Professional Documents
Culture Documents
99
99
170068
research-article2018
Total hip arthroplasty (THA) is a very satisfactory surgical increased revision rates. Common causes of revision THA
procedure for end-stage hip disorders. are wear, loosening, dislocation or instability and infec-
Implant modifications, such as large femoral heads to tion. It should be stressed that a changing pattern of
improve stability, porous metals to enhance fixation and modes of failure has been observed; failures are separated
alternative bearings to improve wear, have been intro- into either early or late, and factors related to failure
duced over the last decade in order to decrease the rate of modes have been identified (Fig. 1). With a more thor-
early and late failures. ough understanding of reasons for failure, of revision tim-
ing and identifiable risk factors, surgeons are better placed
There is a changing pattern of THA failure modes.
to improve their THA outcomes.
The relationship between failure modes and patient- During the past 60 years of THA, it seems that surgeons
related factors, and the time and type of revision are have had to learn from some devastating clinical failures,
important for understanding and preventing short and with patients often having been almost ‘fashion victims’.10
late failure of implants. Despite recent advances in THA technique and improved
The early adoption of innovations in either technique outcomes, we must still pose the question: why do some
or implant design may lead to an increased risk of early THAs fail, leading to technically complicated and expen-
failure. sive revision surgery? This question is of particular impor-
tance nowadays at a time when economic health providers
Keywords: total hip arthroplasty; failure modes; risk factors
are asking challenging questions. With this review we also
aim to show that the relationship between failure modes
Cite this article: EFORT Open Rev 2018;3
and patient-related factors, and the time and type of revi-
DOI: 10.1302/2058-5241.3.170068
sion, are important for understanding and preventing
short and late failure of implants, and that the early adop-
tion of innovations in either technique or implant design
Introduction may lead to an increased risk of early failure. For this pur-
pose, revision rates presented in studies and reported reg-
Since its introduction in the 1960s, total hip arthroplasty
istry data were thoroughly evaluated.
(THA) has proved to be an excellent and reliable treatment
procedure for the end stages of hip pathology, with satis-
factory clinical outcomes at 15- to 20-year follow-up.1-4 Common modes of failure
Following the initial problems which pioneer surgeons
noted in the 1960s and 1970s, such as surgical technique, Wear
structural implant failures and infection, orthopaedic sur- Wear debris production from bearing surfaces is thought
geons in the 1980s faced problems regarding choice of to be the main factor limiting long-term THA survival.
appropriate acetabular and femoral implants and compo- Cobalt-chrome or ceramic heads on old (used mainly by
nent fixation selection.5-8 However, since then it has the Charnley THA group) or conventional polyethylene
become obvious that the long-term survival of a THA is a (PE) liners (developed in order to avoid oxidative degrada-
multifactorial issue. Factors other than the implant, such tion) have been used in the majority of patients, but are
as diagnosis, patient, surgeon and surgical technique, are associated with wear and periprosthetic osteolysis. Alter-
also important in survival. Despite successful outcomes, native bearing surfaces have been developed in order to
THA revision rates have grown steadily in recent years.9 reduce the amount of wear particle production. PE has
Increased life expectancy in a globally aging population is undergone improvements, from the development of
associated with the increased use of THA, resulting in ultrahigh-molecular-weight to crosslinked (XL) PE. These
Downloaded from Bioscientifica.com at 07/14/2024 01:05:38PM
via Open Access. This article is distributed under the terms of the Creative
Commons Attribution-Non Commercial 4.0 International (CC BY-NC 4.0) licence
(https://creativecommons.org/licenses/by-nc/4.0/) which permits
non-commercial use, reproduction and distribution of the work without
further permission provided the original work is attributed.
