Service Manual HBC 80 150 260

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 28

Ice-lined refrigerator

Service Manual

English
Model :
HBC - 80
HBC - 150

Français
HBC - 260

Русский
Contents

Schematic of product appearance..................................... 1


Specification of products................................................... 2
Spare parts of electrical appliance details......................... 3
wiring diagram................................................................... 4
Spare Parts Details of Refrigerating System..................... 6
Exploded View and spare part list..................................... 8
Operation Guide............................................................... 10
Alarm function.................................................................. 14
Special considerations for installation and debugging..... 15
Guide for replacement of key components....................... 16
Others............................................................................... 24
Schematic of product appearance
Handle

Basket
( Optional)
Lifting handle

English
Cabinet

Compressor
indicator

Drainage
hole

Temperature displayer

Hinging

Condenser
Switch

Temperature
Controller

Compressor

Filter drier
Electric control box

1
Specification of products
Product model schematic

H (Haier) B (Ice Liner) C (Refrigerator) -80 (Volume: about 80L)

Specification

Model HBC - 80 HBC-150 HBC-260

Ambient temperature (℃ ) + 5 ~+ 43 + 5 ~+ 43 + 5 ~ + 43

Temperature range of
+2 ~ +8 +2 ~ +8 +2 ~ +8
refrigerating compartment (℃ )

Gross refrigerating volume (L) 80 150 260

Vaccine Storage Capacity (L) 61 122 211

Rated voltage (V) 220V - 240V~ 220V-240V~ 220V-240V~

Rated frequency (Hz) 50 60Hz 50 60Hz 50/60Hz

Refrigerant (g) 60 52 70 60 80

Rated power (W) 110 140 120 150 165

Overall dimension (W x D x H)
788 × 654 × 875 1128 × 654 × 875 1650 × 654 × 940
(mm)

Product Weight (kg) 110 140 205

2
Spare parts of electrical appliance details

No. Special No. Description Qty. Main technical parameters Remarks

Electronic
Working voltage:
1 0274000352A temperature 1
220V-240V~, 50/60Hz
controller

Rated specifications:
2 0075050003 Switch 1
AC 230V, 50/60Hz ,16A

Temperature test range:

English
-50. 0 ℃ ~ 70. 0 ℃
Temperature
3 0274000357 1 resolution: 0. 1℃
displayer
current-carrying is lower
than: 0.2A

Compressor
4 0075050018 1 Rated specification: 230V,
indicator

Working voltage: 220V -


5 0074000172 Axial fan 1
240V, 50/60Hz 0.11/0.1A

Overload Working voltage: AC


6 0274300008B 2
protection switch 230V,5A

3
Refrigerating
compressor

Ice layer sensor

Refrigerating sensor
wiring diagram

Freezing compartment sensor


Indicator light
Ambient temperature sensor

N
L
HBC-80/150(220V 50Hz)

Heating wire

Fan
Freezing compressor
Refrigerating compressor

4
Electronic temperature controller

Condenser fan

Reset overcurrent protector Power switch

L Terminal blocks Heating wire


N

E
English
Ice layer sensor
Refrigerating sensor
Freezing compartment sensor
Ambient temperature sensor
E Refrigerating
N N compressor
L
L C
heating wire
Fan Indicator light
Freezing compressor
Refrigerating compressor

5
Electronic temperature controller
HBC-80/150/260 (220V 50/60Hz)

Condenser fan
Reset overcurrent protector Power switch
L Terminal blocks Heating wire
N
E
Spare Parts Details of Refrigerating System
Details

