Professional Documents
Culture Documents
Neutron Contamination of 10 MV X-rays_ Its Relevance to Treatment Room Door and Maze Design (British Journal of Radiology, Vol. 80, Issue 954) (2007)
Neutron Contamination of 10 MV X-rays_ Its Relevance to Treatment Room Door and Maze Design (British Journal of Radiology, Vol. 80, Issue 954) (2007)
1
Medical Physics Department, Guy’s and St Thomas’ NHS Foundation Trust, St Thomas’ Hospital, London SE1 7EH and
2
Medical Physics Department, Royal Sussex County Hospital, Brighton, East Sussex BN2 5BE, UK
Neutron dosimetry admits that ‘‘insufficient account was taken of the large
photon component that results in the human body when
The impression given by many publications [1] is that exposed to neutrons below 1 MeV’’ [4]. The resulting
the production of neutrons by clinical linear accelerators is difficulties have still not been fully resolved.
only relevant for machines producing X-rays of energy We deliberately avoid the use of radiation weighting
above 10 MV. factors as employed in the maze calculations of Institute of
This can lead to the false conclusion that neutrons are Physics and Engineering in Medicine (IPEM) 75 [5].
unimportant at 10 MV, so much so that at a recent Whenever possible in this publication we quote neutron
installation at St Thomas’ Hospital, London using a direct doses and perform calculations in terms of operational
access door with no maze, the door supplier initially quantities. Ambient dose equivalents H*(10) are relevant
suggested that no neutron protection would be necessary. to survey meter measurements and personal dose equiva-
Neutron dosimetry for radiation protection is difficult. lents Hp(10,0) to personal dosimeter badge measurements.
Terms and quantities relevant to radiation safety are set out The numerical relationship of H*(10) and Hp(10,0) is, to a
in the UK in the Approved Code of Practice of the Ionising good approximation, unity [6] and we use this simplifica-
Radiation Regulations (IRR) [2] and incorporate changes tion. IRR 1999 accepts that such measurements can be used
recommended by the International Commission on to demonstrate legal compliance with effective dose limits.
Radiological Protection (ICRP) in 1991 in Publication 60 [3]. Another area of complexity is the traditional use of
Changes to radiation factors present a particular problem neutron fluence for head leakage and room scatter calcula-
when comparing past measurements or using and adapt- tions as in IPEM 75 and NCRP 79 [7]. To convert fluence to
ing old methods of calculation. Changes to neutron quality ambient or personal dose equivalent [8] knowledge of the
factors introduced by ICRP Publication 60 were not large neutron spectrum is required. We do not have this knowl-
and a neutron survey meter calibrated, say, 20 years ago edge and have attempted to convert fluence calculations to
could retain the same calibration factor if calibration is dose equivalent calculations using information in NCRP 79.
performed to an accuracy of +/2 30% (Private commu- Using the dosimetric approach outlined above, this
nication, Thomas DJ, National Physical Laboratory (NPL) paper investigates neutron leakage when a Precise linear
2004). Even the calibration factor supplied by the NPL for accelerator (Elekta Ltd, Crawley, UK) is generating
the St Thomas’ meter used in this publication has not 10 MV X-rays and quantifies the effect for bunkers with
changed significantly with the introduction of IRR 1999. direct access doors and short mazes.
The impression that neutron effective dose has increased
substantially is given when calculations utilize radiation
weighting factors recommended by ICRP 60. However,
errors were made in estimating the new factors and ICRP
Materials – neutron dosimeters
Neutron personal dosimeter badges
Address correspondence to: Dr Peter Rudd, Medical Physics
Department, Guy’s and St Thomas’ NHS Foundation Trust, St
Neutron monitoring badges were used in high dose
Thomas’ Hospital, Lambeth Palace Road, London SE1 7EH, UK. rate situations and to confirm doses estimated from
E-mail: peter.rudd@gstt.sthames.nhs.uk survey meter readings in low dose rate areas [9].
