from the Maintenance menu button and added it to the Service menu.
3 As countermeasures against use of reagents not authorized by Sysmex, the following changes have
been made.
• It is now possible to use ICODE SLIX2 RFID tags, which feature increased security.
• A message box will now be displayed to inform users if an unauthorized reagent has been set in the
instrument.
• The name of manufacturer and the place of production read from the RFID tag will now be saved in
the IPU database and displayed in the archives.
4 To support the RU-20 Upgrade Kit, the following changes have been made.
• Four analyzers can be connected with the RU-20.
• To connect four analyzers with the RU-20, the preparation volume of 9L/h is not sufficient.
Therefore, it is now possible to select between 9L/h and 12L/h in SystemConfig.
NOTE: The selection radio buttons will not be displayed when the RU-21 is selected.
(3) If XN-3100 or XN-1500 is selected in the initial setup wizard screen, the following screen will
be displayed and instrument ID will be stored.
If XN-3100 is selected
If XN-1500 is selected
(4) It is now possible to select connection with DIA (Digital Image Analyzer) if the system
configuration is XN-1500, XN-3000 (Standalone) or XN-3100 (Standalone), or when upgrading
the system.
Only when XN-1500, XN-3000 (Standalone) or XN-3000 (Stand alone) system configuration is
selected, “Connecting the DIA” button will be displayed in the System Configurator screen.
(5) When the IPU starts up on the XN-1500, the SP-50 will start up together with the analyzer. As
for XN-1500, “SP-50 connection settings” button will be displayed in the System Configurator
screen. (Refer to the picture in (4) listed above.
6 The speed of copying files during program upgrade has been increased.
Remarks
None
ECR 316I050
DVD1XN1X Ver.21.12-00 (European Upgrade Procedure Upgrade Procedure
markets only) (FileDeliverySystem)
Changed Specifications
None
1 Previously, when a sample was switched, patient information was replaced by that of another patient
as follows.
This issue has now been resolved.
Previously, on the XN-9000, patient information was replaced by that of another patient when the user
performed the following operations. This issue has now been resolved.
1) The user placed sample A (Sample No. A and patient ID is A) in position 1 of a sample rack and
performed a system analysis. Sample No. was read on the BT-40.
2) After the sample rack passed through the BT-40, the user moved Sample A from position 1 to a
different position on the same sample rack. The user then placed Sample B in position 1 (Samples
were switched).
3) When samples were switched (as mentioned in item 2 above), analysis information (i.e. analysis
order and patient information) on Sample B was not registered to the host computer.
4) Sample A's patient information was linked to both samples A and B, i.e. both samples A and B had
the same patient information originally recorded for sample A.
2 Previously, the analysis results of a sample number that starts with "QC" could not be displayed. This
issue has now been resolved.
With program versions 00-20 or lower, analysis results which were analyzed in the QC mode were
displayed if a sample has a sample No. that starts "QC" and a lot has not been registered on the
sample.
However, with version 21.00-00, analysis results of such sample were not displayed.
This issue has now been resolved.
Remarks
None
ECR 316G033
DVD1XN1X Ver.21.11-00 (European Upgrade Procedure Upgrade Procedure
market only) (FileDeliverySystem)
Changed Specifications
1 "Blast" flagging sensitivity has been improved. When there are many blasts in the blood, Lymphocyte
might have been classified as Monocyte which lead to missing "Blast?" flag.
"Atypical Lympho" flag has now better classification. In WDF channel when location of Lymphocyte
cluster and Monocyte cluster is very close, there is a possibility to classify Lymphocyte as Monocyte
and generate false "Atypical Lympho?" flag.
"Blast/Abn Lympho?" Q-flag calculation algorithm has been optimized.
1 Previously "Aspiration unit up-down motor error" occurs when running auto rinse after performing QC
measurement by customers or using control mode / calibrator mode in service program. This issue
has been corrected.
Remarks
None
ECR316F086
DVD1XN1X Ver.21.10-00 (patch Upgrade Procedure
program) (European market only)
<Applicable to S/N> <Release Date>
S/N: Not fixed yet May 2016
Resolved issues
1 It has been found that “Aspiration unit up-down motor error" error occurred when auto-rinse is
performed right after running QC on the instrument with program version 00-21. This has been
resolved.
DVD1XN1X Ver.21.10-00 (European Upgrade Procedure Upgrade Procedure
market only) (FileDeliverySystem)
Changed Specifications
1 "error code", "parameter 1" and "parameter 2" can now be displayed on IPU RU error history and can
also be output to SNCS.
2 Priority code has been added to IPU setting.
Set input-output setting of priority code. [0: Do Not Input or Output (Default), 1: Only Input, 2: Only
Output, 3: Input and Output]: (Service Manual P. 5-53)
3 The following data has been added for SNCS.
-Outputting service data and target limits on QC
-Outputting analyzed data for calibration
- Outputting NVRAM data
4 Result with barcode read error can now be performed Auto Validate.
(Service Manual P. 5-60)
5 Mark for “outside of linearity” in PD mode has been changed as follows.
WBC: 4400×10^2/uL
PLT: 500×10^4/uL
PLT-I: 500×10^4/uL
PLT-O: 500×10^4/uL
PLT-F: 500×10^4/uL
WBC-N: 4400×10^2/uL
WBC-D: 4400×10^2/uL
6 Exponential and logarithm formula have been added to rule function.
7 WC1 draining on startup monitoring time has been changed from 0.1 sec to 0.5 sec after starting WC1
draining.
8 Generating and outputting of scattergram for HPC on HPC mode is now available.
9 Infection parameters (AS-LYMP#, AS-LYMP%, RE-LYMP#, RE-LYMP%) can now be masked when
IP message "Blast/Abn.Lympho?" and "Blast?" appear.
10 Action message "Suspect Sample?" can now be turned off in manual mode by service setting.
Service Setting can be changed to "Used" or "Not Used". Default: "Used"
11 Specifications for sampling error have been changed as follows.
Currently, sampling error judgment is performed on first sampling time. It has been changed to
perform during every sampling time.
12 Information for connected reagent can now be output.
13 "Giant Platelet?" flag can be added by changing "PLT Clumps?" algorithm.
Following settings can be changed as follows.
Type A: (Same as version 00-17, High specific, Not judge "Giant Platelet?" )
Type B: (Same as version 00-18, High sensitivity, Not judge "Giant Platelet?" (default))
Type C: (Added by 21-00, Balanced with type A and B, judge "Giant Platelet?")
(Service Manual P. 5-57)
14 Flag judgment for "Dimorphic Population" has been improved.
15 "CELLPACK DFL" has been added to filter function of reagent replacment screen.
16 Notation for program version has been changed as follows.
"##. %%-&&"
##: Incremented when major functional change.
%%: Incremented when major problems are fixed.
&&: Assigned unique number for special order.
17 Previously, there was a possibility that NRBC became falsely high by affecting lipid particles. In order
to resolve this situation, lipid particles can now be discriminated as Ghost when lipid particles are
detected.
18 Previously, there was a possibility that NRBC was included in WBC even when NRBC appeared. In
order to resolve this situation, "WBC Abnormal Scattergram" flag is output and related parameters are
masked with hyphen when NRBC is interfered in WBC area.
19 Previously, there was a possibility that Lymphocyte became falsely low by detecting Lymphocyte was
falsely discriminated as monocyte. This has been resolved.
