Professional Documents
Culture Documents
(Handbook of Exploration Geochemistry 1) W.K. FLETCHER (Eds.)-Analytical Methods in Geochemical Prospecting-Elsevier Science Ltd (1981)
(Handbook of Exploration Geochemistry 1) W.K. FLETCHER (Eds.)-Analytical Methods in Geochemical Prospecting-Elsevier Science Ltd (1981)
VOLUME I
Analytical Methods in
Geochemical Prospecting
by
W.K.FLETCHER
Associate Professor
Department of Geological Sciences
University of British Columbia, Vancouver, B.C., Canada
ELSEVIER/NORTH-HOLLAND INC.
52, Vanderbilt Avenue
New York, N.Y. 10017
tion techniques and the effect they have on precision and accuracy (as well
as the precision and accuracy of the actual analytical method) are of funda-
mental importance to the correct interpretation of analytical data. These
topics are emphasized and discussed in detail.
In conformity with the general objective of the Handbook series, Dr.
Fletcher also provides the theoretical background to the analytical tech-
niques he describes. This volume is therefore also a comprehensive reference
text on analytical methods in exploration geochemistry for the professional
geochemist and university student.
As the editor of the Handbook I am particularly pleased with this, the
first volume. An important facet of the volume — and one that distinguishes
it from other treatises on analytical methods — is the liberal illustration of
the analytical techniques by results from actual exploration geochemical
surveys. The constant reference to a large number of such surveys and the
careful interpretation of the results is an excellent example of the approach
which the Handbook hopes to achieve in all of its volumes.
G.J.S. GOVETT
Sydney, N.S.W., Australia
December, 1980.
PREFACE
W.K. FLETCHER
Vancouver, B.C.
Chapter 1
INTRODUCTION
TABLE 1-1
Census of geochemical samples collected in North America (from A.E.G. Analysis Com-
mittee, 1971, unless otherwise indicated)
(Figs. 1-9, 1-10 and 1-11) with freedom from interferences over a wide range
of compositional variability. The need to handle large numbers of samples
(100—1000 per day), with short turnaround times and at minimum cost,
imposes further constraints on choice of methods and also requires careful
organization of laboratory operations.
10 000
z 1000
o
ioo H
cc
LU
Q_
(Λ
CC
Fig. 1-1. Abundance of some trace elements in soils. Unshaded sections indicate more un-
usual values; abnormally high values found in proximity to ore deposits have been ignored.
(From Mitchell, (1964a, in: F.E. Bear (Editor), Chemistry of the Soil, 2nd ed. © 1964 by
Litton Educational Publishing, Inc. Reprinted by permission of Van Nostrand Reinhold.)
TABLE l-II
Chemical composition (%) of some igneous and sedimentary rocks (based on Ahrens, 1965, unless otherwise indicated)
3
Con- Ultrabasic Basalts Inter- Granites Shales Sand- Lime- Laterite
! 2
stituent rocks mediate stones stones
rocks
HCIO4CU \ *
i
N
- Vx V"*" x
UMW-.\\)^
^ w </·» H I >S-M
Fig. 1-2. Perchloric acid- and EDTA-extractable Cu in stream sediments, Fiji. Both extrac-
tions outline the area of anomalous soils over known mineralization. However, with per-
chloric acid the limit of the third-order anomaly, defined by a distinct break at 70 ppm in
the histogram and indicated on the map by a dashed line, extends upstream of the soil
anomaly and is much more extensive than the EDTA-extractable Cu anomaly. This broad
anomaly is apparently geologically controlled and related to the distribution of un-
mineralized intrusive rocks with slightly raised Cu contents. (From Bradshaw et al., 1974.)
5
advantage such partial, rather than total, analyses of the sample can enhance
geochemical patterns related to mineralization (Fig. 1-2) and also enable the
geochemical dispersion of an element to be studied. Conversely, inappropri-
ate analytical methods can mask geochemical patterns of interest.
The benefits of close liaison between the field and laboratory should be
obvious, ideally the analyst-geochemist should be one person. However, as
noted in a 1968 discussion of "What is a Geochemical Analysis?" (Hansuld
et al., 1969), communication between geologist and analyst is often lacking
to the overall detriment of many geochemical programmes. It is hoped that
this text will provide field-orientated geologists and geochemists with useful
insights into the methods available for analysis of their samples, the princi-
ples behind those methods, and their relative merits and limitations. No
specific analytical methods are described because many excellent publications
already fulfill that purpose. Convenient compilations of analytical methods
used in geochemical prospecting are given by Stanton (1966, 1976), Ward
(1975) and Ward et al. (1963, 1969).
Geochemical considerations
Atomic absorption
Plasma emission
Colorimetry
υ
O
DC-arc emission
XRF
Atomic absorption
Plasma emission
Colorimetry
DC-arc emission
XRF
"4
Atomic absorption
Dry at 110°C Acid Attack
Plasma emission
Colorimetry
DC-arc emission
XRF
Plasma emission
Atomic absorption
Colorimetry
Precipitation
Fig. 1-3. Some of the pathways for preparation, dissolution and analysis of exploration
samples.
Relative pH of Environment
Mobility Acidic Alkaline
Very Mobile
Θ {ä}
S, Ca, Mg, Na
E
Ω. Immobile element Mobile element (hydromorphic)
a anomaly anomaly
fr—7-^^--Soii
W
Fig. 1-5. Idealized separation of mobile and immobile elements in the geochemical land-
scape.
/
1 ^V
LEGEND
Niobium ppm
BACKGROUND VALUES
Formation Threshold Samples
Tokio 7 11
Trinity :
Siltstone 3.5 8
Limestone 3.5 9
Fig. 1-6. Dispersion of Nb in (A) the minus 10/plus 20-, and (B) minus 80- and minus 300-mesh fractions of stream sediments asso-
ciated with kimberlite pipes, Arkansas. (From Gregory and Tooms, 1969.)
9
Fig. 1-7. Distribution of Sn in (A) minus 10/plus 35-mesh till and (B) the minus 10/plus
35-mesh heavy mineral fraction of the till, Mount Pleasant, New Brunswick. (From Szabo
et al., 1975.)
Assuming that the choice of an analytical method has not been severely
curtailed by geochemical considerations, the principal analytical factors that
determine if a particular method is worthy of further consideration are its
sensitivity, freedom from interferences, and its reliability and reproducibil-
ity. Final selection of a method might then be based on organizational and
administrative factors such as the numbers of samples to be analyzed and the
number of elements to be determined; location of the laboratory; capital
and operating costs; availability of supplies; and the skills of the staff.
Analytical sensitivity
The detection limits attainable with AAS, DC-ES and XRF are considered
in relation to average crustal abundances of the elements in Figs. 1-9, 1-10
and 1-11. None of the methods provides adequate sensitivity for all elements
11
Fig. 1-8. Extraction of Cu from stream sediments with (A) cold 6 M hydrochloric acid,
and (B) hot 4 M nitric acid, Coppermine Basalt Belt, Northwest Territories. The cold
extraction gives the best contrast. (From Allan and Hornbrook, 1971.)
KEY
0 - can be determined by flame emission
hydride generation
H n - nitrous oxide - acetylene flaine He
p Be
s
*
- solvent extraction
- cold vapour
B C N 0 F Ne
11Mg mh
ίΑΠ ::,?-.U
P S Cl Ar
11 Sc
1 m\n
jjCrJ s K sH s B Ga Ge am Br Kr
Cs mi
• Sr;:;
La Hf Ta Si Re Os Ir Pt •:Au: Tl W£ Bi Po At Rn
•:Ba:; w
Fr Ra Ac
Ce Pr Nd Pm Sm Eu Gd Tb Dy Ho Er Tm Yb Lu
Th Pa U Np Pu Am Cm Bk Cf Es Fm Md Lw
Fig. 1-9. Analysis of exploration samples by AAS. Elements most suitable for determina-
tion by AAS shown in stipple; bold face letters indicate that the concentration of the ele-
ment can normally be estimated without difficulty after sample decomposition with
strong acids. Small letters indicate that special operating conditions are required or
recommended as shown in the key.
absorption but is superior for some refractory elements (Nb and W) and also
for Bi, Sb and Sn. Comparable and possibly better detection limits than
those attainable with AAS appear to be possible with ICP-ES, although
published data on analyses of exploration samples by this method are still
limited. Detection limits with XRF are generally inferior to those with AAS
or ES. Nevertheless, XRF is capable of determining concentrations at least as
low as 10 ppm for most elements heavier than Ca: it is particularly well
suited to the rapid determination of major elements and the more abundant
trace elements, and to the analysis of heavy mineral separates.
For special determinations the analyst may have to resort to analytical
techniques of exceptional sensitivity. These include neutron activation,
mass spectroscopy and graphite furnace atomic absorption. However, with
the exception of delayed neutron counting in the determination of U
(Garrett and Lynch, 1976), these techniques are normally too time con-
suming and expensive for routine analysis.
13
H He
01ill 111 C N 0 F Ne
Na 11 ÖH Z'pi s Cl Ar
Rb
11 IYI Wtä ΖΓ| gNbi; Tc Ru Rh Pd
3 Cd In | s n | £sP Te 1 Xe
Cs
ill pjLaJlJ Hf Ta nyy* Re Os Ir Pt Au Hg Tl fpbi; 110 Po At Rn
Fr Ra Ac
Ce Pr Nd Pm Sm Eu Gd Tb Dy Ho Er Tm Yb Lu
Th Pa U Np Pu Am Cm Bk Cf Es Fm Md Lw
Fig. 1-10. Analysis of exploration samples by DC-ES. Emission lines of elements in bold
face and stipple are almost invariably present in the DC-arc spectra of geochemical
samples: elements in stipple only are often visible and those in half-stipple are occa-
sionally visible. Remaining elements either cannot be determined by ES or are only very
rarely seen in emission spectra of rocks, soils and sediments.
H He
Li Be B C N 0 F Ne
Fr Ra Ac
iThi: Pa s-us Np Pu Am Cm Bk Cf Es Fm Md Lw
Interferences
Materials collected in geochemical prospecting programmes vary consider-
ably and often unpredictably in their trace element and bulk chemical com-
position. Moreover, their bulk composition is seldom either known or deter-
mined. Consequently, it is usually impossible to match the composition of
standards and samples. This gives rise to a variety of matrix interferences
which often tend to reflect or follow changes in bedrock geology (and are
therefore not always a disadvantage). In addition both ES and X R F are
especially subject to spectral interferences caused by coincident or over-
lapping emission lines.
No analytical method is free of interferences. Nevertheless, compared to
DC-ES and X R F , AAS has considerable advantages in this respect as a result
of the comparative simplicity of absorption compared to emission spectra. In
addition, with AAS the few matrix interferences of sufficient severity to be a
problem in analysis of exploration samples, for example the suppression of
Mo by Ca, are relatively easily avoided by simple modifications to the com-
position of the sample solution. Correction of interferences therefore
involves no additional calculations or data manipulation.
Modification of the sample matrix is also used to reduce interferences in
DC-ES by mixing the sample powder with graphite and spectroscopic buffers
— usually salts of the alkali or alkali earth elements. Even with this precau-
tion interferences, attributable to matrix variability, could still amount t o as
much as ±30 to ±100% for samples of unusual composition unless further
corrections are applied. This is not practical in rapid analysis when the emis-
sion spectra are to be recorded photographically and their intensities esti-
mated visually: such methods are therefore best regarded as semi-quantita-
tive. If a multi-channel direct-reading spectrometer is used for the analysis,
interferences can be corrected by use of internal standards and by simulta-
neous estimation of the bulk composition of the sample. Access to a com-
puter is, however, necessary to calculate the corrections for matrix and spec-
tral interferences with reasonable efficiency.
Compared to flames and DC-arcs, the ICP has several advantageous
characteristics which make its results less susceptible to matrix effects
although spectral interferences are still present, and background emission
and stray light in the spectrometer can be a problem. Because of the
enormous analytical capacity of the ICP-ES, data handling and interference
corrections require at least a small dedicated computer.
In XRF analysis there is a relatively simple relationship between the bulk
composition of a sample, expressed as its mass absorption coefficient, and
the intensity of the characteristic X-rays it emits. Consequently, to correct
for variations in bulk chemical composition, it is necessary to obtain the
relative mass absorption coefficients of samples and standards. A simple,
elegant method of doing this with sufficient reliability for most exploration
analyses has been described by Feather and Willis (1976). If more than one
15
Sample throughput
If samples are digested in test tubes approximately one hundred can be
16
TABLE l-III
The ratio (R) of overall variance to analytical variance and analytical precision (P) for
analyses of minus 80-mesh stream sediments by emission spectroscopy and atomic ab-
sorption. Overall variance was estimated for 20 duplicate pairs of sediments collected
from a single stream draining sandstones, and P by 10 replicate analyses of one of these
samples. Results indicate that sampling errors are significantly greater than analytical
errors when atomic absorption is used. With semi-quantitative emission spectroscopy
analytical errors (except for Sn) are greater than sampling errors (data from Howarth and
Lowenstein, 1971)
Ä1 P2 R P
Cu 6.8 5 - 3 2
Mn 92 14 (3.9) 42
Ni 7.1 8 - 8 8
Pb not detected — 68
Sn not determined 65 641
Zn 37 6 no data
TABLE 1-IV
Evaluation of performance of analytical methods commonly used in exploration geochemistry (see Figs. 1-9, 1-10 and 1-11 for
analytical sensitivities)
TABLE 1-V
Methods of estimation 1970—1971 (from A.E.G. Analysis Committee, 1 9 7 1 )
and Washington (1962), Holman and Durham (1967), Kvalheim (1967), and
Kinson and Belcher (1970).
In making the final choice of an analytical method other factors which
must, of course, be considered include (1) the capital and operating costs of
equipment and the facilities available for servicing and repairing it; (2) costs
and availability of reagents or suitable local substitutes; and (3) the extent to
TABLE 1-VI
Approximate costs of sample preparation and geochemical analysis by commercial labo-
ratories, Vancouver, B.C., Canada, 1979—1980
Special analyses
Sb: solvent extraction/atomic absorption $ 3.00
As: hydride generation/atomic absorption $ 3.00
Au: solvent extraction/atomic absorption $ 3.00
Sn: solvent extraction/atomic absorption $ 3.00
U: fluorimetric $ 3.00
21
which professional staff are essential to set up and supervise the operation of
the equipment (in general, maximum use should be made of non-professional
staff for all routine analysis). A qualitative attempt has been made to
evaluate the principal analytical methods with respect to these and other fac-
tors in Table 1-IV; the impact of these factors on usage of these methods is
reflected in Table 1-V and on analytical cost in Table 1-VI. It should be
noted that the adequacy of equipment maintenance and reagent supply
varies greatly throughout the world: with respect to instrumentation it is
often easier to repair older units locally than state-of-the-art microprocessors.
As an exploration technique, geochemical prospecting has the outstanding
virtue that systematic use of even the simplest equipment can successfully
lead to the discovery of both exposed and concealed orebodies.
REPORTING RESULTS
TABLE 1-VII
Comparison of six stream sediment analyses for nickel (from Hansuld et al., 1969)
Sample description
stream sediment containing 0.5% magnetite
magnetite contains average of 0.28% Ni
80% of magnetite is minus 100 mesh
99% of sample is plus 100 mesh and 96% is plus 80 mesh
:
Emission Spec. : 10 elements = 20 elements Other
Special instructions:
Date:
Sample No. Location Sample No. Location
SAFETY
TABLE 1-VIII
Hazard Precautions
Dust during rock crushing and work in well ventilated area with efficient dust
grinding or sieving extraction system (see p. 48)
Flammable liquids and vapours store minimum quantities in approved cabinets;
(MIBK, benzene, acetone, etc.) work in fume hood; fire fighting equipment
readily available
Toxic vapours (benzene, carbon avoid use if possible; use only in well ventilated
tetrachloride, chloroform, toluene, areas and in fume hoods
etc.)
Radiation: X-rays check equipment for X-rays leakage with a
scintillometer; regular operators wear
dosimeters
radon gas avoid working on high-grade uranium samples
in poorly ventilated areas
1
Guidelines only — consult Hazards in the Chemical Laboratory (Muir, 1977) or similar
manual for detailed information. See p. 63 for use of perchloric acid.
Chapter 2
INTRODUCTION
>- _ _
u C. p x D. p x
' JA CONCENTRATION
" ►
Fig. 2-1. Random and systematic errors. The variation in concentration caused by random
errors is represented by the normal curve with an average value X: μ is the true concentra-
tion of the analyte. A. The dispersion (width) of the normal curve is narrow and sym-
metrical around μ — i.e. X = μ and results are both accurate and precise. B. The disper-
sion is greater but still symmetrical about μ — precision is therefore relatively poor and
although the average value (X) is accurate, this is not necessarily true of individual analy-
ses. C. Dispersion is narrow but a systematic positive error has been introduced (X > μ) —
results are precise but inaccurate. D. Systematic error and poor precision.
26
random errors arising from the variations inherent to any sampling or mea-
surement process, and non-random errors causing systematic negative or
positive deviations from the true result. Accuracy, which is the closeness of a
result to its true value, is dependent on both random and systematic errors,
whereas precision, which is the ability to obtain the same result repeatedly,
is a measure of random errors alone. It follows that results can be precise
without being accurate (Fig. 2-1).
Exploration geochemistry utilizes geochemical patterns, that is natural
variation of element abundances, as a guide to the presence or absence of ore
mineralization. Relative abundances of elements are therefore more impor-
tant than their absolute concentrations and strictly, it is only necessary that
any variability introduced during sample collection and analysis be signifi-
cantly less than the variability sought (Fig. 2-2). Consequently, quality con-
trol in the exploration laboratory, unlike the assay office, usually emphasizes
relative variability (i.e. precision) over accuracy. Development of exploration
geochemistry has therefore followed, and encouraged, development of
simple, rapid analytical methods capable of achieving adequate precision.
2>
<
cc
ill
o
z
o
o
VEIN
Fig. 2-2. Influence of analytical precision on anomaly contrast. Noisy data (A) arising
from random errors in sampling or analysis obscures the anomaly. Although absolute con-
centrations are lower in (B) and (C), the data are less noisy and the anomaly contrast
improved.
27
RANDOM ERRORS
Precision
Pc = — c -x 100%
c
od = y/2oc
28
d = 1.128ac
Md = 0.954a c
where Md is the median value for the difference. oc can be obtained from
each of these relationships but the median (Md) is the most convenient
estimator because it is (1) relatively little affected by wild values; (2) readily
estimated graphically; and (3) corresponds very closely to oc without further
calculation. The following rapid procedure is suggested for estimation of
precision from a minimum of 50 pairs of duplicates (Thompson and
Howarth, 1976).
(1) From the duplicate analyses obtain a list of the means (Xx + X2)/2 and
absolute differences Ι-ΧΊ — X21.
(2) Arrange the list in increasing order of concentration means.
(3) From the first 11 results obtain the mean concentration and median
difference for that group.
(4) Repeat this for each successive group of 11 results, ignoring any
remainder less than 11.
(5) Calculate, or obtain graphically, the linear regression of the median
differences on the means and multiply the intercept and coefficient by 1.048
(i.e. 1/0.954) to obtain σ0 and k, respectively. As an example the variation of
oc with c for the determination of copper in soil samples, based on 50 dupli-
cate analyses, is shown in Fig. 2-3. From the regression oc = σ0 + kc so that
Pc is given by:
KEY
* Individual points
O Interval median
50
-*i_
100 +
150
(X!+X2) / 2
Fig. 2-3. Regression of the median differences of \Χχ — X2 I against average concentration
(Xi + X2)/2 for duplicate pairs of analyses. Vertical dashed lines separate intervals con-
taining eleven pairs of analyses and their median value.
/ s*
— —
•
•
•
»
U 2
X ·*.
•
•
1
10 20 50 100 200 500 1000
(Xi+X2)/2
Fig. 2-4. A precision control chart for P = ±10% at the 95% confidence level. See text for
explanation of data points.
close to the detection limit and values below the limit are set to zero,
recorded as "less than" or are set to the detection limit; and (4) if systematic
differences (bias) arise between batches (Thompson and Howarth, 1976).
Detection limits
100000
190000 .010000
271000 .028000 .001000
3-43900 .052300 .003700 .000100
4095,10 .081460 .008560 .000460 .Ü00010
686189 .302643 .089562 .018535 .002761 .000296 .000023 .G00001 .JCC3ÜC :.:ccccc
717570 .340998 .110870 .025637 .004329 .000641 .000060 .000003 .200300 CC02C2
7-46813 .378655 .133883 .034161 .006460 .000920 .000099 .000008 . JCOCGC 3CCCCC
771232 .415371 .158360 .044133 .009230 .001474 .000181 .00001 7 .2000c: 22-0222
79-4109 .450957 .184061 .055556 .012720 .002250 .000311 .000034 ..'J000C3 222222
81-4698 .485272 .210751 .068406 .017004 .003297 .000606 .000061 . ICCCCb 222300
833228 .518215 .238203 .082641 .022144 .00466 7 .000 7^4 .000106 -ocoo:: 222C21
8-49905 .549716 .266204 .098197 .028194 .006416 .001172 .000173 .JUÜ021 3C0022
864915 .579735 .294555 .114998 .035194 .008693 .001696 .000273 .000036 000004
878423 .608253 .323073 .132953 .043174 .011253 .002386 .000416 ,000060 CCC2C?
890581 .635270 .351591 .151965 .062162 .014446 .0032 73 .000613 .000096 oooc: 2
901523 .660801 .3 79959 .171928 .062134 .018216 .004390 .0008 79 .00014/ 0CCC21
911371 .684873 .408043 .192731 .073113 .022608 .006773 .001230 .000220 000033
920234 .707523 .435726 .214262 .085075 .027658 .007466 .001684 .000 321 000062
928210 .728794 .462906 .236409 .097994 .033400 .009476 .002261 .000468 OCCC79
935389 .748736 .489495 .259058 . 111835 .039859 .011869 •0D2983 .000638 CCC11 ?
941850 .767401 .515419 .282102 .12655 7 .04706 7 .014668 .0038 7 1 .0008 72 000169
947665 .784846 .540617 -.305434 .142112 .055007 .017907 .004961 .001172 C00239 2C0C4?
952899 .801128 .565040 .328952 .158444 .063717 .021617 .00624 7 .001660 3CQ333 200062
957609 .816305 .588649 .352561 .175495 .073190 .025827 .007784 .-.702020 0OC464 202089
961848 .830435 .611414 .376170 .193201 .083421 .030563 .009688 .002696 000611
965663 .843577 .633316 .399694 .211498 .094399 .035849 .011686 .003296 0C0809
969097 .855785 .654342 .422066 .230318 . 106109 .041704 .014102 .004134 001068
972187 .867116 .674487 .446185 .249592 .118530 .048144 .016862 .006131 001366
974968 .877624 .693750 .469015 .269251 .131636 .066183 .019990 .006304 001740
977472 .887358 .712137 .491489 .289227 .146397 .062828 .023509 .00 7673 3C2198 3666
979724 .896369 .729659 .613553 .309454 . 159780 .071086 .027441 .009256 CC2 7->6
981752 .904705 .746330 .635164 .329864 .174748 .079955 .031806 .011076 0C3397 192 3
983577 .912410 .762168 .566281 .350394 .190259 .089434 .036621 .C1314Ö 20-Ί66
985219 .919526 .777192 .576869 .370982 .206273 .099616 .041302 .016496 226263
11 .104662 .005180 .000165 .000003 .000000 .000000 .000000 .000000 .000000 .000000 .000000
12 .113615 .006175 .000206 .000006 .000000 .000000 .000000 .000000 .000000 .000000 .000000 .oooooo
13 .122479 .007249 .000266 .000007 .000000 .000000 .000000 .000000 .000000 .000000 .000000 .oooooo
14 131254 .008401 .000336 .000009 .000000 .000000 .000000 .D00000 .000000 .000000 .000000 .oooooo
15 .139942 .009630 .000416 .000012 .000000 .000000 .000000 .000000 .000000 .000000 .000000 .oooooo
16 .148642 .010933 .000508 .000017 .000000 .000000 .000000 .000000 .000000 .000000 .000000 .000000
17 . 157057 .012309 .000612 .000021 .000001 .000000 .000000 .000000 .000000 .000000 .000000 .000000
18 .165486 .013756 .000729 .000027 .000001 .000000 .000000 .000000 .000000 .000000 .000000 .oooooo
19 .173831 .015274 .000859 .000034 .000001 .000000 .000000 .000000 .000000 .000000 .000000 .oooooo
20 .182093 .016859 .001004 .000043 .000001 .000000 .000000 .000000 .000000 .000000 .000000 .oooooo
21 .190272 .018512 .001162 .000052 .000002 .000000 .000000 .000000 .000000 .000000 .000000 .oooooo
22 .198369 .020229 .001336 .000063 .000002 .000000 .000000 .000000 .000000 .000000 .000000 .000000
23 .206386 .022011 .001525 .000076 .000003 .000000 .000000 .000000 .000000 .000000 .000000 .oooooo
24 .214322 .023854 .001729 .000091 .000004 .000000 .000000 .000000 .000000 .000000 .oooooo .oooooo
25 .222179 .025759 .001951 .000107 .000004 .000000 .000000 .000000 .000000 .000000 .oooooo .oooooo
26 .229957 .027723 .002189 .000125 .000006 .000000 .000000 .000000 .000000 .000000 .oooooo .oooooo
27 .237657 .029746 .002444 .000146 .000007 .000000 .000000 .000000 .000000 .000000 .oooooo .oooooo
28 .245281 .031825 .002717 .000169 .000008 .000000 .000000 .000000 .000000 .000000 .oooooo .oooooo
29 .252828 .033959 .003008 .000194 .000010 .000000 .000000 .000000 .000000 .000000 .oooooo .000000
30 .260300 .036148 .003318 .000223 .000012 .000000 .000000 .000000 .000000 .000000 .oooooo .ocuooo
31 .267697 .038390 .003646 .000254 .000014 .000001 .000000 .000000 .000000 .000000 .oooooo .oooooo
32 .275020 .040683 .003993 .000287 .000016 .000001 .000000 .000000 .000000 .000000 .000000 .oooooo
33 .282269 .043026 .004360 .000325 .000019 .000001 .000000 .000000 .000000 .000000 .oooooo .oooooo
34 .289447 .045418 .004747 .000365 .000022 .000001 .000000 .000000 .000000 .000000 .oooooo .oooooo
35 .296552 .047859 .005154 .000409 .000025 .000001 .000000 .000000 .000000 .000000 .oooooo .000000
36 .303587 .050346 .005581 .000456 .000029 .000002 .000000 .000000 .000000 .000000 .oooooo .oooooo
37 .310551 .052878 .006028 .000507 .000033 .000002 .000000 .000000 .000000 .000000 .000000 .oooooo
38 .317445 .055455 .006497 .000563 .000038 .000002 .000000 .000000 .000000 .000000 .oooooo .oooooo
39 .324271 .058075 .006986 .000622 .000043 .000002 .000000 .000000 .000000 .000000 .oooooo .000000
40 .331028 .060737 .007497 .000686 .000049 .000003 .000000 .000000 .000000 .000000 .000000 .000000
32
Pc%=(^+2k)l00
σ0 =5
z k =0.05
o
CO
Ü
111 \
cc
Q.
20
100 k U·-
CONCENTRATION, c
Fig. 2-5. Variation of precision with concentration for OQ = 5 and k = 0.05. The detection
limit (d = 11) is the concentration corresponding to P = ±100%.
So far only the overall analytical precision, arising from random errors
throughout sample preparation, decomposition and measurement, has been
considered. Occasionally it may be necessary to determine the relative
magnitude of errors generated at each step of the analysis. By including
duplicates or replicates at appropriate stages in the analysis the necessary
estimates can be made from an analysis of variance.
As a simple example Table 2-II tabulates results of duplicate (; = 2) analy-
ses of a sample in each of six (i = 6) batches. We wish to evaluate the
possibility that errors between the batches are significantly greater than
those within batches. First, variance between batches (i>|) and within
batches (S^) is estimated from the corresponding sums of squares, SSB and
SSW, given by the relationships:
TABLE 2-II
Analysis of variance for duplicate analyses from six batches of analyses
n
Means
Q 41.7 39.6 39.9 43.2 42.6 36.6
c 40.6
then:
SSB SSW
and Sw =
«-s N-k
where k is the number of batches (/), n the number of replications (;) and N
the total number of analyses: C is the overall average concentration and C,
the average of the n replications on the ith batch. The F ratio (F) is then
computed as:
Ei _ larger variance
smaller variance
analytical conditions from day to day, or in the relative skill and care of
different analysts.
Analysis of variance, even in its simplest forms, is very versatile. Batches in
the worked example might equally well represent different laboratories or
analysts. Alternatively, by digesting the same sample several times and
analyzing each of the resulting solutions in duplicate, conclusions could have
been drawn about the relative contribution to overall errors of subsampling
and decomposition compared to the measurement step. Having identified the
relative magnitude of errors arising at each stage, efforts to improve overall
precision should focus on the sources of the greatest error — obviously in
Table 2-II considerable reduction of between batch error is needed to reduce
it to the level associated with within batch variations. By extending duplica-
tion to field sampling, total data variance can be partitioned between
regional variability, local (sampling) variability and analytical (laboratory)
variability. The probable significance of the regional patterns can then be
estimated (Plant, 1971; Plant et al., 1975; Chork, 1977; Howarth and
Lowenstein, 1971; Garrett, 1969; Garrett and Goss, 1979). As noted in
Chapter 1, there is little point expending effort to improve laboratory preci-
sion if errors generated during sample collection dominate the data.
Random errors can often be reduced, albeit at the expense of time and
cost, by more careful analysis: for example, by use of volumetric glassware
rather than calibrated test tubes, by more rigorous control of operating con-
ditions or by closer bracketing of unknowns with standards. Sometimes,
particularly in emission spectroscopy, rationing the analyte signal to that of
an internal standard, having the same characteristics in the analytical system
as the analyte, enables the effects of fluctuating conditions to be com-
pensated (Fig. 2-6).
Errors can also be reduced statistically. If the standard deviation for a
single measurement is S, the standard deviation Sa for the average of a series
of n measurements is:
Sa = S/y/n
Laboratory sampling
Si 252.8
Be 249.4
ISi 252.8
cc I Be 249.4
, o
z
UJ
I-
—I 1 1 1 1 1 1 —
i 1 1 1 —
i 1 1 1 1
5 10 15 20
NUMBER OF ARCINGS
Fig. 2-6. Use of an internal standard to correct for- poor reproducibility in the DC-arc.
(Reproduced from Ahrens and Taylor, 1961, Spectrochemical Analysis, 2nd ed., with
permission of Addison-Wesley, Advanced Book Program.)
so that the standard deviation, £,·, for weight Wj can be related to standard
deviation, Sh at weight Wi by:
s,-*· y/Wj/Wi
3
tions. Some examples are described by Howarth and Lowenstein (1971) and
Plant (1971).
With multi-component mixtures in which the abundance of the analyte
differs appreciably between components, the effects of sampling are more
complex. In the simplest case with two-component mixtures the binomial
distribution can be used to estimate the relative standard deviation (R).
Using the nomenclature of Ingamells and Switzer (1973) and Ingamells
(1974):
D _100y/pq B-H
y/n K
w
n
" dour V)3
where w is sample weight in grams, d is density and μ the effective mesh size
in microns. If densities of the mineral species differ n can be obtained more
accurately from:
= WB
+ WH
3
dß(io-V) dH(io-V)3
although this correction will be negligible if B is only present as rare grains
of an accessory mineral. The relationship between volume proportions (p, q)
and weight proportions (p w , qw) with minerals of different densities is given
by:
_ dBpw _ dpq^
dßPw + dHqw dBpw + dHqw
These relationships have been used to derive R for different mesh sizes
and different proportions (q) of an accessory mineral (B) containing 80% of
the element of interest compared to 1 ppm in the gangue (H) (Fig. 2-7).
Densities dB and dH have been taken as 7.0 and 3.0, respectively, and the
model would correspond approximately to a soil or sediment containing rare
grains of cassiterite. The same assumptions and values have been used to esti-
mate R at different sub-sample weights in Table 2-III. It is apparent that
sampling errors increase to a maximum at the concentration at which both
37
100
80
60
o\
cc
20
/'.'
υ
/^
P ' 100 ' ' 200
mesh 200 100 80
Grain Size
Fig. 2-7. The relative error (R%) for an element (X) present as a major constituent (80%)
of a trace mineral. Concentration of the element in the gangue is 1 ppm, sub-sample size
is 1 g, and the densities of the gangue and special mineral are 3 and 7, respectively.
Dashed curves indicate the relative proportions of grains of the special mineral (q) and
gangue (p). Curve A: q = 4 X 10" 4 ^ 7 5 0 p p m X ; B : q = 4 X 10" 5 ^ 7 5 p p m X ; C : q = 2 X
10" 5 ^ 40 ppm X; D: q = 4 X 10" 8 ^ 1 ppm X; E: q = 4 x 10" 6 ^ 8 ppm X; F: q = 2 X
10~ 6 ^ 5 ppm X; G: q = 4 X 10~ 7 ^ 2 ppm X. Solid lines separate fields containing, on
the average, (/) more than five grains of the special mineral; (//) between one and five
grains; and (///) less than one grain.
the accessory mineral and the gangue contribute equally to the overall
content. Under the circumstances described this occurs at 2 ppm, when q ^
5.36 XI0" 7 . At concentrations above and below this the sampling error
decreases. The latter trend, however, would be of little practical significance
if errors from other sources were rising rapidly towards the detection limit of
the analytical method.
