Hindawi

Journal of Healthcare Engineering


Volume 2022, Article ID 2507149, 7 pages
https://doi.org/10.1155/2022/2507149

Research Article
Application of Different Ventilation Modes Combined with
AutoFlow Technology in Thoracic Surgery

Wang Lixian , Yang Yanfang, Cui Chengzong, Jiang Ning, and Guo Yufeng
Cangzhou Central Hospital, Cangzhou, China

Correspondence should be addressed to Wang Lixian; wanglixian@mail.chzu.edu.cn

Received 16 January 2022; Accepted 10 March 2022; Published 28 March 2022

Academic Editor: Deepak Kumar Jain

Copyright © 2022 Wang Lixian et al. This is an open access article distributed under the Creative Commons Attribution License,
which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

To investigate the effect of AutoFlow on airway pressure and hemodynamics in mechanical ventilation constant volume-control
ventilation mode, 100 patients receiving mechanical ventilation were randomly divided into observation group (SIMV-PSV-
PEEP + AutoFlow) and control group (SIMV-PSV-PEEP). The results showed that the peak airway pressure and average airway
pressure decreased with different flow rate settings and automatic flow conversion (P < 0.05). The peak airway pressure and mean
airway pressure decreased with different resistance settings (P < 0.05). With different compliance settings, the peak airway
pressure and average airway pressure decreased after being assisted with an automatic converter (P < 0.05). Adding AutoFlow on
the basis of SIMV-PSV mode can significantly reduce peak inspiratory pressure (PIP), mean airway pressure (Pmean), and airway
resistance (R). There was no significant difference in hemodynamic monitoring results between the observation group and the
control group. It is proved that the SIMV constant volume-controlled ventilation mode combined with AutoFlow can not only
ensure tidal volume but also avoid excessive airway pressure, which has little effect on hemodynamics.

1. Introduction constant pressure, improper pressure adjustment, or in-


creased airway resistance, it is easy to lead to insufficient
At present, ventilator technology tends to mature and the ventilation, causing hypoxemia and carbon dioxide reten-
application level of clinicians is stable. Mechanical venti- tion. The auxiliary ventilation technology combined with
lation has been widely used and has become an important SIMV + AutoFlow can dynamically adjust the air flow
means of clinical treatment and rescue of critically ill pa- according to the patient’s current airway pressure and chest-
tients. With the accumulation of clinical experience and the lung compliance and send the preset tidal volume with a
application of microcomputer technology, some new me- deceleration wave and the lowest pressure, which can make
chanical ventilation modes and technologies have been up for the deficiency of the BiPAP mode [3].
developed and applied, such as: its common features are that
compared with the previous auxiliary ventilation mode, it is 2. Literature Review
closer to the physiological state, more in line with the actual
respiratory mechanical characteristics and pathophysiology Ge et al. believed that thoracoscopy is a new technology
of patients, and more humanized. Under the condition of applied in thoracic surgery in recent years. Under the direct
meeting the ventilation, it aims to make patients comfortable vision of a high-definition television imaging system, the
to the greatest extent, pay more attention to preserving operation system is inserted into the chest through chest wall
patients’ autonomous breathing function, improve man- incision to diagnose and treat chest diseases. This operation
machine coordination as much as possible, minimize pa- has the advantages of small incision, small injury, and fast
tients’ respiratory work, and avoid complications related to recovery and can be widely used in clinics [4]. Campos and
artificial respiration support [1, 2]. Traditional fixed pressure Sharma believed that patients with primary bronchial lung
ventilation mode BiPAP is applied to neonates. Because of cancer (referred to as lung cancer) and chronic obstructive
2 Journal of Healthcare Engineering