https://creativecommons.org/licenses/by-nc/4.0/
Total hip arthroplasty: survival and modes of failure
• DIAGNOSIS
Surgical factors related to dislocation include compo-
• LEVEL OF ACTIVITY nent positioning, failure to restore leg length or offset,
abductor mechanism and capsule insufficiency and poste-
rior approach.17 Component positioning within the “safe”
zone is important (combined anteversion being more
important). However, other factors such as pelvic inclina-
tion and obliquity are equally important.18 Implant-related
IMPLANT
SURGEON AND
• CEMENTED-CEMENTLESS factors that decrease the head-to-neck ratio increase the
TECHNIQUE • MODULARITY risk of dislocation. Larger heads with modern necks
• SURGERY VOLUME BEARING SURFACES
improve the head-to-neck ratio, increase jump distance,
•
• APPROACH • HEAD SIZE
reduce component impingement and increase range of
movement.19 When the implant neck impinges against an
osteophyte, scar tissue, liner, cement or heterotopic ossifi-
cation there is a risk of dislocation.17-19 The aetiology of
Fig. 1 Factors related to total hip arthroplasty failure modes. late instability is often multifactorial and includes PE wear,
longstanding malposition of the components, trauma
and neurological and abductor muscle dysfunction.17,20
improved PEs have shown better wear characteristics in
combination with mainly metallic heads in vitro and in Infection
vivo.11 Hard-on-hard bearings, such as ceramic-on-ceramic
(CoC), with survival rates of approximately 97% at ten In the 1960s and 1970s THA pioneers noted a high inci-
years, or metal-on-metal (MoM), with survival rates of dence of septic complications which were occasionally
approximately 90% at ten years (in the best reports), were lethal.21 Advances in duration of surgery (reduction), sur-
initially considered attractive options due to reduced pro- gical technique, the use of antibiotics and modern oper-
duction of wear particles, but they are not suitable for all ating theatre settings have led to a dramatic reduction of
patients.12 CoC bearings have been found to squeak and infection.21 However, there has recently been a global
break, and they require optimal placement to avoid the increase in infection rates.22 New resistant biofilm-forming
risk of neck-to-socket impingement.13,14 pathogens may be responsible for this, as may THA per-
formed in patients with comorbidities (e.g. obesity, dia-
Aseptic loosening
betes), and the use of immunosuppressive drugs. In
addition, an increased incidence of low-virulence infec-
Various theories have been presented to explain the cause tions has been observed and a number of revision proce-
of aseptic loosening based on observational, experimental dures, previously recorded as ‘revision due to aseptic
and clinical studies.15 The main mechanism seems to be loosening’, are now correctly being documented as
the excess production of wear particles, triggering a pro- infections.22
inflammatory reaction which leads to increased osteoclast
differentiation, macrophage production, linear or focal Reaction to metal debris
osteolysis and aseptic loosening (inflammatory-mediated Recent advances in metallurgy and tribology have led to a
osteolysis). Patient susceptibility to aseptic loosening may renewed interest in the use of MoM bearings in THA. With
be affected by host-, genetic-, surgical- and implant- improved fixation techniques, the concept of large head
related factors (their relative importance is not known). femoral components coupled with thin monoblock ace-
tabular components (close to hip joint geometry) were
Dislocation rapidly adopted for MoM hip arthroplasties. Preliminary
Dislocation is a difficult problem for both the patient and results with these second-generation MoM hip resurfac-
the surgeon and its management is expensive for the ings were excellent, and their use became popular in the
healthcare system. The true incidence of post-operative early 2000s.23 Later, large diameter MoM bearings com-
dislocation varies depending on surgical, patient and bined with cementless stems were used. A decade later,
implant factors (0.3% to 3%).16 Most dislocations occur adverse reactions to metal debris (ARMD) came into focus.
in the immediate post-operative period (30% to 60%).16,17 Metal debris, caused by increased wear of the bearing sur-
Patient risk factors include previous surgery, neuromus- face and/or corrosion of the neck-head taper (trunnion) in
cular disorders, dementia, female gender, inability to stemmed THA Downloaded from Bioscientifica.com at 07/14/2024 01:05:38PM
can lead to local soft-tissue reactions such
via Open Access. This article is distributed under the terms of the Creative
Commons Attribution-Non Commercial 4.0 International (CC BY-NC 4.0) licence
(https://creativecommons.org/licenses/by-nc/4.0/) which permits
233
non-commercial use, reproduction and distribution of the work without
further permission provided the original work is attributed.
https://creativecommons.org/licenses/by-nc/4.0/
FREQUENT CAUSES OF PAIN
Fig. 2 Frequent and rare causes of painful total hip arthroplasty (which may lead to revision surgery).