Increasing sample weight (or grinding) reduces R, but in the circumstances
described even 1-g sub-samples are inadequate (R > 50%) for materials
coarser than 100 mesh containing low, but possibly anomalous tin values, up
to about 20 ppm. Using emission spectroscopy, consuming a few tens of
milligrams of sample, a single analysis or even the average of several analyses
is unlikely to indicate the overall content of an element present as a major
constituent of a few rare mineral grains. One practical solution to this prob-
lem if the rare mineral has a distinctly higher density than the gangue., is to
pan, or otherwise separate, a heavy mineral fraction thereby greatly
38
TABLE 2-III
Effect of sub-sample weight and grain size on sampling error (R) for tin present as rare
grains of cassiterite in a gangue containing 1 ppm
(3) The average sub-sample contains more than 5 special grains. The Poisson
distribution becomes approximately Gaussian and replicate analyses provide
an estimate of the true mean. Relative standard deviation (R) depends on the
size of the sample and number of grains taken for analysis as described previ-
ously.
Three zones corresponding approximately to these conditions, with the
average number of grains (z) fewer than one, between one and five, and
greater than five, are indicated in Fig. 2-7. The probability (Pn) that a certain
number of grains (n) will be found in a particular sub-sample is estimated
from the Poisson distribution:
f
so that the probability (P0) of finding no special grains is e and probabil-
ities for 1, 2, ..., n grains are calculated from:
P =- i - ·P
Xt=H+ Cizi
when H is the content of the element of interest in the gangue, Q the con-
tribution of a single special grain to the overall concentration, and z{ the
number of special grains. Poisson distributions with z = 0.4, 1.0, 3.0 and
10.0 are shown in Fig. 2-8.
Returning to our example of rare grains of cassiterite in a soil or sediment,
one 100-mesh (150 μπι) grain weighs 23.6 μg with a Sn content (C) of
18.9 μg. If the average number of grains in a 1-g sub-sample is z = 1 the
overall content, with 1 ppm in the gangue, is approximately 20 ppm. How-
ever, with z = 1, P0 = 0.3679 and there is therefore a 37% chance of such a
sub-sample analyzing as only 1 ppm Sn (Fig. 2-8). Similarly, at an overall Sn
content of 8.5 ppm (q = 4 X 10" 6 , z = 0.4) the sampling distribution is
strongly skewed towards low values with approximately two thirds of the
samples reporting the background content (1 ppm): no single analysis will
correspond to the overall Sn content but one in twenty analyses will be as
high as 40 ppm. Skewed Poisson distributions are often reported for Au
(Jones and Beaven, 1971) and extreme examples have been reported for
diamonds (Phillips, 1971). The problems of sample size in estimation of Au
content are throughly discussed by Clifton et al. (1969).
Sampling errors can also create problems when data obtained by different
analytical methods, using different size sub-samples, are compared. Ingamells
40
z = 10.0
~c= 190 ppm
z = 3.0
c= 58 ppm
>-
o
z z = 1.0
LU
D 50 ~c= 19 ppm
σ
LU
CO L3^
z = 0.4
c^=8.5ppm Sn
' 7 ' 8 ' 9 Ί θ ' 1 1 '12'13' 1 4 ' 1 5 ' 1 6 ' 1 7 ' 1 8 ' 1 9 '
,
|o>|o|»|Hcol i|2?!S!|:r|o
|i-|co|tf)|^|o)|^|$2|i2|^|q> o N k © oo o N k <£
Sn ppm = H + z
Fig. 2-8. Histograms of the Poisson distribution with different values for the average
number of special grains (z) in a sub-sample. Corresponding concentration of Sn have
been calculated assuming that a 100-mesh grain of cassiterite contains 18.9 μg of Sn and
that there is 1 ppm Sn in the gangue.
SYSTEMATIC ERRORS
oi
3.0 4.0
GOLD ( g / t )
12
SET I R □ nge 3 . 0 5 - 4 . 0 6 g/t
Mean 3.4 3 g/t
§
II
St. dev. 0.24 g/t
10 C.V. 7.0
9
>-
LU
3 6
7Z
σ \
a: °
////////////
4
3 -
w,fe
2
/%;
o 1 1
3 0 5 5 4 0 4.5 5.
GOLD (g/t)
Fig. 2-9. Comparison of duplicate analysis for Au by two analysts using different instru-
ments. (From Allcott and Lakin, 1975.)
42
Contamination
Drift
I I I I I I l_l I I I I
BATCH NUMBER
Fig. 2-10. Control graph, based on the analysis of three laboratory standards, indicating
that systematic between batch variations are present for Cu.
44
\to Woodstock
4.8 km
Fig. 2-11. Spurious Cu anomalies introduced into a soil survey by analytical errors. The
east-west orientation of the anomalies parallels the soil traverse lines and has resulted
from the analysis of the samples in the same order as their collection. (Modified from
Lockhart, 1976.)
Original values
Redetermined values
70
60
10 15 20 25 30 35 40 45 50 55 60 65 70
Sample number
Fig. 2-12. A randomized numbering scheme for geochemical sampling and analysis. The
randomized field numbers have been plotted sequentially to reveal the presence of
systematic analytical errors caused by a leaking standard. (From Plant et al., 1975.)
45
Accuracy
TABLE 2-IV
Relative deviations between laboratory determinations of trace elements in lunar samples
(based on data in Morrison, 1971)
SAMPLE PREPARATION
INTRODUCTION
TABLE 3-1
Grinding equipment
Steel and iron grinding plates Fe, Co, Cr, Cu, Mo, Mn, Ni, V
Alumina ceramic plates l Al, Cu, Fe, Ga, Li, Ti, B, Ba, Co, Mn, Zn, Zr
Tungsten carbide Co, Ti, W
Lubricants Mo
Packaging materials
Polythene Ti, Ba, Zn, Cd
Polypropylene Ti
PVC Ti, Zn, Na, Cd
Brown paper Si
Rubber Zn
1
Thompson and Bankston (1970).
2
Scott and Ure (1972).
48
Fig. 3-1. Rock crushing equipment: in sequence from the foreground are (1) a disc mill
open for cleaning; (2) a jaw crusher; (3) a closed disc mill; and (4) a second jaw crusher.
The equipment is mounted on a hollow pedestal constructed by covering a heavy welded
steel frame with sheet metal: within the pedestal ducts, connecting each piece of equip-
ment to an air exhaust system, extract any escaping dust downwards. The compressor in
the background provides air for blowing equipment clean.
100°/
T3
e
o
J-l
be
Ö
3
Ü- 50
50 100%
P (ground)
Fig. 3-2. Comparison of analytical precision (P%) for trace element concentrations in
ground and unground minus 80-mesh stream sediments. Open symbols: no significant
difference, at the 95% confidence level, between the two populations on the basis of the
Kolmogorov-Smirnov test; solid symbols indicate that populations are significantly
different. (From Howarth and Lowenstein, 1971.)
51
Soils and sediments should be collected in high wet strength paper bags in
which they can be dried without opening the packet. In hot, non-humid
climates it is sufficient to place the bags outside in the sun. Where this is not
practical a simple drying tent can be improvised with four uprights, plastic
sheeting and a kerosene heater. Electric drying cabinets are used in the
laboratory. To avoid baking clay-rich samples, which makes subsequent dis-
aggregation difficult, temperatures should be no greater than 65° C. Koksoy
et al. (1967) reported losses of up to 42% of the Hg content of soils dried at
80° versus 20° C, and other volatiles, such as Se, could probably also be lost.
Contamination of samples which originally had low Hg contents was ob-
served within thirty days of their storage in the same container as an
anomalous sample.
Once dry the sample can be disaggregated and sieved. The object of dis-
aggregation is not to grind the material but to break down aggregates and
clumps to obtain the natural grain size distribution. This is best achieved by
a light pounding with a porcelain pestle and mortar. The sample is then
screened on a silk or nylon sieve mounted in a cardboad or plastic frame —
4|" (11.4 cm) diameter PVC drainpipe is ideal. Stainless steel sieves can also
be used but are expensive and must be treated with caution because silver-
solder is sometimes used in their construction: only sieves with a friction-
flange fitting should be used. Between samples, equipment can be cleaned
with a stiff paint brush or compressed air. Normally only the fines are
retained for analysis. Hoffman (1974), however, has suggested collection of a
coarse fraction (>2 mm) on bitumen coated pebble cards as a guide to sedi-
ment lithologies and provenance.
Mesh sieve sizes are summarized in Table 3-II. In some systems of size
classification the silt-sand boundary is arbitrarily placed at 20 μηι whereas in
others it is at 50 μιη. The latter, at 270 mesh, is about the lower limit of
rapid dry screening and separation of finer fractions requires methods based
on their sedimentation rates. Because dry disaggregation and sieving are not
very efficient, estimates of size fraction distributions so obtained will seldom
agree closely with results from more reliable wet dispersion methods and
wet-sieving or sedimentation.
Detailed schemes for separation of mineral fractions from overburden or
sediment are beyond the scope of this text. However, general remarks on this
topic in the preceeding section apply equally here. The value of panning bulk
samples to obtain reasonably representative sub-samples of rare grains of
heavy minerals, such as cassiterite, was noted in Chapter 2. The scheme used
by the Geological Survey of Sweden for processing tills is summarized in
Fig. 3-3.
Separation of mineral fractions requires considerable effort and the little
additional time required for a rapid visual examination under a binocular
microscope can be very rewarding. In Zambia, the author was able to trace
the sub-outcrop of weathered zinc mineralization by washing the plus 2-mm
52
TABLE 3-II
Approximate equivalent mesh sizes in the British Standard (B.S.), American Society for
Testing Material (ASTM) and Tyler sieve series
2000 8 10 9
1000 16 18 16
850 18 20 20
420 36 40 35
355 44 45 42
300 52 50 48
250 60 60 60
212 72 70 65
180 85 80 80
150 100 100 100
125 120 120 115
90 170 170 170
75 200 200 200
63 240 230 250
53 300 270 270
45 350 325 325
38 — 400 400
Vegetation
Vegetation samples are best collected in brown paper grocery bags. They
should not be packed tightly or placed in closed plastic bags because this will
cause rotting and ultimately produce silage! If the material is to be ashed it
should be remembered that ash content usually amounts to only 3—10% of
the dry weight and an appropriate amount of material collected.
Samples can be dried by hanging the bags in a warm dry room or by using
an oven with forced ventilation. Temperatures above 40° C should not be
used. Dried material can then be ground in a hammer mill equipped with
steel blades and a non-contaminating screen through which the fine, ground
material falls.
Waters
Natural waters may have to be analyzed for trace metals and pH, and less
often for anions (particularly Cl~, SO*", HCOJ, CO3" and F"), temperature,
53
Sampling
Glacial till
| 10 liters
Suction dredge _
and sluice box —H Overflow Waste
Retained on
sluice box
4
Drying
i
Sieving 0 5 mm
Separation with
tetrabromoethane (296 g/cm 3 ) HMatenal<2.96 Waste
I Material >2.96|
Separation with
hand magnet
i
Non-magnetic Weakly magnetic Weakly magnetic
i
Non-magnetic
Magnetite material material material magnetic
material >0.5mm < 5 m n
>3.31 >331 >2.96<3.31 >2.96<3.31
Sub- Kept for further
fraction 1 investigations
Y
Chemical and mineralogical analysis
Fig. 3-3. Flow chart of the stages in heavy mineral separation and processing used by the
Geological Survey of Sweden. (From Brundin and Bergström, 1977.)
The average weights of the subfractions according to this flow sheet are:
conductivity and dissolved oxygen. pH and the anions are usually best deter-
mined as soon as possible at the sample site or shortly after collection. How-
ever, unless colorimetric field procedures are used, determination of trace
metals will usually require storage and shipment of waters to a laboratory.
During storage the very low concentrations of trace metals found in natural
waters are susceptible to changes resulting from bacterial activity, from con-
tamination by their container, and from losses due to adsorption of dis-
solved metals on suspended sediments, on precipitates and on the container
walls.
For field use polyethylene or polypropylene bottles are preferred to glass.
However, several important trace metals are incorporated in plastics during
their manufacture (Table 3-1) and are often present in a leachable form
(Scott and Ure, 1972). Zinc is especially troublesome. To prevent con-
tamination a thorough cleaning of bottles (and caps), first by protracted
soaking in 50% hydrochloric or nitric acids and then by rinsing with distilled
water and aliquots of the water to be collected, is recommended. Marchant
and Klopper (1978) reported that the levels of contamination found after
storing ultrapure water for four years in acid-washed polyethylene, were too
low to be of concern in exploration geochemistry.
Losses by adsorption on container walls have been studied by many
workers and a variety of special procedures have been suggested for individual
elements. Acidification to pH 1 with hydrochloric acid and addition of
50 mg/1 Br is effective for Au (Chao et al., 1968) and preservation by addi-
tion of an acidified solution of potassium permanganate (Jonasson et al.,
Iron
00
/
Acidified
00 Γ
+ / +
+
10 +^ +
+
- +
+ + +-H-
-y
1n ■ i » ■ ....I L _ J _ i, iml , i , , MM
1.0 10 100 1000
Unfiltered Fe m g / l
INTRODUCTION
During decomposition the sample is opened up or decomposed releasing
the elements to be determined. This can be brought about by digestion and
leaching with concentrated or dilute acids; fusion with acid or alkaline
fluxes; liberation of volatile constituents by pyrolysis; and by partial or
selective extraction of constituents with a wide variety of reagents. Strictly,
the almost simultaneous decomposition and excitation of the sample in DC-
ES should also be considered as sample decomposition. Here, however, only
those procedures producing solutions, suitable for analysis by AAS, ICP-ES
or colorimetry, will be considered.
Although decomposition introduces an additional step between sample
preparation and the final analysis, it provides the geochemist with consider-
able freedom to liberate and determine either all or only a particular fraction
of the trace constituents. As discussed in Chapter 1, selection of the appropri-
ate decomposition technique can be a major factor in enhancing anomaly
contrast. Choice of a method will depend largely on the dispersion character-
istics of the element to be determined and on its mode of occurrence within
the sample. Other factors to be considered include: (1) the final method of
analysis and the need to avoid interferences; (2) the desirability of deter-
mining several elements in the same solution; (3) ease of handling, rapidity
and adaptability to processing large numbers of samples; and (4) availability
and costs of reagents and equipment.
The decomposition procedures to be described are summarized in Table
4-1. They fall into two broad groups: strong decompositions capable of
releasing a large proportion of the trace constituents from mineral lattices,
and weak or partial decompositions intended to remove only weakly bonded
elements or those associated with a particular fraction of the sample. Strong
and weak decompositions are often referred to casually as total and cold
extractions, respectively. Neither term can be recommended because the
former are seldom true total attacks (i.e. attacks capable of releasing all of
the metal inside and outside the lattice in whatever form (Lapointe, 1968),
and the efficiency of the latter is very susceptible to even small changes of
ambient temperatures (Ellis et al., 1967).
58
TABLES 4-1
Classification of some decomposition techniques useful in exploration geochemistry
Decomposition Reagents
Strong decompositions
(1) Digestions with hot, usually concen- H N 0 3 , HCl, HCIO4, HF
trated, mineral acids
(2) Fusions:
acid fusions KHSO4, K 2 S 2 0 7
ammonium halide sublimations NH4I, NH4CI
alkaline fusions Na 2 C0 3 , NaOH, L1BO3
oxidative-alkali fusions Na 2 C0 3 or NaOH with ΚΝΟ3 or N a 2 0 2
Partial decompositions
(1) Non-selective decompositions:
cold dilute mineral acids: e.g. 0.1-1.0 N HCl
buffers: e.g. NH 4 -citrate/NH 2 OH · HCl, pH 2-
chelating agents: e.g. 0.05-0.25 M EDTA, pH 4 - 7
(1) Selective decompositions:
(a) removal of exchangeable metals NH 4 -acetate; MgCl2
(b) removal of organic matter H 2 0 2 ;NaOCl
(c) secondary iron and manganese Na-dithionite;
oxides hydrazine;
hydroxylamine hydrochloride;
NH4-oxalate
(d) sulphides KCIO3/HCI; ascorbic acid/H 2 0 2 ;
bromine
STRONG DECOMPOSITIONS
Acid digestion
TABLE 4-II
Physical and chemical properties of mineral acids used in sample decomposition
Property Acid
Fig. 4-1. Sample decomposition equipment. The hot air bath (background), drilled alumi-
nium block (left) and sand bath (right) are all convenient ways of decomposing large
numbers of samples on a single hot plate.
hydrofluoric acid. Lower recoveries were obtained from some pyroxenes and
amphiboles and from most of the rocks studied. Grain size was also found to
influence extraction efficiency.
A similar study of the release of Cr, Co, Cu, Fe, Mn, Ni, V, Zn and Pb
from the products of lateritic weathering, after decomposition with hydro-
fluoric acid, nitric and perchloric acid, and perchloric acid alone, was
reported by Gedeon et al. (1977). Sample mineralogy was examined by
X-ray diffraction, both before and after decomposition, and extraction effi-
ciency compared to total concentrations estimated by XRF. Hydrofluoric
acid left little or no residue and was satisfactory for all elements except Cr
and V. In contrast, residues from the nitric-perchloric decomposition con-
tained quartz, talc, muscovite, feldspars, amphiboles and kaolinite;
depending on the type of material from 50 to 100% of the trace metals were
extracted. Perchloric acid used alone was least satisfactory with regard to
both extraction efficiency and reproducibility.
Clearly, even with strong acid decompositions extraction of lattice-bound
trace elements, from both fresh silicates and the residues of intensive
weathering, is often far from complete. The resulting geochemical patterns
will therefore reflect, in part, mineralogical and lithological differences
among samples rather than absolute differences in their trace element
61
120 -
Vu
I-
X
LU
40-
20- ^^^>Τ*
* Ε
α
η _
α
CO
^ ι^ ° <"> o o
ζ ζ z
ο
X ° ^ I z
g
o o
* Ö
Fig. 4-2. Variation in the extraction of Zn from some common rock-forming minerals
with acid decompositions. (From Foster, 1971.)
100-,
E
aa
co i- co Ü
O
0
<N
CO
CO
O
O
C ix P z
X
O z z o
z 1 X •5 <
o
o X
o X
Fig. 4-3. Variation in the extraction of Zn from dacite, diopside gneiss, diabase and
peridotite with acid decompositions. (From Foster, 1973.)
62
Nitric acid
Nitric acid, dilute or concentrated, is a strong oxidizing agent and an
excellent solvent for most sulphides, except cinnabar, which are oxidized to
sulphates. Lead can be subsequently lost by precipitation of insoluble
lead sulphate. Phosphates and most primary and secondary uranium minerals
are also readily decomposed. Silicates are only partly attacked and organic
matter is not completely destroyed. Organic-rich samples should therefore be
ignited prior to treatment with nitric acid.
Despite incomplete breakdown of organic matter, boiling concentrated
nitric acid in a 5 : 1 (v/w) ratio has been recommended as a general decom-
position procedure for rocks and soils (Ward et al., 1969). The resulting solu-
tion, after dilution, is suitable for the determination of Bi, Cd, Pb, Cu, Zn,
Co and Ni by AAS. Reasons given for the choice of nitric acid, in preference
to mixtures of nitric-perchloric or nitric-hydrochloric acids, were greater
safety and reduced nebulizer corrosion, respectively. A hot dilute nitric acid
leach (4 M at 90° C) has been used to extract Cu and Zn from lake sediments
(Timperley and Allen, 1974); the solution is also suitable for the fluorimetric
determination of U (Smith and Lynch, 1969).
3 HC1 + H N 0 3 -* 2 H 2 0 + NOC1 + Cl 2
Hydrofluoric acid
Hydrofluoric acid is unique in its ability to decompose silicates, Si being
lost as the volatile tetrafluoride. All but the most refractory silicates are dis-
solved and hydrofluoric acid, either alone or with other mineral acids, has
therefore found wide use in many analytical schemes for determination of
major constituents of rocks (Maxwell, 1968; Langmhyr and Paus, 1968).
However, in the exploration laboratory, where rigorous estimation of
absolute metal concentrations is seldom needed, the precautions necessary
in handling this acid make its general use inconvenient except when total
concentrations are required or it is necessary to achieve complete decomposi-
tion of a particular mineral. At temperature below 120°C inexpensive
plastic beakers or bottles, heated on an air- or water-bath, make satisfactory
decomposition vessels. Polytetrafluorethylene (teflon) can be used up to
250° C but at higher temperatures platinum ware is needed.
Hydrofluoric acid is sometimes used both to decompose the sample and
complex the analyte. For example, B can be determined colorimetrically as
its fluorborate complex (BF4) after the sample has been decomposed with
hydrofluoric and sulphuric acids (Stanton and McDonald, 1966). Similarly,
in the estimation of Nb, pyrochlore and other Nb-bearing minerals are
decomposed with hydrofluoric acid; the soluble niobium fluoride complex is
then isolated and estimated by paper chromatography (Hunt et al., 1955).
900
E
aa
z
o
o
LJJ
Q * Smelting with NaOH
CO
o ■ Sintering with K N 0 3 - N a 2 C 0 3
o Complexing with Be(N0 3 ) 2
5000
+" ~+ + + + φΜ + + + + + + + + +~^r-v
+ V + % Vw £ ^ + V ^ Granite + V + V V l
+ + + + ( J ^ + + + + + + + + + + + -(■
Fig. 4-4. Fluoride values in soils estimated by six different extractions. (From Pliiger and
Friedrich, 1973.)
66
Fusions
TABLE 4-III
Applications of bisulphate and pyrosulphate fusions
Element 1
Flux: sample Leach acid Reference
As 4 :1 HC1 1
Bi 3 :1 HN03 1, 2
Bi(col./XRF) 3 :1 HNO3 3
Co 4—5 : 1 HC1 1, 2
Cu 5 :1 HC1 1, 2
Mn 5 :1 H 2 S 0 4 or ΗΝΟ3 1, 2
Mo 4 :1 HC1 4
Nb 20 : 1 tartaric acid 2
Ni 4-5 : 1 HC1 1, 2
Pb 5 :1 HC1 2
Sb 5-7.5 : 1 HC1 or tartaric acid 2, 5, 6
Sb(AAS) 4 :1 HC1 7
V 3-5 : 1 HNO3 1, 2, 8 3 , 9 3
NH4I - NH 3 + HI
which decomposes the sample and converts Sb, Sn and Hg to volatile halides
which sublimate, together with excess ammonium iodide at the cooler, upper
end of the fusion tube. The sublimate is then dissolved with dilute hydro-
chloric acid. In the determination of Sn, cassiterite is readily decomposed;
however, Sn occluded within silicate lattices is n o t released unless silicates are
destroyed by pre-treatment with hydrofluoric acid (Agterdenbos and
Vlogtman, 1972; Welsh and Chao, 1976). Ammonium chloride can be used
in place of ammonium iodide in the determination of Sb, volatile antimony
68
Fig. 4-5. Sample decomposition by sublimation with ammonium iodide using a detacha-
ble condenser (A); and (B) dissolution of the sublimate. (From Pahlavanpour et al.,
1979.)
69
Alkaline fusions
Alkaline fusions, usually with sodium carbonate or lithium metaborate,
appear in many schemes for whole rock analysis (Abbey, 1970; Ingamells,
1970; Maxwell, 1968). In the exploration laboratory their use has generally
been limited to determination of: (1) amphoteric elements of Groups V and
VI of the Periodic Table — these form soluble anions in alkaline media, e.g.
chromate, molybdate and tungstate; (2) uranium; and (3) chloride and
fluoride.
North (1956) first applied alkaline fusion to the colorimetric determina-
tion of W and Mo in geochemical prospecting. The sample is fused, in a
nickel crucible or disposable culture tube, with a flux consisting of sodium
carbonate/potassium nitrate/sodium chloride in a 5 : 1 : 4 ratio (Stanton,
1966; Ward et al., 1963). Potassium nitrate is added as an oxidant and
sodium chloride to decrease the viscosity of the melt. The solidified melt is
leached with water to dissolve molybdate and tungstate together with
aluminate and sodium silicate. Interfering elements, notably Fe, remain in
the residue as their insoluble carbonates. Low recoveries obtained by this
method, compared to pyrosulphate fusion, have been attributed to the diffi-
culty of fully leaching the fused mass with a small volume of water (Stanton
and Hardwick, 1967; Stanton, 1966; Quin and Brooks, 1972).
Chromium is determined colorimetrically, as chromate or with diphenyl-
carbazide, after fusion with a sodium hydroxide/sodium peroxide flux
(Wood and Stanton, 1956—57). Peroxide is used as an oxidant, rather than
potassium nitrate, to avoid interference from nitrate. Manganate, which
would also dissolve in the water leachate and interfere, is reduced to insoluble
manganese dioxide by addition of ethyl alcohol. Most other elements remain
in the residue as insoluble hydroxides.
For the determination of F and Cl samples can be sintered with a 2 : 1
mixture of sodium carbonate and potassium nitrate (Ficklin, 1970; Haynes
and Clark, 1972; Crenshaw and Ward, 1975), or fused with either sodium
carbonate/zinc oxide (Ingram, 1970) or sodium and potassium carbonate
(Hopkins, 1977). The sinter or fused mass is then leached with water and the
leachate analyzed with a specific ion electrode. Sodium hydroxide fluxes
(Kesler et al., 1973; Josephson et al., 1977; Hopkins, 1977) and lithium
metaborate have also been used (Bodkin, 1977). Results obtained with
several of these methods are compared in Fig. 4-4 and examples of their
applications to prospecting are to be found in Plüger and Friedrich (1973),
Farrell (1974), Lalonde (1974, 1976) and Phuphatana et al. (1976). Ficklin
(1975) used a sodium carbonate/potassium carbonate/magnesium oxide flux
in the determination of I.
Fluorimetric determination of U takes advantage of the intense fluores-
cence of uranyl compounds in the presence of fluoride. A powdered sample
(Grimaldi et al., 1954), or a suitable aliquot of a solution containing U
(Smith and Lynch, 1969), is fused in a small platinum crucible or dish with a
70
S £
o «> 620° C •
6 w 30 - ! 650° C
Q CM W~
z o ~** · ·
< d
*
z
c
o
JIT
<
-I W
CQ £
o
Z .2 20 -
LU > ^^e^700°C
g 5
m 2
Q
z<
< CO
Ui 3
£ u. i o - \7500C
</> O
>- 5
H <
α
>
t O v 800° C
(/> DC
z So
C/) 0 "I I I I I I I I
5 10 15 20 25 30 35 40
TOTAL TIME IN FURNACE, minutes
Fig. 4-6. Influence of fusion temperature, with a sodium fluoride/sodium carbonate/
potassium carbonate (9 : 45.5 : 45.5) flux, on sensitivity for the fluorimetric determina-
tion of U. The vertical axis shows the difference in readings between the blank and
0.005 Mg of U. (From Fletcher, 1954.)
71
PARTIAL EXTRACTIONS
Dissolution of sulphides
Several procedures have been suggested for selective liberation of metals
associated with sulphides: these include aqua regia (Warren and Delavault,
1959a, b; Brabec, 1971); a mixture of ascorbic acid and hydrogen peroxide
(Lynch, 1971); potassium chlorate and hydrochloric acid (Olade and
Fletcher, 1974); and various solutions containing bromine (Hausen et al.,
1973; Davis, 1972; Czamanske and Ingamells, 1970; Peachey et al., 1978).
When ascorbic acid/hydrogen peroxide is used sulphides are preferentially
dissolved by their oxidation with hydrogen peroxide, the ascorbic acid pro-
viding a reducing, acidic medium to prevent precipitation of hydrous ferric
oxides. Galena, arsenopyrite, chalcopyrite, pyrite, pyrrhotite, pentlandite,
tetrahedrite, sphalerite and awaruite are all dissolved in the method
described by Lynch (1971). With potassium chlorate/hydrochloric acid the
active oxidizing agent is the nascent chlorine generated: most sulphides are
rapidly dissolved (Dolezal et al., 1968).
Olade and Fletcher (1974) have compared the selectivity of aqua regia,
ascorbic acid/hydrogen peroxide and potassium chlorate/hydrochloric acid
for copper sulphides associated with porphyry copper deposits. Concentra-
tions of copper (Cu x ) and zinc (Zn x ) liberated with these reagents were com-
pared to total contents (Cu t and Z n t ) determined after decomposition with
hydrofluoric acid/nitric acid/perchloric acid. Two criteria based on geo-
chemical behaviour of Cu and Zn were used to evaluate the results: Cu, a
strongly chalcophile element, is probably largely present as sulphide inclu-
sions in most igneous rocks, whereas the more oxyphile Zn occurs predomi-
nantly within silicate lattices. Consequently:
Criterion 1: an efficient copper sulphide selective leach will give a high
Cu x /Zn x ratio in samples containing background Zn values.
Criterion 2: Cu x /Cu t will increase with Cu t as copper sulphide content
increases, until in strongly mineralized samples Cu x equals Cu t within the
limits of analytical error.
Results (Fig. 4-7) show that all three reagents partly meet these criteria.
However, the Cu x /Zn x ratio is greatest with the potassium chlorate/hydro-
chloric acid leach and lowest with aqua regia. Similarly, in samples with low
values of Cu t , Cu x /Cu t is lowest with the potassium chlorate/hydrochloric
acid. On this basis it was concluded that this procedure was least damaging
to silicates and most selective for sulphides. In a later study, Chao and
Sanzolone (1977) investigated the efficiency of potassium chlorate/hydro-
chloric acid for dissolution of individual sulphide minerals. Following the
procedure of Olade and Fletcher (1974), dissolution of galena, cinnabar,
orpiment, stibnite, sphalerite and tetrahedrite was essentially complete,
73
C Aqua regia
o ·
° °
O o O
o ° °° o ° ,
οθ o o °,
2
I-
X ° *<* o a o 0 8 °° °
o o o
0 o o o
B H ?O o - A s c .
l0
K
O
°l • • • · o g
• 1
·· • *· ·
• ••
• • • o 8 °
• ··· • • 6
ÜC 60|
•
UJ
0. •
• o o
• · o
o
o
o
o
o
o •
o
o cS> o o
°8 o o
i
1000 10,000 10 10,000
Water-soluble constituents
Determination of water-soluble constituents in bedrock has included esti-
mation of chloride released from fluid inclusions in freshly crushed plutonic
TABLE 4-IV
Extraction of copper, nickel and cobalt from some silicate minerals with an ascorbic acid/
hydrogen peroxide leach (data from Cameron et al., 1971)
Cu Ni Co
Olivine 67 4 4
Olivine 56 3 4
Clinopyroxene 43 5 3
Orthopyroxene 91 2 0
1
100 X (ascorbic acid/H 2 0 2 )/HF.
75
rocks (Van Loon et al., 1973), and leachable Ca, Mg, Na, K, F and Cl in
mafic and felsic volcanic rocks at the Brunswick No. 12 and Heath Steele
massive sulphide deposits (Goodfellow and Wahl, 1976). Although the
extraction procedure used by Goodfellow and Wahl was extremely simple
(1.0 g of rock powder was stirred with 10 ml of water for one minute) it
successfully outlined anomalous halos extending up to 800 m around the
deposits.
Non-silicate uranium
Under oxidizing, alkaline conditions in carbonate-rich waters U 4 + in
uraninite and pitchblende is oxidized to U 6 + and dissolves as a carbonate
complex — U0 2 (C0 3 )3~ (Langmuir, 1978). Solutions of sodium or ammoni-
um carbonates and hydrogen peroxide are therefore employed as lixiviants in
solution mining of uranium ores. The same principle can be applied to
extraction of U in the laboratory. For example, Ward and Bondar (1979)
were able to distinguish U anomalies attributable t o ore minerals, from
equally anomalous concentrations of lattice held U, by extraction with a 2%
sodium carbonate/5% hydrogen peroxide leach. In addition to its direct
applications to geochemical prospecting, this approach has potential value in
the identification of source rocks, with relatively high content of leachable
(labile) U, favourable to the development of sandstone-type uranium depos-
its.
Fig. 4-8. Influence of pH on the adsorption of cations (Ag+) and anions (ΜοΟ^ ) on
hydrous oxides, clays and related materials. Based on results from Dyck (1971) and Jones
(1957).
77
200 -i
Freely drained soil, UL1(A), extracted for 15 min. at 25°C
0 1 2 3 4 5
PH
Fig. 4-9. Influence of pH and solvent on extraction of Cu from a freely drained soil.
(From Ellis et al., 1967.)
78
^ & tit
Ξ — Ξ Zn
CO 50 H • · Ni
< A A Co
UJ
-1
<·> <·> Pb
UJ
{£
♦ ♦ Cu
5 4 3 2 1 0
pH
Fig. 4-10. Release of metals at different pH values from humic materials. (From
Chowdury and Bose, 1971.)
constant as possible. Ellis et al. (1967) found, for example, that differences
of 10°C in room temperature could change the amount of Cu released with a
citrate/hydroxylamine buffer by as much as 50%. Response surfaces (Fig. 4-
11) from Sorensen et al. (1971) illustrate the influence of extraction time
and sample to solution ratio on extraction of Zn with 0.1 N hydrochloric
acid. The long extraction period before equilibrium is approached is typical
of many partial extractions, other than those involving ion exchange which
proceeds relatively rapidly. Reaction kinetics, i.e. changes in the rate of
release of elements as extraction proceeds (Fig. 4-12), were used by Ellis et
al. (1967) and Warnant et al. (1980) to characterize the distribution of Cu in
soils and sediments.
