Dymind Global Customer Service Center

Dymind Robot

editor-in-chief: Cary, Roy, Jersey


Editor: Levin, Leiter, Danny
Audit:Joe, Jerry

Dymind Global Customer Service Center


Dymind Global Customer Service Center

Catalogue
Syringe Problem.................................................................................................................................. 1
0x230141C0-Failed to leave initial position (Sample syringe)....................................................2
0x230141D0-Failed to return to initial position (Sample syringe)............................................. 7
0x230341C0-Failed to leave initial position (LYSE syringe).......................................................12
0x230341D0-Failed to return to initial position (LYSE syringe)................................................ 17
0x230441C0-Failed to leave initial position (diluent syringe).................................................. 22
0x230441D0-Failed to return to initial position (diluent syringe)............................................25
0x230541C0-Failed to leave initial position (SH syringe)......................................................... 28
0x230541D0-Failed to return to initial position (SH syringe)................................................... 32
0x23014200-Sample syringe action overtime.......................................................................... 36
0x23034200-LYSE syringe action overtime............................................................................... 40
0x23044200-Diluent syringe action overtime.......................................................................... 41
0x23054200-SH syringe action overtime..................................................................................42
Motor Problem.................................................................................................................................. 43
0x230741C0-Horizontal motor failed to leave initial position................................................. 44
0x230741D0-Horizontal motor failed to return to initial position........................................... 56
0x230741F1-Horizontal motor failed to move to initial position.............................................67
0x230841C0-Vertical motor failed to leave initial position...................................................... 78
0x230841D0-Vertical motor failed to return to initial position................................................90
0x230841F1-Vertical motor failed to move to initial position............................................... 102
0x23074190-Horizontal motor photocoupler abnormal........................................................114
0x23074200-Horizontal motor action overtime.....................................................................122
0x23084190-Vertical motor photocoupler abnormal............................................................ 129
0x23084200-Vertical motor action overtime......................................................................... 137
Aperture Problem............................................................................................................................145
0xB1000505-WBC aperture voltage abnormal.......................................................................146
0xB1000506-RBC aperture voltage abnormal........................................................................ 154
0xB1000507-HGB background voltage abnormal.................................................................. 166
Clog Problem................................................................................................................................... 177
0x63026220-Flow cell clog......................................................................................................178
0x63026221-DIFF probe clog.................................................................................................. 182
0xb2004001-background abnormal....................................................................................... 186
0xB2004110-WBC Clog............................................................................................................188
0xB2004210-RBC Clog............................................................................................................. 197
Pressure Problem............................................................................................................................ 210
0x63006030-Positive pressure abnormal (low)......................................................................211
0x63006040-Positive pressure abnormal (high).....................................................................218
0x63006120-failed to build positive pressure........................................................................ 220
0x63016030- vacuum pressure abnormal (high)....................................................................227
0x63016040- vacuum pressure abnormal (low).....................................................................233
0x63016120- Failed to build vacuum pressure.......................................................................235

Dymind Global Customer Service Center


Dymind Global Customer Service Center

Reagent running out Problem.........................................................................................................241


The instruction of how to check and adjust Reagent Detecting Board..................................242
0x64017040-Diluent running out............................................................................................246
0x64047040-Lyse-1 running out or air bubbles in inlet tubing.............................................. 249
0x64037040-Lyse-2 running out or air bubbles in inlet tubing.............................................. 262
0x64027040-Lyse-3 running out or air bubbles in inlet tubing.............................................. 271
0x64057040-R1 running out or air bubbles in inlet tubing.................................................... 277
0x64067040-R2 running out or air bubbles in inlet tubing.................................................... 283
Black Screen Problem......................................................................................................................288
How to solve the issues related to display..............................................................................289
The process of upgrading................................................................................................................ 295
The Process of Upgrading....................................................................................................... 296
WBC Classification Problem............................................................................................................ 303
WBC Results Abnormal Issue.................................................................................................. 304
How to check power supply............................................................................................................ 315
How to check the power supply in DH36/DF50..................................................................... 316
How to check the power supply in DH56/DH76/D5-CRP/D7CRP...........................................318
Counter trigger abnormal............................................................................................................... 321
0x1128D210-Counter trigger abnormal................................................................................. 322
How to connect DMS.......................................................................................................................327
Connect Analyzer to the Data Management Software(Dymind LIS)...................................... 328

Dymind Global Customer Service Center


Dymind Global Customer Service Center

Syringe Problem

1
Dymind Global Customer Service Center
Dymind Global Customer Service Center

0x230141C0-Failed to leave initial position (Sample syringe)

Use Background
When analyzer gives a false alarm - Failed to leave initial position (Sample syringe).
Mechanism: The motor initial position photocoupler always detects that the photocoupler is blocked.

Declaration
26/3/2021, V2.0, For reference only.

Reference Standard
Model Possibility Solution
If we find the syringe block the photocoupler,
then we can check the status of it in
software, Status > Sensor, if it displays Not
Blocked.
A. Clean the photocoupler.
B. Replace the photocoupler.
1. Photocoupler behind the C. Reconnect or replace the photocoupler
sample syringe assembly has wire.
failure.

DH26,
DH31/DH33/DH36/
DH36VET,
DH36X,
A. Reconnect the motor wire.
DF50/DF51/DF52/D
B. Replace the motor wire if it’s open circuit
F50VET,
tested by multi-meter.
DF50CRP
Put the range to Diode, then test the two
ends of the wire from the same color, the
wire is ok if there’s a beep from multi-meter,
or broken.
2. Poor connection of sample
syringe motor wire.

2
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Do a syringe self-test, replace the motor if it


doesn’t move. Service > Syringe Self-test >
Sample Syringe.

3. The motor that controls sample


syringe has failed.

A. Replace syringe assembly body if the


plungers got rusty.

B. Clean the sliding rail and screw rode, then


apply fresh lubricating oil on screw rod.

4. Sample syringe has too large


resistance.

C. Check the O-ring of syringe, replace it if it


got worn.

DH26,DH31/DH33/DH36/DH36VET/DH36X,DF50/DF51/DF52/DF50VET/DF50CRP↑ DH56/UN73/DH76/D5CRP/D7CRP↓

1. Photocoupler behind the If we find the syringe block the photocoupler,


DH56/UN73/DH76,
sample syringe assembly has then we can check the status of it in
D5CRP/D7CRP
failure. software, Status > Sensor, if it displays Not

3
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Blocked.
A. Clean the photocoupler.
B. Replace the photocoupler.
C. Reconnect or replace the photocoupler
wire.

A. Reconnect the motor wire.


B. Replace the motor wire if it’s open circuit
tested by multi-meter.
Put the range to Diode, then test the two
ends of the wire from the same color, the
wire is ok if there’s a beep from multi-meter,
or broken.
2. Poor connection of sample
syringe motor wire.

Do a self-test for the syringe, replace the


motor if it doesn’t move. Self-test > Syringe >
Sample Syringe.

3. The motor controls lyse syringe


is has failure.

A. Replace the syringe assembly without


4. Sample syringe has too large motor if the plunger got rusty.
resistance. B. Apply fresh lubricating oil after cleaning
the sliding rail and screw rode.

4
Dymind Global Customer Service Center
Dymind Global Customer Service Center

C. Check the O-ring of syringe, replace it if


damaged.

If you have any problem with this file, please contact our engineer ASAP, thank you!

Relative spare part code


Model Part Name P/N Picture

Syringe sensor wire 37.02.0037A

DH31/DH33/DH36/DH36VET,
DF50/DF51/DF52/DF50VET

Motor wire 37.02.0034A

DH26/DH31/DH33/DH36/DH
36VET,
DH36X, O-ring for 100ul
/
DF50CRP, sampling syringe
DH56/UN73/DH76/D5CRP/D7
CRP

DF50/DF51/DF52/DF50VET,
O-ring for 250ul
DH56/UN73/DH76, /
sampling syringe
D5CRP/D7CRP

5
Dymind Global Customer Service Center
Dymind Global Customer Service Center

O-ring for Diluent


/
Syringe 1

O-ring for Lyse Syringe 1 /


DH26/DH31/DH33/DH36/DH
36VET,
DH36X,
DF50/DF51/DF52/DF50VET
O-ring for Diluent
/
Syringe 2

O-ring for Lyse Syringe 2 /

Sensor(Vertical) 23.99.0007C

All models above

Syringe motor 24.02.0003A

6
Dymind Global Customer Service Center
Dymind Global Customer Service Center

0x230141D0-Failed to return to initial position (Sample syringe)

Use Background
When analyzer gives a false alarm - Failed to return to initial position (Sample syringe).
Mechanism: The motor initial position photocoupler always detects that the photocoupler is blocked.

Declaration
26/3/2021, V2.0, For reference only.

Reference Standard
Model Possibility Solution
If we find the syringe block the photocoupler,
then we can check the status of it in
software, Status > Sensor, if it displays Not
Blocked.
A. Clean the photocoupler.
B. Replace the photocoupler.
1. Photocoupler behind the C. Reconnect or replace the photocoupler
sample syringe assembly has wire.
failure.

DH26,
DH31/DH33/DH36/
DH36VET,
DH36X,
A. Reconnect the motor wire.
DF50/DF51/DF52/D
B. Replace the motor wire if it’s open circuit
F50VET,
tested by multi-meter.
DF50CRP
Put the range to Diode, then test the two
ends of the wire from the same color, the
wire is ok if there’s a beep from multi-meter,
or broken.
2. Poor connection of sample
syringe motor wire.

7
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Do a syringe self-test, replace the motor if it


doesn’t move. Service > Syringe Self-test >
Sample Syringe.

3. The motor that controls sample


syringe has failed.

A. Replace syringe assembly body if the


plungers got rusty.

B. Clean the sliding rail and screw rode, then


apply fresh lubricating oil on screw rod.

4. Sample syringe has too large


resistance.

C. Check the O-ring of syringe, replace it if it


got worn.

DH26,DH31/DH33/DH36/DH36VET/DH36X,DF50/DF51/DF52/DF50VET/DF50CRP↑ DH56/UN73/DH76/D5CRP/D7CRP↓

1. Photocoupler behind the If we find the syringe block the photocoupler,


DH56/UN73/DH76,
sample syringe assembly has then we can check the status of it in
D5CRP/D7CRP
failure. software, Status > Sensor, if it displays Not

8
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Blocked.
A. Clean the photocoupler.
B. Replace the photocoupler.
C. Reconnect or replace the photocoupler
wire.

A. Reconnect the motor wire.


B. Replace the motor wire if it’s open circuit
tested by multi-meter.
Put the range to Diode, then test the two
ends of the wire from the same color, the
wire is ok if there’s a beep from multi-meter,
or broken.
2. Poor connection of sample
syringe motor wire.

Do a self-test for the syringe, replace the


motor if it doesn’t move. Self-test > Syringe >
Sample Syringe.

3. The motor controls lyse syringe


is has failure.

A. Replace the syringe assembly without


4. Sample syringe has too large motor if the plunger got rusty.
resistance. B. Apply fresh lubricating oil after cleaning
the sliding rail and screw rode.

9
Dymind Global Customer Service Center
Dymind Global Customer Service Center

C. Check the O-ring of syringe, replace it if


damaged.

If you have any problem with this file, please contact our engineer ASAP, thank you!

Relative spare part code


Model Part Name P/N Picture

Syringe sensor wire 37.02.0037A

DH31/DH33/DH36/DH36VET,
DF50/DF51/DF52/DF50VET

Motor wire 37.02.0034A

DH26/DH31/DH33/DH36/DH
36VET,
DH36X, O-ring for 100ul
/
DF50CRP, sampling syringe
DH56/UN73/DH76/D5CRP/D7
CRP

10
Dymind Global Customer Service Center
Dymind Global Customer Service Center

DF50/DF51/DF52/DF50VET,
O-ring for 250ul
DH56/UN73/DH76, /
sampling syringe
D5CRP/D7CRP

O-ring for Diluent


/
Syringe 1

O-ring for Lyse Syringe 1 /


DH26/DH31/DH33/DH36/DH
36VET,
DH36X,
DF50/DF51/DF52/DF50VET
O-ring for Diluent
/
Syringe 2

O-ring for Lyse Syringe 2 /

Sensor(Vertical) 23.99.0007C

All models above

Syringe motor 24.02.0003A

11
Dymind Global Customer Service Center
Dymind Global Customer Service Center

0x230341C0-Failed to leave initial position (LYSE syringe)

Use Background
When analyzer gives a false alarm - Failed to leave initial position (LYSE syringe).
Mechanism: The motor initial position photocoupler always detects that the photocoupler is
blocked.

Declaration
14/4/2021, V2.0, For reference only.

Reference Standard
Model Possibility Solution
If we find the syringe block the photocoupler,
then we can check the status of it in
software, Status > Sensor, if it displays Not
Blocked.
A. Clean the photocoupler.
B. Replace the photocoupler.
1. Photocoupler behind the lyse C. Reconnect or replace the photocoupler
syringe assembly has failure. wire.

DH36X A. Reconnect the motor wire.


B. Replace the motor wire if it’s open circuit
tested by multi-meter.
Method: Put the range to Diode, then test
the two ends of the wire from the same
color, the wire is ok if there’s a beep from
multi-meter, or broken.
2. Poor connection of lyse syringe
motor wire.

12
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Do a self-test for the syringe, replace the


motor if it doesn’t move. Self-test > Syringe >
Lyse Syringe.
3. The motor controls lyse syringe has
failure.

A. Apply fresh lubricating oil after cleaning


the sliding rail and screw rode.

4. Lyse syringe moves with too large


resistance. B. Check the O-ring of syringe, replace it if
damaged.

C. Replace the syringe assembly without


motor.
DH36X↑ DF50CRP↓

If we find the syringe block the photocoupler,


then we can check the status of it in
software, Status > Sensor, if it displays Not
Blocked.
A. Clean the photocoupler.
B. Replace the photocoupler.
1. Photocoupler behind the lyse
C. Reconnect or replace the photocoupler
syringe assembly has failure.
wire.
DF50CRP

A. Reconnect the motor wire.


2. Poor connection of lyse syringe B. Replace the motor wire if it’s open circuit
motor wire. tested by multi-meter.
Method: Put the range to Diode, then test
13
Dymind Global Customer Service Center
Dymind Global Customer Service Center

the two ends of the wire from the same


color, the wire is ok if there’s a beep from
multi-meter, or broken.

Do a self-test for the syringe, replace the


motor if it doesn’t move. Self-test > Syringe >
Lyse Syringe.
3. The motor controls lyse syringe has
failure.

A. Apply fresh lubricating oil after cleaning


the sliding rail and screw rode.

4. Lyse syringe moves with too large


resistance. B. Check the O-ring of syringe, replace it if
damaged.

C. Replace the syringe assembly without


motor.
DF50CRP ↑ DH56,UN73/DH76,D5CRP/D7CRP↓
If we find the syringe block the photocoupler,
then we can check the status of it in
DH56,
1. Photocoupler behind the lyse software, Status > Sensor, if it displays Not
UN73/DH76,
syringe assembly has failure. Blocked.
D5CRP/D7CRP
A. Clean the photocoupler.
B. Replace the photocoupler.

14
Dymind Global Customer Service Center
Dymind Global Customer Service Center

C. Reconnect or replace the photocoupler


wire.

A. Reconnect the motor wire.


B. Replace the motor wire if it’s open circuit
tested by multi-meter.
Method: Put the range to Diode, then test
the two ends of the wire from the same
color, the wire is ok if there’s a beep from
multi-meter, or broken.
2. Poor connection of lyse syringe
motor wire.

Do a self-test for the syringe, replace the


motor if it doesn’t move. Self-test > Syringe >
Lyse Syringe.
3. The motor controls lyse syringe has
failure.

A. Apply fresh lubricating oil after cleaning


the sliding rail and screw rode.

4. Lyse syringe moves with too large


resistance.

B. Check the O-ring of syringe, replace it if


damaged.

15
Dymind Global Customer Service Center
Dymind Global Customer Service Center

C. Replace the syringe assembly without


motor.
If you have any problem with this file, please contact our engineer ASAP, thank you!

Relative spare part code


Model Part Name P/N Picture

Sensor(Vertical) 23.99.0007C

All models above Syringe motor 24.02.0003A

O-ring for Lyse Syringe /

16
Dymind Global Customer Service Center
Dymind Global Customer Service Center

0x230341D0-Failed to return to initial position (LYSE syringe)

Use Background
When analyzer gives a false alarm - Failed to return to initial position (LYSE syringe).
Mechanism: The motor initial position photocoupler always detects that the photocoupler is blocked.

Declaration
14/4/2021, V2.0, For reference only.

Reference Standard
Model Possibility Solution
If we find the syringe block the photocoupler,
then we can check the status of it in
software, Status > Sensor, if it displays Not
Blocked.
A. Clean the photocoupler.
B. Replace the photocoupler.
1. Photocoupler behind the lyse C. Reconnect or replace the photocoupler
syringe assembly has failure. wire.

DH36X A. Reconnect the motor wire.


B. Replace the motor wire if it’s open circuit
tested by multi-meter.
Method: Put the range to Diode, then test
the two ends of the wire from the same
color, the wire is ok if there’s a beep from
multi-meter, or broken.
2. Poor connection of lyse syringe
motor wire.

17
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Do a self-test for the syringe, replace the


motor if it doesn’t move. Self-test > Syringe >
Lyse Syringe.
3. The motor controls lyse syringe has
failure.

A. Apply fresh lubricating oil after cleaning


the sliding rail and screw rode.

4. Lyse syringe moves with too large


resistance. B. Check the O-ring of syringe, replace it if
damaged.

C. Replace the syringe assembly without


motor.
DH36X↑ DF50CRP↓

If we find the syringe block the photocoupler,


then we can check the status of it in
software, Status > Sensor, if it displays Not
Blocked.
A. Clean the photocoupler.
B. Replace the photocoupler.
1. Photocoupler behind the lyse
C. Reconnect or replace the photocoupler
syringe assembly has failure.
wire.
DF50CRP

A. Reconnect the motor wire.


2. Poor connection of lyse syringe B. Replace the motor wire if it’s open circuit
motor wire. tested by multi-meter.
Method: Put the range to Diode, then test
18
Dymind Global Customer Service Center
Dymind Global Customer Service Center

the two ends of the wire from the same


color, the wire is ok if there’s a beep from
multi-meter, or broken.

Do a self-test for the syringe, replace the


motor if it doesn’t move. Self-test > Syringe >
Lyse Syringe.
3. The motor controls lyse syringe has
failure.

A. Apply fresh lubricating oil after cleaning


the sliding rail and screw rode.

4. Lyse syringe moves with too large


resistance. B. Check the O-ring of syringe, replace it if
damaged.

C. Replace the syringe assembly without


motor.
DF50CRP ↑ DH56,UN73/DH76,D5CRP/D7CRP↓
If we find the syringe block the photocoupler,
then we can check the status of it in
DH56,
1. Photocoupler behind the lyse software, Status > Sensor, if it displays Not
UN73/DH76,
syringe assembly has failure. Blocked.
D5CRP/D7CRP
A. Clean the photocoupler.
B. Replace the photocoupler.

19
Dymind Global Customer Service Center
Dymind Global Customer Service Center

C. Reconnect or replace the photocoupler


wire.

A. Reconnect the motor wire.


B. Replace the motor wire if it’s open circuit
tested by multi-meter.
Method: Put the range to Diode, then test
the two ends of the wire from the same
color, the wire is ok if there’s a beep from
multi-meter, or broken.
2. Poor connection of lyse syringe
motor wire.

Do a self-test for the syringe, replace the


motor if it doesn’t move. Self-test > Syringe >
Lyse Syringe.
3. The motor controls lyse syringe has
failure.

A. Apply fresh lubricating oil after cleaning


the sliding rail and screw rode.

4. Lyse syringe moves with too large


resistance.

B. Check the O-ring of syringe, replace it if


damaged.

20
Dymind Global Customer Service Center
Dymind Global Customer Service Center

C. Replace the syringe assembly without


motor.
If you have any problem with this file, please contact our engineer ASAP, thank you!

Relative spare part code


Model Part Name P/N Picture

Sensor(Vertical) 23.99.0007C

All models above Syringe motor 24.02.0003A

O-ring for Lyse Syringe /

21
Dymind Global Customer Service Center
Dymind Global Customer Service Center

0x230441C0-Failed to leave initial position (diluent syringe)

Use Background
When analyzer gives a false alarm - Failed to leave initial position (diluent syringe).
Mechanism: The motor initial position photocoupler always detects that the photocoupler is blocked.

Declaration
9/4/2021, V2.0, For reference only.

Reference Standard
Model Possibility Solution
If we find the syringe block the photocoupler,
then we can check the status of it in
software, Status > Sensor, if it displays Not
Blocked.
A. Clean the photocoupler.
B. Replace the photocoupler.
C. Reconnect or replace the photocoupler
wire.
DH36X

1. Photocoupler behind the


diluent syringe assembly has
failure.
DH36X,
DH56/UN73/DH76,
D5CRP/D7CRP DH56/UN73/DH76/D5CRP/D7CRP

A. Reconnect the motor wire.


B. Replace the motor wire if it’s open circuit
2. Poor connection of diluent
tested by multi-meter.
syringe motor wire.
Method: Put the range to Diode, then test
the two ends of the wire from the same

22
Dymind Global Customer Service Center
Dymind Global Customer Service Center

color, the wire is ok if there’s a beep from


multi-meter, or broken.

Do a self-test for the syringe, replace the


motor if it doesn’t move. Self-test > Syringe >
Diluent Syringe.
DH36X

3. The motor controls diluent


syringe has failure.
DH56/UN73/DH76/D5CRP/D7CRP

A. Apply fresh lubricating oil after cleaning


the sliding rail and screw rode.

4. Diluent syringe moves with too


large resistance.

B. Check the O-ring of syringe, replace it if


damaged.

23
Dymind Global Customer Service Center
Dymind Global Customer Service Center

C. Replace the syringe assembly without


motor if the plunger got rusty.

If you have any problem with this file, please contact our engineer ASAP, thank you!

Relative spare part code


Model Part Name P/N Picture

O-ring for Diluent Syringe /

All models above Sensor(Vertical) 23.99.0007C

Syringe motor 24.02.0003A

24
Dymind Global Customer Service Center
Dymind Global Customer Service Center

0x230441D0-Failed to return to initial position (diluent syringe)

Use Background
When analyzer gives a false alarm - Failed to return to initial position (diluent syringe).
Mechanism: The motor initial position photocoupler always detects that the photocoupler is blocked.

Declaration
9/4/2021, V2.0, For reference only.

Reference Standard
Model Possibility Solution
If we find the syringe block the photocoupler,
then we can check the status of it in
software, Status > Sensor, if it displays Not
Blocked.
A. Clean the photocoupler.
B. Replace the photocoupler.
C. Reconnect or replace the photocoupler
wire.
DH36X

1. Photocoupler behind the


diluent syringe assembly has
failure.
DH36X,
DH56/UN73/DH76,
D5CRP/D7CRP DH56/UN73/DH76/D5CRP/D7CRP

A. Reconnect the motor wire.


B. Replace the motor wire if it’s open circuit
2. Poor connection of diluent
tested by multi-meter.
syringe motor wire.
Method: Put the range to Diode, then test
the two ends of the wire from the same

25
Dymind Global Customer Service Center
Dymind Global Customer Service Center

color, the wire is ok if there’s a beep from


multi-meter, or broken.

Do a self-test for the syringe, replace the


motor if it doesn’t move. Self-test > Syringe >
Diluent Syringe.
DH36X

3. The motor controls diluent


syringe has failure.
DH56/UN73/DH76/D5CRP/D7CRP

A. Apply fresh lubricating oil after cleaning


the sliding rail and screw rode.

4. Diluent syringe moves with too


large resistance. B. Check the O-ring of syringe, replace it if
damaged.

26
Dymind Global Customer Service Center
Dymind Global Customer Service Center

C. Replace the syringe assembly without


motor if the plunger got rusty.

If you have any problem with this file, please contact our engineer ASAP, thank you!

Relative spare part code


Model Part Name P/N Picture

O-ring for Diluent Syringe /

All models above Sensor(Vertical) 23.99.0007C

Syringe motor 24.02.0003A

27
Dymind Global Customer Service Center
Dymind Global Customer Service Center

0x230541C0-Failed to leave initial position (SH syringe)

Use Background
When analyzer gives a false alarm - Failed to leave initial position (SH syringe).
Mechanism: After initialization command is executed, the initial position photocoupler is unblocked.

