Professional Documents
Culture Documents
278065 Astral Service Manual
278065 Astral Service Manual
Astral series
SERVICE
Astral™series
Life support ventilators
ENG
Contents 3
3 Test and calibration procedures .......................................................................... 31
Equipment required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .31
Service software kit - #19688 31
PC and software 31
Measurement tools 31
Test set-up equipment 32
Accessories 32
Connecting the Astral to the service software . . . . . . . . . . . . . . . .32
Test procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .33
3.1 Cleaning the device 33
3.2 Visual inspection 34
External examination 34
Internal examination 34
3.3 External battery interface test 35
3.4 ResBus interface test 35
3.5 Set time and date 35
3.6 Record device settings 36
3.7 Determine ventilation hours 38
3.8 Peep hours check 38
3.9 Learn circuit 39
3.10 Temperature sensor test 40
3.11 Heated flex test 40
3.12 Final assembly leak test 41
3.13 NRV check 41
3.14 Safety valve test 41
3.15 PEEP blower test 41
3.16 Pressure calibration test 41
3.17 Flow calibration test 42
3.18 Quick flow check 42
3.19 Blower test 42
3.20 Outlet pressure model test 42
3.21 Cooling fan test 43
3.22 Alarm test 43
3.23 Maximum pressure test 43
3.24 Remote alarm test 43
3.25 USB port test 43
3.26 Low flow O2 port test 43
3.27 SpO2 port test 44
3.28 LCD test 44
3.29 LED test 44
3.30 Keypad test 44
3.31 Internal battery test 44
3.32 Super capacitor test 44
3.33 Current consumption test 45
3.34 Electrical safety test 45
3.35 Finalise manual check results 46
3.36 Reload device settings 46
Calibration procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .48
3.37 Pressure calibration 48
3.38 Flow calibration 48
3.39 Outlet pressure model 49
Contents 5
Therapy alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
System failures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
Foreword 7
How to use this manual
Each of the sections in this manual have been written with a specific purpose in mind. Use this
guide to help you find the material which is relevant to your task at the moment.
1 General instructions – Outlines requirements when servicing Astral devices. Ensure the
requirements are read prior to servicing the device.
2 Function and operation – Provides technical information describing how the device functions.
3 Testing and calibration procedures – Lists all the test and calibration procedures that are
required as part of servicing the Astral. Also contains information for the cleaning and
disinfection of the device.
4 Preventative maintenance - Provides the preventative maintenance schedule with related kits
and testing.
5 Corrective maintenance - Lists the disassembly / reassembly procedures needed for the major
subassemblies, along with their spare part details. A complete list of Astral spare parts is also
included.
6 Troubleshooting guide - Provides a complete list of Astral error messages along with possible
fixes.
7 System specifications - Lists electrical and mechanical specifications of the device and air
tubing.
Appendix A: ResMed forms - Contains forms to fill out for the two year maintenance and 35,000
hour maintenance. Also includes Service Request Form - fill out this form if you need to return
a device to ResMed for servicing.
Appendix B: Pneumatic block components - Lists the disassembly / reassembly procedures
needed for the pneumatic block parts, along with their spare part details.
Training requirements
The maintenance procedures described in this manual must be performed by competent
technicians who have been trained in the maintenance of Astral devices and are ResMed
certified.
There are three levels of training.The following table shows the training requirements for each
level:
Training level
Training Topics/Modules Level 1 Level 2 Level 3
Device overview
Intended use
Technical specifications
Device operation
Alarms (Basic)
Disassembly/Assembly (excluding
pneumatic block)
Cleaning/ Preventative maintenance
Disassembly/Assembly of pneumatic
block
Definitions
This manual contains special terms and icons that appear in the margins to draw your attention
to specific and important information.
! Caution explains special measures for the safe and effective servicing of the device.
Note: If you have any feedback regarding the Astral service manual please contact Technical
Services, Sydney at TechSupport@resmed.com.au.
Foreword 9
1 General instructions
Warning Prior to disassembly or assembly, make sure that the power is not connected to the
ventilator.
!
General advice
Under no circumstances shall the manufacturer be liable for the damage caused to the device by:
• Servicing or modifications that do not comply with the manufacturer’s procedures or
performed without prior authorization from the manufacturer
• Servicing or modifications performed by unauthorised person.
Power supply
• The device must be connected to the mains supply with an Astral 90W mains supply pack
and cable.
Note: If the Astral power supply requires replacing, the replacement must be recorded in the
appropriate service reporting system. This record must include the serial number details
of the replacement (new) power supply.
• The electrical installation must comply with the local regulation standard.
• If there are any doubts regarding electrical disruption of mains supply the device must run on
the internal battery.
• It is strongly recommended the internal battery is replaced every two years.
Electromagnetic compatibility
• This device many be affected by interference from electromagnetic fields exceeding 20 V/m.
• The user should take the necessary precautions to ensure the specified electromagnetic
compatibility (EMC) limits are not exceeded. In particular, where the device is used in the
presence of short waves or high frequency generators, such as defibrillators, diathermy,
electrosurgery or radiology devices, cell phones etc., interference may occur if the energy
emitted is higher than the levels stipulated in Directive 93/42 EC.
The crossed-out wheeled bin symbol indicates that the product bearing this symbol may not
be disposed of together with general household waste, but instead requires separate disposal.
This requirement for separate disposal is based on the European Directive 2002/96/EC for
electrical and electronic equipment, and the European Directive 2006/66/EC for batteries. You
can hand in the product at a municipal collection point, for example. This reduces the impact on
natural resources and prevents contamination through the release of hazardous substances.
For further information regarding product disposal, please contact your local ResMed office or
visit our website at www.resmed.com
Dispose of used air s and air tubing according to the directives in your country.
1 General instructions 11
2 Function and operation
External device description
Touch screen
Adapter port
Inspiratory port
Handle (to patient)
USB connector
Mini USB
Ethernet
connector
connector
Device on/off
push button
Air inlet
(including DC power
hypoallergenic inlet
filter)
Low flow Data
oxygen input connector
(up to 30 L/min) cover
Top case
assembly
Expiratory
flexible
Screw - main PCB cable*
to chassis
Expiratory valve PCB
Main PCB pressure tube assembly
PEEP extension
tube assembly
Chassis - LP
CPC O2
coupler Expiratory
body and flow sensor
o-ring
Expiratory
interface
seal
O2 inlet seal
Expiratory adapter
Battery connector assembly
Battery cover
Expiratory cover
assembly
* The Expiratory flexible cable is only in Astral 150 devices. It is replaced by the Heated flex cable in newer Astral 150
devices. Heated flex cable functionality is included with SX544-0300 and will only work in devices where the heated
flex cable in installed.