E
a
a
<
en
h-
X
UJ
υ
z
N
300 100
SHAKING PERIOD, min
Fig. 4-11. Response surface showing the effect of different shaking periods and solution
to soil ratios on the amount of Zn extracted with 0.1 N hydrochloric acid from nine
alkaline soils. (From Sorensen et al., 1971.)
80
10000
E
α
α
1000
Ο
(Λ
5
Ο
100
<
CC
I-
X 100000-J B. Background freely drained soil, UL2(A)
LU
UJ
CO
UJ
ζ 10000 -d
<
ο
ζ
<
2
CL
α.
Ο
ο
TIME, minutes
Fig. 4-12. Comparison of the rate of extraction, at 100°C with 0.2 M hydrochloric acid,
of Cu, Fe and Mn from (A) anomalous and (B) background, freely drained A horizon soil
samples, Kilembe, Uganda. Similarities between Cu and Fe curves suggest an association
between Cu and Fe in both sets of samples. (From Ellis et al., 1967.)
TABLE 4-V
Influence of hydroxylamine hydrochloride on the extraction of heavy metals from
stream sediments in relation to their manganese content l . Heavy metals determined by
the dithizone field test (from Canney and Nowlan, 1964)
gesting that under the acidic conditions used in determination of Cu, there
may also be some dissolution of clay minerals and iron oxides.
Cold dilute hydrochloric acid provides an excellent medium for AAS and
is therefore often used as a weak extractant. To avoid erratic results caused
by soil minerals neutralizing the acid, solutions stronger than 0.5 M should
be used (Ellis et al., 1967). Exchangeable metals will be released together
with some of the metals associated with clay minerals, manganese oxides and
organic matter. Sulphides and carbonates are also partly decomposed and the
almost complete release of Pb from Pb-rich soils near sulphide mineraliza-
tion has been attributed to the solubility of galena or the secondary mineral
plumbojarosite (Cameron, 1977; Miller, 1979). Dissolution of iron oxides is
probably limited by low solubility of Fe 3+ .
In geochemical prospecting for U the use of relatively strong decomposi-
tions appears to have been favoured. However, as described on p. 75 for
bedrock geochemistry, there appears to be some merit in using partial
extractants to selectively leach uranium minerals. The observations of Rose
and Keith (1976) corroborate this. Investigating geochemical dispersion in
stream sediments associated with uranium deposits in sandstones of north-
eastern Pennsylvania, they found that an acetic acid/hydrogen peroxide
leach was considerably better than total U (by neutron activation), and
slightly better than extraction with nitric acid, in distinguishing anomalous
sediments (Fig. 4-13). Similar conclusions were reached by Olade and
Goodfellow (1979), using a sodium carbonate/hydrogen peroxide leach to
study distribution of U in sediments of streams draining the Toombstone
Batholith. One advantage of the alkali-oxidizing leach is that the elements,
especially Mn and Fe, that quench the fluorescence of U in its fluorimetric
estimation are insoluble in the leachate.
k • Near prospects
o Background
[-
Ε Γ 3/
07
- / γ • ° /
- #/
O CO <k>/
I
03 o o^4
01 ι/ι i i i i IX 1 1 1 1 1 1 1 1
03 10
Total U (ppm)
Fig. 4-13. Total U and acetic acid/hydrogen peroxide-extractable U in stream sediments,
northeastern Pennsylvania. Sediments related to uranium prospects contain an increased
proportion of readily extractable U. (From Rose and Keith, 1976.)
TABLE 4-VI
Extraction of fluoride from soils with TISAB 0.01 M HCl and 0.037 M FeCl 3 (from
Schwartz and Friedrich, 1973)
TE e x + M a q = TE a q + M e x
Organic matter. Most soils contain from less than 1 to 20% organic matter
(equivalent to up to about 10% organic carbon). Associated trace metals can
85
TABLE 4-VII
Comparison of the action of sodium hypochlorite on release of copper from granodiorite
and peat samples (from Hoffman and Fletcher, 1980)
Cu extracted Cu extracted
(ppm) (%) (ppm) (%)
(1973) have shown, for example, that alkaline solutions containing humic
acids can retain approximately five times the amount of Cu in solution than
the same solutions without humic acids.
Extracts of organic-rich samples and natural waters draining peat lands are
often strongly coloured by organic colloids and soluble organic compounds.
These organic compounds, which are capable of sequestering heavy metals,
must be broken down prior to colorimetric analysis. Oxidation with
hydrogen peroxide or bromine (Mitchell and Smith, 1974) can be used.
Alternatively an aliquot can be evaporated to dryness with nitric acid/
perchloric acid. In the case of natural waters, the difference between total
dissolved metal and the fraction capable of reacting directly with a colori-
metric reagent such as dithizone, provides a rough estimate of the extent of
organometallic complexing.
Providing it is not necessary to remove organic matter as part of a sequen-
tial scheme, chelating agents can be used to extract associated heavy metals.
Ethylenediaminetetracetic acid (EDTA), a strong chelating agent for heavy
metals, is commonly used as 0.05- to 0.25-M solutions of its disodium salt
for this purpose. The potential value of this extraction is illustrated in a com-
parison of the action of 0.1 M EDTA and hot 4 M nitric acid on gytta
(organic-rich) lake sediments (Timperley and Allan, 1974): EDTA was found
to give the more consistent background results for Zn (Fig. 4-14). This was
attributed to the selective action of EDTA on the organic matter whereas the
hot acid also partly decomposed silicates and ferromanganese nodules which
100]
80
601
·%·.
10 20 30 £0 50 60 70 80 20 40 60 80 100 120 U0 160
Fig. 4-14. Analysis of lake sediments. A. EDTA-soluble Cu versus nitric acid-soluble Cu.
B. EDTA-soluble Zn versus nitric acid-soluble Zn. ♦ = samples containing ferromanganese
nodules given abnormally high Zn concentrations with the acid decomposition. (From
Timperley and Allan, 1974.)
87
TABLE 4-VIII
Extraction of copper from anomalous and background peat samples with nitric acid/perchloric acid, 0.5 M hydrochloric acid,
0.05 M EDTA and 1 M ammonium acetate (from Maynard and Fletcher, 1973)
A B A B A B A B
Analytical precision; 2σ level (±%) 3.2 6.8 18.6 5.7 7.7 22.0 n.d.
Cu extracted (ppm) GM 116 19 84 11 73 15 3 n.d
R 26-565 9-32 24-444 2-23 19-307 5-26 2-15 n.d,
Cu extracted (% of total) Av — — 73 67 68 82
n d
R 49-92 13-100 18-100 30-100 <l-8
Threshold (ppm) (GM B + 2σ) 36 38 36 n.d.
Average contrast (GM A /threshold) 3.2 2.2 2.1 n.d.
A = anomalous samples; B = background samples; GM = geometric mean; R = range; Av = average; n.d. = no data.
89
h 400
TOTAL Mn
z
O 15000
CC
LU UJ
O. Q.
10000 200
CC CC
_ EASILY
REDUCIBLE
5000 hlOO
EXCHANGEABLE
Mn
Fig. 4-15. Total, exchangeable and easily reducible Mn in drainage sediments from poorly
drained and freshly drained environments. Note the difference in scale. (From Horsnail et
al., 1969.)
described by Tamm (1922). The reaction is light sensitive and, unless carried
out in strong sunlight or ultra-violet light, dissolves only amorphous and
poorly crystallized oxides (Deb, 1950; McKeague and Day, 1966). According
to LeRiche and Weir (1963), gibbsite and boehmite are only very slightly
decomposed whereas appreciable quantities of organic matter are co-ex-
tracted with the amorphous oxides. This is consistent with the observation
of Hoffman and Fletcher (1980) that ammonium oxalate extracts are
strongly coloured by organic compounds and contain greater concentrations
of the trace elements if organic matter is not first destroyed by a preliminary
treatment with hypochlorite. In carbonate terrains extraction with either
ammonium oxalate or oxalic acid can precipitate calcium oxalate causing
low recoveries for Pb (Alminas and Mosier, 1975).
Reduction with sodium dithionite (hydrosulphite) is more vigorous than
with ammonium oxalate and both amorphous and crystalline iron oxides
dissolve. Many procedures have been described: in that of Mehra and
Jackson (1960) sodium bicarbonate is used to buffer the reaction at the
optimum pH of 7.3 (Fig. 4-16) and sodium citrate (0.3 N) is added to com-
90
-100
-80
O
UJ
-40 ü-o
-20
0
5 6 7 8 9 10 11 12 13
PH
Sequential extractions. After removal of organic matter and oxide phases the
bleached residue, consisting principally of primary and secondary silicate
minerals, can be mineralogically examined and the size fractions separated.
Strong decompositions can then be used to determine their trace element
content. Four sequential extractions schemes, in which organic matter and
hydrous oxide associated trace elements are estimated prior to analysis of
91
500
200
120 180 240 300 360 420 480 120 180 240 300 360 420 480
500
200
k<^^__*
120 180 240 300 360 420 480 60 120 180 240 300 360 420 480
MOLYBDENUM ZINC
z
O
υ
LEGEND
1 ORGANIC FRACTIONS
1A DILUTE HCI
2 AMORPHOUS OXIDES
3 AMORPHOUS Fe OXIDES
4 CRYSTALLINE Fe OXIDES
5 — SILICATE MINERALS
6 TOTAL CONTENT
MANGANESE
Fig. 4-17. Sequential extraction of Mn, Mo and Zn from lake sediments, on a transect
across Capoose Lake, British Columbia, using the extraction scheme of Hoffman and
Fletcher (1979). Details of the extraction appear in Table 4-IX. (Modified from Hoffman
and Fletcher, 1981.)
CO
to
TABLE 4-IX
Four sequential extraction procedures (intermediate extraction steps, to redissolve any precipitates formed, are not included)
Hoffman and Fletcher Gatehouse et al. Rose and Suhr Chao and Theobald
(1979) (1977) (1971) (1976)
the silicate residue, are summarized in Table 4-IX with some typical results
in Figs. 4-17 and 4-18. The sequence of extraction is, of course, critical —
the least aggresive reagents being applied first and organic matter removed
before dissolution of the oxides. This prevents partial release of humic sub-
stances solubilizing metals and influencing the results of subsequent stages in
the procedure (Hoffman and Fletcher, 1980).
The application of selective and sequential extractions to the special prob-
lems of geochemical prospecting in former mining districts has been dis-
cussed by Dijkstraet al. (1979). In these circumstances, with unusual mineral
phases present, results of sequential extractions can be misleading and the
distribution of ore or smelter contaminants in soils and sediments can some-
times be more readily identified as the difference between two extractions,
only one of which is capable of dissolving the contaminant phases.
depth Fe Mn Zn Pb
Fig. 4-18. Distribution of the total amounts of Fe, Mn, Zn, Pb and Cu available in each
10-cm interval of South Comet soils, western Tasmania, using the extraction scheme of
Gatehouse et al. (1977). See Table 4-IX for details of the extraction. (From Gatehouse et
al., 1977.)
94
SUMMARY
enhance anomaly contrast. It should be noted that the more specific the
action of the partial extractant, the more demanding becomes the task of
sampling similar materials at each site. Consequently, the non-selective
partial extractions are of greatest use in routine geochemical prospecting.
They are also well suited, in conjunction with colorimetry, to field tests.
From an anlytical viewpoint the decomposition procedure selected should
lend itself to rapid sample throughput, be compatible with the determina-
tion of a large number of elements on the same solution, and minimize inter-
ferences.
If the elements of interest are present in resistate minerals, decomposi-
tion can often be avoided and the solid sample analyzed directly by DC-ES
or XRF.
Chapter 5
INTRODUCTION
Transmittance
B. lo-»t
Ύ
i m ^
solution-ce
lo-»t
Scattering
Ύ
solution-cell
Ύ
E? κ :
COLORIMETRY
When white light passes through a coloured solution some wavelengths are
selectively absorbed: non-absorbed radiation transmitted by the solution
causes its colour. For example, the purple colour of potassium permanganate
solutions is due to their strong absorption of green, and transmission of
violet and red wavelengths (Fig. 5-2). Solutions of coloured substances can
therefore be characterized by their transmission or absorption curves for visi-
ble light.
99
Solutions of most metallic salts are not usually strongly coloured except
at high concentration (the blue colour of a copper sulphate solution becomes
apparent at about 100 Mg/ml) and are therefore not directly amenable to
trace analysis. Exceptions are determination of trace levels of Mn as
permanganate or Ti as the yellow colour produced with hydrogen peroxide
in an acidic solution. Almost invariably, however, it is necessary to combine
the analyte with an appropriate colorimetric reagent, usually an organic com-
pound, with which it develops a stable coloured complex at very low con-
centrations. An additional step, usually required, is the extraction and con-
centration of the coloured complex from the aqueous phase into an immisci-
ble organic phase. Providing the solution of the coloured substance is suffi-
ciently dilute the relationship between its concentration and absorption is
given by Beer's Law:
simple, inexpensive and suited to field use. Many methods based on visual
colour comparison have therefore been developed for analysis of exploration
samples. Usually the coloured solution in a test tube is matched, against a
uniform white background, with a series of identical test tubes containing
standards of known concentration (see Fig. A-l, Appendix 1). A typical
series might, for example, cover the range 0—1 Mg/ml in 0.2^g increments,
1—10 μg in l.O^g increments and 10—20 μg in steps of 2.0^g. With such
a series an experienced analyst can rapidly match unknowns and standards
and interpolate between them giving results which are quite adequate for
most routine work. Regular preparation of standard series, using the same
batches of reagents used for unknowns, avoids the problem of deterioration
of standards with time and provides a check on methodology and reagent
quality. Any standard series that cannot be arranged in the correct sequence
without reference to its known content should be discarded. Use of artificial
standards, such as coloured plastics, is not recommended unless the real
standards deteriorate very rapidly.
In spectrophotometric analysis concentrations are estimated by measuring
the intensity of those wavelengths most strongly absorbed by the coloured
substance. The instrument is set at zero absorption against a solution blank
and absorption by standards and unknowns determined. Differentiation of
Beer's Law with respect to concentration shows that, ignoring other sources
of error, optimum concentrations for measurements are in the range 0.2—
0.7 absorbance (40—80% absorption): measurements close to 0 or 100%
absorption should be avoided if possible.
A wide variety of spectrophotometers are available. These vary from
simple instruments, using filters to isolate the absorption wavelength, to
more sophisticated instruments with monochromators. Solutions are usually
placed in rectangular glass or quartz cells which have been carefully matched
to have the same absorption characteristics. These cells must be handled with
great care and their optical surfaces kept scrupulously clean. To avoid
repeated handling and changing of cells some spectrophotometers employ
flow cells through which the solutions can be pumped. At least one manufac-
turer produces an instrument with a glass fibre optic cell which is lowered
into the solution to be tested. In general the simplest spectrophotometers are
fully adequate for routine exploration analysis.
Very few colorimetric reagents, 2,2' biquinoline for Cu being one, are
specific for a particular element. Most react to a group of elements and must
be made more selective by controlling reaction conditions with respect to
pH, Eh, time and temperature, and by use of masking agents which prevent
reaction with unwanted elements. Conditions for determination of most of
the elements of interest are treated very thoroughly by Sandell (1959) and
colorimetric reagents that have been used for analysis of exploration mate-
rials are summarized in Table 5-1. As an example of the applications and
problems of colorimetry the uses of dithizone (diphenylthiocarbazone), a
101
TABLE 5-1
Some colorimetric reagents used in analysis of geochemical samples
Ag dithizone 56
catalytic oxidation Mn 2+ 55
As HgCl2 1-3,9-11
silver diethyldithiocarbamate 4,12,75
Molybdenum Blue 4
Cassius gold purple 84
Au Brilliant Green 2,27
4,4'-bisthiobenzophenone (TMK) 28
crystal violet 29
B methylene blue 1,17
Ba sulphate 2,3,13
Be beryllon II 2,14
Bi diethyldithiocarbamate 1 - 3 , 15, 16
Co 2-nitroso-l -napthol 3,19
tri-rc-butylamine 2,20
Cr as chromate 2,3,18
diphenylcarbazide 2,18
Cu 2,2 , -biquinoline 2 , 3 , 21, 77
dithizone 1, 2 2 , 2 3
rubeanic acid 24,76
Neocuproine 25
Ge phenylfluorone 2, 3, 26
Heavy metals dithizone 1,3,72-74
Hg dithizone 2,3,32
I as iodine 30
Mn as permanganate 2 , 3 , 31
Mo dithiol 1,2, 3 3 - 4 1 , 78, 85
thiocyanate 3, 4 2 - 4 4
Nb thiocyanate 3,79
Ni a-furildioxime 2, 3, 4 5 - 4 7
dimethylglyoxime 48
P as phosphomolybdovanadic acid 2,3,49
Pa reduction of phosphomolybdic acid 1, 50,80
Pb dithizone 1-3
Pt reduction of phosphomolybdic acid 1, 50,80
Rhodamine 6G 51
Rare earths (several methods) 52,81
Re thiocyanate 53
103
pH
ION
1 2 3 4 5 6 7 8 9 10
+
Ag
H9 2 +
2+
.2+
Pb2
organic solvent. A few determinations deviate from this sequence. For exam-
ple, in the determination of As, arsine gas is produced by reduction of As
with nascent hydrogen: the arsine is then allowed to react with filter paper
soaked in mercurous chloride giving a coloured spot or, alternatively, is
bubbled through a solution of silver diethyldithiocarbamate (Fig. 5-4).
Chromatographie techniques, especially paper chromatography, can also
be used to separate and estimate concentration of coloured complexes of
trace elements. Theory and application of chromatography in geology have
been described by Ritchie (1964, 1969) and Hunt et al. (1955). Ritchie
TABLE 5-II
Procedure for the colorimetric determination of lead with dithizone (based on Stanton,
1966)
Mercuric chloride
paper
ABSORPTION
0 ring
HEAD
Absorption tube
ARSINE
GENERATOR
Silver
Diethyldithiocarbamate
solution
- Zinc pellets
Fig. 5-4. Determination of As by generation of arsine and its reaction with (A) mercuric
chloride paper, or (B) silver diethyldithiocarbamate dissolved in pyridine or quinoline.
Some ions in some solvents form coloured spots which can be marked on
the chromatogram as soon as it is taken out of the vessel. Other ions, being
colourless on the paper, need to be treated with reagents (by spraying or
dipping) to produce either coloured or fluorescent spots." The solvent is
usually a mixture of a miscible organic solvent and a mineral acid.
Ritchie (1964) gives detailed Chromatographie methods for many trace
metals. However, the only elements routinely determined by chromatog-
raphy in the exploration laboratory are Nb (and Ta) and perhaps U. In the
determination of Nb (Hunt et al., 1955; Stanton, 1966) the sample is decom-
posed with HF and a 0.05-ml aliquot applied to the chromatography paper.
After drying for an hour the paper is stood in a solvent mixture of ethyl
methyl ketone and hydrofluoric acid until the solvent has diffused to the
106
top. The paper is removed, exposed to ammonia vapour for three minutes
and sprayed with tannic acid. Niobium then appears as an orange-yellow
band with a detection limit of about 4 ppm. Rapid Chromatographie methods
for determination of U have been described using potassium ferrocyanide
(Thompson and Lakin, 1957; Hunt et al., 1955; Ward et al., 1963) or l-(2-
pyridylazo)-2-naphthol (Plamondon, 1968); detection limits are 4 ppm and
1 ppm, respectively.
FLUORIMETRY
TABLE 5-III
Comparison of uranium content of sediments determined by fluorimetry, with and
without solvent extraction, after digestion with 4 N nitric acid (data provided by S.J.
Hoffman)
58 282 238
1 3 118
2 5.5 226
5 16 954
0.2 1.0 5200
0.5 1.1 440
0.2 3.9 4300
16.8 62.5 240
0.7 6.8 3850
1.3 11.4 300
8.0 18.8 720
161.0 435.0 665
0.1 10.7 1640
31.2 105.0 680
0.8 370.0 5400
0.1 160.0 405,000
excites fluorescence by both natural dissolved organics and uranyl ions (Fig.
5-5). However, the fluorescence of the latter persists very slightly longer than
that of the organics when the laser is switched off. Consequently, by modu-
lating the laser and tuning the detector, it is possible to measure only the
"ORGANIC"
FLUORESCENCE
URANIUM
FLUORESCENCE
TURBIDIMETRIC METHODS
INTRODUCTION
THEORY
absorption
Fig. 6-1. Excitation of an electron from a lower (£1) to a higher (#2) energy level by
absorption of energy (E^) where E^ = E2 — E\. On collapsing back from E2 to E\ the elec-
tron releases a photon with an energy equal to E± and hence, from Planck's Law, a wave-
length in nanometres (nm) of 1.24/ü^, if E± is in kilovolts.
atom (or ion) has its own unique emission spectrum (Fig. 6-2). Conversely,
excitation of an electron to a higher energy level can result from absorption
of radiant energy of the same wavelength as the emission line for that partic-
ular transition. This is the basis of atomic absorption spectrophotometry.
The ratio of excited (N{) to ground state (N0) atoms at different tempera-
tures is given by:
i Pi (-ΕΛ
N,
where P{ and P0 are statistical weights for the ground and excited state, Ei is
the energy difference between the states, k is the Boltzmann constant, and T
is temperature in °K. As the value of Ei increases (i.e. as wavelength
decreases according to Planck's Law) the smaller will be the fraction of
excited atoms for a given temperature. As temperature increases the number
of excited atoms increases. However, even at high flame temperatures the
proportion of excited atoms remains relatively low (Table 6-1). Atomic
absorption measurements are usually made in flames with temperatures
between 2000 and 3000°C so that virtually all the atoms are in the ground
state (or energy levels very close to the ground state) and N0 is relatively
insensitive to small fluctuations of flame temperature. Also, because only
absorption lines originating in the ground state (or close to the ground state
e.g. Sn 235.48 nm and Zr 360.12 nm) will be sufficiently abundant to cause
measurable absorption, the absorption spectrum is much simpler than the
emission spectrum (Fig. 6-2).
The relationship between the intensity of an incident (I0) and transmitted
(It) light beam, of wavelength λ, passing through an absorbing cloud of
atoms follows Beer's law:
It = I0 exp — (Kxcl)
Ill
10,000
20,000
30,000
40 000 E
υ
50,000
μβο,οοο
^70,000
Fig. 6-2. Grotrian energy level diagram for the Au atom. The wiggly arrows indicate a
transition that is forbidden by the selection rules, a and b are metastable levels. Only
those wavelengths associated with the ground state (i.e. 164.6 nm, 242.8 nm and
267.6 nm) are involved in atomic absorption but the line at 164.6 nm is too low in the
UV to be useful and the 242.8-nm line is usually used for AAS determination of Au.
(Slightly modified from Ahrens and Taylor, 1961, Spectrochemical Analysis, 2nd ed.,
with permission of Addison-Wesley, Advanced Book Program.)
TABLE 6-1
The relationship between temperature, wavelength and the number of excited (Nj) to
neutral (N 0 ) atoms
I\ and P0 are statistical weights derived from the quantum number J such that P =
2 J + 1: e.g. Zn 213.8 arises in the transition 'S0 -+ 'Px so that P0 = 1 and P{ = 3.
INSTRUMENTATION
TABLE 6-II
Crustal abundance of some trace elements compared to their analytical sensitivity by
atomic absorption spectrophotometry
electronic modulation
"o-'t
*γ«
nm i · i n \
[ Ί
I i' ·m
^
>t lo
B.
«o""t
~γ*
i m
Half-silvered
Rotating mirror
mirror
sectors
With a width of 0.001—0.003 nm, atomic absorption lines are too narrow
to be isolated by small monochromators, but if an attainable bandpass of
0.1 nm is used complete absorption of a continuum source by a line
0.002 nm wide only results in a 2% change in signal strength — too small a
difference to be measured accurately. Walsh (1955) realized that this
dilemma could be overcome by having the light source emit only the spec-
trum of the element of interest. Ideally the emitted lines should be sharper
(narrower) than the absorption line of the ground state atoms.
For a few volatile elements, such as Hg, As and Se, electrodeless discharge
lamps are ideal sources of an intense, stable emission. With most elements,
however, the best source is the sealed hollow cathode lamp. This consists of
115
Ni, j j g / m l
Fig. 6-4. Influence of varying lamp current (mA) and slit width (nm) on the sensitivity for
Niat 232.0 nm.
116
Flames
Usually the sample solution is aspirated, by a venturi driven by the pres-
sure of the oxidant gas, into a chamber where the resulting aerosol mixes
with fuel gas before entering the flame through a long, narrow slit. Burners
of this type, known as laminar-flow or pre-mix burners, produce very steady,
stable flames. Aspiration rates with aqueous solutions are usually 3—5 ml/
minute. However, only a small fraction of the solution reaches the flame; the
remainder, consisting of the larger droplets, is trapped by baffles in the
TABLE 6-III
Modifications to routine operating conditions
Situation Action
A. High analyte concentration — to decrease (a) use less sensitive absorption line
sensitivity (b) rotate burner
B. Very low analyte concentrations — to (a) increase slit
improve signal to noise ratio (b) increase lamp current
(c) both (a) and (b)
C. Strongly luminous flames — to improve (a) decrease slit
signal to noise ratio (b) increase lamp current
(c) both (a) and (b)
117
Fig. 6-5. Distribution of Cu and Mo atoms in lean and fuel rich air-acetylene flames. Con-
tours are drawn at intervals of 0.1 absorbance units with maximum absorbance in the
centre: height is measured above the burner slot. (Reprinted with permission from Rann
and Hambly (1965), Distribution of atoms in an atomic absorption flame. Analytical
Chemistry, 37: 879—884. © 1965 American Chemical Society.)
however, these parameters are not too critical in the lean air/acetylene flame
and the best conditions for any one of the elements which are probably most
frequently requested in the exploration laboratory (Cu, Co, Ni, Pb, and Zn)
will usually be satisfactory for the remainder of the group. Considerably
greater care is needed to optimize flame conditions and the position of the
light beam in the red cone or feather of the fuel-rich nitrous oxide/acetylene
flame.
Furnaces
Resistance heated furnaces offer several theoretical advantages over flames
as a means of producing ground-state atoms, notably more efficient utiliza-
tion of the sample solution and a longer residence time for atoms in the fur-
nace than in the flame. This gives sensitivities appreciably better than those
attainable in flames (Table 6-II) and enables very small quantities of material
to be analyzed. Major disadvantages of furnaces, compared to flames, are
more severe interferences and their much lower analytical productivity.
The simplest furnaces are the silica tubes, heated either by resistance wire
(Chu et al., 1972) or by mounting them above an air/acetylene flame
(Thompson and Thomerson, 1974), used for atomization of elements, such
as As, that form gaseous hydrides. The hydride generating system and fur-
nace used by Aslin (1976) to determine As and Sb in geochemical samples is
shown in Fig. 6-6.
Less specialized furnaces for general analytical applications are usually
resistance-heated graphite tubes (Fuller, 1977). Alternatively, a strip (fila-
ment) of an unreactive metal, for example tantalum, enclosed in a chamber
119
Nitrogen
Nitrogen B
Tygon tubing HOLLOW CATHODE LAMP
Plastic
stopper (tight fit)
CONTINUUM LAMP
ΊΓ
Eppendorf tip
Fig. 6-7. Typical absorption curves for the dry, charr and atomize stages of electro-
thermal atomizers. Subtraction of non-atomic absorption, measured with a continuum
source, from total absorption, measured with the hollow cathode lamp, gives the signal
due to the analyte. During drying and charring, smoke and vapour can give large non-
atomic absorption signals.
can be used. The power supply to such furnaces must provide three cycles,
each with variable time and current settings, for drying, charring, and finally
volatilization and atomization of the analyte. At the high temperatures (up to
3000°C) used for atomization, the furnace material would be rapidly con-
sumed by oxidation in air. It is, therefore, arranged for the furnace to be
sheathed with an inert gas (argon or nitrogen) or, to provide a more reducing
environment, hydrogen.
Sample solutions are usually introduced into the furnace with a micro-
pipet (1—100 μΐ) and the solvent evaporated during the drying cycle. The
current is then increased sufficiently to char and burn off organic matter
and, if possible, other matrix constituents. Ideally this step, which produces
smoke, should be completed with no volatilization of the analyte before the
atomization cycle begins. If charring overlaps with atomization, light losses
caused by the smoke must be subtracted from the true absorption signal by
simultaneous background correction. Some typical drying, charring and
atomizing, cycles are shown in Fig. 6-7.
120
Wavelength selection
APPLICATIONS
Furnace techniques are well suited to direct analysis of waters but cannot
compete, either in productivity or freedom from interferences, with flame
AAS for the rapid routine determination of most of the elements of interest
in soils and sediments. Consequently, even although concentration by
solvent extraction may be required (Table 6-IV), most AAS analyses in the
prospecting laboratory involve aspiration of solutions into laminar flow
flames. Notable exceptions are the flameless determination of Hg and the
electrothermal decomposition of gaseous hydrides.
Willis (1975) has described an interesting technique for avoiding the
sample decomposition stage, whereby a 5% suspension of the powdered
sample in water is aspirated directly into the flame: Cu, Ni, Co, Zn and Pb
were determined in stream sediments.
TABLE 6-IV
Application of some solvent extraction schemes to geochemical samples
l
Element Extraction scheme Reference
Ag TOTP-MIBK C h a o e t a l . (1971)
Bratzel et al. (1972)
Au Cl-MIBK Bratzel et al. (1972)
Br-MIBK Tindall (1965)
Thompson et al. (1968)
DBS-toluene Rubeskaet al. (1977)
Be acetylacetone Terashima (1973)
Bi APDC-MIBK Ficklin and Ward (1976)
Mo SCN-MIBK Kim et al. (1974)
Aliquat 336-MIBK Rao (1971)
Pd DBS-toluene Rubeskaet al. (1977)
Pt Sn n -MIBK Stanton and Ramankutti (1977)
Sb Cl-MIBK McHugh and Welsch (1975)
TOPO-MIBK Welsch and Chao (1976)
Sn TOPO-MIBK Welsch and Chao (1976)
Te Br-MIBK Nakagawa and Thompson (1968)
C h a o e t a l . (1978)
Tl and In Br-MIBK Hubert and Lakin (1973)
w SCN-Alamine 336-MIBK K i m e t a l . (1976)
1
Abbreviations: TOTP = triisoctyl thiophosphate; MIBK = methyl isobutylketone, 4-
methyl-2-pentanone; Cl = chloro; Br = bromo; DBS = dibutyl sulphide; APDC = ammo-
nium pyrrolidinedithiocarbamate;SCN = thiocyanate; TOPO = trioctylphosphine oxide.
122
Sensitivity
Of the trace elements in Table 6-II, crustal abundance of twelve (Ba, Be,
Co, Cr, Cu, Mn, Ni, Pb, Rb, Sr, V and Zn) is such that they should be
detectable with routine operating conditions and dilution factors in the
123
Cd scale = 2.5
Α
2.0 η —Α Α
9
1 · — · Cd
I D—D CO
1.6 A
j o o Cu
0 2 4 6 8 10 12 14
pH
Zn scale = 2.5
♦ — ♦ Ni
1.6 H
0—0 Pb
LU Δ—Δ Zn
0 2 4 6 8 10 12 14
PH
Fig. 6-8. Influence of pH on the extraction of metals into MIBK with 1% w/v APDC +
DDDC. (Reprinted with permission from Kinrade and Van Loon, 1974, Solvent extrac-
tion for use with flame atomic absorption. Analytical Chemistry, 46: 1894—1898. ©
1974 American Chemical Society.)
Interferences
Freedom from interferences is often cited as one of the principal advan-
tages of AAS. Nevertheless, if samples and standard solutions differ in bulk
composition a variety of interferences can cause enhancement or depression
of the analyte signal. Principal sources of interferences and some remedial
methods are summarized in Table 6-V.
Most interferences could be overcome either by careful matching of
sample and standard solutions, or by use of the method of additions. Alter-
natively, the analyte could be isolated from interferents by solvent extrac-
tion or some other separatory technique. For example, prior to analysis of
126
TABLE 6-V
Some interferences in the determination of trace elements in geological matrices by flame
atomic absorption spectrophotometry
0.5Ί
g ,
< /
er °· 3 Ί 7
°
CO /
/
CD ' 10 uq/ml Mo
< 02J /
1 / 10^ig/ml Mo +
/ 1 0 0 p g / m l Ca
/
1 1 1 1 1
200 400 600 800 1000
ALUMINIUM, ^g/ml
0.15 0.15
A 40ppm Ba + 10,000 ppm K
• 40 ppm Ba + 2,000 ppm Ca
o
z
<
CO
o
c/>
CO
< 0.05 -4 0.05 H
TABLE 6-VI
Correction for background absorption in lead analysis of a synthetic rock solution (from
Fletcher, 1970)
(ppm) *
cathode lamp H 2 lamp corrected uncorrected corrected
Dilution factor = 50
130
L.JbJ
100 100 100
A% l% A%
Fig. 6-11. Background correction using a continuum source. The profile of the atomic
absorption line is shown in solid black and the continuum is represented, over a slit width
of Δλ, as the area within the dashed line. At the light source the emission intensity of the
atomic line and continuum are balanced at 100%. In the flame atomic absorption is super-
imposed on background absorption so that the intensities reaching the detector are Ia and
Ib) respectively. The readout displays these intensities as a net signal equal to (100 — Ia) —
(100 -Ib).
