Professional Documents
Culture Documents
HPLC Troubleshooting Guide
HPLC Troubleshooting Guide
HPLC troubleshooting
Troubleshooting
There is no standard troubleshooting procedure.
General Pattern:
Locate the problem by ranking possible causes.
Verify the presence of the most probable cause.
If present – fix the problem, otherwise verify the existence of
the next possible cause.
2
HPLC System Components
•Pump
•Injector/ Autosampler
•Column
•Detector
•Data System/Integrator
3
Method vs. System Troubleshooting
System Method
Parameters Parameters
• Flow stability • Flow rate
• Backpressure • Eluent composition
• Clogging • pH &pH modifier (type)
• Detector problems • Injection volume
• Injection suitability • Temperature
• Gradient profile
4
System Parameters
Simple preliminary verification of system setup can save time.
5
Categories of Column and System Problems
I. Pressure
III. Retention
IV. Detection
6
I. Pressure Issues
Column observations Potential Problems
7
Determining the Cause and Correcting
High Back Pressure
Many pressure problems are due to blockages in the system.
8
Column Cleaning
Flush with stronger solvents than your mobile phase
Use at least 25 mL of each solvent for analytical columns
* When using either Hexane or Methylene Chloride the column must be flushed with
Isopropanol before returning to your reversed-phase mobile phase.
9
Prevention Techniques for column problems
Use column protection
➢In-line filters
➢Guard columns
• Filter samples
10
II. Peak Shape Issue
What Are Common Peak Shape Issues?
1. Split peaks
2. Peak tailing
3. Broad peaks
• Many peak shape issues are also combinations - i.e. broad and
tailing or tailing with increased retention
11
Peak Splitting Caused By Disrupted Sample Path
12
Split Peaks from Column Contamination
•Column: StableBond SB-C8, 4.6 x 150 mm, 5 μm Mobile Phase: 60% 25 mM Na2HPO4, pH
3.0 : 40% MeOH Flow Rate: 1.0 mL/min
Temperature: 35°C Detection: UV 254 nm Sample: Filtered OTC Cold Medication: 1.
Pseudoephedrine 2. APAP 3. Chlorpheniramine
13
Split Peaks from Injection Solvent Effects
Sample diluent:
1- 50/50 MeOH/Water 50/50 Buffer/MeOH
2 - 80/20 MeOH/Water Buffer: 20 mM Citrate, pH=4.6
3 - 90/10 MeOH/Water
4 - 95/5 MeOH/Water
5 - 100 MeOH
1
2
3
4
5
Incompatible solvents may cause sample precipitation and column clogging
Different eluent pH and composition may cause peak splitting
14
Peak Tailing, Broadeningand Loss of Efficiency
➢Column contamination
➢Column aging
➢Column loading
➢Extra-column effects
15
Peak Tailing
Identifying Column “Secondary Interactions”
Column: Alkyl-C8, 4.6 x 150 mm, 5μm Mobile Phase: 85% 25 mM Na2HPO4 pH 7.0 : 15% ACN Flow
Rate: 1.0 mL/min
Temperature: 35°C Sample: 1. Phenylpropanolamine 2. Ephedrine 3. Amphetamine 4.
Methamphetamine 5. 5. Phenteramine
10 mM TEA
No TEA USP TF (5%)
USPTF(5%)
1. 1.19
1. 1.29
2. 1.18
2. 1.91
3. 1.20
3. 1.63
4. 1.26
4. 2.35
5. 1.14
5. 1.57
Tip: Mobile phase modifier (TEA) competes with Sample for surface ion exchange sites
at mid-range pH values
16
Low pH Minimizes “Secondary Interactions”
for Amines
Column: Alkyl-C8, 4.6 x 150 mm, 5μm Mobile Phase: 85% 25 mM Na2HPO4 pH 7.0 : 15% ACN Flow
Rate: 1.0 mL/min
Temperature: 35°C Sample: 1. Phenylpropanolamine 2. Ephedrine 3. Amphetamine 4.
Methamphetamine. 5. Phenteramine
pH 3.0
pH 7.0 USP TF (5% )
USP TF (5% ) 4. 1.33
4. 2.35
Tip: Reducing mobile phase pH reduces interactions with silanols and peak tailing.
17
Peak Tailing - Column Contamination
Trick: Reverse Column and Run Sample –If Improved, Possible Cleaning Will Help -No
improvement-Column Damaged and Needs to be Replaced
18
Peak Shape: Fronting Peaks
Causes:
Column Overload
19
Column Overloading
1 ml
5
ml
20
Peak Shape: Broad Peaks
Tip: The extremely low plates for moderately retained peaks are an indication of a very
late eluting peak from a preceding run.
22
III. Changes in Retention Time
Changes in Retention Can Be Chemical or physical
May be caused by:
➢Column aging
➢Column contamination
➢Insufficient equilibration
23
III. Changes in Retention Time
Mobile Phase pH and pH Buffers Why Are These So Important in
HPLC?
pH Effects Ionization
Buffers
RNH3+ RCOOH
RCOO- Base RNH2 Acid
pH=2.3 Base pH=2.3
Acid
pH=6.0 pH=6.0
Absorbance
5 10 15
Retention Time (min.)
Ionization in general decreases hydrophobicity causing a decrease of HPLC
retention.
25
III. Changes in Retention Time
Effect of pH on Acidic Analyte Retention
HPLC Conditions: M obile phase
Column – Zorbax Eclipse XDB-C18, 150 mm x 4.6 mm 30% Acetonitrile
Flow rate – 1.0 mL/min 70% 20mM Na2HPO4 buffer; variable pH
Detection – UV 220nm acidic modifier = perchloric acid
Injection volume – 1mL
26
III. Changes in Retention Time
Effect of pH on Basic Analyte Retention
Chromatographic Conditions
Column: 15 cm x 0.46 cm Zorbax Eclipse XDB-C18 Eluent: 90% Aqueous / 10% MeCN
Buffer: 10 mM Na 2 HPO4 •7H2 O + xH3 PO4 Flow rate: 1 ml/min
Temp: 25o C
A : P y rid in e , pK a
= 5.17
B : 2 -P ic o lin e , p K = 5 .9 6
a
C : 2 , 4 L u t id in e , p K = 6 .7 4
a
D : 4 - e t h y lp yr id in e p K = 5 .8 7
a
E : P h e n y l e th y l a m i n e p K a
=9 .83
F : 2 , 3 d im e t h y l a n i l i n e p K = 4 .7
a
B
C
D
F
pH = 8.04 E
C
B
A
D F
E
pH =7 .04
B C
D F
E
pH =6 .03
B + C C
A D
E
F
p H = 5 .0 8
B + C
A
E
D
F
pH = 4.04
A + B
C
E
p H = 3.0 5 F
C
B
A D
E F
pH = 2.04
27
Importance of pH and Buffers
➢ pH is an effective tool for adjustment of selectivity and retention
column lifetime.
28
IV. Detection Issues
➢ Is it a consequence of technique?
29
Drifting Baselines
➢ Gradient Elution
30
Baseline Noise
OEM Lamp
OEM Lamp
Peak 1 S/N = 15
Peak 2 S/N = 50
Peak 3 S/N = 50
35