20 Following issues have been resolved.
- IG became falsely high when NEUT clusters divided into more than two. This has been resolved.
- IG became falsely low when IG cluster is divied from NEUT clusters. This has been resolved.
21 Following parameters have been added as reportable parameters.
"Delta-He", "RBC-He", "HYPO-He", "HYPER-He", "MicroR" and "MacroR".
Following parameters have been added according to licenses.
"RE-LYMP#" and "RE-LYMP%" (reportable parameter)
"AS-LYMP%L" and "RE-LYMP%L" (research parameter)
22 Grade Value for "Atypical Lympho?" calculation algorithm has been added. Setting can be changed
as follows.
Type A (default)
Type B (added calculation algorithm)
(Service Manual P. 5-57)
23 RE-Lymp discrimination algorithm has been changed by adding fixed gate.
Type A (Automatic gate) (default)
Type B (Fixed gate) (default for EU market)
24 Some characters in Brazilian Portuguese have been fixed.
25 IP message "Giant Platelet?" is now selectable.
(Service Manual P. 5-57)
Resolved issues
1 Previously, there was a possibility that piercer was damaged as tube holder was moved before
monitoring piercer ascending. In order to resolve this situation, sequence of piercer has been
changed to ascend before moving tube holder.
2 Previously, there was a report that result of PLT-O and RBC-O became low by 20%, caused by
lowering of FCM charging DP dispensing volume due to DP dirt. In order to resolve this situation,
cleaning FCM charging DP has been improved by applying CELLCLEAN cleaning for longer time at
shutdown and cleaning sequences.
3 Previously, piercer adjustment screen could not close even when "CLOSE" button was pressed. This
has been resolved.
4 Previously, lot registration information was not inherited when program was updated. This has been
resolved.
5 Previously, there was a possibility that the air gap was not correctly generated on PD mode. In order
to resolve this situation, position of air gaps for changing PD mode have been lowered by 1.4 mm
from the current position.
Remarks
None
ECR: 315G054
DVD1XN1X Ver.21.00-00 Upgrade Procedure Upgrade Procedure
(FileDeliverySystem)
<Applicable to S/N> <Release Date>
S/N: Not fixed yet May 2016
Changed Specifications
1 "error code", "parameter 1" and "parameter 2" can now be displayed on IPU RU error history and can
also be output to SNCS.
2 Priority code has been added to IPU setting.
Set input-output setting of priority code. [0: Do Not Input or Output (Default), 1: Only Input, 2: Only
Output, 3: Input and Output]: (Service Manual P. 5-53)
3 The following data has been added for SNCS.
-Outputting service data and target limits on QC
-Outputting analyzed data for calibration
- Outputting NVRAM data
4 Result with barcode read error can now be performed Auto Validate.
(Service Manual P. 5-60)
5 Mark for “outside of linearity” in PD mode has been changed as follows.
WBC: 4400×10^2/uL
PLT: 500×10^4/uL
PLT-I: 500×10^4/uL
PLT-O: 500×10^4/uL
PLT-F: 500×10^4/uL
WBC-N: 4400×10^2/uL
WBC-D: 4400×10^2/uL
6 Exponential and logarithm formula have been added to rule function.
7 WC1 draining on startup monitoring time has been changed from 0.1 sec to 0.5 sec after starting WC1
draining.
8 Generating and outputting of scattergram for HPC on HPC mode is now available.
9 Infection parameters (AS-LYMP#, AS-LYMP%, RE-LYMP#, RE-LYMP%) can now be masked when
IP message "Blast/Abn.Lympho?" and "Blast?" appear.
10 Action message "Suspect Sample?" can now be turned off in manual mode by service setting.
Service Setting can be changed to "Used" or "Not Used". Default: "Used"
11 Specifications for sampling error have been changed as follows.
Currently, sampling error judgment is performed on first sampling time. It has been changed to
perform during every sampling time.
12 Information for connected reagent can now be output.
13 "Giant Platelet?" flag can be added by changing "PLT Clumps?" algorithm.
Following settings can be changed as follows.
Type A: (Same as version 00-17, High specific, Not judge "Giant Platelet?")
Type B: (Same as version 00-18, High sensitivity, Not judge "Giant Platelet?" (default))
Type C: (Added by 21-00, Balanced with type A and B, judge "Giant Platelet?")
(Service Manual P. 5-57)
14 Flag judgment for "Dimorphic Population" has been improved.
15 "CELLPACK DFL" has been added to filter function of reagent replacment screen.
16 Notation for program version has been changed as follows.
"##. %%-&&"
##: Incremented when major functional change.
%%: Incremented when major problems are fixed.
&&: Assigned unique number for special order.
17 Previously, there was a possibility that NRBC became falsely high by affecting lipid particles. In order
to resolve this situation, lipid particles can now be discriminated as Ghost when lipid particles are
detected.
18 Previously, there was a possibility that NRBC was included in WBC even when NRBC appeared. In
order to resolve this situation, "WBC Abnormal Scattergram" flag is output and related parameters are
masked with hyphen when NRBC is interfered in WBC area.
19 Previously, there was a possibility that Lymphocyte became falsely low by detecting Lymphocyte was
falsely discriminated as monocyte. This has been resolved.
20 Following issues have been resolved.
- IG became falsely high when NEUT clusters divided into more than two. This has been resolved.
- IG became falsely low when IG cluster is divied from NEUT clusters. This has been resolved.
Resolved issues
1 Previously, there was a possibility that piercer was damaged as tube holder was moved before
monitoring piercer ascending. In order to resolve this situation, sequence of piercer has been
changed to ascend before moving tube holder.
2 Previously, there was a report that result of PLT-O and RBC-O became low by 20%, caused by
lowering of FCM charging DP dispensing volume due to DP dirt. In order to resolve this situation,
cleaning FCM charging DP has been improved by applying CELLCLEAN cleaning for longer time at
shutdown and cleaning sequences.
3 Previously, piercer adjustment screen could not close even when "CLOSE" button was pressed. This
has been resolved.
4 Previously, lot registration information was not inherited when program was updated. This has been
resolved.
5 Previously, there was a possibility that the air gap was not correctly generated on PD mode. In order
to resolve this situation, position of air gaps for changing PD mode have been lowered by 1.4 mm
from the current position.
Remarks
None
ECR: 316G057
DVD1XN1X Ver.00-20 build 10 Upgrade Procedure Upgrade Procedure
(SAP markets only) (FileDeliverySystem)
Changed Specifications
1 Differentiation algorithm of neutrophil and eosinocyte in case there are degranulated eosinocytes has
been changed.
* Degranulated eosinocytes could be false-classified as neutrophil.
In addition, a new setting “Display conditions of the analysis results for NEUT and EO” is added in
service settings. The setting enables to set result display setting and low-reliability mark setting.
