Download as pdf or txt
Download as pdf or txt
You are on page 1of 8

Analysis of a functionally graded compliant mechanism surgical

grasper
Jovana Jovanova1, Angela Nastevska1, Mary Frecker2, and Milton E. Aguirre3,
1: Ss Cyril and Methodius University in Skopje, 2: The Pennsylvania State University, 3: Milton
Medical Innovations B.V., Healthcare Technology Start-Up

Abstract— The focus of this paper is to explore the benefits of circumventing the rigid joints and relative motion between
functional grading on the performance of a compliant mating parts, the instrument performance can be increased due
mechanism surgical grasper currently pursuing to the elimination of friction. Past research has documented the
commercialization. The compliant mechanism tool tip is development of tasks, optimal designs and prototypes of
analyzed to illustrate the potential of exploiting functionally compliant mechanism surgical graspers in [2, 3].
graded superelastic behavior of Nickel Titanium alloy in high
stress segments. A finite element model is used to predict International researchers have worked on optimizing MIS
grasping performance. Superelasticity is taken into instrumentation, especially graspers, to improve instrument
consideration using a standard nonlinear shape memory alloy performance and accuracy. Mathematical models, finite
material model. To approximate functional grading, flexible element models and standard pseudo-rigid-body-models are
elements within the compliant mechanism are segmented and typically used for the design of compliant graspers [4-9]. Some
material properties varied to explore the effect on grasping of the benefits of engineered elasticity in the design of surgical
performance. Segmentation analyses show improvements in instruments for laparoscopic and robot-assisted surgery are
maximum stress with minimal compromise to grasping highlighted in [10]. A novel compliant laparoscopic instrument
performance. that combines a scissors and forceps into a single
I. INTRODUCTION multifunctional device is shown in [11].

Shape memory alloys (SMAs) are a group of metallic Some designs incorporate additional actuation and sensor
materials that exhibit unique mechanical properties, such as devices. A superelastic NiTi microgripper with integrated
superelasticity, and, accordingly, are suitable for compliant electromagnetic actuators and piezoelectric force sensors has
mechanisms where high flexibility is required for the been shown in [12]. A powered surgical monobloc tool tip that
application. SMAs undergo solid-to-solid phase can pitch and yaw using corner-filleted flexure hinge-based
transformations to recover very large deformation. The so compliant joints is presented in [13] which has an integrated
called superelastic effect is induced by stress changes during compliant grasper. Origami engineering is applied to generate
external loading and unloading of the material. The most two rotations and one translation for the design of a compliant
common SMA in the biomedical field is Nitinol (NiTi) based gripper as a novel four-degree-of freedom grasper in [14].
on its biocompatibility and good resistance to corrosion and The possibility of having nonhomogeneous mechanical
fatigue. NiTi is advantageous compared to stainless steel and properties in a compliant mechanism in order to obtain flexible
other biocompatible materials because of the superelastic and and stiff regions is possible with functional grading and
shape memory effects leading to higher localized compliance tailoring the moduli of elasticity of NiTi. By allowing each
and higher strain recovery. The benefits of using NiTi in segment to have different composition of nickel and titanium
medical applications, including endoscopic surgery, are during fabrication, it is possible to have different nonlinear
summarized in [1]. material behavior for each of the segments [15, 16]. By using
Endoscopic surgery, also called Minimally Invasive additive manufacturing for fabrication, the superelastic
Surgery (MIS), is a modern surgical technique in which properties of the material could be tailored by controlling the
diagnosis and disease treatment are performed through small composition of nickel and titanium. A direct energy deposition
incisions or natural body openings. These procedures rely on approach to accomplish this is currently being developed to
long slender devices able to access and operate inside the body change the microstructure of the material and its superelastic
with use of cameras. The benefits of MIS procedures have been properties [17, 18]. Different compositions of NiTi have
widely reported, including less post operational complications, different microstructure and therefore different macroscopic
reduced levels of pain, shorter recovery time, and lower risk of material properties. The relationship between the moduli of
infection as compared to traditional open surgery. elasticity of each composition and the fabrication technology
is determined by material characterization. Furthermore the
Compliant surgical graspers have been designed based on mechanical interface between different compositions is still to
the advantages of monolithic design that can enhance be explored. The advantage of using nonlinear elastic material
instrument performance and simplify manufacturing. By with the geometric design is the combined effort to reach

