Dymind Robot V2.0
Dymind Robot V2.0
Dymind Robot
Catalogue
Syringe Problem.................................................................................................................................. 1
0x230141C0-Failed to leave initial position (Sample syringe)....................................................2
0x230141D0-Failed to return to initial position (Sample syringe)............................................. 7
0x230341C0-Failed to leave initial position (LYSE syringe).......................................................12
0x230341D0-Failed to return to initial position (LYSE syringe)................................................ 17
0x230441C0-Failed to leave initial position (diluent syringe).................................................. 22
0x230441D0-Failed to return to initial position (diluent syringe)............................................25
0x230541C0-Failed to leave initial position (SH syringe)......................................................... 28
0x230541D0-Failed to return to initial position (SH syringe)................................................... 32
0x23014200-Sample syringe action overtime.......................................................................... 36
0x23034200-LYSE syringe action overtime............................................................................... 40
0x23044200-Diluent syringe action overtime.......................................................................... 41
0x23054200-SH syringe action overtime..................................................................................42
Motor Problem.................................................................................................................................. 43
0x230741C0-Horizontal motor failed to leave initial position................................................. 44
0x230741D0-Horizontal motor failed to return to initial position........................................... 56
0x230741F1-Horizontal motor failed to move to initial position.............................................67
0x230841C0-Vertical motor failed to leave initial position...................................................... 78
0x230841D0-Vertical motor failed to return to initial position................................................90
0x230841F1-Vertical motor failed to move to initial position............................................... 102
0x23074190-Horizontal motor photocoupler abnormal........................................................114
0x23074200-Horizontal motor action overtime.....................................................................122
0x23084190-Vertical motor photocoupler abnormal............................................................ 129
0x23084200-Vertical motor action overtime......................................................................... 137
Aperture Problem............................................................................................................................145
0xB1000505-WBC aperture voltage abnormal.......................................................................146
0xB1000506-RBC aperture voltage abnormal........................................................................ 154
0xB1000507-HGB background voltage abnormal.................................................................. 166
Clog Problem................................................................................................................................... 177
0x63026220-Flow cell clog......................................................................................................178
0x63026221-DIFF probe clog.................................................................................................. 182
0xb2004001-background abnormal....................................................................................... 186
0xB2004110-WBC Clog............................................................................................................188
0xB2004210-RBC Clog............................................................................................................. 197
Pressure Problem............................................................................................................................ 210
0x63006030-Positive pressure abnormal (low)......................................................................211
0x63006040-Positive pressure abnormal (high).....................................................................218
0x63006120-failed to build positive pressure........................................................................ 220
0x63016030- vacuum pressure abnormal (high)....................................................................227
0x63016040- vacuum pressure abnormal (low).....................................................................233
0x63016120- Failed to build vacuum pressure.......................................................................235
Syringe Problem
1
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Use Background
When analyzer gives a false alarm - Failed to leave initial position (Sample syringe).
Mechanism: The motor initial position photocoupler always detects that the photocoupler is blocked.
Declaration
26/3/2021, V2.0, For reference only.
Reference Standard
Model Possibility Solution
If we find the syringe block the photocoupler,
then we can check the status of it in
software, Status > Sensor, if it displays Not
Blocked.
A. Clean the photocoupler.
B. Replace the photocoupler.
1. Photocoupler behind the C. Reconnect or replace the photocoupler
sample syringe assembly has wire.
failure.
DH26,
DH31/DH33/DH36/
DH36VET,
DH36X,
A. Reconnect the motor wire.
DF50/DF51/DF52/D
B. Replace the motor wire if it’s open circuit
F50VET,
tested by multi-meter.
DF50CRP
Put the range to Diode, then test the two
ends of the wire from the same color, the
wire is ok if there’s a beep from multi-meter,
or broken.
2. Poor connection of sample
syringe motor wire.
2
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH26,DH31/DH33/DH36/DH36VET/DH36X,DF50/DF51/DF52/DF50VET/DF50CRP↑ DH56/UN73/DH76/D5CRP/D7CRP↓
3
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Blocked.
A. Clean the photocoupler.
B. Replace the photocoupler.
C. Reconnect or replace the photocoupler
wire.
4
Dymind Global Customer Service Center
Dymind Global Customer Service Center
If you have any problem with this file, please contact our engineer ASAP, thank you!
DH31/DH33/DH36/DH36VET,
DF50/DF51/DF52/DF50VET
DH26/DH31/DH33/DH36/DH
36VET,
DH36X, O-ring for 100ul
/
DF50CRP, sampling syringe
DH56/UN73/DH76/D5CRP/D7
CRP
DF50/DF51/DF52/DF50VET,
O-ring for 250ul
DH56/UN73/DH76, /
sampling syringe
D5CRP/D7CRP
5
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Sensor(Vertical) 23.99.0007C
6
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Use Background
When analyzer gives a false alarm - Failed to return to initial position (Sample syringe).
Mechanism: The motor initial position photocoupler always detects that the photocoupler is blocked.
Declaration
26/3/2021, V2.0, For reference only.
Reference Standard
Model Possibility Solution
If we find the syringe block the photocoupler,
then we can check the status of it in
software, Status > Sensor, if it displays Not
Blocked.
A. Clean the photocoupler.
B. Replace the photocoupler.
1. Photocoupler behind the C. Reconnect or replace the photocoupler
sample syringe assembly has wire.
failure.
DH26,
DH31/DH33/DH36/
DH36VET,
DH36X,
A. Reconnect the motor wire.
DF50/DF51/DF52/D
B. Replace the motor wire if it’s open circuit
F50VET,
tested by multi-meter.
DF50CRP
Put the range to Diode, then test the two
ends of the wire from the same color, the
wire is ok if there’s a beep from multi-meter,
or broken.
2. Poor connection of sample
syringe motor wire.
7
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH26,DH31/DH33/DH36/DH36VET/DH36X,DF50/DF51/DF52/DF50VET/DF50CRP↑ DH56/UN73/DH76/D5CRP/D7CRP↓
8
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Blocked.
A. Clean the photocoupler.
B. Replace the photocoupler.
C. Reconnect or replace the photocoupler
wire.
9
Dymind Global Customer Service Center
Dymind Global Customer Service Center
If you have any problem with this file, please contact our engineer ASAP, thank you!
DH31/DH33/DH36/DH36VET,
DF50/DF51/DF52/DF50VET
DH26/DH31/DH33/DH36/DH
36VET,
DH36X, O-ring for 100ul
/
DF50CRP, sampling syringe
DH56/UN73/DH76/D5CRP/D7
CRP
10
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DF50/DF51/DF52/DF50VET,
O-ring for 250ul
DH56/UN73/DH76, /
sampling syringe
D5CRP/D7CRP
Sensor(Vertical) 23.99.0007C
11
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Use Background
When analyzer gives a false alarm - Failed to leave initial position (LYSE syringe).
Mechanism: The motor initial position photocoupler always detects that the photocoupler is
blocked.
Declaration
14/4/2021, V2.0, For reference only.
Reference Standard
Model Possibility Solution
If we find the syringe block the photocoupler,
then we can check the status of it in
software, Status > Sensor, if it displays Not
Blocked.
A. Clean the photocoupler.
B. Replace the photocoupler.
1. Photocoupler behind the lyse C. Reconnect or replace the photocoupler
syringe assembly has failure. wire.
12
Dymind Global Customer Service Center
Dymind Global Customer Service Center
14
Dymind Global Customer Service Center
Dymind Global Customer Service Center
15
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Sensor(Vertical) 23.99.0007C
16
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Use Background
When analyzer gives a false alarm - Failed to return to initial position (LYSE syringe).
Mechanism: The motor initial position photocoupler always detects that the photocoupler is blocked.
Declaration
14/4/2021, V2.0, For reference only.
Reference Standard
Model Possibility Solution
If we find the syringe block the photocoupler,
then we can check the status of it in
software, Status > Sensor, if it displays Not
Blocked.
A. Clean the photocoupler.
B. Replace the photocoupler.
1. Photocoupler behind the lyse C. Reconnect or replace the photocoupler
syringe assembly has failure. wire.
17
Dymind Global Customer Service Center
Dymind Global Customer Service Center
19
Dymind Global Customer Service Center
Dymind Global Customer Service Center
20
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Sensor(Vertical) 23.99.0007C
21
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Use Background
When analyzer gives a false alarm - Failed to leave initial position (diluent syringe).
Mechanism: The motor initial position photocoupler always detects that the photocoupler is blocked.
Declaration
9/4/2021, V2.0, For reference only.
Reference Standard
Model Possibility Solution
If we find the syringe block the photocoupler,
then we can check the status of it in
software, Status > Sensor, if it displays Not
Blocked.
A. Clean the photocoupler.
B. Replace the photocoupler.
C. Reconnect or replace the photocoupler
wire.
DH36X
22
Dymind Global Customer Service Center
Dymind Global Customer Service Center
23
Dymind Global Customer Service Center
Dymind Global Customer Service Center
If you have any problem with this file, please contact our engineer ASAP, thank you!
24
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Use Background
When analyzer gives a false alarm - Failed to return to initial position (diluent syringe).
Mechanism: The motor initial position photocoupler always detects that the photocoupler is blocked.
Declaration
9/4/2021, V2.0, For reference only.
Reference Standard
Model Possibility Solution
If we find the syringe block the photocoupler,
then we can check the status of it in
software, Status > Sensor, if it displays Not
Blocked.
A. Clean the photocoupler.
B. Replace the photocoupler.
C. Reconnect or replace the photocoupler
wire.
DH36X
25
Dymind Global Customer Service Center
Dymind Global Customer Service Center
26
Dymind Global Customer Service Center
Dymind Global Customer Service Center
If you have any problem with this file, please contact our engineer ASAP, thank you!
27
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Use Background
When analyzer gives a false alarm - Failed to leave initial position (SH syringe).
Mechanism: After initialization command is executed, the initial position photocoupler is unblocked.
Declaration
7/4/2021, V2.0, For reference only.
Reference Standard
Model Possibility Solution
DF50CRP
28
Dymind Global Customer Service Center
Dymind Global Customer Service Center
4. Sheath fluid syringe moves with B. Check the O-ring of syringe, replace it if it
too large resistance. badly worn.