Resolved issues
None
Remarks
For XN-10/20
ECR 317H094
DVD1XN1X Ver.00-20 build9 Upgrade Procedure Upgrade Procedure
(FileDeliverySystem)
<Applicable to S/N> <Release Date>
XN-1000 SUPPLY PARTS(EU) S/N: A9203 November 2016
and thereafter
XN-1000 SUPPLY PARTS (SA-01_EU) S/N:
A9203 and thereafter
XN-1000 SUPPLY PARTS(CHN) S/N: A9203
and thereafter
XN-1000 SUPPLY PARTS (SA-01_CHN)
S/N: A9203 and thereafter
XN-2000 SUPPLY PARTS(EU) S/N: A2890
and thereafter
XN-2000 SUPPLY PARTS(CHN) S/N: A2890
and thereafter
XN-9000 SUPPLY PARTS(EXP) S/N: A2422
and thereafter
XN-9000 SUPPLY PARTS(US) S/N: A2422
and thereafter
XN-9000 SUPPLY PARTS(CHN) S/N: A2422
and thereafter
XN-9000 SUPPLY PARTS(XN-11/XN-21)
S/N: Not fixed
XN-3000 SUPPLY PARTS(EU) S/N: A2170
and thereafter
XN-3000 SUPPLY PARTS(CHN) S/N: A2170
and thereafter
XN-9100(1) SUPPLY PARTS(US) S/N:
A1026 and thereafter
XN-9100(1) SUPPLY PARTS(EU) S/N:
A1026 and thereafter
XN-9100(1) SUPPLY PARTS(CHN) S/N:
A1026 and thereafter
XN-9100(1) SUPPLY PARTS(XN-11/XN-21)
S/N: A1026 and thereafter
Changed Specifications
None
Resolved issues
1 Previously, when a sample was switched, patient information was replaced by that of another patient
as follows.
This issue has now been resolved.
Previously, on the XN-9000, patient information was replaced by that of another patient when the user
performed the following operations. This issue has now been resolved.
1) The user placed sample A (Sample No. A and patient ID is A) in position 1 of a sample rack and
performed a system analysis. Sample No. was read on the BT-40.
2) After the sample rack passed through the BT-40, the user moved Sample A from position 1 to a
different position on the same sample rack. The user then placed Sample B in position 1 (Samples
were switched).
3) When samples were switched (as mentioned in item 2 above), analysis information (i.e. analysis
order and patient information) on Sample B was not registered to the host computer.
4) Sample A's patient information was linked to both samples A and B, i.e. both samples A and B had
the same patient information originally recorded for sample A.
Remarks
ECR 316G033
DVD1XN1X Ver.00-20 Upgrade Procedure Upgrade Procedure
(FileDeliverySystem)
<Applicable to S/N> <Release Date>
S/N: Not fixed yet November 2015
Changed Specifications
Resolved issues
1 Previously, QC barcodes could not be read if the sample number digit length setting was changed.
This has been resolved.
2 Mis-calculation for NEUT-RI and NEUT-GI has been resolved.
3 Incorrect malaria particle number in the scattergram has been corrected.
4 Previously, there was a possibility that giant platelets became falsely high value when these particles
interfered with WBC on WNR-ch. This has been resolved.
5 Previously, when analysis mode was changed to sampler mode after using micro sample tube,
sampler analysis could not be performed. This has been resolved.
6 Previously, when analysis mode was changed to sampler mode after using micro sample tube,
sampler analysis cannot be performed. This has been resolved.
7 Previously, "Blood aspiration sensor is OFF" message appeared, even when blood aspiration sensor
was ON and service mode setting screen was closed. This has been resolved.
Remarks
None
ECR:315A030
DVD1XN1X Ver.00-19D build 16 Upgrade Procedure Upgrade Procedure
(SAI markets only) (FileDeliverySystem)
Changed Specifications
1 Differentiation algorithm of neutrophil and eosinophil has been changed for the two cases shown
below.
[Case1]
In case of eosinophilia (especially in case of degranulated eosinophil) and the cluster appears nearby
neutrophil cluster, the analyzer falsely classifies the eosinophil cluster as neutrophil cluster.
Therefore, differentiation algorithm has been changed to avoid this false-classification.
[Case2]
In WDFch main scattergram, there is a possibility that the analyzer falsely classifies debris (unknown
particle shown in low-range of SFL) as eosinophil cluster. In this case, Re-differentiation algorithm
does not work correctly.
Therefore, differentiation algorithm has also been improved to avoid this false-classification.
Resolved issues
None
Remarks
For XN-10/20
ECR 318B011
DVD1XN1X Ver.00-19D build 15 Upgrade Procedure Upgrade Procedure
(SAI markets only) (FileDeliverySystem)
Changed Specifications
1 Differentiation algorithm of neutrophil and eosinocyte in case there are degranulated eosinocytes has
been changed.
* Degranulated eosinocytes could be false-classified as neutrophil.
In addition, a new setting “Display conditions of the analysis results for NEUT and EO” is added in
service settings. The setting enables to set result display setting and low-reliability mark setting.
Resolved issues
None
Remarks
For XN-10/20
ECR 317H094
DVD1XN1X Ver.00-19D build 13 Upgrade Procedure Upgrade Procedure
(North America markets only) (FileDeliverySystem)
None
Resolved issues
1 Previously, when a sample was switched, patient information was replaced by that of another patient
as follows.
This issue has now been resolved.
Previously, on the XN-9000, patient information was replaced by that of another patient when the user
performed the following operations. This issue has now been resolved.
1) The user placed sample A (Sample No. A and patient ID is A) in position 1 of a sample rack and
performed a system analysis. Sample No. was read on the BT-40.
2) After the sample rack passed through the BT-40, the user moved Sample A from position 1 to a
different position on the same sample rack. The user then placed Sample B in position 1 (Samples
were switched).
3) When samples were switched (as mentioned in item 2 above), analysis information (i.e. analysis
order and patient information) on Sample B was not registered to the host computer.
4) Sample A's patient information was linked to both samples A and B, i.e. both samples A and B had
the same patient information originally recorded for sample A.
Remarks
For XN-10/20
ECR 316G033
DVD1XN1X Ver.00-19D (North Upgrade Procedure Upgrade Procedure
America markets only) (FileDeliverySystem)
Changed Specifications
Initial version
This program is based on DVD1XN1X(00-19) and added modifications.
1 XN-9100 (CV-55) connection is now possible.
2 12L/h setting on the RU-20 is now available.
Resolved issues
1 Previously, the PDF files bookmarks for built-in IFUs could not be read properly. This has been
resolved.
2 Previously, the IPU for XN-3000 could abnormally shut down in stand-alone mode. This has been
resolved.
3 There was a case that the IPU sent an incorrect communication massage to Host.
This has been resolved.
4 Previously, the sampler analysis could not be performed by switching to sampler mode after
microtubes had been used. This has been resolved.
5 Previously, piercer adjustment screen could not be closed. This issue has been resolved.
Remarks
For XN-10/20
ECR:316D038
DVD1XN1X Ver.00-19E build 5 Upgrade Procedure Upgrade Procedure
(SCH market only) (FileDeliverySystem)
Changed Specifications
None
Resolved issues
1 Differentiation algorithm of neutrophil and eosinocyte in case there are degranulated eosinocytes has
been changed.
* Degranulated eosinocytes could be false-classified as neutrophil.
In addition, a new setting “Display conditions of the analysis results for NEUT and EO” is added in
service settings. The setting enables to set result display setting and low-reliability flag setting.
2 The countermeasure against “Fluorocell RFID tag error” which is frequently occurred when using new
RFID tag on Fluorocell bottles has been added.
Remarks
For XN-10/20
ECR 317K046
DVD1XN1X Ver.00-19E build Upgrade Procedure Upgrade Procedure
3(Chinese market only) (FileDeliverySystem)
Changed Specifications
None
Resolved issues
1 Previously, when a sample was switched, patient information was replaced by that of another patient
as follows.
This issue has now been resolved.