978-1-5386-6380-6/18/$31.00 ©2018 IEEE


superlesticity on demand and tailoring the local material with maximum of 1.26 mm and minimum of 0.0916 mm
behavior for achieving overall controlled performance. element size. The Multifrontal Massively Parallel sparse direct
Solver - MUMPS was used as direct solver from the COMSOL
Design and optimization methods for developing hybrid, solver options.
multi-material, compliant instruments, as well as their
respective fabrication techniques is presented in [19, 20]. Two operational conditions in the finite element analysis
Geometric optimization and material characterization of (FEA) are used to predict grasping performance: the free
functionally graded compliant mechanism has been developed deflection and blocked force conditions at the grasper tip.
and presented in [21]. Exploiting the benefits of contact-aided These distinct loading and boundary conditions are illustrated
compliant mechanism in surgical tool tips is presented in [22]. in Fig. 2. In the free deflection case, the tips of the grasper are
free to move as the center section is displaced in the –x
The remainder of the paper is structured as follows: section
direction. When the grasper is in its operating mode, it comes
II presents the surgical compliant grasper design. Section III in contact with tissue and transfers force to the tissue to
describes the finite element model including superelastic complete a certain surgical task. The grasping force depends
effects. In section IV the functional grading of the compliant on the stiffness of the tissue. In the FEA, the maximum
mechanism is introduced and the potential of tailoring material possible grasping force is modelled by constraining the
properties is explored. Section V concludes the paper and lays vertical displacement of the tips in the y direction while
out the future work for this research. displacing the center section in the –x direction. The reaction
II. COMPLIANT MECHANISM DESIGN OF A SURGICAL force measured at the tip is defined as blocked force (Fb), and
GRASPER TOOL TIP is the maximum gripping force.
The compliant tool tip proposed in this paper has been
analyzed in previous publications [6-9, 23] which focus on
improving the force-feedback capabilities of surgical graspers
through monolithic compliance using a homogeneous material.
Standard graspers lack force feedback because they rely on
conventional rigid-linkages that introduce friction and
backlash. Compliant mechanisms have shown better grasping
performance based on the absence of rigid-linkages and the
ability of flexible members to conform around the shape of
tissue when grasping. In this research, a 5 mm diameter
laparoscopic compliant tool tip is considered. The geometric
properties of the design within this case study remain constant
and performance evaluations are compared to a baseline design
currently pursing commercialization.
The full surgical instrument is illustrated in Fig. 1, which
consists of a compliant tool tip fixated to a conventional 5 mm
diameter laparoscopic actuation system, i.e., handle and shaft.
The tool tip is fabricated in a half-open equilibrium position
and actuated by a central push/pull rod that opens and closes
the device.

Figure 2: Loading and boundary conditions that predict grasper


performance

A nonlinear material model introduces superelastic


behavior in the FEA. The superelastic response is captured by
using a standard nonlinear SMA material model for the
mechanical behavior of NiTi ( ):
Figure 1. Compliant tool tip assembled onto a laparosopic grasper
actuation system.