Declaration
7/4/2021, V2.0, For reference only.

Reference Standard
Model Possibility Solution

If we find the syringe block the photocoupler,


then we can check the status of it in
software, Status > Sensor, if it displays Not
Blocked.
A. Clean the photocoupler.
B. Replace the photocoupler.
C. Reconnect or replace the photocoupler
1. Photocoupler behind the sheath
wire.
fluid syringe assembly has failure.

DF50CRP

A. Reconnect the motor wire.


B. Replace the motor wire if it’s open circuit
tested by multi-meter.
2. Poor connection of sheath fluid
Method: Put the range to Diode, then test
syringe motor wire.
the two ends of the wire from the same
color, the wire is ok if there’s a beep from
multi-meter, or broken.

28
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Do a self-test for the syringe, replace the


motor if it doesn’t move. Self-test > Syringe
and Sampling Mechanism > Sheath Fluid
Syringe.
3. The motor controls sheath fluid
syringe has failure.

A. Apply fresh lubricating oil after cleaning


the sliding rail and screw rode.

4. Sheath fluid syringe moves with B. Check the O-ring of syringe, replace it if it
too large resistance. badly worn.

C. Replace the syringe assembly without


motor if the plunger got rusty.
DF50CRP↑ DH56/UN73/DH76/D5CRP/D7CRP ↓

If we find the syringe block the photocoupler,


DH56,
1. Photocoupler behind the sheath then we can check the status of it in
UN73/DH76,
fluid syringe assembly has failure. software, Status > Sensor, if it displays Not
D5CRP/D7CRP
Blocked.

29
Dymind Global Customer Service Center
Dymind Global Customer Service Center

A. Clean the photocoupler.


B. Replace the photocoupler.
C. Reconnect or replace the photocoupler
wire.

A. Reconnect the motor wire.


B. Replace the motor wire if it’s open circuit
tested by multi-meter.
Method: Put the range to Diode, then test
the two ends of the wire from the same
color, the wire is ok if there’s a beep from
2. Poor connection of sheath fluid multi-meter, or broken.
syringe motor wire.

Do a self-test for the syringe, replace the


motor if it doesn’t move. Self-test > Syringe >
Sheath Fluid Syringe.

3. The motor controls sheath fluid


syringe has failure.

A. Apply fresh lubricating oil after cleaning


the sliding rail and screw rode.

4. Sheath fluid syringe moves with


too large resistance.

30
Dymind Global Customer Service Center
Dymind Global Customer Service Center

B. Check the O-ring of syringe, replace it if it


badly worn.

C. Replace the syringe assembly without


motor if the plunger got rusty.
If you have any problem with this file, please contact our engineer ASAP, thank you!

Relative spare part code


Model Part Name P/N Picture

O-ring for Diluent Syringe /

All models above Sensor(Vertical) 23.99.0007C

Syringe motor 24.02.0003A

31
Dymind Global Customer Service Center
Dymind Global Customer Service Center

0x230541D0-Failed to return to initial position (SH syringe)

Use Background
When analyzer gives a false alarm - Failed to return to initial position (SH syringe).
Mechanism: After initialization command is executed, the initial position photocoupler is unblocked.

Declaration
7/4/2021, V2.0, For reference only.

Reference Standard
Model Possibility Solution

If we find the syringe block the photocoupler,


then we can check the status of it in
software, Status > Sensor, if it displays Not
Blocked.
A. Clean the photocoupler.
B. Replace the photocoupler.
C. Reconnect or replace the photocoupler
1. Photocoupler behind the sheath
wire.
fluid syringe assembly has failure.

DF50CRP

A. Reconnect the motor wire.


B. Replace the motor wire if it’s open circuit
tested by multi-meter.
2. Poor connection of sheath fluid
Method: Put the range to Diode, then test
syringe motor wire.
the two ends of the wire from the same
color, the wire is ok if there’s a beep from
multi-meter, or broken.

32
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Do a self-test for the syringe, replace the


motor if it doesn’t move. Self-test > Syringe
and Sampling Mechanism > Sheath Fluid
Syringe.
3. The motor controls sheath fluid
syringe has failure.

A. Apply fresh lubricating oil after cleaning


the sliding rail and screw rode.

4. Sheath fluid syringe moves with B. Check the O-ring of syringe, replace it if it
too large resistance. badly worn.

C. Replace the syringe assembly without


motor if the plunger got rusty.
DF50CRP↑ DH56/UN73/DH76/D5CRP/D7CRP ↓

If we find the syringe block the photocoupler,


DH56,
1. Photocoupler behind the sheath then we can check the status of it in
UN73/DH76,
fluid syringe assembly has failure. software, Status > Sensor, if it displays Not
D5CRP/D7CRP
Blocked.

33
Dymind Global Customer Service Center
Dymind Global Customer Service Center

A. Clean the photocoupler.


B. Replace the photocoupler.
C. Reconnect or replace the photocoupler
wire.

A. Reconnect the motor wire.


B. Replace the motor wire if it’s open circuit
tested by multi-meter.
Method: Put the range to Diode, then test
the two ends of the wire from the same
color, the wire is ok if there’s a beep from
2. Poor connection of sheath fluid multi-meter, or broken.
syringe motor wire.

Do a self-test for the syringe, replace the


motor if it doesn’t move. Self-test > Syringe >
Sheath Fluid Syringe.

3. The motor controls sheath fluid


syringe has failure.

A. Apply fresh lubricating oil after cleaning


the sliding rail and screw rode.

4. Sheath fluid syringe moves with


too large resistance.

34
Dymind Global Customer Service Center
Dymind Global Customer Service Center

B. Check the O-ring of syringe, replace it if it


badly worn.

C. Replace the syringe assembly without


motor if the plunger got rusty.
If you have any problem with this file, please contact our engineer ASAP, thank you!

Relative spare part code


Model Part Name P/N Picture

O-ring for Diluent Syringe /

All models above Sensor(Vertical) 23.99.0007C

Syringe motor 24.02.0003A

35
Dymind Global Customer Service Center
Dymind Global Customer Service Center

0x23014200-Sample syringe action overtime

Use Background
When analyzer gives a false alarm - Sample syringe action overtime.

Mechanism: Failed to complete the action within the time specified by fluidics.

Declaration
9/4/2021, V2.0, For reference only.

Reference Standard
Model Possibility Solution
A. Apply fresh lubricating oil after cleaning
the sliding rail and screw rode.

DH21,
DH31/DH33/DH
1. Sample syringe moves with too B. Check the O-ring of syringe, replace it if it
36/DH36VET,
large resistance. got worn.
DF50/DF51/DF5
2/DF50VET

C. Replace the syringe assembly or syringe


assembly without motor.
DH21,DH31/DH33/DH36/DH36VET,DF50/DF51/DF52/DF50VET↑ DH36X↓

1. Sample syringe moves with too A. Apply fresh lubricating oil after cleaning
DH36X
large resistance. the sliding rail and screw rode.

36
Dymind Global Customer Service Center
Dymind Global Customer Service Center

B. Check the O-ring of syringe, replace it if it


got worn.

C. Replace the syringe assembly or syringe


assembly without motor.
DH36X↑ DF50CRP↓

A. Apply fresh lubricating oil after cleaning


the sliding rail and screw rode.

1. Sample syringe moves with too


DF50CRP
large resistance.

B. Check the O-ring of syringe, replace it if it


got worn.

C. Replace the syringe assembly or syringe


assembly without motor.
DF50CRP↑ DH56/UN73/DH76,D5CRP/D7CRP↓

DH56/UN73/DH 1. Sample syringe moves with too A. Apply fresh lubricating oil after cleaning

37
Dymind Global Customer Service Center
Dymind Global Customer Service Center

76, large resistance. the sliding rail and screw rode.


D5CRP/D7CRP

B. Check the O-ring of syringe, replace it if


damaged.

C. Replace the syringe assembly or syringe


assembly without motor.

If you have any problem with this file, please contact our engineer ASAP, thank you!
Relative spare part code
Model Part Name P/N Picture
DH26/DH31/DH33/DH36/DH36VET,
DH36X,
O-ring for 100ul
DF50CRP, /
sampling syringe
DH56/UN73/DH76,
D5CRP/D7CRP

DF50/DF51/DF52/DF50VET,
O-ring for 250ul
DH56/UN73/DH76, /
sampling syringe
D5CRP/D7CRP

DH26/DH31/DH33/DH36/DH36VET, O-ring for Diluent


/
DF50/DF51/DF52/DF50VET Syringe 1

38
Dymind Global Customer Service Center
Dymind Global Customer Service Center

DH26/DH31/DH33/DH36/DH36VET, O-ring for Diluent


/
DF50/DF51/DF52/DF50VET Syringe 2

DH26/DH31/DH33/DH36/DH36VET,
O-ring for Lyse
DH36X, /
Syringe 1
DF50/DF51/DF52/DF50VET

DH26/DH31/DH33/DH36/DH36VET,
O-ring for Lyse
DH36X, /
Syringe 2
DF50/DF51/DF52/DF50VET

39
Dymind Global Customer Service Center
Dymind Global Customer Service Center

0x23034200-LYSE syringe action overtime

Use Background
When analyzer gives a false alarm - LYSE syringe action overtime.

Mechanism: Failed to complete the action within the time specified by fluidics.

Declaration
10/4/2021, V2.0, For reference only.

Reference Standard
Model Possibility Solution
A. Apply fresh lubricating oil after cleaning
the sliding rail and screw rode.

DH36X,
DF50CRP,
DH56, 1. Lyse syringe moves with too large
UN73/DH76, resistance.
D5CRP/D7CRP
B. Check the O-ring of syringe, replace it if
damaged.

C. Replace the syringe assembly without


motor.

If you have any problem with this file, please contact our engineer ASAP, thank you!

Relative spare part code


Model Part Name P/N Picture

All models above O-ring for Lyse Syringe /

40
Dymind Global Customer Service Center
Dymind Global Customer Service Center

0x23044200-Diluent syringe action overtime

Use Background
When analyzer gives a false alarm - Diluent syringe action overtime.

Mechanism: Failed to complete the action within the time specified by fluidics.
Declaration
10/4/2021, V2.0, For reference only.

Reference Standard
Model Possibility Solution
A. Apply fresh lubricating oil after cleaning
the sliding rail and screw rode.

DH36X,
DH56/UN73/DH 1. Diluent syringe moves with too
76, large resistance.
D5CRP/D7CRP
B. Check the O-ring of syringe, replace it if
damaged.

C. Replace the syringe assembly without


motor if the plunger got rusty.

Relative spare part code


Model Part Name P/N Picture

All models above O-ring for Diluent Syringe /

41
Dymind Global Customer Service Center
Dymind Global Customer Service Center

0x23054200-SH syringe action overtime

Use Background
When analyzer gives a false alarm - SH syringe action overtime.

Mechanism: Failed to complete the action within the time specified by fluidics.

Declaration
10/4/2021, V2.0, For reference only.

Reference Standard
Model Possibility Solution
A. Apply fresh lubricating oil after cleaning
the sliding rail and screw rode.

DF50CRP,
DH56/UN73/DH 1. Sheath fluid syringe moves with
76, too large resistance.
D5CRP/D7CRP
B. Check the O-ring of syringe, replace it if it
badly worn.

C. Replace the syringe assembly without


motor if the plunger got rusty.

Relative spare part code


Model Part Name P/N Picture

All models above O-ring for Diluent Syringe /

42
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Motor Problem

43
Dymind Global Customer Service Center
Dymind Global Customer Service Center

0x230741C0-Horizontal motor failed to leave initial position

Use Background
When analyzer gives a false alarm: Horizontal motor failed to leave initial position.
Mechanism: The horizontal motor initial position optocoupler always detects that the optocoupler is
blocked.

Declaration
08/9/2020, V1.0, For reference only.

Reference Standard
Model Possibility Solution

Apply some lubricant and move it back


and forth a few times.

1.Horizontal motor has too large


Resistance so it fails to leave the initial
position in the optocoupler area.

DH26
Please press Service<Sensor to test if this
sensor working or not by a paper card, if
there is no response, please
1. Change the sensor
2. Change the sensor cable
2.Horizontal motor initial position
optocoupler malfunctions.

44
Dymind Global Customer Service Center
Dymind Global Customer Service Center

1. Reconnect the motor cable.


2. Change the motor cable.
3. Change the motor

3.Horizontal motor cable has poor


contact/Horizontal motor is damaged.

DH26↑ DH31/DH33/DH36/DH36VET↓

Apply some lubricant and move it back


and forth a few times.

1.Horizontal motor has too large


Resistance so it fails to leave the initial
position in the optocoupler area.

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
there is no response, please
1.Change the sensor
DH31/DH33/D
2.Change the sensor cable
H36/DH36VET 2.Horizontal motor initial position
optocoupler malfunctions.

45
Dymind Global Customer Service Center
Dymind Global Customer Service Center

1.Reconnect the motor cable.


2.Change the motor cable.
3.Change the motor.

3.Horizontal motor cable has poor


contact/Horizontal motor is damaged.

DH31/DH33/DH36/DH36VET↑ DH36X↓
Apply some lubricant and move it back
and forth a few times.

1.Horizontal motor has too large


Resistance so it fails to leave the initial
position in the optocoupler area.

Please press Service<Sensor to test if this


DH36X sensor working or not by a paper card, if
there is no response, please
1. Change the sensor
2. Change the sensor cable

2.Horizontal motor initial position


optocoupler malfunctions.

46
Dymind Global Customer Service Center
Dymind Global Customer Service Center

1. Reconnect the motor cable.


2. Change the motor cable.
3. Change the motor.

3.Horizontal motor cable has poor


contact/Horizontal motor is damaged.

DH36X↑ DF50/DF51/DF52/DF50VET↓
Apply some lubricant and move it back
and forth a few times.

1.Horizontal motor has too large


Resistance so it fails to leave the initial
position in the optocoupler area.

Please press Service<Sensor to test if this


DF50/DF51/DF sensor working or not by a paper card, if
52/DF50VET there is no response, please
1.Change the sensor
2.Change the sensor cable
2.Horizontal motor initial position
optocoupler malfunctions.

47
Dymind Global Customer Service Center
Dymind Global Customer Service Center

1.Reconnect the motor cable.


2.Change the motor cable.
3.Change the motor.

3.Horizontal motor cable has poor


contact/Horizontal motor is damaged.

DF50/DF51/DF52/DF50VET↑ DF50CRP↓
Apply some lubricant and move it back
and forth a few times.

1.Horizontal motor has too large


Resistance so it fails to leave the initial
position in the optocoupler area.

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
DF50CRP
there is no response, please
1.Change the sensor
2.Change the sensor cable

2.Horizontal motor initial position


optocoupler malfunctions.

48
Dymind Global Customer Service Center
Dymind Global Customer Service Center

1. Reconnect the motor cable.


2. Change the motor cable.
3. Change the motor.

3.Horizontal motor cable has poor


contact/Horizontal motor is damaged.

DF50CRP↑ DH73/DH76↓
Apply some lubricant and move it back
and forth a few times.

1.Horizontal motor has too large


Resistance so it fails to leave the initial
position in the optocoupler area.

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
DH73/DH76
there is no response, please
1.Change the sensor
2.Change the sensor cable

2.Horizontal motor initial position


optocoupler malfunctions.

49
Dymind Global Customer Service Center
Dymind Global Customer Service Center

1.Reconnect the motor cable.


2.Change the motor cable.
3.Change the motor.

3.Horizontal motor cable has poor


contact/Horizontal motor is damaged.

DH73/DH76↑ D5-CRP↓
Apply some lubricant and move it back
and forth a few times.

1.Horizontal motor has too large


Resistance so it fails to leave the initial
position in the optocoupler area.

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
there is no response, please
D5-CRP
1.Change the sensor
2.Change the sensor cable

2.Horizontal motor initial position


optocoupler malfunctions.

50
Dymind Global Customer Service Center
Dymind Global Customer Service Center

1.Reconnect the motor cable.


2.Change the motor cable.
3.Change the motor.

3.Horizontal motor cable has poor


contact/Horizontal motor is damaged.

D5-CRP↑ D7-CRP↓
Apply some lubricant and move it back
and forth a few times.

1.Horizontal motor has too large


Resistance so it fails to leave the initial
position in the optocoupler area.

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
there is no response, please
D7-CRP
1.Change the sensor
2.Change the sensor cable

2.Horizontal motor initial position


optocoupler malfunctions.

51
Dymind Global Customer Service Center
Dymind Global Customer Service Center

1.Reconnect the motor cable.


2.Change the motor cable.
3.Change the motor.
3.Horizontal motor cable has poor
contact/Horizontal motor is damaged.

If you have any problem with this file, please contact our engineer ASAP, thank you!

Relative spare part code


Part Name P/N Picture

Horizontal Sampling
68.02.0026A
Motor

DH26 Sensor(Vertical) 23.99.0007C

Sampling Motor
37.02.0145A
Connection Cable

Horizontal Sampling
DH31/DH33/DH36/DH36VET 68.02.0026A
Motor

52
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Sensor(Vertical) 23.99.0007C

Sampling optocoupler
37.02.0036B
connection cable

Horizontal & Vertical Motor


37.02.0035A
Connection Cable

Horizontal Sampling
68.02.0026A
Motor

Sensor(Vertical) 23.99.0007C

DF50/DF51/DF52/DF50VET

Sampling optocoupler
37.02.0036B
connection cable

Horizontal & Vertical Motor


37.02.0035A
Connection Cable

Horizontal Sampling
DF50CRP 68.02.0026A
Motor

53
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Sensor(Vertical) 23.99.0007C

Horizontal &vertical
37.02.0035A
motor connection cable

Sampling optocoupler
37.02.0036B
connection cable

Horizontal Sampling
68.02.0026A
Motor

Sensor(Vertical) 23.99.0007C

Sampling motor
37.01.0042A
DH73/DH76 connection cable

Sampling Optocoupler
37.02.0142A
interchange cable

Sampling Optocoupler
37.02.0051A
connection cable1

54
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Sampling Optocoupler
37.02.0052A
connection cable2

Horizontal Sampling
68.02.0026A
Motor

DH36X/D5-CRP/D7-CRP

Sensor(Vertical) 23.99.0007C

55
Dymind Global Customer Service Center
Dymind Global Customer Service Center

0x230741D0-Horizontal motor failed to return to initial position

Use Background
When analyzer gives a false alarm: Horizontal motor failed to return to initial position.
Mechanism: The horizontal motor initial position optocoupler is always in the unblocked status.

Declaration
08/9/2020, V1.0, For reference only.

Reference Standard
Model Possibility Solution

Apply some lubricant and move it back


and forth a few times.

1.Horizontal motor has too large


Resistance so it fails to return the initial
position in the optocoupler area.

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
DH26 there is no response, please
3. Change the sensor
4. Change the sensor cable

2.Horizontal motor initial position


optocoupler malfunctions.

56
Dymind Global Customer Service Center
Dymind Global Customer Service Center

4. Reconnect the motor cable.


5. Change the motor cable.
6. Change the motor

3.Horizontal motor cable has poor


contact/Horizontal motor is damaged.

DH26↑ DH31/DH33/DH36/DH36VET↓

Apply some lubricant and move it back


and forth a few times.

1.Horizontal motor has too large


Resistance so it fails to return the initial
position in the optocoupler area.

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
there is no response, please
1.Change the sensor
DH31/DH33/D
2.Change the sensor cable
H36/DH36VET

2.Horizontal motor initial position


optocoupler malfunctions.

57
Dymind Global Customer Service Center
Dymind Global Customer Service Center

1.Reconnect the motor cable.


2.Change the motor cable.
3.Change the motor.

3.Horizontal motor cable has poor


contact/Horizontal motor is damaged.

DH31/DH33/DH36/DH36VET↑ DH36X↓
Apply some lubricant and move it back
and forth a few times.

1.Horizontal motor has too large


Resistance so it fails to return the initial
position in the optocoupler area.

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
there is no response, please
3. Change the sensor
4. Change the sensor cable
DH36X

2.Horizontal motor initial position


optocoupler malfunctions.

58
Dymind Global Customer Service Center
Dymind Global Customer Service Center

4. Reconnect the motor cable.


5. Change the motor cable.
6. Change the motor.

3.Horizontal motor cable has poor


contact/Horizontal motor is damaged.

DH36X↑ DF50/DF51/DF52/DF50VET↓
Apply some lubricant and move it back
and forth a few times.

1.Horizontal motor has too large


Resistance so it fails to return the initial
position in the optocoupler area.

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
there is no response, please
1.Change the sensor
DF50/DF51/DF 2.Change the sensor cable
52/DF50VET

2.Horizontal motor initial position


optocoupler malfunctions.

59
Dymind Global Customer Service Center
Dymind Global Customer Service Center

1.Reconnect the motor cable.


2.Change the motor cable.
3.Change the motor.

3.Horizontal motor cable has poor


contact/Horizontal motor is damaged.

DF50/DF51/DF52/DF50VET↑ DF50CRP↓
Apply some lubricant and move it back
and forth a few times.

1.Horizontal motor has too large


Resistance so it fails to return the initial
position in the optocoupler area.

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
there is no response, please
1.Change the sensor
2.Change the sensor cable
DF50CRP

2.Horizontal motor initial position


optocoupler malfunctions.

60
Dymind Global Customer Service Center
Dymind Global Customer Service Center

4. Reconnect the motor cable.


5. Change the motor cable.
6. Change the motor.

3.Horizontal motor cable has poor


contact/Horizontal motor is damaged.

DF50CRP↑ DH73/DH76↓
Apply some lubricant and move it back
and forth a few times.

1.Horizontal motor has too large


Resistance so it fails to return the initial
position in the optocoupler area.

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
there is no response, please
1.Change the sensor
2.Change the sensor cable

DH73/DH76

2.Horizontal motor initial position


optocoupler malfunctions.

61
Dymind Global Customer Service Center
Dymind Global Customer Service Center

1.Reconnect the motor cable.


2.Change the motor cable.
3.Change the motor.

3.Horizontal motor cable has poor


contact/Horizontal motor is damaged.

DH73/DH76↑ D5-CRP↓
Apply some lubricant and move it back
and forth a few times.

1.Horizontal motor has too large


Resistance so it fails to return the initial
position in the optocoupler area.

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
there is no response, please
1.Change the sensor
2.Change the sensor cable

D5-CRP

2.Horizontal motor initial position


optocoupler malfunctions.

1.Reconnect the motor cable.


3.Horizontal motor cable has poor
2.Change the motor cable.
contact/Horizontal motor is damaged.
3.Change the motor.

62
Dymind Global Customer Service Center
Dymind Global Customer Service Center

D5-CRP↑ D7-CRP↓
Apply some lubricant and move it back
and forth a few times.

1.Horizontal motor has too large


Resistance so it fails to return the initial
position in the optocoupler area.

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
there is no response, please
1.Change the sensor
2.Change the sensor cable

2.Horizontal motor initial position


D7-CRP
optocoupler malfunctions.

1.Reconnect the motor cable.


2.Change the motor cable.
3.Change the motor.
3.Horizontal motor cable has poor
contact/Horizontal motor is damaged.

63
Dymind Global Customer Service Center
Dymind Global Customer Service Center

If you have any problem with this file, please contact our engineer ASAP, thank you!

Relative spare part code


Part Name P/N Picture

Horizontal Sampling
68.02.0026A
Motor

DH26 Sensor(Vertical) 23.99.0007C

Sampling Motor
37.02.0145A
Connection Cable

Horizontal Sampling
68.02.0026A
Motor

Sensor(Vertical) 23.99.0007C

DH31/DH33/DH36/DH36VET

Sampling optocoupler
37.02.0036B
connection cable

Horizontal & Vertical Motor


37.02.0035A
Connection Cable

64
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Horizontal Sampling
68.02.0026A
Motor

Sensor(Vertical) 23.99.0007C

DF50/DF51/DF52/DF50VET

Sampling optocoupler
37.02.0036B
connection cable

Horizontal & Vertical Motor


37.02.0035A
Connection Cable

Horizontal Sampling
68.02.0026A
Motor

Sensor(Vertical) 23.99.0007C

DF50CRP

Horizontal &vertical
37.02.0035A
motor connection cable

Sampling optocoupler
37.02.0036B
connection cable

65
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Horizontal Sampling
68.02.0026A
Motor

Sensor(Vertical) 23.99.0007C

Sampling motor
37.01.0042A
connection cable

DH73/DH76

Sampling Optocoupler
37.02.0142A
interchange cable

Sampling Optocoupler
37.02.0051A
connection cable1

Sampling Optocoupler
37.02.0052A
connection cable2

Horizontal Sampling
68.02.0026A
Motor

DH36X/D5-CRP/D7-CRP

Sensor(Vertical) 23.99.0007C

66
Dymind Global Customer Service Center
Dymind Global Customer Service Center

0x230741F1-Horizontal motor failed to move to initial position

Use Background
When analyzer gives a false alarm: Horizontal motor failed to move to initial position.
Mechanism: The horizontal motor initial position optocoupler is always in the unblocked status.

