Professional Documents
Culture Documents
CT Scanner_SternMed
CT Scanner_SternMed
CT Scanner_SternMed
www.sternmed.de
2 © SternMed
WORLD LEADING GANTRY PERFECT X-RAY GENERATOR
TECHNOLOGY SYSTEM
• Higher Rotation Speed • Power of Generator 60 kW
Cytom 16
Multi slice ultra fast CT scanner
© SternMed 3
NON-CONTACT SLIP-RING DESIGN
ADVANTAGE OF AUTO-DOSE
TECHNOLOGY OF LOW-DOSE
4 © SternMed
OUTSTANDING CLINICAL IMAGES
© SternMed 5
APPLICATION PACKAGES
CT rib analysis
Auto-analysis of the rib. One-key segmentation of the target
rib, display 3D view and CPR. Full analysis of the rib makes the
diagnosis quite efficient.
6 © SternMed
TECHNICAL SPECIFICATIONS
Cytom 16 | SternMed multi-slice ultra fast CT Scanner
DATA ACQUISITION SYSTEM SCAN TIME
Detector Spec 21504 UNIT/24 ROWS Scanning time range Axial : 0.5 s , 0.8 s , 1 s , 2 s
Scan vision 500 mm Spiral : 0.5 s , 0.8 s , 1 s
Scanning Type Axial, Dynamic, Spiral, Positioning Max. Continous > 120 s
Reconstruction Matrix 512 x 512 Scanning time
X-RAY HIGH VOLTAGE GEN. Image reconstruction time 40 images/s
Max. output power 60 kW SLICE THICKNESS
Max. Output mA 500 mA Axial & Spiral Scan 0.625, 1.25, 2.5, 5.0, 10 mm
High Voltage Output Range 80 , 100 , 120 , 140 kV Deviation of measured and > 2 mm Thickness: ± 1.0 mm
X-RAY TUBE COMPONENT Nominal < 2 mm Thickness : ± 50%
Anode thermal capacity 5.0 MHU GANTRY
Heat Dissipation Rate 815 kHU/min Gantry tilt Front & Back tilt function
Focus size S: 0.5 x 1.0 mm Tilt Angle indicator
L: 1.0 x 1.0 mm Tilt angle capability ± 30° - Deviation < ± 2°
Tube Focus Technology Dynamic H & V position laser light ind.
COLOR IMAGE DISPLAY Laser Positioning Accuracy ± 1 mm
Monitor 24“ Tube Position for Scanning 0° , 90° , 180° and 270°
Monitor Resolution 1920 x 1200 Gantry Aperture >700 mm
Image Matrix 512 x 512 (Openning Diameter)
SPIRAL SCANNING PARAMETERS EXAMINATION TABLE
Helical Pitches 0.5, 0.75, 1.0, 1.5 Highest Adjusment range > 350 mm
Continus Spiral Scan time 120 s Horizantal movement range > 1600 mm
Scanning range > 160 cm Scanning range > 1600 mm
IMAGE STORAGE AND ARCHIVING Step Accuracy < ± 0.25 mm
DICOM DICOM 3.0 Table Load < 204 kg
WORKING & ENVIRONMENT CONDITIONS Operating Noise < 70 dB
Exam room temperature 20 ~ 26 °C X-RAY GENERATOR DEVICE
Control room temperature 18 ~ 28 °C Voltage Regulation range 80, 100, 120, & 140 kv
Exam room humidity 30 ~ 70 % and accuracy four gears
Control room humidity 20 - 80% Voltage Deviation ± 5%
Atmospheric pressure 700 ~ 1060 hPa
POWER CONDITIONS Current adjusment range 10 mA ~ 500 mA
Nominal Voltage 3 N ~ 380 V and accuracy
Nominal frequency 50 Hz Increment steps 10 mA
Power Capacity 115 kVA Current Deviation ± 20 %
SYSTEM PERFORMANCE Max. Scan layer / single scan 16
SOFTWARE FUNCTIONS
Image Noise > 0.29% ( 28 mGy )
Water Uniformity of CT ± 3 HU • Check Patients
Accuracy of CT Air: -1000 HU ± 10 HU
• Patient Information Management
Water: ± 3HU
• Image viewing and analysis
• Image type setting and printing
High Contrast Resolution 17.51 p/cm @ 0 % MTF • DICOM communucation interface
131 p/cm @ 10 % MTF • Scanning Protocol management
101 p/cm @ 50 % MTF • Task queue management
Lower Contrast Resolution 3.0 mm @ 0.3% (30 mGy) • Enhanced Scanning
Artifacts No in phantom CT Image • Noise Reduction Package
• Automatic speech function
• Log Management
© SternMed 7
SternMed GmbH
Schubertstrasse 31 © SternMed GmbH – All rights reserved.
88214 Ravensburg - Germany
SternMed GmbH reserves the right to make
Tel: +49 751 35 978 0 changes in specifications and features shown
Fax: +49 751 35 978 105 herein, or discontinue the product described
at any time without notice or obligation. Con-
email@sternmed.de tact your SternMed Representative for the
most current information.
Rev. V. 4.0