Frankever Soil Tester

FK-3002-TRREC-N01
Ver 2.0
catalogue
Chapter 1 product introduction
1.1 Product Overview
1.2 functional features
1.3 main parameters
1.4 product selection
1.5 product selection
Chapter 2 usage
2.1 equipment list
2.2 structure description
2.3 sensor connection equipment
2.4 terminal instructions
2.4.1 quick measurement method
2.4.1 buried survey method
Chapter 3 function and operation description
3.1 key description
3.2 introduction to main interface
3.3 operating instructions
3.3.1 system setting description
3.3.2 equipment management description
3.3.3 description of viewing historical data
Chapter 4 instructions for configuration software

1
4.1 setting detection parameters
4.2 importing data
4.3 export data
4.4 clearing equipment data
Chapter 5 description of charging function
Chapter 6 precautions
Chapter 7 common faults and Solutions

Chapter 1 product introduction

1.1 Product Overview


The Dr. earth recorder developed and designed by our company adopts the latest
digital integrated circuit technology and international detection technology to design a
new intelligent handheld detector. The recorder adopts a large-scale full-color LCD
screen, which can display readings in real time. At the same time, it uses the detection
circuit designed by the digital chip of international manufacturers, which can achieve
very high sensitivity and excellent repeatability. Moreover, the recorder integrates
storage, recording and analysis, and can be flexibly connected to the company's soil
nitrogen, phosphorus and potassium sensor, soil conductivity, soil temperature and
humidity and other equipment to measure a variety of soil parameters.
The recorder is widely used in the detection of soil pH, dry farming and
water-saving irrigation, fine agriculture, forestry, geological exploration, plant
cultivation, water conservancy, environmental protection and other fields, as well as
the measurement of pH of various particles.

1.2 functional features


1) It can directly display the measurement results, which is simple and convenient,
low measurement cost and fast measurement speed.
2) Large color display screen and large capacity storage.
3) One click data export is convenient and fast.
4) The probe insertion design ensures accurate measurement and reliable
performance.

1.3 main parameters


Power supply mode Battery powered (5000mAh lithium battery)
Display mode 2.8-inch LCD
data storage 100W data
Charging time ≤8h
Temperature: - 10 ℃ - 50 ℃; Humidity < 95%
work environment
RH, no condensation

2
Connectable equipment 1-4 MODBUS device combinations
communication protocol Modbus RTU protocol
working voltage DC 4.2V
Standby time More than 8h continuous
size 174*88.5*35 mm
weight 284g

1.4 product selection


FK- Company code
3002- Soil recorder housing
TRREC- Soil multi parameter recorder
N01- RS485 (Modbus RTU protocol)
Upload in 4G mode (not
4G-
developed yet)
empty No Bluetooth printing
With Bluetooth printing (not yet
BT
developed)

1.5 product selection

3
Product size:

Equipment dimension drawing (unit: mm) Chapter 2 usage


2.1 equipment list
■ 1 recorder
■ 1 data line
■ warranty card
2.2 structure description

2.3 sensor connection equipment


4
Firstly, take out the aviation plug line, plug the aviation plug adapter line with
one to two line or one to three line, then plug the one to two line or one to three line
branch line with the soil equipment, support the simultaneous access of up to four soil
sensors, and finally connect the male end of the connected aviation plug line with the
aviation plug of Dr. soil recorder.

2.4 terminal instructions


2.4.1 quick measurement method
Select a suitable measuring location, avoid stones and ensure that the steel needle
will not touch hard objects. Throw away the topsoil according to the required
measuring depth, maintain the original tightness of the underlying soil, hold the
sensor tightly and insert it vertically into the soil. Do not shake left and right during
insertion. It is recommended to measure multiple times within a small range of a
measuring point to obtain the average value.

5
2.4.1 buried survey method
Dig the pit with diameter > 20cm vertically, insert the sensor steel needle
horizontally into the pit wall at the given depth, fill the pit tightly and stabilize it for a
period of time, and then measure and record for several days, months or even longer

Chapter 3 function and operation description


3.1 key description

6
picture name explain
Main interface: press and hold the opening sound in the
alarm state and open state. When an alarm occurs, press
and hold to eliminate the alarm sound; Short press up in
Up key
the menu interface to select setting items and add
values; Short press on the password interface to
increase the value
Main interface: long press to close the sound in the
alarm state and open state
Short press down on the menu interface to select setting
Direction down key
items and add values
Short press on the password interface to reduce the
value
Short press right on the main interface to switch the
display interfaces of different devices
Left arrow key
Short press on the menu interface to move the cursor to
the right
Short press right on the main interface to switch the
display interfaces of different devices
Direction right key
Short press on the menu interface to move the cursor to
the right
Host interface: long press to open or close the data
record, short press to enter the password interface;
Confirm key
Short press to confirm the selection in the menu
interface
Short press in the menu interface to return to the
Return key previous interface;
Any interface: long press to return to the main interface.