INTRINSIC CAUSES
INFECTION
EXTRINSIC CAUSES
OSTEOLYSIS
ASEPTIC LOOSENING
VASCULAR DISORDER
OCCULT PERIPROSTHETIC FRACTURE
SPINAL DISORDER
STRESS FRACTURE PERIPHERAL NERVE INJURY
MODULUS OF ELASTICITY MISMATCH PERIPHERAL NEUROPATHY
TROCHANTERIC BURSITIS METABOLIC BONE DISEASE
ILIOPSOAS TENDONITIS HERNIA
NEOPLASMATIC DISEASE
Fig. 3 Intrinsic and extrinsic causes of pain following total hip arthroplasty (which may lead to revision surgery).
revision rate in the younger age groups and for women years, infection continues to be the most common reason
under 55 years a cemented ceramic-on-PE THA combina- for early reoperation (two to three years). Comparing two
tion gives the best results (3.8% at ten years). For male sets of three-year periods (2006 to 2008 and 2012 to
patients aged under 55 years, the revision rate (all bearing 2014), septic revisions have increased from 17.5% to
surfaces) is 10%; 3.5% lower than that of female patients. 32.1%. The third-year reoperation rate decreased during
The best recorded THA combination for younger patients the last observation period (2009 to 2011). From 1979 to
is an all cemented implant with ceramic-on-PE bearing, 2009, revision rates increased with few periods of tempo-
which presents half the revision rate when compared with rary falls. A small reduction in rates was then recorded and
cementless implants with metal-on-PE bearings (at all was followed by a constant rate over the last three years.
time intervals). For older patients all THA combinations Revision rates due to aseptic loosening and osteolysis
show satisfactory revision rates. When mortality and revi- have gradually decreased from 72.3% to 52.2% since the
sion rates are evaluated together, it can safely be predicted beginning of 2000 (2001 to 2005). For the same period,
that for most patients over the age of 75 at index opera- the second most common cause for revision is dislocation.
tion, THAs will be retained for a lifetime. After that infection rates increased, and in 2012 disloca-
For both metal-on-PE and ceramic-on-PE bearings, tion exchanged places with infection, with infection rates
higher failure rates are seen with larger head sizes (36 mm steadily increasing in 2013 from 13.9% to 14.6%. Early
for cemented and above 36 mm for hybrid and cement- overall revision rate due to technical errors is around
less THAs). When a CoC bearing is used, survival rates are 1.7%, but in 2014 the rate increased to 2.3%. In this latest
improved with larger head size. registry report, revision rates for reaction to metal debris,
MoM implants, either resurfacing or stemmed, fail at ‘high and unclear pain’ become apparent. Modes of fail-
higher rates than other bearings, with revision rates rang- ure and revision rates also vary with age. Revision rates
ing from 14% to 27% for the worst implants at ten years. due to loosening and osteolysis are relatively constant
The best examples show lower revisions rates at the level (66% of cases up to 84 years of age and 50.1% for those
of 8% to 9%. This has resulted in a dramatic and sustained patients over 84 years of age). Revision rates due to dislo-
reduction in their use. cation and periprosthetic fracture increased with age
The number of primary THAs performed for hip frac- (more apparent in patients older than 85 years). For the
tures is increasing. Revision rates are similar to those of time-period (2005 to 2014), the most common (44.1%)
THAs performed for other reasons, but mortality rates are cause for reoperation is aseptic loosening. The second
higher. most common cause is infection (19.5%), followed by
The most common cause of revision is aseptic loosen- fracture (13.6%) and then dislocation (12.6%). The revi-
ing, followed by pain. Within the first post-operative year, sion rate due to periprosthetic fracture has doubled com-
dislocation, fracture and infection are the most common pared with the previous ten-year period (1993 to 2002,
reasons for revision, while the revision rate for aseptic from 6.8% to 13.6%). The increased use of cementless
loosening increases over the first ten years. Revision rates stems, which have a greater risk of peri-operative or imme-
due to both aseptic loosening and pain increase with time diate post-operative periprosthetic fractures, is a possible
from surgery, while rates due to dislocation, infection, explanation.
periprosthetic fracture and malalignment are all higher
during the first year and then fall. Both adverse reaction to
particulate debris and osteolysis increase with time.
