IFU OPMI Pico S100.PDF.pdf
IFU OPMI Pico S100.PDF.pdf
About this manual These Instructions for Use are part of the delivery package.
• Carefully read them before using the device.
• Keep them at the site of use of the device.
• Store them for the entire service life of the device.
• Pass them on to every subsequent owner or user of the device.
Orientation aids – The chapter overview at the beginning of these Instructions for Use pro-
vides a summary of all subjects.
– The contents of each chapter are specified in detail at the beginning of
each chapter.
– A list of abbreviations, key words and technical terms in the annex facili-
tates the search for specific terms.
Area of application These Instructions for Use apply to the S100 / OPMI® pico with the following
identification:
Version 7.0
Page 2 G-30-1781-en
S100 / OPMI pico
Version 7.0
G-30-1781-en Page 3
S100 / OPMI pico
Version 7.0
Page 4 G-30-1781-en
S100 / OPMI pico
Chapter overview
Version 7.0
G-30-1781-en Page 5
S100 / OPMI pico
Version 7.0
Page 6 G-30-1781-en
S100 / OPMI pico Safety Measures
Safety Measures
Safety devices..................................................................................19
Version 7.0
G-30-1781-en Page 7
Safety Measures S100 / OPMI pico
Hazard symbols
The following safety information has been incorporated into the Instructions
for Use. Please note this information and be particularly careful in these cases.
WARNING Warning label, which may refer to fatal injuries or severe injuries if pre-
cautions are not followed.
CAUTION Indicating a hazard, which may lead to moderately severe injuries
if risks are not avoided.
NOTE Warning label, which may refer to minor injuries and property damages
if precautions are not followed.
Information symbols
The following information symbols are used in these Instructions for Use:
– Listing
Requirements for an action
• Prompt for action
Result of an action
Additional information and hints
Version 7.0
Page 8 G-30-1781-en
S100 / OPMI pico Safety Measures
Target group
These Instructions for Use are intended for physicians, nurses and other med-
ical and technical staff who prepare, operate or maintain the device after ap-
propriate training. It is the duty of the equipment owner/operator to train and
brief all the operating staff.
Extended service tasks are not part of these Instructions for Use. They will be
performed by staff specially trained for this purpose by ZEISS.
Field of application
Intended use
Normal use
Provision: The S100 / OPMI offers in conjunction with different suspension systems (e.g.
floor stand, wall mount, ceiling mount) and flexible carrier arms, an ergo-
nomic working position for the treatment environment or surgery. The system
can be securely moved into the working position with the hand grips without
any problems.
Version 7.0
G-30-1781-en Page 9
Safety Measures S100 / OPMI pico
Treatment/surgery: The microscope is positioned and roughly focused by the physician or the sur-
geon. The most frequently used functions as the setting of focus, magnifica-
tion and light intensity are manually controlled by the physician or the sur-
geon during treatment or surgery.
A video player is connected to the system for educational purposes or to vis-
ualize the progress of the treatment/surgery.
After the treatment / surgery The physician or the surgeon swings the surgical microscope from the treat-
ment or surgical area into its standby position.
The S100 / OPMI pico can be cleaned and disinfected.
Disposal: For the disposal of electrical and electronic devices, please observe the rele-
vant national legislation and other regulations applicable in your country.
CAUTION Live video images and recorded video sequences, video clips (cut sequences)
and single images must not be used for diagnosis purposes. Visualized im-
ages may contain variations in shape, contrast and color.
The correct use of the device is absolutely vital for safe operation. Therefore,
please thoroughly familiarize yourself with the content of these instructions
for use before starting up the system. Please also observe the instructions for
use of any additional equipment. You can obtain additional information from
ZEISS Service.
• Only operate the device within the scope of its intended use.
• Only personnel who have undergone appropriate training and instruction
are allowed to use this device. It is the responsibility of the customer or
the company operating the device to train and instruct all operating per-
sonnel.
• Please keep these Instructions for Use where they are easily accessible at
all times for the persons operating the system.
Version 7.0
Page 10 G-30-1781-en
S100 / OPMI pico Safety Measures
Legal regulations • Observe the legal regulations for accident prevention and occupational
health and safety applicable in the country concerned.
• This system must not be modified without the manufacturer's approval. If
the system is modified, suitable inspections and testing must be complet-
ed to ensure that it can still be used safely.
The manufacturer is not liable for damage caused by unauthorized per-
sons tampering with the system. Furthermore, this will forfeit any rights
to claim under warranty.
• Modifications and repairs on this device or any equipment operated to-
gether with this device may only be performed by ZEISS SERVICE.
• If required by the regulations and directives applicable in the country of
use, connect the system to a special emergency backup power supply.
Ambient requirements • Do not operate the system contained in the delivery package
– in explosion-risk areas,
– if inflammable anesthetics or volatile solvents such as alcohol, benzine
or similar chemicals are present at a distance of less than 25 cm.
• Do not station or use the instrument in damp rooms. Do not expose the
instrument to water splashes, dripping water or sprayed water.
• Do not place any fluid-filled containers on top of the instrument. Make
sure that no fluids can seep into the instrument.
• Make sure that the installation conditions and the use of the device meet
surgical requirements: See Ambient conditions on Page 257
Electrical safety • Always switch off the system before disconnecting it from / connecting it
to line power, for cleaning its surface or if it will not be used for a pro-
longed period of time.
• Switch off the unit at the power switch if you notice any smoke, sparks or
unusual noise. Do not use the unit until it has been repaired by our service
team.
• Do not use multiple sockets!
• Never open the device! The device contains freely accessible live compo-
nents. If you remove the housing, you run the risk of electric shock.
Version 7.0
G-30-1781-en Page 11
Safety Measures S100 / OPMI pico
• Do not touch the system if your body is electrostatically charged and the
system is not grounded.
• A potential equalization connector (see IEC 60601-1:2005) is provided on
the connector panel. This connector is optional for halogen and Xenon il-
lumination. This connector can be used for placing other active devices at
the same ground potential or for redundant grounding to protective
earth.
• Please note the information on EMC (electromagnetic compatibility) in the
chapter "System Data" on Page 245.
Transport & Service • The device may only be moved over long distances (e.g. relocation, return
for repair etc.) in its original packaging or in special return packaging.
Please contact your dealer or ZEISS Service.
• This system is a high-grade technological product. To ensure optimum
performance and safe working order, we recommend having it checked
by our service representative as part of regular scheduled maintenance.
• In order to prevent any impairment of the device's safety due to age,
wear, etc., the user must ensure that the device is subjected to the neces-
sary safety checks (see section "Care and Maintenance").
Optics and light • Never look into a light source or the sun with the binocular tube, the ob-
jective lens or an eyepiece.
• Never directly look into the light source, e.g. a microscope object lens or
a light guide to avoid phototoxic damages to the eyes.
• Adjust the brightness at the stand. Start with the lowest brightness and
slowly increase it to the required but not more than the allowed bright-
ness.
• With increasing age of the light source, the actual illumination intensity
delivered at the respective setting decreases (this is a normal property of
the system).
Version 7.0
Page 12 G-30-1781-en
S100 / OPMI pico Safety Measures
Version 7.0
G-30-1781-en Page 13
Safety Measures S100 / OPMI pico
Version 7.0
Page 14 G-30-1781-en
S100 / OPMI pico Safety Measures
During use
• Reduce the brightness and duration of light exposure of the surgical area
to the absolute minimum level required to avoid burn injuries.
• Never directly look into the light source, e.g. a microscope object lens or
a light guide to avoid phototoxic damages to the eyes.
• Never leave a device unattended with the light source still switched on.
• Not approved or defective accessories for the device may lead to elevated
leakage currents. Never connect the non approved or defective accesso-
ries to the device.
• Never touch video interfaces while in contact with the patient.
• If you need to change the lamp shortly after it has failed, the lamp will still
be very hot. Use heat-resistant protective gloves when replacing the lamp
to prevent burns.
• If the xenon lamp is used beyond its maximum service life of 500 hours
(see chapter "Maintenance intervals" on Page 222), sudden failure may
occur and interrupt operation. Change to the backup lamp or replace the
lamp module in a timely manner.
• If the halogen lamp is used beyond its maximum service life of 50 hours,
sudden failure may occur and interrupt operation. Replace the halogen
lamp in a timely manner (see chapter "Care and Maintenance" on Chang-
ing the halogen lamp on page 1).
• Do not pull at the power cord or any other connecting cables.
• If a failure occurs which you cannot correct with the aid of the chapter
"What to do in the event of malfunctions attach a sign to the system stat-
ing it is out of order and inform ZEISS Service.
• Never operate the system unattended.
• Always switch off the system
- before disconnecting it from / connecting it to line power
- if it will not be used for a prolonged period of time
- or for cleaning purposes.
Version 7.0
G-30-1781-en Page 15
Safety Measures S100 / OPMI pico
• Always use the master switch to turn off the device if it is not in use.
• Insufficient, incorrect or wrong cleaning or disinfection contrary to the di-
rections in these Instructions for Use may expose the patient or medical
staff to a considerable risk of infection. For more information, please see
"Maintenance of the device" in the chapter "Care and Maintenance".
• If the light guide is not connected at both ends, there may be a risk of fire
or burning when the light source is activated.
• Risk of crushing - mind your fingers!
Fingers may be crushed in the areas marked with the "Risk of crushing"
label.
Do not touch these areas while the system is being moved.
Loss of warranty!
The manufacturer is not liable for damages caused by an unauthorized use
of the device. Furthermore, this will this will invalidate any rights to make a
claim under the warranty.
Version 7.0
Page 16 G-30-1781-en
S100 / OPMI pico Safety Measures
General
This device is optionally equipped with a powerful xenon light source.
Various factors contribute to the risk of burn injuries:
Device-related factors:
– With increasing service life of the lamp, the actual light intensity delivered
at the respective setting decreases. In case of a bulb exchange, illumina-
tion increases to the originally high value.
Surgery-related factors:
– The selected intensity of the light source is a major factor for the risk of
injury. It should always be set to the minimum required for the surgical
procedure to be performed.
– A long surgical procedure increases the risk of injury, in particular if a
standard procedure takes considerably longer than usual.
– Injuries in the peripheral area may be avoided by covering the respective
parts with wet, sterile gauze.
The gauze must be moistened at regular intervals to avoid that the area
dries out or heats up.
– Certain preparations of the surgical field, local vasoconstrictive medica-
tions and incision drapes may also result in a higher risk of injury (drapes
may heat up to varying degrees depending on their color and moisture
content).
– You should also take into account that some areas of the body may be
more sensitive than others.
Patient-related factors:
– The general condition of a patient's health may contribute to the risk of
injury.
– The skin type may also play a major role for the risk of injury.
Version 7.0
G-30-1781-en Page 17
Safety Measures S100 / OPMI pico
Recommendations
Due to the large number of different factors involved and the lack of scientific
publications on this topic, Carl Zeiss cannot provide guidance on acceptable
illumination intensities and exposure durations. However, this system has sev-
eral features that can help the user to reduce the risk of burns.
– The start value of the illumination intensity can be set to a low value.
– The risk of burns can be reduced by constantly by irrigating the illuminated
surgical field and by keeping it moist.
– If used, drapes should also be re-moistened at regular intervals in order to
prevent heat from accumulating underneath the drape.
– Switch off the light when the microscope is not used, and make sure that
it is not pointed at unprotected bare skin.
Please keep in mind that the most published cases with burns refer to the skin
around the incision. The most important measure to avoid burns is covering
the peripheral area with wet, sterile gauze. The area of the incision should be
constantly moistened.
Final remark
ZEISS recommends that the lighting of the surgical field be reduced as far as
is possible for the safety of the patient and for a good microscopic image.
Please refer to the warnings and precautions in this case.
Version 7.0
Page 18 G-30-1781-en
S100 / OPMI pico Safety Measures
Safety devices
Version 7.0
G-30-1781-en Page 19
Safety Measures S100 / OPMI pico
Version 7.0
Page 20 G-30-1781-en
S100 / OPMI pico Safety Measures
Version 7.0
G-30-1781-en Page 21
Safety Measures S100 / OPMI pico
4 Friction symbol
Explanation of the braking effect achieved by turning the friction adjust-
ment knob.
- Left turn: decrease braking effect
- Right turn: increase braking effect
5 Marking the magnification changer
Marks the magnification setting.
1
1.6 6 Labeling magnification changer
Displays the magnification value
2.5
0.4
0.6
1
Version 7.0
Page 22 G-30-1781-en
S100 / OPMI pico Safety Measures
5 6
5
1 6
2
3
11
5 6
1
5
6
2
3
Version 7.0
G-30-1781-en Page 23
Safety Measures S100 / OPMI pico
3 Brightness setting
If the light source is turned on, the brightness can be continuously set by
turning the brightness control of the fiber illumination:
- Turn left: Minimum brightness
- Turn right: Maximum brightness
Version 7.0
Page 24 G-30-1781-en
S100 / OPMI pico Safety Measures
2 1 2 3
4
2 5 1
2
1
1
2 1 2 3
5
4
1
2
Version 7.0
G-30-1781-en Page 25
Safety Measures S100 / OPMI pico
Version 7.0
Page 26 G-30-1781-en
S100 / OPMI pico Safety Measures
1
2
3
Version 7.0
G-30-1781-en Page 27
Safety Measures S100 / OPMI pico
Version 7.0
Page 28 G-30-1781-en
S100 / OPMI pico Safety Measures
5
6
Version 7.0
G-30-1781-en Page 29
Safety Measures S100 / OPMI pico
Version 7.0
Page 30 G-30-1781-en
S100 / OPMI pico Safety Measures
1
2
3
Version 7.0
G-30-1781-en Page 31
Safety Measures S100 / OPMI pico
Version 7.0
Page 32 G-30-1781-en
S100 / OPMI pico Safety Measures
5
6
7
Version 7.0
G-30-1781-en Page 33
Safety Measures S100 / OPMI pico
Version 7.0
Page 34 G-30-1781-en
S100 / OPMI pico Safety Measures
1
2
3
Version 7.0
G-30-1781-en Page 35
Safety Measures S100 / OPMI pico
Version 7.0
Page 36 G-30-1781-en
S100 / OPMI pico Safety Measures
1
2
4
1
3
4
Version 7.0
G-30-1781-en Page 37
Safety Measures S100 / OPMI pico
Version 7.0
Page 38 G-30-1781-en
S100 / OPMI pico Safety Measures
1
2
Version 7.0
G-30-1781-en Page 39
Safety Measures S100 / OPMI pico
3 Fuse label
This plate contains the following information:
– Maximum voltage 115V/ 230V
– Rated current 6.3 A-H / 3.15 A-H
– (Tripping) characteristics T
Version 7.0
Page 40 G-30-1781-en
S100 / OPMI pico Safety Measures
Version 7.0
G-30-1781-en Page 41
Safety Measures S100 / OPMI pico
6 7
9 8
Version 7.0
Page 42 G-30-1781-en
S100 / OPMI pico Safety Measures
Version 7.0
G-30-1781-en Page 43
Safety Measures S100 / OPMI pico
5 Warning label
The xenon light source shall not be used for ophthalmological surgeries.
