Operating Manual

Handpiece D-ACTOR

Part No. 23213_xxxx


Part No. 21700_xxxx
Published: May 2014
Original language: German

Publisher:
STORZ MEDICAL AG
Lohstampfestr. 8
CH-8274 Tägerwilen
Switzerland

24887
Table of Contents

1 System description 6

1.1 Handpiece 6

1.2 Transmitters 7

2 Installation Instructions 8

2.1 Unpacking 8

2.2 Scope of supply 8

2.3 Connecting handpiece 8

2.4 Compatibility 9

3 Operation 10

3.1 Start-up 10

3.2 Function overview 11

3.3 Standard display 12

2 3.3.1 Setting the pressure . . . . . . . . . . . . . . . . . . 12


3.3.2 Setting the frequency . . . . . . . . . . . . . . . . . . 13
3.3.3 Resetting the pulse counter . . . . . . . . . . . . . . . . 13
3.3.4 Changing to main menu . . . . . . . . . . . . . . . . . 13

3.4 Main menu 14


3.4.1 Changing to the options of the main menu . . . . . . . . . . . 14
3.4.2 Menu program selection . . . . . . . . . . . . . . . . . 15
3.4.2.1 Indication mode . . . . . . . . . . . . . . . . . . . 15
3.4.3 Menu nominal value of pulses . . . . . . . . . . . . . . . 18
3.4.4 Menu Information . . . . . . . . . . . . . . . . . . . 19
3.4.4.1 Reset revision pulse counter . . . . . . . . . . . . . . . . 19
3.4.5 Changing to standard display . . . . . . . . . . . . . . . 20

3.5 Functional checks 20

3.6 Standard settings 20

3.7 Treatment 21
Table of Contents

3.7.1 Safety information. . . . . . . . . . . . . . . . . . . 21


3.7.2 Setting parameters . . . . . . . . . . . . . . . . . . 22
3.7.3 Coupling the handpiece . . . . . . . . . . . . . . . . . 22
3.7.4 Triggering pulses . . . . . . . . . . . . . . . . . . . 22

19 920 02 0614
4 Cleaning, Maintenance, Overhaul 23

4.1 Cleaning 23
4.1.1 Changing the transmitters . . . . . . . . . . . . . . . . 24
4.1.2 Cleaning the handpiece . . . . . . . . . . . . . . . . . 25
4.1.3 Cleaning the transmitters . . . . . . . . . . . . . . . . 27

4.2 Overhaul 28
4.2.1 Contents of the D-ACTOR overhaul kit . . . . . . . . . . . . 28
4.2.2 Overhauling the handpiece . . . . . . . . . . . . . . . . 29

4.3 Maintenance 33

4.4 Disposal 33

4.5 Repair 33

4.6 Service life 33

5 Error messages and trouble-shooting 34

6 Accessories and Spare Parts 35

7.1 Symbols and labels 36 3

8 Warranty and Service 37

8.1 Warranty of the D-ACTOR handpiece 37

8.2 Service 37

Table of Contents

19 920 02 0614
Preface
Warning notes

in accordance with liability regulations.


DANGER refers to a situation of acute danger which, if not avoided, could lead to
serious or fatal injury.

DANGER!

The source of the danger is stated here.


These are the possible consequences!
The instructions for avoiding the danger are given here.

WARNING refers to a situation of potential danger which, if not avoided, could lead
to serious injury.

WARNING!

The source of the danger is stated here.


These are the possible consequences!
The instructions for avoiding the danger are given here.

4
CAUTION indicates that incorrect operation could lead to minor injuries.

CAUTION!

The source of the danger is stated here.


These are the possible consequences!
The instructions for avoiding the danger are given here.

ATTENTION indicates that incorrect operation could lead to damage to the device.

ATTENTION!

The source of the danger is stated here.


These are the possible consequences!
The instructions for avoiding the danger are given here.
Preface

19 920 02 0614
Other instructions

NOTE

preceded by the term 'NOTE'.

Instructions for safe use

WARNING !

Incorrect handling of the device.