29
Dymind Global Customer Service Center
Dymind Global Customer Service Center
30
Dymind Global Customer Service Center
Dymind Global Customer Service Center
31
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Use Background
When analyzer gives a false alarm - Failed to return to initial position (SH syringe).
Mechanism: After initialization command is executed, the initial position photocoupler is unblocked.
Declaration
7/4/2021, V2.0, For reference only.
Reference Standard
Model Possibility Solution
DF50CRP
32
Dymind Global Customer Service Center
Dymind Global Customer Service Center
4. Sheath fluid syringe moves with B. Check the O-ring of syringe, replace it if it
too large resistance. badly worn.
33
Dymind Global Customer Service Center
Dymind Global Customer Service Center
34
Dymind Global Customer Service Center
Dymind Global Customer Service Center
35
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Use Background
When analyzer gives a false alarm - Sample syringe action overtime.
Mechanism: Failed to complete the action within the time specified by fluidics.
Declaration
9/4/2021, V2.0, For reference only.
Reference Standard
Model Possibility Solution
A. Apply fresh lubricating oil after cleaning
the sliding rail and screw rode.
DH21,
DH31/DH33/DH
1. Sample syringe moves with too B. Check the O-ring of syringe, replace it if it
36/DH36VET,
large resistance. got worn.
DF50/DF51/DF5
2/DF50VET
1. Sample syringe moves with too A. Apply fresh lubricating oil after cleaning
DH36X
large resistance. the sliding rail and screw rode.
36
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH56/UN73/DH 1. Sample syringe moves with too A. Apply fresh lubricating oil after cleaning
37
Dymind Global Customer Service Center
Dymind Global Customer Service Center
If you have any problem with this file, please contact our engineer ASAP, thank you!
Relative spare part code
Model Part Name P/N Picture
DH26/DH31/DH33/DH36/DH36VET,
DH36X,
O-ring for 100ul
DF50CRP, /
sampling syringe
DH56/UN73/DH76,
D5CRP/D7CRP
DF50/DF51/DF52/DF50VET,
O-ring for 250ul
DH56/UN73/DH76, /
sampling syringe
D5CRP/D7CRP
38
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH26/DH31/DH33/DH36/DH36VET,
O-ring for Lyse
DH36X, /
Syringe 1
DF50/DF51/DF52/DF50VET
DH26/DH31/DH33/DH36/DH36VET,
O-ring for Lyse
DH36X, /
Syringe 2
DF50/DF51/DF52/DF50VET
39
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Use Background
When analyzer gives a false alarm - LYSE syringe action overtime.
Mechanism: Failed to complete the action within the time specified by fluidics.
Declaration
10/4/2021, V2.0, For reference only.
Reference Standard
Model Possibility Solution
A. Apply fresh lubricating oil after cleaning
the sliding rail and screw rode.
DH36X,
DF50CRP,
DH56, 1. Lyse syringe moves with too large
UN73/DH76, resistance.
D5CRP/D7CRP
B. Check the O-ring of syringe, replace it if
damaged.
If you have any problem with this file, please contact our engineer ASAP, thank you!
40
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Use Background
When analyzer gives a false alarm - Diluent syringe action overtime.
Mechanism: Failed to complete the action within the time specified by fluidics.
Declaration
10/4/2021, V2.0, For reference only.
Reference Standard
Model Possibility Solution
A. Apply fresh lubricating oil after cleaning
the sliding rail and screw rode.
DH36X,
DH56/UN73/DH 1. Diluent syringe moves with too
76, large resistance.
D5CRP/D7CRP
B. Check the O-ring of syringe, replace it if
damaged.
41
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Use Background
When analyzer gives a false alarm - SH syringe action overtime.
Mechanism: Failed to complete the action within the time specified by fluidics.
Declaration
10/4/2021, V2.0, For reference only.
Reference Standard
Model Possibility Solution
A. Apply fresh lubricating oil after cleaning
the sliding rail and screw rode.
DF50CRP,
DH56/UN73/DH 1. Sheath fluid syringe moves with
76, too large resistance.
D5CRP/D7CRP
B. Check the O-ring of syringe, replace it if it
badly worn.
42
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Motor Problem
43
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Use Background
When analyzer gives a false alarm: Horizontal motor failed to leave initial position.
Mechanism: The horizontal motor initial position optocoupler always detects that the optocoupler is
blocked.
Declaration
08/9/2020, V1.0, For reference only.
Reference Standard
Model Possibility Solution
DH26
Please press Service<Sensor to test if this
sensor working or not by a paper card, if
there is no response, please
1. Change the sensor
2. Change the sensor cable
2.Horizontal motor initial position
optocoupler malfunctions.
44
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH26↑ DH31/DH33/DH36/DH36VET↓
45
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH31/DH33/DH36/DH36VET↑ DH36X↓
Apply some lubricant and move it back
and forth a few times.
46
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH36X↑ DF50/DF51/DF52/DF50VET↓
Apply some lubricant and move it back
and forth a few times.
47
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DF50/DF51/DF52/DF50VET↑ DF50CRP↓
Apply some lubricant and move it back
and forth a few times.
48
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DF50CRP↑ DH73/DH76↓
Apply some lubricant and move it back
and forth a few times.
49
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH73/DH76↑ D5-CRP↓
Apply some lubricant and move it back
and forth a few times.
50
Dymind Global Customer Service Center
Dymind Global Customer Service Center
D5-CRP↑ D7-CRP↓
Apply some lubricant and move it back
and forth a few times.
51
Dymind Global Customer Service Center
Dymind Global Customer Service Center
If you have any problem with this file, please contact our engineer ASAP, thank you!
Horizontal Sampling
68.02.0026A
Motor
Sampling Motor
37.02.0145A
Connection Cable
Horizontal Sampling
DH31/DH33/DH36/DH36VET 68.02.0026A
Motor
52
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Sensor(Vertical) 23.99.0007C
Sampling optocoupler
37.02.0036B
connection cable
Horizontal Sampling
68.02.0026A
Motor
Sensor(Vertical) 23.99.0007C
DF50/DF51/DF52/DF50VET
Sampling optocoupler
37.02.0036B
connection cable
Horizontal Sampling
DF50CRP 68.02.0026A
Motor
53
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Sensor(Vertical) 23.99.0007C
Horizontal &vertical
37.02.0035A
motor connection cable
Sampling optocoupler
37.02.0036B
connection cable
Horizontal Sampling
68.02.0026A
Motor
Sensor(Vertical) 23.99.0007C
Sampling motor
37.01.0042A
DH73/DH76 connection cable
Sampling Optocoupler
37.02.0142A
interchange cable
Sampling Optocoupler
37.02.0051A
connection cable1
54
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Sampling Optocoupler
37.02.0052A
connection cable2
Horizontal Sampling
68.02.0026A
Motor
DH36X/D5-CRP/D7-CRP
Sensor(Vertical) 23.99.0007C
55
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Use Background
When analyzer gives a false alarm: Horizontal motor failed to return to initial position.
Mechanism: The horizontal motor initial position optocoupler is always in the unblocked status.
Declaration
08/9/2020, V1.0, For reference only.
Reference Standard
Model Possibility Solution
56
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH26↑ DH31/DH33/DH36/DH36VET↓
57
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH31/DH33/DH36/DH36VET↑ DH36X↓
Apply some lubricant and move it back
and forth a few times.
58
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH36X↑ DF50/DF51/DF52/DF50VET↓
Apply some lubricant and move it back
and forth a few times.
59
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DF50/DF51/DF52/DF50VET↑ DF50CRP↓
Apply some lubricant and move it back
and forth a few times.
60
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DF50CRP↑ DH73/DH76↓
Apply some lubricant and move it back
and forth a few times.
DH73/DH76
61
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH73/DH76↑ D5-CRP↓
Apply some lubricant and move it back
and forth a few times.
D5-CRP
62
Dymind Global Customer Service Center
Dymind Global Customer Service Center
D5-CRP↑ D7-CRP↓
Apply some lubricant and move it back
and forth a few times.
63
Dymind Global Customer Service Center
Dymind Global Customer Service Center
If you have any problem with this file, please contact our engineer ASAP, thank you!
Horizontal Sampling
68.02.0026A
Motor
Sampling Motor
37.02.0145A
Connection Cable
Horizontal Sampling
68.02.0026A
Motor
Sensor(Vertical) 23.99.0007C
DH31/DH33/DH36/DH36VET
Sampling optocoupler
37.02.0036B
connection cable
64
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Horizontal Sampling
68.02.0026A
Motor
Sensor(Vertical) 23.99.0007C
DF50/DF51/DF52/DF50VET
Sampling optocoupler
37.02.0036B
connection cable
Horizontal Sampling
68.02.0026A
Motor
Sensor(Vertical) 23.99.0007C
DF50CRP
Horizontal &vertical
37.02.0035A
motor connection cable
Sampling optocoupler
37.02.0036B
connection cable
65
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Horizontal Sampling
68.02.0026A
Motor
Sensor(Vertical) 23.99.0007C
Sampling motor
37.01.0042A
connection cable
DH73/DH76
Sampling Optocoupler
37.02.0142A
interchange cable
Sampling Optocoupler
37.02.0051A
connection cable1
Sampling Optocoupler
37.02.0052A
connection cable2
Horizontal Sampling
68.02.0026A
Motor
DH36X/D5-CRP/D7-CRP
Sensor(Vertical) 23.99.0007C
66
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Use Background
When analyzer gives a false alarm: Horizontal motor failed to move to initial position.
Mechanism: The horizontal motor initial position optocoupler is always in the unblocked status.
Declaration
08/9/2020, V1.0, For reference only.
Reference Standard
Model Possibility Solution
67
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH26↑ DH31/DH33/DH36/DH36VET↓
68
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH31/DH33/DH36/DH36VET↑ DH36X↓
Apply some lubricant and move it back
and forth a few times.
69
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH36X↑ DF50/DF51/DF52/DF50VET↓
Apply some lubricant and move it back
and forth a few times.
70
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DF50/DF51/DF52/DF50VET↑ DF50CRP↓
Apply some lubricant and move it back
and forth a few times.
71
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DF50CRP↑ DH73/DH76↓
Apply some lubricant and move it back
and forth a few times.
DH73/DH76
72
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH73/DH76↑ D5-CRP↓
Apply some lubricant and move it back
and forth a few times.