Make sure that no xenon light enters the patient's eyes.
6 "CAUTION" warning plate
The lamp housing contains an integrated xenon light source.
7 Replace Lamp
The plate graphically shows the three steps involved in changing the lamp.
A - Press the button
B - Pull out the lamp insert module
C - Change bulbs
8 1 2
Version 7.0
Page 44 G-30-1781-en
S100 / OPMI pico Safety Measures
9 "CAUTION" label
Warning of a potential hazard
8 9 10
11
Version 7.0
G-30-1781-en Page 45
Safety Measures S100 / OPMI pico
Version 7.0
Page 46 G-30-1781-en
S100 / OPMI pico Safety Measures
5 4 3 2
Version 7.0
G-30-1781-en Page 47
Safety Measures S100 / OPMI pico
Version 7.0
Page 48 G-30-1781-en
S100 / OPMI pico Safety Measures
1 2
1
2
Version 7.0
G-30-1781-en Page 49
Safety Measures S100 / OPMI pico
Version 7.0
Page 50 G-30-1781-en
S100 / OPMI pico System Overview
System Overview
System overview..............................................................................52
Version 7.0
G-30-1781-en Page 51
System Overview S100 / OPMI pico
System overview
Surgical microscope The surgical microscope (4) enables the illumination and visual magnification
of the field of view. For visual magnification, the surgical microscope can be
equipped with different tubes, eyepieces and objective lenses.
S100 suspension system The S100 comes as a floor stand, ceiling mount or wall mount. It is also avail-
able as a support unit and for adaptation to a Centro stand.
The S100 is used to control and supply the functions of the surgical micro-
scope. It consists of a carrier arm (2), suspension arm (3) and supply unit (1)
with light source. The suspension arm allows rough positioning of the surgical
microscope and moving into standby position.
The S100 floor stand can be positioned in the OR by four controllable
casters (5). The locking tabs of the casters prevent the system from being
moved or rolling away unintentionally.
Light source Depending on the order, the device may be equipped with a halogen, xenon
or LED light source. The halogen, xenon or LED light source is used to illumi-
nate the surgical area. It has a fiber illumination in the surgical microscope,
i. e. the light is directed from the light source via a light guide to the surgical
microscope. The halogen, xenon or LED light source is situated in the lamp
housing of the supply unit (1).
The halogen light source contains two halogen lamps. The second lamp is
used as backup lamp.
The xenon light source produces light whose spectrum is similar to daylight.
Regardless of the brightness setting, the color temperature of the light always
remains the same. Normal daylight film without any additional conversion fil-
ters can therefore be used for photographic documentation.
The xenon light source contains a lamp module with two xenon lamps. The
second lamp is used as backup lamp.
The LED light source has an integrated failure concept. In the event of a defect
in the LED light source, the light intensity drops to 50% and the control lamp
on the lamp housing lights up.
Version 7.0
Page 52 G-30-1781-en
S100 / OPMI pico System Overview
Ceiling-mounted stand
2
Wall-mounted stand
3
1
2 4
Carrier unit
1 3
Centro stand 2
3
1
4
2
1 3 4
Floor stand
Version 7.0
G-30-1781-en Page 53
System Overview S100 / OPMI pico
OPMI pico
Version 7.0
Page 54 G-30-1781-en
S100 / OPMI pico System Overview
5
4
1
2
3
3 2
Version 7.0
G-30-1781-en Page 55
System Overview S100 / OPMI pico
Version 7.0
Page 56 G-30-1781-en
S100 / OPMI pico System Overview
7
8a
3
3
10
1
7
3 8b
3
10
3 8c
Version 7.0
G-30-1781-en Page 57
System Overview S100 / OPMI pico
Version 7.0
Page 58 G-30-1781-en
S100 / OPMI pico System Overview
2
3
Version 7.0
G-30-1781-en Page 59
System Overview S100 / OPMI pico
Fig. 27: Controls for the tubes 180° tiltable tube 45° inclined tube
1 2 1 2
Straight tube
1
Version 7.0
Page 60 G-30-1781-en
S100 / OPMI pico System Overview
Version 7.0
G-30-1781-en Page 61
System Overview S100 / OPMI pico
Use The MORA interface is intended exclusively for use on the OPMI pico surgical
microscope in dentistry.
Description The MORA interface increases the microscope's mobility about its lateral tilt
axis, without changing the user's upright, ergonomic seated posture.
This offers advantages for the treatment of relatively large fields of therapy,
e.g. when several teeth are being treated at once.
A manually operated annular diaphragm permits the depth of field to be
varied.
With maximum lateral tilt of the microscope body and high magnification,
slight vignetting of the field of view may occur.
Version 7.0
Page 62 G-30-1781-en
S100 / OPMI pico System Overview
approx. ±25°
Version 7.0
G-30-1781-en Page 63
System Overview S100 / OPMI pico
1 2 3 4
Version 7.0
Page 64 G-30-1781-en
S100 / OPMI pico System Overview
Use The Endoport endoscope port is intended for use on OPMI pico microscopes
in ENT surgery.
Description The endoscope port of the OPMI pico Endoport is a standardized interface for
endoscope cameras. This port permits an external endoscope camera with
special adapter to be connected using the quick-action locking mechanism.
If you want to equip the surgical microscope with the Endoport accessory,
please observe the requirements for system expansions, see Page 251.
Only attach external endoscope cameras including adapter with a maximum
weight of 1.0 kg to the endoscope port.
The endoscope port always produces an image at infinity; for this reason, the
external endoscope camera's focus must also be set at infinity. The procedure
of focusing the endoscope camera at infinity is a camera-specific procedure
and differs from camera type to camera type.
Microscope settings such as the focusing of the eyepieces do not affect the
endoscope port.
Version 7.0
G-30-1781-en Page 65
System Overview S100 / OPMI pico
1 Carrier arm
2 Suspension arm
The suspension arm is equipped with a switch. This switch activates the
light source when the suspension arm is moved into its working position.
The light source is switched off in the upper rest position.
3 Mobility adjustment screw
You are able to set the lifting mobility of the suspension arm with the ad-
justment screw after the successful balance setting of the microscope
equipment.
4 Mobility adjustment screw
Use this screw to set the mobility of the coupling's rotation.
5 Mobility adjustment screw
Use this screw to set the mobility of the surgical microscope's swing
movement.
6 120° coupling
The 120° coupling improves ergonomic handling.
7 K 0/120 coupling
We recommend using the K 0/120 coupling for gynecological procedures.
This makes it possible to shift the working area downwards by 115 mm.
8 Securing screws
9 Adjustment screw for weight compensation
10 Mobility adjustment screw
You can use this screw to adjust the rotational mobility of the suspension
arm.
11 Mobility adjustment screw
You can use this screw to adjust the rotational mobility of the carrier arm.
Version 7.0
Page 66 G-30-1781-en
S100 / OPMI pico System Overview
2 3 4 5
11
8
10 9 6
2 3 4
8
8
7
8
Version 7.0
G-30-1781-en Page 67
System Overview S100 / OPMI pico
12
12
14
13
12
Version 7.0
Page 68 G-30-1781-en
S100 / OPMI pico System Overview
Attachment of carrier arm flat screen monitor to the S100 floor stand
(option)
Version 7.0
G-30-1781-en Page 69
System Overview S100 / OPMI pico
1 Ventilation grid
Do not cover the ventilation grid!
For example, drapes could be covering the screen. This could cause the in-
tegrated thermal switch to turn off the light source.
2 Flap
The flap is used as the mechanical indicator for the operating status of the
halogen lamps.
– When the flap is closed, the main lamp is operative.
– The open flap indicates that the backup lamp is being used.
3 Manual switching to the backup lamp
Pressing this button swivels in the backup lamp (Page 217)
4 Slide-in lamp module
5 Opening the slide-in lamp module
When you press this button, the slide-in lamp module is partially ejected.
Pull out the lamp module all the way for exchanging the lamp.
6 Brightness setting
If the light source is switched on, the brightness can be infinitely adjusted
by turning the brightness control button for the fiber illumination.
- Left position: Minimum brightness
- Right position: Maximum brightness
3 6
4
5
Version 7.0
Page 70 G-30-1781-en
S100 / OPMI pico System Overview
4
1
5
Version 7.0
G-30-1781-en Page 71
System Overview S100 / OPMI pico
Version 7.0
Page 72 G-30-1781-en
S100 / OPMI pico System Overview
2
3 4
Version 7.0
G-30-1781-en Page 73
System Overview S100 / OPMI pico
4
5
Version 7.0
Page 74 G-30-1781-en
S100 / OPMI pico System Overview
Version 7.0
G-30-1781-en Page 75
System Overview S100 / OPMI pico
Depending on the integrated video camera selected, the supply unit contains
different video control units:
For the integrated SD video camera
– Integrated HD 720p control unit
– Integrated video camera live & streaming control unit
with SW recording license (optional SW license)
The area to the right of the dashed line is displayed with a 90° rotation. It is
located at the level of the connector panel.
1 Config
Use the <config> key to go to the menu.
The configuration menu permits you to set various function sequences to
meet your specific needs. The different possible settings in the configura-
tion of the application program are explained in the chapter "Operation"
(Page 127). The camera is delivered with a default configuration (factory
settings).
2 Indicator light (integrated HD 720p control unit)
The indicator light (LED) is lit when the power switch has been turned on.
3 Indicator lights (integrated video camera live & stream control unit)
The indicator light "power" is lit when the power switch has been turned
on.
The "network" indicator light flashes during network data transmission,
The indicator lamp "record" is lit when recording a video.
4 White bal
White balance for integrated HD 720p control unit (Page 165)
and for the integrated video camera live & streaming control unit
(Page 185)
5 Brightness (integrated HD 720p control unit)
You can use the and keys to adjust the brightness of the video
image.
The configuration menu () gives you the choice between two different
modes: Automatic - Manual.
6 Brightness (integrated video camera live & streaming control unit)
Use the <brightness> key to go to the "Brightness" menu (Page 182)
where you can configure settings.
Version 7.0
Page 76 G-30-1781-en
S100 / OPMI pico System Overview
3 4 6 1
Version 7.0
G-30-1781-en Page 77
System Overview S100 / OPMI pico
7 Image
The <freeze> button allows you to freeze the current live video image and
to display it as a full-screen freeze frame on the video monitor. In the
meantime the text "freeze" appears in the upper left corner.
When you press the <live> key, the current live video image is displayed
again on the monitor. The text "freeze" disappears.
8 Colors (integrated video camera live & streaming control unit)
Use the <colors> key to go to the "Colors" menu (Page 186) where you
can configure the color settings.
9 Live (integrated video camera live & streaming control unit)
Closes all open dialogs, discontinues all actions except <record>. When
you press the <live> key, the current live video image is displayed again
on the monitor.
10 "Select" control keys (cursor keys)
Use can use the control keys to navigate the menu. In addition, the cursor
keys have special functions while an image is being displayed or a video is
being played (Page 80).
11 "Enter" key ( )
Press the <enter> key to confirm a selection or trigger a function. The dif-
ferent functions are described in the respective menu item of the menu.
In addition, textual information for the selected mode is displayed.
Version 7.0
Page 78 G-30-1781-en
S100 / OPMI pico System Overview
8 9 10 11
Version 7.0
G-30-1781-en Page 79
System Overview S100 / OPMI pico
12 Record button
Press the <record> button to start a video recording. All open menus will
be closed and an image display or video playback will be terminated. The
video recording starts.
In the footer bar the text "Rec" appears on the right before USB/LAN and
the recording time appears on the left. The "record" indicator light (3) will
light up.
Press the <record> button to end a video recording. The text "Rec" and
the recording time appears. The "record" indicator light (3) goes out.
13 Capture button
Images can be created in live mode and during video recording or
video playback. Press the <capture> button to trigger the image capture.
The text "Capture" appears on the left in the header bar of the display.
The <Capture> button can be used to create individual images during
video recording or playback.
14 Play button
Press the <play> button to display the last image captured or to play the
last video recorded.
If the <play> button is pressed during video playback, the current display
appears as a still image. The text "Pause" appears on the left in the header
bar of the display.
Press the <play> button again to continue playing the video.
15 Index button
Press the <Index> button to open the index view of the files of the acti-
vated patient. Use the control keys to select a video or image. A blue
frame marks the respective selected file.
Press the <index>, <config> or <live> button, to terminate the index view.
16 Function key (F1)
The <function> key can be programmed individually (Page 190).
17 Function key (F2)
The <function> key can be programmed individually (Page 190).
18 Right cursor key (next image/10% skip forward)
With the right cursor key you can switch to the next image (in the list
below, older) while an image is being displayed, and while a video is being
played you can move forward 10% and continue playing the video from
there.
Version 7.0
Page 80 G-30-1781-en
S100 / OPMI pico System Overview
Version 7.0
G-30-1781-en Page 81
System Overview S100 / OPMI pico
Remote control for the video camera live & streaming control unit (option)
The button functions of the remote control are identical to those of the oper-
ator panel on the device.
The following functions can only be carried out on the HD control unit:
– White bal
– Brightness
– Colors
– Function F1
– Function F2
Version 7.0
Page 82 G-30-1781-en
S100 / OPMI pico System Overview
5 Index 2)
Index view of the files of the activated patient (Page 207)
6 Play
Play the video (Page 207):
Press the <Play> button: Toggles between play video / pause video
Press the <Live> button to stop the playback
7 Record 1) 2)
Start / stop recording (Page 205):
Press the <Record> button: Toggles between start recording /
stop recording
Without SW recording license:
Press the <Record> button: Toggles between streaming ON/OFF
8 Capture 1) 2)
Show (Page 207) / create (Page 205) individual images (photo)
Without SW recording license:
Press the <Capture> button: Toggles between Freeze/Live
9 Menu (Page 176)
1)
Functions without SW recording license
2) Functions in the web interface (see Page 201)
8
3
7
4 6
8 4 5
5
7 2
6
Version 7.0
G-30-1781-en Page 83
System Overview S100 / OPMI pico
Here you will find an overview of the different operating elements of the in-
tegrated HD video camera live & streaming (full HD 1080p) with SW recording
license and its functions.
Version 7.0
Page 84 G-30-1781-en
S100 / OPMI pico System Overview
Press the <config> key of the remote control to go to the menu. From there
you can navigate the menu using the control keys and the "enter" key and this
way configure the integrated video camera live & streaming control unit.
Version 7.0
G-30-1781-en Page 85
System Overview S100 / OPMI pico
10 DVI output
Analog connection for video components.
Version 7.0
Page 86 G-30-1781-en
S100 / OPMI pico System Overview
6 7
1
8
9
10
4
Version 7.0
G-30-1781-en Page 87
System Overview S100 / OPMI pico
Version 7.0
Page 88 G-30-1781-en
S100 / OPMI pico Preparations for use
Connections................................................................................... 108
Connecting the S100 floor stand .............................................................108
Connecting the integrated video camera .................................................110
Connecting the computer to integrated video camera live & streaming Control
unit with LAN connection ........................................................................114
Version 7.0
G-30-1781-en Page 89
Preparations for use S100 / OPMI pico
Version 7.0
Page 90 G-30-1781-en
S100 / OPMI pico Preparations for use
Version 7.0
G-30-1781-en Page 91
Preparations for use S100 / OPMI pico
After you have mounted the OPMI pico, the suspension arm may move
downwards. Prevent the OPMI pico from knocking against anything.