Important considerations
It was found that for 24% of all THA implants available to
Swedish Hip Arthroplasty Registry44 surgeons in the United Kingdom there is no evidence of
Reoperation rates in relation to the total number of pri- clinical effectiveness; these thus may be associated with
mary THAs performed (since 1992) have remained rela- potentially unknown modes of failure. Concern exists
tively stable (12% to 13%) (with a possible under-reporting about the current system of device regulation, and the
of complications). Reoperation rates within two years need for a revised process for introducing new orthopae-
decreased from 3.5% (1992 to 1995) to 1.9% (2000 to dic devices is apparent.45 A higher failure rate of newly
2005). After this, the rate rises but stays at a constant level introduced THA implants with no mid- and long-term
at just over 2%. Patients requiring reoperation became clinical records was also observed in the Australian
older and the proportion of women has increased. These register.41,42
patients are also sicker than those undergoing primary There is wide availability of various cemented designs
surgery and were most commonly operated on due to with long-term proven records of clinical success.41-44
inflammatory joint disease, femoral necrosis and child- However, a surprising number of series have recently
hood hip disorders. The first-year reoperation rate is reported early failures of cemented stems.46 Some designs
mainly due to infection and dislocation. During recent have consistently produced high early failure rates. Others
Downloaded from Bioscientifica.com at 07/14/2024 01:05:38PM
via Open Access. This article is distributed under the terms of the Creative
Commons Attribution-Non Commercial 4.0 International (CC BY-NC 4.0) licence
(https://creativecommons.org/licenses/by-nc/4.0/) which permits
237
non-commercial use, reproduction and distribution of the work without
further permission provided the original work is attributed.
https://creativecommons.org/licenses/by-nc/4.0/
have failed infrequently, but the failures have occurred This article is distributed under the terms of the Creative Commons Attribution-Non
early and with extensive osteolysis. Failures are often mul- Commercial 4.0 International (CC BY-NC 4.0) licence (https://creativecommons.org/
tifactorial and defy a simple explanation based on a single licenses/by-nc/4.0/) which permits non-commercial use, reproduction and distribu-
parameter.46 tion of the work without further permission provided the original work is attributed.
The effect of femoral head size and surgical approach
on the risk of revision for dislocation after THA was evalu- References
ated by the Dutch arthroplasty register.47 Compared with 1. Learmonth ID, Young C, Rorabeck C. The operation of the century: total hip
the posterolateral approach, direct anterior and anterolat- replacement. Lancet 2007;370:1508-1519.
eral approaches reduce the risk of revision for dislocation,
2. Laupacis A, Bourne R, Rorabeck C, et al. The effect of elective total hip
but at the cost of more stem revisions, and revisions for
replacement on health-related quality of life. J Bone Joint Surg [Am] 1993;75-A:1619-1626.
other reasons. With the anterior approach, larger 36-mm
heads increased the risk of revision for other reasons. 3. Morshed S, Bozic KJ, Ries MD, Malchau H, Colford JM Jr. Comparison of
More recently, periprosthetic fractures have increased;41-44 cemented and uncemented fixation in total hip replacement: a meta-analysis. Acta Orthop
this is probably due to the increasing use of cementless 2007;78:315-326.
implants worldwide. For late fractures, age and poor bone 4. Pakvis D, van Hellemondt G, de Visser E, Jacobs W, Spruit M. Is there
stock are predisposing factors. With an aging population, evidence for a superior method of socket fixation in hip arthroplasty? A systematic review.
this failure mechanism will likely continue to increase. Int Orthop 2011;35:1109-1118.
5. Kumar N, Arora GN, Datta B. Bearing surfaces in hip replacement - Evolution and
Conclusions likely future. Med J Armed Forces India 2014;70:371-376.
6. Wang W, Ouyang Y, Poh CK. Orthopaedic implant technology: biomaterials from
As we have come to better understand the failure mecha-
past to future. Ann Acad Med Singapore 2011;40:237-244.
nisms of THA, numerous modifications, such as large fem-
oral heads to improve stability, porous metals to enhance 7. Shah N, Porter M. Evolution of cemented stems. Orthopedics 2005;28(suppl):
fixation and alternative bearings to improve wear, have s819-s825.
been introduced over the last decade in the hope of 8. Callaghan JJ, Brown TD. Back to the future: evolution in hip design. Orthopedics
decreasing the rate of early as well as late failures. Early 1998;21:1007-1008.
failures are most often attributed to either technical errors 9. Kurtz S, Ong K, Lau E, Mowat F, Halpern M. Projections of primary and revision
or early acceptance of alternative surgical techniques or hip and knee arthroplasty in the United States from 2005 to 2030. J Bone Joint Surg [Am]
innovations. Before evidence-based data are available to 2007;89-A:780-785.
justify the risk of their use, care must be taken to make
10. Muirhead-Allwood SK. Lessons of a hip failure. BMJ 1998;316:644.
sure that early adoption of such innovations in either tech-
nique or implant design does not result in an increased 11. Digas G, Karrholm J, Thanner J, Malchau H, Herberts P. Highly cross-
risk of early failure. linked polyethylene in total hip arthroplasty. Randomised evaluation of penetration rate in
cemented and uncemented sockets using radiostereometric analysis. Clin Orthop Relat Res
2004;429:6-16.