The two types of badges employed were of the CR-39 As demonstrated by recent Monte Carlo calculations [11]
type. One difficulty in choosing a suitable badge is these meters can under-respond or over-respond by
determining the consequences of its energy response [6, almost an order of magnitude at certain energies and
10]. In order to determine the leakage rate from the X-ray beam directions. In their normal orientations H*(10) can
head polyallyldiglycol carbonate (PADC) badges were be overestimated by about a factor of 5 at approximately
obtained from the NRPB, Chilton, UK. The NRPB, now 5 keV and underestimated by a factor of 2.5 at 0.025 eV
the Radiation Protection Division of the Health for neutrons in thermal equilibrium with their surround-
Protection Agency, performed calculations to correct ings. However, corrections for NM2 meters in practical
the badge response for intrinsic sheet sensitivity, the spectra in the nuclear industry can be only of the order of
angle of incidence and the neutron spectrum. Luxel dose 15% [12]. Due to the paucity of relevant published
monitoring badges containing a Neutrak 144 neutron spectra no correction has been applied in this paper.
detector were obtained from Landauer Inc., Oxford, UK
for measuring photon and neutron doses within and
external to a treatment room as discussed below.
Landauer reported the doses from these badges in their Neutron head leakage
normal manner. The energy range of these badges is
quoted as ‘‘Thermal neutrons below 0.5 eV and fast Measurements
neutrons from 40 keV to 35 MeV’’. Badges have inherent PADC badges were placed at varying distances and
limitations when measuring low doses outside treatment positions in the isocentric plane and around the head of
rooms. The minimum reported dose for fast neutrons for Elekta 1 (serial number 105635) at St Thomas’ Hospital.
each of the two badges used in this study is quoted as
The collimation system was closed. The photon dose that
0.2 mSv and for Neutrak 144, for thermal neutrons,
would otherwise have been delivered to the isocentre
0.1 mSv.
was 100 Gy.
Edir is the average energy of the direct neutrons after Neutron shielding in the door
passing through the head shielding (MeV) From NCRP 79 the dose equivalent tenth value layer
r is the distance from the X-ray target (m) (TVL) for polyethylene is expected to be 7.15 cm for Eave
S is the internal surface area of the treatment room equal to 0.28 MeV. Unlike photons, it is not clear what
(m2) oblique incidence correction is necessary; therefore, no
W is a dimensionless constant attempt has been made to make such a correction. In
W normally has a value of 5.4. But when we consider order to reduce the transmitted H*(10) to 0.5 mSv per
an opening, we have assumed only half the room annum, 19.9 cm of polyethylene would be required.
scattered neutrons are travelling in the direction out of Atomic International actually installed 22.5 cm of poly-
the treatment room; in such a case we have used a value ethylene to fully fill the void in the door. Calculation
of 2.7. This simple assumption possibly leads to an therefore suggests that the transmitted H*(10) should be
overestimation by a factor of 2 or p in the scattered 0.21 mSv per annum and the instantaneous H*(10) rate,
neutron contribution to the dose [13, 14]. The term 1.25 mSv h21. The attenuation of the steel facing of the
involving the ratio Eave/Edir is needed to convert from a door can be ignored and that of 20 cm of lead is minimal;
fluence based equation. if we used a TVL of 80 cm [15] the instantaneous H*(10)
The power factor 0.735 results from Figure 31a of rate would be reduced to 0.7 mSv h21.
NCRP 79 and is applicable to spectra with average
neutron energies below 1.0 MeV. No account is taken Unattenuated dose at the door due to thermal
here of the effects of ICRP 60 on the accuracy of Figure neutrons
31a. Thermal neutrons scattered in the room are considered
Eave/Edir is derived from Equation (2) which is a separately in NCRP 79. McGinley et al [16] provide an
reformulation of Equation (14) of NCRP 79 to amplify the estimate of the thermal neutron fluence. We derive
effect of selecting different values of W. Equation (3).