pulmonary disease (COPD) had common risk factors, such velocity waveform of deceleration wave and the lowest
as smoking, so many lung cancer patients were diagnosed pressure, so as to avoid the occurrence of barotrauma to the
with COPD in the preoperative pulmonary function as- greatest extent. At the same time, the patient is allowed to
sessment of thoracoscopy [3]. Hahm believe that patients breathe freely in the whole ventilation cycle, and the man-
often have complications such as pulmonary infection and machine coordination is better. If the airway resistance and
atelectasis after thoracoscopic surgery due to low ciliary chest-lung compliance change, the airway pressure shall be
function, which increases the length of hospital stay, so it increased or decreased by 3 cm H2O each time, but not
puts forward higher requirements for nursing work [5]. higher than 5 cm H2O below the set pressure alarm high
Gedik and Alar believed that each pressurization of the limit, otherwise the pressure and tidal volume alarm [11].
oscillating sputum drainage system can force the air flow to Based on the current research, in order to explore the
quickly pass through the lung tissue, stimulate the cilia effect of AutoFlow on airway pressure and hemodynamics in
through the chest wall oscillation, improve the cilia swing mechanical ventilation constant volume control ventilation
frequency, promote the secretion to fall off from the trachea mode, 100 patients receiving mechanical ventilation were
and bronchus and move to the oral cavity, and discharge it randomly divided into observation group (SIMV-PSV-
through normal cough or negative pressure sputum suction PEEP + AutoFlow) and control group (SIMV-PSV-PEEP).
after reaching the oral cavity [6]. Meng et al. believed that
this method is suitable for patients with pulmonary insuf- 3. Materials and Methods
ficiency caused by difficulty in excreting pulmonary secre-
tions or mucus blocking the lungs and can promote airway 3.1. Clinical Data. From January 2019 to January 2021, 100
clearance and sputum excretion or improve bronchial patients with respiratory failure received mechanical ven-
drainage. Therefore, it can reduce the incidence of pul- tilation in the respiratory intensive care unit (RICU) of our
monary infection and atelectasis in patients with lung cancer hospital, including 75 males and 25 females. All patients
complicated with COPD [7]. Habertheuer et al. thought that underwent oral endotracheal intubation or tracheotomy
volume-controlled ventilation (VCV) and pressure-con- after admission to the RICU and were mechanically ven-
trolled ventilation (PCV) are commonly used in mechanical tilated with the Drager EVTA 4 ventilator. The ventilation
ventilation (MV). VCV and PCV have their own advantages mode was SIMV pressure support ventilation (PSV)-positive
and disadvantages. The main advantage of VCV is to ensure end expiratory pressure (PEEP) [12].
tidal volume, while PCV has obvious advantages in reducing
the risk of barotrauma, improving human-machine coor-
dination, reducing patients’ respiratory work, and im- 3.2. Method
proving gas distribution in the lung and ventilation blood 3.2.1. When Transferred to ICU. In a randomized controlled
flow ratio [8]. Park believes that the automatic flow function study, the included children were randomly divided into
of the ventilator is a kind of dual control mode. The dual groups A and B according to the random number table.
control mode attempts to combine the advantages of VCV Group A represents the BiPAP group and group B represents
and PCV. Some studies have compared the differences the SIMV + AutoFlow group. After general anesthesia, the
between other dual control modes and traditional constant subjects of the two groups were transferred to the ICU under
current air supply VCV and PCV. It is considered that the stable respiratory and circulatory conditions, such as body
advantages of a dual control mode are as follows: while temperature, heart rate, blood pressure, oxygen saturation,
ensuring the tidal volume, it avoids the disharmony between central venous pressure (CVP), respiratory rate, tidal vol-
the air supply flow rate caused by the fixed VCV flow rate ume, and peak airway pressure and resistance, and me-
and the needs of patients, improves the man-machine chanical ventilation was given with the German Drager
synchronization, reduces the patient’s respiratory work, EVTA 4 ventilator.
reduces airway pressure, improves gas exchange, and is safer
for patients with unstable ventilation. However, few studies
comprehensively compare the changes of airway pressure, 3.2.2. Breathing Mode Using Ventilator. BiPAP ventilation
flow rate, and other respiratory mechanical indexes after mode parameter setting: determine the tidal volume (VT) by
auxiliary automatic conversion [1]. George et al. believed adjusting the difference between Phigh and Plow according to
that the study of a model lung can set different compliance, the patient’s weight, which is generally calculated as TV � 8 ∼
resistance, and inspiratory flow rate and exclude the in- 10 ml/kg. The preset respiratory rate is 25 ∼ 40 times/min.
fluence of patients’ breathing mode and other clinical Adjust the inspiratory time to determine the inspiratory
conditions. It has good repeatability and is an ideal method respiratory ratio (I:e), usually 1 : 1 or 1 : 2; Phigh is usually 15 ∼
to study respiratory mechanics [9]. Rohde et al. believed that 20 cm H2O, plot is 5 ∼ 7 cm H2O. SIMV + AutoFlow ven-
AutoFlow is an auxiliary ventilation technology combined tilation mode parameter setting: the respiratory rate is
with the SIMV mode. It is not an independent ventilation usually 25 ∼ 40 times/min, VT � 8 ∼ 10 ml/kg, the tidal
mode, but a functional extension of the capacity control volume is set, the breathing ratio is set from 1 : 1 to 1 : 2, and
breathing mode [10]. Dearani et al. believed that after using the positive end expiratory pressure (PEEP) is generally set
AutoFlow, the ventilator will dynamically adjust the air flow at 0 ∼ 3 cm H2O. The support pressure (PSV or PASB) is
according to the patient’s current airway pressure and chest generally adjusted to 10 ∼ 20 cm H2O. After the ventilator is
lung compliance and send it to the preset tidal site with the connected with the patient, the ventilation parameters are
Journal of Healthcare Engineering 3