D5-CRP
73
Dymind Global Customer Service Center
Dymind Global Customer Service Center
D5-CRP↑ D7-CRP↓
Apply some lubricant and move it back
and forth a few times.
74
Dymind Global Customer Service Center
Dymind Global Customer Service Center
If you have any problem with this file, please contact our engineer ASAP, thank you!
Horizontal Sampling
68.02.0026A
Motor
Sampling Motor
37.02.0145A
Connection Cable
Horizontal Sampling
68.02.0026A
Motor
Sensor(Vertical) 23.99.0007C
DH31/DH33/DH36/DH36VET
Sampling optocoupler
37.02.0036B
connection cable
75
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Horizontal Sampling
68.02.0026A
Motor
Sensor(Vertical) 23.99.0007C
DF50/DF51/DF52/DF50VET
Sampling optocoupler
37.02.0036B
connection cable
Horizontal Sampling
68.02.0026A
Motor
Sensor(Vertical) 23.99.0007C
DF50CRP
Horizontal &vertical
37.02.0035A
motor connection cable
Sampling optocoupler
37.02.0036B
connection cable
76
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Horizontal Sampling
68.02.0026A
Motor
Sensor(Vertical) 23.99.0007C
Sampling motor
37.01.0042A
connection cable
DH73/DH76
Sampling Optocoupler
37.02.0142A
interchange cable
Sampling Optocoupler
37.02.0051A
connection cable1
Sampling Optocoupler
37.02.0052A
connection cable2
Horizontal Sampling
68.02.0026A
Motor
DH36X/D5-CRP/D7-CRP
Sensor(Vertical) 23.99.0007C
77
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Use Background
When analyzer gives a false alarm: Vertical motor failed to leave initial position.
Mechanism: The vertical motor initial position optocoupler always detects that the optocoupler is
blocked.
Declaration
08/9/2020, V1.0, For reference only.
Reference Standard
Model Possibility Solution
DH26
78
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH26↑ DH31/DH33/DH36/DH36VET↓
79
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH31/DH33/DH36/DH36VET↑ DH36X↓
DH36X
80
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH36X↑ DF50/DF51/DF52/DF50VET↓
Apply some lubricant and move it back
and forth a few times.
81
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DF50/DF51/DF52/DF50VET↑ DF50CRP↓
DF50CRP↑ DH73/DH76↓
Apply some lubricant and move it back
and forth a few times.
DH73/DH76
83
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH73/DH76↑ D5-CRP↓
Apply some lubricant and move it back
and forth a few times.
84
Dymind Global Customer Service Center
Dymind Global Customer Service Center
D5-CRP↑ D7-CRP↓
Apply some lubricant and move it back
and forth a few times.
D7-CRP
85
Dymind Global Customer Service Center
Dymind Global Customer Service Center
If you have any problem with this file, please contact our engineer ASAP, thank you!
Vertical Sampling
DH26 68.02.0027A
Motor
86
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Sensor(Vertical) 23.99.0007C
Sampling Motor
37.02.0145A
Connection Cable
Vertical Sampling
68.02.0027A
Motor
Sensor(Vertical) 23.99.0007C
DH31/DH33/DH36/DH36VET
Sampling optocoupler
37.02.0036B
connection cable
Vertical Sampling
68.02.0027A
Motor
DF50/DF51/DF52/DF50VET
Sensor(Vertical) 23.99.0007C
87
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Sampling optocoupler
37.02.0036B
connection cable
Vertical Sampling
68.02.0027A
Motor
Sensor(Vertical) 23.99.0007C
DF50CRP
Horizontal &vertical
37.02.0035A
motor connection cable
Sampling optocoupler
37.02.0036B
connection cable
Vertical Sampling
68.02.0027A
Motor
DH73/DH76
Sensor(Vertical) 23.99.0007C
88
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Sampling motor
37.01.0042A
connection cable
Sampling Optocoupler
37.02.0142A
interchange cable
Sampling Optocoupler
37.02.0051A
connection cable1
Sampling Optocoupler
37.02.0052A
connection cable2
Vertical Sampling
68.02.0027A
Motor
DH36X/D5-CRP/D7-CRP
Sensor(Vertical) 23.99.0007C
89
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Use Background
When analyzer gives a false alarm: Vertical motor failed to return to initial position.
Mechanism: The vertical motor initial position optocoupler is always in the unblocked status.
Declaration
08/9/2020, V1.0, For reference only.
Reference Standard
Model Possibility Solution
90
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH26↑ DH31/DH33/DH36/DH36VET↓
91
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH31/DH33/DH36/DH36VET↑ DH36X↓
92
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH36X↑ DF50/DF51/DF52/DF50VET↓
Apply some lubricant and move it back
and forth a few times.
93
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DF50/DF51/DF52/DF50VET↑ DF50CRP↓
DF50CRP
94
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DF50CRP↑ DH73/DH76↓
Apply some lubricant and move it back
and forth a few times.
95
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH73/DH76↑ D5-CRP↓
Apply some lubricant and move it back
and forth a few times.
D5-CRP
96
Dymind Global Customer Service Center
Dymind Global Customer Service Center
D5-CRP↑ D7-CRP↓
Apply some lubricant and move it back
and forth a few times.
97
Dymind Global Customer Service Center
Dymind Global Customer Service Center
If you have any problem with this file, please contact our engineer ASAP, thank you!
Vertical Sampling
68.02.0027A
Motor
Sampling Motor
37.02.0145A
Connection Cable
98
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Vertical Sampling
68.02.0027A
Motor
Sensor(Vertical) 23.99.0007C
DH31/DH33/DH36/DH36VET
Sampling optocoupler
37.02.0036B
connection cable
Vertical Sampling
68.02.0027A
Motor
Sensor(Vertical) 23.99.0007C
DF50/DF51/DF52/DF50VET
Sampling optocoupler
37.02.0036B
connection cable
99
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Vertical Sampling
68.02.0027A
Motor
Sensor(Vertical) 23.99.0007C
DF50CRP
Horizontal &vertical
37.02.0035A
motor connection cable
Sampling optocoupler
37.02.0036B
connection cable
Vertical Sampling
68.02.0027A
Motor
Sensor(Vertical) 23.99.0007C
DH73/DH76
Sampling motor
37.01.0042A
connection cable
Sampling Optocoupler
37.02.0142A
interchange cable
100
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Sampling Optocoupler
37.02.0051A
connection cable1
Sampling Optocoupler
37.02.0052A
connection cable2
Vertical Sampling
68.02.0027A
Motor
DH36X/D5-CRP/D7-CRP
Sensor(Vertical) 23.99.0007C
101
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Use Background
When analyzer gives a false alarm: Horizontal motor failed to move to initial position.
Mechanism: The Vertical motor initial position optocoupler is always in the unblocked status.
Declaration
08/9/2020, V1.0, For reference only.
Reference Standard
Model Possibility Solution
102
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH26↑ DH31/DH33/DH36/DH36VET↓
103
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH31/DH33/DH36/DH36VET↑ DH36X↓
104
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH36X↑ DF50/DF51/DF52/DF50VET↓
Apply some lubricant and move it back
and forth a few times.
105
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DF50/DF51/DF52/DF50VET↑ DF50CRP↓
DF50CRP
106
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DF50CRP↑ DH73/DH76↓
Apply some lubricant and move it back
and forth a few times.
107
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH73/DH76↑ D5-CRP↓
Apply some lubricant and move it back
and forth a few times.
D5-CRP
108
Dymind Global Customer Service Center
Dymind Global Customer Service Center
D5-CRP↑ D7-CRP↓
Apply some lubricant and move it back
and forth a few times.
109
Dymind Global Customer Service Center
Dymind Global Customer Service Center
If you have any problem with this file, please contact our engineer ASAP, thank you!
Vertical Sampling
68.02.0027A
Motor
Sampling Motor
37.02.0145A
Connection Cable
110
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Vertical Sampling
68.02.0027A
Motor
Sensor(Vertical) 23.99.0007C
DH31/DH33/DH36/DH36VET
Sampling optocoupler
37.02.0036B
connection cable
Vertical Sampling
68.02.0027A
Motor
Sensor(Vertical) 23.99.0007C
DF50/DF51/DF52/DF50VET
Sampling optocoupler
37.02.0036B
connection cable
111
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Vertical Sampling
68.02.0027A
Motor
Sensor(Vertical) 23.99.0007C
DF50CRP
Horizontal &vertical
37.02.0035A
motor connection cable
Sampling optocoupler
37.02.0036B
connection cable
Vertical Sampling
68.02.0027A
Motor
Sensor(Vertical) 23.99.0007C
DH73/DH76
Sampling motor
37.01.0042A
connection cable
Sampling Optocoupler
37.02.0142A
interchange cable
112
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Sampling Optocoupler
37.02.0051A
connection cable1
Sampling Optocoupler
37.02.0052A
connection cable2
Vertical Sampling
68.02.0027A
Motor
DH36X/D5-CRP/D7-CRP
Sensor(Vertical) 23.99.0007C
113
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Use Background
When analyzer gives a false alarm: Horizontal motor photocoupler abnormal.
Mechanism: Photocoupler error occurred in the initial position, WBC bath, DIFF bath, RBC bath,
open-sampling position, or auto-sampling position of the horizontal motor.
Declaration
17/8/2020, V1.0, For reference only.
Reference Standard
Model Possibility Solution
DH26
Please press Service<Sensor to test if this
sensor working or not by a paper card, if
there is no response, please
13. Change the sensor
14. Change the sensor cable
2.Horizontal motor initial position
optocoupler malfunctions.
114
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH26↑ DH31/DH33/DH36/DH36VET↓
DH31/DH33/DH36/DH36VET↑ DH36X↓
1.Horizontal motor has too large Apply some lubricant and move it back
DH36X
Resistance and forth a few times.
115
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH36X↑ DF50/DF51/DF52/DF50VET↓
Apply some lubricant and move it back
and forth a few times.
DF50/DF51/DF
52/DF50VET
116
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DF50/DF51/DF52/DF50VET↑ DF50CRP↓
Apply some lubricant and move it back
and forth a few times.
DF50CRP↑ DH73/DH76↓
117
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH73/DH76↑ D5-CRP↓
Apply some lubricant and move it back
and forth a few times.
D5-CRP↑ D7-CRP↓
Apply some lubricant and move it back
and forth a few times.