Possibility of injuries to the person to be treated and the
operating personnel!
Read this operating manual carefully before you start using the
handpiece D-ACTOR.
Read also the separate operating manual for your control device
completely.

For all regulations and instructions for the protection of the person to be treated and
for the use of the control device and the handpiece please read the general Safety
inStrunctiOnS of the operating manual of your control device.

Preface

19 920 02 0614
1 System description

1.1 Handpiece
Radial, low-energy pulses (D-ACTOR) are administered into the body by means of
a freely movable handpiece, and encompass the entire pain zone. Radial pulses are
often also referred to as radial pressure waves – which is indeed physically correct. The
handpiece is placed on the pain zone that has been established by diagnosis.

1 2 3 4

1 Transmitter
2 Transmitter screw
cap
3 Shaft cushion
6 4 Display
5 Handpiece handle
5
6 Handpiece shaft

Fig. 1-1 Handpiece D-ACTOR

The D-ACTOR handpiece contains all the control and display elements required for
operating the device as a standalone unit (see chapter 3 OperatiOn).
6
The D-ACTOR handpiece is also updated immediately after it is connected when there
has been a software update of the system.

NOTE

Pictures of handpiece and pulse transmitters are examples. Individual components


may differ from those shown in the illustration.
System description

19 920 02 0614
1.2 Transmitters
The handpiece is used with different transmitters.
Each of these transmitters generates its own characteristic effect (see user
information).

1 C15, Ø 15 mm
2 DI15, Ø 15 mm
3 F15, Ø 15 mm
4 D20-S, Ø 20 mm
5 D35-S, Ø 35 mm
1 2 6 D20-T, Ø 20 mm

3 4

5 6

Fig. 1-2 Transmitters 7

Replace the transmitter as described in 4.1.1 changing the tranSMitterS.

System description

19 920 02 0614
2 Installation Instructions

2.1 Unpacking
Remove the instrument and accessories from the packaging container. Proceed
with extreme caution.

Check that all items are included in the packaging container and that they are not
damaged.

Contact your supplier or the manufacturer immediately if any items are missing or
damaged.

2.2 Scope of supply


The standard scope of supply of the D-ACTOR handpiece includes the following items:
– Handpiece
– Transmitter
– Open-end spanner
– Hexagonal spanner 2.5 mm
– Projectile (2 pcs.)
– Guide tube
– Operating manual
8 – Case with foam insert
Optional accessories, see chap. 6 acceSSOrieS and Spare partS

2.3 Connecting handpiece


The control device has a connector for the handpiece D-ACTOR. It is located on the
rear side of the device.

1
1

2
Installation instructions

1 red dots
1 connector for the D-ACTOR handpiece 2 outside of the plug body

Fig. 2-1 Fig. 2-2 Connecting the handpiece

19 920 02 0614
Insert the plug of the handpiece into the handpiece connector on the control
device.

While inserting, make sure that the red dots on the plug and connector are lined
up above one another.

Gently pushing the plug into the connector will immediately lock the connection
and will prevent the connector from disengaging automatically when the cable is
pulled.

Place the handpiece into the handpiece holder.

locking function, allowing the plug to be pulled out of the handpiece connector.

2.4 Compatibility
The STORZ MEDICAL D-ACTOR handpiece is allowed to be operated on the following
control devices:
– STORZ MEDICAL D-ACTOR 50
– STORZ MEDICAL D-ACTOR 100
– STORZ MEDICAL D-ACTOR 200
– STORZ MEDICAL CELLACTOR SC1 BT
– STORZ MEDICAL D-100

Installation instructions

19 920 02 0614
3 Operation
The handpiece operates using the embedded display. The settings can be adjusted
manually or pre-programmed settings can be adopted.

On the display can be changed into the following modes:


– Standard display
– Main menu

WARNING !

Malfunction of the device or its components.


Various injuries are possible!
Before starting treatment, it is essential to perform the functiOnal
checkS described in chapter 3.5.

3.1 Start-up
NOTE
To start up the control device, please refer to the separate operating manual for
your control device.

Connect the D-ACTOR handpiece to the control device (see chapter 2.3 cOnnecting
10
handpiece).

