RSL3000 (1)

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

Mo: 7202822212, 9998755736

Office:A-1,2,2ndFloor,TejendraComplex,
Opp. CMC-Power House, Odhav Road, Odhav,
Ahmedabad-382415. Ph. 079 22870063
E-mail excellentcalahm@gmail.coM
CertificateNo.CC-3050 Web: www.excellentcal.com

CALIBRATIONCERTIFICATE
Cali. Date: 12-12-2023 ULR No. CC300219000002758F C.R.F.No.: EX/2023/06/22
Due Date: 11-12-2024 Certi. No.: EX/M/2022/04/4/113 Discipline: Mechanical
Receipt Date: 12-12-2023 Call. WA. : EX/CWP/M/029 Parameter: Pressure
Issue Date: 12-12-2023 Ref. IS. DKD-R 6-1 Format No. EX/CR/PG

CUSTOMER
M/s, Immense Industry 107AshutoshSoc, Environmental Temp. 25± 5°C
B/hUtkarshPetrolPump, Condition % Rh Lessthan 75% Rh
Karelibaug,Vadodara. Location of calibration At Site
Condition of Item Satisfactory

2.Description of Item
Name :- Bolt Torqueing tool Range :- 0- 10000 psi
I.D No. :- ENERPACBOLT TORQUING TOOL Resolution 10 Bar
Make / Model :- ENERPAC S3000 Specified Accu. :- ---
Sr.No. :- 23A0714 Location wrench :- Bolttorqueingtool
Type :- HYDRULIC
.
3. Master Instruments Details
Name :- Digital Pressure Gauge
I.D No. :- EX/EQP/78
Certificate No. :- VIS/23-24/M-184
Certified By :-VijayInstrumentationServices
Cali. Validity :-10-12-2025

4.Calibration Result of UUC Zero Set Point Error = 0.000


Reading Observed By Master Gauge Std.
Set Value Oil Exp.UNC
Average ± Error Hysteresis
UUC UP Down Ap
+- % Rdg
Reading
Pind M1 M2 Mk, Pkid-Mm terse/
0.0 0.00 0.00 0.00 0-00 0.00 #
120.0 120.89 120.81 120.85 -0.85 0.01 0.88
190.0 190.90 190.84 190.90 -0.88 0.06 0.83
350.0 351.78 350.84 350.86 -0.86 0.02 0.82
530.0 530.80 530.84 530.82 -0.82 0.03 0.83
700.0 700.44 700.40 700.78 -0.78 0.04 0.85
Mo: 7202822212, 9998755736
Office:A-1,2,2ndFloor,TejendraComplex,
Opp. CMC-Power House, Odhav Road, Odhav,
Ahmedabad-382415. Ph. 079 22870063
E-mail excellentcalahm@gmail.coM
CertificateNo.CC-3050 Web: www.excellentcal.com

Thereporteduncertaintyistheexpandeduncertaintyinmeasurementobtainedbymultiplyingthe standarduncertaintyby
thecoveragefactor k=2, whichcorresponds toa coverageprobabilityofapproximately 95% for normaldistribution.
Note:-
1. Next Calibrationdue date is reportedas per customer request.
2. Thiscertificate refers onlythe particular item calibrated at site/laboratory.
3. The calibrationresults reportedare valid at the the time of and under the stated conditions of measurement.
4. Thecertificateshouldnot be reproducedexcept,infullwithoutthe prior permissionfromthe TechnicalManager