LOGIQ S8 XDclear 2.0 FibroScan Quick Card
LOGIQ S8 XDclear 2.0 FibroScan Quick Card
0
with FibroScan VCTE
Quick Card
FibroScan ® VCTE ™ by Echosens is an integrated
module on the LOGIQ™ S8 XDclear™ 2.0 ultrasound
system, located in the console under the operator
panel. There are two probes supported for acquiring
fibrosis measurements – the M+ and XL+ probes.
This guide is meant to assist the user in performing
quick and reproducible measurements.
LOGIQ S8 XDclear 2.0 Quick Card
1. The FibroScan module is located in the console under the operator 3. Acquire an image of the rt lobe of the liver or select an image of the rt
panel. FibroScan probes can be plugged directly into this module in lobe from the clipboard; freeze; activate the FibroScan module on the
either of the available ports. operator panel. The monitor will automatically display a dual screen
with ultrasound image on the left.
4. Place the FibroScan probe in the mid axillary line approximately at the
level of the xiphoid process – holding the probe perpendicular, using
adequate pressure and positioned over the liver.
2
LOGIQ S8 XDclear 2.0 Quick Card
5. Select the correct probe and place on the patient. Look at graphics to 6. Keep the probe perpendicular to the skin surface.
confirm correct pressure and good quality acquisition. Utilize the pre-measurement indicators covered on page 4 to know
when conditions are optimal to acquire measurements. When all the
conditions are met — “3 P rule” — measurement can be triggered:
Position of the probe over liver
Pressure
Perpendicularity
There are two probes available: M+ and XL+. Choose which to use based
on guidelines and the Automatic probe indicator.
7. Press the measure button on the probe to acquire the liver measurement.
Repeat this 10 times. Check the probe perpendicularity along both axes.
Probe recommendation is based on real time SCD (Skin-to-liver Capsule Check the elastograms. The shear wave propagation should be visible
Distance) and operates in real time as soon as the probe detects ultra- throughout the depth of the elastogram. The sides of the black strip
sound signals (probe in contact with patient skin). representing the shear wave must be parallel. Results
are available on screen and in the worksheet.
Note: If the probes are changed during the FibroScan exam, any mea-
surements acquired from previous probe will be discarded.
3
LOGIQ S8 XDclear 2.0 Quick Card
4
LOGIQ S8 XDclear 2.0 Quick Card
5
LOGIQ S8 XDclear 2.0 Quick Card
• Liver is located too deep below Find a window where the liver is closer to
skin’s surface the skin’s surface
x • Can lead to over-estimation
of stiffness
6
LOGIQ S8 XDclear 2.0 Quick Card
1
Bousier et al. Hepatology 2012.
7
LOGIQ S8 XDclear 2.0 Quick Card
Imagination at work
© 2017 General Electric Company – All rights reserved.
GE Healthcare reserves the right to make changes in specifications and features shown herein,
or discontinue the product described at any time without notice or obligation. Contact your GE
Healthcare representative for the most current information. GE, the GE Monogram and LOGIQ
are trademarks of General Electric Company. GE Healthcare, a division of General Electric
Company. FibroScan technology has been developed by Echosens. FibroScan, VCTE and CAP
are trademarks of Echosens (www.echosensclinicallibrary.com).
GE Medical Systems, Inc., doing business as GE Healthcare.
March 2017
DOC1964241