LOGIQ S8 XDclear 2.

0
with FibroScan VCTE
Quick Card
FibroScan ® VCTE ™ by Echosens is an integrated
module on the LOGIQ™ S8 XDclear™ 2.0 ultrasound
system, located in the console under the operator
panel. There are two probes supported for acquiring
fibrosis measurements – the M+ and XL+ probes.
This guide is meant to assist the user in performing
quick and reproducible measurements.
LOGIQ S8 XDclear 2.0 Quick Card

1. The FibroScan module is located in the console under the operator 3. Acquire an image of the rt lobe of the liver or select an image of the rt
panel. FibroScan probes can be plugged directly into this module in lobe from the clipboard; freeze; activate the FibroScan module on the
either of the available ports. operator panel. The monitor will automatically display a dual screen
with ultrasound image on the left.

2. Patient should be fasting 3 hours prior to exam and be positioned


supine with the right arm abducted over the head with a slight
bend of the torso to the left side to open up the intercostal space
for better access.

4. Place the FibroScan probe in the mid axillary line approximately at the
level of the xiphoid process – holding the probe perpendicular, using
adequate pressure and positioned over the liver.

2
LOGIQ S8 XDclear 2.0 Quick Card

5. Select the correct probe and place on the patient. Look at graphics to 6. Keep the probe perpendicular to the skin surface.
confirm correct pressure and good quality acquisition. Utilize the pre-measurement indicators covered on page 4 to know
when conditions are optimal to acquire measurements. When all the
conditions are met — “3 P rule” — measurement can be triggered:
Position of the probe over liver
Pressure
Perpendicularity

There are two probes available: M+ and XL+. Choose which to use based
on guidelines and the Automatic probe indicator.

7. Press the measure button on the probe to acquire the liver measurement.
Repeat this 10 times. Check the probe perpendicularity along both axes.
Probe recommendation is based on real time SCD (Skin-to-liver Capsule Check the elastograms. The shear wave propagation should be visible
Distance) and operates in real time as soon as the probe detects ultra- throughout the depth of the elastogram. The sides of the black strip
sound signals (probe in contact with patient skin). representing the shear wave must be parallel. Results
are available on screen and in the worksheet.
Note: If the probes are changed during the FibroScan exam, any mea-
surements acquired from previous probe will be discarded.
3
LOGIQ S8 XDclear 2.0 Quick Card

Interpreting the FibroScan Pre-Measure Guidance:


1. Probe/Pressure Indictor: Indicator of adequate probe to skin
pressure and which probe is active.

2. Time Motion Mode: Reflected ultrasound graphic display of depth


over time. Look for homogeneous smooth pattern indicating good 1 6
liver positioning. 4

3. Amplitude Mode: Display of ultrasound signal strength. Looking


for a good consistent signal strength indicating good probe
positioning with adequate gel coupling.
2 3
4. Liver Targeting Tool: When both of these bars are green, this
5
indicates overall good condition, perpendicularity and contact.

5. Shear Wave Propagation Map: Indicates the shear wave


generated in a ratio of how far/how fast.

6. Probe Calibration Indicator: These are seen only when there is


an action needed by the operator. They will indicate the status of
the probe selected for the exam.

4
LOGIQ S8 XDclear 2.0 Quick Card

This is what I got This is what happened What to do to correct

• A Mode Line is not linear Apply more ultrasound gel

• Not enough gel

• TM mode shows presence of vessel Find a different probe position that


produces homogeneous TM mode and
• A mode shows disturbance and linear A mode
is not linear

• Probe is positioned over area of heterogeneous Reposition your probe to achieve a


structures in the liver homogeneous TM mode and linear A
mode

5
LOGIQ S8 XDclear 2.0 Quick Card

This is what I got This is what happened What to do to correct

• Rib bounce called an A wave Find a larger intercostal space

• Example shows over-estimation


of liver stiffness

• Rib bounce called an E wave Find a larger intercostal space

• Will overestimate liver stiffness

• Shear wave must be parallel

• Liver is located too deep below Find a window where the liver is closer to
skin’s surface the skin’s surface
x • Can lead to over-estimation
of stiffness

6
LOGIQ S8 XDclear 2.0 Quick Card

FibroScan Result Window


• Stiffness (E) 1.7-75 kPa
• CAP (dB/m) 100-400 dB/m
Controlled Attenuation Parameter is a measure of signal
attenuation rate
• Median of all valid measurements within the current exam
which is updated in real time after each measurement
• IQR (kPa or dB/m) of all valid measurements within the
current exam
• IQR/med. (%) = IQR/Median Ratio
Must remain < 30% for E only if median E > 7.1 kPa1
Reflects the dispersion of stiffness (E) measurements.
There is no criteria for CAP.
• Number of valid and invalid measurements

1
Bousier et al. Hepatology 2012.
7
LOGIQ S8 XDclear 2.0 Quick Card

FibroScan Probe Warning Indicators


= No Compatible Probe Connected = Probe Calibration
The probe contains tiny mechanical parts that
shift slightly over time. When it is time to
calibrate the probe, this icon will appear in the
information window during the examination.
Operator can continue to perform exams. It is
strongly recommended that you contact GE
Service representative when this icon appears.

= Replace Probe = Electromagnetic Disturbances

Imagination at work
© 2017 General Electric Company – All rights reserved.
GE Healthcare reserves the right to make changes in specifications and features shown herein,
or discontinue the product described at any time without notice or obligation. Contact your GE
Healthcare representative for the most current information. GE, the GE Monogram and LOGIQ
are trademarks of General Electric Company. GE Healthcare, a division of General Electric
Company. FibroScan technology has been developed by Echosens. FibroScan, VCTE and CAP
are trademarks of Echosens (www.echosensclinicallibrary.com).
GE Medical Systems, Inc., doing business as GE Healthcare.
March 2017
DOC1964241

You might also like