No. Special No. Description Qty Main technical parameters Remarks


HBC-80/150
0274000418 1 Model: HXK 70AT, 220V - 240V, 50Hz
Compressor
HBC-260 Model: FMXA9C, 220V - 240V,
1 0060705994A 1
Compressor 50/60Hz
HBC-80/150 Model: FMXA6C, 220V - 240V,
0274002031 1
Compressor 50/60Hz
HBC-80 Φ6×0.71 Steel pipe surge rustproof +
0270700425 1
Condenser Φ1. 5 iron wire
HBC-150 Φ6×0.71 Steel pipe surge rustproof +
2 0270700421 1
Condenser Φ1. 5 iron wire
HBC-260 Φ6×0.71 Steel pipe surge rustproof +
0270700621 1
Condenser Φ1. 5 iron wire
3 0074180001 Dry Filter 1 10g
HBC-80
0270700660 Capillary 1 Refrigerant flowrate:9.8
tube
4 HBC-
150/260
0270700657 1 Refrigerant flowrate:10.5
Capillary
tube
HBC-80 Inside
0270700606 1 TP2M φ8*0.75
Evaporator insulation layer
HBC-150
5 0270700505 1 TP2M φ8*0.75
Evaporator
HBC-260
0270700620 1 TP2M φ8*0.75
Evaporator

6
Refrigerating system

5 2
4

English
3

1-Compressor 2-Condenser 3-Drying filter 4-Capillary 5-Evaporator

The refrigerating system of Haier vaccine refrigerator adopts direct cooling method.
The refrigeration effect is realized by compressor working which is under the control of thermostat.
Specifically, the compressor works and forces the refrigerant to flow through the condenser, dry filter
and evaporator and finally return to the compressor. This is a complete cooling cycle.

7
Exploded View and spare part list

8
Exploded view & Spare part list

No. Special No. Name Remarks

1 0274000357 Solar display screen

2 0070814623D Handle Assembly

3 0070816334 Hinging ASSEMBLY

Upper/ Lower Hinge

English
4 0070203942B 0070204349B
Cover

HBC - 80: 0270700425 HBC-150:


5 0270700421 Condenser
HBC - 260: 0270700621

6 0074180001 Filter

HBC - 80, HBC - 150: 0274000418


7 HBC - 260: 0060705994A Compressor
HBC – 80/ 150(60Hz): 0274002031

8 0075050003 Power Switch

9 0270400390 Power cord socket

10 0075050018 Power pilot lamp

11 0274000352A Temperature Controller

1-19 0074000172 Fan

13 0270200838 FAN, HOUSING

0270103889C
HBC-80:1EA
Or
15 Storage basket HBC-150:3EA
0270202884
HBC-260:5EA

9
Operation Guide
Temperature Controller

•Temperature controller panel

•Indication description

Power supply indication

Alarm indication
Refrigeration indication

Lock indication

Defrost indication

• Power supply indication: The controller is powered on.


• Alarm indication: When the controller is in the alarm state, the symbol flashes to alert the user.
• Refrigeration indication: This symbol is displayed when the refrigeration is outputting.
• Lock the indication. This symbol displays when the controller is locked.
• Defrost indication. This symbol displays when the controller is defrosting. This product does not
have this function.
•Controller lockout, unlocking and temperature viewing
If there is no key operation for the 60s, the controller will automatically lockout and the buzzer will
beep for approximately 0.5 seconds.
If there is no key operation for 2 minutes, the controller will automatically turn off the display, and only
the power flag flashes.
Pressing button "SET" and " " for 5 seconds can unlock the locked controller locked or lockout it,
and the buzzer beeps approximately for about 0.5 seconds. When the controller is locked, it cannot
be operated until it is unlocked.
Press "SET"and " " key under unlock state for 1s, the freezing chamber’s display window flashes
to display the ambient temperature (before the failure of the ambient temperature sensor), and will

10
automatically log-out if there is no button operation in 2 seconds.
Press " " and " " key under unlock state for 1s, the refrigerating chamber’s display window flashes
to display the ambient temperature, and will automatically log-out if there is no button operation in 2
seconds.
•Parameter settings
1.Under unlock state, press the "SET" key for more than 5s , the display screen will display the
parameter code" PS "(password);
2.Press "SET" key to enter password setting;
3.Use key " " and " " to roll the number to "-15";
4.Press key "SET" to confirm the password;