The adjustment screw (1) on the suspension arm of the S100 suspension
system only acts as a brake but has no locking function.
Without coupling
• Tighten adjustment screw (1) (braking function).
• Insert the shaft (5) into the the suspension arm's holding fixture.
• Screw in and tighten the safety screws (2) and (4).
• The safety screw (2) contains a hex socket head screw (3), which is an ad-
ditional safety screw. Tighten it.
Including 120° coupling
• Tighten adjustment screw (1)(braking function).
• Insert the shaft (6) into the mount of the suspension arm.
• Screw in and tighten the safety screws (2) and (4).
• The safety screw (2) contains a hex socket head screw (3). This is an addi-
tional safety screw. Tighten it.
• Insert shaft (5) into the holding fixture of the coupling. Position and tight-
en the safety screws (7) and (8).
• Securing screw (8) contains a hex socket head screw (9) which is an addi-
tional securing screw. Tighten it.
Including K 0/120 coupling
We recommend using the K 0/120 coupling for gynecological procedures.
This makes it possible to shift the working area downwards by 115 mm.
• Tighten adjustment screw (1)(braking function).
• Insert the shaft (10) into the the suspension arm's holding fixture.
• Screw in securing screws (2) and (4). Tighten it.
• Securing screw (2) contains a hex socket head screw (3) which is an addi-
tional securing screw. Tighten it.
• Insert shaft (5) into the holding fixture of the coupling. Position and tight-
en the safety screws (11).
Version 7.0
Page 92 G-30-1781-en
S100 / OPMI pico Preparations for use
120° coupling
K 0/120 coupling 9 8 7 6
10 5
11
Version 7.0
G-30-1781-en Page 93
Preparations for use S100 / OPMI pico
Move the suspension arm into its upper position before dismantling the
surgical microscope. While the surgical microscope is dismounted, the
suspension arm may shoot up by its own spring force despite tightened
locking screw.
Without coupling
• Tighten adjustment screw (1)(braking function).
• Hold the surgical microscope securely while loosening the safety screws.
• Loosen safety screw (3).
• Loosen and remove the safety screws (2) and (4).
• Remove the surgical microscope.
Including 120° coupling
• Tighten adjustment screw (1)(braking function).
• Hold onto the surgical microscope firmly.
• Loosen safety screw (9).
• Loosen and remove the safety screws (7) and (8).
• Remove the surgical microscope.
• Hold onto the surgical microscope firmly.
• Loosen safety screw (3).
• Loosen and remove the safety screws (2) and (4).
• Remove the coupling.
Including K 0/120 coupling
• Tighten adjustment screw (1)(braking function).
• Hold onto the surgical microscope firmly.
• Loosen safety screw (11).
• Remove the surgical microscope.
• Hold onto the surgical microscope firmly.
• Loosen safety screw (3).
• Loosen and remove the safety screws (2) and (4).
• Remove the coupling.
Version 7.0
Page 94 G-30-1781-en
S100 / OPMI pico Preparations for use
120° coupling
K 0/120 coupling 9 8 7 6
10 5
11
Version 7.0
G-30-1781-en Page 95
Preparations for use S100 / OPMI pico
1 2 3 4 5
Version 7.0
Page 96 G-30-1781-en
S100 / OPMI pico Preparations for use
Five handgrips are available for the OPMI pico surgical microscope (see
Page 58).
The handgrips can be positioned as required by reinserting (2) or pressing
them together (1).
• Select the handgrip suitable for your application.
• Mount the handgrip supplied with the instrument. Firmly tighten the
screw(s).
Version 7.0
G-30-1781-en Page 97
Preparations for use S100 / OPMI pico
Make sure that the light guide is not stretched or bent when the micro-
scope is turned or tilted.
Version 7.0
Page 98 G-30-1781-en
S100 / OPMI pico Preparations for use
Prerequisites Your OPMI pico has been equipped with the "Endoport" accessory com-
ponent:
Your endoscope camera, including adapter, does not weigh more than
1.0 kg.
• Focus the endoscope camera at infinity.
The procedure of focusing the endoscope camera at infinity is a camera-
specific procedure and differs from camera type to camera type.
Microscope settings such as the focusing of the eyepieces do not affect
the endoscope port.
• Remove the cover from the endoscope port.
• Attach your endoscope camera to the endoscope port and allow the
quick-action lock to snap in.
• Connect the video cable of the endoscope camera to the endoscope sys-
tem. Please also observe the information contained in the user's manual
on the endoscope system.
• Adjust the orientation of the image on the video monitor. To achieve this,
turn the endoscope camera in the endoscope port as required.
Version 7.0
G-30-1781-en Page 99
Preparations for use S100 / OPMI pico
2
7
3
8
4
Version 7.0
Page 100 G-30-1781-en
S100 / OPMI pico Preparations for use
5 6 7
approx. ± 25°
Version 7.0
G-30-1781-en Page 101
Preparations for use S100 / OPMI pico
Requirement The MORA interface with lateral port for documentation accessories has
been mounted on the surgical microscope.
The method described below can also be used for other accessories. The op-
erating principle of the accessories is described in the relevant Instructions for
Use.
Attaching documentation • Loosen knurled ring (2). An arrow with the labeling "open" is located on
accessories the shaft next to knurled ring (2).
• Remove dust cover (3) and store it in a safe place.
• Slide the accessories (4) in the image exit port holding fixture .
The mount of the image exit port is fitted with guide projections. Acces-
sory (4) has the corresponding grooves.
• Carefully turn the accessory module until the guide projections fit into the
grooves,
• and slide the accessory in the mount as far as it will go.
• Screw knurled ring (2) onto accessory module (3).
• Firmly tighten knurled ring (2).
If the accessory is mounted with the connector pointing vertically up-
wards, the image of the connected camera will be displayed with a 90°
rotation.
• To avoid this, you can mount the accessory rotated by 90° so that the con-
nector (incl. camera) points horizontally to the back.
• Note the effect of annular diaphragm (1) swung into the beam path.
1 2 3 4
Version 7.0
Page 102 G-30-1781-en
S100 / OPMI pico Preparations for use
2
3
4
Version 7.0
G-30-1781-en Page 103
Preparations for use S100 / OPMI pico
Adjustment instructions
– In order to be able to quickly set/adjust the surgical microscope, it is ad-
visable to save the interocular distance as well as the refraction values for
the different users so they can be preset within the scope of the prepara-
tions.
– Do not wear multifocal or varifocal glasses during the adjustment process
or when working with the microscope later on.
Such glasses make a proper diopter adjustment ring setting impossible
and result in unsatisfactory imaging.
– Optimum image clarity in the eyepieces and at the optical port for docu-
mentation equipment (photo, video) can only be achieved through careful
focusing and precisely adjusting the binocular tube.
Adjusting the interocular • Move the microscope to a vertical position above a flat object, e.g. a sheet
distance of paper with writing.
• Adjust the distance of the eyepieces at the binocular tube to your interoc-
ular distance so that the two eyepiece images (object and edge of visual
field) merge into one.
Adjusting the eyepieces The following procedure must be performed separately for each eyepiece in
the order described.
• Set the microscope to minimum magnification.
• Set the diopter adjustment ring at the eyepiece to zero position.
• Look through the eyepiece and move the microscope body to focus the
image.
• Set the microscope to maximum magnification and adjust the fine focus-
ing system until the image is sharply defined.
• Then reset the microscope to minimum magnification without changing
the working distance.
• Set the diopter adjustment ring at the eyepiece to its maximum value (plus
direction).
• Look through the eyepiece and slowly turn the diopter adjustment ring to
the minus direction until the image is sharply defined.
• Repeat the entire procedure for the second eyepiece.
Version 7.0
Page 104 G-30-1781-en
S100 / OPMI pico Preparations for use
The microscope is now adjusted in such a way that the image is sharply
defined over the complete magnification range without any renewed fo-
cusing during the magnification setting.
If this is not the case, repeat the procedure.
Adjusting the eyecups • Adjust the eyecups of the eyepieces in such a way that the entire visual
field can be seen.
– Viewing with eyeglasses: Rotate eyecups inwards.
– Viewing without eyeglasses: Rotate eyecups outwards.
Version 7.0
G-30-1781-en Page 105
Preparations for use S100 / OPMI pico
Please observe the following points when relocating the suspension system:
• Turn off the device at the power switch.
• Remove the power plug from the wall socket.
• Disconnect the video cable between the video modules (e.g. video moni-
tor, video recorder) and the camera control unit (CCU).
• Adjust the flat screen (optional) to a secure transport position.
• Do not transport the device by pulling or pushing on the optional flat
screen!
• Avoid any uncontrolled swiveling of the optional flat screen.
• Swing the suspension arm into the transport position.
• Unlock all arrestors of the casters.
• Hold the system by the marked position on the stand column and carefully
push it along. Make sure that the system is positioned on level ground.
• Press at least two of the arrestors and make sure that the suspension sys-
tem is no longer able to roll away by itself.
The mass may be underestimated as the suspension system can be smoothly
moved. Therefore, move the suspension system slowly and carefully!
Version 7.0
Page 106 G-30-1781-en
S100 / OPMI pico Preparations for use
Version 7.0
G-30-1781-en Page 107
Preparations for use S100 / OPMI pico
Connections
• Ensure that the rated voltage indicated in the display window (3) (only for
halogen/xenon) corresponds with the rated voltage for the network at the
site of the installation.
• Ensure that the unit has been switched off at the power switch before
plugging in or unplugging connectors at the power input socket (4).
• Make sure that power switch (1) is OFF.
• Plug the supplied power cord on the supply unit into the power inlet sock-
et (4).
Version 7.0
Page 108 G-30-1781-en
S100 / OPMI pico Preparations for use
• Secure the power cord using the strain relief device (2).
• If necessary, connect the potential equalization connector (5) of the sys-
tem with the potential equalization contact at the site of use.
Xenon
1 3
4
5
2
LED
3
2 5
Version 7.0
G-30-1781-en Page 109
Preparations for use S100 / OPMI pico
• Slide the dust cover (3) on the system cable (camera side). Make sure that
the thread faces the connector plug on the camera side.
• Connect the camera head (1) of the surgical microscope to the camera
control unit (CCU) (5) using the included system cable (4).
• Screw the dust cover (3) into the thread (2).
Make sure not to screw on the dust cover (3) crookedly. This may damage
the thread!
• Firmly tighten screws on the camera head connector.
Version 7.0
Page 110 G-30-1781-en
S100 / OPMI pico Preparations for use
Version 7.0
G-30-1781-en Page 111
Preparations for use S100 / OPMI pico
2 FBAS
VIDEO OUT
In the HD video control unit the video signal is scaled down in the SD video
output (FBAS VIDEO OUT) for SD devices.
Version 7.0
Page 112 G-30-1781-en
S100 / OPMI pico Preparations for use
2 HDMI 2 HDMI
VIDEO OUT VIDEO OUT
2 DVI 2 DVI
VIDEO OUT VIDEO OUT
2 YPbPr
VIDEO OUT
Version 7.0
G-30-1781-en Page 113
Preparations for use S100 / OPMI pico
Connecting the computer to integrated video camera live & streaming Control
unit with LAN connection
You can connect the integrated video camera live & streaming control unit
with LAN connection to a computer . Observe the conditions for system ex-
tension (see Page 251).
The network cable is available as an accessory (see Accessories).
Optional software is not included in our delivery package.
System requirements
Computer with LAN connection: Network at least 100 Mbit
Requirements for connecting the network computer:
– You will need administrator rights for the network computer.
– You will need the Ethernet data of the archiving PC to configure the
network share.
– You can obtain the data required for this purpose from the appropri-
ate IT administrator.
Video software
• Connect the computer with the corresponding network cable with the
LAN connection on the video unit (on the stand).
Version 7.0
Page 114 G-30-1781-en
S100 / OPMI pico Preparations for use
Version 7.0
G-30-1781-en Page 115
Preparations for use S100 / OPMI pico
Ergonomic positioning the carrier arm flat screen monitor and instrument tray
In order to ergonomically position carrier arm flat screen monitor (2) and in-
strument tray (3), loosen the adapter (1), adjust side and height and screw it
tight again:
Instrument tray • Hold the instrument tray with one hand and unscrew the four locking
screws at the adapter (1) with a few turns using the hex key until the
adapter (1) can easily be moved.
• Position instrument tray (3) in the required height (max. 800 mm above
the floor) and uniformly tighten (2.1 Nm) all four locking screws of
adapter (1) using the hex key.
• When attaching adapter (1), make sure the two components of the
adapter (1) are separated by the same distance (4) from top to bottom.
Carrier arm flat screen monitor • Hold the carrier arm flat screen monitor with one hand and unscrew the
four locking screws at the adapter (1) with a few turns using the hex key
until the adapter (1) can easily be moved.
• Position the carrier arm flat screen monitor (2) in the required height (max.
1230 mm above the floor) and uniformly tighten (2.1 Nm) all four locking
screws of adapter (1) using the hex key.
• When attaching adapter (1), make sure the two components of the
adapter (1) are separated by the same distance (4) from top to bottom.
Version 7.0
Page 116 G-30-1781-en
S100 / OPMI pico Preparations for use
1
3 max.
1230 mm
max.
800 mm
1 4
4x
Version 7.0
G-30-1781-en Page 117
Preparations for use S100 / OPMI pico
Mounting the flat screen monitor on the carrier arm flat screen monitor
Version 7.0
Page 118 G-30-1781-en
S100 / OPMI pico Preparations for use
11 2 3
1 1
4 5 6
Version 7.0
G-30-1781-en Page 119
Preparations for use S100 / OPMI pico
Version 7.0
Page 120 G-30-1781-en
S100 / OPMI pico Preparations for use
Version 7.0
G-30-1781-en Page 121
Preparations for use S100 / OPMI pico
After mounting the carrier arm flat screen monitor, you can adjust the friction
of its joints as required. The factory setting of the joints is suitable for a me-
dium load capacity which meets the requirements of most users. If one joint
is set too stiffly or too freely, it is possible to adjust the friction.
Adjusting the tilt function of the carrier arm flat screen monitor
• Check the tilt function of the monitor joint (upward/downward move-
ment).
• Use the tool supplied to increase or reduce the friction of the joint as re-
quired.
Version 7.0
Page 122 G-30-1781-en
S100 / OPMI pico Preparations for use
Adjusting the swivel movement of the carrier arm flat screen monitor
If the carrier arm flat screen monitor does not remain in place in the position
required, the following readjustments can be performed.
• Remove plastic cover (1) at the joint.
• Tighten the adjustment screw (2) clockwise until the required friction is
achieved.
• Reattach plastic cover (1).
• Check if the securing screw (3) is firmly tightened.
