Download as pdf or txt
Download as pdf or txt
You are on page 1of 3

6211

Methane
Approved by Standard Methods Committee, 2010. Editorial revisions, 2020.

6211 A. Introduction

1. Occurrence and Significance Methane also is produced from wastewater and may be
present in sewers and wastewater treatment plants (see Section
Methane (CH4) is a colorless, odorless, tasteless combustible 2720).
gas occasionally found in groundwaters. Escape of this gas from
water may cause an explosive atmosphere not only in a utility’s 2. Selection of Method
tanks, pumphouses, and other facilities, but also on the consumer’s
property, particularly where water is sprayed through poorly ven- The combustible-gas indicator method (B) offers the advan-
tilated spaces, such as public showers. tages of simplicity, speed, and great sensitivity. The volumetric
The explosive limits of CH4 in air are 5% to 15% by volume. At method (C) can be made more accurate for concentrations of 4
sea level, a 3.95% CH4 concentration in air theoretically could be to 5 mg/L and higher, but will not be satisfactory for very low
reached in a poorly ventilated space sprayed with hot (68 °C) water concentrations. The volumetric method also can be applied to dif-
having a CH4 concentration of only 0.7 mg/L. At higher water tem- ferentiate between CH4 and other gases, as when a water supply
peratures, the vapor pressure of water is so great that no explosive is contaminated by liquid petroleum gas or other volatile combus-
mixture can form. At lower barometric pressures, the theoretical tible materials.
hazardous concentration of methane in water will be reduced pro- Methane also may be determined with the gas chromatograph
portionately. In an atmosphere of N2 or other inert gas, at least as described in Section 2720 C. This method permits differentia-
12.8% O2 must be present for there to be an explosion hazard. tion between H2 and CH4, and/or its higher homologs.

6211 B. Combustible-Gas Indicator Method

1. General Discussion c. Minimum detectable concentration: The limit of sensitivity


of the test is approximately 0.2 mg/L.
a. Principle: An equilibrium according to Henry’s law is estab- d. Sampling: If the water is supersaturated with CH4, a rep-
lished between CH4 in solution and the partial pressure of CH4 in resentative sample cannot be obtained unless the water is under
the gas phase above the solution. The partial pressure of CH4 can sufficient pressure to keep all of the gas dissolved. Operate wells
be determined with a combustible-gas indicator. The operation long enough to ensure sampling water coming directly from the
of the instrument is based on the catalytic oxidation of a com- aquifer. Representative samples can be expected only when the
bustible gas on a heated platinum filament that is made a part of well is equipped with a pump operating at sufficient submergence
a Wheatstone bridge. The heat generated by the oxidation of the to ensure that no gas escapes from the water.
gas increases the electrical resistance of the filament. The result- e. Quality control (QC): The QC practices considered to be
ing imbalance of the electrical circuit causes deflection of a mil- an integral part of each method are summarized in Table 6020:1.
liammeter that may be calibrated in terms of percentage of CH4 or
percentage of the lower explosive limit of the gas sampled. 2. Apparatus
b. Interference: Small amounts of ethane usually are associ-
ated with CH4 in natural gas and presumably would be present a. Combustible-gas indicator: Connect a 3-way stopcock to the
in water that contains methane. Hydrogen gas has been observed inlet to zero instrument on atmospheric air immediately before
in well waters and would behave similarly to CH4 in this pro- obtaining sample reading. For laboratory use, replace the suction
cedure. Hydrogen sulfide may interfere if the pH of the water bulb with a filter pump throttled to draw gas through the instru-
is low enough for an appreciable fraction of the total sulfide to ment at a rate of approximately 600 mL/min. See Figure 6211:1.
exist in the un-ionized form. The vapors of combustible oils also b. Laboratory filter pump.
may interfere. In general, these interferences are of no practical c. Glass bottle, 4-L, fitted with a 2-hole rubber stopper. Extend
importance because primary interest is in calculating the explo- inlet tube to within 1 cm of bottom. End outlet tube approximately
sion hazard to which all combustible gases and vapors contribute. 1 cm below stopper. Use metal or glass tubes, each fitted with
Interference due to H2S can be reduced by the addition of solid stopcocks or with short (approximately 5-cm) lengths of rubber
NaOH to the container before sampling. tubing and pinchcocks. The entire assembly should be capable

https://doi.org/10.2105/SMWW.2882.119 1
6211 METHANE - B. Combustible-Gas Indicator Method

where:
P = partial pressure of CH4 (kPa),
Vg = volume of gas phase (mL),
T = temperature (°C),
V1 = volume of liquid phase (mL), and
H = Henry’s law constant (kPa/mole CH4/mole of water).