English
5.The keys " " and " " to roll the parameters (SP1, SP2, SP3, AH1, AH2, AH3, AL3, r0, r1, r2, r3,
Rd1, Rd2, Rd3, CA0, CA1, CA2, CA3, CA4, t0, t1, t2, Ad, AE, Aln, PCH);
6.Press key "SET" to confirm the parameter and display the parameter value;
7.Use the keys " " and " " to enter the value of the parameter;
8.Press the key "SET" to temporarily save the modified parameter value and return to the display
parameter;
9.When modifying other parameters, repeat steps 5 - 8;
10.Press the key "SET" for more than 5 seconds, save the modified parameter value and return to
normal condition.
•Factory default parameter list

11
Default
Parameters Description minimum maximum Unit
Pch=1 Pch=3

SP2 Control Temperature 2 r1 r0 ℃ 5 4.5

SP3 Control Temperature 3 r3 r2 ℃ -15 -15

r0 SP2 can set maximum temperature r1 85.0 ℃ 8.0 8.0

r1 SP2 can set minimum temperature -45.0 r0 ℃ 2.0 2.0

r2 SP3 can set maximum temperature r3 85.0 ℃ 0.0 0.0

r3 SP3 can set minimum temperature -45.0 r2 ℃ -20.0 -20.0

Rd1 SP1 control temperature difference 0 10.0 ℃ 0.2 0.2

Rd2 SP2 control temperature difference 0 10.0 ℃ 0.2 0.2

Rd3 SP3 control temperature difference 0 10.0 ℃ 1 1

CA0 The ice temperature calibration target -10.0 10 ℃ 5.5/5.2 5.3

CA1 Ice Sensor Calibration Temperature -5.0 5.0 ℃ 0 0

Refrigerating storage sensor


CA2 -5.0 5.0 ℃ 0 0
calibration temperature

Freezing Sensor Calibration


CA3 -5.0 5.0 ℃ 0 0
Temperature

Ambient Temperature Sensor


CA4 -5.0 5.0 ℃ 0 0
Calibration Temperature

t0 Start - Up Delay Start 0 12 Min 2 2

t1 Time Delay Start 0 60 Min 5 5

t2 Time of Delay for Starting Freezing 0 60 Min 2 2

Ad Alarm Delay 0 255 Min 180 180

Delay for Canceling beeping and re-


AE 0 255 Min 240 240
alarm

ALn Buzzer’s energy supply 0 1 - 1 1

PCH Product Model Selection 1 5 - 1 3

HBC-80/150: PCH=1
Caution
HBC-260: PCH=3
HBC-80/150:PCH=1 (For 60Hz products :CAO=5.2)

12
•Fault code
Displayed symbol Symbol description
Er1 Ice storage sensor fault
Er2 Refrigerating sensor fault
Er3 Freezing sensor fault
Er4 Ambient temperature sensor fault
AH1 Ice high temperature alarm
AL1 Ice low temperature alarm
AH2 Refrigerating high temperature alarm
AL2 Refrigerating low temperature alarm
•Wiring Diagram

English
Ice layer sensor

Refrigerating sensor

Freezing compartment sensor

Ambient temperature sensor

N
L
heating wire

Fan
Freezing compressor
Refrigerating compressor

Electronic temperature controller


Temperature indicator

5.0 ℃
REFRIGERATOR

Temperature displaying area Solar receiving area


The display shows the real-time temperature in the refrigerator. The display is powered by solar
energy, no extra circuits need to be connected. The display only shows the temperature inside the
refrigerator and does not control the refrigeration system of the product.
During the operation, the solar energy receiving area cannot be blocked, otherwise temperature will
not be displayed on the temperature display area.
In the case of weak light, the display will flicker or the display will be unclear. It can be displayed
normally by using a flashlight or other light source for about 3 seconds.

13
Alarm function

Displayed
Symbol description Alarm mode Alarm clear mode
symbol

Code flicker
Er1 Ice storage sensor fault Maintenance of ice storage sensor
+ buzzing

Refrigerating sensor Code flicker


Er2 Maintenance of refrigerating sensor
fault + buzzing

Code flicker
Er3 Freezing sensor fault Maintenance of freezing sensor
+ buzzing

Ambient temperature Code flicker Maintenance of ambient temperature


Er4
sensor fault + buzzing sensor

Automatically clear or repair the


Ice high temperature Code flicker
AH1 sensor after the temperature is
alarm + buzzing
lowered