Switch on the control device.

NOTE
If the control unit is changed to which the handpiece is connected, so an adap-
tion process starts. This process takes about 2 minutes and will be visible on the
display of the handpiece.
Operation

19 920 02 0614
Operation
11

Standard display Setting pressure


Setting frequency
Resetting shock counter
D-ACTOR Display
Main menu Menu: Program selection Ref.no. 1 - Indication 1 Treatment step 1
Treatment step 2
Ref.no. 2 - Indication 2 Treatment step 1
Treatment step 2
Ref.no. n - Indication n Treatment step 1
Treatment step 2
Menu: nominal value of pulses setting nominal value of pulses
Function overview

Fig. 3-1 Function overview


Info 1 - Reset
Menu: Information revision shock counter
Info 2 - Total puls counter - HP
Info 3 - Total operating hours - HP
Info 4 - Total puls counter - CD
Info 5 - Total operating hours - CD
Info 6 - Software version - HP

CD = control device
Info 7 - Software version - CD

HP handpiece

19 920 02 0614
Info 8 - Hardware number - HP
Info 9 - Hardware number - CD
Info 10 - Boot load number - HP
3.2 Info 11 - Boot load number - CD
3.3 Standard display
After switching on the display shows the last set values.

Press the trigger button 5 once to get to the standard display.

– Now you can set the energy level and the frequency, reset the treatment puls
counter, change to the main menu and release puls waves.

1 2 1 Reduce pressure
2 Increase pressure
6
3 Reduce frequency
4 Increase frequency
7 5 Trigger button
10 11
8 6 Display pressure
7 Nominal pulse
9 counter
3 8 pulse counter
4
9 Display frequency
10 Key combination with
12 double function
(see text)
11 Key combination
5
(change to the menu)

Fig. 3-1 Standard display

3.3.1 Setting the pressure


The pressure can be selected in steps from 1,0 to 5,0 bar.

Set the pressure by pressing the buttons 1 (reduce pressure) and 2


(increase pressure).
– The selected nominal value of the pressure (6) is shown on the display.

By pressing the particular button once the pressure will be increased or reduced in
single steps.

By pressing and holding the particular button after the second step the pressure will
be regulated faster.
Operation

19 920 02 0614
3.3.2 Setting the frequency
The pulse frequency can be selected in steps from 1,0 Hz to 21 Hz.

Set the frequency by pressing the buttons 3 (reduce frequency) and 4


(increase frequency).
– The selected nominal value of the frequency (9) is shown on the display.

By pressing the particular button once the frequency will be increased or reduced in
single steps.

By pressing and holding the particular button after the second step the frequency will
be regulated faster.

3.3.3 Resetting the pulse counter


For the standard display and for the menu the key combination 10 has different func-
tions.

Press the key combination 10 (buttons 1 and 3 ) simultaneously to


reset the pulse counter to „0“ while you are in the standard display.

3.3.4 Changing to main menu

Press the key combination 11 (buttons 2 and 4 ) simultaneously to


change to the main menu, while you are in the standard display.
13
Standard display Main menu

= 11

11 Key combination (2+4)


(Change to main menu)

Fig. 3-2 Changing to main menu


Operation

19 920 02 0614
3.4 Main menu
From the main menu the operator can change into the following options:
– Program selection
– nominal value of pulses
– Informations

3.4.1 Changing to the options of the main menu


Press button 1 to show the previous option.

Press button 2 to show the next option.

Press button 3 to change from menu to the corresponding option in the main
menu.

Main menu Main menu Main menu


Program selection Nominal value SW Informations

1 2 1 2 1 2

14

4 4 4

Program selection Nominal value SW Informations

3 3 3
Operation

Fig. 3-3 Main menu with options

19 920 02 0614
3.4.2 Menu program selection
With the menu program selection a reference number can be selected, which is
related to an indication.

parameters for the pressure, frequency, nominal value of the pulses and the applicator
to be used for this application in your application brochure.
Press button 1 to show the previous reference number.