Values for Henry’s constant are as follows:

Henry’s Henry’s
Temperature (°C) Constant, H* Temperature (°C) Constant, H*
0 2.265 40 5.261
5 2.625 45 5.577
10 3.010 50 5.846
15 3.413 60 6.342
20 3.804 70 6.749
25 4.181 80 6.911
30 4.544 90 7.013
Figure 6211:1. Combustible gas indicator circuit and flow diagram. 35 4.926 100 7.106
Multiply given values by 106.
*

of holding a low vacuum for several hours. Determine volume of


assembly by filling with water and measuring volume, or weight, For most determinations, it may be assumed that atmospheric
of water contained. pressure is 100 kPa, and that the temperature is 20 °C. The con-
centration of CH4 in the sample is then given by:
3. Reagent
 V − V1 
mg CH 4 /L = Rf 6.7 0 + 0.24
Sodium hydroxide (NaOH) pellets.  V1 

4. Procedure where:

a. Rough estimation of CH4 concentration: Fill bottle about R = scale reading,


f = factor depending on instrument used,
half full of water, using a rubber tube connecting sampling
V0 = total volume of sample bottle (mL), and
tap and inlet tube, with outlet tube open. With both inlet
V1 = volume of water sampled (mL).
and outlet tubes closed, shake bottle vigorously for approx-
imately 15 s and let stand for approximately 1 min. Sample If the instrument reads directly in percentage of methane,
gas phase by withdrawing gas from the outlet, leaving inlet f = 1.00. If the instrument reads in percentage of the lower explo-
open to admit air. If the needle swings rapidly to a high level sive limit of CH4, f = 0.05. For instruments that require additional
on the meter and then drops to zero, the CH 4-air mixture is factors, consult the manufacturer. For example, one commercial
too rich to burn; take a smaller sample for the final test. If instrument with a scale that reads in percentage of the lower
needle deflection is too small to be read accurately, take a explosive limit of combustible gases requires an additional factor
larger volume of water. of 0.77 for CH4. Hence, the value of f in the above equation would
b. Accurate determination: If the water contains H2S, add be 0.77 × 0.05, or 0.0385.
approximately 0.5 g NaOH pellets to empty bottle to suppress For more accurate work, or in locations where normal baromet-
interference. Evacuate bottle, using filter pump. Fill bottle not ric pressure is significantly <100 kPa, use the equation:
more than three-quarters full by connecting inlet tube to sampling
cock, with outlet tube closed. After collecting desired volume of  V − V1 8900 
water, let bottle fill with air through inlet tube. Close inlet cock, mg CH 4 /L = RBf 19.277 0 + 
 TV1 H 
shake bottle vigorously for 60 s, and let stand for at least 2 h.
Sample gas phase through outlet tube with inlet cock open. Take
reading as rapidly as possible before the entering air has diluted where:
sample appreciably. Measure volume of water sampled. B = barometric pressure, kPa,

5. Calculation and other symbols are as above.

The weight of CH4 (w), in mg, in the sample is given by the 6. Accuracy
equation:
The accuracy of the determination is limited by the accuracy
1.928 Vg 890 V  of the instrument used. Errors of approximately 10% may be
w = p  + 1
 expected. Calibration of instrument on known CH4–air mixtures
 T + 273 H 
will improve accuracy.

https://doi.org/10.2105/SMWW.2882.119 2
6211 METHANE - C. Volumetric Method

Bibliography

Rossum JR, Villarruz PA, Wade JA Jr. A new method for determining
methane in water. J Amer Water Works Assoc. 1950;42(4):413–415.

6211 C. Volumetric Method

1. General Discussion 2. Methane Determination

a. Principle: If CH4 is slowly mixed with an excess of O2 in the See Section 2720 B for a description of apparatus, reagents,
presence of a platinum coil heated to yellow incandescence, most procedure, calculation, and precision and bias.
of the CH4 will be converted to CO2 and H2O in a smooth reaction. Use percentage of CH4 found by this method with Henry’s law
Several passes of the mixed gases may be needed to burn sub- to obtain the CH4 concentration in original sample. Substitute CH4
stantially all the CH4. An excess of O2 is mixed with the sample percentage for R (scale reading) and f = 1 in the calculation given
before passage through the assembly. By differential absorption under 6211 B.5.
and volumetric changes the product CO2 is measured.
b. Interference: Low-boiling hydrocarbons other than ethane Bibliography
and vapors from combustible oils interfere. These substances,
however, are not likely to be present in water in sufficiently high Dennis LM, Nichols ML. Gas analysis. New York (NY): Macmillan Co.,
concentration to affect the results significantly. 1929.
c. Minimum detectable concentration: This method is not satis- Haldane JS, Graham JI. 1935. Methods of air analysis. London (UK):
factory for determining CH4 in water where the concentration is Charles Griffin & Co.; 1935.
<2 mg/L. Buswell AM, Larson TE. Methane in ground waters. J Amer Water Works
d. Sampling: Collect sample as directed in 6211 B and Assoc. 1937;29:1978.
Berger LB, Schrenk HH. Bureau of Mines Haldane gas analysis apparatus.
observe the same precautions to obtain representative samples
U.S. Bureau of Mines Information Circ. No. 7017; 1938.
(6211 B.1d). Omit NaOH pellets and fill sample bottle with water Larson TE. Properties and determination of methane in ground waters.
up to 90% of capacity. J Amer Water Works Assoc. 1938;30:1828.
e. Quality control (QC): The QC practices considered to be
an integral part of each method are summarized in Table 6020:1.

Published online: August 27, 2018


Revised: March 17, 2020
https://doi.org/10.2105/SMWW.2882.119 3

You might also like