Ice low temperature Code flicker Overfreezing of water tank caused


AL1
alarm + buzzing by too low ambient temperature

Automatically clear or repair the


Refrigerating high Code flicker
AH2 sensor after the temperature is
temperature alarm + buzzing
lowered

Refrigerating low Code flicker Check heating function, repair


AL2
temperature alarm + buzzing sensor

Automatically clear or repair the


Freezing high Code flicker
AH3 sensor after the temperature is
temperature alarm + buzzing
lowered

Check if the ambient temperature


Freezing low Code flicker is too low, causing the temperature
AL3
temperature alarm + buzzing inside the refrigerator to be too low,
or repair the sensor

14
Special considerations for installation and debugging
1. Read the instructions carefully before installing the equipment;
2. The freezer should be placed in a cool, well-ventilated place with no direct sunlight. The ground
should be flat, and the distance between each side of the refrigerator should be greater than 15cm to
ensure good heat dissipation. Keep the floor clean and prevent blockage of the front and rear cabins.
3. HBC-80/HBC-150 has no casters. When moving the product, please use the handles on both sides
of the cabinet to lift the cabinet to avoid forced dragging.
4. After connecting the power cord, you need to install the power cord fixing brack. Refer to the
installation instructions of the power cord bracket and install it according to the step requirements.

English
5.About 7 minutes after power-on, the front compressor operation indicator lamp will start (green). If
the local voltage is unstable, it needs to cooperate with the pressurizer, otherwise the equipment will
be damaged.

15
Guide for replacement of key components
Troubleshooting for temperature controller

Display Phenomenon 1 Phenomenon 2 Reasons Solution

Display screen
Compressor
show Err, the
run 15 minutes, The controller is
ERR alarm bell Change the controller
stop 8 minutes, broken
twinkle and
so the cycle
buzzer ring

The inner temp.


Display screen Check the temperature’s
EHi is higher than
show Ehi position
85 ℃

The inner temp.


Display screen Check the temperature’s
ELo is lower than
show Elo position
-50℃

1. if the refrigerator
start not long time ,It is
normal.
The inner temp.
2. check whether the
Alarm bell is higher than
improper setting of
HA twinkle,buzzer the setting
high temperature alarm
rings value of high
parameters
temp.alarm.
3. Find the cause of
high temperature in the
refrigerator and solve it.

1. Check if the
The inner temp. environment temp.is
Alarm bell is lower than lower than 0℃
Al twinkle,buzzer the setting 2. Check if the low temp.
rings value of low alarm setting value is
temp.alarm. wrong
3. other reasons

The
Alarm bell
show Power was
twinkle,buzzer Reconnect the power
screen failed.
rings
is dark

16
How to deal with the high temp. Alarm

When the temperature is higher than


the set value of AH over 10 hours. Or
the temperature is higher 10 degree. N
Then the high alarm start.(factory
default is 8℃ )

English
Y

Alarm appear(The buzzer ring)

Check the compressor whether is N


was working or not ? check compressor

Check the fan motor whether is


N
check the fan motor
was working or not?

N Use a hot blower to heat the gasket


Check the door gasket whether it was ,so that it can back to the original
sealing good or not? shape.

It must be refrigerant leakage, Check


leakage points and recharge

17
How to deal with the low temp. alarm

When the temperature is lower than the


set value of AL over Ad hours.
Then the low alarm start.(factory default is N
2℃

Alarm appear(The buzzer ring)

N Check the inner temperature


Check the compressor whether it can whether it is lower than Set
stop working? temperature?

Check the inner temperature whether it is N


lower than Set temperature check the temperature controller

Check the ambient temperature


whether it is lower than 5 ℃

18
Servicing the vaccine refrigerator
(1) Check the whole circuit of vaccine refrigerator for disconnection, damaged wire and short circuit.
(2) Check the insulation of vaccine refrigerator. In normal conditions, the resistance is higher than
2 MΩ;if it is lower than 2MΩ, please make a further check of the refrigerator’s resistance. Set the
voltmeter at R1 Ohm position, insert its two styli respectively into L and N terminals and read the
indicated value. The normal value is from several ohms to scores of ohm. If the resistance is zero,
it means that short circuit occurs in the vaccine refrigerator or short circuit occurs due to poor
performance of the compressor/fan motor. In this case, please investigate the cause and repair. If
the resistance is too high or infinite, it means that open circuit occurs. In this case, please check and
clear the fault.
After the above checks, make sure that the vaccine refrigerator is free of electrical fault and turn it on.