Declaration
08/9/2020, V1.0, For reference only.

Reference Standard
Model Possibility Solution

Apply some lubricant and move it back


and forth a few times.

1.Horizontal motor has too large


Resistance so it fails to move to the initial
position in the optocoupler area.

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
DH26 there is no response, please
5. Change the sensor
6. Change the sensor cable

2.Horizontal motor initial position


optocoupler malfunctions.

67
Dymind Global Customer Service Center
Dymind Global Customer Service Center

7. Reconnect the motor cable.


8. Change the motor cable.
9. Change the motor

3.Horizontal motor cable has poor


contact/Horizontal motor is damaged.

DH26↑ DH31/DH33/DH36/DH36VET↓

Apply some lubricant and move it back


and forth a few times.

1.Horizontal motor has too large


Resistance so it fails to move to the initial
position in the optocoupler area.

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
there is no response, please
1.Change the sensor
DH31/DH33/D
2.Change the sensor cable
H36/DH36VET

2.Horizontal motor initial position


optocoupler malfunctions.

68
Dymind Global Customer Service Center
Dymind Global Customer Service Center

1.Reconnect the motor cable.


2.Change the motor cable.
3.Change the motor.

3.Horizontal motor cable has poor


contact/Horizontal motor is damaged.

DH31/DH33/DH36/DH36VET↑ DH36X↓
Apply some lubricant and move it back
and forth a few times.

1.Horizontal motor has too large


Resistance so it fails to move to the initial
position in the optocoupler area.

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
there is no response, please
5. Change the sensor
6. Change the sensor cable
DH36X

2.Horizontal motor initial position


optocoupler malfunctions.

69
Dymind Global Customer Service Center
Dymind Global Customer Service Center

7. Reconnect the motor cable.


8. Change the motor cable.
9. Change the motor.

3.Horizontal motor cable has poor


contact/Horizontal motor is damaged.

DH36X↑ DF50/DF51/DF52/DF50VET↓
Apply some lubricant and move it back
and forth a few times.

1.Horizontal motor has too large


Resistance so it fails to move to the initial
position in the optocoupler area.

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
there is no response, please
1.Change the sensor
DF50/DF51/DF 2.Change the sensor cable
52/DF50VET

2.Horizontal motor initial position


optocoupler malfunctions.

70
Dymind Global Customer Service Center
Dymind Global Customer Service Center

1.Reconnect the motor cable.


2.Change the motor cable.
3.Change the motor.

3.Horizontal motor cable has poor


contact/Horizontal motor is damaged.

DF50/DF51/DF52/DF50VET↑ DF50CRP↓
Apply some lubricant and move it back
and forth a few times.

1.Horizontal motor has too large


Resistance so it fails to move to the initial
position in the optocoupler area.

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
there is no response, please
1.Change the sensor
2.Change the sensor cable
DF50CRP

2.Horizontal motor initial position


optocoupler malfunctions.

71
Dymind Global Customer Service Center
Dymind Global Customer Service Center

7. Reconnect the motor cable.


8. Change the motor cable.
9. Change the motor.

3.Horizontal motor cable has poor


contact/Horizontal motor is damaged.

DF50CRP↑ DH73/DH76↓
Apply some lubricant and move it back
and forth a few times.

1.Horizontal motor has too large


Resistance so it fails to move to the initial
position in the optocoupler area.

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
there is no response, please
1.Change the sensor
2.Change the sensor cable

DH73/DH76

2.Horizontal motor initial position


optocoupler malfunctions.

72
Dymind Global Customer Service Center
Dymind Global Customer Service Center

1.Reconnect the motor cable.


2.Change the motor cable.
3.Change the motor.

3.Horizontal motor cable has poor


contact/Horizontal motor is damaged.

DH73/DH76↑ D5-CRP↓
Apply some lubricant and move it back
and forth a few times.

1.Horizontal motor has too large


Resistance so it fails to move to the initial
position in the optocoupler area.

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
there is no response, please
1.Change the sensor
2.Change the sensor cable

D5-CRP

2.Horizontal motor initial position


optocoupler malfunctions.

1.Reconnect the motor cable.


3.Horizontal motor cable has poor
2.Change the motor cable.
contact/Horizontal motor is damaged.
3.Change the motor.

73
Dymind Global Customer Service Center
Dymind Global Customer Service Center

D5-CRP↑ D7-CRP↓
Apply some lubricant and move it back
and forth a few times.

1.Horizontal motor has too large


Resistance so it fails to move to the initial
position in the optocoupler area.

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
there is no response, please
1.Change the sensor
2.Change the sensor cable

2.Horizontal motor initial position


D7-CRP
optocoupler malfunctions.

1.Reconnect the motor cable.


2.Change the motor cable.
3.Change the motor.
3.Horizontal motor cable has poor
contact/Horizontal motor is damaged.

74
Dymind Global Customer Service Center
Dymind Global Customer Service Center

If you have any problem with this file, please contact our engineer ASAP, thank you!

Relative spare part code


Part Name P/N Picture

Horizontal Sampling
68.02.0026A
Motor

DH26 Sensor(Vertical) 23.99.0007C

Sampling Motor
37.02.0145A
Connection Cable

Horizontal Sampling
68.02.0026A
Motor

Sensor(Vertical) 23.99.0007C

DH31/DH33/DH36/DH36VET

Sampling optocoupler
37.02.0036B
connection cable

Horizontal & Vertical Motor


37.02.0035A
Connection Cable

75
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Horizontal Sampling
68.02.0026A
Motor

Sensor(Vertical) 23.99.0007C

DF50/DF51/DF52/DF50VET

Sampling optocoupler
37.02.0036B
connection cable

Horizontal & Vertical Motor


37.02.0035A
Connection Cable

Horizontal Sampling
68.02.0026A
Motor

Sensor(Vertical) 23.99.0007C

DF50CRP

Horizontal &vertical
37.02.0035A
motor connection cable

Sampling optocoupler
37.02.0036B
connection cable

76
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Horizontal Sampling
68.02.0026A
Motor

Sensor(Vertical) 23.99.0007C

Sampling motor
37.01.0042A
connection cable

DH73/DH76

Sampling Optocoupler
37.02.0142A
interchange cable

Sampling Optocoupler
37.02.0051A
connection cable1

Sampling Optocoupler
37.02.0052A
connection cable2

Horizontal Sampling
68.02.0026A
Motor

DH36X/D5-CRP/D7-CRP

Sensor(Vertical) 23.99.0007C

77
Dymind Global Customer Service Center
Dymind Global Customer Service Center

0x230841C0-Vertical motor failed to leave initial position

Use Background
When analyzer gives a false alarm: Vertical motor failed to leave initial position.
Mechanism: The vertical motor initial position optocoupler always detects that the optocoupler is
blocked.

Declaration
08/9/2020, V1.0, For reference only.

Reference Standard
Model Possibility Solution

Apply some lubricant and move it back


and forth a few times.

1.Vertical motor has too large


Resistance so it fails to leave the initial
position in the optocoupler area.

DH26

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
there is no response, please
7. Change the sensor
2.Vertical motor initial position 8. Change the sensor cable
optocoupler malfunctions.

78
Dymind Global Customer Service Center
Dymind Global Customer Service Center

10. Reconnect the motor cable.


11. Change the motor cable.
12. Change the motor

3.Vertical motor cable has poor


contact/Vertical motor is damaged.

DH26↑ DH31/DH33/DH36/DH36VET↓

Apply some lubricant and move it back


and forth a few times.

1.Vertical motor has too large


Resistance so it fails to leave the initial
position in the optocoupler area.

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
2.Vertical motor initial position
there is no response, please
optocoupler malfunctions.
DH31/DH33/D 1.Change the sensor
H36/DH36VET 2.Change the sensor cable

79
Dymind Global Customer Service Center
Dymind Global Customer Service Center

1.Reconnect the motor cable.


2.Change the motor cable.
3.Change the motor.

3.Vertical motor cable has poor


contact/Vertical motor is damaged.

DH31/DH33/DH36/DH36VET↑ DH36X↓

Apply some lubricant and move it back


and forth a few times.

1.Vertical motor has too large


Resistance so it fails to leave the initial
position in the optocoupler area.

DH36X

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
2.Vertical motor initial position
there is no response, please
optocoupler malfunctions.
7. Change the sensor
8. Change the sensor cable

80
Dymind Global Customer Service Center
Dymind Global Customer Service Center

10. Reconnect the motor cable.


11. Change the motor cable.
12. Change the motor.

3.Vertical motor cable has poor


contact/Vertical motor is damaged.

DH36X↑ DF50/DF51/DF52/DF50VET↓
Apply some lubricant and move it back
and forth a few times.

1.Vertical motor has too large


Resistance so it fails to leave the initial
position in the optocoupler area.
DF50/DF51/DF
52/DF50VET

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
2.Vertical motor initial position
there is no response, please
optocoupler malfunctions.
1.Change the sensor
2.Change the sensor cable

81
Dymind Global Customer Service Center
Dymind Global Customer Service Center

1.Reconnect the motor cable.


2.Change the motor cable.
3.Change the motor.

3.Vertical motor cable has poor


contact/Vertical motor is damaged.

DF50/DF51/DF52/DF50VET↑ DF50CRP↓

Apply some lubricant and move it back


and forth a few times.

1.Vertical motor has too large


Resistance so it fails to leave the initial
position in the optocoupler area.
DF50CRP

Please press Service<Sensor to test if this


2.Vertical motor initial position
sensor working or not by a paper card, if
optocoupler malfunctions.
there is no response, please
82
Dymind Global Customer Service Center
Dymind Global Customer Service Center

1.Change the sensor


2.Change the sensor cable

10. Reconnect the motor cable.


11. Change the motor cable.
12. Change the motor.

3.Vertical motor cable has poor


contact/Vertical motor is damaged.

DF50CRP↑ DH73/DH76↓
Apply some lubricant and move it back
and forth a few times.

1.Vertical motor has too large


Resistance so it fails to leave the initial
position in the optocoupler area.

DH73/DH76

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
2.Vertical motor initial position
there is no response, please
optocoupler malfunctions.
1.Change the sensor
2.Change the sensor cable

83
Dymind Global Customer Service Center
Dymind Global Customer Service Center

1.Reconnect the motor cable.


2.Change the motor cable.
3.Change the motor.

3.Vertical motor cable has poor


contact/Vertical motor is damaged.

DH73/DH76↑ D5-CRP↓
Apply some lubricant and move it back
and forth a few times.

1.Vertical motor has too large


Resistance so it fails to leave the initial
D5-CRP position in the optocoupler area.

2.Vertical motor initial position Please press Service<Sensor to test if this


optocoupler malfunctions. sensor working or not by a paper card, if

84
Dymind Global Customer Service Center
Dymind Global Customer Service Center

there is no response, please


1.Change the sensor
2.Change the sensor cable

1.Reconnect the motor cable.


2.Change the motor cable.
3.Change the motor.

3.Vertical motor cable has poor


contact/Vertical motor is damaged.

D5-CRP↑ D7-CRP↓
Apply some lubricant and move it back
and forth a few times.

1.Vertical motor has too large


Resistance so it fails to leave the initial
position in the optocoupler area.

D7-CRP

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
2.Vertical motor initial position
there is no response, please
optocoupler malfunctions.
1.Change the sensor
2.Change the sensor cable

85
Dymind Global Customer Service Center
Dymind Global Customer Service Center

1.Reconnect the motor cable.


2.Change the motor cable.
3.Change the motor.

3.Vertical motor cable has poor


contact/Vertical motor is damaged.

If you have any problem with this file, please contact our engineer ASAP, thank you!

Relative spare part code


Part Name P/N Picture

Vertical Sampling
DH26 68.02.0027A
Motor

86
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Sensor(Vertical) 23.99.0007C

Sampling Motor
37.02.0145A
Connection Cable

Vertical Sampling
68.02.0027A
Motor

Sensor(Vertical) 23.99.0007C

DH31/DH33/DH36/DH36VET

Sampling optocoupler
37.02.0036B
connection cable

Horizontal & Vertical Motor


37.02.0035A
Connection Cable

Vertical Sampling
68.02.0027A
Motor

DF50/DF51/DF52/DF50VET

Sensor(Vertical) 23.99.0007C

87
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Sampling optocoupler
37.02.0036B
connection cable

Horizontal & Vertical Motor


37.02.0035A
Connection Cable

Vertical Sampling
68.02.0027A
Motor

Sensor(Vertical) 23.99.0007C

DF50CRP

Horizontal &vertical
37.02.0035A
motor connection cable

Sampling optocoupler
37.02.0036B
connection cable

Vertical Sampling
68.02.0027A
Motor

DH73/DH76

Sensor(Vertical) 23.99.0007C

88
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Sampling motor
37.01.0042A
connection cable

Sampling Optocoupler
37.02.0142A
interchange cable

Sampling Optocoupler
37.02.0051A
connection cable1

Sampling Optocoupler
37.02.0052A
connection cable2

Vertical Sampling
68.02.0027A
Motor

DH36X/D5-CRP/D7-CRP

Sensor(Vertical) 23.99.0007C

89
Dymind Global Customer Service Center
Dymind Global Customer Service Center

0x230841D0-Vertical motor failed to return to initial position

Use Background
When analyzer gives a false alarm: Vertical motor failed to return to initial position.
Mechanism: The vertical motor initial position optocoupler is always in the unblocked status.

Declaration
08/9/2020, V1.0, For reference only.

Reference Standard
Model Possibility Solution

Apply some lubricant and move it back


and forth a few times.

1.Vertical motor has too large


Resistance so it fails to return the initial
position in the optocoupler area.

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
DH26 there is no response, please
9. Change the sensor
10. Change the sensor cable

2.Vertical motor initial position


optocoupler malfunctions.

90
Dymind Global Customer Service Center
Dymind Global Customer Service Center

13. Reconnect the motor cable.


14. Change the motor cable.
15. Change the motor

3.Vertical motor cable has poor


contact/Vertical motor is damaged.

DH26↑ DH31/DH33/DH36/DH36VET↓

Apply some lubricant and move it back


and forth a few times.

1.Vertical motor has too large


Resistance so it fails to return the initial
position in the optocoupler area.

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
there is no response, please
DH31/DH33/D 1.Change the sensor
H36/DH36VET 2.Change the sensor cable

2.Vertical motor initial position


optocoupler malfunctions.

91
Dymind Global Customer Service Center
Dymind Global Customer Service Center

1.Reconnect the motor cable.


2.Change the motor cable.
3.Change the motor.

3.Vertical motor cable has poor


contact/Vertical motor is damaged.

DH31/DH33/DH36/DH36VET↑ DH36X↓

Apply some lubricant and move it back


and forth a few times.

1.Vertical motor has too large


Resistance so it fails to return the initial
position in the optocoupler area.

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
there is no response, please
9. Change the sensor
DH36X 10. Change the sensor cable

2.Vertical motor initial position


optocoupler malfunctions.

92
Dymind Global Customer Service Center
Dymind Global Customer Service Center

13. Reconnect the motor cable.


14. Change the motor cable.
15. Change the motor.

3.Vertical motor cable has poor


contact/Vertical motor is damaged.

DH36X↑ DF50/DF51/DF52/DF50VET↓
Apply some lubricant and move it back
and forth a few times.

1.Vertical motor has too large


Resistance so it fails to return the initial
position in the optocoupler area.

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
there is no response, please
DF50/DF51/DF 1.Change the sensor
52/DF50VET 2.Change the sensor cable

2.Vertical motor initial position


optocoupler malfunctions.

93
Dymind Global Customer Service Center
Dymind Global Customer Service Center

1.Reconnect the motor cable.


2.Change the motor cable.
3.Change the motor.

3.Vertical motor cable has poor


contact/Vertical motor is damaged.

DF50/DF51/DF52/DF50VET↑ DF50CRP↓

Apply some lubricant and move it back


and forth a few times.

1.Vertical motor has too large


Resistance so it fails to return the initial
position in the optocoupler area.

DF50CRP

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
there is no response, please
1.Change the sensor
2.Change the sensor cable
2.Vertical motor initial position
optocoupler malfunctions.

94
Dymind Global Customer Service Center
Dymind Global Customer Service Center

13. Reconnect the motor cable.


14. Change the motor cable.
15. Change the motor.

3.Vertical motor cable has poor


contact/Vertical motor is damaged.

DF50CRP↑ DH73/DH76↓
Apply some lubricant and move it back
and forth a few times.

1.Vertical motor has too large


Resistance so it fails to return the initial
position in the optocoupler area.

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
DH73/DH76
there is no response, please
1.Change the sensor
2.Change the sensor cable

2.Vertical motor initial position


optocoupler malfunctions.

95
Dymind Global Customer Service Center
Dymind Global Customer Service Center

1.Reconnect the motor cable.


2.Change the motor cable.
3.Change the motor.

3.Vertical motor cable has poor


contact/Vertical motor is damaged.

DH73/DH76↑ D5-CRP↓
Apply some lubricant and move it back
and forth a few times.

1.Vertical motor has too large


Resistance so it fails to return the initial
position in the optocoupler area.

D5-CRP

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
there is no response, please
1.Change the sensor
2.Vertical motor initial position
2.Change the sensor cable
optocoupler malfunctions.

96
Dymind Global Customer Service Center
Dymind Global Customer Service Center

1.Reconnect the motor cable.


2.Change the motor cable.
3.Change the motor.

3.Vertical motor cable has poor


contact/Vertical motor is damaged.

D5-CRP↑ D7-CRP↓
Apply some lubricant and move it back
and forth a few times.

1.Vertical motor has too large


Resistance so it fails to return the initial
position in the optocoupler area.

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
D7-CRP there is no response, please
1.Change the sensor
2.Change the sensor cable

2.Vertical motor initial position


optocoupler malfunctions.

97
Dymind Global Customer Service Center
Dymind Global Customer Service Center

1.Reconnect the motor cable.


2.Change the motor cable.
3.Change the motor.

3.Vertical motor cable has poor


contact/Vertical motor is damaged.

If you have any problem with this file, please contact our engineer ASAP, thank you!

Relative spare part code


Part Name P/N Picture

Vertical Sampling
68.02.0027A
Motor

DH26 Sensor(Vertical) 23.99.0007C

Sampling Motor
37.02.0145A
Connection Cable

98
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Vertical Sampling
68.02.0027A
Motor

Sensor(Vertical) 23.99.0007C

DH31/DH33/DH36/DH36VET

Sampling optocoupler
37.02.0036B
connection cable

Horizontal & Vertical Motor


37.02.0035A
Connection Cable

Vertical Sampling
68.02.0027A
Motor

Sensor(Vertical) 23.99.0007C

DF50/DF51/DF52/DF50VET

Sampling optocoupler
37.02.0036B
connection cable

Horizontal & Vertical Motor


37.02.0035A
Connection Cable

99
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Vertical Sampling
68.02.0027A
Motor

Sensor(Vertical) 23.99.0007C

DF50CRP

Horizontal &vertical
37.02.0035A
motor connection cable

Sampling optocoupler
37.02.0036B
connection cable

Vertical Sampling
68.02.0027A
Motor

Sensor(Vertical) 23.99.0007C

DH73/DH76

Sampling motor
37.01.0042A
connection cable

Sampling Optocoupler
37.02.0142A
interchange cable

100
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Sampling Optocoupler
37.02.0051A
connection cable1

Sampling Optocoupler
37.02.0052A
connection cable2

Vertical Sampling
68.02.0027A
Motor

DH36X/D5-CRP/D7-CRP

Sensor(Vertical) 23.99.0007C

101
Dymind Global Customer Service Center
Dymind Global Customer Service Center

0x230841F1-Vertical motor failed to move to initial position

Use Background
When analyzer gives a false alarm: Horizontal motor failed to move to initial position.
Mechanism: The Vertical motor initial position optocoupler is always in the unblocked status.

Declaration
08/9/2020, V1.0, For reference only.

Reference Standard
Model Possibility Solution

Apply some lubricant and move it back


and forth a few times.

1.Vertical motor has too large


Resistance so it fails to move to the initial
position in the optocoupler area.

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
DH26 there is no response, please
11. Change the sensor
12. Change the sensor cable

2.Vertical motor initial position


optocoupler malfunctions.

102
Dymind Global Customer Service Center
Dymind Global Customer Service Center

16. Reconnect the motor cable.


17. Change the motor cable.
18. Change the motor

3.Vertical motor cable has poor


contact/Vertical motor is damaged.

DH26↑ DH31/DH33/DH36/DH36VET↓

Apply some lubricant and move it back


and forth a few times.

1.Vertical motor has too large


Resistance so it fails to move to the initial
position in the optocoupler area.

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
there is no response, please
DH31/DH33/D 1.Change the sensor
H36/DH36VET 2.Change the sensor cable

2.Vertical motor initial position


optocoupler malfunctions.

103
Dymind Global Customer Service Center
Dymind Global Customer Service Center

1.Reconnect the motor cable.


2.Change the motor cable.
3.Change the motor.

3.Vertical motor cable has poor


contact/Vertical motor is damaged.

DH31/DH33/DH36/DH36VET↑ DH36X↓

Apply some lubricant and move it back


and forth a few times.

1.Vertical motor has too large


Resistance so it fails to move to the initial
position in the optocoupler area.

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
there is no response, please
11. Change the sensor
DH36X 12. Change the sensor cable

2.Vertical motor initial position


optocoupler malfunctions.

104
Dymind Global Customer Service Center
Dymind Global Customer Service Center

16. Reconnect the motor cable.


17. Change the motor cable.
18. Change the motor.

3.Vertical motor cable has poor


contact/Vertical motor is damaged.

DH36X↑ DF50/DF51/DF52/DF50VET↓
Apply some lubricant and move it back
and forth a few times.

1.Vertical motor has too large


Resistance so it fails to move to the initial
position in the optocoupler area.

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
there is no response, please
DF50/DF51/DF 1.Change the sensor
52/DF50VET 2.Change the sensor cable

2.Vertical motor initial position


optocoupler malfunctions.

105
Dymind Global Customer Service Center
Dymind Global Customer Service Center

1.Reconnect the motor cable.


2.Change the motor cable.
3.Change the motor.

3.Vertical motor cable has poor


contact/Vertical motor is damaged.

DF50/DF51/DF52/DF50VET↑ DF50CRP↓

Apply some lubricant and move it back


and forth a few times.

1.Vertical motor has too large


Resistance so it fails to move to the initial
position in the optocoupler area.

DF50CRP

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
there is no response, please
1.Change the sensor
2.Change the sensor cable
2.Vertical motor initial position
optocoupler malfunctions.

106
Dymind Global Customer Service Center
Dymind Global Customer Service Center

16. Reconnect the motor cable.


17. Change the motor cable.
18. Change the motor.

3.Vertical motor cable has poor


contact/Vertical motor is damaged.

DF50CRP↑ DH73/DH76↓
Apply some lubricant and move it back
and forth a few times.

1.Vertical motor has too large


Resistance so it fails to move to the initial
position in the optocoupler area.

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
DH73/DH76
there is no response, please
1.Change the sensor
2.Change the sensor cable

2.Vertical motor initial position


optocoupler malfunctions.

107
Dymind Global Customer Service Center
Dymind Global Customer Service Center

1.Reconnect the motor cable.


2.Change the motor cable.
3.Change the motor.

3.Vertical motor cable has poor


contact/Vertical motor is damaged.

DH73/DH76↑ D5-CRP↓
Apply some lubricant and move it back
and forth a few times.

1.Vertical motor has too large


Resistance so it fails to move to the initial
position in the optocoupler area.

D5-CRP

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
there is no response, please
1.Change the sensor
2.Vertical motor initial position
2.Change the sensor cable
optocoupler malfunctions.

108
Dymind Global Customer Service Center
Dymind Global Customer Service Center

1.Reconnect the motor cable.


2.Change the motor cable.
3.Change the motor.

3.Vertical motor cable has poor


contact/Vertical motor is damaged.

D5-CRP↑ D7-CRP↓
Apply some lubricant and move it back
and forth a few times.

