Table of Contents Product functions: Please confirm the function of purchased product Please note that

EW-BW56K011 EW-BW16W011 Warning ※During the measurement process, do not use devices that generate
1 Safety precautions 3 6 Operating
Operating Instructions instructions 14 Hypertension flash ● ● Please do not allow children and people who can't electromagnetic interference near this product, such as mobile
2 Parts identification 6 express their self-awareness to use this device.
6.1 Height sensing prompt Memory number 90 times 60 times phones, as the strong radiation field generated by such devices will
3 How to take a proper While measuring with this device, do not use the interfere with the normal use of the product.
Digital Blood Pressure Monitor function 14 Movement warning ● ● unit on an arm where a catheter hasbeen inserted.
measurement 8 Forbid Otherwise, this product may cause mis-operation or inaccurate
6.2 Method of measurement 16 Irregular pulse ● ● Otherwise, it may cause an accident.
Model NO. EW-BW56K011 EW-BW16W011 4 Precautions in 6.3 Call-up and deletion of Persons with serious obstructions of blood flow in the arm measurement.
Backlight ● —
measurement 9 measured values 18 should consult a doctor before using this instrument. ※In the case of common arrhythmia (such as atrial premature beat,
Comparison of averages ● ● Failure to do so may result in deterioration of physical condition. ventricular premature beat and atrial fibrillation), the measurement
5 Before use 11 6.4 About each instruction 19
Height sensing prompt ● — If the device doesn't work or senses abnormality, please will cause errors.
5.1 Inserting batteries 11 Be sure to
7 Others 22 observe stop using it and send it for inspection and repair. ※The main unit is not likely to cause interference to electronic devices
5.2 Wrapping the cuff 12 Otherwise, there is a risk of fire, electric shock or burns.
7.1 After use 22 around (such as TV, computer, mobile phone).
7.2 Care and maintenance 22 1 Safety precautions Please do not disassemble, repair or transform the ※There may be a risk when the air cuff is in the state of persistent
7.3 FAQ 24 device without permission. hyperinflation. (causing discomfort such as blood stasis or paralysis
7.4 Troubleshooting 27 Do not Otherwise, it may cause an accident or malfunction. in the wrist.) If such abnormality occurs, please remove the cuff.
※ Before use, please read these [Safety Precautions] carefully for proper disassemble
Height
7.5 Specification 29 use. It is recommended to use under the direction of a doctor. ※If the original parts are replaced by parts not provided by the
manufacturer, measurement errors may be caused.
Sensing

※ To prevent users from suffering harm or loss due to improper use,


M/R
please refer to the [Safety Precautions] and use this device correctly. Caution ※Disposal suggestion: The disposal method of the main unit, cuff and
Do not judge the measurement results by yourself or batteries shall be subject to the relevant environmental protection
For safety reasons, be sure to observe these safety precautions. take medicine and other treatments on such basis. regulations of the city.
EW-BW
M/R
Product name: Digital Blood Pressure Monitor •It may cause physical discomfort.
56K011 Model: EW-BW56K011, EW-BW16W011 ※For the potential electromagnetic interference or other interference
•For the measurement results, please consult a professional
Structure and composition: This product is composed of a main unit and indicates the contents of caution and warning. doctor to explain the measured blood pressure value. between this device and other devices, see the manufacturer's
EW-BW •For those with peripheral circulation disorder caused by statement on electromagnetic compatibility for details.
16W01 a cuff Forbid
1 some diseases, the blood pressure measured on the wrist ※The brightness of the screen is often 5-10 s.
Scope of application: This product can measure diastolic blood pressure, Warning: Content that may cause death or serious and upper arm may vary significantly.
systolic blood pressure and pulse rate of human body by oscillometry, injury.
and its values are for diagnostic reference. Please do not use it for purpose other than blood
pressure measurement. Description of markings
The wrist circumference that can be measured by this device is about Caution: Content that may cause personal injury •Otherwise, it may cause an accident or malfunction.
12.5 cm~22.0 cm.
Contraindications: None
and material loss. If the device needs to be repaired for various reasons,
indicates direct current
Be sure to please contact the After-sales Service Office designated
Product object: It is suitable for adults with a wrist circumference of
observe by Panasonic Appliances (China) Co., Ltd. to consult the Caution! Consult the accompanying documents
12.5 cm~22.0 cm. However, users warned or prohibited in the indicates prohibited indicates no maintenance methods or ask for maintenance information.
Thank you for purchasing our products. Operating Instructions and prohibited by doctors are excluded. It is behaviors disassembly Protection against electric shock
Please read these in structions carefully not suitable for newborns. Type B applied part (main unit of the device)
before using this product, and save this Version: NO.01
manual for future use. Issued date: March 2023 1 2 3 4 5