119
Dymind Global Customer Service Center
Dymind Global Customer Service Center
If you have any problem with this file, please contact our engineer ASAP, thank you!
Sensor(Vertical) 23.99.0007C
DH31/DH33/DH36/DH36VET
DF50/DF51/DF52/DF50VET
Sampling optocoupler
37.02.0036B
connection cable
Sensor(Vertical) 23.99.0007C
DF50CRP
Sampling optocoupler
37.02.0036B
connection cable
120
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Sensor(Vertical) 23.99.0007C
Sampling Optocoupler
37.02.0142A
interchange cable
DH73/DH76
Sampling Optocoupler
37.02.0051A
connection cable1
Sampling Optocoupler
37.02.0052A
connection cable2
121
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Use Background
When analyzer gives a false alarm: Horizontal motor action overtime.
Mechanism: Failed to complete the action within the time specified by fluidics.
Declaration
17/8/2020, V1.0, For reference only.
Reference Standard
Model Possibility Solution
DH26
DH26↑ DH31/DH33/DH36/DH36VET↓
Apply some lubricant and move it back
and forth a few times.
122
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH31/DH33/DH36/DH36VET↑ DH36X↓
Apply some lubricant and move it back
and forth a few times.
DH36X
DH36X↑ DF50/DF51/DF52/DF50VET↓
Apply some lubricant and move it back
and forth a few times.
DF50/DF51/DF
52/DF50VET
DF50/DF51/DF52/DF50VET↑ DF50CRP↓
123
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DF50CRP
DF50CRP↑ DH73/DH76↓
Apply some lubricant and move it back
and forth a few times.
DH73/DH76
DH73/DH76↑ D5-CRP↓
1.Horizontal motor has too large Apply some lubricant and move it back
D5-CRP
Resistance and forth a few times.
124
Dymind Global Customer Service Center
Dymind Global Customer Service Center
D5-CRP↑ D7-CRP↓
Apply some lubricant and move it back
and forth a few times.
D7-CRP
If you have any problem with this file, please contact our engineer ASAP, thank you!
125
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Horizontal Sampling
68.02.0026A
Motor
Sampling Motor
37.02.0145A
Connection Cable
Horizontal Sampling
68.02.0026A
Motor
Sensor(Vertical) 23.99.0007C
DH31/DH33/DH36/DH36VET
Sampling optocoupler
37.02.0036B
connection cable
Horizontal Sampling
DF50/DF51/DF52/DF50VET 68.02.0026A
Motor
126
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Sensor(Vertical) 23.99.0007C
Sampling optocoupler
37.02.0036B
connection cable
Horizontal Sampling
68.02.0026A
Motor
Sensor(Vertical) 23.99.0007C
DF50CRP
Horizontal &vertical
37.02.0035A
motor connection cable
Sampling optocoupler
37.02.0036B
connection cable
Horizontal Sampling
68.02.0026A
DH73/DH76 Motor
127
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Sensor(Vertical) 23.99.0007C
Sampling motor
37.01.0042A
connection cable
Sampling Optocoupler
37.02.0142A
interchange cable
Sampling Optocoupler
37.02.0051A
connection cable1
Sampling Optocoupler
37.02.0052A
connection cable2
Horizontal Sampling
68.02.0026A
Motor
DH36X/D5-CRP/D7-CRP
Sensor(Vertical) 23.99.0007C
128
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Use Background
When analyzer gives a false alarm: Vertical motor photocoupler abnormal.
Mechanism: Photocoupler error occurred in the initial position or aspirate position of the vertical
motor.
Declaration
09/9/2020, V1.0, For reference only.
Reference Standard
Model Possibility Solution
129
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH26↑ DH31/DH33/DH36/DH36VET↓
DH31/DH33/DH36/DH36VET↑ DH36X↓
130
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH36X↑ DF50/DF51/DF52/DF50VET↓
Apply some lubricant and move it back
and forth a few times.
131
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DF50/DF51/DF52/DF50VET↑ DF50CRP↓
DF50CRP↑ DH73/DH76↓
132
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH73/DH76
DH73/DH76↑ D5-CRP↓
Apply some lubricant and move it back
and forth a few times.
133
Dymind Global Customer Service Center
Dymind Global Customer Service Center
D5-CRP↑ D7-CRP↓
Apply some lubricant and move it back
and forth a few times.
134
Dymind Global Customer Service Center
Dymind Global Customer Service Center
If you have any problem with this file, please contact our engineer ASAP, thank you!
Sensor(Vertical) 23.99.0007C
DH26
Sampling Motor
37.02.0145A
Connection Cable
Sensor(Vertical) 23.99.0007C
DH31/DH33/DH36/DH36VET
Sampling optocoupler
37.02.0036B
connection cable
135
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Sampling optocoupler
37.02.0036B
connection cable
Sensor(Vertical) 23.99.0007C
DF50CRP
Sampling optocoupler
37.02.0036B
connection cable
Sensor(Vertical) 23.99.0007C
Sampling Optocoupler
37.02.0142A
interchange cable
DH73/DH76
Sampling Optocoupler
37.02.0051A
connection cable1
Sampling Optocoupler
37.02.0052A
connection cable2
136
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Use Background
When analyzer gives a false alarm: Vertical motor action overtime.
Mechanism: Failed to complete the action within the time specified by fluidics.
Declaration
09/9/2020, V1.0, For reference only.
Reference Standard
Model Possibility Solution
Apply some lubricant and move it back
and forth a few times.
DH26
Please change the whole Vertical
assembly.
DH26↑ DH31/DH33/DH36/DH36VET↓
137
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH31/DH33/D
H36/DH36VET 2.Action loses steps caused by mechanical
ageing of the Vertical assembly.
DH31/DH33/DH36/DH36VET↑ DH36X↓
DH36X
DH36X↑ DF50/DF51/DF52/DF50VET↓
DF50/DF51/DF 1.Vertical motor has too large Apply some lubricant and move it back
52/DF50VET Resistance and forth a few times.
138
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DF50/DF51/DF52/DF50VET↑ DF50CRP↓
DF50CRP
DF50CRP↑ DH73/DH76↓
139
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH73/DH76
Please change the whole Vertical
assembly.
DH73/DH76↑ D5-CRP↓
Apply some lubricant and move it back
and forth a few times
D5-CRP
Please change the whole Vertical
assembly.
D5-CRP↑ D7-CRP↓
1.Vertical motor has too large Apply some lubricant and move it back
D7-CRP
Resistance and forth a few times
140
Dymind Global Customer Service Center
Dymind Global Customer Service Center
If you have any problem with this file, please contact our engineer ASAP, thank you!
Vertical Sampling
68.02.0027A
Motor
DH26
Sensor(Vertical) 23.99.0007C
141
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Sampling Motor
37.02.0145A
Connection Cable
Vertical Sampling
68.02.0027A
Motor
Sensor(Vertical) 23.99.0007C
DH31/DH33/DH36/DH36VET
Sampling optocoupler
37.02.0036B
connection cable
Vertical Sampling
68.02.0027A
Motor
Sampling optocoupler
37.02.0036B
connection cable
142
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Vertical Sampling
68.02.0027A
Motor
Sensor(Vertical) 23.99.0007C
DF50CRP
Horizontal &vertical
37.02.0035A
motor connection cable
Sampling optocoupler
37.02.0036B
connection cable
Vertical Sampling
68.02.0027A
Motor
Sensor(Vertical) 23.99.0007C
DH73/DH76
Sampling motor
37.01.0042A
connection cable
143
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Sampling Optocoupler
37.02.0142A
interchange cable
Sampling Optocoupler
37.02.0051A
connection cable1
Sampling Optocoupler
37.02.0052A
connection cable2
Vertical Sampling
68.02.0027A
Motor
DH36X/D5-CRP/D7-CRP
Sensor(Vertical) 23.99.0007C
144
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Aperture Problem
145
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Use Background
When analyzer gives a false alarm - WBC aperture voltage abnormal.
Mechanism: The WBC aperture voltage is out of the allowable range [1.45, 2.05] V, which can be
checked in Self-test > Other self-test > WBC aperture voltage.
Declaration
31/3/2021, V2.0, For reference only.
Reference Standard
Model Possibility Solution
Check if there’s any diluent in RBC chamber.
A. If there is no, then check the diluent box,
the syringe, the valve connected to syringe.
B. If there is, then do a blank test, confirm
whether the problem is no diluent dispense
into WBC chamber or WBC chamber leakage.
Replace LV6, isolation chamber, LV10
respectively to confirm.
2. Some bubbles go through WBC Replace LV1, LV7, the tube connected, and
aperture or nearby. the connector respectively to confirm.
146
Dymind Global Customer Service Center
Dymind Global Customer Service Center
148
Dymind Global Customer Service Center
Dymind Global Customer Service Center
expired.
149
Dymind Global Customer Service Center
Dymind Global Customer Service Center
150
Dymind Global Customer Service Center
Dymind Global Customer Service Center
expired.
If you have any problem with this file, please contact our engineer ASAP, thank you!
Relative spare part code
Model Part Name P/N Picture
151
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH31/DH33/DH36/
Mainboard 22.01.0023A
DH36VET
DH31/DH33/DH36/
DH36VET, WBC bath assembly(with Ruby) 20.01.0143A
DH36X
152
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Mainboard 22.01.0072A
DH26
DH36X,
Isolation Chamber with
DH56/UN73/DH76, 60.01.0012C
Strainer
D5CRP/D7CRP
Mainboard 22.01.0001C
DH56/UN73/DH76,
WBC bath assembly(with Ruby) 20.01.0007A
D5CRP/D7CRP
153
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Use Background
When analyzer gives a false alarm - RBC aperture voltage abnormal.
Mechanism: The RBC aperture voltage is out of the allowable range [1.65, 2.25] V, which can be
checked in Self-test > Other self-test > RBC aperture voltage.
Declaration
31/3/2021, V2.0, For reference only.
Reference Standard
Model Possibility Solution
154
Dymind Global Customer Service Center
Dymind Global Customer Service Center
155
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH36X
156
Dymind Global Customer Service Center
Dymind Global Customer Service Center
157
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DF50/DF51/DF5
2/DF50VET Check the tube, connectors and valves
connected to RBC aperture. Replace LV9,
LV18, the tube connected, and the
connectors respectively to locate it.