1.Vertical motor has too large


Resistance so it fails to move to the initial
position in the optocoupler area.

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
D7-CRP there is no response, please
1.Change the sensor
2.Change the sensor cable

2.Vertical motor initial position


optocoupler malfunctions.

109
Dymind Global Customer Service Center
Dymind Global Customer Service Center

1.Reconnect the motor cable.


2.Change the motor cable.
3.Change the motor.

3.Vertical motor cable has poor


contact/Vertical motor is damaged.

If you have any problem with this file, please contact our engineer ASAP, thank you!

Relative spare part code


Part Name P/N Picture

Vertical Sampling
68.02.0027A
Motor

DH26 Sensor(Vertical) 23.99.0007C

Sampling Motor
37.02.0145A
Connection Cable

110
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Vertical Sampling
68.02.0027A
Motor

Sensor(Vertical) 23.99.0007C

DH31/DH33/DH36/DH36VET

Sampling optocoupler
37.02.0036B
connection cable

Horizontal & Vertical Motor


37.02.0035A
Connection Cable

Vertical Sampling
68.02.0027A
Motor

Sensor(Vertical) 23.99.0007C

DF50/DF51/DF52/DF50VET

Sampling optocoupler
37.02.0036B
connection cable

Horizontal & Vertical Motor


37.02.0035A
Connection Cable

111
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Vertical Sampling
68.02.0027A
Motor

Sensor(Vertical) 23.99.0007C

DF50CRP

Horizontal &vertical
37.02.0035A
motor connection cable

Sampling optocoupler
37.02.0036B
connection cable

Vertical Sampling
68.02.0027A
Motor

Sensor(Vertical) 23.99.0007C

DH73/DH76

Sampling motor
37.01.0042A
connection cable

Sampling Optocoupler
37.02.0142A
interchange cable

112
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Sampling Optocoupler
37.02.0051A
connection cable1

Sampling Optocoupler
37.02.0052A
connection cable2

Vertical Sampling
68.02.0027A
Motor

DH36X/D5-CRP/D7-CRP

Sensor(Vertical) 23.99.0007C

113
Dymind Global Customer Service Center
Dymind Global Customer Service Center

0x23074190-Horizontal motor photocoupler abnormal

Use Background
When analyzer gives a false alarm: Horizontal motor photocoupler abnormal.
Mechanism: Photocoupler error occurred in the initial position, WBC bath, DIFF bath, RBC bath,
open-sampling position, or auto-sampling position of the horizontal motor.

Declaration
17/8/2020, V1.0, For reference only.

Reference Standard
Model Possibility Solution

Apply some lubricant and move it back


and forth a few times.

1.Horizontal motor has too large


Resistance

DH26
Please press Service<Sensor to test if this
sensor working or not by a paper card, if
there is no response, please
13. Change the sensor
14. Change the sensor cable
2.Horizontal motor initial position
optocoupler malfunctions.

114
Dymind Global Customer Service Center
Dymind Global Customer Service Center

DH26↑ DH31/DH33/DH36/DH36VET↓

Apply some lubricant and move it back


and forth a few times.

1.Horizontal motor has too large


Resistance

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
there is no response, please
1.Change the sensor
DH31/DH33/D
2.Change the sensor cable
H36/DH36VET

2.Horizontal motor initial position


optocoupler malfunctions.

DH31/DH33/DH36/DH36VET↑ DH36X↓
1.Horizontal motor has too large Apply some lubricant and move it back
DH36X
Resistance and forth a few times.

115
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
there is no response, please
13. Change the sensor
14. Change the sensor cable

2.Horizontal motor initial position


optocoupler malfunctions.

DH36X↑ DF50/DF51/DF52/DF50VET↓
Apply some lubricant and move it back
and forth a few times.

1.Horizontal motor has too large


Resistance

DF50/DF51/DF
52/DF50VET

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
2.Horizontal motor initial position
there is no response, please
optocoupler malfunctions.
1.Change the sensor
2.Change the sensor cable

116
Dymind Global Customer Service Center
Dymind Global Customer Service Center

DF50/DF51/DF52/DF50VET↑ DF50CRP↓
Apply some lubricant and move it back
and forth a few times.

1.Horizontal motor has too large


Resistance

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
there is no response, please
1.Change the sensor
2.Change the sensor cable
DF50CRP

2.Horizontal motor initial position


optocoupler malfunctions.

DF50CRP↑ DH73/DH76↓

117
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Apply some lubricant and move it back


and forth a few times.

1.Horizontal motor has too large


Resistance

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
there is no response, please
1.Change the sensor
DH73/DH76 2.Change the sensor cable

2.Horizontal motor initial position


optocoupler malfunctions.

DH73/DH76↑ D5-CRP↓
Apply some lubricant and move it back
and forth a few times.

1.Horizontal motor has too large


D5-CRP Resistance

2.Horizontal motor initial position Please press Service<Sensor to test if this


118
Dymind Global Customer Service Center
Dymind Global Customer Service Center

optocoupler malfunctions. sensor working or not by a paper card, if


there is no response, please
1.Change the sensor
2.Change the sensor cable

D5-CRP↑ D7-CRP↓
Apply some lubricant and move it back
and forth a few times.

1.Horizontal motor has too large


Resistance

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
D7-CRP
there is no response, please
1.Change the sensor
2.Change the sensor cable

2.Horizontal motor initial position


optocoupler malfunctions.

119
Dymind Global Customer Service Center
Dymind Global Customer Service Center

If you have any problem with this file, please contact our engineer ASAP, thank you!

Relative spare part code


Part Name P/N Picture

DH26 Sensor(Vertical) 23.99.0007C

Sensor(Vertical) 23.99.0007C

DH31/DH33/DH36/DH36VET

DF50/DF51/DF52/DF50VET

Sampling optocoupler
37.02.0036B
connection cable

Sensor(Vertical) 23.99.0007C

DF50CRP

Sampling optocoupler
37.02.0036B
connection cable

120
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Sensor(Vertical) 23.99.0007C

Sampling Optocoupler
37.02.0142A
interchange cable

DH73/DH76

Sampling Optocoupler
37.02.0051A
connection cable1

Sampling Optocoupler
37.02.0052A
connection cable2

DH36X/D5-CRP/D7-CRP Sensor(Vertical) 23.99.0007C

121
Dymind Global Customer Service Center
Dymind Global Customer Service Center

0x23074200-Horizontal motor action overtime

Use Background
When analyzer gives a false alarm: Horizontal motor action overtime.
Mechanism: Failed to complete the action within the time specified by fluidics.

Declaration
17/8/2020, V1.0, For reference only.

Reference Standard
Model Possibility Solution

Apply some lubricant and move it back


and forth a few times.

1.Horizontal motor has too large


Resistance

DH26

2.Action loses steps caused by mechanical


ageing of the horizontal assembly.

DH26↑ DH31/DH33/DH36/DH36VET↓
Apply some lubricant and move it back
and forth a few times.

DH31/DH33/D 1.Horizontal motor has too large


H36/DH36VET Resistance

122
Dymind Global Customer Service Center
Dymind Global Customer Service Center

2.Action loses steps caused by mechanical


ageing of the horizontal assembly.

DH31/DH33/DH36/DH36VET↑ DH36X↓
Apply some lubricant and move it back
and forth a few times.

1.Horizontal motor has too large


Resistance

DH36X

2.Action loses steps caused by mechanical


ageing of the horizontal assembly.

DH36X↑ DF50/DF51/DF52/DF50VET↓
Apply some lubricant and move it back
and forth a few times.

1.Horizontal motor has too large


Resistance

DF50/DF51/DF
52/DF50VET

2.Action loses steps caused by mechanical


ageing of the horizontal assembly.

DF50/DF51/DF52/DF50VET↑ DF50CRP↓

123
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Apply some lubricant and move it back


and forth a few times.

1.Horizontal motor has too large


Resistance

DF50CRP

2.Action loses steps caused by mechanical


ageing of the horizontal assembly.

DF50CRP↑ DH73/DH76↓
Apply some lubricant and move it back
and forth a few times.

1.Horizontal motor has too large


Resistance

DH73/DH76

2.Action loses steps caused by mechanical


ageing of the horizontal assembly.

DH73/DH76↑ D5-CRP↓
1.Horizontal motor has too large Apply some lubricant and move it back
D5-CRP
Resistance and forth a few times.

124
Dymind Global Customer Service Center
Dymind Global Customer Service Center

2.Horizontal motor initial position


optocoupler malfunctions.

D5-CRP↑ D7-CRP↓
Apply some lubricant and move it back
and forth a few times.

1.Horizontal motor has too large


Resistance

D7-CRP

2.Action loses steps caused by mechanical


ageing of the horizontal assembly.

If you have any problem with this file, please contact our engineer ASAP, thank you!

125
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Relative spare part code


Part Name P/N Picture

Horizontal Sampling
68.02.0026A
Motor

DH26 Sensor(Vertical) 23.99.0007C

Sampling Motor
37.02.0145A
Connection Cable

Horizontal Sampling
68.02.0026A
Motor

Sensor(Vertical) 23.99.0007C

DH31/DH33/DH36/DH36VET

Sampling optocoupler
37.02.0036B
connection cable

Horizontal & Vertical Motor


37.02.0035A
Connection Cable

Horizontal Sampling
DF50/DF51/DF52/DF50VET 68.02.0026A
Motor

126
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Sensor(Vertical) 23.99.0007C

Sampling optocoupler
37.02.0036B
connection cable

Horizontal & Vertical Motor


37.02.0035A
Connection Cable

Horizontal Sampling
68.02.0026A
Motor

Sensor(Vertical) 23.99.0007C

DF50CRP

Horizontal &vertical
37.02.0035A
motor connection cable

Sampling optocoupler
37.02.0036B
connection cable

Horizontal Sampling
68.02.0026A
DH73/DH76 Motor

127
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Sensor(Vertical) 23.99.0007C

Sampling motor
37.01.0042A
connection cable

Sampling Optocoupler
37.02.0142A
interchange cable

Sampling Optocoupler
37.02.0051A
connection cable1

Sampling Optocoupler
37.02.0052A
connection cable2

Horizontal Sampling
68.02.0026A
Motor

DH36X/D5-CRP/D7-CRP

Sensor(Vertical) 23.99.0007C

128
Dymind Global Customer Service Center
Dymind Global Customer Service Center

0x23084190-Vertical motor photocoupler abnormal

Use Background
When analyzer gives a false alarm: Vertical motor photocoupler abnormal.
Mechanism: Photocoupler error occurred in the initial position or aspirate position of the vertical
motor.
Declaration
09/9/2020, V1.0, For reference only.

Reference Standard
Model Possibility Solution

Apply some lubricant and move it back


and forth a few times.

1.Vertical motor has too large


Resistance

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
DH26 there is no response, please
15. Change the sensor
16. Change the sensor cable

2.Vertical motor initial position


optocoupler malfunctions.

129
Dymind Global Customer Service Center
Dymind Global Customer Service Center

DH26↑ DH31/DH33/DH36/DH36VET↓

Apply some lubricant and move it back


and forth a few times.

1.Vertical motor has too large


Resistance

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
there is no response, please
DH31/DH33/D 1.Change the sensor
H36/DH36VET 2.Change the sensor cable

2.Vertical motor initial position


optocoupler malfunctions.

DH31/DH33/DH36/DH36VET↑ DH36X↓

Apply some lubricant and move it back


and forth a few times.

1.Vertical motor has too large


DH36X
Resistance

130
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
there is no response, please
15. Change the sensor
16. Change the sensor cable

2.Vertical motor initial position


optocoupler malfunctions.

DH36X↑ DF50/DF51/DF52/DF50VET↓
Apply some lubricant and move it back
and forth a few times.

1.Vertical motor has too large


Resistance

DF50/DF51/DF Please press Service<Sensor to test if this


52/DF50VET sensor working or not by a paper card, if
there is no response, please
1.Change the sensor
2.Change the sensor cable
2.Vertical motor initial position
optocoupler malfunctions.

131
Dymind Global Customer Service Center
Dymind Global Customer Service Center

DF50/DF51/DF52/DF50VET↑ DF50CRP↓

Apply some lubricant and move it back


and forth a few times.

1.Vertical motor has too large


Resistance

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
there is no response, please
DF50CRP 1.Change the sensor
2.Change the sensor cable

2.Vertical motor initial position


optocoupler malfunctions.

DF50CRP↑ DH73/DH76↓

132
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Apply some lubricant and move it back


and forth a few times.

1.Vertical motor has too large


Resistance

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
there is no response, please
1.Change the sensor
2.Change the sensor cable

DH73/DH76

2.Vertical motor initial position


optocoupler malfunctions.

DH73/DH76↑ D5-CRP↓
Apply some lubricant and move it back
and forth a few times.

1.Vertical motor has too large


D5-CRP
Resistance

133
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
there is no response, please
1.Change the sensor
2.Change the sensor cable

2.Vertical motor initial position


optocoupler malfunctions.

D5-CRP↑ D7-CRP↓
Apply some lubricant and move it back
and forth a few times.

1.Vertical motor has too large


Resistance

Please press Service<Sensor to test if this


sensor working or not by a paper card, if
D7-CRP
there is no response, please
1.Change the sensor
2.Change the sensor cable

2.Vertical motor initial position


optocoupler malfunctions.

134
Dymind Global Customer Service Center
Dymind Global Customer Service Center

If you have any problem with this file, please contact our engineer ASAP, thank you!

Relative spare part code


Part Name P/N Picture

Sensor(Vertical) 23.99.0007C

DH26

Sampling Motor
37.02.0145A
Connection Cable

Sensor(Vertical) 23.99.0007C

DH31/DH33/DH36/DH36VET

Sampling optocoupler
37.02.0036B
connection cable

DF50/DF51/DF52/DF50VET Sensor(Vertical) 23.99.0007C

135
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Sampling optocoupler
37.02.0036B
connection cable

Sensor(Vertical) 23.99.0007C

DF50CRP

Sampling optocoupler
37.02.0036B
connection cable

Sensor(Vertical) 23.99.0007C

Sampling Optocoupler
37.02.0142A
interchange cable

DH73/DH76

Sampling Optocoupler
37.02.0051A
connection cable1

Sampling Optocoupler
37.02.0052A
connection cable2

DH36X/D5-CRP/D7-CRP Sensor(Vertical) 23.99.0007C

136
Dymind Global Customer Service Center
Dymind Global Customer Service Center

0x23084200-Vertical motor action overtime

Use Background
When analyzer gives a false alarm: Vertical motor action overtime.
Mechanism: Failed to complete the action within the time specified by fluidics.

Declaration
09/9/2020, V1.0, For reference only.

Reference Standard
Model Possibility Solution
Apply some lubricant and move it back
and forth a few times.

1.Vertical motor has too large


Resistance

DH26
Please change the whole Vertical
assembly.

2.Action loses steps caused by mechanical


ageing of the Vertical assembly.

DH26↑ DH31/DH33/DH36/DH36VET↓

Apply some lubricant and move it back


and forth a few times.

1.Vertical motor has too large


Resistance

137
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Please change the whole Vertical


assembly.

DH31/DH33/D
H36/DH36VET 2.Action loses steps caused by mechanical
ageing of the Vertical assembly.

DH31/DH33/DH36/DH36VET↑ DH36X↓

Apply some lubricant and move it back


and forth a few times.

1.Vertical motor has too large


Resistance

DH36X

Please change the whole Vertical


assembly.

2.Action loses steps caused by mechanical


ageing of the Vertical assembly.

DH36X↑ DF50/DF51/DF52/DF50VET↓
DF50/DF51/DF 1.Vertical motor has too large Apply some lubricant and move it back
52/DF50VET Resistance and forth a few times.

138
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Please change the whole Vertical


assembly.

2.Action loses steps caused by mechanical


ageing of the Vertical assembly.

DF50/DF51/DF52/DF50VET↑ DF50CRP↓

Apply some lubricant and move it back


and forth a few times.

1.Vertical motor has too large


Resistance

DF50CRP

Please change the whole Vertical


assembly.

2.Action loses steps caused by mechanical


ageing of the Vertical assembly.

DF50CRP↑ DH73/DH76↓
139
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Apply some lubricant and move it back


and forth a few times

1.Vertical motor has too large


Resistance

DH73/DH76
Please change the whole Vertical
assembly.

2.Action loses steps caused by mechanical


ageing of the Vertical assembly.

DH73/DH76↑ D5-CRP↓
Apply some lubricant and move it back
and forth a few times

1.Vertical motor has too large


Resistance

D5-CRP
Please change the whole Vertical
assembly.

2.Action loses steps caused by mechanical


ageing of the Vertical assembly.

D5-CRP↑ D7-CRP↓
1.Vertical motor has too large Apply some lubricant and move it back
D7-CRP
Resistance and forth a few times
140
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Please change the whole Vertical


assembly.

2.Action loses steps caused by mechanical


ageing of the Vertical assembly.

If you have any problem with this file, please contact our engineer ASAP, thank you!

Relative spare part code


Part Name P/N Picture

Vertical Sampling
68.02.0027A
Motor

DH26

Sensor(Vertical) 23.99.0007C

141
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Sampling Motor
37.02.0145A
Connection Cable

Vertical Sampling
68.02.0027A
Motor

Sensor(Vertical) 23.99.0007C

DH31/DH33/DH36/DH36VET

Sampling optocoupler
37.02.0036B
connection cable

Horizontal & Vertical Motor


37.02.0035A
Connection Cable

Vertical Sampling
68.02.0027A
Motor

DF50/DF51/DF52/DF50VET Sensor(Vertical) 23.99.0007C

Sampling optocoupler
37.02.0036B
connection cable

142
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Horizontal & Vertical Motor


37.02.0035A
Connection Cable

Vertical Sampling
68.02.0027A
Motor

Sensor(Vertical) 23.99.0007C

DF50CRP

Horizontal &vertical
37.02.0035A
motor connection cable

Sampling optocoupler
37.02.0036B
connection cable

Vertical Sampling
68.02.0027A
Motor

Sensor(Vertical) 23.99.0007C
DH73/DH76

Sampling motor
37.01.0042A
connection cable

143
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Sampling Optocoupler
37.02.0142A
interchange cable

Sampling Optocoupler
37.02.0051A
connection cable1

Sampling Optocoupler
37.02.0052A
connection cable2

Vertical Sampling
68.02.0027A
Motor

DH36X/D5-CRP/D7-CRP

Sensor(Vertical) 23.99.0007C

144
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Aperture Problem

145
Dymind Global Customer Service Center
Dymind Global Customer Service Center

0xB1000505-WBC aperture voltage abnormal

Use Background
When analyzer gives a false alarm - WBC aperture voltage abnormal.
Mechanism: The WBC aperture voltage is out of the allowable range [1.45, 2.05] V, which can be
checked in Self-test > Other self-test > WBC aperture voltage.

Declaration
31/3/2021, V2.0, For reference only.

Reference Standard
Model Possibility Solution
Check if there’s any diluent in RBC chamber.
A. If there is no, then check the diluent box,
the syringe, the valve connected to syringe.
B. If there is, then do a blank test, confirm
whether the problem is no diluent dispense
into WBC chamber or WBC chamber leakage.
Replace LV6, isolation chamber, LV10
respectively to confirm.

1. There’s no diluent in the WBC


DH26/DH31/D
chamber or unstable liquid level.
H33/DH36/DH
36VET

2. Some bubbles go through WBC Replace LV1, LV7, the tube connected, and
aperture or nearby. the connector respectively to confirm.

146
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Replace another box diluent.

3. Diluent get contaminated or


expired.

Replace the WBC aperture or chamber.

4. The WBC aperture has failure.

Replace the WBC chamber.

5. The electrode of WBC chamber has


failure.

Wear an anti-static wrist and replace the


6. The main board has failure.
main board, keep the software of the
147
Dymind Global Customer Service Center
Dymind Global Customer Service Center

machine (host) the same as IPU(OPERATOR


SOFTWARE), otherwise it will not work.
DH26/DH31/DH33/DH36/DH36VET↑ DH36X↓
Check whether there is any diluent in RBC
chamber.
A. If there is no, then check the diluent box,
the syringe, the valve connected to syringe.
B. If there is, then do a blank test, confirm
whether the problem is no diluent dispense
into WBC chamber or WBC chamber leakage.
Replace LV6, isolation chamber, LV5, LV15,
1. There’s no diluent in the WBC LV16 respectively to confirm.
chamber or unstable liquid level.

DH36X Replace LV11, LV17, the tube connected, and


the connector respectively to confirm.

2. Some bubbles go through WBC


aperture or nearby.

3. Diluent get contaminated or Replace another box diluent.

148
Dymind Global Customer Service Center
Dymind Global Customer Service Center

expired.

Replace the WBC aperture or chamber.

4. The WBC aperture has failure.

Replace the WBC chamber.

5. The electrode of WBC chamber has


failure.

Wear an anti-static wrist and replace the


main board, keep the software of the
6. The main board has failure.
machine (host) the same as IPU(OPERATOR
SOFTWARE), otherwise it will not work.
DH36X↑ DH56/UN73/DH76/D5CRP/D7CRP↓

Check if there’s any diluent in RBC chamber,


A. If there is no, then check the diluent box,
the syringe, the valve connected to syringe.
DH56/UN73/D B. If there is, then do a blank test, confirm
1. There’s no diluent in the WBC
H76, the problem is no diluent dispense into WBC
chamber or unstable liquid level.
D5CRP/D7CRP chamber or WBC chamber leakage. Replace
LV8, isolation chamber, LV15, LV21(no need
to check for DH56/D5CRP), LV25 respectively
to locate it.

149
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Replace LV5, LV23, LV27, the tube connected,


and the connectors respectively to confirm it.

2. Some bubbles go through WBC


aperture or nearby.

3. Diluent get contaminated or Replace another box diluent.

150
Dymind Global Customer Service Center
Dymind Global Customer Service Center

expired.

Replace the WBC aperture or chamber.

4. The WBC aperture has failure.

Replace the WBC chamber.

5. The electrode of WBC chamber has


failure.

Wear an anti-static wrist and replace the


main board, keep the software of the
6. The main board has failure.
machine (host) the same as IPU(OPERATOR
SOFTWARE), otherwise it will not work.

If you have any problem with this file, please contact our engineer ASAP, thank you!
Relative spare part code
Model Part Name P/N Picture

Three-way valve (L) - SMC 23.99.0002A

All models above

Two-way valve (L) - SMC 23.99.0001A

151
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Three-way valve (S) - SMC 23.99.0004A

Two-way valve (S) - SMC 23.99.0003A

WBC aperture 24.11.0001A

T420-1 Connector 53.05.0003A

DH26/ Isolation Chamber with


DH31/DH33/DH36/ Strainer(Upper port 68.02.0022A
DH36VET blocked)

DH31/DH33/DH36/
Mainboard 22.01.0023A
DH36VET

DH36X Mainboard 22.01.0199A

DH31/DH33/DH36/
DH36VET, WBC bath assembly(with Ruby) 20.01.0143A
DH36X

152
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Mainboard 22.01.0072A

DH26

WBC bath assembly(with Ruby) 20.01.0459A

DH36X,
Isolation Chamber with
DH56/UN73/DH76, 60.01.0012C
Strainer
D5CRP/D7CRP

Mainboard 22.01.0001C

DH56/UN73/DH76,
WBC bath assembly(with Ruby) 20.01.0007A
D5CRP/D7CRP

Y230-1 Connector 53.05.0011A

153
Dymind Global Customer Service Center
Dymind Global Customer Service Center

0xB1000506-RBC aperture voltage abnormal

Use Background
When analyzer gives a false alarm - RBC aperture voltage abnormal.
Mechanism: The RBC aperture voltage is out of the allowable range [1.65, 2.25] V, which can be
checked in Self-test > Other self-test > RBC aperture voltage.

Declaration
31/3/2021, V2.0, For reference only.

Reference Standard
Model Possibility Solution

Check the diluent box, the syringe, the valve


connected to syringe. Replace LV2, LV5, LV6,
isolation chamber, LV11 respectively to
locate it.

1. There’s no diluent in the RBC


chamber or unstable liquid level.
DH26/DH31/DH
33/DH36/DH36
VET

Check the tube, connectors and valves


2. Some bubbles go through RBC connected to RBC aperture. Replace LV1, LV7,
aperture or nearby. the tube connected, and the connectors
respectively to locate it.

154
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Replace another box diluent.

3. Diluent get contaminated or


expired.

Replace the RBC aperture or chamber.

4. The RBC aperture has failure.

Replace the RBC chamber.

