ID1 Plus / ID1 Plus-A weighing terminals Contents
Contents Page
1 Checklists ......................................................................................................................................... 2
1.1 Maintenance checklist ......................................................................................................................... 2
1.2 Service checklist ................................................................................................................................. 3
4 Troubleshooting ................................................................................................................................ 9
6 Settings ............................................................................................................................................ 14
6.1 Configuring terminal and weighing platform ........................................................................................... 14
6.2 Preparing the identcard (ID1 Plus only) ................................................................................................. 14
1 Checklists
• Check that transport locks of the weighing platform have been released properly and are in the correct position.
• Check supports of the weighing platform.
• Check that the load plate is free and has no contact with the surroundings.
• Check maximum capacity and linearity of the weighing platform.
• Check play of all stops and limits.
• Check that the lever system is free on all sides.
• Check knife edges, bearings and flexible bearings.
2 Spare parts
Cable connections
for example cable A to connector A
Item Designation Part No. ID1 Plus Part No. ID1 Plus-A
1 Cover, compl., with display window and keypad 00 505 326 00 505 326
5 Power cable
- Euro 00 504 247 00 504 247
- GB 00 504 248 00 504 248
- DK 00 504 249 00 504 249
- CH 00 504 250 00 504 250
- USA 00 504 251 00 504 251
- ZA 00 504 252 00 504 252
- AUS, RA 00 504 253 00 504 253
Item Designation Part No. ID1 Plus Part No. ID1 Plus-A
27 Verification set (for updating Identcard bracket) 22 000 386 22 000 386
3 Error displays
ERR ID1 Error terminal Check all plug-in connec- EPROM main board
tions and EPROM for firm ID1 Plus
seating.
ERR I-F Error interface Check attached peripherals EPROM main board
and connections. ID1 Plus
Check EPROM.
ERR BRU Error weighing platform Check connection cable of Measuring cell
measuring cell.
If error is redisplayed: Replace faulty component, see section 5, Corrective maintenance or service manual of attached
weighing platform.
4 Troubleshooting
Terminal can not be switched on • Power cable not plugged in • Establish connection to power
supply
• No line voltage • Check power cable
• Fuse blown • Check microfuses on main board
and replace if necessary (see 5.4)
• Display switched off • Switch on display
• Main board faulty • Change main board (see 5.6)
After terminal switched on, • Weighing platform not attached to • Attach weighing platform
- PLUG IN - appears terminal
• Connection cable of weighing • Replace connection cable
platform faulty
• Fuse blown • Check microfuses on main board
and replace if necessary (see 5.4)
• Measuring cell faulty • Replace measuring cell
(see service manual of attached
weighing platform)
• Ribbon cable for weighing platform • Plug in weighing platform cable
connection on main board (item H
in exploded view diagram) not
attached
• Analog board faulty • Change analog board (see 5.9)
• Point cable faulty • Change point cable
After terminal switched on, no • Transport safeguards of weighing • Release transport safeguards
weight display platform not released (see operating instructions of
attached weighing platform)
• Mechanical system touching • Adjust frame and lever system (see
somewhere service manual of attached weighing
platform)
• Main board faulty • Change main board (see 5.6)
• Analog board faulty • Change analog board (see 5.9)
• Point cable faulty • Change point cable
No data transfer to measuring cell • Fuse blown • Check microfuses on main board
and replace if necessary (see 5.4)
• Main board faulty • Change main board (see 5.6)
• Measuring cell faulty • Replace measuring cell
(see service manual of attached
weighing platform)
• Analog board faulty • Change analog board (see 5.9)
• Point cable faulty • Change point cable
No entry via keypad possible • Keypad faulty • Replace keypad (see 5.8)
• Main board faulty • Change main board (see 5.6)
• Analog board faulty • Change analog board (see 5.9)
• Point cable faulty • Change point cable
Note
See also operating instructions of ID1 Plus or ID1 Plus-A, section "What if...?".
5 Corrective maintenance
Caution
Before opening the instrument:
• Disconnect power plug.
• Unplug attached weighing platform and peripherals.
Fuse ratings
SI2 SI 1 slow blow 2.5 A - H 250 V
SI 2 slow-blow 1.25 A - L 250 V
SI3 SI 3 slow-blow 315 mA - L 250 V
SI 5 slow-blow 315 mA - L 250 V
SI5
SI1
• Reinstall main board, see section 5.10.
Caution
The battery must not be changed as improper handling can lead to an explosion hazard.
Disposal of the board with battery only by the manufacturer!
5.7 Checking and changing the transformer, setting the power supply voltage
Checking the transformer
• Change main board, see section 5.3.
• Check voltages on secondary side.
S2 S10
S1, S2 0.95 A 12 V ± 5 % 13.8 V
V V S4, S5 0.225 A 3.2 V ± 5 % 4.0 V
0 5
5
2
1
1
S5, S6 0.225 A 3.2 V ± 5 % 4.0 V
7 6 5 4 3 2 1 S7, S9 0.13 A 32 V ± 5 % 36.5 V
S8, S10 0.17 A 10 V ± 5 % 11.6 V
2 1
ST5
Pin assignment
ST 3 L1 (brown)
ST 4 N (blue)
ST 5
ST 6
ST 7
} transformer
ST 9 shield bonding
6 Settings
All other information on the ID1 Plus-A terminal is contained in the operating instructions for the ID1 Plus-A.
For the ID1 Plus terminal, the service mode settings are described in the service manual for the service mode of the connected
weighing platform.
Service mode operation
Program blocks are selected the same way in service mode
= YES = NO
as in master mode.
identcode.
p
p p
p
p p
p p
p p
p p
p p
pp
p
p pp
p
pp
p pp
ppp p
• Assemble identcard.
2 • With certified scales, seal screw (3).
7 Software compatibility
The ID1 Plus weighing terminal is compatible with all weighing platforms of the F, K, N and D series.
With the ID1 Plus-A terminal: see ID1 Plus-A operating instructions, section "Technical data".
8 Pin assignment
D H
D RXD+, receiving loop of the weighing platform
F
F RXD–, receiving loop of the weighing platform
E G
G Protective ground