Press button 2 to show the next reference number.

reach the indication mode.

Press button 3 to change from menu to the option program selection in the main
menu.

1 Previous reference number


2 Next reference number
3 back to program selection
1 2 4
5 Display - reference number
5

3 4

15

Fig. 3-4 Menu program selection

3.4.2.1 Indication mode


Over main menu, menu program selection and reference number you reach indication
mode.

pulses as well as the proposed applicator are shown. You can adopt them or change
them.

Menu program selection

reach the indication mode starting with treatment step 1.


Operation

19 920 02 0614
Selecting treatment step
You can choose between two treatment steps.
Press button 1 to show the previous treatment step.

Press button 2 to show the next treatment step.

of the parameters will be stored.

Press button 3 to change from indication mode to the option program selection in
the main menu.

Usually begin with treatment step 1.

Display parameters

You can press button 1 (reduce pressure) and 2 (increase pressure) to


change the pressure.

You can press button 3 (reduce frequency) and 4 (increase frequency)


to change the frequency.

Press the trigger button 5 to begin with treatment.

Treatment will be performed.

After the nominal value of pulses from treatment step 1 is reached treatment stops
automatically. The display returns to the level treatment steps however to treatment
step 2.
16
After the nominal value of pulses from treatment step 2 is reached treatment stops
automatically. The display automatically returns to the standard display (see chapter
3.3 Standard diSplay).
Operation

19 920 02 0614
Program selection

10 key combination (1+3)


10 (change to standard display)
11 key combination (2+4)
(change to main menu)

Treatment step 1 Treatment step 2

1 2 1 2

10 11 10 11

3 4 3 4

3 back to program
selection
17

1 2 1 2
1 Reduce pressure
10 11 10 11
2 Increase pressure
3 3 Reduce frequency
3 4 4
4 In crease frequency
5 Trigger button
5 (more functions
see. text)

Fig. 3-5 Indication mode


Operation

19 920 02 0614
-
quency, nominal value of pulses and the applicator to be used for this application
which are recommended by users.

In the menu parameter selection the display changes between the selected reference
number with treatment step and the value of the pulse counter every 2 sec.

display changes every


2 sec

Fig. 3-6 Automatic chance of the display of the parameter

3.4.3 Menu nominal value of pulses


With the menu nominal valve of pulses the number of shots can be set after which
the handpiece stops release of pulses automatically.
18

Press button 1 to reduce the value of pulses in steps of 50 shots.

Press button 2 to increase the value of pulses in steps of 50 shots.

to the option nominal value of pulses in the main menu. (see Fig. 3 - 7).

Press button 4 to cancel the selection and return to the menu nominal value of
pulses (see Fig. 3 - 7).

1 reduce nominal value of pulses


2 increase nominal value of pulses
3 back to nominal value of pulses
1 2 4
5 Display - nominal value of pulses
5

3 4
Operation

Fig. 3-7 Menu nominal value of pulses

19 920 02 0614
3.4.4 Menu Information
Press button 1 to show the previous info menu point.

Press button 2 to show the next info menu point.

Press button 3 to change to the option information in the main menu.

1 previous information number


2 next information number
3 bach to information
1 2 4 reset button (for pulse counter
- information number 1)
5 5 information number
6 6
3 4

Fig. 3-8 Menu Information

In the menu information the following information can be read out and the pulse
counter can be resetted
1 Reset pulse counter (see chapter 3.3.3 reSetting the pulSe cOunter)
2 Total pulse counter of the handpiece
3 Operating hours of the handpiece
19
4 Total pulse counter of the device
5 Operating hours of the device
6 Software version - handpiece
7 Software version - device
8 Hardware number - handpiece
9 Hardware number - device
10 Boot software number - handpiece
11 Boot software number - device

3.4.4.1 Reset revision pulse counter

Under info point number in the menu information the pulse counter is displayed.
The display shows the pulse number.
Press button 4 and afterwords simultaneously the buttons 1 and 3 to reset the
revision pulse counter.
Operation

19 920 02 0614
3.4.5 Changing to standard display

Press the key combination 10 (buttons 1 and 3 ) simultaneously while


you are in the main menu or any other menu to change to the standard display.

