Professional Documents
Culture Documents
Steam Sterilization Validation 4
Steam Sterilization Validation 4
TECHNICAL NOTE
Sanjeev patel
sanjeev@gampservices.com
A Brief Overview of Steam Sterilization Cycles:
Cycle development is the process of determining the physical properties
to be met in a sterilization cycle that will be used to reproducibly and
consistently sterilize the product, component, and/or equipment in a
defined loading pattern.
2. Once the cycle type is determined you must decide on the key cycle
parameters: sterilization temperature and sterilizer time.
3. Once you have determined your cycle type and parameters, you
must validate the cycle to ensure the load will be sterile.
5. Typically, at least 10 BIs are used per cycle. If all BIs show no growth
(i.e. all the bacteria are destroyed) in three successive cycles, you
have validated the cycle.
All loads cannot be run with the same sterilization cycle. This is why
autoclaves are equipped to run various cycle types.
Once the cycle type is determined you must decide on the key
cycle parameters: sterilization temperature and sterilization time.
The vast majority of loads sterilized in the laboratory setting are
sterilized at 121.1°C/250°F for 30-90 minutes.
If your load contains items that are sensitive to heat, the default
cycle temperature of 121.1°C/250°F may be too hot, and you might
need to consider a lower temperature. Using lower sterilization
temperatures will result in longer sterilization times. This can be
calculated by using the F0 relationship:
F0 is the accumulated lethality in minutes at 121.1°C/250°F.
t = the amount of time in minutes of the sterilization phase of the sterilization
cycle.
T = the sterilization temperature in °C.
The F0 equation solves to:
Let’s take an example by running the numbers for a sterilization cycle using
110°C as the sterilization temperature. To solve for t (time), plug in an F0 value
of 15 along with the desired T (110°C) on the right side of the equation:
t = 193 minutes
As you can see, lower temperature cycles take more time to achieve
sterilization. As a comparison, at 121.1°C, this cycle would theoretically take
15 minutes to achieve sterilization.