5. The electrode of RBC chamber has


failure.

Wear an anti-static wrist and replace the


main board, keep the software of the
6. The main board has failure.
machine (host) the same as IPU(OPERATOR
SOFTWARE), otherwise it will not work.
DH26/DH31/DH33/DH36/DH36VET↑ DH36X↓

155
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Check the diluent box, the syringe, the valve


connected to syringe. Replace LV7, LV15,
LV16, isolation chamber, LV12, LV4
respectively to locate it.

1. There’s no diluent in the RBC


chamber or unstable liquid level.

DH36X

Check the tube, connectors and valves


2. Some bubbles go through RBC connected to RBC aperture. Replace LV1,
aperture or nearby. LV17, the tube connected, and the
connectors respectively to locate it.

156
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Replace another box diluent.

3. Diluent get contaminated or


expired.

Replace the RBC aperture or chamber.

4. The RBC aperture has failure.

Replace the RBC chamber.

5. The electrode of RBC chamber has


failure.

157
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Wear an anti-static wrist and replace the


main board, keep the software of the
6. The main board has failure.
machine (host) the same as IPU(OPERATOR
SOFTWARE), otherwise it will not work.
DH36X↑ DF50/DF51/DF52/DF50VET↓

Check the diluent box, the syringe, the valve


connected to syringe. Replace LV1, LV5, LV7,
isolation chamber, LV13 to locate it.

1. There’s no diluent in the RBC


chamber or unstable liquid level.

DF50/DF51/DF5
2/DF50VET Check the tube, connectors and valves
connected to RBC aperture. Replace LV9,
LV18, the tube connected, and the
connectors respectively to locate it.

2. Some bubbles go through RBC


aperture or nearby.

3. Diluent get contaminated or Replace another box diluent.

158
Dymind Global Customer Service Center
Dymind Global Customer Service Center

expired.

Replace the RBC aperture or chamber.

4. The RBC aperture has failure.

Replace the RBC chamber.

5. The electrode of RBC chamber has


failure.

Wear an anti-static wrist and replace the


main board, keep the software of the
6. The main board has failure.
machine (host) the same as IPU(OPERATOR
SOFTWARE), otherwise it will not work.
DF50/DF51/DF52/DF50VET↑ DF50CRP↓

Check the diluent box, the syringe, the valve


connected to syringe. Replace LV13, LV21,
isolation chamber to locate it.

1. There’s no diluent in the RBC


DF50CRP
chamber or unstable liquid level.

159
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Check the tube, connectors and valves


connected to RBC aperture. Replace LV14,
LV23, the tube connected, and the
connectors respectively to locate it.

2. Some bubbles go through RBC


aperture or nearby.

Replace another box diluent.

3. Diluent get contaminated or


expired.

Replace the RBC aperture or chamber.

4. The RBC aperture has failure.

5. The electrode of RBC chamber has Replace the RBC chamber.

160
Dymind Global Customer Service Center
Dymind Global Customer Service Center

failure.

Wear an anti-static wrist and replace the


main board, keep the software of the
6. The main board has failure.
machine (host) the same as IPU(OPERATOR
SOFTWARE), otherwise it will not work.
DF50CRP ↑ DH56/UN73/DH76/D5CRP/D7CRP↓

Check the diluent box, the syringe, the valve


connected to syringe. Replace LV16, LV26,
isolation chamber, LV4, LV7, LV8 to locate it.

1. There’s no diluent in the RBC


DH56/UN73/DH
chamber or unstable liquid level.
76,
D5CRP/D7CRP

Replace LV6, LV24, LV28, the tube connected,


2. Some bubbles go through WBC and the connectors respectively to locate it.

161
Dymind Global Customer Service Center
Dymind Global Customer Service Center

aperture or nearby.

Replace another box diluent.

3. Diluent get contaminated or


expired.

Replace the RBC aperture or chamber.

4. The RBC aperture has failure.

Replace the RBC chamber.

5. The electrode of RBC chamber has


failure.

Replace the main board, and keep the


software of the machine (host) the same
6. The main board has failure.
with IPU(OPERATOR SOFTWARE), otherwise
it will not work.

162
Dymind Global Customer Service Center
Dymind Global Customer Service Center

If you have any problem with this file, please contact our engineer ASAP, thank you!

Relative spare part code


Model Part Name P/N Picture

Three-way valve (L) - SMC 23.99.0002A

Two-way valve (L) - SMC 23.99.0001A

Three-way valve (S) - SMC 23.99.0004A

All models above

Two-way valve (S) - SMC 23.99.0003A

RBC aperture 24.11.0002A

T420-1 Connector 53.05.0003A

DH26/DH31/DH33/DH36/DH Isolation Chamber with


36VET, Strainer(Upper port 68.02.0022A
DF50/DF51/DF52/DF50VET blocked)

163
Dymind Global Customer Service Center
Dymind Global Customer Service Center

DH31/DH33/DH36/DH36VET,
DH36X, RBC bath assembly(with
20.01.0144A
DF50/DF51/DF52/DF50VET, Ruby)
DF50CRP

DH36X Main Board 22.01.0199A

RBC bath assembly(with


20.01.0460A
Ruby)

DH26

Main Board 22.01.0072A

DH36X,
DF50CRP, Isolation Chamber with
60.01.0012C
DH56/UN73/DH76, Strainer
D5CRP/D7CRP

DH31/DH33/DH36/DH36VET Mainboard 22.01.0023A

DF50/DF51/DF52/DF50VET,
Mainboard 22.01.0021C
DF50CRP

DH56/UN73/DH76,
Mainboard 22.01.0001C
D5CRP/D7CRP

164
Dymind Global Customer Service Center
Dymind Global Customer Service Center

RBC bath assembly(with


20.01.0008A
Ruby)

Y230-1 Connector 53.05.0011A

165
Dymind Global Customer Service Center
Dymind Global Customer Service Center

0xB1000507-HGB background voltage abnormal

Use Background
When analyzer gives a false alarm - HGB background voltage abnormal.
Mechanism: The HGB background voltage is not in the allowable range [4.2~4.8]V.

Declaration
26/3/2021, V2.0, For reference only.

Reference Standard
Model Possibility Solution
Go to Gain Settings to set the HGB blank voltage
to 4.5±0.1V.

1. HGB blank voltage jump and


goes out of range.

DH26/DH31/D
H33/DH36/DH
36VET Check if there’s any diluent in RBC chamber.
A. If there is no, then check the diluent box, the
syringe, the valve connected to syringe.
B. If there is, then do a blank test, confirm the
2. There is no diluent or a little
problem is no diluent dispense into WBC chamber
leakage or overfill in WBC
or WBC chamber leakage. Replace LV6, isolation
chamber.
chamber, LV10 respectively to locate it.
Don’t forget to clean the outer-wall of WBC
chamber and HGB unit by non-dust cloth after fix
the overfill problem.

166
Dymind Global Customer Service Center
Dymind Global Customer Service Center

3. The HGB unit has failure, LED


Replace the HGB unit only.
light is off.
A. Do a cleanser soak for WBC chamber, clean the
aperture manually;
B. Clean the inner and outer-wall of WBC chamber
and HGB unit with Non-dust cloth;
C. Replace the WBC chamber.

4. The HGB unit or WBC chamber


get dirty.

Wear an anti-static wrist and replace the main


board, keep the software of the machine (host)
5. The main board has failure.
the same as IPU(OPERATOR SOFTWARE),
otherwise it will not work.
DH26/DH31/DH33/DH36/DH36VET↑ DH36X↓

Go to Gain Settings to set the HGB blank voltage


DH36X 1. HGB blank voltage jump and
to 4.5±0.1V.

167
Dymind Global Customer Service Center
Dymind Global Customer Service Center

goes out of range.

Check if there’s any diluent in RBC chamber.


A. If there is no, then check the diluent box, the
syringe, the valve connected to syringe.
B. If there is, then do a blank test, confirm the
problem is no diluent dispense into WBC chamber
or WBC chamber leakage. Replace LV5, LV6,
isolation chamber, LV12, LV15, LV16 respectively
to locate it.
Don’t forget to clean the outer-wall of WBC
chamber and HGB unit by non-dust cloth after fix
the overfill problem.
2. There is no diluent or a little
leakage or overfill in WBC
chamber.

3. The HGB unit has failure, light


Replace the HGB unit only.
emitter is off.
A. Do a cleanser soak for WBC chamber, clean the
aperture manually;
4. The HGB unit or WBC chamber
B. Clean the inner and outer-wall of WBC chamber
get dirty.
and HGB unit with Non-dust cloth;
C. Replace the WBC chamber.

168
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Wear an anti-static wrist and replace the main


board, keep the software of the machine (host)
5. The main board has failure.
the same as IPU(OPERATOR SOFTWARE),
otherwise it will not work.
DH36X↑ DF50/DF51/DF52/DF50VET↓
Go to Gain Settings to set the HGB blank voltage
to 4.5±0.1V.

1. HGB blank voltage jump and


goes out of range.

DF50/DF51/DF
52/DF50VET Check if there’s any diluent in RBC chamber.
A. If there is no, then check the diluent box, the
syringe, the valve connected to syringe.
B. If there is, then do a blank test, confirm the
2. There is no diluent or a little
problem is no diluent dispense into WBC chamber
leakage or overfill in WBC
or WBC chamber leakage. Replace LV7, LV14,
chamber.
isolation chamber, LV19 respectively to locate it.
Don’t forget to clean the outer-wall of WBC
chamber and HGB unit by non-dust cloth after fix
the overfill problem.

169
Dymind Global Customer Service Center
Dymind Global Customer Service Center

3. HGB unit get damaged and no


Replace the HGB unit.
LED light at all.
A. Do a cleanser soak for WBC chamber;
B. Clean the inner and outer-wall of WBC chamber
and HGB unit with Non-dust cloth;
C. Replace the WBC chamber.

4. The HGB unit or WBC chamber


get dirty.

Wear an anti-static wrist and replace the main


board, keep the software of the machine (host)
5. The main board damaged.
the same as IPU(OPERATOR SOFTWARE),
otherwise it will not work.
DF50/DF51/DF52/DF50VET↑ DF50CRP↓
Go to Gain Settings to set the HGB blank voltage
DF50CRP 1. HGB blank voltage jump and
to 4.5±0.1V. Setup > Gain settings.

170
Dymind Global Customer Service Center
Dymind Global Customer Service Center

goes out of range.

Check if there’s any diluent in RBC chamber.


A. If there is no, then check the diluent box, the
syringe, the valve connected to syringe.
B. If there is, then do a blank test, confirm the
problem is no diluent dispense into WBC chamber
or WBC chamber leakage. Replace LV12, isolation
chamber, LV22 and LV16 respectively to locate it.
Don’t forget to clean the outer-wall of WBC
chamber and HGB unit by non-dust cloth after
fixing the overfill problem.

2. There is no diluent or a little


leakage or overfill in WBC
chamber.

3. The HGB unit has failure, LED


Replace the HGB unit only.
light is off.

171
Dymind Global Customer Service Center
Dymind Global Customer Service Center

A. Do a cleanser soak for WBC chamber, clean the


aperture manually;
B. Clean the inner and outer-wall of WBC chamber
and HGB unit with Non-dust cloth;
C. Replace the WBC chamber.

4. The HGB unit or WBC chamber


get dirty.

Wear an anti-static wrist and replace the main


board, keep the software of the machine (host)
5. The main board has failure.
the same as IPU(OPERATOR SOFTWARE),
otherwise it will not work.
DF50CRP↑ DH56/UN73/DH76/D5CRP/D7CRP↓
Go to Gain Settings to set the HGB blank voltage
to 4.5±0.1V.

1. HGB blank voltage jump and


goes out of range

DH56/UN73/D
H76, Check if there’s any diluent in RBC chamber,
D5CRP/D7CRP A. If there is no, then check the diluent box, the
syringe, the valve connected to syringe.
B. If there is, then do a blank test, confirm the
2. There is no diluent or a little problem is no diluent dispense into WBC chamber
leakage or overfill in WBC or WBC chamber leakage. Replace LV8, isolation
chamber. chamber, LV15, LV21(DH56&D5CRP don’t have),
LV25 respectively to locate it.
Don’t forget to clean the outer-wall of WBC
chamber and HGB unit by non-dust cloth after fix
the overfill problem.

172
Dymind Global Customer Service Center
Dymind Global Customer Service Center

3. The HGB unit has failure, LED


Replace the HGB unit only.
light is off.
A. Do a cleanser soak for WBC chamber, clean the
aperture manually;
B. Clean the inner and outer-wall of WBC chamber
and HGB unit with Non-dust cloth;
C. Replace the WBC chamber.

4. The HGB unit or WBC chamber


get dirty.

Wear an anti-static wrist and replace the main


board, keep the software of the machine (host)
5. The main board has failure.
the same as IPU(OPERATOR SOFTWARE),
otherwise it will not work.

If you have any problem with this file, please contact our engineer ASAP, thank you!

173
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Relative spare part code


Model Part Name P/N Picture

Three-way valve (L) - SMC 23.99.0002A

Two-way valve (L) - SMC 23.99.0001A

All models above

Three-way valve (S) - SMC 23.99.0004A

Two-way valve (S) - SMC 23.99.0003A

HGB unit 20.01.1209A

WBC Bath Assembly(with


DH26 20.01.0459A
Ruby)

Main Board 22.01.0072A

DH31/DH33/DH36/DH36VE,
DH36X,
HGB unit 68.02.0024A
DH56/UN73/DH76,
D5CRP/D7CRP

174
Dymind Global Customer Service Center
Dymind Global Customer Service Center

DH36X,
DF50/DF51/DF52/DF50VET,
Isolation Chamber with
DF50CRP, 60.01.0012C
Strainer
DH56/UN73/DH76,
D5CRP/D7CRP

Isolation Chamber with


DH26/DH31/DH33/DH36/DH Strainer(Upper port 68.02.0022A
36VET blocked)

WBC bath assembly(with


20.01.0143A
Ruby)

DH31/DH33/DH36/DH36VET

Mainboard 22.01.0023A

HGB unit 68.02.0025A

DF50/DF51/DF52/DF50VET, WBC Bath Assembly(with


20.01.0121A
DF50CRP Ruby)

Mainboard 22.01.0021C

DH56/UN73/DH76, WBC bath assembly(with


20.01.0007A
D5CRP/D7CRP Ruby)

175
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Mainboard 22.01.0001C

176
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Clog Problem

177
Dymind Global Customer Service Center
Dymind Global Customer Service Center

0x63026220-Flow cell clog

Use Background
When “Flow cell Clog” is prompted in the area of problem on the analyzer interface and the RBC
counting result of the analyzer is abnormal of the repeatability analyzers are not satisfied.

During the counting, flow cell is clog, which slows down the fluid from passing through cell flow,
increases the sheath pressure sensor voltage, and triggers the clogging alarm.

Declaration
9/4/2021, V1.0, For reference only.

Reference Standard
Module Possibility Solution
DH36 N/A N/A
We could flash the flow cell manually.
(Power off the analyzer, then aspirate the diluent out)
We could get these videos of how to do maintenance for
analyzer in our FTP account, as following:
Website: http://159.138.130.90/
User name: S-user03
Password: PVns6bPL
The blood clot or
some big particle
block the flow cell

For
DF50/
DF51/
DF52/
DF50VET
a. Check the process as shown below, and replace LV11
b. Check the connector (T9\T10\J13) and replace it

Hydraulic system
analysis

178
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Replace the shealth pressure sensor, then to check if the results


are normal (as following)

The sheath
For DF50/ pressure sensor
DF52 damaged and give
the false alarm

We could flash the flow cell manually.


(Power off the analyzer, then aspirate the diluent out)
We could get these videos of how to do maintenance for
analyzer in our FTP account, as following:
Website: http://159.138.130.90/
User name: S-user03
Password: PVns6bPL

The blood clot or


some big particle
block the flow cell

For
DF50CRP

c. Check the process as shown below, and replace LV20


d. Check the connector (T70/T29/J13) and replace it

Hydraulic system
analysis

179
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Replace the shealth pressure sensor, then to check if the results


are normal (as following)

The sheath
For pressure sensor
DF50CRP damaged and give
the false alarm

We could flash the flow cell manually.


(Power off the analyzer, then aspirate the diluent out)
We could get these videos of how to do maintenance for
analyzer in our FTP account, as following:
Website: http://159.138.130.90/
User name: S-user03
Password: PVns6bPL
The blood clot or
some big particle
block the flow cell

For
DH56/
D5-CRP
DH76/
D7-CRP a. Check the process as shown below, and replace LV18
b. Check the connector (T70\T29\J2) and replace it

Hydraulic system
analysis

180
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Replace the shealth pressure sensor, then to check if the results


are normal (as following)

For
DH56/ The sensor
D5-CRP damaged and give
DH76/ the false alarm
D7-CRP

If you have any problem with this file, please contact our engineer ASAP, thank you!

Relative spare part code


Part Name P/N Picture

Sheath flow pressure


23.99.0019A
sensor

All

2-way valve (L) – (SMC) 20.01.1197A

181
Dymind Global Customer Service Center
Dymind Global Customer Service Center

0x63026221-DIFF probe clog

Use Background
When “DIFF probe Clog” is prompted in the area of problem on the analyzer interface and the RBC
counting result of the analyzer is abnormal of the repeatability analyzers are not satisfied.

During the counting, flow cell is clog, which slows down the fluid from passing through cell flow,
increases the sheath pressure sensor voltage, and triggers the clogging alarm.

Declaration
4/9/2021, V1.0, For reference only.

Reference Standard
Module Possibility Solution
DH36 N/A N/A
We could flash the flow cell manually.
(Power off the analyzer, then aspirate the diluent out)
We could get these videos of how to do maintenance for
analyzer in our FTP account, as following:
Website: http://159.138.130.90/
User name: S-user03
Password: PVns6bPL
The blood clot or
some big particle
block the flow cell

For DF50/
DF52

e. Check the process as shown below, and replace LV11


f. Check the connector (T9\T10\J13) and replace it

Hydraulic system
analysis

182
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Replace the shealth pressure sensor, then to check if the results


are normal (as following)

The sheath
For DF50/ pressure sensor
DF52 damaged and give
the false alarm

We could flash the flow cell manually.


(Power off the analyzer, then aspirate the diluent out)
We could get these videos of how to do maintenance for
analyzer in our FTP account, as following:
Website: http://159.138.130.90/
User name: S-user03
Password: PVns6bPL

The blood clot or


some big particle
block the flow cell

For
DF50CRP

g. Check the process as shown below, and replace LV20


h. Check the connector (T70/T29/J13) and replace it

Hydraulic system
analysis

183
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Replace the shealth pressure sensor, then to check if the results


are normal (as following)

The shealth
For pressure sensor
DF50CRP damaged and give
the false alarm

We could flash the flow cell manually.


(Power off the analyzer, then aspirate the diluent out)
We could get these videos of how to do maintenance for
analyzer in our FTP account, as following:
Website: http://159.138.130.90/
User name: S-user03
Password: PVns6bPL
The blood clot or
some big particle
block the flow cell

For
DH76/
D7-CRP

c. Check the process as shown below, and replace LV18


d. Check the connector (T70\T29\J2) and replace it

Hydraulic system
analysis

184
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Replace the shealth pressure sensor, then to check if the results


are normal (as following)

For
The sensor
DH76/
damaged and give
D7-CRP
the false alarm

If you have any problem with this file, please contact our engineer ASAP, thank you!

Relative spare part code


Part Name P/N Picture

Sheath flow pressure


23.99.0019A
sensor

All

2-way valve (L) – (SMC) 20.01.1197A

185
Dymind Global Customer Service Center
Dymind Global Customer Service Center

0xb2004001-background abnormal

Use Background
When analyzer gives a false alarm –background abnormal, and we can’t remove it.

The background counting result of each parameter upon starup is not within the specified range

Declaration
15/9/2020, V1.0, For reference only.

Reference Standard
Module Possibility Solution

1.The reagents are expired Replace the reagents, prime more times to check if it
or contaminated back to normal

A. Clean the chambers manually, especially the


aperture(Refer to SOP for specific steps)

All modules
2.The hydraulics system
B. Perform cleanser soak the chamber manually. (If
dirty
the problem is still existed, we could replace a new
chamber)

186
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Check if there’s bubbles inside the tube of syringe,


replace the O-ring of the syringe or the syringe body
assembly.
3.There are bubbles
generated by the syringe

All modules

A. Check if the shielding cover loose and fix it


4.External interference
B.Move to another room if there’s big power
machine nearby, or big noisy

If you have any problem with this file, please contact our engineer ASAP, thank you!

Relative spare part code


Part Name P/N Picture

According to modules,
All modules Please purchase a set of / /
service kit

187
Dymind Global Customer Service Center
Dymind Global Customer Service Center

0xB2004110-WBC Clog

Use Background
When “WBC Clog” is prompted in the area of problem on the analyzer interface and the WBC
counting result of the analyzer is abnormal of the repeatability analyzers are not satisfied.
Principle of problem: During the counting some or all of the apertures are clogged, which slows
down the fluid from passing through the apertures, increases the aperture voltage, and triggers
the clogging alarm.

Declaration
25/3/2021 , V1.0, For reference only.

Reference Standard
modules Possible cause Solution
Check if the tubing as shown below is loose or replace the
corresponding tubing, valves (LV1, LV7, LV8, LV9 ) and pump1.

1.The tubing is
damaged or
loose

For
DH36/ A. Go to service > clean > click the WBC chamber button in the
DH31/ analyzer software interface
DH33/
DH36-VET

2. Aperture
clogging

188
Dymind Global Customer Service Center
Dymind Global Customer Service Center

B. Go to the service > Maintenance > click the flush Apertures and
then the zap Aperture button.

a. Go to the service > Maintenance > click the


Empty WBC Chamber button

For
DH36/
2.Aperture
DH31/
clogging
DH33/
DH36-VET

C. After front steps b. Take a syringe with diluted cleaner, then


are finished, if connect it at back of counting chamber,
clogging persists, then pull slowly for 3 times. If the issue is
we need to back still existed, we could repalce new
flash WBC chamber.
counting chamber Tips: Do not too fast, or the tubes would be
disconnected

189
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Observe if the chamber mixing is normal. If any problem occurs,


please check the corresponding tubing and components
(LV10/LV8/LV9/pump1), perform the solution for “Aperture
clogging”.

For
DH36/
3. Failed to mix
DH31/
evenly
DH33/
DH36-VET

Check if the tubing as shown below is loose or replace the


corresponding tubing, valves (LV11, LV17, LV18) and pump1.

1.The tubing is
damaged or
loose

For DH36X

A. Go to service > clean > click the WBC chamber button in the
analyzer software interface

2. Aperture
clogging

190
Dymind Global Customer Service Center
Dymind Global Customer Service Center

B. Go to the service > Maintenance > click the zap Aperture


button.

a. Go to the service > Maintenance > click


the Empty WBC Chamber button

2. Aperture
For DH36X
clogging

C. After front steps


b. Take a syringe with diluted cleaner, then
are finished, if
connect it at back of counting chamber,
clogging persists, we
then pull slowly for 3 times. If the issue
need to back flash
is still existed, we could repalce new
WBC counting
chamber.
chamber
Tips: Do not too fast, or the tubes would
be disconnected

191
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Observe if the chamber mixing is normal. If any problem occurs,


please check the corresponding tubing and components (LV6/ air
pump2), perform the solution for “Aperture clogging”.

3. Failed to mix
evenly

Observe if the chamber mixing is normal. If any problem occurs,


For DH36x
please check the corresponding tubing and components
(LV5/LV15/LV16/LV18/pump1), perform the solution for “Aperture
clogging”.

4. Failed to
drain

192
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Check if the tubing as shown below is loose


or replace the corresponding tubing,
valves (LV27/LV29) and pump 2 .

A. For the process


of counting

1.The tubing is
Check if the tubing as shown below is loose
damaged or
or replace the corresponding tubing, valves
loose
(LV5/LV23/LV29) and pump 2 .

For B. For the process


DH56/ of WBC chamber
D5-CRP cleaning
DH76/
D7-CRP

Check if the tubing as shown below is loose


or replace the corresponding tubing,
valves (LV17/LV27) and air pump 3.