Changing to standard display is possible from any view of the display.


Exception: pulse release is active.

3.5 Functional checks


Perform the following functional checks after the system has been installed:
Check the D-ACTOR module and the handpiece for any signs of damage.

Put the handpiece into operation.

Set the energy level to 2 bar.

Reset the treatment pulse counter. on the display of the handpiece.

Release individual pulses in single pulse mode.

Release pulses in continuous pulse mode (pulse frequency 5 Hz/15 Hz).

Check that the triggered pulses are correctly counted on the treatment pulse counter.

3.6 Standard settings


20 Before each treatment, make sure that the number of pulses and the actual energy
value are set to zero.

NOTE
Set the nominal value counter to the required value. The '-' symbol appears if zero

Start the D-ACTOR treatment at a pressure of 2 bar and a frequency of 5 Hz.

NOTE

The pulse frequency beginning with 1 Hz and also the energy level beginning with
1.0 bar can be increased in steps up to maximum value depending on the control
device.
Operation

19 920 02 0614
3.7 Treatment

3.7.1 Safety information

Before using the device, the user must make sure it is functioning safely and in proper
condition.
Each time after the device has been transported, make sure that all functional
checks have been performed on the device before you start treatment. Read about
this in chapter 3.5 functiOnal checkS.

CAUTION !

Handpiece not positioned correctly.


Impairment to health due to ineffective treatment!

position always corresponds to the treatment zone.


Make sure that the treatment is only administered by users
authorized for this purpose. (For the conditions please read the
operating manual of your control device.)

chapter general Safety infOrMatiOn in the operating manual of your control device
is not permitted!

21
CAUTION !

Malfunction of the device or its components


Various injuries are possible!
Immediately comply with all status and error messages which
appear during the treatment.

CAUTION!

Over extended periods, the noise of the pulses can be per-


ceived as unpleasant!
Offer ear protection to the person to be treated.
Recommendation: The user should also wear ear protection.
Operation

19 920 02 0614
3.7.2 Setting parameters
Treatment should always start at a low energy level. This also applies to resuming
treatment after an interruption. The pulse energy should be increased gradually
during treatment. The low levels are used less for therapy and more for familiarising
the person to be treated.
Select a low energy level and frequency
OR
Load an indication and administer the treatment according to the
recommendations of the treatment parameters stored in the system)

NOTE

The selection of energy levels is based on the opinion of the operator adminis-
tering treatment. The maximum energy level used during treatment must not
cause the person to be treated undue pain under any circumstances.

3.7.3 Coupling the handpiece


D-ACTOR

area and to the transmitter.

Avoid excessive pressure of the transmitter to the patient‘s skin. Excessive pressure
is not needed for the success of the treatment.
22
V-ACTOR

and to the V-ACTOR II transmitter.

3.7.4 Triggering pulses


Once all necessary preparations have been taken, it is possible to start the treatment.
Cleaning, Maintenance, Overhaul

Make sure that the pulse counter is at zero and a low energy level has been set.

Press the trigger button on the handpiece.

Press the trigger button again to stop release of pulses.

19 920 02 0614
4 Cleaning, Maintenance, Overhaul

4.1 Cleaning
Regular cleaning ensures perfect hygiene and operation of the D-ACTOR handpiece.
The handpiece, in particular the transmitter, must be thoroughly cleaned and disin-
fected after each therapy session.

CAUTION !

Flammable and volatile cleaning and disinfecting agents may


generate an explosive atmosphere.

Disconnect the handpiece from the control unit before starting any
cleaning or maintenance work.

ATTENTION !

its tubing.

Clean the coupling gel off the handpiece using a cleaning agent that is suitable for
surface cleaning.
23
Disinfect the handpiece with an alcohol-based disinfectant that is suitable for
surface cleaning (follow the manufacturer’s instructions).