To explore the advantages of the superelastic behavior of


NiTi, a 3D finite element model of the surgical grasper
compliant tool tip has been developed and analyzed. COMSOL =
Multiphysics was used as FEA modeling environment. The (1)
structural mechanics module was used to model the grasper
with the nonlinear structural material module implemented
with the fully coupled approach. Automatic physics controlled
mesh was implemented with 149931 free tetrahedral elements
where is the Young’s modulus of the austenite phase of The free deflection condition was analyzed for two cases:
NiTi, is the Poisson’s ratio, is the material response ratio the entire compliant mechanism consisting of linear elastic
between tension and compression, is the maximum residual material, and the entire compliant mechanism consisting of the
strain, is the starting stress value for the forward phase nonlinear superelastic material. To examine the deformation
transformation, is the final stress value for the forward modes of the grasper, it was deformed to a complete closed
position with a prescribed horizontal displacement of 0.7mm,
phase transformation, is the starting stress value for the which required an input actuation force of 50N. The
reverse phase transformation, and is the final stress value deformation and the von Mises stress of the 3D FEA models
for the reverse phase transformation. This nonlinear material are presented in Fig. 4 and Fig. 5. The maximum stress in the
model, Fig. 3, has been developed in [24], and successfully linear elastic case is 660MPa, while in the superelastic case it
implemented for compliant mechanism design in [25]. The is 520MPa. The maximum vertical displacement of the grasper
combined effort of using superelastic material and the tip in the free deflection condition for the nonlinear superelastic
geometry properties of compliant mechanisms to exploit high material is 5.819mm, which compared to the linear elastic
localized flexibility has been shown in [26, 27]. grasper is 5.822mm; these two are nearly the same value, i.e.
the total free deflection is the same for both the linear and
nonlinear superelastic materials.

Figure 3: SMA nonlinear material model. The model incorporates the


critical stresses for transformation obtained from uniaxial tension-
compression test.

The critical parameters used to evaluate the benefits of


functional grading are the linear region corresponding to the
higher Young’s modulus ( ) - nitinol austenite phase -
marked with blue in Fig. 3, and the transformational stress
( ) that initiates the transformation from austenite to Figure 4. Free deflection and von Mises stress of the grasper with the linear
martensite - marked with yellow in Fig. 3. The values material. The deformation modes and output performance parameters
prescribed in the nonlinear material model are based on (Delta_tip and Von Mises stress) remained consistent in the free deflection
experimental material characterization of NiTi bulk specimen of the grasper assuming linear elastic and nonlinear superelastic NiTi
in uniaxial loading and unloading, which were fabricated using material.
additive manufacturing [17]. The parameters important for
analyses in this work are given in Table I.

TABLE I.

NiTi material properties


NiTi
70 [GPa] 15.1[GPa] 500 [MPa]

III. FEA MODEL OF COMPLIANT SURGICAL GRASPER WITH


LINEAR ELASTIC AND NONLINEAR SUPERELASTIC NITINOL
The operational loading conditions used to predict grasping
performance are analyzed assuming constant geometry and
material properties listed in Table 1. For comparison, the
analyses are modeled first with linear elastic NiTi material with
constant Young’s modulus , and then superlesticity was
introduced using the standard nonlinear material model for
NiTi. Superelasticity is expected to be advantageous in areas
where there are high strains. Therefore the stresses are
expected to be reduced in these thin sections compared to the
linear elastic material. Figure 5. Free deflection and von Mises stress of the grasper with
nonlinear superelastic NiTi material
The blocked force condition was also considered. As the value for the linear elastic case is 3800 MPa, while in the
grasper is actuated, the fixed constraint at the tip creates a nonlinear superelastic case is 2500 MPa.
reaction force, or blocked force. The stress distributions of the
grasper are shown in Fig. 6 for the linear elastic material and The blocked force and the maximum stress have been
in Fig. 7 for the nonlinear superelastic material. analyzed over the range of input displacement from 0.05mm
to 0.3mm, as shown in Figure 8.

Figure 8. Maximal von Mises stress and reaction force in the linear
elastic and nonlinear superelastic grasper

It is noticeable that the maximum stress values for the


nonlinear material model are lower than in the linear material
case. This clearly shows the benefit of superelasticity in
reducing stress in the surgical grasper. The blocked force
Figure 6. Von Mises stress for linear elastic material NiTi compliant surgical remains almost the same in the two cases, which shows that the
grasper grasping performance is maintained, even though the peak
stresses are reduced. The same grasping performance, but with
less stress would result in a longer product lifetime and reduced
risk of failure.
In the next section we explore localizing the superelasticity
and further improving the performance through functional
grading of the material.