2 Function and operation 13
Remote alarm
wire harness
SpO2 connector
wire harness
Thermistor
probe DC power
assembly (in connector
air inlet seal) wire harness
Power button
wire harness
Screw - chassis
to top case Data
connector
cover
Outlet tube
& seal
Pneumatic
block
Technical description
The Astral device provides continuous or intermittent ventilatory support for patients weighing
more than 5 kg (11 lb) who require mechanical ventilation. The Astral device is intended to be
used in home, institution/hospital and portable applications for both invasive and non-invasive
ventilation.
There are two versions of the Astral ventilator:
• Astral 100
• Astral 150
The following table details the differences between Astral 100 and Astral 150
Internal battery
Double limb
Preset programmes 2 4
Apnea ventilation
Therapy
Manual breath button
Astral
Main Controller Pneumatic Controller
Internal
External Patient
Battery
Interfaces
Pneumatic Ambient
Main micro- Main Blower
Power In Power Block HAC micro- temperature
processor Control
processor sensor
Patient Circuit
Atmospheric
User Interface Safety PEEP Blower
User Interface pressure Patient
Control Circuit Control
sensor interface
Airpath nebulizer
Outlet
Pressure
sensor
watertraps
Double circuit
pneumatic valve Single circuit
pneumatic
valve Tubes
Proximal Proximal
Expiratory
pressure Peep blower I/E Valve exhalation
Flow sensor
sensors valve
Exhalation
valve
Optional link
Optional Mandatory
Pneumatic link
Component Component
Electrical link
The Astral ventilator comprises three complex sub-assemblies: main PCB, a pneumatic block
(including a sensor PCB) and a top case assembly.
Main PCB
The main PCB hosts four subsystems:
• the Power Block;
• the Main Controller;
• the Pneumatic Controller;
• the Hardware Alarm Controller.
Power Block
The power block provides for conversion of the available DC supply to the required internal
operating voltages for use by the main blower, PEEP blower, electro-valves, electronic
subsystems, user interface (including LCD) and external interfaces.
The power block utilises external power when available, or the internal battery, when external
power is unavailable. When external power is available, the power block provides for
recharging of the internal battery.
The internal power rails are segregated between core ventilation rails and peripheral device
rails to prevent peripheral circuit failures from halting ventilation.
Pneumatic Controller
The Pneumatic Controller is comprised of an independent microprocessor, main & PEEP blower
control, ambient temperature and pressure sensors, and safety circuit.
The Pneumatic Controller provides control and monitoring of all pneumatic functions including:
• delivery of target inhalation pressure
• delivery of target inhalation flow
• control of exhalation pressure
• sequencing of valves
• acquisition and conditioning of all sensor signals
• derivation of patient measures
• leak compensation
• circuit compensation
Within the Pneumatic Controller a hardware circuit is implemented, such that a single
command from the Hardware Alarm Controller will force all pneumatic controls to their safe
state (“Safe states” on page 28).
2 3 4
1 2 1
3
4
The following diagram shows the main airpath components, the position of the control valves
and status of subassemblies when the device is powered off.
The following diagram shows the device connected with a single circuit adapter.
1/ Inspiration
At the start of inspiration, the flow control valve is deactivated, enabling the non-return valve
to open as wide as needed to satisfy the inspiratory flow. The I:E valve is also activated
routing outlet pressure to the expiratory valve in order to fully close it.
If the prescribed therapy specifies a positive expiratory pressure, the PEEP blower is driven to
the required set point, in preparation for the expiratory phase.
Oxygen management
Up to 30 L/min supplemental oxygen can be added into Astral’s oxygen inlet at the rear of the
device. Once ventilation has started, low pressure oxygen is added and mixed with ambient air
at the inlet to the main blower. The oxygen flow rate is controlled by the user before it
reaches the inside of the Astral. Monitoring supplemental oxygen is achieved using:
• integrated FiO2 sensor
• Low FiO2 and High FiO2 alarms, available when the oxygen sensor is installed
Power management
The Astral device can be used with four different power sources:
• AC Mains power
• Astral External battery (#27918)
• Internal battery
• External DC power supply (eg, car 12V power outlet)
For more information on power supplies and sources see the Clinical Guide.
Note: If the Astral power supply requires replacing, the replacement must be recorded in
the appropriate service reporting system. This record must include the serial number
details of the replacement (new) power supply
• power alarms:
Power disconnected when the power source is changed from an external source to the
internal battery
Using internal battery when the device is powered on and is using battery power
Automatic shutdown
Forced shutdown
When a forced shutdown of the device is required complete the following steps.
1 Disconnect any external power source (such as AC mains or external battery).
2 Press and hold the green on/off button and the alarm mute/reset button for at least 10
seconds. After 10 seconds the alarm bar will flash yellow.
3 Release both buttons. The Astral will then power off.
PC and software
• ResMed Service Software, available on the Technical services website
Note: Ensure you check the Technical Services website to confirm the latest version of
service software available for use with this product.After installation of the Service
Software ensure you install the Astral device driver.
• A computer running Windows XP/7 (32/64 bit), with at least three USB ports
Measurement tools
• Electrical safety analyser e.g. Rigel 288
• FlowAnalyser (PF-300 or equivalent)
The accuracy requirements of the flow analyser are as follows:
Measurement Range Accuracy
Low flow ± 20 L/min ±2% or 0.05 L/min
High flow ± 300 L/min ±2% or 0.1 L/min
Differential pressure ± 150 mbar ±0.75% or 0.1 mbar
Atmospheric pressure 0 - 1150 mbar ±1% or 5 mbar
Temperature 0 - 100° C ±1.75% or ±0.5°C
Note: A ±2% accuracy is acceptable for Low flow and High flow during preventative
maintenance procedures. An accuracy of ±1.75% is required for all other testing and
calibration
Note: When viewing values on the flow analyser screen disregard any negative sign (-) when
reading absolute value.
Note: Zero calibration must be performed on the flow analyser before every Astral service.
Accessories
• Remote alarm (#27902)
• Remote alarm cable (#27904)
• Nonin XPOD® oximeter Type 3012
• Astral External Battery (#27918)
• Oxygen sensor (#27935)
Caution If the target flow cannot be achieved in any of the following test and calibration
procedures complete the “3.19 Blower test” on page 42 to help locate the source of the
! device issue.
USB cable
Ethernet cable
Note: Instructions for setting up the ethernet connection to the PC can be found in the
Service Software User Guide.
3 Connect the Astral to the power outlet and press the green power button turn it on.
4 Allow the Astral to complete the self test.
6 Press the Refresh button (highlighted above) to enable the service software to recognise the
device.
7 The unit details will then be displayed at the bottom of the screen beside the Refresh button.
The calibration and tests in the Xplorer bar at the left of the screen will also change to
indicate the procedures specific for that product.