2 Parts identification Note: All diagrams in this specification are for EW-BW56K011 Display ※ This figure is a full display representation 3 How to take a proper measurement 4 Precautions in measurement Measurement instructions
Pulse mark Battery indicator Any blood pressure measurement will be influenced by the posture
Please confirm the product Indication of the detected Relax for 4~5 minutes before measurement and be sure to take For correct measurement, please read the following key points
Main unit Structure and composition: The product is composed pulse during measurement
Indication when the battery is the measurement in a relaxed state. carefully and observe the exact environment and posture. Please of the user and his/her physical condition.
out of power The posture and physical condition of the user will affect the refer to P8 (How to take a proper measurement). For those with poor blood circulation in their hands and feet*, and
of a main unit and a cuff. Systolic blood Diastolic
pressure blood pressure measurement. peripheral circulation disorder, such as those suffering diabetes, liver
blood pressure Measuring time
EW-BW56K011 Cuff Indicates the diastolic blood Systolic blood Diastolic blood
Indicates the diastolic disease, nephropathy, arteriosclerosis, hypertension, etc., the wrist
pressure value blood pressure value Measuring posture and place The blood pressure is constantly changing throughout the day. blood pressure and the upper arm blood pressure may vary significantly.
Positioning label Memory mark Height sensing Therefore, take the period with the least movement after getting
Therefore, you shall not judge by yourself, and must consult a doctor.
When the measured value is up and before sleeping as the measuring time.
Can help to adjust Align the height Wrap the * People who have poor blood circulation in their hands and feet due
Velcro fastener stored, this mark will flash the wrist height Lean on the of the main unit cuff around to basic diseases.
When the measured value /min.
(only EW-BW56K011 Within 1 hour •Before breakfast Before sleep •Relax for
is called up, this mark is back of a chair to the heart. the bare skin after getting up •After urination 4~5 mins
has this function)
Display displayed and relax •Before taking medicine Cuff wrapping and measuring posture
Indicates Pulse rate (in case of taking antihypertensive
during drugs, etc.) Wrapping method
Battery Indicates the measurem Indicates the number of pulses
memory number •Relax for 4~5 mins Do not wrap the cuff The cuff is wrapped too
cover Height
Sensing
Movement warning around your clothes loose
Indication of possible body movement during
Relax your wrist, Prompt of measuring place and measuring time The pressure of the cuff cannot The pressure of the cuff cannot
Average put your arm on
M/R
the measurement Blood pressure is affected by climate, emotion, exercise, sleep time, be transmitted to the blood be transmitted to the blood
Indicates the average of the Irregular pulse warning the table, with food and other conditions. In addition, blood pressure tends to rise
stored measurement value vessels, and the blood pressure vessels, and no proper
Memory/ The change of heart rate can be detected in palm up because of blood vessel contraction due to low outdoor temperature will rise. measurement can be taken.
Call-up button On/Off button the measurement in winter.
Measuring posture
No measurement •Outdoor and cold places
shall be taken in •After exercise or showering Do not lean forward Do not talk
Backlight display case of the right When the abdomen is Talking will raise blood
•Within 1 hour after a meal
Pouch Batteries (Only EW-BW56K011 has this function) circumstances •After drinking stressed, the blood pressure.
Do not move •Shortly after drinking caffeinated pressure will rise.
About backlight function: The device has backlight function. When
your wrist, put beverages or smoking Do not touch the monitor
measuring and calling up the measured value, the backlight displays. Bend inward or shake
2 AAA alkaline your other hand •When feeling restless body
batteries for under your arm When feeling the sign of urination hard It will cause confusion of
※If the display is facing the light, sometimes
test run or defecation So no proper measurement measurement data.
Systolic blood Diastolic blood
the user cannot see the display clearly when Keep your At a can be taken. So no proper measurement
looking from the right angle. Meanwhile, the feet on the
According to the need, sometimes the specifications of products display is facing away from the light, and it is comfortable Measure in a room with stable temperature, and in a non-anxious can be taken.
ground side room and calm state, take five deep breaths, and relax for 4~5 mins before Keep your hands,fingertips,
and parts will be changed without prior statement. easy to see the display from an oblique angle.
by side temperature measurement. wrist and body still Align the height of the cuff
The LCD display of EW-BW56K011 device is As the blood pressure is affected by various conditions, it is difficult Subtle changes in blood
/min.
shown in the right picture, which is displayed to the heart.
to judge whether or not the blood pressure value is true if only one pressure cannot be detected.
in black characters on a bright background. measurement is taken. The height of 10 cm will cause
So no proper measurement an error of 7 mmHg.
It is necessary to measure 2 to 3 times to confirm the change. can be taken.
6 7 8 9 10