Previously, on the XN-9000, patient information was replaced by that of another patient when the user
performed the following operations. This issue has now been resolved.
1) The user placed sample A (Sample No. A and patient ID is A) in position 1 of a sample rack and
performed a system analysis. Sample No. was read on the BT-40.
2) After the sample rack passed through the BT-40, the user moved Sample A from position 1 to a
different position on the same sample rack. The user then placed Sample B in position 1 (Samples
were switched).
3) When samples were switched (as mentioned in item 2 above), analysis information (i.e. analysis
order and patient information) on Sample B was not registered to the host computer.
4) Sample A's patient information was linked to both samples A and B, i.e. both samples A and B had
the same patient information originally recorded for sample A.
Remarks
None
ECR 316G033
DVD7XN1J Ver.00-19E build 2 (Only Upgrade Procedure Upgrade Procedure
Chinese market) (FileDeliverySystem)
Changed Specifications
Resolved issues
1 Previously, output function for reagent information to SNCS when starting up instrument and replacing
reagent has been added, however instrument ID and reagent information were not output correctly.
This has been resolved.
Remarks
None
ECR:316C110
DVD5XN1X Ver.00-19C (Patch Upgrade Procedure Upgrade Procedure
Program) (FileDeliverySystem)
Changed Specifications
Resolved issues
1 It was observed that XN/XN-L sent incorrect data to Host Computer when special conditions were
met. This has been resolved.
Remarks
None
ECR:315G028
DVD1XN1X Ver.00-19C build2 Upgrade Procedure Upgrade Procedure
(FileDeliverySystem)
<Applicable to S/N> <Release Date>
XN-1000 SUPPLY PARTS(EU) S/N: A9203 November 2016
and thereafter
XN-1000 SUPPLY PARTS (SA-01_EU) S/N:
A9203 and thereafter
XN-1000 SUPPLY PARTS(CHN) S/N: A9203
and thereafter
XN-1000 SUPPLY PARTS (SA-01_CHN)
S/N: A9203 and thereafter
XN-2000 SUPPLY PARTS(EU) S/N: A2890
and thereafter
XN-2000 SUPPLY PARTS(CHN) S/N: A2890
and thereafter
XN-9000 SUPPLY PARTS(EXP) S/N: A2422
and thereafter
XN-9000 SUPPLY PARTS(US) S/N: A2422
and thereafter
XN-9000 SUPPLY PARTS(CHN) S/N: A2422
and thereafter
XN-9000 SUPPLY PARTS(XN-11/XN-21)
S/N: Not fixed
XN-3000 SUPPLY PARTS(EU) S/N: A2170
and thereafter
XN-3000 SUPPLY PARTS(CHN) S/N: A2170
and thereafter
XN-9100(1) SUPPLY PARTS(US) S/N:
A1026 and thereafter
XN-9100(1) SUPPLY PARTS(EU) S/N:
A1026 and thereafter
XN-9100(1) SUPPLY PARTS(CHN) S/N:
A1026 and thereafter
XN-9100(1) SUPPLY PARTS(XN-11/XN-21)
S/N: A1026 and thereafter
None
Resolved issues
1 Previously, when a sample was switched, patient information was replaced by that of another patient
as follows.
This issue has now been resolved.
Previously, on the XN-9000, patient information was replaced by that of another patient when the user
performed the following operations. This issue has now been resolved.
1) The user placed sample A (Sample No. A and patient ID is A) in position 1 of a sample rack and
performed a system analysis. Sample No. was read on the BT-40.
2) After the sample rack passed through the BT-40, the user moved Sample A from position 1 to a
different position on the same sample rack. The user then placed Sample B in position 1 (Samples
were switched).
3) When samples were switched (as mentioned in item 2 above), analysis information (i.e. analysis
order and patient information) on Sample B was not registered to the host computer.
4) Sample A's patient information was linked to both samples A and B, i.e. both samples A and B had
the same patient information originally recorded for sample A.
Remarks
ECR 316G033
DVD1XN1X Ver.00-19 Upgrade Procedure Upgrade Procedure
(FileDeliverySystem)
<Applicable to S/N> <Release Date>
S/N: Not fixed yet April 2015
Changed Specifications
1 WBC eliminating area malfunction on Malaria analysis algorithm has been revised.
2 Not reflecting target values for QC file registration has been revised.
3 Output values for WPC-GR#, WPC-MO# and WPC-LY# which are calculated from WPC ch have
been revised.
4 Brazil Portuguese wording has been revised.
5 Previously, RBC low value and HGB low value samples were regarded as blank samples. This has
been resolved.
6 Previously, unnecessary flag messages were displayed when ID read error occurred on reanalysis.
This has been resolved.
7 Previously, short sample occurred even when sample was sufficient volume. This has been resolved
8 Previously, XN with SA-01 responded that wrong sample No. Attribute was sent to host computer.
This has been resolved.
Remarks
None
ECR:313L045
DVD5XN1X Ver.00-18B (Patch Upgrade Procedure Upgrade Procedure
Program) (FileDeliverySystem)
Changed Specifications
Resolved issues
1 It was observed that XN/XN-L sent incorrect data to Host Computer when special conditions were
met. This has been resolved.
Remarks
None
ECR:315G028
DVD1XN1X Ver.00-18 Upgrade Procedure Upgrade Procedure
(FileDeliverySystem)
<Applicable to S/N> <Release Date>
S/N: Not fixed yet October 8, 2014
Changed Specifications
1 Previously, there was a possibility that the False Negative (FN) ratio for judgment of “PLT Clumps?”
on low PLT sample may be high when PLT-F ch or WDF ch are not used on CBC measurement. In
order to resolve this, the algorithm has been enhanced.
2 Research parameters and reportable parameters will now be displayed in separate screens. (North
American markets only)
When “Display research use only parameters” check box is selected in IPU setting, “Lab Only (RUO)”
tab is added to the main screen.
1 Previously, there was a possibility that the “Aged sample?” was displayed instead of “Blast/Abnormal
Lymp” even when the “Aged sample?” setting was not used. This has been resolved.
Remarks
None
ECR:314E078
DVD5XN1X Ver.00-17C (Patch Upgrade Procedure Upgrade Procedure
Program) (FileDeliverySystem)
Changed Specifications
Resolved issues
1 It was observed that XN/XN-L sent incorrect data to Host Computer when special conditions were
met. This has been resolved.
Remarks
None
ECR:315G028
DVD5XN1X Ver.00-17B (Patch Upgrade Procedure Upgrade Procedure
Program) (FileDeliverySystem)
Changed Specifications
Resolved issues
1 It was observed that XN/XN-L sent incorrect data to Host Computer when special conditions were
met. This has been resolved.
Remarks
None
ECR:315G028
DVD1XN1X Ver.00-17A Upgrade Procedure Upgrade Procedure
(FileDeliverySystem)
<Applicable to S/N> <Release Date>
S/N: Not fixed yet November 27, 2014
Changed Specifications
1 The following function for “PLT Clumps?” in version 00-17 has been enhanced in program version 00-
17A.
Previously, there was a possibility that the PLT coagulated sample was missed in CBC mode or
CBC+DIFF mode. In order to resolve this, the algorithm has been enhanced.
