CALIBRATION-REPORT-TEMPLATE-FOR-GENERAL-RADIOGRAPHY-STATIONARY (1)
CALIBRATION-REPORT-TEMPLATE-FOR-GENERAL-RADIOGRAPHY-STATIONARY (1)
CALIBRATION-REPORT-TEMPLATE-FOR-GENERAL-RADIOGRAPHY-STATIONARY (1)
• Page 1 of 5 •
MECHANICAL
PARAMETERS TRIAL 1 TRIAL 2 TRIAL 3 MEAN
CHECKS
a) Misalignment in side A (cm,
%)
b) Misalignment in side B (cm,
%)
1. COLLIMATION c) Misalignment in side C (cm,
ASSESSMENT %)
(Provide images d) Misalignment in side D (cm,
of the test after %)
calibration e) Total misalignment (cm, %)
report) f) The x – ray beam
perpendicularity (degrees)
<<State the source image receptor distance (cm), x
g) Settings – ray field size (cm2), kVp, mA, and s settings
here>>
Misalignment in side B
Misalignment in side C
• Page 2 of 5 •
Misalignment in side D
SETTINGS: <<State the source image receptor distance (cm), x – ray field size (cm2), kVp, mA, and s settings
here>>
• Page 3 of 5 •
<<INSERT PICTURES <<INSERT PICTURES <<INSERT PICTURES
HERE>> HERE>> HERE>>
SETTINGS: <<State the source image receptor distance (cm), x – ray field size (cm2), kVp, mA, and s settings
here>>
• Page 4 of 5 •
alignment
Generator kVp Accuracy
● For < 100 kVp: actual kVp shall be within ± 6 % of the set kVp
checks
● For > 100 kVp: actual kVp shall be within ± 6 % of the set kVp
Exposure time
● For exposure time < 0.1 s, actual time shall be within ± 20 % of the set time
accuracy
● For exposure time > 0.1 s, actual time shall be within ± 10 % of the set time
• Page 5 of 5 •