Version 7.0
G-30-1781-en Page 123
Preparations for use S100 / OPMI pico
Version 7.0
Page 124 G-30-1781-en
S100 / OPMI pico Preparations for use
Version 7.0
G-30-1781-en Page 125
Preparations for use S100 / OPMI pico
Version 7.0
Page 126 G-30-1781-en
S100 / OPMI pico Operation
Operation
Version 7.0
G-30-1781-en Page 127
Operation S100 / OPMI pico
Prerequisites The correct rated voltage has been set on the suspension system (please
refer to Page 108).
The included power cord (floor stand) is connected (please refer to
Page 108).
The light guide is connected (please refer to Page 98).
Proceed as follows to power the system up and down:
• Press the power switch (1) on the suspension system.
Before starting to work with the system, you have to perform a function test
(see next page). This permits you to check the system configuration and user-
specific settings and to reconfigure them, if necessary.
1
Version 7.0
Page 128 G-30-1781-en
S100 / OPMI pico Operation
Version 7.0
G-30-1781-en Page 129
Operation S100 / OPMI pico
Version 7.0
Page 130 G-30-1781-en
S100 / OPMI pico Operation
The main and backup lamp are intact in the halogen light source.
Main and backup lamp are intact in the xenon light source.
In case of exceeding the maximum service life of 500h of the main
lamp and the backup lamp is in operation, have a backup lamp module
available for replacement purposes as precaution (see Page 226). For
the order number, please see the chapter "Device data".
With the LED light source, the amber LED is not illuminated.
It is possible to switch on or off the halogen, xenon or LED light source
by moving up and down the suspension arm.
The light source is turned on within the preset working area.
The light source is turned off in the upper resting position.
Version 7.0
G-30-1781-en Page 131
Operation S100 / OPMI pico
Typical procedure
Starting work
Version 7.0
Page 132 G-30-1781-en
S100 / OPMI pico Operation
Ending work
Version 7.0
G-30-1781-en Page 133
Operation S100 / OPMI pico
Test pattern
A test pattern (color bars) is displayed if the camera head is not connected
when you switch on the system.
• Switch off the system again, connect the camera head and switch the sys-
tem back on.
Menu structure
Menu display
– The menu text is displayed on the monitor in light green.
A light green arrow is displayed in front of a selected line.
Version 7.0
Page 134 G-30-1781-en
S100 / OPMI pico Operation
Version 7.0
G-30-1781-en Page 135
Operation S100 / OPMI pico
In the installation menu you can install device settings that are seldom
changed and that require the camera to be restarted.
• Press the <config> key for approximately 5 seconds.
This will take you to the following menu structure:
Version 7.0
Page 136 G-30-1781-en
S100 / OPMI pico Operation
During the initial start of the surgical microscope with integrated HD video
camera, the sub menu "APPLICATION VARIANT" will be displayed for the first
time (see Page 163).
The application is set in this menu.
ZEISS Service adjusts this setting for you during initial start-up. You can read-
just this setting in the "INSTALLATION MENU". see “Opening the installation
menu (config)” on Page 155.
The configuration menu allows you to make user-specific settings, such as se-
lecting a user or setting the brightness / color saturation.
• Press the <config> key to go to the configuration menu.
The following menu appears on the monitor:
USER User 1 - 4
BRIGHTNESS AUTOMATIC - LOW AUTO -
MANUAL
LIGHTMETER SMALL SPOT - LARGE SPOT -
INTEGRAL
ORIENTATION STANDARD - MIRROR - FLIP -
ROTATE
ADVANCED ...
SERVICE MENU ...
EXIT
Version 7.0
G-30-1781-en Page 137
Operation S100 / OPMI pico
• You can exit the configuration menu by selecting menu item "Exit" or by
pressing the <config> or <live> key. If changes have been made, a prompt
is displayed for each user on exit from the configuration menu as to
whether the settings should be saved. Answer this prompt with "YES" or
"NO".
An existing "freeze" state ends when the configuration menu is activated.
Version 7.0
Page 138 G-30-1781-en
S100 / OPMI pico Operation
Version 7.0
G-30-1781-en Page 139
Operation S100 / OPMI pico
Proceed as follows to select a user and to load and activate the respective con-
figuration:
• Press the <config> key to access the configuration menu.
• In the Configuration menu, select the menu item "User" with the con-
trol keys and press the <enter> key.
The following menu appears:
USER SELECTION
USER 1
USER 2
USER 3
USER 4
CANCEL
Version 7.0
Page 140 G-30-1781-en
S100 / OPMI pico Operation
Brightness submenu
In the "BRIGHTNESS" sub menu, you can select between three different expo-
sure controls: AUTOMATIC - LOW LIGHT AUTO - MANUAL
AUTOMATIC In this mode the camera adjusts the brightness of the video image automati-
cally to the value specified by the <brightness> keys and . Exposure con-
trol affects the exposure time of the camera.
The nominal brightness is set in the AUTOMATIC mode. The camera controls
the exposure time in the range between 1/50(60)s and 1/10000s. The expo-
sure time currently selected is not displayed on the monitor. The nominal
brightness can be adjusted between 0 and 100%. Changes are made in 5%
increments.
A required value in the range between 50% and 70% usually provides op-
timum exposure results.
LOW AUTO This mode has the same function as AUTOMATIC mode, however, exposure
times longer than 1/50 (60)s are allowed, which can result in blurred move-
ments.
MANUAL In this mode the exposure time selected is displayed when the control
keys are pressed. The setting range extends from 1/6 to 1/10000.
– Increments for PAL: 1/10000, 1/4000, 1/2000, 1/1000, 1/500, 1/250, 1/
100, 1/50, 1/25, 1/12 and 1/6
– Increments for NTSC: 1/10000, 1/4000, 1/2000, 1/1000, 1/500, 1/250, 1/
120, 1/60, 1/30, 1/15 and 1/8
Greater exposure times can result in blurred movements.
The set time disappears from the screen again if the <brightness> key or
was not pressed longer than 2 seconds.
Version 7.0
G-30-1781-en Page 141
Operation S100 / OPMI pico
BRIGHTNESS MODE
AUTOMATIC 80%*
LOW LIGHT AUTO 80%*
MANUAL 1/16*
CANCEL
Version 7.0
Page 142 G-30-1781-en
S100 / OPMI pico Operation
INTEGRAL: The exposure is measured and averaged across the full video image. This
measurement is recommended for surgical fields that are fully and evenly lit.
LARGE SPOT: The exposure is measured in an area in the image center. This measurement
is ideal for working with an eclipsed surgical field edge (resulting from a re-
duced illuminated field diameter).
SMALL SPOT: The exposure is measured in a very small area in the image center. This meas-
urement is suitable for work in narrow channels when an extremely small illu-
minated field diameter is being used.
However, if the object of interest is not located in the image center, the
SMALL SPOT setting usually does not provide the desired exposure result. In
this case, select either a larger area (INTEGRAL or LARGE SPOT) or use the
manual exposure mode.
Version 7.0
G-30-1781-en Page 143
Operation S100 / OPMI pico
Proceed as follows to select a exposure measurement type and to load the re-
spective configuration:
• Press the <config> key to access the configuration menu.
• Select the menu item "LIGHTMETER" in the configuration menu using the
control keys and press the <enter> key.
The following menu appears:
LIGHTMETER
SMALL SPOT
LARGE SPOT
INTEGRAL
CANCEL
Version 7.0
Page 144 G-30-1781-en
S100 / OPMI pico Operation
Orientation submenu
The "ORIENTATION" submenu allows setting the screen orientation. To do
this, proceed as follows:
• Press the "config" key to access the configuration menu.
• In the configuration menu, use the control keys to select menu item
"ORIENTATION" and press the "enter" key.
The following menu appears:
ORIENTATION
STANDARD Standard
MIRROR Mirror the image vertically*
FLIP Mirror the image horizontally*
ROTATE Turn the image through 180°*
CANCEL
The comments marked with * are only provided for information purposes
and do not appear on the monitor.
• Use the control keys to select the required mode.
The selection is displayed with an arrow.
• Press the "enter" key to save the changes.
The selection is saved and you return to the configuration menu.
• If you do not want to save any changes, select menu item "CANCEL" to
return to the configuration menu. As a result, the image orientation on the
screen will be reset to the setting present when the submenu was opened.
Version 7.0
G-30-1781-en Page 145
Operation S100 / OPMI pico
Advanced submenu
In the "ADVANCED" submenu, you can set the settings that are seldom used.
Proceed as follows to make changes:
• Press the <config> key to access the configuration menu.
• Select the menu item "ADVANCED" in the Configuration menu using the
control keys and press the <enter> key.
The following menu appears:
RED ±0
BLUE ±0
CHROMA ±0
COLOR ANGLE ±0
BLACK LEVEL ±0
ZOOM LEVEL 1.0 x
PEAK/AVERAGE 0
BACK
Version 7.0
Page 146 G-30-1781-en
S100 / OPMI pico Operation
CHROMA The camera's color saturation is intensified or reduced using the "Chroma"
setting.
COLOR ANGLE With the "Color Angle" setting (frequently also: HUE), you can change the
color angle as well as the hues of the camera.
BLACK LEVEL You can set the blackness value of the camera with the "Black Level" setting.
Reducing the black level generates a higher degree of contrast. Increasing a
value means that dark details are easier to recognize.
ZOOM LEVEL The camera image can be magnified electronically using the "Zoom Level" set-
ting.
Version 7.0
G-30-1781-en Page 147
Operation S100 / OPMI pico
PEAK / AVERAGE The setting "Peak / Average" is used to modify the lighting control character-
istics. A lower value results in peak value-based lighting control, which re-
sponds precisely to changes in the project brightness. A higher value sets
mean value-based lighting control, which responds less sensitively to changes
in brightness.
RED submenu In this submenu, you can set the red value of the camera in a range from -20
to +20. To do this, proceed as follows:
• Select menu item "RED" in the "ADVANCED" menu using the control
keys and press the "enter" key.
• Select the "SETTINGS" menu item with the control keys.
The selected choice is displayed with an arrow.
• Change the red value using the control keys.
• Press the "enter" key to save the changes.
The selection is saved and you return to the "ADVANCED USER SET-
TINGS" menu.
• If you do not want to save your changes, return to the "ADVANCED USER
SETTINGS" menu via "CANCEL". The red value is set as it was when the
submenu was accessed.
The red value setting also influences the camera's white balance. see “Per-
forming the white balance (white bal)” on Page 165.
BLUE submenu In this submenu, you can set the blue value of the camera in a range from
-20 to +20. To do this, proceed as follows:
• Select the "BLUE" menu item in the "USER MENU ADVANCED" menu us-
ing the control keys and press the "enter" key.
• Select the "SETTINGS" menu item with the control keys.
The selected choice is displayed with an arrow.
• Change the blue value using the control keys.
• Press the "enter" key to save the changes.
The selection is saved and you return to the "ADVANCED USER SET-
TINGS" menu.
• If you do not want to save your changes, return to the "ADVANCED USER
SETTINGS" menu via "CANCEL". The blue value is set as it was when the
submenu was accessed.
Version 7.0
Page 148 G-30-1781-en
S100 / OPMI pico Operation
The blue value setting also influences the camera's white balance. see “Per-
forming the white balance (white bal)” on Page 165.
CHROMA submenu In this submenu, you can set the color saturation of the camera in a range
from -10 to +10. To do this, proceed as follows:
• Select the "CHROMA" menu item in the "USER MENU ADVANCED" menu
using the control keys and press the "enter" key.
• Use the control keys to select "SETTINGS".
The selection is displayed with an arrow.
• Change the color saturation using the control keys.
A chroma setting in the range of ± 0 usually provides the best color im-
pression. Chroma setting is most effective with strongly saturated (bright)
colors. The effect is low with weakly saturated colors.
• Press the "enter" key to save the changes.
The selection is saved and you return to the "ADVANCED USER SET-
TINGS" menu.
If you do not want to save your changes, return to the "ADVANCED USER SET-
TINGS" menu via "CANCEL". As a result, color saturation will be reset to the
setting present when the submenu was opened.
COLOR ANGLE submenu In this submenu, you can set the color angle (hue) of the camera in a range
from -10 to +10. To do this, proceed as follows:
• Select the "COLOR ANGLE" menu item in the "USER MENU ADVANCED"
menu using the control keys and press the "enter" key.
• Use the control keys to select "SETTINGS".
The selection is displayed with an arrow.
• Change the color angle using the control keys.
• Press the "enter" key to save the changes.
The selection is saved and you return to the "ADVANCED USER SET-
TINGS" menu.
If you do not want to save your changes, return to the "ADVANCED USER SET-
TINGS" menu via "CANCEL". As a result, the color angle will be reset to the
setting present when the submenu was opened.
BLACK LEVEL submenu This submenu allows setting the black level of the camera in a range from -10
to +10. To do this, proceed as follows:
Version 7.0
G-30-1781-en Page 149
Operation S100 / OPMI pico
• Select the "BLACK LEVEL" menu item in the "USER MENU ADVANCED"
menu using the control keys and press the "enter" key.
• Use the control keys to select "SETTINGS".
The selection is displayed with an arrow.
• Change the black level value using the control keys.
• Press the "enter" key to save the changes.
The selection is saved and you return to the "ADVANCED USER SET-
TINGS" menu.
If you do not want to save your changes, return to the "ADVANCED USER SET-
TINGS" menu via "CANCEL". The black level is set as it was when the submenu
was accessed.
ZOOM LEVEL submenu In this submenu, you can set the zoom level of the camera in a range from
1.0x to 2.0x. To do this, proceed as follows:
• Select the "ZOOM LEVEL" menu item in the "USER MENU ADVANCED"
menu using the control keys and press the "enter" key.
• Use the control keys to select "SETTINGS".
The selection is displayed with an arrow.
• Change the zoom level using the control keys.
• Press the "enter" key to save the changes.
The selection is saved and you return to the "ADVANCED USER SET-
TINGS" menu.
If you do not want to save your changes, return to the "ADVANCED USER SET-
TINGS" menu via "CANCEL". The zoom level is set as it was when the sub-
menu was accessed.
PEAK / AVERAGE submenu In this submenu, you can set the exposure control characteristics of the
camera in a range from 0 to +8. To do this, proceed as follows:
• Select menu item "PEAK / AVERAGE" from the "ADVANCED USER SET-
TINGS" menu using the control keys and press the "enter" key.
• Use the control keys to select "SETTINGS".
The selection is displayed with an arrow.
• Change the value using the control keys.
Version 7.0
Page 150 G-30-1781-en
S100 / OPMI pico Operation
Version 7.0
G-30-1781-en Page 151
Operation S100 / OPMI pico
SERVICE submenu
You can set the settings that are seldom used in the "SERVICE MENU" sub-
menu. Proceed as follows to make changes:
• Press the <config> key to access the configuration menu.
• In the Configuration menu, use the control keys to select the menu
item "SERVICE MENU" and press <enter>.
The following menu appears:
SERVICE MENU
BACK TO CONFIG
REMOTE SETUP You can configure the device in this sub menu so that a selected function
(please refer to REMOTE FUNCTION) is controllable via a foot switch or a
fundus imaging system. The sub menu contains the following areas:
– REMOTE FUNCTION
– SWITCHING MODE
– AUDIO CONFIRMATION
You can assign one of the functions described below to the integrated HD
video camera under "REMOTE FUNCTION". This function is then triggered
when the foot switch is activated or by swiveling in and out of a fundus im-
aging system.