Bluetooth printing (not


Long press to realize one touch Bluetooth printing
yet developed)

Shutdown state: press and hold for 2S, and the


equipment will enter the startup state;
Power key
Main interface status: press and hold 4S, and the
equipment shuts down,

7
3.2 introduction to main interface

Serial name explain


numbe
r
one time Display current time (manual adjustment)
two call the police As shown in the figure, it indicates that the general alarm sound is on (the
buzzer alarm can be turned off separately). When it is displayed, it
indicates that the general alarm is on
three USB socket This flag is displayed when USB is inserted
four data record As shown in the figure, it means to turn on the recording data (you can set
to turn on or turn off the recording data). When ■ is displayed, it means
to turn off the recording data
5five Buzzer As shown in the figure, it indicates that the buzzer is normally
enabled. When the buzzer is disabled, it displays:

When the alarm is off, the buzzer cannot be activated


six Electric quantity Displays the current remaining power
seven Access device type Type of current access device (access device can be set)
eight Equipment status Displays the online / offline status of the current device
nine address Address of access device
ten Baud rate Baud rate of access device

eleven Real time value Displays the actual feature values currently detected
twelve Proportion The ratio of the current measured value to the range of this sensor

8
3.3 operating instructions
1. In the shutdown state, long press the 2S confirmation key, the equipment
detects whether the buzzer is in normal use, and the equipment enters the main
interface
2. On the main page, briefly press the confirm key to enter the password
interface. The password defaults to 0000. Press the cycle key to move the cursor to
confirm, and click the confirm key to enter the menu interface. As shown in Figure 1:

图1
3.3.1 system setting description
Move the cursor to the corresponding icon and click OK to enter the setting

interface

9
3.3.2 equipment management description
Note: when multiple sensors are connected at the same time, the sensor addresses
shall not conflict, so as to avoid confusion when the configuration software views the
data curve.

10
3.3.3 description of viewing historical data
This operation can view the historical data on the device and import the data stored on
the device into the computer for viewing. Please refer to section 4.2 and 4.3 for
specific operations.

Chapter 4 instructions for configuration software

After the software is installed, it will be generated on the desktop“ ”Icon,


double click to open the software.
4.1 setting detection parameters

Connect the device to the computer through the USB data cable, open the
software and automatically read the device information. After editing the device
parameters, click "write device information" to write the parameters into the device.

11
It should be noted that after the modified parameter, check the modify parameter, and
then click write device information to write the parameter to the device. This function
is set to facilitate the wrong modification of other parameters when modifying
individual parameters.

Software version: factory default, can be viewed and cannot be modified.


Battery power: the current remaining power of the tester.
Number of items stored: the number of records currently stored by the tester (the
total number of records of all sensors).
Current time: check the clock time indicated by the tester to determine whether time
calibration is required.
Allow recording: set the storage function of the tester on or off.
Recording interval: set the recording interval of the tester (60 seconds by default).
Current record status: the open status of data record. It is closed by default.

4.2 importing data


Connect the recorder to the computer through USB cable, and then open the recorder
configuration software. The software will automatically read "read device
information", and then click "stored data reading" to import the data stored in the
tester into the configuration software.

12
Click data to view the stored historical data in time.

4.3 export data


Select the export format (txt / XLS / PDF) in the toolbar to save the data export to the
specified path (taking PDF as an example), and you will be prompted that the export
is successful.

13
4.4 clearing equipment data

Click "device" in the toolbar and click "clear device stored data" in the drop-down
menu to clear the device data. After clearing, the software prompts "data cleared
successfully".

Chapter 5 description of charging function

When the equipment is shut down, connect the charger to the recorder and the other
end to the AC220V AC power supply. After the detector is fully charged, the power
display is full and can be used normally.

Chapter 6 precautions

1. Prevent the machine from falling from a height or being subjected to severe
vibration.

2. Please use the recorder in strict accordance with the instructions, otherwise the test
results may be inaccurate or the product may be damaged.

Chapter 7 common faults and Solutions


Fault phenomenon Possible cause of failure Treatment method
The voltage is too low Please charge in time
Please contact your dealer or
Unable to power on crash
manufacturer for repair
Circuit failure Please contact your dealer or

14
manufacturer for repair
No response from Please contact your dealer or
Circuit failure
inserted sensor manufacturer for repair
Contact your dealer or manufacturer
Sensor failure
Inaccurate display
Replace the sensor
Long term uncalibration Please calibrate in time
The battery is completely Replace the RTC battery and reset the
discharged time
Time display error
Strong electromagnetic
Reset time
interference
When the instrument Contact your dealer or manufacturer
is tested normally Replace the sensor
Sensor failure
The interface displays
full scale

15