IV. SEGMENTED FEA MODEL FOR FUNCTIONALLY GRADED


SURGICAL GRASPER
Functional grading could lead to tailored stiffness
characteristics that can further improve grasping performance.
The tailored modulus of elasticity and critical transformation
stress program the material to go into the superelastic behavior
as required by the design. To improve the design performance
of the compliant grasper, the aim is to produce sufficient force
while reducing the peak stress. Reducing the peak stresses in
the surgical grasper is expected to result in better reliability as
well as longer lifetime of the device. To approximate
functional grading, the compliant members are segmented to
explore the potential of localizing the superelasticity. The
segmentation is focused on the thin parts of the compliant
grasper, where the highest stresses are expected to occur.
Figure 7. Von Mises stress for nonlinear superelastic material NiTi The compliant segments of a half-symmetry model of the
compliant surgical grasper
grasper are divided into three segments each, coinciding with
From the figures showing the von Mises stress in both cases, the regions where highest stress expected, based on the
it can be concluded that the maximum stresses are much previous FEA. The segmentation is shown in Fig. 9. In order
lower in the nonlinear material grasper. The maximum stress to make systematic changes in the parameters, three cases are
considered, as numbered and shown in Fig. 9.
In Fig. 10 it can be observed that localizing the
superelasticity in various segments reduces the maximum
stress in all cases, but there is not much difference in the
maximum stress among the three segmented cases considered.
However, the blocked force results in Fig.11 show that cases
13 and 23 exhibit more blocked force than the other cases,
particularly at higher input displacements.
Design 13 was selected for further analysis because the
superelastic regions correspond to areas of high stress in the
free deflection case; the stress distribution can been seen in Fig.
12. By introducing the superelastic material in model 13, the
peak von Mises stress is reduced in the grasper, compared to
both the linear elastic and nonlinear superelastic case. The
blocked force for this design is 0.9 N, which is higher than the
linear elastic case (0.76 N) and the nonlinear superelastic case
(0.67N). This indicates that the blocked force has not been
reduced with the segmentation, and the grasper maintains its
performance.

Figure 9. Segmented half grasper model

To account for the segmentation, the grasper is modeled


with linear elastic material, but for each of the segments
marked in blue, the nonlinear superelastic material behavior is
assigned, using the properties in Table I. The maximum
stresses and the blocked forces at the tip are simulated for the
same prescribed input displacement, as shown in Fig. 10 and
Fig. 11.

Figure 12. Free deflection and von Mises stress with nonlinear superelastic
material in segments for case 13

In order to improve the grasping performance and reduce the


maximum stress even further, the material properties of the
Figure 10. Maximum stress in segmented model for prescribed displacement NiTi could be tailored during fabrication. The next set of cases
in blocked force condition. explores this possibility through analysis of the compliant
grasper with different nonlinear material behavior.
In order to identify the influence of the material properties
to the grasper performance, five different nonlinear material
curves were created with parametric variation of the values
given in Table I. The superelastic nonlinear material from
Table I is called Material A in Table II. The critical
transformation stress was varied in Material B and
Material C, while keeping the modulus of elasticity constant as
in Material A. Then the transformation stress was kept constant
as in Material A, while varying the NiTi austenite modulus of
elasticity , for Material D and Material E. All materials are
marked with different colors to match the results later in this
Figure 11. Blocked force at the tip in segmented model for prescribed section.
displacement.
TABLE II. decrease of the modulus of elasticity, Fig. 14, which means that
NiTi material properties variations the performance of the grasper changes. Even though the
stresses are reduced, as shown in Fig. 13, further exploration is
Material [GPa] [GPa] [MPa] color
needed to maintain the grasping blocked force.
Material A 70 15.1 500
However, the change in the critical transformation
Material B 70 15.1 250 stress showed potential to maintain the grasper
performance even when the material properties vary. Material
Material C 70 15.1 375 A was kept as the basic nonlinear superelastic material, with
Material D 55 15.1 500 = 500 MPa. For parameterization, Material B was created
by reducing the critical stress to = 250 MPA, and Material
Material E 40 15.1 500 C is in between with = 375 MPa, while keeping the
modulus of elasticity in all phases the same for all three
The grasping performance was calculated for all five materials A, B and C.
nonlinear materials through the blocked force in operating The influence of the critical transformation stress on
mode condition and the maximum overall stress, and compared the grasping performance is shown in Fig. 15 and Fig. 16, with
to the linear elastic material case. the comparisons of the maximum overall stress and the reaction
To identify the parametric influence of the change of forces for the nonlinear superelastic materials A, B and C and
modulus of elasticity while the critical transformation stress is the linear elastic material case.
constant, Material A has modulus of elasticity =70GPa, then
Material D and Material E have modulus of elasticity
=55GPa and =40GPa, accordingly. The maximum overall
stress and the blocked forces for the nonlinear materials A, D
and E are shown in the graphs in Fig. 13 and Fig. 14.