1.The tubing is
C. For the process
damaged or
of back flashing
loose

193
Dymind Global Customer Service Center
Dymind Global Customer Service Center

A. Go to service > clean > click the WBC chamber button in the
software interface

B. Go to the service > Maintenance > click the flush Apertures and
then the zap Aperture button.

2.Aperture
clogging

For a. Go to the service > Maintenance > click the


DH56/ Empty WBC Chamber button
D5-CRP C. After front steps
DH76/ are finished, if
D7-CRP clogging persists,
we need to back
flash RBC counting
chamber

b. Take a syringe with diluted cleaner, then


connect it at back of counting chamber,
push slowly then pull slowly for 3 times. If
the issue is still existed, weneed to
C. After front steps replace a new chamber.
are finished, if Tips: Do not too fast, or the tubes would be
2.Aperture clogging persists, disconnected
clogging we need to back
flash RBC counting
chamber

194
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Observe if the chamber mixing is normal. If any problem occurs,


please check the corresponding tubing and components
(LV16/pump1), perform the solution for “Aperture clogging”.

3.Failed to mix
evenly

For
DH56/
D5-CRP
DH76/
D7-CRP Observe if the chamber draining is normal. If any problem occurs,
please check the corresponding tubing and components
(LV21/LV25/pump2 /pump3), perform the solution for “Aperture
clogging”.

4. Failed to
drain

195
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Relative spare parts code


Parts Name P/N Picture

2-way valve(L)-
23.99.0001A
(SMC)

3-way valve(S)-
23.99.0004A
(SMC)

2-way valve(S)-
23.99.0003A
(SMC)

All modules

Vacuum Pump 23.99.0008A

WBC Bath
20.01.0143A
Assembly(with Ruby)

Isolation Chamber
68.02.0023A
(Upper port blocked)

If you have any problem with this file, please contact our engineer ASAP, thank you!

196
Dymind Global Customer Service Center
Dymind Global Customer Service Center

0xB2004210-RBC Clog

Use Background
When “RBC Clog” is prompted in the area of problem on the analyzer interface and the RBC
counting result of the analyzer is abnormal of the repeatability analyzers are not satisfied.

Principle of problem: During the counting some or all of the apertures are clogged, which slows
down the fluid from passing through the apertures, increases the aperture voltage, and triggers
the clogging alarm.

Declaration
16/7/2020, V1.0, For reference only.

Reference Standard
modules Possible Solution
cause
Check if the tubing as shown below is loose or replace the
corresponding tubing, valves (LV1, LV7, LV8, LV9) and pump 1 .

1. The
tubing is
damaged or
loose

For
DH36/
A. Go to service > clean > click the RBC chamber button in the
DH31/
analyzer software interface
DH33/
DH36-VET

2. Aperture
clogging

197
Dymind Global Customer Service Center
Dymind Global Customer Service Center

B. Go to the service > Maintenance > click the flush Apertures and
then the zap Aperture button.

c. Go to the service > Maintenance > click


the Empty RBC Chamber button

For
DH36/
2. Aperture
DH31/
clogging
DH33/
DH36-VET

C. After front steps b. Take a syringe with diluted cleaner, then


are finished, if connect it at back of counting chamber,
clogging persists, then pull slowly for 3 times.
we need to back Tips: Do not too fast, or the tubes would be
flash RBC counting disconnected
chamber

198
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Observe if the chamber mixing is normal. If any problem occurs,


please check the corresponding tubing and components
(LV11/LV8/LV9/pump1), perform the solution for “Aperture
clogging”.

For
DH36/
3. Failed to
DH31/
mix evenly
DH33/
DH36-VET

Check if the tubing as shown below is loose or replace the


corresponding tubing, valves (LV11, LV17, LV18) and pump1.

1.The tubing
is damaged
or loose

For DH36X

A. Go to service > clean > click the RBC chamber button in the
analyzer software interface

2. Aperture
clogging

199
Dymind Global Customer Service Center
Dymind Global Customer Service Center

B. Go to the service > Maintenance > click the zap Aperture button

a. Go to the service > Maintenance > click


the Empty RBC Chamber button

2. Aperture
For DH36X
clogging

C. After front steps


are finished, if c. Take a syringe with diluted cleaner, then
clogging persists, connect it at back of counting chamber,
we need to back then pull slowly for 3 times. If the issue
flash RBC counting is still existed, we could repalce new
chamber chamber.
Tips: Do not too fast, or the tubes would be
disconnected

200
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Observe if the chamber draining is normal. If any problem occurs,


please check the corresponding tubing and components (LV7/ air
pump 2 ), perform the solution for “Aperture clogging”.

3. Failed to
mix evenly

Observe if the chamber mixing is normal. If any problem occurs,


For DH36X
please check the corresponding tubing and components
(LV15/LV16/LV18/pump1), perform the solution for “Aperture
clogging”.

4. Failed to
drain evenly

201
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Check if the tubing as shown below is loose or replace the


corresponding tubing, valves(LV1/LV9/LV18/ LV15/LV16) and
pump1 .

1.The tubing
is damaged
or loose

A. Go to service > clean > click the RBC chamber button in the
analyzer software interface

For
DF50/
DF51/
DF52/
DF50VET

2.Aperture B. Go to the service > Maintenance > click the flush Apertures and
clogging then the zap Aperture button.

202
Dymind Global Customer Service Center
Dymind Global Customer Service Center

a. Go to the service > Maintenance > click


the Empty RBC Chamber button

C. After front steps b. Take a syringe with diluted cleaner, then


are finished, if connect it at back of counting chamber,
2. Aperture clogging persists, push slowly then pull slowly for 3 times.
clogging we need to back If the issue is still existed, weneed to
flash RBC counting replace a new chamber.
chamber Tips: Do not too fast, or the tubes would be
disconnected

For
DF50/
DF51/
DF52/
DF50VET

Observe if the chamber mixing is normal. If any problem occurs,


please check the corresponding tubing and components
(LV13/LV15/LV16/pump1), perform the solution for “Aperture
clogging”.

3.Failed to
mix evenly

203
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Check if the tubing as shown below is loose or replace the


corresponding tubing, valves (LV14, LV23, LV25) and pump1.

1.The tubing
is damaged
or loose

A. Go to service > clean > click the RBC chamber button in the
analyzer software interface

For
DF50CRP

B. Go to the service > Maintenance > click the zap Aperture button.
2. Aperture
clogging

204
Dymind Global Customer Service Center
Dymind Global Customer Service Center

a. Go to the service > Maintenance > click


the Empty RBC Chamber button

C. After front steps


are finished, if c. Take a syringe with diluted cleaner, then
2. Aperture clogging persists, connect it at back of counting chamber,
clogging we need to back then pull slowly for 3 times. If the issue
flash RBC counting is still existed, we could repalce new
chamber chamber.
Tips: Do not too fast, or the tubes would be
disconnected

For
DF50CRP

Observe if the chamber mixing is normal. If any problem occurs,


please check the corresponding tubing and components
(LV13/pump2), perform the solution for “Aperture clogging”.

3. Failed to
mix evenly

205
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Observe if the chamber mixing is normal. If any problem occurs,


please check the corresponding tubing and components
(LV21/LV25/pump1), perform the solution for “Aperture clogging”.

For 4. Failed to
DF50CRP drain

Check if the tubing as shown below is loose


or replace the corresponding tubing,
valves (LV28/LV29) and pump 2

A. For the process


of counting

For
Check if the tubing as shown below is loose
DH56/ 1.The tubing
or replace the corresponding tubing, valves
D5-CRP is damaged or
(LV6/LV24/LV29) and pump 2 .
DH76/ loose
D7-CRP
B. For the process
of WBC chamber
cleaning

Check if the tubing as shown below is loose


C. For the process
or replace the corresponding tubing,
of back flashing
valves (LV17/LV28) and air pump 3.

206
Dymind Global Customer Service Center
Dymind Global Customer Service Center

A. Go to service > clean > click the RBC chamber button in the
software interface

B. Go to the service > Maintenance > click the flush Apertures and
then the zap Aperture button.

For
DH56/
2.Aperture
D5-CRP
clogging
DH76/
D7-CRP a. Go to the service > Maintenance > click
the Empty RBC Chamber button

C. After front steps


are finished, if
clogging persists,
we need to back
b. Take a syringe with diluted cleaner, then
flash RBC counting
connect it at back of counting chamber,
chamber
push slowly then pull slowly for 3 times.
If the issue is still existed, weneed to
replace a new chamber.
Tips: Do not too fast, or the tubes would be
disconnected
207
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Observe if the chamber mixing is normal. If any problem occurs,


please check the corresponding tubing and components
(LV16/pump1), perform the solution for “Aperture clogging”.

3.Failed to
mix evenly

For
DH56/
D5-CRP Observe if the chamber draining is normal. If any problem occurs,
DH76/ please check the corresponding tubing and components
D7-CRP (LV21/LV25/pump2 /pump3), perform the solution for “Aperture
clogging”.

4. Failed to
drain

208
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Relative spare parts code


Parts Name P/N Picture

2-way valve(L)-
23.99.0001A
(SMC)

3-way valve(S)-
23.99.0004A
(SMC)

All modules 2-way valve(S)-


23.99.0003A
(SMC)

Vacuum Pump 23.99.0008A

RBC Bath
20.01.0144A
Assembly(with Ruby)

If you have any problem with this file, please contact our engineer ASAP, thank you!

209
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Pressure Problem

210
Dymind Global Customer Service Center
Dymind Global Customer Service Center

0x63006030-Positive pressure abnormal (low)

Use Background
When analyzer gives a false alarm –positive pressure abnormal (low) , and we can’t remove it.
Alarm principle: The host queries the pressure in the positive-pressure chamber. The current
pressure is less than the lower limit (5.5KPa) of the attached

Declaration
25/3/2020, V1.0, For reference only.

Reference Standard
Module Possibility Solution
A. Check if the tube of yellow part is blocked or folded;
B. Check if LV8, LV 9, pump1 are abnormal. ( it’s better to
replace new one to check)
C. Clean the swab by cleanser. ( we could follow the steps of
SOP to operate)

1.Hydraulic analysis

For A. Check pressure


DH36/ sensor ( replace new
DH31/ one)
DH33/ B. Check mainboard
DH36-VET

2.Hardware analysis

211
Dymind Global Customer Service Center
Dymind Global Customer Service Center

the new analyzer has not been used for a long time, the
ambient temperature in same places is too high, the
membrane of pump is too dry to work abnormally
(as following figure)

For
DH36/
3.special situation How to solve it: we just back flash the pump1 by diluent .(as
DH31/
for new analyzer following figure)
DH33/
DH36-VET

A. Check if the tube of yellow part is blocked or folded;


B. Check if pump2 are abnormal. ( it’s better to replace new
one to check)

1.Hydraulic analysis

For
DH36x
A. Check pressure sensor ( replace new one)
B. Check mainboard

2.Hardware analysis

212
Dymind Global Customer Service Center
Dymind Global Customer Service Center

A. Check if the tube of yellow part is blocked or folded;


B. Check if LV15, LV 16, pump1 are abnormal. ( it’s better to
replace new one to check)
C. Clean the swab by cleanser. ( we could follow the steps of
SOP to operate)
1.Hydraulic analysis

For
DF50/ A. Check pressure
DF51/ sensor ( replace new
DF52/ one)
DF50VET B. Check mainboard

2.Hardware analysis

3. special situation Same as DH36


for new analyzer

A. Check if the tube of red part is blocked or folded;


B. Check if pump2 are abnormal. ( it’s better to replace new
one to check)

For
1.Hydraulic analysis
DF50CRP

213
Dymind Global Customer Service Center
Dymind Global Customer Service Center

A. Check pressure
sensor ( replace new
one)
B. Check mainboard

For
2.Hardware analysis
DF50CRP

1.Check if the tube


of yellow part is
blocked or folded;

2.Check if pump3
For (air pump) are
DH56/ abnormal. ( it’s
DH76/ better to replace
D5-CRP/ new one to check)
D7-CRP
3.replace air filter

4.check if the
positive pressure
sensor is normal.

214
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Relative spare part code


Part Name P/N Picture

mainboard 22.01.0023A

Liquid detective board 22.01.0022B

DH31/DH33/DH36/
3-way valve(S)-(SMC) 23.99.0004A
DH36-VET

Vacuum Pump 23.99.0008A

Probe Wipe Assembly 68.02.0020A

Mainboard with SD card


22.01.0199A
(without battery)

DH36X

Reagent Detecting Board 22.01.0204A

215
Dymind Global Customer Service Center
Dymind Global Customer Service Center

DH36X Air pump 24.01.0002A

Main Board 22.01.0021C

Liquid Detection Board 22.01.0022B

DF50/DF51/DF52/
3-way valve(S)-(SMC) 23.99.0004A
DF50VET

Vacuum Pump 23.99.0008A

Probe Wipe Assembly 68.02.0020A

Mainboard with SD card


DF50CRP 22.01.0021C
(without battery)

216
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Liquid Detection Board 20.01.0006B

DF50CRP

Air pump 24.01.0002A

Liquid Detecting Board 20.01.0006B

DH56/ DH76/
D5-CRP/
D7-CRP Liquid Detecting Board 20.01.0024B

Air pump 24.01.0002A

If you have any problem with this file, please contact our engineer ASAP, thank you!

217
Dymind Global Customer Service Center
Dymind Global Customer Service Center

0x63006040-Positive pressure abnormal (high)

Use Background
When analyzer gives a false alarm - HGB background voltage abnormal, and we can’t remove it.

Alarm principle: The host queries the pressure in the positive-pressure chamber. The current
pressure is over the upper limit (9KPa) of the attached parameter.

Declaration
20/8/2020, V1.0, For reference only.

Reference Standard
Module Possibility Solution

1.Air pump driver


chip is in the
on-state of
For DH36/
breakdown, and the
DH36X/
air pump keeps
DF50/
working;
DF50CRP

2.Air pump is
damaged

For DH76 N/A N/A


218
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Relative spare part code


Part Name P/N Picture

DH31/DH33/
Mainboard with SD card(without
DH36/ 22.01.0023A
battery)
DH36VET

Mainboard with SD card(without


DH36X 22.01.0199A
battery)

DF50/DF51/
DF52/ Mainboard with SD card(without
22.01.0021C
DF50VET battery)
&DF50CRP

DH31/DH33/
DH36/
DH36VET
Vacuum Pump 23.99.0008A
&DF50/DF51
/DF52/
DF50VET

DH36X
Air Pump 23.99.0012A
&DF50CRP

If you have any problem with this file, please contact our engineer ASAP, thank you!

219
Dymind Global Customer Service Center
Dymind Global Customer Service Center

0x63006120-failed to build positive pressure

Use Background
When analyzer gives a false alarm – Failed to build positive pressure, and we can’t remove it.
Alarm principle: The pressure built for positive pressure chamber is 50% higher or 50% lower than
the target pressure(8KPa)

Declaration
25/3/2020, V1.0, For reference only.

Reference Standard
Module Possibility Solution
D. Check if the tube of yellow part is blocked or folded;
E. Check if LV8, LV 9, pump1 are abnormal. ( it’s better to
replace new one to check)
F. Clean the swab by cleanser. ( we could follow the steps of
SOP to operate)

1.Hydraulic analysis

For C. Check pressure


DH36/ sensor ( replace new
DH31/ one)
DH33/ D. Check mainboard
DH36-VET

2.Hardware analysis

220
Dymind Global Customer Service Center
Dymind Global Customer Service Center

the new analyzer has not been used for a long time, the
ambient temperature in same places is too high, the
membrane of pump is too dry to work abnormally
(as following figure)

For
DH36/
3.special situation How to solve it: we just back flash the pump1 by diluent .(as
DH31/
for new analyzer following figure)
DH33/
DH36-VET

C. Check if the tube of yellow part is blocked or folded;


D. Check if pump2 are abnormal. ( it’s better to replace new
one to check)

1.Hydraulic analysis

For
DH36x
A. Check pressure sensor ( replace new one)
B. Check mainboard

2.Hardware analysis

221
Dymind Global Customer Service Center
Dymind Global Customer Service Center

D. Check if the tube of yellow part is blocked or folded;


E. Check if LV15, LV 16, pump1 are abnormal. ( it’s better to
replace new one to check)
F. Clean the swab by cleanser. ( we could follow the steps of
SOP to operate)
1.Hydraulic analysis

For
DF50/ C. Check pressure
DF51/ sensor ( replace new
DF52/ one)
DF50VET D. Check mainboard

2.Hardware analysis

3. special situation Same as DH36


for new analyzer

C. Check if the tube of red part is blocked or folded;


D. Check if pump2 are abnormal. ( it’s better to replace new
one to check)

For
1.Hydraulic analysis
DF50CRP

222
Dymind Global Customer Service Center
Dymind Global Customer Service Center

C. Check pressure
sensor ( replace new
one)
D. Check mainboard

For
2.Hardware analysis
DF50CRP

1.Check if the tube


of yellow part is
blocked or folded;

2.Check if pump3
For (air pump) are
DH56/ abnormal. ( it’s
D5-CRP better to replace
DH76/ new one to check)
D7-CRP
3.replace air filter

4.check if the
positive pressure
sensor is normal.

223
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Relative spare part code


Part Name P/N Picture

mainboard 22.01.0023A

Liquid detective board 22.01.0022B

DH31/DH33/DH36/
3-way valve(S)-(SMC) 23.99.0004A
DH36-VET

Vacuum Pump 23.99.0008A

Probe Wipe Assembly 68.02.0020A

Mainboard with SD card


22.01.0199A
(without battery)

DH36X

Reagent Detecting Board 22.01.0204A

224
Dymind Global Customer Service Center
Dymind Global Customer Service Center

DH36X Air pump 24.01.0002A

Main Board 22.01.0021C

Liquid Detection Board 22.01.0022B

DF50/DF51/DF52/
3-way valve(S)-(SMC) 23.99.0004A
DF50VET

Vacuum Pump 23.99.0008A

Probe Wipe Assembly 68.02.0020A

Mainboard with SD card


DF50CRP 22.01.0021C
(without battery)

225
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Liquid Detection Board 20.01.0006B

DF50CRP

Air pump 24.01.0002A

Liquid Detecting Board 20.01.0006B

DH56/ DH76/
D5-CRP/
D7-CRP Liquid Detecting Board 20.01.0024B

Air pump 24.01.0002A

If you have any problem with this file, please contact our engineer ASAP, thank you!

226
Dymind Global Customer Service Center
Dymind Global Customer Service Center

0x63016030- vacuum pressure abnormal (high)

Use Background
When analyzer gives a false alarm – vacuum pressure abnormal(high), and we can’t remove it.
Alarm principle: The host queries the pressure in the negative pressure chamber. The current
pressure is higher than the lower limit (-25KPa) of the attached parameter.

Declaration
25/3/2020, V1.0, For reference only.

Reference Standard
Module Possibility Solution
A. Check if the tube of yellow part is blocked or folded;
B. Check if LV8, LV 9, pump1 are abnormal. ( it’s better to
replace new one to check)

1.Hydraulic analysis

For
DH36/
A. Check pressure sensor
DH31/
( replace new one)
DH33/
B. Check mainboard
DH36-VET

2.Hardware analysis

A. Check if the tube of red part is blocked or folded;


For
1.Hydraulic analysis B. Check if LV18, pump1 are abnormal. ( it’s better to
DH36X
replace new one to check)

227
Dymind Global Customer Service Center
Dymind Global Customer Service Center

A. Check pressure sensor ( replace new one)


B. Check mainboard

For
2.Hardware analysis
DH36X

A. Check if the tube of yellow part is blocked or folded;


B. Check if LV15, LV 16, pump1 are abnormal. ( it’s better to
replace new one to check)

1.Hydraulic analysis

For
DF50/
DF51/ A. Check pressure sensor
DF52/ ( replace new one)
DF50VET B. Check mainboard

2.Hardware analysis

E. Check if the tube of red part is blocked or folded;


For
1.Hydraulic analysis F. Check if LV25, pump1 are abnormal. ( it’s better to
DF50CRP
replace new one to check)

228
Dymind Global Customer Service Center
Dymind Global Customer Service Center

E. Check pressure sensor


( replace new one)
F. Check mainboard

For
2.Hardware analysis
DF50CRP

1.Check if the tube


of yellow part is
blocked or folded;

For 2.Check if pump2


DH56/ (liquid pump) are
D5-CRP abnormal. ( it’s
DH76/ better to replace
D7-CRP new one to check)

3.check if the
positive pressure
sensor is normal.

229
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Relative spare part code


Part Name P/N Picture

mainboard 22.01.0023A

Liquid detective board 22.01.0022B

DH31/DH33/DH36/
DH36-VET

3-way valve(S)-(SMC) 23.99.0004A

Vacuum Pump 23.99.0008A

Mainboard with SD card


22.01.0199A
(without battery)

DH36X
Reagent Detecting Board 22.01.0204A

Vacuum Pump 23.99.0008A

230
Dymind Global Customer Service Center
Dymind Global Customer Service Center

DH36X 3-way valve(S)-(SMC) 23.99.0004A

Main Board 22.01.0021C

Liquid Detection Board 22.01.0022B

DF50/DF51/DF52/
DF50VET

3-way valve(S)-(SMC) 23.99.0004A

Vacuum Pump 23.99.0008A

Mainboard with SD card


22.01.0021C
(without battery)

DF50CRP

Liquid Detection Board 20.01.0006B

231
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Vacuum Pump 23.99.0008A

DF50CRP 3-way valve(S)-(SMC) 23.99.0004A

Liquid Detecting Board 20.01.0006B

Liquid Detecting Board 20.01.0024B


DH56/ DH76/
D5-CRP/
D7-CRP

Vacuum Pump 23.99.0008A

2-way valve(S)-(SMC) 23.99.0003A

If you have any problem with this file, please contact our engineer ASAP, thank you!

232
Dymind Global Customer Service Center
Dymind Global Customer Service Center

0x63016040- vacuum pressure abnormal (low)

Use Background
When analyzer gives a false alarm – vacuum pressure abnormal(low), and we can’t remove it.
Alarm principle: The host queries the pressure in the negative pressure chamber. The current
pressure is less than the lower limit (-30KPa) of the attached parameter.

Declaration
25/3/2020, V1.0, For reference only.

Reference Standard
Module Possibility Solution

1.negative-pressure
pump driver chip is
in the on-state of
breakdown, and the
For DH36/ negative pressure
DH36X/ pump keeps
DF50/ working;
DF50CRP
2.Pump is damaged

3.The mainboard is
damaged

233
Dymind Global Customer Service Center
Dymind Global Customer Service Center

For DH76 N/A N/A

Relative spare part code


Part Name P/N Picture
DH31/DH33/DH36/D mainboard 22.01.0023A
H36VET

DH36X mainboard 22.01.0199A

DF50/DF51/DF52/DF Main Board 22.01.0021C


50VET

DF50CRP Reagent Detecting Board 22.01.0204A

DH31/DH33/DH36/D Vacuum Pump 23.99.0008A


H36VET,
DF50/DF51/DF52/DF
50VET

If you have any problem with this file, please contact our engineer ASAP, thank you!

234
Dymind Global Customer Service Center
Dymind Global Customer Service Center

0x63016120- Failed to build vacuum pressure

Use Background
When analyzer gives a false alarm – Failed to build vacuum pressure, and we can’t remove it.
Alarm principle: The pressure built for negative-pressure chamber is 50% higher or 50% lower than
the target pressure (-30KPa)

Declaration
25/3/2020, V1.0, For reference only.

Reference Standard
Module Possibility Solution
C. Check if the tube of yellow part is blocked or folded;
D. Check if LV8, LV 9, pump1 are abnormal. ( it’s better to
replace new one to check)

1.Hydraulic analysis

For
DH36/
C. Check pressure sensor
DH31/
( replace new one)
DH33/
D. Check mainboard
DH36-VET

2.Hardware analysis

C. Check if the tube of red part is blocked or folded;


For
1.Hydraulic analysis D. Check if LV18, pump1 are abnormal. ( it’s better to
DH36X
replace new one to check)

235
Dymind Global Customer Service Center
Dymind Global Customer Service Center

A. Check pressure sensor ( replace new one)


B. Check mainboard

For
2.Hardware analysis
DH36X

C. Check if the tube of yellow part is blocked or folded;


D. Check if LV15, LV 16, pump1 are abnormal. ( it’s better to
replace new one to check)

1.Hydraulic analysis

For
DF50/
DF51/ C. Check pressure sensor
DF52/ ( replace new one)
DF50VET D. Check mainboard

2.Hardware analysis

G. Check if the tube of red part is blocked or folded;


For
1.Hydraulic analysis H. Check if LV25, pump1 are abnormal. ( it’s better to
DF50CRP
replace new one to check)

236
Dymind Global Customer Service Center
Dymind Global Customer Service Center

G. Check pressure sensor


( replace new one)
H. Check mainboard

For
2.Hardware analysis
DF50CRP

1.Check if the tube


of yellow part is
blocked or folded;

For 2.Check if pump2


DH56/ (liquid pump) are
D5-CRP abnormal. ( it’s
DH76/ better to replace
D7-CRP new one to check)

3.check if the
positive pressure
sensor is normal.