5 Before use Caution 6 Operating 6.1 Height sensing prompt function


Appropriate height Inappropriate height
1. Wrap the cuff around the inside of the Positioning label instructions (only EW-BW56K011 has this function)
5.1 Inserting batteries Always follow the cautions printed on the batteries. wrist with the palm up

Main unit height


Check the polarities of the batteries and install them This function is to help you place the device in an appropriate position
correctly. Align the positioning label with the wrist for measurement (roughly parallel to the heart). When the measuring
Take out the batteries with insufficient power immediately. root and wrap the cuff. posture and measuring environment (such as the height of the table,
1. Remove the battery cover. If this device will not be used for a long time, take out the The cuff cannot be bent in the opposite and the height of the chair) cannot provide an appropriate position,
batteries from the device. direction. Velcro the use of this function can lead to a better measurement. The main unit is The main unit is
(Otherwise, the cuff may be broken) Judge the position of the device wrapped on the wrist from the wrist The main unit is approximately higher than your lower than your
When replacing batteries, new batteries of the same type must fastener level with your heart heart heart
2. Check the polarities
Be sure to
be used and all 2 batteries shall be replaced at the same time. angle and remind you of the proper measuring posture through the
wrist position indicator. After the mark in this range is
of the batteries, and install observe
Please use the batteries within the expiration date.
Failure to comply with the above matters may result in injury
2. Adjustment position When measuring Please do not use this function in the following situations and then displayed, it indicates that the
them. use it every time you confirm your heart height. position is high.