English
Electric leakage of the compressor

(1) Electric leakage of the thermostat


(2) Electric leakage of the fan
(3) Electric leakage of damaged wire
If the electric power cut because of the vaccine refrigerator leakage, try to disconnect the compressor\
controller\fan\power line and so on one by one. Once the power circuit is normal when you disconnect
the one, you can conclude the unit is leaky.

Troubleshooting for refrigerant leak


(1) Touch the cover of the condenser; after continuous running, the temperature in the upper part of
the cover is the highest, and that of the lower part is close to the room temperature. The temperature
is associated with the ambient temperature. The temperature is low in winter; so the temperature of
the condenser is low as well, and only a small part of the condenser is warm. The temperature is high
in summer; so the temperature of the condenser is high as well, and the whole condenser is warm. If
the condenser is not warm, it may be a result from refrigerant leak.
(2) The temperature of the high-pressure exhaust pipe is high; when you touch it with hand in
summer, you may feel hot. It feels hot even in winter. The temperature of the low-pressure intake
pipe is low. Dew drops may appear on the pipe wall in summer, and you may feel cool when touching
it. In winter, you may feel cooler when touching it. If the exhaust pipe is not hot, and the intake pipe is
not cool, then it may be a result of refrigerant leak.
(3) Under normal working conditions, the temperatures around the surface of the evaporator of the
vaccine storage freezer are almost the same, and the frosts are of the same thickness. If there is
no frost on the surface of the evaporator (the whole inner wall of the chest vaccine storage freezer)
or only in a small upper part of the evaporator surface, then the cooling system has failed, such as
insufficient refrigerant or refrigerant leak
(4) Check the pipeline and determine that there is no leak, especially at the joints. The refrigerant
containing a small amount of refrigerant oil is of good permeability. In case of leak, there are oil stains
at the leak point.

19
(5) When the vaccine storage freezer is working normally, the high-pressure liquid refrigerant flows
into the low-pressure evaporator through the capillary, and the sounds of air flow and water flow could
be heard. Open the door of the vaccine storage freezer, approach the evaporator, and you may hear
the sounds as mentioned. If there are no sounds or there are only occasional sounds, it indicates that
there is refrigerant leak or insufficient refrigerant in the system.

Analysis of poor refrigeration effect and solutions


(1) Check the setting value of thermostat. If it is not in the range of 2℃ ~5℃ , adjust it to 3℃ .
(2) Check whether the back and top of vaccine refrigerator are too close to the wall, whether the
vaccine refrigerator is exposed to direct sunlight or close to a heat source, whether the room
temperature is too high and whether the condenser is clogged by a piece of cloth, paper or dirt. Any
of the above problems will affect the condenser’s heat dispersion and further reduce its refrigerating
capacity.
(3) Check whether there are too many vaccines in the vaccine refrigerator. If hot vaccines or too
many vaccines are put in the refrigerator, cold air circulation would be affected. Hot vaccines must not
be put in the refrigerator until they get cool.
(4) Check whether the refrigerator door has been closed well, whether the door sealing strip is too
dirty and whether the door is sealed well. Remove dirt and apply a little talcum powder.
(5) Frequent opening and extended opening of the refrigerator door will reduce the refrigerating
capacity.

How to handle poor sealing of the sealing strip


If poor sealing is caused by too much dirt on the sealing strip, clean the sealing strip with a cloth and
lukewarm water. After cleaning, apply a little talcum powder. If there is still a gap, blow the sealing
strip with an electric dryer to heat it and maintain the temperature between 50 ℃ and 60 ℃ , then
press around the sealing strip by hand to recover it. If this method is useless, wedge a piece of
sponge (about 10 mm in width, 5 mm in thickness and a little longer than the gap) into the position
between the bottom of sealing strip and door mat.