Test procedures
• Actichlor Plus
• Isopropanol
• CaviCide*
• Mikrozid*
* Suitable for cleaning on a monthly basis only
Caution When using a cleaning agent with a high alcohol content refer to the appropriate
Material Safety Data Sheet to ensure safe handling of the cleaning agent.
!
Note: The Astral can be decontaminated using Hydrogen Peroxide (H2O2) vapor or ozone.
This can be done up to 15 times with H2O2 vapor or 7 times with ozone throughout
the expected service life of the Astral (8 years). Decontamination should involve 15
minutes gassing, then 30 minutes dwell, followed by aeration (the length of aeration
depends on the room size).
For information regarding cleaning of multi-patient use devices refer to the Astral
Clinical Manual.
Caution Do not wash the air filter. The air filter is not washable or reusable.
!
External examination
Visually examine the Astral for external damage such as:
• deformation or cracking of the upper or lower cases
• damaged components such as the mains power inlet, oxygen inlet and SpO2 or FiO2
connectors
• damage to the user interface or bottom decal
Where there is external damage, it is recommended that an internal examination is conducted.
Internal examination
Whenever the unit is opened as part of a preventative or corrective maintenance a visual internal
examination should also be performed.
Inspect the unit for the following recording any irregularities and fixing them as appropriate:
• there is no contamination within the air inlet or outlet tube, or elsewhere within the device
• all mechanical components are seated correctly
• all electrical components are mounted correctly and secured in position
• there are no loose wires or loose terminals on electrical components or connections
• there are no loose pieces of wire or other objects trapped in the PCB
• there is no dust on the isolation barriers
Note: If there is dust on the isolation barriers, gently remove the dust using an ESD brush.
(The isolation barriers are the dark green strips surrounding the islands of circuits
around the rear connectors).
3 If the DC LED does not light up determine if the fault lies with the power inlet (on the chassis)
or with the main PCB.
3 If the AC power source LED does not light up determine if the fault lies with the power inlet
(on the chassis) or with the main PCB.
6 At this point, you can select the name of the service agent handling the device and enter the
Service Record (SR) and Repair numbers, so that this information will be included with the
device settings details.
Note: To enter new service agent details, click on the image in the "Select service agent"
window. A cursor will then appear in the "Service Agent" field, and the name can be
entered.
Once the service agent details are entered, any service agent names entered for that
device will be available for selection in the drop down menu.
If an existing session was selected, the "Select service agent" window will already be
populated.
Warning If the device settings cannot be recorded, reset to the default settings after completing
the device tests. Notify the customer that the device has been reset to default settings.
! Refer to page 50 for instructions on resetting to default settings.
4 If the ventilation hours displayed is > 35,000 the pneumatic block must be replaced.
Note: When examining the ventilation hours to determine if the pneumatic block should be
replaced, consider if the device will reach the 35,000 hour benchmark well before the
next scheduled maintenance.
2 If the value of the PEEP hours displayed is > 35,000 the pneumatic block must be replaced.
Note: When examining the PEEP hours to determine if the pneumatic block should be
replaced, consider if the device will reach the 35,000 hour benchmark well before the
next scheduled maintenance.
6 The Select Circuit type screen is then displayed. Select the Double Circuit for the Astral 150.
Select Single with leak for the Astral 100 (or if completing the second Learn Circuit for Astral
150 devices).
7 The Learn Circuit screen will then be displayed. Press the Start button.
8 Follow the prompts to attach and test the circuit.
9 If the Learn Circuit was successful you will then be prompted to select a required ventilation
mode. Select any mode, as this mode will be changed once the original device settings are
reloaded.
The following icons are used to report the Learn Circuit results:
Icon Description
OK Learn Circuit has passed
Warning!
13 If the device fails the Learn circuit examine the results on the device screen to determine the
location of the issue. Press the Confirm button and service the device as is required.
Further Learn Circuit troubleshooting information can be found in Section 6 Troubleshooting
guide, where a Learn circuit troubleshooting flow chart is provided.
If you are using a probe to test the unit ensure you test the ground pin of each port. The ground
pins are as follows:
Warning If the device settings cannot be restored, reset to the default settings after completing
! the device tests. Notify the customer that the device has been reset to default settings.
Refer to page 50 for instructions on resetting to default settings.
2 Ensure the expiratory adapter present in the unit at the beginning of the service has been
reinstalled in the device.
3 Fill out the ResMed Service Request Report in Appendix A if the device needs to be returned
to a ResMed service centre.
Note: If the Astral needs to be returned to ResMed the log files should be captured for
investigation. Contact TechSupport@resmed.com.au for the steps required to capture
the log files.
1 From the patient home screen press and hold the padlock icon for three seconds.
2 The following window will then be displayed:
2 Once in the Information menu hold down the Mute button for at least 10 seconds.
3 When the blue bar is visible at the top of the screen you have entered Service mode.
Note: Do not refer to the device Service mode when determining ventilation hours. In some
versions of the device software the ventilation hours are incorrectly described as
"Patient hours".
Caution If the target flow cannot be achieved in any of the test and calibration procedures
complete the “3.19 Blower test” on page 42 to help locate the source of the device
! issue.
After determining the issue with the device and replacing any faulty components, complete the
following calibration and test procedures in the order given below, prior to returning the device
to the customer.
3.11 Heated flex test Only required for Astral 150 devices with SX544-0300
and above where the heated flex cable is present
3.12 Final assembly leak test Required if the device has been opened.
3.37 Pressure calibration Required if the pneumatic block, main PCB or sensor PCB
was replaced
3.38 Flow calibration Only required if any part of the pneumatic block is
replaced -(not if the entire pneumatic block was replaced)
3.39 Outlet pressure model Only required if Pressure calibration and/or Flow
calibration have been completed.
3.13 NRV check Required if the pneumatic block has been replaced or
components within the pneumatic block replaced.
3.14 Safety valve test Required if the pneumatic block has been replaced or
components within the pneumatic block replaced.
3.17 Flow calibration test Only required if the flow calibration has been completed.
3.21 Cooling fan test Required when the cooling fan is replaced.
3.24 Remote alarm test Only required if the main PCB has been replaced.
3.25 USB port test Required if main PCB has been replaced or if there is a
fault in data transfer.
3.26 Low flow O2 port test Only required when the low pressure O2 adapter has been
disconnected from the pneumatic block or the chassis.
3.27 SpO2 port test Only required if the main PCB has been replaced.
3.28 LCD test Only required if the main PCB or top case has been
replaced.
3.34 Electrical safety test Required when the device has been opened.
The following flow chart outlines the regular maintenance procedures required every two years
or when the Vent hours or PEEP hours exceed 35,000 hours:
Note: When examining the ventilation hours to determine if the pneumatic block should be
replaced, consider if the device will reach the 35,000 hour benchmark well before the
next scheduled maintenance.