80
70
60
50
40
30
20
10
0\- O^
-10
4 6
■r-ψ-^-ψ: 12
Θ 10
ΡΡΦ Pb in solution
Electrothermal atomization
(Ficklin and Ward, 1976); Cd (Gong and Suhr, 1976); Te (Corbett and
Godbeer, 1977; Chao et al., 1978); Mo (Kontas, 1976); and precious metals
(Fryer and Kerrich, 1978). Other references and a discussion of the deter-
mination of individual elements are to be found in Fuller (1977).
In marked contrast to analysis of solids, a rapidly expanding literature on
application of flameless AAS to water samples has developed in response to
environmental problems. Natural fresh waters contain a wide range of dis-
solved ions and constitute a complex, non-uniform analytical matrix. Conse-
quently, background absorption and inter-element interferences, although
less severe than with solids, are still present. Nevertheless, multi-element
studies by Edmunds et al. (1973) and Barnard and Fishman (1973) have
shown that for rapid screening and anomaly detection, direct injection and
comparison with standard calibration curves could be adequate if accom-
panied by simultaneous background correction. Fordham (1978) obtained
more accurate results by the method of additions: however, only sixty deter-
minations per day were possible.
Hydride generation
tion volumes and gas flow rates. Greenland and Campbell (1976), deter-
mining Te in rocks, found it difficult to maintain maximum sensitivity
without constant attention to the equipment. They, therefore, opted to use
less sensitive conditions for routine analysis: even so, as little as 5 ppb Te
could be detected with a 0.25-g sample.
As usual, early reports on the absence of interferences proved illusory.
Interferences reported by various workers do, however, differ somewhat,
apparently depending on reagent strengths and volumes and order of mixing.
In a detailed study, Smith (1975) generated the hydrides of As, Bi, Ge, Sb,
Se and Te from dilute hydrochloric acid with sodium borohydride. Using an
argon/hydrogen flame he found: Cu, Ag, Au, Pt, Pd, Rh, Ru, Ni and Co
always suppressed hydride generation — probably due to their preferential
reduction and precipitation, resulting in coprecipitation or adsorption of the
analyte, and consumption of reductant; and the hydride-forming elements
mutually interfere. Belcher et al. (1975) found that suppression of the evolu-
tion of arsine and stibine, from 0.1 M hydrochloric acid, by Co, Ni, Zn, Fe,
Bi, Cd and Cu could be almost eliminated by addition of EDTA. In more
strongly acidic media (5 M HC1), Aggett and Aspell (1976) found no inter-
ference from up to a 10,000-fold excess of Cu on the evolution of arsine.
Problems with high reagent blanks for As, Sn and Sb in sodium borohydride
have been reported.
Rapid methods of analysis of geochemical samples for As, Sb, Se and Te
have been described (Table 6-VII). With hot acid decompositions it is neces-
sary to prevent volatilization of As and Se by maintaining them in their
highest valency state, which must then be reduced to the trivalent state for
quantitative hydride generation. Large amounts of nitric acid remaining after
decomposition cannot be tolerated and should be removed by evaporation
with perchloric acid. Interference studies suggest that Fe, Cu, Ni, Mn and Co
are most likely to be troublesome in geological materials. Aslin (1976) found
that Ni suppressed evolution of arsine from 1.5 M hydrochloric acid and
both Ni and Ag suppressed stibine; the interferences could, however, be con-
trolled by addition of EDTA and it then became possible to determine
2 ppm Sb in the sulphide standard SU-1. Thompson and Thoresby (1977)
and Thompson et al. (1978b) found no significant interferences in the evolu-
tion of arsine from 5 M hydrochloric acid. However, the latter authors did
find it necessary to avoid interference from Cu in the hydride generation of
Bi, Se and Te by their coprecipitation and separation on lanthanum
hydroxide. Greenland and Campbell (1976) also reported interference from
Cu, in excess of 200 ppm, in the determination of Te.
A it
I Differential
I Amplifier
Fig. 6-13. Design of a spectrometer using pressure broadening of the 253.7-nm Hg line to
correct for non-atomic absorption in the determination of Hg. A pressure-broadened Hg
spectrum, emitted by a low-pressure mercury lamp, passes through an absorption
chamber and is then split into two beams, one of which falls directly on a photomultiplier
(PMA). The other beam passes through a cell saturated with mercury vapour so that the
central portion of the 253.7-nm Hg line is completely absorbed leaving only the wings of
the line to reach the reference photomultiplier (PMR). When a vapour causing non-atomic
absorption is pumped into the absorption chamber, output from both PMA and PMR is
proportionally reduced: atomic absorption due to Hg only reduces the signal fromPM A .
(Modified from Barringer, 1966.)
Ultra-violet
(?fi Lamp
/\
Photocell Photocell
4 pump
Glass w o o l '
mm m W^A
Palladium chloride
on glass wool
Silica gel
tm
Fig. 6-14. A double-beam mercury vapour meter. The sample stream is split into two
halves, one of which passes through glass wool loaded with palladium chloride, to remove
Hg but not constituents causing non-atomic absorption, before entering the reference
absorption cell on the right side of the instrument. Output from the two photomultipliers
is compared. (From James and Webb, 1964.)
Infra-Red
Heater Lamp
" I I I i \ | | | I i I \ \ V
Hg°
(U Absorption Cell
Quartz Window
Air
Gas Washing
Bottle
: Ä ■
ym
Fig. 6-15. Cold vapour generation and determination of Hg by AAS. In the gas washing
bottle Hg 2+ is reduced to Hg° with Sn 2+ and then swept into the absorption cell by a
stream of air. The IR heater lamp prevents condensation of water vapour in the cell.
138
EMISSION SPECTROSCOPY
INTRODUCTION
THEORY
x 5040 _
log = -logp e + 2.5 log T - — - - V - 6 . 1 8
1 —x 1
where log pe, the electron pressure, has a value of 10" 4 t o 10~ 3 atmospheres
in the carbon arc (Boumans, 1966).
The difference between t h e ionization potential (V) and the apparent
ionization potential (V) is small and need not concern us for the purpose of
illustrating the relative effects of temperature and ionization potential on the
degree of ionization (Table 7-1). It is apparent from the Saha equation that
for many elements both atom and ion lines will be present in their arc spec-
141
TABLE 7-1
Percent ionization of Na, Ca, Cu and Zn at different temperatures in the DC carbon arc.
Calculated using the Saha equation assuming an electron pressure of 4.1 X 10~4 atm
Na 5.14 5.49 16 89 99
Ca 6.11 5.82 7 79 99
Cu 7.72 8.13 <1 2 40
Zn 9.39 9.06 <1 <1 10
V = ionization potential.
V = apparent ionization potential from Boumans (1966).
trum and for those elements that are relatively readily ionized there will be
a pronounced weakening of their atomic spectrum. In the ICP, which oper-
ates at temperatures above 5000° K, ion lines can assume considerable
analytical importance (see Fig. A-2, Appendix 1). It should be noted that
when an atom is ionized by loss of one electron, the resulting ion spectrum
will resemble the atomic emission spectrum of the element whose atomic
number is one less.
In the DC-arc the sample powder is volatilized directly. Sample composi-
tion, particularly the content of readily ionized elements, therefore has a
considerable influence on arc temperatures, on the rates of volatilization of
individual elements, and on the distribution of excitation states in the
resulting atom and ion populations. Furthermore, conditions in the arc
column will progressively change as the relatively more volatile elements are
lost first, by selective volatilization, and arc temperatures rises (Fig. 7-3).
The proportions of emitting species can therefore vary considerably, both
between samples and at different stages in the arcing process. In quantitative
analysis these variations must be controlled, with spectroscopic buffers, and
monitored and corrected for using internal standards (Fig. 2-6).
readily excited — notably the alkalies and alkali-earths. Chemical and ioniza-
tion interferences are similar to those described for AAS.
Analysis by flame emission spectroscopy has almost entirely been replaced
by AAS and is seldom used in the exploration laboratory. However, it should
be remembered that most atomic absorption spectrophotometers are readily
converted to operate in the flame emission mode. Consequently, if no
hollow cathode lamp is available, quantitative determination of Li and other
alkalies is still possible providing precautions are taken to prevent chemical
and ionization interferences.
Equipment
The basic components of a spectrographic system (Fig. 7-1) consist of an
excitation source, usually a DC-arc, and its power supply; the spectrograph
itself, with either a prism or diffraction grating; and a camera to record spec-
photomultipliers
behind exit slits
on focal circle
lens source
source
Y
slit A
F7 \
—— . v
«^^OOnm
prism
— ■ — ^ 0 0 nm
camera or
plateholder
Fig. 7-2. External optical arrangements to give uniform illumination of the slit by
focussing the image of the source at the grating. In (A) a lens immediately in front of the
slit focusses the image of the source at the spectrograph's internal collimating lens; in (B)
focussing the image of the source on an adjustable diaphragm allows any portion of the
arc to be selected for analysis.
With practice visual comparison of line intensities becomes very rapid and
is entirely adequate for semi-quantitative analysis. Manual densitometric
measurement is both considerably slower and unjustified when the magni-
tude of errors arising from other sources is considered. The additional
expense of recording densitometers or microphotometers is therefore seldom
warranted and a simple projector comparator will suffice for estimating line
intensities. The comparator should provide a magnification of about 20X, to
avoid operator eyestrain, and be designed to permit rapid alignment of the
sample and reference film or plate.
Heiz (1973) and Walthall (1973) have automated plate reading using a
sophisticated high-speed scanning microphotometer capable of measuring
the intensity of up to 400 lines in only 70 seconds. In a semi-automatic
system, Tait and Coats (1976) transferred digitized densitometric readings
directly from a manually operated microphotometer to punch tape for com-
puter calculation of element concentrations and correction of interferences.
Analytical throughput was comparable to that with visual estimation but
results were more reliable. It seems probable tht attempts to automate plate
reading, where semi-quantitative visual estimation is regarded as inadequate,
will be superceded by the use of ICP direct-reading spectrometers.
145
Operating conditions
Apart from the optical arrangement of the spectrograph the main variables
to be considered are excitation conditions and choice of analytical lines. Five
semi-quantitative procedures used in analysis of exploration samples are
compared in Table 7-II. All five employ uniform illumination of the slit with
anode excitation and the electrodes are of similar shapes and sizes, except
that a hemispherical or conical cathode is used in methods 1 and 4 in an
attempt to reduce arc wandering. Other workers have apparently found flat
cathodes equally satisfactory.
Principal differences between the methods are the duration of excitation,
ranging from 5 seconds to total consumption of the sample at 120 seconds,
and the use or absence of spectroscopic buffers. Total consumption of the
sample has the advantage of eliminating variations in line intensity caused by
different rates of element volatilization in samples of different bulk composi-
tion. On the other hand, background is increased and the line to background
ratio becomes less favourable for volatile elements that are released at an
early stage (Fig. 7-3).
Sample admixtures with graphite and buffers are used to improve and
maintain arc conditions. Addition of graphite alone, usually in a 1 : 1 to 1 :
4 ratio, results in a smoother burn and higher arc temperatures, thereby
reducing differences between volatilization rates of different elements and
increasing the intensity of lines of involatile elements. Avni et al. (1972)
found matrix effects, in standard rocks ranging from granite to peridotite,
were appreciably reduced when samples were mixed 1 : 5 with graphite and
only the central 2 mm of a 6-mm arc gap recorded. A disadvantage is greatly
increased cyanogen (CN) band emission caused by the combination of car-
bon and atmospheric nitrogen in the arc.
Addition of an alkali salt (L1CO3, NaCl, Na 2 C0 3 , NaF) as a spectroscopic
buffer, together with graphite, lowers arc temperature, enhances selective
volatilization and reduces CN emission. Still more important, differences in
bulk composition between samples are reduced and background variations
and matrix effects on line intensities are significantly diminished. For exam-
ple, with a lithium carbonate—carbon buffer Nichol and Henderson-Hamilton
(1965) found that matrix effects on analyte concentration were less than
twofold even between such compositional extremes as a siliceous soil and a
ferruginous laterite. In contrast, method 5, with no addition of graphite or
buffer, is subject to considerable matrix effects — though results are still
acceptable for many purposes. Tennant (1967) has described an A1203/
CaC0 3 /K 2 C0 3 ( 7 : 3 : 1 ) buffer which increases selective volatilization,
lowers general background and greatly increases sensitivity for volatile ele-
ments. Sensitivity for Mo is also improved (Delavault and Marshall, 1973).
To determine concentrations ranging from less than 10 ppm up to 1%
several analyte lines of differing sensitivities may have to be used for each
146
TABLE 7-II
Operating conditions for five DC-arc semi-quantitative methods of analysis of geochemi-
cal samples (all dimensions in millimetres unless otherwise indicated) 1
2
Reference Spectro- D Wave- Film/ N ζ Anode
graph (nm/mm) length plate
(nm) a b c
Myers et al. Wadsworth 225.0
0.524 4"X10" 38 68 4" ^ m w a ^
(1961) grating 475.0 United Carbon
3170
Nichol and Hilger and 280.0 6.15 2.91 5.00
2"X10" 12 14
Henderson- Watts E492 495.0 Ringsdorff
Hamilton RW402 Graphite
(1965)
Dorrzapf Ebert 230.0 y thin wall
0.500 4"X20" 68
(1973) 3.4 m 470.0 Ultra No. 3170
(Mark III)
Tait and Hilger 225.0 3.2 1.6 6.0
0.500 24 20
Coats 3.4 m 350.0 Johnson-Matthey
(1976) Ebert Spec-pure
ARL 232.0 35-mm
— 0.695 22 24 6.15 3.0 5.0
1.5 m 474.0 film
1
D = dispersion; N - number of spectra per film; Z = number of elements determined;
T = exposure time.
2
a: anode diameter, b: diameter of crater, c: depth of crater.
3
tode Sample Arc T Arc Slit Comments
charge gap (s) current (μιη)
e (mm) (A)
c a t h o d e d i a m e t e r , e: shape,
imple: buffer r a t i o s .
Fig. 7-3. Volatilization curves for various metals, and variations of voltage, current and
electrode gap with a DC-arc. (From Timperley, 1974.)
Ag 338.29 2 Mo 319.40 1 V
328.07 1 Mn 317.04 1 Fe
281.62 2
As 278.02 2
234.98 1 Cu Nb 313.08 1
228.81 1 309.44 1
295.09 2
Au 267.60 2 SiO
242.80 1 SiO Ni 349.30 1
341.48 1 Zr
B 249.77 1 Fe : SiO
249.68 2 SiO P 255.49 2 Fe
255.33 1 Fe
Ba 455.40 1
307.16 2 Pb 287.33 3
283.31 1
Be 332.13 2 266.32 2
313.04 3
234.86 1 Zr Sb 287.79 1 Cr
259.81 2
Bi 306.77 1 Fe
289.80 2 Mn Sn 326.23 2
317.50 1
Cd 346.62 2 Fe 284.00 3 Mn
326.11 3
228.80 1 Sr 460.73 1 Fe, Mn
Co 345.35 1 Ti 398.98 1
340.51 2 Ti 337.76 2
Cr 428.97 3 V 437.92 1
427.48 2 Ti 318.54 1
425.44 1 Ho
W 294.70
Cu 327.40 2
324.75 1 Mn Zn 334.50 Ti
282.43 3
Zr 339.20 Fe
Ga 294.36 1 Fe
2874.2
Li 323.26 1 Ti
Mn 403.45 2
280.11
TABLE 7-IV
Composition of two standard mixtures to give 1% of each element
Mix A Mix B
compound g compound g
identify and mark all analytical lines. Alternatively line identification can be
simplified by construction of a set of cardboard or plastic wavelength guides,
showing both analytical lines and prominent major element lines, to fit the
comparator screen. Films are positioned against the wavelength guide by
aligning the most readily identifiable major element lines. Prominent lines
and groups of lines suitable for this purpose are shown in Table 7-V.
Occasionally unknown lines will have to be identified. This task is greatly
facilitated if films are available in which the pure spectra of individual ele-
ments alternate with the spectrum of a typical geochemical sample.
Analysis of unknowns
TABLE 7-V
Some prominent lines in geochemical spectra
Mg 1 line 285.21
Si 1 line 288.16
Fe 3 lines 309.99, 310.03 310.07
CN bandhead 359.04
CN bandhead 388.34
Ca 2 lines 393.37, 396.85
Al 2 lines 394.40, 396.15
152
(Fig. 7-1). Output from the photo multiplier, which is dependent on incident
light intensity and hence on element concentration, is fed to a voltmeter or
other electronic readout system.
With only a few photomultipliers it would be feasible to convert volts to
concentration units manually. However, for efficient utilization of a direct
reader, with perhaps as many as 60 analytical channels, concentrations must
be calculated automatically and most direct readers are therefore interfaced
to small dedicated computers. Ideally the computer should also be pro-
grammed to correct for background, matrix effects and spectral interferences
thus improving precision and accuracy of the data. Various aspects of the
application of direct reading spectrometers to determination of trace ele-
ments in geochemical matrices have been discussed by Cruft and Giles
(1967), Tennant and Sewell (1969), Scott et al. (1969), Foster (1970),
Timperley (1974) and De Pablo Galan (1975).
Initial standardization of the direct-reading spectrometer involves firing of
synthetic or natural standards of known major and trace element composi-
tion. Calibration curves, either of line intensity versus concentration or
intensity ratio to an internal standard versus concentration, are then stored
in the computer. Depending on the concentration range of interest and the
sensitivity of the analytical lines, the calibration curves can be either
approximately linear, or exhibit a pronounced flattening at low concentra-
tions because of background or the presence of an interfering line (Fig. 7-4).
100
without correction
corrected
1 10 100 1000
CONCENTRATION, ppm
Fig. 7-4. On a log-log plot the presence of background or an interfering line can cause the
slope of the calibration line to flatten at low concentrations (dashed line). Subtraction of
the intensity obtained with the same matrix, in the absence of the analyte, extends the
lower limit of the linear range (solid line).
154
Matrix effects
Matrix effects can be reduced in several ways most of which are normally
used, either singly or in combination, in quantitative spectroscopy and are
fully described by Ahrens and Taylor (1961). Use of spectroscopic buffers
has already been mentioned on p. 145. Other important methods of
reducing matrix effects include the use of internal standards and calculation
of matrix correction factors from the bulk chemical composition of the
sample.
Introduction of an internal standard into the sample allows variations in
line intensities, due to matrix effects or arc instability, to be monitored and
corrected providing behaviour of the internal standard is similar to that of
the analyte (Fig. 2-6). Their respective volatilization rates are particularly
important in this respect. Indium is generally suitable as an internal standard
for volatile elements (Ga, Pb, Ag, Cu, Tl, Ge, Sn, Zn) and Pd for involatiles
(Zr, Sc, Y, Nd, La, Sr, Ba, V, Ni, B, Co, Cr, Mo, Mn, Ti). Suitability of an
internal standard should, however, be checked by comparing volatilization
curves under actual operating conditions. This can be done by programming
the direct reader to present results for each channel At short intervals of time
throughout the duration of the burn. Timperley (1974) has presented a
series of curves prepared in this way (Fig. 7-3).
By analyzing natural and synthetic standards of known major and trace
element composition it is possible to calculate correction factors for matrix
effects of major elements on each other and on trace elements. For example,
De Pablo Galan (1975) prepared 13 synthetic matrices, with variable Si0 2 ,
A1203, Fe 2 0 3 , CaO, MgO, Na 2 0 and K 2 0 contents representing a wide range
of geological materials, containing trace element concentrations from 1 to
4600 ppm. By analyzing these matrices and solving a series of simultaneous
equations, correction factors (x) were obtained for the effect of the con-
155
Fig. 7-5. The tape machine. A = stock roll; B = folding device; C = safety device; D = per-
forating device; E = sample hopper; F = pressure-suction device for removal of excess
sample powder; G = spark gap; H = air blast; K = pulling wheels; and L = waste bin. Sam-
ple powder is continually added to the adhesive tape via the hopper (E) and then passes
through the spark gap where a new portion of the sample is continually excited.
156
Background
l.b - .
1 4- //
z
LU
1.2 -
1.0-
1 //
/ ^ / /
(0 0.8-
>^^~\ Diabase W-1 ^ /y/Synthetic //
3 < ^ — \ ^>^ * ^ / ^ - b a s i c rock //
0.6-
o f V ^ ^ - - ^ " G i r a n i t e G-1^/^-->^ /
cc
O
(J
0.4 -<
\/y\^<zzzri^_^Z^
^ Synthetic acidic rock
^^
0.2-
<
0 1 1 1 1 1 1 1 1 1—i—
250 260 270 280 290 300 320 340
WAVELENGH, nm
Fig. 7-6. Relationship between background intensity and wavelength in a DC-arc for
several different rock types. (From Tennant and Sewell, 1969.)
157
line (spectral) interference and only the background estimate from the inter-
ference-free side of the line is used. With some direct-reading spectrometers
the entrance slit of the spectrograph can be automatically oscillated to pro-
vide a display of the line profile and adjacent spectrum.
Spectral interferences
PLASMA SOURCES
tures much higher than those attainable in flames or arcs. Microwave plasmas
have been used to analyze the volatiles released by inductive heating of soils
and sediments (Meyer and Lam Shang Leen, 1973). However, the greatest
interest in plasmas has focussed on the general analytical capabilities of the
inductively coupled (radio-frequency) plasma in conjunction with the direct-
reading spectrometer. In contrast to DC-arc spectroscopy the sample is
usually introduced into the plasma in solution: the technique therefore has
all the advantages and disadvantages associated with dissolution of geochemi-
cal samples.
The first reports on the use of the ICP as an excitation source for trace ele-
ment analysis were published by Greenfield et al. (1964) and Wendt and
Fassel (1965). Theory and analytical applications have been reviewed by
Greenfield et al. (1975b, c), Fassel and Kniseley (1974a, b), Fassel (1978)
and Boumans (1979). State of the art information regularly appears in the
ICP Information Newsletter edited by R.M. Barnes (Department of Chem-
istry, University of Massachusetts, Amherst, MA 01002, U.S.A.).
Various designs of plasma cells have been used but modifications of that
first described by Greenfield et al. (1964) are generally regarded as most
satisfactory. These consist of two concentric quartz tubes surrounding a glass
sample injector (Fig. 7-7). Argon to form the plasma passes through the con-
centric quartz tubes, flow in the outer tube being tangential to produce a
vortex which both stabilizes the plasma and keeps it away from the tube
walls. (Consumption of argon, at 10—20 1/minute, is the major operating
cost of the ICP.) The upper part of the plasma cell is surrounded by a few
turns of an induction coil powered by a high-frequency generator: crystal-
controlled generators operating at 27 MHz and providing up to 2 kW of
power appear to be becoming standard. Greenfield et al. (1975b, c), how-
ever, prefer a more powerful (up to 15 kW) free-running generator operating
at lower frequencies (7 MHz).
High-frequency currents in the induction coil generate a magnetic field
around the plasma cell (Fig. 7-7). If the plasma gas is then seeded with elec-
trons using a Telsa coil, the electrons and ions move in closed annular paths
within the cell. Heating occurs from collisions between charged particles to
give temperatures of the order of 7000—10,000°K. After the plasma has
formed the sample aerosol is injected into its centre, creating a doughnut-
shaped plasma in which the sample passes through the hole without pene-
trating the skin of the plasma. This impenetrability of the plasma results in it
being relatively insensitive to injection of aerosols and hence able to tolerate
high sample loadings without loss of stability or quenching.
Although temperatures in the axial channel of the plasma fireball are
lower than those within the plasma, they are still much higher than tempera-
159
60 -
A. > - 3 5 7 . 8 6 9 nm
Excitation energy =3.46 eV /^^Nv
Q 20 mm / ^^
Z
3
O
GC
O
40 H
A-*"' >
/ y 15mm \
*
o
/ ' ■
<
CO
\UJ 2 0 - /
z
/^ #' -"♦~~ ^N. 10 mm
" V
mr—^^^f''
ψ:°—-~
^* >-♦
n
v'-'-Magnetic
field
/ ^ 15 mm
20-
'/ \ \
Q 'i
{
N
\l0mm
\
Z
•D
i* / \♦ \
o J" \
QC
O
15-
\ \
o* \\ \•
<
oo \\
X
UJ
10- 20^Tun ^
\
z
-1
s' \\ \
5- B. /> =267.716 nm \\ \ \\ \
Argon
cooling
Excitation energy \ v
12.6 eV V >
gas \ Λ
Sample 0
Plasma aerosol
Argon AEROSOL Ar FLOW RATE, I min -
Fig. 7-8. Influence of the aerosol argon flow rate on emission of (A) the atomic Cr line at
357.869 nm; and (B) the ionic Cr line at 267.716 nm, at three observation heights in an
inductively coupled plasma. (From Berman and McLaren, 1978a.)
tures attainable in flames or arcs and are more than adequate to vaporize the
solute and dissociate molecular species. However, the background continuum
in the plasma fireball is intense and for analytical purposes it is necessary to
view the spectrum at some point in the tail flame where background is low
but temperatures are still sufficiently high to prevent recombination of
atoms. Because the axial channel is cooler than the surrounding plasma the
problems of self-absorption and self-reversal — encountered with flames and
arcs in which temperatures decrease towards the outer fringes — are absent
and the plasma behaves as an optically thin source.
160
Temperature variations along the tail flame of the plasma can be dramati-
cally illustrated by aspirating a solution containing Ba (see Fig. A-2, Appen-
dix 1). Emission from the yellow (Ba 585.37) and blue (Ba 493.4) ion lines
is well separated from the continuum whereas the most intense emission of
the green (Ba 553.5) atom line occurs immediately above the continuum and
again, though somewhat weaker, higher in the plasma. As discussed below,
the response of ion and atoms lines to operating conditions in the plasma
can differ considerably. Boumans (1979) has listed the most sensitive lines
for seventy elements; approximately half of them are ion lines.
Principal variables controlled by the analyst are: (1) the argon and aerosol
flow rate, (2) the generator power fed to the plasma, and (3) the height in
the tail flame at which the emission is viewed. Generally the objective will be
to establish optimum compromise conditions for determination of as many
elements as possible, rather than to optimize for a single element. The prob-
lem of establishing compromise conditions has been discussed by Berman
and McLaren (1978a) who present a series of line to background ratios under
varying operating conditions for atom and ion lines. For atomic lines of Pb,
Mn and Cr the line to background ratio increased with viewing height and
aerosol flow rate up to about 1.0 1/minute (Fig.7-8). In contrast, with ion
lines the optimum line to background ratio occurs lower in the plasma and
with lower aerosol flow rates. Increasing power input to the plasma increases
line intensity but also causes a rapid rise in background so that line to back-
ground ratios are less favourable at maximum power (Fig.7-9). On the basis
-- A - 357.869 nm
Aerosol Ar
0.95 I min"1
— A = 267.716 nm fi
Aerosol Ar /
>* 150 0.80 I min"1 '
n
<
z
UJ
I-
Z 50
of these results, it was concluded that the best compromise conditions would
be obtained with relatively low argon flow rates and observation of ion lines
at a height of about 15 mm. These conditions should also minimize non-
spectroscöpic interferences.
Kirkbright and Ward (1974), comparing ICPs and flames as emission
sources, have noted that in addition to its higher temperature and lower self
absorption, the plasma is also characterized by both longer particle residence
times and freedom from the intense band spectra of CH, OH, C and CO
found in flames (or CN found in DC-arcs). Also, because the free electron
density is much greater than sample particle density, the degree of ionization
is essentially unaffected by sample introduction.
In practical terms the longer residence times and higher temperatures
should ensure complete volatilization of the solute and dissociation of
molecular species, thereby increasing sensitivity and eliminating band spectra
and interferences caused by formation of refractory compounds. This is
borne out by the virtual absence of the Ca-Al and Ca-P interferences which
are so troublesome in flames (Fig. 7-10; see also p. 127). Ionization inter-
ferences also appear to be less than with other sources presumably due to the
high electron densities (Larson et al., 1975). As a further advantage, the
combination of increased sensitivity and, at high analyte concentrations,
lower self-absorption gives the plasma a linear working range extending over
five orders of magnitude compared to three for flames. It is therefore possi-
ble to determine both trace and major elements on the same solution
without dilution.
In the absence of significant chemical and ionization interferences the
most serious analytical problems, particularly near an elements detection
limit, have been associated with spectral interferences from overlapping and
coincident emission lines or from background. Coincident or overlapping
spectral lines are not, of course, restricted to plasma sources: assuming
spectrometer resolution is adequate they are best avoided by careful selec-
tion of lines. If this is not possible, appropriate correction factors can be
computed by measuring the signal produced in the analyte channel by a
known concentration of interferent (Marciello and Ward, 1978). Tables of
line interferences in the ICP have been published by Parsons et al. (1980).
Correction of background is particularly important if low signal to back-
ground intensities are to be measured. Under these circumstances small
changes in background emission can severely degrade analytical performance.
Variations in intensity of background emission between samples and stan-
dards can result either from changes in the proportions of molecular species
emitting band spectra or from light scattering, causing stray light, in the
spectrometer. Stray light is most troublesome when strong emitters, such as
Ca, are abundant in the sample. Apart from minimizing stray light with
improved spectrographs and gratings, the most promising approach to this
problem appears to be measurement of background intensity adjacent to the
162
1.0 -e
£--*■-—r- ICP
» -*
=> (Ca 393.4 nm)
0.8 -
s.
<
0.6 -
AAS
or Air Acetyl ene <§L
h-
0.4 -
Z (Ca 422.7 nm) ^s.
LU
CC
<
a. 0.2 - v§)
o. A,
< —,
0 L-// .— 1
0 100 500 1000 □ 3.5 7o HCI0 4
PHOSPHORUS/CALCIUM RATIO
• 500 ppm Ca
CO
2.0 n z A 100 ppm Sr
AAS LU
♦ 100 ppm Ti
N 2 0-Acetylene I-
z
(Ca 422.7 nm) ($-
< LU
υ >
LAJ
ICP
z (Ca 393.4 nm)
LU 0.5 H
cc
<
Q.
Q.
< -//- —\ 1 —r~ ~v
0 10 30 100 300 1000 0.91 0.90 0.89 0.88 | 0.87 0.86
POTASSIUM/CALCIUM RATIO 400.875
WAVELENGTH, nm
Fig. 7-10. Comparison of chemical and ionization interferences in flames and the induc-
tively coupled plasma. A. Chemical interference of P on Ca. B. Ionization interference of
K on Ca. (From Ediger and Wilson, 1979.)
Fig. 7-11. Some wavelength profiles showing interelement effects on emission at the W
400.875 nm line in an inductively coupled plasma, a = 3.5% HCIO4 background; plus, · =
500 ppm Ca — broadband scatter; * - 100 ppm Sr — line reflection or scatter; and ♦ =
100 ppm Ti — spectral overlap from 400.893 nm. (Reprinted with permission from
McQuaker et al., 1979, Calibration of an inductively coupled plasma-atomic emission
spectrometer for analysis of environmental matrices, Anal. Chem.y 5 1 : 888—895. © 1979
American Chemical Society.)
analyte line (Fassel, 1978). This can be done by oscillating the spectrometer's
entrance slit to display a scan of the line profile (Fig. 7-11).
Detection limits reported by various workers using compromise conditions
and pneumatic nebulizers are summarized in Table 7-VL Appreciably lower
limits are obtained with ultrasonic nebulizers; these, however, are unsuited
to rapid sample throughput. The stability of the plasma enables relative stan-
dard deviations of about 5% to be achieved.
163
TABLE 7-VI
Detection limits in solution ^ g / m l ) with the ICP *
Element 1 2 3
As 0.039
B 0.1 0.001
Ba 0.0001 0.001 0.0003
Be 0.0002
Cd 0.1 0.002
Co 0.004 0.02 0.029
Cr 0.001 0.02 0.004
Cu 0.01 0.003
Fe 0.04 0.004
Ga 0.022
Mn 0.003 0.0006
Mo 0.02 0.004
Nb 0.001
Ni 0.003 0.01 0.017
P 0.17
Pb 0.008 0.07 0.037
Sb 0.050
Sn 0.1 0.029
Sr 0.02 0.0001
Te 0.19
Ti 0.005 0.03 0.002
V 0.002 0.009 0.002
w
Zn 0.01 0.1
0.15
0.003
1
Reference: 1 = Scott et al. (1974): detection limit twice standard deviation of back-
ground signal; 2 = Greenfield et al. (1975c): compromise conditions; 3 = McQuaker et
al. (1979): means over a one-year period.
Analysis of fresh waters with the ICP appear to present few problems and
has rapidly gained acceptance. Considerably less information is available on
application of the ICP to trace element determinations in soils, sediments
and geological materials, although the topic has been reviewed by Berman
and McLaren (1978b).