158
Dymind Global Customer Service Center
Dymind Global Customer Service Center
expired.
159
Dymind Global Customer Service Center
Dymind Global Customer Service Center
160
Dymind Global Customer Service Center
Dymind Global Customer Service Center
failure.
161
Dymind Global Customer Service Center
Dymind Global Customer Service Center
aperture or nearby.
162
Dymind Global Customer Service Center
Dymind Global Customer Service Center
If you have any problem with this file, please contact our engineer ASAP, thank you!
163
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH31/DH33/DH36/DH36VET,
DH36X, RBC bath assembly(with
20.01.0144A
DF50/DF51/DF52/DF50VET, Ruby)
DF50CRP
DH26
DH36X,
DF50CRP, Isolation Chamber with
60.01.0012C
DH56/UN73/DH76, Strainer
D5CRP/D7CRP
DF50/DF51/DF52/DF50VET,
Mainboard 22.01.0021C
DF50CRP
DH56/UN73/DH76,
Mainboard 22.01.0001C
D5CRP/D7CRP
164
Dymind Global Customer Service Center
Dymind Global Customer Service Center
165
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Use Background
When analyzer gives a false alarm - HGB background voltage abnormal.
Mechanism: The HGB background voltage is not in the allowable range [4.2~4.8]V.
Declaration
26/3/2021, V2.0, For reference only.
Reference Standard
Model Possibility Solution
Go to Gain Settings to set the HGB blank voltage
to 4.5±0.1V.
DH26/DH31/D
H33/DH36/DH
36VET Check if there’s any diluent in RBC chamber.
A. If there is no, then check the diluent box, the
syringe, the valve connected to syringe.
B. If there is, then do a blank test, confirm the
2. There is no diluent or a little
problem is no diluent dispense into WBC chamber
leakage or overfill in WBC
or WBC chamber leakage. Replace LV6, isolation
chamber.
chamber, LV10 respectively to locate it.
Don’t forget to clean the outer-wall of WBC
chamber and HGB unit by non-dust cloth after fix
the overfill problem.
166
Dymind Global Customer Service Center
Dymind Global Customer Service Center
167
Dymind Global Customer Service Center
Dymind Global Customer Service Center
168
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DF50/DF51/DF
52/DF50VET Check if there’s any diluent in RBC chamber.
A. If there is no, then check the diluent box, the
syringe, the valve connected to syringe.
B. If there is, then do a blank test, confirm the
2. There is no diluent or a little
problem is no diluent dispense into WBC chamber
leakage or overfill in WBC
or WBC chamber leakage. Replace LV7, LV14,
chamber.
isolation chamber, LV19 respectively to locate it.
Don’t forget to clean the outer-wall of WBC
chamber and HGB unit by non-dust cloth after fix
the overfill problem.
169
Dymind Global Customer Service Center
Dymind Global Customer Service Center
170
Dymind Global Customer Service Center
Dymind Global Customer Service Center
171
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH56/UN73/D
H76, Check if there’s any diluent in RBC chamber,
D5CRP/D7CRP A. If there is no, then check the diluent box, the
syringe, the valve connected to syringe.
B. If there is, then do a blank test, confirm the
2. There is no diluent or a little problem is no diluent dispense into WBC chamber
leakage or overfill in WBC or WBC chamber leakage. Replace LV8, isolation
chamber. chamber, LV15, LV21(DH56&D5CRP don’t have),
LV25 respectively to locate it.
Don’t forget to clean the outer-wall of WBC
chamber and HGB unit by non-dust cloth after fix
the overfill problem.
172
Dymind Global Customer Service Center
Dymind Global Customer Service Center
If you have any problem with this file, please contact our engineer ASAP, thank you!
173
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH31/DH33/DH36/DH36VE,
DH36X,
HGB unit 68.02.0024A
DH56/UN73/DH76,
D5CRP/D7CRP
174
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH36X,
DF50/DF51/DF52/DF50VET,
Isolation Chamber with
DF50CRP, 60.01.0012C
Strainer
DH56/UN73/DH76,
D5CRP/D7CRP
DH31/DH33/DH36/DH36VET
Mainboard 22.01.0023A
Mainboard 22.01.0021C
175
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Mainboard 22.01.0001C
176
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Clog Problem
177
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Use Background
When “Flow cell Clog” is prompted in the area of problem on the analyzer interface and the RBC
counting result of the analyzer is abnormal of the repeatability analyzers are not satisfied.
During the counting, flow cell is clog, which slows down the fluid from passing through cell flow,
increases the sheath pressure sensor voltage, and triggers the clogging alarm.
Declaration
9/4/2021, V1.0, For reference only.
Reference Standard
Module Possibility Solution
DH36 N/A N/A
We could flash the flow cell manually.
(Power off the analyzer, then aspirate the diluent out)
We could get these videos of how to do maintenance for
analyzer in our FTP account, as following:
Website: http://159.138.130.90/
User name: S-user03
Password: PVns6bPL
The blood clot or
some big particle
block the flow cell
For
DF50/
DF51/
DF52/
DF50VET
a. Check the process as shown below, and replace LV11
b. Check the connector (T9\T10\J13) and replace it
Hydraulic system
analysis
178
Dymind Global Customer Service Center
Dymind Global Customer Service Center
The sheath
For DF50/ pressure sensor
DF52 damaged and give
the false alarm
For
DF50CRP
Hydraulic system
analysis
179
Dymind Global Customer Service Center
Dymind Global Customer Service Center
The sheath
For pressure sensor
DF50CRP damaged and give
the false alarm
For
DH56/
D5-CRP
DH76/
D7-CRP a. Check the process as shown below, and replace LV18
b. Check the connector (T70\T29\J2) and replace it
Hydraulic system
analysis
180
Dymind Global Customer Service Center
Dymind Global Customer Service Center
For
DH56/ The sensor
D5-CRP damaged and give
DH76/ the false alarm
D7-CRP
If you have any problem with this file, please contact our engineer ASAP, thank you!
All
181
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Use Background
When “DIFF probe Clog” is prompted in the area of problem on the analyzer interface and the RBC
counting result of the analyzer is abnormal of the repeatability analyzers are not satisfied.
During the counting, flow cell is clog, which slows down the fluid from passing through cell flow,
increases the sheath pressure sensor voltage, and triggers the clogging alarm.
Declaration
4/9/2021, V1.0, For reference only.
Reference Standard
Module Possibility Solution
DH36 N/A N/A
We could flash the flow cell manually.
(Power off the analyzer, then aspirate the diluent out)
We could get these videos of how to do maintenance for
analyzer in our FTP account, as following:
Website: http://159.138.130.90/
User name: S-user03
Password: PVns6bPL
The blood clot or
some big particle
block the flow cell
For DF50/
DF52
Hydraulic system
analysis
182
Dymind Global Customer Service Center
Dymind Global Customer Service Center
The sheath
For DF50/ pressure sensor
DF52 damaged and give
the false alarm
For
DF50CRP
Hydraulic system
analysis
183
Dymind Global Customer Service Center
Dymind Global Customer Service Center
The shealth
For pressure sensor
DF50CRP damaged and give
the false alarm
For
DH76/
D7-CRP
Hydraulic system
analysis
184
Dymind Global Customer Service Center
Dymind Global Customer Service Center
For
The sensor
DH76/
damaged and give
D7-CRP
the false alarm
If you have any problem with this file, please contact our engineer ASAP, thank you!
All
185
Dymind Global Customer Service Center
Dymind Global Customer Service Center
0xb2004001-background abnormal
Use Background
When analyzer gives a false alarm –background abnormal, and we can’t remove it.
The background counting result of each parameter upon starup is not within the specified range
Declaration
15/9/2020, V1.0, For reference only.
Reference Standard
Module Possibility Solution
1.The reagents are expired Replace the reagents, prime more times to check if it
or contaminated back to normal
All modules
2.The hydraulics system
B. Perform cleanser soak the chamber manually. (If
dirty
the problem is still existed, we could replace a new
chamber)
186
Dymind Global Customer Service Center
Dymind Global Customer Service Center
All modules
If you have any problem with this file, please contact our engineer ASAP, thank you!
According to modules,
All modules Please purchase a set of / /
service kit
187
Dymind Global Customer Service Center
Dymind Global Customer Service Center
0xB2004110-WBC Clog
Use Background
When “WBC Clog” is prompted in the area of problem on the analyzer interface and the WBC
counting result of the analyzer is abnormal of the repeatability analyzers are not satisfied.
Principle of problem: During the counting some or all of the apertures are clogged, which slows
down the fluid from passing through the apertures, increases the aperture voltage, and triggers
the clogging alarm.
Declaration
25/3/2021 , V1.0, For reference only.
Reference Standard
modules Possible cause Solution
Check if the tubing as shown below is loose or replace the
corresponding tubing, valves (LV1, LV7, LV8, LV9 ) and pump1.
1.The tubing is
damaged or
loose
For
DH36/ A. Go to service > clean > click the WBC chamber button in the
DH31/ analyzer software interface
DH33/
DH36-VET
2. Aperture
clogging
188
Dymind Global Customer Service Center
Dymind Global Customer Service Center
B. Go to the service > Maintenance > click the flush Apertures and
then the zap Aperture button.
For
DH36/
2.Aperture
DH31/
clogging
DH33/
DH36-VET
189
Dymind Global Customer Service Center
Dymind Global Customer Service Center
For
DH36/
3. Failed to mix
DH31/
evenly
DH33/
DH36-VET
1.The tubing is
damaged or
loose
For DH36X
A. Go to service > clean > click the WBC chamber button in the
analyzer software interface
2. Aperture
clogging
190
Dymind Global Customer Service Center
Dymind Global Customer Service Center
2. Aperture
For DH36X
clogging
191
Dymind Global Customer Service Center
Dymind Global Customer Service Center
3. Failed to mix
evenly
4. Failed to
drain
192
Dymind Global Customer Service Center
Dymind Global Customer Service Center
1.The tubing is
Check if the tubing as shown below is loose
damaged or
or replace the corresponding tubing, valves
loose
(LV5/LV23/LV29) and pump 2 .
1.The tubing is
C. For the process
damaged or
of back flashing
loose
193
Dymind Global Customer Service Center
Dymind Global Customer Service Center
A. Go to service > clean > click the WBC chamber button in the
software interface
B. Go to the service > Maintenance > click the flush Apertures and
then the zap Aperture button.