Why the vaccine refrigerator cannot be used at low temperature

When ambient temperature is below 0℃ , the lubricating ability of refrigeration oil in the compressor
will reduce and the compressor will be difficult to start or cannot start or may be damaged.

How to know the compressor motor is at fault


It is easy to find that the motor is at fault by checking the resistance of its coil. Specifically, connect
both styli of ohmmeter respectively to the “common” and “run” terminals to measure the resistance of
coil. If the resistance value is infinite, it means open circuit of coil. If the resistance value is much less
than that of the motor, it means that short circuit occurs in the motor or the winding is burned..

20
Fault analysis and servicing of the vaccine refrigerator

Fault Causes Solutions


1. The fuse is blown2. The plug, socket and 1. Check for short circuit: if short circuit occurs,
cable are in bad connection. The cable is clear this fault; if there is no short circuit,
broken3. The voltage is too low, 10-15% less replace the fuse with a proper one2. Check and
than rated voltage4. The thermostat setting is not reconnect, or replace them3. Use continuously
correct5The thermostat is at fault6. The starting until the voltage rises to rated voltage or install
The compressor
relay or thermal protection device is at fault or a voltage stabilizer4. Reset the thermostat5.
cannot start
damaged7. The wire of compressor’s starting coil Replace the thermostat6. Replace the starting
or running coil is broken8. The compressor motor relay or thermal protection device7. Measure
is burned or the rotor and starter are locked9. with the R×1 position of a multimeter; if the coil

English
High-pressure gas valve leaks or the shaft is is broken, replace the compressor8. Replace
locked the compressor9. Replace the compressor
1. There is an excessive distance between the
thermostat probe and evaporator
1. Move the probe to be close to the inner wall
2. The thermostat is at fault
2. Replace the thermostat
The compressor 3. There are too many vaccines in the vaccine
3. Take out some vaccines, reduce the times
starts frequently and refrigerator; the door is opened and closed
of door opening and shorten the time of door
keeps operating for frequently or is kept open for an unnecessarily
opening
a very long time, but long time
4. Install the vaccine refrigerator in a proper
the temperature drop 4. Ambient temperature and humidity are too high;
well-ventilated location
inside the vaccine airflow is impeded
5. defrost
refrigerator is at very 5.The frost is thick on evaporater
6Check or refill refrigerant
low speed 6. Refrigerant is insufficient or refrigerant leakage
7. Replace the compressor
occurs
8. Replace the dry filter with a new one
7. The compressor efficiency reduces
8. The dry filter is clogged
The compressor
1.The temperature setting value of thermostat is 1. Adjust the temperature setting value to a
keeps running
too lowThe contact of thermostat is stuck correct value
without stop,
2.The probe of thermostat in an improper position 2. Repair or replace
the temperature
causes maladjustmentThe compressor efficiency 3. Move the probe to a proper position4.
inside the vaccine
reduces Replace the compressor
refrigerator is too low
1.Install a voltage stabilizer and adjust it to rated
The compressor
1.The voltage is too low value
buzzes but cannot
2.The starting relay is at fault 2.Replace the starting relay
start, the thermal
3.The starting coil of starting motor is broken 3.Replace the compressor
protection device
4.The compressor does not run 4.The shaft piston in the compressor is clogged
trips repeatedly
5.replace it