Wrist height sensing meter


or damage due to battery heating, cracking or liquid leakage When wrapping the cuff, by the right hand
align the positioning label •Use by family members or many people Need to lower the wrist angle
Appropriate height
3. Install the battery cover Batteries shall be kept out of the reach of children. (If the battery
with the central part of the • Multiple places for measurements ( display)
After the mark in this range is
displayed, it indicates that the
inside of the wrist.
Install it in the opposite direction to 1. is swallowed by mistake, please see a doctor immediately) Height setting position is low.
Otherwise, it may cause an accident or injury. Need to raise the wrist angle
1 In the power-off state, continuously
※Do not open the battery cover during use to prevent electric shock caused press the「M/R」button for more If wrist height sensing is not set, please High
[Battery life]
adjust the height to the heart position
Batteries last for approx. 300 times, which is 3 times/day for 3 months
by touching the batteries. 3. Press the velcro fastener than 3 seconds.
[ ] and [ ] flash.
Height
(standard position) for measurement.
Standard
firmly, with no gap in the middle
sensing
position Low
(when using a pack of new Panasonic alkaline batteries). (Under the When it is not pressed continuously, the
condition that the room temperature is 23°C, 170mmHg pressured, and When the positioning label is misplaced, please adjust the alignment of memory interface will appear.
the circumference of wrist is about 17.0cm.) 5.2 Wrapping the cuff M/R
the main unit. At this time, please press the button
If the ambient temperature is low, the battery life will be shortened. When loosening the velcro fastener, it shall be uncovered parallel to the and start the operation again. Cancel the height setting
As the accompanying batteries are used to confirm the product functions ※Be sure to use a professional cuff front side. Do not pull or drag twistedly by force.
at the time of purchase, their service life may not reach 300 times. (Otherwise, the cuff may break) 2 Align the height of the main unit to the heart 1 In the power-off state, continuously press the「M/R」
The service life of non-alkaline batteries may be very short, so be sure to It can be wrapped on either left or right arm for measurement. Fine-tune the wrist angle so that the wrist is at
the correct height. button for more than 3 seconds.
use alkaline batteries.
Take the measurement of left arm as an example. [ ] and [ ] are displayed. (The height sensing border lights up)
[Battery replacement] The measurement results of the left and right arms may vary to Notice 3 Press the「M/R」button. When it is not pressed continuously, the memory interface will appear.
different extent due to different conditions of each user. Therefore, [ ] and [ ] are displayed. At this time, please press the button and start the operation again.
When the battery is low, the [ ] mark will be displayed. please measure with the same arm every time. Loosely wrapped cuff can easily lead to high blood pressure or failure to The height sensing border lights up.
The batteries can be used before [UO1]. measure the blood pressure. 2 Press the「M/R」button again for more than 3 seconds.
After [UO1], please replace 2 new batteries at the same time.
4 When the setting is completed, press the button. The height sensing border disappears and the setting is cancelled.

Notice In case of changes of the measuring place or the user, 3 When the setting is completed, press the button.
please conduct the above operation again.
11 12 13 14 15
6.2 Method of measurement 6.3 Call-up and deletion of measured values (Example) 6.4 About each instruction Reading of [Blood Pressure Indication]
(comparison example) WHO/ISH blood pressure
Keep calm when measuring, so that your body is in a state of 3 Press the「M/R」button Systolic blood Diastolic blood
Call up the recorded measured value Systolic blood Diastolic blood As the basis of correct measurement, there are movement warning and irregular For the Systolic • Diastolic Blood blood
Systolic classification *1
natural relaxation. again to compare with
Measured and average value for confirmation. pulse warning signals. After the signal is displayed, the correct measured value Pressure value, the WHO/ISH (World pressure Hypertension
1 In the power-off state, press the
Systolic blood Diastolic blood
the average value. value this may not be obtained. Health Organization/International (mild, moderate, severe)
[ ] button. time Average Society of Hypertension) blood
The average value is to
average all the recorded
1 In the power-off state, press After the measurement, re-measurement is recommended.
(Please rest for 4~5 mins before re-measurement) pressure classification table shall
Normal high
EW-BW56K011 Average
the「M/R」button to call up /min.
be used as a benchmark to judge Normal
measured values. If the measured values are recorded, the signals are also recorded together.
•After the display is fully lit, the value (daily
the measured value to be whether it is within the hypertension
confirmed.
/min.
Systolic blood Diastolic blood
height sensing is displayed. blood pressure value) range or a normal range, and the user Diastolic
Please refer to P14 for the appropriate •Every time it is pressed, the Movement warning
Latest /min.
is prompted by a chart. blood pressure
display position new measured value will be measured During measurement, in case of hand or
After the wrist is adjusted to the proper Under inflation
(during measurement) (example)
4 After pressing the button,turn it off and take off the cuff. displayed. (When recorded value wrist movement, and when the cuff detects excess pressure, the
•If the button is not pressed, it will automatically turn only once, only this measured mark will flash.
position, it will be pressurized for /min.