Component Procedure Interval

Handpiece shaft and clean and disinfect daily


cushion or after 20,000 pulses

Cleaning, Maintenance, Overhaul


Guide tube clean from inside with every day
brush

Transmitters and O-rings clean in ultrasonic bath after each treatment or


and disinfect contact with a person to
be treated

Guide tube, projectile replace after 1,000,000 pulses


and O-rings (overhaul the handpiece)

Table 4-1 Cleaning interval

19 920 02 0614
4.1.1 Changing the transmitters

NOTE
The handpiece is equipped with a projectile safety catch device in order to prevent the
projectile from being ejected in case a pulse is released while the transmitter and the
transmitter screw cap are removed.
The safety catch is also activated if the transmitter screw cap is not screwed on tightly, if the
sealing ring between the cap and the transmitter is missing or if two sealing rings (old and
new) come together at the rear of the transmitter.

CAUTION !

Danger of injury due to pulse triggering when handpiece is


open.

Disconnect the handpiece from the control unit before starting any
cleaning or maintenance work.

this device has to be replaced. Please send the handpiece for


reparation.

Disconnect the handpiece from the


1
control device.

24

Unscrew the transmitter screw cap


2
from the handpiece.

Remove the transmitter insert.


3
Cleaning, Maintenance, Overhaul

4 Insert the transmitter insert into the


corresponding transmitter screw cap.

5 Screw the transmitter screw cap onto

19 920 02 0614
6 After replacing the transmitter, make
sure that the handpiece cap and the

NOTE
Make sure that the two cap parts of the D20 and the D35 transmitters are

shaft.
Check the screw connection of the transmitter screw cap and cap parts during
prolonged treatment phases

4.1.2 Cleaning the handpiece


NOTE
After cleaning, the handpiece must be dry before it can be reassembled.

components.

25
Disconnect the handpiece from the control
1 device.

Unscrew the transmitter screw cap from


2 the handpiece.

3
Clean the coupling gel off the handpiece Cleaning, Maintenance, Overhaul
using a cleaning agent that is suitable for
surface cleaning.
Disinfect the handpiece with an alcohol-
based disinfectant that is suitable for
surface cleaning (follow the manufacturer’s
instructions).

19 920 02 0614
4 Unscrew the shaft from the handpiece and
pull it out of the handpiece handle (5).
Use the supplied open-end spanner (4) for
this purpose.

6 Clean the guide tube with a brush in order


to ensure perfect projectile movement.

26
Cleaning, Maintenance, Overhaul

19 920 02 0614
4.1.3 Cleaning the transmitters
Unscrew the transmitter screw cap and remove the transmitter insert from the
transmitter screw cap.

Clean all of the parts under running water.

NOTE
The transmitter insert with the D20-S, D20-T and D35-S can only be dismantled
and the sealing rings can only be removed using special tools. You should avoid
doing this because it could result in damage to the transmitter. It is not necessary
for cleaning purposes.

27

Cleaning, Maintenance, Overhaul


It is also recommended that
the transmitter be cleaned and
disinfected in an ultrasonic bath.
For this purpose, use only instrument
disinfectants for heat-sensitive,
reusable devices.

Clean and disinfect not only the transmitter insert but also the transmitter screw
cap with the usual alcohol-based cleaning agents and disinfectants.

Dry the transmitter and transmitter screw cap before screwing them together.

19 920 02 0614
Push the insert into the front cap and screw the two cap parts together until

NOTE
Make sure that the two cap parts of the D20 and the D35 transmitters are

shaft.
Check the screw connection of the transmitter screw cap and cap parts during
prolonged treatment phases

4.2 Overhaul
Pulses are generated mechanically. Due to the effects of friction, the handpiece com-
ponents are continuously exposed to mechanical stress, which will cause minor wear.

NOTE
The D-ACTOR handpiece should be overhauled about every 1,000,000 pulses.
This can be done quickly and easily by the user of the device. All that is required is
the overhaul kit, which includes all required wear parts.