The following options are available:
Version 7.0
Page 152 G-30-1781-en
S100 / OPMI pico Operation
– FREEZE / LIVE
If the current live video image is displayed and the foot switch is pressed,
the live video image is frozen and displayed on the monitor as a freeze
frame. With renewed pressing of the foot switch the live video image is
displayed again.
– USER1 / USER2
(or Endodonty/Treatment or Anterior/Posterior)
You can switch between user 1 and user 2 by pressing the foot switch. If
the current user is neither user 1 nor user 2, user 1 is switched on with the
first activation of the foot switch.
– USER3 / USER4
You can switch between user 3 and user 4 by pressing the foot switch. If
the current user is neither user 3 nor user 4, user 1 is switched on with the
first activation of the foot switch.
– INVERTED / STANDARD (is only required with a special tube)
If the current live video image is displayed and the foot switch is pressed,
the live video image is frozen and displayed on the monitor as a freeze
frame. With renewed pressing of the foot switch the live video image is
displayed again.
You can select under "SWITCHING MODE" if a remote function shall be con-
trolled via a foot switch or a fundus imaging system. The following control
types are available:
– FOOTSWITCH / AUX
Function control is performed via a foot switch.
– FUNDUS SYSTEM
The function is controlled via a fundus imaging system. This setting is only
available under the "OPHTHALMOLOGY" application.
Version 7.0
G-30-1781-en Page 153
Operation S100 / OPMI pico
FACTORY SETTINGS This submenu allows resetting the user and camera settings to the factory set-
tings. To do this, proceed as follows:
• In the "SERVICE ", use the control keys to select menu item "FACTORY
SETTINGS" and press the "enter" key.
• Use the control keys to select one of the following items:
– RESET USER 1
– RESET ALL USERS
– RESET CAMERA
After selection of one of the menu items, a message informs you that
the stored data will be overwritten.
• Press the "enter" key to overwrite the settings.
The system returns to the SERVICE MENU after approx. 2 seconds.
• If you do not want to save any changes, select menu item "CANCEL" to
return to the SERVICE MENU.
VERSION INFORMATION This submenu shows the current versions of the hardware and software com-
ponents as well as the serial numbers of the camera and CCU. Proceed as fol-
lows:
• Select menu item "VERSION INFORMATION" and press the "enter" key.
The following information is displayed:
– SOFTWARE (software version of the CCU)
– S/N CCU (serial number of the CCU)
• Press the "enter" key to return to the SERVICE MENU.
Version 7.0
Page 154 G-30-1781-en
S100 / OPMI pico Operation
You can use the installation menu to make device settings that are seldom
changed and that require a restart of the camera.
• Press the <config> key for approx. 5 seconds to access the installation
menu.
The following menu appears on the monitor:
INSTALLATION MENU
Version 7.0
G-30-1781-en Page 155
Operation S100 / OPMI pico
720P Select this setting format if your display device is 720p compatible. This mode
is similar to the 1080p progressive standard, but its maximum resolution is
only 1280 x 720 pixels.
1080P* Select this setting if your display device is 1080p compatible. The image is set
up using real full frames. This prevents any interline flicker, and the image is
sharply defined and stable. The maximum resolution is 1920 x 1080 pixels.
1080I* Use the 1080i (interlace) mode if your display device supports the 1080i
standard. The image is set up by two consecutive half images. The delayed
image setup can cause slight interline flicker. The maximum resolution is 1920
x 1080 pixels.
As a requirement, you should have connected the HD compatible OPMI pico
with a 1080p version and a 1080p or 1080i compatible display device. We
recommend the use of a 1080p compatible display device.
Version 7.0
Page 156 G-30-1781-en
S100 / OPMI pico Operation
HDTV MODE
720P
1080P*
1080I*
CANCEL
Version 7.0
G-30-1781-en Page 157
Operation S100 / OPMI pico
When you chose "NO", the camera resolution is not changed and you
go back to the installation menu.
• If you do not want to save your changes, return to the installation menu
via "CANCEL". The video signal is set as it was when the sub menu was
accessed.
Version 7.0
Page 158 G-30-1781-en
S100 / OPMI pico Operation
CANCEL
Version 7.0
G-30-1781-en Page 159
Operation S100 / OPMI pico
LETTERBOX Use this setting for display devices that are not compatible with the 16:9 wide
screen format. A black bar will appear at the top and bottom of the screen
and the image will be displayed in the 16:9 format.
SIDE-CUT In this mode, the sides of the image will be "cut", leaving the center of the
image to be displayed with a 4:3 aspect ratio.
Proceed as follows to select a format and to load the respective configuration:
• Press the "config" key for approx. 5 seconds to access the installation
menu.
• Select menu item "SDTV FORMAT MODE" in the installation menu using
the control keys and press the "enter" key.
The following menu appears:
SDTV FORMAT
LETTERBOX
SIDE-CUT
CANCEL
Version 7.0
Page 160 G-30-1781-en
S100 / OPMI pico Operation
PAL
NTSC
CANCEL
Version 7.0
G-30-1781-en Page 161
Operation S100 / OPMI pico
LIGHT SOURCE
In the "LIGHT SOURCE" submenu, you can select between XENON / LED and
HALOGEN.
• Press the "config" key for approx. 5 seconds to access the installation
menu.
• Select menu item "LIGHT SOURCE" from the installation menu using the
control keys and press the "enter" key.
The following menu appears:
LIGHT SOURCE
XENON / LED
HALOGEN
CANCEL
Version 7.0
Page 162 G-30-1781-en
S100 / OPMI pico Operation
Application submenu
In the "APPLICATION" submenu, you can select between five different appli-
cations. Each application affects the user profile and the associated camera
settings.
The application selection is displayed when the device is switched on for the
first time and can be changed at any time. The selected application is dis-
played briefly during start-up. The following applications can be selected:
DENTISTRY This application is designed for displaying the best image in the lower or upper
tooth segment. The "ENDODONTY" and "TREATMENT" users have optimized
and defined settings for the brightness control. This is why the "BRIGHTNESS"
and "LIGHTMETER" menu items are not available to these users.
NEURO/SPINE This application is designed for displaying the image in the neurological area.
The difference refers to the camera settings as well as to the not selectable
image orientation.
ENT This application id for the imaging of the neck, nose, ear area. The differences
can be found in regard of the camera settings and the adjusted image section.
OPHTHALMOLOGY This application is designed for displaying the image in the ophthalmological
area. The "ANTERIOR" and "POSTERIOR" users are available for displaying the
best image in the anterior or posterior eye segment. Both users have fixed set-
tings for the brightness control. This is why the "BRIGHTNESS" and "LIGHT-
METER" menu items are not available to these users.
Version 7.0
G-30-1781-en Page 163
Operation S100 / OPMI pico
APPLICATION VARIANT
OPHTHALMOLOGY
DENTISTRY
NEURO/SPINE
ENT
CANCEL
Version 7.0
Page 164 G-30-1781-en
S100 / OPMI pico Operation
In the white balancing procedure, the system adjusts the video signal in such
a way that white areas in the surgical field are also white on the video image.
• Press the <start> key to activate the white balance procedure.
The following message is displayed:
"PLEASE PUT A PIECE OF WHITE PAPER UNDER THE MICROSCOPE"
• Place a suitable white object (e. g., a piece of white paper with a matte
surface) under the surgical microscope.
• Carry out the white balance procedure using a medium brightness level.
• Press the <start> key to start the white balance procedure.
During the white balance procedure (display: "WHITE BALANCE IN
PROGRESS") the key functions are locked.
If the white balance procedure is successfully completed, the following will be
displayed: "WHITE BALANCE SUCCESSFUL". Otherwise, the following mes-
sage is displayed on the monitor: "WHITE BALANCE FAILED". Check the set-
tings described above and repeat the white balance procedure.
The white balance is saved for the user currently selected. When the user is
changed, the white balance values pertaining to that user are loaded.
After the white balance procedure is completed, all keys are enabled again.
The determined values are saved and are available to all users each time the
device is restarted.
White balance is adjusted corresponding to the set light type. The adjustment
depends on the "LIGHT SOURCE" setting in the "INSTALLATION MENU" and
the BLUE and RED settings in the "USER MENU ADVANCED".
Version 7.0
G-30-1781-en Page 165
Operation S100 / OPMI pico
The RED and BLUE values have an influence on the camera colors even after
the white balance procedure.
Recommended settings for a halogen light source
"LIGHT SOURCE" - "HALOGEN"
RED BLUE
Adjustment for a weak to medium illumi- +4 -8
nation (default)
Adjustment for a bright halogen illumina- +2 -4
tion
Without halogen light adjustment 0 0
RED BLUE
No adjustment needed (every illumina- 0 0
tion level)
Version 7.0
Page 166 G-30-1781-en
S100 / OPMI pico Operation
Under "brightness" you can adjust the brightness of the video image in 5%
increments between 0% and 100% using the and keys. The value set
appears in the upper left corner of the monitor.
In the configuration menu, you can select between three different modes in
the brightness menu: AUTO - LOW AUTO - MANUAL.
AUTO
In AUTO mode the camera adjusts the brightness of the video image automat-
ically to the value specified by the and control keys. Exposure control
affects the exposure time of the camera.
The nominal brightness is set in the AUTO mode. The camera controls the ex-
posure time in the range between 1/50(60)s and 1/10000s accordingly. The
nominal brightness can be adjusted between 0 and 100%. Changes are made
in 5% increments.
A required value in the range between 50% and 70% usually provides op-
timum exposure results.
LOW AUTO
This has the same function as AUTO Mode, however, exposure times longer
than 1/50(60)s are allowed, which can result in blurred movements.
MANUAL
In MANUAL Mode the selected exposure time is displayed when the con-
trol keys are pressed. The setting range goes from 1/6 to
1/10000.
– Increments for PAL: 1/10000, 1/4000, 1/2000, 1/1000, 1/500, 1/250, 1/
100, 1/50, 1/25, 1/12 and 1/6
– Increments for NTSC: 1/10000, 1/4000, 1/2000, 1/1000, 1/500, 1/250, 1/
120, 1/60, 1/30, 1/15 and 1/8
Greater exposure times can result in blurred movements.
The set time disappears from the screen again if the or control key was
not pressed longer than 2 seconds.
Version 7.0
G-30-1781-en Page 167
Operation S100 / OPMI pico
Use the <freeze> key to freeze the current live video image and to display it
on the video monitor as a full-screen still image. In the meantime the text
"freeze" appears in the upper left corner.
When you press the <live> key, the current live video image is displayed again
on the monitor. The text "freeze" disappears.
If the current live video image is being displayed and the switch connected to
the remote socket (foot switch) is pressed, and if FREEZE/LIVE is set under
SERVICE MENU - REMOTE SETUP - REMOTE FUNCTION in the CONFIG MENU,
the live video image is frozen and displayed as a still image on the monitor.
This corresponds to the "freeze" function. If you press the connected switch
(foot control panel) while the freeze frame is being displayed, the live video
image appears on the monitor again. This corresponds to the "live"function.
Version 7.0
Page 168 G-30-1781-en
S100 / OPMI pico Operation
Version 7.0
G-30-1781-en Page 169
Operation S100 / OPMI pico
Test pattern
A test pattern (color bars) is displayed if the camera head is not connected
when you switch on the system.
Switch off the system again, connect the camera head and switch the system
back on.
Menu structure
Menu display
The selected menu, submenu or the function is displayed in light blue.
Selected tabs appear in white letters.
Variants
The operation and the menu differ depending on whether you have a SWQ
recoding license or not. Functions that are available only with a SW re-
cording license are identified with a "*".
Version 7.0
Page 170 G-30-1781-en
S100 / OPMI pico Operation
Main menu*
Menu
Patient*
Menu
Settings
Submenu
Camera
- Brightness
- Image
Version 7.0
G-30-1781-en Page 171
Operation S100 / OPMI pico
- Color
- Profiles
Submenu
Recording*/Streaming
Submenu
System
- General
- Software
Version 7.0
Page 172 G-30-1781-en
S100 / OPMI pico Operation
- Microscope
- Network
- Video
Version 7.0
G-30-1781-en Page 173
Operation S100 / OPMI pico
Version 7.0
Page 174 G-30-1781-en
S100 / OPMI pico Operation
Version 7.0
G-30-1781-en Page 175
Operation S100 / OPMI pico
Main menu*
*5
1 Header bar
Mode: No video, Live, Image, Capture, Freeze, Pause, Play,
Forward, Rewind
Patient*: Name of the active patient, short term memory mode or
media name of the image display or video playback
Time: Time setting
2 Menu bar
Menu name is displayed
3 Patient* (only with SW recording license)
See Page 178
4 Settings (see Page 181)
Camera (see Page 182)
Recording* (with SW recording license) /
Streaming (without SW recording license) (see Page 188)
System (see Page 190)
Version 7.0
Page 176 G-30-1781-en
S100 / OPMI pico Operation
5 Footer bar
Recording medium: USB*, LAN*, Streaming, (not available).
During video recording, the information "Rec" or "Smart Rec"*
(when smart recording is active) appears next to the recording me-
dium and after it the duration of the current recording.*
Remaining recording time of the recording medium (e.g. 346:53:06)*
Version 7.0
G-30-1781-en Page 177
Operation S100 / OPMI pico
Patient*
Patients can be managed in the menu "Patient". The "Patient" menu is avail-
Patient able only with a SW recording license.
• Select the "Patient" menu in the Main menu.
The list of patients appears
2 3
1 4
Version 7.0
Page 178 G-30-1781-en
S100 / OPMI pico Operation
Variant 1: Select "Last name + first name" in "Patient data mode" (see
Page 189).
A virtual keyboard appears prompting you to enter the last name.
• Please enter the patient's last name.
The maximum text length is 15 characters.
• To save the entry, select the "OK" key.
A virtual keyboard appears prompting you to enter a first name (for
the requirement, see “Patient data mode*” on Page 189).
• Please enter the patient's first name.
The maximum text length is 15 characters.
• To save the entry, select the "OK" key.
Variant 1: Select "Last name + first name" in "Patient data mode" (see
Page 189).
A virtual keypad showing the previous user name is displayed.
• Please change the patient's last name.
The maximum text length is 15 characters.
• To save the entry, select the "OK" button.
The virtual keyboard appears with the patient's first name .
• Change the patient's first name.
The maximum text length is 15 characters.
Version 7.0
G-30-1781-en Page 179
Operation S100 / OPMI pico
Version 7.0
Page 180 G-30-1781-en
S100 / OPMI pico Operation
Settings
1 Camera
See Page 182
2 Recording (with SW recording license) /
Streaming (without SW recording license)
See Page 188
3 System
See Page 190
Version 7.0
G-30-1781-en Page 181
Operation S100 / OPMI pico
Camera
In the "Camera" submenu you can adjust the brightness, image, color and
profiles.