2 Channel for Discrete test and “PLT Clumps?” flagging judgment has been changed as follows.
CBC + DIFF + WNR + WDF
Resolved issues
Remarks
None
ECR:314J042
DVD1XN1X Ver.00-17 Upgrade Procedure Upgrade Procedure
(FileDeliverySystem)
<Applicable to S/N> <Release Date>
S/N: Not fixed yet July 28, 2014
Changed Specifications
Resolved issues
1 Previously, there was a possibility that the "Out of XX (reagent)" error occurred erroneously for DCL.
In order to resolve this situation, bubble sensor setting has been changed for Cellpack DCL
2 Previously, there was a possibility that the QC data was abnormal on radar chart even the QC data
was normal on QC Chart. This has been resolved. Program changed bubble count threshold from 6 to
12 bubbles.
3 Previously, there was a possibility that the edge of the micro tube holder was erroneously detected as
micro tube body. In order to resolve this situation, micro tube detecting sequence has been changed.
4 Previously, there was a possibility that blood aspiration sensors were adjusted incorrectly, as some
blood sample for sensor adjusting was erroneously drained. In order to resolve this situation, blood
sensor adjustment sequence has been changed. Program now discards current blood (2 uL) and
uses the next 2 uL for adjusting.
Remarks
None
ECR:313I003
DVD1XN1X Ver.00-16 Upgrade Procedure Upgrade Procedure
(FileDeliverySystem)
<Applicable to S/N> <Release Date>
S/N: Not fixed yet March 17, 2014
Changed Specifications
1 Analysis algorithm for HPC Mode has been improved for Europe and China specifications. (This
modification was applied for North America market with ver] 00-15.)
2 Switching from WBC-N to WBC-D has been applied when specific WBC aggregation occurred.
3 Action message "Sample Mixing Failure" for the sample insufficient mixing has been added.
4 Display and output in Portuguese has been added to Europe and North America specifications.
5 "PRBC" flag has been changed to "pRBC."
6 Not using Folder sharing setting has been added for the method of image data output in ASTM
format. (Service Setting)
7 PNG file format setting has been added for the method of image data output in DPS format. (Service
Setting)
8 Sub scattergram has been added when outputing in CSV output.
9 Low reliability mark will not be attached to the sample with the sample with low TNC value.
10 Built-in IFU has been applied to AP specifications.
Resolved issues
1 Previously, SP-10 built-in IFU was not functioned properly. This has been resolved.
2 Previously, scattergrams and histogram could not viewed in host computer when the data included
"\","/", ":", "<", ">" and "|" in ASTM format. This has been resolved.
3 Previously, background check for BF mode or hsA mode was performed when Autorinse was
performed in sampler analysis. This has been resolved.
4 Previously, there was a possibility that blood sample splattered when removing blood sample from
back side of test tube cap. This has been resolved by not performing this process when open mode.
5 Previously, there was a possibility that WC2 was overflown on PD mode. This has been resolved.
6 Previsouly, there was a possibility that remaining lyse solution could not detected on RR-10. This has
been resolved.
7 Previously, there was a possibility that blood sample was aspirated to rinse cup when rinsing the
piercer before analyzing RET-ch and PLT-F ch. As a result, RBC-O and PLT-O became lower and
calibration could not be performed.
8 Previously, incorrect data was transmitted when outputing histogram data in DPS format. This has
been resolved.
9 Previously, there was a possibility that output marks for host computer did not disappear when
outputting to host comuter. This has been resolved.
Remarks
None
ECR:313G047
DVD1XN1X Ver.00-15 Upgrade Procedure Upgrade Procedure
(FileDeliverySystem)
<Applicable to S/N> <Release Date>
S/N: Not fixed yet November 5, 2013
Changed Specifications
Resolved issues
1 There was a possibility that WBC-N value might be falsely low. This problem has been corrected.
2 There was a possibility that the sample aspiration sensor could not detect short sample error and the
result was falsely low. To resolve this situation, the judgment condition has been changed.
3 There was a possibility that the IPU program could not start if setting file was corrupt. This problem
has been corrected.
4 The function to highlight the analysis mode field in the explorer screen when "Analysis result is high"
error occurs was not implemented. This is now being implemented.
5 There was a possibility that text including apostrophe (') could not be properly registered on IPU
program. This problem has been corrected.
6 Previously, "WBC Abn Scattergram" flagging for BASO# did not work in HPC mode. To resolve this
situation, this function has been implemented.
Remarks
None
ECR:313B037
DVD1XN1X Ver.00-14 Upgrade Procedure Upgrade Procedure
(FileDeliverySystem)
<Applicable to S/N> <Release Date>
S/N: Not fixed yet. April 18, 2013
Changed Specifications
1 Discrimination performance for NRBC high value sample on WDF channel has been improved.
2 Discrimination for Neutrophil and Eosinophil has been improved.
3 Service setting for removing "PLT Abnormal Distribution" in PLT-F analysis has been added.
4 Service setting for changing PLT judgment criteria of Action message and PLT switching has been
added.
5 Service setting to add SNCS file for X-barM QC plot has been added. (When performing "saved as
SNCS format" or when Auto Output is set, the file is saved as
"XbarM_[AnalyzerID][PlotDateTime]_[Channel].txt under the location of "C:\Program
Files\Sysmex\SNCS\SendFolder\XbarM\".)
6 When discrimination error occurred for WBC scattergram, Service setting for masking related
parameter has been added.
7 Reagent remaining volume indicator for DCL has been applied with the countermeasure against the
HGB unexpected high value.
8 In order to prevent from fluctuation of dye solution volume caused by air bubbles on DP, Setting
sequence has been improved.
Remarks
None
ECR: 312J024
DVD1XN1X Ver.00-13 Upgrade Procedure Upgrade Procedure
(FileDeliverySystem)
<Applicable to S/N> <Release Date>
S/N: Not fixed yet. February 18, 2013
Changed Specifications
1 The color of Abnormal Lympho area on WPC scattergram has been changed from red to gray when
"Abn Lympho?" is not flagged.
2 Malaria detection function has been added.
3 hsA Mode (High Sensitivity Analysis) has been added.
4 Data can be stored as FCS format (Flow Cytometry Data File Format Standards).
5 Positive/Negative judgment level criteria change for PLT Clumps is available by channels by service
setting.
6 When there are specification differences are between left main unit and right main unit on XN-2000 or
XN-3000, the retest judgment and PLT-I/O switching judgment will be performed to be equal between
left and right main units.
7 MCHC criteria for "Turbidity/HGB Interference?" has been added to service setting.
8 "*****000 10^6/L" has been added to RBC-BF unit setting.
9 On SNCS remote maintenance output function, outputting "reagent electric conductivity calibration"
has been added.
10 Initial value for "NRBC low count correction" has been changed to "0: Do Not Correct."
11 MicroR >= 10% of PLT low reliability judgment has been changed to MicroR >= 5%.
<NOTE>As for item 2 and 3 above, those functions become available when license registration is
completed.
