Professional Documents
Culture Documents
ZN5N1ME-02_ND-500
ZN5N1ME-02_ND-500
ZN5N1ME-02_ND-500
Nitrogen Detector
ND-500
Instruction Manual
CAUTION
To use the unit correctly, read this instruction manual before use.
Keep the manual at hand when operating the unit.
Nittoseiko Analytech
EC DECLARATION OF CONFORMITY
We hereby declare that the following equipment complies with the essential requirements
of:
_______________________
Name : Kaoru Hayashi
Position : General Manager
Date : 16 . 07 . 2021
DD.MM.YYYY
UK Declaration of Conformity (DoC)
Object of Declaration
Product name: Nitrogen Detector
Model name: ND-500
Manufacturer
Name: NITTOSEIKO ANALYTECH CO., LTD.
Address: 7-10-1 Chuo-rinkan, Yamato, Kanagawa 242-0007, Japan
(Authorized Representative)
(Name: a1-envirosciences Ltd )
(Address: Suite2, Ground Floor,910 Birchwood Boulevard, Birchwood
Warrington WA3 7QN United Kingdom )
We declare that the DoC is issued under our sole responsibility and that the object of the
declaration described above is in conformity with the following relevant UK legislation:
Electromagnetic Compatibility Regulations 2016
Electrical Equipment (Safety) Regulations 2016
The Restriction of the Use of Certain Hazardous Substances in Electrical
and Electronic Equipment Regulations 2012
The following UK designated standards and other technical specifications have been applied.
EN61000-6-4:2019
EN61000-6-2:2019
EN61010-1:2010/A1:2019
EN IEC 63000:2018
Signature: ____________________
Name: Kaoru Hayashi
Position / Title: General Manager
Introduction 1
Introduction
Thank you for purchasing our nitrogen detector Model ND-500. This unit measures
nitrogen in liquid and gas samples.
This manual is the instruction manual for the nitrogen detector Model ND-500. It also
includes the explanation of the high vacuum diaphragm-type dry vacuum pump DTU-20.
To use the unit efficiently and safely, read this manual and the attached instruction
manual from ULVAC KIKO and fully understand the functions and the operation methods
before use.
This instruction manual describes the parts of the unit, specifications, and guarantee
provisions.
The NSX-5000 system consists of an injector, an electric furnace, and a detector. The
units are combined according to each measurement. Read the instruction manual of each
component unit. For measurement operations, refer to the operation manual.
Some unit combinations may not be supported depending on the system configuration.
When changing the configuration of the units, you may need to update the application
software and the unit software.
Before Use
(3) Though this manual was prepared carefully, contact local distributor when mistakes,
omissions, or missing pages are found.
The application software in this manual may be different depending on the version
of the software.
(4) For any consequences of used results, we do not take any responsibility regardless
of (3).
(5) Assign specific operators and supervisors when using this unit.
(6) Always follow the contents in this manual. When problems or damages occur by
neglecting the contents in this manual, we do not guarantee the system even within
the guarantee period.
Nittoseiko Analytech has the copyrights of this manual and the unit software.
Microsoft and Windows are registered trademarks of Microsoft Corporation.
Other company and product names described in this manual are their trademarks and
registered trademarks.
ND-500 (NSX-5000)
2 Introduction
About Composition
About Indications
ND-500 (NSX-5000)
Introduction 3
About Terms
ND-500 (NSX-5000)
4 Introduction
ND-500 (NSX-5000)
SAFETY PRECAUTIONS 1
SAFETY PRECAUTIONS
Read this instruction manual,the operation manual, and the instruction manuals supplied
with parts carefully before use.
Always read safety precautions before use.
Security cannot be assured if the unit is operated without following the instructions in
this manual. Keep the manual and the instruction manuals supplied with parts at hand.
This instruction manual uses several indications to highlight critical precautions. The
items with these indications are important for the safe use of the unit. Always observe
them. The following indications are used for safety precautions.
WARNING
Information under this sign explains critical operations which contain the
risks of death or serious casualties caused by wrong handing.
CAUTION
Information under this sign explains critical operations which contain the
risks of human casualties or property damages caused by wrong handing.
Other Indications
Besides the safety precaution indications, this instruction manual uses the following
indication to show an important or useful thing for measurement with this unit. The items
with this indication are important for the safe use of the unit. Be sure to read them before
use.
POINT
Information under this sign explains important items and tips for
measurement with this unit.
ND-500 (NSX-5000)
2 SAFETY PRECAUTIONS
Labels
The following label is attached to the unit. Always follow the instructions on the labels.
Cautions
Read the following cautions carefully to use the unit correctly and safely.
Installation
ND-500 (NSX-5000)
SAFETY PRECAUTIONS 3
WARNING Humidity
Install the unit in a place where humidity is 80%RH or less with no
condensation.
Failure to do so may cause condensation, leading to a fire or electric shock.
Obtain a humidity environment free from condensation using an air
conditioner.
WARNING Dust
Install the unit in a place free from much dust.
Failure to do so may cause a fire or electric shock. Clean the area around
the unit regularly. If the inlet or outlet of the unit is blocked by dust, the
inside will become hot, which may cause a fault or significantly shorten the
life.
WARNING Fire
Install the unit in a place where fire is not used.
Failure to do so may lead to a fire.
WARNING Altitude
Install the unit in a place where the altitude is below 2000 meters to avoid
unstable operation, a fire, or an electric shock.
CAUTION Ventilation
Ventilate the room by an air conditioner, a ventilating fan, etc.
Failure to do so may damage health by the gas generated from chemicals
or exhaust gas from the unit.
ND-500 (NSX-5000)
4 SAFETY PRECAUTIONS
CAUTION Combustibles
Do not place combustibles around the unit.
If a fire occurs, more damage will be caused.
CAUTION Vibration
Install the unit in a place free from strong vibration or continuous weak
vibration. Failure to do so may cause unstable operation or a fault and lead
to an accident.
ND-500 (NSX-5000)
SAFETY PRECAUTIONS 5
CAUTION Transportation
When moving the unit, firmly hold it with hands and do not incline it.
Bumping the unit or falling it causes not only unit breakage but also injuries.
Usage
When performing measurement or other operations with the unit, observe the following
precautions.