Figure 15. Maximum stress for different critical transformation stress


(Material A = 500 MPa; Material B = 250 MPa; Materia _C = 375 MPa)

Figure 13. Maximum stress for different austenite modulus of elasticity


(Material A =70GPa; Material D =55GPa; Material E =40GPa)

Figure 16. Reaction force for different critical transformation stress


(Material_A = 500 MPa; Material_B = 250 MPa; Material_C = 375 MPa)

The change of critical transformation stress, reduces the


overall stress, Fig. 15, but gives the possibility to adjust and
Figure 14. Reaction force for different austenite modulus of elasticity
control the reaction force to maintain the grasper’s
(Material A =70GPa; Material D =55GPa; Material E =40GPa) performance. Fig. 16 shows that for different prescribed
displacements the blocked force varies, but follows the trend
It can be seen that the change in austenite modulus of as in the linear elastic material case. It can be noted that
elasticity changes the grasping performance. The reaction force material C creates higher reaction force for the lower
shows a trend of lowering the value drastically with the prescribed displacement range, but then it becomes equal to
material A, and follows the trend of the linear elastic case for force slightly exceeds the reaction force of the linear case over
the most of the displacement range. Material B shows a trend the lower displacement range, but then between 0.25 mm to 0.3
of limiting the maximal reaction force that could be beneficial mm, the functionally graded grasper exhibits the same grasping
to surgical procedures sensitive to tissue damaging by applying force as the linear elastic one.
an extensive force.
By fine tuning the local material properties in particular
high flexibility regions using functional grading, it is possible
to maintain the grasping performance of the surgical grasper,
while reducing the maximum overall stress. To further explore
the functional grading, a combination of different nonlinear
material curves (Table II) in different segments of the surgical
grasper (Fig. 9) was modeled and analyzed for the blocked
force condition. Since the change in the critical transformation
stress provides better grasping performance, the functional
grading is modeled as a linear elastic material grasper with
nonlinear materials A, B and C in segments 1, 2, and 3. The
designs are shown schematically in Table III.
TABLE III. (Refer to Table II for color legend) Figure 17. Maximum overall stress for functionally graded NiTi grasper
Functionally graded designs Design line (Designs 1, 2, 3, and 4)

Design 1: 1B_2C_3A

Design 2: 1B_2A_3C
Figure 18. Reaction force for functionally graded NiTi grasper
(Designs 1, 2, 3, and 4)

These results demonstrate the potential of functional


grading, as the material properties could be tailored so that the
Design 3: 1C_2C_3A surgical grasper could maintain the grasping force and
experience reduced stresses, resulting in better overall
performance of the surgical tool. By localizing the functional
grading, the superelastic effect is exploited only when
necessary.