4.2.1 Contents of the D-ACTOR overhaul kit


28
– 2 projectiles
– 2 guide tubes
– 2 socket sealing rings
– 1 O-ring guide

The overhaul kit can be ordered from your dealer quoting part number 17212.
Cleaning, Maintenance, Overhaul

NOTE
The sealing rings, the projectile and the guide tube must always be replaced each
time the handpiece is overhauled. Observe the O-ring Guide when selecting the
sealing rings to be used. It is contained in the overhaul kit.

19 920 02 0614
4.2.2 Overhauling the handpiece

CAUTION !

Flammable and volatile cleaning and disinfecting agents may


generate an explosive atmosphere.

Disconnect the handpiece from the control unit before starting any
cleaning or maintenance work.

ATTENTION !
An open-end spanner has to be used to release or to assemble handpiece
shaft when overhauling the handpiece.

Place the handpiece on a dry, clean and dust-free surface.

Disconnect the handpiece from the control


1 device.

Unscrew the transmitter screw cap from


2 the handpiece

29

Unscrew the shaft from the handpiece and


pull it out of the handpiece handle (4).
Use the supplied open-end spanner (3) for
this purpose.

Cleaning, Maintenance, Overhaul


4

19 920 02 0614
5 the shaft. If necessary, use a thin metal
rod or the supplied hexagonal spanner as
a pulling tool by inserting it through the
openings in the guide tube.

6 the handpiece handle to hold back the


projectile. To remove the projectile, hold
the handpiece handle with its opening
pointing down. Gently knock the handle
against the table surface until the projectile
falls out. In the event that the projectile
breaks apart due to overloading, a
fragment of the projectile may be left
behind inside the guide tube.

Dispose of the used guide tube and the


7 used projectile

30
8
F15

C15
Cleaning, Maintenance, Overhaul

DI15

Dispose of the detachable sealing rings of the C15, DI15, F15 transmitters and the
sealing ring on the shaft.

19 920 02 0614
Clean the shaft, the transmitter (including
9
transmitter screw cap using a disinfectant
containing alcohol. These are reused after
cleaning.

NOTE
D20-S
The transmitter insert of the D20-S, D20-T
and D35-S can only be dismantled and
D20-T the sealing rings can be removed only
with special tools. You should avoid doing
this because it could result in damage
to the transmitter. It is not necessary for
D35-S cleaning purposes.

31

Now take out of the overhaul kit the new


sealing rings for the C15, F15, DI15 and
C15
transmitters and for the shaft and install
them. Observe the O-ring Guide for this

Cleaning, Maintenance, Overhaul


purpose. It is contained in the overhaul kit.
F15

DI15

19 920 02 0614
Take out the new guide tube and the new
10
projectile from the overhaul kit.

Insert the guide tube into the opening in


11 the shaft by pressing in until the stop.
Make sure that the end of the guide tube
where the two air slots are located is in the
direction of the handpiece handle.

12 guide tube.

32
Screw the shaft into the handpiece until
13

onto the table and tighten the shaft using


the open-end spanner. It must no longer
be possible to unscrew the shaft by hand.
Cleaning, Maintenance, Overhaul

Screw the transmitter screw cap with the


14
shaft.
Make sure that the two cap parts of the
D20 and the D35 transmitters are screwed

Carry out a functional check of the handpiece (see chapter 3.5 functiOnal checkS).

19 920 02 0614
NOTE
Pictures of handpiece and transmitters are examples. Individual components may
differ from those shown in the illustration.

Reset the overhaul pulse counter (see chapter 3.3.3 reSetting the pulSe cOunter).

4.3 Maintenance
Preventive maintenance is not necessarily required.

4.4 Disposal

When disposing of this product, no special measures have to be observed. Please

the service life of the handpiece, please return the instrument to STORZ MEDICAL AG.

4.5 Repair
33
Repair work on defective instruments must only be carried out by personnel suitably
authorised by STORZ MEDICAL. Only original STORZ MEDICAL parts may be used for
this purpose. The personnel suitably authorised can be from STORZ MEDICAL or be
representatives of STORZ MEDICAL agencies and dealers.