Remarks
None
ECR: 312I007
DVD1XN1X Ver.00-12 Upgrade Procedure Upgrade Procedure
(FileDeliverySystem)
<Applicable to S/N> <Release Date>
January 18, 2013
XN-1000 SUPPLY PARTS (EU) S/N: A1280 and thereafter
XN-1000 SUPPLY PARTS (AP) S/N: A1102 and thereafter
XN-1000 SUPPLY PARTS (CHN) S/N: A1321 and thereafter
XN-2000 SUPPLY PARTS (EU) S/N: A1226 and thereafter
XN-2000 SUPPLY PARTS (AP) S/N: A1028 and thereafter
XN-2000 SUPPLY PARTS (CHN) S/N: A1121 and thereafter
XN-3000 SUPPLY PARTS (AP) S/N:
XN-3000 SUPPLY PARTS (CHN) S/N:
A1070 and thereafter
XN-3000 SUPPLY PARTS (EU) S/N:
XN-3000 SUPPLY PARTS (US) S/N:
XN-9000 SUPPLY PARTS (CHN) S/N:
XN-9000 SUPPLY PARTS (EXP) S/N: A1116 and thereafter
XN-9000 SUPPLY PARTS (US) S/N:
Changed Specifications
1 Flagging for "Blasts/Abn Lymph?" has been improved.
2 Discrimination performance for WBC/NRBC on WNR ch has been improved.
Remarks
None
ECR: 312I006
DVD1XN1X Ver.00-11 Upgrade Procedure Upgrade Procedure
(FileDeliverySystem)
<Applicable to S/N> <Release Date>
September 24, 2012
XN-1000 SUPPLY PARTS (EU) S/N: A1280 and thereafter
XN-1000 SUPPLY PARTS (AP) S/N: A1062 and thereafter
XN-1000 SUPPLY PARTS (CHN) S/N: A1101 and thereafter
XN-2000 SUPPLY PARTS (EU) S/N: A1207 and thereafter
XN-2000 SUPPLY PARTS (AP) S/N: A1028 and thereafter
XN-2000 SUPPLY PARTS (CHN) S/N: A1051 and thereafter
XN-3000 SUPPLY PARTS (AP) S/N:
XN-3000 SUPPLY PARTS (CHN) S/N:
A1070 and thereafter
XN-3000 SUPPLY PARTS (EU) S/N:
XN-3000 SUPPLY PARTS (US) S/N:
XN-9000 SUPPLY PARTS (CHN) S/N:
XN-9000 SUPPLY PARTS (EXP) S/N: A1116 and thereafter
XN-9000 SUPPLY PARTS (US) S/N:
Changed Specifications
1 Discrimination for HFLC particles has been improved when particles appear in high florescent area on
WDF channel.
2 Discrimination for "Atypical Lymph?" on WDF channel has been improved.
3 In order to improve reliability for PLT-F, a rule for "PLT Abn Scattergram" has been added.
4 Previously, there was a possibility that BASO# became falsely high by invading lymphocyte particles
into BASO area when analyzing CLL or DLBCL samples. In order to resolve this situation, "WBC Abn
Scattergram" will be notified in such a case.
5 The program now supports XN-3000 Standalone mode. (In XN-3000 Standalone mode, only ASTM
format will be supported.)
6 A research parameter "PLT-F2" has been added. (A parameter which has one more digit than PLT-F
and has high reproducibility with low values.)
7 Previously, when Rerun and Reflex are judged at the same time, only Rerun was performed. In order
to resolve this situation, the priority of the rule action has been modified.
8 Previously, there was a possibility that Hgb value was incorrect when analyzing high bilirubin sample
due to SLS capability. In order to resolve this situation, a research parameter "HGB-O" analyzing with
FCM has been added.
9 BCR5342H (NEC made) barcode reader is now supported. (Only for software. Hardware modification
will be available.)
10 An Analysis Stop Setting, when an order inquiry was set for a parameter which would be analyzed on
defective analyzer only, has been added.
11 Some instrument configurations have been added so that more combinations with B1, B3 and B4 can
be connected to IPU. (EU market only)
12 A service setting has been added to enable connecting with CELLPACK DCL temporarily when
installing analyzer or RU-20 is defective.
13 For printing list, a white color background screen can now be selected.
Remarks
None
ECR: 312G038
DVD1XN1X Ver.00-10 Upgrade Procedure
<Applicable to S/N> <Release Date>
XN-1000 SUPPLY PARTS (EU) S/N: June 23, 2012
Not fixed yet.
XN-1000 SUPPLY PARTS (AP) S/N:
Not fixed yet.
XN-1000 SUPPLY PARTS (CHN) S/N:
Not fixed yet.
XN-2000 SUPPLY PARTS (EU) S/N:
Not fixed yet.
XN-2000 SUPPLY PARTS (AP) S/N:
Not fixed yet.
XN-2000 SUPPLY PARTS (CHN) S/N:
Not fixed yet.
Changed Specifications
1 When low specific gravity eosinophil appeared on WDF ch, position of neutrophil cluster and
eosinophil cluster were very close. As a result, eosinophil was discriminated as neutrophil. This
discrimination has been improved.
2 There was a possibility that reticulocyte count was affected by appearing RBC with Howell-Jolly body.
In order to resolve this, rules to judge as "RET Abnormal Scattergram.".
3 Changing PLT Clumps rule as follows
- To improve flag detection, CBC rule for PLT Clumps was removed when WDF channel or PLT
channel was used.
- Previously, PLT Clumps was judged by PLT-F rule when PLT-F channel and WDF channel were
used. To improve flag detection for PLT aggregation, WDF channel rule has been added.
4 The discrimination between WBC cluster and NRBC cluster when lots of tumorous lymphocytes were
observed like CLL sample has been improved.
5 Instrument's configuration has been changed so that additional instruments can connect to IPU.
6 Up to three analyzers can now be connected to RU-20.
7 Display digit for HGB on XN-21 or XN-11 can be selected with SystemConfigurator as follows.
- Current display digit for HGB: ***.* [g/dL] or **** [g/L]
- Added display digit for HGB: ***.** [g/dL] or **** .*[g/L]
8 IFU for XN-21/XN-11 will now be displayed as a built-in manual on XN-21/XN-11.
9 BL-180/185 (KEYENCE made) barcode reader is now supported. (Only for software. Hardware
modification will be available.)
10 Scattergram and histogram research parameters can now be output in ASTM Format.
11 Registered user number has been changed from 20 to 100.
12 Specifications for HPC Mode have been changed as follows.
- HPC Mode results will not be used for delta check.
- PLT has been added to numeric value and graph on HPC tab.
- HPC# will now be displayed when analysis mode is "Control" even not in QC mode.
13 The status of Feed-in arm left position sensor will now be displayed on Sampler status screen. (The
SNS213 sensor status will be displayed as No.13 on sensor status screen.
14 The Mixing method of PLT-F channel sample has been streamlined as a countermeasure against
PLT-F high value on QC.
15 The following sequences have been changed.
- Previously, piercer needle pierced into bottom of sample tube. As a result, sample tubes came up
when piercer ascends. In order to resolve this, on open mode, piercer will ascend with closing tube
clamp.
- Previously, Syringe was not in limit position after completing Shipping 1. In order to resolve this,
syringe will return to its limit position when shipping sequences 1, 2 or 3 are completed.
Remarks
None
ECR: 312D066
DVD1XN1X Ver.00-09 Upgrade Procedure
<Applicable to S/N> <Release Date>
XN-1000 SUPPLY PARTS (EU) S/N: June 11, 2012
A1177 and thereafter
XN-1000 SUPPLY PARTS (AP) S/N:
A1052 and thereafter
XN-1000 SUPPLY PARTS (CHN) S/N:
A1001 and thereafter
XN-2000 SUPPLY PARTS (EU) S/N:
A1158 and thereafter
XN-2000 SUPPLY PARTS (AP) S/N:
A1019 and thereafter
XN-2000 SUPPLY PARTS (CHN) S/N:
A1001 and thereafter
Changed Specifications
1 The judgment rule of PLT Channel Error has been changed to eliminate poor detection of PLT
channel abnormality.