In one of the few published studies Scott and Kokot (1975) compared
determination of Cu, Pb, Zn, Co and Ni with the ICP and AAS. One-gram
soil samples were digested with nitric-perchloric acids and diluted to 50 ml
with water for aspiration into a 27-MHz, 1-kW argon plasma at 1 ml/minute.
When viewed 20 cm above the load coil no problems were encountered in
determination of Cu (324.7 nm) and Zn (213.8 nm) which had linear calibra-
tions over four orders of magnitude. Linear calibrations were also obtained
164
for Ni (351.5 nm) and Co (345.3 nm) but results were lower than those by
AAS: this appeared to be due to interferences in the flame rather than the
plasma. In the case of Pb, high Ca content of the soils caused spectral back-
ground in the vicinity of the most sensitive line (405.8 nm) and it was neces-
sary to apply a correction. Scott et al. (1976) obtained a detection limit of
0.1 ppm U in solution with the line at 378.28 nm. Motooka et al. (1979)
eliminated major element interferences and enhanced sensitivity for trace
elements (Ag, Au, Bi, Cd, Cu, Pb and Zn), after their extraction from geologi-
cal materials using potassium chlorate/hydrochloric acid (see p. 72), by sol-
vent extraction into MIBK with Aliquat 336. Although the plasma is some-
what less stable with MIBK than with aqueous solutions, results on standard
samples were well within the limits necessary for exploration purposes.
Dahlquist and Knoll (1978), McQuaker et al. (1979) and Floyd et al.
(1980) have discussed problems encountered in analysis of soils and similar
materials. Spectral wavelength scans indicated that broad band scatter and
spectral overlaps, caused by varying amounts of major concomitants, give
rise to significant interferences for most elements at low concentrations
(Fig. 7-11). Contribution of these interferences to the analyte signal can be
corrected using linear or second-order polynomial regressions. As reported
by other workers, the effects of chemical and ionization interferences in the
105-
ί
95
-
°/0 U p t a k e R a t e Γ
<
g
υ
100 -
<KJ ►
L
^90 UJ
UJ
CC
i H
► '
Q.
3
95- i
- |
% Recovery
90-
1 1 1 80
7.0 14.0
% HCI0 4
Fig. 7-12. Effect of perchloric acid content on uptake rate and recovery (when calibration
is carried out in 3.5% perchloric acid) in the inductively coupled plasma. Recoveries are
mean effects found for Cd, Cr, Cu, Fe, Mg, Mn, Mo, Pb and Zn. Bars indicate spread.
(Reprinted with permission from McQuaker et al., 1 9 7 9 , Calibration of an inductively
coupled plasma-atomic emission spectrometer for analysis of environmental matrices,
Anal. Chem., 5 1 : 888—895. © 1 9 7 9 American Chemical Society.)
165
X-RAY FLUORESCENCE
INTRODUCTION
present these instruments are only able to measure trace element concentra-
tions in excess of several hundred parts per million. They are therefore most
useful for determination of elements with background values in this range,
for example Ba and Sr (Ball et al., 1979; Grout and Gallagher, 1980) or for
on-site inspection of drill-core, pit faces or outcrops where the presence of
"invisible" mineralization or anomalous values is suspected. Friedrich et al.
(1974) have described a sled mounted system for remote analysis of manga-
nese nodules on the ocean floor. Development of portable instruments, their
use and associated problems have been described by Bowie et al. (1965),
Bowie (1968), Gallagher (1967, 1969), Garson and Bateson (1967); and Von
Alfthanetal. (1980).
Other specialized geochemical applications of XRF, beyond the scope of
this general text, include its use to determine trace elements after their pre-
concentration from solution by ion exchange. With natural waters this
involves collection of particulate and dissolved metals on ion-exchange
loaded filters (Blasius et al., 1972). Alternatively, trace elements can be
co-precipitated from natural waters or analytical digests, the precipitate
collected as a thin film on a filter paper and analyzed by XRF (Campbell et
al., 1966; Green et al., 1971; Watanabe et al., 1972). Stanton (1976), fol-
lowing procedures described by Luke (1968), gives rapid methods for deter-
mination of Bi; As, Se and Te; and Au, Pa and Pt by co-precipitation and
XRF.
THEORY
Numerous texts have described the properties of X-rays and the nature of
their interactions with matter — the reader is especially referred to Jenkins
and De Vries (1970). In this section the fundamental principles underlying
X-ray spectrometry are only briefly described.
Excitation of X-rays
\ = hc/V=1.24/V
169
l/X = k(Z-o)2
where k and σ are constants. A corollary to Moseley's Law is that more
energetic primary radiation is necessary to produce X-rays from the heavier
elements. For example, the minimum potentials needed to excite Ka spectra
for Cu (Z = 29), Sn (Z = 50) and U (Z = 92) are 8.98, 29.19, and 115.9 kV,
170
TABLE 8-1
K spectrum excitation potentials and wavelengths for selected elements (based on
Jenkins and De Vries, 1970)
are therefore generally favoured for analysis. However, for heavy elements
the excitation potential of the K spectra may exceed the instrument power
and safety rating. One of the more intense lines in the L spectrum is then
chosen. M spectrum lines are very weak.
The actual measured intensity or fluorescent yield for a particular transi-
tion is always less than predicted. This results from the Auger effect, which
involves the internal re-absorption of the secondary X-ray photon and causes
serious reduction of X-ray intensity for light elements.
μλ = Σ (μΖλ · Wz)
where μΖχ and Wz are the mass absorption coefficient and weight fraction
for each element (Z) in the material.
The mass absorption coefficient includes both photo-electric absorption
(r) and scatter (σ) so that μ = r + o. However, photo-electric absorption is
quantitatively far greater than scatter, so μ ^ r. A plot of μ for an element as
a function of wavelength shows several major breaks, or absorption edges,
each one corresponding to the minimum potential required to excite the
spectrum associated with a particular shell. For example, in Fig. 8-2 the K
absorption edge for Ni at 0.149 nm corresponds to the energy (8.33 kV) just
necessary to eject an electron from the K-shell of a Ni atom. Wavelengths
slightly longer than the K absorption edge have lower energies, cannot eject a
K-shell electron, and are, therefore, only subject to the photo-electric
absorption associated with emission of the L and M spectra. There is there-
fore an abrupt decrease of μ at the absorption edge. Maximum absorption of
incident X-rays, and therefore most efficient excitation of secondary X-rays,
occurs with wavelengths just shorter than the targets absorption edge: thus,
172
Fig. 8-2. Relationship between wavelength and the mass absorption coefficients of Ni and
Co in the vicinity of their K-shell absorption edges. Mass absorption coefficients of Ni and
Co for CU-KQ, are approximately 50 and 350, respectively.
in Fig. 8-2, Cu-Ka is strongly absorbed by Co (but not Ni) and would be an
ideal primary radiation for excitation of Co-Ka. Between absorption edges μ
is proportional to λ 3 and Z 4 .
Scattering of X-rays by the target is usually much less important than
photo-electric absorption. Nevertheless, scattering cannot be ignored as it
contributes to background radiation reaching the detector and also has some
very useful analytical applications. Total scatter is a combination of coherent
(Rayleigh) and incoherent (Compton) scatter. Rayleigh scatter, which is the
phenomenon involved in X-ray diffraction, results from elastic collisions
between the incident X-rays and target electrons. The primary radiation is
scattered without a change in wavelength. In contrast, Compton scatter
involves inelastic collisions in which the primary X-ray photons loose some
of their energy to the electrons. As a result the scattered X-rays are at a
slightly longer wavelengths than the primary beam. The change in wave-
length (Xd) is given by:
Xd = (1 — cos φ)
mc
where h = Planck's constant, m = rest mass of an electron, c = velocity of
light, and φ = angle between the primary beam and the scattered beam. For
most spectrometers φ is approximately 90° so that the wavelength shift
between coherently and incoherently scattered lines is about 0.0024 nm
(Fig. 8-3). The intensity of the scattered radiation is proportional to Z" 3 t o " 2 .
Consequently, as shown in Fig. 8-3 for Mo-Ka radiation scattered from
teflon, silica and lead discs, scattering of radiation is most intense from
light matrices and can be used as an estimate of μ (Fig. 8-11).
173
KocCompton
26 24 22 20 18 16 12
DEGREES 2Θ
Fig. 8-3. Relationship between the intensity of coherent (Rayleigh) and incoherent
(Compton) scattering of Mo tube lines with sample discs having low (teflon), medium
(silica) and high (lead) mass absorption coefficients. Note that the Compton peaks are too
small to be detected with the lead disc and that the general intensity of background also
decreases as the mass absorption coefficient increases. See also Fig. 8-11.
INSTRUMENTATION
A. WAVELENGTH DISPERSIVE
(sample
Goniometer arm
Detector
(sample
°o°
Electronics,
o w Amplifier,
o
Multi-Channel
Analyzer
C. PORTABLE UNIT WITH
ISOTOPE SOURCE
Nal Crystal Filter pair on
rotating mount
Electronics m
<s~ Photomultiplier fc^l
Safety
shutter
analysis becomes possible. Jenkins and De Vries (1970) and Woldseth (1973)
should be consulted for more detailed information on wavelength- and
energy-dispersive X-ray spectrometers, respectively.
Primary radiation sources
In commercially available X-ray fluorescence units sample fluorescence is
invariably excited in one of three ways: with an X-ray continuum generated
in a X-ray tube; with monochromatic X-rays produced by exposing a suitable
target to an X-ray continuum; or with λ-rays emitted by a radio-isotope.
X-ray tubes
When sufficiently energetic electrons strike a target they loose their
energy in an irregular step wise fashion and an X-ray continuum (or Brems-
strahlung) is emitted. The hardest radiation emitted, i.e. the short-wave-
length end of the continuum (X min ), is related to the energy of the electron
beam by Planck's Law:
Xmin = 1.24/V
where V is the voltage in kilovolts acrsss the X-ray tube. Peak intensity of
the continuum is at approximately 1.5 X min . The overall intensity of the con-
tinuum increases with tube voltage and filament current, and with the
atomic number (Z) of the target. If the energy of the electron beam exceeds
the excitation potential for the target's X-ray spectrum, the target element's
characteristic lines will be superimposed on the continuum.
In practice, primary X-rays are generally produced in a sealed vacuum
tube containing a W filament (the cathode) and a target which forms the
anode. A high-voltage generator, stabilized to better than 0.5%, is required to
operate the tube. A current through the filament produces electrons which
are accelerated by the tube potential towards the anode. On striking the tar-
get the electrons lose their energy producing the X-ray continuum and a
large amount of heat. The target must therefore be cooled and fabricated
from materials with good thermal characteristics. Suitable elements are Au,
W, Ag, Mo, Cu and Cr. X-rays leave the tube through a side window made
from a thin Be sheet.
Targets of heavy elements such as W produce the most intense X-ray con-
tinuum and are, therefore, often favoured for routine trace element analysis.
However, a characteristic line is best excited by wavelengths just shorter than
the corresponding absorption edge. The shortest wavelengths of the con-
tinuum are therefore not necessarily the most desirable, and the peak inten-
sity of the continuum and the presence of characteristic target lines may be
important. For example, a Mo target emitting Mo-Ktt at 0.071 nm is very
effective in exciting Rb-Ka and Sr-Ka with absorption edges at 0.0816 and
0.0770 nm, respectively.
176
X-ray tubes are normally operated close to their maximum power rating in
order to produce both the most intense continuum and t o excite the K spec-
tra of heavy elements. Occasionally, to prevent line interferences or reduce
background, it may be helpful to operate below the minimum excitation
potential for the K spectrum of a particular element.
Finally, the use of primary X-rays to excite secondary monochromatic
fluorescence, which is then used to excite characteristic radiation from the
sample, should be mentioned. The secondary fluorescence is, of course,
much less intense than the primary continuum source. However, provided
the target is chosen so that the monochromatic radiation lies on the short
wavelength side of and close to the absorption edge associated with the
characteristic line of interest, adequate intensities can be obtained. The tech-
nique is particularly useful in energy dispersive spectrometry for which the
high background scatter produced by a continuum source is unacceptable.
Isotope sources
Radio-isotopes can provide suitable primary sources when either back-
ground radiation is a limiting factor (as in energy-dispersive methods) or the
instrument must be compact and/or portable. Source strength is generally a
few millicuries. The relevant properties and analytical applications of several
useful isotopes are summarized by Gallagher (1969).
Dispersion of X-rays
Before its intensity can be measured the characteristic line must be sepa-
rated from other secondary X-rays and from scattered primary radiation.
Three methods are employed: (1) wavelength dispersion by X-ray diffrac-
tion; (2) filters with absorption edges bracketing the analytes characteristic
line; and (3) energy discrimination with semi-conductor detectors. Descrip-
tion of the last method is deferred to p. 1 8 1 .
Wavelength dispersion
In the wavelength-dispersive spectrometer the principal components of the
dispersive system are the primary and secondary collimators, the analyzing
crystal, and the movable goniometer arm carrying the detector (Fig. 8-4).
The collimators are a series of parallel metal plates arranged, rather like
Venetian blinds, to ensure that an essentially parallel beam of X-rays reaches
the crystal and detector and thereby exclude unwanted wavelengths.
Increasing the collimation, by decreasing the gap between the plates or
increasing their length, improves resolution by decreasing the range of wave-
lengths reaching the detector but also decreases line intensity.
The collimated beam of X-rays falling on the crystal is diffracted according
to Bragg's Law so that, if λ is in nanometres:
ηλ = 2d - sin Θ
177
or about 2d. Thus a LiF crystal, reflecting on the (200) crystal plane with
d = 0.2014 nm, has a X m a x of 0.4028 nm and potassium is the lightest ele-
ment that can be determined.
The ability of the spectrometer to resolve adjoining lines depends on its
dispersive power and the line widths. Dispersive power, the angular separa-
tion of adjoining lines, is given by:
d0/,dX = n / 2 d - cos Θ
TABLE 8-II
d-spacings of some diffraction crystals used in X-ray fluorescence (based on Jenkins and
De Vries, 1970, and manufacturer's literature)
K series L series
heaviest elements towards the shorter-wavelength end (low 20) of the spec-
trometer. Since λ is proportional 1/Z2 this is also a region where character-
istic lines are relatively closely spaced thereby increasing the possibility of
spectral interference. Consequently, providing some loss in line intensity can
be tolerated, it may be desirable to improve resolution by using increased
(finer) coUimation. Conversely, light elements with low fluorescent yields are
best determined using coarse coUimation in conjunction with crystals dif-
fracting their characteristic lines at large values of 20.
Filters
The simplest methods of isolating a particular characteristic line is with
balanced filters having absorption edges immediately adjacent to and on
either side of the analyte's characteristic line. A typical example would be
the use of a Ni-Co filter pair to isolate Cu-Ka which is transmitted by the Ni,
but not the Co, filter (Fig.8-2). The filters must be carefully matched, by
adjusting their thickness, so that they transmit equal intensities except for
wavelengths lying between their absorption edges. Readings are taken first
with one filter in position and then the other, the difference in readings
giving the intensity of the characteristic line. If the element to be determined
is a major constituent, for example Fe, giving rise to almost all the short
wavelength radiated emitted by the sample, it may only be necessary to iso-
late the characteristic line using the long-wavelength filter.
Use of filters is restricted to compact, portable XRF units, a purpose for
which their simplicity is ideally suited (Fig. 8-4). Generally such instruments
can be supplied with sets of interchangable filter discs, each containing
several filter pairs chosen by the analyst.
Detectors
The detector has the dual role of both measuring X-ray intensities and
sorting them on the basis of their energies. Measurement of X-ray intensities
is dependent on the ability of X-rays to ionize matter, releasing electrons and
creating a pulse which can be amplified and measured. Sorting them on the
basis of their energy is possible because the voltage (v) of the pulse produced
by an X-ray photon entering the detector is proportional to the energy of
the incident X-ray photon (e0), that is:
v °c e0/ei
where ex is the ionization potential for the detector. Three types of detectors
are in general use: gas-filled proportional counters, scintillation counters, and
semi-conductor detectors.
chamber, containing an inert gas, which X-rays enter through a thin mylar
window. The thinness (less than 6 μιη) of the window makes it impractical
to exclude air or prevent gas leakage. The inert gas, usually argon mixed with
methane, is therefore allowed to flow continuously through the chamber.
On entering the detector an X-ray photon ionizes inert gas atoms (e{ ^
20 eV) to produce primary ion pairs:
the average number of primary ion pairs depending on the energy (e0) of
the photon. The potential applied to the anode causes the electrons to
accelerate towards it, resulting in collisions between accelerated electrons
and inert gas atoms that produce secondary ion pairs and give a cascade
effect known as gas amplification. In the region of proportional response the
voltage of the amplified pulse produced by the counter remains proportional
to e0. At higher anode voltages this relationship breaks down and the detector
behaves as a geiger counter.
Scintillation counters
In a typical scintillation counter X-rays penetrating a Be window are
absorbed by a thallium-activated sodium iodide crystal causing it to phos-
phoresce in the blue region of the visible spectrum. Some of the emitted
light photons strike a photomultiplier, mounted in optical continuity with
the crystal, to produce a voltage pulse. The amplitude of the pulse depends
on the efficiency of the detector, for which the effective e{ is about 50 eV,
and on counter voltage.
n = e0/ei
where n is the number of electrons initially produced when the X-ray photon
enters the detector. Like radioactive decay, the phenomenon is random and
n has a Poisson distribution so that the standard deviation on = \fn. With
typical values of n the Poisson distribution approximates a normal, Gaussian
distribution and counter resolution (r) is defined as:
2.35σ„Χ100 w X 100
r(%) =
Natural Peak
VOLTAGE
Fig. 8-5. X-ray pulse amplitude distribution of a detector. Detector resolution (%) is
defined as 100 X w/υ By setting the detector electronics to a lower threshold voltage (Vi)
together with a window (V2 — Vi), pulses lying below V\ and above V2 can be rejected.
181
Semi-conductor detectors
The development of semi-conductor detectors is a relatively new field
which has advanced rapidly and several energy-dispersive X-ray analyzing
systems are now available. Their outstanding advantage, compared to wave-
length-dispersive spectrometers, is that true simultaneous multi-element
determinations become practical with a relatively inexpensive instrument. A
serious disadvantage is that the detector must be maintained at low tempera-
tures, using liquid nitrogen, to prevent its deterioration.
In X-ray spectrometry the semi-conductor detector is a lithium-drifted
silicon crystal between two metal electrodes. A high voltage applied across
the electrodes produces a large electric field in the detector. When an X-ray
photon enters the detector ionization occurs and the resulting ions migrate
to the electrodes. The effective ionization potential is much lower (about
3 eV) for semi-conductor detectors than for gas proportional or scintillation
counters. Consequently, their resolution is much greater. For example,
resolution of Cu-K^ with a scintillation counter, a proportional counter and
with a semi-conductor detector, would be about 65%, 15% and 5%, respec-
tively. Thus, with a semi-conductor detector it is possible to divide the X-ray
spectrum emitted by a sample into a number of relatively narrow energy
channels; the X-ray intensity within each channel is stored, processed and
displayed by a multi-channel analyzer.
Energy-dispersive spectrometry permits a much simpler sample-detector
geometry than can be used with wavelength dispersion (Fig. 8-4). This con-
siderably improves counting efficiency, permitting use of less intense
primary sources. However, rather than measuring only a narrow band of
X-rays dispersed by an analyzing crystal, the semi-conductor must handle all
incoming X-ray photons emitted by the sample. Consequently, a particular
characteristic line may account for only a small proportion of the total num-
ber of counts — perhaps a few hundred or a few thousand out of up to
50,000 cps — registered by the detector. Under these conditions it is desira-
ble to reduce the contribution of background radiation as much as possible.
The largest single source of background is normally the back-scattered con-
tinuum and characteristic radiation from the X-ray tube. Background can
therefore be greatly reduced if an isotope source or monochromatic radia-
tion, generated by bombarding a secondary target with primary X-rays, is
used to excite the sample.
182
Instrument conditions
o o o o o o o q o r- N n n ^ m Φ N o <j> o
do odd d ö ό ö ό o o d d ό ό ό ό ό^
Mill I I 1I I I I I I I I I I I I I J L
ZnNi
w Dy
__|
Ba
I
Sn Ag
I I
Mo
I
Sr
i
As GalCul Fe Cr Ti
I I I ιΠ ii I I I I
CaK Cl S P Si AI Mg Na
I I I I I
Th PbAuW Dy Ba Sn Ag Mo
LIF(220)
- TOPAZ ~
-LIF(200)-
_E.D.D.;r_
CRYSTAL
PE
--+ GYPSUM* rt
* KAP-
S.C.-
F.C.
DETECTOR
COLLIMATOR COARSE-
- AIR
MEDIUM -VACUUM -
proportional counter, in conjuction with LiF (for Fe, Ti and Ca), PET (Al
to K) and TLAP (Na and Mg) crystals, provide optimum conditions.
The principal remaining operating conditions relate to the detector, in
particular to the detector potential and the settings for pulse height analysis.
Optimum detector voltage should be selected for each element following the
manufacturer's instructions — generally this involves setting the goniometer
20 position for the peak to be measured and with a sample in place gradually
increasing the detector voltage. The count rate should initially increase
rapidly until a plateau is reached. Optimum voltage is near the middle of this
plateau. Hutchison (1974) gives an excellent description of optimizing
counter voltage and pulse height analysis with the Philips 1540 spectrometer.
Pulse height analysis often seems to be a subject of some confusion and,
since its misapplication is probably worse than no use at all, it warrants a
rather detailed description. It will be remembered from p. 178 that the
voltage of the pulse produced by both proportional and scintillation detectors
is proportional to the energy (e0) of the X-ray photon. Pulses corresponding
to X-ray photons having different energies can therefore be sorted and
rejected by setting the detector electronics to a lower threshold value and a
window (Fig. 8-5). Only pulses having energies lying above the threshold and
within the window are recorded.
It remains to account for the unwanted pulses registered by the detector.
These usually arise as a result of either harmonic repetition of the X-ray
spectrum or from fluorescence of the components of the spectrometer.
From the Bragg equation higher-order (n > 1) lines of heavy elements can
coincide with or overlap a first-order analytical line. However, their energies
will be very different. For example, first-order Mn-Ka (0.2103 nm) is partly
overlapped by the second-order reflection of Au-L^2; their respective energies
are 5.88 kV and 11.57 kV. Because of the large energy differences involved,
pulse height analysis with a flow proportional counter can completely
separate the lines. With the poorer resolution of a scintillation counter some
overlap remains — the calculations to demonstrate this are given in a worked
example by Jenkins and De Vries (1970). The intensity of second-order and
higher lines is relatively weak so the harmonic overlap is unlikely to be a
problem unless light elements are to be determined in heavy matrices.
Instrument fluorescence can arise if the sample emits or scatters X-rays at
wavelengths shorter than the absorption edges associated with the con-
stituents of the dispersing crystal or some other component of the spectrom-
eter: these can then fluoresce and become a source of extraneous pulses.
Crystal fluorescence is not a problem with the LiF crystal. It might, how-
ever, for example be encountered in the measurement of P-Ka with a
gypsum crystal — P-Ka, Ca-K^ and S-Ka would all reach the detector.
Unwanted pulses can also arise within an argon-filled proportional
detector if an X-ray photon, giving rise to a natural peak (i.e. the peak to be
measured), has sufficient energy (e0) to eject an electron from the argon
184
ee = e0 — eA
when eA is the energy associated with the argon K-shell absorption edge. The
escape electron produces a pulse on the low-energy side of the natural peak,
the separation of the two peaks increasing as the energy difference between
the escape electron and eA decreases. In practice, escape peaks are encoun-
tered for elements K to Ni with argon-filled detectors.
With sodium iodide scintillation detectors escape peaks should be asso-
ciated with ejection of K- or L-shell iodine electrons with absorption edges at
0.0374 nm and 0.2389 nm, respectively. However, the scintillation detector
would not normally be used at wavelengths longer than Fe-Ka (0.1936 nm)
and only elements heavier than La will give rise to the K-shell escape elec-
trons.
Counting strategy
oR=ER= y/RjT
and, since variances are additive, the standard deviation (a a ) for the analyte
signal (E a ), with Ra = Rp — Rh, is:
tfa = >/σΓ+ ol
■m+
T ' τ b ρ
E(%)
V¥kl·'-** X 100
If counting time is the same at the background and peak positions (i.e. T p
T b = Γ/2) this can be rearranged to give:
_ 100^2 V E p + Rl
E(%)
—
^ρ ^b
As usual the detection limit is defined as the background signal plus two
standard deviations: in this case Rh and a b , respectively. However, as noted
above, two measurements must be made — one at the peak (Rp) and one at a
background position (Rh). The standard deviation for the analyte (a a ) is
therefore greater than the standard deviation for background ( a b ) alone, by
a factor of:
186
At the detection limit Ä p ^ Rh so that this factor is close to \/2 and the
detection limit becomes the concentration giving a net count rate (Rp ■ A b )
equal to:
? 0,000 360,000
1,000
- 100
10
20 30 40 50 60 70 Θ0 90 100
Ca Zn Zr Sn Nd Lu Hg Th
Atomic number
Fig. 8-7. Detection limits (ppm) by XRF at the 95% confidence level: based on 100 sec-
onds counting time for pressed powder pellets. Broken lines join results between which
unmeasured elements occur. (From Leake et al., 1969.)
187
TABLE 8-III
Analysis of geochemical samples by X-ray fluorescence
A. Single-element methods
Elements Reference
Sample preparation
β-α
mjtjb.
n . v Secondary X-rays
J
Primary X-rays '
Fig. 8-8. Sample particle and mineralogical effects in XRF. In a sample consisting of
rounded grains in a finer matrix, characteristic X-rays with a penetration depth (x), less
than the diameter of the grains, will underestimate the concentrations of constituents
contributed by the grains. To avoid this effect samples should be ground until x >> grain
diameter.
189
grinding to 50 μιη or less will prevent particle size effects with wavelengths
shorter than about 0.3 nm. Where grinding is not feasible and particle size
effects are suspected, Jenkins (1970) has suggested a correction based on the
intensity ratio of an element's K and L, or L and M lines.
As a check on the amount of sample required, varying amounts of powder
(or pellets of increasing thickness) can be loaded into the holder until no
further increase in intensity is observed. Feather and Willis (1976) found
that with pellets of pure quartz, 40 mm in diameter, 8 g of material sufficed
for Mo-Ka (0.071 nm) whereas 20 g was required with Ba-Ka (0.039 nm).
Grinding of a few tens of grams of material in a ring mill will therefore
usually provide an ample quantity of sufficiently fine powder.
Matrix corrections
where Ι8χ and Ιχχ are intensities (cps) on the standard and sample respec-
tively, μ8λ and μχχ are the corresponding mass absorption coefficients, and
Cs is the concentration of the analyte in the standard. Brown et al. (1973)
describe an iterative procedure, employing successive approximations of the
sample's mass absorption coefficient and major element content, for accu-
rate analysis of pelletized samples without resort to fusion methods.
Unfortunately, if only trace elements are of interest the bulk composition
of samples will seldom be known and some other method of estimating
matrix effects is therefore needed. The relationships between scattered radia-
tion and mass absorption (p. 172) provide several methods of making the
necessary corrections; before describing these it is useful to consider the
X-ray absorption spectrum of silicates as presented by Hower (1959).
190
'a = / l - / l b
7 7
K ib ib
191
Ca Absorption Edge
280
z z
ο 220 H Ti Absorption Edge- o
t-
Q.
OC
a.
OC
o O
</>
Fe Absorption Edge </>
03
<
03
<
O Relative Absorption
c/>
03
<
Absolute Absorption
0.20 0.30
1
Fe Absorption Edge
z
o
Q. Ca Absorption
18
O Ti Absorption 9
c/>
m
< Edge
LU
-· ♦ ♦ * *-
>
5 1.4·
_l
UJ
Diabase
Shale
Ounite
Granite
~r
0.15 0.20
WAVELENGTH, nm
Fig. 8-9. The X-ray absorption spectrum of typical silicates. A. Absolute mass absorption
coefficient of a dolerite and its absorption relative to that of alumina. B. Absorption
relative to alumina for a granite, diabase, shale and dunite. (From Hower, 1959.)
Because of its simplicity, Levinson (1975) has recommended the ratio meth-
od for analysis of exploration samples: using a 50-kV generator and 20-sec-
ond counting he obtained detection limits of 50 ppm for several elements.
Lower limits are achievable with more powerful generators and longer
counting times. In adopting the ratio method careful consideration should be
192
3- B
2-
Q. Y IFe 7„Zn
0-1 0-1
JL
/ ■
· 1 -7 0-1
1- ▲ 1 -7 2-6
+ 4-7 60
r~
2.0
1
Fig. 8-10. Calibration curves for Pb in ores. A. Absolute calibration, with %Pb plotted
against the intensity of Pb-L a , showing the wide scatter caused by the effect of varying
Fe and Zn contents on the mass absorption coefficient. B. Scatter has been considerably
reduced using the ratio method and plotting %Pb against the ratio of Pb-L a to intensity of
scattered radiation at 0.06 nm. (Reprinted with permission from Anderman and Kemp,
1958, Scattered X-rays as internal standards in X-ray emission spectroscopy, Anal. Chem.,
30: 1306—1309. © 1958 American Chemical Society.)
F
50-
e 2 <VX
40-
Magnetites x\
ί 30
(0
y^Fe-mica
XBCR
10-
>XVGSP
0
c) 10 20 30
10 5
cps MoK Compton
Fig. 8-11. Mass absorption coefficient for Sr-K a versus Compton scattering of the Mo-K a
tube line using a LiF(200) crystal. The slight curvature, that results from the difficulty in
resolving the small Compton peak from the coherent peak in samples with a high mass
absorption coefficient (Fig. 8-3), could be reduced using a LiF(220) crystal. (Data pro-
vided by R.L. Armstrong.)
C - λ
.^*0·09. r
8
λ ^s0.0
Using a silver-target tube the range can be extended up to Mo-K a and would
also include the La lines of Hf to U (Feather and Willis, 1976). Measurement
of Compton scatter is therefore a simple and rapid method of correcting
mass absorption differences for many elements of exploration interest.
Principal shortcomings of the original Compton scattering method, as
subsequently elaborated by Reynolds (1967), were:
(1) Samples with μ 0 .ο9 l e s s than 7 could not be handled because of the
difficulty of achieving infinite thickness in the sample cup.
(2) Samples with μ 0 .09 greater than 20 were excluded because the
Compton peak, lying on the shoulder of the more intense Mo-K a , could not
be reliably measured; and
194
(3) Mass absorption coefficients could not be extrapolated across the Fe-K
absorption edge to provide corrections for the lighter trace elements.
Solutions to each of these problems were suggested. For light, organic-rich
matrices Ryland (1964) has shown that Compton scatter can be success-
fully measured using the softer radiation of a chromium-target tube. With
high mass absorption coefficients, which would be expected in heavy mineral
concentrates, gossans or other iron-rich materials, separation of the Mo-Ka
Compton peak from the Mo-Ka peak can be greatly improved by replacing
the LiF(200) crystal with a LiF(220) crystal. This provides better resolution
with some loss of intensity. The separation can be further improved by
installing an yttrium oxide filter, for which Reynolds (1967) provides
detailed instructions, at the detector entrance slit. Mass absorption correc-
tions can then be made up to values of at least 80 for Mo.09·
The remaining problem, that of extrapolating μχ values across major ele-
ment absorption edges, was also dealt with by Reynolds (1967) with respect
to the estimation of Mn, Cr, V, Ti and Sc on the long-wavelength side of the
Fe-K absorption edge. A somewhat different approach, which also allows for
enhancement of the Ka lines of a trace element as a result of their excitation
by Fe-K radiation, has been described by Giauque et al. (1977).
It will have been recognized from the foregoing descriptions of the back-
ground ratio and Compton scatter methods that they have much in common.
In fact, the latter is a special case of the former. On this basis Feather and
Willis (1976) have proposed a simple, elegant method of background and
matrix correction based on determination of background intensity at a single
point. This can be either a Compton peak or, where this cannot be used
some other interference-free background intensity (Ih) on the short-wave-
length side of the Fe-K absorption edge. Their method is well suited to
exploration samples and therefore warrants a fuller description. The basic
premises are:
(1) Background results from scattering of the continuum emitted by the
X-ray tube and its intensity is therefore inversely proportional to the mass
absorption coefficient (Fig. 8-11).
(2) Ratios of mass absorption coefficients measured at any two wave-
lengths between major element absorption edges are constant (Hower,
1959).
(3) There is therefore a constant relationship between background mea-
sured at any two positions within the same wavelength region (Fig. 8-12).
These relationships can be used to estimate the background (7lb) and net
peak intensity (7a) for any analyte line. The first step is to use blank pellets
of pure compounds, covering a range of mass absorption values, to relate
background intensity at each line (7lb) to that (Ih) at the position chosen to
monitor background (Fig. 8-13). There is a residual component (R) to these
plots — probably arising from scatter within the spectrometer — so that true
background 7 lb , which equals mlh when m is the slope in Fig. 8-13, must be
195
'MgO
150
Al 2 0 3 >
\~
(Λ
c Z MgO
D UJ
O I-
Ü z
100 A Q AI2(W^ ;,
o NaCI .y z3
O CaCOo*^
KCI« s
AC
>Ti02
V205 v ^ T i 0 2 Si
ω >^Fe 2 03 B
3 Cr290 V
3: < u0
o <^
Fe 2 0 3 _l *l
cc
O
Ag X-ray tube LiF(220) <c
o j R
H I Residual
< υ
(0 I
QQ
100 200 300 400
L. · ; : t... I
BACKGROUND AT 35°, counts/s INTERFERENCE-FREE BACKGROUND INTENSITY, l b
Fig. 8-12. The linear relationship between two backgrounds at different wavelengths,
illustrated using blank pellets having different mass absorption coefficients. (From
Feather and Willis, 1976.)