Note: After completing the Learn circuit you may wish to examine Astral Information menu or
the device error log (via the service software) if you are concerned with device
function.
4 Preventative maintenance 55
End of preventative maintenance test procedures
After the replacement of the components required as part of the two year preventative
maintenance the following test procedures must be performed in the following order:
If the device required replacement of the pneumatic block, cooling fan or any faulty components,
ensure you complete the required tests listed in “Calibration and test procedures required when
parts are replaced” on page 52.
Caution Printed circuit boards can be damaged by static electricity. Please read the instructions
“Electrostatic discharge precautions” on page 11 before attempting to replace any
! PCBs.
Thread-forming screws are used throughout the device. When replacing a thread-
forming screw, turn the screw anti-clockwise until it fits into the previously cut thread.
Otherwise the original thread may be stripped out
Equipment required
• Adjustable or pre-setable Torque screwdriver (such as the Lindstrom PS501-2 Pre Set
Precision Torque Screwdrivers - 7-70 cNm[Ncm], or the Stahlwille 760/7.5 Torsiometer
Torque Screwdriver - 15-75 cNm)
• Screwdriver bit, TX8
• Screwdriver bit, TX10
Note: The screw driver blade needs to be at least 65 mm long, with a diameter no larger than
5 mm.
Caution Torque control is required when fixing screws to the external cases.
!
5 Corrective maintenance 57
Opening the Astral
Note: Before opening the unit ensure the device is shutdown (not in standby). To shutdown,
remove mains power from the device. Press the green On/Off switch at the rear of the
device and follow the instructions on the screen.
1 Remove the filter from the rear of the unit.
2 Turn the unit upside down. Push on the button and pull away the expiratory valve cover, as
shown below:
3 Remove the two screws from the cover of the oxygen sensor enclosure (Z2053 - part of
screw kit #19651), and remove the cover (if there is an oxygen sensor present in the unit).
7 Remove the screw at the base of the battery compartment (Z2053) then remove the eight
screws from the bottom case assembly:
8 The bottom case assembly (#19611) can now be pulled away from the unit and replaced if
required. (If replacing the bottom case also remove the expiratory flow sensor, membrane and
filter from the case.)
Note: The bottom case assembly includes the 8 way connector for the battery and base
decals for both the Astral 100 and Astral 150. The base decals for the US (#19645) are
available separately.
If replacing the bottom case the serial number of the device needs to be retained. For
instructions regarding installation of a serial number decal onto a new bottom case
refer to tech note 1063224.
Caution Ensure the correct base decal is placed on the unit when the bottom case is replaced.
!
5 Corrective maintenance 59
Battery
Spare part: Internal battery - 19620
Battery cover - 19619
Caution Ensure there is no O2 supply connected and there are not high levels of O2 inside the
! unit prior to replacing the battery.
Note: Before opening the unit ensure the device is shutdown (not in standby). To shutdown,
remove mains power from the device. Press the green On/Off switch at the rear of the
device and follow the instructions on the screen.
The battery is replaced every two years as part of the two year maintenance procedure. The
battery should also be replaced when there is a noticeable reduction in usage time.
1 Using a Tx10 screwdriver bit, remove the screw (Z2053) from the battery cover.
Battery cover
6 Screw the battery cover into place, using a Tx10 bit with 0.2 ± 0.05 Nm torque.
Battery storage
When storing Astral batteries ensure:
• the battery is recharged every six months
• the battery is not stored in a depleted charge state. Charge the battery to 50% when
storing.
5 Corrective maintenance 61
Top case
Spare part: Top case assembly - 19606
Top case assembly (USA) - 19773
Top case assembly (Fukuda) - 19803
Top case decal (150) - 19643
Top case decal (100) - 19644
Top case decal (100SC) - 19794
Note: When installing a new top case on the device ensure you record the serial number of
the new top case in the device service record. The top case serial number can be found
on the flexible cable - see the image on the following page.
To replace the top case, the expiratory adapter membrane must be removed from the device. For
information regarding the expiratory membrane refer to “Expiratory flow sensor, membrane and
adapter” on page 88.
1 To disconnect the top case flexible cable from the main PCB gently pull up on the top case
flexible cable, to avoid damaging the main PCB connector and the top case connector.
Air inlet
3 Pull the chassis away from the top case, taking care to feed the flexible cable and connector
through the slot in the chassis to avoid damage.
4 Take the replacement top case and stick the appropriate decal (Astral 150 or Astral 100) in
the recess on the case, in the location shown:
Caution Ensure the correct product type decal is stuck to the top case.
!
5 Corrective maintenance 63
5 Turn the top case over and place it on the work surface as shown below:
6 Take the chassis and place it over the top case, while at the same time feeding the flexible
cable through the slot in the chassis.
7 Lift the flexible cable to visually align the flexible cable connector with the PCB connector.
8 Once they are aligned, gently press the connector into place. Do not force the connector.
9 Gently rock the connector stiffener to ensure it is fully inserted. The connector should be
firmly held in place.
3
2
Forced shutdown
Where the touch screen is unresponsive and you need to power off the device, complete the
following steps.
1 Disconnect any external power source (such as AC mains or external battery).
2 Press and hold the green on/off button and the alarm mute/reset button for at least 10
seconds. After 10 seconds the alarm bar will flash yellow.
3 Release both buttons. The Astral will then power off.
5 Corrective maintenance 65
Cooling fan
Spare part: Cooling fan assembly - 19610
1 Disconnect the fan cable from the connector on the main PCB.
Fan connector
1 Move the fan assembly away from the motor capacitor to access the motor capacitor cable.
2 Pull the capacitor from its slot in the chassis.
3 Disconnect the motor capacitor cable from the main PCB.
4 Take the new motor capacitor and slot it into the chassis.
5 Connect the capacitor cable to the main PCB.
6 Tuck the excess capacitor cable into the slot shown below, so it does not obstruct the screw
boss (circled) in the chassis.
5 Corrective maintenance 67
Buzzer assembly
Spare part: Buzzer assembly - 19607
1 Disconnect the buzzer cable from the connector on the main PCB.
2 Pull the buzzer assembly away from the rear of the device chassis.
3 Fit the new buzzer to the chassis. Begin by pressing firmly on the left side of the buzzer
housing then press the rest of the buzzer into place, ensuring the entire perimeter of the
buzzer is pressed fully onto the chassis.
4 Take the new buzzer assembly and connect it to the main PCB.
5 Route the cable from the buzzer assembly under the cable holder on the chassis and connect
it to the main PCB.
Cable holder
Note: When installing a new pneumatic block in the device ensure you record the serial
number of the new pneumatic block in the device service record.