The settings in the "Brightness", "Image" and "Color" tabs directly affect the
profile set in "Profiles". Different values can be saved in the respective tabs for
each profile. The settings that are configured are saved automatically. When
the profile is changed (see "Active Profile" Page 187) the settings saved for
this profile are loaded.
Brightness
In the "Brightness" tab you can configure different exposure and brightness
settings.
Use the <Brightness> key to go directly to the "Brightness" tab.
2
3
1
4
5
6
1 Mode
In "Mode" you can choose between three different exposure controls: Au-
tomatic - Low light - Manual
Automatic In this mode the camera adjusts the brightness of the video image auto-
matically to the value specified in the "Auto level". Exposure control af-
fects the exposure time of the camera.
In "Automatic" mode the target brightness is specified and the camera ad-
justs the shutter speed in a range between 1/50 (60) s and 1/10000 s.
There is no screen output regarding the shutter speed actually set. The
target value can be set between 0 and 100%. The change is made in 5%
increments.
Version 7.0
Page 182 G-30-1781-en
S100 / OPMI pico Operation
A target value in a range from 50% to 70% usually provides the best ex-
posure results.
Low light This mode has the same function as the "Automatic" mode, but exposure
times greater than 1/50 (60) s are allowed, which can result in blurred
movements.
Manual In this mode you can set the exposure time under "Exposure time". The
setting range extends from 1/6 to 1/10000.
– Increments for PAL: 1/10000, 1/4000, 1/2000, 1/1000, 1/500,
1/250, 1/100, 1/50, 1/25, 1/12 and 1/6
– Increments for NTSC: 1/10000, 1/4000, 1/2000, 1/1000, 1/500,
1/250, 1/120, 1/60, 1/30, 1/15 and 1/8
2 Lightmeter
In "Lightmeter" you can choose between three types of exposure meas-
urement: Small spot, large spot or integral.
The lightmeter function is active only when the brightness is to set "Auto-
matic" or "Little light".
Small spot The exposure is measured in a very small area in the image center. This
measurement is suitable for work in narrow channels when an extremely
small illuminated-field diameter is being used.
Large spot The exposure is measured in an area in the image center. This measure-
ment is ideal for working with an eclipsed surgical field edge (resulting
from a reduced illuminated-field diameter).
Integral The exposure is measured and averaged across the full video image. This
measurement is recommended for surgical fields that are fully and evenly
lit.
If the object of interest to you is not in the image center, in most cases the
"Small spot" setting will not deliver the desired exposure result. Select ei-
ther a larger area ("Integral" or "Large spot") or manual exposure.
Version 7.0
G-30-1781-en Page 183
Operation S100 / OPMI pico
3 Black level
In "Black level" you can set the black level of the camera in a range from
-10 to +10.
4 Peak / Average
In "Peak / Average", you can set the exposure control characteristics of the
camera between 0 and +8.
5 Exposure time
In "Exposure time" you can set the exposure time of the camera. See also
"Mode" (1).
6 Auto level
In Auto level you can set the target brightness of the camera. See also
"Mode" (1).
Version 7.0
Page 184 G-30-1781-en
S100 / OPMI pico Operation
Image
In the "Image" tab you can set the display configurations.
1 2
1 Zoom
In "Zoom" you can set the zoom value of the camera in a range from 1.0 x
to 2.0 x in 0.1 increments.
2 Orientation
In "Orientation" you can set the image orientation.
Standard Standard
Mirror Mirror the image vertically
Flip Flip the image horizontally
Rotate Rotate the image 180°
3 White balance
With "Start white balance" and the <white bal> key you can start white
balancing.
In the white balancing procedure, the system adjusts the video signal in
such a way that white areas in the surgical field also appear in white on
the video image.
• Switch off the color filter and point the surgical microscope to a suit-
able white object (e.g. a sheet of white paper with a matte surface).
• Carry out the white balance procedure using a medium brightness lev-
el.
• Press the "Start white balance" key or the <white bal> button.
• If the white balancing was concluded successfully, a message appears.
• Confirm the notification with "OK".
Version 7.0
G-30-1781-en Page 185
Operation S100 / OPMI pico
Color
You can make the color settings in the "Color" tab.
Use the <colors> key to go to the "Color" tab.
2
1
3
4
1 Red
In "Red" you can set the red value of the camera in a range from -20 to
+20.
The setting of the red value also affects the white balance of the camera
(to perform white balancing, see Page 185).
2 Blue
In "Blue" you can set the blue value of the camera in a range from
-20 to +20.
The setting of the blue value also affects the white balancing of the
camera (to perform white balancing, see Page 185).
3 Color saturation
In "Color saturation" you can set the color saturation of the camera in a
range from -10 to +10.
A color saturation setting in a range of ±0 usually provides the best color
impression. The effect of the color saturation setting is the strongest for
heavily saturated (vivid) colors. The effect is small with low-saturation
colors.
4 Hue
In "Hue" you can set the hue of the camera in a range from
-10 to +10.
Version 7.0
Page 186 G-30-1781-en
S100 / OPMI pico Operation
Profiles
You can select the current profile in the "Profile" tab.
1 Active profile
A profile can be selected in "Active profile". Every profile can be renamed
and reset.
Rename Profile (Rename):
• Select the <Rename> button.
A virtual keyboard is displayed. This virtual keyboard can be used to
rename the profile (max. 15 characters).
• You can select the characters using the control keys . Use the <En-
ter> or to adopt the selected character.
• You can delete a character by pressing "Delete".
• Press "OK" to accept a change. You can cancel the action by pressing
"Cancel".
Reset profile:
• Select the <Reset> button.
• Confirm the warning message with "OK" to reset the profile.
The profile is reset to factory settings.
Version 7.0
G-30-1781-en Page 187
Operation S100 / OPMI pico
Recording* / Streaming
Streaming
1 4*
*2 5*
*3 6*
1 Video quality
Here you can set the recording quality of the video to be saved. Saving a
higher video quality requires more memory space.
2 Image format*
The memory format of the still images is set here. Saving an image in
TIFF requires more memory than JPEG format. With the TIFF format, the
file is saved without data loss and is not compressed. With the JPEG
format, the file is compressed with a little data loss.
3 Recording mode*
If "Smart Recording" is active, the video image of the preceding 30 sec-
onds is recorded when the <record> button is pressed. This recording
from the past is recorded in parallel at the start of the video recording. The
recording time should be at least 30 seconds in order for the preceding 30
seconds also be to written completely in the video file.
If the recording is ended too early, not all the data can be recorded.
Version 7.0
Page 188 G-30-1781-en
S100 / OPMI pico Operation
4 Storage mode*
Here you set the output medium for the images or videos.
Without a SW recording license only "Streaming" mode is available.
In the "LAN" storage mode, a "Standard Windows Network Share" pro-
tocol is used.
In the "Streaming" storage mode, the "Real Time Protocol" (RTP) is used
and the connection is established using the "Real-Time Streaming Pro-
tocol" (RTSP).
5 Patient data mode*
Here you determine the mode of the file name.
You can enter the patient name in "Last name + first name". In ID you can
assign an ID yourself for the patient using numerical characters.
6 File mode*
When "Short Term Memory Mode" (default) is active, the data is not as-
signed to a patient, but is filed in the folder according to the date and
time.
Files that are older than 3 days are deleted automatically. In "Short Term
Memory Mode" you cannot switch the patient. Deactivate this mode in
order to be able to switch patients.
If "Short Term Memory Mode" is deactivated, patient mode is active. Your
files are then assigned to the active patient.
Version 7.0
G-30-1781-en Page 189
Operation S100 / OPMI pico
System
General
In the "general" tab you can set general configurations.
4
1
5
2
3
1 Language
In "Language" you can select a language.
2 Function key F1 and Function key F2
In "Function key F1" and "Function key F2" you can assign to the keys
<function F1> and <function F2> the following functions:
Key <function F1>:
– "Zoom ON/OFF"
– "Smart Recording ON/OFF"*
Key <function F2>:
– "Switch camera profile"
– "Switch storage mode"* /
"Streaming ON/OFF" without SW recording license
Version 7.0
Page 190 G-30-1781-en
S100 / OPMI pico Operation
The "Zoom ON/OFF" setting of the <function F1> key will show visible
changes only if the zoom value is set greater than 1.0 (see Page 185).
3 Remote function (foot switch)
In "Remote Function" you can assign the following functions to a foot
switch:
– Video recording, which corresponds to the triggering of the button
<Record>* / "Streaming ON/OFF" without SW recording license
– Image capture, which corresponds to the triggering of the button
<Capture>* / "Freeze"/"Live" without SW recording license
The assigned function is triggered when the foot switch is pressed.
4 Sounds
In "Sounds" you can activate or deactivate the notification or acknowledg-
ment sounds.
Message: Notification in case of information and error messages
Confirmation: Confirmation when the key function is performed
5 Date / Time
In "Date / Time" you can adjust the date and the time.
Setting the date:
• Select "Date".
A virtual keyboard appears.
• Delete the appropriate characters (backwards) by pressing the "De-
lete" button.
• Enter the date using the virtual keyboard.
Year: Month: Day
Setting the time:
• Select "Time".
A virtual keyboard appears.
• Delete the appropriate characters (backwards) by pressing the "De-
lete" button.
• Enter the time using the virtual keyboard.
Hours: Minutes
Version 7.0
G-30-1781-en Page 191
Operation S100 / OPMI pico
Software
In the "Software" tab you can obtain information on the software.
1
3
2
1 Version
"Version" shows you the installed software version.
2 Special
Software Update
The software can be updated via a USB medium. Old system configura-
tions are retained in this process. After this, a restart is required.
Do not switch the device off during a software update!
Factory settings
Resetting the factory settings
Service
Only for service staff.
3 Export
Log Files
Automatically generated protocols are saved as files on the USB medium.
They can be sent to ZEISS Service or an authorized partner to evaluate mal-
functions.
Open-Source Licenses
Creates a license file on the USB medium. A zip file containing notes and
the licenses is saved in the USB medium.
Version 7.0
Page 192 G-30-1781-en
S100 / OPMI pico Operation
1 2
1 Treatment
In "Treatment" you can choose between four different treatment settings.
Switching the type of treatment causes all the profiles to be reset to the
new default values.
Each type of treatment affects the profiles and thus the camera settings
associated with them The profiles associated with the type of treatment
selected have optimized and specified settings for the brightness control.
The following types of treatment can be selected:
– Dental: This type of treatment is designed for the optimum image set-
ting in the upper and lower tooth section area.
– ENT: This type of treatment is designed to display images in the ear,
nose and throat area.
– Gynecology: This type of treatment is designed to display images in
the gynecological area.
– Neuro: This type of treatment is designed to display images in the neu-
rological area.
2 Export / Import
In "Export / Import" you can export or import the camera settings or all
settings to or from a USB medium.
Version 7.0
G-30-1781-en Page 193
Operation S100 / OPMI pico
Version 7.0
Page 194 G-30-1781-en
S100 / OPMI pico Operation
Network
In the "Network" tab you can set your network configurations.
*2
1 3
1 IP Settings
MAC Address
The MAC Address (Media Access Control Address) is the hardware address
of each individual network adapter that serves as a unique identification
of the device in a computer network if required. The unique MAC address
is required to create a SW recording license.
DHCP activated:
Network link via a dynamic IP address "DHCP"
A dynamic IP address is an IP address which is automatically assigned to
the requesting device by an address service (DCHP server) available in the
network.
After applying the activated DHCP, you can read out the values trans-
mitted from the DHCP server using the buttons IP Address, Subnet Mask
and Standard Gateway.
DHCP deactivated:
Network connection via a static IP address.
Static IP addresses are fixed IP addresses, permanently assigned to a
system. You can obtain the data required for this purpose (IP address,
subnet mask and standard gateway) from your IT Administrator.
• Enter the data and use "Apply" to activate the changes.
For more detailed information on the network configuration, see "Config-
uring the Network Connection (default settings)" onPage 197 .
2 Network share*
The recorded video and image data can be saved in a "network folder"“
in a computer on the network.
Version 7.0
G-30-1781-en Page 195
Operation S100 / OPMI pico
Version 7.0
Page 196 G-30-1781-en
S100 / OPMI pico Operation
– You can use access rights (user and password) to the network server or
network computer to control access to the network share.
– Share only the minimum data fields required (e.g. only for the daily re-
quirement or not all patient data).
– Use a separate share with a separate patient area for each S100 / OPMI
pico with SW recording license to prevent problems among the different
devices and users.
– You can change and adjust the network share any time.
– After using the network share, you can disable it again in the network
server.
– Install the current security updates in the network server.
The integrated HD video camera live & streaming is not an archiving system.
Therefore to protect your videos and images, you should secure a proper ar-
chiving system.
*2
1 3
In the "Network" tab (Main menu > Menu settings > System submenu > Net-
work tab) you can configure the surgical microscope for a network.
The following settings are preinstalled settings. These settings can be
changed by the IT Administrator any time that the network environment re-
quires it.
1 IP settings (static)
IP address: 192.168.0.11
Subnet mask: 255.255.255.0
Standard gateway: 192.168.0.1
Version 7.0
G-30-1781-en Page 197
Operation S100 / OPMI pico
2 Network share
Host IP: 192.168.0.10 (external computer)
Share name: recording
Work group: (or domain name)
User name: zeiss
Password: recording
3 Streaming
Client IP: 192.168.0.255 (broadcast or single client IP address)
Port: 5004 (normally does not have to be changed)
You can stream to a single receiver (client IP) by indicating a single IP ad-
dress. But the stream can also be transmitted to several receivers at the
same time if you use the broadcast address of your network area as the
client IP. In addition, each receiver must set up the connection via RTSP.
Use broadcast only in a cable-connected network and not to WLAN re-
ceivers.
If streaming is active in the S100 / OPMI pico, the following URL can be
used in the video software (additional software may be required):
URL: rtsp://192.168.0.11:8554/video
You only have to adjust the IP address or replace the name configured in
the Domain Name System (DNS).
You can read out the IP address of the surgical microscope using the "IP
Address" button even if DHCP is active (see Page 195).
4 Web control / Web interface
URL: http://192.168.0.11
(Port 80 must not be entered)
You can read out the IP address of the surgical microscope using the "IP
Address" button even if DHCP is active (see Page 195).
If a name has been saved in your Domain Name System (DNS), you can
also use this name instead of the IP address.
If you have any questions, contact your IT Administrator.
Version 7.0
Page 198 G-30-1781-en
S100 / OPMI pico Operation
Video
In the "Video" tab you can determine the display and recording options.
1 4
2 5
1 HDTV format
You can select the video camera's output signal and adjust it to your dis-
play device here.
720p
Select this setting if your display device is 720p compatible. The image is
set up using real full frames. This prevents any interline flicker, and the
image is sharply defined and stable. The resolution is 1280 x 720 pixels.
1080p
Select this setting if your display device is 1080p compatible. The image is
set up using real full frames. This prevents any interline flicker, and the
image is sharply defined and stable. The resolution is 1920 x 1080 pixels.
Without a SW recording license, streaming is transmitted with 720p even
if your camera and monitor are set to 1080p.