2 A border line has been added between two particles areas at PLT-Fch.
3 WBC classification accuracy of sample that shows high NRBC value has been improved.
4 A compensation function for NRBC false high value has been added when NRBC shows exceedingly
low value.
5 Some instrument configurations have been added so that more combination with B3 and B4 can be
connected to IPU.
6 The specification has been changed so that PLT-I is applied as PLT regardless of switching condition
on Analyzer A2/B3 that operate PLT-I/O switching.
7 A service setting has been added so that parameter unit settings on IPU are matched with the ASTM
format.
8 A function has been added so that only a specific barcode can be read in case multiple barcode labels
are attached on a sample tube. (Unset barcode symbol in the IPU will be ignored.)
9 SysmexInsight function has been implemented. (Only for U.S. markets)
10 Air bubble draining in Whole Blood Aspiration syringe during priming sequence has been improved.
11 Sent pulse at fine adjustment of optic axis has been reduced by changing adjustment time from 29
sec to 24 sec.
12 SD bar has been displayed on Whole Blood parameters of browser main screen of HPC mode
analysis data.
13 The color of patelite status has been standardized as follows. Green: Normal status Red: Abnormal status.
14 The specifications of LED lighting of XN-1000 (SA-01) have been standardized with XN-1000. XN-
2000. XN-3000 and XN-9000.
15 The usability of GP customize function has been changed to match to the general operation method
of draw tools.
16 Error monitoring timing and detailed information of sheath motor error have been added.
17 The condition of [Cleaning is required.] and [Cleaning is required (Warning).] has been changed.
18 The program has been changed so that the compressor will be stopped regardless of error
occurrence when analyzer has not been operated for a specified time.
19 Adobe plug-in has been removed from the IPU PC.
20 Progress status information will now be displayed during program upgrading process.
21 Non-translated messages have been translated into multiple languages.
22 A message will now be displayed in case of the parameter DB loading is failed during IPU start up.
23 Residual pulses at initialization of WB Asp. Motor, RBC sheath motor and FCM sheath motor will now
be stored.
Remarks
None
ECR: 311L057
DVD1XN1X Ver.00-08 Upgrade Procedure
<Applicable to S/N> <Release Date>
XN-1000 SUPPLY PARTS (EU) S/N: February 15, 2012
A1165 and thereafter
XN-1000 SUPPLY PARTS (AP) S/N:
A1052 and thereafter
XN-1000 SUPPLY PARTS (CHN) S/N:
A1001 and thereafter
XN-2000 SUPPLY PARTS (EU) S/N:
A1156 and thereafter
XN-2000 SUPPLY PARTS (AP) S/N:
A1019 and thereafter
XN-2000 SUPPLY PARTS (CHN) S/N:
A1001 and thereafter
Added Functions
1 The program now supports CELLCLEAN AUTO.
2 The program now supports HPC mode.
3 The program now supports XN-21/XN-11 models.
4 The program now supports SA-01 sampler.
5 The program now supports output for RU-20 SNCS as follows.
-Adding shutdown report
-Adding event history
-Adding error report
6 Action message "Difference between PLT and PLT-F. Check the results." has been added.
7 Revised some messages in German program.
8 The program now supports color sensor.
Changed Specifications
1 Rule for the "PLT channel error" has been changed.
2 The algorithm for IPF/WBC discrimination on PLT/F channel has been added.
3 "PLT Clumps?" rules have been changed. (Added for American market Only. Already available in
other markets by earlier version.)
4 Ghost discrimination algorithm for PLT-F channel has been changed. (Added for American market
Only. Already available in other markets by earlier version.)
5 The algorithm for improving NEUT/EO discrimination on WDF channel has been added. Added for
American market Only. Already available in other markets by earlier version.)
6 Shutdown and cleaning are now recorded individually on maintenance record.
7 To prevent rinse cup drain line clogging, improved cleaning after measurement and before entering
sleep mode status have been implemented.
8 Error detection method for "FCM sheath motor error" and "RBC sheath motor error" has been
changed by checking remaining pulse number.
9 Explorer now will no longer store dummy results when detecting a sample without blood sample
detect on sampler L-J QC.
10 Cumulative tab setting on the Data Browser Screen has been changed to show all data or last data.
11 Setting for monitor leakage sensor has been added as follows.
- Adding not monitoring sensor
- If a "Water leak" error appears as yellow in the status (warning) screen in service mode, the error
continues and the error will be cleared once the error is recovered. If the error appears as red in
status screen, the error can only be cleared by turning off the instrument to resolve.
12 Combination of XN modules connected to one IPU has been expanded.
13 Conditions for stopping alarm have been changed if error occurs while logging ON. To stop alarm, IPU
screen can be touched using touch panel function or clicking with mouse.
Remarks
None
ECR: 311J096
DVD1XN1X Ver.00-07 Upgrade Procedure
<Applicable to S/N> <Release Date>
XN-1000 SUPPLY PARTS (EU) S/N: December 8, 2011
A1165 and thereafter
XN-1000 SUPPLY PARTS (AP) S/N:
A1052 and thereafter
XN-1000 SUPPLY PARTS (CHN) S/N:
A1001 and thereafter
XN-2000 SUPPLY PARTS (EU) S/N:
A1156 and thereafter
XN-2000 SUPPLY PARTS (AP) S/N:
A1019 and thereafter
XN-2000 SUPPLY PARTS (CHN) S/N:
A1001 and thereafter
Added Functions
1 A copy / paste function has been added, on a record basis, for Rerun/Reflex/Comment Rule,
Validation Rule and Output Rule on the Rule screen.
2 The program now supports traditional Chinese and Korean languages.
3 Backup and restore tools for the IPU have now been loaded.
4 Service settings have been added to enable a sample number to be modified when a barcode is read
as follows:
- Suppress/Padding setting for specified digits
- Fetching setting for specified digits
5 An open/close button has been added on the top right corner of Rule screen to allow users to hide a
flow chart (Rule Execution Order). Displaying of the Rule Execution order chart will be toggled with a
click of the button.
6 The program now supports Windows 7 SP1.
7 When outputting in CSV format, users can now select outputting scattergrams in white background.
8 A service setting has been added to allow users to select the letter size for printing.
9 The program now supports the function of transmitting CELLPACK DST reagent information to WAM
IF.
10 The specifications have been changed to display a progress dialogue box when "GP Customize" icon
is clicked in the Menu screen.
11 Previously, if the IPU computer is switched OFF abruptly, the layout setting files for IPU screens may
be corrupted, and if these files are read, there was a possibility that the IPU could not be started. In
order to resolve this, the program has been changed to show the initial screen if the layout setting file
is corrupted.
Changed Specifications
1 To make more user friendly, the layout for the Instructions For Use screen has been changed.
2 The functions for RU icons have been changed to make more user friendly as follows:
- The reagent replacement screen is now displayed when clicking the Reagent icon.
- The reagent replacing icon is now displayed while replacing reagent on the RU.
3 To prompt users to enter a reagent code, a caption has been added to the screen displayed when
reagent is replaced on the RU.
4 As there is a time lag between when the Execute button is pressed in the Reagent Replacement
screen and when reagent replacement is started, users are unable to tell whether reagent
replacement is in progress or not. To resolve this, the message of reagent replacement in progress
and its display method have been changed.
