Model 9437
Low and High Level Dissolved Oxygen Monitor
A compact, competitively
priced low level dissolved
oxygen monitoring system
Fastresponse
— reacts to rapid changes in plant operation
2
Low and High Level Dissolved Oxygen Monitor
Models 9437 DS/9437–EN Rev. H
Sensor System
The sensing system employs the well proven 9435-300
disposable capsular sensor mounted in a compact flow cell
which also houses a Pt1000 temperature sensor.
The sample flow path in the cell is designed so that rapid
response is ensured while membrane erosion by particles in
sample is avoided.
3
Low and High Level Dissolved Oxygen Monitor
Models 9437 DS/9437–EN Rev. H
Interconnecting Cable
(Maximum length 30m)
Wall-mounting
Transmitter
9435-300 Sensor
Panel-mounting
Transmitter
4
Low and High Level Dissolved Oxygen Monitor
Models 9437 DS/9437–EN Rev. H
Specification — Flowcell
Mounting
Vertically, using the built-in fixing bracket Salinity correction
Preset within the range 0 to 80ppt
Measuring ranges
Programmable within the ranges 0 to 20.0μg/kg Barometric pressure correction
Preset within the range 500 to 800mm Hg
and 0 to 20mg/kg
Sample flow
Scaling
100 to 500ml/min
μg/kg, mg/kg or ppb, ppm
Sample pressure
Accuracy
Maximum 2bar
±5% of reading or ±1μg/kg, whichever is the greater
Sample temperature
Response time
5 to 55°C (41 to 131°F)
90% of a step change in 1 minute
Sensor ambient temperature
Resolution
0 to 55°C (32 to 131°F)
0.1 μg/kg
Stability
±5% of reading or ±1μg/kg per week, whichever is the greater
Temperature compensation
5 to 55°C automatic using Pt1000 resistance thermometer
5
Low and High Level Dissolved Oxygen Monitor
Models 9437 DS/9437–EN Rev. H
Specification — Transmitter
Serial communication
Set Points and Relays
RS422/RS485 (optional, with one current output signal)
No. of set points
Two
Power Sup ply
Set point adjustment
Voltage requirements
Programmable
100 to 130V or
Set point hysteresis 200 to 260V 50/60 Hz
±1% of FSD (fixed)
Power Consumption
Local set point annunciation < 10VA
Red LED
Error due to power supply variations
No. of relays Less than ±2% for +6% –20% variation from nominal supply
Two voltage
6
Low and High Level Dissolved Oxygen Monitor
Models 9437 DS/9437–EN Rev. H
7
Low and High Level Dissolved Oxygen Monitor
Models 9437 DS/9437–EN Rev. H
Installation Options
The adjacent Figure A shows the recommended system with a
sample isolation valve V1 and a flow meter, where the dissolved
oxygen sensor remains in the flowcell during calibration in air.
An extra valve (V2) is fitted to drain the sample from the flowcell Tundish
prior to calibration. Flowcell
Figure B shows a simple system with a sample line isolation Flowmeter Drain
valve (V1). In this arrangement the oxygen sensor is removed
from the flowcell for calibration in air.
V1
V2
Drain
V1 = Sample Isolating Valve
A – Recommended System V2 = Sample Drain Valve
Tundish
Flowcell
Drain
V1
B – Simple System
Compression fitting
suitable for 6mm O.D.
47. or 1/4 in. O.D. tubing
(1.80 (To be specified when
5) ordering)
Sample
Inlet
8
Low and High Level Dissolved Oxygen Monitor
Models 9437 DS/9437–EN Rev. H
Overall Dimensions
Dimensions in mm (in.)
160 (6.3) 68 (2.68)
69 (2.72)
Wall-mounting Transmitter
Dimensions in mm (in.)
96 92 +0.8
–0
Panel Cut-out (3.62 +0.03)
(3.78) –0
Panel-mounting Transmitter
9
Low and High Level Dissolved Oxygen Monitor
Models 9437 DS/9437–EN Rev. H
Electrical Connections
1 2 3 1 2 3
1 2 3 4 5 6 7 + – N L
4 5 6 4 5 6
Earth Stud
Wall-mounting Transmitter
Retransmission Positive 1
1 0V
output Negative 2
2 Rx–
3 RS485 Serial 2nd retransmission
1 1 3 Rx+ Interface (Optional) output available
Normally Closed 4 2 2
4 Tx–
3 3 – when RS485 serial
Relay 1 Common 5 4 4 5 Tx+ + communications not
Normally Open 6 5 5
used.
Normally Closed 7 TBA 6
7
6
7
TBB 6 Link
Relay 2 8 8
7 Green Temperature Compensator
Common 8 9 9 8 White
Normally Open 9 10 10
9 –ve (Black)
11 11
10 12 12 10 +ve (Red)
Neutral 11
11
Mains supply DO Sensor
Live 12
12
Earth
Earth stud (on case)
Panel-mounting Transmitter
10
Low and High Level Dissolved Oxygen Monitor
Models 9437 DS/9437–EN Rev. H
Ordering Information
Dissolved Oxygen Monitor 9437/00 X X X X X
Standard
Standard 6mm fitting 0
Flow cell (6mm fitting) less capsule 1
Standard 1/4 in. fitting 2
Flow cell (1/4 in. fitting) less capsule 3
Special 9
Enclosure
Wall-mounting 0
Panel-mounting 1
Output
Current output only 0
Current output + serial data interface (Modbus protocol) 1
2 current outputs 2
Special 9
Cable length
3m (9.75 ft) 0
5m (16.25 ft) 1
10m (32.4 ft) 2
20m (65 ft) 3
30m (97.5 ft) 4
Special 9
Language
English 0
German 1
French 2
Spanish 3
Special 9
11
Contact us
12.2010
Process Automation We reserve the right to make technical changes or
modify the contents of this document without prior
Oldends Lane
notice. With regard to purchase orders, the agreed
DS/9437–EN Rev. H
Stonehouse particulars shall prevail. ABB does not accept any
Gloucestershire GL10 3TA responsibility whatsoever for potential errors or
UK possible lack of information in this document.
USA
Tel: +1 215 674 6000
Fax: +1 215 674 7183
www.abb.com