2.Aperture
clogging
194
Dymind Global Customer Service Center
Dymind Global Customer Service Center
3.Failed to mix
evenly
For
DH56/
D5-CRP
DH76/
D7-CRP Observe if the chamber draining is normal. If any problem occurs,
please check the corresponding tubing and components
(LV21/LV25/pump2 /pump3), perform the solution for “Aperture
clogging”.
4. Failed to
drain
195
Dymind Global Customer Service Center
Dymind Global Customer Service Center
2-way valve(L)-
23.99.0001A
(SMC)
3-way valve(S)-
23.99.0004A
(SMC)
2-way valve(S)-
23.99.0003A
(SMC)
All modules
WBC Bath
20.01.0143A
Assembly(with Ruby)
Isolation Chamber
68.02.0023A
(Upper port blocked)
If you have any problem with this file, please contact our engineer ASAP, thank you!
196
Dymind Global Customer Service Center
Dymind Global Customer Service Center
0xB2004210-RBC Clog
Use Background
When “RBC Clog” is prompted in the area of problem on the analyzer interface and the RBC
counting result of the analyzer is abnormal of the repeatability analyzers are not satisfied.
Principle of problem: During the counting some or all of the apertures are clogged, which slows
down the fluid from passing through the apertures, increases the aperture voltage, and triggers
the clogging alarm.
Declaration
16/7/2020, V1.0, For reference only.
Reference Standard
modules Possible Solution
cause
Check if the tubing as shown below is loose or replace the
corresponding tubing, valves (LV1, LV7, LV8, LV9) and pump 1 .
1. The
tubing is
damaged or
loose
For
DH36/
A. Go to service > clean > click the RBC chamber button in the
DH31/
analyzer software interface
DH33/
DH36-VET
2. Aperture
clogging
197
Dymind Global Customer Service Center
Dymind Global Customer Service Center
B. Go to the service > Maintenance > click the flush Apertures and
then the zap Aperture button.
For
DH36/
2. Aperture
DH31/
clogging
DH33/
DH36-VET
198
Dymind Global Customer Service Center
Dymind Global Customer Service Center
For
DH36/
3. Failed to
DH31/
mix evenly
DH33/
DH36-VET
1.The tubing
is damaged
or loose
For DH36X
A. Go to service > clean > click the RBC chamber button in the
analyzer software interface
2. Aperture
clogging
199
Dymind Global Customer Service Center
Dymind Global Customer Service Center
B. Go to the service > Maintenance > click the zap Aperture button
2. Aperture
For DH36X
clogging
200
Dymind Global Customer Service Center
Dymind Global Customer Service Center
3. Failed to
mix evenly
4. Failed to
drain evenly
201
Dymind Global Customer Service Center
Dymind Global Customer Service Center
1.The tubing
is damaged
or loose
A. Go to service > clean > click the RBC chamber button in the
analyzer software interface
For
DF50/
DF51/
DF52/
DF50VET
2.Aperture B. Go to the service > Maintenance > click the flush Apertures and
clogging then the zap Aperture button.
202
Dymind Global Customer Service Center
Dymind Global Customer Service Center
For
DF50/
DF51/
DF52/
DF50VET
3.Failed to
mix evenly
203
Dymind Global Customer Service Center
Dymind Global Customer Service Center
1.The tubing
is damaged
or loose
A. Go to service > clean > click the RBC chamber button in the
analyzer software interface
For
DF50CRP
B. Go to the service > Maintenance > click the zap Aperture button.
2. Aperture
clogging
204
Dymind Global Customer Service Center
Dymind Global Customer Service Center
For
DF50CRP
3. Failed to
mix evenly
205
Dymind Global Customer Service Center
Dymind Global Customer Service Center
For 4. Failed to
DF50CRP drain
For
Check if the tubing as shown below is loose
DH56/ 1.The tubing
or replace the corresponding tubing, valves
D5-CRP is damaged or
(LV6/LV24/LV29) and pump 2 .
DH76/ loose
D7-CRP
B. For the process
of WBC chamber
cleaning
206
Dymind Global Customer Service Center
Dymind Global Customer Service Center
A. Go to service > clean > click the RBC chamber button in the
software interface
B. Go to the service > Maintenance > click the flush Apertures and
then the zap Aperture button.
For
DH56/
2.Aperture
D5-CRP
clogging
DH76/
D7-CRP a. Go to the service > Maintenance > click
the Empty RBC Chamber button
3.Failed to
mix evenly
For
DH56/
D5-CRP Observe if the chamber draining is normal. If any problem occurs,
DH76/ please check the corresponding tubing and components
D7-CRP (LV21/LV25/pump2 /pump3), perform the solution for “Aperture
clogging”.
4. Failed to
drain
208
Dymind Global Customer Service Center
Dymind Global Customer Service Center
2-way valve(L)-
23.99.0001A
(SMC)
3-way valve(S)-
23.99.0004A
(SMC)
RBC Bath
20.01.0144A
Assembly(with Ruby)
If you have any problem with this file, please contact our engineer ASAP, thank you!
209
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Pressure Problem
210
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Use Background
When analyzer gives a false alarm –positive pressure abnormal (low) , and we can’t remove it.
Alarm principle: The host queries the pressure in the positive-pressure chamber. The current
pressure is less than the lower limit (5.5KPa) of the attached
Declaration
25/3/2020, V1.0, For reference only.
Reference Standard
Module Possibility Solution
A. Check if the tube of yellow part is blocked or folded;
B. Check if LV8, LV 9, pump1 are abnormal. ( it’s better to
replace new one to check)
C. Clean the swab by cleanser. ( we could follow the steps of
SOP to operate)
1.Hydraulic analysis
2.Hardware analysis
211
Dymind Global Customer Service Center
Dymind Global Customer Service Center
the new analyzer has not been used for a long time, the
ambient temperature in same places is too high, the
membrane of pump is too dry to work abnormally
(as following figure)
For
DH36/
3.special situation How to solve it: we just back flash the pump1 by diluent .(as
DH31/
for new analyzer following figure)
DH33/
DH36-VET
1.Hydraulic analysis
For
DH36x
A. Check pressure sensor ( replace new one)
B. Check mainboard
2.Hardware analysis
212
Dymind Global Customer Service Center
Dymind Global Customer Service Center
For
DF50/ A. Check pressure
DF51/ sensor ( replace new
DF52/ one)
DF50VET B. Check mainboard
2.Hardware analysis
For
1.Hydraulic analysis
DF50CRP
213
Dymind Global Customer Service Center
Dymind Global Customer Service Center
A. Check pressure
sensor ( replace new
one)
B. Check mainboard
For
2.Hardware analysis
DF50CRP
2.Check if pump3
For (air pump) are
DH56/ abnormal. ( it’s
DH76/ better to replace
D5-CRP/ new one to check)
D7-CRP
3.replace air filter
4.check if the
positive pressure
sensor is normal.
214
Dymind Global Customer Service Center
Dymind Global Customer Service Center
mainboard 22.01.0023A
DH31/DH33/DH36/
3-way valve(S)-(SMC) 23.99.0004A
DH36-VET
DH36X
215
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DF50/DF51/DF52/
3-way valve(S)-(SMC) 23.99.0004A
DF50VET
216
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DF50CRP
DH56/ DH76/
D5-CRP/
D7-CRP Liquid Detecting Board 20.01.0024B
If you have any problem with this file, please contact our engineer ASAP, thank you!
217
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Use Background
When analyzer gives a false alarm - HGB background voltage abnormal, and we can’t remove it.
Alarm principle: The host queries the pressure in the positive-pressure chamber. The current
pressure is over the upper limit (9KPa) of the attached parameter.
Declaration
20/8/2020, V1.0, For reference only.
Reference Standard
Module Possibility Solution
2.Air pump is
damaged
DH31/DH33/
Mainboard with SD card(without
DH36/ 22.01.0023A
battery)
DH36VET
DF50/DF51/
DF52/ Mainboard with SD card(without
22.01.0021C
DF50VET battery)
&DF50CRP
DH31/DH33/
DH36/
DH36VET
Vacuum Pump 23.99.0008A
&DF50/DF51
/DF52/
DF50VET
DH36X
Air Pump 23.99.0012A
&DF50CRP
If you have any problem with this file, please contact our engineer ASAP, thank you!
219
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Use Background
When analyzer gives a false alarm – Failed to build positive pressure, and we can’t remove it.
Alarm principle: The pressure built for positive pressure chamber is 50% higher or 50% lower than
the target pressure(8KPa)
Declaration
25/3/2020, V1.0, For reference only.
Reference Standard
Module Possibility Solution
D. Check if the tube of yellow part is blocked or folded;
E. Check if LV8, LV 9, pump1 are abnormal. ( it’s better to
replace new one to check)
F. Clean the swab by cleanser. ( we could follow the steps of
SOP to operate)
1.Hydraulic analysis
2.Hardware analysis
220
Dymind Global Customer Service Center
Dymind Global Customer Service Center
the new analyzer has not been used for a long time, the
ambient temperature in same places is too high, the
membrane of pump is too dry to work abnormally
(as following figure)
For
DH36/
3.special situation How to solve it: we just back flash the pump1 by diluent .(as
DH31/
for new analyzer following figure)
DH33/
DH36-VET
1.Hydraulic analysis
For
DH36x
A. Check pressure sensor ( replace new one)
B. Check mainboard
2.Hardware analysis
221
Dymind Global Customer Service Center
Dymind Global Customer Service Center
For
DF50/ C. Check pressure
DF51/ sensor ( replace new
DF52/ one)
DF50VET D. Check mainboard
2.Hardware analysis
For
1.Hydraulic analysis
DF50CRP
222
Dymind Global Customer Service Center
Dymind Global Customer Service Center
C. Check pressure
sensor ( replace new
one)
D. Check mainboard
For
2.Hardware analysis
DF50CRP
2.Check if pump3
For (air pump) are
DH56/ abnormal. ( it’s
D5-CRP better to replace
DH76/ new one to check)
D7-CRP
3.replace air filter
4.check if the
positive pressure
sensor is normal.