ND-500 (NSX-5000)
6 SAFETY PRECAUTIONS
ND-500 (NSX-5000)
SAFETY PRECAUTIONS 7
ND-500 (NSX-5000)
8 SAFETY PRECAUTIONS
Operation
ND-500 (NSX-5000)
Table of Contents 1
Table of Contents
Introduction ...................................................................................................................... 1
Before Use ................................................................................................................ 1
About This Manual .................................................................................................... 2
SAFETY PRECAUTIONS ................................................................................................ 1
Indications of Safety Precautions .............................................................................. 1
Other Indications ....................................................................................................... 1
Warnings and Cautions............................................................................................. 2
Labels ............................................................................................................... 2
Cautions............................................................................................................ 2
Section 1: Names and Functions of the Unit................................................................. 1-1
1-1. Names and Functions of ND-500 Front .......................................................... 1-1
1-2. Names and Functions of ND-500 Rear ........................................................... 1-2
1-3. Names and Functions of ND-500 Inside ......................................................... 1-4
1-4. Names and Functions of DTU-20 Front .......................................................... 1-5
Section 2: Parts Check ................................................................................................. 2-1
2-1. ND-500 Parts .................................................................................................. 2-1
2-2. Parts for Vertical Furnace ............................................................................... 2-2
2-2-1. Parts of ND-500 Gas Dryer Set for Vertical Furnace ........................... 2-2
2-2-2. Parts of ND-500 Combustion Tube Set (Nonaqueous) for Vertical
Furnace ............................................................................................... 2-2
2-2-3. Parts of ND-500 Combustion Tube Set (Aqueous) for Vertical
Furnace ............................................................................................... 2-3
2-2-4. Parts of Manual Injection Port Set for ND-500 ..................................... 2-4
2-3. Parts of DTU-20 Vacuum Pump Set ............................................................... 2-4
2-4. Parts of Acid Gas Removal Set ...................................................................... 2-4
2-5. Other Parts ..................................................................................................... 2-5
Section 3: Installation ................................................................................................... 3-1
3-1. Installation Conditions..................................................................................... 3-1
3-2. Installation Space ........................................................................................... 3-1
3-3. Power Preparation .......................................................................................... 3-2
3-3-1. Power .................................................................................................. 3-2
3-3-2. Grounding ............................................................................................ 3-2
3-4. Cable Connection ........................................................................................... 3-3
ND-500 (NSX-5000)
2 Table of Contents
ND-500 (NSX-5000)
Table of Contents 3
ND-500 (NSX-5000)
4 Table of Contents
ND-500 (NSX-5000)
Names and Functions of the Unit 1-1
Fig. 1-1 shows the ND-500 front and Table 1-1 shows the names and functions.
ND-500 (NSX-5000)
1-2 Names and Functions of the Unit
Fig. 1-2 shows the ND-500 rear and Table 1-2 shows the names and functions.
1
2
3
9
7
10
11
4
5
6
ND-500 (NSX-5000)
Names and Functions of the Unit 1-3
ND-500 (NSX-5000)
1-4 Names and Functions of the Unit
Fig. 1-3 shows the ND-500 inside and Table 1-3 shows the names and functions.
2 4
3
ND-500 (NSX-5000)
Names and Functions of the Unit 1-5
Fig. 1-4 shows the DTU-20 front and Table 1-4 shows the names and functions.
4 3 6
ND-500 (NSX-5000)
1-6 Names and Functions of the Unit
ND-500 (NSX-5000)
Parts Check 2-1
ND-500 (NSX-5000)
2-2 Parts Check
Table 2-2 shows the parts of the ND-500 gas dryer set for the vertical furnace.
Table 2-2. Parts List of ND-500 Gas Dryer Set for Vertical Furnace
Table 2-3. shows the parts of the ND-500 combustion tube set (nonaqueous) for vertical
furnace.
Table 2-3. Parts list of ND-500 Combustion Tube Set (Nonaqueous) for Vertical
Furnace
ND-500 (NSX-5000)
Parts Check 2-3
Table 2-4. shows the parts of the ND-500 combustion tube set (aqueous) for vertical
furnace.
Table 2-4. Parts list of ND-500 Combustion Tube Set (Aqueous) for Vertical Furnace
ND-500 (NSX-5000)
2-4 Parts Check
Table 2-5 shows the parts list of the manual injection port set for ND-500. Only aqueous
samples can be injected manually. A microsyringe is required additionally.
Table 2-5. Parts List of Manual Injection Port Set for ND-500
Table 2-6 shows the parts list of the DTU-20 vacuum pump set. DTU-20 is as per our
special specifications.
Table 2-7 shows the parts list of the acid gas removal set. When acid gas is generated
through sample combustion, use this acid gas removal set to remove the acid gas before
it is injected into the detector.
ND-500 (NSX-5000)
Parts Check 2-5
ND-500 (NSX-5000)
2-6 Parts Check
ND-500 (NSX-5000)
Installation 3-1
Section 3: Installation
This section describes necessary space, power, and gas lines for ND-500 installation.
It also describes the installation with the acid gas removal set.
Install the unit indoors to meet the following conditions for the long-term stable use. Refer
to “SAFETY PRECAUTIONS” in this manual for details.
• Room temperature is 15 to 35 degrees Celsius. Temperature change is small.
• The unit is not exposed to direct wind from an air conditioner.
• Humidity is under 80%RH. (No condensation)
• No much dust
• No direct sunlight
• Fire is not used.
• The room is ventilated.
• No corrosive gas
• No flammable materials around the unit
• No strong vibration and continuous weak vibration
• No strong electromagnetic field
• The surface for installation is flat and stable.
• Altitude is less than 2000 meters.
ND-500 (NSX-5000)
3-2 Installation
3-3-1. Power
3-3-2. Grounding
The ND-500 power cable is a 3-line type cable including a grounding wire. Prepare either
of the following appropriate outlets with ground on the table, and securely ground the
power cable. Select an outlet with ground according to the power supply voltage in your
area. If you cannot prepare the following outlet with ground, prepare a grounded outlet
and a grounded cable that satisfy the rating and unit power consumption conditions.
• 100/115 V Rating: 125 VAC, 15 A (JIS C 8303)
• 230/240 V Rating: 250 VAC, 16 A (SE type)
ND-500 (NSX-5000)
Installation 3-3
Connect the communication cable connector (MAIN UNIT) on the rear of the ND-500 unit
and one of communication cable connectors (ACCESSORIES) 1 to 8 on the rear of the
electric furnace with the attached communication cable.
Connect the attached power cable to the power connector (a.c. POWER) on the rear of
the ND-500 unit.
Connect the vacuum pump power cable to the ND-500 rear pump outlet (VACUUM
PUMP).
For ND-500, you need to prepare O2 gas for ozone generation and as dry gas. The gas
line prepared for the electric furnace will be used for O2 gas by branching it, so there is
no need to prepare new gas supply lines.
ND-500 (NSX-5000)
3-4 Installation
ND-500 (NSX-5000)
Installation 3-5
Branch the O2 gas supplied to the gas purification filter and connect a line to the ND-500
unit by the following procedure while referring to “Fig. 3-1. Line Connection”. Refer to
“Details of A” to connect the fluororesin tubes.
(1) Connect the stainless tube of the 2. three-way connector to the 1. reducer above
the gas purification filter to which O2 gas is supplied to branch O2 gas.