2 SDTV format
Here you can choose between a "Letterbox" and a "Sidecut" display when
connected to a display unit with Y/C analog input (monitor, projector, etc.)
mostly in 4:3 format.
Letterbox
Black bars will appear at the top and bottom of the screen and the image
will be displayed in the 16:9 format.
Sidecut
The image is cut off on the sides and only the middle of the image is dis-
played in an aspect ratio of 4:3.
Version 7.0
G-30-1781-en Page 199
Operation S100 / OPMI pico
Version 7.0
Page 200 G-30-1781-en
S100 / OPMI pico Operation
Web interface
With the help of the web interface you can use a web browser to trigger the
control buttons <Record>, <Capture> and <Live> from any computer.
• Start the web interface by entering the URL in the address line of your
browser (see Page 198).
The start screen appears.
The link displayed on the start screen and the URL facilitate the configuration
and starting of the streaming application of your computer.
5
1 Header bar
See "Main menu" on Page 176
2 Upper selection bar
Patient or Index*
Record
Capture
Live
3 Menu bar
The menu name is displayed.
4 Lower selection bar
For patient: Select, New, Rename, Delete
For index: Show/Play, Delete
5 Footer bar
See "Main menu" on Page 176
Version 7.0
G-30-1781-en Page 201
Operation S100 / OPMI pico
Version 7.0
Page 202 G-30-1781-en
S100 / OPMI pico Operation
When web control is deactivated, the Patient/Index, Record, Capture and Live
buttons are not displayed.
Without a SW recording license, the Patient/Index button is not displayed
during streaming.
A change cannot be made to the Patient menu in Short Term Memory Mode.
Only the index is available.
Version 7.0
G-30-1781-en Page 203
Operation S100 / OPMI pico
1 2
You need the SW recording license to be able to use all the functions such as
recording and patient management.
The SW recording license is available from your dealer. Your MAC address
(Page 195) is required for this purpose. You will receive a USB medium with
the SW recording license.
• Insert the USB medium with the license file into the USB port.
• Open the Microscope tab (Main menu > Settings menu > System sub-
menu > Microscope tab).
• Press the "Import license" button (3).
• Enter "OK" to confirm the notification.
"Installed" appears after the button. The SW recording license is active im-
mediately. All functions can be used without restarting.
Version 7.0
Page 204 G-30-1781-en
S100 / OPMI pico Operation
• Always check the following points (without the patient) before using the
device:
– The monitor is switched on.
– The network cable connection is established or a USB medium is con-
nected, is detected and has sufficient memory capacity.
– The correct profile was selected (see Page 187).
– The correct patient has been selected (see Page 178) or "Short Term
Memory Mode" is activated (see Page 188).
The name of the patient or "Short Term Memory Mode" (1) appears in
the header bar.
Recording HD videos
• To start video recording, press the <Record> button on the remote con-
trol, the video unit, the foot switch (if configured) or the web interface.
During video recording, "Rec" (2), the duration of the recording (3)
e.g. 00:00:08 and the remaining recording time (4) e.g. (348:26:22)
appear in the footer bar. The "record" indicator light on the control
unit lights up with an amber color.
While recording a video the photo function (Capture) can be used to
create additional images.
Version 7.0
G-30-1781-en Page 205
Operation S100 / OPMI pico
3
2 4
Version 7.0
Page 206 G-30-1781-en
S100 / OPMI pico Operation
The patient whose data you wish to view is activated (see Page 178) or
"Short Term Memory Mode" is activated (see Page 188).
The name of the patient or "Short Term Memory Mode" (1) appears in the
header bar.
• Press the <Index> button of the remote control, the video unit or the web
interface.
The "Index" is opened and the existing data is displayed.
• Button (2) shows various texts, depending on the data type that was se-
lected:
- "Show" for images,
- "Play" for videos.
• Use the keys to select the file that you want to view.
• Use the keys to select the "Show "or "Play" button and press the
<Enter> or key.
The selected image or video is displayed or played.
• Press the <Index> or <Live> button of the remote control, video unit or
web interface to end the image display or video playback.
The image display or video playback is ended and the "Index" is displayed.
Version 7.0
G-30-1781-en Page 207
Operation S100 / OPMI pico
The patient whose data you wish to delete is activated (see Page 178) or
"Short Term Memory Mode" is activated (see Page 188).
The name of the patient or "Short Term Memory Mode" (1) appears in the
header bar.
• Press the <Index> button of the remote control, the video unit or the web
interface.
The "Index" is opened and the existing data is displayed.
• Use the keys to select the file that you want to delete.
• Use the keys to select the "Delete" button (2) and press the <Enter>
or key.
• Confirm the notification with "OK" to delete the selected filed.
The selected file is deleted and the "Index" is displayed.
Version 7.0
Page 208 G-30-1781-en
S100 / OPMI pico What to do in the event of malfunctions
Troubleshooting ............................................................................210
For your safety ........................................................................................210
Malfunctions of the system......................................................................210
Malfunctions on S100 stands with video (optional)..................................210
Malfunctions of S100 suspension systems with video and image storage op-
tion .........................................................................................................216
Version 7.0
G-30-1781-en Page 209
What to do in the event of malfunctions S100 / OPMI pico
Troubleshooting
• If a failure occurs which you cannot correct with the aid of the chapter
"What to do in the event of malfunctions“, attach a sign to the system
stating it is out of order and contact our service representative.
Version 7.0
Page 210 G-30-1781-en
S100 / OPMI pico What to do in the event of malfunctions
Version 7.0
G-30-1781-en Page 211
What to do in the event of malfunctions S100 / OPMI pico
Version 7.0
Page 212 G-30-1781-en
S100 / OPMI pico What to do in the event of malfunctions
Version 7.0
G-30-1781-en Page 213
What to do in the event of malfunctions S100 / OPMI pico
Version 7.0
Page 214 G-30-1781-en
S100 / OPMI pico What to do in the event of malfunctions
Version 7.0
G-30-1781-en Page 215
What to do in the event of malfunctions S100 / OPMI pico
Malfunctions of S100 suspension systems with video and image storage op-
tion
Version 7.0
Page 216 G-30-1781-en
S100 / OPMI pico What to do in the event of malfunctions
The lamp housing contains a backup lamp which can be manually swung into
the illumination beam path when the first lamp fails. Open flap (1) indicates
that the backup lamp is operative.
Version 7.0
G-30-1781-en Page 217
What to do in the event of malfunctions S100 / OPMI pico
The xenon light source has a lamp module containing two xenon lamps. The
second lamp is used as backup lamp. Manually swivel in the backup lamp after
a failure of the first lamp or if the maximum service life of the first lamp of
500h (see chapter "Maintenance intervals" on Page 222) is exceeded. The
backup lamp is in operation if the red segment (3) is lit in button (2).
Version 7.0
Page 218 G-30-1781-en
S100 / OPMI pico What to do in the event of malfunctions
1
2
Version 7.0
G-30-1781-en Page 219
What to do in the event of malfunctions S100 / OPMI pico
Version 7.0
Page 220 G-30-1781-en
S100 / OPMI pico Care and Maintenance
System maintenance......................................................................222
Maintenance intervals..............................................................................222
Safety check ............................................................................................223
Replacing the halogen lamp ....................................................................224
Exchanging the lamp module of the xenon illumination system ..............226
Changing the fuse - halogen illumination system .....................................228
Version 7.0
G-30-1781-en Page 221
Care and Maintenance S100 / OPMI pico
System maintenance
Maintenance intervals
Version 7.0
Page 222 G-30-1781-en
S100 / OPMI pico Care and Maintenance
Safety check
Version 7.0
G-30-1781-en Page 223
Care and Maintenance S100 / OPMI pico
Version 7.0
Page 224 G-30-1781-en
S100 / OPMI pico Care and Maintenance
1
2
3
7 8 9
Version 7.0
G-30-1781-en Page 225
Care and Maintenance S100 / OPMI pico
Version 7.0
Page 226 G-30-1781-en
S100 / OPMI pico Care and Maintenance
1
3
Version 7.0
G-30-1781-en Page 227
Care and Maintenance S100 / OPMI pico
Fuse rating
For 100 V...125 V~ T 6,3 A/H 250 V
For 220 V...240 V~ T 3,15 A/H 250 V
Version 7.0
Page 228 G-30-1781-en
S100 / OPMI pico Care and Maintenance
Cleaning
Prevention of fogging
To protect the eyepiece optics from fogging, we recommend using an anti-
fogging agent. Anti-fogging agents provided by eyecare professionals for use
with eyeglass lenses are also suitable for Zeiss eyepieces.
• Please observe the Instructions for Use supplied with each anti-fogging
agent.
Anti-fogging agents do not only ensure fog-free eyepiece optics. They also
clean the eyepiece optics and protect them from dirt, grease, dust, fluff and
fingerprints.
Version 7.0
G-30-1781-en Page 229
Care and Maintenance S100 / OPMI pico
Sterilization
The asepsis sets available from ZEISS contain rubber caps and hand grips that
can be sterilized in autoclaves. For detailed information on sterilization, please
see the enclosed instructions "Preparation of resterilizable products" for the
respective asepsis set.
You can also use disposable sterile drapes to cover the system.
• When fitting the sterile drape, ensure there is sufficient slack to allow for
movement of the microscope carrier and surgical microscope.
• Do not cover the ventilation opens - this ensures the lamps are cooled ad-
equately and no lamp fails.
Disinfection
Version 7.0
Page 230 G-30-1781-en
S100 / OPMI pico Care and Maintenance
Version 7.0
G-30-1781-en Page 231
Care and Maintenance S100 / OPMI pico
Note on disposal
User information
on the disposal of electrical and electronic devices
This symbol means that the product must not be disposed of as normal do-
mestic waste.
The correct disposal of electrical or electronic devices helps to protect the en-
vironment and to prevent potential hazards to the environment and/or human
health which may occur as a result of improper handling of the devices con-
cerned.
For detailed information on the disposal of the product, please contact your
local dealer or the device manufacturer or its legal successor. Please also note
the manufacturer's current information on the Internet. In the event of resale
of the product or its components, the seller is required to inform the buyer
that the product must be disposed of in accordance with the applicable na-
tional regulations currently in force.
Version 7.0
Page 232 G-30-1781-en
S100 / OPMI pico System Data
System Data
Version 7.0
G-30-1781-en Page 233
System Data S100 / OPMI pico
Technical data
The S100 / OPMI pico has no essential performance within the meaning of
IEC 60601-1: 2005 + A1:2012.
The fundamental functions within the meaning of the intended use were
tested by the manufacturer for their stability.
Version 7.0
Page 234 G-30-1781-en
S100 / OPMI pico System Data
Component Properties
Magnification Manual magnification changer,
Five settings where
γ = 0.4x, γ = 0.6x, γ = 1.0x, γ = 1.6x, γ = 2.5x
Focusing Objective lens with fixed focal distance: Focusing oc-
curs by employing the working distance to the ob-
ject
Focusing objective lens: Variably adjustable working
distance, focus range 13 mm.
Varioscope 100: Variably adjustable working dis-
tance, focus range 100 mm.
Objective lens Objective lens with fixed focal distance:
Optionally f = 200 mm, f = 250 mm, f = 300 mm,
f = 350 mm or f = 400 mm
Focusing objective lens:
Optionally f =200 mm, f = 250 mm or f = 300 mm
Varioscope 100:
AA = 200 - 300 mm
Tubes / Eyepieces Optional binocular sloped, straight or 180° swiveling
tube.
10x widefield eyepieces, alternatively: 12.5 x
Weight approx. 1.7 kg (without CCU, Endoport, tube, object
lens, eyepieces and MORA interface)
Version 7.0
G-30-1781-en Page 235
System Data S100 / OPMI pico
Component Features
Suspension arm Length: 600 mm
Swivel angle about carrier arm: ± 150°
Lifting range: ± 300 mm
Carrier arm Length: 500 mm
Stand height 1730 mm
Base 625 x 625 mm
(dimensions)
Weight approx. 90 kg
(including
microscope)
± 300 mm
1730 mm
360° ± 150°
1470 mm
Version 7.0
Page 236 G-30-1781-en
S100 / OPMI pico System Data
Component Properties
Suspension arm Length: 950 mm
Vertical lift: ± 400 mm
Turning angle around carrier arm: 150°
Carrier arm Length: 500 mm
Turning angle around stand column: ±105°
Weight Approx. 45 kg
(including
microscope)
520 mm
± 400 mm
±105° ±150°
Version 7.0
G-30-1781-en Page 237
System Data S100 / OPMI pico
Components Properties
Suspension arm Length: 950 mm
Turning angle around carrier arm: ±150°
Vertical lift: ± 400 mm
Carrier arm Length: 500 mm
Turning angle around stand column: ±163°
Weight Approx. 55 kg
(including
microscope)
500 mm 950 mm
425 mm
± 400 mm
± 163° ± 150°
Version 7.0
Page 238 G-30-1781-en
S100 / OPMI pico System Data
Component Properties
Suspension arm Length: 950 mm
Turning angle around carrier arm: 150°
Vertical lift: ± 400 mm
Carrier arm Length: 500 mm
Weight Approx. 30 kg
(including microscope
and halogen lighting)
500 mm 950 mm
425 mm
± 400 mm
±150°
Version 7.0
G-30-1781-en Page 239
System Data S100 / OPMI pico
Component Properties
Suspension arm Length: 950 mm
Turning angle around carrier arm: ±150°
Vertical lift: ± 400 mm
Carrier arm Length: 500 mm
Weight Approx. 35 kg
(including microscope and
lighting)
275 mm
± 400 mm
± 150°
Version 7.0
Page 240 G-30-1781-en
S100 / OPMI pico System Data
Component Properties
Suspension arm Length: 950 mm
Turning angle around carrier arm: 150°
Vertical lift: ± 400 mm
Carrier arm Length: 500 mm
Weight (including Approx. 25 kg
microscope)
275 mm
± 400 mm
± 150°
Version 7.0
G-30-1781-en Page 241
System Data S100 / OPMI pico
Component Properties
Suspension arm Length: 950 mm
Turning angle around carrier arm: 150°
Vertical lift: ± 400 mm
Carrier arm Length: 500 mm
Weight (including Approx. 25 kg
microscope)
275 mm
± 400 mm
± 150°
Version 7.0
Page 242 G-30-1781-en
S100 / OPMI pico System Data
Component Properties
Power connection Only connect the system to wall outlets which
are equipped with a properly connected pro-
tective ground conductor.