Design 4: 1B_2C_3B
V. CONCLUSION
The comparisons of the maximum overall stress and the This paper introduces nonlinear material properties in a
reaction forces for the functionally graded designs are surgical compliant grasper tool tip. One of the main concerns
presented in Fig. 17 and Fig. 18.
with compliant mechanisms is the possibility of fatigue failure
The results in Fig. 17 confirm once again the stress reduction when high stresses and high strains occur repeatedly. The
in the compliant mechanism grasper, as all 4 designs exhibit compliant grasper takes advantage of the superelastic behavior
lower overall maximum stress than the linear elastic case. On of nitinol, as it reduces the peak stress in the compliant
average the stress reduction is 30%. Fig. 18 compares the mechanism while maintain the grasping force, resulting in
blocked force of all four functionally graded designs with the
equivalent performance and longer lifetime of the device.
linear elastic case materials. Design 4 exhibits 40% lower
reaction force at the tip, while all other designs are with similar Comparison analyses were presented of the grasper
performance with a variation of 8%. Designs 1 and 3 perform performance when fabricated with linear elastic NiTi material
almost the same, and they produce slightly higher reaction and nonlinear superelastic material. A parametric study
force at lower input displacements, but then between 0.2mm showed the influence of the critical transformation stress and
and 0.25 mm they begin to exhibit slightly lower reaction force the austenite modulus of elasticity to the grasping
than the linear elastic material. Design 2 shows that the reaction performance. When the nonlinear material is introduced only
to the segments with high stress expected, overall stress [18] B. Bimber et al, "Anisotropic microstructure and superelasticity of
additive manufactured NiTi alloy bulk builds using laser directed
reduction occurs. This shows potential in functional grading, energy deposition," Materials Science and Engineering a-Structural
where the nonlinear material is used locally. This approach Materials Properties Microstructure and Processing, vol. 674, pp. 125-
showed that the grasper performance could be improved, as 134, 2016.
[19] Aguirre, Milton E., and Mary Frecker. "Design and optimization of
with fine tuning of material properties in different segments of hybrid compliant narrow-gauge surgical forceps." ASME 2010
the grasper, the grasping force is kept the same while the Conference on Smart Materials, Adaptive Structures and Intelligent
maximum overall stress is reduced Systems, Department of Mechanical and Nuclear Engineering,
Future research includes further analyses for material Pennslyvania State University, University Park, PA, Sept. 2010.
[20] Aguirre, M. E., et al. "Fabrication and strength-based design of a meso
tailoring based on manufacturing parameterization and/or forceps." Proceedings of ASME IDETC/CIE (2009): 1-9.
geometrical constraints, and on specific requirements for the [21] Jovanova J., Frecker M, Hamilton R.F., Palmer T.A., Target shape
blocked force. This type of performance analyses could lead optimization of functionally graded shape memory alloy compliant
mechanisms, Journal of Intelligent Material Systems and Structures,
to an optimized design of the compliant grasper with October 2017, doi.org/10.1177/1045389X17733057
functional grading for a particular surgical task or application. [22] Aguirre, Milton E., and Mary Frecker. "Design of a Multi-Contact-
Aided Compliant Mechanism." ASME 2011 International Design
Engineering Technical Conferences and Computers and Information in
REFERENCES Engineering Conference. American Society of Mechanical Engineers,
[1] L. Lecce, ed, “Shape memory alloy engineering: for aerospace, 2011.
structural and biomedical applications”, Elsevier, 2014. [23] Aguirre, Milton, et al. "Technology Demonstrator for Compliant
[2] Addis, M., Aguirre, M.*, Frecker, M., Haluck, R., Mathew, A., Pauli, Statically Balanced Surgical Graspers." Journal of Medical Devices 9.2