Eave =Edir ~(Sz0:24W 4pr2 )=(SzW 4pr2 ) ð2Þ TR~L(1)r12 (Ft =Fdir ) (Y 4p=S) ð3Þ
for the primary barriers, shaded in Figure 1, which are of 2.2 cm depth at the isocentre in a 10 cm square field.
density 3500 kg m23. The Elekta 4 bunker is constructed Table 2 compares measurements and calculated values.
of concrete of density 2350 kg m23. Vertical primary Repeat readings, made by interchanging the survey
barriers, shaded in Figure 2, and approximately one- meters, suggest that the values were reasonably repro-
third of the surface area of the treatment room ceiling are ducible at each position. Measurement 1A was made
lined with steel plates. The Elekta 4 maze is fitted with a with a 6 MV beam and indicates that meter A appears
door of 4 mm lead, 36 mm MDF and 8 mm plywood to to respond to photons. This was unexpected.
reduce the annual photon H*(10) to below 1 mSv. Both Measurements 1, at 10 MV, indicated meter A readings
bunkers are dry lined with plasterboard. much higher than both meter B readings and the
calculated values. The apparent ability of meter A to
detect photons at 6 MV would not explain such a large
Calculations discrepancy. At positions further along the maze meter A
again appears to over-read but with decreasing effect.
We have used a slightly modified version of the For measurements in the range 1 to 30 mSv h21, meter
original Kersey method [5, 19]. A reads 14% higher on average than the Elekta meter.
The ratio of measured/calculated H*(10) was of the
H(di)~DRL(1)106 (r12 =R2 )(Sm=So)1=10(1z(di3)=5) ð4Þ order of 0.9 for meter A and 0.8 for meter B. The accuracy
of the measurements at position 7 is affected by the
resolution of the digital scale. Calculation and measure-
where
ment indicate that, at the end of the maze, the annual
H(di) is the H*(10) rate at position di in the
value of H*(10) would be of the order of only 0.1 mSv for
maze (mSv h21)
neutrons. Had the maze been 7 m long the neutron
di is the distance along the maze (m)
H*(10) dose could have been of the order of 1 to 1.5 mSv,
DR is the photon dose rate at the isocentre (Gy h21) dependent on using the measured or calculated values.
L(1) and r1 are as above (Sv Gy21 and m)
R is the distance from the isocentre to the maze
entrance (m)
Sm is the minimum cross-section of the maze (m2) Measurements at St Thomas’ Hospital
So is a an area constant (6 m2) Measurements were generally made with the same
106 is a constant to convert Sv to mSv conditions as at the Royal Sussex County Hospital but
We have assumed that below a di value of 3 m the limited studies were made to determine the effect of fully
attenuation term becomes 1/10(di/3). The distances R and di opening the multileaf collimator (MLC)/jaw system and
have been measured in accordance with Figures 1 and 2 gantry angle. Initially measurements were made using
which are consistent with the diagrams shown in IPEM 75. meter A. Dr David Thomas of the National Physical
The values of Sm and R at the Royal Sussex County bunker Laboratory, Teddington kindly agreed to perform some
and the Elekta 4 bunker at St Thomas’ are, respectively, independent neutron measurements with two Harwell
4.1 m2, 4.1 m and 5.4 m2 and 7.1 m. Calculated values are Model 0949 survey meters. At the NPL visit meter A
shown in Tables 2 and 3. No allowance was made for the measurements were made only at the maze entrance.
presence of the door in the Elekta 4 maze.
Results
Measurements at Royal Sussex County Hospital The nominal 10 MV photon dose rate at the isocentre was
Measurements were made using meters A and B. The 450 monitor units per minute; this equates to 270 Gy h21.
meters were placed side by side in the maze. At certain Table 3 compares measured and calculated values. We
positions the meters were interchanged and the readings present only the average results of the two NPL meters.
repeated. The digital scale of the meter was read on the Measurement 1A was made with the 6 MV beam and again
999.9 mSv h21 setting, which involved 34 s measuring demonstrates that meter A appears to respond to photons.