off after 30 seconds. value will be displayed) Systolic blood Diastolic blood
When the measured value is within the
measurement automatically. (Even when After measurement, the mark flashes and displays.
it is not in the appropriate position, 2 Press the button to turn Previous new When a large abnormal movement is detected during measurement, it hypertension range, the value will flash.
automatic inflation starts after about Notice off the power supply. measured
/min.
will prompt with [U12] and [ ], and automatically stop detection. (for about 6 seconds)
7 seconds) value •Systolic blood pressure:
Flashing of When the and marks are displayed during measurement or at the •If the button is not
mark when pulse pressed, it will automatically above 140 mmHg (18.7 kPa)
EW-BW16W011 end of the measurement
Number of records (No.)
/min.

is detected P19 [About each instruction] turn off after 30 seconds. Irregular pulse warning •Diastolic blood pressure:
After the display is fully lit, the cuff will be /min.
above 90 mmHg (12.0 kPa)
When recorded only once, the average value is expressed as [ ]
pressurized for measurement automatically. When deleting a measured value Measure the significant abnormal change
When the maximum times are exceeded, the first measured values will be
Note: When the pressurization value is more End of measurement (example) deleted in order to record new measured values. You can delete all recorded measured values of pulse spacing. After measurement, the [ ] mark flashes *1 1999 World Health Organization-International Society
than 280 mmHg, the cuff will automatically Systolic blood Diastolic blood
Press the button for about 5 seconds, to switch measuring methods (the average value will also be deleted) and then displays. of Hypertension Guidelines for the Management of
exhaust. of mmHg and kPa. It is mmHg by factory setting. (7.5 mmHg=1 kPa) This mark is not to inform arrhythmia. Hypertension
They cannot be deleted one by one *This function is a reference function set to ensure the accuracy of the
The measured value [ ] flashes. When
recording the measured value, please go Suggestions ① In the power-off state, press the「M/R」button. measured value. Even if the [ ] mark is not displayed, in order to
•Display the average value ensure the accuracy of the measurement, you are recommended to
to step 2 during the flashing process. Compared with the first measurement, the second measurement is more
Flash /min. measure for 2~3 times in a relaxed state.
② Long press the「M/R」button
2 Press the「M/R」button to stable and the blood pressure will also drop. Especially for people who are
(for more than 3 seconds)
*If the mark appears frequently, you should consult your doctor about
store the measured value. Display measured value easily to get nervous, the second measurement is more accurate. During your health.
•[ ] is displayed. the second measurement, the measured value with smaller value is usually •Display content flashing
Please do not judge the measurement results by yourself and take
Number of times that the measured value can be recorded: recorded. For the second measurement, please relax for 4~5 mins after the ③ When flashing, press the「M/R」
any treatment on such basis. Please follow the doctor's advice.
EW-BW56K011 90 times first measurement button again and the flash
EW-BW16W011 60 times disappears. *In order to prevent the confusion between the name of the product
When the [ ] flashes, the measured value cannot be ④ Press the button to turn off auxiliary function and “Arrhythmia”, adjust “Arrhythmia warning” to
recorded if the「M/R」button is not pressed. the power supply. “Irregular pulse warning”
16 17 18 19 20