Author Information
1School of Health Sciences, Faculty of Medicine, University of Thessalia, University
12. Varnum C. Outcomes of different bearings in total hip arthroplasty - implant survival,
revision causes, and patient-reported outcome. Dan Med J 2017;64:1-21.
General Hospital of Larissa, Mezourlo Region, 41110 Larissa, Greece.
2Orthopaedic Department, University General Hospital of Larissa, Greece. 13. Mai K, Verioti C, Ezzet KA, et al. Incidence of ‘squeaking’ after ceramic-on-
ceramic total hip arthroplasty. Clin Orthop Relat Res 2010;468:413-417.
Correspondence should be sent to: T. Karachalios, School of Health Sciences, 14. Traina F, De Fine M, Bordini B, Toni A. Risk factors for ceramic liner fracture after
Faculty of Medicine, University of Thessalia, University General Hospital of total hip arthroplasty. Hip Int 2012;22:607-614.
Larissa, Mezourlo Region, 41110 Larissa, Greece. 15. Jiang Y, Jia T, Wooley PH, Yang SY. Current research in the pathogenesis of aseptic
Email: kar@med.uth.gr implant loosening associated with particulate wear debris. Acta Orthop Belg 2013;79:1-9.
16. Bozic KJ, Kurtz SM, Lau E, et al. The epidemiology of revision total hip
ICMJE Conflict of interest statement arthroplasty in the United States. J Bone Joint Surg [Am] 2009;91-A:128-133.
None declared. 17. Ullmark G. The unstable total hip arthroplasty. EFORT Open Rev 2017;1:83-88.
18. Seagrave KG, Troelsen A, Malchau H, Husted H, Gromov K. Acetabular cup
Funding statement
position and risk of dislocation in primary total hip arthroplasty. Acta Orthop 2017;88:10-17.
No benefits in any form have been received or will be received from a commercial
party related directly or indirectly to the subject of this article. 19. Girard J. Femoral head diameter considerations for primary total hip arthroplasty.
Orthop Traumatol Surg Res 2015;101(suppl):S25-S29.
Licence 20. Pulido L, Restrepo C, Parvizi J. Late instability following total hip arthroplasty.
© 2018 The author(s) Clin Med Res 2007;5:139-142.
Downloaded from Bioscientifica.com at 07/14/2024 01:05:38PM
via Open Access. This article is distributed under the terms of the Creative
Commons Attribution-Non Commercial 4.0 International (CC BY-NC 4.0) licence
(https://creativecommons.org/licenses/by-nc/4.0/) which permits
238 non-commercial use, reproduction and distribution of the work without
further permission provided the original work is attributed.
https://creativecommons.org/licenses/by-nc/4.0/
Total hip arthroplasty: survival and modes of failure
21. Fitzgerald RH Jr. Total hip arthroplasty sepsis. Prevention and diagnosis. Orthop Clin 35. Fevang BTS, Lie SA, Havelin LI, Engesaeter LB, Furnes O. Improved results
North Am 1992;23:259-264. of primary total hip replacement. Acta Orthop 2010;81:649-659.
22. Kurtz SM, Lau E, Schmier J, et al. Infection burden for hip and knee arthroplasty 36. Herberts P, Malchau H. How outcome studies have changed total hip arthroplasty
in the United States. J Arthroplasty 2008;23:984-991. practices in Sweden. Clin Orthop Relat Res 1997;344:44-60.
23. Zywiel MG, Sayeed SA, Johnson AJ, Schmalzried TP, Mont MA. State of the 37. Herberts P, Malchau H. Long-term registration has improved the quality of hip
art in hard-on-hard bearings: how did we get here and what have we achieved? Expert Rev replacement: a review of the Swedish THR Register comparing 160,000 cases. Acta Orthop
Med Devices 2011;8:187-207. Scand 2000;71:111-121.