adjusted appropriately according to the specific condition Table 1: Respiratory mechanics results of two groups (x ± s).
[13]. Number of
Group PIP Pmean R VT
cases
Observation
3.2.3. Offline Standard. When the patient is awake and 50 28 ± 5 27 ± 4 40 ± 18 589 ± 70.1
group
cooperative, spontaneous breathing is stable, cough and Control group 50 15 ± 4 9 ± 4 26 ± 14 574 ± 79.9
swallowing reflexes are recovered, the respiratory tract is P value <0.01 <0.05 <0.05 >0.05
unobstructed, circulation is stable, and the internal envi-
ronment is balanced. It is expected that there is no possibility
of reoperation and endotracheal intubation, and the offline
steps can be implemented when the blood gas analysis Pao Table 2: Blood gas and hemodynamic results of two groups.
is > 60 mmHg, PaCO2 is < 50 mm Hg, and FiO2 is ≤ 40%.
Number of
Group HR ABPM PH PaO2
cases
Observation
3.2.4. Offline Steps. Firstly, reduce the inhaled oxygen group
50 95 ± 11 74 ± 12 6.9 ± 0.1 87 ± 13
concentration to less than 40%, and then, gradually adjust Control group 50 92 ± 10 72 ± 10 6.9 ± 0.1 91 ± 16
the support pressure to less than 5cmho. The respiratory rate P value >0.05 >0.05 >0.05 >0.05
in the SIMV mode is gradually reduced to less than 5 times
per minute, the Phigh in the BIPAP mode is gradually re-
duced, the Plow is synchronously reduced, and the reduction
can be stopped when it reaches 3 cm H2O. When the dif- 4. Results
ference between Phigh and Plow is reduced to 5 ∼ 7 cm H2O, it
can be changed to CPAP or maintained at 5 cm H2O and the The respiratory mechanics results of the two groups are
PASB can be synchronously reduced to less than 5 cm H2O. shown in Table 1. Table 1 shows that PIP and Pmean in the
Observe for 30 minutes after the trial shutdown and measure observation group decreased significantly and R decreased
the arterial blood gas analysis (including arterial blood gas significantly compared with the control group. There was no
pH value, arterial blood oxygen partial pressure, arterial significant difference in VT between the observation group
blood carbon dioxide partial pressure, blood oxygen satu- and the control group. The results of hemodynamics and
ration, and BE value). The patient’s blood pressure, heart arterial blood gas analysis of the two groups are shown in
rate, and respiration are stable. Endotracheal intubation was Tables 2 and 3. It can be seen from Tables 2 and 3 that there is
pulled out, and oxygen was absorbed by a mask. After no significant difference in arterial blood gas analysis results
tracheotomy, patients were removed from the machine, and hemodynamic monitoring results between the two
oxygen was directly absorbed by gas cut masks to ensure groups [15].
normal blood oxygen saturation [14]. A total of 100 children were enrolled. The longest
transfer time was 137 minutes, and the longest use time of
ventilator was 40 hours and the shortest was 5 hours. Five
3.3. Observation Indicators. Monitoring indicators include cases died recently after operation, including 3 cases of low
the following: (1) record the general information of patients cardiac output syndrome, 1 case of arrhythmia, and 1 case of
after admission to the ICU, including patient name, gender, pulmonary infection.
age, weight, hospitalization number, admission diagnosis, Two independent samples t-test was used for the general
operation status, operation name, and cardiopulmonary data of the two groups. Among the 100 patients observed,
bypass time and (2) after admission to the ICU, each patient there was no significant difference in age, weight, transit
was continuously monitored for basic vital signs, including time, blocking time, vital signs, and ventilator monitoring
body temperature, blood pressure, heart rate, pulse, respi- indexes between the two groups (P > 0.05).
ratory rate, and blood oxygen saturation. After admission, The intragroup comparison showed that there was sig-
routine arterial blood gas analysis was given. Vital signs, nificant difference in peak airway pressure at different time
blood gas (pH, PaCO2, and PaO2), and ventilator indexes, points in patients with the same mechanical ventilation
including respiratory rate (f ), tidal volume (VT), peak mode with time (P < 0.05). It can be seen from the figure that
airway pressure (Ppeak), lung compliance (c), airway resis- the peak airway pressure of children applying the
tance (R), and minute ventilation (MV), were recorded at SIMV + AutoFlow mechanical ventilation mode is lower
five time points of admission (t0), 30 minutes (t1), one hour than that of the BiPAP group and shows a downward trend
(t2), two hours (t3), and three hours (t4). (see Figure 1 for the results) [16]. At the same time, the
alveolar compliance was different at different time points
(P > 0.05) (see Figure 2 for the results).
3.4. Statistical Treatment. All data were processed by PEMs The delivered tidal volume remained unchanged before
3.1 medical statistics software package, the observation data and after auxiliary automatic conversion (500 ± 1.3 ml vs,
were expressed by X ± s, and the paired t-test was used for 500 ± 1.2 ml, P > 0.05). After auxiliary automatic conversion,
comparison between groups. P < 0.05 was statistically peak airway pressure and average airway pressure decreased
significant. (P < 0.05) (see Figure 3).
4 Journal of Healthcare Engineering