7 Others 7.1 After use Product care 7.3 FAQ Questions Answers Questions Answers
Why is the blood The blood pressure is constantly changing even After A prompt that an irregular pulse is detected
Storage of the blood pressure monitor Please do not repair, disassemble and transform the device by yourself. Questions Answers in one day, and it will also vary due to the cuff measurement during the measurement.
pressure value
Do not drop the main unit on the ground to avoid collision or strong impact. Why does the Whether the height of the main unit is at the same measured wrapping method and measuring posture. [ ] mark is Even if an irregular pulse is not displayed.
(Otherwise, it will cause malfunction) systolic or diastolic level as the heart. different every Therefore, please conduct each measurement displayed → Please relax for 4~5 mins before re-measuring.
1 Remove the main unit from Please do not take the device from a place below 0 °C and use it immediately. blood pressure → When the main unit is lower than the heart, the
measured blood pressure will be higher.
time? under the same conditions.
The blood pressure of people on antihypertensive During The hand and wrist move during the measurement.
the wrist and wrap the cuff Keep it in a warm environment for at least 1 hour. value fluctuate?
The cuff is too loose. drugs is sometimes subject to great changes due measurement → After measurement, the [ ] mark is displayed.
as shown in the figure. (Otherwise, it will cause inflation failure)
Avoid storing the device in an environment with high temperature, humidity → Wrap the cuff around your wrist. to their drug effects. [ ] mark Please relax for 4~5 mins before re-measuring.
Whether you move or talk during the measurement. When 2 consecutive measurements are made, the flashes Even if the hand or wrist does not move during
or direct sunlight.
Whether you feel uneasy or anxious. user must relax for 4~5 mins in between. measurement, there will be a prompt of body
2 Put it in the pouch. (Otherwise, it will cause malfunction)
Do not bend the cuff in the opposite direction. The cuff is only used to wrap
→ Please take a deep breath for 5~6 times before Why are the Both the left and right wrists can be measured. movement when the hand is exerting force.
the measurement, so as to relax your body and measured values However, the measurement results of the left and
the wrist for blood pressure measurement. stabilize your blood pressure. Please relax and then re-measure.
of the left and right wrists may vary to different extent due to
Do not apply excessive force on the product or leave it all over (Otherwise, the cuff may be broken) In addition, please relax for 4~5 mins before right wrists different conditions of each user. Therefore, please If the problems are not solved after confirmation, please stop using the
the place. (Otherwise, it will cause malfunction) Cannot be contaminated with foreign matter or corrosive liquid. measuring. different? measure with the same wrist every time. device immediately and consult a doctor.
Please do not touch the button when putting it in the pouch. (Otherwise, it will cause malfunction)
If so, the batteries will run out quickly. Why is the blood The blood pressure changes during 24 hours. In addition,due When the hand or It is a function that indicates that the hand or wrist
Do not allow dust or foreign matter to enter.
pressure measured to climate, emotion, exercise and other reasons, sometimes the wrist is moved, is bent, or the cuff exerts excessive pressure during
7.2 Care and maintenance (Otherwise, it will cause malfunction) it may be subject to great changes. Especially in hospitals,
in hospital different but the [ ] measurement.
Take out the batteries when the device is not in use for a long time. there is the so-called [white-coat hypertension], which is that mark is not Movement with no influence on the cuff (like gentle
from that measured
(for more than 1 month) (otherwise, it will cause liquid leakage and damage the blood pressure measured in the hospital is sometimes displayed? elbow bending) may not be prompted.
Cleaning frequency/maintenance cycle at home?
to components). higher than that measured at home because of nervousness If you have any doubt about the measurement,
Be sure to cut off the power supply. or uneasiness. please re-measure.
When the blood pressure measured at home is higher than
1. Cleaning of main unit According to the The [ ] warning is only a reference prompt for
that in the hospital, please pay attention to the following
Visually confirm the appearance of the main unit. When there is dirt, for cleaning, [ ] prompt, measuring whether the posture is correct or not.
items:
take the batteries out of the unit and then gently wipe it with cloth that is soaked when the hand After the [ ] warning mark is displayed, if the
Whether the height of the main unit is at the same
in soapy water or warm water then wrung; and wrist are measured value is still high or low, please re-measure.
level as the heart.
2. Cleaning of cuff placed in the In case of the above case, it may be due to the
→ When the main unit is lower than the heart, the
When there is dirt on the cuff, please gently wipe it with cloth that is soaked in following factors.
soapy water or warm water then wrung; measured blood pressure will be higher. correct position
The cuff is too loose. Whether a proper measurement is taken
3. To prevent infection, the cuff shall be cleaned regularly for measurement, (wrapping method of cuff, posture, etc.)
4. Cleaning precautions → Wrap the cuff around your wrist. will the measured It is measured directly after smoking
Before cleaning the device, please take the batteries out; Whether you feel uneasy or anxious. value be higher It is measured when emotionally unstable
Do not wash the device or cuff with water; → Please take a deep breath for 5~6 times before the than usual? It is measured in cold environment
Please keep the device away from dust or water vapor; measurement, so as to relax your body and stabilize your Constitution with small pulse change
Do not try to clean the cuff by washing machine or rubbing; blood pressure. → If you have any doubt about the measured value,
Do not use alcohol, gasoline or thinner for cleaning, otherwise it will cause In addition, please relax for 4~5 mins before measuring. please re-measure after relaxing for 4~5 mins.
cracking or discoloration.
21 22 23 24 25