Fig. 8-13. jEstimation of the true spectral peak background intensity (I\h), and the total
background intensity (B - J l h + R), from a plot of the intensity measured at an interference-
free background position (Jb) and the spectral peak position using blank pellets of pure
compounds representing a range of values for the mass absorption coefficient. (From
Feather and Willis, 1976.)
J lb + R = mlh + R
It should be noted that with typical silicate matrices R will be negligibly
small if 7b is measured at a Compton peak. Net peak intensity (/a) at the
analyte line is given by:
/.=/p-(/ib+Ä)=/p-(mib+Ä)
r .
1
= [* _Jp-(m/b+Ä)
ratio
■»rat.ir» ~ — -
mlh
Line interferences
Summary
TABLE 8-IV
Line interferences in X-ray fluorescence analysis of geochemical samples (based largely
on Leake et al., 1969; Webber and Newburry, 1971; and Feather and Willis, 1976)
Lines overlaps can often be reduced by using a LiF(220) crystal rather than LiF(200).
197
low mass absorption coefficients (Leake and Peachey, 1973). For these
materials and for trace elements, such as Ba and Sr, which occur in most geo-
chemical samples well above their detection limits by XRF, the ease of sam-
ple preparation is a considerable advantage when information on the total
content of an element is required. XRF is also ideal for rapid determination
of major, rock-forming elements when information on the bulk chemical
composition of samples is required.
Feather and Willis (1976), using a multi-channel spectrometer with a 160-
position automatic sample changer, were able to analyze 25,500 samples for
three elements in 54 working days. However, a multi-channel spectrometer
represents a major capital investment (Table 1-IV) and analytical throughput
with the less expensive sequential spectrometers, in which the goniometer
arm is programmed to move the detector from one peak position to the
next, would be appreciably lower. It is, therefore, of considerable interest
that advances in energy-dispersive spectrometry have reached the state that,
using backscatter corrections for mass absorption variations, quite reasonable
multi-element results can be obtained. Hansel and Martell (1977), for exam-
ple, determined Ni, Cu, W, Pb, Bi, Nb, Ag, Cd and Sn in stream sediments
ground to pass 325 mesh. Detection limits were from 5 to 20 ppm and varia-
tions in sample mass absorption, due to varying Fe contents, were corrected
using the intensity of the Compton scattered tube peak. Sixty samples per
day could be analyzed. Other examples of the use of energy-dispersive X-ray
fluorescence in analysis of exploration samples are given by Von Alfthan et
al. (1980), Clayton and Packer (1980) and Kramar and Puchelt (1980). The
instrument described by Von Alfthan et al. uses gas proportional, rather than
semiconductor detectors, and is portable.
Chapter 9
ELECTROCHEMICAL METHODS
INTRODUCTION
Polarographic analysis was developed almost fifty years ago and applied in
the early days of exploration geochemistry to determination of Bi, Cd, Cu,
Pb and Zn (Hawkes and Webb, 1962). However, these methods were soon
superceded by faster, simpler colorimetric methods so that, with the excep-
tion of pH measurement and several non-routine determinations, they fell
into disfavour. Despite considerable improvements in polarography and
related techniques such as anodic stripping voltametry, which have found
important applications in water chemistry (Whitney and Risby, 1975), this
generally remains true. In contrast, a relatively recent development — the
specific ion electrode — has greatly simplified estimation of the halogens,
especially F, and thereby promoted investigation of their geochemical behav-
iour and application t o exploration.
Non-routine electrochemical determinations include conductivity, dis-
solved oxygen and Eh. Equipment and methods are described in most texts
on analysis of natural waters (e.g. Golterman, 1970) and, since they are only
required occasionally by the exploration geochemist, will not be discussed
further except to mention the use of conductivity measurements to estimate
concentrations of F", Cl", NOJ and SOl~ in natural waters after their separa-
tion by ion chromatography (Smee et al., 1978). Bölviken et al. (1973) have
described an instrument for in-situ measurement of pH and Eh in diamond
drill holes. Although perhaps n o t strictly electrochemical, an ingenious
method for determination of gaseous Hg by measuring changes in the
resistance caused by adsorption of Hg on a thin gold film (McNerney et al.,
1972) also deserves mention here. The instrument is portable and, with an
absolute sensitivity better than 0.05 ng Hg, has been used to determine the
Hg content of soil gases for mineral exploration purposes (McNerney and
Buseck, 1973). The remainder of this chapter is devoted to pH and specific
ion electrodes.
200
THEORY
fill hole
Pt wire
Mercurous chloride
1
Wg
(calomel)
sented as:
Ag; AgCl, 0.1 M HC1II glass II sample solution I saturated KC1, Hg2Cl2; Hg
sensing electrode reference electrode
>
E
MOO
O
Q_
4*
Q
O
cc /
/
1-300-
/
—i-
iTi 10'' 10" 10 10"' 10" J
0.0002 0.002 0.02 0.2 20
FLUORIDE CONCENTRATION
Fig. 9-2. Response of the fluoride electrode to fluoride activity. At activities above
10~ 5 M response is Nernstian: at lower concentrations the slope flattens due to con-
tamination of the reagent solution with F~ and release of F " from the electrode's
lanthanum fluoride sensing crystal.
Fig. 9-3. Relationship between the activity coefficient (7) of an ion and the ionic strength
of a solution. Calculated from the Debey-Huckel expression:
-log Ίι = Azfy/T/il + äßVT)
where A and B are constants determined by the solvent (in this case water at 15°C), a,· is
the effective diameter of the ion, zt is its charge and / is the ionic strength of the solution.
Values used were: A = 0.5;J3 = 0.3262 X lO" 8 ;^· = 2;^· X 10 8 = 6.
203
response ion activity (A) rather than concentration (C) has been referred to.
In very dilute solutions the distinction is negligible. However, the two are
related by: A = jC where 7, the activity coefficient, decreases as the total
ionic strength of the solution increases (Fig. 9-3). Variations in ionic strength
therefore affect the results when the electrode is used to estimate concentra-
tion. In practice this problem is often overcome by preparing both sample
solutions and standards in a total ionic strength adjustment buffer (TISAB)
to eliminate such variations. For a discussion of the significance of the activ-
ity coefficient and its estimation the reader should consult Garrels and
Christ (1965).
Other factors requiring consideration in electrode systems are sensitivity,
specificity (i.e. freedom from interferences) and their speed of response.
These are considered with respect to individual electrodes in the following
pages.
DETERMINATION OF pH
determination are best kept under field conditions or only gently air-dried.
Before pH can be determined dry soils must be moistened by addition of
water or a salt solution. This introduces problems in so far as soil colloids
interact with the solutions to give different pH values with different solution
to soil ratios or when salt solutions, rather than water, are used. If water is
added pH increases as the water to soil ratio increases (Table 9-1). Usually
some definite water to soil ratio between 1 : 1 and 10 : 1 is used or, in a
more time-consuming procedure, the soil is brought to its moisture satura-
tion point by careful addition of water until the surface of the wet soil
glistens (Jackson, 1958; Peech, 1965). Addition of a salt solution, for exam-
ple 0.01 M calcium chloride, has the advantage of approximating soil-water
composition for many soils. This reduces ion exchange effects and the
system is much less suceptible to changes of solution to soil ratios than when
water alone is used. Junction potentials are also reduced. A disadvantage is
that it is not possible to continue with the same solution and determine its
conductivity.
Occasionally different pH values are obtained depending on whether the
reference electrode (but not the pH electrode) is placed in the soil suspen-
sion or in the clear supernatant liquid. This phenomenon, known as the sus-
pension effect, can be avoided by consistently keeping the reference elec-
trode at the same depth in the supernatant solution.
The variations obtained in measuring the pH of a single soil (Table 9-1)
illustrates the practical difficulties in measuring any theoretically "correct"
pH. For exploration purposes a simple, rapid procedure using a definite
water to soil ratio and measuring pH to ±0.2 units will generally suffice. To
avoid breakage of electrodes or changes in their response due to scratches,
robust electrodes protected by a plastic sleeve are most useful for mea-
surements in soil sluries.
TABLE 9-1
Effect of soil to water ratio and presence of salts on estimation of pH
Treatment pH
(soil : water)
soil A soil B
1 : 0.5 - 8.1
1:1 4.8 8.1
1:2 4.9 8.2
1:5 5.1 8.5
1 : 10 5.0 8.6
1 : 2 0.01 M CaCl2 4.3 7.3
1 : 25 1.0MKC1 3.9 8.3
205
Specific ion electrodes are available for at least sixteen ionic species and
many more can be determined using electrodes as endpoint detectors in titra-
tions. However, only a few of the available electrodes (Table 9-II) are likely
to be of interest in the exploration laboratory, and of these only the fluoride
electrode has already achieved widespread usage.
Measurements with electrodes can be made either directly or by methods
of addition. Direct measurements, which are the simplest and most rapid,
involve construction of a calibration curve on semi-log paper using standard
solutions (Fig. 9-2). Calibration should be checked regularly and a buffer
solution is usually added to both unknowns and standards to swamp out any
differences in ionic strength, adjust the pH to the optimum range and con-
trol interferences. Many specific ion meters are provided with logarithmic
concentration scales to simplify direct readout of results in concentration
units.
Methods of addition, in which a known amount of the analyte is added to
the unknown, avoid the need for frequent re-calibration of the electrode and
overcome the problems of close matching of ionic strengths and composi-
tions of standards and unknowns. However, the initial concentration of the
analyte must be known approximately so that the addition can be arranged
to give a two- to five-fold increase. Also the standard addition must not
TABLE 9-11
Characteristics of some specific ion electrodes (based on Orion Research literature)
change either the ionic strength (or else ion activities and junction potentials
would also change) or the proportion, if any, of complexed analyte ions. The
first requirement is met by addition of a small volume of a relatively con-
centrated solution of the analyte to a much larger, accurately known volume
of the unknown. Providing strong complexing agents are present in large
excess this also causes no significant change in the proportion of complexed
ions.
Calculation of results is straightforward. The original potential (Εχ) of the
system is given by:
Ε^Εχ+S log C0
iiC0V0 CsSVsS \
E22=EX+Sx log
8 —-5—^- + -
\V0+VS V0+Vj
However, if we make Vs negligible compared to V0 (V0/Vs> 100 : 1), by
adding a small volume of a relatively concentrated standard, the expression
simplifies t o :
E2=Ex+Slog (co + ~ )
AE = S log(C 0 + AC)
or:
A E = S l o g ( l + AC/C 0 )
single addition or, more reliably, by making several standard additions and
plotting (Z — l) versus AC to obtain an average slope. Incremental additions
to the reagent blank provide good estimates of the blank that can then be
used to obtain a linear, blank-corrected, calibration curve at lower concentra-
tions than would otherwise be possible (Smith and Manaham, 1973;
McQuaker and Gurney, 1977).
If we do not arrange for Vs to be negligible compared to V0, the calcula-
tions become relatively cumbersome and it is much simpler to plot results on
Gran's paper — a special volume corrected semi-antilog graph paper named
after one of the principal exponents of linear titrations (Orion Research,
1970). An example of a Gran's plot is shown in Fig. 9-4: the paper is availa-
ble from Orion Research, 380 Putnam Avenue, Cambridge, MA 02139 U.S.A.
Specific ion electrodes can also be used as endpoint detectors for indirect
determination of ions by titration. Many such determinations are feasible;
two of potential interest here are analysis of natural waters or leachates for
Cl" or SO*". Chloride can be determined directly with a Cl~-sensing elec-
trode, but at levels below 5 Mg/ml is best determined indirectly using a Ag/
AgS electrode to follow its titration with silver nitrate (Haynes and Clark,
1972; Orion Research, 1970). Sulphate is determined by titration of a
methanolic solution with lead perchlorate using a lead (Pb/PbS) electrode:
Fluoride
Boron
cr NaCl 1.6482
Cr K2Cr2Ch 2.828
Co CoCl2 · 6 H 2 0 4.037
Cu Cu 1.000 polished Cu wire: dissolve in 10 ml
CuS0 4 • 5 H 2 0 3.929 nitric acid
F" NaF 2.210
Fe Fe 1.000 pure Fe wire: dissolve in 5 ml nitric
acid
Au Au 1.000 dissolve in aqua-regia
Pb Pb(N0 3 ) 2 1.599
Li LiCl 6.109
Mg MgO 1.658 dissolve in 10 ml nitric acid
Mn MnS0 4 2.7486 dry at 180°C; dissolves in water
Hg HgCl2 1.354
Mo Na 2 Mo0 4 · 2 H 2 0 2.522
Ni NiS0 4 ■6 H 2 0 4.477
215
APPENDIX 2 (continued)
* Unless otherwise indicated the compound dissolves in water. For atomic absorption
spectrophotometry, standard solution are normally prepared in dilute (^10%) hydro-
chloric or nitric acid. Hydrochloric acid should not be used for stock solutions of ele-
ments, e.g. Ag and Pb, with insoluble chlorides.
REFERENCES
Abbey, S., 1970. Analysis of rocks and minerals by atomic absorption spectroscopy, 3.
A lithium-fluorborate scheme for seven major elements. Geol. Surv. Can., Paper, 70-
23: 20 pp.
Abbey, S., 1977. Studies in "standard samples" for use in general analysis of silicate rocks
and minerals, 5. 1977 edition of "usable" values. Geol. Surv. Can., Paper, 77-34:
31pp.
A.E.G. Analysis Committee, 1971. Committee Report. Assoc. Explor. Geochem. Newsl.,
July 1971: 3 - 5 .
Aggett, J. and Aspell, A.C., 1976. The determination of arsenic (III) and total arsenic by
atomic-absorption spectroscopy. Analyst, 100: 341—347.
A.G.R.G., 1962. Determination of sulphate in natural water. Appl. Geochem. Res. Group,
Tech. Comm., No. 27: 2 pp.
Agterdenbos, J. and Vlogtman, J., 1972. Determination of and differentiation between
cassiterite and silicate bound tin in silicate rocks containing traces of tin. Talanta, 19:
1295-1300.
Ahrens, L.H., 1965. Distribution of the Elements in Our Planet. McGraw-Hill, New York,
N.Y., 110 pp.
Ahrens, L.H. and Taylor, S.R., 1961. Spectrochemical Analysis. Addison-Wesley, London,
2nd ed., 454 pp.
Allan, J.E., 1961. The use of organic solvents in atomic absorption spectrophotometry.
Spectrochim. Ada, 17: 467—473.
Allan, R.J. and Hornbrook, E.H.W., 1971. Exploration geochemistry evaluation study in
a region of continuous permafrost, Northwest Territories, Canada. In: R.W. Boyle
(Editor), Geochemical Exploration. Can. Inst. Min. Metall, Spec. Vol., 11: 53—66.
Allcott, G.H. and Lakin, H.W., 1975. The homogeneity of six geochemical exploration
reference samples. In: I.L. Elliott and W.K. Fletcher (Editors), Geochemical Explora-
tion 1974. Else vier, Amsterdam, pp. 659—681.
Allcott, G.H. and Lakin, H.W., 1978. Tabulation of geochemical data furnished by 109
laboratories for six exploration reference samples. U.S. Geol. Surv., Open File Rep.,
78-163: 199 pp.
Alminas, H.V. and Mosier, EX., 1975. Oxalic-acid leaching of rock, soil and stream-sedi-
ment samples as an anomaly-accentuation technique. U.S. Geol. Surv., Open File Rep.,
76-275: 25 pp.
Almond, H., 1953a. Field method for the determination of traces of arsenic in soils. Anal.
Chem., 25: 1 7 6 6 - 1 7 6 7 .
Almond, H., 1953b. Determination of traces of cobalt in soils. Anal. Chem., 25: 166—
167.
Almond, H., 1953c. A field method for the determination of manganese in soils adapted
from methods described in the literature. U.S. Geol. Surv., Open File Rep.
218
Almond, H., 1955. Rapid field and laboratory method for the determination of copper in
soil and rock. U.S. Geol. Surv., Bull, 1036-A: 1—8.
Almond, H., Crowe, H.E. and Thompson, C.E., 1955. Rapid determination of germanium
in coal, soil and rock. U.S. Geol. Surv., Bull., 1036-B: 9—17.
Aly, M.M., Wassef, S.N. and Hathout, M.H., 1977. The colorimetric estimation and the
distribution of thorium in Egyptian beach sands. Chem. Erde, 36: 336—342.
Analytical Methods Committee, 1959. Notes on perchloric acid and its handling in ana-
lytical work. Analyst, 84: 214—216.
Anderman, G. and Kemp, J.W., 1958. Scattered X-rays as internal standards in X-ray
emission spectroscopy. Anal. Chem., 30: 1306—1309.
Anderson, J.U., 1963. An improved pretreatment for mineralogical analysis of samples
containing organic matter. Clays Clay Miner., 10: 380—388.
Anderson, J.U. and O'Connor, G.A., 1972. Production of permanganate ion by sodium
hypochlorite treatment to remove soil organic matter. Soil Sei. Soc. Am., Proc., 36:
973-975.
Aslin, G.E.M., 1976. The determination of arsenic and antimony in geological materials
by flameless atomic absorption spectrometry. J. Geochem. Explor., 6: 321—330.
Avni, R., Harel, A. and Brenner, I.B., 1972. A new approach to the spectrochemical anal-
ysis of silicate rocks and minerals. Appl. Spectrosc., 26: 641—645.
Azzaria, L.M. and Webber, G.R., 1969. Mercury analysis in geochemical exploration. Can.
Inst. Min. Metall, Bull, 62: 5 2 1 - 5 3 0 .
Baker, W.E., 1965. Rapid colorimetric procedures in geochemical prospecting for molyb-
denum. Proc. Australas. Inst. Min. Metall, 214: 125—134.
Ball, T.K., Booth, S.J., Nickless, E.F.P. and Smith, R.T., 1979. Geochemical prospecting
for baryte and celestite using a portable radioisotope fluorescence analyser. J. Geo-
chem. Explor., 1 1 : 277—284.
Band, R.B. and Wilkinson, N.M., 1972. Interferences in the determination of mercury in
mineralized samples by the wet reduction—flameless atomic absorption method. J.
Geochem. Explor., 1: 195—198.
Barakso, J.J., 1967. Geochemical field kit for the determination of trace amounts of
molybdenum. Econ. Geol, 62: 732—736 (Discussions, 63: 85, 572, 695—696).
Barakso, J.J. and Tornocai, C , 1970. A mercury determination method and its use for
exploration in British Columbia. Can. Inst. Min. Metall, Bull, 63: 501—505.
Barnard, W.M. and Fishman, M.J., 1973. Evaluation of the use of the heated graphite
atomizer for the routine determination of trace metals in water. At. Absorpt. NewsL,
12: 1 1 8 - 1 2 4 .
Barringer, A.R., 1966. Interference-free spectrometer for high-sensitivity mercury analy-
ses of soils, rocks and air. Inst. Min. Metall, Trans., Sect. B, 75: 120—124.
Belcher, R., Bogdanski, S.L., Henden, E. and Townshend, A., 1975. Elimination of inter-
ferences in the determination of arsenic and antimony by hydride generation using
molecular emission cavity analysis (MECA). Analyst, 100: 522—523.
Berenice, G., Da Silva, C.T.C.B. and Mello, V.N.A., 1975. A method of interpreting
results obtained by a six-step semiquantitative spectrographic procedure. App. Spec-
trosc, 29: 269—271.
Bergseth, H., 1975. Determination of zinc in soils by X-ray fluorescence spectrometry
involving a modified background ratio method. Analyst, 100: 96—98.
Berman, S.S. and McLaren, J.W., 1978a. Establishment of compromise conditions for
multielement analysis by inductively coupled plasma emission spectrometry: a prelimi-
nary report. Appl Spectrosc, 32: 372—377.
Berman, S.S. and McLaren, J.W., 1978b. Application of the inductively coupled plasma
to geochemical analysis. Paper presented at Geoanalysis 78, Ottawa, Ont.
219
Billings, G.K., 1965. Light scattering in trace-element analysis by atomic absorption. At.
Absorpt. NewsL, 4 : 357—361.
Blasius, M.B., Kerkhoff, S.J., Wright, R.S. and Cothern, C.R., 1972. Use of X-ray fluores-
cence to determine trace metals in water resources. Water Resour. Bull, 8: 704—714.
Bloom, H., 1955. A field method for the determination of ammonium citrate-soluble
heavy metals in soils and alluvium. Econ. GeoL, 55: 533—541.
Bloom, H., 1962. Field methods for the determination of nickel using dimethylglyoxime.
Econ. GeoL, 57: 5 9 5 - 6 0 4 .
Bloom, H., 1966. A field method for the determination of silver in soils and rocks using
dithizone. Econ. GeoL, 6 1 : 189—197.
Bodkin, J.B., 1977. Determination of fluorine in silicates by use of an ion-selective elec-
trode following fusion with lithium metaborate. Analyst, 102: 409—413.
Bölviken, B., Logn, O. and Uddu, O., 1973. Instrument for in situ measurements of pH,
Eh and self-potential in diamond drill holes. In: M.J. Jones (Editor), Geochemical
Exploration 1972. Institution of Mining and Metallurgy, London, pp. 417—422.
Bond, A.M., 1970. Use of ammonium fluoride in determination of zirconium and other
elements by atomic absorption spectrometry in the nitrous oxide-acetylene flame.
Anal. Chem., 42: 932—935.
Boumans, P.W.J.M., 1966. Theory of Spectrochemical Excitation. Hilger and Watts,
London, 383 pp.
Boumans, P.W.J.M., 1979. Inductively coupled plasma-atomic emission spectroscopy: its
present and future position in analytical chemistry. ICP Inf. NewsL, 5: 181—209.
Bowden, P., 1964. A geochemical field method for determining tungsten in soils and
stream sediments. Analyst, 89: 771—774.
Bowie, S.H.U., 1968. Portable X-ray fluorescence analysers in the mining industry. Min.
Mag., 118: 2 3 0 - 2 3 9 .
Bowie, S.H.U., Darnley, A.G. and Rhodes, J.R., 1965. Portable radioisotope X-ray
fluorescence analyser. Inst. Min. Metall., Trans., Sect. B, 74: 361—379.
Boyle, R.W., 1976. Report of retiring president R.W. Boyle to Annual Meeting of the
Association of Exploration Geochemists, Fredericton, N.B., April 23, 1976. J. Geo-
chem. Explor., 6: 389—395.
Brabec, D., 1971. Aqua regia extractable vs. total copper and zinc content of granitic
rocks. Trans. Soc. Min. Eng. AIME, 250: 94—97.
Bradshaw, P.M.D., Thomson, I., Smee, B.W. and Larsson, J.O., 1974. The application of
different analytical extractions and soil profile sampling in exploration geochemistry.
J. Geochem. Explor., 3: 209—225.
Bratzel, M.P., Chakrabarti, C.L., Sturgeon, R.E., Mclntyre, M.W. and Agemian, H., 1972.
Determination of gold and silver at parts-per-billion or lower levels in geological and
metallurgical samples by atomic absorption spectrometry with a carbon rod atomizer.
Anal. Chem., 44: 3 7 2 - 3 7 4 .
Brobst, D.A. and Ward, F.N., 1965. A turbidimetric test for barium and its geologic
application in Arkansas. Econ. GeoL, 60: 1020—1040.
Brown, E., Skougstad, M.W. and Fishman, M.J., 1970. Methods for Collection and Anal-
ysis of Water Samples for Dissolved Minerals and Gases. Techniques of Water-Resources
Investigations of the U.S. Geological Survey, Book 5. U.S. Geological Survey, Washing-
ton, D.C., Chapter A l , 160 pp.
Brown, G.C., Hughes, D.J. and Esson, J., 1973. New X.R.F. data retrieval techniques and
the application to U.S.G.S. standard rocks. Chem. GeoL, 1 1 : 223—229.
Brundin, N.H. and Bergström, J., 1977. Regional prospecting for ores based on heavy
minerals in glacial till. J. Geochem. Explor., 7: 1—19.
Brundin, N.H. and Nairis, B., 1972. Alternative sample types in regional geochemical
prospecting. J. Geochem. Explor., 1: 7—46.
220
Chao, T.T., Ball, J.W. and Nakagawa, H.M., 1971. Determination of silver in soils, sedi-
ments and rocks by organic chelate extraction and atomic absorption spectrophotom-
etry. Anal. Chim. Acta, 54: 77—81.
Chao, T.T., Jenne, E.A. and Heppting, L.M., 1968. Prevention of adsorption of trace
amounts of gold by containers. U.S. Geol. Suru., Prof. Paper, 600-D: 16—19.
Chao, T.T., Sanzolone, R.F. and Hubert, A.E., 1978. Flame and flameless atomic-absorp-
tion determination of tellurium in geological materials. Anal. Chim. Acta, 96: 251—
257.
Chork, C.Y., 1977. Seasonal, sampling and analytical variations in stream sediment
surveys. J. Geochem. Explor., 7: 31—48.
Chaudury, A.N. and Böse, B.B., 1971. Role of "humus matter" in the formation of geo-
chemical anomalies. In: R.W. Boyle (Editor), Geochemical Exploration. Can. Inst.
Min. Metall, Spec. Vol., 11: 410—415.
Chu, R.C., Barron, G.P. and Baumgarner, A.W., 1972. Arsenic determination at sub-
microgram levels by arsine evolution and flameless atomic absorption spectrophoto-
metric technique. Anal. Chem., 44: 1476—147$.
Cioni, R., Mazzucotelli, A. and Ottonello, G., 1976a. Interference effects in the deter-
mination of barium in silicates by flame atomic-absorption spectrophotometry.
Analyst, 101: 9 5 6 - 9 6 0 .
Cioni, R., Mazzucotelli, A. and Ottonello, G., 1976b. Matrix effects in the flameless
atomic absorption determination of trace amounts of barium in silicates. Anal. Chim.
Acta, 82: 4 1 5 - 4 2 0 .
Clark, L.J. and Axley, J.H., 1955. Molybdenum determination in soils and rocks with
dithiol. Anal. Chem., 27: 2000—2003.
Clark, N.H. and Pyke, J.G., 1972. The determination of uranium in exploration samples
by X-ray emission spectrometry. Anal. Chim. Acta, 58: 234—237.
Clayton, C G . and Packer, T.W., 1980. Some applications of energy dispersive X-ray
fluorescence analysis in mineral exploration, mining and process control. In: C.S.
Barrett, D.E. Leyden, J.B. Newkirk, P.K. Predecti and C O . Ruud (Editors), Advances
in X-ray Analysis, Vol. 23. Plenum, New York, N.Y., pp. 1—13.
Clifton, H.E., Hunter, R.E., Swanson, F.J. and Phillips, R.L., 1969. Sample size and
meaningful gold analysis. U.S. Geol Suru., Prof. Paper, 625-C: 17 pp.
Cogger, N., 1974. An absorptiometric method for the determination of small amounts of
tin in rocks and ores with toluene-3, 4-dithiol. Inst. Geol. Sei. Inf. Ser., Rep., 1 3 : 8 pp.
Cogger, N., 1976. An extraction-spectrophotometric method for the determination of W
in geological materials. Anal. Chim. Acta, 84: 143—148.
Corbett, J.A. and Godbeer, W.C., 1977. The determination of tellurium in weathered
outcrop, mineralized and barren rock. Anal. Chim. Acta, 9 1 : 211—219.
Craven, C.A.U., 1954. Statistical estimation of the accuracy of assaying. Inst. Min. Metall,
Trans., Sect. B, 63: 5 5 1 - 5 6 3 .
Crenshaw, G.L. and Ward, F.N., 1975. Determination of fluorine in soils and rocks by
known-increment addition and selective-ion electrode detection. In: F.N. Ward (Editor),
New and Refined Methods of Trace Analysis Useful in Geochemical Exploration. U.S.
Geol. Suru., Bull, 1408: 7 7 - 8 4 .
Cruft, E.F. and Giles, D.L., 1967. Direct reading emission spectrometry as a geochemical
tool. Econ. Geol, 62: 406—411.
Cruz, R.B. and Van Loon, J . C , 1974. A critical study of the application of graphite-fur-
nace non-flame atomic absorption spectrometry to the determination of trace base
metals in complex heavy-matrix sample solutions. Anal. Chim. Acta, 72: 231—243.
Czamanske, G.K. and Ingamells, C O . , 1970. Selective chemical dissolution of sulphide
minerals: a method of mineral separation. Am. Mineral, 55: 2131—2134.
222
Dahlquist, R.L. and Knoll, J.L., 1978. Inductively coupled plasma-emission spectrometry:
analysis of biological materials and soils for major, trace and ultra-trace elements.
Appl. Spectrosc, 32: 1—29.
Dalton, E.F. and Malanoski, A.J., 1971. Note on the determination of arsenic by atomic
absorption by arsine into an argon-hydrogen entrained air flame. At. Absorpt Newsl,
10: 9 2 - 9 3 .
Danielsson, A., 1967. Spectrochemical analysis for geochemical prospecting. In: A.H.
Gillieson (Editor), XIII Colloquium Spectroscopium Internationale. Adam Hilger,
London, pp. 311—323.
Danielsson, A. and Sundkvist, G., 1959. The tape machine, II. Applications using differ-
ent kinds of isoformations. Spectrochim. Ada, 15: 126—133.
Danielsson, A., Lundgren, F. and Sundkvist, G., 1959. The tape machine, I. A new tool
for spectrochemical analysis. Spectrochim. Acta, 15: 122—125.
Davenport, P.H. and Nichol, I., 1973. Bedrock geochemistry as a guide to areas of base-
metal potential in volcano-sedimentary belts of the Canadian Shield. In: M.J.. Jones
(Editor), Geochemical Exploration 1972. Institution of Mining and Metallurgy, London,
pp. 45—57.
Davies, B.E., 1974. Loss-on-ignition as an estimate of soil organic matter. Soil Sei. Soc.
Am.,Proc, 38: 150—151.
Davies, R.L, Cheshire, M.V. and Graham-Bryce, I.J., 1969. Retention of low levels of
copper by humic acid. J. Soil Sei., 20: 65—71.
Davis, C.E.S., 1972. Analytical methods used in the study of an ore intersection from
Lunnon Shoot, Kambalda. Econ. Geol., 67: 1091—1092.
Deb, B.C., 1950. The estimation of free iron oxides in soils and clays and their removal.
J. Soil Sei., 1: 2 1 2 - 2 2 0 .
Delavault, R., 1977. Simplified field test for copper. J. Geochem. Explor., 8: 537—540.
Delavault, R.E. and Marshall, D.B., 1973. Detection of molybdenum in rocks using
Tennant's bead-forming buffer. Can. J. Spectrosc, 18: 10—12.
De Pablo Galan, L., 1975. Direct-reading emission spectroscopy analysis of geochemical
samples. In: I.L. Elliott and W.K. Fletcher (Editors), Geochemical Exploration 1974.
Elsevier, Amsterdam, pp. 707—720.
Dijkstra, S., Van den Hul, H.J. and Bill, E., 1979. Experiments on the usefulness of some
selected chemical quantities in geochemical exploration in a former mining district. In:
J.R. Watterson and P.K. Theobald (Editors), Geochemical Exploration 1978. Associa-
tion of Exploration Geochemists, Rexdale, Ont., pp. 283—288.
Dolezal, J., Povondra, P. and Sulcek, Z., 1968. Decomposition Techniques in Inorganic
Analysis (English Edition). Iliffe Books, London, 224 pp.
Dorrzapf, A.F., 1973. Spectrochemical analysis — argon-oxygen DC arc method for
silicate rocks. U.S. Geol. Surv., J. Res., 1: 559—562.
Dyck, W., 1971. The adsorption and coprecipitation of silver on hydrous oxide of iron
and manganese. Geol. Surv. Can., Paper, 70-64: 23 pp.
Ediger, R.D., 1973. A review of water analysis by atomic absorption. At. Absorpt. Newsl,
12: 1 5 1 - 1 5 7 .
Ediger, R.D., 1975. Atomic absorption analysis with the graphite furnace using matrix
modification. At. Absorpt. Newsl, 14: 127—130.
Ediger, R.D. and Wilson, D.L., 1979. The performance of an inductively coupled plasma
on the model 5000 atomic absorption spectrophotometer. At. Absorpt. Newsl, 18:
41-45.
Edmond, C.R., 1969. Direct determination of fluoride in phosphate rock samples using
the specific ion electrode. Anal Chem., 41: 1327—1328.
Edmunds, W.M., Giddings, D.R. and Morgan-Jones, M., 1973. The application of flame-
less atomic absorption in hydrogeochemical analysis. At. Absorpt. Newsl, 12: 45—49.
223
El-Kholy, H.K., Burridge, J.C. and Scott, R.O., 1975. A tripleflow gas-sheathed DC arc
for spectrochemical analysis. Anal. Chim. Ada, 74: 247—252.
Ellis, A.J., Tooms,· J.S., Webb, J.S. and Bicknell, J.V., 1967. Application of solution
experiments in geochemical prospecting. Inst. Min. Metall., Trans., Sect. B, 76: 25—39.