5 To make more user friendly, the following changes have been made to the manual analysis screen:
Previously, when the OK button is pressed even once in the manual analysis screen, it was
impossible to return to the barcode reading mode. To resolve this, a barcode read setting checkbox
has been added to allow users to return to the barcode reading mode.
The checked status of checkboxes in the manual analysis screen will remain unchanged even after
the IPU program is closed.
To encourage users to remove caps from the sample tubes before performing an analysis, the icon for
the manual analysis button in the control menu has been changed to the image of an uncapped
sample tube, flashing in white color.
6 If set to not allow users to edit rules in Service Mode Settings, [Register Rule], [Modify Rule] and
[Delete Rule] icons in the Rule screen will be hidden, not disabled.
7 The following errors have been deleted from the Repeat Rule screen, as sampler analysis would be
stopped if any of the errors occurs.
"Failed to read sample number during manual analysis" error has been deleted from the Repeat Rule
screen.
8 Color of the status indicator LED when the analyzer is being restarted has been changed from green
to red.
9 The rule for "PLT Clumps?" flag has been changed. (Except for the North America region.)
10 The order of IP messages for Data Printer (DP) printing format settings in IPU Setting have been
modified to match the order in the analysis data list window.
11 Chinese screen display has been changed based on requests from local offices.
12 An icon showing required sample volume or cap open/close has been added to the manual analysis
screen for the Body Fluid mode.
13 Previously, errors that do not occur due to the configuration of the instrument were displayed in gray
in Repeat Rule tab in the Rule screen. The specification has now been changed to hide those errors.
14 Previously, when one reagent was shared by multiple instruments (XN-2000, XN-3000 or XN-9000), if
one of the instruments was in the sleep status and reagent replacement was to be performed in
another instrument, reagent was not replaced until the sleep status was canceled. It has been
changed to automatically cancel the sleep status for all instruments to replace reagent.
15 To make more user friendly, the positions of display switching buttons are now stored in the IPU and
will be restored when the IPU is restarted.
16 The specifications for WAM IF material records have been changed as follows:
- "The starting date of using reagent" has been changed to "The starting date and time of using
reagent".
- "The opened date of reagent" has been changed to "Installed date of reagent".
17 Error messages for when Fluorocell reagents have run out have been changed as follows:
- "Replace Fluorocell" message is deleted .
- "Out of Fluorocell" message is now displayed when replacement is needed.
18 In XN-2000 and XN-3000, when the analysis module on the right-hand side was switched OFF, the
tube grabber came in contact with the sample tube detection unit of the sampler module. In order to
resolve this, the sequence program has been changed to not lower the tube grabber to the lowest
position when the instrument is being shut down; by forwarding the sample set unit a little to hook the
tube grabber.
19 The sequence program has been changed to prevent syringes from getting stuck, by increasing motor
driving current.
20 The sequence program has been changed to prevent false tube detection errors at home position by
chattering of signal.
Remarks
None
ECR: 311E068
DVD1XN1X Ver.00-06 Upgrade Procedure
<Applicable to S/N> <Release Date>
XN-1000 SUPPLY PARTS (US) S/N: September 1, 2011
A1007 and thereafter
XN-1000 SUPPLY PARTS (EU) S/N:
A1076 and thereafter
XN-1000 SUPPLY PARTS (AP) S/N:
A1052 and thereafter
XN-1000 SUPPLY PARTS (CHN) S/N:
A1001 and thereafter
XN-2000 SUPPLY PARTS (US) S/N:
A1002 and thereafter
XN-2000 SUPPLY PARTS (EU) S/N:
A1082 and thereafter
XN-2000 SUPPLY PARTS (AP) S/N:
A1019 and thereafter
XN-2000 SUPPLY PARTS (CHN) S/N:
A1001 and thereafter
Changed Specifications
1 Calendar Control is now supported using a touch panel.
2 "NRBC-X" and "NRBC-Y" have been added to the browser screen service tabs.
3 Algorithm for WDF channel NEUT/EO discretion has been improved. (Exclude for American Markets)
Remarks
None
ECR: 311E067
DVD1XN1X Ver.00-05 Upgrade Procedure Upgrade Procedure (NA)
<Applicable to S/N> <Release Date>
XN-1000 SUPPLY PARTS (US) S/N: August 8, 2011
A1007 and thereafter
XN-1000 SUPPLY PARTS (EU) S/N:
A1063 and thereafter
XN-1000 SUPPLY PARTS (AP) S/N:
A1052 and thereafter
XN-1000 SUPPLY PARTS (CHN) S/N:
A1001 and thereafter
XN-2000 SUPPLY PARTS (US) S/N:
A1002 and thereafter
XN-2000 SUPPLY PARTS (EU) S/N:
A1029 and thereafter
XN-2000 SUPPLY PARTS (AP) S/N:
A1019 and thereafter
XN-2000 SUPPLY PARTS (CHN) S/N:
A1001 and thereafter
Changed Specifications
1 Reagent replacement function for Argentina has been added.
2 XN-3000 has been added to SystemConfig tool.
3 Conforms to PeteLite.
4 Zero padding/zero suppress function has been added when reading barcode label.
5 Inverse rotation function for barcode reading (Retrying barcode reading with inverse rotation on
barcode read error.) has been added. (This function can be set by service setting.)
6 Nicknames for XN-9000 analyzers have been set as "XN-9000-1", "XN-9000-2", "XN-9000-3" from
the right to left order.
7 Text in screens for Italian and Portuguese have been revised.
8 Open mode is now set as default mode when switching to BF mode. (Non-open mode is the default
mode when switching to Whole blood mode or Low WBC mode.)
9 Required sample volume is now displayed on manual analysis screen for BF mode.
10 "uL" has been added to WBC-BF unit setting.
11 The plot data of X-BarM is now included in the SNCS output.
12 The error reports are no longer output to SNCS when logging on with Sysmex user.
13 Preparation process is now performed when switching from non-PD mode to PD mode.
14 Initial focus (initial mouse pointer position) in now positioned out of edit box.
15 All characters in the edit box are now selected when edit box for numbers are selected when using
software keyboard.
16 Shortcut key assigns which are not supported by IPU program (Ctrl+W, Ctrl+Alt+ arrow key,
Ctrl*shift+B, Ctrl+Shift+C) are now invalid.
17 Some screen designs (scroll bar, radio button, check box or tab) have been changed.
18 All targets/limits are now loaded regardless of analyzer type when reading QC assay value.
19 Backup and restore of patient information can now be performed after shutting down the analyzer.
20 Encodes and languages on CSV output are now followed by OS locale.
21 Supply related error messages for Reservoir Tank have been deleted.
22 WPC related error messages (WPC sampling error, WPC channel error) that appear on Repeat rule
screen are now displayed under the RET and PLT-F related error messages.
23 Particles in SSC-FL and SSC-FSC areas on WPC channel scattergram are now displayed in red
color.
24 Ghost discrimination algorithm for PLT-F channel has been changed. (excludes North American
markets)
25 In order to improve barcode reading performance, number of matching scans on barcode read has
been changed to two times.
26 Some parts of the "PLT Clumps?" rule have been changed. (excludes North American markets)
27 Condition of PLT channel error has been changed.
Remarks
None
ECR: 311E066
DVD1XN1X Ver.00-04
<Applicable to S/N> <Release Date>
S/N: Not fixed yet.
Changed Specifications
Remarks
This is the initial version released to the markets.