(2) Connect the 3. black fluororesin tube 4/2, 0.9 m, which is the O2 gas line to the
electric furnace, to the 2. three-way connector and connect the opposite end to the
oxygen gas inlet (GAS IN O2) on the rear of the electric furnace.
(3) Connect the 4. fluororesin tube 4/2, 1.6 m, which is the O2 gas supply line, to 2.
three-way connector and connect the opposite end to the oxygen gas inlet (GAS IN
O2) on the rear of the ND-500 unit.
ND-500 rear
2. A 3. 4.
Gas
purification
Details of A
Front ferrule Nut
Ar gas
Insert
O2 gas Back ferrule Fluororesin tube
ND-500 (NSX-5000)
3-6 Installation
The connection of the ND-500 unit and the vertical furnace is described. For the
procedure for setting a combustion tube, refer to the instruction manual of the vertical
furnace.
The procedure for connecting combustion tube outlet and gas tube dryer lines is
described. Connect the combustion tube outlet and the gas tube dryer by the following
procedure while referring to “Fig. 3-2. Line Connection of Combustion Tube Outlet and
Gas Tube Dryer”.
(1) Open the door of the electric furnace and pass the 1. gas tube dryer though the
detector connection hole of the electric furnace. Let out about the two-thirds of the
gas tube dryer toward the rear.
(2) Set 2. 6/4 quick joint on the outer tube connector of the gas tube dryer in the electric
furnace.
(3) Connect 3. fluororesin tube 4/2, 0.9 m to 2. 6/4 quick joint, pass it through the
detector connection hole of the electric furnace, and let it out toward the rear.
(4) Set 4. fluororesin tube 6/4, 30 mm to the gas tube dryer end and set 5. connector
6 (For aqueous samples, set the L-shaped joint 6.)
(5) Set 6. female ball joint 18 (or trap tube 18 of the combustion tube bottom for
aqueous samples) to 5. connector 6, connect it to the combustion tube outlet, and
fix them with 7. clip 18 for ball joint.
1.
3.
2. 4. 5. 6. 7.
Fig. 3-2. Line Connection of Combustion Tube Outlet and Gas Tube Dryer
ND-500 (NSX-5000)
Installation 3-7
The procedure for connecting gas tube dryer and buffer tank lines is described. Connect
the gas tube dryer, buffer tank, and the ND-500 unit by the following procedure while
referring to “Fig. 3-3. Line Connection of Gas Tube Dryer and Buffer Tank”.
(1) Set 1. fitting to the rear of the electric furnace and fix 2. gas tube dryer, which
protrudes from the detector connection hole, to 3. fluororesin tube 4/2, 0.9 m.
(2) Set 4. 6/4 quick joint with socket to the outer tube connector of the fixed gas tube
dryer.
(3) Connect “4” to 5. fluororesin tube protruding from the oxygen gas outlet (EXHAUST
O2) on the rear of the ND-500 unit.
(4) Set 6. fluororesin tube 6/4, 30 mm to the gas tube dryer end and set 7. 6/4 L
shape joint.
(6) Diagonally cut one end of 9. fluororesin tube 4/2, 0.45 m and set it to the pass-
through side of the buffer tank. When setting lines to the buffer tank, use the sleeve
inside the nut to fix it firmly. The tube end should be 5 mm away from the top surface
of the buffer tank not to clog the end. Connect the other end of 9. fluororesin tube
4/2, 0.45 m to 7. 6/4 L shape joint.
(7) Connect 10. fluororesin tube 4/2 protruding from 10. sample gas inlet (GAS IN
SAMPLE) on the rear of the ND-500 unit to the non-pass-through side of the buffer
tank.
ND-500 (NSX-5000)
3-8 Installation
6. 7. 4. 9.
Cut diagonally
1. 8.
2.
3. 10.
5.
Fig. 3-3. Line Connection of Gas Tube Dryer and Buffer Tank
ND-500 (NSX-5000)
Installation 3-9
The setting of the ozone scrubber is described. Set the ozone scrubber behind the front
door of the unit by referring to “Fig. 3-4. Ozone Scrubber Setting”.
(1) Open the front door of the ND-500 unit, set 1. ozone scrubber exhaust inlet (thin)
upward, and set the ozone scrubber to the setting metals.
(2) Connect 2. thin silicon tube behind the front door of the ND-500 unit to the thin
exhaust inlet of the ozone scrubber. Fix it with the small hose band.
(3) Connect 3. thick silicon tube behind the front door of the ND-500 unit to the thick
exhaust outlet of the ozone scrubber. Fix it with the large hose band.
2.
1.
3.
ND-500 (NSX-5000)
3-10 Installation
The connection of ND-500 and DTU-20 is described. Connect the ND-500 unit and the
DTU-20 vacuum pump by the following procedure while referring to “Fig. 3-5. Vacuum
Pump Connection”.
(1) Fix 2. suction tube for vacuum pump to 1. connector at the end of the tube protruding
from the pump suction port (VACUUM PUMP) on the rear of the ND-500 unit with 3.
large hose band.
(2) Cut the suction tube for vacuum pump to a proper length and connect it to 4. suction
tube connection port of the DTU-20 vacuum pump, and fix it with 5. hose band.
1.
6. 4. 5.
3.
2.
ND-500 (NSX-5000)
Installation 3-11
By connecting the male ball joint to the exhaust line of ND-500, a gas leak of the entire
system can be checked. For simultaneous measurement of nitrogen/sulfur, set the male
ball joint to the SD-500 exhaust line. Refer to “Fig. 3-6. Setting of Male Ball Joint” and set
the male ball joint to the ND-500 exhaust line as follows.
(2) Connect 3. fluororesin tube 4/2 coming out from sample gas outlet (EXHAUST
SAMPLE) and 2. fluororubber connector 6/3.
1
2
ND-500 (NSX-5000)
3-12 Installation
The acid gas removal set is an apparatus designed to remove acid gas, which is an
interference element when a sample is combusted and nitrogen is detected by the ND-
500 unit. Use this set when measuring samples such as seawater samples and samples
with much sulfur content.
Assemble the sample bottle of the acid gas removal set by the following procedure while
referring to “Fig. 3-7 Assembly of Sample Bottle”.
(1) Cut the fluororesin tube 4/2, 0.9m in half and set 1. fluororubber connector 6/3 to
each end.
(2) Connect “1” to IN (where the blowing tube is extended downward) and OUT on 2.
sample bottle cap.
(3) Insert “2” into 3. sample bottle and fix it with 4. sample bottle clamp.
1.
IN OUT
2.
4.
3.
ND-500 (NSX-5000)
Installation 3-13
The connection of the assembled sample bottle to the combustion tube of the vertical
furnace is described. Connect the sample bottle between the combustion tube and the
gas tube dryer by the following procedure while referring to “Fig. 3-8 Connection of Acid
Gas Removal Set to Vertical Furnace”.
(1) Connect the combustion tube outlet and the gas tube dryer by referring to “3-6-1.