periods. A member of staff was present in the maze to Measurements at positions 1 to 3 in the range 9 to
operate the start button on the monitors but otherwise 73 mSv h21 indicated that meter A reads 20% lower on
retired close to the maze entrance whilst the beam was average than the NPL meters and the ratio of measured/
running. The gantry was pointed vertically down and the calculated H*(10) was of the order of 1.59 for meter A and
collimator system closed down to produce a minimal size 2.03 for the NPL meters. The raised value of the ratio for
field. Consequently personal doses were kept very low. measurement 4 might be indicative of backscatter of
Measurements were made along the centre of the maze at neutrons from the door and the reduction for measure-
approximately 1.2 m above floor level. An attempt was ments 5 and 5A suggests neutron attenuation by the door
made to determine if the X-ray pulse affected monitor by a factor of approximately 2. Further readings were made
readings by running the beam at 6 MV. at the control desk side of the treatment door with meter A
to determine the effect of opening the collimator system
and gantry angle. With the gantry vertical and with the
field opened to a 40 cm square the NM2B reading was
Results
reduced from 1.7 to 1.4 mSv h21; a reduction would be
The nominal photon dose rate at the isocentre was 400 expected since the collimation system contributes neutrons
monitor units per minute; this equates to 240 Gy h21 at when in the photon beam path. Directing a 40 cm square
Table 2. Comparison of measured and calculated H*(10) rates in the maze at the Royal Sussex County Hospital.
Measurement di (m) Measured (mSv h21) Calc (mSv h21) Measurement/calculation
field to a man size phantom positioned at the isocentre loaded. Atomic International supplied 22.5 cm of poly-
yielded NM2B readings in the range of 0.7 to 2.3 mSv h21 ethylene, sufficient to fill the void in the door. The
depending on gantry angle. The variation with gantry angle instantaneous neutron H*(10) rate outside the door was
was not expected, since neutron emission from the head is in good agreement with that calculated. Such agreement
often assumed to be reasonably symmetric. The average may have been fortuitous. It would of course have
reading was, perhaps fortuitously, close to that at the been interesting to determine how the dose rate varied
vertical gantry angle position. If we use the NPL reading of with the thickness of polyethylene, but this was not
2.3 mSv h21, the anticipated annual H*(10) dose equivalent practicable. Boron loaded polyethylene was not used and
due to neutrons is 0.43 mSv. The photon H*(10) has been this appeared to have had no detrimental impact.
determined to be of the order of 0.93 mSv. The total annual Information on the ability of pure polyethylene to
H*(10) is therefore expected to be 1.35 mSv; this is within the attenuate thermal neutrons appears to be contradictory.
design criteria of 1.5 mSv for 100% occupancy. Meter A appeared to produce erroneous results in
fields where neutron H*(10) rates are higher than
100 mSv h21. The fact that it responds to 6 MV photon
Discussion and conclusions fields suggests the delay circuit designed to eliminate the
fast photon pulse is not working correctly. At dose rates
IPEM 75 quotes a value of L(1) of 0.002% Sv Gy21, encountered at the end of real mazes, meter A agreed
based on data applicable to accelerators produced by with the other three meters to better than 26%. This gives
Varian, Palo Alto, USA. The Elekta Precise Planning a modicum of confidence in the use of the St Thomas’
Guide quoted a neutron dose equivalent of 0.03% meter. However, NM2B and Leake type meters have
Sv Gy21, a factor of seven times that measured. This similar energy responses and could be under-or over-
value appears to be consistent with Table 6.1 in IPEM 75 estimating the true H*(10) rates. It is unlikely that all the
which depicts neutron leakage limits if a radiation neutrons are thermalized in typical mazes and therefore
weighting factor of 10 is used. In subsequent correspon- it is unlikely that our results significantly underestimate
dence with Elekta Ltd this was revised downwards to dose rates. Equation (4) appears to predict measured
0.01% Sv Gy21. We used our measured value of 0.0045% H*(10) doses within a factor of approximately 2. Neutron
Sv Gy21 for calculations. H*(10) rates along mazes of the order of 7 m or less in
Calculation of neutron shielding in the direct entry length are not insignificant if the aim is to reduce total
room door suggested that we use 20 cm of polyethylene photon and neutron dose to the order of 1.5 mSv per
and tradition dictated that 25 mm of this be boron annum at 100% occupancy.