1 If needed, move the cooling fan and the motor capacitor out of the way of the pneumatic
block.
2 Disconnect the PEEP extension tube (at the front of the unit) from the chassis.
3 Turn the unit around and disconnect the following cables from the main PCB:
• 10 pin PAFS cable running from the pneumatic block
• 8 pin VAMPS cable running from the pneumatic block
Note: Disengage the PAFS and VAMPS cable by pressing in the small tab on the connector
while gently pulling the connector away from the main PCB.
• Thermistor cable
• Blower cable
Blower cable
Thermistor cable
5 Corrective maintenance 69
4 Use a Tx8 screwdriver bit to remove the two screws (Z2039 - part of screw kit #19651) that
attach the low pressure O2 adapter to the pneumatic block. Hold the low pressure O2 adapter
while removing the screws, as shown below:
5 Gently lift upwards on the pneumatic block and disconnect the low pressure O2 adapter.
6 While pulling upwards on the pneumatic block the inlet assembly (with thermistor) and outlet
tube will also be removed from the chassis.
Outlet tube
Inlet assembly
7 Remove the inlet assembly and outlet tube from the pneumatic block.
Note: For information regarding the replacement of the outlet tube refer to page 73.
8 Take the new pneumatic block assembly and install the inlet assembly. Ensure it is installed
correctly, as shown below:
10 Place the pneumatic block over the chassis and connect the blower cable to the PCB.
11 While lowering the pneumatic block connect the low pressure O2 adapter.
Note: Ensure the thermistor is sitting under the low pressure tube.
Thermistor
12 While lowering the pneumatic block ensure the inlet assembly and outlet tubes slide into
position and are sitting flush with the chassis.
5 Corrective maintenance 71
Thermistor, inlet assembly
Spare part: Inlet assembly (seal and grid) - 19621
Thermistor - 19609
1 The inlet assembly (with thermistor) is removed from the device along with the pneumatic
block. See page 70 for further details.
2 Remove the inlet assembly from the pneumatic block.
3 Take the inlet assembly and pull the thermistor out from the inlet seal.
4 Take the new thermistor and insert it in the hole at the side of the inlet seal.
5 Just the tip of the thermistor should be sitting in the area between the outer grid ring and
inner grid ring, as shown below:
6 Part of the red thermistor probe should also be visible on the outside of the inlet seal.
7 Refit the inlet assembly to the pneumatic block. See page 70 for further details.
1 The outlet tube is removed from the device along with the pneumatic block. See page 70 for
further details.
2 Remove outlet tube from the pneumatic block.
3 Install the new outlet tube, with the O2 sensor interface facing upwards, as shown below.
Ensure the silicone seal has not been twisted out of position during assembly.
5 Corrective maintenance 73
Low pressure O2 adapter
1 The low pressure O2 adapter is easily replaced when the pneumatic block has been removed
from the device. See “Pneumatic block” on page 69 for further details.
2 Pull the low pressure O2 adapter from the CPC connector.
3 Take the new adapter and insert it on the CPC connector, ensuring it has been inserted
completely onto the adapter port.
With both the top and bottom cases removed the Tee tubing can be replaced.
1 Turn over the chassis to expose the main PCB.
2 Disconnect the Tee tubing from the chassis and the ports on the main PCB.
3 Fit the new Tee tubing to the chassis and the main PCB. Ensure each port is fully inserted into
the tubes to ensure effective sealing.
5 Corrective maintenance 75
Heated flex cable
Spare part: Heated flex cable - 19793
With both the top and bottom cases removed the heated flex cable can be replaced.
1 Remove the expiratory flow sensor from the chassis.
6 Feed the tab at the end of the cable so it is sitting under the main PCB.
7 Re-insert the flow sensor into the chassis (it will sit over the heated flex cable).
8 While holding the flow sensor in place turn the chassis back over. Bend over the heated flex
cable and insert the PCB into the sensor (through the hole in the chassis) and connect the end
of the cable to the main PCB.
5 Corrective maintenance 77
Main PCB
Spare parts: Main PCB (Astral 100) - 19653
Main PCB (Astral 150) - 19600
Expiratory flexible cable - 19602
PEEP extension tube - part of kit 19603
Note: When installing a new main PCB in the device ensure you record the serial number of
the new main PCB in the device service record.
The device serial number and product code also need to be installed on the new main
PCB. See “Installing the device serial number and product code” on page 82 for
installation procedures.
Note: If the main PCB is replaced and the device settings could not be recorded the internal
battery must also be replaced. After installation check the Battery install date. If the
following date is displayed (1/1/2050) please contact your local ResMed office for
assistance.
With the both the top and bottom cases removed, the main PCB can be replaced.
1 Disconnect all components currently connected to the top of the main PCB, following the
instructions provided in this manual.
2 Turn over the chassis to expose the PCB.
3 Disconnect the Tee tubing (circled) from the chassis and the main PCB.
4 Disconnect the expiratory flexible cable or the heated flex cable from the main PCB.
6 Use the Tx8 screwdriver bit to remove the following three screws (Z2039).
5 Corrective maintenance 79
8 Take the new main PCB and fit it on the chassis, using the board locators (highlighted below)
to ensure it is in the correct position.
9 Ensure the PEEP extension tube assembly and flexible cable are not pinned under the PCB.
13 Connect the oxygen sensor plug cable to the main PCB and insert the cable into ribs on the
chassis.
After the main PCB has been replaced the following behaviour is observed once the device is
powered on:
• System Error 75 will be displayed, as the pressure sensors have not yet been calibrated.
Recalibration of the device will remove the error message.
• If System Error 65 is displayed after replacing the main PCB contact
TechSupport@resmed.com.au.
5 Corrective maintenance 81
Installing the device serial number and product code
If the main PCB has been replaced the device serial number and product code needs to be
installed.
1 Once the device has been completely reassembled connect the unit to the PC and open the
service software, following the instructions “Connecting the Astral to the service software”
on page 32.
2 Select Device Settings in the XPlorer bar, or click on the Device Settings Dialog icon at the
top of the screen.
Note: It will take a few seconds for the device settings to be displayed.
3 The settings will be displayed.
4 Enter the unit’s serial number (found on the base of the unit) and select the appropriate
product code.
7 Click the Start session button . Any existing records for the unit will be displayed on
screen. Ensure the most recent record is highlighted.
Note: If no previously recorded session exists for the unit begin a new session.
8 Complete the calibration and test procedures required when the main PCB has been replaced
as outlined in Section 3 Test and calibration procedures.
9 Reload the clinical settings to the unit following the procedures from “3.36 Reload device
settings” on page 46.
10 Finally, reset the device time and date following the procedures from “3.5 Set time and date”
on page 35.