Line Connection of Combustion Tube Outlet and Gas Tube Dryer”.
(2) Install 1. sample bottle holder on the left side of the vertical furnace and set 2.
sample bottle assembled in “3-9-1. Assembly of Sample Bottle”.
(3) Set 3. connector 6/4 to the fluororesin tube 4/2 connected to IN (on the blowing
tube side) on the sample bottle and put the tube in the electric furnace through 4.
hole on the left side of the vertical furnace.
(4) Set 5. 6/4 L shape joint to the fluororesin tube 4/2 connected to OUT on the
sample bottle and put the tube in the electric furnace through 4.
(5) Remove the connector 6 from 6. female ball joint 18 and connect “3”.
(6) Remove the connector 6 from 7. fluororesin tube 6/4, 30 mm connected to the
gas tube dryer end and connect “5”.
(7) Connect “6” to the combustion tube outlet, and fix it with 8. clip 18 for ball joint.
4. 3. 6. 8.
2.
1. 7. 5.
ND-500 (NSX-5000)
3-14 Installation
ND-500 (NSX-5000)
Measurement 4-1
Section 4: Measurement
This section describes the principle and parameters for nitrogen measurement with the
ND-500 unit. For operation as a system, read the operation manual. Measurement with
the acid gas removal set is also described.
This section describes the sulfur measurement principle for the vertical furnace. Fig. 4-1
is the measurement principle diagram using the vertical furnace.
Argon gas whose flow is controlled by the mass flow controller is used as the carrier gas.
A sample is injected by the argon carrier gas into a high-temperature (600 to 1100
degrees Celsius) combustion tube at a constant rate. Most of the nitrogen compound in
the sample changes into NO through pyrolysis and combustion oxidation by O2 gas
supplied to the outer combustion tube.
After sample injection, the change valve changes the gases of the inner and outer
combustion tubes to completely combust the sample in the inner combustion tube with
O2 gas.
Generated NO passes through the gas tube dryer to remove moisture and reacts with
ozone in the cell. Then, the following oxidation reaction occurs and an emission of 590
to 2500 nm is observed.
NO + O3 → NO2 + O2 + h
This emitted light is received by the photomultiplier tube (PMT) and AREA value is
obtained. Using the relation between the area value and the absolute volume
(calibration curve) obtained from calibration curve solutions, the nitrogen content of the
sample is calculated.
ND-500 (NSX-5000)
4-2 Measurement
PC Controlled
injection Mass flow
Buffer tank controller
Printer
Ar
Combustion
tube
O2
Change
Electric valve
Photomultiplier furnace
tube Cell
ND-500 (NSX-5000)
Measurement 4-3
Due to the detection principle, any substance other than NO that reacts with ozone to
generate chemiluminescence gives positive interference and any substance that
removes excitation energy from excited NO2 gives negative interference.
(1) Carbonyl compounds of iron and nickel react with ozone to generate
chemiluminescence, which gives positive interference. However, it is unlikely that
these substances exist in sample gas after combustion. Therefore, there are no
practical problems.
(2) Olefins (such as ethylene and trimethyl ethylene) and sulfide compounds (such as
dimethyl sulfide and hydrogen sulfide) react with ozone to generate
chemiluminescence, which gives positive interference. However, the wavelengths
of these substances are different from the wavelength of chemiluminescence
caused by NO and the ND-500 unit uses optical filters to detect only arbitrary
wavelengths. Therefore, the wavelengths of these substances are not detected
basically.
(3) A sample that contains chlorine produces chlorine gas through combustion and
may give positive interference. It is necessary to use the acid gas removal set to
trap chlorine gas.
(4) Moisture in sample gas gives negative interference. This system uses a gas tube
dryer to remove moisture from the sample gas after combustion.
(5) Carbon dioxide in sample gas gives negative interference at a fixed rate if the
concentration is stable. However, by the reduced pressure method, the pressure
within the chemiluminescence cell is reduced and intervention in molecules is
decreased. Therefore, the negative interference of carbon dioxide can be
decreased to an irrelevant level.
ND-500 (NSX-5000)
4-4 Measurement
To use new platinum catalyst, oxidation treatment needs to be done before use for
measurement. Fill it into the combustion tube, and do oxidation treatment by the
following procedure.
(1) Fill new quartz wool and platinum catalyst into the combustion tube by referring to
the operation manual.
(2) Set the combustion tube by referring to the instruction manual of the electric
furnace.
(3) Without connecting the female ball joint 18 or trap tube 18 of the combustion
tube bottom, prebake the tube for about 5 hours with the following settings.
Electric furnace Inlet: 800 degrees Celsius, Outlet: 800 degrees Celsius, Ar gas:
300 mL/min, O2 gas: 300 mL/min
ND-500 (NSX-5000)
Measurement 4-5
The acid gas removal set is an apparatus designed to remove acid gas, which is an
interferer when a sample is combusted and nitrogen is detected by the ND-500 unit.
Use this set when measuring samples such as seawater samples and samples with
much sulfur content.
4-4-1. Preparations
This section describes preparations of the acid gas removal set for measurement. While
the temperature of the electric furnace has not risen, set the sample bottle between the
combustion tube and the dryer tube by referring to “3-9-1. Assembly of Sample Bottle”
and “3-9-2. Connection (Vertical Furnace)”. Make preparations before measurement by
the following procedure.
(1) Dissolve 0.42 g of sodium hydrogen carbonate (NaHCO3) with pure water to be 50
mL of 0.1 mol/L sodium hydrogen carbonate solution.
(2) Remove the two fluororubber connectors 6/3 connected to the sample bottle cap,
and remove the sample bottle from the sample bottle holder.
(3) Remove the sample bottle clamp and cap, pour 50 mL of 0.1 mol/L sodium
hydrogen carbonate solution into the sample bottle, and set the sample bottle cap
and clamp.
(4) Set the assembled sample bottle to the sample bottle holder, and connect the
fluororubber connectors 6/3, which are supplied with gas from the electric furnace,
to the reagent bottle. For details on connection, refer to “3-10-2. Connection
(Vertical Furnace)”.
(5) Supply gas and check that air bubbles are generated continuously and properly in
the sample bottle.
(6) Connect the fluororubber connectors φ6/3, which supplies gas to the tube dryer, to
the reagent bottle. For details on connection, refer to “3-10-2. Connection (Vertical
Furnace)”.
ND-500 (NSX-5000)
4-6 Measurement
If the system is left with no gas flow to the electric furnace, the sodium hydrogen
carbonate solution in the sample bottle may flow backward. After measurement, perform
shutdown operations on the acid gas removal set by the following procedure.
(1) While gas is flowing to the electric furnace, remove the two fluororubber
connectors 6/3 connected to the sample bottle cap, and remove the sample bottle
from the sample bottle holder.
(2) Remove the sample bottle clamp and cap, and dispose of the sodium hydrogen
carbonate solution in the sample bottle.