223
Dymind Global Customer Service Center
Dymind Global Customer Service Center
mainboard 22.01.0023A
DH31/DH33/DH36/
3-way valve(S)-(SMC) 23.99.0004A
DH36-VET
DH36X
224
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DF50/DF51/DF52/
3-way valve(S)-(SMC) 23.99.0004A
DF50VET
225
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DF50CRP
DH56/ DH76/
D5-CRP/
D7-CRP Liquid Detecting Board 20.01.0024B
If you have any problem with this file, please contact our engineer ASAP, thank you!
226
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Use Background
When analyzer gives a false alarm – vacuum pressure abnormal(high), and we can’t remove it.
Alarm principle: The host queries the pressure in the negative pressure chamber. The current
pressure is higher than the lower limit (-25KPa) of the attached parameter.
Declaration
25/3/2020, V1.0, For reference only.
Reference Standard
Module Possibility Solution
A. Check if the tube of yellow part is blocked or folded;
B. Check if LV8, LV 9, pump1 are abnormal. ( it’s better to
replace new one to check)
1.Hydraulic analysis
For
DH36/
A. Check pressure sensor
DH31/
( replace new one)
DH33/
B. Check mainboard
DH36-VET
2.Hardware analysis
227
Dymind Global Customer Service Center
Dymind Global Customer Service Center
For
2.Hardware analysis
DH36X
1.Hydraulic analysis
For
DF50/
DF51/ A. Check pressure sensor
DF52/ ( replace new one)
DF50VET B. Check mainboard
2.Hardware analysis
228
Dymind Global Customer Service Center
Dymind Global Customer Service Center
For
2.Hardware analysis
DF50CRP
3.check if the
positive pressure
sensor is normal.
229
Dymind Global Customer Service Center
Dymind Global Customer Service Center
mainboard 22.01.0023A
DH31/DH33/DH36/
DH36-VET
DH36X
Reagent Detecting Board 22.01.0204A
230
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DF50/DF51/DF52/
DF50VET
DF50CRP
231
Dymind Global Customer Service Center
Dymind Global Customer Service Center
If you have any problem with this file, please contact our engineer ASAP, thank you!
232
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Use Background
When analyzer gives a false alarm – vacuum pressure abnormal(low), and we can’t remove it.
Alarm principle: The host queries the pressure in the negative pressure chamber. The current
pressure is less than the lower limit (-30KPa) of the attached parameter.
Declaration
25/3/2020, V1.0, For reference only.
Reference Standard
Module Possibility Solution
1.negative-pressure
pump driver chip is
in the on-state of
breakdown, and the
For DH36/ negative pressure
DH36X/ pump keeps
DF50/ working;
DF50CRP
2.Pump is damaged
3.The mainboard is
damaged
233
Dymind Global Customer Service Center
Dymind Global Customer Service Center
If you have any problem with this file, please contact our engineer ASAP, thank you!
234
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Use Background
When analyzer gives a false alarm – Failed to build vacuum pressure, and we can’t remove it.
Alarm principle: The pressure built for negative-pressure chamber is 50% higher or 50% lower than
the target pressure (-30KPa)
Declaration
25/3/2020, V1.0, For reference only.
Reference Standard
Module Possibility Solution
C. Check if the tube of yellow part is blocked or folded;
D. Check if LV8, LV 9, pump1 are abnormal. ( it’s better to
replace new one to check)
1.Hydraulic analysis
For
DH36/
C. Check pressure sensor
DH31/
( replace new one)
DH33/
D. Check mainboard
DH36-VET
2.Hardware analysis
235
Dymind Global Customer Service Center
Dymind Global Customer Service Center
For
2.Hardware analysis
DH36X
1.Hydraulic analysis
For
DF50/
DF51/ C. Check pressure sensor
DF52/ ( replace new one)
DF50VET D. Check mainboard
2.Hardware analysis
236
Dymind Global Customer Service Center
Dymind Global Customer Service Center
For
2.Hardware analysis
DF50CRP
3.check if the
positive pressure
sensor is normal.
237
Dymind Global Customer Service Center
Dymind Global Customer Service Center
mainboard 22.01.0023A
DH31/DH33/DH36/
DH36-VET
DH36X
Reagent Detecting Board 22.01.0204A
238
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DF50/DF51/DF52/
DF50VET
DF50CRP
239
Dymind Global Customer Service Center
Dymind Global Customer Service Center
If you have any problem with this file, please contact our engineer ASAP, thank you!
240
Dymind Global Customer Service Center
Dymind Global Customer Service Center
241
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Use Background
When we need to check if the voltage of Reagent detecting board abnormal,
which may lead to the machine alarm ”Lyse running out or air bubbles in inlet tubing” .
When we install the new Reagent Detecting Board and double check the voltage.
Version Description
It’s the first version of this file, which is published on Feb 10th ,2020 by Dymind service
team.
The instruction of how to check and adjust Reagent Detecting Board-V1.0
Reference Standard
1. Type one
2. Type two
243
Dymind Global Customer Service Center
Dymind Global Customer Service Center
3. Type three
244
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Measurement Method
1. Drain the lyse inside the tube or insert an empty tube inside the lot, which
should be the same type.
2. Measure the voltage of the certain lyse according to the test point mentioned
above by multimeter like below.
Place one pen on the test point and the other on the screw or metal cover. Then check if the
voltage meet the requirement.
3. If the voltage doesn’t meet the requirement, remove the glue on the adjust
point if there is, then adjust it by adjusting the corresponding adjust point introduced
above.
Normally, the indicator would be always green if it’s filled with lyse. If the tube is filled with
bubbles, the indicator would be flash. If the tube is empty, the indicator would be off.
245
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Use Background
When analyzer gives a false alarm - Diluent running out, but container have enough.
Mechanism: Diluent float switch detects no reagent.
Declaration
16/7/2020, V1.0, For reference only.
Reference Standard
For all
modules
3.Reconnect the diluent cap assembly connector.
5.If you don’t have the diluent assembly, you can use a
clip just for emergency! Don't forget to replace a new
one otherwise it will cause a really bad problem.
If you have any problem with this file, please contact our engineer ASAP, thank you!
246
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Additional information
247
Dymind Global Customer Service Center
Dymind Global Customer Service Center
248
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Use Background
When analyzer gives a false alarm: Lyse-1 running out or air bubbles in inlet tubing.
Mechanism: The time with empty reagent in the specified time period of time sequence testing
exceeds the specified threshold value by liquid detective board.
Declaration
17/8/2020, V1.0, For reference only.
Reference Standard
DH26
249
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH26↑ DH31/DH33/DH36/DH36VET↓
DH31/DH33/
DH36/DH36V
ET
250
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH31/DH33/DH36/DH36VET↑ DH36X↓
DH36X
251
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH36X↑ DF50/DF51/DF52/DF50VET↓
DF50/DF51/D
F52/DF50VET
2.Too many bubbles detected by liquid
detective board due to valve2 leakage or
syringe broken.
252
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DF50/DF51/DF52/DF50VET↑ DF50CRP↓
DF50CRP
253
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DF50CRP↑ DH73/DH76↓
DH73/DH76
254
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH73/DH76↑ D5-CRP↓
255
Dymind Global Customer Service Center
Dymind Global Customer Service Center
D5-CRP
256
Dymind Global Customer Service Center
Dymind Global Customer Service Center
D5-CRP↑ D7-CRP↓
D7-CRP
257
Dymind Global Customer Service Center
Dymind Global Customer Service Center
If you have any problem with this file, please contact our engineer ASAP, thank you!
258
Dymind Global Customer Service Center
Dymind Global Customer Service Center
259
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH36
260
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH76
Lyse Syringe
20.01.0040A
Assembly(2.5ml)
D5-CRP/D7-CRP
261
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Use Background
When analyzer gives a false alarm: Lyse-2 running out or air bubbles in inlet tubing.
Mechanism: The time with empty reagent in the specified time period of time sequence testing
exceeds the specified threshold value by liquid detective board.
Declaration
17/8/2020, V1.0, For reference only.
Reference Standard
262
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DF50/DF51/DF52/DF50VET↑ DF50CRP↓
263
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DF50CRP↑ DH73/DH76↓
264
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH73/DH76↑ D5-CRP↓
265
Dymind Global Customer Service Center
Dymind Global Customer Service Center
D5-CRP
266
Dymind Global Customer Service Center
Dymind Global Customer Service Center
D5-CRP↑ D7-CRP↓
D7-CRP
267
Dymind Global Customer Service Center
Dymind Global Customer Service Center
268
Dymind Global Customer Service Center
Dymind Global Customer Service Center
If you have any problem with this file, please contact our engineer ASAP, thank you!
269
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DF50
DH76
Lyse Syringe
20.01.0040A
Assembly(2.5ml)
D5-CRP/D7-CRP
270
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Use Background
When analyzer gives a false alarm: Lyse-3 running out or air bubbles in inlet tubing.
Mechanism: The time with empty reagent in the specified time period of time sequence testing
exceeds the specified threshold value by liquid detective board.
Declaration
18/8/2020, V1.0, For reference only.
Reference Standard
DH73/DH76
3.The tube from valve1 to lyse bottle or
the tube from valve1 to syringe broken or
leakage.
271
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH73/DH76↑ D5-CRP↓
272
Dymind Global Customer Service Center
Dymind Global Customer Service Center
D5-CRP↑ D7-CRP↓
273
Dymind Global Customer Service Center
Dymind Global Customer Service Center
274
Dymind Global Customer Service Center
Dymind Global Customer Service Center
If you have any problem with this file, please contact our engineer ASAP, thank you!
275
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH76
Lyse Syringe
20.01.0040A
Assembly(2.5ml)
D5-CRP/D7-CRP
276
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Use Background
When analyzer gives a false alarm: R1 running out or air bubbles in inlet tubing.
Mechanism: The time with empty reagent in the specified time period of time sequence testing
exceeds the specified threshold value by liquid detective board.
Declaration
18/8/2020, V1.0, For reference only.
Reference Standard
277
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH36X↑ DF50CRP↓
DF50CRP
DF50CRP↑ D5-CRP↓
278
Dymind Global Customer Service Center
Dymind Global Customer Service Center
D5-CRP
D5-CRP↑ D7-CRP↓
279
Dymind Global Customer Service Center
Dymind Global Customer Service Center
D7-CRP
280
Dymind Global Customer Service Center
Dymind Global Customer Service Center
If you have any problem with this file, please contact our engineer ASAP, thank you!