(Discussion, 76: 216—217; and 77 : 136.)
Fabbi, B.P. and Moore, W.J., 1970. Rapid X-ray fluorescence determination of sulfur in
mineralized rocks from the Bingham Mining District, Utah. Appl. Spectrosc, 24: 426—
428.
Farmer, V.C. and Mitchell, B.D., 1963. Occurrence of oxalates in soil clays following
hydrogen peroxide treatment. Soil Sei., 96: 221—229.
Farrell, B.L., 1974. Fluorine, a direct indicator of fluorite mineralization in local and
regional geochemical surveys. J. Geochem. Explor., 3: 227—244.
Fassel, V.A., 1978. Quantitative elemental analyses by plasma emission spectroscopy.
Science, 202: 183—191.
Fassel, V.A. and Kniseley, R.N., 1974a. Inductively coupled plasma-optical emission
spectroscopy. Anal. Chem., 46: 1110A—1120A.
Fassel, V.A. and Kniseley, R.N., 1974b. Inductively coupled plasmas. Anal. Chem., 46:
1155A—1164A.
Faye, G.H., 1975. Description of ultramafic rock (ore) samples UM-1, UM-2, UM-4: a
correction. Chem. Geol, 15: 235—237.
Feather, C.E. and Willis, 1976. A simple method for background and matrix correction of
spectral peaks in trace element determination by X-ray fluorescence spectrometry.
X-Ray Spectrom., 5: 41—48.
Feldman, C , 1974. Preservation of dilute mercury solutions. Anal. Chem., 46: 99—102.
Fernandez, F.J., 1973. Atomic absorption determination of gaseous hydrides using
sodium borohydride reduction. At. Absorpt. Newsl., 12: 93—97.
Fernandez, F.J. and Manning, D.C., 1971. The determination of arsenic at sub-microgram
levels by atomic absorption spectrophotometry. At. Absorpt. Newsl., 10: 86—88.
Ficklin, W.H., 1970. A rapid method for the determination of fluoride in rocks and soils,
using an ion-selective electrode. U.S. Geol. Surv., Prof. Paper, 700-C: 186—188.
Ficklin, W.H., 1975. Ion-selective electrode determination of iodine in rocks and soils.
U.S. Geol. Surv., J. Res., 3: 753—755.
Ficklin, W.H. and Ward, F.N., 1976. Flameless atomic absorption determination of
bismuth in soils and rocks. U.S. Geol. Surv., J. Res., 4: 217—220.
Flanagan, F.J., 1969. U.S. Geological Survey standards, II. First compilation of data for
the new USGS rocks. Geochim. Cosmochim. Acta, 33: 81—120.
Fletcher, K., 1970. Some applications of background correction to trace metal analysis of
geochemical samples by atomic-absorption spectrophotometry. Econ. Geol., 65: 588—
589.
Fletcher, M.H., 1954. Collected papers on methods of analysis for uranium, 8. A study of
critical factors in the "direct" fluorimetric determination of uranium. U.S. Geol. Surv.,
Bull., 1006: 5 1 - 6 8 .
Floyd, M.A., Fassel, V.A. and D'Silva, A.P., 1980. Computer-controlled scanning meno-
chromator for the determination of 50 elements in geochemical and environmental
samples by inductively coupled plasma-atomic emission spectrometry. Anal. Chem.,
52:2168-2172.
Fordham, A.W., 1978. Correction of matrix effects in the analysis of stream waters by
flameless atomic absorption spectrophotometry. J. Geochem. Explor., 10: 41—51.
Foster, J.R., 1970. The application of direct-reading spectroscopy to the analysis of trace
metals in soils and stream sediments. Proc. Br. Ceram. Soc, 16: 25—31.
Foster, J.R., 1971. The reduction of matrix effects in atomic absorption analysis and the
efficiency of selected extractions in rock-forming minerals. In: R.W. Boyle (Editor),
Geochemical Exploration. Can. Inst. Min. Metall, Spec. Vol., 11: 554—560.
224
Foster, J.R., 1973. The efficiency of various digestion procedures on the extraction of
metals from rocks and rock-forming minerals. Can. Inst. Min. Metall., Bull., 66: 85—
92.
Frant, M.S. and Ross, J.W., Jr., 1968. Use of a total ionic strength adjustment buffer for
electrode determination of fluoride in water supplies. Anal. Chem., 40: 1169—1171.
Fratta, M., 1974. AAS determination of ppb amounts of thallium in silicate rocks. Can.
J. Spectrosc, 19: 33—37.
Friedrich, G.H., Kunzendorf, H. and Plüger, W.L., 1974. Shipborne geochemical investiga-
tions of deep-sea manganese-nodule deposits in the Pacific using a radio-isotope energy
dispersive X-ray system. J. Geochem. Explor., 3: 303—317.
Friedrich, G.H., Plüger, W.L., Hilmer, E.F. and Abu-Abed, I., 1973. Flameless atomic
absorption and ion-sensitive electrodes as analytical tools in copper exploration. In:
M.J. Jones (Editor), Geochemical Exploration 1972. Institution of Mining and
Metallurgy, London, pp. 437—445.
Fryer, B.J. and Kerrich, R., 1978. Determination of precious metals at ppb levels in rocks
by combined wet chemical and flameless atomic absorption method. At. Absorpt.
Newsl, 17: 4 - 6 .
Fuller, C.W., 1977. Electrothermal Atomization for Atomic Absorption Spectrometry.
Analytical Sciences Monograph No. 4. Chemical Society, London, 127 pp.
Gallagher, M.J., 1967. Determination of molybdenum, iron, and titanium in ores and
rocks by portable radioisotope X-ray fluoresence analyser. Inst. Min. Metall., Trans.,
Sect.B, 76: 1 5 5 - 1 6 4 .
Gallagher, M.J., 1969. Portable X-ray spectrometers for rapid ore analysis. In: Ninth
Commonwealth Mining and Metallurgical Congress. Institution of Mining and Metal-
lurgy, London, pp. 1—39.
Garreis, R.M. and Christ, C.L., 1965. Solutions, Minerals and Equilibria. Harper and Row,
New York, N.Y., 450 pp.
Garrett, R.G., 1969. The determination of sampling and analyical errors in exploration
geochemistry. Econ. Geoi, 64: 568—569.
Garrett, R.G., 1975. Copper and zinc in Proterozoic acid volcanic rocks as a guide to
exploration in the Bear Province. In: I.L. Elliott and W.K. Fletcher (Editors), Geo-
chemical Exploration 1974. Elsevier, Amsterdam, pp. 371—388.
Garrett, R.G. and Goss, T.I., 1979. The evaluation of sampling and analytical variation in
regional geochemical surveys. In: P.J. Watterson and P.K. Theobald (Editors), Geo-
chemical Exploration 1978. Association of Exploration Geochemists, Rexdale, Ont.,
pp. 3 7 1 - 3 8 3 .
Garrett, R.G. and Hornbrook, E.H.W., 1976. The relationship between zinc and organic
content in centre-lake bottom sediments·. J. Geochem. Explor., 5: 31—38.
Garrett, R.G. and Lynch, J.J., 1976. A comparison of neutron activation delayed neutron
counting versus fluorimetric analysis in large scale geochemical exploration for urani-
um. In: Exploration for Uranium Deposits. Proceedings, Symposium on Exploration
of Uranium Ore Deposits Vienna, 1976. International Atomic Energy Agency, Vienna,
pp. 321—334.
Garson, M.S. and Bateson, J.H., 1967. Possible use of the P.I.F. analyser in geochemical
prospecting for tin. Inst. Min. Metall., Trans., Sect. B, 76: 165—166.
Gatehouse, S., Russell, D.W. and Van Moort, J.C., 1977. Sequential soil analysis in
exploration geochemistry. J. Geochem. Explor., 8: 483—494.
Gedeon, A.Z., Butt, C.R.M., Gardner, K.A. and Hart, M.K., 1977. The applicability of
some geochemical analytical techniques in determining "total" compositions of some
lateritized rocks. J. Geochem. Explor., 8: 283—303.
Giauque, R.D., Garrett, R.B. and Goda, L.Y., 1977. Energy dispersive X-ray fluorescence
spectrometry for determination of twenty-six trace and two major elements in geo-
chemical specimens. Anal. Chem., 48: 62—67.
225
Goertzen, J.O. and Oster, J.D., 1972. Potentiometric titration of sulphate in water and
soil extracts using a lead electrode. Soil Sei. Soc. Am., Proc, 34: 691—693.
Golterman, H.L., 1970. Methods for Chemical Analysis of Fresh Waters. IBP Handbook
No. 8. Blackwell, Oxford, 166 pp.
Gong, H. and Suhr, M.H., 1976. The determination of cadmium in geological materials
by flameless atomic absorption. Anal. Chim. Acta, 8 1 : 297—303.
Goodfellow, W.D. and Wahl, J.L., 1976. Water extracts of volcanic rocks — detection of
anomalous halos at Brunswick No. 12 and Heath Steele B-zone massive sulphide depos-
its. J. Geochem. Explor., 6: 35—58.
Goodman, R.J., 1973. Rapid analysis of trace amounts of tin in stream sediments, soils
and rocks by X-ray fluorescence analysis. Econ. Geol, 68: 275—278.
Gorsuch, T.T., 1970. The Destruction of Organic Matter. Pergamon, Oxford, 152 pp.
Govett, G.J.S. and Whitehead, R.E., 1973. Errors in atomic absorption spectrophoto-
metric determination of Pb, Zn, Ni and Co in geologic materials. J. Geochem. Explor.,
2: 1 2 1 - 1 3 2 .
Green, T.E., Campbell, W.J. and Law, S.L., 1971. Application of a combined ion
exchange paper X-ray spectrographic method to geochemical exploration. In: R.W.
Boyle (Editor), Geochemical Exploration. Can. Inst. Min. Metall, Spec. Vol., 11: 516.
Greenfield, S., Jones, I.LI. and Berry, C.T., 1964. High-pressure plasmas as spectroscopic
emission sources. Analyst, 89: 713—720.
Greenfield, S., McGeachin, H.McD. and Smith, P.B., 1975a. Plasma emission sources in
analytical spectroscopy, I. Talanta, 22: 1—15.
Greenfield, S., McGeachin, H.McD. and Smith, P.B., 1975b. Plasma emission sources in
analytical spectroscopy, III. Talanta, 23: 1—14.
Greenfield, S., Jones, I.LI., McGeachin, H.McD. and Smith, P.B., 1975c. Automatic multi-
sample simultaneous multi-element analysis with a H.F. plasma torch and direct
reading spectrometer. Anal. Chim. Acta, 74: 225—245.
Greenland, L.P. and Campbell, E.Y., 1976. Rapid determination of nanogram amounts of
tellurium in silicate rocks. Anal. Chim. Acta, 87: 323—328.
Gregory, P. and Tooms, J.S., 1968. Geochemical prospecting for kimberlites. Q. Colo.
ach. Mines, 64: 2 6 5 - 3 0 5 .
Grifitts, W.R., Ward, F.N. and Alminas, H.V., 1976. A simple spot test for molybdenum
minerals. Econ. Geol., 7 1 : 1595.
Grimaldi, F.S. and Schnepfe, M.M., 1971. Determination of iodine in the p.p.m. range in
rocks. Anal. Chim. Acta, 53: 181—184.
Grimaldi, F.S., Ward, F.N. and Fuyat, R.K., 1954. Collected papers on methods of analy-
sis for uranium. U.S. Geol. Surv., Bull., 1006: 184 pp.
Grout, A. and Gallagher, M.J., 1980. Barium determination in rock and overburden by
portable XRF spectrometer. Inst. Min. Metall, Trans., Sect. B, 89: 130—133.
Gunton, J.E. and Nichol, I., 1975. Chemical zoning associated with the Ingerbelle—Copper
Mountain mineralization, Princeton, British Columbia. In: I.L. Elliott and W.K.
Fletcher (Editors), Geochemical Exploration 1974. Elsevier, Amsterdam, pp. 297—
312.
Hadeishi, T. and McLaughlin, R.D., 1966. Hyperfine Zeeman effect atomic absorption
spectrometer for mercury. Science, 174: 404—407.
Hall, G.E.M., 1979. A study of the stability of uranium in waters collected from various
geological environments in Canada. In: Current Research, Part A. Can. Geol. Surv.,
Paper, 79-1A: 361—365.
Hannaker, P. and Hughes, T.C., 1977. Multielement trace analysis of geological materials
with solvent extraction and flame atomic absorption spectrometry. Anal. Chem., 49:
1485-1488.
226
Hannaker, P. and Hughes, T.C., 1978. Improved solvent extraction scheme for geochemi-
cal trace analysis using flame atomic absorption. J. Geochem. Explor., 10: 169—180.
Hansel, J.M. and Martell, L.J., 1977. Automated energy dispersive X-ray determination of
trace elements in stream sediments. Los Alamos Sei. Lab., Informal Rep., LA-6869-
MS: 8 pp.
Hansuld, J.A., Mannard, G.W., Lakin, H.W., Canney, F.C., Salmon, M.L. and Weber, G.R.,
1969. What is a geochemical analysis? — a panel discussion. Q. Colo. Sch. Mines, 64:
5-26.
Harden, G. and Tooms, J.S., 1964. Efficiency of the potassium bisulphate fusion in geo-
chemical analysis. Inst. Min. Metall., Trans., Sect. B, 74: 129—141.
Hatch, W.R. and Ott, W.L., 1968. Determination of sub-microgram quantities of mercury
by atomic absorption spectrophotometry. Anal. Chem., 40: 2085—2087.
Hausen, D.M., Ahlrichs, J.W. and Odekirk, J.R., 1973. Application of sulphur and nickel
analyses to geochemical prospecting. In: J. Jones (Editor), Geochemical Exploration
1972. Institution of Mining and Metallurgy, London, pp. 13—24.
Hawkes, H.E., 1963. Dithizone field tests. Econ. Geol., 58: 579—586.
Hawkes, H.E. and Webb, J.S., 1962. Geochemistry in Mineral Exploration. Harper and
Row, New York, N.Y., 415 pp.
Haynes, S.J. and Clark, A.H., 1972. A rapid method for the determination of chlorine in
silicate rocks using ion-selective electrodes. Econ. Geol., 67: 378—382.
Head, P.C. and Nicholson, R.A., 1973. A cold vapour technique for the determination of
mercury in geological materials involving its reduction with tin (II) chloride and collec-
tion on gold wire. Analyst, 98: 53—56.
Heffernan, B.J., Archibold, R.O. and Vickers, T.J., 1967. Determination of tin in
cassiterite ores by atomic absorption spectrophotometry. Proc. Australas. Inst. Min.
Metall, 223: 6 5 - 6 9 .
Heiz, A.W., 1964. A gas jet for DC arc spectroscopy. U.S. Geol. Surv., Prof. Paper, 475D:
176-178.
Heiz, A.W., 1965. The problem of automatic plate reading and computer interpretation
for spectrochemical analysis. U.S. Geol. Surv., Prof. Paper, 525-B: 160—164.
Heiz, A.W., 1973. Spectrochemical computer analysis-instrumentation. U.S. Geol. Surv.,
J. Res., 1: 475—482. j
Herbert, A.J. and Young, G.G., 1977. Portable field water sample filtration unit. Anal.
Chem., 49: 672—673.
Hill, W.E. Jr., 1975. The use of analytical standards to control assaying projects. In: I.L.
Elliott and W.K. Fletcher (Editors), Geochemical Exploration 1974. Elsevier, Amster-
dam, pp. 651—657.
Hoffman, S.J., 1974. Pebble cards —a record of the coarse fraction of stream sediments
for geochemical exploration. J. Geochem. Explor., 3: 387—388.
Hoffman, S.J. and Fletcher, W.K., 1979. Extraction of Cu, Zn, Mo, Fe and Mn from soils
and sediments using a sequential procedure. In: J.R. Watterson and P.K. Theobald
(Editors), Geochemical Exploration 1978. Association of Exploration Geochemists,
Rexdale, Ont., pp. 289—299.
Hoffman, S.J. and Fletcher, W.K., 1980. Organic matter scavenging of copper, zinc,
molybdenum, iron and manganese, estimated by sodium hypochlorite extraction
(pH 9.5). Abstr., 8th Int. Geochem. Explor. Symp., Hanover, 1980.
Hoffman, S.J. and Fletcher, W.K., 1981. Detailed lake sediment geochemistry of anoma-
lous lakes on the Nechako Plateau, central British Columbia: comparison of trace
metal distributions in Capoose and Fish Lakes. J. Geochem. Explor., 14: 221—244.
Hoffman, S.J. and Waskett-Myers, M.J., 1974. Determination of molybdenum in soils and
sediments with a modified zinc dithiol procedure. J. Geochem. Explor., 3: 61—66.
Holman, R.H.C., 1956-57. A method for determining readily-soluble copper in soil and
alluvium — introducing white spirit for dithizone. Inst. Min. Metall., Trans., Sect. B,
66: 7 - 1 6 .
227
Holman, R.H.C., 1963. Field and laboratory methods used by the Geological Survey of
Canada in geochemical surveys, 2. A method for determining readily-soluble copper in
soil and alluvium. Geol. Surv. Can., Paper, 63-67: 5 pp.
Holman, R.H.C. and Durham, C.C., 1967. A mobile spectrographic laboratory. Geol.
Surv. Can., Paper, 66-35: 15 pp.
Hopkins, D.M., 1977. An improved ion-selective electrode method for the rapid deter-
mination of fluorine in rocks and soils. U.S. Geol. Surv., J. Res., 5: 589—593.
Horsnail, R.F., Nichol, I. and Webb, J.S., 1969. Influence of variations in secondary
environment on the metal content of drainage sediments. Q. Colo. Sch. Mines, 64:
307-322.
Horton, R. and Lynch, J.J., 1975. A geochemical field laboratory for the determination
of some trace elements in soil and water samples. Geol. Surv. Can., Paper, 75-1, Part A:
213-214.
Howarth, R.J., 1977. Automated generation of randomised sample submittal schemes for
laboratory analysis. Comput. Geosci., 3: 327—334.
Howarth, R.J. and Lowenstein, P.L., 1971. Sampling variability of stream sediments in
broad-scale regional geochemical reconnaissance. Inst. Min. Metall., Trans., Sect. B, 80:
363-372.
Howarth, R.J. and Martin, L., 1979. Computer-based techniques in the compilation,
mapping and interpretation of exploration geochemical data. In: P.J. Hood (Editor),
Geophysics and Geochemistry in the Search for Metallic Ores. Geol. Surv. Can., Econ.
Geol Rep., 3 1 : 545—574.
Howarth, R.J. and Thompson, M., 1976. Duplicate analysis in geochemical practice, II.
Examination of proposed method and examples of its use. Analyst, 101: 699—709.
Hower, J., 1959. Matrix corrections in the X-ray spectrographic trace element analysis of
rocks and minerals. Am. Mineral, 44: 19—32.
Hubert, A.E. and Lakin, H.W., 1973. Atomic absorption determination of thallium and
indium in geologic materials. In: M.J. Jones (Editor), Geochemical Exploration 1972.
Institution of Mining and Metallurgy, London, pp. 383—387.
Huffman, C , Rahill, R.L., Shaw, V.E. and Norton, D.R., 1972. Determination of mer-
cury in geologic materials by flameless atomic absorption spectrometry. U.S. Geol
Surv., Prof. Paper, 800-C: 203—207.
Hulanicki, A., Lewandowski, R. and Lewenstam, A., 1976. Elimination of ionic inter-
ferences in the determination of sulphates in water using the lead-sensitive ion-selec-
tive electrode. Analyst, 101: 939—942.
Hulanicki, A., Trojanowicz, M. and Krawczyk, T.K.V., 1977. Determination of copper in
water by means of chalcocite copper ion-selective electrode. Water Res., 1 1 : 627—630.
Hunt, E.C., North, A.A. and Wells, R.A., 1955. Application of paper Chromatographie
methods of analysis to geochemical prospecting. Analyst, 80: 172—194.
Hunt, E.C., Stanton, R.E. and Wells, R.A., 1959-60. Field determination of beryllium in
soils for geochemical prospecting. Inst. Min. Metall, Trans., Sect. B, 69: 361—369.
Husler, J., 1972. Determination of niobium in rocks, ores and alloys by atomic-absorp-
tion spectrophotometry. Talanta, 19: 863—869.
Hutchison, C.S., 1974. Laboratory Handbook of Petrographic Techniques. Wiley-Inter-
science, New York, N.Y., 527 pp.
Hutton, R.C., Ottaway, J.M., Rains, T.C. and Epstein, M.S., 1977. Determination of
barium in calcium carbonate rocks by carbon furnace atomic emission spectrometry.
Analyst, 102: 429.
Ingamells, C O . , 1970. Lithium metaborate flux in silicate analysis. Anal Chim. Ada, 52:
323-334.
Ingamells, C O . , 1974. New approaches to geochemical analysis and sampling. Talanta,
21: 1 4 1 - 1 5 5 .
228
Ingamells, C O . and Switzer, P., 1973. A proposed sampling constant for use in geochemi-
cal analysis. Talanta, 20: 547—568.
Ingram, B.L., 1970. Determination of fluoride in silicate rocks without separation of
aluminum using a specific ion electrode. Anal. Chem., 42: 1825—1827.
Irving, 1977. Dithizone. Analytical Sciences Monograph No. 5. Chemical Society, London,
106 pp.
Iskander, I., Syers, J.K., Jacobs, L.W., Keeney, D.R. and Gilmour, J.T., 1972. Determina-
tion of total mercury in sediments and soils. Analyst, 97: 388—393.
Jackson, M.L., 1958. Soil Chemical Analysis. Prentice-Hall, New York, N.Y., 498 pp.
Jagner, D. and Pavlova, V., 1972. A standard addition titration method for the determina-
tion of fluorine in silicate rocks. Anal. Chim. Ada, 60: 153—158.
James, C.H., 1970. A rapid method for calculating the statistical precision of geochemical
prospecting analysis. Inst. Min. Metall, Trans., Sect. B, 79: 91—96.
James, C.H. and Webb, J.S., 1964. Sensitive mercury vapour meter for use in geochemical
prospecting. Inst. Min. Metall, Trans., Sect. B, 73: 633—641.
James, G.W., 1977. Parts-per-million determinations of uranium and thorium in geologic
samples by X-ray spectrometry. Anal. Chem., 49: 967—968.
Jardine, M.A., 1963. Field and laboratory methods used by the Geological Survey of
Canada in geochemical surveys, 4. A laboratory method for determining antimony in
soils and rocks. Geol. Suru. Can., Paper, 63-29: 11 pp.
Jenkins, R., 1970. Rapid Solutions to the Particle Size Problems in X-ray Fluorescence
Spectrometry. Philips, Eindhoven, 4 pp.
Jenkins, R. and De Vries, J.L., 1970. Practical X-ray Spectrometry. Philips, Eindhoven,
190 pp.
Jonasson, I.R., 1974. Some comments on interferences by Cu(II) ions and Ag(I) ions in
the wet reduction-flameless atomic absorption determination of mercury. J. Geochem.
Explor., 3: 77—81.
Jonasson, I.R., Lynch, J.J. and Trip, L.J., 1973. Field and laboratory methods used by
the Geological Survey of Canada in geochemical surveys, 12. Mercury in ores, rocks,
soils, sediments and water. Geol. Surv. Can., Paper, 73-21: 21 pp.
Jones, L.H.P., 1957. The solubility of molybdenum in simplified systems and aqueous
soil suspension. J. Soil Sei., 8: 313—327.
Jones, M.P. and Beaven, C.H.J., 1971. Sampling of non-Gaussian mineralogical distribu-
tions. Inst. Min. Metall, Trans., Sect. B, 80: 316—323.
Josephson, M., Cook, E.B.T. and Dixon, K., 1977. A rapid method for the determination
of fluoride in geologic samples. Rep. Natl. Inst. Metall, S. Afr., 1886: 8 pp.
Lakin, H.W. and Nakagawa, H.M., 1965. A spectrophotometric method for the deter-
mination of traces of gold in geologic materials. U.S. Geol. Surv., Prof. Paper, 525-C:
168—171.
Lakin, H.W., Stevens, R.E. and Almond, H., 1949. Field method for the determination of
zinc in soils. Econ. Geol, 44: 296—306.
Lalonde, J.P., 1974. Research in geochemical prospecting methods for fluorite deposits,
Madoc area, Ontario. Geol Surv. Can., Paper, 73-38: 56 pp.
Lalonde, J.P., 1976. Fluorine — an indicator of mineral deposits. Can. Min. Metall Bull,
69: 110—122.
230
Lynch, J.J. and Mihailov, G., 1963. Field and laboratory methods used by the Geological
Survey of Canada in Geochemical surveys, 3. Method for determining arsenic. Geol.
Surv. Can., Paper, 63-8: 12 pp.
Lynch, J.J., Garrett, R.G. and Jonasson, I.R., 1973. A rapid estimation of organic carbon
in silty lake sediments. J. Geochem. Explor., 2: 171—174.
McDonald, A.J. and Stanton, R.E., 1962. The spectrophotometric determination of tin in
soil with gallein. Analyst, 87: 600—602.
McHugh, J.B. and Welsch, E.P., 1975. Atomic absorption determination of antimony in
geological materials. In: F.N. Ward (Editor) New and Refined Methods of Trace Analy-
sis Useful in Geochemical Exploration. U.S. Geol. Surv., Bull., 1408: 5—11.
McKeague, J.A. and Day, J.H., 1966. Dithionite and oxalate-extractable Fe and Al as aids
in differentiating various classes of soils. J. Soil Sei., 46: 13—22.
MacKenzie, R.C., 1954. Free iron-oxide removal from soils. J. Soil Sei., 5: 167—172.
McNerney, J.J. and Buseck, P.R., 1973. Geochemical exploration using mercury vapour.
Econ. Geol., 68: 1 3 1 3 - 1 3 2 0 .
McNerney, J.J., Buseck, P.R. and Hanson, R.C., 1972. Mercury detection by means of
thin gold films. Science, 178: 611—612.
McQuaker, N.R. and Gurney, M., 1977. Determination of total fluoride in soil and vegeta-
tion using an alkali fusion-seleotive ion electrode technique. Anal. Chem., 49: 53—86.
McQuaker, N.R., Kluckner, P.D. and Chang, G.N., 1979. Calibration of an inductively
coupled plasma-atomic emission spectrometer for the analysis of environmental mate-
rials. Anal. Chem., 5 1 : 888—895.
Mahaffey, E.J., 1974. A spectrophotometric method for the determination of rhenium in
geologic materials. J. Geochem. Explor., 3: 53—59.
Marchant, J.W. and Klopper, B.C., 1978. Microgram metal contamination of water/nitric
acid after four years in linear polyethylene containers. J. Geochem. Explor., 9: 103—
107.
Marciello, L. and Ward, A.F., 1978. Interelement corrections for spectral line inter-
ferences. Jarrell-Ash Plasma Newsl., 1: 12—13.
Marinenko, J., May, I. and Dinnin, J.I., 1972. Determination of mercury in geologic mate-
rials by flameless atomic absorption spectrometry. U.S. Geol. Surv., Prof. Paper,
800-B: 1 5 1 - 1 5 5 .
Marranzino, A.P. and Wood, W.H., 1956. Multiple-unit fusion rack. Anal. Chem., 28:
273-274.
Marshall, N.J., 1964. Rapid determination of molybdenum in geochemical samples using
dithiol. Econ. Geol., 59: 142—148.
Marshall, N.J., 1968. Rapid determination of molybdenum in geochemical samples using
dithiol. Econ. Geol, 6 3 : 291.
Marshall, N.J., 1978. Colorimetric determination of arsenic in geochemical samples. J.
Geochem. Explor., 10: 307—313.
Martin, T.D., Kopp, J.F. and Ediger, R.D., 1975. Determining selenium in water, waste-
water, sediment, and sludge by flameless atomic absorption spectroscopy. At. Absorpt.
Newsl, 14: 1 0 9 - 1 1 6 .
Maruta, T., Takeuchi, T. and Suzuki, M., 1972. Atomic-absorption and emission inter-
ferences on barium. Anal. Chim. Ada, 58: 452—455.
Maxwell, J.A., 1968. Rock and Mineral Analysis. Interscience, New York, N.Y., 584 pp.
Maynard, D.E. and Fletcher, W.K., 1973. Comparison of total and partial extractable
copper in anomalous and background peat samples. J. Geochem. Explor., 2: 19—24.
Mehra, O.P. and Jackson, M.L., 1960. Iron oxide removal from soils and clays by a
dithionite-citrate system buffered with sodium bicarbonate. In: Clays and Clay Miner-
als, Proceedings of the Seventh National Conference. Pergamon, Oxford, pp. 317—327.
Meier, A.L., 1980. Flameless atomic-absorption determination of gold in geological mate-
rials. J. Geochem. Explor., 13: 77—85.
232
Meyer, W.T. and Lam Shang Leen, Y.C.Y., 1973. Microwave-induced argon plasma emis-
sion system for geochemical trace analysis. In: M.J. Jones (Editor), Geochemical
Exploration 1972. Institution of Mining and Metallurgy, London, pp. 325—335.
Miller, J.K., 1979. Geochemical dispersion over massive sulphides within the continuous
permafrost zone, Bathurst Norsemines, Canada. In: Prospecting in Areas of Glaciated
Terrain. Institution of Mining and Metallurgy, London, pp. 101—109.
Miller, J.K. and Cazalet, P.C.D., 1979. Barium geochemistry in Irish mineral exploration.
In: Prospecting in Areas of Glaciated Terrain. Institution of Mining and Metallurgy,
London, pp. 16—21.
Minkkinen, P., 1975. A method for the correction of the background absorption in silver
analysis of calcareous samples. At. Absorpt. Newsl., 14: 71—72.
Mitchell, B.D. and Smith, F.L., 1974. The removal of organic matter from soil extracts
by bromine oxidation. J. Soil Sei., 25: 239—241.
Mitchell, B.D., Smith, B.F.L. and de Endredy, A.S., 1971. The effect of buffered sodium
dithionite solution and ultrasonic agitation on clays. Isr. J. Chem., 9: 45—52.
Mitchell, R.L., 1964a. Trace elements in soils. In: F.E. Bear (Editor), Chemistry of the
Soil. American Chemical Society, Monograph Series No. 160. Reinhold, New York,
N.Y., 2nd ed., pp. 320—368.
Mitchell, R.L., 1964b. The Spectrochemical Analysis of Soil, Plants and Related Mate-
rials. Commonwealth Bureau of Soils, Harpenden, Tech. Comm. No. 44A, 225 pp.
Morrison, G.H., 1971. Evaluation of lunar elemental analyses. Anal. Chem., 4 3 : 22A—
31A.
Morrison, G.H. and Freiser, H., 1957. Solvent Extraction in Analytical Chemistry. Wiley,
New York, N.Y., 269 pp.
Motooka, J.M., Mosier, E.L., Sutley, S.J. and Viets, J.G., 1979. Induction-coupled plasma
determination of Ag, Au, Bi, Cd, Cu, Pb and Zn in geologic materials using a detective
extraction technique — preliminary investigation. Appl. Spectrosc, 33: 456—460.
Muir, G.D. (Editor), 1977. Hazards in the Chemical Laboratory. Chemical Society,
London, 489 pp.
Mulford, C.E., 1966. Solvent extraction techniques for atomic absorption spectroscopy.
At. Absorpt. Newsl., 5: 88—90.
Myers, A.T. and Wood, W.H., 1960. Ceramic mills in a paint mixer for preparation of
multiple rock samples. Appl. Spectrosc, 14: 136—138.
Myers, A.T., Havens, R.G. and Dunton, P.J., 1961. A spectrochemical method for the
semi-quantitative analysis of rocks, minerals and ores. U.S. Geol. Surv., Bull., 1084-1:
207-229.
Nowlan, G.A., 1970. A field method for the determination of cold-extractable nickel in
stream sediments and soils. U.S. Geol Surv., Prof. Paper, 700-B: 177—180.
Olade, M. and Fletcher, K., 1974. Potassium chlorate-hydrochloric acid; a sulfide selec-
tive leach for bedrock geochemistry. J. Geochem. Explor., 3: 337—344.
Olade, M.A. and Fletcher, W.K., 1975. Primary dispersion of rubidium and strontium
around porphyry copper deposits, Highland Valley, British Columbia. Econ. Geol.,
70: 1 5 - 2 1 .
Olade, M. and Fletcher, K., 1976. Distribution of sulphur, and sulphide iron and copper
in bedrock associated with porphyry copper deposits, Highland Valley, British
Columbia. J. Geochem. Explor., 5: 21—30.
Olade, M.A. and Goodfellow, W.D., 1979. Lithogeochemistry and hydrogeochemistry of
uranium and associated elements in the Toombstone Batholith, Yukon, Canada. In:
J.R. Watterson and P.K. Theobald (Editors), Geochemical Exploration 1978. Associa-
tion of Exploration Geochemists, Rexdale, Ont., pp. 407—428.
Ooghe, W. and Verbeek, F., 1974. Atomic absorption spectrometry of the lanthanides in
minerals and ores. Anal. Chim. Acta, 73: 87—95.
Orion Research, 1970. Gran's plots and other schemes. Orion Res. Inc. Newsl., 11 (11-12):
49-55.