To replace the chassis, the following parts need to be removed from the device:
1 Once the chassis has been replaced stick the air inlet decal to the side of the chassis, in the
location shown below:
Caution Ensure you attach the air inlet decal that is appropriate for the device region.
!
The air inlet decals for each region and the languages displayed on these decals are as follows:
5 Corrective maintenance 83
2 After placing the air inlet decal on the device, two more decals need to be placed at the rear
of the chassis. Place the "Power On/Off" decal and the "O2" decals in the locations shown
below:
The oxygen sensor is supplied with the Astral 150, and is available as an optional accessory for
the Astral 100.
1 Turn the Astral over and remove the two screws (Z2053) from the cover of the oxygen
sensor enclosure, using a Tx10 screwdriver bit.
5 Corrective maintenance 85
4 Take the replacement oxygen sensor. Examine the seal on the oxygen sensor to confirm it is
in good condition.
Oxygen sensor with original seal Oxygen sensor with new lip seal
Note: Leak through the flow ventilation seal could affect ventilation accuracy.
5 Ensure the polarity features on the connector are oriented correctly before gently inserting the
connector into the chassis. (The red cable from the oxygen sensor is the cable closest to the
front of the Astral.)
Note: Pins can be damaged if alignment is incorrect.
Connector orientation
Caution Prior to fitting the oxygen sensor cover ensure there are no foreign objects trapped
within the oxygen sensor well.
!
6 With the cable connected, insert the sensor into the unit.
7 Push down on the tab to secure it into place. If it is a tight fit, wiggle the tab side to side a
little as you push downwards.
Cables tucked
to the side
Tab
9 Fit the oxygen sensor cover to the bottom case and refit the two screws (Z2053), using a
Tx10 bit, with 0.2 ± 0.05 Nm torque.
Note: The sensor cover will not go on correctly if the sensor is not installed properly.
5 Corrective maintenance 87
Expiratory flow sensor, membrane and adapter
Spare parts: Expiratory flow sensor - 27936
Pressure sensor filter - 27937
Expiratory adapter membrane - 19614
Expiratory adapter cover (ROW) - 19649
Expiratory adapter cover (USA) - 19613
Double limb adapter - 27940
Single limb valve adapter - 27941
Single limb leak adapter - 27942
Expiratory flexible cable - 19602
Note: For details regarding the replacement of the expiratory flexible cable refer to page 51.
For details regarding the replacement of the heated flexible cable refer to page 76.
1 Turn the Astral over.
2 Press and hold the eject button while pulling forwards on the cover.
3 The adapter is now visible. Lift up the adapter to remove it from the device.
Note: When the expiratory adapter membrane is removed, the pressure sensor filter may stick
to the membrane.
If the filter needs replacing, remove the filter from the membrane and place the new
filter in the chassis, as shown below:
5 Corrective maintenance 89
5 Sitting in the base of the chassis is the expiratory flow sensor. Use the tab at the side of the
sensor to remove it from the chassis.
6 Take the replacement expiratory flow sensor and insert it in the chassis, taking care not to
bend the sensor pins.
7 Refit the expiratory adapter membrane to the chassis, followed by the adapter.
Note: If installing a new expiratory adapter membrane ensure an expiratory pressure sensor
filter is installed in the unit. See page 89 for details.
8 Slide the adapter cover into place over the expiratory adapter and allow it to "click" into
place.
10 Turn the dial clockwise until it clicks.
Caution Do not wash the air filter. The air filter is not washable or reusable.
!
1 Remove the filter from the rear of the device by turning it in an anti-clockwise direction then
pulling it away from the device.
2 Pull out the air filter and muffler from the cover and discard the filter (and muffler if required).
3 Insert the new air filter into the muffler and insert both into the filter cover.
4 Refit the filter cover to the Astral, turning it in a clockwise direction until it is securely in
place.
Caution Only use genuine ResMed spare parts, because the filter affects the flow characteristics
of the device, which in turn affects the pressure supplied to the patient.
!
5 Corrective maintenance 91
Closing the unit
1 Before fitting the bottom case to the unit confirm:
• the three screws which attach the chassis to the top case are present
• the inlet assembly is present
• the expiratory pressure sensor filter and expiratory flow sensor are present
• the thermistor is present and in the correct position
• the PEEP extension tube is connected to the chassis
• the cooling fan is present and oriented correctly
• the motor capacitor cable is not obstructing the screw boss
Expiratory flow
sensor (under
membrane)
Third screw
(under
membrane)
Expiratory
pressure sensor
filter
Motor capacitor
cable and
screw boss
Thermistor
Cooling fan
Inlet assembly
2 Fit the bottom case to the unit. Using a small amount of force, push down on the case to
ensure the device is sealed and the case engages on the chassis.
3 Examine the fan mount assembly and confirm it is clearly visible against the bottom case.
92 Astral™ series service manual
4 Fit the nine screws in the following order (screws 1-8 are Z2051, screw 9 is Z2053), using a
Tx10 bit with 0.2 ± 0.05 Nm torque.
8 4
1
5
6
3 2
7
5 Refit the oxygen sensor. Details for refitting the oxygen sensor can be found on page 86.
6 Fit the oxygen sensor cover to the bottom case and refit the two screws (Z2053), using a
Tx10 bit, with 0.2 ± 0.05 torque.
Caution Ensure there is no O2 supply connected and there are not high levels of O2 inside the
! unit prior to replacing the battery.
5 Corrective maintenance 93
8 Refit the battery cover to the unit, as shown below, and slide it into place.
9 Screw the battery cover into place (using screw Z2053), using a Tx10 bit, with 0.2 ± 0.05
torque.
10 If removed previously, fit the expiratory adapter to the unit.
11 Slide the adapter cover into place over the expiratory adapter.
13 Turn the dial clockwise until it clicks.
14 Turn the unit over.
15 Insert the inlet filter and twist it until it "clicks" into place.
Note: If the Astral external battery is replaced, the replacement must be recorded in the
appropriate service reporting system. This record must include the serial number details
of the replacement (new) external battery.
To determine the amount of charge left in the Astral external battery press the On button. The
Battery Level Check indicator will be displayed for five seconds.
Charging LEDS
Flashing green Charging
It is recommended that the external battery is returned for servicing every two years. At the two
year service the battery inside the external battery should be replaced.
Caution When installing a new battery, discard any battery that has been dropped or shows
signs of being dropped, such as dents on the battery casing.
!
Note: When the battery inside Astral external battery is replaced, the replacement must be
recorded in the appropriate service reporting system. This record must include the serial
number details of the replacement (new) internal battery and the external battery into
which it was placed.