Table 3: Blood gas and hemodynamic results of the two groups.


Group Number of cases PaCO2 CVP CO SV
Observation group 50 43 ± 10 14.8 ± 3.1 7.6 ± 0.9 65 ± 10.4
Control group 50 41 ± 9 15.4 ± 3.0 7.8 ± 0.8 65 ± 8.9
P value >0.05 >0.05 >0.05 >0.05

50
17.5

Airway pressure (cm water column)


40
Estimated Marginal Means

17.0

30
16.5

20
16.0

10
15.5

0
0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 The airway peak pressure Mean airway pressure
Time (h)
Auxiliary white dynamic flow converter before
Observation group
The control group Auxiliary white dynamic variable flow

Figure 1: Change in the trend of peak airway pressure in the Figure 3: Peak and average airway pressure before and after
SIMV + AutoFlow group at different time points. auxiliary automatic conversion.

settings, the peak airway pressure and average airway


34
pressure decreased after assisted automatic flow conversion
32 (P < 0.05) (see Figures 8 and 9).
Estimated Marginal Means

30
28 5. Discussion
26 Mechanical ventilation modes can be divided into two
24
categories: constant pressure ventilation and constant vol-
ume ventilation (SIMV). Compared with constant pressure
22 ventilation, the advantage of constant volume ventilation is
20 that it can ensure the preset tidal volume; the doctor presets
the VTdelivered by the command breathing mode, and then,
18
the ventilator delivers the predetermined VT regardless of
16 resistance and compliance. The proximal airway pressure of
1.0 1.5 2.0 2.5 3.0 3.5 4.0
patients changes. If the airway resistance increases or the
Time (h) respiratory compliance decreases, it can cause high peak
airway pressure and plateau pressure, resulting in excessive
Observation group
The control group expansion of localized alveoli and barotrauma [17]. The data
of this group showed that after SIMV combined with
Figure 2: Change in the trend of lung compliance in the AutoFlow, the peak inspiratory pressure and average airway
SIMV + AutoFlow group at different time points. pressure decreased significantly and the airway resistance
decreased, indicating that AutoFlow can improve respira-
On the premise that the conveying moisture volume tory mechanics to a certain extent. AutoFlow itself is not an
remains unchanged (P > 0.05), with different flow rate set- independent ventilation mode. As a pressure regulation
tings, the peak airway pressure and average airway pressure function extension of the volume-controlled breathing
decreased after auxiliary automatic flow conversion mode, it combines the advantages of pressure-controlled
(P < 0.05) (see Figures 4 and 5); with different resistance ventilation and volume-controlled ventilation when used in
settings, the peak airway pressure and average airway combination with the SIMV mode. The working principle is
pressure decreased after auxiliary automatic flow conversion as follows: at the beginning of the ventilator, four consec-
(P < 0.05) (see Figures 6 and 7); with different compliance utive experimental breaths with a pressure of 10 cm H2O are
Journal of Healthcare Engineering 5