237
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Relative spare part code


Part Name P/N Picture

mainboard 22.01.0023A

Liquid detective board 22.01.0022B

DH31/DH33/DH36/
DH36-VET

3-way valve(S)-(SMC) 23.99.0004A

Vacuum Pump 23.99.0008A

Mainboard with SD card


22.01.0199A
(without battery)

DH36X
Reagent Detecting Board 22.01.0204A

Vacuum Pump 23.99.0008A

238
Dymind Global Customer Service Center
Dymind Global Customer Service Center

DH36X 3-way valve(S)-(SMC) 23.99.0004A

Main Board 22.01.0021C

Liquid Detection Board 22.01.0022B

DF50/DF51/DF52/
DF50VET

3-way valve(S)-(SMC) 23.99.0004A

Vacuum Pump 23.99.0008A

Mainboard with SD card


22.01.0021C
(without battery)

DF50CRP

Liquid Detection Board 20.01.0006B

239
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Vacuum Pump 23.99.0008A

DF50CRP 3-way valve(S)-(SMC) 23.99.0004A

Liquid Detecting Board 20.01.0006B

Liquid Detecting Board 20.01.0024B


DH56/ DH76/
D5-CRP/
D7-CRP

Vacuum Pump 23.99.0008A

2-way valve(S)-(SMC) 23.99.0003A

If you have any problem with this file, please contact our engineer ASAP, thank you!

240
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Reagent running out


Problem

241
Dymind Global Customer Service Center
Dymind Global Customer Service Center

The instruction of how to check and adjust Reagent Detecting Board

Use Background
 When we need to check if the voltage of Reagent detecting board abnormal,
which may lead to the machine alarm ”Lyse running out or air bubbles in inlet tubing” .
 When we install the new Reagent Detecting Board and double check the voltage.

Version Description
It’s the first version of this file, which is published on Feb 10th ,2020 by Dymind service
team.
The instruction of how to check and adjust Reagent Detecting Board-V1.0
Reference Standard
1. Type one

Model Slot Tes Adjus Indicato Standard


t Point t point r
1101/110 LH R12 VR2 D3 1.5±
2 0.05V
1103/110 DIFF_I R18 VR3 D1 1.5±
5 0.05V
DIFF_I R6 VR1 D2 1.5±
I 0.05V
242
Dymind Global Customer Service Center
Dymind Global Customer Service Center

2. Type two

Model Slot Tes Adjus Indicato Standard


t Point t point r
1001/1003 LYE-1 TP3 VR2 D3 1.5±0.05V
1104/1108 LYC-1 TP3 VR2 D3 1.5±0.05V
LYC-2 TP1 VR3 D1 1.5±0.05V

243
Dymind Global Customer Service Center
Dymind Global Customer Service Center

3. Type three

Model Slot Tes Adjus Indicato Standard


t Point t point r
1101/110 LYA-3 TP1 VR3 D1 1.5±
2 0.05V
1103/110 LYA-2 TP2 VR1 D2 1.5±
5 0.05V
1107 LYA-1 TP3 VR2 D3 1.5±
0.05V
1102/110 R1 TP7 VR6 D6 2.0±0.1V
7 R2 TP6 VR4 D4 2.0±0.1V
1109 1109R TP4 VR5 D5 1.5±0.05V
1
1109R TP6 VR4 D4 1.5±0.05V
2
LYD-1 TP5 VR7 D7 1.5±0.05V
LYD-2 TP7 VR6 D6 1.5±0.05V

244
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Measurement Method
1. Drain the lyse inside the tube or insert an empty tube inside the lot, which
should be the same type.
2. Measure the voltage of the certain lyse according to the test point mentioned
above by multimeter like below.

Place one pen on the test point and the other on the screw or metal cover. Then check if the
voltage meet the requirement.
3. If the voltage doesn’t meet the requirement, remove the glue on the adjust
point if there is, then adjust it by adjusting the corresponding adjust point introduced
above.

Normally, the indicator would be always green if it’s filled with lyse. If the tube is filled with
bubbles, the indicator would be flash. If the tube is empty, the indicator would be off.

245
Dymind Global Customer Service Center
Dymind Global Customer Service Center

0x64017040-Diluent running out

Use Background
When analyzer gives a false alarm - Diluent running out, but container have enough.
Mechanism: Diluent float switch detects no reagent.

Declaration
16/7/2020, V1.0, For reference only.

Reference Standard

1.Shake the diluent box.


(Prevent bad sensor contact)

2.Take out the diluent cap assembly then move the


sensor at the bottom more times.

For all
modules
3.Reconnect the diluent cap assembly connector.

4.Change the diluent cap assembly.

5.If you don’t have the diluent assembly, you can use a
clip just for emergency! Don't forget to replace a new
one otherwise it will cause a really bad problem.

If you have any problem with this file, please contact our engineer ASAP, thank you!
246
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Additional information

First of all, this part broken won’t generate alarm due to


the mechanism of diluent sensor, but sometimes the
tube from diluent to valve has lots of bubbles because
this part broken and it will effect our result eventually, a
but after changing valve、syringe、even tubes still can’t
fix this problem, if you meet this problem, please better
consider this part. b

Relative spare part code


Part Name P/N Picture

For all human analyzers Diluent Cap Assembly 20L 20.01.0037A

For all veterinary analyzers Diluent Cap Assembly 10L 20.01.0454A

LNS-1 Locking bolt for panel


For all modules 53.05.0023A
installation

247
Dymind Global Customer Service Center
Dymind Global Customer Service Center

CCLR-1 Color-coded locking


For all modules 53.05.0028A
ring

For all modules Integrated locking ring 53.05.0022A

248
Dymind Global Customer Service Center
Dymind Global Customer Service Center

0x64047040-Lyse-1 running out or air bubbles in inlet tubing

Use Background
When analyzer gives a false alarm: Lyse-1 running out or air bubbles in inlet tubing.
Mechanism: The time with empty reagent in the specified time period of time sequence testing
exceeds the specified threshold value by liquid detective board.

Declaration
17/8/2020, V1.0, For reference only.

Reference Standard

1.The lid of the bottle has been screwed on


too tightly or the connector broken.

2.Too many bubbles detected by liquid


detective board due to valve3 leakage or
syringe broken.

DH26

3.The tube from valve3 to lyse bottle or


the tube from valve3 to syringe broken or
leakage.

4.The bubbles was not too much, less than


the normal values. But machine still gived a
error, so the detective module was broken.

249
Dymind Global Customer Service Center
Dymind Global Customer Service Center

DH26↑ DH31/DH33/DH36/DH36VET↓

1.The lid of the bottle has been screwed on


too tightly or the connector broken.

2.Too many bubbles detected by liquid


detective board due to valve3 leakage or
syringe broken.

DH31/DH33/
DH36/DH36V
ET

3.The tube from valve3 to lyse bottle or


the tube from valve3 to syringe broken or
leakage.

a.Readjust the liquid detective board.


Please refer to the instruction of how to
check and adjust Reagent Detecting Board
4.The bubbles was not too much, less than
the normal values. But machine still gived a
error, so the detective board is too
sensitive, or the detective sensor was
b.Change the
broken.
detective board.

250
Dymind Global Customer Service Center
Dymind Global Customer Service Center

DH31/DH33/DH36/DH36VET↑ DH36X↓

1.The lid of the bottle has been screwed on


too tightly or the connector broken.

2.Too many bubbles detected by liquid


detective board due to valve2 leakage or
syringe broken.

DH36X

3.The tube from valve2 to lyse bottle or


the tube from valve2 to syringe broken or
leakage.

4.The bubbles was not too much, less than


the normal values. But machine still gived a
error, so the detective module was broken.

251
Dymind Global Customer Service Center
Dymind Global Customer Service Center

5.In old machine, if the tube from Lyse to


valve has lots of bubbles, it perhaps are
these part broken and it will cause the
alarm. If you can’t fix this problem after
a
changing valve 、 syringe 、 even tubes,
please better consider these parts.

DH36X↑ DF50/DF51/DF52/DF50VET↓

1.The lid of the bottle has been screwed on


too tightly or the connector broken.

DF50/DF51/D
F52/DF50VET
2.Too many bubbles detected by liquid
detective board due to valve2 leakage or
syringe broken.

3.The tube from valve2 to lyse bottle or


the tube from valve2 to syringe broken or
leakage.

252
Dymind Global Customer Service Center
Dymind Global Customer Service Center

a.Readjust the liquid detective board.


Please refer to the instruction of how to
check and adjust Reagent Detecting Board
4.The bubbles was not too much, less than
the normal values. But machine still gived a
error, so the detective board is too
sensitive, or the detective sensor was b.Change the
broken. detective board.

DF50/DF51/DF52/DF50VET↑ DF50CRP↓

1.The lid of the bottle has been screwed on


too tightly or the connector broken.

DF50CRP

2.Too many bubbles detected by liquid


detective board due to valve2 leakage or
syringe broken.

253
Dymind Global Customer Service Center
Dymind Global Customer Service Center

3.The tube from valve2 to lyse bottle or


the tube from valve2 to syringe broken or
leakage.

4.The bubbles was not too much, less than


the normal values. But machine still gived a
error, so the detective module was broken.

DF50CRP↑ DH73/DH76↓

1.The lid of the bottle has been screwed on


too tightly or the connector broken.

DH73/DH76

2.Too many bubbles detected by liquid


detective board due to valve3 leakage or
syringe broken.

254
Dymind Global Customer Service Center
Dymind Global Customer Service Center

3.The tube from valve3 to lyse bottle or


the tube from valve3 to syringe broken or
leakage.

a.Readjust the liquid detective board.


Please refer to the instruction of how to
check and adjust Reagent Detecting Board
4.The bubbles was not too much, less than
the normal values. But machine still gived a
error, so the detective board is too
sensitive, or the detective sensor was b.Change the
broken. detective board.

5.In old machine, if the tube from Lyse to


valve has lots of bubbles, it perhaps are
these part broken and it will cause the
alarm. If you can’t fix this problem after
a
changing valve 、 syringe 、 even tubes,
please better consider these parts.

DH73/DH76↑ D5-CRP↓

255
Dymind Global Customer Service Center
Dymind Global Customer Service Center

1.The lid of the bottle has been screwed on


too tightly or the connector broken.

2.Too many bubbles detected by liquid


detective board due to valve3 leakage or
syringe broken.

D5-CRP

3.The tube from valve3 to lyse bottle or


the tube from valve3 to syringe broken or
leakage.

a.Readjust the liquid detective board.


Please refer to the instruction of how to
check and adjust Reagent Detecting Board
4.The bubbles was not too much, less than
the normal values. But machine still gived a
error, so the detective board is too
sensitive, or the detective sensor was b.Change the
broken. detective board.

256
Dymind Global Customer Service Center
Dymind Global Customer Service Center

5.In old machine, if the tube from Lyse to


valve has lots of bubbles, it perhaps are
these part broken and it will cause the
alarm. If you can’t fix this problem after
changing valve 、 syringe 、 even tubes,
a
please better consider these parts.

D5-CRP↑ D7-CRP↓

1.The lid of the bottle has been screwed on


too tightly or the connector broken.

D7-CRP

2.Too many bubbles detected by liquid


detective board due to valve3 leakage or
syringe broken.

257
Dymind Global Customer Service Center
Dymind Global Customer Service Center

3.The tube from valve3 to lyse bottle or


the tube from valve3 to syringe broken or
leakage.

a.Readjust the liquid detective board.


Please refer to the instruction of how to
check and adjust Reagent Detecting Board
4.The bubbles was not too much, less than
the normal values. But machine still gived a
error, so the detective board is too
sensitive, or the detective sensor was b.Change the
broken. detective board.

5.In old machine, if the tube from Lyse to


valve has lots of bubbles, it perhaps are
these part broken and it will cause the
alarm. If you can’t fix this problem after
changing valve 、 syringe 、 even tubes, a
please better consider these parts.

If you have any problem with this file, please contact our engineer ASAP, thank you!

258
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Relative spare part code


Part Name P/N Picture

3-way valve(S)-(SMC) 23.99.0004A

3-way valve(B)-(SMC) 23.99.0002A

LNS-4 Locking bolt for panel


53.05.002A
installation

For all modules

Integrated locking ring 53.05.0022A

CCLR-4 Color-coded locking


53.05.003A
ring

Bimetallic Tubing 63.01.0006A

Syringe Assembly with


DH26 20.01.0070A
Motor

259
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Lyses Detection Sensor 43.11.0010A

Lyses Detection Sensor


60.01.0109A
Fixed block

Syringe Assembly with Motor 20.01.0070A

DH36

Liquid Detecting Board 22.01.0022B

1004 Lyse Syringe


20.01.0840A
Assembly(10ml & 2.5ml)

DH36X Lyses Detection Sensor 43.11.0010A

Lyses Detection Sensor


60.01.0109A
Fixed block

DF50 Syringe assembly with motor 20.01.0124A

260
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Liquid Detecting Board 22.01.0022B

DF50CRP Lyse Syringe Assembly 20.01.0140A

Lyse Syringe Assembly(2.5ml) 20.01.0040A

DH76

Liquid Detecting Board 20.01.0024B

Lyse Syringe
20.01.0040A
Assembly(2.5ml)

D5-CRP/D7-CRP

Liquid Detecting Board 20.01.0024B

261
Dymind Global Customer Service Center
Dymind Global Customer Service Center

0x64037040-Lyse-2 running out or air bubbles in inlet tubing

Use Background
When analyzer gives a false alarm: Lyse-2 running out or air bubbles in inlet tubing.
Mechanism: The time with empty reagent in the specified time period of time sequence testing
exceeds the specified threshold value by liquid detective board.

Declaration
17/8/2020, V1.0, For reference only.

Reference Standard

1.The lid of the bottle has been screwed on


too tightly or the connector broken.

2.Too many bubbles detected by liquid


detective board due to valve3 leakage or
syringe broken.
DF50/DF51/D
F52/DF50VET

3.The tube from valve3 to lyse bottle or the


tube from valve3 to syringe broken or
leakage.

262
Dymind Global Customer Service Center
Dymind Global Customer Service Center

a.Readjust the liquid detective board.


Please refer to the instruction of how to
check and adjust Reagent Detecting Board
4.The bubbles was not too much, less than
the normal values. But machine still gived a
error, so the detective board is too
sensitive, or the detective sensor was b.Change the
broken. detective board.

DF50/DF51/DF52/DF50VET↑ DF50CRP↓

1.The lid of the bottle has been screwed on


too tightly or the connector broken.

2.Too many bubbles detected by liquid


detective board due to valve3 leakage or
syringe broken.
DF50CRP

3.The tube from valve3 to lyse bottle or


the tube from valve3 to syringe broken or
leakage.

263
Dymind Global Customer Service Center
Dymind Global Customer Service Center

4.The bubbles was not too much, less than


the normal values. But machine still gived a
error, so the detective module was broken.

5.In old machine, if the tube from Lyse to


valve has lots of bubbles, it perhaps are
these part broken and it will cause the
alarm. If you can’t fix this problem after
changing valve 、 syringe 、 even tubes,
please better consider these parts.
a

DF50CRP↑ DH73/DH76↓

1.The lid of the bottle has been screwed on


DH73/DH76
too tightly or the connector broken.

264
Dymind Global Customer Service Center
Dymind Global Customer Service Center

2.Too many bubbles detected by liquid


detective board due to valve2 leakage or
syringe broken.

3.The tube from valve2 to lyse bottle or


the tube from valve2 to syringe broken or
leakage.

a.Readjust the liquid detective board.


Please refer to the instruction of how to
check and adjust Reagent Detecting Board
4.The bubbles was not too much, less than
the normal values. But machine still gived a
error, so the detective board is too
sensitive, or the detective sensor was b.Change the
broken. detective board.

5.In old machine, if the tube from Lyse to


valve has lots of bubbles, it perhaps are
these part broken and it will cause the
alarm. If you can’t fix this problem after a
changing valve 、 syringe 、 even tubes,
please better consider these parts.
b

DH73/DH76↑ D5-CRP↓

265
Dymind Global Customer Service Center
Dymind Global Customer Service Center

1.The lid of the bottle has been screwed on


too tightly or the connector broken.

2.Too many bubbles detected by liquid


detective board due to valve2 leakage or
syringe broken.

D5-CRP

3.The tube from valve2 to lyse bottle or


the tube from valve2 to syringe broken or
leakage.

a.Readjust the liquid detective board.


Please refer to the instruction of how to
check and adjust Reagent Detecting Board
4.The bubbles was not too much, less than
the normal values. But machine still gived a
error, so the detective board is too
sensitive, or the detective sensor was b.Change the
broken. detective board.

266
Dymind Global Customer Service Center
Dymind Global Customer Service Center

5.In old machine, if the tube from Lyse to


valve has lots of bubbles, it perhaps are
these part broken and it will cause the
alarm. If you can’t fix this problem after a
changing valve 、 syringe 、 even tubes,
please better consider these parts.
b

D5-CRP↑ D7-CRP↓

1.The lid of the bottle has been screwed on


too tightly or the connector broken.

D7-CRP

2.Too many bubbles detected by liquid


detective board due to valve2 leakage or
syringe broken.

267
Dymind Global Customer Service Center
Dymind Global Customer Service Center

3.The tube from valve2 to lyse bottle or


the tube from valve2 to syringe broken or
leakage.

a.Readjust the liquid detective board.


Please refer to the instruction of how to
check and adjust Reagent Detecting Board
4.The bubbles was not too much, less than
the normal values. But machine still gived a
error, so the detective board is too
sensitive, or the detective sensor was b.Change the
broken. detective board.

5.In old machine, if the tube from Lyse to


valve has lots of bubbles, it perhaps are
these part broken and it will cause the
alarm. If you can’t fix this problem after
changing valve 、 syringe 、 even tubes,
please better consider these parts. a

268
Dymind Global Customer Service Center
Dymind Global Customer Service Center

If you have any problem with this file, please contact our engineer ASAP, thank you!

Relative spare part code


Part Name P/N Picture

3-way valve(S)-(SMC) 23.99.0004A

3-way valve(B)-(SMC) 23.99.0002A

LNS-5 Locking bolt for panel


53.05.0027A
installation

For all modules Integrated locking ring 53.05.0022A

CCLR-5 Color-coded locking 553.05.0032A


ring

LNS-2 Locking bolt for panel


53.05.0024A
installation

CCLR-2 Color-coded locking


53.05.0029A
ring

269
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Bimetallic Tubing 63.01.0006A

Syringe assembly with motor 20.01.0124A

DF50

Liquid Detecting Board 22.01.0022B

DF50CRP Lyse Syringe Assembly 20.01.0140A

Lyse Syringe Assembly(2.5ml) 20.01.0040A

DH76

Liquid Detecting Board 20.01.0024B

Lyse Syringe
20.01.0040A
Assembly(2.5ml)

D5-CRP/D7-CRP

Liquid Detecting Board 20.01.0024B

270
Dymind Global Customer Service Center
Dymind Global Customer Service Center

0x64027040-Lyse-3 running out or air bubbles in inlet tubing

Use Background
When analyzer gives a false alarm: Lyse-3 running out or air bubbles in inlet tubing.
Mechanism: The time with empty reagent in the specified time period of time sequence testing
exceeds the specified threshold value by liquid detective board.

Declaration
18/8/2020, V1.0, For reference only.

Reference Standard

1.The lid of the bottle has been screwed on


too tightly or the connector broken.

2.Too many bubbles detected by liquid


detective board due to valve1 leakage or
syringe broken.

DH73/DH76
3.The tube from valve1 to lyse bottle or
the tube from valve1 to syringe broken or
leakage.

a.Readjust the liquid detective board.


Please refer to the instruction of how to
check and adjust Reagent Detecting Board
4.The bubbles was not too much, less than
the normal values. But machine still gived a
error, so the detective board is too
sensitive, or the detective sensor was b.Change the
broken. detective board.

271
Dymind Global Customer Service Center
Dymind Global Customer Service Center

5.In old machine, if the tube from Lyse to


valve has lots of bubbles, it perhaps are
these part broken and it will cause the
alarm. If you can’t fix this problem after a
changing valve 、 syringe 、 even tubes,
please better consider these parts.
b

DH73/DH76↑ D5-CRP↓

1.The lid of the bottle has been screwed on


too tightly or the connector broken.

2.Too many bubbles detected by liquid


D5-CRP detective board due to valve1 leakage or
syringe broken.

3.The tube from valve1 to lyse bottle or


the tube from valve1 to syringe broken or
leakage.

272
Dymind Global Customer Service Center
Dymind Global Customer Service Center

a.Readjust the liquid detective board.


Please refer to the instruction of how to
check and adjust Reagent Detecting Board
4.The bubbles was not too much, less than
the normal values. But machine still gived a
error, so the detective board is too
sensitive, or the detective sensor was b.Change the
broken. detective board.

5.In old machine, if the tube from Lyse to


valve has lots of bubbles, it perhaps are
these part broken and it will cause the
alarm. If you can’t fix this problem after
changing valve 、 syringe 、 even tubes,
please better consider these parts. a

D5-CRP↑ D7-CRP↓

1.The lid of the bottle has been screwed on


D7-CRP
too tightly or the connector broken.

273
Dymind Global Customer Service Center
Dymind Global Customer Service Center

2.Too many bubbles detected by liquid


detective board due to valve1 leakage or
syringe broken.

3.The tube from valve1 to lyse bottle or


the tube from valve1 to syringe broken or
leakage.

a.Readjust the liquid detective board.


Please refer to the instruction of how to
check and adjust Reagent Detecting Board
4.The bubbles was not too much, less than
the normal values. But machine still gived a
error, so the detective board is too
sensitive, or the detective sensor was b.Change the
broken. detective board.

5.In old machine, if the tube from Lyse to


valve has lots of bubbles, it perhaps are
these part broken and it will cause the
alarm. If you can’t fix this problem after
a
changing valve、syringe、even tubes,
please better consider these parts.

274
Dymind Global Customer Service Center
Dymind Global Customer Service Center

If you have any problem with this file, please contact our engineer ASAP, thank you!

Relative spare part code


Part Name P/N Picture

3-way valve(S)-(SMC) 23.99.0004A

3-way valve(B)-(SMC) 23.99.0002A

LNS-5 Locking bolt for panel


53.05.0027A
installation

For all modules Integrated locking ring 53.05.0022A

CCLR-5 Color-coded locking 553.05.0032A


ring

LNS-2 Locking bolt for panel


53.05.0024A
installation

CCLR-2 Color-coded locking


53.05.0029A
ring

275
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Bimetallic Tubing 63.01.0006A

Lyse Syringe Assembly(2.5ml) 20.01.0040A

DH76

Liquid Detecting Board 20.01.0024B

Lyse Syringe
20.01.0040A
Assembly(2.5ml)

D5-CRP/D7-CRP

Liquid Detecting Board 20.01.0024B

276
Dymind Global Customer Service Center
Dymind Global Customer Service Center

0x64057040-R1 running out or air bubbles in inlet tubing

Use Background
When analyzer gives a false alarm: R1 running out or air bubbles in inlet tubing.
Mechanism: The time with empty reagent in the specified time period of time sequence testing
exceeds the specified threshold value by liquid detective board.

Declaration
18/8/2020, V1.0, For reference only.

Reference Standard

1.The lid of the bottle has been screwed on


too tightly or the connector broken.

2.Too many bubbles detected by liquid


DH36X detective board due to valve9 leakage or
syringe broken.

3.The tube from valve9 to lyse bottle or


the tube from valve9 to syringe broken or
leakage.

277
Dymind Global Customer Service Center
Dymind Global Customer Service Center

4.The bubbles was not too much, less than


the normal values. But machine still gived a
error, so the detective part may be broken

DH36X↑ DF50CRP↓

1.The lid of the bottle has been screwed on


too tightly or the connector broken.

2.Too many bubbles detected by liquid


detective board due to valve8 leakage or
syringe broken.

DF50CRP

3.The tube from valve8 to lyse bottle or


the tube from valve8 to syringe broken or
leakage.