7.4 Troubleshooting Failures Cause and solution 7.5 Specifications


[Production date] See the label
List of products and accessories:
Please confirm before consulting or repairing Power supply DC3V 2 AAA alkaline batteries (LR03)
Low battery. EW-BW56K011 EW-BW16W011
Failures Cause and solution Display Digital LCD ※ If stored and used outside the specified temperature or humidity range,
Please refer to P11 Main unit 1 set 1 set
Method of the specified performance specifications may not be met.
Even if the The polarities of the batteries are installed measurement Oscillographic system Cuff 1 pc 1 pc
On/Off button ※ Product specifications are subject to change without further notice.
incorrectly. Please refer to P11 Static pressure measurement range: (0.0~37.3) kPa ((0 ~ 280) mmHg) Operating Instructions 1 copy 1 copy
is pressed, there Battery power is low. Please refer to P11 Pressurization to more than 280 mmHg. Pulse measurement range: (30~160) beats/min
is no display Is the measuring posture correct? (P8,P10) Measurement Measurement range of blood pressure: Caution: Test-run batteries 2 pcs 2 pcs
range If the device needs to be repaired for various reasons, please contact
Continuous Please stop using it, and entrust a sales store to Pressure decreased suddenly. Systolic blood pressure shall be at least (8.0~29.3) kPa ((60~220) mmHg) Pouch 1 pc 1 pc
Is the measurement method or posture correct? Diastolic blood pressure shall be at least (5.3~17.3) kPa ((40~130) mmHg) the After-sales Service to consult the maintenance methods or ask for
inflation inspect and repair it.
(P8,P10) Pressure: within ±0.4 kPa (±3 mmHg) maintenance information. ※ The accompanying batteries are for test-run. They are used to
Accuracy
Battery power is low. Is the cuff wrapped correctly? (P13) Pulse rate: within ±5% confirm the function of the device at the time of purchase and
is displayed
Please prepare new batteries (P11) Please relax for 4~5 mins. Do not shake you Normal working Temperature: 5 °C ~ 40 °C
environment their service life will be short.
hands, fingers, wrists, etc. and measure again. Relative humidity: 15% ~ 80%
The display The pulse cannot be detected correctly. (The Storage and
disappears The battery has been used up. transportation Temperature: -20°C ~ 55°C Relative humidity: no more than 93%
Please refer to P11 mark flashes several times or does not flash at all.) environment
during inflation Not inflate or inflate but failed to measure properly. Atmospheric
Is the cuff wrapped correctly? (P13) pressure for 80 kPa ~ 105 kPa
Please do not repair it yourself, and be sure to operation
Main unit
entrust the service shop designated by Panasonic During measurement, the measured values vary Atmospheric
damage pressure for
for inspection and repair. significantly. transportation 50 kPa ~ 105 kPa
(Large range of movement, many times of and storage
movements, large wrist or body movements, etc.) Measurable
Please relax for 4~5 mins. Do not shake your hands, circumference Approx. 12.5cm ~ 22.0cm
of wrist
fingers, wrists, etc. and measure again. (P16)
Weight EW-BW56K011: Approx. 116 g (with cuff, excluding batteries)
Press the Memory/Call-up button during measurement.
Avoid pressing this button during measurement. EW-BW16W011: Approx.113 g (with cuff, excluding batteries)
Dimensions L(89mm) x W(64mm )x H(34mm)
Main unit malfunction. (main unit)
Please do not repair it yourself, and be sure to Protection
against electric Internal power supply equipment (Type B applied part) Panasonic Appliances (China)Co., Ltd
entrust the service shop designated by Panasonic
for inspection and repair. shock ©Panasonic Appliances (China)Co., Ltd Issued date: 2023.3.1
Safety Equipment inapplicable to usage in flammable anesthetic gas mixed with Printed in China W9030BW56101
classification air or flammable anesthetic gas mixed with oxygen or nitrous oxide

26 27 28 29 30