(3) Wash the sample bottle and the tip of the sample bottle cap with tap water, and
then rinse them with pure water.
(4) Set the sample bottle cap and clamp to the empty sample bottle, set the sample
bottle on the sample bottle holder, and connect the two fluororubber connectors
6/3 to restore the original state. For details on connection, refer to “3-10-2.
Connection (Vertical Furnace)”.
(5) After checking that the sample bottle is empty, stop gas flow.
ND-500 (NSX-5000)
Measurement 4-7
4-5. Parameters
There are ND-500 parameters set in a method or a schedule. For parameters which are
set in a method, they are reflected into the whole schedule. For parameters which are
set in a schedule, they are reflected every step.
This section describes ND-500 parameters which are set in a method. Table 4-1 lists
the parameters. For details on how to check parameters, refer to the operation manual.
The parameters do not have to be reset every time. Some parameters may not be used
depending on the system. They are not displayed. PMT sensitivity and detection mode
are set for each step. For details on how to set, refer to the operation manual.
ND-500 (NSX-5000)
4-8 Measurement
Parameter Details
(1) O2 flow
Set O2 gas flow for ozone generation and dry gas. When sample volume is large,
increase O2 gas flow used as dry gas. Table 4-2 shows sample volume and O2
gas flow settings. In the case of aqueous samples, the maximum sample volume
is 100 µL.
ND-500 (NSX-5000)
Measurement 4-9
1.
Detection start 5 seconds
waiting time
ND-500 (NSX-5000)
4-10 Measurement
1.
Integration start
wait time
2. Measurement time
ND-500 (NSX-5000)
Measurement 4-11
This section describes ND-500 parameters which are set in a schedule. PMT sensitivity
and a detection mode can be set in a schedule every measurement step. The details of
PMT sensitivities and detection modes are as follows. Refer to the operation manual for
setting methods.
(1) PMT sensitivity
Select the sensitivity of the detector according to the concentration range for
measurement. “High sensitivity” is a setting for the measurement of trace range
and “Low sensitivity” is a setting for measurement of higher-concentration range. It
may be necessary to select not only the PMT sensitivity but also an appropriate
sample volume, depending on the concentration range. Table 4-3 shows PMT
sensitivity and sample volume settings. Further dilution may be necessary
because the optimal sample volume may vary depending on the unit configuration,
the sample matrix, and other factors. To measure gas, select an appropriate PMT
sensitivity and sample volume based on the absolute volume (nitrogen content).
Liquid
Electric Nitrogen Absolute
PMT sensitivity Sample
furnace concentration volume (ng)
volume (μL)
(μg/mL)
0.03 to 50 100
High 3 to 5000
0.2 to 250 20
Vertical
5 to 1000 20
furnace
Low 10 to 2000 10 100 to 20000
50 to 10000 2
(2) Mode
To detect start and end points of signal processing, select “Auto” or “Timer”. When
measuring samples in a trace range, select “Timer” because the peak start and
end sometimes may not be detected automatically.
Auto: Peak is automatically detected by judging the start and end points based on
the slope of signal strength.
Timer: Peak is detected by processing signals with set start and end times.
ND-500 (NSX-5000)
4-12 Measurement
ND-500 (NSX-5000)
Troubleshooting 5-1
Section 5: Troubleshooting
This section describes the troubleshooting of the ND-500 unit. When problems are not
solved even by taking countermeasures, contact our local distributor.
Cause Countermeasure
□ The power cable is Connect the power cable firmly to the power
disconnected from the connector.
power connector.
□ The power plug is not Connect the power plug firmly to the outlet.
inserted properly into the
outlet.
□ Power is not supplied to Check that power is supplied to the outlet by using
the outlet. another instrument.
□ The power fuse is blown. Take off the fuse holder of the unit with a flathead
screwdriver. Check the conduction between both ends
with a tester. If there is no conduction, change the
fuse with new one.
ND-500 (NSX-5000)
5-2 Troubleshooting
Cause Countermeasure
□ The power of the unit is off. Turn on the power of the unit by referring to “5-1. No
Power to the Unit”.
□ The USB cable is Connect the USB cable firmly to the communication
disconnected from the connector of the electric furnace.
communication connector
of the electric furnace.
□ The communication cable Connect the communication cable firmly to the
is disconnected from the communication cable connector.
communication cable
connector.
□ The USB or Change the USB or communication cable with new
communication cable is one.
broken.
□ Communication is stopped. Restart the whole system. When communication does
not recover even by restarting the system, contact our
local distributor.
□ ND-500 is not selected in Select ND-500 in the equipment configuration in the
the equipment application software.
configuration in the
application software.
Cause Countermeasure
□ The gas cylinder valve is Open the gas cylinder valve.
closed.
□ Remaining gas is Change the gas cylinder.
insufficient.
□ The gas supply valve is Open the gas supply valve.
closed.
□ The pressure of supplied Adjust the pressure of supplied gas appropriately.
gas is inappropriate.
□ Lines are disconnected. Connect the lines firmly.
□ A line is clogged. Check for any clogging in the lines. Replace any
clogged parts.
□ The power switch is off. Turn on the power switch.
□ The gas status in the PC Turn on the gas status in the PC window.
window is off.
ND-500 (NSX-5000)
Troubleshooting 5-3
Cause Countermeasure
□ No schedule is set to Click [Prepare] in a schedule to set the schedule to
“Measurement schedule”. “Measurement schedule”. Parameters are transmitted
into the unit and the unit is ready after stabilization.
□ All steps in the schedule Add a new step into the schedule.
are already performed.
□ There is no communication Select “File” - “Connect” of the menu to reconnect to
with the electric furnace. it.
□ O2 gas is not supplied. Supply O2 gas.
□ The ozonizer is off. Turn on the ozonizer.
□ The cell temperature is not Wait until the temperature of the cell rises to 36 to 40
stable. C.
□ The cooler temperature is Wait until the temperature of the cooler is cooled to 2
not stable. to 6 C.
□ The baseline is unstable. Wash glass parts, such as the combustion tube and
□ The baseline is high. the female ball joint, and change the quartz wool and
filler. If the baseline is still unstable, change the gas
purification filter and check the baseline.
When there is no improvement even by taking the
countermeasures above, contact our local distributor.
5-5. No Peak
Cause Countermeasure
□ The temperature of the Check “Temp./Flow” in the PC window, and check
electric furnace has not that the temperature of the electric furnace has risen
risen. to the setting temperature. If the electric furnace is off,
turn on it.
□ The combustion tube and Connect the combustion tube and the female ball joint
the female ball joint are not with the clip 18 (long) for ball joint.
connected.
□ Samples are not injected Check the condition of the injector and syringe.
correctly.
□ Carrier gas is leaking. Check for gas leakage. Refer to “Gas Leak Check” of
the operation manual for details.