4.6 Service life

Cleaning, Maintenance, Overhaul


The D-ACTOR handpiece should be overhauled after around every 1 million pulses
(see chapter 4.2 Overhaul).
Provided this interval is observed, the average expected service life is approx. 5 million
pulses for the handpiece and 1 million pulses for the transmitters.
Exceeding the service life can be expected to result in a failure of the devices.
In this case, no warranty claims shall be accepted on the basis of the information
given in chapter 8 Warranty and Service.
For information about the service life of your control device, please refer to the sepa-
rate operating manual for your control device.

19 920 02 0614
5 Error messages and trouble-shooting

Fault description Possible cause Corrective actions

No power output Leaks in handpiece cable Check the cable and tube
or cable not properly connections and replace
connected them, if necessary

Blocked or worn projectile Dismantle the handpiece

Overhaul the handpiece

Handpiece defective Replace the handpiece

No projectile Install a projectile

Guide tube installed the Turn the guide tube


wrong way round around

Irregular frequency 2 projectiles Remove one projectile

Leak in the handpiece Defective or missing red Send in handpiece or


connector O-ring on plug inform Service

34

pictographic
Cause Corrective actions
error messages

deactivated menu function

The device combination


Error messges and trouble-shooting

doesn't support menu


function

pulse counter reached perform overhaul of the


limit for overhaul handpiece
(always the actual value of (s. chapter 4.2 Overhaul)
the total pulse counter is
displayed )

19 920 02 0614
6 Accessories and Spare Parts

Handpiece D-ACTOR
Handpiece D-ACTOR 21700.0001
Handpiece D-ACTOR LT 23213.0001
D-ACTOR overhaul kit 17212
Red sealing sleeve 19475

Transmitters

C15 transmitter 19222

F15 transmitter 21356

DI15 transmitter 21374

D20-S transmitter 21004

D20-T transmitter 21125

D35-S transmitter 21122

Other accessories

D-ACTOR handpiece operating manual 24887 35

O-ring guide 23031

Accessories ans Spare Parts

19 920 02 0614
7
Handpiece D-ACTOR
Compressed air input 1.0 - 5.0 bar
Ambient temperature
10 – 40 °C
during operation
Ambient temperature
0 – 60 °C
during storage and transport
Ambient air pressure
800 - 1,060 hPa
during operation
Ambient air pressure
500 - 1,060 hPa
during storage and transport
Air humidity
5 – 95%, non-condensing
Storage and transport
Air humidity
5 – 95%, non-condensing
operation
Weight 480 g

Protection against the ingress of water IPX0

Subject to technical changes

NOTE
36
When the product is distributed to third parties, the following must be
observed:

– The complete device documentation must be delivered together with the


product.
– The product may only be exported to a foreign country when the product
and the corresponding indications are allowed there.

This device complies with the applicable standards EN 60601-1, CAN CSA-C22.2 No.
601.1, UL Std. No 60601-1.

For information about conformity with directives, please refer to the separate operat-
ing manual for your control device.

7.1 Symbols and labels


Label Name

It is essential to read the operating manual!

Table 7-2 Labelling handpiece

19 920 02 0614
8 Warranty and Service

8.1 Warranty of the D-ACTOR handpiece


The D-ACTOR handpiece is a wear part. We will replace new handpieces that have
performed up to 1 million pulses at no charge to the customer upon the customer fur-
nishing adequate proof that the defect is due to defects in material or workmanship.
Transport costs and the risk of loss during the shipping of returned products shall be
borne by the customer.
Warranty claims will only be accepted if the handpiece is returned in its complete and
original state, cleaned and in the original packaging if at all possible (without trans-

Missing components will be replaced subject to charge.


Transmitters and overhaul kits are not covered by the handpiece’s warranty.

ATTENTION !
M

unauthorised personnel will relieve the manufacturer of its liability and


responsibility for safe system operation. This will automatically void the
warranty even before the end of the warranty period.

37
8.2 Service
Should you have any further questions or require additional information, please feel
free to contact your dealer.

Warranty ans Service

19 920 02 0614
Warranty ans Service

38

19 920 02 0614