281
Dymind Global Customer Service Center
Dymind Global Customer Service Center
282
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Use Background
When analyzer gives a false alarm: R2 running out or air bubbles in inlet tubing.
Mechanism: The time with empty reagent in the specified time period of time sequence testing
exceeds the specified threshold value by liquid detective board.
Declaration
18/8/2020, V1.0, For reference only.
Reference Standard
283
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH36X↑ DF50CRP↓
DF50CRP
284
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DF50CRP↑ D5-CRP↓
D5-CRP
D5-CRP↑ D7-CRP↓
285
Dymind Global Customer Service Center
Dymind Global Customer Service Center
D7-CRP
If you have any problem with this file, please contact our engineer ASAP, thank you!
286
Dymind Global Customer Service Center
Dymind Global Customer Service Center
287
Dymind Global Customer Service Center
Dymind Global Customer Service Center
288
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Declaration
2/4/2021 , V1.0, For reference only.
Reference Standard
modules Issue Description Solutions
1.Flower screen A. Check if the cable of the LCD is loose;
B. If the the problem is still existed, please replace LCD
screen.
2.Ground color
abnormal
Display
abnormal
3.Horizontal or
vertical line
Touch 1.No reaction at A. There is static with protective film, please tear it off;
malfunction all B. Program the software to calibrate the touch screen
289
Dymind Global Customer Service Center
Dymind Global Customer Service Center
1.No reaction at
all
Touch
malfunction
2.Just receive the
feedback of shift
button in screen
interface
290
Dymind Global Customer Service Center
Dymind Global Customer Service Center
1.Status
indicator is red
Fail to enter
system
1.Status
indicator is red
Fail to enter
system
2.Status
indicator is off
291
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Fail to enter
system
292
Dymind Global Customer Service Center
Dymind Global Customer Service Center
5.Stuck with a
short message
If you have any problem with this file, please contact our engineer ASAP, thank you!
According to
the module
mainboard
of your
analyzer
All modules
293
Dymind Global Customer Service Center
Dymind Global Customer Service Center
294
Dymind Global Customer Service Center
Dymind Global Customer Service Center
The process of
upgrading
295
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Declaration
31/3/2021 , V1.0, For reference only.
Reference Standard
modules Steps Content
All of the upgrading files are in our FTP acount (if you don’t know
what’s FTP, please conact to our after-sale service engineer.) ;
296
Dymind Global Customer Service Center
Dymind Global Customer Service Center
1.The routine
of how to get
the upgrading
files
For DH36/
DF50/
DF50CRP/
DF50VET
2.The process
of upgrading
B. Unpack the upgrading file and you wiil get the upgrading
file(XXXXX.tar.bz) and insert the USB disk with the
upgrading fils(XXXXX.tar.bz)
For DH36/
DF50/ 2.The process
DF50CRP/ of upgrading
DF50VET
298
Dymind Global Customer Service Center
Dymind Global Customer Service Center
For DH36/
DF50/ 2.The process
DF50CRP/ of upgrading
DF50VET
All of the upgrading files are in our FTP acount (if you don’t know
what’s FTP, pleace conact to our after-sale service engineer.) ;
For DH76/
1.The routine
DH73
of how to get
DH56/
the upgrading
D5-CRP/
files
D7-CRP
299
Dymind Global Customer Service Center
Dymind Global Customer Service Center
1.The routine
of how to get
the upgrading
files
For DH76/
DH73
DH56/
D5-CRP/
D7-CRP
B. Click the "Open File" button, then select and open the upgrade
2.Upgrading
file:
the main
"upgrade_1101_vxxxx_XXXXXXXXXXXX.tar" in the popup dialog
system
box.
As below picture shows:
Note: In the name of the upgrade file, "vxxxx" indicates the version
of the upgrade file, and "XXXXXXXXXXXX" indicates the release date
of the upgrade file.
300
Dymind Global Customer Service Center
Dymind Global Customer Service Center
C. Click "Upgrade".
The upgrade confirmation dialog box will pop up.
For DH76/
DH73 2.Upgrading
DH56/ the main
D5-CRP/ system
D7-CRP
301
Dymind Global Customer Service Center
Dymind Global Customer Service Center
B. Keep clicking "Next," and then select "Install" in the screen and
follow next steps as shown below.
For DH76/
DH73 3. Upgrading
DH56/ the software
D5-CRP/ interface
D7-CRP
If you have any problem with this file, please contact our engineer ASAP, thank you!
302
Dymind Global Customer Service Center
Dymind Global Customer Service Center
WBC Classification
Problem
303
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Use Background
When analyzer gives a false alarm – No error code
Mechanism: The particle ratio of DIFF count/WBC total count is less than 0.5 or more than 2 in
five-part machine if the DIFF results are blocked by machine.
Declaration
20/3/2021, V1.0, For reference only.
Reference Standard
Module Scatter Gram & Histogram Possibilities & Solutions
DH26/DH31/
DH33/DH36/
DH36VET/DH The impedance testing process is
36X not smooth while testing.
Maintenance needs to be done.
Or the gain of WBC get weak.
1. Do cleanser
soaking by machine.
2. Clean WBC
aperture manually.
3. Do gain
calibration by normal
level control.
304
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DF50/DF51/D
The lyse LYC-1 doesn’t make any
F52/DF50VET
function during testing process.
/DF50CRP
1. Wrong lyse is
connected.
2. LYC-1 is bad,
which need to replace.
3. There’s many
bubbles filled in the lyse
tube or no lyse. Check
the lyse bottle, lyse tube,
valve NO.2, lyse syringe,
titanium tube above
WBC chamber.
305
Dymind Global Customer Service Center
Dymind Global Customer Service Center
306
Dymind Global Customer Service Center
Dymind Global Customer Service Center
307
Dymind Global Customer Service Center
Dymind Global Customer Service Center
308
Dymind Global Customer Service Center
Dymind Global Customer Service Center
anti-static wrist!!!
309
Dymind Global Customer Service Center
Dymind Global Customer Service Center
310
Dymind Global Customer Service Center
Dymind Global Customer Service Center
311
Dymind Global Customer Service Center
Dymind Global Customer Service Center
If you have any problem with this file, please contact our engineer ASAP, thank you!
DF50/DF51/DF52/DF50V
ET,
DF50CRP, Pinch Valve 23.99.0005A
DH56/DH73/DH76,
D5CRP/D7CRP
312
Dymind Global Customer Service Center
Dymind Global Customer Service Center
DH56/DH73/DH76,
LYA-3 lyse Cap Assembly 20.01.0034A
D5CRP/D7CRP
Diluents&Lyses connector /
313
Dymind Global Customer Service Center
Dymind Global Customer Service Center
314
Dymind Global Customer Service Center
Dymind Global Customer Service Center
315
Dymind Global Customer Service Center
Dymind Global Customer Service Center
2. Take out the rear door by screw driver and find these positions.
316
Dymind Global Customer Service Center
Dymind Global Customer Service Center
4. Testing picture
5. Last step, please don’t forget to test the grounding line cable!
More information, please refer to our service manual.
317
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Declaration
2/4/2021 , V1.0, For reference only.
Reference Standard
1. Prepare these items below.
3. Take out the mainboard and driver board to find front board.
318
Dymind Global Customer Service Center
Dymind Global Customer Service Center
319
Dymind Global Customer Service Center
Dymind Global Customer Service Center
4. Testing picture
5. Last step, please don’t forget to test the grounding line cable!
More information, please refer to our service manual.
If you have any problem with this file, please contact our engineer ASAP, thank you!
320
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Counter trigger
abnormal
321
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Use Background
When analyzer gives a false alarm - Counter trigger abnormal
Declaration
9/4/2021, V2.0, For reference only.
Reference Standard
Model Possibility Solution
1. The
UN73/DH76/
counter trigger
D7CRP
is dirty or rusty
C. Remove those 5 screws, the rack fixing plate and the loading
motor cable.
322
Dymind Global Customer Service Center
Dymind Global Customer Service Center
D. Lift the plate and remove the two screws beneath the plate to
release the auto loading feeding assembly.
323
Dymind Global Customer Service Center
Dymind Global Customer Service Center
324
Dymind Global Customer Service Center
Dymind Global Customer Service Center
I.Install back the plate and screws. Be careful about the trigger
position.
J. For a correct assembly of the rack fixing plate, please use a rack
as the picture shows and then adjust the screws.
K. Finally, you can check if the problem was solved using the Auto
Venous Whole Blood mode with 4 empty racks.
UN73/DH76/ 2. The A. Replace the auto-loading feeding assembly.
325
Dymind Global Customer Service Center
Dymind Global Customer Service Center
D7CRP auto-loading
feeding
assembly is
damaged.
If you have any problem with this file, please contact our engineer ASAP, thank you!
326
Dymind Global Customer Service Center
Dymind Global Customer Service Center
327
Dymind Global Customer Service Center
Dymind Global Customer Service Center
Notice
If you don’t have DMS, Pls download it from our FTP server with GENERAL ACCOUNT in DMS folder.
DYMIND General
Website http://159.138.130.90/
User name G-user01
Password n6m2K1qr
Declaration
1) 03/02/2021, V1.0, For reference only.
2) This DMS Software can only used for transferring date from analyzer to PC or Laptop, it can’t do the
bidirectional LIS function, please seek the guideline of engineer in third-part LIS field if you want to
implement this function.
3) If you need LIS protocol, please download from the link above in LIS folder.
Reference Standard
1.Connect the Analyzer to PC through LAN Port.
2.Find the network setting of our analyzer in ethernet and close all firewall and anti-virus software!
328
Dymind Global Customer Service Center
Dymind Global Customer Service Center
3.Set the IP address of the PC as below. The IP address of the PC is set to 10.0.0.102 from Internet
Protocol Version 4
4.Go to Setup menu of the analyzer to do Host Communication and LIS Communication setting.
5.Open the DMS software and set the 5600 in LIS communication and LIS Server.
329
Dymind Global Customer Service Center
Dymind Global Customer Service Center
6.Run cmd to check the connection between the analyzer and the PC.
a)Press combination key [Windows+R] to open the Run window, and input cmd
330
Dymind Global Customer Service Center
Dymind Global Customer Service Center
8.Click Comm to send the report data from analyzer to LIS Client Software.
If you have any problem with this file, please contact our engineer ASAP, thank you!
331