□ Sample concentration is Check the peak with a standard sample.
too low.
□ The vacuum pump is not Check that the power switch of the vacuum pump is
operating. on.
ND-500 (NSX-5000)
5-4 Troubleshooting
Causes Countermeasures
□ The pressure of supplied Adjust the pressure of supplied gas appropriately.
gas is inappropriate.
□ The gas flow setting is Set an appropriate gas flow.
inappropriate.
□ Impurities are contained in If the problem occurs after changing the gas cylinder,
gas. change it with another one.
□ The temperature setting of Set an appropriate temperature for the electric
the electric furnace is furnace.
inappropriate.
□ The injector setting is Set an injector appropriately.
inappropriate.
□ Gas is leaking from the Check for gas leakage from the combustion tube
combustion tube connection part. Refer to “Gas Leak Check” of the
connection part. operation manual for details.
□ The combustion tube is Wash the combustion tube. After washing, rinse the
contaminated. combustion tube thoroughly with pure water and dry it
before use.
□ Quartz wool or filler in the Change the quartz wool or filler.
combustion tube is
deteriorated.
□ The inside of the detector While the ozonizer remains operating, flow O2 gas at
is contaminated. 200 mL/min for one day and wash the inside.
When there is no improvement, contact our local
distributor.
□ The combustion tube and Connect the combustion tube outlet and the female
the female ball joint are not ball joint firmly with the clip 18 (long) for ball joint.
connected properly.
□ The disk filter is Change it with a new one. When the connected
contaminated. fluororesin tube is contaminated, wash it or change it.
ND-500 (NSX-5000)
Troubleshooting 5-5
Cause Countermeasure
□ “Auto” mode is used for If trace measurement is run in “Auto” mode, the
trace measurement. measurement end may be difficult to detect. Select
“Timer” mode for trace measurement.
□ Sample concentration is Set “PMT sensitivity” to “Low”. When “Low” is already
high. selected, the measurement range may be exceeded.
Reduce the sample volume or dilute the sample
before remeasurement.
ND-500 (NSX-5000)
5-6 Troubleshooting
ND-500 (NSX-5000)
Error Messages and Measures 6-1
ND-500 (NSX-5000)
6-2 Error Messages and Measures
ND-500 (NSX-5000)
Error Messages and Measures 6-3
ND-500 (NSX-5000)
6-4 Error Messages and Measures
ND-500 (NSX-5000)
Maintenance and Inspection 7-1
Table 7-1 shows the daily inspection items. If any problems are found, solve the
problem and re-check the unit. When problems cannot be solved or they may be inside
the unit, contact our local distributor.
The details of each daily inspection item and maintenance are as follows.
ND-500 (NSX-5000)
7-2 Maintenance and Inspection
Check the female ball joint φ18 for soot contamination caused by incomplete
combustion. If it is contaminated, wash the female ball joint φ18 as follows.
(1) Wipe off contamination on the female ball joint φ18 with absorbent cotton
moistened with toluene.
(2) Mix acetone with pure water in the proportion of one to ten to prepare washing
solution, and immerse the part in the solution for 12 hours or more.
(3) Thoroughly rinse the part with pure water and fully dry it.
Check of the Trap Tube of the Combustion Tube Bottom (Aqueous Sample
Measurement)
Visually check that water has not accumulated in the trap tube φ18 of the combustion
tube bottom and that the tube is not contaminated. If water has accumulated, dispose of
the water. Then, thoroughly rinse the part with pure water and fully dry it. If it is
contaminated, remove the contamination. Then, thoroughly rinse the part with pure
water and fully dry it.
The color of the inner tube of the gas tube dryer is translucent at the start of use.
However, it gradually turns brown as the number of measurements increases. If the
inner tubes are significantly discolored by incomplete combustion, they cannot be used
anymore. During daily inspection, visually check that the color of the inner tube of the
gas tube dryer has not turned dark brown or been significantly discolored. When it is
dark brown or significantly discolored, change the dryer with new one.
Also visually check that the inner tube of the gas tube dryer is not clogged. A clogged
inner tube may cause incomplete combustion. Change the dryer with new one.
Change the gas tube dryer as follows.
(1) Disconnect the φ6/4 two fluororesin tubes, 30 mm from the outlet and inlet of the
gas tube dryer.
(2) Disconnect the φ6/4 two quick joints from the gas tube dryer.
(3) Connect the disconnected parts to a new gas tube dryer and assemble the lines to
restore the original state. For the installation of a gas tube dryer, refer to “3-6.
Lines for Vertical Furnace”.
ND-500 (NSX-5000)
Maintenance and Inspection 7-3
Check the disk filter inside the ND-500 unit for contamination. When it is contaminated,
change it with a new one.
Check the vacuum pump for abnormalities. If the connected suction tube or the
connector is loose, firmly connect it. If the pump emits abnormal noise or odor, contact
our local distributor.
ND-500 (NSX-5000)
7-4 Maintenance and Inspection
Table 7-2 shows the periodical inspection items. If any problems are found, solve the
problem and re-check the unit. When problems cannot be solved or they may be inside
the unit, contact our local distributor.
ND-500 (NSX-5000)
Maintenance and Inspection 7-5
The details of each periodical inspection item and maintenance are as follows.
Visually check the unit for contamination and dust every two weeks. If there is
contamination and dust, clean the unit.
(1) Mix acetone with pure water in the proportion of one to ten to prepare washing
solution, and immerse the female ball joint φ18 in the solution for 12 hours or more.
(2) Thoroughly rinse the part with pure water and fully dry it.
Washing of the Trap Tube of the Combustion Tube Bottom (Aqueous Sample
Measurement)
Wash the trap tube φ18 of the combustion tube bottom every month as follows.
(1) Wash the trap tube φ18 of the combustion tube bottom with tap water.
(2) Thoroughly rinse the part with pure water and fully dry it.
ND-500 (NSX-5000)
7-6 Maintenance and Inspection
Measure the calibration curve solution every three to four weeks and check the area
value. If the area value sharply changes, change the platinum catalyst with a new one
or regenerate it. If there is no improvement after regeneration, change the catalyst with
a new one. Change the platinum catalyst with a new one or regenerate it as follows.
(1) Take out the combustion tube from the electric furnace and remove the quartz
wool and platinum catalyst.
(2) Thoroughly rinse the combustion tube with pure water and fully dry it.
(3) Fill new quartz wool and platinum catalyst into the combustion tube by referring to
the operation manual.
(4) Set the combustion tube to the electric furnace by referring to the instruction
manual of the electric furnace.
(5) Without connecting the female ball joint φ18 or trap tube of the combustion tube
φ18 bottom, prebake the tube for about 5 hours with the following settings. Electric
furnace Inlet: 800 degrees Celsius, Outlet: 800 degrees Celsius, Ar gas: 300
mL/min, O2 gas: 300 mL/min
(1) Remove the combustion tube from the electric furnace and transfer the platinum
catalyst into another container.