60 70

55
Airway pressure (cm water column)

Airway pressure (cm water column)


60
50

45 50

40
40
35
30
30

25 20

20
10
30 35 40 45 50 55 R1 R2 R3 R4 R5 R6
Time (min)
Auxiliary white dynamic flow converter before
Auxiliary white dynamic flow converter before Auxiliary white dynamic variable flow
Auxiliary white dynamic variable flow
Figure 6: Changes in peak airway pressure before and after
Figure 4: Changes in peak airway pressure after auxiliary auto- auxiliary automatic conversion under different resistance settings.
matic conversion under different flow rate settings.

12.5 14
Peak airway pressure (cm water column)

12.0 13
Airway pressure (cm water column)

11.5 12

11.0 11

10.5
10

10.0
9
9.5
8
9.0
R1 R2 R3 R4 R5 R6
30 35 40 45 50 55 Auxiliary white dynamic variable flow
Time (min) Auxiliary white dynamic flow converter before
Auxiliary white dynamic flow converter before Figure 7: Changes in mean airway pressure before and after
Auxiliary white dynamic variable flow
auxiliary automatic conversion under different resistance settings.
Figure 5: Changes in mean airway pressure after auxiliary auto-
matic conversion under different flow rate settings.
the preset inspiratory upper limit pressure level, but each
performed, and the current airway pressure and lung-chest adjustment range is less than 3 cm H2O, otherwise the
compliance of the patient are continuously measured by a pressure and tidal volume alarm. The main advantages of
microcomputer. According to the volume-pressure rela- SIMV combined with AutoFlow are as follows: good man-
tionship, calculate the suction pressure required for the next machine coordination can reduce or avoid the use of sed-
ventilation to reach the preset tidal volume, automatically atives or muscle relaxants; it can automatically adjust the
adjust the preset suction pressure level (usually adjusted to inspiratory pressure according to the monitoring indicators
75% of the calculated value), and send the preset tidal of lung function and meet the preset tidal volume with the
volume with the velocity waveform of deceleration wave and lowest airway pressure, which is conducive to the prevention
the lowest pressure so as to make the actual tidal volume of barotrauma. The waveform of air flow velocity is a de-
consistent with the preset tidal volume and avoid the oc- celeration wave. When the airway is blocked, it can reduce
currence of air pressure injury to the greatest extent [18, 19]. eddy current, so as to reduce pressure consumption and
At the same time, the patient is allowed to breathe freely reduce inspiratory peak pressure [20, 21].
throughout the ventilation cycle. The inspiratory pressure AutoFlow is suitable for some patients with respiratory
level can be automatically adjusted within the range of failure (such as acute respiratory distress) and high airway
5 cm H2O when the end expiratory airway pressure reaches resistance (such as critical asthma). According to the actual
6 Journal of Healthcare Engineering

pressure and increase lung compliance. Peak airway pressure


45 decreases and lung compliance increases with time. Using
Peak airway pressure (cm water column)

the BIPAP mode and the SIMV + AutoFlow breathing mode,


40
there was no significant difference in arterial blood gas, tidal
volume, minute ventilation, and airway resistance.
35
SIMV + AutoFlow is a safe and reliable mechanical venti-
lation mode for patients with congenital heart disease and
30
deformity after correction.
25
Data Availability
20
The data used to support the findings of this paper are
available from the corresponding author upon request.
15

C1 C2 C3 C4 C5 C6 C7 Ethical Approval
Auxiliary white dynamic variable flow In this manuscript, all procedures comply with the inter-
Auxiliary white dynamic flow converter before
national code of ethical conduct for biomedical research
Figure 8: Changes in peak airway pressure before and after involving human subjects (CIOM) and the Declaration of
auxiliary automatic conversion under nonreturn compliance Helsinki. The experimental protocol was approved by the
modification. Institutional Review Committee (IRB) of Cangzhou Central
Hospital.
12.5
Consent
12.0
Auxiliary white dynamic variable flow

All participants provided written informed consent.