Orion Research, 1975. R7 Sulphate in natural waters. In: Analytical Methods Guide.
Orion Research, 7th ed., 33 pp.
Orion Research, 1976. Fluoride in drinking water. Orion Res. Inc., Appl. Bull., 5A: 2 pp.
Pahlavanpour, B., Thompson, M. and Walton, S.J., 1979. The determination of tin in
geochemical samples by inductively coupled plasma atomic emission spectrometry.
J. Geochem. Explor., 12: 45—55.
Parslow, G.R., 1979. Infrared absorption as a consistent method of correcting matrix
effect(s) of pellets used in the fluorometric determination of uranium. Chem. Geol,
27: 3 4 3 - 3 5 3 .
Parslow, G.R. and Dwairi, I., 1976. Precision and accuracy in the fluorometric determina-
tion of uranium in lake sediment. In: C.E. Dunn (Editor), Uranium in Saskatchewan.
Sask. Geol. Soc, Spec. Publ, 3: 169—191.
Parslow, G.R. and Dwairi, I., 1977. Extraction of uranium ions from water — a rapid and
convenient method using prepackaged ion-exchange resins. J. Geochem. Explor., 8:
541-547.
Parsons, M.L., Foster, A. and Anderson, D., 1980. An Atlas of Spectral Interferences in
ICP Spectroscopy. Plenum, New York, N.Y., 644 pp.
Peachey, D., 1976. Extraction of copper from ignited soil samples. J. Geochem. Explor.,
5: 1 2 9 - 1 3 4 .
Peachey, D. and Allen, B.P., 1977. An investigation into the selective dissolution of sul-
fide phases from stream sediments and soils. J. Geochem. Explor., 8: 571—577.
Peachey, D. and Vickers, B.P., 1978. An investigation into the determination of Pb in
geochemical exploration samples. Inst. Geol. Sei., Inf. Ser. Rep., 48: 8 pp.
Peachey, D., Cooper, D.C. and Allen, B.P., 1978. Rapid colorimetric determination of
copper in drill-sludge samples. Inst. Min. Metall, Trans., Sect. B, 87: 136—139.
Peachey, D., Roberts, J.L. and Scott-Baker, J., 1973. Rapid colorimetric determination
of phosphorus in geochemical survey samples. J. Geochem. Explor., 2: 115—120.
Peech, M., 1965. Hydrogen-ion activity. In: C.A. Black et al. (Editors), Methods of Soil
Analysis, Part 2. American Society of Agronomy, Madison, Wise, pp. 914—926.
Pettijohn, F.J., 1957. Sedimentary Rocks. Harper and Row, New York, N.Y., 718 pp.
Phillips, R., 1971. A method for estimating the grade of diamond deposits. Inst. Min.
Metall, Trans., Sect. B, 80: 357—362.
234
Phuphatana, A., Carlson, E.H. and Manus, R.W., 1976. The fluoride ion-specific elec-
trode as applied to fluorspar deposits in southwest Texas and southern Illinois. Econ.
GeoL, 7 1 : 6 6 1 - 6 6 5 .
Plamondon, J., 1968. Rapid determination of uranium in geochemical samples by paper
chromatography. Econ. GeoL, 6 3 : 76—79.
Plant. J., 1971. Orientation studies on stream-sediment sampling for a regional geochemi-
cal survey in northern Scotland. Inst. Min. Metall, Trans., Sect. B, 80: 324—345.
Plant, J., 1973. Random numbering system for geochemical samples. Inst. Min. Metall.,
Trans., Sect. B, 82: 64—65.
Plant, J., Jeffery, K., Gill, E. and Fage, C , 1975. The systematic determination of accu-
racy and precision in geochemical exploration data. J. Geochem. Explor., 4: 467—486.
Plüger, W.L. and Friedrich, G.H., 1973. Determination of total and cold-extractable
fluoride in soils and stream sediments with an ion-sensitive fluoride electrode. In: M.J
M.J. Jones (Editor), Geochemical Exploration 1972. Institution of Mining and Metal-
lurgy, London, pp. 423—429.
Price, G.R., Ferretti, R.J. and Schwartz, S., 1953. Fluorophotometric determination of
uranium. Anal. Chem., 25: 322—331.
Quin, B.F. and Brooks, R.R., 1972. The rapid determination of tungsten in soils, rocks
and vegetation. Anal. Chim. Ada, 58: 301—309.
Rann, C.S. and Hambly, A.N., 1965. Distribution of atoms in an atomic absorption
flame. Anal. Chem., 37: 879—884.
Rao, P.D., 1971. Determination of molybdenum in geological materials. At. Absorpt.
Newsi, 10: 1 1 8 - 1 1 9 .
Rashid, M.A. and Leonard, J.D., 1973. Modifications in the solubility and precipitation
behaviour of various metals as a result of their interaction with sedimentary humic
acid. Chem. GeoL, 1 1 : 8 9 - 9 7 .
Reid, W.R., 1969. Mineral staining tests. Colo. Sch. Mines, Miner. Ind. Bull., 12(3): 1—20.
Reynolds, R . C , 1963. Matrix corrections in trace element analysis by X-ray fluorescence:
estimation of the mass absorption coefficient by Compton scattering. Am. Mineral.,
48: 1 1 3 3 - 1 1 4 3 .
Reynolds, R . C , 1967. Estimation of mass absorption coefficients by Compton Scattering:
improvements and extensions of the method. Am. Mineral., 82: 1493—1502.
Ritchie, A.S., 1964. Chromatography in Geology. Elsevier, Amsterdam, 185 pp.
Ritchie, A.S., 1969. Recent advances in the Chromatographie analysis of geologic mate-
rials. Q. Colo. Sch. Mines, 64: 427—436.
Robbins, J . C , 1973. Zeeman spectrometer for measurement of atmospheric mercury
vapour. In: M.J. Jones (Editor), Geochemical Exploration 1972. Institution of Mining
and Metallurgy, London, pp. 315—323.
Robbins, J . C , 1978. Field technique for the measurement of uranium in natural waters.
Can. Inst. Min. Metall., Bull., 7 1 : 61—67.
Roberts, J.L., 1971. An improved colorimetric determination of vanadium in geochemical
prospecting samples. Talanta, 18: 1070—1072.
Rooney, R.C. and Woolley, J.F., 1978. Interference of calcium on barium as a means of
assessing atomic-absorption spectrophotometers. Analyst, 103: 1100—1103.
Rose, A.W., 1975. The mode of occurrence of trace elements in soils and stream sedi-
ments applied to geochemical exploration. In: I.L. Elliott and W.K. Fletcher (Editors),
Geochemical Exploration 1974. Elsevier, Amsterdam, pp. 699—705.
Rose, A.W. and Keith, M.L., 1976. Reconnaissance geochemical techniques for detecting
uranium deposits in sandstones of northeastern Pennsylvania. J. Geochem. Explor., 6:
119-137.
235
Rose, A.W. and Suhr, N.H., 1971. Major element content as a means of allowing for back-
ground variation in stream-sediment geochemical exploration. In: R.W. Boyle (Editor),
Geochemical Exploration. Can. Inst. Min. Metall., Spec. Vol., 11: 587—593.
Rose, A.W., Hawkes, H.E. and Webb, J.S., 1979. Geochemistry in Mineral Exploration.
Academic Press, London, 658 pp.
Rose, E.R., 1969. A progress report on experiments with chemical field tests for the
detection of the rare-earth elements cerium and yttrium. Geol. Surv. Can., Paper, 69-
15: 13 pp.
Rose, E.R., 1976. A field test for rare earth elements. Geol. Survey Can., Paper, 75-16:
10 pp.
Rubeska, I., Koreckova, J. and Weiss, D., 1977. The determination of gold and palladium
in geological materials by atomic absorption after extraction with dibutyl sulphide.
At. Absorpt. Newsl., 16: 1—3.
Rundle, L., Holmes, K.A. and Peachey, D., (?). The determination of arsenic by the silver
diethydithiocarbamate procedure. Inst. Geol. Sei., Geochem. Method, 16.
Ryland, A.L., 1964. A general approach to the X-ray spectroscopic analysis of samples of
low atomic number. Abstr., Am. Chem. Soc. Annu. Meet.
Sherman, G.D., McHargue, J.S. and Hodgkiss, W.S., 1942. Determination of active
manganese in soil. Soil Sei., 54: 253—257.
Sholkovitz, EJML, 1976. Flocculation of dissolved organic and inorganic matter during the
mixing of river water and seawater. Geochim. Cosmochim. Ada, 40: 831—848.
Sierra, J. and Leon, J.M., 1967. Mechanical apparatus for fusing samples in the determina-
tion of tin in geochemical surveys. Inst. Min. Metall, Trans., Sect. B, 76: 210—211.
Sighinolfi, G.P., 1972. Determination of beryllium in standard rock samples by flameless
atomic absorption spectroscopy. At. Absorpt. Newsl., 1 1 : 96.
Sinclair, A.J., Fletcher, W.K., Price, B.J., Bentzen, A. and Wong, S.S., 1977. Minor ele-
ments in pyrites from some porphyry-type deposits, British Columbia. Trans. Soc.
Min. Eng. AIME, 262: 9 4 - 1 0 0 .
Smee, B.W., Hall, G.E.M. and Koop, D.J., 1978. Analysis of fluoride, chloride, nitrate and
sulfate in natural waters using ion chromatography. J. Geochem. Explor., 10: 245—
258.
Smith, A.E., 1975. Interferences in the determination of elements that form volatile
hydrides with sodium borohydride using atomic-absorption spectrophotometry and
the argon-hydrogen flame. Analyst, 100: 300—306.
Smith, A.Y., 1964. Field and laboratory methods used by the Geological Survey of
Canada in geochemical surveys, 5. Cold extractable "heavy metal" in soil and alluvium.
Geol. Surv. Can., Paper, 63-49: 9 pp.
Smith, A.Y., 1967. Field and laboratory methods used by the Geological Survey of
Canada in geochemical surveys, 9. Tin in soils and stream sediments. Geol. Surv. Can.,
Paper, 67-50: 11 pp.
Smith, A.Y. and Lynch, J.J., 1969. Field and laboratory methods used by the Geological
Survey of Canada in geochemical surveys, 11. Uranium in soil, stream sediment and
water. Geol. Surv. Can., Paper, 69-40: 9 pp.
Smith, A.Y. and Washington, R.A., 1962. A mobile chemical laboratory for trace-ele-
ment analysis. Geol. Surv. Can., Paper, 62-21: 11 pp.
Smith, M.J. and Manaham, S.E., 1973. Copper determination in water by standard addi-
tion potentiometry. Anal. Chem., 45: 836—839.
Sorensen, R.C., Oelsligle, D.D. and Knudsen, D., 1971. Extraction of Zn, Fe and Mn
from soils with 0.1 N hydrochloric acid as affected by soil properties, solution: soil
ratio and length of extraction period. Soil Sei., I l l : 352—359.
Stanton, R.E., 1962. The field determination of zinc with dithizone in the presence of
aluminium. Geochem. Prospect. Res. Centre, Tech. Comm., 19: 9 pp.
Stanton, R.E., 1964. The field determination of arsenic in soils and sediments. Econ.
Geol., 59: 1 5 9 9 - 1 6 0 2 .
Stanton, R.E., 1966. Rapid Methods of Trace Analysis for Geochemical Applications.
Edward Arnold, London, 103 pp.
Stanton, R.E., 1970a. The colorimetric determination of molybdenum in soils, sediments,
and rocks by zinc dithiol. Proc. Australas. Inst. Min. Metall., 235: 101—102.
Stanton, R.E., 1970b. The colorimetric determination of tungsten in soils, sediment and
rocks by zinc dithiol. Proc. Australas. Inst. Min. Metall., 236: 59—60.
Stanton, R.E., 1971a. The determination of bismuth in soils, sediments, and rocks. Proc.
Australas. Inst. Min. Metall., 240: 113—114.
Stanton, R.E., 1971b. The colorimetric determination of thorium in soils, sediments and
rocks. Proc. Australas. Inst. Min. Metall., 239: 101—103.
Stanton, R.E., 1975. The colorimetric determination of palladium and platinum. Lab.
Pract., 24: 5 2 5 - 5 2 7 .
Stanton, R.E., 1976. Analytical Methods Used in Geochemical Exploration. Edward
Arnold, London, 54 pp.
Stanton, R.E. and Coope, J.A., 1958-59. Modified field test for the determination of
small amounts of nickel in soils and rocks. Inst. Min. Metall, Trans., Sect. B, 68: 9—
14.
237
Stanton, R.E. and Hardwick, A.J., 1967. The colorimetric determination of molybdenum
in soils and sediments by zinc dithiol. Analyst, 92: 387—390.
Stanton, R.E. and Härdwick, A.J., 1971. The colorimetric determination of vanadium in
soils, sediments, and rocks for use in geochemical exploration. Proc. Australas. Inst.
Min. Metall, 240: 1 1 3 - 1 1 4 .
Stanton, R.E. and McDonald, A.J., 1961-62a. Field determination of cobalt in soil and
sediment samples. Inst. Min. Metall., Trans., Sect. B, 71: 511—516.
Stanton, R.E. and McDonald, A.J., 1961-62b. Field determination of tin in geochemical
soil and stream sediment surveys. Inst. Min. Metall., Trans., Sect. B, 7 1 : 27—29.
Stanton, R.E. and McDonald, A.J., 1961-62c. Field determination of antimony in soil
and sediment samples. Inst. Min. Metall., Trans., Sect. B, 71: 517—522.
Stanton, R.E. and McDonald, A.J., 1962. The spectrophotometric determination of anti-
mony using brilliant green. Analyst, 87: 299—301.
Stanton, R.E. and McDonald, A.J., 1964. The determination of gold in soil with brilliant
green. Analyst, 89: 767—770.
Stanton, R.E. and McDonald, A.J., 1965. The determination of selenium in soils and sedi-
ments with 3,3'-diaminobenzidene. Analyst, 90: 497—499.
Stanton, R.E. and McDonald, A.J., 1966. The colorimetric determination of boron in
soils, sediments and rocks with methylene blue. Analyst, 9 1 : 775—778.
Stanton, R.E. and Ramankutti, S., 1977. The determination of platinum by atomic
absorption spectrophotometry. J. Geochem. Explor., 7: 73—76.
Stanton, R.E., Mockler, M. and Newton, S., 1973. The colorimetric determination of
molybdenum in organic-rich soil. J. Geochem. Explor., 2: 37—40.
Stary, J., 1964. The Solvent Extraction of Metal Chelates. MacMillan, New York, N.Y.,
240 pp.
Steger, H.F., 1976. Chemical phase-analysis of ores and rocks. Talanta, 23: 81—87.
Stubbs, M.F., 1968. Modification of dithiol tests for the field determination of molyb-
denum. Analyst, 9 3 : 59—60.
Sutcliffe, P., 1976. Determination of molybdenum in geological materials by atomic-
absorption spectrophotometry. Analyst, 1 0 1 : 949—955.
Szabo, N.L., Govett, G.J.S. and Lajtai, E.Z., 1975. Dispersion trends of elements and
indicator pebbles in glacial till around Mt. Pleasant, New Brunswick. Can. J. Earth
Sei., 12: 1 5 3 4 - 1 5 5 6 .
Tait, B.A.R. and Coats, J.S., 1976. A rapid method for the analysis of trace elements in
rocks, stream sediments and soil by photographic emission spectrography using a semi-
automated method of plate evaluation. Rep. Inst. Geol. Sei., 76-11: 30 pp.
Tamm, O., 1922. Eine Methode zur Bestimmung der anorganischen Komponenten des
Gelkomplexes im Boden. Medd. Statens Skogsförskningsinst. (Sweden), 19: 387—404.
Tennant, W.C., 1967. General spectrographic technique for the determination of volatile
trace elements in silicates. Appl. Spectrosc, 2 1 : 282—285.
Tennant, W.C. and Sewell, J.R., 1969. Direct reading spectrochemical determination of
trace elements in silicates incorporating automatic background correction. Geochim.
Cosmochim. Ada, 33: 640—645.
Terashima, S., 1973. Atomic absorption analysis of beryllium and vanadium in rocks.
Jpn. Anal, 22: 1 3 1 7 - 1 3 2 3 .
Terashima, S., 1976. The determination of arsenic in rocks, sediments and minerals by
arsine generation and atomic absorption spectrometry. Anal. Chim. Ada, 86: 43—51.
Thompson, A.J. and Thoresby, P.A., 1977. Determination of arsenic in soil and plant
materials by atomic-absorption spectrophotometry with electrothermal atomization.
Analyst, 102: 9—16.
Thompson, C.E., 1967. A spectrophotometric method for the determination of traces of
platinum and palladium in geologic materials. U.S. Geol. Surv., Prof. Paper, 575-D:
236-238.
238
Van Loon, J.C., 1 9 7 1 . Laboratory construction and laboratory and field evaluation of
thermoplastic chloride-selective electrodes with liquid filling solution and solid-solid
connections. Anal. Chim. Ada, 5 4 : 23—28.
Van Loon, J.C., 1 9 7 2 . The determination of small amounts of m o l y b d e n u m in soils by
atomic absorption spectrophotometry for geochemical exploration purposes. At.
Absorpt. Newsi, 1 1 : 60—62.
Van Loon, J.C., Kesler, S.E. and Moore, C M . , 1 9 7 3 . Analysis of water-extractable chlo-
ride in rocks by use of a selective ion electrode. In: M.J. Jones (Editor), Geochemical
Exploration 1972. Institution of Mining and Metallurgy, London, pp. 429—434.
Varley, J.A. and Chin, P.Y., 1 9 7 0 . Determination of water soluble sulphate in acidic sul-
phate soils by atomic-absorption spectroscopy. Analyst, 9 5 : 592—595.
Vaughn, W.W., 1 9 6 7 . A simple mercury vapour detector for geochemical prospecting.
U.S. Geol. Surv., Circ, 5 4 0 : 8 p p .
239
Vaughn, W.W. and McCarthy, J.H., 1964. An instrumental technique for the determina-
tion of submicrogram concentrations of mercury in soils, rocks and gas. U.S. Geol.
Surv., Prof. Paper, 501-D: 1 2 3 - 1 2 7 .
Viets, J.G., 1978. Determination of silver, bismuth, cadmium, copper, lead and zinc in
geologic materials by atomic absorption spectrometry with tricaprylyl methyl am-
monium chloride. Anal. Chem., 50: 1097—1101.
Vijan, P.N., Rayner, A.C., Sturgis, D. and Wood, G.R., 1976. A semi-automated method
for the determination of arsenic in soil and vegetation by gas-phase sampling and
atomic absorption spectrometry. Anal. Chim. Ada, 82: 329—336.
Von Alfthan, C , Rautala, P. and Rhodes, J.R., 1980. Applications of a new multi-
element portable X-ray spectrometer to materials analysis. In: C.S. Barrett, D.E.
Leyden, J.B. Newkirk, P.K. Predecti and C O . Ruud (Editors), Advances in X-ray
Analysis, Vol. 23. Plenum, New York, N.Y., pp. 27—36.
Walsh, A., 1955. The application of atomic absorption spectra to chemical analysis. Spec-
trochim. Ada, 7: 108—117.
Walthall, F.G., 1973. Spectrochemical computer analysis-program description. U.S. Geol.
Surv., J. Res., 2: 61—71.
Ward, F.N., 1951a. Determination of molybdenum in soils and rocks. Anal. Chem., 23:
788.
Ward, F.N., 1951b. A field method for the determination of tungsten in soils. U.S. Geol.
Surv., Circ., 119: 4 pp.
Ward, F.N. (Editor), 1975. New and Refined Methods of Trace Analysis Useful in Geo-
chemical Exploration. U.S. Geol. Surv., Bull., 1408: 105 pp.
Ward, F.N. and Bailey, E.H., 1960. Camp- and sample-site determination of traces of mer-
cury in soils and rocks. Trans. AIME, 217: 343—350.
Ward, F.N. and Bondar, W.F., 1979. Analytical methodology in the search for metallic
ores. In: P.J. Hood (Editor), Geophysics and Geochemistry in the Search for Metallic
Ores. Geol. Surv. Can., Econ. Geol. Rep., 3 1 : 365—383.
Ward, F.N. and Crowe, H.E., 1956. Colorimetric determinations of traces of bismuth in
rocks. U.S. Geol. Surv., Bull, 1036-1: 1 7 3 - 1 7 9 .
Ward, F.N. and Lakin, H.W., 1954. Determination of traces of antimony in soils and
rocks. Anal. Chem., 26: 1168—1173.
Ward, F.N. and Marranzino, A.P., 1955. Field determination of microgram quantities of
niobium in rocks. Anal. Chem., 27: 1325—1328.
Ward, F.N., Lakin, H.W., Canney, F.C. et al., 1963. Analytical Methods Used in Geo-
chemical Exploration by the U.S. Geological Survey. U.S. Geol. Surv., Bull., 1152:
100 pp.
Ward, F.N., Nakagawa, H.M., Harms, T.F. and VanSickle, G.H., 1969. Atomic-Absorption
Methods of Analyses Useful in Geochemical Exploration. U.S. Geol. Surv., Bull.,
1289: 45 pp.
Warnant, P., Martin, H. and Herbillion, A.J., 1980. Kinetics of the selective extraction of
iron oxides in geochemical samples — association between Fe and Cu in acid brown
soils. Abstr. 8th Int. Geochem. Explor. Symp., Hanover, 1980.
Warren, H.V. and Delavault, R.E., 1958. Rubeanic acid field test for copper in soils and
sediments. Min. Eng., 10: 1186—1188.
Warren, H.V. and Delavault, R.E., 1959a. Readily extractable copper in eruptive rocks as
a guide for prospecting. Econ. Geol, 54: 1291—1297.
Warren, H.V. and Delavault, R.E., 1959b. Aqua regia extractable copper and zinc in
plutonic rocks in relation to ore deposits. Inst. Min. Metall, Trans., Sect. B, 69: 495—
504.
Warren, H.V., Delavault, R.E., Peterson, G.R. and Fletcher, K., 1971. The copper, zinc
and lead content of trout livers as an aid in the search for favourable areas to prospect.
240
In: R.W. Boyle (Editor), Geochemical Exploration. Can. Inst. Min. Metall., Spec. Vol.,
11: 4 4 4 - 4 5 0 .
Watanabe, H., Berman, S. and Rüssel, D.S., 1972. Determination of trace metals in water
using X-ray fluorescence spectrometry. Talanta, 19: 1363—1375.
Watling, R.J., Davis, G.R. and Meyer, W.T., 1973. Trace identification of mercury com-
pounds as a guide to sulphide mineralization at Keel, Eire. In: M.J. Jones (Editor),
Geochemical Exploration 1972. Institution of Mining and Metallurgy, London,
pp. 59—69.
Wauchope, R.D., 1976. Atomic absorption determination of trace quantities of arsenic:
application of a rapid arsine generation technique to soil, water and plant samples.
At. Absorpt. NewsL, 15: 64—67.
Webber, G.R. and Newburry, M.L., 1971. X-ray fluorescence determination of minor
and trace elements in silicate rocks. Can. Spectrosc, 16: 3—7.
Weissberg, B.G., 1971. Determination of mercury in soils by flameless atomic absorption
spectrometry. Econ. Geol., 66: 1042—1047.
Welsh, E.P. and Chao, T.T., 1975. Determination of trace amounts of antimony in geolog-
ical materials by atomic absorption spectrometry. Anal. Chim. Acta, 76: 65—69.
Welsh, E.P. and Chao, T.T., 1976. Determination of trace amounts of tin in geological
materials by atomic absorption spectrometry. Anal. Chim. Ada, 82: 337—342.
Wendt, R.H. and Fassel, V.A., 1965. Induction-coupled plasma spectrometric excitation
source. Anal. Chem., 37: 920—922.
Wenrich-Verbeek, K.J., 1976. Water and stream-sediment sampling techniques for use in
uranium exploration. U.S. Geol. Surv., Open-file Rep., 76-77: 30 pp.
Wenrich-Verbeek, K.J., 1977. Uranium and coexisting element behaviour in surface
waters and associated sediments with varied sampling techniques for uranium explora-
tion. J. Geochem. Explor., 8: 337—355.
Whitehead, N.E. and Brooks, R.R., 1969. A comparative evaluation of scintillometric,
geochemical and biogeochemical methods of prospecting for uranium. Econ. Geol.,
64: 5 0 - 5 6 .
Whitney, R.G. and Risby, T.H., 1975. Selected Methods in the Determination of First
Row Transition Metals in Natural Fresh Water. Pennsylvania State University Press,
University Park, Pa., 94 pp.
Willis, J.B., 1975. Atomic absorption spectrometric analyses by direct introduction of
powders into the flame. Anal. Chem., 47: 1752—1758.
Windham, R.L., 1972. Simple device for compensation of broadband absorption inter-
ference in flameless atomic-absorption determination of mercury. Anal. Chem., 44:
1334-1336.
Woldseth, R., 1973. X-ray Energy Spectrometry. Kevex Corporation, Burlingame, Calif.
Wolfe, W.J., 1975. Zinc abundance in Early Precambrian volcanic rocks: its relationship
to exploitable levels of zinc in sulphide deposits of volcanic-exhalative origin. In: I.L.
Elliott and W.K. Fletcher (Editors), Geochemical Exploration 1974. Elsevier, Amster-
dam, pp. 261—278.
Wood, G.A., 1959. A rapid method for the determination of small amounts of tin in soils.
In: Symposium de Exploracion Geoquimica, International Geological Congress, XX
Session, Mexico 1956, pp. 461—474.
Wood, G.A. and Stanton, R,E., 1956-57. A rapid method for the determination of
chromium in soils for use in geochemical prospecting. Inst. Min. Metall., Trans.,
Sect.B, 66: 3 3 1 - 3 4 0 .
SUBJECT INDEX
Cadmium in water, 1 2 4
abundance, 1 1 3 mobility, 7
atomic absorption, 1 1 3 , 1 2 3 , 1 2 4 , sampling errors, 4 0
125, 1 2 6 , 1 3 1 , 132 standard, 2 1 4
—, interferences, 1 2 6 , 131 X-ray fluorescence, 1 9 4 , 1 9 6
—, solvent extraction for, 1 2 3 , 125 Cinnabar, 7 2
colorimetry, 1 0 4 Clays, 7 5 , 7 6 , 77, 8 2 , 8 4
decomposition for, 62 Cobalt
emission spectroscopy, 1 1 , 1 4 9 abundance, 2, 1 1 3
in water, 1 2 4 , 125 atomic absorption, 1 1 3 , 1 1 8 , 1 2 1 ,
mobility, 7 1 2 2 , 1 2 4 , 1 2 5 , 129, 131
polarography, 1 9 9 —, interferences, 1 2 9 , 131
standard, 2 1 4 —, solvent extraction for, 1 2 4 , 125
X-ray fluorescence, 197 colorimetry, 1 0 2 , 1 0 4
Calcium decomposition for, 5 9 , 6 0 , 6 2 , 6 6 , 67
abundance, 3 emission spectroscopy, 1 4 9 , 1 5 4 , 1 6 3 ,
atomic absorption, 127 164
emission spectroscopy, 141 extraction with ascorbic acid/hydro-
in volcanic rocks, 75 gen peroxide, 4 6 , 74
in waters, 54 in humic material, 78
mobility, 7 in igneous rocks, 59
standard, 2 1 4 in laterite, 60
water-soluble, 75 in manganese oxides, 87
Calomel electrode, 2 0 0 in rock-forming minerals, 5 9 , 6 6 , 74
Carbonates, 7 8 , 8 2 , 131 in silicate lattices, 5 9 , 74
Cassiterite, 3 6 , 37, 3 8 , 3 9 , 5 1 , 67 in sulphides, 74
decomposition for, 67 in ultramafic standards, 4 6 , 74
sampling problems, 36, 37, 3 8 , 39 in water, 1 2 4
Census mobility, 7
analytical methods used, 2 0 standard, 2 1 4
materials sampled, 2 X-ray fluorescence, 1 7 2 , 1 9 6
Chalcopyrite, 7 2 , 7 3 Cold extractions, 57, 7 1 , 9 4 ; see also
Chloride Partial extractions
ion chromatography, 1 9 9 Colorimetry, 4, 5, 6, 16, 18, 67, 7 4 , 80
mobility, 7 Bloom total heavy metal test, 8 1 , 101
specific ion electrode, 2 0 5 , 2 0 7 , 2 0 9 determination
standard, 2 1 4 —, arsenic, 1 0 4 , 105
Chlorine, 6 9 , 7 4 , 7 5 , 2 0 9 - , lead, 1 0 4 , 2 1 3
Chromatography —, niobium, 1 0 5 , 1 0 6
paper, 1 0 4 , 1 0 5 , 1 0 6 —, uranium, 1 0 5 , 1 0 6
--, for niobium, 1 0 5 , 1 0 6 paper chromatography, 1 0 4 , 1 0 5 , 106
—, for uranium, 1 0 5 , 1 0 6 principles, 97, 9 8 , 99
ion, 199 reagents, 9 9 , 1 0 0 , 1 0 2 , 1 0 3
Chromium —, dithizone, 8 1 , 8 6 , 1 0 0 , 1 0 1 , 1 0 4 ,
abundance, 2, 1 1 3 213
atomic absorption, 1 1 3 , 1 2 2 , 1 2 4 , 126 standards, 9 9 , 1 0 0 , 2 1 3
—, interferences, 126 Conductivity, 199
—, solvent extraction for, 1 2 4 Contamination, 17, 4 1 , 4 2 , 4 3 , 4 4 , 4 7 ,
colorimetry, 1 0 2 48, 5 1 , 5 4 , 58, 210, 219
decomposition for, 6 0 , 69 chloride, 2 0 9
emission spectroscopy, 1 4 9 , 1 5 4 , 1 5 9 , control graphs, 4 3 , 4 4
160 from grinding, 48
244
—, solvent extraction for, 124, 125 extraction of, 58, 8 1 , 84, 87, 92
colorimetric determination, 101, 102, readily reducible, 87, 89
104,213 stability, 77
decomposition for, 60, 62, 64, 67 Massive sulphides, weathering, 78
emission spectroscopy, 16, 149, 154, Mercury
160, 163,164 abundance, 113
in humics, 78 amalgamation on gold, 135, 199
in latentes, 60 atomic absorption, 11, 64, 109, 113,
in soils, 93 114, 116, 121, 1 3 4 - 1 3 7
in waters, 54, 124, 125 ~", by pyrolysis, 48, 135
mobility, 6 —, cold vapour method, 135, 136, 137
ores, 6, 190, 192 contamination during storage, 51
partial extraction, 80, 89 decompositions for, 64, 66, 67, 135,
polarography, 199 136
sequential extraction, 93 detectors, 136, 137, 199
specific ion electrode, 205, 207 electrochemical determination, 199
standard, 214 emission spectroscopy, 11
X-ray fluorescence, 197 loss
Lepidocrocite, 87 —, during drying, 51
Limonite, 59 —, during grinding, 48
Lithium —, during ashing, 71
abundance, 2 in soil gases, 199
emission spectroscopy, 149 in waters, preservation, 54
flame emission, 142 specific ion electrode, 205
in water, 54 spectrometer, 136
standard, 215 standard, 215
MIBK, 121, 123, 124, 125, 126
Mafic minerals, 75 Microwave plasma, 158
Magnesium, 3, 7, 75, 127, 186, 215 Mineral
Magnetite, 126 phases, 9
Manganese separates, 7, 9, 48
abundance, 2, 3, 113 stability, 6
atomic absorption, 16, 113, 120, 122 staining tests, 97
colorimetry, 99, 102 zoning, 74
decomposition for, 59, 60, 67 Mobility of elements, 6, 7, 75
emission spectroscopy, 16, 149, 154, Molybdenite, 73
160 Molybdenum
extraction, 80, 8 1 , 87, 9 1 , 92, 93 abundance, 2, 113
in drainage sediments, 8 1 , 87 atomic absorption, 11, 113, 117, 118,
in igneous rocks, 59 121, 123, 126, 127, 132
in lake sediments, 91 —, interferences, 126, 127
in latentes, 60 —, solvent extraction for, 121, 123
in rock-forming minerals, 59 colorimetry, 64, 69, 102
in soils, 80, 87, 93 decompositions for, 63, 64, 66, 67, 69
in waters, 54 emission spectroscopy, 145, 149, 154,
mobility, 7 159
nodules, 86, 168 in lake sediments, 91
standard, 215 mobility, 7
X-ray fluorescence, 168, 194, 196 standard, 215
Manganese oxides, 9, 58, 76, 82 X-ray fluorescence, 189, 196
accumulation of trace elements, 87 Molybdite, 62
exchangeable, 87, 89 Moseley's Law, 169
Muscovite, 60
249
in lake sediments, 86, 87, 62, 91 partial extractions, 72, 73, 74, 80, 86,
in latentes, 60 87,91,93
in manganese oxides, 87 polarography, 199
in porphyry copper deposits, 72 stain test, 51, 52
in sediments, 63 X-ray fluorescence, 187
in silicate lattices, 9, 59 Zirconium
in soils, 78, 93 abundance, 2, 113
in sulphides, 72, 73, 74 atomic absorption, 110, 113, 124, 126
in waters, 54, 124, 125 emission spectroscopy, 139, 159
mobility, 6, 7 mobility, 7
negative anomalies, 78 X-ray fluorescence, 196