Table 3. Comparison of measured and calculated H*(10) rates in the maze of the Elekta 4 bunker at St Thomas’ Hospital
Measurement di (m) Measured (mSv h21) Calc (mSv h21) Measurement/calculation
Neutron dosimetry for radiation protection is dose equivalent quantities for neutrons. Journal of the ICRU;
fraught with difficulties. This paper does not purport Volume 1 No 3. Ashford, UK: Nuclear Technology
to be a definitive text on the subject. However, it Publishing, 2001.
indicates that neutron doses cannot be neglected when 7. US National Council on Radiation Protection and
considering the safety of direct access doors or short Measurements. Neutron contamination from medical elec-
tron accelerators, NCRP Report 79. Bethesda, MD, 1984.
mazes at 10 MV.
8. International Commission on Radiological Protection.
Conversion coefficients for use in radiological protection
against external radiation. ICRP Publication 74. Oxford, UK:
Acknowledgments Pergamon, 1996.
9. Powell NL, Newing A, Bullen MA, Sims C, Leaton SF. A
We are grateful for the loan of the NM2B survey meter radiation safety survey on a Clinac-20 linear accelerator.
from Elekta Ltd, Crawley. We are indebted to Dr. David Phys Med Biol 1987;32:707–18.
Thomas, NPL, Teddington for the measurements per- 10. Tanner RJ, Thomas DJ, Bartlett DT, Hager LG, Horwood N,
formed at St Thomas’ and Dr R. Tanner, NRPB, Chilton Taylor GC. Effect of the energy dependence of neutron
for the analysis of the PADC neutron badge results. personal dosemeters routinely used in the UK on the
Angela Daynes of Guy’s and St Thomas’ ably assisted in accuracy of dose estimation, Report NRPB-W25. Chilton,
the measurements made on Elekta 4, Tony Greener UK: National Radiological Protection Board, 2002.
provided helpful comments on the paper and Philip 11. Bartlett DT, Tanner RJ, Tagziria H, Thomas DJ. Response
Sands kindly prepared the figures of the bunkers. characteristics of neutron survey instruments, Report
R333(rev). Chilton, UK: National Radiological Protection
Board, 2002.
References 12. Tanner RJ, Bartlett DT, Hager LG, Jones LN, Molinos C,
1. Greene D, Williams PC. Linear accelerators for radiation Roberts NJ, et al. Practical implications of neutron survey
therapy. Bristol, UK: Institute of Physics Publishing, 1997. instrument performance. Radiat Prot Dosim 2004;110:763–7.
2. HSE. Work with ionising radiation, Ionising Radiations 13. Hayes DP. Letter to the Editor. Med Phys 2003;30:1283.
Regulations 1999, Approved Code of Practice and 14. McGinley. Letter to the Editor. Med Phys 2003;30:1284.
Guidance. London, UK: HSE Books, 2000. 15. Barish RJ. Evaluation of a new high-density shielding
3. International Commission on Radiological Protection. ICRP material. Health Phys 1993;64:412–16.
publication 60. Oxford, UK: Pergamon, 1991. 16. McGinley PH, Miner MS, Mitchum ML. A method for
4. International Commission on Radiological Protection. calculating the dose due to capture gamma rays in
Relative biological effectiveness (RBE), quality factor (Q), accelerator mazes. Phys Med Biol 1995;40:1467–73.
and radiation weighting factor (wR). ICRP Publication 92. 17. Gibson JAB, Piesch E. Neutron monitoring for radiological
Oxford, UK: Pergamon, 2003. protection, Technical Report Series No. 252. Vienna,
5. Stedeford B, Morgan HM, Mayles WPM, editors. The Austria: International Atomic Energy Agency, 1985.
design of radiotherapy treatment room facilities, Report 18. Bewley DK. The physics of radiology of fast neutron beams.
No. 75. York, UK: Institute of Physics and Engineering in Bristol, UK: Adam Hilger, 1989.
Medicine, 1997. 19. Kersey RW. Estimation of neutron and gamma radiation
6. International Commission on Radiation Units and doses in the entrance mazes of SL 75–20 linear accelerator
Measurements. ICRU Report 66. Determination of operational treatment rooms. Medicamundi 1979;3:151–5.