CALIBRATION CERTIFICATE
Cali. Date: 15-02-2024 ULR No. CC300019000002501F Carafano.: EX/2024/02/7
Due Date: 14-02-2025 Certi. No.: EX/M/2022/02/8/011 Discipline: Mechanical
Receipt Date: 15-02-2024 Call. WA. : EX/CWP/M/028 Parameter: Pressure
Issue Date: 15-02-2024 Ref. IS. DKD-R 6-1 Format No. EX/CR/PG

CUSTOMER
M/s, Immense Industry Environmental Temp. 25± 5°C
107AshutoshSoc,B/hUtkarshPetrolPump, Condition % Rh Lessthan 75% Rh
Karelibaug,Vadodara. Location of calibration At Site
Condition of Item Satisfactory

2.Description of Item
Name :- Pressure Gauge Range :- 0- 10000 psi
I.D No. :- EN 837-1 Resolution 10 Bar
Make / Model :- Bolttorqueingpump Specified Accu. :- ---
Sr.No. :- 54102/9001TT8U Location wrench :- Nut Torquing Application
Type :- Analog
.
3. Master Instruments Details
Name :- Digital Pressure Gauge
I.D No. :- EX/EQP/02
Certificate No. :- VIS/24-25/M-134
Certified By :-VijayInstrumentationServices
Cali. Validity :-15-05-2025

4.Calibration Result of UUC Zero Set Point Error = 0.000


Reading Observed By Master Gauge Std. Exp.UNC
Set Value Oil ± Error
Average Hysteresis +- % Rdg
UUC UP Down Reading Ap

Pind M1 M2 Mk, Pkid-Mm. terse/


0.0 0.00 0.00 0.00 0-00 0.00 #
140.0 140.98 140.96 140.97 -0.97 0.02 0.83
280.0 280.93 280.87 280.90 -0.90 0.06 0.83
420.0 420.89 420.82 420.86 -0.86 0.07 0.83
560.0 560.84 560.78 560.81 -0.81 0.06 0.83
700.0 700.78 700.78 700.78 -0.78 0.00 0.83
Mo: 7202822212, 9998755736
Office:A-1,2, 2ndFloor,TejendraComplex,
Opp. CMC-Power House, Odhav Road, Odhav,
Ahmedabad-382415. Ph. 079 22870063
E-mail excellentcalahm@gmail.coM
Certificate No. CC-3000 Web: www.excellentcal.com

Thereporteduncertaintyis the expandeduncertaintyin measurementobtainedby multiplyingthe standarduncertaintyby


the coverage factor k=2, whichcorresponds to a coverage probability of approximately 95% for normal distribution.
Note:-
1. Next Calibrationdue date is reported as per customer request.
2. Thiscertificate refers onlythe particular item calibrated at site/laboratory.
3. The calibrationresults reported are valid at the the time of and under the stated conditions of measurement.
4. Thecertificateshouldnot be reproducedexcept,infullwithoutthe prior permissionfromthe TechnicalManager