5 Corrective maintenance 95
1 Turn over the Astral external battery pack and remove the screw indicated below:
2 Slide back the battery cover and lift it away from the unit, exposing the Astral battery inside.
3 Remove the Astral battery from inside the external battery pack.
4 Insert the new Astral battery into the external battery, oriented as shown below:
5 Refit the cover to the base of the external battery and screw it into place.
5 Corrective maintenance 97
19653 Main PCB (Astral 100)
19608 Motor capacitor
19632 NRV assembly
19650 O2 sensor cover (ROW)
19617 O2 sensor cover (USA)
19616 O2 sensor plug and o-ring
19622 Outlet tube (with seal)
19652 PEEP assembly kit
19623 Pneumatic block assembly
19637 Pneumatic block sensor PCB
19626 Pneumatic block silicone kit
27937 Pressure sensor filter
19651 Screw kit, Final assembly (5 x Z2039, 9 x Z2051, 4 x Z2053, 3 x Z2055)
19640 Screw kit, Pneumatic block (8 x Z2054, 4 x Z2039)
19641 Single limb valve kit
19642 Double limb kit
19603 Silicone tubes kit (Tee tubing)
27942 Single limb leak adapter (Expiratory adapter)
27941 Single limb valve adapter
19609 Thermistor
19606 Top case assembly
19803 Top case assembly (Fukuda)
19773 Top case assembly (USA)
19643 Top case decal (Astral 150)
19644 Top case decal (Astral 100)
19794 Top case decal (Astral 100SC)
19638 Top lid
19627 Volute assembly
Note: The screws used in the Astral are not available as separate spare parts - they are only
available in screw kits #19651 and #19640.
Learn circuit
When troubleshooting a device fault it is highly recommended that you run the Learn Circuit.
The Learn Circuit takes the device through a number of tests determining the impedance and
compliance characteristics of the circuit. It also checks and calibrates the oxygen cell and
expiratory flow sensors.
The resulting information available after the completion of the Learn Circuit provides valuable
information regarding possible faulty components. A breakdown of the tests run during the
Learn circuit is provided in the flow chart below:
6 Troubleshooting guide 99
For the steps required to perform the Learn Circuit refer to “3.9 Learn circuit” on page 39.
Once the Learn circuit is complete you can examine the device screen to help determine the
location of issues with the device.
The following icons are used to report the Learn circuit results:
Icon Description
OK Learn Circuit has passed
Warning!
Problem/Cause Action
The following message is displayed on the LCD: Battery 1 fault (Battery 1 indicates the external battery
connected directly to the Astral)
Alarm priority: Low
Description: External battery 1 is connected but faulty
The following message is displayed on the LCD: Battery 2 fault (Battery 2 indicates the external battery
connected directly to Battery 1)
Alarm priority: Low
Description: 2 external batteries are connected and Battery 2 is faulty
• Circuit impedence does not match actual circuit • Run Learn circuit
• Pressure sensor fault (proximal or outlet pressure • Run pressure calibration test to determine location of
sensor) faulty sensor.
• Replace affected subassembly (main PCB or sensor PCB)
• Flow sensor fault • Run Quick flow test to determine location of faulty
sensor.
• Replace affected subassembly
The following message is displayed on the LCD: Critically low internal battery
Alarm priority: High
Description: Very low battery charge while device is powered internally or externally
• Low charge on battery • Check battery charge & connect unit to external power
source
• Foreign object in cooling fan inlet or outlet • Remove object from cooling fan inlet or outlet
• Incorrect expiratory adapter installed for circuit type • Connect appropriate adapter
• Peep valve control line not reconnected • Connect peep valve control line
• Peep valve pressure sensor failure • Replace pneumatic block (Level 2 technician)
• Replace sensor PCB (Level 3 technician)
Run Learn Circuit to determine any issues with the device. If the device successfully completed the Learn Circuit run
the following tests to determine the location of the device issue:
• Alarm test
• LED test
• Check oxygen sensor and connections • Fix connections or replace oxygen sensor as appropriate
• Replace main PCB
• View alarm details prior to this reset to determine cause for restart
Problem/Cause Action
• Main PCB has been replaced and pressure sensors • Recalibrate device
have not yet been calibrated
• Cables from pneumatic block not connected properly • Connect pneumatic block cables to main PCB
to main PCB
• Blower cable from pneumatic block not connected to • Reconnect blower cable to main PCB
main PCB
• Motor capacitor cable not connected to main PCB • Reconnect motor capacitor cable to main PCB
• Issue with cable connection to sensor PCB • Examine cable connections to sensor PCB and fix as
appropriate (Level 3 technician)
• Faulty PEEP blower or peep valve pressure sensor in • Replace pneumatic block(Level 2 technician)
pneumatic block • Replace PEEP assembly (Level 3 technician)
a. The International ventilator standard indicates that Paediatric patient type is intended to be used for a patient receiving less
than 300 mL, however Astral permits adjustment of ’Vt’ setting parameter up to 500 mL for cases where ’Vt’ is set such
that it compensates for leak in the breathing circuit.
b. Limits are the sum of device and circuit impedance.
c. Individual configurations may be more sensitive.
Note: The manufacturer reserves the right to change these specifications without notice.
A list of symbols which appear on the device can be found in the device clinical manual.
Refer to the clinical manual for Astral standards compliance details.
Monitoring
This section specifies the monitored parameters of the Astral device. Monitored therapy parameters are only
displayed during ventilation. Other parameters (eg, battery charge level and oximetry data) are monitored even
when not ventilating.
The Astral device monitors the following parameters:
Time parameters
Parameter Updates Units Range Resolution Accuracy
Pressure parameters
Parameter Updates Units Range Resolution Accuracy
SpO2 Once per second % SpO2 0 to 100 1 Refer to Nonin Xpod 3012
technical specifications at
www.nonin.com
Pulse rate Once per second 1/min 18 to 321 1 Refer to Nonin Xpod 3012
technical specifications at
www.nonin.com
Notes:
• All flow and volumes are measured at BTPS conditions.
• To achieve specified accuracies, a successful Learn Circuit must be performed prior to measurement
testing. Specified accuracies exclude mask compliance.
• Pressure may be displayed in cm H2O, hPa or mbar. The conversion factor of the pressure units is 1.
This means the displayed pressure will have the same values regardless of the unit selected.