(2) Wash the platinum catalyst with pure water several times while being careful not to
crush it.
ND-500 (NSX-5000)
Maintenance and Inspection 7-7
(4) Fill quartz wool and the regenerated platinum catalyst into the combustion tube by
referring to the operation manual. Fill new quartz wool.
(5) Set the combustion tube to the electric furnace by referring to the instruction
manual of the electric furnace.
(6) Without connecting the female ball joint φ18 or the trap tube φ18 for combustion
tube bottom, set a glass beaker under the combustion tube outlet. The glass
beaker prevents water drops from dropping onto the unit when the platinum
catalyst is not fully dry.
(7) Set the electric furnace temperature and gas flow to the values used for usual
measurement and prebake the tube for three to four hours.
Check the condition of the buffer tank every month. When the buffer tank is
contaminated, wash it as follows. When an O-ring is damaged or twisted, change it with
new one.
(1) Detach the buffer tank from the ND-500 unit, loosen the nut, and disconnect the
fluororesin tube φ4/2.
(2) Remove the buffer tank cap from the body. Refer to “Fig. 7-1. Cap Removal”. The
buffer tank cap is fixed so that gas does not easily leak. Hold the body tightly,
apply a little power, and move the cap gradually.
ND-500 (NSX-5000)
7-8 Maintenance and Inspection
(3) Carefully wipe the cap rear, the O-ring, and the inside of the body with water or
ethanol-soaked paper for laboratory use.
(4) When an O-ring is damaged or twisted, change it with new one. Hitch the tweezers
tip to the O-ring of the cap rear to remove the O-ring by referring to “Fig. 7-2. O-
ring Removal”, and set a new one.
(5) Carefully wipe water or ethanol remaining in the surface with new paper and dry it
naturally.
(6) After drying, set the body upside down, and assemble the parts tightly by evenly
and closely pushing the cap edge not to break the connector of the cap. Refer to
“Fig. 7-3. Cap Setting”.
ND-500 (NSX-5000)
Maintenance and Inspection 7-9
(7) Set the buffer tank to the unit. For the setting, refer to “3-6. Lines for Vertical
Furnace”.
After about 800 hours or 6 months from the start of usage, change the ozone scrubber.
Refer to “3-7. Ozone Scrubber Setting”.
After about 6000 hours from the start of usage, overhaul the vacuum pump. An overhaul
of the vacuum pump is conducted by our local distributer. For details, contact our local
distributor.
After about 15000 hours from the start of usage, change the bearing of the vacuum
pump. Our local distributer takes off the vacuum pump and changes the bearing. For
details, contact our local distributor.
ND-500 (NSX-5000)
7-10 Maintenance and Inspection
When measuring samples that contain much salt using the acid gas removal set,
conduct the periodical inspection items for the acid gas removal set, which are shown in
Table 7-3, in addition to “7-1. Daily Inspection and Maintenance” and “7-2. Periodical
Inspection and Maintenance”. If any problems are found, solve the problem and re-
check the unit. When problems can not be solved or they may be in the unit, contact our
local distributor.
Table 7-3. Periodical Inspection Items for Acid Gas Removal Set
* The frequency of change varies depending on the measurement time, the frequency
of use, and sample volume.
ND-500 (NSX-5000)
Maintenance and Inspection 7-11
The details of each periodical inspection item and maintenance are as follows.
When measuring samples that contain much salt, such as seawater, maintain the
combustion tube and filler once in every two days. Since the maintenance frequency is
high, we recommend that you prepare two or more sets of combustion tubes and
platinum catalysts and use them alternately. The maintenance procedure is as follows.
(1) Take out the combustion tube from the electric furnace and remove the quartz
wool and platinum catalyst.
(3) Thoroughly rinse the combustion tube with tap water and then pure water and fully
dry it.
(1) Mix acetone with pure water in the proportion of one to ten to prepare washing
solution, and wash the female ball joint φ18.
(2) Thoroughly rinse the part with pure water and fully dry it.
Washing of the Trap Tube of the Combustion Tube Bottom (Aqueous Sample
Measurement)
Wash the trap tube φ18 of the combustion tube bottom every month as follows.
(1) Wash the trap tube φ18 of the combustion tube bottom with tap water.
(2) Thoroughly rinse the part with pure water and fully dry it.
ND-500 (NSX-5000)
7-12 Maintenance and Inspection
Store the ND-500 unit as follows when it is not used for long time.
The cautions for storing the unit on a table are as follows.
• Disconnect the power cable, gas lines, and acid gas removal kit lines.
• Remove the connection cable connecting the unit to the electric furnace.
• Remove the both lines of the gas inlet and outlet of the gas purification filter and
put the closing plugs.
• Seal tube ends with parafilm and store them to avoid the inside contamination.
• Put the cover over the whole unit.
ND-500 (NSX-5000)
Disposal of the Unit 8-1
ND-500 (NSX-5000)
8-2 Disposal of the Unit
ND-500 (NSX-5000)
Specifications 9-1
Section 9: Specifications
This section describes ND-500 and DTU-20 specifications. Some parts of DTU-20 are as
per our specifications.
Item Specifications
Model ND-500
Measurement Chemiluminescence method *1 (decompression method)
method
Gas O2 gas (99.7% or higher purity, 0.3±0.1 MPa)
Dimensions About 220 (W) × 375 (D) × 500 (H) mm
Weight About 22 kg
Power AC100/115 V 50/60 Hz, 300 VA
AC230/240 V 50/60 Hz, 330 VA
Fuse 3.15 A
Interface MAIN UNIT communication cable
Temperature 15 to 35 C
Humidity 80%RH or less (No condensation)
Overvoltage CategoryⅡ (IEC61010-1)
category
Pollution degree 2 (IEC61010-1)
ND-500 (NSX-5000)
9-2 Specifications
Item Specifications
*1
Model DTU-20
Systems High vacuum diaphragm-type dry vacuum pump
Exhaust velocity 20/23 (50/60Hz) L/min
Ultimate pressure 200 Pa
Protective function 1, 80 W, 4P, capacitor run
with thermal protector (automatic recovery type)
Air supply/exhaust O.D.10×I.D. 6 (RC1/8)
tube
Dimensions About 161 (W) × 327 (D) × 217 (H) mm
Weight About 7.5 kg
Power 100/115/230 VAC±10 % 50/60 Hz *2
Full‐load current AC100 V: 1.46/1.46 A (50/60 Hz)
AC115 V: 1.30/1.30 A (50/60 Hz)
AC230 V: 0.67/0.68 A (50/60 Hz)
Temperature 5 to 40 C *3
Humidity 80%RH or less (No condensation)
*1. The specifications of the following parts were changed to ours at the shipment
from the manufacturer.
• The power cable was changed to one for the service outlet for our unit.
• The gas ballast valve was changed to a connector.