11.5

11.0 Conflicts of Interest


10.5 The authors declare that they have no conflicts of interest.
10.0
References
9.5
[1] M Park, H. J Ahn, J. A Kim et al., “Driving pressure during
9.0 thoracic surgery: a randomized clinical trial,” Anesthesiology,
vol. 130, no. 3, pp. 385–393, 2019.
8.5 [2] Y. Wang, S. Shi, Q. Zheng, Y. Jin, and Y. Dai, “Application of
3-dimensional printing technology combined with guide
C1 C2 C3 C4 C5 C6 C7 plates for thoracic spinal tuberculosis,” Medicine, vol. 100,
Auxiliary white dynamic flow converter before no. 6, Article ID e24636, 2021.
Auxiliary white dynamic variable flow [3] J. H. Campos and A. Sharma, “Predictors of hypoxemia
Figure 9: Changes of mean airway pressure before and after during one-lung ventilation in thoracic surgery: is oxygen
Fuchuan automatic conversion under different compliance reserve index (ori) the answer?” Journal of Cardiothoracic and
settings. Vascular Anesthesia, vol. 34, no. 2, pp. 423–425, 2020.
[4] W. Y Ge, R Jing, L. H Pan et al., “Clinical impact of type 2
diabetes mellitus on outcomes after one-lung ventilation
situation of the patient, on the premise of avoiding air during thoracic surgery: a retrospective cohort study,” Annals
pressure injury, the preset inspiratory upper limit pressure of Palliative Medicine, vol. 9, no. 6, pp. 2455–2465, 2020.
level should not be too low; otherwise, it can be difficult to [5] T. S. Hahm, H. Jeong, and H. J. Ahn, “Systemic oxygen
reach the preset tidal volume because the range of inspiratory delivery during one-lung ventilation: comparison between
pressure automatically adjusted by the computer is too small propofol and sevoflurane anaesthesia in a randomised con-
[22]. The data of this group show that there is no significant trolled trial,” Journal of Clinical Medicine, vol. 8, no. 9, p. 1438,
difference in hemodynamic monitoring parameters between 2019.
the observation group and the control group after being [6] S. E. Gedik and T. Alar, “Protective measures undertaken
combined with AutoFlow, indicating that SIMVP combined during chest tube thoracostomy in covid-19 outbreak,” Indian
Journal of Thoracic and Cardiovascular Surgery, vol. 37,
with AutoFlow has little effect on hemodynamics [23–25].
no. 10228, pp. 1–4, 2020.
[7] B. Meng, K. Wu, Y. Wang, S. Zhang, X. Zhou, and Y. Ding,
6. Conclusion “Effect of retrograde autologous priming based on minia-
turized cardiopulmonary bypass in children undergoing open
Compared with BiPAP, the SMV + AutoFlow mechanical heart surgery,” Medicine, vol. 99, no. 5, Article ID e18801,
ventilation mode can significantly reduce peak airway 2020.
Journal of Healthcare Engineering 7

