Hitachi 912 HbA1c

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

Enzymatic HbA1c Assay Kit

Hitachi 912 parameters


Analysis
Select Test: [A1c] Analyzer Cycle Time: 10 sec Diluent Water [0]
Test name: [A1c] Assay/Time/Point : [2 Point End] [10] [ 15] [25] [0] [0]
Application Code [*] Wave(2nd/Primary) [800][700]
Sample Volume Class 1 Class 2
Normal [25] [0.0] [0] [25] [0.0] [0]
Decrease [25] [0.0] [0] [25] [0.0] [0]
Increase [25] [0.0] [0] [25] [0.0] [0]
Reagent(R1) T1 [112] [0] [*] [28]
Reagent(R2) T2 [48] [0] [*] [28]
Reagent(R3) T3 [70] [0] [*] [28]
Reagent(R4) T4 [0] [0] [*] [0]
Class 1 Class 2 Twin test [ ]
Absorbance Limit [32000] [32000] Increase
Prozone Limit [-32000] [32000] [Lower]
Calibration
Calib Type [Linear] [ ]
point [2] Span Point [2]
Weight [0]
Automatic Calibration
Time out Change Over
Blank [0]hr Lot [] SD Limit [999.9]
Span [0]hr Bottle [] Duplicate Limit [10] % [32000] ABS
2 Point [0]hr Sensitivity Range [-99999] - [999999]
Full [0]hr S1 ABS Range [-32000] - [32000]
Range
Application Code [*] Unit [%]
Report name [A1c] Data Mode [On Board]
Control Interval [0]
Ins.Factor(Y=aX+b) a=[1.0] b=[0.0]
Sample Type [Whole blood] Repeat range [-99999] [999999]
Class 1 Technical range [4] [16] Class 2 Technical range [4] [ 16]
Others
Standard (1) (2) (3) (4) (5) (6)
Calib.Code [*] [*] [0] [0] [0] [0]
Concentration [*.*] [*.*] [0] [0] [0] [0]
Position [*.*] [*.*] [0] [0] [0] [0]
Sample Volume [25] [25] [0] [0] [0] [0]
Diluted S.Volume [0] [0] [0] [0] [0] [0]
Diluent Volume [0] [0] [0] [0] [0] [0]