• The exhaust outlet connector was changed to a silencer.
*2. Select 100, 115, or 230 VAC as the power supply voltage.
The unit cannot be used with a power supply voltage different from the unit
specifications.
ND-500 (NSX-5000)
Specifications 9-3
Item Specifications
Samples Nonaqueous samples: Nitrogen compound in petroleum
and petrochemicals *1
Aqueous samples: Nitrogen compound in solution *1
Measurement Two-step change (High sensitivity and low sensitivity) *2
range
Measurement Nonaqueous samples: 3 to 20000 ng (to 10000 µg/mL) *3
range Aqueous samples: 1 to 10000 ng (to 5000 µg/mL) *3
Quantitative lower Liquid (nonaqueous): 0.03 µg/mL *4
limit Liquid (aqueous): 0.01 µg/mL *4
Sample volume Liquid (nonaqueous): 1 to 200 µL *5
Liquid (aqueous): 1 to 100 µL *5
Gas: 1 to 25 mL *5*6
Measurement time 4 minutes per measurement *5
Catalyst Platinum catalyst (oxidation catalyst)
*1. Samples need to be combustible under 1100 degrees Celsius and have no
interference.
(excluding nitrogen compounds, such as nitrogen (N2) and azo compounds)
*3. Select an optimal sample volume based on the sample concentration and a
matrix.
*4. It depends on the sample volume, the matrix, the purity of reagents or gases,
and the condition of the unit.
*5. It depends on the unit configuration, method parameters, sample volume, and
concentration.
*6. For gas sample measurement, a gas injector of the GI-500 series is required.
ND-500 (NSX-5000)
9-4 Specifications
This section shows the repeatability of the vertical furnace. To achieve this repeatability,
conduct measurement with optimal parameters, such as electric furnace temperature and
syringe capacity. These also depend on the purity of gases and reagents to use and the
condition of the unit.
Table 9-4 shows the repeatability (nonaqueous samples) when the ASC-550L unit is used.
The data represents the repeatability when pyridine/toluene is measured five times
repeatedly.
Table 9-5 shows the repeatability (aqueous samples) when the ASC-550L unit is used.
The data represents the repeatability when ammonium sulfate solution is measured five
times repeatedly. When a sample is injected manually, it is sampled manually. Therefore,
the repeatability at sampling strongly affects measurement repeatability.
ND-500 (NSX-5000)
Specifications 9-5
Table 9-6 shows the repeatability (aqueous salt-containing samples) when ASC-550L is
used. The data represents the repeatability when the acid gas removal set is connected
and ammonium sulfate/3% sodium chloride solution is measured five times repeatedly.
When a sample is injected manually, it is sampled manually. Therefore, the repeatability
at sampling strongly affects measurement repeatability.
Table 9-7 shows the repeatability when the CRI-500V unit is used. The data represents
the repeatability when dibutyl disulfide/toluene is measured five times repeatedly.
Because of manual sampling, the repeatability at sampling strongly affects measurement
repeatability.
ND-500 (NSX-5000)
9-6 Specifications
Table 9-8 shows the repeatability when GI-500 series is used. The data represents the
repeatability when pyridine/hexane is measured five times repeatedly. Because of
manual sampling, the repeatability at sampling strongly affects measurement
repeatability.
ND-500 (NSX-5000)
Guarantee Provisions 10-1
Guarantee Terms
1. Guarantee period is one year from the date of delivery.
2. If the product is used in Japan and is used under normal usage during the
guarantee period and a problem occurs, Nittoseiko Analytech will repair
defective parts or replace them with new ones or equivalents free of charge.
The replaced old parts will be owned by Nittoseiko Analytech.
3. In the following cases, repairs will be charged even within the guarantee period:
(1) Failure due to improper handling or misuse
(2) Failure due to handling, use, storage, or breakage by storage that deviates
from the design and specification conditions
(3) Failure and damage due to transportation after delivery, dropping or
vibration during movement
(4) Failure caused by modification and repair by a person other than a
serviceman commissioned by the Company
(5) Replenishment of consumables and naturally consumed parts*
(6) Failure and damage due to accidental force such as fire, flood,
earthquakes, lightning, salt damage, abnormal voltage, and other natural
disasters
(7) Failure due to use in an environment outside the specifications
(8) Failure due to use with options and consumables other than certified
products
(9) Insufficient maintenance management of cleaning, maintenance,
inspection, and use of consumables other than those specified by the
Company.
(10) Other failures that are not considered to be our responsibility
* Items listed in " Consumables" at the end of the section are consumables.
4. Judgment as to whether or not failures correspond to free repair depends on
the company or a person commissioned by the company.
5. The contact for the failures shall be Nittoseiko Analytech or its agent.
6. Methods of returning defective products will be decided in consultation with the
contact place each time.
ND-500 (NSX-5000)
10-2 Guarantee Provisions
ND-500 (NSX-5000)
Guarantee Provisions 10-3
Guarantee Terms
1. If you purchased a product from a Nittoseiko Analytech distributor outside
Japan, please contact the distributor for inquiries about the guarantee period
and terms as well as for failures.
Please note that Nittoseiko Analytech may not support even if you contact us
directly.
2. The preceding paragraph does not apply to inquiries regarding damages
(limited to those for which Nittoseiko Analytech may not be exempted in whole
or in part) related to Product Liability Act applicable to Nittoseiko Analytech.
ND-500 (NSX-5000)
10-4 Guarantee Provisions
ND-500 (NSX-5000)
Consumables 11-1
ND-500 (NSX-5000)
11-2 Consumables
11-2-1. Consumables for ND-500 Gas Dryer Set for Vertical Furnace
Table 11-2 lists consumables for ND-500 gas dryer set for the vertical furnace.
Table 11-2. Consumables List of ND-500 Gas Dryer Set for the Vertical Furnace
ND-500 (NSX-5000)
Consumables 11-3
Table 11-3 lists the consumables for combustion tube set (nonaqueous) for ND-500 with
vertical furnace.
Table 11-3. Consumables for Combustion Tube Set (Nonaqueous) for ND-
500 with Vertical Furnace
ND-500 (NSX-5000)
11-4 Consumables
Table 11-4 lists the consumables for combustion tube set (aqueous) for ND-500 with
vertical furnace.
Table 11-4. Consumables for Combustion Tube Set (Aqueous) for ND-500
with Vertical Furnace
ND-500 (NSX-5000)
Consumables 11-5
Table 11-5 lists the consumables for the manual injection port set for ND-500.
Table 11-5. Consumables List for Manual Injection Port Set for ND-500
2 Silencer DTU20S 1 pc
ND-500 (NSX-5000)
11-6 Consumables
Table 11-7 lists consumables for the acid gas removal set.
ND-500 (NSX-5000)
Consumables 11-7
ND-500 (NSX-5000)
11-8 Consumables
ND-500 (NSX-5000)