21
1.The voltage is too high
1.Install a voltage stabilizer and adjust it to rated
2.The overload protection device is not in good
value
After short time condition and trips earlier
2.Replace the overload protection device
of operation, the 3.The contact of starter is stuckShort circuit
3.Replace the starting relay
overload protection occurs in the compressor
4.Replace the compressor
device cuts off 4.Mechanical failure occurs in the compressor
5.Replace the compressorIncrease the space
5.The temperature is too high around the
for heat dispersion
compressor
1.The floor is not solid
2.The vaccine refrigerator is not stable and level
1.Reinforce the floor
3.When the compressor is running, the friction
2.Keep the vaccine refrigerator stable
between pipes and the refrigerator body causes
There is loud 3.Move pipes a little away from the refrigerator
resonance
noise when the body to avoid friction
4.The fixing screws of compressor are loose
compressor is 4.Tighten the screws to a proper extent
5.Vibration pad fixing the compressor is too tight,
running 5.Adjust the tightness of vibration pad or replace
too loose or is aged
vibration pad
6. There is loud noise inside the compressor
6. Replace the compressor
or the vibration-absorbing suspension spring is
broken
1.Carry out leak detection, repair, weld, and add
No frost on the 1.Severe refrigerant leak
the refrigerant
evaporator 2. The capillary or the system is blocked by dirt
2. Clean the capillary or replace the filter
1.There is a little leak in the refrigerator
There is insufficient 1.Carry out leak detection, repair, weld, and add
2.It is blocked by dirt, but not severely
frost on the the refrigerant
3.Reduced refrigerating capacity of the
evaporator 2. Clean the capillary and the filter
compressor
The evaporator is 1. The vaccine stored in the vaccine storage 1. Clean the vaccine and dry it, and then place
frosted rapidly, or is freezer contains water (not dry) into the vaccine storage freezer
with too thick frost, 2. There are too many vaccines stored in the 2. Only appropriate number of vaccines are
or is iced vaccine storage freezer stored
Electric leakage 1.The refrigerator body is not grounded 1.Ground the refrigerator body as stated
occurs on the 2.The compressor terminal contacts the 2.Replace the compressor
refrigerator body and refrigerator shell and causes short circuit 3.Check carefully step by step. If insulation is
you will be tingled 3.The components of electrical system get wet, seriously damaged, replace the components;
when touching by and thus their insulating ability reduces, resulting remove wet components and put them in a
hand in electric leakage drying box to dry them
1.Ambient humidity is too high or the insulating
1.Control ambient humidity and replace the wire
Electric leakage ability of insulator reduces
of fan
occurs in the fan 2.Short circuit occurs between coils or coils are
2.Replace the fan
burned
1.The fan blade is locked or the fan support is 1.Remove any obstacle around the fan or
The fan makes loud
loose tighten the support
noise
2. The clearance of fan axle is too big 2.Replace the fan

22
1.Check joints for disconnection 1.Replace the fan
The fan does not run
2. Check whether the fan is burned 2.Replace the fan
1.The thermostat has been used for an
Electric leakage 1.Replace the thermostat
excessively long time
occurs in the 2.Remove dust with a brushKeep the thermostat
2.The environment is too dirty
thermostat dry and control ambient humidity
3. Ambient humidity is too high
The thermostat is at
The temperature-sensing probe is damaged Replace the thermostat
fault
Bad connection
between the
The position of probe is improper or changes
temperature- Reinstall the probe
during transport

English
sensing probe and
evaporator

23
Others
Transporting and Transferring Conditions
•The inclination angle shall be no greater than 45 ° when handling the refrigerator.
•When transferring the refrigerator, pay attention to use the special unloading vehicle or truck. It is
strictly prohibited from sliding down or dropping directly from high place.
•When handling the refrigerator, be careful not to trip over it, preventing personal injury or damage to
the refrigerator.
•Do not attempt to pull or carry the equipment by the door handle, preventing personal injury or
damage to the refrigerator.

Installation environment
•Ambient temperature: ≤ 43℃ .The ideal temperature is from 18℃ to 32℃ . Air - conditioning system
shall be used for extreme weather condition.
•Ambient humidity: below 85% Rh.
•Avoid large quantities of dust.
•Avoid machine sway or vibration.
•Working position of the refrigerator: below 2000m.
•It should be placed in bright place without direct solar radiation.

Packing Dismantling
•Cut the packing belt outside the packing case with a blade.
•It can be lifted by multiple people to open the package.
•Remove the bottom bracket.
•The inclination angle of the refrigerator should be no
Caution
more than 45 ° in the above process.
•Take care to avoid unnecessary personal injury in this
process.

24
25
English
Certificate of Quality

checker:

Manufacturer: Qingdao Haier Biomedical Co., Ltd.


Address: Haier Industrial Park, Economic Technology Development Zone.
Qingdao P.R.China
Web:www.haiermedical.com
Revision Date:02/2022
Version:2rd,2022
Dedicated code:0270501299
V13026