ENGINEERING CHANGE RECORD

No: 317C110 Revision R2


Program Update to Version 00-07 (For Sysmex markets
Title:
only)[FCA#1629]
Included ECR No: 317F002
Priority: Important
Date Released: October 30, 2017
Model: CA-600 Series

Revision Information:
R2:
 Version Upgrade Procedure has been revised to Rev 1.2. (Correcting descriptions in “5.3
Re-setting of Flag Threshold” steps (5) and (8).)
 Serial number information for new production has been added for “Applicable To” section.
(2017.10.30)

R1: Adding information related to FCA#1629 in “Applicable To”, “Summary”, “Action To Be


Taken” and “Parts Quantity Changes” sections. (2017.7.18)

Applicable To:
 For FCA#1629, please refer to iFCA which will be provided by S-Corp Quality Assurance
division. (2017.7.18)
 Serial number information for new production:
CA-600 Supply Parts (EXP) S/N: A5747, A5748, A5798 and thereafter
CA-600 Supply Parts (CHN) S/N: A5558-A5609, A5772-A5776, A5806 and thereafter
(2017.10.30)

Summary:
CA-600 series program upgrade (Ver.00-07)
FCA#1629 will be carried out in this program version. Details will be provided as iFCA from S-
Corp. Quality Assurance division to your responsible division or person once parts and shipping
information become available. (2017.7.18)

Detailed Information:
The CA-600 series program has been upgraded to Ver.00-07 to resolve program issues as
follows.

[Resolved program issues]

(1) Previously, when “Barcode Scanner Driver Error” occurred, sample ID for the same single
sample could be read several times, assigning the same sample ID to several different
sample orders. When starting the analysis with this condition (Barcode Scanner Driver
Error) after returning to stand-by status, dialog to announce sample ID’s repetition was not
displayed, resulting in restarting of analysis with the wrong sample ID.
This issue has been resolved. When CA-600 is connected to barcode reader, now sample
ID repetition will be checked at the start of analysis and at the start of analysis after adding
orders. (TW#70179)

*Please refer to "CA-600 Program Version History" “CA-600 Version Upgrade Procedure
(Ver.00-07)” for details.

Action To Be Taken:

1. Please refer to the attached pages of updated SYSMEX documentation for applicable
information. These updated pages will be included in the next Manual update.

2. FCA#1629 will be carried out in this program version. Please update the program to
Ver.00-07 once available. (KIT NO.470 [P/N: AP363965] will be provided by free-of-
charge.) Details will be announced as iFCA from S-Corp. Quality Assurance division to your
responsible division or person once parts and shipping information become available.
(2017.7.18)

Reference Document:
None

Parts Quantity Changes:


Old New Lead
No. Description Status Category
Qty Qty Time*1
BD335959 CARD_ASSY NO.18(MEMORY) 1 1 Changed 90 days MTO
AP363965 KIT NO.470*2 0 1 New item - Unsalable
*1: NOTE: Lead time is only an approximate delivery time for your reference.
*2: This part is FCA part, and therefore, Unsalable.

Definition of Status in “Parts Quantity Changes” section


New item Newly registered item
Available Currently also used on other models
Modified Code no. remains unchanged however some minor modification was done to the item
Quantity changed Code no. and item characteristics remain unchanged, but item quantity was changed
Not used Not used on the specified unit/assembly/model however may be used on other
Discontinued unit/assembly/model
No longer being used on any Sysmex products coming from production

Attachments:
CA-600 Version Upgrade Procedure (Ver.00-07)(Rev 1.2)
CA-600 Program Version History (Ver.00-07)

End Of Document

Person in Charge: Chinatsu Niimi


Reviewed by: Takeshi Uemura, Junko Horitaka, Mark Dorfner
Authorized by: Tamiko Fujita