E., and Gopal, J., 2012. Development of Tasks and Evaluation of a (2015): 020926.
Prototype Forceps for NOTES. Journal of the Society of [24] Auricchio F., Taylor R. L., Lubliner J., 1997, "Shape-memory alloys:
Laparoendoscopic Surgeons, 16(1), pp. 95-104 macromodeling and numerical simulations of the superelastic
[3] Aguirre, M., G. Hayes, R. Meirom, M. Frecker, C. Muhlstein, and J. Behavior," Computational Methods in Applied Mechanical
Adair, 2011. Optimal Design and Fabrication of Narrow-Gauge Engineering, 146: pp.281–312.
Compliant Forceps. ASME Journal of Mechanical Design, 133(8) [25] Jiening Liu, , Benjamin Hall, Mary Frecker, Edward W Reutzel, 2013,
[4] Tolou, Nima, and Just L. Herder. "Concept and modeling of a "Compliant articulation structure using superelastic NiTiNOL," Smart
statically balanced compliant laparoscopic grasper." ASME Paper No. materials and structures, V22 No. 094018, doi:10.1088/0964-
DETC2009-86694 (2009). 1726/22/9/094018
[5] Lamers, A. J., Juan Andrés Gallego Sánchez, and Just L. Herder. [26] Jovanova, Jovana, et al. "Target Shape Optimization of Functionally
"Design of a statically balanced fully compliant grasper." Mechanism Graded Shape Memory Alloy Compliant Mechanism." ASME 2016
and machine theory 92 (2015): 230-239. Conference on Smart Materials, Adaptive Structures and Intelligent
[6] Ananthasuresh GK, Kota S (1995) Designing compliant mechanisms, Systems. American Society of Mechanical Engineers, 2016.
Mechanical Engineering, 117(11)93/6. [27] Jovanova, Jovana, and Mary Frecker. "Two Stage Design of Compliant
[7] Canfield S, Edinger B, Frecker M, and Koopman G (1999) Design of Mechanisms With Superelastic Compliant Joints." ASME 2017
Piezoelectric Inchworm Actuator and Compliant End-Effector for Conference on Smart Materials, Adaptive Structures and Intelligent
Minimally Invasive Surgery, Proceedings SPIE Smart Strucutres and Systems. American Society of Mechanical Engineers, 2017.
Materials, March, 1999, Newport Beach, CA, Paper 3668-77.
[8] Herder, J. L. (2001). Energy-free systems, Theory, conception and
design of statically balanced spring mechanisms. Delft, Delft University
of Technology. PhD.
[9] Tolou N, Herder JL, Concept and Modeling of a Statically Balanced
Compliant Laparoscopic Grasper. 2009. Proceedings of the ASME
Design Engineering Technical Conference. San Diego, CA, USA.
[10] Kota, S., et al. "Design and application of compliant mechanisms for
surgical tools." Journal of biomechanical engineering 127.6 (2005):
981-989.
[11] Frecker, Mary I., Katherine M. Powell, and Randy Haluck. "Design of
a multifunctional compliant instrument for minimally invasive
surgery." Journal of biomechanical engineering 127.6 (2005): 990-993.
[12] Kim, Deok-Ho, et al. "A superelastic alloy microgripper with
embedded electromagnetic actuators and piezoelectric force sensors: a
numerical and experimental study." Smart materials and structures
14.6 (2005): 1265.
[13] Chandrasekaran, Karthik, and Asokan Thondiyath. "Design of a Two
Degree-of-Freedom Compliant Tool Tip for a Handheld Powered
Surgical Tool." Journal of Medical Devices 11.1 (2017): 014502.
[14] Salerno, Marco, et al. "A novel 4-DOF origami grasper with an SMA-
actuation system for minimally invasive surgery." IEEE Transactions
on Robotics 32.3 (2016): 484-498.
[15] Hamilton, R.F., Bimber, B.A., Taheri Andani, M. & Elahinia, M.
2017, "Multi-scale shape memory effect recovery in NiTi alloys
additive manufactured by selective laser melting and laser directed
energy deposition", Journal of Materials Processing Tech, vol. 250, pp.
55-64.
[16] M. Elahinia et al, "Fabrication of NiTi through additive manufacturing:
A review," Progress in Materials Science, vol. 83, pp. 630-663, 2016.
[17] Hamilton, R.F., T.A. Palmer, B.A. Bimber, 2015 "Spatial
Characterization of the Thermal Induced Phase Transformation
Throughout As-Deposited Additive Manufactured Ni-Ti Bulk Builds,"
Scripta Materialia, 101: pp. 56-59

You might also like