[8] A. Habertheuer, T. Richards, F. Sertic et al., “Stratification risk choledocholithiasis by ercp,” Medicine, vol. 100, no. 8, Article
analysis in bridging patients to lung transplant on ecmo: the ID e24730, 2021.
stable risk score,” The Annals of Thoracic Surgery, vol. 110, [24] L. Xu, X. Li, M. Song, L. Xu, and X. Wu, “Clinical application
no. 4, pp. 1175–1184, 2020. of accelerated rehabilitation surgery in elderly patients with
[9] I. George, M. Salna, S. Kobsa et al., “The rapid transformation colorectal cancer,” Medicine, vol. 99, no. 41, Article ID e22503,
of cardiac surgery practice in the coronavirus disease 2019 2020.
(COVID-19) pandemic: insights and clinical strategies from a [25] L. Lin, M. Sun, Y. Gao, G. Chai, and L. Xie, “Modeling and
center at the epicenter,” The Annals of Thoracic Surgery, clinical application of incision space in facial contour sur-
vol. 110, no. 4, pp. 1108–1118, 2020. gery,” Journal of Craniofacial Surgery, vol. 31, no. 2, p. 1, 2019.
[10] S. Rohde, C. Antonides, M. Dalinghaus, R. Muslem, and
A. Bogers, “Clinical outcomes of paediatric patients supported
by the berlin heart excor: a systematic review,” European
Journal of Cardio-Thoracic Surgery, vol. 56, no. 5, pp. 830–839,
2019.
[11] J. A. Dearani, T. K. Rosengart, M. B. Marshall et al., “In-
corporating innovation and new technology into cardiotho-
racic surgery,” The Annals of Thoracic Surgery, vol. 107, no. 4,
pp. 1267–1274, 2019.
[12] S. M Stokes, N. N Massarweh, J. R Stringham, and
T. K Varghese, “Clinical-pathologic correlation and guideline
concordance in resectable non-small cell lung cancer,” The
Annals of Thoracic Surgery, vol. 108, no. 1, pp. 837–844, 2019.
[13] L. Bertolaccini, E. Prisciandaro, G. Sedda, L. Girelli, and
L. Spaggiari, “89p long-term clinical outcomes and prognostic
factors of upfront surgery as a first-line therapy in patho-
logical n2 nsclc,” Journal of Thoracic Oncology, vol. 16, no. 4,
p. S744, 2021.
[14] P. A. J. Beckers, M. I. M. Versteegh, T. J. Van Brakel et al.,
“Multicenter phase ii clinical trial of isolated lung perfusion in
patients with lung metastases,” The Annals of Thoracic Sur-
gery, vol. 108, no. 1, pp. 167–174, 2019.
[15] H. Miyata, K. Sugimura, M. Motoori et al., “Clinical features
of metastasis from superficial squamous cell carcinoma of the
thoracic esophagus,” Surgery, vol. 166, no. 6, pp. 1033–1040,
2019.
[16] Y. Shimada, Y. Kudo, H. Furumoto et al., “Computed to-
mography histogram approach to predict lymph node me-
tastasis in patients with clinical stage IA lung cancer,” The
Annals of Thoracic Surgery, vol. 108, no. 4, pp. 1021–1028,
2019.
[17] P. Hemmati, H. V. Schaff, J. A. Dearani, R. C. Daly, B. D. Lahr,
and A. Lerman, “Clinical outcomes of surgical unroofing of
myocardial bridging in symptomatic patients,” The Annals of
Thoracic Surgery, vol. 108, no. 1, pp. 452–457, 2019.
[18] N. Ma, R. Lu, D. Zhao et al., “Left atrial appendage fibrosis and
3-year clinical outcomes in atrial fibrillation after endoscopic
ablation: a histologic analysis,” The Annals of Thoracic Sur-
gery, vol. 109, no. 1, pp. 69–76, 2020.
[19] K. Guan, B. Liu, M. Wang et al., “Principles of allergen im-
munotherapy and its clinical application in China: contrasts
and comparisons with the USA,” Clinical Reviews in Allergy
and Immunology, vol. 57, no. 1, pp. 128–143, 2019.
[20] R. He, Z. Wang, W. Shi et al., “Exosomes in hepatocellular
carcinoma microenvironment and their potential clinical
application value,” Biomedicine & Pharmacotherapy, vol. 138,
no. 6, p. 111529, 2021.
[21] C. Montemayor, P. A. R. Brunker, and M. A. Keller, “Banking
with precision,” Current Opinion in Hematology, vol. 26, no. 6,
pp. 480–487, 2019.
[22] J. Pei, X. Zhao, A. S. Patchefsky et al., “Clinical application of
RNA sequencing in sarcoma diagnosis,” Medicine, vol. 98,
no. 25, Article ID e16031, 2019.
[23] Y. Zhang, Z. Gong, and S. Chen, “Clinical application of
enhanced recovery after surgery in the treatment of

You might also like