24. Liow MHL, Kwon YM. Metal-on-metal total hip arthroplasty: risk factors for 38. Kurtz S, Mowat F, Ong K, et al. Prevalence of primary and revision total hip
pseudotumours and clinical systematic evaluation. Int Orthop 2017;41:885-892. and knee arthroplasty in the United States from 1990 through 2002. J Bone Joint Surg [Am]
25. Esposito CI, Wright TM, Goodman SB, Berry DJ, Clinical, Biological and 2005;87-A:1487-1497.
Bioengineering Study Groups from Carl T. Brighton Workshop. What is the trouble 39. Kurtz SM, Ong KL, Schmier J, et al. Primary and revision arthroplasty surgery
with trunnions? Clin Orthop Relat Res 2014;472:3652-3658. caseloads in the United States from 1990 to 2004. J Arthroplasty 2009;24:195-203.
26. Reito A, Lainiala O, Elo P, Eskelinen A. Prevalence of failure due to adverse 40. Melvin JS, Karthikeyan T, Cope R, Fehring TK. Early failures in total hip
reaction to metal debris in modern, medium and large diameter metal on metal arthroplasty - a changing paradigm. J Arthroplasty 2014;29:1285-1288.
replacements. The effect of novel screening methods: systematic review and Metaregression 41. No authors listed. Australian Orthopaedic Association National Joint
analysis. PLoS One 2016;11:e0147872. Replacement Registry (AOANJRR): 2013 Annual Report. https://aoanjrr.sahmri.com/
27. Günther KP, Schmitt J, Campbell P, et al. Consensus statement “Current documents/10180/127202/Annual%20Report%202013?version=1.2&t=1385685288617
evidence on the management of metal-on-metal bearings”-April 16, 2012. Hip Int 2013;23: (date last accessed 7 March 2018).
2-5. 42. No authors listed. Australian Orthopaedic Association National Joint Replacement
28. Miettinen SSA, Mäkinen TJ, Kostensalo I, et al. Risk factors for intraoperative Registry (AOANJRR): 2017 Annual Report. https://aoanjrr.sahmri.com/annual-reports-2017
calcar fracture in cementless total hip arthroplasty. Acta Orthop 2016;87:113-119. (date last accessed 7 March 2018).
29. Sidler-Maier CC, Waddell JP. Incidence and predisposing factors of periprosthetic 43. No authors listed. National joint registry for England, Wales, Northern Ireland
proximal femoral fractures: a literature review. Int Orthop 2015;39:1673-1682. and the Isle of Man. 2017 Annual Report. http://www.njrcentre.org.uk/njrcentre/
30. Liu KL, Wu WT, Wang JH, et al. When and how do prosthetic hips fail after total Reports,PublicationsandMinutes/Annualreports/tabid/86/Default.aspx(date last accessed
hip arthroplasties?-A retrospective study. J Formos Med Assoc 2016;115:786-793. 7 March 2018).
31. Prokopetz JJZ, Losina E, Bliss RL, et al. Risk factors for revision of primary total 44. No authors listed. Swedish Hip Arthroplasty Register Annual Report 2014. https://
hip arthroplasty: a systematic review. BMC Musculoskeletal Disorders 2012;13:251: registercentrum.blob.core.windows.net/shpr/r/Annual-report-2014-BJv-q8pil.pdf (date
32. Cherian JJ, Jauregui JJ, Banerjee S, Pierce T, Mont MA. What host factors last accessed 7 March 2018).
affect aseptic loosening after THA and TKA? Clin Orthop Relat Res 2015;473:2700-2709. 45. Kynaston-Pearson F, Ashmore AM, Malak TT, et al. Primary hip replacement
33. Schwartz FH, Lange J. Factors that affect outcome following total joint prostheses and their evidence base: systematic review of literature. BMJ 2013;347:f6956.
arthroplasty: a review of the recent literature. Curr Rev Musculoskelet Med 2017;10: 46. Barrack RL. Early failure of modern cemented stems. J Arthroplasty 2000;15:1036-1050.
346-355. 47. Zijlstra WP, De Hartog B, Van Steenbergen LN, Schreurs BW, Nelissen
34. Abdulkarim A, Ellanti P, Motterlini N, Fahey T, O’Byrne JM. Cemented RGHH. Effect of femoral head size and surgical approach on risk of revision for dislocation
versus uncemented fixation in total hip replacement: a systematic review and meta-analysis after total hip arthroplasty. An analysis of 166231 procedures in the Dutch Arthroplasty
of randomised trials. Orthop Rev 2013;5:34-44. Register (LROI). Acta Orthop Scand 2017;88:395-401.