Ti (Inspiration time) (A)CV, V-SIMV (mandatory 0.1 sec (Adult) ± (20ms + 5% of setting)
breaths: 0.3 to 3.0 sec 0.05 sec (Paed)
P(A)CV, P-SIMV (mandatory
breaths): 0.2 to 5.0 sec
P(A)C: 0.3 to 4.0 sec
Apnoea Ti When Volume Breath option is 0.1 sec (Adult) ±(20 ms + 5% of setting)
set to Ti. 0.05 sec (Paed)
If Apnoea Response is (A)CV
+ Alarm: 0.3 to 3 sec
If Apnoea Response is P(A)CV
+ Alarm: 0.2 to 5 sec
a. The International ventilator standard indicates that Paediatric patient type is intended to be used for a patient receiving less
than 300 mL, however Astral permits adjustment of ’Vt’ setting parameter up to 500 mL for cases where ’Vt’ is set such
that it compensates for leak in the breathing circuit
SmartStart? Yes No
Problem description:
Effect on user:
Service details
ResMed reference number Distributor reference number (if applicable) Signature
AF305-009-01
Cleaning the device: Pass Fail Maximum pressure test: Pass Fail
Visual inspection: Pass Fail LED test: Pass Fail
External battery test: Pass Fail Keypad test: Pass Fail
ResBus interface test: Pass Fail Internal battery test: Pass Fail
Internal inspection and dusting: Pass Fail Max. pressure test: Pass Fail
Visual inspection: Pass Fail Low flow O2 port test: Pass Fail
Outlet tube
Inlet assembly
2 Turn the pneumatic block over and remove the three screws (Z2054 [Tx10] - part of kit
#19640) from the top lid.
Note: When installing a new sensor PCB in the device ensure you record the serial number of
the new sensor PCB in the device service record.
1 With the top lid removed the pneumatic block sensor PCB is visible.
2 Remove the four screws (Z2039 [Tx8] - part of kit #19640) from the PCB using a Tx8
screwdriver bit.
3 Disconnect the blue and yellow cables from the top of the sensor PCB.
4 Gently lift up the PCB to expose the cables connected underneath.
5 Disconnect the PEEP turbine cable and the yellow electrovalve cable from the PCB.
Grommet
Note: With the sensor PCB removed the grommet (part of kit #19626) can be replaced if
required.
7 Remove the two cable harnesses (with grommet) from the sensor PCB.
Note: At this point cable harness 1 (10 pin PAFS cable - part #19635) and cable harness 2 (8
pin VAMPS cable - part #19636) can be replaced.
8 Take the new sensor PCB and connect the cable harnesses (with grommet).
9 Connect the PEEP turbine cable and the yellow electrovalve cable to the PCB base.
Note: Do not bend the cable headers when connecting the cables to the PCB.
10 Carefully push the sensor PCB into place on the assembly, ensuring the pressure and flow
sensors (on the underside of the PCB) engage with the seal volute sensor.
11 Slot the grommet into place in the pneumatic block chassis.
12 Tuck the PEEP turbine cable and the yellow electrovalve cable next to the silicone tubing.
13 Confirm the silicone is sitting correctly under the PCB and that no wires are obstructing the
screw boss.
Cables tucked
next to tubing
Screw boss
2
4
15 Reconnect the blue and yellow cables to the top of the sensor PCB. Push the blue and yellow
cables away from the screw boss, towards the PCB, to avoid pinching the cables when the
top case in installed.
With the sensor PCB removed the PEEP control tube can be replaced.
1 Lift up the blower cable and pull the PEEP control tube away from the blower.
2 Pull the PEEP control tube off the pressure port.
3 Take the new PEEP control tube and insert it onto the pressure port.
4 Lift up the blower cable. Wind the tube around the blower, ensuring it sits behind the clip
next to the blower.
5 Ensure the tubing sits in the groove circled below, with the elbow outside of the pneumatic
block box.
Blower cable
Clip
Tube attached to
pressure port
Note: When installing a new PEEP turbine in the device ensure you record the serial number of
the new PEEP blower in the device service record.
1 With the sensor PCB removed the PEEP assembly can be accessed.
2 To pull the PEEP assembly away from the pneumatic block, pull up on the PEEP inlet/EV
moulded tube and the volute sensor seal.
3 Take the new PEEP assembly and slot it into the volute assembly, as shown below:
Pressure port
5 Push the volute sensor seal onto the five areas on the volute assembly shown below:
1 Turn the pneumatic block over and remove the five screws (Z2054 [Tx10] - part of kit
#19640) from the bottom lid of the Astral housing and remove the lid from the assembly.
2 Disconnect the moulded NRV inlet elbow from the NRV assembly.
Note: At this point the moulded NRV inlet elbow (part of kit #19626) can be replaced if
required.
3 Rotate the NRV anti-clockwise to unlock it from the volute.
Note: The force to unlock the NRV is quite high until the clips are disengaged.
6 The pressure port on the NRV should be aligned with the moulded NRV inlet elbow.
7 Fit the moulded NRV inlet elbow to the NRV assembly.
8 Using a Tx10 bit refit the five screws in the following order:
4
3
1 5
Note: When installing a new volute assembly in the device ensure you record the serial
number of the new blower in the device service record.
1 With the sensor PCB and the PEEP turbine removed from the pneumatic block the volute
assembly can be replaced.
2 Turn the pneumatic block over and remove the five screws (Z2054 [Tx10] - part of kit
#19640) from the bottom lid of the Astral housing and remove the lid from the assembly.
3 Disconnect the moulded NRV inlet elbow from the NRV assembly.
EV-NRV tube
6 Remove the EV-NRV tube from the electrovalve with the yellow cable.
7 Remove the main blower heat sink from the top of the blower.
8 Remove the integrated seal from the air outlet and the port at the rear.
Note: At this point the moulded EV-NRV tube (part of kit #19626) can be replaced if required.
12 Insert the new volute assembly into the pneumatic box block, taking care to feed through the
moulded NRV inlet elbow through the hole in the housing.
13 Fit the integrated seal to the air outlet and the port at the rear, and ensure the seal is sitting
correctly on the aluminium housing.
14 Connect the EV-NRV tube to the electrovalve with the yellow cable.
Retaining clip
16 Turn the assembly over and confirm the volute is sitting on the groove of the seal (as
highlighted below):
4
3
1 5
1 With the volute assembly removed the aluminium housing can be replaced.
2 Remove the integrated seal from the aluminium housing.
Note: At this point the integrated seal (part of kit #19626) can be replaced if required.
3 Take the new aluminium housing and insert the integrated seal. When installing the seal
ensure it is seated correctly in the pneumatic box block.
4
3
1 5
Note: The bottom lid (#19639) can be replaced if required. If replacing the bottom lid the
serial number of the pneumatic block needs to be retained. For instructions regarding
installation of the serial number decal onto a new bottom lid refer to tech note
1063224.
3 Confirm that all cables and tubes are in the correct place and will not be crushed by the top
lid (particularly the PEEP control tube and PEEP turbine cable).
2
3
6 Refit the outlet tube and inlet assembly to the pneumatic block prior to installing the
pneumatic block in the unit. Steps for refitting both the inlet assembly and outlet tube can be
found on page 71.
Note: The screws used in the pneumatic block are not available as separate spare parts - they
are only available in screw kit #19640.