4.The bubbles was not too much, less than


the normal values. But machine still gived a
error, so the detective part may be broken

DF50CRP↑ D5-CRP↓

278
Dymind Global Customer Service Center
Dymind Global Customer Service Center

1.The lid of the bottle has been screwed on


too tightly or the connector broken.

2.Too many bubbles detected by liquid


detective board due to valve31 leakage or
syringe broken.

D5-CRP

3.The tube from valve31 to lyse bottle or


the tube from valve31 to syringe broken or
leakage.

4.The bubbles was not too much, less than


the normal values. But machine still gived a
error, so the detective part may be broken

D5-CRP↑ D7-CRP↓

279
Dymind Global Customer Service Center
Dymind Global Customer Service Center

1.The lid of the bottle has been screwed on


too tightly or the connector broken.

2.Too many bubbles detected by liquid


detective board due to valve31 leakage or
syringe broken.

D7-CRP

3.The tube from valve31 to lyse bottle or


the tube from valve31 to syringe broken or
leakage.

4.The bubbles was not too much, less than


the normal values. But machine still gived a
error, so the detective part may be broken

280
Dymind Global Customer Service Center
Dymind Global Customer Service Center

5.In old machine, if the tube from Lyse to


valve has lots of bubbles, it perhaps are
these part broken and it will cause the
alarm. If you can’t fix this problem after a
changing valve 、 syringe 、 even tubes,
please better consider these parts.
b

If you have any problem with this file, please contact our engineer ASAP, thank you!

Relative spare part code


Part Name P/N Picture

3-way valve(S)-(SMC) 23.99.0004A

Bimetallic Tubing 63.01.0006A

For all module

1.0mm Teflon Tubing 63.01.0014A

CCLR-3 Color-coded locking


53.05.0030A
ring

281
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Integrated locking ring 53.05.0022A

LNS-3 Locking bolt for panel


53.05.0025A
installation

282
Dymind Global Customer Service Center
Dymind Global Customer Service Center

0x64067040-R2 running out or air bubbles in inlet tubing

Use Background
When analyzer gives a false alarm: R2 running out or air bubbles in inlet tubing.
Mechanism: The time with empty reagent in the specified time period of time sequence testing
exceeds the specified threshold value by liquid detective board.

Declaration
18/8/2020, V1.0, For reference only.

Reference Standard

1.The lid of the bottle has been screwed on


too tightly or the connector broken.

2.Too many bubbles detected by liquid


DH36X detective board due to valve8 leakage or
syringe broken.

3.The tube from valve8 to lyse bottle or


the tube from valve8 to syringe broken or
leakage.

283
Dymind Global Customer Service Center
Dymind Global Customer Service Center

4.The bubbles was not too much, less than


the normal values. But machine still gived a
error, so the detective part may be broken

DH36X↑ DF50CRP↓

1.The lid of the bottle has been screwed on


too tightly or the connector broken.

2.Too many bubbles detected by liquid


detective board due to valve7 leakage or
syringe broken.

DF50CRP

3.The tube from valve7 to lyse bottle or


the tube from valve7 to syringe broken or
leakage.

4.The bubbles was not too much, less than


the normal values. But machine still gived a
error, so the detective part may be broken

284
Dymind Global Customer Service Center
Dymind Global Customer Service Center

DF50CRP↑ D5-CRP↓

1.The lid of the bottle has been screwed on


too tightly or the connector broken.

2.Too many bubbles detected by liquid


detective board due to valve32 leakage or
syringe broken.

D5-CRP

3.The tube from valve32 to lyse bottle or


the tube from valve32 to syringe broken or
leakage.

4.The bubbles was not too much, less than


the normal values. But machine still gived a
error, so the detective part may be broken

D5-CRP↑ D7-CRP↓

285
Dymind Global Customer Service Center
Dymind Global Customer Service Center

1.The lid of the bottle has been screwed on


too tightly or the connector broken.

2.Too many bubbles detected by liquid


detective board due to valve32 leakage or
syringe broken.

D7-CRP

3.The tube from valve32 to lyse bottle or


the tube from valve32 to syringe broken or
leakage.

4.The bubbles was not too much, less than


the normal values. But machine still gived a
error, so the detective part may be broken

If you have any problem with this file, please contact our engineer ASAP, thank you!

286
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Relative spare part code


Part Name P/N Picture

3-way valve(S)-(SMC) 23.99.0004A

For all module Bimetallic Tubing 63.01.0006A

1.0mm Teflon Tubing 63.01.0014A

287
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Black Screen Problem

288
Dymind Global Customer Service Center
Dymind Global Customer Service Center

How to solve the issues related to display

Declaration
2/4/2021 , V1.0, For reference only.

Reference Standard
modules Issue Description Solutions
1.Flower screen A. Check if the cable of the LCD is loose;
B. If the the problem is still existed, please replace LCD
screen.

2.Ground color
abnormal

Display
abnormal

3.Horizontal or
vertical line

Touch 1.No reaction at A. There is static with protective film, please tear it off;
malfunction all B. Program the software to calibrate the touch screen

289
Dymind Global Customer Service Center
Dymind Global Customer Service Center

C. Check if the cable of the touch screen is loose.


D. Replace the touch screen.

1.No reaction at
all

A. Program the software to calibrate the touch screen

Touch
malfunction
2.Just receive the
feedback of shift
button in screen
interface

B.If the problem is still existed, please replace the mainboard.

290
Dymind Global Customer Service Center
Dymind Global Customer Service Center

A. Check if the DIP switch on the mainboard is on, then to


power off the analyer and turn it to NAND and restard

1.Status
indicator is red
Fail to enter
system

B. Check if the cable from mainboard to screen is loose.


C. The mainboard is damaged. (check if D79,D90 and D52
are flashing)

1.Status
indicator is red

Fail to enter
system

A. No power output from the socket and the fuse inside


power switch is blown, please replace it.

2.Status
indicator is off

291
Dymind Global Customer Service Center
Dymind Global Customer Service Center

B. The power supply is damaged, please replace it .


C. The mainboard is completely damaged, please replac the
mainboard.

A. SD card connection is loose, please power off the analyzer


and reconnct it, if the problem is still existed, please
replace the SD card by same version one.
B. The SD card house on the mainboard is damaged, please
replace the mainboard.
3. White screen
is stuck in login
interface.

Fail to enter
system

4. Blank white A.The cable of LCD is loose, please reconnect it


screen with B.The LCD screen is damaged, please replace it and perform
nothing screen calibration.

292
Dymind Global Customer Service Center
Dymind Global Customer Service Center

A. Delete the data manually, here is steps as following:


“service” ----“data cleanup”--- choos the time range and
content---- apply.

5.Stuck with a
short message

If you have any problem with this file, please contact our engineer ASAP, thank you!

Relative spare part code


Part Name P/N Picture

According to
the module
mainboard
of your
analyzer
All modules

LCD Screen 31.02.0001B

293
Dymind Global Customer Service Center
Dymind Global Customer Service Center

LCD Connection Cable 37.02.0045A

Touch Screen 31.11.0004A

294
Dymind Global Customer Service Center
Dymind Global Customer Service Center

The process of
upgrading

295
Dymind Global Customer Service Center
Dymind Global Customer Service Center

The Process of Upgrading

Declaration
31/3/2021 , V1.0, For reference only.

Reference Standard
modules Steps Content
All of the upgrading files are in our FTP acount (if you don’t know
what’s FTP, please conact to our after-sale service engineer.) ;

For DH36/ 1.The routine


DF50/ of how to get
DF50CRP/ the upgrading
DF50VET files

296
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Noted: Don't skip to upgrade, please choose correct files.

1.The routine
of how to get
the upgrading
files

A. Login the machine by “service” authority


User name: service
Password: dm_ser0620

For DH36/
DF50/
DF50CRP/
DF50VET

2.The process
of upgrading

B. Unpack the upgrading file and you wiil get the upgrading
file(XXXXX.tar.bz) and insert the USB disk with the
upgrading fils(XXXXX.tar.bz)

Note: The upgrading file should be kept in the root


directory of the USB disk.not(XXXXX.zip)
297
Dymind Global Customer Service Center
Dymind Global Customer Service Center

C. Come into service interface to click “upgrade” button.

D. Choose the upgrade files as below picture shows, then click


“Yes”.

For DH36/
DF50/ 2.The process
DF50CRP/ of upgrading
DF50VET

E. Choose as below picture shows:

298
Dymind Global Customer Service Center
Dymind Global Customer Service Center

F. Just wait until the upgrading process finished:

For DH36/
DF50/ 2.The process
DF50CRP/ of upgrading
DF50VET

G. Finally restart our analyzer.

All of the upgrading files are in our FTP acount (if you don’t know
what’s FTP, pleace conact to our after-sale service engineer.) ;

For DH76/
1.The routine
DH73
of how to get
DH56/
the upgrading
D5-CRP/
files
D7-CRP

299
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Noted: Don't skip to upgrade, please choose correct files.

1.The routine
of how to get
the upgrading
files

A. Double click file folder “host”, you could find


"upgradestudio.zip", please uncompress "upgradestudio.zip" to
get "upgradestudio.exe", then double click "upgradestudio.exe" to
launch the upgrade tool.
It takes about half a minute for the upgrade tool to connect to the
machine.

For DH76/
DH73
DH56/
D5-CRP/
D7-CRP
B. Click the "Open File" button, then select and open the upgrade
2.Upgrading
file:
the main
"upgrade_1101_vxxxx_XXXXXXXXXXXX.tar" in the popup dialog
system
box.
As below picture shows:

Note: In the name of the upgrade file, "vxxxx" indicates the version
of the upgrade file, and "XXXXXXXXXXXX" indicates the release date
of the upgrade file.

300
Dymind Global Customer Service Center
Dymind Global Customer Service Center

C. Click "Upgrade".
The upgrade confirmation dialog box will pop up.

For DH76/
DH73 2.Upgrading
DH56/ the main
D5-CRP/ system
D7-CRP

D. After it is successfully upgraded, power off the device and then


close theupgrade tool.

301
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Note: When upgrading the user interface, we have to exit the


current version software; otherwise the upgrade will not be
successfully.
A.Double click "setup.exe", then select language at below interface

B. Keep clicking "Next," and then select "Install" in the screen and
follow next steps as shown below.

For DH76/
DH73 3. Upgrading
DH56/ the software
D5-CRP/ interface
D7-CRP

If you have any problem with this file, please contact our engineer ASAP, thank you!
302
Dymind Global Customer Service Center
Dymind Global Customer Service Center

WBC Classification
Problem

303
Dymind Global Customer Service Center
Dymind Global Customer Service Center

WBC Results Abnormal Issue

Use Background
When analyzer gives a false alarm – No error code
Mechanism: The particle ratio of DIFF count/WBC total count is less than 0.5 or more than 2 in
five-part machine if the DIFF results are blocked by machine.

Declaration
20/3/2021, V1.0, For reference only.

Reference Standard
Module Scatter Gram & Histogram Possibilities & Solutions

The impedance testing process is


not smooth while testing.
Maintenance needs to be done.
1. Do cleanser
soaking by machine.
2. Clean WBC
aperture manually.

DH26/DH31/
DH33/DH36/
DH36VET/DH The impedance testing process is
36X not smooth while testing.
Maintenance needs to be done.
Or the gain of WBC get weak.
1. Do cleanser
soaking by machine.
2. Clean WBC
aperture manually.
3. Do gain
calibration by normal
level control.

304
Dymind Global Customer Service Center
Dymind Global Customer Service Center

The lyse LYE-1 doesn’t make any


function during testing process.
1. Wrong lyse is
connected.
2. Lyse is bad,
which need to replace.
3. No or few lyse
get into WBC chamber.
Check the lyse bottle,
lyse tube, valve
no.2(DH36X), no.3(other
models), lyse syringe,
titanium tube above
WBC chamber.
DH26/DH31/DH33/DH36/DH36VET/DH36X↑ DF50/DF51/DF52/DF50VET/DF50CRP↓

It’s control material which tested


in sample analysis interface, the
algorithm between sample and
QC is different. Test control in QC
analysis interface, sample in
sample analysis interface.

DF50/DF51/D
The lyse LYC-1 doesn’t make any
F52/DF50VET
function during testing process.
/DF50CRP
1. Wrong lyse is
connected.
2. LYC-1 is bad,
which need to replace.
3. There’s many
bubbles filled in the lyse
tube or no lyse. Check
the lyse bottle, lyse tube,
valve NO.2, lyse syringe,
titanium tube above
WBC chamber.

305
Dymind Global Customer Service Center
Dymind Global Customer Service Center

This case is the same as above,


WBC total count is very high, HGB
is high. Pay attention to lyse
LYC-1.

The lyse LYC-2 doesn’t make any


function during testing process.
1. Wrong lyse is
connected.
2. LYC-2 is bad,
which need to replace.
3. There’s many
bubbles filled in the lyse
tube or no lyse. Check
the lyse bottle, lyse tube,
valve NO.3, lyse syringe.
The gain of optical unit got weak
or flow-cell got dirty.
1. Clean the
inner-wall and outer-wall
of flow-cell
2. Adjust the
gain value, increase
LS&MS gain value of
both DIFF&Baso.
The gain of optical unit got weak
or flow-cell got dirty.
1. Clean the
inner-wall and outer-wall
of flow-cell
2. Adjust the
gain value, increase
LS&MS gain value of
both DIFF&Baso.

306
Dymind Global Customer Service Center
Dymind Global Customer Service Center

The gain of optical unit got weak


or flow-cell got dirty.
1. Clean the
inner-wall and outer-wall
of flow-cell
2. Adjust the
gain value, increase
LS&MS gain value of
both DIFF&Baso.
3. Replace the
optical unit if the scatter
gram is still the same as
previous.
Do never touch or operate the
optical unit without wearing an
anti-static wrist!!!
The sample got leakage in DIFF
testing channel.
1. Pinch valve
no.12(other models),
no.19(DF50CRP)
2. Valve
no.17(other models),
no.24(DF50CRP)
3. Bubble mixing
valve no.19(other
models), no.12
(DF50CRP)
The sample got leakage in
WBC/BASO testing channel.
1. Pinch valve
no.12(other models),
no.19(DF50CRP)
2. Valve
no.17(other models),
no.24(DF50CRP)
3. Bubble mixing
valve no.19(other
models), no.12
(DF50CRP)

307
Dymind Global Customer Service Center
Dymind Global Customer Service Center

There is small particle in the


scatter gram. The sample got
leakage in DIFF&WBC/BASO
testing channel.
1. Pinch valve
no.12(other models),
no.19(DF50CRP)
2. Valve
no.17(other models),
no.24(DF50CRP)
3. Bubble mixing
valve no.19(other
models), no.12
(DF50CRP)
There is no particle in the scatter
gram, there’s something wrong
with signal receiving or
processing process.
1. The cable of
optical unit loose, we can
shut down the machine
and reconnect all the
cables of optical unit.
2. Replace
another optical unit and
update the gain setting.
Do never touch or operate the
optical unit without wearing an
anti-static wrist!!!
The HS signal got influenced
during testing process.
1. Confirm the
grounding voltage is
under requirement.(≤
5V)
2. Shutdown the
machine and reconnect
the cables of HS signal
board.
3. Replace the
optical unit.
Do never touch or operate the
optical unit without wearing an

308
Dymind Global Customer Service Center
Dymind Global Customer Service Center

anti-static wrist!!!

The signal received by the


amplifier board or HS receiver
board is unstable.
1. We can try to
clean the flow-cell, then
adjust the gain setting
value.
2. Replace the
optical unit if it’s still the
same as previous.
Do never touch or operate the
optical unit without wearing an
anti-static wrist!!!
The signal received by the
amplifier board or HS receiver
board is unstable.
1. We can try to
clean the flow-cell, then
adjust the gain setting
value.
2. Replace the
optical unit if it’s still the
same as previous.
Do never touch or operate the
optical unit without wearing an
anti-static wrist!!!
DF50/DF51/DF52/DF50VET/DF50CRP↑ DH56/DH73(UN73)/DH76/D7CRP↓
Lyse LYA-3 doesn’t make function
during sample preparation
process in DIFF chamber.
1. Wrong lyse is
connected.
DH56/DH73(
2. LYA-3 is bad,
UN73)/DH76/
which need to replace.
D7CRP
3. There’s many
bubbles filled in the lyse
tube or no lyse. Check
the lyse bottle, lyse tube,
connectors on rear cover,

309
Dymind Global Customer Service Center
Dymind Global Customer Service Center

valve no.1, lyse syringe.

Lyses don’t make functions


during testing process.
1. Confirm if
LYA-1 & LYA-3 are
connected wrong.
2. Confirm is
LYA-1 & LYA-3 are bad,
which need need to
replace.
3. Confirm the
lyse adding channel from
bottle, lyse tube,
connectors on rear cover,
corresponding valve, lyse
syringe, tube connected
chambers.
The gain of optical unit got weak
or flow-cell got dirty.
1. Clean the
inner-wall and outer-wall
of flow-cell
2. Adjust the
gain value, increase
LS&MS gain value.
3. Replace the
optical unit if the scatter
gram is still the same as
previous.
Do never touch or operate the
optical unit without wearing an
anti-static wrist!!!

310
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Machine get no signal from WBC


impedance counting process.
1. Clean the
WBC aperture manually
by syringe.
2. Replace
bubble mixing valve
no.15
3. Replace waste
discharging valve no.25
or no.21
4. Replace
counting valve no.27
Some part impact the testing
process.
1. Clean the
inner-wall and outer-wall
of flow-cell, DIFF
chamber.
2. Confirm the
sample preparation
process is ok. Replace
valve no.14, 30, 22, 18,
12, 11 accordingly.
3. Replace the
optical unit.
Do never touch or operate the
optical unit without wearing an
anti-static wrist!!!
There is no particle in the scatter
gram, and something wrong with
signal receiving or processing
process.
1. The cable of
optical unit may loose,
we can shut down the
machine and reconnect
all the cables of optical
unit.
2. Replace
another optical unit and
update the gain setting.
Do never touch or operate the

311
Dymind Global Customer Service Center
Dymind Global Customer Service Center

optical unit without wearing an


anti-static wrist!!!

If you have any problem with this file, please contact our engineer ASAP, thank you!

Relative spare part code


Model Part Name P/N Picture

Three-way valve (S) - SMC 23.99.0004A

All models above

Two-way valve (S) - SMC 23.99.0003A

DF50/DF51/DF52/DF50V
ET,
DF50CRP, Pinch Valve 23.99.0005A
DH56/DH73/DH76,
D5CRP/D7CRP

DH31/DH33/DH36/DH36 LYE-1 lyse connection


68.02.0028A
VET tube(Green)

Lyse-1 Lyse Connection


DH36X 20.01.0036A
Tube(Green)

312
Dymind Global Customer Service Center
Dymind Global Customer Service Center

LYC-1 lyse connection


68.02.0029A
tube(Green)
DF50/DF51/DF52/DF50V
ET,
DF50CRP
LYC-2 lyse connection
68.02.0030A
tube(Blue)

LYA-1 lyse Cap Assembly 20.01.0036A

DH56/DH73/DH76,
LYA-3 lyse Cap Assembly 20.01.0034A
D5CRP/D7CRP

Diluents&Lyses connector /

DF50/DF51/DF52 Optical components 20.01.0187A

DF50VET Optical components 20.01.0347A

DF50CRP Optical components 20.01.0292A

313
Dymind Global Customer Service Center
Dymind Global Customer Service Center

DH56/DH73/DH76 Optical components 20.01.0019A

D5CRP/D7CRP Optical components 20.01.1156A

314
Dymind Global Customer Service Center
Dymind Global Customer Service Center

How to check power


supply

315
Dymind Global Customer Service Center
Dymind Global Customer Service Center

How to check the power supply in DH36/DF50

1. Prepare these items below.

2. Take out the rear door by screw driver and find these positions.

316
Dymind Global Customer Service Center
Dymind Global Customer Service Center

3. Setting the digital multimeter as shown below.

4. Testing picture

5. Last step, please don’t forget to test the grounding line cable!
More information, please refer to our service manual.

317
Dymind Global Customer Service Center
Dymind Global Customer Service Center

How to check the power supply in DH56/DH76/D5-CRP/D7CRP

Declaration
2/4/2021 , V1.0, For reference only.

Reference Standard
1. Prepare these items below.

2. Take out the top cover by screw driver

3. Take out the mainboard and driver board to find front board.

318
Dymind Global Customer Service Center
Dymind Global Customer Service Center

319
Dymind Global Customer Service Center
Dymind Global Customer Service Center

3. Setting the digital multimeter as shown below.

4. Testing picture

5. Last step, please don’t forget to test the grounding line cable!
More information, please refer to our service manual.
If you have any problem with this file, please contact our engineer ASAP, thank you!

320
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Counter trigger
abnormal

321
Dymind Global Customer Service Center
Dymind Global Customer Service Center

0x1128D210-Counter trigger abnormal

Use Background
When analyzer gives a false alarm - Counter trigger abnormal

Mechanism: Failed to complete the action within the time.

Declaration
9/4/2021, V2.0, For reference only.

Reference Standard
Model Possibility Solution

A. Remove the base panel cover

B. Remove those 8 screws and 2 cables to remove the auto-load


tray

1. The
UN73/DH76/
counter trigger
D7CRP
is dirty or rusty

C. Remove those 5 screws, the rack fixing plate and the loading
motor cable.

322
Dymind Global Customer Service Center
Dymind Global Customer Service Center

D. Lift the plate and remove the two screws beneath the plate to
release the auto loading feeding assembly.

E. Please remove the 2 screws for fixing the belt.

F. Take out the auto-loading feeding assembly and remove 2 screws


to take out the 2 counter triggers.

323
Dymind Global Customer Service Center
Dymind Global Customer Service Center

G. Check if the following screws on counter trigger get rusty or


dirty. If it is, please clean it or replace a new one

H. Check if the following screw is extended too long. If it is, please


loosen the screw to make the extended part backward.

324
Dymind Global Customer Service Center
Dymind Global Customer Service Center

I.Install back the plate and screws. Be careful about the trigger
position.

J. For a correct assembly of the rack fixing plate, please use a rack
as the picture shows and then adjust the screws.

K. Finally, you can check if the problem was solved using the Auto
Venous Whole Blood mode with 4 empty racks.
UN73/DH76/ 2. The A. Replace the auto-loading feeding assembly.

325
Dymind Global Customer Service Center
Dymind Global Customer Service Center

D7CRP auto-loading
feeding
assembly is
damaged.

Relative spare part code


Model Part Name P/N Picture

Auto Loading Feeding


UN73/DH76/ D7CRP 20.01.0079A
Assembly

If you have any problem with this file, please contact our engineer ASAP, thank you!

326
Dymind Global Customer Service Center
Dymind Global Customer Service Center

How to connect DMS

327
Dymind Global Customer Service Center
Dymind Global Customer Service Center

Connect Analyzer to the Data Management Software(Dymind LIS)

Notice
If you don’t have DMS, Pls download it from our FTP server with GENERAL ACCOUNT in DMS folder.

DYMIND General

Website http://159.138.130.90/
User name G-user01
Password n6m2K1qr

Declaration
1) 03/02/2021, V1.0, For reference only.
2) This DMS Software can only used for transferring date from analyzer to PC or Laptop, it can’t do the
bidirectional LIS function, please seek the guideline of engineer in third-part LIS field if you want to
implement this function.
3) If you need LIS protocol, please download from the link above in LIS folder.

Reference Standard
1.Connect the Analyzer to PC through LAN Port.

2.Find the network setting of our analyzer in ethernet and close all firewall and anti-virus software!

328
Dymind Global Customer Service Center
Dymind Global Customer Service Center

3.Set the IP address of the PC as below. The IP address of the PC is set to 10.0.0.102 from Internet
Protocol Version 4

4.Go to Setup menu of the analyzer to do Host Communication and LIS Communication setting.

5.Open the DMS software and set the 5600 in LIS communication and LIS Server.

329
Dymind Global Customer Service Center
Dymind Global Customer Service Center

6.Run cmd to check the connection between the analyzer and the PC.

a)Press combination key [Windows+R] to open the Run window, and input cmd

B) b) Input ping 10.0.0.101 command


As is shown below, analyzer software can be connected to the main unit normally

As is shown below, analyzer software fails to be connected to the main unit.

7.Check the LIS connection status icon.


If the connection is ok, the LIS connection status icon and Comm icon will become active.

330
Dymind Global Customer Service Center
Dymind Global Customer Service Center

8.Click Comm to send the report data from analyzer to LIS Client Software.

If you have any problem with this file, please contact our engineer ASAP, thank you!

331
Dymind Global Customer Service Center