1.Product synopsis
Mini-6K with No.1 rotor centrifuge appearance of the novel, Smart multi-purpose,Equipped with
two types of centrifugal rotor and a variety of in vitro,Apply to2.0ml,1.5ml,0.5ml,0.2ml Centrifuge
tube and PCR use 0.2ml,8 row centrifugal tube。Humanized design, flip the switch function, open
cover automatically stop(According to turn up and cover, open the lid is stopped), and has the
function of electronic timing. Transparent round cover, rotor, fully embodies the people for this, the
new idea of service research.

2.Basic instructions
 2.1 Structure diagram

Centrifugal cover

Rotor

Open button

Panel

 2.2 Change the turn

1
1.Index-finger to unload turn will yank the rotor buckle inside card as shown in figure A1, holding
the lower edge of the rotor vertical lift the rest of the fingers, as shown in figure B after take out the
rotor.
2. Then use another rotor on the rotor vertical press the button, button until the rotor stuck rotor, if
you want to change back, please repeat the above method.

 2.3 operation panel and the display screen

 2.4operating buttons
Long time and speed up or down key, keys rapidly rising or falling, is
set to the highest time 99 minutes 59s.

 2.5 operation guide


1、Time and speed setting
a) open the power switch, the screen will show "8" individually, instrument into the initialization,
and with "di" sound;
b) After 1 seconds ,time display window number 010 is the last run time for 10 seconds;
c) according to the or set Time values increase or decrease, and 10, one hundred, in
turn, can be increase or decrease.
d) Then Cover the centrifugal cover,the Mini centrifuge is running。
2、Instrument operation

2
a) before using this machine, it is necessary to determine the position of the switch position is in the
"I", according to the above method setup time and speed, close the battery cover the centrifugal
pressure centrifugal cover top, hear the ringing sound of "pa", suggests that the centrifugal cover has
on the safety clasp, instrument began to run. Instrument stop running when running time is over, after
waiting for motor stalling, then press the open hand, micro centrifugal cover up bounce. If you want
to stop halfway to run, but will the power switch to "0" position

 Warning:
1、Turn not completely stopped before cannot open centrifugal cover;
2 、 Centrifugal tube should be equal symmetrical placed, each centrifugal tube weight must not
exceed 3.0 g;
3 、 After startup if it is found that the noise is too large or there are abnormal noise, should
immediately stop and check the reason;
4、Using 8 x 1.5 ml turn does not allow to 1.5 ml and 0.5 ml and 0.2 ml tube to mix a variety of
specifications.

3. Safety tips
1、Prohibited without turn operation instrument;
2、It is forbidden to install the centrifugal tube in turn add samples, in order to avoid corrosion and
head, etc.;
3、Prohibited in the process of centrifugal mobile instrument;
4、The machine should be placed in a horizontal, solid work platform.

4. Clean
With dry cloth and PH = 7 plus or minus 1 non corrosive liquid detergent, clean surface of
instruments, to ensure that equipment to run the instrument after all parts completely dry, don't use
instrument immersed in a liquid or liquid wet instrument.

3
5.Parameters
Model Mini-4K Mini-6K Mini-7K Mini-10K

Speed 4000rpm 6000rpm 7000rpm 10000rpm


The relative centrifugal
860g 2000g 2600g 5400g
force (RCF)

Time range 0s~99min59s

8×2.0ml/1.5ml/0.5ml/0.2ml tube
Sample
(additional 0.5ml and 0.2ml tube carriers)
Capacity
2 ×8×0.2ml PCR tube strip
Working
45db 55db
noise

Overall dimensions 176x160x121(mm)

Weight ≤1.5kg
The input
220V/110V 50-60HZ
power supply

4
Packing List

No. Name Specification Quantity Remarks

1 Mini centrifuge Mini series 1

2 Instructions 1

3 Certificate 1

4 0.2ml adapter 8

5 0.5ml adapter 8

6 Fuse some

5
BIOBASE GROUP
2# building, No.9 Gangxing Road, High-tech Zone, Jinan City, Shandong Province,
China
Tel: +86-531-81219803/01
Fax: +86-531-81219804
Inquiry: export@biobase.com
Complaints: customer_support@biobase.cc
After-sales service: service_sd@biobase.cc; service_ivd@biobase.cc
Web: www.biobase.cc/www.meihuatrade.com / www.biobase.com