Electrical outputs HD Video output HDMI
video control units
Video output YPbPr
Video output BNC
Remote Socket
External switching contact input
Electrical outputs video Video output YPbPr
camera live &
LAN input/output
streaming control unit
Video output port Y/C
Video output port DVI
Remote Socket
External switching contact input
Halogen lighting system
Rated voltage (115 V): 100 V - 125 V
(230 V): 220 V - 240 V
Rated frequency 50 - 60 Hz
Current consumption (115 V): max. 250 VA
(230 V): max. 250 VA
Fiber optic lighting Halogen lamp, 12 V, 100 W,
backup lamp can be manually swiveled into beam
path
Fuses Fuse link
115 V ~ T 6.3 A/H 250 V
230 V ~ T 3.15 A/H 250 V
Xenon lighting system
Rated voltage (115 V): 100 V - 125 V
(230 V): 220 V - 240 V
Rated frequency 50 - 60 Hz
Version 7.0
G-30-1781-en Page 243
System Data S100 / OPMI pico
Component Properties
Current consumption (115 V): max. 350 VA
(230 V): max. 350 VA
Fiber optic illumination Xenon short-arc reflector lamp,
Color temperature: approx. 5000 K
Rated power: approx. 180 W
Backup lamp can be manually swung into beam
path
Fuses Automatic circuit breaker
LED lighting system
Rated voltage 100 V - 240 V
Maximum input 250 VA
Power frequency range 50 Hz - 60 Hz
Fuses Automatic circuit breaker
Version 7.0
Page 244 G-30-1781-en
S100 / OPMI pico System Data
Version 7.0
G-30-1781-en Page 245
System Data S100 / OPMI pico
Electromagnetic interference
Version 7.0
Page 246 G-30-1781-en
S100 / OPMI pico System Data
Version 7.0
G-30-1781-en Page 247
System Data S100 / OPMI pico
The device S100 / OPMI picois intended to be operated in one of the electro-
magnetic environments specified below. The customer or the user of the de-
vice S100 / OPMI pico is responsible for ensuring that it is operated in such an
environment.
Note 1 At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2 These guidelines may not apply in all situations. Electromagnetic propagation is influ-
enced by absorption and reflection by structures, objects and persons.
Version 7.0
Page 248 G-30-1781-en
S100 / OPMI pico System Data
Version 7.0
G-30-1781-en Page 249
System Data S100 / OPMI pico
The device S100 / OPMI pico is intended for use in an electromagnetic envi-
ronment in which RF disturbances are controlled. The customer or the user of
the device S100 / OPMI pico can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communi-
cation equipment (transmitters) and the S100 / OPMI pico - depending on the
output power of the communication equipment as specified below.
d = 1, 2 P d = 1, 2 P d = 2, 3 P
Version 7.0
Page 250 G-30-1781-en
S100 / OPMI pico System Data
System upgrades
The device S100 / OPMI pico can be extended with the accessories offered by
ZEISS (see ordering data/accessories) to form an already tested system. This
system combination has been approved by ZEISS.
Additional devices that are connected to electronic medical devices must
comply with the relevant standards and guidelines
(for example IEC 60950 for data processing devices).
Furthermore, any configurations must comply with normative requirements
for electronic medical devices (see IEC 60601-1:2005, chapter 16; in some
countries IEC 60601-1-1:2000 may still be valid).
The one who is connecting additional devices to electronic medical devices
acts as system configurer and is responsible for the system complying with the
normative requirements for electronic medical devices.
Please note that local legislation takes priority over the above-mentioned nor-
mative requirements. If you have any questions, please contact your local
dealer or ZEISS Service.
Ordering data
Operate the device only with accessories included in the scope of delivery and
recommended by ZEISS. The contact partners in charge of your country are
listed on the Internet on the following website:
www.zeiss.com/med
Illumination
Description Order number
Halogen lamp, 12 V, 100 W, 380079-9040-000
Version 7.0
G-30-1781-en Page 251
System Data S100 / OPMI pico
Illumination
Description Order number
Tinted UV protective eyewear 000000-0480-028
N-xenon lamp module incl. 2 xenon lamps in trans- 304977-9036-700
port case for Superlux 180
Couplings
Description Order number
120° coupling for S100 suspension system 000000-1080-351
K 0/120 coupling for use during gynecological ex- 305256-9903-000
aminations
Tubes
Description Order number
Sleeve for inclined and straight tubes 305542-0107-000
Straight binocular tube, f = 170 mm 303765-0000-000
45° inclined binocular tube 303784-0000-000
180° swiveling binocular tube 303791-0000-000
180° swiveling binocular tube, f = 200 mm 303792-0000-000
Foldable tube f170/f260 incl. rotators and 303771-9020-000
PROMAG
(not for use with a mouth switch)
Eyepieces
Description Order number
10x push-in widefield eyepiece 305542-0000-000
12.5x push-in widefield eyepiece, asph. 305543-9901-000
Objective lens
Description Order number
Focusing objective lens, f = 200 mm 000000-1159-867
Focusing objective lens, f = 250 mm 000000-1160-302
Focusing objective lens, f = 300 mm 000000-1176-319
Objective lens with fixed focal distance, 305132-0000-000
f = 200 mm
Version 7.0
Page 252 G-30-1781-en
S100 / OPMI pico System Data
Objective lens
Description Order number
Objective lens with fixed focal distance, 305705-0000-000
f = 250 mm
Objective lens with fixed focal distance, 305137-0000-000
f = 300 mm
Objective lens with fixed focal distance, 305707-0000-000
f = 350 mm
Objective lens with fixed focal distance, 305146-0000-000
f = 400 mm
Hand grips
Description Order number
Handgrip, complete, rear 000000-1058-010
Handgrip, complete 000000-1057-543
Comfort handgrip for OPMI pico 303335-9006-000
Comfort T handgrip 305929-8300-000
Splash protection
Description Order number
Drape adapter ring for VisionGuard 000000-1366-769
VisionGuard replacement lenses (sterile, 20/Box) 306001-0000-000
Vision Guard lens protectors (non-sterile, 10/box) 306002-0000-000
Asepsis
Description Quanti- Order number
ty
Sterile handgrip sleeves for OPMI pico 2 pairs 305810-9004-000
Sterile handgrips (rotatable through 360°) pack of 305810-9016-000
2
Asepsis caps 12 mm pack of 305810-9002-000
(resterilizable caps) 6
Asepsis caps 22 mm pack of 305810-9001-000
(resterilizable caps) 6
Asepsis caps for PD 180° tiltable tube pack of 305810-9003-000
(resterilizable caps) 6
Version 7.0
G-30-1781-en Page 253
System Data S100 / OPMI pico
Asepsis
Description Quanti- Order number
ty
Asepsis caps 27 mm pack of 305810-9008-000
(resterilizable caps) 6
Asepsis caps 49 mm pack of 305810-9007-000
(resterilizable caps) 2
Video accessories
Video accessories can be found in the separate product overview G-30-1888.
Foot switch
Remote control
Description Order number
Remote control 308203/-2210-000
2 x LR03 alkaline batteries AAA size ---
Version 7.0
Page 254 G-30-1781-en
S100 / OPMI pico System Data
Accessories
Description Order number
Beam splitter 50 301513-9902-000
Angled lens without lateral output for documenta- 000000-1099-230
tion accessories
Angled lens with lateral output for 000000-1079-239
documentation accessories
Rotating adapter 301007-0000-000
(Rotating adapter for binocular tubes)
Adapter with double iris diaphragm 303354-0000-000
Stereo coobserver tube with 2 pivot joints 000000-1063-869
Long coobservation tube with two 302706-0000-000
pivot joints
Version 7.0
G-30-1781-en Page 255
System Data S100 / OPMI pico
These upgrade kits require a service call which must be coordinated with your
local ZEISS service partner.
Version 7.0
Page 256 G-30-1781-en
S100 / OPMI pico System Data
Environmental conditions
During operation
Version 7.0
G-30-1781-en Page 257
System Data S100 / OPMI pico
Symbol Meaning
This way up
Indicates the correct upright position of the
package.
Fragile, handle with care
Keep dry
Do not stack
1
Permissible range of relative humidity
Humidity limits for storage and operation
Permissible range of air pressure
Air pressure limits for storage and operation
Version 7.0
Page 258 G-30-1781-en
S100 / OPMI pico System Data
Regulatory information
.
RoHS compliance The product is RoHS compliant in accordance with
Directive 2011/65/EU.
Version 7.0
G-30-1781-en Page 259
System Data S100 / OPMI pico
Devices S100 / OPMI® pico (Carrier unit for external lighting) (000000-1095-094)
S100 / OPMI® pico (Atmos support unit) (000000-1080-349)
Version 7.0
Page 260 G-30-1781-en
S100 / OPMI pico Indexes
Indexes
Term Explanation
Apochromatic optics Optical system with maximum correction of chro-
matic aberration
Drapes Sterile covers for surgical microscopes
FlexioStill Adapter for the attachment of digital compact
cameras to ZEISS surgical microscopes
FlexioMotion Adapter for the attachment of camcorders to ZEISS
surgical microscopes
Interlace (i) In interlace mode, the even and uneven-numbered
lines are alternatively set up. Todays video stand-
ards are based on this procedure.
Progressive (p) In Progressive mode, images are displayed in full
frames. This way, the line structure cannot be rec-
ognized and interline flicker is prevented.
Radiation intensity Radiation incident or emitted at right angles per
unit time and unit area.
VisionGuard Lens guard for ZEISS surgical microscopes
Version 7.0
G-30-1781-en Page 261
Indexes S100 / OPMI pico
List of abbreviations
Term Explanation
AG Aktiengesellschaft (German acronym for "stock
company")
BEV German acronym for "special emergency backup
power supply"
BNC Bayonet Neill Concelmann
(coaxial bayonet-locking connector for high fre-
quencies, named after its inventor)
CCU Camera Control Unit
(Video camera control unit)
CE Communauté Européenne (European Community)
- The manufacturer declares that the device com-
plies with the directives of the European Union.
DIN Deutsches Institut für Normung (German stand-
ards association)
D Diopter (dioptric power of optical systems)
DVI Digital Video Interface
(digital interface for video data)
EMC Electromagnetic Compatibility - defines non-inter-
ference of electrical and electronic devices with
their environment
EN European Norm (European Standard)
H Breaking capacity of the instrument fuse
HD High Definition
(providing high resolution)
HDMI High Definition Multimedia Interface
(digital interface for video and audio data)
RF Radio frequency
i interlace (see technical term index)
IEC International Electrotechnical Commission
IOL Intraocularlens - artificial lens for the eye
Version 7.0
Page 262 G-30-1781-en
S100 / OPMI pico Indexes
Term Explanation
JPEG JPEG File Interchange Format
(file format for the storage of images)
LED Light Emitting Diode
(LED)
NTSC National Television Systems Committee
(US institution defining the first color television
standard)
OR Surgical operation
OPMI Surgical microscope
PAL Phase Alternating Line
(color encoding system in analog TV systems)
p progressive (see technical term index)
PD Pupillary distance
SD Standard Definition
SDTV Standard Definition TeleVision
(TV standard providing a lower resolution than
HDTV, e.g. PAL or NTSC)
T* T* multi-coating - minimizes reflection
TIFF Tagged Image File Format
(file format for the storage of images)
UV UltraViolet
VESA Video Electronics Standards Association
(Device to mount flat screen monitors on the wall
or ceiling in accordance with the VESA standard.
Y/C Analog interface that transmits separate brightness
signals (luminance Y) and color signals (chromi-
nance C).
YPbPr Color model that transmits brightness information
Y and color difference information Pb and Pr sep-
arately.
Version 7.0
G-30-1781-en Page 263
Indexes S100 / OPMI pico
Version 7.0
Page 264 G-30-1781-en
S100 / OPMI pico
Keyword index
Symbols
..............................................................................................................188
"Enter" key .........................................................................................78, 82
Numerics
1080 Mode
Interlace ......................................................................................156
Progressive ..................................................................................156
1080p ....................................................................................................199
720 Mode ..............................................................................................156
720p ......................................................................................................199
Version 7.0
G-30-1781-en Page 265
S100 / OPMI pico
Disposal ......................................................................................231
insertion ......................................................................................231
Black level submenu HD Video live & streaming ......................................184
Black Level submenu, HD video ...............................................................149
Blue submenu, HD video .........................................................................148
Brightness .................................................................................................76
brightness
HD video .....................................................................................167
HD Video submenu .....................................................................141
submenu HD Video live & streaming ............................................182
Brightness control .....................................................................................72
Brightness setting .....................................................................................70
Burn injuries caused by illumination intensity ............................................17
Version 7.0
Page 266 G-30-1781-en
S100 / OPMI pico
Version 7.0
G-30-1781-en Page 267
S100 / OPMI pico
Version 7.0
Page 268 G-30-1781-en
S100 / OPMI pico
Export .........................................................................................192
Export / Import ............................................................................193
exposure time .............................................................................184
Function key F1 ...........................................................................190
General .......................................................................................190
HDTV format ...............................................................................199
hue .............................................................................................186
Image .........................................................................................185
Image format ..............................................................................188
Index ...................................................................................207, 208
IP settings ...................................................................................195
Language ....................................................................................190
Lightmeter ..................................................................................183
Main menu .................................................................................176
mode ..........................................................................................182
Network share ............................................................................195
Orientation .................................................................................185
Peak / Average ............................................................................184
Profiles ........................................................................................187
Recording ...................................................................................188
Recording license ........................................................................194
Recording mode ..........................................................................188
SDTV format ...............................................................................199
Settings .......................................................................................181
Software .....................................................................................192
Special ........................................................................................192
Streaming ...........................................................................188, 196
Version .......................................................................................192
Video ..........................................................................................199
Video quality ...............................................................................188
Video standard ............................................................................200
Web interface .............................................................................196
White balance .............................................................................185
Zoom ..........................................................................................185
HD Video live & streaming settings .........................................................181
HD Video menu
brightness ...................................................................................141
configuration menu ....................................................................155
HDTV format ..........................................................................................199
HDTV sub menu SD Video ......................................................................156
Hex key ..................................................................................................116
Hue submenu HD Video live & streaming ................................................186
Version 7.0
G-30-1781-en Page 269
S100 / OPMI pico
Lamp failure
Halogen ......................................................................................217
xenon ..........................................................................................218
LAN input/output ......................................................................................86
LED lighting system ...................................................................................21
Legal regulations ......................................................................................11
Letterbox ........................................................................................160, 199
Light guide socket ........................................................................71, 74, 75
Light guide, connection socket ...........................................................54, 75
Light guide, socket .............................................................................71, 74
Light Source submenu, HD video ............................................................162
Lightmeter sub menu SD Video ...............................................................143
Line fuses ................................................................................................228
Live ...........................................................................................................82
live ............................................................................................................78
Live video image of HD video ..................................................................168
Locking screws .......................................................................................116
Locking tab ...............................................................................................66
Low Light Auto (semiautomatic brightness adjustment) ..........................167
Version 7.0
Page 270 G-30-1781-en
S100 / OPMI pico
Version 7.0
G-30-1781-en Page 271
S100 / OPMI pico
Version 7.0
Page 272 G-30-1781-en
S100 / OPMI pico
Version 7.0
G-30-1781-en Page 273
S100 / OPMI pico
Version 7.0
Page 274 G-30-1781-en
S100 / OPMI pico
Version 7.0
G-30-1781-en Page 275
S100 / OPMI pico
Version 7.0
Page 276 G-30-1781-en
S100 / OPMI pico
Version 7.0
G-30-1781-en Page 277
Carl Zeiss Meditec AG
Goeschwitzer Strasse 51-52 Fax: + 49 (0)7364 - 20 4823
07745 Jena E-Mail: surgical@meditec.zeiss.com
Germany Internet: www.zeiss.com/med